Abstract
Nitazenes are highly potent synthetic opioids increasingly detected in illicit drug markets and associated with significant public-health and forensic challenges. Rapid identification tools, such as lateral-flow immunoassay (LFA) test strips, are widely used in harm-reduction and forensic settings; however, their analytical performance for detecting nitazene-class opioids remains insufficiently characterized. This systematic review aims to evaluate the analytical performance of nitazene LFA test strips when used on seized drug materials and laboratory-prepared solutions, compared with mass spectrometry (MS) reference methods (liquid chromatography-tandem mass spectrometry (LC-MS/MS), liquid chromatography-quadrupole time-of-flight mass spectrometry (LC-QTOF-MS), and gas chromatography-mass spectrometry (GC-MS)). Outcomes included sensitivity, specificity, cross-reactivity, limits of detection, and operational interferences. Adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses - Diagnostic-Test Accuracy (PRISMA-DTA) guidelines and applying Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) framework, five eligible studies were synthesized, including laboratory evaluations and field-paired analyses. Only studies evaluating LFAs on drug materials - not biological samples - were included. Our results showed that two studies provided paired LFA-mass-spectrometry datasets. Under concentrated preparation conditions (10 mg/1 mL), LFAs demonstrated complete agreement with MS. Sensitivity declined at higher dilution (10 mg/5 mL), while specificity remained high. Analytical detectability varied substantially across nitazene analogues, with N-desethyl-metonitazene exhibiting the greatest sensitivity. Matrix effects, caffeine adulteration (~300 µg/mL), solvent concentrations >10% acetonitrile, and elevated temperatures all reduced line intensity or hindered wicking.
In conclusion, Nitazene LFA test strips show potential value as preliminary material-based screening tools in forensic and harm-reduction applications, but should not be interpreted as confirmatory. Their use requires standardized protocols, conservative interpretation rules, and mandatory mass-spectrometric confirmation. Large, independent, multi-site analytical validation studies are needed to establish reliability, optimize field use, and support integration into drug-checking programs.
Keywords: forensic medicine, forensics, nitrazine test, test strips, toxicology
Introduction and background
New synthetic opioids from the nitazene class have quickly appeared in the illegal drug trade. This makes things very hard for forensic labs, drug-checking services, and public health monitoring programs [1,2]. These molecules generated from benzimidazole demonstrate exceptional potency, frequently surpassing that of fentanyl, and persist in powders, counterfeit tablets, and paraphernalia residues linked to both fatal and non-fatal intoxications [3]. It is hard to find nitazenes on a regular basis since they have different structures, and it is easy to produce new analogues. This is especially true because many nitazenes are not included in typical opioid immunoassays [4].
Lateral-flow immunoassays (LFAs), often called rapid test strips, are becoming more popular for preliminary screening of illegal substances in harm-reduction and forensic settings because they are cheap, easy to carry, and give quick visual results [5].
However, the LFA products that are currently available were originally designed for regular opioids, and their antibodies may not work well with the chemically different nitazene family. This raises issues about false-negative and false-positive results, especially when the strips are used on complicated street-level powders that have more than one adulterant and excipient [6, 7]. To guarantee dependable substance identification, the analytical efficacy of nitazene LFAs must be assessed against high-specificity confirmatory techniques, including liquid chromatography-tandem mass spectrometry (LC-MS/MS) and gas chromatography-mass spectrometry (GC-MS), which are regarded as the standard methodologies in forensic toxicology [8, 9]. However, current research exhibits significant variability in sample preparation, analogue makeup, dilution techniques, ambient conditions, and interpretation guidelines, hindering the capacity to draw conclusive determinations regarding their trustworthiness [10,11,12].
This systematic review consolidates the existing information about the analytical efficacy of nitazene lateral-flow test strips when utilized on confiscated illicit substances or laboratory-synthesized powders, employing MS-based methodologies as the benchmark standard. The review analyzes sensitivity, specificity, analogue-level detectability, cross-reactivity, matrix interferences, and operational factors pertinent to forensic and harm-reduction practices. The goal is to make it clear what nitrazine LFAs can and can't do as preliminary material-based screening methods and to find important gaps that need more analytical validation.
Review
Methodology
Study Design and Search Strategy
This systematic review was conducted in accordance with the PRISMA-DTA framework, with the “target condition” defined as the presence of nitazene-class synthetic opioids in seized drug materials or laboratory-prepared powders, rather than any clinical or diagnostic condition. The index tests were lateral-flow immunoassay (LFA) strips intended for nitazene or non-fentanyl opioid detection, and the reference standards were mass-spectrometric methods, including LC-MS/MS, liquid chromatography-quadrupole time-of-flight mass spectrometry (LC-QTOF-MS), high-resolution mass spectrometry (HRMS), and GC-MS. A thorough search was conducted in the following databases: PubMed/MEDLINE, Embase, Scopus, Web of Science, and Google Scholar, encompassing papers from November 2015 to November 2025. The search was enhanced by gray literature sources, such as ChemRxiv, medRxiv, Center for Forensic Science Research and Education (CFSRE)/Novel Psychoactive Substances (NPS) Discovery, DrugsData, Institute of Environmental Science and Research (ESR) New Zealand analytical reports, and manufacturer white papers. Search queries included both restricted vocabulary and free-text words about nitazenes, synthetic opioids, immunoassay testing, and analytical validation. An example of a search string is: (nitazene OR benzimidazole opioid OR isotonitazene OR metonitazene OR protonitazene OR "non-fentanyl opioid*") AND (lateral flow OR immunoassay OR "test strip*" OR LFA) AND (analytical accuracy OR sensitivity OR specificity OR "limit* of detection") AND (LC-MS OR LC-MS/MS OR GC-MS OR HRMS). We manually went through the reference lists of all the studies that were included to find more records. Duplicate records were removed in EndNote (Clarivate Plc, London, UK), followed by additional de-duplication and title/abstract screening in Covidence (Covidence, Melbourne, Australia). No machine-learning-based exclusion tools were used; all automated decisions were manually verified.
Eligibility Criteria
Studies were eligible for inclusion if they met all the following criteria: evaluated a lateral-flow immunoassay designed to detect nitazenes or related non-fentanyl synthetic opioids. Tested seized materials, drug powders, tablet residues, or prepared solutions (biological samples were excluded). Employed mass spectrometry as a reference standard or provided sufficient analytical validation data (e.g., limit of detection (LOD), cross-reactivity). Reported outcomes related to sensitivity, specificity, signal intensity, LOD, cross-reactivity, repeatability, or interference effects. Our exclusion criterion was to exclude editorials, commentaries, opinion pieces, animal studies, studies lacking reproducible methodological description, and studies addressing only fentanyl or non-nitazene opioids without nitazene data.
Study Selection
Two reviewers independently screened titles, abstracts, and full texts. Discrepancies were resolved by discussion and consensus. Automated filtering tools were used to remove duplicate digital records, and all automated exclusions were subsequently verified by manual screening to ensure accuracy. A PRISMA flow diagram (Figure 1) was constructed to document all stages of search, screening, and inclusion. All included studies evaluated LFAs on powders, seized materials, or laboratory-prepared solutions; none evaluated clinical specimens.
Figure 1. PRISMA Flow Diagram .
PRISMA 2020 flow diagram for new systematic reviews, which included searches of databases and registers only.
LFA: lateral-flow immunoassay
Data Extraction
Data were extracted using a structured template that captured key methodological and analytical variables from each study, including the type of nitazene analogue tested, the LFA brand and lot number, sample preparation methods, dilution ratios, cut-off thresholds, and interpretation rules. Information related to environmental and operational conditions - such as temperature, solvent composition, and potential interference - was also recorded. For studies that employed mass-spectrometric confirmation, details of the reference method and timing were extracted, and paired results were used to reconstruct true-positive, false-positive, false-negative, and true-negative outcomes when available. Additional analytical parameters, including limits of detection, cross-reactivity, and lot-to-lot variability, were collected to support narrative synthesis. All extracted data were independently reviewed by two investigators to ensure accuracy and consistency.
Quality Assessment, Data Synthesis, and Analysis
Methodological quality was evaluated using the QUADAS-2 tool, tailored for material-based analytical testing. The domains assessed included: Sample selection, Execution of the index test, Execution of the reference standard, Flow and timing, and Funding or manufacturer involvement. Studies lacking MS confirmation for negative samples or involving manufacturer funding were rated as high risk. Because complete paired datasets were available from only two studies, a narrative structured synthesis was performed. Sensitivity and specificity were calculated with Clopper-Pearson exact 95% confidence intervals for paired studies. Quantitative pooling through bivariate random-effects meta-analysis was planned but not conducted due to insufficient data. Analytical characteristics - including LOD, cross-reactivity, solvent effects, excipient interference (e.g., caffeine), lot variability, and temperature-based changes - were summarized descriptively across studies.
Results
Study Selection and Characteristics
Five papers passed the inclusion criteria and were used in the final synthesis: De Vrieze (2024) [1], Marland (2025) [2], BTNX/Chicago Recovery Alliance (CRA) White Paper (2024) [3], ESR NZ (2024) [4], and Edward Sisco (2024) [5]. These studies combined laboratory evaluations, bench validation panels, and field-based diagnostic assessments of nitazene lateral-flow immunoassay (LFA) test strips. Table 1 displays research conducted in academia and manufacturer-affiliated settings, utilizing the BTNX Rapid Response™ (BTNX Inc., Pickering, Canada) brand for all tests. Two studies (De Vrieze 2024 [1] and BTNX/CRA 2024 [3]) explicitly compared LFA results to a mass-spectrometry reference (LC-QTOF-MS or GC/MS), but the other three (Marland 2025 [2], ESR NZ 2024 [4], and Sisco 2024 [5]) focused on analytical performance, cross-reactivity, and interferences without diagnostic pairing. All tests used the manufacturer's competitive immunoassay rule (absence of test line = positive), with read times ranging from 5 to 10 minutes. Lot numbers varied between experiments, indicating possible lot-to-lot variability (Table 1).
Table 1. Study characteristics and index-test conduct.
All LFAs are competitive (absence of test line = positive). Read time ≤ 10 min per manufacturer's instructions. Flow and timing indicate the interval between the strip test and MS confirmation.
BTNX Rapid Response™ by BTNX Inc., Pickering, Canada. ImageJ is free software for basic scientific image analysis (https://imagej.net/ij/).
Abbreviations: LFA = lateral-flow assay; LOD = limit of detection; MS = mass spectrometry; LC-QTOF-MS = liquid chromatography-quadrupole time-of-flight MS; COI = conflict of interest; CFSRE = Center for Forensic Science Research and Education; GC = gas chromatography; TP = true positive; lot A/B = lot A and lot B
| Study (year) | Setting/sample type | Brand/lot | Funding/conflicts | Read time & positivity rule | Reader blinding/method | Sample preparation/dilution | Flow & timing vs MS | Reference standard | Notes |
| De Vrieze 2024 [1] | Laboratory + 6 authentic powders | BTNX Rapid Response™ Lots A/B | Independent academic (no COI) | Competitive LFA (“absence of T line = positive”); 5 & 10 min | 2 visual + ImageJ (blinded) | Water suspension; some undissolved | Same sample tested ≤ 1 h by LC-QTOF | LC-QTOF-MS | 6/6 authentic TP; no negatives. |
| Marland 2025 [2] | Lab; 36 nitazenes, 93 adulterants, 3 seized heroin | BTNX lots DOAB24050015/DOAB24020005 | Academic (no COI) | ≤ 10 min; competitive | Dual visual readers | Tap water; LOD by analogue | Same-day triplicates (no MS) | — | Caffeine fading (~300 µg/mL); LOD 250 ng–100 µg/mL. |
| BTNX/CRA 2024 White Paper [3] | Field; street samples (USA) | BTNX Rapid Response™ | Manufacturer sponsored | Competitive; faint line = negative | Field visual readers | 10 mg / 1 mL and 10 mg / 5 mL | Strip then GC/MS ≤ 24 h | GC/MS (DrugsData) | n = 11 paired; prelim sens 100 / 87.5%. |
| ESR NZ 2024 [4] | Service lab; drug checking (NZ) | BTNX lot DOAB23120001 | Govt funded (no COI) | Dip 15 s; read ≤ 10 min | Visual single reader | 1 mL water; serial dilutions | Same sample; no MS | — | Poor solubility for etonitazene noted. |
| Sisco (2024) [5] | Bench lab validation; cross-reactivity panel | BTNX Rapid Response™ | Manufacturer / CFSRE collab | 5 min; competitive | Visual only | Dilutions 0.25–10 µg/mL | None | — | Expanded LOD (~30 analogues); repeatability CV < 15%. |
Analytical Accuracy (Paired Data)
Two trials yielded paired data appropriate for diagnostic accuracy assessment. Table 2 summarizes that the CRA (2024) [3] field evaluation encompassed 11 paired street samples validated by GC/MS. Employing a 10 mg/1 mL procedure, the strip demonstrated 100% sensitivity (95% CI 63.1-100) and 100% specificity (95% CI 29.2-100). Upon diluting the identical samples to 10 mg/5 mL, sensitivity decreased to 87.5% (95% CI 47.3-99.7), although specificity persisted at 100%. The sole false negative was observed in the higher-dilution procedure (sample Y4 319). The De Vrieze (2024) [1] laboratory assessment of six genuine powders indicated total agreement with LC-QTOF-MS, yielding six true positives and no false negatives. Due to the limited paired sample size and broad confidence intervals, pooled meta-analysis was not conducted; however, both datasets exhibit exceptionally high accuracy under concentrated conditions (Table 2).
Table 2. Analytical accuracy vs MS comparator.
CRA row Y4 332 excluded (no GC/MS). Exact 95 % CIs use the Clopper–Pearson method.
Abbreviations: TP = true positive; FP = false positive; FN = false negative; TN = true negative; CI = confidence interval; Sens = sensitivity; Spec = specificity; GC = gas chromatography; MS: mass spectrometry; CRA = Chicago Recovery Alliance; LC-QTOF-MS = liquid chromatography-quadrupole time-of-flight mass spectrometry
| Dataset | Protocol | Unit of analysis | n (paired) | TP | FP | FN | TN | Sensitivity (95 % CI) | Specificity (95 % CI) | Flow/timing | Positivity rule | Reference |
| CRA (USA) [3] | 10 mg/1 mL | Per sample | 11 | 8 | 0 | 0 | 3 | 100 % (63.1–100) | 100 % (29.2–100) | ≤ 24 h same aliquot | Faint line = negative | GC/MS (DrugsData) |
| CRA (USA) [3] | 10 mg/5 mL | Per sample | 11 | 7 | 0 | 1 | 3 | 87.5 % (47.3–99.7) | 100 % (29.2–100) | ≤ 24 h same aliquot | Faint line = negative | GC/MS (DrugsData) |
| De Vrieze 2024 [1] | Lab/authentic | Per-sample | 6 | 6 | — | 0 | — | All TP (no FN) | — | ≤ 1 h | Faint line = negative | LC-QTOF-MS |
Analogue-Level Detection and Cross-Reactivity
Analytical sensitivity showed significant variation among nitazene analogues (Table 3). Isotonitazene demonstrated a limit of detection (LOD) ranging from 2000 to 3000 ng/mL, with uniform cross-reactivity at or below 3000 ng/mL, as reported by De Vrieze (2024) [1], Marland (2025) [2], and Sisco (2024) [5]. Protonitazene necessitated concentrations of 3000-4500 ng/mL, while N-pyrrolidino etonitazene required 1300 ng/mL, although the latter exhibited incomplete solubility in water. N-desethyl metonitazene exhibited the highest sensitivity, identified at approximately 250 ng/mL, while desnitazene derivatives (metodesnitazene, etodesnitazene) sometimes did not respond even at 100 µg/mL. Aggregated findings from investigations revealed that 24 out of 33 nitazene analogues were identified at concentrations of <9000 ng/mL (Table 3).
Table 3. Analogue-level limits of detection (LOD) and cross-reactivity.
Analogue LODs compiled from three lab evaluations (Marland [2], De Vrieze [1], v5). Cross-reactivity indicates any visible test-line loss at ≤ 3000 ng/mL.
Abbreviations: LOD = limit of detection; conc = concentration; ESR: Environmental Science and Research, New Zealand
| Analogue | Lot ID | Best LOD (ng/mL) | Cross-reactive ≤ 3000 ng/mL? | Read time | Source/notes |
| Isotonitazene | A & B | 2000–3000 | Yes | 5–10 min | Lot-to-lot variation shown (De Vrieze; v5). |
| Protonitazene | A | 3000–4500 | Yes | 5 min | Detected consistently (Marland; v5). |
| N-pyrrolidino etonitazene | A | 1300 | Yes | 5 min | Poor solubility in water (ESR). |
| N-desethyl metonitazene | A | 250 | Yes (low LOD) | 5 min | Detected easily (Marland). |
| Metodesnitazene / Etodesnitazene | A | > 100 000 | Mixed / often missed | 10 min | Detected only at high conc.; lot dependent. |
| All analogues (aggregate) | — | ≤ 9000 for 24/33 | — | — | Cross-reactivity confirmed on triplicate runs (De Vrieze; version 5). |
Interferences, Matrix Effects, and Operational Findings
Experimental and field data revealed multiple sources of potential influence (Table 4). Caffeine, a prevalent adulterant, resulted in significant attenuation of the test line at around 300 µg/mL, hence posing a danger of false negatives, especially in heroin mixes documented by Marland (2025) [2] and CRA (2024) [3]. Organic solvent concentrations exceeding 10% acetonitrile inhibited adequate wicking, and temperatures surpassing 37 °C diminished line intensity by approximately 10%. Most techniques alleviated these effects by restricting solvent concentration and preserving storage at ambient temperature (15-30 °C). Training in interpretation was advised, highlighting either the manufacturer's guideline (faint line = negative) or a harm-reduction alternative (faint line = positive) to mitigate false reassurance (Table 4).
Table 4. Interferences, matrix effects and interpretation.
Matrix and environmental interference data compiled from lab and manufacturer evaluations; read time = interval from immersion to interpretation.
Abbreviations: ACN = acetonitrile; MeOH = methanol; T-line = test line; CRA = Chicago Recovery Alliance; ESR = Environmental Science and Research, New Zealand
| Factor | Threshold / effect | Read time | Mitigation | Evidence |
| Caffeine | Fades T-line ≈ 300 µg/mL → false negative risk | 5 min | Use ≤ 5 mg sample / 1 mL protocol | Marland 2025 [2]; CRA [3] heroin mixes. |
| Organic solvents > 10 % ACN | Wicking failure (no control line) | 10 min | Avoid high ACN; prefer ≤ 5 % MeOH | ESR 2024 [4] tech report. |
| Storage > 37 °C | Band intensity ↓ ~10 % | 5 min | Store 15–30 °C | BTNX [3] stability data (v5). |
| Interpretation rule | Competitive: absence of T = positive | 5 min | Train users; optionally “faint line = positive” in harm-reduction | De Vrieze 2024 [1]; Marland 2025 [2]. |
Risk of Bias and Applicability
The quality assessment with QUADAS-2 is illustrated in Table 5. De Vrieze (2024) [1] exhibited moderate overall risk, chiefly because of the limited sample size and absence of negative controls. Marland (2025) [2] and Sisco (2024) [5] were classified as high risk due to the absence of a reference standard, functioning solely as analytical validations. The CRA (2024) white paper [3] was assessed as high risk due to its manufacturer financing, convenience sampling, and lack of blinded evaluations. Environmental Science and Research (ESR) New Zealand (2024) [4] demonstrated a significant danger for analogous reasons. These restrictions collectively diminish trust in aggregated accuracy estimates while offering significant preliminary evidence regarding test feasibility and performance. Table 5.
Table 5. QUADAS-2 risk of bias and applicability.
QUADAS-2 domains applied to each study. “Flow & timing” judges same-sample testing interval; “Funding bias” added for transparency.
Abbreviations: RoB = risk of bias; DTA = diagnostic-test accuracy; MS = mass spectrometry; LC-QTOF = liquid chromatography-quadrupole time-of-flight; LOD = limit of detection
| Study | Selection | Index test | Reference standard | Flow & timing | Applicability | Funding bias | Overall RoB | Key limitations |
| De Vrieze 2024 [1] | Low/Mod. | Low (dual read + ImageJ blinded) | Low (LC-QTOF) | Unclear (no negatives) | Some concerns (lab→field) | Low | Moderate | Small n; no specificity. |
| Marland 2025 [2] | High (bench panel) | Low | High (no MS) | High | Some concerns | Low | High | Not DTA; only LOD panel. |
| BTNX/CRA 2024 [3] | High (convenience street sample) | Unclear (field visual) | Moderate (GC/MS) | Unclear | Some concerns | High | High | Gray literature; no blinding. |
| ESR New Zealand 2024 [4] | High (service report) | Low | High (no MS) | Unclear | Some concerns | Low | High | Not DTA study. |
| Sisco (2024) [5] | High (manufacturer bench) | Low | High (no MS) | High | Low concern (analytical) | High | High | Analytical validation only. |
Discussion
Summary of Key Findings
This systematic review presents the inaugural synthesis of analytical performance data for nitazene lateral-flow immunoassay (LFA) test strips in comparison with mass-spectrometry reference methods. Adhering to PRISMA-DTA standards [13] and employing QUADAS-2 [14], the review demonstrates that the LFAs show very good analytical sensitivity under concentrated sample conditions, with reduced sensitivity at higher dilutions. The paired datasets exhibited complete agreement with MS at 10 mg/1 mL [12], decreasing to 87.5% at 10 mg/5 mL, while maintaining 100% specificity. These findings are consistent with established performance patterns in fentanyl test strips (FTS), where analyte concentration, matrix composition, and temperature substantially influence visual thresholds [15]. Similarly, specific excipients (e.g., caffeine ≥300 µg/mL) and solvents (>10% acetonitrile) were observed to affect wicking and diminish line intensity, underscoring the importance of standardized preparation protocols.
Interpretation and Comparison With the Existing Literature
Forensic case reports and toxicological investigations have linked isotonitazene, metonitazene, and etonitazepyne with severe and fatal intoxications [16,17,18,19,20]. Nitazenes represent a structurally diverse class of potent synthetic opioids, distinct from fentanyl but exhibiting comparable or greater μ-opioid receptor activity [21,22]. Detecting this chemically varied class via immunoassay is challenging, as structural heterogeneity influences antibody binding affinity and may result in variable cross-reactivity [23-26].
The dilution-related decline in performance observed in this review aligns with studies [23] and [26], which showed that lower analyte concentrations can reduce line intensity and overall accuracy in fentanyl test strips [23,26,27,28,29]. Studies also highlighted lot- and instruction-dependent variability, underscoring the necessity of rigorous quality-control procedures [27]. Forensic/toxicology evidence-based evaluations similarly demonstrated strong agreement between field LFA testing and mass-spectrometric confirmation when standardized dilution and interpretation procedures were applied [17-20]. From a mechanistic perspective, receptor-binding data and metabolic variability among nitazene analogues [22,23] provide plausible explanations for inconsistent immunoassay reactivity. While LC-MS/MS can achieve robust multi-analyte detection [19], lateral-flow technology will inherently remain constrained by antibody specificity and structural cross-reactivity.
Strengths and Limitations
A key strength of this review is its methodological rigor, including dual-reviewer screening, structured data extraction, and formal quality assessment using QUADAS-2 [14]. Incorporation of both laboratory-controlled evaluations and field-paired datasets supports a broader applicability of findings. However, the evidence base remains constrained by small paired-sample sizes (n < 20), limited blinding in one study, and incomplete MS verification of negative LFA results. Additional limitations include lot-to-lot variability, non-standardized storage and preparation conditions, and matrix-related interferences. Similar challenges are well documented in prior LFA analytical-accuracy assessments [15,16,27].
Implications for Practice Policy
These findings bear practical relevance for harm-reduction and forensic screening, where rapid material-based testing can assist in identifying nitazene-containing substances [24,25]. Nitazene LFAs may be used as preliminary screening tools when accompanied by clear interpretive guidance and subsequent confirmatory analysis (e.g., Fourier-transform infrared spectroscopy (FTIR), Raman, or LC-MS) [30,31,32,33,34,35]. Consistent with implementation studies for fentanyl strips [24,25,34,35], accessible and reliable screening methods can support safer decision-making among people who use drugs, such as testing small doses, avoiding solitary use, and carrying naloxone. Given the extreme potency of nitazenes - often exceeding fentanyl by 20-40 fold - the consequences of false-negative screening are substantial. Programs should employ conservative reading protocols (“any absent or faint line = positive”), utilize strict sample-preparation ratios (≤5 mg/1 mL), avoid organic solvents and elevated temperatures, and ensure operator training. Regular lot verification is vital for sustaining analytical reliability [33,34,36]. Integration of nitazene screening into drug-checking systems should emphasize that LFAs supplement - but do not replace - confirmatory laboratory testing [37].
Future Research Directions
Future work should focus on expanding and strengthening the evidence base through large-scale, blinded, multi-site evaluations that incorporate full MS confirmation of both positive and negative samples [38]. Further research is needed to characterize cross-reactivity within the rapidly evolving nitazene family, as structural modifications significantly influence antibody affinity and assay responsiveness [37,38]. Controlled laboratory experiments should explore operational variables - including temperature, dilution ratios, solvent composition, and matrix effects - to define standardized protocols and reference conditions [39]. Comparative assessments across manufacturers and production lots would support the development of formal quality-assurance frameworks for harm-reduction programs [40]. Translational research linking analytical performance to behavioral and overdose-prevention outcomes remains essential [41]. Additionally, the development of multiplex or hybrid screening systems combining LFAs with FTIR, Raman, or LC-MS/MS may provide scalable and contextually relevant approaches to address the ongoing expansion of nitazene distribution [41].
Conclusions
This systematic review demonstrates that nitrazine lateral-flow immunoassays exhibit promising analytical performance when applied to seized drug materials, particularly under concentrated preparation conditions and standardized operational procedures. Although the strips consistently detected several high-potency nitazene analogues, their performance decreased with greater dilution, variable matrices, and environmental interferences. The evidence base remains limited by small paired sample sizes, incomplete negative controls.
Despite these limitations, nitazene LFAs may serve as practical preliminary screening tools within harm-reduction and forensic drug-checking systems, provided that results are interpreted conservatively and always confirmed using validated laboratory methods such as LC-MS/MS or GC-MS. To ensure reliability and broader applicability, future work must include large, blinded, multi-site evaluations with full mass-spectrometric confirmation of both positive and negative samples, expanded analogue panels, and systematic assessments of lot variability and operational factors. Improved standardization and analytical validation will be essential for integrating nitrazine LFAs responsibly into drug-monitoring and public-health surveillance frameworks.
Disclosures
Conflicts of interest: In compliance with the ICMJE uniform disclosure form, all authors declare the following:
Payment/services info: All authors have declared that no financial support was received from any organization for the submitted work.
Financial relationships: All authors have declared that they have no financial relationships at present or within the previous three years with any organizations that might have an interest in the submitted work.
Other relationships: All authors have declared that there are no other relationships or activities that could appear to have influenced the submitted work.
Author Contributions
Concept and design: Abdulkreem Al-Juhani, Abdulelah Alasmari, Omar O. Aljohani, Hatoon Hakeem, Rodan Desoky, Muhannad A. Badghaish, Naif Aljohani, Fai A. Alanazi
Acquisition, analysis, or interpretation of data: Abdulkreem Al-Juhani, Abdulelah Alasmari, Omar O. Aljohani, Hatoon Hakeem, Rodan Desoky, Muhannad A. Badghaish, Naif Aljohani, Fai A. Alanazi
Drafting of the manuscript: Abdulkreem Al-Juhani, Abdulelah Alasmari, Omar O. Aljohani, Hatoon Hakeem, Rodan Desoky, Muhannad A. Badghaish, Naif Aljohani, Fai A. Alanazi
Critical review of the manuscript for important intellectual content: Abdulkreem Al-Juhani, Abdulelah Alasmari, Omar O. Aljohani, Hatoon Hakeem, Rodan Desoky, Muhannad A. Badghaish, Naif Aljohani, Fai A. Alanazi
Supervision: Abdulkreem Al-Juhani
References
- 1.Nitazene test strips: a laboratory evaluation. De Vrieze LM, Stove CP, Vandeputte MM. Harm Reduct J. 2024;21:159. doi: 10.1186/s12954-024-01078-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Evaluation of nitazene immunoassay test strips for rapid in-situ detection of nitazene and nitazene analogs in illicit drug samples. Marland V, Nisbet L, Nic Daéid N. Harm Reduct J. 2025;22:137. doi: 10.1186/s12954-025-01287-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.BTNX Rapid Response Nitazene Test Strip 2000 ng/mL | BTNX Inc. March. [ Dec; 2025 ]. 2024. https://cloudfront.exchangesupplies.org/pdfs/BTNX%20Rapid%20Response%20Nitazene%20White%20Paper%202024-03%20V3%201%20%281%29.pdf https://cloudfront.exchangesupplies.org/pdfs/BTNX%20Rapid%20Response%20Nitazene%20White%20Paper%202024-03%20V3%201%20%281%29.pdf
- 4.Marsh DR, Wall C, Haren L, et al. Drug checking technical report: Part 2 - additional testing of opioids with BTNX rapid response nitazene test strips | ESR - Science for Communities. New Zealand: Institute of Environmental Science and Research Ltd. [ Dec; 2025 ]. 2024. https://research.esr.cri.nz/articles/report/Drug_Checking_Technical_Report_-Nitazene_Test_Strips_Part_2-Opioids/27320973/1/files/50050719.pdf https://research.esr.cri.nz/articles/report/Drug_Checking_Technical_Report_-Nitazene_Test_Strips_Part_2-Opioids/27320973/1/files/50050719.pdf
- 5.Beyond fentanyl test strips: investigating other urine drug test strips for drug checking applications. Sisco E, Appley MG, Pyfrom EM, et al. https://doi.org/10.1016/j.forc.2024.100594. Forensic Chem. 2024;40 doi: 10.1016/j.forc.2024.100594. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Comprehensive method to detect nitazene analogues and xylazine in wastewater. Keller EL, Peake B, Simpson BS, White JM, Gerber C. Environ Sci Pollut Res Int. 2025 doi: 10.1007/s11356-025-36425-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Validation of a lateral flow chromatographic immunoassay for the detection of fentanyl in drug samples. Park JN, Sherman SG, Sigmund V, Breaud A, Martin K, Clarke WA. https://doi.org/10.1016/j.drugalcdep.2022.109610. Drug Alcohol Depend. 2022;240:109610. doi: 10.1016/j.drugalcdep.2022.109610. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Nitazenes (CCENDU Drug Alert) | Canadian Centre on Substance Use and Addiction. 2022. https://www.ccsa.ca/en/nitazenes-ccendu-drug-alert https://www.ccsa.ca/en/nitazenes-ccendu-drug-alert
- 9.Mu-opioid receptor selective superagonists produce prolonged respiratory depression. Malcolm NJ, Palkovic B, Sprague DJ, et al. iScience. 2023;26:107121. doi: 10.1016/j.isci.2023.107121. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.The emergence of nitazenes: a new chapter in the synthetic opioid crisis. Caprari C, Ferri E, Rossetti P, Gregori A, Citti C, Cannazza G. Arch Toxicol. 2025;99:3877–3896. doi: 10.1007/s00204-025-04102-3. [DOI] [PubMed] [Google Scholar]
- 11.Rapid GC-MS method for screening seized drugs in forensic investigations: optimization and validation. Askar AM, Al Ali AY, Khalifa MK, Salem AA, Alkhuwaildi BM, Shah I. Front Chem. 2025;13:1559279. doi: 10.3389/fchem.2025.1559279. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Nitazenes: review of comparative pharmacology and antagonist action. Stangeland M, Dale O, Skulberg AK. https://doi.org/10.1080/15563650.2025.2504133. Clin Toxicol (Phila) 2025;63:393–406. doi: 10.1080/15563650.2025.2504133. [DOI] [PubMed] [Google Scholar]
- 13.Preferred Reporting Items for a Systematic Review and Meta-analysis of Diagnostic Test Accuracy Studies: The PRISMA-DTA Statement. McInnes MD, Moher D, Thombs BD, et al. JAMA. 2018;319:388–396. doi: 10.1001/jama.2017.19163. [DOI] [PubMed] [Google Scholar]
- 14.QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Whiting PF, Rutjes AW, Westwood ME, et al. Ann Intern Med. 2011;155:529–536. doi: 10.7326/0003-4819-155-8-201110180-00009. [DOI] [PubMed] [Google Scholar]
- 15.Lateral flow (immuno)assay: its strengths, weaknesses, opportunities and threats. A literature survey. Posthuma-Trumpie GA, Korf J, van Amerongen A. Anal Bioanal Chem. 2009;393:569–582. doi: 10.1007/s00216-008-2287-2. [DOI] [PubMed] [Google Scholar]
- 16.Lateral flow assays. Koczula KM, Gallotta A. Essays Biochem. 2016;60:111–120. doi: 10.1042/EBC20150012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Isotonitazene quantitation and metabolite discovery in authentic forensic casework. Krotulski AJ, Papsun DM, Kacinko SL, Logan BK. J Anal Toxicol. 2020;44:521–530. doi: 10.1093/jat/bkaa016. [DOI] [PubMed] [Google Scholar]
- 18.Metonitazene in the United States-Forensic toxicology assessment of a potent new synthetic opioid using liquid chromatography mass spectrometry. Krotulski AJ, Papsun DM, Walton SE, Logan BK. Drug Test Anal. 2021;13:1697–1711. doi: 10.1002/dta.3115. [DOI] [PubMed] [Google Scholar]
- 19.Analytical characterization of "etonitazepyne," a new pyrrolidinyl-containing 2-benzylbenzimidazole opioid sold online. Blanckaert P, Balcaen M, Vanhee C, et al. Drug Test Anal. 2021;13:1627–1634. doi: 10.1002/dta.3113. [DOI] [PubMed] [Google Scholar]
- 20.Intoxications in an Australian Emergency Department involving 'nitazene' benzylbenzimidazole synthetic opioids (etodesnitazene, butonitazene and protonitazene) Schumann JL, Syrjanen R, Alford K, et al. J Anal Toxicol. 2023;47:0–9. doi: 10.1093/jat/bkac062. [DOI] [PubMed] [Google Scholar]
- 21.Metabolic characterization of the new benzimidazole synthetic opioids - nitazenes. Jadhav GR, Fasinu PS. Front Pharmacol. 2024;15:1434573. doi: 10.3389/fphar.2024.1434573. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Synthesis, chemical characterization, and μ-opioid receptor activity assessment of the emerging group of "nitazene" 2-benzylbenzimidazole synthetic opioids. Vandeputte MM, Van Uytfanghe K, Layle NK, St Germaine DM, Iula DM, Stove CP. ACS Chem Neurosci. 2021;12:1241–1251. doi: 10.1021/acschemneuro.1c00064. [DOI] [PubMed] [Google Scholar]
- 23.An assessment of the limits of detection, sensitivity and specificity of three devices for public health-based drug checking of fentanyl in street-acquired samples. Green TC, Park JN, Gilbert M, et al. Int J Drug Policy. 2020;77:102661. doi: 10.1016/j.drugpo.2020.102661. [DOI] [PubMed] [Google Scholar]
- 24.Fentanyl test strips as an opioid overdose prevention strategy: findings from a syringe services program in the Southeastern United States. Peiper NC, Clarke SD, Vincent LB, Ciccarone D, Kral AH, Zibbell JE. Int J Drug Policy. 2019;63:122–128. doi: 10.1016/j.drugpo.2018.08.007. [DOI] [PubMed] [Google Scholar]
- 25.High willingness to use rapid fentanyl test strips among young adults who use drugs. Krieger MS, Yedinak JL, Buxton JA, et al. Harm Reduct J. 2018;15:7. doi: 10.1186/s12954-018-0213-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Testing the test strips: laboratory performance of fentanyl test strips. Halifax JC, Lim L, Ciccarone D, Lynch KL. Harm Reduct J. 2024;21:14. doi: 10.1186/s12954-023-00921-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Assessment of two brands of fentanyl test strips with 251 synthetic opioids reveals "blind spots" in detection capabilities. Hayes KL, Lieberman M. Harm Reduct J. 2023;20:175. doi: 10.1186/s12954-023-00911-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Selectivity and sensitivity of urine fentanyl test strips to detect fentanyl analogues in illicit drugs. Bergh MS, Øiestad ÅML, Baumann MH, Bogen IL. Int J Drug Policy. 2021;90:103065. doi: 10.1016/j.drugpo.2020.103065. [DOI] [PubMed] [Google Scholar]
- 29.Evaluating the sensitivity, stability, and cross-reactivity of commercial fentanyl immunoassay test strips. Rodriguez-Cruz SE. J Forensic Sci. 2023;68:1555–1569. doi: 10.1111/1556-4029.15332. [DOI] [PubMed] [Google Scholar]
- 30.Validation of an analytical method for quantitation of metonitazene and isotonitazene in plasma, blood, urine, liver and brain and application to authentic postmortem casework in New York City. Pardi J, Ford S, Cooper G. J Anal Toxicol. 2023;47:648–655. doi: 10.1093/jat/bkad062. [DOI] [PubMed] [Google Scholar]
- 31.Scientific Working Group for Forensic Toxicology (SWGTOX) standard practices for method validation in forensic toxicology. J Anal Toxicol. 2013;37:452–474. doi: 10.1093/jat/bkt054. [DOI] [PubMed] [Google Scholar]
- 32.ANSI/ASB Standard 036 for method validation in forensic toxicology has replaced SWGTOX's version. LeBeau MA. J Anal Toxicol. 2020;44:414. doi: 10.1093/jat/bkz115. [DOI] [PubMed] [Google Scholar]
- 33.Assessing the limit of detection of Fourier-transform infrared spectroscopy and immunoassay strips for fentanyl in a real-world setting. McCrae K, Tobias S, Grant C, Lysyshyn M, Laing R, Wood E, Ti L. Drug Alcohol Rev. 2020;39:98–102. doi: 10.1111/dar.13004. [DOI] [PubMed] [Google Scholar]
- 34.Fentanyl test strip use and overdose risk reduction behaviors among people who use drugs. Vickers-Smith RA, Gelberg KH, Childerhose JE, et al. JAMA Netw Open. 2025;8:0. doi: 10.1001/jamanetworkopen.2025.10077. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.A lot testing protocol for quality assurance of fentanyl test strips for harm reduction applications. Fernando H, Amate A, Hayes KL, et al. Harm Reduct J. 2024;21:152. doi: 10.1186/s12954-024-01058-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Highlighting variability in fentanyl test strip instructions using thematic content analysis. Piercey CJ, Pince CL, Karoly HC. Harm Reduct J. 2025;22:110. doi: 10.1186/s12954-025-01252-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.The use of confidence or fiducial limits. Clopper CJ, Pearson ES. Biometrika. 1934;26:404–413. [Google Scholar]
- 38.Perspectives on rapid fentanyl test strips as a harm reduction practice among young adults who use drugs: a qualitative study. Goldman JE, Waye KM, Periera KA, Krieger MS, Yedinak JL, Marshall BD. Harm Reduct J. 2019;16:3. doi: 10.1186/s12954-018-0276-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Fentanyl test strip distribution acceptability and use among substance-related emergency department visits. Kannam A, Cowett A, Caruso K, et al. Acad Emerg Med. 2024;31:925–928. doi: 10.1111/acem.14934. [DOI] [PubMed] [Google Scholar]
- 40.Clinical toxicity of nitazene detections in two Australian emergency department toxicosurveillance systems. Isoardi KZ, Alfred S, Weber C, et al. Drug Alcohol Rev. 2026;45:0. doi: 10.1111/dar.13998. [DOI] [PubMed] [Google Scholar]
- 41.Prospective cohort study of fentanyl test strip use and distribution in three states: the stay safe study protocol and implementation recommendations for researchers. Childerhose JE, Gelberg K, Vickers-Smith R, et al. Harm Reduct J. 2025;22:146. doi: 10.1186/s12954-025-01277-x. [DOI] [PMC free article] [PubMed] [Google Scholar]