Improved Myocardial Sodium Quantification at 7 T Using Interleaved 23Na/ 1H pTx MRI With Motion and Anatomy‐Based B1 Correction

Laurent Ruck 1,✉, Nico Egger 1, Benedikt Zobler 2, Judith Schirmer 1, Sophia Nagelstraßer 1, Andreas Bitz 3, Tanja Platt 4, Simon Konstandin 5, Christoph Kopp 2, Michael Uder 1, Armin Michael Nagel 1,4
PMCID: PMC13156449  PMID: 41840447

ABSTRACT

Purpose

To improve accuracy and repeatability of myocardial 23Na quantification in cardiac MRI at 7 T by combining retrospective respiratory and cardiac motion correction with a novel anatomy‐based B1 bias field correction.

Methods

In this study, a dual‐nuclear interleaved 23Na/1H MRI sequence at 7 T was applied. Here, 1H MR data enabled automated segmentation of myocardium and blood pool via nnUNet and facilitated respiratory and cardiac motion correction for both contrasts using 1H navigators. For 23Na MRI, a novel anatomy‐based B1 bias field correction was developed, estimating the transmit and receive (B1) field bias from 1H‐derived segmentations and synthetic 23Na images. All correction methods were validated through realistic simulations and in vivo studies. Repeatability was assessed in 10 healthy subjects, and the proposed B1 correction was compared to phantom‐based methods.

Results

Respiratory and cardiac motion correction reduced quantification errors and improved SNR compared to conventional gating. The anatomy‐based B1 correction effectively mitigated errors caused by B1 inhomogeneities and outperformed phantom‐based methods in terms of repeatability. The measured apparent tissue sodium concentration of the myocardium after all corrections (49.5 ± 4.7 mM) was consistent with literature values. Combined motion and B1 corrections improved repeatability, reducing the coefficient of repeatability from 4.9 mM (11.2%) without corrections to 2.0 mM (4.0%) after corrections.

Conclusion

The combination of retrospective motion correction and anatomy‐based B1 bias field correction enables repeatable myocardial sodium quantification at 7 T. The interleaved acquisition facilitates robust segmentation and efficient use of 1H MRI for motion and B1 correction, providing a scalable framework for future clinical research studies.

Keywords: cardiac sodium (23Na) MRI, interleaved dual‐nuclear MRI, respiratory and cardiac motion correction, transmit and receive (B1) correction, ultrahigh field strength (7 T)

1. Introduction

Sodium ions (Na+) play a crucial role in cardiomyocyte excitability and thus in the contractile function of the human heart [1]. Accordingly, alterations of the myocardial tissue sodium concentration (TSC) can serve as indicators for pathological changes [2]. For instance, elevated myocardial TSC has been observed in patients with myocardial infarction [3] and primary hyperaldosteronism [4]. In this context, sodium (23Na) MRI enables noninvasive in vivo measurement of the myocardial TSC [1].

However, 23Na MRI faces significant challenges due to the lower MR sensitivity of 23Na compared to hydrogen (1H), resulting in a signal‐to‐noise ratio (SNR) approximately 6000 times lower in myocardial tissue [1]. To address this, larger voxel sizes, longer acquisition times, and ultrahigh field strengths such as 7 T are commonly used to improve SNR [5]. Due to the limited spatial resolution of 23Na MRI, additional high‐resolution 1H MR images are required for accurate myocardial and blood pool segmentation, enabling partial volume correction (PVC) [6], and reliable TSC quantification [7].

Conventionally, 23Na and 1H MRI are acquired separately, often using different pulse sequences and sometimes even at different field strengths, which increases scan time and introduces susceptibility to physiological changes or subject motion between scans [7]. To overcome these limitations, we employed a time‐efficient, dual‐nuclear interleaved 23Na/1H pTx MR sequence [8] enabling quasi‐simultaneous 23Na and 1H MR data acquisition, while ensuring homogenous 1H excitation via parallel transmission (pTx) [9].

In cardiac 23Na MRI, physiological motion presents additional challenges, as both breathing and cardiac motion can corrupt TSC quantification [7]. While methods such as respiratory self‐gating [7, 10] and external cardiac logging enable retrospective motion‐resolved reconstructions, they reduce acquisition efficiency by discarding large portions of the data. Motion correction using interleaved acquired 23Na and 1H acquisitions has previously been demonstrated for rigid head movements in 23Na brain MRI [11], but is not directly applicable to physiological cardiac and respiratory motion due to temporal resolution constraints and the nonrigid contraction of the heart. Thus, in this work, we implemented a new motion correction approach for interleaved cardiac 23Na/1H MRI, combining rigid respiratory and nonrigid cardiac motion correction based on high‐resolution 1H data.

Another factor influencing the 23Na quantification is the spatial variation of the B1 field generated by the radio frequency (RF) coil. The combined 23Na/1H torso RF coil employed in this work exhibits nonuniform 23Na transmit (B1 +) and receive (B1 −) field profiles [8], which are additionally influenced by body shape and subject positioning. While B1 + can be measured in vivo [12], B1 − must typically be estimated from phantom data [10] or electromagnetic RF field simulations [7], potentially introducing errors. To reduce these effects, we developed a novel transmit and receive (B1) bias field correction for cardiac 23Na MRI. It utilizes a synthetic 23Na prior image, derived from 1H segmentations of the myocardium and blood, to estimate and correct for low‐frequency B1 related signal bias.

The proposed correction methods were first validated in 23Na/1H simulations incorporating realistic respiratory and cardiac motion [13], as well as B1 bias field effects. To assess the impact of motion and B1 corrections on the repeatability of myocardial TSC quantification in vivo, a repeatability study was performed with 10 healthy subjects.

2. Methods

2.1. Simulations

The 4D extended cardiac‐torso (XCAT) simulation phantom [14] is a computational model that simulates realistic anatomical structures and physiological motions based on real human imaging data. It enables the generation of anatomical compartment masks (1 mm isotropic spatial resolution) of the human torso, in particular the heart, during different timepoints of the respiratory and cardiac cycle (Figure 1B). In this study, respiratory and cardiac motion were simulated separately: the heart remained in the end‐diastolic phase during respiratory motion simulation, and in the fully‐exhaled state during cardiac motion simulation. For the respiratory simulation, the maximal heart motion range was set to 18.4 mm in SI, 2.4 mm in AP, and 0 mm in LR direction [15], with a cycle duration of 3.6 s. The cardiac simulation used a 1 s cycle (heart rate of 60 bpm, systole: 400 ms, diastole: 600 ms) (see Figure 1A). Both simulations were sampled at a temporal resolution of 60 ms, yielding 50 distinct cardiac and 60 distinct respiratory motion states, which are repeated periodically.

FIGURE 1.

FIGURE 1

Simulation scheme: XCAT simulation with realistic respiration and heartbeat, sampled at a temporal resolution of 60 ms. A respiratory cycle of 3.6 s was chosen, while the heartbeat was simulated with a cycle duration of 1 s (A). The motion states for respiration and heartbeat were simulated independently. The resulting anatomical XCAT masks (B, only representative states shown here) of all distinct motion states were then used as inputs for 23Na and 1H MR simulations (C). Specific projections of each simulated k‐space were selected—based on their temporal alignment with the respiratory and cardiac cycle—and subsequently combined into a single composite k‐space.

For each motion state, individual 23Na and 1H MR simulations of the radial k‐space data were performed using binary masks from the XCAT phantom. The MR simulations [16] incorporated 23Na and 1H specific tissue parameters (see Tables S1 and S2) as well as sequence parameters similar to the used interleaved sequence (Section 2.2). To simulate the influence of the physiological motion, 23Na and 1H projections were selected from the MR simulations of the different respiratory/cardiac motion states—based on the acquisition timestamps within the sequence scheme—and assembled into a composite radial k‐space for each nucleus (Figure 1C).

2.2. Measurements

2.2.1. Measurement Setup

Measurements were performed at a 7 T whole‐body MR system (MAGNETOM Terra.X, Siemens Healthineers, Erlangen, Germany) using a combined 23Na/1H RF coil setup (Rapid Biomedical, Rimpar, Germany) [8]. The setup consists of a 23Na RF volume coil and two 4Tx/8Rx 1H torso arrays allowing for interleaved 23Na/1H acquisitions and the use of pTx for 1H MRI.

In order to assess the repeatability of the myocardial TSC quantification, 10 healthy subjects (3 female, 27.1±3.5 years, BMI: 24.2±2.3kg/m2) were examined twice using the described MR imaging protocol consisting of the interleaved 23Na/1H pTx sequence and 23Na B1 + mapping. After the first measurement block, the subjects stood up and were then repositioned within the coil. Venous blood samples were collected from each subject before the measurement and the Na+ concentration in the blood serum as well as hematocrit was determined in the laboratory. All measurements were approved by the local ethics committee and written informed consent was obtained from all subjects prior to the examination.

2.2.2. Interleaved 23Na/ 1H pTx Sequence

An interleaved 23Na/1H pTx sequence [8] with a density‐adapted 3D radial [17] readout was utilized to acquire both 23Na and 1H MR data within one measurement: (∆x 23Na)3 = (6 mm)3, (∆x 1H)3 = (2 mm)3, FA23Na = 82°, FA1H = 10°, TR23Na = 60 ms, TE23Na = 1.15 ms, TA = 15 min. A full overview of all sequence parameters is provided in ref. [8]. For 1H excitation, a universal cardiac phase shim [18] was employed. As the data were acquired during free breathing, navigators were added to the sequence to enable retrospective respiratory and cardiac self‐gating [10] (Figure 2A). The navigators consisted of 10 additional data points of the 1H k‐space center signal, called k0, which were acquired before each 1H projection. The radial projections of both nuclei were acquired in a 3D golden angle scheme [19] to allow motion‐sorted reconstruction of the data. For validation purposes, an additional electrocardiogram was recorded simultaneously to the interleaved sequence for 3 of the 10 subjects.

FIGURE 2.

FIGURE 2

Respiratory and cardiac motion correction algorithm for interleaved acquired 23Na/1H MRI. For the extraction of the respiratory and cardiac signal, additional 1H navigators (called k0) were added to the interleaved acquisition scheme (A). For the respiratory motion correction (B), the 1H navigators are utilized to extract a respiratory signal (I). Based on the filtered respiratory signal, the acquired 23Na and 1H data are assigned into eight different respiratory bins allowing to perform a motion‐sorted reconstruction (II). The motion‐sorted 1H images are then rigidly co‐registered to the exhaled image (here: Bin 1) within the heart region (red outline), resulting in three translational and three rotational transformation parameters for each bin (III). Finally, these parameters are used to correct the 1H as well as the 23Na data directly in the k‐space (IV, reconstructed images shown). The cardiac motion correction (C) is performed after applying the respiratory motion correction on the 23Na and 1H data. Similar to the respiratory motion correction, the cardiac motion signal is retrieved from the 1H navigators (I). Based on the maxima of the cardiac motion signal, the 23Na and 1H projections are sorted into ten different cardiac phases allowing to reconstruct cardiac motion‐sorted images (II). The ten cardiac phases are clustered into a diastolic and systolic phase based on the 1H SSIM within the heart region (III). Then the diastolic and systolic 1H images were B1 bias corrected and the systolic 1H image was nonrigidly co‐registered to the diastolic image. Subsequently, the same deformation field was applied to the systolic 23Na image. Adding up the diastolic and co‐registered systolic 23Na and 1H images results in cardiac motion‐corrected 23Na and 1H images (IV).

All data were reconstructed on a standalone PC (Intel Core i7‐9700K CPU 8 cores @ 3.60 GHz with 128 GB RAM) using a custom‐written MATLAB (The Math‐Works, Natick, MA, USA; version 2020b) script, including the application of a Hamming filter, a nonuniform fast Fourier transform [20] as well as a correction for gradient field nonlinearities [21]. 23Na images were zero‐filled to 2 mm isotropic resolution to match the 1H MRI. For analyses with myocardium and blood masks, simulated 23Na images were zero‐filled to 1 mm isotropic to match the XCAT mask resolution.

2.3. Motion Gating and Corrections

To extract respiratory and cardiac motion during the interleaved 23Na/1H acquisition, the k0 navigator of the 1H signal was used. Ten central k‐space samples (Tdwell=7.6μs) per 1H projection were averaged, and gradient timing mismatches were corrected [22]. As 1H data were acquired with alternating TRs (13.08 and 20.76 ms), the navigator signal was interpolated to a uniform 15 ms resolution for easier processing.

2.3.1. Respiratory Motion Gating and Correction

Respiratory motion was estimated by filtering the k0 navigator signals of all 16 1H receive channels using a moving average filter with a 1.5 s window. This suppresses cardiac‐related signal fluctuations, as their cycle duration typically lies below this threshold (heart rates > 40 bpm). Based on the navigator signal of each channel, images corresponding to the fully‐inhaled and fully‐exhaled states, that is, bins 1 and 8, were reconstructed and rigidly co‐registered within the heart region using Elastix [23]. The channel leading to the maximum displacement in the superior–inferior (SI) direction between these two respiratory phases was selected as the most sensitive to respiratory motion and was used as the final respiratory navigator channel (Figure 2A—I).

The extracted navigator signal was used to bin both 1H and 23Na projections into Nresp respiratory states, each containing an equal number of projections. These bins were reconstructed to generate Nresp respiratory‐gated images (Figure 2A—II). All gated 1H images were then rigidly co‐registered to the fully‐exhaled state, and the resulting transformation parameters (three translations, three rotations, Figure 2A—III) were applied to the corresponding k‐space subsets of both 1H and 23Na data [11, 24], yielding respiratory motion–corrected images reconstructed in the exhaled state (Figure 2A—IV).

2.3.2. Cardiac Motion Gating and Correction

Cardiac motion correction was performed after respiratory motion correction. For each 1H receive channel, the frequency power spectrum of the k0 navigator signal was computed and denoised using a Savitzky–Golay filter [25] (sgolayfilt in MATLAB's Signal Processing Toolbox). Within the expected cardiac frequency range (0.5–2 Hz), the dominant peak was identified using MATLAB's findpeaks function (Signal Processing Toolbox), and its prominence (pcard), center frequency (fcard), and width (wcard) were extracted. The total power of the remaining spectrum (Snoise) was calculated, and the channel with the highest prominence‐to‐noise ratio (pcard/Snoise) was selected.

This channel's navigator signal was bandpass‐filtered using a subject‐specific Butterworth filter [26] with cutoffs fcut=fcard±2·wcard. The resulting peaks served as cardiac triggers (Figure 2B—I), which were validated against ECG signals in three subjects (Figure S1).

1H and 23Na projections were sorted into 10 cardiac phases based on the time since the last trigger. Structural similarity indices [27] (SSIM) of cardiac‐gated 1H images (Figure 2B—II) were used for k‐means clustering [28, 29] into systolic and diastolic phases. The clustered projections were reconstructed into diastolic and systolic gated 1H and 23Na images (Figure 2B—III). Subsequently, nonrigid registration between the B1 bias corrected (see Section 2.7) systolic and diastolic 1H images was performed using Elastix [23] to estimate cardiac motion fields from the 1H images. These motion fields were then applied to the systolic 23Na and 1H images. Finally, averaging of the diastolic and motion‐corrected systolic images yielded cardiac motion–corrected 23Na and 1H images in the diastole (Figure 2B—IV).

Respiratory motion‐sorted reconstruction into 8 bins required ∼1.5 h, with motion parameters estimated in 3 min. Cardiac motion‐sorted reconstruction into 10 and 2 bins took ∼2 h and ∼1 h. Cardiac motion correction was executed in ∼3 min.

2.4. Segmentation

For the in vivo data, initial manual segmentations of the left myocardium and blood pool were performed on the interleaved acquired 1H images with B1 bias correction (see Section 2.7) but without any motion corrections. These 20 manual segmentations (10 subjects, each measured twice) were then used to train a 3D nnUNet [30] with both 23Na and 1H images as inputs, enabling automatic prediction of myocardium and blood pool for motion‐corrected 23Na and 1H images. The predicted masks of the blood pool were truncated above the myocardial voxel with the highest spatial position in the superior–inferior direction to exclude the great vessels in the evaluation.

2.5. Partial Volume and Relaxation Correction

To improve the accuracy of the 23Na quantification, a PVC [6] was applied. Point‐spread functions (PSFs), including T 2* values from literature (Table S1), were convolved with myocardial and blood masks to obtain region‐spread functions (RSFs). These were used to correct for the 23Na signal spillover between compartments.

Additionally, relaxation effects on the 23Na signals were corrected using the formula

Scorr(x→)=Suncorr(x→)r·exp−TET2,s*+(1−r)·exp−TET2,l*·1−cos(FA)exp−TRT11−exp−TRT1 (1)

and the corresponding relaxation parameters for myocardium and blood (see Table S1). The flip angle (FA) used in this formula was the subject‐specific median FA within the blood pool, measured with the actual flip‐angle imaging (AFI) sequence (see Section 2.7).

2.6. TSC Quantification

The sodium concentration cNa+,blood in whole blood was calculated using the blood serum sodium concentration cNa+,serum and hematocrit values obtained from laboratory analysis:

cNa+,blood=(1−hematocrit)·cNa+,serum (2)

Next, the ventricular blood pool served as an internal reference for 23Na MRI to derive quantitative myocardial apparent tissue sodium concentrations (aTSC) by linear calibration:

aTSCmyo=cNa+,blood·S‾myoS‾blood (3)

The term “apparent” accounts for potentially remaining relaxation weighting, as suggested by Stobbe and Beaulieu [31].

2.7. B1 Correction

To correct for 23Na signal variations caused by the spatially varying B1 + and B1 − field distributions of the employed RF coil, we developed a novel B1 bias field correction method. Previous approaches assumed a homogeneous signal distribution across the entire organ [32, 33], which is sufficient for qualitative 1H images but inadequate for quantitative 23Na MRI due to differing sodium concentrations in the blood pool and myocardium. In order to address this, we incorporated RSFs of the blood pool and myocardium into our B1 bias correction algorithm:

  1. RSFs of myocardium and blood were computed from 1H segmentations, assuming homogeneous signal and accounting for partial volume blurring.

  2. A synthetic prior image was generated by adding the RSFs with an assumed myo‐blood ratio rassumed.

  3. The prior and original 23Na images were masked to the heart and low‐pass filtered using imgaussfilt3 (Image Processing Toolbox) with standard deviation σ23Na = 10 mm (optimization in Figure S2).

  4. The B1 bias field was estimated as
    B1,bias(x→)=Sorig,filt(x→)Sprior,filt(x→)≈sin(FA(x→))·B1−(x→) (4)

    yielding an approximation of the low‐frequency bias, arising from the combined effects of B1 + and B1 − inhomogeneities.

  5. The B1 bias correction was applied voxel‐wise in the heart:
    SB1,corr(x→)=Suncorr(x→)B1,bias(x→) (5)
  6. PVC was then applied to the mean myocardium and blood 23Na signals of the B1 corrected image to obtain the measured myo‐blood ratio rmeasured.

Because the true ratio is unknown, steps (2–6) were iterated for rassumed∈ [0.3, 0.7] (step size 0.005), and the value yielding the smallest difference rmeasured−rassumed was selected. For the 1H B1 correction, only steps (2–5) were used, with a homogeneous support region instead of a synthetic prior and a Gaussian kernel with standard deviation σ1H = 12 mm. A detailed overview of the implemented B1 bias corrections can be found in Figure S3. The iterative 23Na B1 bias correction of the in vivo measurements took around 5 min per subject and the 1H B1 bias correction below 1 s.

In addition, fast B1 + mapping for the 23Na RF coil was performed using the AFI [34] sequence ((Δ x)3 = (12 mm)3, FA = 60°, TR1/TR2 = 12/48 ms, TE = 1.05 ms, TA = 3 min, randomized xy‐spoiling [35]). The resulting FA maps were smoothed using a Gaussian kernel (σAFI = 14 mm) and employed to estimate the mean absolute FA within the blood pool for relaxation correction (Equation 1). The commonly used double angle method requires a long TR (∼250 ms) and two different flip angles (45° and 90°), reducing acquisition efficiency and potentially introducing errors from tissue‐specific relaxation [36]. This is not the case for the AFI sequence using only one pulse for excitation as well as shorter TRs (12 and 48 ms), which enables more efficient k‐space sampling (Figure S4). Accelerating the AFI acquisition from 15 to 3 min had minimal impact on the FA maps (Figure S4).

The proposed B1 bias field correction method was compared to two other commonly used methods: (1) B1 − and B1 + maps derived from phantom measurements [8] and (2) phantom‐derived B1 − map combined with in vivo B1 + maps obtained from the AFI measurements.

2.8. Evaluation Procedures

23Na and 1H simulations were performed 20 times with different spatially distributed noise approximating the in vivo SNR. For simulated 23Na MRI, the voxel‐wise SNR was computed from the mean and standard deviation across all 20 simulations [37]. The blood SNR was obtained by averaging over the entire blood pool, and myo–blood signal ratios were computed by dividing the mean myocardium 23Na signal by the mean 23Na blood signal, calculated using masks from the corresponding ground truth state. Additionally, the normalized root mean square error (NRMSE) relative to the ground truths was calculated within the heart. The ground truths were defined as simulated images representing the fully‐exhaled state for respiratory motion, the end‐diastolic state for cardiac motion, and a B1 bias–free image for B1 correction.

To evaluate the 23Na B1 bias correction across different myocardial TSC values, simulations were performed for TSCs between 30.1 and 55.9 mM (blood fixed at 81 mM), each repeated 20 times with different noise distributions. A B1 bias field was generated from phantom‐measured B1 +/B1 − maps [8] using B1,bias(x→)≈sin(FA(x→))·B1−(x→) and applied to the GT images to create biased images. Three heart positions were simulated by shifting the bias field relative to the images. The 23Na B1 bias correction was then applied, and corrected and uncorrected images were compared to the GT using NRMSE and myo‐blood ratios.

For in vivo 23Na MRI, SNR values were estimated for myocardium and blood using background noise region‐of‐interests [37]:

SNRmyo/blood=meanmyo/bloodstdnoise·4−π2 (6)

This analysis was performed only for rigid respiratory motion correction, as it is unsuitable for nonrigid cardiac motion correction, which alters noise properties.

To evaluate the repeatability of the in vivo myocardial aTSC quantification, the coefficient of repeatability (CR) was calculated using the formula

CR=2.77·110∑i=110∑k=12aTSCi,k−aTSC‾i2 (7)

with aTSC‾i=aTSCi,1+aTSCi,22 as described in ref. [38]. The relative CR (rCR) was obtained by dividing the CR by the global mean aTSC value. Additionally, the results were visualized using Bland–Altman plots [39].

3. Results

3.1. Validation of Respiratory and Cardiac Motion Correction in Simulation Study

The respiratory motion correction was validated using 23Na and 1H simulations containing realistic respiratory motion. Representative reconstructed images are shown in Figure 3A. The 1H images demonstrate that gating data into more respiratory bins (e.g., 8 instead of 2)—resulting in fewer projections per bin—leads to reduced respiratory motion blurring compared to the uncorrected image. However, this also increases undersampling artifacts and reduces SNR.

FIGURE 3.

FIGURE 3

Validation of the respiratory motion correction in simulation study (respiratory motion: 18.1/2.4/0 mm in SI/AP/LR directions; no cardiac motion: End‐diastolic motion state used). Exemplary 23Na and 1H images are shown in (A). The use of respiratory gating to reduce respiratory motion effects (B, C) results in a decrease in SNR (D). Motion correction shows improved compensation of the respiratory motion compared to respiratory gating (lower NRMSE and myo‐blood ratio closer to GT), while showing no decrease in SNR (D). Performing the respiratory motion correction with eight respiratory bins resulted in the lowest NRMSE (B) and a myo‐blood ratio closest to the GT (C).

In contrast, the proposed respiratory motion correction further decreases quantification errors in 23Na MRI, achieving lower NRMSE values (B) and myo‐blood ratios (C) closer to the GT (fully‐exhaled motion state) than respiratory gating, while maintaining the original SNR (D). Based on the NRMSE analysis, using 8 respiratory bins for motion correction resulted in the lowest NRMSE (B), indicating the best quantitative agreement to the GT. Consequently, eight respiratory bins were used for the respiratory motion correction of the in vivo data.

Cardiac motion correction was validated in a simulation study with realistic cardiac motion (Figure 4). Both diastolic‐gated as well as motion‐corrected images (A) showed a reduction in NRMSE (B) compared to the uncorrected image and 23Na myo‐blood signal ratios (C) closer to the GT, demonstrating decreased quantification errors caused by cardiac motion. In addition, applying the cardiac motion correction increased the SNR by 25% compared to the diastolic‐gated images (D).

FIGURE 4.

FIGURE 4

Validation of the cardiac motion correction in simulation study (no respiratory motion: Fully‐exhaled motion state used; cardiac motion: 60 bpm, systole: 0–400 ms, diastole: 400–1000 ms). Exemplary 23Na and 1H images are shown in (A). Cardiac diastolic gating and motion correction into the diastolic phase reduces the NRMSE (B) as well as the deviation of the myo‐blood ratio to the GT (C) compared to the uncorrected image. Cardiac motion correction enables this while providing a higher SNR than cardiac gating (D).

3.2. Validation of B1 Bias Field Correction in Simulation Study

The B1 bias field correction was validated in a simulation study incorporating three different B1 bias field distributions within the heart (Figure 5). After applying the proposed B1 bias correction, the 23Na signal appeared more homogeneous across the heart (A). NRMSE values (B) within the heart region as well as myocardium‐blood ratios of the 23Na signal (C) confirmed that the B1 bias correction effectively reduced quantification errors for all three heart positions and all simulated myocardial TSC values (30.1–55.9 mM). The 23Na B1 bias correction also demonstrated robust performance in simulations with myocardial T*2,short deviating from the assumed 3 ms (Figure S5—I) as well as in simulations with hypertrophied myocardium (Figure S6B—II).

FIGURE 5.

FIGURE 5

Validation of the B1 bias field correction in simulation study (no respiratory motion: Fully‐exhaled motion state used; no cardiac motion: End‐diastolic motion state used). Note that since the B1 bias field correction was applied only within the heart, the B1 corrected images were masked accordingly. After applying B1 bias correction, the 23Na signal distribution appeared more homogeneous within the heart (A, images shown for 43 mM myocardial aTSC). Across all simulated TSC values and all three positions, the B1 correction consistently reduced NRMSE over the heart (B). Additionally, for every position and TSC value, the myo‐blood ratios (masks see Figure 4) after B1 correction were closer to the GT than those obtained without correction (C).

3.3. In Vivo Application of Respiratory and Cardiac Motion Correction

Figure 6 illustrates the impact of respiratory motion correction on in vivo 23Na and 1H MRI for two subjects with different breathing depths. In the deep‐breathing subject (I; 24.4 mm displacement), uncorrected images (A) show strong blurring and 1H signal dropouts, whereas these effects are less pronounced for the subject with average breathing (II). Respiratory gating into eight bins (B) reduces blurring but uses only one‐eighth of the data, lowering SNR. In contrast, respiratory motion correction (C) aligns all projections to the exhaled state, improving sharpness while preserving SNR. Across all 20 measurements, motion correction increased SNR by factors of 3.56±0.37 in the myocardium and 3.57±0.36 in the blood pool compared to gated exhaled state images. Difference maps (D) highlight the stronger effect for the deep‐breathing subject.

FIGURE 6.

FIGURE 6

Influence of respiratory motion gating and correction on 23Na and 1H images for the in vivo measurements of two subjects with different breathing depth. For the subject with deep breathing (I), the uncorrected image (A) exhibits severe motion blurring and even signal dropouts in the 1H MRI (yellow arrow). With a more average breathing depth (II), these artifacts are less pronounced. Respiratory‐gated reconstruction (B) reduces motion‐induced blurring in both subjects and visualizes the respiratory displacement between the fully‐inhaled and fully‐exhaled bins. Applying respiratory motion correction (C) diminishes motion‐related errors while preserving higher SNR than respiratory gating. The quantitative impact of the respiratory motion correction is evident in the difference maps (D) and is particularly pronounced for the subject with deep breathing (I).

Figure 7 illustrates the influence of cardiac motion on the 23Na and 1H images. Note that respiratory motion correction into the fully‐exhaled state has already been applied to the 23Na and 1H k‐space data prior to addressing cardiac motion. Compared to electrocardiograms recorded during measurements of three subjects, the cardiac self‐gating signal derived from the interleaved 1H data showed comparable trigger consistency (Figure S1).

FIGURE 7.

FIGURE 7

Influence of cardiac motion gating and correction on 23Na and 1H images for one in vivo measurement. Note that the cardiac gating and motion correction is applied after respiratory motion correction in the fully‐exhaled state. The correct assignment of the 23Na and 1H data into a systolic and diastolic phase is demonstrated by the increased width (black bar for reference) of the left ventricle in the diastolic compared to the systolic phase (B). The cardiac motion‐corrected image (C) displays good accordance to the cardiac‐gated diastolic image. The difference between the cardiac motion‐corrected and uncorrected image (A) shows quantitative changes in the region of the myocardium and blood pool (D).

Cardiac gating divided the data into 10 distinct phases, which were then clustered to reconstruct images corresponding to systolic and diastolic cardiac phases (B). The correct assignment of these phases is evident from the dilation of the left ventricle, appearing contracted during systole compared to diastole. Applying a nonrigid motion correction to the systolic image results in a cardiac motion‐corrected image in the diastolic phase (C). Compared to the uncorrected 23Na image (A), signal differences are visible in the myocardium and blood of the left ventricle (D).

3.4. In Vivo Application of B1 Bias Field Correction

Figure 8 shows the results of the B1 bias field corrections, applied to in vivo 23Na and 1H MRI of one subject. For 1H MRI, a homogeneous mask was created by thresholding and used as prior information (B). This allows obtaining a rough estimate of the 1H bias field (D), although some anatomical structures remain visible. Nonetheless, the B1 corrected 1H image (C) displays a more homogeneous signal distribution across the entire body compared to the uncorrected image (A). In contrast, for 23Na MRI we used a synthetic prior image, based on anatomical 1H segmentations (B). The estimated 23Na B1 bias field (D) exhibits a smooth profile, which does not contain anatomical structures and only corrects the low‐frequency variations of the B1 bias field arising from B1 + and B1 − inhomogeneities. The B1 corrected 23Na image (C) shows a more homogeneous 23Na signal distribution across the heart.

FIGURE 8.

FIGURE 8

Influence of B1 bias field correction on 23Na and 1H images for one subject. (A) shows the respiratory and cardiac motion‐corrected 23Na and 1H image before B1 bias correction. For 1H MRI, a simple homogeneous support region (B) was used as anatomical prior to extract the 1H B1 bias field (D). After correction (C), the 1H image exhibits a more homogeneous signal distribution within the heart. For 23Na MRI, the prior information was generated based on 1H segmentations of the myocardium and blood (B). The estimated 23Na B1 bias field contained no anatomical structures (D). The B1 correction compensates for the signal loss on the top left of the heart (white arrow), leading to a more homogeneous signal distribution over the heart (C). Note that since the 23Na B1 bias field correction was applied only within the heart, the B1 corrected 23Na image (C) and the 23Na bias field (D) were masked accordingly.

3.5. Quantification and Repeatability Study

Figure 9 shows the motion‐ and B1‐corrected 23Na and 1H images of one subject, along with the corresponding segmentations of myocardium and blood pool predicted by the nnUNet. Due to the interleaved acquisition scheme, both images are intrinsically aligned.

FIGURE 9.

FIGURE 9

Overview of interleaved 23Na and 1H MRI of one subject, after application of respiratory and cardiac motion correction as well as B1 corrections. Since the 23Na B1 bias field correction was performed only within the heart region, the B1 corrected 23Na image (A) was masked accordingly. 1H images (B) are overlaid with myocardium (red) and blood pool (blue) segmentations predicted by the nnUNet (C). Note that the blood masks are constrained to the uppermost voxel of the myocardium in SI direction (cut‐off visible in coronal and sagittal view). Due to the interleaved acquisition scheme, both 23Na and 1H MR images are intrinsically co‐registered.

Figure 10B compares the repeatability of different B1 correction methods (Figure S7), with all other corrections applied consistently. Using phantom‐derived B1 +/B1 − maps resulted in worse repeatability (rCR = 10.0%) than no correction (rCR = 8.9%). A subject‐specific in vivo B1 + map combined with the phantom B1 − map improved rCR to 5.1%, though outliers were observed (e.g., subject 8, Figure 10A). The best repeatability was achieved using the proposed anatomy‐based bias field correction (rCR = 4.0%).

FIGURE 10.

FIGURE 10

Repeatability study: Influence of B1 correction methods (A, B) applied in combination with respiratory and cardiac motion correction as well as PVC and relaxation correction. The aTSC values of measurement 1 and 2 are shown in different box plots for each correction method (A). While the use of in vivo B1 + maps improved repeatability, the proposed anatomy‐based B1 bias field correction achieved the best repeatability (B). Myocardial aTSC values of the in vivo repeatability study after application of the applied correction methods (C). For B1 correction, the proposed B1 bias field method was used. Each correction method is applied cumulatively from left to right, with each additional method building upon the previous ones. B1 bias field correction and partial volume correction had the largest impact on the quantitative aTSC values, while respiratory and cardiac motion correction resulted in smaller aTSC changes. The coefficients of repeatability (D) show that respiratory and cardiac motion correction, the proposed B1 correction and the B1 + dependent relaxation correction improved repeatability, while PVC led to a slight reduction. The overall improvement in repeatability is also visible in the Bland–Altman plots of the aTSC values for no corrections and all correction applied (E).

Figure 10C shows subject‐wise myocardial aTSC values from the repeatability study (Table S3). Respiratory motion correction impacted aTSC in only a few subjects (e.g., 5 and 9), while cardiac motion correction had a small effect overall. In contrast, B1 bias correction increased mean aTSC by 16%. PVC led to a reduction of 6%, while relaxation correction slightly increased it by 3%. After all corrections, the mean myocardial aTSC across all subjects was 49.5 ± 4.7 mM.

Figure 10D summarizes the impact of each correction step on repeatability. Without any corrections, the CR was 4.9 mM (rCR = 11.2%). Applying respiratory motion correction reduced the CR to 3.5 mM (rCR = 8.1%), and adding cardiac motion correction further improved it to 2.8 mM (rCR = 6.3%). The most substantial improvement resulted from B1 bias field correction, lowering the CR to 1.8 mM (rCR = 3.6%). PVC caused a slight increase to 2.1 mM (rCR = 4.3%), while subsequent relaxation correction yielded a final CR of 2.0 mM (4.0%) with a 95% confidence interval of 1.4–3.7 mM (rCR = 2.8%–7.3%). The overall improvement in repeatability is also illustrated in the Bland–Altman plots (Figure 10E).

4. Discussion

In this work, we demonstrated the feasibility of retrospective respiratory and cardiac motion correction combined with a novel anatomy‐based B1 bias field correction for interleaved 23Na/1H cardiac MRI at 7 T. These methods significantly improved the repeatability of myocardial aTSC quantification.

4.1. Respiratory and Cardiac Motion Correction

Breath‐holding is not feasible in 23Na MRI due to low SNR and long scans, so respiratory motion is usually recorded and data is retrospectively gated using external devices like respiratory belts [10]. Conventional ECG‐based cardiac gating is unreliable at 7 T due to magnetohydrodynamic effects, though acoustic gating [7] can mitigate this at the cost of additional hardware.

Respiratory and cardiac motion can be extracted directly from 1H MR data without external devices [40]. In 23Na MRI, self‐gating has mainly been applied for respiratory motion [10], while cardiac gating remains challenging due to low SNR. To overcome this, we incorporated 1H navigators into the interleaved sequence. These navigators enabled respiratory and cardiac self‐gating from 1H data, which could be directly applied to the 23Na data due to the interleaved acquisition. This achieved finer respiratory resolution than prior 23Na MRI studies [10] and successfully incorporated cardiac self‐gating for 23Na. However, gating reduces acquisition efficiency and SNR, being particularly problematic for 23Na MRI.

Instead, we applied retrospective respiratory and cardiac motion correction, fully utilizing all 23Na data. Extending prior work correcting rigid head movements in brain 23Na/1H data [11], respiratory motion was corrected in k‐space assuming rigid heart motion. In contrast, cardiac motion was addressed through nonrigid registration of diastolic and systolic 1H images, with corresponding deformation fields applied to 23Na images.

Simulations confirmed that the implemented respiratory and cardiac motion corrections reduced quantitative errors and improved SNR compared to motion gating. In vivo, respiratory motion correction decreased quantitative errors and especially preserved 1H image quality during deep breathing (Figure 6A). The SNR gain over gating was confirmed, and measured respiratory displacements (Table S3) matched literature values [41, 42]. While cardiac motion correction had a smaller quantitative impact, it nonetheless contributed to improved repeatability.

Overall, this integrated motion correction strategy represents a significant advance, enabling more reliable and efficient myocardial sodium quantification at 7 T. By fully leveraging interleaved 23Na/1H data and avoiding the SNR penalties of gating, our method paves the way for improved clinical research applications of cardiac 23Na MRI.

4.2. B1 Correction

Myocardial TSC quantification is also affected by the nonuniform B1 field distribution of the RF coil used in this work [8]. While the B1 + transmission field can be measured in vivo, the B1 − receive field is typically estimated via phantom scans or electromagnetic field simulations based on representative body models [43], both potentially introducing errors.

For B1 bias correction of the qualitative 1H MRI, we assumed a homogeneous support region following previous works that treated the organ signal as uniform [32, 33]. While this B1 correction is simple, fast, and relatively robust, it reduces the contrast between myocardium and blood. However, for our application, the quality of the B1 corrected 1H images was sufficient to enable nnUNet based segmentation as well as estimation of the motion fields for cardiac motion correction.

To address B1 inhomogeneities in quantitative 23Na MRI, we proposed a novel anatomy‐based B1 bias field correction approach. Our method aims to estimate the low‐frequency B1 bias field by comparing the measured 23Na image with a synthetic one.

However, unlike the previous method assuming a homogenous support region, we incorporate anatomical information: Segmentations from interleaved 1H images are used to synthesize a 23Na image with assumed constant TSCs, including PSF convolution. Low‐pass filtering of both the original and synthetic images then isolates the B1 bias field.

With the true myocardial sodium concentration unknown, we iteratively tested assumed myo‐blood ratios. B1 correction and PVC were applied for each, selecting the assumed ratio best matching post‐correction measured ratio as optimal. Simulations confirmed effective B1 correction across varied simulated myocardial‐to‐blood TSC ratios and B1 fields (Figure 5). However, PVC errors may affect the selection of the optimal ratio and thus B1 estimates. Another limitation is the reliance on accurate anatomical segmentations. Although Gaussian filtering reduces high‐frequency errors, segmentation inaccuracies can still affect the correction (Figure S6—A/BII). Nevertheless, applying erosion or dilation to the in vivo myocardial segmentations had little effect on repeatability and thus on the performance of the in vivo B1 correction (Figure S8).

Lacking electromagnetic simulations for the coil used, we compared our method to two alternatives: (1) phantom‐derived B1 − and B1 + maps and (2) phantom B1 − combined with in vivo B1 + maps. Phantom‐based correction performed worse than no correction, highlighting the importance of subject‐specific B1 correction. Using subject‐specific B1 + maps improved repeatability but remained inferior to our method. This may be because our method accounts for subject‐specific anatomy and coil positioning—both of which affect the B1 field distribution—whereas using a phantom‐based B1 − map does not. Applying the AFI method to measure the 23Na B1 + maps in vivo results in minor deviations of the measured B1 + field depending on the T 1 relaxation time of the tissue (Figure S9A). While an additional correction of these effects led to a slight reduction of the quantified aTSC values, the coefficients of repeatability were not influenced (Figure S9B). While the AFI method was suited to be a good choice for the application in the human torso, depending on the application, other B1 + mapping methods might be beneficial [12].

In summary, the proposed method outperforms the other tested methods in terms of repeatability. Moreover, it eliminates the need for additional B1 + mapping—provided the flip angle has negligible influence on the 23Na T1 relaxation correction (Equation 1). Additionally, it avoids the need for a specialized phantom with body‐matched dielectric properties to measure the B1 − field, which may not be available at all sites.

4.3. Quantification and Repeatability

Lott et al. [7] quantified the myocardial TSC in four healthy subjects using respiratory and cardiac gating, B0 and B1 correction, as well as PVC. Similarly, cardiac and respiratory motion gating/correction had minor effects on aTSC values in our study. However, our 23Na volume coil [8] exhibits greater B1 field inhomogeneity than theirs [10], which likely explains the larger impact of the B1 correction observed here. Additionally, PVC reduced myocardial mean aTSC by only 6% in our work, compared to 30% reported by Lott et al. This discrepancy may be partly attributed to errors in the affine coregistration between the 1H and 23Na images acquired at 3 and 7 T, respectively. Our interleaved acquisition scheme intrinsically aligns 23Na and 1H images, avoiding such registration issues. Inaccuracies in the nnUNet‐predicted masks used in this study may also have contributed to the observed differences (Figure S8). Regarding the interleaved acquisition scheme, phantom measurements showed no relevant influence on the 23Na MR signal [8].

In our study, a mean myocardial aTSC of 49.5±4.7mM was measured for the entire myocardium of the left ventricle. This is in good accordance with extracellular and intracellular volume based estimates (43mM [1]) as well as measured values reported by Ouwerkerk et al. [44] (free wall: 45±4mM, septum: 56±13mM), which were calculated from the original concentrations in μmol/g using a specific gravity of 1.05g/mL for myocardium [44]. Lott et al. reported a lower value of 29±3mM. In addition to coregistration errors, this discrepancy may also be partially caused by residual relaxation weighting due to their shorter TR of 21 ms. Notably, the exact myocardial aTSC values are currently unknown, and our results further demonstrate that segmentation inaccuracies (Figure S6) and deviations in true relaxation times (Figure S5) can influence quantification.

To further improve segmentation consistency, we trained a nnUNet to predict myocardium and blood pool masks using both contrasts. This approach enabled automatic segmentation of motion‐corrected images of the training data and may reduce inter‐reader variability, indicated by a high average DICE similarity score between the segmentations of both measurements (Figure S10). Future work should evaluate the nnUNet's performance on unseen data and expand training to include patient datasets.

Evaluation of the in vivo repeatability study showed that the implemented respiratory and cardiac motion correction combined with the proposed B1 correction substantially improved repeatability. PVC slightly reduced it, likely because its segmentation‐derived correction factors amplify segmentation differences. After all corrections, myocardial aTSC variation, caused by measurement noise and post‐processing, is expected to remain below 2.0 mM with 95% confidence. In addition, the time‐efficient interleaved sequence reduced the total scan time to 18 min—significantly shorter than previous 7 T cardiac 23Na MRI protocols requiring at least 47 min [7, 43]. Together, the achieved precision and reduced scan time support reliable application in future patient studies.

4.4. Limitations

This study has several limitations. Cardiac motion correction was restricted to two phases (diastole and systole), which may not fully capture the complexity of cardiac motion. Advanced reconstruction techniques [45, 46] reducing undersampling artifacts could enable the inclusion of more cardiac phases, further mitigating cardiac motion effects. If working properly, the usage of a simultaneously logged ECG signal for the cardiac binning could improve binning precision, as the ECG provides a higher temporal resolution of 2.5 ms compared to the ≈15 ms of the self‐gated cardiac navigator. Moreover, respiratory motion was assumed rigid, despite containing nonrigid components. Future approaches incorporating nonrigid respiratory models may further improve the correction. Patients may show irregular breathing and heart rates. While respiratory motion can be extracted and corrected regardless of regularity (Figure S11), cardiac irregularities may require adapting the subject‐specific Butterworth filter or using a broader cardiac frequency window (0.5–2 Hz). For severe heart‐rate variations, cardiac binning could alternatively rely on ECG or pulse‐meter signals.

The proposed 23Na B1 bias correction assumed a homogeneous myocardial sodium concentration, which is reasonable for conditions such as Conn syndrome or chronic kidney disease. In myocardial infarction, however, infarcted tissue has altered sodium levels, which may introduce errors in the B1 bias correction. In such cases, we recommend extending the algorithm to use three segmentations—healthy myocardium, infarcted myocardium, and blood—and to optimize both corresponding myo‐blood ratios.

PVC only accounted for the myocardium and blood pool, omitting surrounding tissues. Although Lott et al. [7] suggested that their influence is minor, this simplification may still introduce small quantification errors. The 2 mm isotropic spatial resolution of the interleaved 1H MRI may also limit the mitigation of partial volume effects.

Due to specific absorption rate and RF power constraints, a relatively long TE of 1.15 ms had to be used resulting in partial signal loss of the short T 2* component. Moreover, the TR of 60 ms introduced residual T 1 weighting. Both relaxation effects were corrected using literature values; however, deviations from the assumed relaxation times can lead to errors (Figure S5—II). Thus, future studies could improve accuracy by measuring T 2* and T 1 in vivo. Theoretically, quantification may be affected by additional spin‐3/2 specific relaxation effects during the relatively long RF pulse of 2 ms [31, 36]. However, spin‐3/2 dynamic simulation [47, 48] for the assumed relaxation times showed that the resulting transverse magnetization at TE = 1.15 ms depends only marginally on the RF pulse characteristics. Moreover, relaxation effects are adequately corrected by the applied biexponential relaxation model. Nevertheless, excitation‐related signal losses may become more significant in tissues with shorter relaxation times and/or quadrupolar splitting and should be investigated for the specific application.

5. Conclusion

This work presents a comprehensive framework for quantitative myocardial 23Na MRI using a dual‐nuclear interleaved 23Na/1H sequence at 7 T. Respiratory and cardiac motion correction were implemented based on the interleaved 1H data to retrospectively correct the 23Na images. Additionally, we introduced a novel anatomy‐based 23Na B1 bias field correction using 1H‐derived segmentations. These corrections substantially improved measurement repeatability (CR = 2.0 mM, rCR = 4.0%) and yielded a myocardial aTSC of 49.5±4.7mM, consistent with values reported in literature. Combined with the time‐efficient interleaved acquisition and automated segmentation using a nnUNet model, this pipeline provides a robust and scalable foundation for patient‐specific myocardial sodium quantification.

Funding

This work was supported by Deutsche Forschungsgemeinschaft (449552397, 509149993) (TRR 374).

Supporting information

Table S1: Tissue parameters for 23Na simulation. Assumed values are marked by (*). The myocardial TSC of 43 mM was derived from 41 μmol/g [1] by applying a specific gravity of 1.05g/mL for myocardium [44].

Table S2: Tissue parameters for 1H simulation. Assumed values are marked by (*).

Figure S1: Comparison of cardiac self‐gating signal (SG), derived from the interleaved 1H data, and simultaneously acquired electrocardiogram (ECG) for three subjects. Due to the filtering in the post‐processing of the SG signal, the triggers of SG tend to occur in mid‐diastole, while scanner‐predicted ECG triggers (M1, red circles) mark the end of the diastolic phase. Additionally, the minima of the ECG signal were used as triggers (M2, blue circles). For subject 6 (a), the histogram of the temporal difference ΔT between consecutive triggers showed good agreement between SG and ECG independent of the selected triggers. However, for subjects 7 (b) and 8 (c) scanner‐based ECG triggering (M1) resulted in trigger errors. In contrast, minima based ECG trigger (M2) and SG showed consistent triggering for all three subjects.

Figure S2: Influence of the standard deviation σ of the Gaussian low‐pass filter in the 23Na B1 bias correction (see Figure S3). For this optimization, the phantom B1 bias field and the simulated 23Na GT of a simulation with myocardial TSC of 43 mM (see top row of Figure 5A) was used and Gaussian noise was added. In this case, the assumed myo‐blood ratio for the 23Na prior information rassumed was exactly set to match the simulation. Then, the in Figure S3 described procedure was performed using different σ in the low‐pass filtering step. Finally, the obtained B1 corrected 23Na images were compared to the 23Na GT via NRMSE. Too low σ does not suppress noise sufficiently, while too high σ loses some of the spatial low‐frequency components of the B1 bias field. The optimal σ was found for σ = 10 mm, which was used for the 23Na B1 bias correction.

Figure S3: Overview of the implemented B1 bias field correction for 1H and 23Na MRI. For 1H MRI (A) a homogeneous support region was created by thresholding. The B1 uncorrected 1H image and the supporting region were then low‐pass filtered using a Gaussian filter with standard deviation σ1H=12mm and divided to obtain the estimated 1H B1 bias field. The B1 uncorrected 1H image was divided by the estimated 1H B1 bias field resulting in a B1 corrected 1H image. Using a support region as anatomical prior leads to some errors, however, since 1H images are only used for segmentation minor errors can be neglected. In contrast, for 23Na MRI (B) the algorithm was designed more complex to reduce quantitative errors. Here, the 1H based segmentations of myocardium and blood pool were convoluted with the T 2* dependent corresponding point‐spread functions (PSF) to obtain region‐spread functions (RSF) of both tissues. These RSF were then summed up to the prior information rassumed·RSFmyo+RSFblood using an assumed myo‐blood ratio rassumed. Subsequently, the B1 uncorrected 23Na image and the 23Na prior information were masked and Gaussian filtered with σ23Na=10mm. Division of both yields the estimated 23Na B1 bias field, which was used to correct the B1 uncorrected 23Na image within the heart region, yielding a B1 corrected 23Na image. Partial volume correction—based on the mean signals in myocardium and blood—of the B1 corrected 23Na image allows to quantify the myo‐blood ratio rmeasured after the application of the B1 bias correction, which is then compared to the input rassumed. The entire procedure is repeated iteratively for different rassumed and the one showing the lowest absolute difference of rassumed and rmeasured is selected as the optimal choice.

Figure S4: Comparison of the double angle and actual flip‐angle imaging method for 23Na B1 + mapping in one healthy subject. The double angle method [5] was based on the acquisition of two images with FA = 45° and FA = 90° using a TR = 250 ms to ensure full T 1 relaxation. The AFI method used a FA = 60° and TR1 = 12 ms and TR2 = 48 ms. For both methods the total acquisition time was set to 15 min. The outline of the heart was drawn on an additional 1H image (A). The FA determined using the DAM (B) is higher compared to the AFI method (C). This is likely caused by the T 1 bias of the AFI sequence. In general, the AFI appears smoother and showed no high‐frequency FA variations within the heart. When accelerating the AFI measurement from 15 to 3 min, an additional Gaussian filter (imgaussfilt3 in Matlab) with standard deviation of 14 mm was applied to the FA map. The resulting FA map (D) showed a similar FA distribution as for the 15 min AFI measurement (C).

Figure S5: Influence of deviations of the myocardial T*2,short to the assumed value of 3 ms. The influence was evaluated for the B1 bias correction (I) as well as entire quantification procedure, including B1 bias correction, partial volume correction and relaxation correction. Simulations were performed for a myocardial TSC of 43 mM and myocardial T*2,short times, which were shorter (2.4 ms) or longer (3.6 ms) than or the same as the 3 ms assumed for all corrections. The B1 bias correction was evaluated for three different B1 bias field distributions as shown in Figure 5. B1 bias correction was almost not influenced by the deviating T*2,short times as B1 corrected images showed low and stable aTSC deviations relative to the B1 free GT images and lower deviations than the B1 uncorrected images across all three B1 distributions and all three simulated T*2,short times (I). When considering the entire quantification procedure, stronger aTSC deviations are visible, mainly caused by errors in the performed relaxation correction of the T 2* decay. Over all three simulated T*2,short values, the quantified aTSC showed deviations between −23.4% and +17.1% for the B1 uncorrected images, while this reduced to −7.4% and 12% after B1 correction.

Figure S6: Influence of imperfect segmentations for normal (A) as well as hypertrophied heart (B), which was additionally simulated. In order to evaluate the influence of segmentation errors, we eroded (spherical erosion with one voxel radius) and dilated (spherical dilation with one voxel diameter) the original myocardial masks. This resulted in an increase/decrease of the myocardial volume of −/+25% and −/+14% for the normal and hypertrophied heart, respectively (I). The 23Na B1 bias correction was evaluated for the three different B1 bias field distributions as shown in Figure 5. For both normal and hypertrophied heart, a thinner segmented myocardium results in a slight overcorrection when applying the B1 correction, while a too thick segmented myocardium leads to a slight undercorrection (II). Over all three used mask configurations and B1 bias field distributions, for the normal heart (A) the quantified aTSC showed deviations between −22.9% and +14.5% for the B1 uncorrected images, which were reduced to −4.3% and +10.5% after B1 correction (III). For the hypertrophied heart (B), deviations for the B1 uncorrected images ranged between −15.5% and +11.5% and between −8.4% and +6.8% after B1 correction.

Table S3: Quantitative results of the repeatability study in 10 healthy subjects. Listed are subject‐specific characteristics (gender, age, height, weight, BMI), physiological parameters derived during reconstruction (respiratory and cardiac frequency, cardiac peak width, maximal respiratory displacement in the superior–inferior direction), as well as tissue sodium concentrations (aTSC) for blood (measured from blood samples) and myocardium (measured by 23Na MRI).

Figure S7: Influence of different B1 correction methods on 23Na MRI. The B1 corrections were applied after respiratory and cardiac motion correction. Three different B1 correction methods were compared: (1) B1 + and B1 − field measured in phantom; (2) B1 + field measured in vivo using the AFI sequence and B1 − field measured in phantom; (3) proposed B1 bias field correction. Note that since the B1 bias field correction was applied only within the heart, the B1 bias corrected 23Na image was masked accordingly (D). 23Na signal across the blood pool displayed a homogenous signal distribution for the B1 bias field correction (D), which is not the case for the other two methods (B, C). Compared to the uncorrected image (A), B1 bias field correction increased the 23Na signal, in particular in superior left regions of the heart.

Figure S8: Influence of myocardial masks size on quantitative myocardial aTSC as well as repeatability. All corrections were applied as in Figure 10C, with B1 bias field correction and partial volume correction in particular depending on the segmentations. The original myocardial masks were eroded or dilated using a spherical structuring element with a radius of 1 voxel. The relative change in myocardial mask volume compared to the original mask, averaged across all 10 subjects, was: –30% ± 3% for erosion and +30% ± 3% after dilation. (A) shows an overlay of the modified myocardial (red) and blood pool (blue) masks for subject 1. Reducing the myocardial mask size led to an increase in measured myocardial aTSC values, whereas dilation resulted in a decrease (B). The changes in global mean myocardial aTSC were: +7.6% for the eroded masks and −5.2% for the dilated masks. There were no relevant influences of the masks on the repeatability (C).

Figure S9: Influence of the AFI approximation on flip angles measured in myocardium and blood. The AFI method assumes TR1/2≪T1. Since this is not fulfilled for the applied 23Na AFI measurements (TR1=12ms,TR2=48ms), this leads to deviations of the measured flip angle compared to the true flip angle (A). These deviations depend on the T 1 relaxation time of the tissue and are therefore higher in myocardium than in blood. The found relation in (A) allows to correct this effect for the different compartments assuming specific T 1 times (T1,myo=30ms,T1,blood=49.5ms). For the method using the AFI for the estimation of the B1 + distribution and the B1 − map from the phantom (left side in B), this reduced the myocardial aTSC on average by 2.2 mM and slightly worsened the repeatability from CR = 5.1% to 5.3%. When applying the B1 bias correction (right side in B), the AFI measurement was only used to determine the flip angle for the T 1 relaxation correction. The average aTSC was decreased by −0.9 mM, while the repeatability slightly improved from 4.0% to 3.9%.

Figure S10: Assessment of the variability of the segmentations predicted by the nnUNet. For the evaluation of the DICE score the 1H image of the second measurement was registered to the 1H image of the first measurement. The segmentations of measurement 2 were than transformed accordingly. DICE scores for the myocardial and blood segmentation between both measurements (A) indicate good accordance. The mean DICE score across all subjects and both compartments was 0.89±0.04. Relative volume changes of the myocardial and blood masks (B) ranged between −11.1% and 9.1%.

Figure S11: Validation of respiratory motion correction for different breathing patterns. In this measurement of one healthy subject (male, 28 years, 83 kg), the subject was instructed to change breathing patterns: (A) normal breathing, (B) shallow breathing, (C) deep breathing, (D) random breathing. Each pattern was performed for around 3:45 min. The 1H based self‐gating navigator signal (top) captures all four breathing patterns and exhibits amplitude variations associated with different breathing depths. For the entire measurement the respiratory motion correction algorithm was applied as described in the paper, such that all acquired projections were binned into eight different respiratory bins and corrected into the fully‐exhaled state. Here, bin 1 corresponds to the fully‐inhaled state, while bin 8 represent the fully‐exhaled state. For better assessment of the respiratory motion correction for the different breathing patterns, uncorrected and respiratory motion‐corrected 23Na and 1H images of each phase (A–D) were reconstructed as well as of the entire measurement (E). Note that the shown images do not contain cardiac motion correction and B1 correction was only applied for 1H, not 23Na MRI. Applying respiratory motion correction, 23Na and 1H images of all phases are aligned in the fully‐exhaled state, while motion artifacts are especially visible in the uncorrected images for deep and random breathing. Differences between uncorrected and respiratory corrected images are more pronounced for deep and random breathing compared to normal or shallow breathing as expected.

MRM-96-173-s001.docx (4.2MB, docx)

Acknowledgments

We gratefully acknowledge Manuela Linß for conducting the blood sampling prior to the MRI acquisitions and Tobias Wilferth for performing the pulse simulations. This work was supported by the Deutsche Forschungsgemeinschaft under grant numbers 449552397 and 509149993 (TRR 374). Open Access funding enabled and organized by Projekt DEAL.

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.

References

  • 1. Bottomley P. A., “Sodium MRI in Human Heart: A Review,” NMR in Biomedicine 29, no. 2 (2016): 187–196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Madelin G. and Regatte R. R., “Biomedical Applications of Sodium MRI In Vivo,” Journal of Magnetic Resonance Imaging 38, no. 3 (2013): 511–529. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Ouwerkerk R., Bottomley P. A., Solaiyappan M., et al., “Tissue Sodium Concentration in Myocardial Infarction in Humans: A Quantitative 23Na MR Imaging Study1,” Radiology 248, no. 1 (2008): 88–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Christa M., Weng A. M., Geier B., et al., “Increased Myocardial Sodium Signal Intensity in Conn's Syndrome Detected by 23Na Magnetic Resonance Imaging,” European Heart Journal Cardiovascular Imaging 20, no. 3 (2019): 263–270. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Gast L. V., Platt T., Nagel A. M., and Gerhalter T., “Recent Technical Developments and Clinical Research‐Applications of Sodium (23Na) MRI,” Progress in Nuclear Magnetic Resonance Spectroscopy 138‐139 (2023): 1–51. [DOI] [PubMed] [Google Scholar]
  • 6. Niesporek S. C., Hoffmann S. H., Berger M. C., et al., “Partial Volume Correction for In Vivo 23Na‐MRI Data of the Human Brain,” NeuroImage 112 (2015): 353–363. [DOI] [PubMed] [Google Scholar]
  • 7. Lott J., Platt T., Niesporek S. C., et al., “Corrections of Myocardial Tissue Sodium Concentration Measurements in Human Cardiac (23) Na MRI at 7 Tesla,” Magnetic Resonance in Medicine 82, no. 1 (2019): 159–173. [DOI] [PubMed] [Google Scholar]
  • 8. Ruck L., Egger N., Wilferth T., et al., “Interleaved 23Na/1H MRI of the Human Heart at 7 T Using a Combined 23Na/1H Coil Setup and 1H Parallel Transmission,” Magnetic Resonance in Medicine 94, no. 1 (2025): 231–241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Ibrahim T. S., Lee R., Baertlein B. A., Kangarlu A., and Robitaille P.‐M. L., “Application of Finite Difference Time Domain Method for the Design of Birdcage RF Head Coils Using Multi‐Port Excitations,” Magnetic Resonance Imaging 18, no. 6 (2000): 733–742. [DOI] [PubMed] [Google Scholar]
  • 10. Platt T., Umathum R., Fiedler T. M., et al., “In Vivo Self‐Gated 23Na MRI at 7 T Using an Oval‐Shaped Body Resonator,” Magnetic Resonance in Medicine 80, no. 3 (2018): 1005–1019. [DOI] [PubMed] [Google Scholar]
  • 11. Wilferth T., Müller M., Gast L. V., et al., “Motion‐Corrected 23Na MRI of the Human Brain Using Interleaved 1H 3D Navigator Images,” Magnetic Resonance in Medicine 88 (2022): 309–321. [DOI] [PubMed] [Google Scholar]
  • 12. Lommen J., Konstandin S., Krämer P., and Schad L. R., “Enhancing the Quantification of Tissue Sodium Content by MRI: Time‐Efficient Sodium B1 Mapping at Clinical Field Strengths,” NMR in Biomedicine 29, no. 2 (2016): 129–136. [DOI] [PubMed] [Google Scholar]
  • 13. Wissmann L., Santelli C., Segars W. P., and Kozerke S., “MRXCAT: Realistic Numerical Phantoms for Cardiovascular Magnetic Resonance,” Journal of Cardiovascular Magnetic Resonance 16, no. 1 (2014): 1–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Segars W. P., Sturgeon G., Mendonca S., Grimes J., and Tsui B. M., “4D XCAT Phantom for Multimodality Imaging Research,” Medical Physics 37, no. 9 (2010): 4902–4915. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Wang Y., Riederer S. J., and Ehman R. L., “Respiratory Motion of the Heart: Kinematics and the Implications for the Spatial Resolution in Coronary Imaging,” Magnetic Resonance in Medicine 33, no. 5 (1995): 713–719. [DOI] [PubMed] [Google Scholar]
  • 16. Ruck L., Mennecke A., Wilferth T., et al., “Influence of Image Contrasts and Reconstruction Methods on the Classification of Multiple Sclerosis‐Like Lesions in Simulated Sodium Magnetic Resonance Imaging,” Magnetic Resonance in Medicine 89, no. 3 (2023): 1102–1116. [DOI] [PubMed] [Google Scholar]
  • 17. Nagel A. M., Laun F. B., Weber M. A., Matthies C., Semmler W., and Schad L. R., “Sodium MRI Using a Density‐Adapted 3D Radial Acquisition Technique,” Magnetic Resonance in Medicine 62, no. 6 (2009): 1565–1573. [DOI] [PubMed] [Google Scholar]
  • 18. Egger N., Nagelstraßer S., Wildenberg S., et al., “Accelerated B 1+ B _1^+ Mapping and Robust Parallel Transmit Pulse Design for Heart and Prostate Imaging at 7 T,” Magnetic Resonance in Medicine 92, no. 5 (2024): 1933–1951. [DOI] [PubMed] [Google Scholar]
  • 19. Chan R. W., Ramsay E. A., Cunningham C. H., and Plewes D. B., “Temporal Stability of Adaptive 3D Radial MRI Using Multidimensional Golden Means,” Magnetic Resonance in Medicine 61, no. 2 (2009): 354–363. [DOI] [PubMed] [Google Scholar]
  • 20. Fessler J. A. and Sutton B. P., “Nonuniform Fast Fourier Transforms Using Min‐Max Interpolation,” IEEE Transactions on Signal Processing 51, no. 2 (2003): 560–574. [Google Scholar]
  • 21. Doran S. J., Charles‐Edwards L., Reinsberg S. A., and Leach M. O., “A Complete Distortion Correction for MR Images: I. Gradient Warp Correction,” Physics in Medicine and Biology 50, no. 7 (2005): 1343–1361. [DOI] [PubMed] [Google Scholar]
  • 22. Grimm R., Block K. T., Kiefer B., and Hornegger J., “Bias Correction for Respiration Detection in Radial 3D Gradient‐Echo Imaging,” in Proceedings of the 19th Annual Meeting of ISMRM, Montreal, Canada (2011), 2677.
  • 23. Klein S., Staring M., Murphy K., Viergever M. A., and Pluim J. P., “Elastix: A Toolbox for Intensity‐Based Medical Image Registration,” IEEE Transactions on Medical Imaging 29, no. 1 (2009): 196–205. [DOI] [PubMed] [Google Scholar]
  • 24. Ruck L., Egger N., Kopp C., et al., “Respiratory Motion Corrected Interleaved 23Na/1H pTx MRI of the Human Heart at 7 Tesla,” in Proceedings of the 33rd Annual Meeting of ISMRM, Honolulu, USA (2025).
  • 25. Orfanidis S. J., Introduction to Signal Processing (Prentice Hall, 1995). [Google Scholar]
  • 26. Butterworth S., “On the Theory of Filter Amplifiers,” Wireless Engineering 7, no. 6 (1930): 536–541. [Google Scholar]
  • 27. Wang Z., Bovik A. C., Sheikh H. R., and Simoncelli E. P., “Image Quality Assessment: From Error Visibility to Structural Similarity,” IEEE Transactions on Image Processing 13, no. 4 (2004): 600–612. [DOI] [PubMed] [Google Scholar]
  • 28. MacQueen J., “Multivariate Observations,” in Proceedings of the 5th Berkeley Symposium on Mathematical Statistics and Probability (1967), 281–297.
  • 29. Hartigan J. A. and Wong M. A., “Algorithm AS 136: A k‐Means Clustering Algorithm,” Applied Statistics 28, no. 1 (1979): 100–108. [Google Scholar]
  • 30. Isensee F., Jaeger P. F., Kohl S. A., Petersen J., and Maier‐Hein K. H., “nnU‐Net: A Self‐Configuring Method for Deep Learning‐Based Biomedical Image Segmentation,” Nature Methods 18, no. 2 (2021): 203–211. [DOI] [PubMed] [Google Scholar]
  • 31. Stobbe R. W. and Beaulieu C., “Residual Quadrupole Interaction in Brain and Its Effect on Quantitative Sodium Imaging,” NMR in Biomedicine 29, no. 2 (2016): 119–128. [DOI] [PubMed] [Google Scholar]
  • 32. Lachner S., Ruck L., Niesporek S. C., et al., “Comparison of Optimized Intensity Correction Methods for (23)Na MRI of the Human Brain Using a 32‐Channel Phased Array Coil at 7 Tesla,” Zeitschrift für Medizinische Physik 30, no. 2 (2020): 104–115. [DOI] [PubMed] [Google Scholar]
  • 33. Murakami J. W., Hayes C. E., and Weinberger E., “Intensity Correction of Phased‐Array Surface Coil Images,” Magnetic Resonance in Medicine 35, no. 4 (1996): 585–590. [DOI] [PubMed] [Google Scholar]
  • 34. Yarnykh V. L., “Actual Flip‐Angle Imaging in the Pulsed Steady State: A Method for Rapid Three‐Dimensional Mapping of the Transmitted Radiofrequency Field,” Magnetic Resonance in Medicine 57, no. 1 (2007): 192–200. [DOI] [PubMed] [Google Scholar]
  • 35. Maggioni M. B., Krämer M., and Reichenbach J. R., “Optimized Gradient Spoiling of UTE VFA‐AFI Sequences for Robust T1 Estimation With B1‐Field Correction,” Magnetic Resonance Imaging 82 (2021): 1–8. [DOI] [PubMed] [Google Scholar]
  • 36. Kordzadeh A., Duchscherer J., Beaulieu C., and Stobbe R., “Radiofrequency Excitation–Related 23Na MRI Signal Loss in Skeletal Muscle, Cartilage, and Skin,” Magnetic Resonance in Medicine 83, no. 6 (2020): 1992–2001. [DOI] [PubMed] [Google Scholar]
  • 37. Dietrich O., Raya J. G., Reeder S. B., Reiser M. F., and Schoenberg S. O., “Measurement of Signal‐To‐Noise Ratios in MR Images: Influence of Multichannel Coils, Parallel Imaging, and Reconstruction Filters,” Journal of Magnetic Resonance Imaging 26, no. 2 (2007): 375–385. [DOI] [PubMed] [Google Scholar]
  • 38. Barnhart H. X. and Barboriak D. P., “Applications of the Repeatability of Quantitative Imaging Biomarkers: A Review of Statistical Analysis of Repeat Data Sets,” Translational Oncology 2, no. 4 (2009): 231–235. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Bland J. M. and Altman D., “Statistical Methods for Assessing Agreement Between Two Methods of Clinical Measurement,” Lancet (London, England) 327, no. 8476 (1986): 307–310. [PubMed] [Google Scholar]
  • 40. Pang J., Sharif B., Fan Z., et al., “ECG and Navigator‐Free Four‐Dimensional Whole‐Heart Coronary MRA for Simultaneous Visualization of Cardiac Anatomy and Function,” Magnetic Resonance in Medicine 72, no. 5 (2014): 1208–1217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Kawaji K., Spincemaille P., Nguyen T. D., et al., “Direct Coronary Motion Extraction From a 2D Fat Image Navigator for Prospectively Gated Coronary MR Angiography,” Magnetic Resonance in Medicine 71, no. 2 (2014): 599–607. [DOI] [PubMed] [Google Scholar]
  • 42. McLeish K., Hill D. L., Atkinson D., Blackall J. M., and Razavi R., “A Study of the Motion and Deformation of the Heart due to Respiration,” IEEE Transactions on Medical Imaging 21, no. 9 (2002): 1142–1150. [DOI] [PubMed] [Google Scholar]
  • 43. Scheipers A. K., Grimm J. A., Losch J., et al., “Quantitative Abdominal Sodium MRI Combined With 32‐Channel Proton pTx MRI at 7 Tesla in a Large Field‐Of‐View,” Magnetic Resonance in Medicine 94 (2025): 1930–1945. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44. Ouwerkerk R., Weiss R. G., and Bottomley P. A., “Measuring Human Cardiac Tissue Sodium Concentrations Using Surface Coils, Adiabatic Excitation, and Twisted Projection Imaging With Minimal T2 Losses,” Journal of Magnetic Resonance Imaging 21, no. 5 (2005): 546–555. [DOI] [PubMed] [Google Scholar]
  • 45. Feng L., Axel L., Chandarana H., Block K. T., Sodickson D. K., and Otazo R., “XD‐GRASP: Golden‐Angle Radial MRI With Reconstruction of Extra Motion‐State Dimensions Using Compressed Sensing,” Magnetic Resonance in Medicine 75, no. 2 (2016): 775–788. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Lustig M., Donoho D., and Pauly J. M., “Sparse MRI: The Application of Compressed Sensing for Rapid MR Imaging,” Magnetic Resonance in Medicine 58, no. 6 (2007): 1182–1195. [DOI] [PubMed] [Google Scholar]
  • 47. Stobbe R. and Beaulieu C., “In Vivo Sodium Magnetic Resonance Imaging of the Human Brain Using Soft Inversion Recovery Fluid Attenuation,” Magnetic Resonance in Medicine 54, no. 5 (2005): 1305–1310. [DOI] [PubMed] [Google Scholar]
  • 48. Wilferth T., Mennecke A., Huhn K., et al., “Influence of Residual Quadrupolar Interaction on Quantitative Sodium Brain Magnetic Resonance Imaging of Patients With Multiple Sclerosis,” Investigative Radiology 58, no. 10 (2023): 730–739. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1: Tissue parameters for 23Na simulation. Assumed values are marked by (*). The myocardial TSC of 43 mM was derived from 41 μmol/g [1] by applying a specific gravity of 1.05g/mL for myocardium [44].

Table S2: Tissue parameters for 1H simulation. Assumed values are marked by (*).

Figure S1: Comparison of cardiac self‐gating signal (SG), derived from the interleaved 1H data, and simultaneously acquired electrocardiogram (ECG) for three subjects. Due to the filtering in the post‐processing of the SG signal, the triggers of SG tend to occur in mid‐diastole, while scanner‐predicted ECG triggers (M1, red circles) mark the end of the diastolic phase. Additionally, the minima of the ECG signal were used as triggers (M2, blue circles). For subject 6 (a), the histogram of the temporal difference ΔT between consecutive triggers showed good agreement between SG and ECG independent of the selected triggers. However, for subjects 7 (b) and 8 (c) scanner‐based ECG triggering (M1) resulted in trigger errors. In contrast, minima based ECG trigger (M2) and SG showed consistent triggering for all three subjects.

Figure S2: Influence of the standard deviation σ of the Gaussian low‐pass filter in the 23Na B1 bias correction (see Figure S3). For this optimization, the phantom B1 bias field and the simulated 23Na GT of a simulation with myocardial TSC of 43 mM (see top row of Figure 5A) was used and Gaussian noise was added. In this case, the assumed myo‐blood ratio for the 23Na prior information rassumed was exactly set to match the simulation. Then, the in Figure S3 described procedure was performed using different σ in the low‐pass filtering step. Finally, the obtained B1 corrected 23Na images were compared to the 23Na GT via NRMSE. Too low σ does not suppress noise sufficiently, while too high σ loses some of the spatial low‐frequency components of the B1 bias field. The optimal σ was found for σ = 10 mm, which was used for the 23Na B1 bias correction.

Figure S3: Overview of the implemented B1 bias field correction for 1H and 23Na MRI. For 1H MRI (A) a homogeneous support region was created by thresholding. The B1 uncorrected 1H image and the supporting region were then low‐pass filtered using a Gaussian filter with standard deviation σ1H=12mm and divided to obtain the estimated 1H B1 bias field. The B1 uncorrected 1H image was divided by the estimated 1H B1 bias field resulting in a B1 corrected 1H image. Using a support region as anatomical prior leads to some errors, however, since 1H images are only used for segmentation minor errors can be neglected. In contrast, for 23Na MRI (B) the algorithm was designed more complex to reduce quantitative errors. Here, the 1H based segmentations of myocardium and blood pool were convoluted with the T 2* dependent corresponding point‐spread functions (PSF) to obtain region‐spread functions (RSF) of both tissues. These RSF were then summed up to the prior information rassumed·RSFmyo+RSFblood using an assumed myo‐blood ratio rassumed. Subsequently, the B1 uncorrected 23Na image and the 23Na prior information were masked and Gaussian filtered with σ23Na=10mm. Division of both yields the estimated 23Na B1 bias field, which was used to correct the B1 uncorrected 23Na image within the heart region, yielding a B1 corrected 23Na image. Partial volume correction—based on the mean signals in myocardium and blood—of the B1 corrected 23Na image allows to quantify the myo‐blood ratio rmeasured after the application of the B1 bias correction, which is then compared to the input rassumed. The entire procedure is repeated iteratively for different rassumed and the one showing the lowest absolute difference of rassumed and rmeasured is selected as the optimal choice.

Figure S4: Comparison of the double angle and actual flip‐angle imaging method for 23Na B1 + mapping in one healthy subject. The double angle method [5] was based on the acquisition of two images with FA = 45° and FA = 90° using a TR = 250 ms to ensure full T 1 relaxation. The AFI method used a FA = 60° and TR1 = 12 ms and TR2 = 48 ms. For both methods the total acquisition time was set to 15 min. The outline of the heart was drawn on an additional 1H image (A). The FA determined using the DAM (B) is higher compared to the AFI method (C). This is likely caused by the T 1 bias of the AFI sequence. In general, the AFI appears smoother and showed no high‐frequency FA variations within the heart. When accelerating the AFI measurement from 15 to 3 min, an additional Gaussian filter (imgaussfilt3 in Matlab) with standard deviation of 14 mm was applied to the FA map. The resulting FA map (D) showed a similar FA distribution as for the 15 min AFI measurement (C).

Figure S5: Influence of deviations of the myocardial T*2,short to the assumed value of 3 ms. The influence was evaluated for the B1 bias correction (I) as well as entire quantification procedure, including B1 bias correction, partial volume correction and relaxation correction. Simulations were performed for a myocardial TSC of 43 mM and myocardial T*2,short times, which were shorter (2.4 ms) or longer (3.6 ms) than or the same as the 3 ms assumed for all corrections. The B1 bias correction was evaluated for three different B1 bias field distributions as shown in Figure 5. B1 bias correction was almost not influenced by the deviating T*2,short times as B1 corrected images showed low and stable aTSC deviations relative to the B1 free GT images and lower deviations than the B1 uncorrected images across all three B1 distributions and all three simulated T*2,short times (I). When considering the entire quantification procedure, stronger aTSC deviations are visible, mainly caused by errors in the performed relaxation correction of the T 2* decay. Over all three simulated T*2,short values, the quantified aTSC showed deviations between −23.4% and +17.1% for the B1 uncorrected images, while this reduced to −7.4% and 12% after B1 correction.

Figure S6: Influence of imperfect segmentations for normal (A) as well as hypertrophied heart (B), which was additionally simulated. In order to evaluate the influence of segmentation errors, we eroded (spherical erosion with one voxel radius) and dilated (spherical dilation with one voxel diameter) the original myocardial masks. This resulted in an increase/decrease of the myocardial volume of −/+25% and −/+14% for the normal and hypertrophied heart, respectively (I). The 23Na B1 bias correction was evaluated for the three different B1 bias field distributions as shown in Figure 5. For both normal and hypertrophied heart, a thinner segmented myocardium results in a slight overcorrection when applying the B1 correction, while a too thick segmented myocardium leads to a slight undercorrection (II). Over all three used mask configurations and B1 bias field distributions, for the normal heart (A) the quantified aTSC showed deviations between −22.9% and +14.5% for the B1 uncorrected images, which were reduced to −4.3% and +10.5% after B1 correction (III). For the hypertrophied heart (B), deviations for the B1 uncorrected images ranged between −15.5% and +11.5% and between −8.4% and +6.8% after B1 correction.

Table S3: Quantitative results of the repeatability study in 10 healthy subjects. Listed are subject‐specific characteristics (gender, age, height, weight, BMI), physiological parameters derived during reconstruction (respiratory and cardiac frequency, cardiac peak width, maximal respiratory displacement in the superior–inferior direction), as well as tissue sodium concentrations (aTSC) for blood (measured from blood samples) and myocardium (measured by 23Na MRI).

Figure S7: Influence of different B1 correction methods on 23Na MRI. The B1 corrections were applied after respiratory and cardiac motion correction. Three different B1 correction methods were compared: (1) B1 + and B1 − field measured in phantom; (2) B1 + field measured in vivo using the AFI sequence and B1 − field measured in phantom; (3) proposed B1 bias field correction. Note that since the B1 bias field correction was applied only within the heart, the B1 bias corrected 23Na image was masked accordingly (D). 23Na signal across the blood pool displayed a homogenous signal distribution for the B1 bias field correction (D), which is not the case for the other two methods (B, C). Compared to the uncorrected image (A), B1 bias field correction increased the 23Na signal, in particular in superior left regions of the heart.

Figure S8: Influence of myocardial masks size on quantitative myocardial aTSC as well as repeatability. All corrections were applied as in Figure 10C, with B1 bias field correction and partial volume correction in particular depending on the segmentations. The original myocardial masks were eroded or dilated using a spherical structuring element with a radius of 1 voxel. The relative change in myocardial mask volume compared to the original mask, averaged across all 10 subjects, was: –30% ± 3% for erosion and +30% ± 3% after dilation. (A) shows an overlay of the modified myocardial (red) and blood pool (blue) masks for subject 1. Reducing the myocardial mask size led to an increase in measured myocardial aTSC values, whereas dilation resulted in a decrease (B). The changes in global mean myocardial aTSC were: +7.6% for the eroded masks and −5.2% for the dilated masks. There were no relevant influences of the masks on the repeatability (C).

Figure S9: Influence of the AFI approximation on flip angles measured in myocardium and blood. The AFI method assumes TR1/2≪T1. Since this is not fulfilled for the applied 23Na AFI measurements (TR1=12ms,TR2=48ms), this leads to deviations of the measured flip angle compared to the true flip angle (A). These deviations depend on the T 1 relaxation time of the tissue and are therefore higher in myocardium than in blood. The found relation in (A) allows to correct this effect for the different compartments assuming specific T 1 times (T1,myo=30ms,T1,blood=49.5ms). For the method using the AFI for the estimation of the B1 + distribution and the B1 − map from the phantom (left side in B), this reduced the myocardial aTSC on average by 2.2 mM and slightly worsened the repeatability from CR = 5.1% to 5.3%. When applying the B1 bias correction (right side in B), the AFI measurement was only used to determine the flip angle for the T 1 relaxation correction. The average aTSC was decreased by −0.9 mM, while the repeatability slightly improved from 4.0% to 3.9%.

Figure S10: Assessment of the variability of the segmentations predicted by the nnUNet. For the evaluation of the DICE score the 1H image of the second measurement was registered to the 1H image of the first measurement. The segmentations of measurement 2 were than transformed accordingly. DICE scores for the myocardial and blood segmentation between both measurements (A) indicate good accordance. The mean DICE score across all subjects and both compartments was 0.89±0.04. Relative volume changes of the myocardial and blood masks (B) ranged between −11.1% and 9.1%.

Figure S11: Validation of respiratory motion correction for different breathing patterns. In this measurement of one healthy subject (male, 28 years, 83 kg), the subject was instructed to change breathing patterns: (A) normal breathing, (B) shallow breathing, (C) deep breathing, (D) random breathing. Each pattern was performed for around 3:45 min. The 1H based self‐gating navigator signal (top) captures all four breathing patterns and exhibits amplitude variations associated with different breathing depths. For the entire measurement the respiratory motion correction algorithm was applied as described in the paper, such that all acquired projections were binned into eight different respiratory bins and corrected into the fully‐exhaled state. Here, bin 1 corresponds to the fully‐inhaled state, while bin 8 represent the fully‐exhaled state. For better assessment of the respiratory motion correction for the different breathing patterns, uncorrected and respiratory motion‐corrected 23Na and 1H images of each phase (A–D) were reconstructed as well as of the entire measurement (E). Note that the shown images do not contain cardiac motion correction and B1 correction was only applied for 1H, not 23Na MRI. Applying respiratory motion correction, 23Na and 1H images of all phases are aligned in the fully‐exhaled state, while motion artifacts are especially visible in the uncorrected images for deep and random breathing. Differences between uncorrected and respiratory corrected images are more pronounced for deep and random breathing compared to normal or shallow breathing as expected.

MRM-96-173-s001.docx (4.2MB, docx)

Data Availability Statement

The data that support the findings of this study are available on request from the corresponding author. The data are not publicly available due to privacy or ethical restrictions.