Abstract
Post-operative infections remain a challenge in implant surgery, leading to prolonged treatment, increased costs, and implant failure. Localized antibiotic-delivery coatings are a promising strategy to reduce infection risk while maintaining biocompatibility. Here, we evaluate cold atmospheric pressure plasma (CAP) spraying as a solvent-free method to deposit polyvinyl alcohol (PVA) layers containing amikacin on stainless steel and to identify plasma parameters that control release and antibacterial activity. A 3 × 3 factorial design varied nozzle distance (15, 20, 25 mm) and speed (10, 15, 20 cm/s). Surface morphology was assessed by optical microscopy, amikacin release quantified by HPLC-HRMS, and antibacterial activity tested against Staphylococcus aureus and Escherichia coli using Kirby–Bauer disc diffusion. Two-way ANOVA with Tukey post hoc tests and nonparametric validation were applied. Cold plasma spraying speed significantly affected drug release, whereas distance and the interaction term were not significant. Lower spraying speeds produced thicker, more porous coatings with greater cumulative release and larger inhibition zones. Drug release profiles were best described by Weibull and first-order models, showing an initial burst followed by sustained release. These findings indicate that CAP spraying enables solvent-free fabrication of antibiotic-loaded PVA coatings with tunable release, and optimizing spraying speed improves coating mass, drug delivery, and antibacterial performance.
Keywords: cold atmospheric plasma, polyvinyl alcohol, amikacin release, medical implants, drug release kinetics, sustained release
1. Introduction
Post-operative infections associated with metal implants remain a major challenge in clinical practice, often leading to prolonged hospitalization, increased treatment costs, and, in severe cases, implant failure or removal [1]. To mitigate these risks, localized antibiotic delivery directly from implant surfaces has attracted significant attention as a strategy to provide high local drug concentrations while minimizing systemic side effects [2].
Biodegradable polymers, particularly polyvinyl alcohol (PVA) (Figure 1, right), have gained interest due to their excellent biocompatibility, hydrophilicity, and ability to sustain and control drug release [3,4]. However, traditional coating techniques such as dip-coating or solvent casting present significant limitations [5]. Dip-coating, for instance, frequently results in non-uniform film thickness and poor adhesion, increasing the risk of delamination. Furthermore, the elevated temperatures often required during post-deposition processing can introduce surface defects and compromise coating quality [6].
Figure 1.
Chemical structures of amikacin (left) and polyvinyl alcohol (PVA, right).
Cold atmospheric pressure plasma (CAP) deposition offers a promising solvent-free alternative that allows precise control over surface properties, thin-film formation, and coating adhesion, while providing a rapid, clean, and environmentally friendly process [7]. CAP has been widely used to tailor surface energy, improve polymer-substrate adhesion, and control thin-film morphology [8]. Nevertheless, despite extensive use of cold atmospheric plasma for surface modification, its application in plasma-assisted deposition of biodegradable, antibiotic-loaded polymer coatings with controlled drug release has been comparatively underexplored. Moreover, the influence of CAP parameters, particularly nozzle distance and plasma flow speed, on coating structure, drug-release kinetics, and antibacterial efficacy is not yet fully understood, especially in the context of plasma-assisted deposition of biodegradable, drug-loaded polymer coatings.
The present study aims to optimize CAP deposition of PVA–amikacin (Figure 1, left) coatings on stainless steel substrates to achieve sustained antibiotic release and enhanced antibacterial activity. A 3 × 3 factorial design was employed to systematically investigate the effects of plasma nozzle distance and spraying speed on coating surface morphology, drug release kinetics, and antimicrobial performance. We hypothesize that specific CAP parameter combinations will produce coatings with desirable surface roughness and porosity, leading to more effective and prolonged antibiotic release. These insights could inform the development of next-generation biodegradable antimicrobial coatings for medical implants.
2. Materials and Methods
2.1. Materials
Substrate: Stainless steel foil (AISI 304, Hasberg Schneider GmbH, Rudolf-Stratz-Straße 183233 Bernau am Chiemsee, Germany) with thicknesses of 100 µm or 200 µm (depending on availability).
Poly(vinyl alcohol) 99% (Parteck® SRP 80, EMPROVE® ESSENTIAL, Cat. No. 141439) was obtained from Merck KGaA (Darmstadt, Germany).
Amikacin (≥99% purity) was obtained from Sintez Pharmaceutical Company (Kurgan, Russia).
2.2. Methods
2.2.1. Sample Preparation
The foil was cut into 15 × 15 cm pieces and pre-cleaned with ethanol. Before plasma spraying, the substrate was cleaned with pure argon plasma at a distance of 25 mm and a nozzle speed of 20 cm/s. The drug-polymer mixture was deposited onto the foil using cold atmospheric plasma (CAP).
2.2.2. Plasma Coating Procedure
Plasma spraying was conducted using a Cold Atmospheric Plasma (CAP) system (Plasma Coat PCU3D) with a plasma current of 100 A and voltage of 26 V. Argon was used as the carrier gas at a flow rate of 20 L/min, with the particle feed rate set to maximum. Before coating, stainless-steel substrates were pre-treated with pure argon plasma at a fixed distance of 25 mm and nozzle speed of 20 cm/s.
The coating was applied under varying nozzle speeds (10, 15, and 20 cm/s) and spraying distances (15, 20, and 25 mm), creating nine unique combinations (see Table 1). Each condition was replicated with a specific substrate thickness (100 μm or 200 μm).
Table 1.
Plasma spraying parameters and corresponding coating mass on stainless steel foil substrates. Lower nozzle speed and shorter spraying distance increased coating mass deposition. Substrate Mass = 23.41 ± 8.48 g; Substrate Thickness = 144 ± 52.7 µm (mean ± SD, n = 9).
| Number of Sample | Spraying Distance (mm) |
Nozzle Speed (cm/s) |
Substrate Mass (g) |
Substrate Thickness(μm) | Coating Mass (mg) |
|---|---|---|---|---|---|
| 1 | 15 | 10 | 34.48 | 200 | 490 |
| 2 | 20 | 10 | 34.77 | 200 | 490 |
| 3 | 25 | 10 | 18.03 | 100 | 370 |
| 4 | 15 | 15 | 34.92 | 200 | 330 |
| 5 | 20 | 15 | 17.64 | 100 | 300 |
| 6 | 25 | 15 | 17.80 | 100 | 280 |
| 7 | 15 | 20 | 17.79 | 100 | 240 |
| 8 | 20 | 20 | 17.59 | 100 | 150 |
| 9 | 25 | 20 | 17.68 | 100 | 200 |
2.2.3. Sample Storage
All coated and plasma-treated samples were stored in zipper storage bags at ambient temperature to prevent microbial contamination and moisture exposure prior to characterization and testing.
2.3. Characterization
2.3.1. Surface Roughness
Surface morphology and topography were analyzed using an optical microscope (DM 6000M, Leica Microsystems, Wetzlar, Germany) operated in dark-field mode. To enhance depth of focus and visualize surface features with varying height, a focus stacking technique was employed. For each sample, a series of 20–25 images was captured at different focal planes depending on surface roughness (Figure 2). These images were processed using Helicon Focus 5.3 Pro software to generate a single high-clarity 2D image.
Figure 2.
Optical microscopy images (×5) of PVA–amikacin coatings deposited on stainless steel by cold atmospheric pressure plasma spraying at different nozzle distances and spraying speeds. (A) Sample 1: 15 mm, 10 cm/s. (B) Sample 2: 20 mm, 10 cm/s. (C) Sample 3: 25 mm, 10 cm/s. (D) Sample 4: 15 mm, 15 cm/s. (E) Sample 5: 20 mm, 15 cm/s. (F) Sample 6: 25 mm, 15 cm/s. (G) Sample 7: 15 mm, 20 cm/s. (H) Sample 8: 20 mm, 20 cm/s. (I) Sample 9: 25 mm, 20 cm/s. Lower nozzle distances and speeds produced rougher, more porous coatings, whereas higher speeds and longer distances yielded smoother, thinner films.
2.3.2. Antimicrobial Activity
Antibiotic efficiency was evaluated using the Kirby–Bauer disc diffusion method. Samples were placed on agar plates inoculated with either Gram-positive (ATCC 6538) or Gram-negative (Escherichia coli ATCC 8739) bacteria (Figure 3 and Figure 4). Plates were incubated at 37 °C for 24–48 h, after which zones of inhibition were measured in millimeters [9].
Figure 3.
Representative agar diffusion assay against Staphylococcus aureus ATCC 6538 for PVA–amikacin coatings prepared under different cold atmospheric pressure plasma (CAP) conditions. Plates № 1–№ 9 correspond to Samples 1–9 in Table 1: (1) 15 mm, 10 cm/s; (2) 20 mm, 10 cm/s; (3) 25 mm, 10 cm/s; (4) 15 mm, 15 cm/s; (5) 20 mm, 15 cm/s; (6) 25 mm, 15 cm/s; (7) 15 mm, 20 cm/s; (8) 20 mm, 20 cm/s; (9) 25 mm, 20 cm/s. Each Petri dish shows three replicate coated squares per condition.
Figure 4.
Representative agar diffusion assay against Escherichia coli ATCC 8739 for PVA–amikacin coatings prepared under different CAP conditions. Film numbers № 1–№ 9 correspond to the same coating preparations as in Figure 3 and Table 1: (1) 15 mm, 10 cm/s; (2) 20 mm, 10 cm/s; (3) 25 mm, 10 cm/s; (4) 15 mm, 15 cm/s; (5) 20 mm, 15 cm/s; (6) 25 mm, 15 cm/s; (7) 15 mm, 20 cm/s; (8) 20 mm, 20 cm/s; (9) 25 mm, 20 cm/s.
2.3.3. Drug Release Profile
In vitro release profiles of amikacin were determined by immersing coated samples in the HPLC mobile phase. Antibiotic concentrations in the samples were quantified using HPLC-HRMS as described in Section 2.3.4, and cumulative release was plotted as drug release percentage versus time.
The release percentage (%R) was calculated using the equation:
where is the mass of amikacin released at time t, and is the total amount of amikacin incorporated in the coating (determined from complete extraction at the end of the experiment).
2.3.4. HPLC-HRMS Analysis of Amikacin
Quantitative analysis of amikacin in the release medium was performed using a Vanquish UHPLC system coupled to a Q Exactive HF Orbitrap high-resolution mass spectrometer (Thermo Fisher Scientific, Hanna-Kunath-Str. 11, 28199 Bremen, Germany). Separation was achieved using a Zorbax SB-C8 column (2.1 × 100 mm, 1.8 µm; Agilent, USA). The mobile phases consisted of (A) 0.1% formic acid in Milli-Q water and (B) HPLC-grade acetonitrile.
The gradient program was as follows:
| Time | Mobile Phase Composition (% B) |
| 0–5.0 min | 100% B |
| 5.0–7.5 min | linear gradient to 0% B |
| 7.5–9.0 min | held at 0% B |
| 9.0–9.001 min | ramp to 100% B |
| 9.001–11.0 min | re-equilibration at 100% B |
The flow rate was set at 0.45 mL/min with a 2.0 μL injection volume. The column was maintained at 25 °C and the autosampler at 12 °C. Mass spectrometry was performed in positive electrospray ionization (ESI) mode with the following parameters: sheath gas 50, sweep gas 1, auxiliary gas 30 (arbitrary units); spray voltage 4.0 kV; capillary temperature 280 °C; in-source CID energy 80 eV.
Data acquisition was done in two modes. First, full scan mode was used over the mass range of 50–750 m/z with an AGC target of 3 × 106 and a resolution of 15,000. Second, Parallel Reaction Monitoring was conducted for amikacin at m/z 586.2930, with a normalized collision energy of 30, a resolution of 30,000, an AGC target of 5 × 105, and an isolation window of 4 m/z. Quantification was based on a five-point calibration curve (0.1–10 µg/mL), with R2 > 0.999.
2.4. Data Analysis
The dataset comprised 68 observations, each representing drug concentration at defined speed (10, 15, or 20 cm/s) and distance (15, 20, or 25 mm) combinations. Each speed group included 22–24 samples distributed across distance levels.
To describe the release kinetics, experimental data were fitted to multiple mathematical models commonly applied to drug release studies [10,11]: zero-order , first-order , Higuchi , Hixson–Crowell , Korsmeyer–Peppas , Weibull , linear, logarithmic, polynomial (second-order), and exponential rise-to-maximum models.
For each plate, all models were tested and compared using goodness-of-fit criteria (coefficient of determination , residual sum of squares [RSS], and Akaike Information Criterion [AIC]). The model with the highest R2 and lowest AIC was selected as the best-fit description of the release profile [12,13] (see Appendix B).
Statistical analyses were performed in Google Colaboratory using Python 3.12.13. Data processing and organization were conducted using pandas (v2.2.2). Two-way ANOVA models were constructed using statsmodels (v0.14.6) to evaluate the effects of nozzle speed, spraying distance, and their interaction on drug release. Post hoc pairwise comparisons between speed groups were performed using Tukey’s Honestly Significant Difference (HSD) test where appropriate. Because assumption checks indicated violations of normality and homogeneity of variance, nonparametric validation was performed using the Kruskal–Wallis test followed by Mann–Whitney U tests with Bonferroni correction. Statistical analyses were conducted using SciPy (v1.16.3) and statsmodels (v0.14.6). Data visualization was performed using matplotlib (v3.10.0) and seaborn (v0.13.2). Statistical significance was defined as p < 0.05.
2.5. Data and Material Availability
All data, materials, and protocols used in this study are available from the corresponding author upon request. There are no restrictions on data or material availability.
3. Results
Stainless steel foil (AISI 304) was used as the substrate material. Each foil was pre-coated with a PVA release layer and cut into uniform 2 cm × 2 cm squares to standardize surface area across all samples.
Polymers and Antibiotics: PVA (Sigma-Aldrich Chemie GmbH, Eschenstr. 5, 82024 Taufkirchen, Germany, Cat. No. 141439) was employed as a sacrificial release layer in film preparation. Amikacin was incorporated into the PVA matrix at a 1:2 ratio to provide antimicrobial function.
Schematic overview of the experimental workflow is shown in Figure 5.
Figure 5.
Schematic overview of the experimental workflow.
Resulted stainless steel foil with applied PVA+amikacin is shown in Figure 6.
Figure 6.

Stainless steel foil and applied PVA + amikacin.
3.1. Coating Mass and Plasma Parameters
Plasma spraying produced coatings with masses ranging from 150 to 490 mg, depending on nozzle speed and spraying distance (Table 1). Lower spraying speeds (10 cm/s) at shorter distances (15–20 mm) resulted in the highest coating mass (490 mg), while higher speeds (20 cm/s) and longer distances (25 mm) yielded the lowest mass (150–200 mg). Substrate thickness (100 µm vs. 200 µm) had no apparent influence on mass deposition.
Coating mass values represent single measurements per condition; however, reproducibility was assessed indirectly through the consistency of drug release profiles and statistical analysis across multiple observations (n = 68).
Although a general decrease in coating mass with increasing distance is observed, minor deviations (e.g., between Samples 8 and 9) indicate that distance has a weaker and non-monotonic influence compared to nozzle speed, consistent with the non-significant ANOVA result.
3.2. Surface Morphology and Roughness
Optical microscopy revealed distinct morphological differences among coatings (Figure 2). At short nozzle distances (15 mm) and low speeds (10 cm/s), surfaces appeared rougher and more porous, while longer distances (25 mm) produced thinner, smoother films. Depth profiles confirmed that rougher surfaces corresponded to increased porosity and higher surface area (Figure 7).
Figure 7.
3D surface renderings from optical microscopy (×5) of the same PVA–amikacin coatings as in Figure 2, illustrating the effect of plasma parameters on surface roughness and porosity. (A–I) Coating conditions correspond to those in Figure 2.
3.3. Drug Release Profiles
Drug release from the PVA–amikacin coatings was well described by the Weibull model for all nine plates (Table 2). The fits were excellent, with coefficients of determination in the range values ranging from 0.902 to 0.996 and AIC values between −29.8 and −54.6. Plates 1 and 2 showed the best agreement with the model (both = 0.996; AIC = −54.0 and −54.6, RSS = 0.00565 and 0.0053, respectively). Plates 3, 5, 7, and 8 also presented very good fits ( = 0.982–0.993, AIC = −35.9 to −45.8), while Plates 4, 6, and 9 still showed acceptable fits despite slightly lower values and higher RSS (0.0288–0.1167).
Table 2.
Comparison of drug release kinetic models for PVA–amikacin films.
| Plate | Best-Fit Model | AIC | RSS | Equation (Fractional Release) | |
|---|---|---|---|---|---|
| 1 | Weibull | 0.996 | −54.0 | 0.00565 | |
| 2 | Weibull | 0.996 | −54.6 | 0.0053 | |
| 3 | Weibull | 0.985 | −44.8 | 0.0018 | |
| 4 | Weibull | 0.902 | −29.8 | 0.1167 | |
| 5 | Weibull | 0.985 | −45.8 | 0.0158 | |
| 6 | Weibull | 0.978 | −34.4 | 0.0288 | |
| 7 | Weibull | 0.982 | −35.9 | 0.0233 | |
| 8 | Weibull | 0.993 | −41.8 | 0.0101 | |
| 9 | Weibull | 0.940 | −32.9 | 0.0359 |
For all coatings, the characteristic time parameter of the Weibull function fell within = 63.7–264.5 min, indicating slow, prolonged release. The shape parameter was >1 for eight of the nine plates (1.03–5.00), consistent with predominantly sigmoidal release kinetics and suggesting complex, diffusion-dominated transport rather than simple first-order release. Plates 1, 6, and 8, in particular, displayed high values (2.35, 3.99, and 5.00), corresponding to a pronounced acceleration phase in the middle of the release profile. Plate 9 was the only sample with < 1 ( = 0.688), indicative of a more pronounced initial burst followed by a progressively slower release phase.
Experimentally, the coatings released approximately 45–80% of their loaded amikacin over the studied period (Figure 8), demonstrating sustained release behavior. All release profiles were biphasic, with an initial burst phase followed by a slower, diffusion-controlled release consistent with the Weibull fits.
Figure 8.
Experimental drug release profiles with best-fit kinetic model curves for Plates 1–9.
3.4. Antimicrobial Activity
Kirby–Bauer assays confirmed functional antibacterial performance (Table 3, Figure 3 and Figure 4). Samples with higher cumulative release (e.g., Plates 1–2) generated larger inhibition zones (37–40 mm for S. aureus, ~37 mm for E. coli). In contrast, Plates 6 and 9, with lower release (~30–45%), produced smaller inhibition zones (28–32 mm).
Table 3.
Antimicrobial activity (Kirby–Bauer method) of coated PVA–amikacin films against Staphylococcus aureus ATCC 6538 and Escherichia coli ATCC 8739. Data are means ± SD (n = 3).
| Substance Name | Sample Number | Zone of Inhibition, mm | |||
|---|---|---|---|---|---|
| Repetition | Average | ||||
| 1 | 2 | 3 | |||
|
Staphylococcus aureus ATCC 6538 |
1 | 37.0 | 37.0 | 37.0 | 37.0 ± 0.00 |
| 2 | 35.0 | 35.0 | 35.0 | 35.0 ± 0.00 | |
| 3 | 30.0 | 30.0 | 30.0 | 30.0 ± 0.00 | |
| 4 | 35.0 | 35.0 | 35.0 | 35.0 ± 0.00 | |
| 5 | 32.0 | 33.0 | 33.0 | 32.6 ± 0.58 | |
| 6 | 28.0 | 28.0 | 30.0 | 28.6 ± 1.15 | |
| 7 | 35.0 | 37.0 | 37.0 | 36.3 ± 1.15 | |
| 8 | 35.0 | 35.0 | 35.0 | 35.0 ± 0.00 | |
| 9 | 40.0 | 40.0 | 40.0 | 40.0 ± 0.00 | |
|
Escherichia coli ATCC 8739 |
1 | 34.0 | 35.0 | 35.0 | 34.7 ± 0.58 |
| 2 | 38.0 | 38.0 | 38.0 | 38.0 ± 0.00 | |
| 3 | 30.0 | 32.0 | 32.0 | 31.3 ± 1.15 | |
| 4 | 33.0 | 33.0 | 35.0 | 33.7 ± 1.15 | |
| 5 | 35.0 | 35.0 | 35.0 | 35.0 ± 0.00 | |
| 6 | 28.0 | 32.0 | 30.0 | 30.0 ± 2.00 | |
| 7 | 35.0 | 35.0 | 35.0 | 35.0 ± 0.00 | |
| 8 | 35.0 | 35.0 | 35.0 | 35.0 ± 0.00 | |
| 9 | 33.0 | 30.0 | 30.0 | 31.0 ± 1.73 | |
3.5. Correlation of Drug Release with Antimicrobial Activity
To directly compare release kinetics with antibacterial activity, cumulative release percentages at 60 min and 480 min were paired with inhibition zone diameters against S. aureus and E. coli (Table 4). Average inhibition zones were calculated as the means of the two bacterial strains. A scatter plot (Figure 9) was generated to visualize the relationship between release percentage and inhibition zone size.
Table 4.
Comparison of release percentage at 60 min and 480 min with inhibition zone diameters.
| Plate | Release % at 60 min | Release % at 480 min | S. aureus Zone (mm) | E. coli Zone (mm) | Average Zone (mm) |
|---|---|---|---|---|---|
| 1 | 15 | 99.21 | 37.0 | 34.67 | 35.8 |
| 2 | 29 | 100.00 | 35.0 | 37.00 | 36.00 |
| 3 | 11 | 57.27 | 30.0 | 31.33 | 30.8 |
| 4 | 74 | 100.00 | 35.0 | 33.67 | 34.3 |
| 5 | 36 | 95.84 | 32.7 | 35.00 | 33.8 |
| 6 | 2 | 83.05 | 28.7 | 30.00 | 29.3 |
| 7 | 47 | 100.00 | 36.3 | 35.00 | 35.7 |
| 8 | 0.09 | 100.00 | 35.0 | 35.00 | 35.00 |
| 9 | 48 | 94.24 | 40.0 | 31.00 | 35.50 |
Figure 9.
Correlation between release percentage and inhibition zone size for Plates 1–9.
3.6. Statistical Analysis
Two-way ANOVA revealed a significant main effect of speed on drug release (F(2,63) = 5.60, p = 0.0058), whereas distance (F(2,63) = 1.41, p = 0.252) and the speed × distance interaction (F(4,63) = 0.64, p = 0.638) were not significant (Figure 10 (left); Table 5). Post hoc Tukey analysis showed that drug release at 20 cm/s was significantly lower than at 10 cm/s (p = 0.0053) and 15 cm/s (p = 0.0465), while no difference was observed between 10 and 15 cm/s (Figure 10 (right); Table A1, see Appendix A.2).
Figure 10.
Effect of plasma parameters on drug release. (Left) Interaction plot showing mean concentration (±95% CI) across nozzle distance (15, 20, 25 mm) for each speed (10, 15, 20 cm/s). Lines are approximately parallel, consistent with a non-significant speed × distance interaction [F(4,63) = 0.64, p = 0.638]. Error bars in the panel denote 95% confidence intervals. (Right) Boxplots of concentration grouped by speed with individual observations overlaid. Distributions at 20 cm/s (green) are shifted lower than at 10 (blue) and 15 cm/s (red), supporting the significant main effect of speed (F(2,63) = 5.60, p = 0.0058) and Tukey contrasts (20 cm/s < 10 cm/s, p = 0.0053; 20 cm/s < 15 cm/s, p = 0.0465; 10 cm/s vs. 15 cm/s, n.s.).
Table 5.
Two-way ANOVA results for the effects of nozzle speed and distance on drug release. ** indicates statistical significance at p < 0.01.
| Source | Sum of Squares | df | F | p-Value |
|---|---|---|---|---|
| Speed | 9715.20 | 2 | 5.60 | 0.0058 ** |
| Distance | 2441.90 | 2 | 1.41 | 0.252 |
| Speed × Distance | 2211.21 | 4 | 0.64 | 0.638 |
| Residual | 54,660.18 | 63 | - | - |
Assumption checks indicated violations of normality (Shapiro–Wilk, p < 0.05) and homogeneity of variance (Levene’s test, p = 0.005). To confirm robustness, nonparametric testing was performed. The Kruskal–Wallis test identified significant group differences (H = 17.78, p < 0.001), and Mann–Whitney U tests with Bonferroni correction confirmed that 20 cm/s differed significantly from both 10 cm/s (p < 0.001) and 15 cm/s (p = 0.0017), while 10 vs. 15 cm/s remained non-significant (Table A2, see Appendix A.2).
Together, these results demonstrate that nozzle speed is the dominant factor governing drug release, while spraying distance has no independent or interactive effect.
4. Discussion
4.1. Coating Mass and Plasma Parameters
Our results show that plasma spraying can produce coatings with masses ranging from 150 to 490 mg, depending on the nozzle speed and spraying distance (Table 1). The greatest deposition was achieved with lower spraying speeds (10 cm/s) at short distances (15–20 mm), while substantially lower coating mass was produced with higher speeds (20 cm/s) and longer distances (25 mm). This relationship can be attributed to particle residence time: at lower nozzle speeds, plasma-entrained particles have more time to settle and adhere to the substrate, whereas higher speeds reduce contact time and deposition efficiency. Similar trends have been reported in the plasma spraying of polymers and ceramics, where nozzle speed and stand-off distance are critical factors governing film thickness and uniformity. Interestingly, substrate thickness (100 µm vs. 200 µm) had no effect on deposition, indicating that heat dissipation and stability were sufficient in both cases. These findings suggest that nozzle dynamics, rather than substrate characteristics, are the primary determinants of coating yield in CAP [14].
Unlike solvent-cast PVA films, which typically form homogeneous and relatively dense matrices, CAP-deposited coatings exhibit heterogeneous, plasma-generated architectures characterized by variable porosity and surface roughness. Optical microscopy confirmed that plasma parameters directly control these features, with low nozzle speeds producing rough, porous coatings and high speeds yielding thinner, smoother films.
This structural heterogeneity is reflected in the release kinetics. The predominance of Weibull behavior with β > 1 suggests complex transport mechanisms involving combined diffusion and structural effects, rather than the simple Fickian diffusion commonly reported for solvent-cast systems.
Therefore, CAP deposition introduces an additional level of control over drug release by tailoring microstructure through processing conditions, which is not readily achievable with conventional solvent-based methods.
4.2. Surface Morphology and Roughness
Microscopic analysis revealed distinct morphological differences (Figure 2 and Figure 7). At short nozzle distances and low speeds, the surfaces appear rougher and more porous, while long distances yield smoother films with reduced porosity. Optical microscopy depth analysis confirmed that the increased roughness corresponds to a higher surface area, a property known to influence drug loading capacity and release kinetics. Previous works on plasma-modified polymer coatings support this observation, as increased porosity enhances fluid penetration and accelerates drug diffusion [15]. In the context of antimicrobial implants, rougher surfaces may also improve tissue integration [16], though excessively porous films could compromise coating integrity. Thus, controlling the plasma parameters offers a route to balancing adhesion, porosity, and release performance.
4.3. Drug Release Profiles
Drug release studies demonstrated sustained biphasic release from all coatings (Table 2, Figure 8). Each profile showed an initial burst phase followed by a slower, diffusion-controlled stage, consistent with the typical behavior of PVA matrices, where rapid release originates from surface-associated drug while prolonged release is governed by diffusion through the hydrated polymer bulk [17,18,19]. Experimentally, the coatings released approximately 45–80% of their loaded amikacin over the studied period, indicating sustained but partial drug depletion from the films. Drug release from the PVA–amikacin coatings was well described by the Weibull model for all nine plates (Table 2). The fits were excellent, with coefficients of determination ranging from 0.902 to 0.996 and AIC values between −29.8 and −54.6.
For some samples (e.g., Plate 4), the Weibull model provided a comparatively lower fit (R2 ≈ 0.90), suggesting that more complex or multi-phase release mechanisms may be present; however, a unified modeling approach was retained to allow consistent comparison across conditions.
The kinetic parameters also reflected the influence of the plasma spraying conditions. At a fixed nozzle speed of 10 cm/s (Plates 1–3), increasing the spraying distance from 15 to 25 mm shifted the profiles toward slower, more extended release: t50 values were 136.98, 120.69 and 208.15 min, respectively, with Plate 3 (25 mm) showing the largest scale parameter (264.5 min) and the slowest release (t80 = 361 min). At a fixed distance of 15 mm (Plates 1, 4, 7), increasing nozzle speed from 10 to 20 cm/s reduced coating mass (490 → 330 → 240 mg) and accelerated early release (t10 = 61.45 → 7.47 → 19.38 min), while altering the shape parameter ( = 2.35 → 1.05 → 1.34), consistent with more heterogeneous coating microstructures. The slowest, most sustained release was observed for Plate 3 (25 mm, 10 cm/s), whereas Plate 4 (15 mm, 15 cm/s) released 80% of its load in ~100 min, representing the fastest profile. Overall, coating mass broadly correlated with total release time, but deviations such as Plate 8 (low mass, 150 mg, yet very high = 5) suggest that microstructural features generated by specific plasma parameter combinations also play a critical role in governing drug transport. These observations highlight that CAP spraying offers a powerful handle to tune antibiotic release kinetics, but also that careful optimization of both distance and speed is required to avoid excessive burst release while maintaining a therapeutically relevant, sustained delivery window. Comparable sustained or controlled-release behaviour has been reported for PVA-based films and hydrogels prepared by conventional solvent-based methods, including simple PVA matrices and PVA blends with chitosan, gelatin, hyaluronic acid, or other polysaccharides for ocular and wound-healing applications [20,21,22,23,24,25,26,27].
4.4. Antimicrobial Activity
Kirby–Bauer assays confirmed the antibacterial functionality of the PVA–amikacin films (Table 3; Figure 3 and Figure 4). Coatings with a higher release (e.g., Plates 1–2) yielded larger inhibition zones (~35–40 mm), while those with a lower release (~30–45%) produced smaller zones (28–32 mm). This direct relationship supports the biological relevance of the release profiles. Notably, both S. aureus and E. coli were effectively inhibited, demonstrating broad-spectrum efficacy. Previous studies have highlighted amikacin’s suitability for implant-associated infections due to its stability and activity against resistant strains [28]. Comparable local delivery approaches include amikacin-loaded porous PLA/PVA matrices, which generate inhibition zones of ~7–9.5 mm against S. aureus and E. coli while providing multi-day release, and amikacin-coated 3D-printed stainless-steel implants, which produce inhibition areas of 0.9–3.0 cm2 (≈10–20 mm equivalent diameter) and maintain antibacterial activity for up to 7 days [29,30]. Encapsulation of amikacin into degradable PLA-based microparticles has also been shown to support prolonged release at concentrations exceeding MIC values for common pathogens [31]. In this context, our CAP-deposited PVA coatings achieve relatively large inhibition zones after 24 h, indicating at least comparable short-term antibacterial efficiency, although direct comparison is limited by differences in sample geometry, drug loading, and assay conditions [29,30]. Our results further suggest that plasma-deposited PVA coatings can serve as an efficient local delivery platform, potentially reducing the need for systemic antibiotic administration and minimizing associated side effects.
Although the chemically defined medium was optimized for the present system, its applicability to other thraustochytrid species remains to be evaluated, and future studies should investigate its performance across diverse strains.
4.5. Correlation of Drug Release with Antimicrobial Activity
Correlation analysis revealed moderate to strong positive associations between release percentage (at 60 and 480 min) and inhibition zone diameters (Figure 9; Table 4). The strongest correlations were observed for cumulative release at 480 min, which aligns with antibacterial outcomes measured at 24 h. This suggests that early release kinetics critically shape antimicrobial efficacy in vitro [32]. The weaker correlations at 60 min reflect the fact that, while measurable, initial burst release is not always sufficient to fully explain inhibition zone sizes, which integrate drug exposure over time [33,34]. These findings highlight the importance of sustained release for maintaining antibacterial activity, suggesting that plasma-optimized coatings capable of extended release will be the most clinically effective.
4.6. Statistical Analysis
Statistical evaluation confirmed that nozzle speed was the dominant factor governing drug release [35], while spraying distance had no significant main or interaction effect (Figure 10; Table 5). ANOVA identified significant differences between speeds, with 20 cm/s producing consistently lower release compared to 10 and 15 cm/s. Tukey and Mann–Whitney tests confirmed these trends, though no difference was observed between 10 and 15 cm/s. The robustness of these findings across parametric and nonparametric approaches underscores the reliability of speed as the key controllable factor. From a mechanical perspective, higher speeds reduce residence time and deposition efficiency, resulting in thinner coatings with less drug loading and reduced release. These insights provide practical guidance for optimising the plasma process, emphasizing speed control over distance adjustment [36].
4.7. Broader Implications and Future Directions
Collectively, our findings support the feasibility of CAP as a versatile platform for producing antibiotic-loaded biodegradable coatings with tunable properties [37]. Unlike dip-coating or solvent casting, plasma spraying eliminates the need for solvents, minimizes processing defects, and enables precise control over film morphology and drug distribution. This makes it especially attractive for clinical translation, where reproducibility and biocompatibility are critical [38]. Future research should explore (i) incorporating different antibiotics or combinations to address resistant pathogens, (ii) multilayer coatings to decouple burst and sustained release phases, (iii) in vivo evaluation of antimicrobial efficacy and biocompatibility, and (iv) scale-up strategies for coating complex implant geometries. By addressing these areas, plasma-assisted deposition could advance the development of next-generation antimicrobial implants with improved infection control and patient outcomes.
5. Conclusions
This study demonstrates that cold atmospheric pressure plasma (CAP) deposition is an effective solvent-free approach for fabricating polyvinyl alcohol amikacin coatings with tunable structural and functional properties. Using a 3 × 3 factorial design (nozzle speed: 10–20 cm/s; distance: 15–25 mm), coating masses in the range of 150–490 mg were obtained, confirming strong process–structure relationships.
Nozzle speed was identified as the dominant parameter governing coating mass and drug release (two-way ANOVA: F(2,63) = 5.60, p = 0.0058), whereas spraying distance had no statistically significant effect (p = 0.252). Lower speeds produced thicker, more porous coatings with higher cumulative release, while higher speeds resulted in thinner films with reduced drug loading. Drug release profiles were consistently described by the Weibull model (R2 = 0.902–0.996; AIC = −29.8 to −54.6), with characteristic times τ = 63.7–264.5 h and shape parameters β = 1.03–5.00, indicating complex, non-Fickian transport mechanisms.
All coatings exhibited sustained, biphasic release behavior, delivering approximately 45–80% of the incorporated amikacin over the studied period. Antibacterial assays confirmed effective activity against Staphylococcus aureus and Escherichia coli, with inhibition zones ranging from ~28 to 40 mm, correlating with cumulative drug release.
Overall, CAP deposition enables precise control of coating architecture and drug release behavior through process parameters, providing a versatile platform for localized antibiotic delivery on medical implants. Future work should focus on in vivo validation, extension to other antimicrobial agents, and scaling the process for complex implant geometries. By directly linking plasma process parameters to functional drug delivery outcomes, this work establishes CAP deposition as a fundamentally new paradigm for designing antimicrobial coatings, moving beyond passive materials toward actively engineered, performance-driven implant surfaces.
Acknowledgments
Rinat Iskakov (Chemical & Biochemical Engineering Department, Satbayev University)—for mentorship, Nemkayeva Renata—for help with microscopy images obtaining.
Appendix A
Appendix A.1. Assumption Checks
Although the primary analysis was conducted using two-way ANOVA, assumption checks indicated deviations from normality and homogeneity of variance. Shapiro–Wilk tests showed non-normal distributions across speed groups (W = 0.65–0.85, p < 0.01), while Levene’s test revealed unequal variances (F(2,65) = 5.73, p = 0.005).
Appendix A.2. Nonparametric Validation
To ensure robustness, nonparametric analyses were performed. The Kruskal-Wallis test confirmed significant differences among speed groups (H = 17.78, df = 2, p < 0.001). Pairwise Mann-Whitney U tests with Bonferroni correction showed that 20 cm/s differed significantly from 10 cm/s (p < 0.001) and 15 cm/s (p = 0.0017), while no difference was observed between 10 and 15 cm/s (p = 0.224). These findings are consistent with the ANOVA and Tukey HSD results, supporting the conclusion that nozzle speed is the key determinant of drug release.
Table A1.
Tukey’s HSD post hoc comparisons for the effect of Speed on drug release.
| Comparison (Speed, cm/s) | Mean Difference | 95% CI (Lower, Upper) |
p-Value | Significant (p < 0.05) |
|---|---|---|---|---|
| 10 vs. 15 | –6.79 | –27.06, 13.48 | 0.703 | No |
| 10 vs. 20 | –27.33 | –47.60, –7.05 | 0.0053 | Yes |
| 15 vs. 20 | –20.53 | –40.81, –0.26 | 0.0465 | Yes |
Table A2.
Nonparametric tests for the effect of Speed on drug release.
| Test | Statistic | df | p-Value | Interpretation |
|---|---|---|---|---|
| Sharipo-Wilk (normality) | W = 0.65–0.85 (per group) | - | p < 0.01 | Normality violated |
| Levene’s test (equal variances) | F = 5.73 | 2.65 | 0.005 | Homogeneity violated |
| Kruskal-Wallis test | H = 17.78 | 2 | 0.00014 | Significant differences across groups |
| Mann-Whitney U (10 vs. 15) | U = 347.5 | - | 0.224 | Not significant |
| Mann-Whitney U (10 vs. 20) | U = 476.5 | - | 0.00011 | Significant |
| Mann-Whitney U (15 vs. 20) | U = 441.0 | - | 0.0017 | Significant |
Appendix A.3. Correlation of Early Drug Release with Antimicrobial Activity
To further examine the relationship between early release and antibacterial performance, cumulative release percentages at 60 min were compared with inhibition zone diameters against S. aureus, E. coli, and their average values. Linear regression analysis indicated weak to moderate correlations, with R2 values ranging from 0.03 to 0.25, suggesting that short-term release alone does not fully explain antibacterial activity (Figure A1).
Figure A1.
Correlation between cumulative release at 60 min and inhibition zone size. (A) Average inhibition zone (mean of S. aureus and E. coli), (B) S. aureus inhibition zone, (C) E. coli inhibition zone. Dashed lines represent linear regression fits with equations and R2 values shown in each panel.
Appendix B
Comparative Fitting of Alternative Kinetic Models
AIC values were not calculated for these empirical models; comparison is based on RSS and R2.
| Plate | Model | Equation | RSS | R2 |
| 1 | Linear | (0.016x + 4.974) | 325.61 | 0.5788 |
| Logarithmic | (5.254\ln(x) − 13.413) | 112.79 | 0.8541 | |
| Polynomial | (−4.006 × 10^{−5}x^2 + 0.074x − 0.056) | 47.87 | 0.9381 | |
| Exponential | (25.09(1−e^{−0.0046x})) | 40.21 | 0.9480 | |
| 2 | Linear | (0.009x + 3.782) | 101.70 | 0.5862 |
| Logarithmic | (3.074\ln(x) − 7.057) | 19.82 | 0.9194 | |
| Polynomial | (−2.292 × 10^{−5}x^2 + 0.042x + 0.904) | 10.75 | 0.9563 | |
| Exponential | (14.96(1−e^{−0.0055x})) | 2.30 | 0.9906 | |
| 3 | Linear | (0.006x + 1.214) | 24.96 | 0.7162 |
| Logarithmic | (1.757\ln(x) − 4.739) | 14.16 | 0.8390 | |
| Polynomial | (−1.182 × 10^{−5}x^2 + 0.023x − 0.271) | 0.78 | 0.9911 | |
| Exponential | (9.17(1−e^{−0.0030x})) | 3.37 | 0.9617 | |
| 4 | Linear | (0.003x + 3.119) | 36.36 | 0.3427 |
| Logarithmic | (1.402\ln(x) − 2.079) | 8.28 | 0.8503 | |
| Polynomial | (−1.032 × 10^{−5}x^2 + 0.018x + 1.823) | 17.93 | 0.6759 | |
| Exponential | (6.60(1−e^{−0.0164x})) | 5.31 | 0.9039 | |
| 5 | Linear | (0.007x + 2.607) | 43.65 | 0.6239 |
| Logarithmic | (2.134\ln(x) − 4.880) | 7.13 | 0.9385 | |
| Polynomial | (−1.476 × 10^{−5}x^2 + 0.028x + 0.753) | 5.92 | 0.9490 | |
| Exponential | (10.46(1−e^{−0.0053x})) | 1.69 | 0.9854 | |
| 6 | Linear | (0.001x + 0.796) | 9.55 | 0.2049 |
| Logarithmic | (0.502\ln(x) − 1.064) | 5.97 | 0.5027 | |
| Polynomial | (−3.613 × 10^{−6}x^2 + 0.006x + 0.342) | 7.29 | 0.3931 | |
| Exponential | (2.20(1−e^{−0.0114x})) | 4.89 | 0.5926 | |
| 7 | Linear | (0.005x + 3.408) | 75.88 | 0.3312 |
| Logarithmic | (1.949\ln(x) − 3.798) | 22.63 | 0.8005 | |
| Polynomial | (−1.525 × 10^{−5}x^2 + 0.027x + 1.492) | 35.60 | 0.6862 | |
| Exponential | (8.87(1−e^{−0.0112x})) | 11.48 | 0.8988 | |
| 8 | Linear | (0.003x + 0.600) | 11.04 | 0.6251 |
| Logarithmic | (0.898\ln(x) − 2.390) | 10.16 | 0.6548 | |
| Polynomial | (−6.636 × 10^{−6}x^2 + 0.013x − 0.234) | 3.42 | 0.8840 | |
| Exponential | (5.06(1−e^{−0.0026x})) | 5.21 | 0.8232 | |
| 9 | Linear | (0.001x + 1.228) | 1.99 | 0.5758 |
| Logarithmic | (0.383\ln(x) − 0.089) | 1.18 | 0.7482 | |
| Polynomial | (−2.608 × 10^{−6}x^2 + 0.005x + 0.900) | 0.81 | 0.8273 | |
| Exponential | (2.52(1−e^{−0.0115x})) | 1.65 | 0.6478 |
Author Contributions
Conceptualization, A.K., D.R. and D.T.; methodology, Z.S., A.K., D.R. and D.T.; investigation, Z.S., A.K. and M.K.S.; formal analysis, A.K.; data curation, A.K.; resources, D.T., D.R. and B.D.; writing—original draft preparation, A.K.; writing—review and editing, A.K., B.D., D.R. and D.T.; visualization, A.K.; supervision, D.R. and D.T.; project administration, D.T. All authors have read and agreed to the published version of the manuscript.
Data Availability Statement
The data supporting the findings of this study are available from the corresponding author upon reasonable request. Although one of the authors is affiliated with a commercial company, the company had no restrictions on data availability, data interpretation, or publication of the results.
Conflicts of Interest
Author Dias Tastanbekov was employed by the company Akyldy Food Supplements. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Funding Statement
This research received no external funding. This research is an independent study.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Dadi N., Radochová B., Vargová J., Bujdáková H. Impact of Healthcare-Associated Infections Connected to Medical Devices-An Update. Microorganisms. 2021;9:2332. doi: 10.3390/microorganisms9112332. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Tiwari A., Sharma P., Vishwamitra B., Singh G. Review on Surface Treatment for Implant Infection via Gentamicin and Antibiotic Releasing Coatings. Coatings. 2021;11:1006. doi: 10.3390/coatings11081006. [DOI] [Google Scholar]
- 3.Georgieva N., Bryaskova R., Tzoneva R. New Polyvinyl alcohol-based hybrid materials for biomedical application. Mater. Lett. 2012;88:19–22. doi: 10.1016/j.matlet.2012.07.111. [DOI] [Google Scholar]
- 4.Jain D., Carvalho E., Banerjee R. Biodegradable hybrid polymeric membranes for ocular drug delivery. Acta Biomater. 2010;6:1370–1379. doi: 10.1016/j.actbio.2009.11.001. [DOI] [PubMed] [Google Scholar]
- 5.Yunus Y., Mahadzir N.A., Mohamed Ansari M.N., Tg Abd Aziz T.H., Mohd Afdzaluddin A., Anwar H., Wang M., Ismail A.G. Review of the Common Deposition Methods of Thin-Film Pentacene, Its Derivatives, and Their Performance. Polymers. 2022;14:1112. doi: 10.3390/polym14061112. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Moreau D. Ph.D. Thesis. Université Paris Sciences et Lettres; Paris, France: 2016. Design and Characterization of Hydrogel Films and Hydrogel-Ceramic Composites for Biomedical Applications. (In English) [Google Scholar]
- 7.Kravanja K.A., Finšgar M. A Review of Techniques for the Application of Bioactive Coatings on Metal-Based Implants to Achieve Controlled Release of Active Ingredients. Mater. Des. 2022;217:110653. doi: 10.1016/j.matdes.2022.110653. [DOI] [Google Scholar]
- 8.Rezaei H., Matin A.A., Mohammadnejad M. Cold Atmospheric Plasma Treated 3D Printed Polylactic Acid Film; Application in Thin Film Solid Phase Microextraction of Anticancer Drugs. Talanta. 2024;266:125064. doi: 10.1016/j.talanta.2023.125064. [DOI] [PubMed] [Google Scholar]
- 9.Ermukhambetova A., Berillo D. Green synthesis of silver nanoparticles using paper wasp’s hydrolysate with antibacterial activity. Results Surf. Interfaces. 2023;11:100114. doi: 10.1016/j.rsurfi.2023.100114. [DOI] [Google Scholar]
- 10.Costa P., Lobo J.M.S. Modeling and comparison of dissolution profiles. Eur. J. Pharm. Sci. 2001;13:123–133. doi: 10.1016/s0928-0987(01)00095-1. [DOI] [PubMed] [Google Scholar]
- 11.Dash S., Murthy P.N., Nath L., Chowdhury P. Kinetic modeling on drug release from controlled drug delivery systems. Acta Pol. Pharm. 2010;67:217–223. [PubMed] [Google Scholar]
- 12.Burnham K.P., Anderson D.R. Model Selection and Multimodel Inference. Springer; Berlin/Heidelberg, Germany: 2004. [DOI] [Google Scholar]
- 13.Motulsky H., Christopoulos A. Fitting Models to Biological Data Using Linear and Nonlinear Regression. Oxford University Press; Oxford, UK: 2004. [DOI] [Google Scholar]
- 14.Lorenzo-Bañuelos M., Díaz A., Rodríguez D., Cuesta I.I., Fernández A., Alegre J.M. Influence of Atmospheric Plasma Spray Parameters (APS) on the Mechanical Properties of Ni-Al Coatings on Aluminum Alloy Substrate. Metals. 2021;11:612. doi: 10.3390/met11040612. [DOI] [Google Scholar]
- 15.Rajesh K., Rangaswamy M., Zhang C., Haldar S., Kumarasamy M., Agarwal A., Roy P., Lahiri D. Surface Modified Metallic Orthopedic Implant for Sustained Drug Release and Osteocompatibility. ACS Appl. Bio Mater. 2019;2:4181–4192. doi: 10.1021/acsabm.9b00443. [DOI] [PubMed] [Google Scholar]
- 16.Wu B., Chen X., Lu H., Zhao X., Wang G. Recent Progress in the Surface Modification of Bone Implant Using Plasma Spraying Technique. China Surf. Eng. 2021;34:1–11. doi: 10.11933/j.issn.1007-9289.20210410001. [DOI] [Google Scholar]
- 17.Otarbayeva S., Berillo D. Poly(Vinyl Alcohol) Drug and PVA–Drug–Surfactant Complex Organogel with Dimethyl Sulfoxide as a Drug Delivery System. Gels. 2024;10:753. doi: 10.3390/gels10110753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Liu Z., Zhu Y., Liu X., Yeung K., Wu S. Construction of poly (vinyl alcohol)/poly (lactide-glycolide acid)/vancomycin nanoparticles on titanium for enhancing the surface self-antibacterial activity and cytocompatibility. Colloids Surf. B Biointerfaces. 2017;151:165–177. doi: 10.1016/j.colsurfb.2016.12.016. [DOI] [PubMed] [Google Scholar]
- 19.Shen J., Burgess D. Accelerated in vitro release testing of implantable PLGA microsphere/PVA hydrogel composite coatings. Int. J. Pharm. 2012;422:341–348. doi: 10.1016/j.ijpharm.2011.10.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Rogojanu A., Rusu E., Olaru N., Dobromir M., Dorohoi D.O. Development and characterization of poly(vinyl alcohol) matrix for drug release. Dig. J. Nanomater. Biostructures. 2011;6:809–818. doi: 10.12775/equil2011.026. [DOI] [Google Scholar]
- 21.Wang L.C., Chen X.G., Yu L.J., Li P.W. Controlled drug release through carboxymethyl-chitosan/poly(vinyl alcohol) blend films. Polym. Eng. Sci. 2007;47:1373–1379. doi: 10.1002/pen.20823. [DOI] [Google Scholar]
- 22.Pavaloiu R.D., Stoica-Guzun A., Stroescu M., Jinga S.I., Dobre T. Composite films of poly(vinyl alcohol)–chitosan–bacterial cellulose for drug controlled release. Int. J. Biol. Macromol. 2014;68:117–124. doi: 10.1016/j.ijbiomac.2014.04.040. [DOI] [PubMed] [Google Scholar]
- 23.Ghezzi M., Ferraboschi I., Fantini A., Pescina S., Padula C., Santi P., Sissa C., Nicoli S. Hyaluronic acid–PVA films for the simultaneous delivery of dexamethasone and levofloxacin to ocular tissues. Int. J. Pharm. 2023;638:122911. doi: 10.1016/j.ijpharm.2023.122911. [DOI] [PubMed] [Google Scholar]
- 24.Suhail M., Shih C.M., Liu J.Y., Hsieh W.C., Lin Y.W., Lin I.L., Wu P.C. Synthesis of glutamic acid/polyvinyl alcohol based hydrogels for controlled drug release: In-vitro characterization and in-vivo evaluation. J. Drug Deliv. Sci. Technol. 2022;75:103715. doi: 10.1016/j.jddst.2022.103715. [DOI] [Google Scholar]
- 25.Das S., Behera A., Habibullah S., Pattnaik G., Mohanty B. Moxifloxacin-loaded PVA-chitosan composite films as potential ocular drug delivery systems: A comprehensive characterization and efficacy assessment. Int. J. Biol. Macromol. 2025;296:139726. doi: 10.1016/j.ijbiomac.2025.139726. [DOI] [PubMed] [Google Scholar]
- 26.Ali S., Ranjha N.M., Ahmad B., Khan A.A., Hassan F.U., Aziz T., Alharb M., Alshammari A., Alasmari A.F., Alharbi M.E. Sustained release of drug facilitated through chemically crosslinked polyvinyl alcohol-gelatin (PVA-GE) hydrogels: A sustainable biomedical approach. Pol. J. Chem. Technol. 2023;25:56–65. doi: 10.2478/pjct-2023-0017. [DOI] [Google Scholar]
- 27.Abdollahi S., Chamchangi M.A., Raoufi Z., Heidari F. Dual-layer alginate hydrogel dressings with chitosan nanofibers for enhanced wound healing, infection prevention, and controlled drug release. Int. J. Biol. Macromol. 2025;307:142033. doi: 10.1016/j.ijbiomac.2025.142033. [DOI] [PubMed] [Google Scholar]
- 28.Ramirez M.S., Tolmasky M.E. Amikacin: Uses, Resistance, and Prospects for Inhibition. Molecules. 2017;22:2267. doi: 10.3390/molecules22122267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Zhang C., Poudel I., Mita N., Kang X., Annaji M., Lee S., Panizzi P., Shamsaei N., Fasina O., Babu R.J., et al. Amikacin coated 3D-printed metal devices for prevention of postsurgical infections (PSIs) Pharmaceutics. 2025;17:911. doi: 10.3390/pharmaceutics17070911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Glinka M., Filatova K., Kucińska-Lipka J., Šopík T., Domincová Bergerová E., Mikulcová V., Wasik A., Sedlařík V. Antibacterial porous systems based on polylactide loaded with amikacin. Molecules. 2022;27:7045. doi: 10.3390/molecules27207045. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Glinka M., Filatova K., Kucińska-Lipka J., Bergerová E.D., Wasik A., Sedlařík V. Encapsulation of amikacin into microparticles based on low-molecular-weight poly(lactic acid) and poly(lactic acid-co-polyethylene glycol) Mol. Pharm. 2021;18:2986–2996. doi: 10.1021/acs.molpharmaceut.1c00193. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Ávila P., Moliz T., Leal J., Baca P., Valverde M., Hernández M. Novel Ti surface coated with PVA hydrogel and chitosan nanoparticles with antibacterial drug release: An experimental in vitro study. Clin. Implant Dent. Relat. Res. 2024;26:427–441. doi: 10.1111/cid.13305. [DOI] [PubMed] [Google Scholar]
- 33.Lv J., Li X., Yin H., Wang L., Pei Y., Lv X. Controlled release of vancomycin hydrochloride from a composite structure of polymeric films and porous fibers on implants. Chem. Eng. J. 2017;325:601–610. doi: 10.1016/J.CEJ.2017.05.118. [DOI] [Google Scholar]
- 34.Sirc J., Kubínová Š., Hobzova R., Stránská D., Kozlík P., Bosáková Z., Mareková D., Holan V., Sykova E., Michalek J. Controlled gentamicin release from multi-layered electrospun nanofibrous structures of various thicknesses. Int. J. Nanomed. 2012;7:5315–5325. doi: 10.2147/IJN.S35781. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Schwenk A., Nutsch G., Gruner H. Modified Nozzle for the Atmospheric Plasma Spraying; Proceedings of the Thermal Spray 2003: Proceedings from the International Thermal Spray Conference; Orlando, FL, USA. 5–8 May 2003; [DOI] [Google Scholar]
- 36.Martin-Salvador P., Verschueren R., De Beer T., Kumar A. CFD-DEM model of a cold plasma assisted fluidized bed powder coating process. Front. Chem. Eng. 2024;6:1347313. doi: 10.3389/fceng.2024.1347313. [DOI] [Google Scholar]
- 37.Rajesh K., Khatua C., Singh P., Roy P., Keshri A., Lahiri D. Microsphere embedded hydroxyapatite coating on metallic implant for sustained drug release in orthopedic applications. J. Drug Deliv. Sci. Technol. 2024;98:105840. doi: 10.1016/j.jddst.2024.105840. [DOI] [Google Scholar]
- 38.Olayiwola B., O’Neill F., Frewen C., Kavanagh D., O’Hara R., O’Neill L. Cold Plasma Deposition of Tobramycin as an Approach to Localized Antibiotic Delivery to Combat Biofilm Formation. Pathogens. 2024;13:326. doi: 10.3390/pathogens13040326. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data supporting the findings of this study are available from the corresponding author upon reasonable request. Although one of the authors is affiliated with a commercial company, the company had no restrictions on data availability, data interpretation, or publication of the results.