Abstract
Background: Adult hearing loss is common, yet hearing-aid uptake remains inadequate due to systemic and financial barriers in traditional audiologist-based fitting models. Over-the-counter (OTC) and self-fitting strategies have emerged to improve accessibility, allowing users to select and adjust their devices independently. However, concerns remain that reducing professional involvement may compromise electroacoustic accuracy, speech-in-noise outcomes, and overall patient satisfaction. Methods: A comprehensive literature search across PubMed, Web of Science, CENTRAL, Scopus, and Embase was conducted up to 1 June 2026. Parallel-group and crossover randomized controlled trials (RCTs) comparing self-fitting with professional-fitting were included. The primary outcome was patient-reported hearing benefit. Standardized mean differences (SMDs) and risk ratios (RRs) were pooled using a random-effects model. Results: Seven RCTs involving 1092 participants were included. The analysis revealed no statistically significant difference between self-fitting and professional-fitting strategies regarding patient-reported hearing benefit (SMD: 0.01, 95% CI [−0.44, 0.46], p = 0.97, I2 = 89.2%, very low certainty of evidence). Similarly, no significant differences were observed across secondary outcomes, including speech-in-noise performance (SMD: 0.11, 95% CI [−0.25, 0.47]; p = 0.55), hearing loss (p = 0.12), hearing-aid satisfaction (p = 0.31), device adherence (p = 0.65), or the need for follow-up adjustment (RR: 0.46, 95% CI [0.12, 1.71]; p = 0.25). Safety outcomes were sparse and did not demonstrate a statistically significant difference between the fitting approaches; however, the evidence was imprecise and inconclusive. Conclusions: Self-fitting hearing aids showed no clear differences from professional best-practice fitting across several patient-reported and functional outcomes in adults with uncomplicated mild-to-moderate hearing loss. However, the certainty of evidence was very low because of risk of bias, substantial heterogeneity, imprecision, and the small number of trials. These findings are therefore hypothesis-generating and suggest that well-designed self-fitting pathways may have potential as an accessible alternative for selected users, but larger, rigorous, and longer-term trials are needed before broader clinical implementation can be recommended.
Keywords: over the counter, self-fitting, audiologist fitting, hearing loss, speech-in-noise test, APHAB, systematic review, meta-analysis
1. Introduction
Adult hearing loss is a common condition associated with impaired communication, social isolation, and decreased quality of life [1,2]. Other consequences of untreated hearing loss include accelerated cognitive and psychosocial decline [3,4]. Hearing aids remain the most effective intervention for patients with mild-to-moderate sensorineural hearing loss (SNHL) [5]. However, despite the well-established clinical benefits, the overall hearing-aid uptake remains inadequate, and a large portion of the affected population remains without adequate care [6,7].
Traditional hearing-aid delivery typically consists of extensive professional evaluation, individualized device selection, acoustic fitting, real-ear verification, and follow-up adjustments [8]. This professional model can provide optimized electroacoustic amplification and patient-centered fitting. Still, it is often limited by high out-of-pocket costs, limited access to audiology services, travel burdens, and time constraints [9]. Accordingly, these systemic and financial challenges limit device uptake, especially among adults with mild-to-moderate hearing loss who may delay or avoid seeking professional hearing care [10].
To address these challenges, over-the-counter (OTC) and self-fitting strategies have been proposed as alternative service models. These models can allow users to independently select, physically fit, and adjust their devices through preset programs, smartphone apps, or automated self-fitting algorithms [11,12]. However, while these models can reduce the need for in-person professional care and extend rehabilitation to underserved populations, there remain some concerns [13]. Removing or reducing professional involvement may compromise electroacoustic fitting accuracy, objective aided benefits, speech-in-noise outcomes, overall patient satisfaction, and long-term device adherence [14].
Recent randomized controlled trials (RCTs) have compared self-fitting, OTC-style hearing-aid models with audiologist-fitted or best-practice clinical models [15,16,17,18,19,20,21,22]. However, these trials vary in their methodological design, self-fitting approaches, follow-up durations, and whether the professional fitting arm reflects clinical best practice. Accordingly, the current evidence on the comparative efficacy of self-fitting versus professional-fitting devices remains uncertain. Therefore, this systematic review and meta-analysis aim to synthesize RCTs evaluating self-fitting versus professional-fitting hearing aids in adults with mild-to-moderate hearing loss using OTC-style hearing-aid models.
2. Methodology
2.1. Review Protocol and Reporting Style
Ethical approval was not required as this meta-analysis relied exclusively on previously published studies. The study design and reporting adhered to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [23]. The review protocol was registered in the PROSPERO database (registration ID: CRD420261445071).
2.2. Information Sources and Search Strategy
A comprehensive systematic literature search was conducted across five electronic databases, including PubMed, Scopus, the Cochrane Central Register of Controlled Trials (CENTRAL), Embase, and Web of Science (WoS), from inception to 1 June 2026. The search strategy combined terms related to adult hearing loss, hearing aids, OTC, self-fitting delivery models, and professional or audiologist-based fitting. The exact search strategies that were used for each database are depicted in Table S1. No restrictions were imposed regarding language, country of origin, or publication status.
To enhance the completeness of the systematic search, the reference lists of all included studies were manually reviewed, and trial registries (i.e., ClinicalTrials.gov and ICTRP) were searched. Additional sources (i.e., ResearchGate and medRxiv) were also examined to identify potentially eligible unpublished studies. When required, corresponding authors were contacted to obtain missing information or to clarify study details.
2.3. Eligibility Criteria
We included parallel-group and crossover RCTs according to the following PICO framework:
Population: Adults aged 18 years or older with perceived or confirmed mild-to-moderate hearing loss who were candidates for hearing-aid use.
Intervention: OTC-style fitting strategies, including self-fitting, app-based self-fitting, preset-based self-fitting, consumer-decides, person-fit, or participant-directed hearing-aid fitting models in which participants independently selected, fitted, adjusted, or programmed the hearing aid without full individualized audiologist best-practice fitting.
Comparator: Audiologist professional-fit or best-practice hearing-aid fitting, including hearing-aid selection, programming, counseling, verification, or adjustment delivered by a hearing-care professional.
Outcomes: The primary outcome was self-reported hearing-aid benefit, measured using the abbreviated profile of hearing aid benefit (APHAB), PHAB, Glasgow hearing aid benefit profile (GHABP), or an equivalent score. Secondary outcomes included speech-in-noise performance, hearing loss, hearing-aid satisfaction, hearing-aid use adherence, device retention/willingness to keep hearing aids, need for follow-up adjustment, and adverse events.
We excluded non-randomized studies, observational studies, conference abstracts, protocols, reviews, editorials, and studies without a directly eligible comparator. For studies with multiple relevant intervention arms, eligible OTC/self-fit arms were combined when clinically appropriate to generate a single comparison against the audiologist-fit group. Specifically, self-fitting or consumer-directed arms were classified under the OTC/self-fit group, whereas professional or best-practice fitting arms were classified under the audiologist-fit group.
2.4. Study Selection Process
Records were managed and screened using an electronic platform. Two investigators (HA and NA) independently screened studies in a two-stage process, beginning with title and abstract screening, followed by full-text assessment of potentially eligible articles. Disagreements were resolved through consultation with a third reviewer (EA). The same approach was applied during risk of bias assessment (YYA and AMA) and statistical analysis (EA and AMA). When studies reported multiple eligible outcome measures or follow-up assessments, the measure and time point most directly corresponding to the prespecified outcome were selected, while avoiding duplicate inclusion of participants from the same trial. Where multiple publications reported the same trial or cohort, including extension or follow-up reports, they were treated as a single study and data were collated accordingly.
2.5. Data Extraction
A standardized Excel extraction form was developed and pilot-tested using a subset of eligible studies. Extracted data were grouped into three categories. First, study characteristics included author, year, country, study design, trial setting, recruitment period, total sample size, follow-up duration, hearing-aid device type, fitting approach, comparator fitting protocol, and key details of the intervention and professional fitting procedures. Second, baseline participant characteristics included age, sex, prior hearing-aid experience, proportion of hearing-aid-naïve participants, hearing-loss severity, pure-tone average definition and values, baseline speech-in-noise or word-recognition performance, and baseline patient-reported hearing difficulty. Third, outcome data included all primary and secondary outcomes described above.
Dichotomous outcomes were extracted as the number of events and the total number of randomized or analyzed participants. Continuous outcomes were extracted as means and standard deviations (SDs). When means and SDs were not directly reported, standard conversion methods were applied. Outcomes reported as medians and interquartile ranges were converted to means and SDs using the methods of Wan et al. [24]. When studies reported 95% confidence intervals around a mean without SDs, SDs were calculated algebraically from the confidence interval limits according to the Cochrane Handbook [25]. For change-from-baseline outcomes in which the SD of change was not reported, the SD was imputed using baseline and final SDs with a conservative correlation coefficient of 0.5, following Follmann et al. [26]. When two eligible intervention arms required combination into a single OTC/self-fit group, pooled sample sizes, means, and SDs were calculated using standard combined-group formulae.
2.6. Risk of Bias and Certainty of Evidence
The risk of bias for each included randomized trial was assessed using the revised Cochrane Risk of Bias 2 (RoB-2) tool [27]. The assessment considered bias arising from the randomization process, deviations from intended interventions, missing outcome data, measurement of the outcome, and selection of the reported result. For crossover trials, additional considerations relevant to period, carryover, and allocation sequence were considered where applicable. Disagreements were resolved through discussion and consensus.
The certainty of evidence for each major outcome was evaluated using the Grading of Recommendations Assessment, Development and Evaluation (GRADE) framework [28,29]. The certainty of evidence was rated across the domains of risk of bias, inconsistency, indirectness, imprecision, and publication bias. All judgments were justified and documented.
2.7. Meta-Analysis and Data Synthesis
All statistical analyses were performed using Stata/SE software, version 19 (StataCorp, College Station, TX, USA). Dichotomous outcomes were analyzed using risk ratios (RRs), while continuous outcomes were analyzed using standardized mean differences (SMDs) as the included studies used different scales; all effect estimates were reported with 95% confidence intervals (CIs). Because the included studies differed in hearing-aid technology, fitting algorithms, degree of professional involvement, and follow-up duration, clinical heterogeneity was anticipated a priori. Random-effects meta-analysis was therefore selected as the primary approach. Leave-one-out sensitivity analyses and Galbraith plots were used to explore the influence of individual studies on statistical heterogeneity; these analyses were interpreted as exploratory and were not used to justify exclusion of studies solely on the basis of statistical significance. For dichotomous outcomes with zero events in one arm, a continuity correction of 0.5 was applied. Statistical heterogeneity was assessed using the Chi-squared test and quantified using the I2 statistic. Significant heterogeneity was defined as a chi-squared p-value < 0.01 or an I2 value > 50%. In case of significant heterogeneity, a leave-one-out sensitivity analysis was conducted to explore the source of heterogeneity. Publication bias was not assessed because all meta-analyzed outcomes included fewer than 10 trials [30].
For crossover trials, outcome data were extracted according to the reporting format of the original studies. Because participant-level paired data and within-participant correlation coefficients were not available in the published reports, the available condition-level summary statistics were used in the meta-analysis without introducing an assumed correlation coefficient. This limitation was considered when interpreting the certainty of the evidence. Given the small number of eligible RCTs, crossover trials were retained in the primary analysis rather than excluded in a separate sensitivity analysis, as excluding them would further reduce the already limited evidence base.
3. Results
3.1. Summary of the Systematic Review Search
A total of 1484 records were retrieved from the initial literature search across five databases. Before screening, 1185 references were removed, including 821 duplicates identified by Covidence and 364 records marked as ineligible by Covidence automation tools. The remaining 299 records were screened, of which 277 were excluded after failing to meet the inclusion criteria. Twenty-two full-text reports were assessed for eligibility, and 14 were excluded for various reasons (Table S2). Finally, six trials reported across seven reports were included in the quantitative and qualitative synthesis [16,17,18,19,20,21,22], while one additional trial was included in the qualitative synthesis only [15] (Figure 1). Multiple publications arising from the same randomized cohort were treated as a single trial for purposes of participant counting and meta-analysis; specifically, De Sousa et al. 2023 [20] and De Sousa et al. 2024 [19] represent reports from the same randomized cohort and were not counted as independent trials.
Figure 1.

PRISMA flow chart of the study screening and selection process.
3.2. Summary of Study Characteristics and Baseline Participant
Seven RCTs reported across eight articles, involving 1092 participants, were included in this review [15,16,17,18,19,20,21,22]. All were conducted in the United States [15,16,17,18,21,22], except for one trial from South Africa [19,20]. All trials enrolled adults with predominantly mild-to-moderate hearing loss. Wu et al. randomized 295 participants, of whom 245 completed follow-up. After excluding the OTC+ arms and combining technology levels within the OTC and AUD groups, 165 participants contributed to the meta-analysis (OTC, n = 82; AUD, n = 83). A key characteristic of the included evidence was variation in the OTC or self-fitting delivery model, ranging from commercially available self-fitting OTC devices and app-based fitting algorithms to simulated OTC models based on preset selection or consumer-directed fitting. Comparator arms generally involved audiologist-fitted or conventional best-practice care, commonly incorporating audiometric thresholds, NAL-NL2 targets, real-ear verification, counseling, and follow-up support. Follow-up duration ranged from short field-use periods of 2 weeks to longer assessments extending up to 8 months. Detailed study designs and baseline participant characteristics are summarized in Table 1 and Table 2, respectively. Finally, detailed descriptions of self-fitting and best-practice-fitting strategies across the included trials are provided in Tables S3 and S4, respectively.
Table 1.
Summary characteristics of the included RCTs.
| Study ID | Study Design | Country | Total Participants | Hearing Loss Definition/PTA | Prior Hearing-Aid Experience | Main Inclusion Criteria | Follow-Up Duration | Attrition/Loss to Follow-Up | Primary Outcome |
|---|---|---|---|---|---|---|---|---|---|
| Baltzell et al. 2025 [21] | Within-subject crossover RCT | USA | 21 | Four-frequency pure-tone average (PTA) at 05, 1, 2, and 4 kHz; at least one threshold > 20 dB HL with limits at specific frequencies | 17/21 new users; 4/21 experienced users | Adults ≥ 18 years old, able to understand English, with mild-to-moderate hearing loss and normal middle ear function | 2-week field-use period for each intervention condition | 2 of the 23 eligible participants were lost to follow-up/did not complete | Self-reported benefit (APHAB) and objective speech-in-noise benefit (QuickSIN) |
| De Sousa et al. 2023 [20] | Parallel group RCT | South Africa | 64 | PTA at 05, 10, 20, and 40 kHz; excluded if thresholds < 20 dB HL at all frequencies or >80 dB HL at two or more frequencies | Mixed: AF 6/32 experienced; SF 11/32 experienced; remainder new users | Adults ≥ 18 years old with self-reported mild to moderate hearing loss and no history of outer/middle ear disease in the past 90 days | 2 weeks and 6 weeks post-fitting | 4 of the 68 randomized participants were excluded or lost to follow-up | Self-reported hearing aid benefit measured via the APHAB questionnaire |
| De Sousa et al. 2024 [19] | Long-term extension of De Sousa 2023 RCT | South Africa | 44 | PTA at 05, 10, 20, and 40 kHz, consistent with the original trial’s audiometric criteria | Mixed: AF 26 new/6 experienced; SF 21 new/10 experienced | Participants from the original 6-week non-inferiority trial who were not lost to follow-up | Approximately 8 months post-fitting | 20 of the original 64 participants were lost to long-term follow-up due to nonresponse | Self-reported hearing aid benefit measured using the APHAB and the IOI-HA questionnaires |
| Humes et al. 2017 [18] | Parallel group RCT | USA | 154 | PTA (500, 1000, 2000 Hz) averaged ~26–28 dB HL; high-frequency PTA (1000, 2000, 4000 Hz) averaged ~38–39 dB HL | Required no prior hearing-aid experience | Adults 55–79 years old, native English speakers, MMSE-2 SV > 25, no prior HA experience, with symmetrical age-related hearing loss | 6-week trial (Session 3) followed by an optional 4-week trial for CD and placebo groups | 9 participants out of 163 withdrew after their initial session (154 analyzed) | Aided benefit measured by the 66-item Profile of Hearing Aid Benefit (PHAB) |
| Humes et al. 2025 [17] | Parallel group RCT | USA | 584 | Better-ear 4-frequency PTA (500, 1000, 2000, 4000 Hz) ≤ 50 dB HL | Required never used/tried hearing aids | Adults 50–79 years, no HA experience, English speakers, HHIE score > 4, MoCA score ≥ 23, able to pay $650 | 6 weeks and 6 months post-fitting | 119 participants out of 584 withdrew or were lost to follow-up by the 6-month mark | Global score from the Profile of Hearing Aid Benefit (PHAB) |
| Wu et al. 2025 [16] | Two-site randomized factorial trial: 3 service models × 2 technology levels | USA | 165 | PTA (05, 1, 2, and 4 kHz) between 25 and 65 dB HL | Required no previous hearing-aid experience | Adults aged 55 to 85 years with bilateral sensorineural hearing loss, native English speakers, and no previous HA experience | 6 to 7 weeks post-fitting | 50 participants withdrew out of 295 randomized trials (245 completed) | Glasgow Hearing Aid Benefit Profile (GHABP) administered using ecological momentary assessment (EMA) |
| Yellamsetty and Lewis 2024 [15] | Two-site repeated-measures randomized crossover clinical trial | USA | 40 | WHO classification for 500–4000 Hz: slight (<25 dB HL), mild (26–40 dB HL), and moderate (41–55 dB HL) | Mostly experienced: 36/40 experienced users; 4/40 new users | Adults aged 18 to 80 with mild-to-moderate HL, computer/English proficiency, no significant neurological deficits, and with a preference for ≥1 year HA experience | 2 weeks of field use per condition (4 weeks total) | 0 reported lost to follow-up (all 40 participants evaluated) | Efficacy assessed via objective (REM), behavioral (QuickSIN), and subjective (APHAB) measures |
| Zheng et al. 2025 [22] | Parallel group RCT | USA | 64 | Air-conduction threshold between 25- and 60 dB HL with an air-bone gap < 15 dB; PTA assessed for 05, 1, 2, and 4 kHz | Mixed: pro-fit 11 nonusers/21 HA users; self-fit 12 nonusers/20 HA users | Native-English speakers ≥ 18 years old with mild-to-moderate sensorineural hearing loss and normal cognitive and middle ear functions | 1 month | 7 participants out of 71 randomized trials discontinued or were excluded | Aided benefit evaluated by the QuickSIN test, APHAB, SSQ12, and objective REAG/O measures |
AF, audiologist fitting; APHAB, Abbreviated Profile of Hearing Aid Benefit; CD, consumer decides; dB, decibel; EMA, ecological momentary assessment; GHABP, Glasgow Hearing Aid Benefit Profile; HA, hearing aid; HHIE, Hearing Handicap Inventory for the Elderly; HL, hearing level; IOI-HA, International Outcome Inventory for Hearing Aids; kHz, kilohertz; MMSE-2 SV, Mini-Mental State Examination, Second Edition, Standard Version; MoCA, Montreal Cognitive Assessment; PHAB, Profile of Hearing Aid Benefit; PTA, pure-tone average; QuickSIN, Quick Speech-in-Noise test; RCT, randomized controlled trial; REAG/O, real-ear aided gain/output; REM, real-ear measurement; SF, self-fitted; SSQ12, Speech, Spatial and Qualities of Hearing Scale-12; USA, United States of America; WHO, World Health Organization.
Table 2.
Baseline characteristics of the participants.
| Study ID | Number of Patients in Each Group | Age (Years), Mean (SD) | Gender (Male), N. (%) | Prior Hearing-Aid Use, N. (%) | Hearing-Aid Naïve, N. (%) | PTA, Mean (SD) | ||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Self-Fitting | Best Practice-Fitting | Self-Fitting | Best Practice-Fitting | Self-Fitting | Best Practice-Fitting | Self-Fitting | Best Practice-Fitting | Self-Fitting | Best Practice-Fitting | Self-Fitting | Best Practice-Fitting | |
| Baltzell et al. 2025 [21] | 21 | 66.2 (13.4) | 13 (61.9%) | 4 (19.1%) | 17 (80.9%) | 37.0 (10.1), PTA 0.5/1/2/4 kHz | ||||||
| De Sousa et al. 2023 [20] | 32 | 32 | 62.0 (13.1) | 65.3 (14.9) | 15 (46.9%) | 18 (56.3%) | 11 (34.4%) | 6 (18.8%) | 21 (65.6%) | 26 (81.3%) | 38.4 (11.7), PTA 0.5/1/2/4 kHz | 41.8 (13.6), PTA 0.5/1/2/4 kHz |
| De Sousa et al. 2024 [19] | 23 | 21 | NR | NR | NR | NR | 10/32 (31.3%), parent baseline | 6/32 (18.8%), parent baseline | 21/32 (65.6%), parent baseline | 26/32 (81.3%), parent baseline | 38.4 (11.7), parent baseline PTA 0.5/1/2/4 kHz | 41.8 (13.6), parent baseline PTA 0.5/1/2/4 kHz |
| Humes et al. 2017 [18] | 51 | 53 | 68.0 (6.2) | 69.9 (5.6) | 27 (49.1%) | 28 (52.8%) | 0 | 0 | 55/55 (100%) | 53/53 (100%) | 28.9 (8.6), bilateral PTA | 28.7 (7.7), bilateral PTA |
| Humes et al. 2025 [17] | 394 | 190 | 69.4 (6.4) | 68.4 (6.9) | NR arm-specific; across fitting groups, males ranged 53.9–58.4% | 0 | 0 | 394 (100%) | 190 (100%) | Better-ear PTA4 32.0 (9.2), calculated; worse-ear PTA4 36.7 (9.7), calculated | Better-ear PTA4 32.1 (9.0); worse-ear PTA4 36.7 (9.6) | |
| Wu et al. 2025 [16] | 82 | 83 | 68.5 (7.1) | 66.9 (6.6) | 39 (47.6%) | 42 (50.6%) | 0 | 0 | 82 (100%) | 83 (100%) | PTA 0.5/1/2/4 kHz: 35.6 (5.7), calculated | PTA 0.5/1/2/4 kHz: 36.5 (7.1), calculated |
| Yellamsetty and Lewis 2024 [15] | 40 | 40 | 59.33 (12.3) | 59.33 (12.3) | 21 (52.5%) | 21 (52.5%) | 36 (90.0%) | 36 (90.0%) | 4 (10.0%) | 4 (10.0%) | 41.45 (10.89), 4-frequency PTA per ear | 41.45 (10.89), 4-frequency PTA per ear |
| Zheng et al. 2025 [22] | 32 | 32 | 58.0 (13.7) | 62.0 (11.2) | 15 (46.9%) | 13 (40.6%) | 20 (62.5%) | 21 (65.6%) | 12 (37.5%) | 11 (34.4%) | PTA 0.5/1/2/4 kHz: 32.1 (9.0) | PTA 0.5/1/2/4 kHz: 35.1 (8.7) |
kHz, kilohertz; N, number; NR, not reported; PTA, pure-tone average; PTA4, four-frequency pure-tone average; SD, standard deviation.
3.3. Risk of Bias and Certainty of Evidence
For parallel-group RCTs, two trials were judged to have a low overall risk of bias [17,20], while another two raised some concerns [16,18], and Zheng et al. showed a high overall risk of bias [22] (Figure 2A). The main drivers of concern ratings were bias due to missing outcome data and bias in the measurement of outcomes. For randomized crossover trials, Yellamsetty and Lewis [15] showed a low overall risk of bias, whereas Baltzell et al. [21] raised some concerns (Figure 2B). Additionally, the detailed GRADE certainty of evidence assessment is outlined in Table 3.
Figure 2.

Quality assessment of risk of bias in the included randomized trials. Panel (A) presents the risk-of-bias judgments for parallel-group randomized trials using the RoB 2 tool [16,17,18,20,22], and Panel (B) presents the risk-of-bias judgments for randomized crossover trials [15,21]. Green indicates low risk, yellow indicates some concerns, and red indicates high risk of bias.
Table 3.
GRADE evidence profile.
| Outcomes | Participants (Studies) | Risk of Bias | Inconsistency | Indirectness | Imprecision | Publication Bias | Overall Certainty of Evidence | Effect Estimates |
|---|---|---|---|---|---|---|---|---|
| Patient-reported hearing benefit | 1002 (6 RCTs) | serious a | very serious b | not serious | serious d | not assessed f | ⊕◯◯◯ Very low | SMD 0.01 higher (0.44 lower to 0.46 higher) |
| Speech-in-noise/speech recognition | 354 (4 RCTs) | serious a | serious c | not serious | serious d | not assessed f | ⊕◯◯◯ Very low | SMD 0.11 higher (0.25 lower to 0.47 higher) |
| Hearing loss | 734 (3 RCTs) | serious a | serious c | not serious | serious d | not assessed f | ⊕◯◯◯ Very low | SMD 0.22 lower (0.50 lower to 0.06 higher) |
| Hearing-aid satisfaction | 333 (3 RCTs) | serious a | very serious b | not serious | serious d | not assessed f | ⊕◯◯◯ Very low | SMD 0.33 lower (0.96 lower to 0.30 higher) |
| Device use/adherence | 700 (3 RCTs) | serious a | very serious b | not serious | serious d | not assessed f | ⊕◯◯◯ Very low | SMD 0.12 lower (0.62 lower to 0.39 higher) |
| Need for follow-up adjustment | 333 (3 RCTs) | serious a | very serious b | not serious | serious d | not assessed f | ⊕◯◯◯ Very low | RR 0.46 (0.12 to 1.71) |
| Any adverse event | 955 (4 RCTs) | serious a | not serious | not serious | very serious e | not assessed f | ⊕◯◯◯ Very low | RR 0.82 (0.21 to 3.17) |
RCT: randomized controlled trial; RR: risk ratio; SMD: standardized mean difference. For crossover trials, participants were counted once in the participants column. Explanations: a. Downgraded for risk of bias because several contributing trials had some concerns or a high overall risk of bias, mainly related to missing outcome data, outcome measurement, deviations from intended interventions, or selection of the reported result. b. Downgraded two levels for very substantial heterogeneity (I2 > 75%) with variability in the direction and magnitude of study effects. c. Downgraded for substantial heterogeneity (I2 between 50% and 75%) with clinically and statistically variable study effects. d. Downgraded for imprecision because the confidence interval crossed the null and included a potentially important benefit and/or harm. e. Downgraded two levels for imprecision because events were rare and the confidence interval was very wide, including both appreciable benefit and harm. f. Publication bias was not assessed because fewer than 10 studies contributed to each outcome.
3.4. Primary Outcome: Patient-Reported Hearing Benefit
There was no significant difference between the two groups (SMD: 0.01, 95% CI [−0.44, 0.46], p = 0.97, I2 = 89.2%) (Figure 3A). Leave-one-out sensitivity analysis showed that the pooled effect remained statistically non-significant after omitting each study (Figure S1). However, the Galbraith plot suggested important between-study inconsistency, with Baltzell et al. and Wu et al. appearing as potential outliers (Figure S2).
Figure 3.

Forest plots comparing self-fitting and best practice-fitting for primary and secondary effectiveness outcomes. The forest plots display standardized mean differences for (A) patient-reported hearing benefit, (B) speech-in-noise/speech recognition, and (C) hearing loss [16,17,18,20,21,22].
4. Secondary Outcomes
4.1. Speech-in-Noise/Speech Recognition
There was no significant difference between the two groups (SMD: 0.11, 95% CI [−0.25, 0.47], p = 0.55, I2 = 64.34%) (Figure 3B). Leave-one-out sensitivity analysis showed that the pooled effect remained statistically non-significant after omitting each study (Figure S3). The Galbraith plot did not identify any clear outliers, although the individual effects varied in direction, which explains the moderate heterogeneity (Figure S4).
4.2. Hearing Loss
There was no significant difference between the two groups (SMD: −0.22, 95% CI [−0.50, 0.06], p = 0.12, I2 = 64.24%) (Figure 3C). Sensitivity analysis suggested that this result was partly influenced by Humes et al. 2025 [17], as after omitting this study, the pooled estimate became statistically significant (SMD: −0.38, 95% CI [−0.62, −0.14], p = 0.002), favoring self-fitting (Figure S5). The Galbraith plot did not show a clear outlier (Figure S6).
4.3. Hearing-Aid Satisfaction
There was no significant difference between the two groups (SMD: −0.33, 95% CI [−0.96, 0.30], p = 0.31, I2 = 87.29%) (Figure 4A). Leave-one-out sensitivity analysis indicated that this finding was sensitive to the exclusion of De Sousa et al. 2023 [20], as omitting this study made the pooled estimate statistically significant (SMD: −0.65, 95% CI [−0.89, −0.40], p < 0.001), favoring best-practice fitting (Figure S7). The Galbraith plot contributed to this instability, showing De Sousa et al. 2023 [20] and Wu et al. 2025 [16] as the main contributors to heterogeneity (Figure S8).
Figure 4.

Forest plots comparing self-fitting and best practice-fitting for use-related, patient-reported, and safety outcomes. The forest plots display standardized mean differences for (A) hearing-aid satisfaction and (B) device use/adherence, and risk ratios for (C) need for follow-up adjustment and (D) any adverse event [16,17,18,20].
4.4. Device Use/Adherence
There was no significant difference between the two groups (SMD: −0.12, 95% CI [−0.62, 0.39], p = 0.65, I2 = 84.70%) (Figure 4B). Sensitivity analysis showed that the pooled estimate became statistically significant after omitting De Sousa et al. 2023 [20] (SMD: −0.40, 95% CI [−0.56, −0.23], p < 0.001), favoring best-practice fitting (Figure S9). The Galbraith plot identified De Sousa et al. 2023 [20] and Humes et al. 2025 [17] as the principal contributors to heterogeneity (Figure S10).
4.5. Need for Follow-Up Adjustment
There was no significant difference between the two groups (RR: 0.46, 95% CI [0.12, 1.71], p = 0.25, I2 = 83.59%) (Figure 4C). Leave-one-out sensitivity analysis showed that the result remained statistically non-significant after omitting each study (Figure S11). The Galbraith plot suggested that De Sousa et al. 2023 [20] was the main contributor to heterogeneity (Figure S12).
4.6. Any Adverse Event
There was no statistically significant difference between the self-fitting and best-practice fitting groups in the incidence of any adverse event (RR: 0.82, 95% CI [0.21, 3.17], p = 0.78, I2 = 0%). However, the evidence was sparse and imprecise, and the wide confidence interval does not exclude clinically important differences in either direction (Figure 4D).
5. Discussion
Across the seven included RCTs and 1092 patients, there was no statistically significant difference between self-fitting and best-practice fitting for patient-reported hearing benefit, speech-in-noise/speech-recognition performance, hearing loss, hearing-aid satisfaction, device use/adherence, need for follow-up adjustment, or adverse events. These findings may suggest that, under trial conditions and in selected adult patients, self-fitting models can achieve outcomes similar to those of professional-fitting models. However, the certainty of the evidence was very low across outcomes due to risk-of-bias concerns, considerable heterogeneity, imprecision, and the small number of trials included.
The absence of statistically significant differences across several outcomes should not be interpreted as evidence that professional fitting is unnecessary or that self-fitting is equivalent to professional fitting. Given the very low certainty of evidence and the limitations of the available trials, the findings only indicate that a clear advantage of professional fitting was not demonstrated in the populations and conditions studied. This is consistent with the rationale for the OTC hearing-aid introduction: to improve access and affordability by allowing adults with perceived mild-to-moderate hearing loss to obtain hearing aids without a mandatory medical examination, prescription, or audiologist-fitting adjustment [31,32]. Also, the FDA has recently granted authorization for an OTC hearing aid software that enables personalized amplification based on the user’s hearing profile, which contributes to the ongoing regulatory and technological transition toward the independent or semi-independent fitting of hearing devices [33].
Within the included evidence, De Sousa et al. showed comparable short-term outcomes between a self-fitting OTC device and audiologist fitting [20]. Baltzell et al. found that a self-fitted OTC intervention was not inferior to a clinician-fitted intervention for self-reported benefit and speech-in-noise outcomes [21], and Zheng et al. reported no clear patient-reported advantage of professional fitting over self-fitting for an OTC device [22]. Related self-fitting and OTC studies also suggest that autonomous fitting methods can provide a clinically meaningful benefit when device design, fitting algorithms, and user guidance are adequate [34]. Thus, these findings raise the possibility that optimized self-fitting strategies could provide an alternative to routine professional fitting for a subset of appropriately selected users. However, given the very low certainty of evidence, this possibility should be considered hypothesis-generating rather than evidence of established clinical equivalence. In that model, audiologists remain essential. Still, their role shifts from an obligatory first-line fitting for all users toward diagnostic triage, counseling, troubleshooting, and escalation for users who fail or are unsuitable for self-fitting.
Given the small number of eligible trials and the limited number of studies contributing to several outcomes, further subgroup analyses by fitting model, device technology, or follow-up duration were not considered sufficiently reliable or informative. Accordingly, the pooled estimates should be interpreted in the context of substantial clinical and methodological heterogeneity, and the findings should not be generalized across all self-fitting technologies or service-delivery models. Baltzell et al. evaluated an app-guided self-fitting device in a crossover design [21]. De Sousa et al. tested the Lexie Lumen self-fitting hearing aid against an audiologist-fitted version of the same device, with long-term outcomes reported separately [19,20]. Humes et al. evaluated a consumer-decides model and later compared self-fit/person-fit methods with audiologist best-practice fitting in a large multisite trial [17,18]. Wu et al. compared different service models and technology levels [16], while Zheng et al. compared self-fit and professional-fit modes of an OTC device [22]. Therefore, this may indicate that, for some adults with uncomplicated mild-to-moderate hearing loss, well-designed self-fitting pathways can approximate the patient-perceived benefit of professional fitting. However, the high heterogeneity suggests that this replacement potential is likely affected by the self-fitting strategy, and an optimal method remains unknown.
Moreover, the lack of a significant difference in speech-in-noise or speech-recognition outcomes also supports the possibility that professional fitting may not always be required to achieve routine functional benefit. Speech understanding in noise is one of the most clinically relevant problems for adults with sensorineural hearing loss and is not fully captured by simple audibility measures [35,36]. Although professional fitting would be expected to improve target matching through audiometric programming, real-ear verification, fine-tuning, and counseling, this did not translate into a consistent pooled advantage. These findings suggest that a clear pooled advantage of professional fitting was not demonstrated for the speech outcomes assessed; however, they do not establish that automated or user-guided fitting provides equivalent speech-in-noise or speech-recognition benefits. However, laboratory speech tests may not fully reflect real-world listening conditions, where competing talkers, fatigue, cognitive load, and changing environments can affect communication success [35,37]. Future trials should therefore test whether optimized self-fitting strategies remain comparable to professional fitting using real-world speech understanding, listening effort, ecological momentary assessment, and daily-use data.
Importantly, the applicability of these findings to routine real-world hearing-aid use remains uncertain. Most included trials evaluated relatively short-term outcomes in selected adults with uncomplicated mild-to-moderate hearing loss and under trial conditions that may not fully represent the complexity of everyday listening environments. Important real-world outcomes, including sustained long-term adherence, listening effort in complex acoustic environments, device management, troubleshooting, and help-seeking behavior, remain insufficiently studied for self-fitting pathways. Therefore, the absence of significant differences in the available trial-based outcomes should not be interpreted as evidence that self-fitting and professional fitting provide equivalent long-term real-world experiences. Future research should prioritize pragmatic trials embedded within routine clinical and community settings, with longer follow-up and ecologically relevant measures such as ecological momentary assessment, objectively measured daily device use, listening effort, device-management burden, and patterns of help-seeking and professional support.
Furthermore, findings on satisfaction and device use/adherence can support the idea that hearing outcomes are shaped by implementation rather than by professional involvement alone. Hearing-aid satisfaction depends on perceived sound quality, physical comfort, ease of use, expectations, cosmetic acceptance, cost, support availability, and perceived value [38,39]. Hence, the replacement potential of self-fitting devices depends less on simply removing audiologists from the pathway and more on whether self-fitting systems can reliably reproduce the components that drive real-world use. An optimized self-fitting pathway with clear instructions, adaptive fine-tuning, remote support, and easy return or escalation options may be more clinically relevant.
Supporting this, the need for follow-up adjustment also did not differ significantly. In professionally fitted pathways, follow-up may be actively scheduled or encouraged as part of best-practice care [40]. In contrast, in self-fitting pathways, adjustment may depend on whether users recognize a problem, know how to seek help, and have access to support. Thus, a lower rate of follow-up adjustment in a self-fitting arm may reflect fewer problems. Still, it may also reflect reduced access, lower expectations, or unmet needs that remain unrecognized. Conversely, a higher rate in a professional arm may reflect more proactive care rather than poorer initial fitting [40,41]. Accordingly, if self-fitting is to replace routine professional fitting for selected users, future studies should distinguish between patient-requested adjustment, clinician-recommended adjustment, successful troubleshooting, unmet adjustment needs, and device discontinuation, because these represent different aspects of service quality.
The temporal and technological evolution of hearing-aid devices also has implications for the generalizability of these findings. Newer self-fitting algorithms, smartphone-based interfaces, automated personalization, and remote-support models may perform differently from earlier approaches included in this review. Conversely, improvements in professional fitting technologies and service delivery may also influence the magnitude of the difference between fitting pathways. Thus, the absence of a pooled difference should not be interpreted as evidence that self-fitting and professional fitting are equivalent across all current or future technologies. Future trials should use contemporary devices, clearly characterize the fitting algorithm and level of professional support, and report outcomes in a manner that permits comparison across technological generations and service models.
This review has several strengths. It focused on RCTs, reducing confounding compared with observational evidence. It included both parallel-group and crossover RCTs, assessed commercially available self-fitting OTC devices, and examined a broad set of clinically relevant outcomes. Risk of bias was assessed using RoB 2; certainty was judged using the GRADE framework; and sensitivity analyses were used to explore the influence of individual studies.
However, several limitations should also be acknowledged. First, the number of included trials was small, and most were conducted in the United States, limiting generalizability across geography and health systems. Second, several outcomes included only three or four trials, and publication bias could not be reliably assessed. Third, follow-up was short in many studies, limiting conclusions about long-term device retention, sustained use, and late adverse events. Fourth, some trials experienced attrition, missing outcome data, or concerns about the measurement and reporting of outcomes. Fifth, the self-fitting interventions varied extensively, ranging from app-based in situ algorithms to preset selection and consumer-decides approaches. At the same time, professional-fitting strategies also differed. Sixth, the outcome assessment scores differed across the included trials, and the use of SMD allowed pooling across instruments but may have affected the interpretation of the results. An additional methodological limitation is that two included trials used a crossover design, but participant-level paired data and within-participant correlation estimates were not available from the published reports. Consequently, the available condition-level summary data were used, which may not fully account for the statistical dependence inherent to crossover designs. This limitation contributes to uncertainty in the corresponding pooled estimates and was considered in the overall assessment of evidence certainty. Seventh, the selected nature of trial participants and predominantly short follow-up periods limit the generalizability of the findings to routine real-world use. Participants enrolled in RCTs may be more motivated, digitally capable, or clinically suitable for self-fitting than users encountered in routine clinical or community settings, who may have greater variation in hearing-loss characteristics, digital skills, support needs, and ability to manage hearing aids independently. Therefore, the findings may be applicable primarily to selected uncomplicated users and should not be generalized to all adults with hearing loss. An additional limitation is variation in professional contact and follow-up support across trials. These implementation details were inconsistently reported, precluding reliable quantitative analysis. Thus, observed differences may partly reflect variation in service support rather than fitting strategy alone. A further limitation is the temporal and technological evolution of self-fitting and OTC hearing aids across studies. The small number of eligible RCTs and substantial variation in devices, fitting algorithms, and service models precluded reliable subgroup or meta-regression analyses by publication year or technology generation. Because publication year is also an imperfect proxy for technological advancement, the pooled findings should not be interpreted as demonstrating stable equivalence across technological eras. Finally, given these limitations, the current certainty of the evidence remains very low, warranting careful interpretation.
For appropriately selected adults with uncomplicated mild-to-moderate hearing loss, self-fitting hearing aids may have potential as an access-oriented fitting pathway. However, the current evidence is insufficient to establish equivalence or non-inferiority to professional fitting and therefore does not support replacing routine professional fitting as a general practice. Professional care remains important for users who fail or are unsuitable for self-fitting, prefer clinician support, have more severe hearing loss, or have complex audiologic needs.
Accordingly, future research should move beyond asking only whether self-fitting works and should instead identify which self-fitting strategy can safely and consistently replace routine professional fitting, and for which patient subgroup. Large pragmatic non-inferiority or equivalence RCTs are needed to compare contemporary commercially available OTC/self-fitting devices with clearly defined audiologist best-practice fitting. Follow-up should extend to at least 6–12 months because early benefit does not necessarily predict sustained adherence. Future trials should also stratify participants by age, prior hearing-aid experience, degree of hearing loss, digital literacy, socioeconomic status, and cognitive status. Finally, future research can test optimized replacement strategies, including fully autonomous self-fitting, OTC plus remote support, OTC plus optional real-ear verification, OTC plus audiology triage, and stepped referral models. Such designs can determine whether professional fitting is best reserved for selected cases rather than routinely required for all users.
6. Conclusions
The available randomized evidence did not demonstrate clear differences between self-fitting hearing aids and best-practice professional fitting for patient-reported benefit, speech outcomes, hearing loss, satisfaction, adherence, follow-up adjustment, or adverse events, although the evidence for adverse events was sparse and imprecise. However, the certainty of evidence was very low, and substantial heterogeneity, risk of bias, imprecision, and the small number of trials limit confidence in these findings. Therefore, the results should be considered hypothesis-generating rather than evidence that self-fitting can broadly replace professional fitting. Self-fitting may have potential as an accessible option for selected adults with uncomplicated mild-to-moderate hearing loss; however, larger, rigorous, and longer-term trials across diverse clinical and service-delivery settings are needed to establish its effectiveness, safety, and appropriate role relative to professional care.
Supplementary Materials
The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/audiolres16050139/s1. Table S1: Search strategy; Table S2: Excluded records in full-text screening; Table S3: Characteristics of the OTC/self-fitting interventions; Table S4: Characteristics of the audiologist-fitted/conventional-care comparators. Figure S1: Leave-one-out sensitivity analysis for patient-reported hearing benefit; Figure S2: Galbraith plot for patient-reported hearing benefit; Figure S3: Leave-one-out sensitivity analysis for speech-in-noise/speech recognition; Figure S4: Galbraith plot for speech-in-noise/speech recognition; Figure S5: Leave-one-out sensitivity analysis for hearing handicap; Figure S6: Galbraith plot for hearing handicap; Figure S7: Leave-one-out sensitivity analysis for hearing-aid satisfaction; Figure S8: Galbraith plot for hearing-aid satisfaction; Figure S9: Leave-one-out sensitivity analysis for device use/adherence. Figure S10: Galbraith plot for device use/adherence; Figure S11: Leave-one-out sensitivity analysis for need for follow-up adjustment; Figure S12: Galbraith plot for need for follow-up adjustment.
Author Contributions
E.A. contributed to study conception, study design, data collection, data analysis, writing the manuscript’s original draft, and reviewing the manuscript for editorial and intellectual content. H.A., N.A., Y.Y.A., A.M.A., M.Y.H.I. and D.A.A. contributed to the literature review, data collection, writing the manuscript’s original draft, and reviewing the manuscript for editorial and intellectual content. A.A.-Z. contributed to study design, data analysis, supervision, and reviewing the manuscript for editorial and intellectual content. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
Not applicable.
Informed Consent Statement
Not applicable.
Data Availability Statement
All data are available within the manuscript and can be obtained from the corresponding authors upon a reasonable request.
Conflicts of Interest
The authors declare no conflicts of interest.
Funding Statement
This research received no external funding.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Shukla A., Harper M., Pedersen E., Goman A., Suen J.J., Price C., Applebaum J., Hoyer M., Lin F.R., Reed N.S. Hearing Loss, Loneliness, and Social Isolation: A Systematic Review. Otolaryngol.—Head Neck Surg. 2020;162:622–633. doi: 10.1177/0194599820910377. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Podury A., Jiam N.T., Kim M., Donnenfield J.I., Dhand A. Hearing and Sociality: The Implications of Hearing Loss on Social Life. Front. Neurosci. 2023;17:1245434. doi: 10.3389/fnins.2023.1245434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Amieva H., Ouvrard C. Does Treating Hearing Loss in Older Adults Improve Cognitive Outcomes? A Review. J. Clin. Med. 2020;9:805. doi: 10.3390/jcm9030805. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Livingston G., Huntley J., Sommerlad A., Ames D., Ballard C., Banerjee S., Brayne C., Burns A., Cohen-Mansfield J., Cooper C., et al. Dementia Prevention, Intervention, and Care: 2020 Report of the Lancet Commission. Lancet. 2020;396:413–446. doi: 10.1016/S0140-6736(20)30367-6. Erratum in Lancet 2023, 42, 1132. https://doi.org/10.1016/S0140-6736(23)02043-3 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Ferguson M.A., Kitterick P.T., Chong L.Y., Edmondson-Jones M., Barker F., Hoare D.J. Hearing Aids for Mild to Moderate Hearing Loss in Adults. Cochrane Database Syst. Rev. 2017;2017:CD012023. doi: 10.1002/14651858.CD012023.pub2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.He P., Wen X., Hu X., Gong R., Luo Y., Guo C., Chen G., Zheng X. Hearing Aid Acquisition in Chinese Older Adults with Hearing Loss. Am. J. Public Health. 2018;108:241–247. doi: 10.2105/AJPH.2017.304165. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Chien W., Lin F.R. Prevalence of Hearing Aid Use among Older Adults in the United States. Arch. Intern. Med. 2012;172:292–293. doi: 10.1001/archinternmed.2011.1408. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Anderson M.C., Arehart K.H., Souza P.E. Survey of Current Practice in the Fitting and Fine-Tuning of Common Signal-Processing Features in Hearing AIDS for Adults. J. Am. Acad. Audiol. 2018;29:118–124. doi: 10.3766/jaaa.16107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Dillard L.K., Der C.M., Laplante-Lévesque A., De Wet S., Thorne P.R., McPherson B., de Andrade V., Newall J., Ramos H.D., Kaspar A., et al. Service Delivery Approaches Related to Hearing Aids in Low- and Middle-Income Countries or Resource-Limited Settings: A Systematic Scoping Review. PLoS Glob. Public Health. 2024;4:e0002823. doi: 10.1371/journal.pgph.0002823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Blustein J., Weinstein B.E., Chodosh J. Over-the-Counter Hearing Aids: What Will It Mean for Older Americans? J. Am. Geriatr. Soc. 2022;70:2115–2120. doi: 10.1111/jgs.17781. [DOI] [PubMed] [Google Scholar]
- 11.Sabin A.T., Van Tasell D.J., Rabinowitz B., Dhar S. Validation of a Self-Fitting Method for Over-the-Counter Hearing Aids. Trends Hear. 2020;24:2331216519900589. doi: 10.1177/2331216519900589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Keidser G., Convery E. Self-Fitting Hearing Aids: Status Quo and Future Predictions. Trends Hear. 2016;20:2331216516643284. doi: 10.1177/2331216516643284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Convery E., Keidser G., Hickson L., Meyer C. Factors Associated with Successful Setup of a Self-Fitting Hearing Aid and the Need for Personalized Support. Ear Hear. 2019;40:794–804. doi: 10.1097/AUD.0000000000000663. [DOI] [PubMed] [Google Scholar]
- 14.Irace A.L., Sharma R.K., Reed N.S., Golub J.S. Smartphone-Based Applications to Detect Hearing Loss: A Review of Current Technology. J. Am. Geriatr. Soc. 2021;69:307–316. doi: 10.1111/jgs.16985. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Yellamsetty A., Lewis R.M. Evaluation of Outcomes in a Clinical Trial: Comparing Self-Fit Hearing Aids and Hearing Aids Fit with Best Practices. Front. Audiol. Otol. 2024;2:1397604. doi: 10.3389/fauot.2024.1397604. [DOI] [Google Scholar]
- 16.Wu Y.H., Stangl E., Branscome K., Oleson J., Ricketts T. Hearing Aid Service Models, Technology, and Patient Outcomes: A Randomized Clinical Trial. JAMA Otolaryngol.—Head Neck Surg. 2025;151:684–692. doi: 10.1001/jamaoto.2025.1008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Humes L.E., Dhar S., Meskan M., Pitman A., Singh J. A Multisite Randomized Controlled Trial Comparing the Effectiveness of Two Self-Fit Methods to the Best-Practices Method of Hearing Aid Fitting. J. Speech Lang. Hear. Res. 2025;68:2080–2103. doi: 10.1044/2024_JSLHR-24-00423. [DOI] [PubMed] [Google Scholar]
- 18.Humes L.E., Rogers S.E., Quigley T.M., Main A.K., Kinney D.L., Herring C. The Effects of Service-Delivery Model and Purchase Price on Hearing-Aid Outcomes in Older Adults: A Randomized Double-Blind Placebo-Controlled Clinical Trial. Am. J. Audiol. 2017;26:53–79. doi: 10.1044/2017_AJA-16-0111. Erratum in Am. J. Audiol. 2019, 28, 730. https://doi.org/10.1044/2019_AJA-19-0 . [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.De Sousa K.C., Manchaiah V., Moore D.R., Graham M.A., Swanepoel D.W. Long-Term Outcomes of Self-Fit vs. Audiologist-Fit Hearing Aids. JAMA Otolaryngol.—Head Neck Surg. 2024;150:765–771. doi: 10.1001/jamaoto.2024.1825. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.De Sousa K.C., Manchaiah V., Moore D.R., Graham M.A., Swanepoel D.W. Effectiveness of an Over-the-Counter Self-Fitting Hearing Aid Compared with an Audiologist-Fitted Hearing Aid: A Randomized Clinical Trial. JAMA Otolaryngol.—Head Neck Surg. 2023;149:522–530. doi: 10.1001/jamaoto.2023.0376. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Baltzell L.S., Kokkinakis K., Li A., Yellamsetty A., Teece K., Nelson P.B. Validation of a Self-Fitting Over-the-Counter Hearing Aid Intervention Compared with a Clinician-Fitted Hearing Aid Intervention: A Within-Subjects Crossover Design Using the Same Device. Trends Hear. 2025;29:23312165251328055. doi: 10.1177/23312165251328055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Zheng Y., Sepanski J., Li L. Self-Fit vs. Professional-Fit Over-the-Counter Hearing Aids: A Randomized Clinical Trial. J. Am. Acad. Audiol. 2026;37:54–71. doi: 10.3766/jaaa.250047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Page M.J., McKenzie J.E., Bossuyt P.M., Boutron I., Hoffmann T.C., Mulrow C.D., Shamseer L., Tetzlaff J.M., Akl E.A., Brennan S.E., et al. The PRISMA 2020 Statement: An Updated Guideline for Reporting Systematic Reviews. Syst. Rev. 2021;10:89. doi: 10.1186/s13643-021-01626-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Wan X., Wang W., Liu J., Tong T. Estimating the Sample Mean and Standard Deviation from the Sample Size, Median, Range and/or Interquartile Range. BMC Med. Res. Methodol. 2014;14:135. doi: 10.1186/1471-2288-14-135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Higgins J.P.T., Li T., Deeks J.J. Cochrane Handbook for Systematic Reviews of Interventions. 2nd ed. Wiley; Hoboken, NJ, USA: 2019. Choosing Effect Measures and Computing Estimates of Effect; pp. 143–176. [DOI] [Google Scholar]
- 26.Follmann D., Elliott P., Suh I., Cutler J. Variance Imputation for Overviews of Clinical Trials with Continuous Response. J. Clin. Epidemiol. 1992;45:769–773. doi: 10.1016/0895-4356(92)90054-Q. [DOI] [PubMed] [Google Scholar]
- 27.Sterne J.A.C., Savović J., Page M.J., Elbers R.G., Blencowe N.S., Boutron I., Cates C.J., Cheng H.-Y., Corbett M.S., Eldridge S.M., et al. RoB 2: A Revised Tool for Assessing Risk of Bias in Randomised Trials. BMJ. 2019;366:l4898. doi: 10.1136/BMJ.L4898. [DOI] [PubMed] [Google Scholar]
- 28.Guyatt G.H., Oxman A.D., Kunz R., Vist G.E., Falck-Ytter Y., Schünemann H.J. Rating Quality of Evidence and Strength of Recommendations: What Is “Quality of Evidence” and Why Is It Important to Clinicians? BMJ. 2008;336:995. doi: 10.1136/BMJ.39490.551019.BE. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Guyatt G.H., Oxman A.D., Vist G.E., Kunz R., Falck-Ytter Y., Alonso-Coello P., Schünemann H.J. Rating Quality of Evidence and Strength of Recommendations: GRADE: An Emerging Consensus on Rating Quality of Evidence and Strength of Recommendations. BMJ Br. Med. J. 2008;336:924. doi: 10.1136/BMJ.39489.470347.AD. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Lin L., Chu H. Quantifying Publication Bias in Meta-Analysis. Biometrics. 2018;74:785–794. doi: 10.1111/biom.12817. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Food and Drug Administration, Department of Health and Human Services(HHS) Medical Devices; Ear, Nose, and Throat Devices; Establishing Over-the-Counter Hearing Aids. Volume 87. Food and Drug Administration; Silver Spring, MD, USA: 2022. pp. 50698–50762. [Google Scholar]
- 32.Blazer D.G., Domnitz S., Liverman C.T. Hearing Health Care for Adults: Priorities for Improving Access and Affordability. The National Academies Press; Washington, DC, USA: 2016. pp. 1–324. [DOI] [PubMed] [Google Scholar]
- 33.Pflaum C. FDA Authorizes First Over-the-Counter Hearing Aid Software. US Food and Drug Administration; Silver Spring, MD, USA: 2024. [Google Scholar]
- 34.Keidser G., Convery E. Outcomes with a Self-Fitting Hearing Aid. Trends Hear. 2018;22:2331216518768958. doi: 10.1177/2331216518768958. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Pichora-Fuller M.K., Kramer S.E., Eckert M.A., Edwards B., Hornsby B.W.Y., Humes L.E., Lemke U., Lunner T., Matthen M., Mackersie C.L., et al. Hearing Impairment and Cognitive Energy: The Framework for Understanding Effortful Listening (FUEL) Ear Hear. 2016;37:5S–27S. doi: 10.1097/AUD.0000000000000312. [DOI] [PubMed] [Google Scholar]
- 36.Killion M.C., Niquette P.A., Gudmundsen G.I., Revit L.J., Banerjee S. Development of a Quick Speech-in-Noise Test for Measuring Signal-to-Noise Ratio Loss in Normal-Hearing and Hearing-Impaired Listeners. J. Acoust. Soc. Am. 2004;116:2395–2405. doi: 10.1121/1.1784440. Erratum in J. Acoust. Soc. Am. 2006, 119, 1888. https://doi.org/10.1121/1.2166610 . [DOI] [PubMed] [Google Scholar]
- 37.Mattys S.L., Davis M.H., Bradlow A.R., Scott S.K. Speech Recognition in Adverse Conditions: A Review. Lang. Cogn. Process. 2012;27:953–978. doi: 10.1080/01690965.2012.705006. [DOI] [Google Scholar]
- 38.McCormack A., Fortnum H. Why Do People Fitted with Hearing Aids Not Wear Them? Int. J. Audiol. 2013;52:360–368. doi: 10.3109/14992027.2013.769066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Chisolm T.H., Johnson C.E., Danhauer J.L., Ports L.J.P., Abrams H.B., Lesner S., McCarthy P.A., Newman C.W. A Systematic Review of Health-Related Quality of Life Hearing Aids: Final Report of the American Academy of Audiology Task Force on the Health-Related Quality of Life Benefits of Amplication in Adults. J. Am. Acad. Audiol. 2007;18:151–183. doi: 10.3766/jaaa.18.2.7. [DOI] [PubMed] [Google Scholar]
- 40.National Institute for Health and Care Excellence . Quality Statement 6: Follow-Up Audiology Appointment|Hearing Loss in Adults|Quality Standards. NICE; London, UK: 2019. [Google Scholar]
- 41.National Institute for Health and Care Excellence . Hearing Loss in Adults: Assessment and Management. NICE Guideline NG98. NICE; London, UK: 2020. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All data are available within the manuscript and can be obtained from the corresponding authors upon a reasonable request.