Abstract
Background and objectives: Patient-specific skeletal implants are designed through a pipeline of imaging, computational planning and, increasingly, artificial intelligence. Each pillar has its own validation literature, but no review has asked what evidence supports each one in implant design itself. We mapped that evidence from the biomedical-indexed literature in this paper. Methods: A systematic review was conducted under a protocol committed before retrieval (PRISMA 2020, PRISMA-S, SWiM). MEDLINE, Europe PMC, Crossref and OpenAlex were searched on 2 and 3 September 2026 for primary studies from 2015 onwards applying artificial intelligence, an evaluated imaging pipeline or computational planning to an implant whose geometry derives from the individual patient’s anatomy. Engineering indexes were unavailable, so this review maps biomedical-indexed literature, and a bounded probe was used to estimate what that omits. Two artificial intelligence screeners from different vendors screened every record; every clinical study was appraised twice, independently, by assessors from different vendors. Claims about validation quality are reported on the studies read in full. All screening, charting and appraisals were machine-performed and no human-verified subsample exists. Results: Of 5059 records, 639 studies were included: 373 technical studies, 179 clinical series, 75 comparative studies and 12 randomised trials. Computational planning was applied in 601, an evaluated imaging step in 219 and artificial intelligence in 36. Randomised evidence came from 12 small trials, 10 of them craniomaxillofacial, which evaluated the delivered implant rather than the planning method; virtual surgical planning was used in 11 of them. When restricting to the four studies that measure deviation from the virtual plan, with smaller meaning better, patient-specific implants deviated less from the plan (standardised mean difference −1.21, 95% CI −1.69 to −0.74). Revision or removal was less frequent (3 of 135 against 25 of 149; risk ratio 0.28, 95% CI 0.11 to 0.72); operative time, complications and infection gave intervals compatible with a difference either way. No randomised trial was at low risk of bias, and 69 of 75 comparative studies were at serious risk, mostly of unadjusted confounding, so all estimates are of very low certainty. Among technical studies read in full, independent validation material was used in fewer than half, and files were shared in about one in fourteen. Of 36 artificial intelligence studies, 4 were evaluated on independent data, 2 were used for a real patient, none were clinically comparative, and 8 of 17 cranial studies relied on AutoImplant family benchmarks. Conclusions: The design pipeline is unequally validated: planning has clinical evidence concentrated in one region, computational methods are widely applied but rarely validated independently or shared, and artificial intelligence has benchmark evidence for cranial defect completion and little else. Outcome studies of patient-specific implants should be read as evidence about whole pipelines, and future studies should evaluate the components.
Keywords: patient-specific implant, custom implant, additive manufacturing, artificial intelligence, deep learning, finite element analysis, topology optimisation, virtual surgical planning, systematic review
1. Introduction
Implants whose geometry is derived from the individual patient’s imaging are now routine in cranioplasty, orbital and midface reconstruction, mandibular and maxillary reconstruction, pelvic and limb-salvage oncology and complex revision arthroplasty, and are spreading into primary joint replacement, spinal reconstruction and foot and ankle surgery [1,2,3,4]. The clinical literature is large and has been reviewed repeatedly by anatomical region: cranioplasty [5], orbital reconstruction [6], orthognathic surgery [7], maxillary reconstruction [8], mandibular reconstruction [9,10], the pelvis [3], the spine [1,11] and segmental long-bone defects [12]. Those reviews ask whether a patient-specific implant performs at least as well as a conventional one. They rarely ask how the implant was designed and whether the methods that produced it have themselves been validated.
The design pipeline has three technical pillars. The first is imaging: acquisition, segmentation and surface modelling produce the anatomy the implant is fitted to, and errors here propagate into every downstream step. The second is computational planning: virtual surgical planning positions the resection and the implant, statistical shape modelling or mirroring restores the missing anatomy, parametric or automated design generates the geometry, topology or lattice optimisation tunes stiffness and mass, and finite element analysis is used to verify the result before manufacture. The third and most recent is artificial intelligence, applied to segmentation, to defect and shape completion, to design generation and to prediction of design parameters or outcomes; the AutoImplant challenges made automatic cranial implant design a benchmark task for deep learning [13], and generative approaches have followed [14]. Each pillar has an engineering validation literature of its own, with reporting and credibility frameworks such as the risk-based credibility framework for computational models [15], the TRIPOD + AI and CLAIM guidelines for prediction models and imaging AI [16,17], and reviews of statistical shape model accuracy [18]. Clinical papers on patient-specific implants tend to invoke these methods as though their accuracy in the implant-design setting were settled.
No review, to our knowledge, has set out the evidence for each pillar in the specific setting of patient-specific skeletal implant design: which methods have been applied, for which regions and implant types, at which tier of evidence (in silico, bench, cadaver or phantom, retrospective imaging dataset, clinical series, comparative clinical study, randomised trial), against which comparators, with what outcomes, and with what risk of bias or technical credibility. A registry search run for this review found many region-specific registrations on patient-specific implant outcomes and a few on single methods, but none covering the three pillars across skeletal regions with evidence tiers. This review fills that gap for the literature indexed in biomedical databases. Its questions are what methods within each pillar have been applied and evaluated, at what tier and against what comparator, what outcomes they report, whether artificial intelligence methods have been evaluated on independent data or in clinical use, how credible the clinical and technical studies are, and where the evidence stops. The boundary matters and is stated at the outset: finite element, optimisation and machine learning work on implant design is published partly in engineering venues that the biomedical indexes do not cover, and the engineering indexes were not available for this review. What follows is therefore a map of what the biomedical literature contains, not a census of the field, and Section 3.8 measures how much sits outside it.
2. Materials and Methods
2.1. Design and Registration
This is a systematic review with a structured narrative synthesis and an evidence map, reported per PRISMA 2020 [19], with the search reported per PRISMA-S [20] and the synthesis per the Synthesis Without Meta-analysis guideline [21]. The protocol was written and committed to a version-controlled repository on 2 September 2026 before any record was retrieved or screened; eight PubMed hit-count queries, returning counts only, were run before the protocol to size the concept blocks and are archived with it. Every subsequent change is recorded in a dated amendments table in the protocol (Text S2). A PROSPERO record was prepared from the protocol but was not submitted, so the review was not prospectively registered.
2.2. Eligibility Criteria
We included primary studies published from 1 January 2015 to the search date, in any language, concerning a skeletal implant, that is a device implanted in or onto bone in orthopaedic, spinal, cranial, craniomaxillofacial or jaw reconstruction, including joint replacement components, tumour endoprostheses, cranioplasty plates, orbital and midface plates, mandibular and maxillary reconstruction implants, spinal cages and vertebral body replacements, and patient-specific subperiosteal or reconstruction jaw implants. Two concept criteria had to hold. First, the implant, or a load-bearing part of it, had to be patient-specific: its geometry derived from the individual patient’s imaging or anatomy. Off-the-shelf implants, including 3D-printed but non-individualised designs, did not qualify. Second, at least one of three pillars had to be applied to the implant’s design or planning and evaluated: artificial intelligence (machine or deep learning at any step of the pipeline); an evaluated imaging pipeline (segmentation accuracy for the design, the effect of imaging protocol or modality, registration or mirroring accuracy, statistical or template shape modelling for the design); or computational planning (finite element or other biomechanical simulation of the specific implant, topology, lattice or shape optimisation, parametric or automated design, or virtual surgical planning in which the implant itself is planned). Studies had to report at least one measured quantity about the method or the implant (geometric deviation or fit, mechanical response, design or production time, or a clinical outcome).
Eligible study types were in silico studies, bench or mechanical testing, cadaveric or phantom studies, retrospective imaging-dataset studies, clinical series of at least five patients, comparative non-randomised studies and randomised trials. Preprints carrying a full report were eligible and, where a journal version was also in the corpus, the preprint was treated as a duplicate of it (amendment A3); anatomy-derived jaw implants were eligible, but implants customised only by parameter choice without anatomy-derived geometry were not (amendment A2). We excluded case reports and series of fewer than five patients (counted in the flow, not charted), reviews, editorials, letters, protocols and conference abstracts, non-human studies, studies of patient-specific instrumentation or guides without a patient-specific implant, standard endosseous dental implants and prosthodontic restorations, external prostheses and orthoses, non-skeletal implants, and tissue-engineering studies without an implant designed for a specific patient.
2.3. Information Sources and Search
MEDLINE via PubMed (E-utilities), Europe PMC (title and abstract fields) and Crossref (a bounded relevance sweep, disclosed as supplementary rather than as a full-retrieval source) were searched by script on 2 September 2026, with raw responses archived. OpenAlex was in the protocol but could not be searched on 2 September: its API now meters a daily budget, and every request was refused (amendment A1). The prepared OpenAlex queries were run on 3 September 2026 once the budget had reset as a documented supplementary run (amendment A9); records not already in the corpus entered the same rule layer, dual triage and charting, and their counts are folded into the database arm of the flow and listed separately in Text S3. The main search was the intersection of four concept blocks, individualisation (patient-specific, custom, personalised, 3D-printed and synonyms), device, skeletal region and method (the union of the three pillars’ vocabularies), limited to publication dates from 2015 and with no outcome, language or study-type filter, so as to protect recall; a supplementary strategy used explicit patient-specific-implant phrases. Full strings, hit counts and retrieval counts are in Text S3. PROSPERO and OSF Registries were searched for competing reviews, and published systematic and scoping reviews on the topic were identified to position this review (searches/registry-and-prior-reviews.md in the repository). Reference lists of systematic and scoping reviews identified by the search or the registry search were chased backward through the NCBI reference links because the Europe PMC reference service was under maintenance on the run date; reference lists are indexed only for articles deposited in PubMed Central, so this chase is partial and is reported as such. Embase, Scopus, Web of Science, IEEE Xplore and CINAHL were unavailable to the project, a limitation declared in the protocol; the Crossref sweep is supplementary discovery of unknown recall, not a substitute for them. Because that omission bears directly on what this review can claim, its size was measured rather than asserted. Two checks were run at revision and are reported in Section 3.8. The first asks what one broader index already contributed by counting how many included studies entered only through the OpenAlex arm and were invisible to the three biomedical sources. The second runs the review’s own concept blocks against engineering-type sources in OpenAlex, identifies the records absent from this review’s corpus, screens a random sample of them against the eligibility criteria, and turns the observed eligible proportion into an estimate, with a confidence interval, of how many eligible studies the engineering route would add and of which kind they would be.
2.4. Selection
Selection had three layers, each with published per-record reasons. A rule layer excluded records by explicit, reproducible text rules: no individualisation term, no device term, a non-human indexing signal without an accompanying human signal (the exception, an implementation detail not stated in the committed rule, retained 138 records carrying both signals; amendment A4), a review, editorial, letter, comment or protocol publication type or a review-type title (these were retained separately for citation chasing), or the Case Reports publication type. A seeded random sample of 25 rule-excluded records was re-screened against the full criteria to bound the false-exclusion rate. The remaining records were then classified independently by two AI screeners from different vendors, a GPT-based agent run through the Codex command-line interface and a Claude model run as Claude Code subagents, each reading the same written instruction file (Text S4) and blind to the other, as to include, exclude or uncertain; a record was retained unless both excluded it (recall-union). Outputs were reconciled against the input identifier set so that no record could be dropped silently, and the full cross-tabulation, raw agreement and Cohen’s kappa are reported. Records were deduplicated by identifier at retrieval; both independent reviews of the first draft found that the same article could survive under a second identifier (a DOI-keyed copy beside the PubMed copy) or as a preprint beside its journal version, and such copies were excluded at assessment as duplicate reports, with the copy carrying full text then the PubMed copy retained (amendments A3 and A6). Retained records, together with citation-chase candidates screened in the same way, were assessed against the criteria at full record by an AI charter from open full text where it could be retrieved (PubMed Central, Europe PMC, Unpaywall) and otherwise from the abstract, with the text source recorded per record; a record with neither an abstract nor retrievable text was held as unresolved whatever the charter returned. All screening and assessment were AI-performed under the written instructions.
2.5. Data Charting
One row per report was charted by the AI charter from the written charting instructions (Text S4): identifiers, year, country, anatomical region, implant type, material and manufacturing route; the pillars applied, with method detail for each (artificial intelligence task and model family; imaging modality and method; computational method); the highest evidence tier reached; numbers of patients treated and of analysed units (cases, images, specimens, models or designs); comparator and its type; headline results by outcome domain (geometric accuracy or fit, mechanical, design or production time, clinical) with units; follow-up; for artificial intelligence studies, whether evaluation used data independent of development, whether any clinical or prospective evaluation exists, and whether code or data are public; and a verbatim sentence supporting the tier assignment. Funding source, conflict-of-interest declarations and industry authorship, prespecified but omitted from the charting form, were charted in a separate pass from the records with full text (amendment A7). A field the text did not settle was charted as unclear, never as absent. Three targeted re-adjudication passes, each from a written instruction file (Text S4), re-examined the records first charted as unclear, the jaw and dental records against amendment A2, and the evaluation-data source of every artificial intelligence study; their outputs were applied as corrections. A seeded random sample of 60 assessed records was charted a second time, independently, by a Claude model from the same instructions, and agreement on the key fields is reported; disagreements were adjudicated by the orchestrating system against the source text, and the adjudicated values were recorded as corrections separate from the charters’ output. Counts in this manuscript are computed by script from the charting file. A verification script checks that every numeric literal in the text and tables is a value the charting data produce; it bounds numerical drift but does not check that a value sits in the right sentence, which the independent reviews and the authors’ reading address separately.
2.6. Appraisal
Two different questions run through this review, and they need different evidence, so the target of every appraisal and every pooled estimate is stated before the tools are named. The first question is whether a delivered patient-specific implant performs better than a conventional one. Randomised and comparative clinical studies answer that, and RoB 2, ROBINS-I and GRADE are the right instruments for it. The second question, the one this review was written to ask, is whether the imaging, planning and artificial intelligence methods that produced the implant have themselves been validated. No clinical outcome study answers that, because every arm of such a study receives a whole pipeline rather than one component of it. Risk-of-bias appraisal and certainty rating in this review therefore describe the delivered-implant comparison and nothing more; the validation status of the design methods is assessed separately against the credibility items below, and the two are never combined into a single judgement. Clinical studies were appraised with the tool matching their design: RoB 2 for randomised trials [22], ROBINS-I for comparative non-randomised studies [23] and the JBI checklist for case series for non-comparative clinical series [24]. Technical studies (in silico, bench, cadaver or phantom, imaging dataset, artificial intelligence development) were charted against a six-item credibility set derived from the risk-based credibility framework for computational models [15] and from TRIPOD + AI and CLAIM [16,17]: validation data or specimens independent of those used to develop or tune the method; quantitative comparison against a physical experiment or a ground truth; uncertainty or sensitivity analysis; reproducibly specified inputs (imaging protocol, material properties, loads and boundary conditions, or training data description); availability of code, model or design files; and evaluation on more than one anatomy or specimen. Items are reported as counts, not as a score, because the six are not interchangeable and an aggregate would hide which one failed. Each item is coded ‘yes’ where the text establishes it, ‘no’ where the text establishes its absence, and ‘unclear’ where the text settles neither; silence is ‘unclear’ and never ‘no’, which is what allows the abstract-only route to be separated from a genuine deficiency in Section 3.5. Two further coders, from different vendors, applied the same six items independently to a seeded random sample of 90 technical studies so that the reproducibility of the set could be estimated rather than assumed; per-item agreement is reported in Section 3.5.
The six items were chosen to cover what is common to simulation, imaging and learned models, and they do not cover what TRIPOD + AI, CLAIM and PROBAST + AI ask specifically of prediction models. For the 36 artificial intelligence studies, a second, prediction-specific set was therefore charted separately: the named benchmark or private dataset the model was evaluated on, the risk that evaluation data were not independent of development data, whether calibration was reported, which discrimination or error metrics were reported, whether missing data handling was described, whether the sample size was justified, and whether validation was external in the TRIPOD + AI sense. Calibration is recorded as not applicable where a model outputs a geometry rather than a probability, because recording a false deficiency would be as misleading as missing a real one. PROBAST + AI itself was not applied: it appraises risk of bias in a single prediction model against an intended clinical use, and most of these studies report a reconstruction benchmark with no stated clinical use to appraise against. That is a limitation of what can be claimed here, not a judgement that the tool is unsuitable. Every clinical study was appraised twice. The first assessment was made by a GPT-based assessor and the second, independently and from the same written instruction file (Text S4), by a Claude-based assessor from a different vendor, which was blinded to the first assessor’s judgements and was instructed not to read them. Agreement on the overall judgement is reported per tool in Section 3.7, together with the full distribution of each assessor’s judgements, and every disagreement is listed per study in Table S4. The first draft of this review carried a single appraisal, and the duplicate assessment was added in response to peer review. Certainty of evidence was rated per outcome for the clinical comparisons that could be pooled or narratively synthesised (geometric accuracy, operative time, complications), using the GRADE domains of risk of bias, inconsistency, indirectness, imprecision and publication bias [25], starting from low certainty for the non-randomised body of evidence and from high for randomised trials, with each downgrade named (Section 3.7).
2.7. Synthesis
The synthesis is structured and narrative, grouped by pillar and then by anatomical region, with an evidence map of the pillar by tier and tables of included studies. Within each group, studies are presented in order of tier (randomised, comparative, series, then technical tiers) and, among artificial intelligence studies, those with evaluation on independent data are described first; results are summarised as counts of studies per category and as the range of reported values in their own units, without transformation, and without vote counting on direction or statistical significance. Heterogeneity is explored descriptively by tier, region, comparator class and whether the record was charted from full text.
The two kinds of claim this review makes are supported by different evidence and are reported on different denominators. Whether a study exists, what it studied, in which region, at which tier and with which pillar can be read from an abstract, so the evidence map uses all included studies and says so. Whether a study validated its method independently, compared it against a physical experiment, specified its inputs reproducibly or shared its files usually cannot be read from an abstract, so every claim of that kind is reported primarily on the studies read in full, with the abstract-only studies given separately rather than pooled with them. For each credibility item we also report the share met among only those records where the item could be read at all, which separates a method a study did not use from a method its abstract did not mention. The same split is applied to the risk-of-bias judgements, which for abstract-only records rest on absence of information and are labelled as such. The protocol permitted a random-effects meta-analysis only if at least three comparative clinical studies reported the same outcome against the same class of comparator with extractable estimates. That condition was tested on a dedicated extraction of arm-level estimates from the comparative studies and trials with full text (amendment A8): where it held, a DerSimonian–Laird random-effects model gave a pooled mean difference (operative time, minutes) or standardised mean difference (geometric accuracy, because measures differ), with tau-squared, I-squared and a t-based 95% prediction interval, for the delivered patient-specific implant against its conventional comparator; such an estimate is evidence about the implant as delivered, not about one design method against another, and is labelled as such. Membership of each pool was taken from the charted comparator class and, for accuracy, from the measure (deviation of the delivered result from the preoperative plan). Sensitivity analyses were a Hartung–Knapp confidence interval, leave-one-out estimates and, where ten or more studies were pooled, an Egger regression test for small-study effects; with fewer studies, no small-study test was attempted. At revision, three further analyses were added. Each pool was examined within an anatomical region. For deviation from the plan, two adjudicators from different vendors read each pooled paper’s methods to establish what its accuracy measure actually compared the delivered result against and whether a smaller value meant a better result; a study entered the primary estimate only where both adjudicators found a virtual-plan reference and a smaller-is-better direction, and the quoted sentence behind each decision is in Text S6. For the binary outcomes, events and denominators in both arms, the absolute risk difference and the number needed to treat are reported beside the risk ratio, because a ratio alone cannot be interpreted without the underlying risk. No minimal important difference has been established for deviation from plan in this setting; a deviation of about 2 mm is often treated as clinically acceptable in craniomaxillofacial reconstruction, but that threshold is a convention rather than a validated one, so differences are described in the measured units, and no study is called clinically important on the strength of a pooled standardised difference. Where the condition failed, the disposition is reported. The first draft’s decision not to pool although the condition was met was a post hoc deviation and was withdrawn on independent review.
2.8. Use of Artificial Intelligence and Verification
The search code, the rule layer, both title–abstract screening passes, the full-record assessment and charting, the second-charter sample, the output pipeline, the figures, the manuscript drafting and two rounds of independent adversarial pre-review (four reports) were performed with large language model agents (Claude Fable 5.1 and Claude Opus 5, Anthropic, San Francisco, CA, USA; GPT-5.6-sol via the Codex CLI, OpenAI, San Francisco, CA, USA) on 2 and 3 September 2026, orchestrated by human authors under the committed protocol; a contemporaneous log of which system did which stage on which date is shown in Text S5. Counts, meta-analyses and figures were computed with scripts in Python 3.14 (Python Software Foundation, Wilmington, DE, USA) using NumPy 2.5, SciPy 1.18 and Matplotlib 3.11. A first second-screening pass run with a smaller Claude model was superseded the same day by a pass with Claude Opus 5 at the authors’ direction; the superseded outputs are archived and did not enter the reported flow. The authors reviewed the instruction files, the reconciliation logic, the adjudicated corrections and every section of this manuscript and are accountable for its content. No AI system is reported as an author.
3. Results
3.1. Study Selection
The database arm retrieved 7772 records: 3207 from MEDLINE via PubMed across the two strategies, 1687 from Europe PMC, 2000 from the bounded Crossref sweep and 878 from the OpenAlex queries run on 3 September (amendment A9). After deduplication by identifier, 5059 unique records remained (2713 duplicates removed). The rule layer excluded 2682: 1850 with no individualisation term, 182 with no device term, 40 with a non-human signal and no human signal, 410 review, editorial or protocol records (set aside for citation chasing) and 200 records indexed as case reports. The seeded audit of 25 rule-excluded records found no wrongly excluded records. The sample was drawn from the 2566 records the rule layer excluded in the main run of 2 September before the supplementary OpenAlex run of 3 September added a further 116 rule exclusions to give the 2682 total; the audit therefore bounds the main run only, and the supplementary exclusions are unaudited. Because 25 records is a small sample, the exact one-sided 97.5% upper bound on the false-exclusion proportion is 13.7%, which would permit up to 353 wrongly excluded records among the 2566 audited exclusions. The audit is therefore a check that the rule layer was not grossly wrong, not a demonstration that recall was adequate, and the reader should carry that uncertainty into every count in this review. The 2377 remaining records entered dual AI triage. On the main run, the two screeners agreed on 516 inclusions and 1300 exclusions, with raw three-class agreement 0.893 and Cohen’s kappa 0.781, and for inclusion against non-inclusion agreement 0.947 and kappa 0.866; on the OpenAlex increment they agreed on 95 inclusions, with three-class agreement 0.879 and kappa on inclusion 0.874 (full cross-tabulations in Text S1). Under the recall-union rule, 924 records were retained. The citation-chasing arm covered 32 systematic and scoping reviews whose 836 references were available through the NCBI links; 728 were new after deduplication, 144 passed the rule layer, 89 were excluded by both screeners, and 55 were retained (kappa on inclusion 0.592). Of the 979 reports sought, full text was obtained and verified against the record title for 414; 505 were assessed from the abstract alone, and 60 had neither an abstract nor retrievable text and were held as unresolved, whatever the charter returned (Figure 1). After full-record assessment, 273 reports were excluded: 107 with no patient-specific implant, 31 without an evaluated design method, 30 case reports or series below five patients, 17 without a measured quantity, 15 by source type, 10 non-human, 3 non-skeletal, and 60 copies of a report already assessed (25 preprint, repository or version copies of a journal article and 35 copies of the same article under a second identifier; amendments A3 and A6, with the grouping rerun over all assessed records after independent review found the first pass incomplete). Sixty-seven reports remain unresolved: 47 of the 60 records without text (the other 13 were second copies of assessed reports and are counted among the duplicates) and 20 reports with text whose eligibility, patient denominator or patient-specificity it did not settle; 639 studies were included. Two measurements describe how reliable the charting is. The first shows agreement between the two charters on the seeded 60-record sample: eligibility 87% (kappa 0.70), evidence tier 87% (kappa 0.84), region 90% (kappa 0.86) and pillars applied 83% (kappa 0.73). Agreement, however, only shows that two readers coincide, not that either is right, so at revision a reference standard was built for the same sample. Two adjudicators from different vendors read each record’s own text, decided each field themselves and recorded the sentence that decided it. Where both adjudicators reached the same value, that value is the reference standard, which was available for between 83% and 100% of records depending on the field. Against it, the primary charter was correct for eligibility in 89.3% of records, evidence tier 89.3%, region 87.5%, pillars 88.7%, whether artificial intelligence evaluation was independent 100%, whether it was clinical 98.3%, and availability of code or data 85.0%; the second charter was correct in 71.7% to 98.3% across the same fields. On 15 records, the two adjudicators did not agree, so no reference standard exists there, and those records are listed for resolution by the authors. This represents agreement between independent machine readings of the same text, and it is not a human gold standard; no human-verified subsample exists for this review, and Section 4.3 says what that leaves open. The 68 field-level disagreements across 33 records were adjudicated against the source text, and those adjudications, the format repairs, the re-charting of records whose eligibility or patient denominator was questioned, the re-assessment of the 92 records first charted as unclear, the dental-boundary and AI-validation re-adjudications and the duplicate exclusions are all recorded as corrections separate from the charters’ output (Text S1). Of the included studies, 295 were charted from verified full text and 344 from the abstract only (clinical tiers 119 and 147; technical tiers 176 and 197); 16 are preprints. An audit of every cached full text, prompted by the second assessor noticing one, found 14 retrievals that had returned a repository or publisher landing page rather than the article. Such a page reprints the title and therefore passes a title-match guard while containing none of the paper. Those records were reclassified as abstract-only, which is what they always were, and two of them, having no abstract either, moved to unresolved; that is why 639 studies are included here where the submitted version reported 641.
Figure 1.

PRISMA 2020 flow. Parallel database and citation-chasing identification arms show deduplication, rule-based screening and dual independent AI triage before merging at report retrieval and eligibility assessment. Reports unavailable in full text were assessed from abstracts. Residual duplicate reports remain in both the assessed and excluded totals. Every junction is asserted against the machine-readable flow before rendering.
3.2. Evidence Map
The included studies applied computational planning in 601 cases, an evaluated imaging step in 219 and artificial intelligence in 36; 396 applied computational planning alone, 182 imaging with computational planning, 25 imaging alone, 7 artificial intelligence alone and 29 artificial intelligence with one or both of the other pillars (Table 1, Figure 2). By tier, 195 studies were in silico, 89 bench, 53 cadaver or phantom, 36 imaging-dataset, 179 non-comparative clinical series, 75 comparative non-randomised clinical studies and 12 randomised trials; 266 studies carry a clinical evaluation, and 373 are technical. The craniomaxillofacial region accounts for 290 studies, the cranial vault for 92, the pelvis and acetabulum for 55, the knee for 48, other lower-limb sites for 40, the spine for 35, other upper-limb sites for 25, the hip for 23, the shoulder for 7 and the chest wall for 5, with 14 multi-region and 5 other. Clinical evaluation concentrates in the craniomaxillofacial region, where 164 of 290 studies are clinical, and 10 of the 12 randomised trials sit, whereas the knee (38 of 48 technical), the spine (31 of 35), the hip (22 of 23), the cranial vault (57 of 92) and the shoulder (5 of 7) are evaluated mainly in silico, on the bench or on imaging datasets. Publications rose from 14 studies in 2015 to 78 in 2021 and 98 in 2025, with 76 already in 2026 at the search date. The 266 clinical studies reported 7444 treated patients, 5524 of them in the craniomaxillofacial region; median analysed units were 3 for in silico and bench studies, 10 for cadaver or phantom, 71 for imaging-dataset, 12 for clinical series, 36 for comparative studies and 25.5 for randomised trials. Every count in this review is a count of studies, and studies are of very unequal size, so the two should not be read as the same thing. Among the technical studies, the median number of analysed units is 3, and half analyse three or fewer, which means a large part of the finite element and optimisation literature consists of single-anatomy demonstrations; among clinical studies, the median is 18, and almost none analyse three or fewer; among artificial intelligence studies it is 161.5, because a benchmark dataset supplies many cases at once. A region with many small technical studies therefore looks busier on the evidence map than a region with fewer but larger ones, and the map should be read as showing where work has been done rather than how much evidence it carries. We have not weighted the map by size, because the unit of analysis differs between tiers; an anatomy, a specimen, a scan and a patient are not commensurable, and a weighted count would imply a comparability the data do not support. A comparator was present in 533 studies: an alternative design or method in 222, a conventional implant in 153, a ground truth in 117, no treatment in 17, a manual design in 14 and another comparator in 10. Titanium was the stated material in 219 studies, and 233 did not state one; 158 of the 195 in silico studies designed an implant that was never manufactured, 147 studies used metal additive manufacturing and 49 used polymer additive manufacturing. A public-code statement was charted in 13 studies (4 of them with data as well) and a public-data statement in 39; 181 stated that nothing was available and 412 made no statement.
Table 1.
Evidence map by pillar and tier.
| Pillar Combination | In Silico | Imaging Dataset | Bench | Cadaver or Phantom | Clinical Series | Clinical Comparative | Randomised Trial | Total |
|---|---|---|---|---|---|---|---|---|
| Artificial intelligence | 0 | 6 | 1 | 0 | 0 | 0 | 0 | 7 |
| Artificial intelligence + Computational planning | 5 | 6 | 3 | 2 | 1 | 0 | 0 | 17 |
| Artificial intelligence + Imaging | 0 | 5 | 1 | 0 | 0 | 0 | 0 | 6 |
| Artificial intelligence + Imaging + Computational planning | 1 | 3 | 2 | 0 | 0 | 0 | 0 | 6 |
| Computational planning | 177 | 4 | 59 | 17 | 93 | 39 | 7 | 396 |
| Imaging | 0 | 1 | 1 | 6 | 9 | 7 | 1 | 25 |
| Imaging + Computational planning | 12 | 11 | 22 | 28 | 76 | 29 | 4 | 182 |
| Total | 195 | 36 | 89 | 53 | 179 | 75 | 12 | 639 |
Figure 2.

Evidence map. Included studies by pillar combination and highest evidence tier. Cell annotations are study counts derived from the canonical dataset. The sequential viridis scale is designed to remain distinguishable for common forms of colour-vision deficiency.
3.3. Artificial Intelligence in Implant Design
Thirty-six studies applied machine or deep learning to a step of the design pipeline (Table 2); 25 of them were charted from the abstract only. Their tasks, which overlap, were shape completion of a skeletal defect in 17, generation of the implant geometry in 15, segmentation for design in 9, prediction of design parameters in 6, prediction of an outcome such as stress, displacement or alignment in 5 and other tasks in 3. Seventeen concerned the cranial vault, where the AutoImplant benchmark datasets have made automatic cranioplasty design a standard task: convolutional, U-Net, conditional generative adversarial, masked-autoencoder and latent-diffusion models reconstruct the missing skull from the defective CT and report Dice coefficients above 0.9 with 95th-percentile Hausdorff distances between about 1 and 2 mm on held-out challenge cases [26,27,28,29]. Five were craniomaxillofacial, including an nnU-Net segmentation of the skull, mandible, teeth and mandibular canal for computer-assisted surgery evaluated on 153 internal test scans and a separate external set of 10 (the preprint was treated as a duplicate of the journal article) [30], an encoder–decoder network for maxillary and zygomatic defect completion, developed on single-centre scans and evaluated on 2377 synthetically defected test volumes, with one printed case checked for physical fit [31], and a neural-network surrogate that predicts plate stress and displacement from design parameters in place of repeated finite element runs [32]. Four were knee, three spinal, and two each hip and chest wall, with one each for another lower-limb site, a multi-region pipeline and another site.
The evidence tier is low. Twenty of the 36 were imaging-dataset studies, 7 bench, 6 in silico and 2 cadaver or phantom; one was a clinical series, in which a model predicting postoperative lumbar lordosis and pelvic tilt was used alongside the design of patient-specific spinal rods for 34 patients, without the text establishing that the model determined the rod design [33]. Under the review’s definition, evaluation on data independent of the development source (a different centre, an external public benchmark or a prospective set) was established for 4 studies (2 on another centre’s data, 1 on a public benchmark and 1 prospectively); 16 evaluated on a split of their own source (10 by a patient-level or random split, 5 on synthetically generated defects and 1 by a split not otherwise described), and 16 gave no usable description of where the evaluation data came from. Clinical use of a model was established for one study: a cranioplasty pipeline that evaluated its network on 1930 internally generated test volumes and then compared the automated with the manual design in one representative patient, whose implant was manufactured and implanted, reporting asymmetry indices of 2.22% for the automated and 2.05% for the manual design [34]. Code or model files were stated to be public in 6 studies and data in 9. No artificial intelligence study was comparative at the clinical tier, and none were randomised.
Table 2.
Artificial intelligence studies.
| Study | Region | Task | Model | Analysed Units | Independent Evaluation | Clinical Evaluation | Headline Accuracy Result | Code or Data |
|---|---|---|---|---|---|---|---|---|
| Chaubey et al. (2025) [35] | chest_wall | design_generation | 3D U-Net with EfficientNet-B0 encoder | 160 | internal_only | no | DSC 0.2524; Hausdorff distance 148.90 mm | code |
| Shash et al. (2025) [36] | chest_wall | outcome_prediction | LR/Ridge/SVR/DT/RF/FCN/LightGBM | 8 | internal_only | no | AI model accuracy > 99.9% | unclear |
| Karlagina et al. (2022) [37] | cmf | other | U-Net | 7 | unclear | no | . | data |
| Dot et al. (2022) [30] | cmf | segmentation | nnU-Net | 163 | yes | no | test vDSC/sDSC: upper skull 96.22%/96.92%, mandible 94.19%/97.92%, upper teeth 94.83%/98.87%, lower teeth 94.38%/98.53%, mandibular canal 81.59%/97.90% | both |
| Beisekenov et al. (2025) [32] | cmf | parameter_prediction | feed-forward ANN + Bayesian network | . | internal_only | no | held-out test MAE 7.8 MPa for peak plate stress, 0.062 mm for displacement, and 6.4 MPa for screw stress | both |
| Juneja et al. (2026) [31] | cmf | shape_completion + design_generation | 3D encoder–decoder CNN | 2377 | internal_only | no | DSC 0.778; HD95 3.453 mm | unclear |
| Carlos et al. (2026) [38] | cmf | segmentation | nnU-Net | . | unclear | no | Pseudo Dice near 1 | unclear |
| Morais et al. (2018) [39] | cranial | shape_completion + design_generation | volumetric convolutional denoising autoencoder | . | unclear | no | . | data |
| Oldřich et al. (2019) [40] | cranial | segmentation | encoder–decoder CNN | . | unclear | no | . | unclear |
| Li et al. (2021) [41] | cranial | segmentation | voxel rearrangement network | . | unclear | no | . | code |
| Li et al. (2021) [26] | cranial | shape_completion + design_generation | CNN | 167 | internal_only | no | . | code |
| Sulakhe et al. (2022) [42] | cranial | shape_completion + design_generation | 3D conditional GAN | . | unclear | no | . | unclear |
| Wodzinski et al. (2022) [27] | cranial | shape_completion + design_generation | modified U-Net | . | internal_only | no | Dice 0.91; boundary Dice 0.94; 95th-percentile Hausdorff distance 1.53 mm | both |
| Wu et al. (2023) [34] | cranial | shape_completion + design_generation | 3D U-Net | 1930 | yes | yes | ACVAI 2.22% vs. 2.05% manual; PCVAI 2.14% vs. 6.03% manual | data |
| S et al. (2024) [43] | cranial | shape_completion + design_generation | 3D U-Net with vision transformer | . | unclear | no | higher Dice than 3D U-Net and attention 3D U-Net | unclear |
| Kesornsri et al. (2024) [44] | cranial | shape_completion + design_generation | ConvNeXt | . | unclear | no | Dice 0.7969 ± 0.13 | unclear |
| Zhang et al. (2024) [45] | cranial | design_generation + parameter_prediction | neural network | . | unclear | no | . | unclear |
| Wodzinski et al. (2024) [28] | cranial | shape_completion | latent diffusion + VQ-VAE + GAN | . | yes | no | Dice > 0.94 on SkullBreak and >0.96 on SkullFix | unclear |
| Wodzinski et al. (2024) [46] | cranial | shape_completion | masked autoencoder | . | internal_only | no | . | unclear |
| Wodzinski et al. (2025) [29] | cranial | shape_completion | deep neural network | . | yes | no | DSC 0.94; bDSC 0.94; HD95 1.31 versus 0.84/0.76/2.43 baseline | data |
| Juneja et al. (2025) [47] | cranial | shape_completion + design_generation | unclear | . | internal_only | no | DSC 0.889; Hausdorff distance 1.856 mm; SSIM 0.897 | unclear |
| Salem et al. (2026) [48] | cranial | shape_completion + design_generation | 3D U-Net | 40 | internal_only | no | mean Dice 0.901; Boundary Dice 0.908; HD95 1.52 mm | data |
| Nguyen et al. (2026) [49] | cranial | shape_completion | 3D U-Net | 250 | internal_only | no | Dice 0.94; Hausdorff distance 3.8 mm | unclear |
| Guo et al. (2026) [50] | cranial | segmentation + design_generation + parameter_prediction + outcome_prediction | BASNet + XGBoost + U-Net + conditional diffusion | . | unclear | no | segmentation accuracy 94.7%; porosity error 3.2 ± 1.8%; elastic modulus error 4.1 ± 2.3% | unclear |
| Banis et al. (2025) [51] | hip | segmentation + other | UNETR | 1300 | unclear | no | FEA-DIC total displacement difference 0.17 ± 0.09 mm | unclear |
| Yáñez González-Cuéllar et al. (2026) [52] | hip | outcome_prediction | dual-CNN | 392 | unclear | no | mean absolute error reduced approximately 30% vs. random forest | unclear |
| Kanagalingam et al. (2023) [53] | knee | design_generation | unclear | 2 | unclear | no | mean surface deviation −0.10 mm for plate A and −0.11 mm for plate B | none |
| Burge et al. (2023) [54] | knee | segmentation + shape_completion + design_generation + other | CNN + U-Net + SSM | 165 | internal_only | no | femur model RMSE 0.87 (SD 0.20) mm and component RMSE 0.85 (SD 0.21) mm, OUH ≥3 mm in 6/84; tibia model RMSE 0.80 (SD 0.15) mm and component RMSE 0.82 (SD 0.27) mm, OUH ≥3 mm in 1/81 | unclear |
| Grothues et al. (2023) [55] | knee | parameter_prediction | artificial neural network | 673 | internal_only | no | prediction error below each parameter’s standard deviation; regression had the lowest errors | unclear |
| Guezou-Philippe et al. (2025) [56] | knee | segmentation + shape_completion | 2D U-Net + SSM | 40 | internal_only | no | segmentation Dice 98.5% (SD 0.8%); full-bone RMSE 1.0 mm (SD 0.3); implant-fit RMSE 0.9 mm (SD 0.5) | none |
| He et al. (2023) [57] | lower_limb_other | segmentation | nnU-Net | . | internal_only | no | DSC > 0.95; HD95 < 0.6 mm | unclear |
| Mazzocchetti et al. (2024) [58] | multi | shape_completion | point-cloud completion networks | 71 | internal_only | no | PointAttN accuracy 1.4468 mm; Chamfer distance 3.2731 mm; F-score 0.8929 | data |
| Zhou et al. (2025) [59] | other | outcome_prediction | neural network + random forest | 80 | internal_only | no | prediction error 2–6% for nonlinear machine-learning models | unclear |
| Biswas et al. (2020) [60] | spine | parameter_prediction | artificial neural network | . | unclear | no | . | unclear |
| Von Atzigen et al. (2022) [61] | spine | parameter_prediction | stereo neural network | . | unclear | no | . | unclear |
| Ngo et al. (2026) [33] | spine | outcome_prediction | unclear | 34 | unclear | unclear | postoperative-minus-predicted lumbar lordosis −3.8° and pelvic tilt −0.1° at 2 years | unclear |
cmf, craniomaxillofacial; ACVAI and PCVAI, anterior and posterior cranial vault asymmetry index; ANN, artificial neural network; bDSC, boundary Dice similarity coefficient; CNN, convolutional neural network; DSC, Dice similarity coefficient (vDSC, volumetric; sDSC, surface); DT, decision tree; FCN, fully connected neural network; FEA-DIC, finite element analysis compared with digital image correlation; GAN, generative adversarial network; HD95, 95th-percentile Hausdorff distance; LR, linear regression; MAE, mean absolute error; OUH, maximum over- or underhang of the implant edge; RF, random forest; RMSE, root-mean-square error; SD, standard deviation; SSIM, structural similarity index; SSM, statistical shape model; SVR, support vector regression; VQ-VAE, vector-quantised variational autoencoder. U-Net, nnU-Net, UNETR, ConvNeXt, EfficientNet, BASNet, PointAttN, XGBoost and LightGBM are the architecture or software names used by the original authors. Region, task and evaluation labels (e.g., shape_completion, internal_only) are the charting vocabulary defined in Text S4; internal_only, evaluated only on a split of the development source; a full stop (.) indicates not reported.
Because these 36 studies do quite different things, they are also reported by task rather than as one set (Table S7). Shape completion accounts for 17 of them, design generation 15, segmentation 9, prediction of a design parameter 6 and prediction of an outcome 5, and the tasks overlap where a study segments an anatomy and then generates an implant from it. Evaluation independent of the development source was established for 3 of the 17 shape-completion studies, 1 of the 15 design-generation studies, 1 of the 9 segmentation studies and none of the 11 prediction studies, so the maturity of the subfields differs and the aggregate figure conceals that.
The cranial studies were tabulated by the dataset they evaluated on, because this literature reuses a small number of public benchmarks and reuse carries a leakage risk. Two coders from different vendors read each study. Across the 17 cranial studies, the named evaluation data were SkullFix in 6, SkullBreak in 4, AutoImplant in 2, another named public dataset in 5 (most often MUG500+), institutional data in 5, and synthetically generated defects in 7; 5 named no evaluation dataset at all. Eight of the seventeen, or 47.1%, evaluate on data from the AutoImplant family, which means a substantial part of what is known about automatic cranial implant design rests on one benchmark lineage, and a method that performs well there has not thereby been shown to generalise. On the risk that evaluation data were not independent of development data, the two coders agreed on low risk for 2 of the 36 studies and on high risk for 2, agreed that 17 could not be judged from the text, and differed on 11; the dominant finding is that most of these papers do not describe how their evaluation cases relate to their training cases, which is a reporting failure rather than a demonstrated defect. Against the prediction-model reporting items, calibration is not an applicable concept for 26 of the 36, whose output is a geometry rather than a probability, and among those where it does apply, it was reported in 2 and not reported in 4; handling of missing data was described in a minority and a sample size was justified in fewer still. PROBAST + AI was not applied for the reason given in Section 2.6.
3.4. Imaging Pipeline Evaluations
An imaging step was itself evaluated in 219 studies. The most common was mirroring of the contralateral anatomy to generate the target shape (99 studies, 72 of them clinical, mostly orbital and midface reconstruction), followed by registration of the postoperative result to the plan or of intraoperative to preoperative data (65, 49 clinical), segmentation accuracy for the design (13, all technical), template or atlas fitting (12), statistical shape modelling (10, all technical) and the effect of the imaging protocol or modality on the implant (6). The clinical imaging studies mostly report the geometric accuracy of the implanted result against the virtual plan: of the 266 clinical studies, 181 report an accuracy measure, typically a mean surface or landmark deviation of the order of 1 mm for orbital, orthognathic and mandibular implants, and 85 report a design, planning or operative time. Only 6 studies, analysing between 10 and 83 units each, examined how acquisition parameters propagate into implant geometry, and the 13 segmentation-accuracy studies evaluated segmentation against a reference without carrying the error through to an implant. Statistical shape models appeared as design tools in 13 studies and as imaging evaluations in 10, all below the clinical tier. Of the studies with an imaging pillar, 131 were charted from full text and 88 from the abstract.
3.5. Computational Planning
Computational planning was applied in 601 studies. Finite element analysis was the most frequent method (312 studies, 281 technical and 31 clinical), followed by virtual surgical planning of the implant (261, of which 223 clinical), parametric design (113), topology optimisation (49), automated design (43), lattice optimisation (42), shape optimisation (21), musculoskeletal modelling (16) and statistical-shape-model-driven design (13). Table 3 summarises the technical evidence by region and method. The finite element literature is dominated by single-anatomy studies: median analysed units were 3 in silico, and of the 373 technical studies, 284 report a mechanical result, usually peak von Mises stress, displacement or stiffness under a static load case, whereas 104 report a geometric accuracy result. Against the six credibility items, the distributions among the 373 technical studies were as follows: validation data or specimens independent of development—119 yes, 181 no, 73 unclear; quantitative comparison against a physical experiment or ground truth—182 yes, 179 no, 12 unclear; uncertainty or sensitivity analysis—187 yes, 142 no, 44 unclear; inputs specified reproducibly—174 yes, 17 no, 182 unclear; code, model or design files available—17 yes, 134 no, 222 unclear; and more than one anatomy or specimen evaluated—149 yes, 183 no, 41 unclear.
Table 3.
Technical evidence by region and method.
| Region | Pillar | Method | Studies | Tier Breakdown | Analysed Units Range |
|---|---|---|---|---|---|
| chest_wall | computational | fea | 3 | In silico: 2; Bench: 1 | 1 to 8 |
| chest_wall | computational | parametric_design | 1 | Bench: 1 | 3 to 3 |
| chest_wall | computational | ssm_design | 1 | Cadaver or phantom: 1 | 288 to 288 |
| cmf | computational | automated_design | 7 | In silico: 2; Imaging dataset: 1; Bench: 1; Cadaver or phantom: 3 | 1 to 21 |
| cmf | computational | fea | 107 | In silico: 73; Imaging dataset: 2; Bench: 28; Cadaver or phantom: 4 | 1 to 2377 |
| cmf | computational | lattice_opt | 9 | In silico: 3; Bench: 6 | 1 to 8 |
| cmf | computational | musculoskeletal_model | 4 | In silico: 2; Bench: 2 | 1 to 7 |
| cmf | computational | parametric_design | 23 | In silico: 12; Bench: 9; Cadaver or phantom: 2 | 1 to 40 |
| cmf | computational | shape_opt | 9 | In silico: 8; Bench: 1 | 2 to 25 |
| cmf | computational | ssm_design | 3 | Imaging dataset: 3 | 19 to 35 |
| cmf | computational | topology_opt | 23 | In silico: 10; Imaging dataset: 1; Bench: 10; Cadaver or phantom: 2 | 1 to 35 |
| cmf | computational | vsp | 17 | In silico: 6; Imaging dataset: 2; Bench: 1; Cadaver or phantom: 8 | 1 to 160 |
| cmf | imaging | mirroring | 12 | In silico: 1; Imaging dataset: 4; Bench: 5; Cadaver or phantom: 2 | 3 to 160 |
| cmf | imaging | other | 11 | In silico: 3; Bench: 4; Cadaver or phantom: 4 | 1 to 40 |
| cmf | imaging | protocol_effect | 1 | Cadaver or phantom: 1 | 10 to 10 |
| cmf | imaging | registration | 2 | Cadaver or phantom: 2 | 10 to 36 |
| cmf | imaging | segmentation_accuracy | 3 | In silico: 1; Imaging dataset: 1; Bench: 1 | 5 to 163 |
| cmf | imaging | ssm | 3 | Imaging dataset: 3 | 19 to 35 |
| cmf | imaging | template | 2 | Bench: 2 | 30 to 30 |
| cranial | computational | automated_design | 15 | In silico: 2; Imaging dataset: 5; Bench: 4; Cadaver or phantom: 4 | 1 to 1930 |
| cranial | computational | fea | 27 | In silico: 14; Imaging dataset: 1; Bench: 10; Cadaver or phantom: 2 | 1 to 144 |
| cranial | computational | lattice_opt | 4 | In silico: 1; Bench: 3 | 1 to 6 |
| cranial | computational | parametric_design | 9 | In silico: 4; Imaging dataset: 1; Bench: 4 | 1 to 30 |
| cranial | computational | shape_opt | 5 | In silico: 5 | 1 to 6 |
| cranial | computational | topology_opt | 1 | Bench: 1 | . |
| cranial | computational | vsp | 6 | Bench: 2; Cadaver or phantom: 4 | 1 to 11 |
| cranial | imaging | mirroring | 9 | In silico: 1; Imaging dataset: 1; Bench: 3; Cadaver or phantom: 4 | 1 to 20 |
| cranial | imaging | other | 5 | Imaging dataset: 2; Bench: 1; Cadaver or phantom: 2 | 2 to 30 |
| cranial | imaging | registration | 7 | Imaging dataset: 1; Bench: 1; Cadaver or phantom: 5 | 1 to 15 |
| cranial | imaging | segmentation_accuracy | 5 | Imaging dataset: 3; Bench: 1; Cadaver or phantom: 1 | 23 to 23 |
| cranial | imaging | template | 3 | Imaging dataset: 1; Bench: 1; Cadaver or phantom: 1 | 1 to 1 |
| hip | computational | automated_design | 3 | Bench: 2; Cadaver or phantom: 1 | 1 to 1300 |
| hip | computational | fea | 19 | In silico: 10; Bench: 8; Cadaver or phantom: 1 | 1 to 1300 |
| hip | computational | lattice_opt | 10 | In silico: 3; Bench: 7 | 1 to 9 |
| hip | computational | musculoskeletal_model | 3 | In silico: 2; Bench: 1 | 1 to 57 |
| hip | computational | parametric_design | 8 | In silico: 3; Bench: 4; Cadaver or phantom: 1 | 1 to 1300 |
| hip | computational | ssm_design | 1 | Imaging dataset: 1 | . |
| hip | computational | topology_opt | 4 | In silico: 1; Bench: 3 | 1 to 9 |
| hip | imaging | other | 2 | Bench: 2 | 1 to 1 |
| hip | imaging | ssm | 1 | Imaging dataset: 1 | . |
| knee | computational | automated_design | 3 | Imaging dataset: 2; Bench: 1 | 2 to 165 |
| knee | computational | fea | 30 | In silico: 25; Bench: 4; Cadaver or phantom: 1 | 1 to 28 |
| knee | computational | lattice_opt | 2 | Bench: 2 | 9 to 24 |
| knee | computational | musculoskeletal_model | 4 | In silico: 4 | 4 to 28 |
| knee | computational | parametric_design | 15 | In silico: 9; Imaging dataset: 2; Bench: 3; Cadaver or phantom: 1 | 1 to 673 |
| knee | computational | ssm_design | 2 | Imaging dataset: 2 | 40 to 165 |
| knee | computational | topology_opt | 3 | In silico: 2; Bench: 1 | 1 to 24 |
| knee | computational | vsp | 3 | Imaging dataset: 1; Bench: 1; Cadaver or phantom: 1 | 6 to 19 |
| knee | imaging | mirroring | 1 | In silico: 1 | 3 to 3 |
| knee | imaging | other | 4 | In silico: 1; Bench: 2; Cadaver or phantom: 1 | 1 to 24 |
| knee | imaging | registration | 1 | Cadaver or phantom: 1 | 6 to 6 |
| knee | imaging | segmentation_accuracy | 2 | Imaging dataset: 2 | 40 to 165 |
| knee | imaging | ssm | 2 | Imaging dataset: 2 | 40 to 165 |
| lower_limb_other | computational | automated_design | 2 | Bench: 2 | 1 to 2 |
| lower_limb_other | computational | fea | 22 | In silico: 15; Bench: 6; Cadaver or phantom: 1 | 1 to 720 |
| lower_limb_other | computational | lattice_opt | 6 | Bench: 5; Cadaver or phantom: 1 | 1 to 6 |
| lower_limb_other | computational | musculoskeletal_model | 2 | In silico: 2 | 1 to 1 |
| lower_limb_other | computational | parametric_design | 10 | In silico: 5; Imaging dataset: 1; Bench: 4 | 1 to 140 |
| lower_limb_other | computational | shape_opt | 1 | In silico: 1 | 720 to 720 |
| lower_limb_other | computational | ssm_design | 2 | In silico: 1; Imaging dataset: 1 | 4 to 140 |
| lower_limb_other | computational | topology_opt | 5 | In silico: 1; Bench: 4 | 1 to 6 |
| lower_limb_other | computational | vsp | 2 | In silico: 1; Cadaver or phantom: 1 | 3 to 12 |
| lower_limb_other | imaging | mirroring | 1 | Bench: 1 | . |
| lower_limb_other | imaging | other | 4 | Bench: 3; Cadaver or phantom: 1 | 1 to 1 |
| lower_limb_other | imaging | segmentation_accuracy | 1 | Imaging dataset: 1 | . |
| multi | computational | automated_design | 1 | Imaging dataset: 1 | 71 to 71 |
| multi | computational | fea | 3 | In silico: 1; Bench: 2 | 2 to 60 |
| multi | computational | lattice_opt | 1 | In silico: 1 | 6 to 6 |
| multi | imaging | mirroring | 1 | Cadaver or phantom: 1 | 3 to 3 |
| other | computational | fea | 5 | In silico: 3; Bench: 1; Cadaver or phantom: 1 | 80 to 80 |
| other | computational | parametric_design | 2 | In silico: 1; Bench: 1 | 80 to 80 |
| other | computational | shape_opt | 1 | Bench: 1 | . |
| other | computational | vsp | 1 | In silico: 1 | . |
| other | imaging | other | 1 | Bench: 1 | . |
| pelvis_acetabulum | computational | automated_design | 1 | In silico: 1 | 15 to 15 |
| pelvis_acetabulum | computational | fea | 25 | In silico: 19; Bench: 6 | 1 to 10 |
| pelvis_acetabulum | computational | lattice_opt | 2 | In silico: 1; Bench: 1 | 2 to 2 |
| pelvis_acetabulum | computational | musculoskeletal_model | 2 | In silico: 2 | 1 to 1 |
| pelvis_acetabulum | computational | parametric_design | 5 | In silico: 4; Bench: 1 | 1 to 6 |
| pelvis_acetabulum | computational | ssm_design | 2 | Imaging dataset: 2 | 90 to 90 |
| pelvis_acetabulum | computational | topology_opt | 5 | In silico: 4; Bench: 1 | 1 to 3 |
| pelvis_acetabulum | computational | vsp | 5 | Imaging dataset: 2; Bench: 1; Cadaver or phantom: 2 | 10 to 111 |
| pelvis_acetabulum | imaging | mirroring | 1 | Imaging dataset: 1 | 13 to 13 |
| pelvis_acetabulum | imaging | other | 3 | In silico: 2; Imaging dataset: 1 | 1 to 13 |
| pelvis_acetabulum | imaging | registration | 3 | Imaging dataset: 1; Cadaver or phantom: 2 | 10 to 111 |
| pelvis_acetabulum | imaging | ssm | 2 | Imaging dataset: 2 | 90 to 90 |
| shoulder | computational | fea | 4 | In silico: 3; Bench: 1 | 1 to 19 |
| shoulder | computational | parametric_design | 1 | In silico: 1 | 19 to 19 |
| shoulder | computational | vsp | 1 | Cadaver or phantom: 1 | 10 to 10 |
| shoulder | imaging | other | 1 | In silico: 1 | 2 to 2 |
| spine | computational | automated_design | 6 | In silico: 4; Cadaver or phantom: 2 | 7 to 1080 |
| spine | computational | fea | 26 | In silico: 20; Bench: 4; Cadaver or phantom: 2 | 1 to 24 |
| spine | computational | lattice_opt | 5 | In silico: 3; Bench: 1; Cadaver or phantom: 1 | 4 to 24 |
| spine | computational | parametric_design | 10 | In silico: 7; Bench: 2; Cadaver or phantom: 1 | 1 to 1080 |
| spine | computational | shape_opt | 2 | In silico: 2 | 3 to 1080 |
| spine | computational | ssm_design | 1 | In silico: 1 | 10 to 10 |
| spine | computational | topology_opt | 5 | In silico: 2; Bench: 2; Cadaver or phantom: 1 | 2 to 16 |
| spine | computational | vsp | 1 | Cadaver or phantom: 1 | 5 to 5 |
| spine | imaging | other | 4 | Bench: 1; Cadaver or phantom: 3 | 3 to 18 |
| spine | imaging | ssm | 1 | In silico: 1 | 10 to 10 |
| upper_limb_other | computational | automated_design | 1 | In silico: 1 | 6 to 6 |
| upper_limb_other | computational | fea | 10 | In silico: 6; Bench: 1; Cadaver or phantom: 3 | 1 to 19 |
| upper_limb_other | computational | musculoskeletal_model | 1 | Bench: 1 | 1 to 1 |
| upper_limb_other | computational | parametric_design | 4 | In silico: 2; Cadaver or phantom: 2 | 1 to 19 |
| upper_limb_other | computational | shape_opt | 2 | In silico: 1; Bench: 1 | 1 to 6 |
| upper_limb_other | computational | ssm_design | 1 | Cadaver or phantom: 1 | 2 to 2 |
| upper_limb_other | computational | vsp | 2 | Cadaver or phantom: 2 | 5 to 20 |
| upper_limb_other | imaging | mirroring | 2 | Cadaver or phantom: 2 | 2 to 11 |
| upper_limb_other | imaging | other | 2 | Cadaver or phantom: 2 | 5 to 20 |
| upper_limb_other | imaging | protocol_effect | 3 | Imaging dataset: 1; Cadaver or phantom: 2 | 20 to 20 |
| upper_limb_other | imaging | registration | 3 | In silico: 1; Cadaver or phantom: 2 | 6 to 20 |
| upper_limb_other | imaging | segmentation_accuracy | 2 | Imaging dataset: 1; Cadaver or phantom: 1 | 20 to 20 |
| upper_limb_other | imaging | ssm | 1 | Cadaver or phantom: 1 | 2 to 2 |
cmf, craniomaxillofacial; fea, finite element analysis; lattice_opt, shape_opt and topology_opt, lattice, shape and topology optimisation; multi, multi-region; protocol_effect, effect of the imaging protocol or modality on the implant; ssm, statistical shape model; ssm_design, design driven by a statistical shape model; vsp, virtual surgical planning. Region and method labels are the charting vocabulary defined in Text S4.
Those totals mix two different things and should not be read as they stand, because an item recorded as unclear in an abstract-only record says nothing about the study. The primary reading is therefore restricted to the studies read in full, and within those to the records where the item could be read at all. On that basis, independent validation material was used in 48.8% of the 168 full-text records where the item could be read, against 28.0% of the 132 readable abstract-only records; a physical or ground-truth comparison in 58.5% of 176 against 42.7% of 185; an uncertainty or sensitivity analysis in 64.4% of 174 against 48.4% of 155; inputs specified reproducibly in 94.1% of 170 against 66.7% of only 21 readable; files available in 7.2% of 138 against 53.8% of only 13 readable; and more than one anatomy in 44.5% of 173 against 45.3% of 159. Two of these deserve care rather than a headline. Reproducible input specification and file availability are unclear in 89.3% and 93.4% of abstract-only records, so the abstract percentages rest on a handful of records, and the full-text figures are the only usable ones; the apparently higher file availability in abstracts is due to selection, since an abstract mentions shared data only when the authors treat it as a selling point. The substantive finding survives the restriction and is a full-text finding: among technical studies that can actually be read, independent validation material is used in fewer than half, a physical comparison in about three in five, and code, model or design files are available in about one in fourteen.
How reproducible are these judgements? Two further coders from different vendors applied the same six items independently to a seeded random sample of 90 technical studies. They agreed with each other on 75.6% to 98.9% of records per item (kappa 0.32 to 0.91) and with the original coding on 51.1% to 88.9%. Agreement was highest for file availability (98.9% between the new coders) and for reproducible inputs (86.7%) and lowest for independent validation material and for uncertainty analysis, where the new coders matched the original coding only about half the time. Those two items are exactly the ones carrying the review’s strongest claims about validation, so the counts above should be read as approximate, with the direction more reliable than the precise figure. The recurring source of disagreement is what counts as varying an input and whether comparing a computed result against published experimental values from another study is a physical comparison. Within the 281 technical finite element studies, inputs were specified reproducibly in 126 (8 no, 147 unclear), independent validation material was used in 76 (163 no, 42 unclear), a physical or ground-truth comparison was made in 108 (166 no, 7 unclear) and files were available in 6 (98 no, 177 unclear); the precise reading is therefore that input specification is usually present where it can be read, whereas independent validation, physical comparison and file availability are usually absent where they can be read. Topology, lattice and automated design studies were almost entirely technical (46 of 49, 39 of 42 and 40 of 43), with 3 clinical studies each; virtual surgical planning was the only computational method whose evidence lies mainly at the clinical tier.
3.6. Clinical Evidence
Clinical evaluation of a patient-specific implant designed with an evaluated pillar was reported in 266 studies: 179 non-comparative series (Table S1), 75 comparative non-randomised studies and 12 randomised trials (Table 4). The comparative studies concentrated in the craniomaxillofacial region (60 of 75), with 5 pelvic and acetabular, 4 cranial, 2 upper-limb and 1 each spinal, hip, knee and shoulder; 49 compared against a conventional implant, 22 against an alternative design or planning method, 3 against a manual design and 1 against another comparator. The 12 randomised trials randomised between 10 and 58 patients each (median 25.5); 10 were craniomaxillofacial (patient-specific orthognathic or maxillary osteosynthesis plates against conventional plates or splint-based planning in 5, orbital implants in 3, a maxillary ridge augmentation mesh in 1, mandibular fracture plates in 1), one compared a patient-specific acetabular plate with intraoperative contouring and one a custom osteosynthesis plate for the distal humerus with conventional plating. The implant was designed by virtual surgical planning in 11 trials and by mirroring-based mesh adaptation without a stated planning comparison in one orbital trial; an imaging step was itself evaluated in 5. Reported outcomes were operative time or blood loss, accuracy of the transferred plan, occlusal or orbital position, complications and, in the distal humerus trial, bone union and elbow function; serious adverse events were not consistently reported, and their absence from a headline result is not evidence of no difference. Outside the craniomaxillofacial region the clinical evidence is predominantly non-comparative but not exclusively so: the cranial vault has 31 series and 4 comparative studies, the pelvis and acetabulum 17 series, 5 comparative studies and 1 randomised trial, other upper-limb sites 4 series, 2 comparative studies and 1 randomised trial, other lower-limb sites 11 series, the knee 9 series and 1 comparative study, the spine 3 series and 1 comparative study, the shoulder 1 series and 1 comparative study, the hip 1 comparative study and multi-region series 9. Of the 266 clinical studies, 224 report a clinical outcome, 181 a geometric accuracy outcome and 85 a time outcome; follow-up was stated in 159. By text source, a clinical outcome was charted for 113 of the 119 clinical studies with full text and 111 of the 147 charted from abstracts, an accuracy outcome for 75 and 106, and a time outcome for 53 and 32.
Table 4.
Randomised trials and comparative clinical studies (non-comparative series are listed in Table S1; a clinical-result cell without figures reproduces the authors’ qualitative statement and is not an effect estimate).
| Study | Design | Region | Implant Type | Pillars | Patients | Comparator Type | Clinical Result | Follow-Up |
|---|---|---|---|---|---|---|---|---|
| Raisian et al. (2017) [62] | RCT | cmf | orbital titanium mesh | imaging | 10 | manual_design | preoperative vertical dystopia or diplopia resolved in 2/5 intervention patients | 4 months |
| Schneider et al. (2019) [63] | RCT | cmf | pre-bent orthognathic fixation plate | comp | 21 | conventional_implant | . | . |
| Li et al. (2021) [64] | RCT | cmf | maxillary osteosynthesis plate | comp | 58 | alternative_method | infection 1/27 vs. 1/31; no significant difference in serious adverse events | within 1 week |
| Sánchez-Jáuregui et al. (2022) [65] | RCT | cmf | maxillary osteosynthesis plate | comp | 30 | alternative_method | . | . |
| Van der Wel et al. (2023) [66] | RCT | cmf | maxillary osteosynthesis plate | comp | 58 | conventional_implant | . | 1 year |
| Bevini et al. (2025) [67] | RCT | cmf | orthognathic fixation implant | comp | 20 | alternative_method | no surgical complications [authors’ qualitative statement; no figures] | 1 month |
| Sheta et al. (2025) [68] | RCT | cmf | orbital wall implant | imaging + comp | 28 | conventional_implant | residual enophthalmos 1/14 vs. 3/14; residual diplopia 0/14 vs. 1/14 | 3 months |
| Alshikh et al. (2026) [69] | RCT | cmf | maxillary ridge augmentation mesh | comp | 14 | conventional_implant | gained bone volume 499.47 ± 80.46 mm3 PEEK vs. 370.82 ± 51.69 mm3 titanium (p = 0.004); graft loss 174.83 ± 40.78 mm3 vs. 127.73 ± 35.55 mm3 (p = 0.040); mesh exposure 1/14 sites vs. 2/14 sites | 6 months |
| El Morsy et al. (2026) [70] | RCT | cmf | orbital floor implant | imaging + comp | 24 | alternative_method | complications: 3 in titanium group and 0 in zirconia group | . |
| Abdellatif et al. (2026) [71] | RCT | cmf | mandibular fracture fixation plates | imaging + comp | 26 | conventional_implant | wound dehiscence 0/13 vs. 1/13; immediate occlusal disharmony 0/13 vs. 2/13 | 4 weeks |
| Maini et al. (2018) [72] | RCT | pelvis_acetabulum | acetabular fixation plate | comp | 25 | conventional_implant | total blood loss lower than conventional contouring [authors’ qualitative statement; no figures] | . |
| Shuang et al. (2016) [73] | RCT | upper_limb_other | custom osteosynthesis plate | imaging + comp | 13 | conventional_implant | bone union 13/13; good or excellent elbow function 83.1% vs. 71.4%; no wound infections | 10.6 months mean (range 6–13 months) |
| Ciocca et al. (2015) [74] | Comparative | cmf | mandibular reconstruction plate | comp | 10 | conventional_implant | . | . |
| Tarsitano et al. (2016) [75] | Comparative | cmf | customised mandibular reconstruction plate | imaging + comp | 30 | conventional_implant | . | . |
| Tarsitano et al. (2016) [76] | Comparative | cmf | mandibular reconstruction plate | comp | 40 | conventional_implant | microvascular complications 0/20 vs. 2/20; mean hospital stay 13.8 vs. 17 days | . |
| Zimmerer et al. (2016) [77] | Comparative | cmf | orbital reconstruction implant | comp | 195 | conventional_implant | individualisation and navigation provided clinical benefit [authors’ qualitative statement; no figures] | 12 weeks |
| De Maesschalck et al. (2017) [78] | Comparative | cmf | mandibular reconstruction plate | imaging + comp | 18 | conventional_implant | CAS and freehand reconstruction were morphologically comparable except nonaffected-side axial angle (1.0° vs. 2.9°; p = 0.03) | . |
| Kim et al. (2017) [79] | Comparative | cmf | pre-bent orbital-wall mesh | imaging | 82 | conventional_implant | . | . |
| Fan et al. (2017) [80] | Comparative | cmf | orbital Medpor-titanium implant | imaging | 56 | manual_design | postoperative clinical evaluation superior to control [authors’ qualitative statement; no figures] | . |
| Kim et al. (2018) [81] | Comparative | cmf | orbital floor and medial wall implant | imaging | 28 | alternative_method | diplopia improved in 9/10 patients; moderate-to-severe enophthalmos improved in 18/18 patients | 6 months |
| Rückschloß et al. (2019) [82] | Comparative | cmf | maxillary patient-specific fixation implant | comp | 18 | alternative_method | . | . |
| Kim et al. (2019) [83] | Comparative | cmf | orbital wall implant | imaging + comp | 56 | alternative_method | . | . |
| Liu et al. (2020) [84] | Comparative | cmf | maxillary osteosynthesis plate | comp | 18 | alternative_method | . | . |
| Sigron et al. (2020) [85] | Comparative | cmf | hybrid orbital floor mesh | imaging + comp | 22 | conventional_implant | postoperative stay 4.6 (3.9) d vs. 3.8 (3.0) d; no significant difference | . |
| Lee et al. (2020) [86] | Comparative | cmf | mandibular reconstruction plate | comp | 126 | conventional_implant | hardware complications 5.5% vs. 15.5%; p = 0.091; customised plates independently predicted fewer overall complications p = 0.03 | 23.5 months mean |
| Badiali et al. (2020) [87] | Comparative | cmf | mandibular osteosynthesis plate | comp | 22 | alternative_method | system successful in 22/22 patients; outcome clinically satisfactory in 22/22 | 14 months average (range 5–29 months) |
| Rahimov et al. (2020) [88] | Comparative | cmf | customised orbital plate | imaging + comp | 77 | conventional_implant | residual diplopia 4/42 (9.5%) vs. 12/35 (34.3%) at 6 months | 6 months |
| Yang et al. (2021) [89] | Comparative | cmf | maxillary or mandibular reconstruction plate | comp | 33 | conventional_implant | plate removal 0/17 vs. 2/16; no major postoperative adverse events in either group | CT follow-up 1.8 ± 0.8 months vs. 17.3 ± 11.2 months |
| Olate et al. (2021) [90] | Comparative | cmf | mandibular angle contour implant | comp | 21 | conventional_implant | infections in both groups; p > 0.05 | 6 months |
| Karanxha et al. (2021) [91] | Comparative | cmf | maxillary and mandibular fixation plates | comp | 16 | alternative_method | . | . |
| Neuhaus et al. (2021) [92] | Comparative | cmf | patient-specific total TMJ prosthesis | imaging + comp | 21 | alternative_method | mouth opening 21.69 ± 2.80 mm preop to 36.40 ± 1.25 mm; pain 6.18 ± 0.74 to 1.06 ± 0.52 | 15.3 ± 3.0 months mean |
| Knitschke et al. (2021) [93] | Comparative | cmf | custom osteosynthesis plate | comp | 104 | conventional_implant | bone R0 margins 95.1% VSP vs. 96.8% non-VSP (p = 0.516); soft-tissue R0 margins 63.4% vs. 73.0% (p = 0.470) | . |
| Sigron et al. (2021) [94] | Comparative | cmf | hybrid patient-specific orbital floor mesh | imaging + comp | 30 | conventional_implant | complete diplopia resolution 9/12 vs. 5/7; no major complications | 6 months mean (range 1 day to almost 2 years) |
| Kang et al. (2021) [95] | Comparative | cmf | pre-bent midface reconstruction plate | imaging + comp | 34 | conventional_implant | dental implantation 50.0% vs. 11.1%; all patients satisfied with facial symmetry | . |
| Möllmann et al. (2021) [96] | Comparative | cmf | mandibular reconstruction plate | comp | 94 | conventional_implant | complications 33/46 PSI vs. 31/48 CR; no significant difference | . |
| Harding et al. (2022) [97] | Comparative | cmf | custom mandibular osteosynthesis plate | comp | 55 | conventional_implant | . | . |
| Timoshchuk et al. (2022) [98] | Comparative | cmf | orbital reconstruction implant | comp | 85 | conventional_implant | complications 8.3% vs. 26.2% (p = 0.08) | . |
| Jones et al. (2022) [99] | Comparative | cmf | maxillary osteosynthesis plate | comp | 82 | alternative_method | . | . |
| Liokatis et al. (2022) [100] | Comparative | cmf | midfacial osteosynthesis implant | imaging + comp | 8 | alternative_method | no major complications; plate removal 2/8 | 13–79 months |
| Rodríguez-Arias et al. (2022) [101] | Comparative | cmf | mandibular reconstruction plate | comp | 37 | conventional_implant | 5% fewer complications with CAD/CAM; flap survival 100% | . |
| Chepurnyi et al. (2022) [102] | Comparative | cmf | orbital wall patient-specific implant | imaging + comp | 50 | manual_design | no implant malposition or enophthalmos; diplopia 13.0% vs. 11.1% at 3 months | 3 months |
| Cassoni et al. (2022) [103] | Comparative | cmf | maxillary fixation plate | imaging + comp | 10 | conventional_implant | . | 1 month |
| Spalthoff et al. (2022) [104] | Comparative | cmf | mandibular reconstruction plate | imaging + comp | 83 | alternative_method | surgery postponed in 16/83; 15/16 postponed cases were complex, p = 0.001 | . |
| Kerkfeld et al. (2022) [105] | Comparative | cmf | orthognathic plates and facial augmentation implants | imaging + comp | 30 | alternative_method | PEEK implant exposure 2/8; no other PEEK complications | at least 9 months |
| Knitschke et al. (2022) [106] | Comparative | cmf | jaw reconstruction plate | comp | 133 | conventional_implant | incomplete osseous union 26.6% PSI vs. 8.7% non-PSI at patient level; adjusted OR 3.518 (95% CI 1.223–10.124), p = 0.02 | 54 months mean to final scan (range 2–216) |
| Landau Prat et al. (2023) [107] | Comparative | cmf | orbital floor implant | imaging | 27 | conventional_implant | primary or vertical gaze diplopia improved in both groups (p = 0.03) | 1.1 years mean |
| Kim et al. (2023) [108] | Comparative | cmf | orthognathic fixation plate | comp | 12 | conventional_implant | plate exposure requiring replacement 1/6 customised patients | 6 months |
| Pietzka et al. (2023) [109] | Comparative | cmf | maxillary osteosynthesis plate | comp | 126 | conventional_implant | dental root injuries 10/651 screws in conventional cohort vs. 0/773 PSI screws (p < 0.001) | 13 months mean |
| Pietzka et al. (2023) [110] | Comparative | cmf | orbital implant | imaging + comp | 75 | conventional_implant | intraoperative revision 26.6% vs. 13%; postoperative revision 11.1% vs. 0% PSI | . |
| Kuehle et al. (2023) [111] | Comparative | cmf | maxillary osteosynthesis plate | imaging + comp | 50 | conventional_implant | . | first postoperative week |
| Block et al. (2024) [112] | Comparative | cmf | mandibular reconstruction plate | imaging + comp | 29 | alternative_method | . | . |
| Merta et al. (2024) [113] | Comparative | cmf | mandibular osteotomy fixation plate | comp | 53 | conventional_implant | no difference in skeletal stability between VSP-PSI and conventional mini-plate groups [authors’ qualitative statement; no figures] | . |
| Consorti et al. (2024) [114] | Comparative | cmf | customised orbital implant | imaging + comp | 52 | conventional_implant | clinical improvements and complications similar between groups [authors’ qualitative statement; no figures] | . |
| Krasovsky et al. (2024) [115] | Comparative | cmf | orbital reconstruction implant | imaging + comp | 23 | conventional_implant | . | . |
| D’Alpaos et al. (2024) [116] | Comparative | cmf | orbital floor implant | imaging + comp | 23 | conventional_implant | no implant-related complications or reoperations in 13 PSI patients | 30.69 months mean PSI; 25.30 months mean control |
| Ettinger et al. (2024) [117] | Comparative | cmf | mandibular reconstruction plate | comp | 130 | conventional_implant | . | . |
| Yildirim et al. (2026) [118] | Comparative | cmf | bimaxillary orthognathic fixation implant | comp | 46 | alternative_method | no conversion to splint-based transfer 0/46 | ≤4 weeks |
| Le et al. (2025) [119] | Comparative | cmf | mandibular reconstruction plate | comp | 18 | conventional_implant | . | . |
| Mohamed et al. (2025) [120] | Comparative | cmf | mandibular reconstruction plate | comp | 265 | conventional_implant | hardware-related complications in 72/265; stock plates vs. patient-specific implants HR 1.80, 95% CI 1.11–2.91, p = 0.0171 | ≥1 year |
| Chen et al. (2025) [121] | Comparative | cmf | absorbable orbital reconstruction plate | imaging + comp | 94 | manual_design | no infection, implant displacement, or vision deterioration; superior diplopia resolution p = 0.023 and ocular mobility restoration p = 0.049 | 6–11 months |
| Ren et al. (2025) [122] | Comparative | cmf | absorbable orbital wall plate | imaging + comp | 106 | conventional_implant | residual diplopia 9.2% vs. 31.6%; no implant-related infection or displacement in absorbable group | 6 months minimum |
| Ismailov et al. (2026) [123] | Comparative | cmf | pre-contoured mandibular reconstruction plate | comp | 12 | other | hardware destabilisation 1/12 | . |
| Zoabi et al. (2026) [124] | Comparative | cmf | orbital floor implant | imaging + comp | 23 | alternative_method | orbital volume 108.6% to 99.5% with mirroring and 107.3% to 101.1% with virtual reduction; diplopia resolved in most cases | . |
| Adell-Gómez et al. (2026) [125] | Comparative | cmf | orthognathic fixation plates | imaging + comp | 90 | conventional_implant | mandibular plates replaced intraoperatively in 1/15 PSI patients | . |
| Troise et al. (2026) [126] | Comparative | cmf | orbital floor mesh | comp | 52 | conventional_implant | persistent diplopia 1/24 (4%) vs. 8/28 (28.6%); p = 0.028 | 12 months |
| Yang et al. (2026) [127] | Comparative | cmf | mandibular reconstruction plate | imaging + comp | 111 | conventional_implant | plate-related complications 43/110; plate exposure 24/110 | 59.8 months mean |
| Mäntynen et al. (2026) [128] | Comparative | cmf | orthognathic fixation plate | imaging + comp | 40 | alternative_method | reoperation for postoperative malocclusion 1/40 | . |
| Prikule et al. (2026) [129] | Comparative | cmf | orbital floor mesh | imaging + comp | 74 | conventional_implant | reoperation 5/34 control vs. 0/40 3D, p = 0.017; persistent diplopia 4/34 vs. 1/40 at 6 months | 6 months |
| Aguilar et al. (2026) [130] | Comparative | cmf | customised maxillary fixation plate | comp | 64 | alternative_method | instability 28.6% minimally invasive vs. 33.3% conventional; OR 1.23, 95% CI 0.32–4.65 | 6 months |
| Rouhiainen et al. (2026) [131] | Comparative | cmf | maxillary fixation plate | comp | 59 | conventional_implant | . | 12.4–13.2 months median |
| De Oliveira Gorla et al. (2026) [132] | Comparative | cmf | orbital implant | comp | 10 | alternative_method | clinical success 6/9; persistent diplopia 3/9 | . |
| Bertram et al. (2026) [133] | Comparative | cmf | patient-specific maxillary osteosynthesis plate | imaging + comp | 26 | conventional_implant | . | . |
| Desai et al. (2019) [134] | Comparative | cranial | cranioplasty implant | comp | 30 | alternative_method | infection 13.3%; patient satisfaction excellent in 69%; 1 resurgery in alginate-impression group | . |
| Baldia et al. (2022) [135] | Comparative | cranial | cranioplasty prosthesis | imaging | 70 | alternative_method | 3D-printed prostheses had better cosmetic scores than hand-moulded and bone-impression prostheses, p < 0.001 | . |
| Kim et al. (2023) [136] | Comparative | cranial | cranioplasty implant | imaging | 35 | conventional_implant | complications 2/12 vs. 6/23; reoperation 2/12 vs. 5/23 | . |
| Kim et al. (2026) [137] | Comparative | cranial | cranioplasty flap | imaging | 36 | conventional_implant | CSF collection 0/16 modified vs. 2/20 control; haemorrhage 0/16 vs. 2/20 control | 1 month |
| Zhao et al. (2024) [138] | Comparative | hip | custom hip joint prosthesis | comp | 33 | conventional_implant | 33/33 had no infection, deep vein thrombosis, or wound nonunion; implants stable without loosening or osteolysis at 6 months | 6 months |
| Hu et al. (2022) [139] | Comparative | knee | strut-type bone-defect prosthesis | comp | 10 | conventional_implant | MSTS 28.4 ± 1.8 versus 26.2 ± 2.8, p = 0.176; all 5 prosthesis patients achieved bone graft fusion | 24.7 ± 2.4 months mean |
| Liu et al. (2019) [140] | Comparative | pelvis_acetabulum | pelvic reconstruction implant | comp | 38 | alternative_method | lower tumour recurrence but higher implant loosening than common 3D-printed template | . |
| Pu et al. (2021) [141] | Comparative | pelvis_acetabulum | sacroiliac reconstruction prosthesis | comp | 24 | conventional_implant | MSTS-93 24.1 ± 2.8 vs. 18.9 ± 2.6; complications 1/12 vs. 1/12; disease-free survival 9/12 vs. 9/12 | 21 months median for 3D-printing and 20 months median for control; range 6–90 months |
| Hu et al. (2024) [142] | Comparative | pelvis_acetabulum | hemipelvic endoprosthesis | comp | 32 | conventional_implant | MSTS 24.3 ± 1.8 vs. 21.8 ± 2.0; HHS 79.8 ± 5.2 vs. 75.3 ± 3.5 | 41.6 months mean |
| Zhang et al. (2024) [143] | Comparative | pelvis_acetabulum | acetabular augment/triflange cup | comp | 109 | conventional_implant | . | 6.0 ± 1.3 years augment; 2.8 ± 0.6 years triflange |
| Huang et al. (2025) [144] | Comparative | pelvis_acetabulum | hemipelvic prosthesis | imaging + comp | 28 | alternative_method | implant-related complications: breakage 4/28, aseptic loosening 4/28, dislocation 2/28; MSTS-93 18.2 ± 5.3 | 36.2 months mean |
| Guo et al. (2022) [145] | Comparative | shoulder | scapular fixation plate | comp | 21 | conventional_implant | no complications in 21/21; follow-up functional outcome similar between groups | . |
| Lv et al. (2025) [146] | Comparative | spine | artificial vertebral body | comp | 21 | conventional_implant | 2-yr VAS 0.7 ± 0.6 vs. 1.8 ± 0.6; fusion 100% vs. 38.9%; rod fracture 0/3 vs. 6/18 | 38.5 months mean |
| Van Doremalen et al. (2021) [147] | Comparative | upper_limb_other | contoured clavicle fixation plate | imaging + comp | 14 | conventional_implant | discomfort-related implant removal 0/6 vs. 5/7; p = 0.012; infection 1/7 | 4 years minimum |
| Gong et al. (2023) [148] | Comparative | upper_limb_other | distal radius tumour endoprosthesis | comp | 22 | alternative_method | subluxation rate significantly lower with 0° than biological-angle design | 60 months median (range 19–82) |
cmf, craniomaxillofacial; comp, computational planning; imaging, evaluated imaging step; CAD/CAM, computer-aided design and manufacturing; CAS, computer-assisted surgery; CI, confidence interval; CR, conventional reconstruction plate; CSF, cerebrospinal fluid; CT, computed tomography; HHS, Harris Hip Score; HR, hazard ratio; MSTS, Musculoskeletal Tumor Society score; OR, odds ratio; PEEK, polyetheretherketone; PSI, patient-specific implant; R0, microscopically clear resection margin; RCT, randomised controlled trial; TMJ, temporomandibular joint; VAS, visual analogue scale; VSP, virtual surgical planning. Values are reproduced as reported by the original authors. Region and comparator labels (e.g., pelvis_acetabulum, conventional_implant) are the charting vocabulary defined in Text S4; a full stop (.) indicates not reported.
The prespecified meta-analysis condition was tested on a dedicated extraction of arm-level estimates, one row per outcome, from the 40 included comparative studies and trials with full text, of which 26 reported at least one listed outcome for both arms (Table 5; the outcome-by-study disposition is Table S6). Membership was taken from the charted comparator class (conventional implant) and, for accuracy, from the measure (deviation of the delivered result from the preoperative plan; clinical symmetry measures were not pooled), with one measure per study, preferring an overall or total deviation. Deviation from the plan was extractable from 7 studies (1 randomised, 6 comparative; 243 patients), and pooling those 7 gives a standardised mean difference of −0.72 (95% CI −1.32 to −0.13) favouring the patient-specific arm, with I-squared 76%. That estimate was the one reported in the submitted version, and examining it at revision showed that it should not be read, because the pool mixed measures of different things. Two adjudicators read each of the seven papers’ methods. In one study the measure compares postoperative with preoperative anatomy and never references the plan at all, and in a second the adjudicators disagreed about whether it does; both are therefore outside a pool defined as deviation from plan. In two further studies, the deviation is reported as a signed value with zero as the target, so the arm that sat closer to the plan carries the numerically larger mean: one of these entered the submitted pool contributing +0.32, that is, in favour of the conventional implant, although its own authors report the patient-specific arm as more accurate. Restricting the pool to the five studies whose measure references the virtual plan gives −0.85 (−1.69 to −0.01) with I-squared 78%, and restricting it further to the four in which a smaller value also means a better result gives −1.21 (−1.69 to −0.74) with I-squared 0%. The heterogeneity that made the original estimate hard to interpret was therefore largely an artefact of pooling incommensurable measures rather than a property of the studies. We report −1.21 from 4 studies as the primary estimate, all of it craniomaxillofacial, and the per-study adjudication with its quoted evidence is in Text S6. The certainty of this estimate remains very low, and its direction should not be read as established: four small non-randomised or single-trial comparisons agreeing with one another is a weak basis for a claim about accuracy. Operative time was extractable from 11 studies (1 randomised, 10 comparative; 384 patients): the pooled mean difference was −11.8 min (95% CI −30.5 to 6.9) with I-squared 93%, a Hartung–Knapp interval of −46.3 to 22.7, a 95% prediction interval of −74.6 to 51.0 min and leave-one-out estimates from −24.0 to −7.3 (Figure 3); direction varied within and between regions, with pelvic reconstructions and one mandibular study longer in the patient-specific arm, and the Egger intercept was −1.26 (p = 0.601). Revision or removal of the implant was extractable from 8 comparative studies (284 patients), none of them randomised: the pooled risk ratio was 0.28 (95% CI 0.11 to 0.72) favouring the patient-specific arm, with I-squared 0%, a Hartung–Knapp interval of 0.12 to 0.66 and a prediction interval of 0.09 to 0.91. In absolute terms, 3 of 135 patients in the patient-specific arms were revised or had the implant removed (2.2%) against 25 of 149 in the conventional arms (16.8%), a difference of 14.6 percentage points (95% CI 8.1 to 21.1), or about 146 fewer revisions per 1000 patients treated, which corresponds to one revision avoided for every 7 patients. Those absolute figures come entirely from non-randomised studies in which the two arms were assigned by period, surgeon or case complexity without adjustment, so the certainty is very low and the true difference could be substantially smaller, absent or in the opposite direction. Any complication was extractable from 8 studies (1 randomised; 242 patients), risk ratio 1.05 (95% CI 0.8 to 1.36), 41 of 126 against 41 of 116 patients, an absolute difference of −2.8 percentage points (95% CI −14.7 to 9.1); and infection from 6 (2 randomised; 133 patients), risk ratio 0.94 (95% CI 0.3 to 2.99), 3 of 66 against 3 of 67, a difference of 0.1 percentage points (95% CI −7.0 to 7.1). Neither interval excludes a difference in either direction, and with 6 to 8 small studies and few events, these are findings of imprecision rather than evidence that the arms are alike. The same applies to operative time, where the interval runs from 30.5 min saved to 6.9 min lost: the review cannot tell whether patient-specific implants shorten operating time and does not show that they do not. Blood loss and function were reported for both arms in 5 and 4 studies respectively but without extractable dispersion, so they were not pooled. All pooled estimates describe the delivered patient-specific implant against a conventional one; no study compared one design pipeline against another on the same implant type.
Table 5.
Prespecified meta-analysis test (amendment A8): outcomes reported for both arms against a conventional comparator.
| Outcome | Studies with Extractable Estimates | Pooled Estimate (95% CI) | 95% Prediction Interval | I-Squared | Sensitivity (HKSJ CI; Leave-One-Out Range; Small-Study Test) | Disposition |
|---|---|---|---|---|---|---|
| accuracy | 7 (RCT 1, comparative 6) | −0.72 (−1.32 to −0.13), standardised mean difference (Hedges g) | −2.6 to 1.16 | 76% | HKSJ −1.42 to −0.03; leave-one-out −0.94 to −0.53; small-study test not performed (fewer than 10 studies) | pooled; 9 included studies reported the outcome for both arms; 7 against a comparator charted as a conventional implant and measured as deviation of the delivered result from the plan |
| complications any | 8 (RCT 1, comparative 7) | 1.05 (0.8 to 1.36), risk ratio (pooled on the log scale) | 0.75 to 1.45 | 0% | HKSJ 0.8 to 1.37; leave-one-out 0.63 to 1.11; small-study test not performed (fewer than 10 studies) | pooled; 9 included studies reported the outcome for both arms; 8 against a comparator charted as a conventional implant |
| infection | 6 (RCT 2, comparative 4) | 0.94 (0.3 to 2.99), risk ratio (pooled on the log scale) | 0.18 to 4.85 | 0% | HKSJ 0.51 to 1.74; leave-one-out 0.77 to 1.39; small-study test not performed (fewer than 10 studies) | pooled; 8 included studies reported the outcome for both arms; 6 against a comparator charted as a conventional implant |
| operative time | 11 (RCT 1, comparative 10) | −11.8 (−30.5 to 6.9), mean difference (minutes) | −74.6 to 51.0 | 93% | HKSJ −46.3 to 22.7; leave-one-out −24.0 to −7.3; Egger intercept −1.26, p = 0.601 | pooled; 14 included studies reported the outcome for both arms; 13 against a comparator charted as a conventional implant |
| revision or removal | 8 (RCT 0, comparative 8) | 0.28 (0.11 to 0.72), risk ratio (pooled on the log scale) | 0.09 to 0.91 | 0% | HKSJ 0.12 to 0.66; leave-one-out 0.22 to 0.33; small-study test not performed (fewer than 10 studies) | pooled; 9 included studies reported the outcome for both arms; 8 against a comparator charted as a conventional implant |
| blood loss | . | not pooled | . | . | . | 5 included studies reported the outcome for both arms; 4 against a comparator charted as a conventional implant; 0 with extractable estimates against a conventional comparator, below the prespecified minimum of 3 |
| function | . | not pooled | . | . | . | 4 included studies reported the outcome for both arms; 4 against a comparator charted as a conventional implant; 0 with extractable estimates against a conventional comparator, below the prespecified minimum of 3 |
CI, confidence interval; HKSJ, Hartung–Knapp–Sidik–Jonkman; I-squared, percentage of variability attributable to between-study heterogeneity; RCT, randomised controlled trial. Mean differences and standardised mean differences (Hedges g) below 0, and risk ratios below 1, favour the patient-specific implant. The accuracy row is the prespecified seven-study pool; the primary estimate, restricted to the four studies that measure deviation from the virtual plan with smaller meaning better, is −1.21 (95% CI −1.69 to −0.74; I-squared 0%; Section 3.6 and Text S6). A full stop (.) indicates not applicable.
Figure 3.

Forest plot of operative time. Difference in operative time (minutes) between the patient-specific implant arm and the conventional comparator arm in the studies with extractable arm-level estimates (amendment A8), with the DerSimonian–Laird random-effects pooled estimate and its 95% prediction interval. Squares are individual studies with 95% confidence intervals (grey, comparative non-randomised studies; blue, the randomised trial); the orange diamond is the pooled estimate and the orange bar its 95% prediction interval; the dashed vertical line marks no difference. Studies, top to bottom: Le 2025 [119], Knitschke 2021 [93], Pu 2021 [141], Sigron 2020 [85], Lv 2025 [146], Sigron 2021 [94], Shuang 2016 [73], Kim 2023 [108], van Doremalen 2021 [147], Möllmann 2021 [96] and Hu 2024 [142]. The comparison concerns the delivered implant, not a design method.
3.7. Risk of Bias, Technical Credibility and Certainty
All 266 clinical-tier studies were appraised twice, independently, by assessors from different vendors. The two agreed on the overall judgement for 75.0% of the 12 randomised trials, 86.7% of the 75 comparative studies and, for the case series, 32.4% exactly and 60.3% within one item of the ten. Where they differed on the comparative studies, the second assessor was the more severe: the first recorded 2 studies at low and 4 at moderate risk, with 69 at serious; the second recorded none below serious and placed 4 at critical. Both therefore reach the same conclusion about the body of evidence: that almost every non-randomised comparison is at serious risk of bias from confounding, and neither supports a claim that any particular study is sound. Agreement on the case-series checklist is poor, and the median item count reported below should be read as an indication rather than a measurement. Cohen’s kappa is not quoted for the comparative studies because one category accounts for almost every judgement, which drives kappa towards zero despite high observed agreement; the full distribution of each assessor’s judgements is given instead, and every disagreement is listed per study in Table S4. The figures below are the first assessor’s, with the second assessor’s in Table S4 (Table 6; per-study judgements in Table S4). Of the 12 randomised trials, 10 were at some concerns and 2 at high risk of bias under RoB 2; none were at low risk because none reported a prespecified analysis plan (selection of the reported result was at some concerns in all 12) and randomisation was adequately described in 2. Of the 75 comparative non-randomised studies, 69 were at serious, 4 at moderate and 2 at low overall risk under ROBINS-I; confounding was serious in 68, because patient-specific and conventional implants were assigned by period, surgeon or case complexity without adjustment, and 68 gave no information on selection of the reported result. The 179 non-comparative series met a median of 5 of the 10 JBI items; 100 met five or fewer, and 35 met eight or more, and the items least often met were consecutive inclusion (35 series) and complete inclusion (31), against 169 that reported outcomes clearly. Judgements for abstract-only records rest on absence of information and are labelled as such in Table S4. Funding and conflict statements could be read only in the full-text records: a funding source was stated in 34 studies as public, in 4 as industry, in 5 as mixed and in 19 as none, a conflict of interest was declared in 9 and denied in 57, and an author with a manufacturer or commercial design-service affiliation was identified in 5 (Table S5); for the remaining full-text records, no statement was found, and abstract-only records are unclear. Technical credibility is summarised in 3.5 and per study in Table S3. Certainty of evidence per outcome is set out in Table 7: very low for deviation from the plan, because the body of evidence is mostly non-randomised, at serious risk of bias, measured in different ways across studies and indirect to the review question; very low for operative time, with very serious inconsistency added; very low for revision or removal, any complication and infection, where events are few, definitions differ, and reporting is incomplete; and very low for the clinical outcomes outside the craniomaxillofacial region, where comparative and randomised studies exist but are sparse and unpooled.
Table 6.
Risk of bias of the clinical-tier studies.
| Tool (Tier) | Studies Appraised | Overall Judgement Distribution |
|---|---|---|
| RoB 2 (randomised trials) | 12 | high: 2; some concerns: 10 |
| ROBINS-I (comparative non-randomised) | 75 | low: 2; moderate: 4; serious: 69 |
| JBI case series (non-comparative series) | 179 | items met, median 5 of 10; 5 or fewer items met: 100; 8 or more: 35; distribution 0/10: 3; 1/10: 18; 2/10: 32; 3/10: 18; 4/10: 17; 5/10: 12; 6/10: 19; 7/10: 25; 8/10: 20; 9/10: 9; 10/10: 6 |
Table 7.
Certainty of evidence per clinical outcome (GRADE domains; the comparison is the delivered patient-specific implant against a conventional comparator).
| Outcome | Studies (Participants) | Effect | Risk of Bias | Inconsistency | Indirectness | Imprecision | Publication Bias | Certainty |
|---|---|---|---|---|---|---|---|---|
| Deviation of the delivered implant from the preoperative plan (patient-specific versus conventional) | 4 (81) | −1.21 (−1.69 to −0.74); prediction −2.26 to −0.17; standardised mean difference (Hedges g) | serious: the three comparative studies at serious risk under ROBINS-I; the trial at some concerns under RoB 2 | not serious: I-squared 0% once the pool is restricted to commensurable measures; surface and landmark measures still differ across studies | serious: the comparison evaluates the implant as delivered, not a design method; accuracy measures differ across studies | serious: four small studies with 81 participants in total | not assessed (fewer than ten studies; small-study tests underpowered) | Very low |
| Operative time (patient-specific versus conventional) | 11 (384) | −11.8 (−30.5 to 6.9); prediction −74.6 to 51.0; mean difference (minutes) | serious: confounding by period, surgeon and case complexity | very serious: I-squared about 90%, prediction interval spans large savings and large excesses; direction varies within and between regions | serious: implant as delivered, not design method | serious: confidence interval includes no difference | Egger test reported in Table 5 (ten or more studies); no downgrade applied because the test is weak with this heterogeneity | Very low |
| Any complication (patient-specific versus conventional) | 8 (242) | 1.05 (0.8 to 1.36); prediction 0.75 to 1.45; risk ratio (pooled on the log scale) | serious: confounding by period, surgeon and case complexity; selective reporting of events | not serious statistically (I-squared 0%) but event definitions differ | serious: implant as delivered, not design method | serious: interval includes both benefit and harm | not assessed (fewer than ten studies) | Very low |
| Infection (patient-specific versus conventional) | 6 (133) | 0.94 (0.3 to 2.99); prediction 0.18 to 4.85; risk ratio (pooled on the log scale) | serious: confounding by period, surgeon and case complexity; selective reporting of events | not serious statistically (I-squared 0%) but event definitions differ | serious: implant as delivered, not design method | very serious: few events, interval from a third to three times | not assessed (fewer than ten studies) | Very low |
| Revision or removal of the implant (patient-specific versus conventional) | 8 (284) | 0.28 (0.11 to 0.72); prediction 0.09 to 0.91; risk ratio (pooled on the log scale) | serious: confounding by period, surgeon and case complexity; selective reporting of events | not serious statistically (I-squared 0%) but event definitions differ | serious: implant as delivered, not design method | serious: few events; interval excludes no effect but rests on eight non-randomised studies | not assessed (fewer than ten studies) | Very low |
| Clinical outcomes of computationally designed implants outside the craniomaxillofacial region (cranial, pelvic, limb and spinal) | mostly non-comparative series; a minority of comparative studies and two randomised trials (pelvis, distal humerus), no pooled outcome | not pooled | serious: comparative studies at serious risk from confounding; series rarely report consecutive or complete inclusion | not estimable (no pooled outcome) | serious | very serious | not assessed | Very low |
Domain-level judgements. RoB 2 (12 trials): randomisation, low 2, some concerns 10; deviations from intended interventions, low 2, some concerns 9, high 1; missing outcome data, low 5, some concerns 6, high 1; measurement of the outcome, low 5, some concerns 7; selection of the reported result, some concerns 12. ROBINS-I (75 studies): confounding, low 1, moderate 5, serious 68, no information 1; selection of participants, low 9, moderate 19, serious 16, no information 31; classification of interventions, low 75; deviations from intended interventions, low 9, moderate 5, serious 1, no information 60; missing data, low 17, moderate 5, serious 10, no information 43; measurement of outcomes, low 9, moderate 64, serious 1, no information 1; selection of the reported result, moderate 6, serious 1, no information 68. JBI checklist items met (179 series): Q1 96, Q2 84, Q3 77, Q4 35, Q5 31, Q6 83, Q7 116, Q8 169, Q9 61, Q10 102 (item wording in Text S4). JBI, Joanna Briggs Institute; RoB 2, Cochrane risk-of-bias tool for randomised trials, version 2; ROBINS-I, Risk Of Bias In Non-randomised Studies of Interventions.
Starting level: The starting level was high for randomised bodies and low for non-randomised or mixed bodies (the non-randomised studies carry most of the weight). Deviation of the delivered implant from the preoperative plan: Primary estimate from the four studies that measure deviation from the virtual plan with smaller meaning better (Section 3.6 and Text S6); the prespecified seven-study pool, −0.72 (−1.32 to −0.13), is given in Table 5. The pooled standardised mean difference favours the patient-specific arm, but its applicability to the design methods is indirect. Operative time: Direction differs by region; no summary claim is warranted. Any complication: Event outcomes were pooled from a single extraction pass; reporting of adverse events was incomplete across the corpus. Infection: Event outcomes were pooled from a single extraction pass; reporting of adverse events was incomplete across the corpus. Revision or removal of the implant: Event outcomes were pooled from a single extraction pass; reporting of adverse events was incomplete across the corpus. Clinical outcomes of computationally designed implants outside the craniomaxillofacial region: Comparative and randomised evidence exists outside the craniomaxillofacial region but is sparse and unpooled; the series predominate.
3.8. How Much Lies Outside the Searched Databases?
The review searched biomedical indexes and OpenAlex and not Embase, Scopus, Web of Science, IEEE Xplore or CINAHL. Two measurements bound what that omits.
The first uses the review’s own data. OpenAlex, the one source that is not biomedical-only, was the sole route to 68 of the 639 included studies, or 10.6%. Those 68 are overwhelmingly technical: 57 sit at the in silico, bench, cadaver or imaging-dataset tiers against 11 at clinical tiers, and 11 carry an artificial intelligence pillar. Adding one broader index therefore changed the technical map appreciably and the clinical map very little.
The second measurement probes the literature the review did not reach. The review’s own concept blocks were run against engineering-type sources in OpenAlex, retrieving 989 unique records, of which 379 were already in this review’s corpus, and 610 were not. A random sample of 40 of the 610 was screened against the eligibility criteria: 2 were clearly eligible, 34 clearly not, and 4 could not be settled from the title and abstract. Taking only the clear includes, 5.0% (95% CI 1.4% to 16.5%) of the absent records would have been eligible, which projects to about 31 further studies (95% CI 8 to 101); counting the uncertain records as eligible raises that to about 92 (43 to 177). Both estimates are small beside 639, and both consist of technical work: the two clear includes were a topology-optimised mandibular plate and an accuracy study of additively manufactured bone plates. The probe found no sign of a hidden body of randomised or comparative clinical evidence in engineering venues.
The conclusion we draw is specific. The review’s clinical findings, which rest on randomised and comparative studies indexed in MEDLINE, are unlikely to be changed by the missing databases. Its technical counts are an undercount, probably by something between 1% and 15%, concentrated in finite element and optimisation work, so statements such as “finite element analysis appeared in 312 studies” should be read as a floor. This is why the title and the aims of this review describe the biomedical-indexed literature rather than the field as a whole.
4. Discussion
4.1. Principal Findings
Across 639 studies published since 2015, the evidence for the three pillars of patient-specific implant design is deep in one place and shallow almost everywhere else. Virtual surgical planning, usually with mirroring or registration, is the only method whose evidence reaches the randomised tier, and it does so mostly in craniomaxillofacial surgery, in small trials that evaluate the delivered implant rather than the planning method. Pooled across the four studies that measure deviation from the virtual plan in a direction where smaller is better, the delivered patient-specific implant deviated less from its plan than a conventional implant did, and across eight comparative studies it was revised or removed less often, namely 3 of 135 against 25 of 149 patients; operative time, any complication and infection gave intervals too wide to tell whether a difference exists. Every estimate is of very low certainty, every one of them comes from craniomaxillofacial surgery, and none compare one design method with another. Finite element analysis is the most frequent method in the corpus but is applied overwhelmingly in single-anatomy in silico studies; where the text can be read, inputs are usually specified, but independent validation material, comparison against a physical experiment and availability of model or design files are usually absent. Topology, lattice and shape optimisation, automated design, musculoskeletal modelling and statistical shape modelling remain almost entirely technical. Artificial intelligence contributes 36 studies, none comparative at the clinical tier, 2 in which a model was used in the care of a real patient, 4 with evaluation on data independent of the development source, and 6 with public code; favourable offline reconstruction or segmentation metrics are not evidence of clinical performance, and on the present literature no claim about clinical benefit from artificial intelligence in implant design can be supported at all; its strongest evidence is benchmark performance for cranial defect completion, a task made tractable by the AutoImplant datasets. Outside the craniomaxillofacial region, comparative and randomised evidence for computationally designed patient-specific implants exists in the pelvis, the cranial vault and the upper limb but is sparse, and most of the clinical evidence there rests on series that do not report consecutive or complete inclusion.
4.2. Relation to Prior Reviews
The region-specific reviews and registrations identified in the registry search ask whether patient-specific implants perform as well as conventional ones in cranioplasty, the orbit, orthognathic surgery, the jaws, the pelvis, the spine and long bones [1,3,5,6,7,8,9,11,12]. Their positive answers are compatible with the present map and with its pooled plan-deviation estimate: the same craniomaxillofacial trials and comparative studies populate our clinical tier. What those reviews do not show is that the implants they evaluate were produced by a pipeline whose components are validated to very different degrees and that outside their regions the same components have little clinical evaluation. The nearest AI-oriented review on three-dimensional templating for arthroplasty [149] and the deep learning cranial implant design study that frames that task as a benchmark problem [14] are each confined to one task; the present review places those tasks in the wider design pipeline and shows that the deficit of independent and clinical evaluation is general. Reviews of statistical shape model accuracy [18] and the credibility framework for computational models [15] provide the standards against which the technical studies here fall short.
4.3. Strengths and Limitations
The review’s strengths are procedural: a protocol committed before retrieval with every deviation dated; a recall-oriented four-block search with a rule layer whose exclusions are published and audited; two independent screeners from different vendors under a recall-union rule with the full cross-tabulation reported; charting from a written instruction file, with an independent second charter on a seeded sample and every adjudicated correction recorded separately; a prespecified meta-analysis test run on a dedicated extraction with sensitivity analyses; machine-generated counts checked by a verification script; and two rounds of independent adversarial pre-review, whose findings and their disposition ship with the package. The limitations are substantial and several are specific to the workflow. Screening, assessment, charting and appraisal were performed by AI systems. At revision, every clinical study was appraised a second time by an assessor from a different vendor, the six credibility items were coded again on a sample by two further coders, and a reference standard was built for the charting sample by two adjudicators reading the source text. Those additions measure the problem rather than remove it. Against the adjudicated reference standard, the primary charter was correct on 85% to 100% of records depending on the field, which is reassuring but represents agreement between independent machine readings of the same text, not verification by a person. No human-verified subsample exists for this review, and both screeners read the same instruction file, so a blind spot written into that file would be invisible to every check reported here. The credibility items proved only moderately reproducible, with the two items carrying the strongest claims matching the original coding about half the time, and agreement on the case-series checklist was poor. A reader should treat the direction of these findings as better supported than any individual count and should treat the evidence map as a machine-generated artefact that has been checked by machines. The two rounds of independent review found residual duplicate reports, an unevenly applied dental boundary, overstated descriptions of three artificial intelligence studies, clinical-tier rows without a verified patient denominator, records charted from a title alone, and a pooled outcome that mixed constructs; all were corrected and the corrections are recorded, but they show that the machine-generated corpus needed independent checking and that further errors of the same kind may remain. Full text was available for fewer than half of the reports sought, so 344 of the 639 included studies were charted from the abstract, which inflates the unclear categories (material, manufacturing, funding, availability and, above all, the credibility items), can misstate a tier or an evaluation source in either direction, and limits the risk-of-bias judgements for abstract-only clinical studies to the absence of information; the artificial intelligence set is the most affected, with 25 of 36 charted from abstracts. Sixty reports had neither an abstract nor retrievable text: 47 of these are held unresolved and the other 13 were excluded as second copies of reports already assessed, so they are counted among the duplicate exclusions rather than among the unresolved. A first full-text retrieval attached the wrong articles to most records through an incorrect link type; the defect was caught by the second charters, the affected outputs were discarded and every text was re-retrieved with a title-match guard, but the episode illustrates the fragility of automated retrieval and is reported for that reason. Embase, Scopus, Web of Science, IEEE Xplore and CINAHL were unavailable, OpenAlex was searched only in a supplementary run a day after the main search, the Crossref sweep was to provide supplementary discovery of unknown recall, and the citation chase covered only reference lists indexed in PubMed Central. The review’s coverage and gap claims are therefore claims about the biomedical indexes and OpenAlex, not about the engineering literature as a whole, which is why the title says so. Section 3.8 measures the shortfall rather than asserting it: one broader index was the sole route to 10.6% of the included studies, almost all technical, and a probe of engineering-indexed records absent from this corpus projects roughly 31 further eligible studies (95% CI 8 to 101), again technical. That is consistent with what is known more generally about database selection in systematic reviews, where a substantial minority of eligible records are found in only one database, and the records missed by a narrow search are not a random sample of the literature [150,151], although the effect of extending beyond one index on pooled effect estimates is often modest [152]. Applied here, that means the descriptive proportions in the evidence map are conditional on the databases searched and the technical counts are a floor, while the clinical findings, which rest on MEDLINE-indexed randomised and comparative studies, are unlikely to move. The underlying problem, that a structure inferred from an incompletely observed network can differ systematically from the whole, has been formalised in network science, where how far a sampled subgraph preserves the properties of the full graph depends both on how much is missing and on whether it is missing at random [153]. That work treats random and non-uniform random sampling, and the omission here is neither: it is correlated with discipline, venue and indexing age, which is the less forgiving case, so the bound we report rests on the empirical screening in Section 3.8 rather than on an analytical result. The rule-layer audit bounds the false-exclusion rate only loosely. Sixty-seven reports remain unresolved and 16 included studies are preprints. The pooled estimates rest on between 6 and 11 studies from a single extraction pass (a first pass that returned one outcome per study was found incomplete by the third independent review and was repeated per outcome), compare the delivered implant rather than any design method, and are very low certainty; publication bias could not be assessed for the smaller pools and the small-study test on the operative-time pool is weak under its heterogeneity. Finally, the review counts studies, not implants or patients, and a region with many small series may appear better evidenced than a region with a single large registry.
4.4. Implications and Research Agenda
Three implications follow. For the interpretation of the clinical literature, the accuracy and complication figures reported for patient-specific implants are properties of an entire pipeline whose upstream components are rarely evaluated in the same study; a clinical series that reports a 1 mm deviation from plan says nothing about whether a different segmentation, a different mirroring procedure or a different finite element model would have produced a different implant, and the pooled plan-deviation and revision advantages found here are evidence about delivered implants, not about which pipeline produced them. For engineering studies, the credibility items that were most often unmet where they could be read—independent validation material, comparison against a physical experiment and availability of files—are exactly the items that make an in silico result transferable to the next patient, and the reporting frameworks that already exist for computational models and for imaging AI could be adopted without waiting for a design-specific standard. For artificial intelligence, the field has benchmark-level evidence for cranial shape completion and almost nothing else: the next studies needed are not further architectures on the same datasets but evaluation on independent centres, prospective use in design workflows with the manual design as comparator, and release of models and data under governance. Comparative clinical studies of design methods, in which the same implant type is produced by two pipelines, would answer the question that the clinical trials of implants cannot.
5. Conclusions
The evidence for artificial intelligence, imaging and computational planning in patient-specific skeletal implant design is unevenly distributed across a large body of literature. Virtual surgical planning with mirroring or registration has randomised and comparative clinical evidence, concentrated in craniomaxillofacial surgery and evaluating the delivered implant rather than the method; pooled across the studies with extractable estimates, delivered patient-specific implants deviate less from their plan and are revised less often than conventional ones, at very low certainty. Finite element analysis and optimisation methods are applied widely, validated against independent material or physical experiments in a minority of the studies that can be read, and shared almost never. Artificial intelligence has benchmark evidence for cranial defect completion, little evaluation on independent data and almost no clinical use. The clinical evidence is also weak in itself, and this should carry as much weight as the effect estimates: no randomised trial reached low risk of bias, and almost every non-randomised comparison was at serious risk from confounding, because patient-specific and conventional implants were assigned by period, surgeon or case complexity and the analyses did not adjust for it. Clinicians reading outcome studies of patient-specific implants should read them as evidence about pipelines whose components are unequally validated, and investigators should direct the next studies at those components rather than at the implants alone.
Acknowledgments
During the preparation of this manuscript, the authors used Claude (Opus 5 and Fable 5.1, Anthropic) and GPT-5.6-sol (OpenAI, via the Codex command-line interface) for the purposes of literature screening, full-record assessment and data charting, risk-of-bias appraisal, analysis code, figure generation and drafting, as described in Section 2.8 and logged in Text S5. The authors have reviewed and edited the output and take full responsibility for the content of this publication.
Abbreviations
The following abbreviations are used in this manuscript:
| AI | Artificial intelligence |
| API | Application programming interface |
| CAD/CAM | Computer-aided design/computer-aided manufacturing |
| CI | Confidence interval |
| CINAHL | Cumulative Index to Nursing and Allied Health Literature |
| CLAIM | Checklist for Artificial Intelligence in Medical Imaging |
| CLI | Command-line interface |
| CNN | Convolutional neural network |
| CSF | Cerebrospinal fluid |
| CT | Computed tomography |
| DOI | Digital object identifier |
| DSC | Dice similarity coefficient |
| GPT | Generative pre-trained transformer |
| GRADE | Grading of Recommendations Assessment, Development and Evaluation |
| HD95 | 95th-percentile Hausdorff distance |
| HHS | Harris Hip Score |
| HKSJ | Hartung–Knapp–Sidik–Jonkman (confidence-interval method) |
| HR | Hazard ratio |
| JBI | Joanna Briggs Institute |
| MSTS/MSTS-93 | Musculoskeletal Tumor Society (rating system), 1993 revision |
| NCBI | National Center for Biotechnology Information |
| OSF | Open Science Framework |
| PEEK | Polyether ether ketone |
| PRISMA | Preferred Reporting Items for Systematic Reviews and Meta-Analyses |
| PRISMA-S | PRISMA extension for reporting literature searches |
| PROSPERO | International Prospective Register of Systematic Reviews |
| PSI | Patient-specific implant |
| R0 | Clear (microscopically negative) resection margin |
| RCT | Randomised controlled trial |
| RMSE | Root mean square error |
| RoB 2 | Cochrane Risk-of-Bias tool, version 2 (randomised trials) |
| ROBINS-I | Risk Of Bias In Non-randomised Studies—of Interventions |
| SD | Standard deviation |
| SSM | Statistical shape model(ling) |
| SWiM | Synthesis Without Meta-analysis |
| TMJ | Temporomandibular joint |
| TRIPOD + AI | Transparent Reporting of a multivariable prediction model for Individual Prognosis Or Diagnosis + Artificial Intelligence |
| VAS | Visual analogue scale |
| VSP | Virtual surgical planning |
Supplementary Materials
The following supporting information can be downloaded at https://www.mdpi.com/article/10.3390/bioengineering13101149/s1, Text S1: Extended methods: search, selection, retrieval, charting agreement and corrections; Text S2: Prospective protocol with dated amendments; Text S3: Full search strategies (PRISMA-S); Text S4: Instructions given to the AI screeners, charters and assessors; Text S5: Contemporaneous log of AI use; Text S6: Adjudication of the plan-deviation accuracy pool; Table S1: All included studies; Table S2: Studies excluded at assessment; Table S3: Technical credibility items; Table S4: Risk of bias per clinical-tier study; Table S5: Funding, conflict of interest and industry authorship; Table S6: Outcome-by-study disposition for the meta-analysis test; Table S7: Artificial intelligence studies: evaluation data, leakage risk and prediction-model reporting. References [13,15,16,17,19,20,21,22,23,24,25,71,89,96,103,108,111,119] are cited in the Supplementary Materials.
Author Contributions
Conceptualisation, A.B. and M.M.; methodology, R.D.D. and A.L.; software, V.T.; validation, A.B., M.M. and C.S.; formal analysis, C.S. and C.-A.G.; investigation, V.T. and A.D.; resources, C.S. and A.L.; data curation, V.T. and A.D.; writing—original draft preparation, R.A. and V.T.; writing—review and editing, R.D.D. and C.-A.G.; visualisation, R.A.; supervision, M.E.P.; project administration, M.E.P. All authors have read and agreed to the published version of the manuscript.
Institutional Review Board Statement
Not applicable (published literature only).
Informed Consent Statement
Not applicable.
Data Availability Statement
The data presented in this study are available in the article and its Supplementary Materials. The search code, raw search responses, screening and charting outputs, instruction files, adjudicated corrections, analysis code, verification suite and the adversarial review reports are available on request from the corresponding author.
Conflicts of Interest
V.T. and A.L. are employed by Eden Health SRL, C.-A.G. by CG Tech Labs SRL and A.D. by Humans Nexus Ltd. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Funding Statement
This research received no external funding.
Footnotes
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.
References
- 1.Wallace N., Schaffer N.E., Aleem I.S., Patel R. 3D-printed Patient-specific Spine Implants: A Systematic Review. Clin. Spine Surg. 2020;33:400–407. doi: 10.1097/BSD.0000000000001026. [DOI] [PubMed] [Google Scholar]
- 2.Levesque J.N., Shah A., Ekhtiari S., Yan J.R., Thornley P., Williams D.S. Three-dimensional printing in orthopaedic surgery: A scoping review. EFORT Open Rev. 2020;5:430–441. doi: 10.1302/2058-5241.5.190024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Zoltan J., Popescu D., Sanei S.H.R. A systematic review of follow-up results of additively manufactured customized implants for the pelvic area. Expert Rev. Med. Devices. 2023;20:233–244. doi: 10.1080/17434440.2023.2183839. [DOI] [PubMed] [Google Scholar]
- 4.Pai A.G., Prabhu S.S. Patient-specific implants for intraoral and maxillofacial reconstruction: A scoping review on customization and fabrication methods. Maxillofac. Plast. Reconstr. Surg. 2025;47:28. doi: 10.1186/s40902-025-00485-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Nolden E.L., Guimarães Carvalho B.K., Barkovskij-Jakobsen K.S., Wenning A.S., Szentes B.L., Agócs G., Németh Z., Kivovics M., Hegyi P., Köles L., et al. Comparative Safety and Efficacy of Patient-Specific Versus Hand-Molded Implants in Cranioplasty: A Systematic Review and Meta-Analysis. J. Clin. Med. 2025;14:8655. doi: 10.3390/jcm14248655. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Kotecha S., Ferro A., Harrison P., Fan K. Orbital reconstruction: A systematic review and meta-analysis evaluating the role of patient-specific implants. Oral Maxillofac. Surg. 2023;27:213–226. doi: 10.1007/s10006-022-01074-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Diaconu A., Holte M.B., Berg-Beckhoff G., Pinholt E.M. Three-Dimensional Accuracy and Stability of Personalized Implants in Orthognathic Surgery: A Systematic Review and a Meta-Analysis. J. Pers. Med. 2023;13:125. doi: 10.3390/jpm13010125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.D’Alpaos D., Badiali G., Ceccariglia F., Nosrati A., Tarsitano A. Use of CAD/CAM Workflow and Patient-Specific Implants for Maxillary Reconstruction: A Systematic Review. J. Clin. Med. 2026;15:647. doi: 10.3390/jcm15020647. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Shergill K.S., Aulakh A.S., Robibo E., Kürten C., Tran K.L., Dial H.S., Prisman E. The utility of patient-specific surgical plates in free flap mandibular reconstruction: A systematic review and meta-analysis. Oral Oncol. 2025;168:107497. doi: 10.1016/j.oraloncology.2025.107497. [DOI] [PubMed] [Google Scholar]
- 10.Barr M.L., Haveles C.S., Rezzadeh K.S., Nolan I.T., Castro R., Lee J.C., Steinbacher D., Pfaff M.J. Virtual Surgical Planning for Mandibular Reconstruction With the Fibula Free Flap: A Systematic Review and Meta-analysis. Ann. Plast. Surg. 2020;84:117–122. doi: 10.1097/SAP.0000000000002006. [DOI] [PubMed] [Google Scholar]
- 11.Hirase T., Vemu S.M., Boddapati V., Ling J.F., So M., Saifi C., Marco R.A.W., Bird J.E. Customized 3-dimensional-printed Vertebral Implants for Spinal Reconstruction After Tumor Resection: A Systematic Review. Clin. Spine Surg. 2024;37:31–39. doi: 10.1097/BSD.0000000000001462. [DOI] [PubMed] [Google Scholar]
- 12.Buist M.L., IJpma F.F.A., van Wees S., Meesters A.M.L., Assink N., Jutte P.C. Patient-specific implants for joint-sparing reconstruction of segmental tibial and femoral defects: Systematic literature review and introduction of a classification framework. Bone Jt. Res. 2026;15:647–661. doi: 10.1302/2046-3758.156.BJR-2025-0565.R1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Li J., Pimentel P., Szengel A., Ehlke M., Lamecker H., Zachow S., Estacio L., Doenitz C., Ramm H., Shi H., et al. AutoImplant 2020-First MICCAI Challenge on Automatic Cranial Implant Design. IEEE Trans. Med. Imaging. 2021;40:2329–2342. doi: 10.1109/TMI.2021.3077047. [DOI] [PubMed] [Google Scholar]
- 14.Memon A.R., Shi H., Memon T.R., Egger J., Chen X. Deep learning-based automatic cranial implant design through direct defect shape prediction and its comparison study. Med. Biol. Eng. Comput. 2025;63:2815–2826. doi: 10.1007/s11517-025-03363-5. [DOI] [PubMed] [Google Scholar]
- 15.Morrison T.M., Hariharan P., Funkhouser C.M., Afshari P., Goodin M., Horner M. Assessing Computational Model Credibility Using a Risk-Based Framework: Application to Hemolysis in Centrifugal Blood Pumps. ASAIO J. 2019;65:349–360. doi: 10.1097/MAT.0000000000000996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Collins G.S., Moons K.G.M., Dhiman P., Riley R.D., Beam A.L., Van Calster B., Ghassemi M., Liu X., Reitsma J.B., van Smeden M., et al. TRIPOD+AI statement: Updated guidance for reporting clinical prediction models that use regression or machine learning methods. BMJ. 2024;385:e078378. doi: 10.1136/bmj-2023-078378. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Tejani A.S., Klontzas M.E., Gatti A.A., Mongan J.T., Moy L., Park S.H., Kahn C.E., CLAIM 2024 Update Panel Checklist for Artificial Intelligence in Medical Imaging (CLAIM): 2024 Update. Radiol. Artif. Intell. 2024;6:e240300. doi: 10.1148/ryai.240300. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Patil A., Kulkarni K., Xie S., Bull A.M.J., Jones G.G. The accuracy of statistical shape models in predicting bone shape: A systematic review. Int. J. Med. Robot. 2023;19:e2503. doi: 10.1002/rcs.2503. [DOI] [PubMed] [Google Scholar]
- 19.Page M.J., McKenzie J.E., Bossuyt P.M., Boutron I., Hoffmann T.C., Mulrow C.D., Shamseer L., Tetzlaff J.M., Akl E.A., Brennan S.E., et al. The PRISMA 2020 statement: An updated guideline for reporting systematic reviews. BMJ. 2021;372:n71. doi: 10.1136/bmj.n71. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Rethlefsen M.L., Kirtley S., Waffenschmidt S., Ayala A.P., Moher D., Page M.J., Koffel J.B., PRISMA-S Group PRISMA-S: An extension to the PRISMA Statement for Reporting Literature Searches in Systematic Reviews. Syst. Rev. 2021;10:39. doi: 10.1186/s13643-020-01542-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Campbell M., McKenzie J.E., Sowden A., Katikireddi S.V., Brennan S.E., Ellis S., Hartmann-Boyce J., Ryan R., Shepperd S., Thomas J., et al. Synthesis without meta-analysis (SWiM) in systematic reviews: Reporting guideline. BMJ. 2020;368:l6890. doi: 10.1136/bmj.l6890. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Sterne J.A.C., Savović J., Page M.J., Elbers R.G., Blencowe N.S., Boutron I., Cates C.J., Cheng H.Y., Corbett M.S., Eldridge S.M., et al. RoB 2: A revised tool for assessing risk of bias in randomised trials. BMJ. 2019;366:l4898. doi: 10.1136/bmj.l4898. [DOI] [PubMed] [Google Scholar]
- 23.Sterne J.A., Hernán M.A., Reeves B.C., Savović J., Berkman N.D., Viswanathan M., Henry D., Altman D.G., Ansari M.T., Boutron I., et al. ROBINS-I: A tool for assessing risk of bias in non-randomised studies of interventions. BMJ. 2016;355:i4919. doi: 10.1136/bmj.i4919. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Munn Z., Barker T.H., Moola S., Tufanaru C., Stern C., McArthur A., Stephenson M., Aromataris E. Methodological quality of case series studies: An introduction to the JBI critical appraisal tool. JBI Evid. Synth. 2020;18:2127–2133. doi: 10.11124/JBISRIR-D-19-00099. [DOI] [PubMed] [Google Scholar]
- 25.Guyatt G.H., Oxman A.D., Vist G.E., Kunz R., Falck-Ytter Y., Alonso-Coello P., Schünemann H.J., GRADE Working Group GRADE: An emerging consensus on rating quality of evidence and strength of recommendations. BMJ. 2008;336:924–926. doi: 10.1136/bmj.39489.470347.AD. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Li J., von Campe G., Pepe A., Gsaxner C., Wang E., Chen X., Zefferer U., Tödtling M., Krall M., Deutschmann H., et al. Automatic skull defect restoration and cranial implant generation for cranioplasty. Med. Image Anal. 2021;73:102171. doi: 10.1016/j.media.2021.102171. [DOI] [PubMed] [Google Scholar]
- 27.Wodzinski M., Daniol M., Socha M., Hemmerling D., Stanuch M., Skalski A. Deep learning-based framework for automatic cranial defect reconstruction and implant modeling. Comput. Methods Programs Biomed. 2022;226:107173. doi: 10.1016/j.cmpb.2022.107173. [DOI] [PubMed] [Google Scholar]
- 28.Wodzinski M., Kwarciak K., Daniol M., Hemmerling D. Improving deep learning-based automatic cranial defect reconstruction by heavy data augmentation: From image registration to latent diffusion models. Comput. Biol. Med. 2024;182:109129. doi: 10.1016/j.compbiomed.2024.109129. [DOI] [PubMed] [Google Scholar]
- 29.Wodzinski M., Daniol M., Hemmerling D. Automatic skull reconstruction by deep learnable symmetry enforcement. Comput. Methods Programs Biomed. 2025;263:108670. doi: 10.1016/j.cmpb.2025.108670. [DOI] [PubMed] [Google Scholar]
- 30.Dot G., Schouman T., Dubois G., Rouch P., Gajny L. Fully automatic segmentation of craniomaxillofacial CT scans for computer-assisted orthognathic surgery planning using the nnU-Net framework. Eur. Radiol. 2022;32:3639–3648. doi: 10.1007/s00330-021-08455-y. [DOI] [PubMed] [Google Scholar]
- 31.Juneja M., Kharbanda M., Pandey N., Sudhir A., Poddar A., Kaur H., Prakash P., Jaiswal M.K., Jindal P., Breedon P. MaxI-Net: A 3D AI Framework for CBCT-Based Maxillofacial Defect Reconstruction and Patient-Specific Implant Generation with Biomechanical Validation. Bioengineering. 2026;13:619. doi: 10.3390/bioengineering13060619. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Beisekenov N., Azamatov B., Sadenova M., Dogadkin D., Kaliyev D., Rudenko S., Syrnev B. Data-Driven Design and Additive Manufacturing of Patient-Specific Lattice Titanium Scaffolds for Mandibular Bone Reconstruction. J. Funct. Biomater. 2025;16:350. doi: 10.3390/jfb16090350. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Ngo J., Athreya K., Ehrlich B., Sayrs L., Morphew T., Halvorson S., Aminian A. Use of machine learning predictive models in sagittal alignment planning in adolescent idiopathic scoliosis surgery. Eur. Spine J. 2026;35:23–31. doi: 10.1007/s00586-025-09420-9. [DOI] [PubMed] [Google Scholar]
- 34.Wu C.T., Yang Y.H., Chang Y.Z. Creating high-resolution 3D cranial implant geometry using deep learning techniques. Front. Bioeng. Biotechnol. 2023;11:1297933.:1297933. doi: 10.3389/fbioe.2023.1297933. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Chaubey G., Singh A., Farooq A., Mishra D. RibCageImp: A Deep Learning Framework for 3D Ribcage Implant Generation; Proceedings of the Sixteen Indian Conference on Computer Vision, Graphics and Image Processing; Mandi, India. 17–20 December 2025; pp. 1–9. [DOI] [Google Scholar]
- 36.Shash Y.H., Elden R.H. Integration of finite element method and artificial intelligence for evaluating PEEK composites in rib cage reconstruction process under impact conditions. J. Mater. Sci. Mater. Med. 2025;37:17. doi: 10.1007/s10856-025-06972-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Karlagina Y., Shelemanov A., Zilberman V., Zernitckaia E., Lovushkina E., Zavialova A., Rogachev K., Afanasiev N., Kolobov Y., Manokhin S., et al. Laser-induced antibacterial coating on the surface of an individual titanium membrane designed by a neural network. Opt. Quantum Electron. 2022;55:775. doi: 10.21203/rs.3.rs-2252495/v1. [DOI] [Google Scholar]
- 38.Ayala Ortín C. Bachelor’s Thesis. Polytechnic University of Cartagena; Cartagena, Spain: 2026. Design, Optimization, and Virtual Validation of Personalized Mandibular Prostheses Using Automatic Segmentation with nnU-Net. [Google Scholar]
- 39.Morais A.R.A. Master Thesis. Universidade do Minho; Braga, Portugal: 2018. Automated Computer-Aided Design of Cranial Implants: A Deep Learning Approach. [Google Scholar]
- 40.Kodym O., Španěl M., Herout A. Segmentation of Defective Skulls from CT Data for Tissue Modelling. arXiv. 2019 doi: 10.48550/arxiv.1911.08805. [DOI] [Google Scholar]
- 41.Li J., Pepe A., Gsaxner C., Jin Y., Egger J. Learning to Rearrange Voxels in Binary Segmentation Masks for Smooth Manifold Triangulation. arXiv. 2021 doi: 10.48550/arxiv.2108.05269. [DOI] [Google Scholar]
- 42.Sulakhe H., Li J., Egger J., Goyal P. CranGAN: Adversarial Point Cloud Reconstruction for patient-specific Cranial Implant Design. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2022;2022:603–608. doi: 10.1109/EMBC48229.2022.9871069. [DOI] [PubMed] [Google Scholar]
- 43.Resmi S., Singh R.P., Palaniappan K. Automatic Skull Shape Completion of Defective Skulls Using Transformers for Cranial Implant Design. Procedia Comput. Sci. 2024;235:3305–3314. doi: 10.1016/j.procs.2024.04.312. [DOI] [Google Scholar]
- 44.Kesornsri T., Asawalertsak N., Tantisereepatana N., Manowongpichate P., Lohwongwatana B., Puncreobutr C., Achakulvisut T., Vateekul P. CraNeXt: Automatic Reconstruction of Skull Implants With Skull Categorization Technique. IEEE Access. 2024;12:84907–84922. doi: 10.1109/access.2024.3415173. [DOI] [Google Scholar]
- 45.Zhang J., Zhuang Y., Feng C., Li X., Chen K., Han L., Wang Y., Zhu X., Yang M., Yin G., et al. Inverse design of skull osteoinductive implants with multi-level pore structures through machine learning. J. Mater. Chem. B. 2024;12:9991–10003. doi: 10.1039/d4tb01104j. [DOI] [PubMed] [Google Scholar]
- 46.Wodzinski M., Hemmerling D., Daniol M. Automatic Cranial Defect Reconstruction with Self-Supervised Deep Deformable Masked Autoencoders. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 2024;2024:1–4. doi: 10.1109/EMBC53108.2024.10781853. [DOI] [PubMed] [Google Scholar]
- 47.Juneja M., Poddar A., Kharbanda M., Sudhir A., Gupta S., Joshi P., Goel A., Fatma N., Gupta M., Tarkas S., et al. SCAI-Net: An AI-driven framework for optimized, fast, and resource-efficient skull implant generation for cranioplasty using CT images. Comput. Biol. Med. 2025;194:110504. doi: 10.1016/j.compbiomed.2025.110504. [DOI] [PubMed] [Google Scholar]
- 48.Salem M., Wael O., Attallah M.M., Elkaseer A. Tailoring Patient-Specific Cranial Implants for Bone Reconstruction via End-to-End Deep Learning Image-to-Print Approach. IEEE Access. 2026;14:25677–25690. doi: 10.1109/access.2026.3663618. [DOI] [Google Scholar]
- 49.Nguyen L.H., Phung M.N., Nguyen H.T., Nguyen C.T.K., Hoang H.H. Optimizing 3D UNet Parameters for Cranial Defect Reconstruction. Appl. Sci. 2026;16:4763. doi: 10.3390/app16104763. [DOI] [Google Scholar]
- 50.Guo W. Doctoral Dissertation. University of Alberta; Edmonton, AB, Canada: 2026. Data-Driven Design of Customized Biomimetic Cranial Implants: Integrating Microstructural Analysis, Mechanical Modeling, and Additive Manufacturing. [DOI] [Google Scholar]
- 51.Banis G., Kousiatza C., Sarkiri M., Touloupas Κ., Spiliotopoulos A., Kontopoulos P., Stergiadou S., Hantes M., Zibis A., Malizos K., et al. Does the ratio of stem-to-femur length matter? A validated finite element analysis. J. Exp. Orthop. 2025;12:e70484. doi: 10.1002/jeo2.70484. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Yáñez González-Cuéllar J.A., Moscol Albanil I.D.P., Ojeda C., Franco-Martínez F., Díaz Lantada A., Solórzano-Requejo W. Dual-CNN-based AI framework for patient-specific design of short femoral stems. Comput. Methods Programs Biomed. 2026;285:109521. doi: 10.1016/j.cmpb.2026.109521. [DOI] [PubMed] [Google Scholar]
- 53.Kanagalingam S., Dalton C., Champneys P., Boutefnouchet T., Fernandez-Vicente M., Shepherd D.E.T., Wimpenny D., Thomas-Seale L.E.J. Detailed design for additive manufacturing and post processing of generatively designed high tibial osteotomy fixation plates. Prog. Addit. Manuf. 2023;8:409–426. doi: 10.1007/s40964-022-00342-2. [DOI] [Google Scholar]
- 54.Burge T.A., Jeffers J.R.T., Myant C.W. Applying machine learning methods to enable automatic customisation of knee replacement implants from CT data. Sci. Rep. 2023;13:3317. doi: 10.1038/s41598-023-30483-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Grothues S., Becker A.K., Hohlmann B., Radermacher K. Parameter-based patient-specific restoration of physiological knee morphology for optimized implant design and matching. Biomed. Tech. 2023;68:537–544. doi: 10.1515/bmt-2023-0017. [DOI] [PubMed] [Google Scholar]
- 56.Guezou-Philippe A., Clavé A., Maguet E., Maintier L., Garraud C., Fouefack J.R., Burdin V., Stindel E., Dardenne G. Fully automated workflow for designing patient-specific orthopaedic implants: Application to total knee arthroplasty. PLoS ONE. 2025;20:e0325587. doi: 10.1371/journal.pone.0325587. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.He J., Guo Z.X., Zeng Y.X., Su X.Y., Xian P.G. Pre-planning of Individualized Talus Implants Based on Computed Tomography-Automated Segmentation. Preprints. 2023:2023070273. doi: 10.20944/preprints202307.0273.v1. [DOI] [Google Scholar]
- 58.Mazzocchetti S., Spezialetti R., Bevini M., Badiali G., Lisanti G., Salti S., Di Stefano L. Neural shape completion for personalized Maxillofacial surgery. Sci. Rep. 2024;14:19810. doi: 10.1038/s41598-024-68084-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Zhou Y., Helgason B., Ferguson S.J., Persson C. Optimization of primary screw stability in Trabecular bone using neural network-based models. Comput. Methods Programs Biomed. 2025;264:108720. doi: 10.1016/j.cmpb.2025.108720. [DOI] [PubMed] [Google Scholar]
- 60.Biswas J.K., Dey S., Karmakar S.K., Roychowdhury A., Datta S. Design of Patient Specific Spinal Implant (Pedicle Screw Fixation) using FE Analysis and Soft Computing Techniques. Curr. Med. Imaging. 2020;16:371–382. doi: 10.2174/1573405614666181018122538. [DOI] [PubMed] [Google Scholar]
- 61.von Atzigen M., Liebmann F., Hoch A., Miguel Spirig J., Farshad M., Snedeker J., Fürnstahl P. Marker-free surgical navigation of rod bending using a stereo neural network and augmented reality in spinal fusion. Med. Image Anal. 2022;77:102365. doi: 10.1016/j.media.2022.102365. [DOI] [PubMed] [Google Scholar]
- 62.Raisian S., Fallahi H.R., Khiabani K.S., Heidarizadeh M., Azdoo S. Customized Titanium Mesh Based on the 3D Printed Model vs. Manual Intraoperative Bending of Titanium Mesh for Reconstructing of Orbital Bone Fracture: A Randomized Clinical Trial. Rev. Recent Clin. Trials. 2017;12:154–158. doi: 10.2174/1574887112666170821165206. [DOI] [PubMed] [Google Scholar]
- 63.Schneider D., Kämmerer P.W., Hennig M., Schön G., Thiem D.G.E., Bschorer R. Customized virtual surgical planning in bimaxillary orthognathic surgery: A prospective randomized trial. Clin. Oral Investig. 2019;23:3115–3122. doi: 10.1007/s00784-018-2732-3. [DOI] [PubMed] [Google Scholar]
- 64.Li B., Wei H., Jiang T., Qian Y., Zhang T., Yu H., Zhang L., Wang X. Randomized Clinical Trial of the Accuracy of Patient-Specific Implants versus CAD/CAM Splints in Orthognathic Surgery. Plast. Reconstr. Surg. 2021;148:1101–1110. doi: 10.1097/PRS.0000000000008427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Sánchez-Jáuregui E., Baranda-Manterola E., Ranz-Colio Á., Bueno de Vicente Á., Acero-Sanz J. Custom made cutting guides and osteosynthesis plates versus CAD/CAM occlusal splints in positioning and fixation of the maxilla in orthognathic surgery: A prospective randomized study. J. Craniomaxillofac. Surg. 2022;50:609–614. doi: 10.1016/j.jcms.2022.05.010. [DOI] [PubMed] [Google Scholar]
- 66.van der Wel H., Kraeima J., Spijkervet F.K.L., Schepers R.H., Jansma J. Postoperative skeletal stability at the one-year follow-up after splintless Le Fort I osteotomy using patient-specific osteosynthesis versus conventional osteosynthesis: A randomized controlled trial. Int. J. Oral Maxillofac. Surg. 2023;52:679–685. doi: 10.1016/j.ijom.2022.09.005. [DOI] [PubMed] [Google Scholar]
- 67.Bevini M., Zavalloni G., Federico L., Centra M., Gulotta C., Mirabile F.M.C., Cercenelli L., Incerti Parenti S., Marcelli E., Tarsitano A., et al. Single-jaw patient-specific implants in bimaxillary orthognathic surgery: Randomized cross-controlled comparison between maxilla-guided and mandible-guided approach. J. Craniomaxillofac. Surg. 2025;53:632–641. doi: 10.1016/j.jcms.2024.12.022. [DOI] [PubMed] [Google Scholar]
- 68.Sheta M.S., Elsaadany W.H. The accuracy of reconstruction of orbital wall fractures using prebent mesh versus patient specific implants: A randomized clinical trial. BMC Oral Health. 2025;25:1719. doi: 10.1186/s12903-025-07082-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Alshikh B.A.A., Eid M.H., Amin A., Elsholkamy M. Assessment of Efficacy of Prebent Titanium Mesh versus Customized Poly-Ether-Ether-Ketone Mesh for Three-Dimensional Augmentation in Deficient Maxillary Ridge: A Randomized Clinical Trial. Preprint. 2026 doi: 10.21203/rs.3.rs-8841053/v1. [DOI] [Google Scholar]
- 70.El Morsy O.A., Mekhemar S., Mounir M., El Assiouty S., Mounir S. Assessment of Accuracy and Orbital Volume Using Patient-Specific Titanium Implant Versus Patient-Specific Zirconia Implant for Orbital Floor Reconstruction in Blowout Fractures: A Randomized Clinical Trial. J. Craniofac. Surg. 2026;ahead of print doi: 10.1097/SCS.0000000000012936. [DOI] [PubMed] [Google Scholar]
- 71.Abdellatif A., Hatem H., Elhadidy M., Omara M. Precision of virtual surgical planning employing screw-holes locator and patient-specific implants versus conventional plate osteosynthesis for parasymphyseal mandibular fracture: A double-blinded randomized control trial. BMC Oral Health. 2026;26:1348. doi: 10.1186/s12903-026-09303-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Maini L., Verma T., Sharma A., Sharma A., Mishra A., Jha S. Evaluation of accuracy of virtual surgical planning for patient-specific pre-contoured plate in acetabular fracture fixation. Arch. Orthop. Trauma. Surg. 2018;138:495–504. doi: 10.1007/s00402-018-2868-2. [DOI] [PubMed] [Google Scholar]
- 73.Shuang F., Hu W., Shao Y., Li H., Zou H. Treatment of Intercondylar Humeral Fractures With 3D-Printed Osteosynthesis Plates. Medicine. 2016;95:e2461. doi: 10.1097/MD.0000000000002461. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Ciocca L., Marchetti C., Mazzoni S., Baldissara P., Gatto M.R., Cipriani R., Scotti R., Tarsitano A. Accuracy of fibular sectioning and insertion into a rapid-prototyped bone plate, for mandibular reconstruction using CAD-CAM technology. J. Craniomaxillofac. Surg. 2015;43:28–33. doi: 10.1016/j.jcms.2014.10.005. [DOI] [PubMed] [Google Scholar]
- 75.Tarsitano A., Ciocca L., Scotti R., Marchetti C. Morphological results of customized microvascular mandibular reconstruction: A comparative study. J. Craniomaxillofac. Surg. 2016;44:697–702. doi: 10.1016/j.jcms.2016.03.007. [DOI] [PubMed] [Google Scholar]
- 76.Tarsitano A., Battaglia S., Crimi S., Ciocca L., Scotti R., Marchetti C. Is a computer-assisted design and computer-assisted manufacturing method for mandibular reconstruction economically viable? J. Craniomaxillofac. Surg. 2016;44:795–799. doi: 10.1016/j.jcms.2016.04.003. [DOI] [PubMed] [Google Scholar]
- 77.Zimmerer R.M., Ellis E., Aniceto G.S., Schramm A., Wagner M.E., Grant M.P., Cornelius C.P., Strong E.B., Rana M., Chye L.T., et al. A prospective multicenter study to compare the precision of posttraumatic internal orbital reconstruction with standard preformed and individualized orbital implants. J. Craniomaxillofac. Surg. 2016;44:1485–1497. doi: 10.1016/j.jcms.2016.07.014. [DOI] [PubMed] [Google Scholar]
- 78.De Maesschalck T., Courvoisier D.S., Scolozzi P. Computer-assisted versus traditional freehand technique in fibular free flap mandibular reconstruction: A morphological comparative study. Eur. Arch. Otorhinolaryngol. 2017;274:517–526. doi: 10.1007/s00405-016-4246-4. [DOI] [PubMed] [Google Scholar]
- 79.Kim Y.C., Jeong W.S., Park T.K., Choi J.W., Koh K.S., Oh T.S. The accuracy of patient specific implant prebented with 3D-printed rapid prototype model for orbital wall reconstruction. J. Craniomaxillofac. Surg. 2017;45:928–936. doi: 10.1016/j.jcms.2017.03.010. [DOI] [PubMed] [Google Scholar]
- 80.Fan B., Chen H., Sun Y.J., Wang B.F., Che L., Liu S.Y., Li G.Y. Clinical effects of 3-D printing-assisted personalized reconstructive surgery for blowout orbital fractures. Graefes Arch. Clin. Exp. Ophthalmol. 2017;255:2051–2057. doi: 10.1007/s00417-017-3766-y. [DOI] [PubMed] [Google Scholar]
- 81.Kim Y.C., Min K.H., Choi J.W., Koh K.S., Oh T.S., Jeong W.S. Patient-specific puzzle implant preformed with 3D-printed rapid prototype model for combined orbital floor and medial wall fracture. J. Plast. Reconstr. Aesthet. Surg. 2018;71:496–503. doi: 10.1016/j.bjps.2017.11.016. [DOI] [PubMed] [Google Scholar]
- 82.Rückschloß T., Ristow O., Müller M., Kühle R., Zingler S., Engel M., Hoffmann J., Freudlsperger C. Accuracy of patient-specific implants and additive-manufactured surgical splints in orthognathic surgery—A three-dimensional retrospective study. J. Craniomaxillofac. Surg. 2019;47:847–853. doi: 10.1016/j.jcms.2019.02.011. [DOI] [PubMed] [Google Scholar]
- 83.Kim M.J., Jeong W.S., Kim Y.H., Kim H., Cho H., Kim Y., Choi J.W. Clinical Application of a Specific Simulation Software for 3-Dimensional Orbital Volume Modeling for Orbital Wall Reconstruction. Ann. Plast. Surg. 2019;83:48–54. doi: 10.1097/SAP.0000000000001830. [DOI] [PubMed] [Google Scholar]
- 84.Liu K., Sun H., Zhang L., Li B., Chakraborty S., Wang X. Do patient-specific cutting guides and plates improve the accuracy of maxillary repositioning in hemifacial microsomia? Br. J. Oral Maxillofac. Surg. 2020;58:590–596. doi: 10.1016/j.bjoms.2020.02.021. [DOI] [PubMed] [Google Scholar]
- 85.Sigron G.R., Rüedi N., Chammartin F., Meyer S., Msallem B., Kunz C., Thieringer F.M. Three-Dimensional Analysis of Isolated Orbital Floor Fractures Pre- and Post-Reconstruction with Standard Titanium Meshes and “Hybrid” Patient-Specific Implants. J. Clin. Med. 2020;9:1579. doi: 10.3390/jcm9051579. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Lee Z.H., Alfonso A.R., Ramly E.P., Kantar R.S., Yu J.W., Daar D., Hirsch D.L., Jacobson A., Levine J.P. The Latest Evolution in Virtual Surgical Planning: Customized Reconstruction Plates in Free Fibula Flap Mandibular Reconstruction. Plast. Reconstr. Surg. 2020;146:872–879. doi: 10.1097/PRS.0000000000007161. [DOI] [PubMed] [Google Scholar]
- 87.Badiali G., Bevini M., Ruggiero F., Cercenelli L., Lovero E., De Simone E., Rucci P., Bianchi A., Marchetti C. Validation of a patient-specific system for mandible-first bimaxillary surgery: Ramus and implant positioning precision assessment and guide design comparison. Sci. Rep. 2020;10:13317. doi: 10.1038/s41598-020-70107-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Rahimov C.R., Ahmadov S.G., Rahimli M.C., Farzaliyev I.M. Three-Dimensional Diagnosis in Orbital Reconstructive Surgery. Ann. Maxillofac. Surg. 2020;10:3–9. doi: 10.4103/ams.ams_183_19. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Yang W.F., Choi W.S., Wong M.C., Powcharoen W., Zhu W.Y., Tsoi J.K., Chow M., Kwok K.W., Su Y.X. Three-Dimensionally Printed Patient-Specific Surgical Plates Increase Accuracy of Oncologic Head and Neck Reconstruction Versus Conventional Surgical Plates: A Comparative Study. Ann. Surg. Oncol. 2021;28:363–375. doi: 10.1245/s10434-020-08732-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Olate S., Uribe F., Huentequeo-Molina C., Goulart D.R., Sigua-Rodriguez E.A., Alister J.P. Mandibular Angle Contouring Using Porous Polyethylene Stock or PEEK-based Patient Specific Implants. A Critical Analysis. J. Craniofac. Surg. 2021;32:242–246. doi: 10.1097/SCS.0000000000006926. [DOI] [PubMed] [Google Scholar]
- 91.Karanxha L., Rossi D., Hamanaka R., Giannì A.B., Baj A., Moon W., Del Fabbro M., Romano M. Accuracy of splint vs splintless technique for virtually planned orthognathic surgery: A voxel-based three-dimensional analysis. J. Cranio-Maxillofac. Surg. 2021;49:1–8. doi: 10.1016/j.jcms.2020.11.008. [DOI] [PubMed] [Google Scholar]
- 92.Neuhaus M.T., Zeller A.N., Jehn P., Lethaus B., Gellrich N.C., Zimmerer R.M. Intraoperative real-time navigation and intraoperative three-dimensional imaging for patient-specific total temporomandibular joint replacement. Int. J. Oral Maxillofac. Surg. 2021;50:1342–1350. doi: 10.1016/j.ijom.2021.02.020. [DOI] [PubMed] [Google Scholar]
- 93.Knitschke M., Bäcker C., Schmermund D., Böttger S., Streckbein P., Howaldt H.P., Attia S. Impact of Planning Method (Conventional versus Virtual) on Time to Therapy Initiation and Resection Margins: A Retrospective Analysis of 104 Immediate Jaw Reconstructions. Cancers. 2021;13:3013. doi: 10.3390/cancers13123013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Sigron G.R., Barba M., Chammartin F., Msallem B., Berg B.I., Thieringer F.M. Functional and Cosmetic Outcome after Reconstruction of Isolated, Unilateral Orbital Floor Fractures (Blow-Out Fractures) with and without the Support of 3D-Printed Orbital Anatomical Models. J. Clin. Med. 2021;10:3509. doi: 10.3390/jcm10163509. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Kang Y.F., Lv X.M., Qiu S.Y., Ding M.K., Xie S., Zhang L., Cai Z.G., Shan X.F. Virtual Surgical Planning of Deep Circumflex Iliac Artery Flap for Midface Reconstruction. Front. Oncol. 2021;11:718146. doi: 10.3389/fonc.2021.718146. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Möllmann H.L., Apeltrath L., Karnatz N., Wilkat M., Riedel E., Singh D.D., Rana M. Comparison of the Accuracy and Clinical Parameters of Patient-Specific and Conventionally Bended Plates for Mandibular Reconstruction. Front. Oncol. 2021;11:719028. doi: 10.3389/fonc.2021.719028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Harding J., Hartsfield J.K., Mian A.S., Allan B.P., Naoum S., Lee R.J.H., Goonewardene M.S. Accuracy of mandibular proximal segment position using virtual surgical planning and custom osteosynthesis plates. Int. J. Oral Maxillofac. Surg. 2022;51:219–225. doi: 10.1016/j.ijom.2021.04.002. [DOI] [PubMed] [Google Scholar]
- 98.Timoshchuk M.A., Murnan E.J., Chapple A.G., Christensen B.J. Do Patient-Specific Implants Decrease Complications and Increase Orbital Volume Reconstruction Accuracy in Primary Orbital Fracture Reconstruction? J. Oral Maxillofac. Surg. 2022;80:669–675. doi: 10.1016/j.joms.2021.12.006. [DOI] [PubMed] [Google Scholar]
- 99.Jones J.P., Amarista F.J., Jeske N.A., Szalay D., Ellis E. Comparison of the Accuracy of Maxillary Positioning With Interim Splints Versus Patient-Specific Guides and Plates in Executing a Virtual Bimaxillary Surgical Plan. J. Oral Maxillofac. Surg. 2022;80:827–837. doi: 10.1016/j.joms.2022.01.006. [DOI] [PubMed] [Google Scholar]
- 100.Liokatis P., Malenova Y., Fegg F.N., Haidari S., Probst M., Boskov M., Cornelius C.P., Troeltzsch M., Probst F.A. Digital planning and individual implants for secondary reconstruction of midfacial deformities: A pilot study. Laryngoscope Investig. Otolaryngol. 2022;7:369–379. doi: 10.1002/lio2.753. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Rodríguez-Arias J.P., Tapia B., Pampín M.M., Morán M.J., Gonzalez J., Barajas M., Del Castillo J.L., Navarro Cuéllar C., Cebrian J.L. Clinical Outcomes and Cost Analysis of Fibula Free Flaps: A Retrospective Comparison of CAD/CAM versus Conventional Technique. J. Pers. Med. 2022;12:930. doi: 10.3390/jpm12060930. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102.Chepurnyi Y., Zhukovtseva O., Kopchak A., Kanura O. Clinical application of automated virtual orbital reconstruction for orbital fracture management with patient-specific implants: A prospective comparative study. J. Craniomaxillofac. Surg. 2022;50:686–691. doi: 10.1016/j.jcms.2022.05.006. [DOI] [PubMed] [Google Scholar]
- 103.Cassoni A., Manganiello L., Barbera G., Priore P., Fadda M.T., Pucci R., Valentini V. Three-Dimensional Comparison of the Maxillary Surfaces through ICP-Type Algorithm: Accuracy Evaluation of CAD/CAM Technologies in Orthognathic Surgery. Int. J. Environ. Res. Public Health. 2022;19:11834. doi: 10.3390/ijerph191811834. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Spalthoff S., Nejati-Rad N., Rahlf B., Jehn P., Gellrich N.C., Lentge F., Korn P. Time is crucial in malignant tumor cases: Speeding up the process of patient-specific implant creation. Front. Oncol. 2022;12:904343. doi: 10.3389/fonc.2022.904343. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Kerkfeld V., Schorn L., Depprich R., Lommen J., Wilkat M., Kübler N., Rana M., Meyer U. Simultaneous PSI-Based Orthognathic and PEEK Bone Augmentation Surgery Leads to Improved Symmetric Facial Appearance in Craniofacial Malformations. J. Pers. Med. 2022;12:1653. doi: 10.3390/jpm12101653. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Knitschke M., Yonan M., Roller F.C., Pons-Kühnemann J., Attia S., Howaldt H.P., Streckbein P., Böttger S. Osseous Union after Jaw Reconstruction with Fibula-Free Flap: Conventional vs. CAD/CAM Patient-Specific Implants. Cancers. 2022;14:5774. doi: 10.3390/cancers14235774. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Landau Prat D., Massarwa S., Zohar A., Priel A., Sagiv O., Zloto O., Ben Simon G.J. Patient-Specific Orbital Implants Vs. Pre-Formed Implants for Internal Orbital Reconstruction. Semin. Ophthalmol. 2023;38:365–370. doi: 10.1080/08820538.2023.2166353. [DOI] [PubMed] [Google Scholar]
- 108.Kim S.H., Lee S.M., Park J.H., Yang S., Kim J.W. Effectiveness of individualized 3D titanium-printed Orthognathic osteotomy guides and custom plates. BMC Oral Health. 2023;23:255. doi: 10.1186/s12903-023-03000-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109.Pietzka S., Fink J., Winter K., Wilde F., Schramm A., Ebeling M., Kasper R., Sakkas A. Dental Root Injuries Caused by Osteosynthesis Screws in Orthognathic Surgery-Comparison of Conventional Osteosynthesis and Osteosynthesis by CAD/CAM Drill Guides and Patient-Specific Implants. J. Pers. Med. 2023;13:706. doi: 10.3390/jpm13050706. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.Pietzka S., Wenzel M., Winter K., Wilde F., Schramm A., Ebeling M., Kasper R., Scheurer M., Sakkas A. Comparison of Anatomical Preformed Titanium Implants and Patient-Specific CAD/CAM Implants in the Primary Reconstruction of Isolated Orbital Fractures-A Retrospective Study. J. Pers. Med. 2023;13:846. doi: 10.3390/jpm13050846. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111.Kuehle R., Scheurer M., Bouffleur F., Fuchs J., Engel M., Hoffmann J., Freudlsperger C. Accuracy of Patient-Specific Implants in Virtually Planned Segmental Le Fort I Osteotomies. J. Clin. Med. 2023;12:6038. doi: 10.3390/jcm12186038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Block O.M.D. Doctoral Thesis. Georg-August-Universität Göttingen; Göttingen, Germany: 2024. Vergleichende Analyse Zweier Operationsplanungsverfahren für die Unterkieferrekonstruktion mit Freien Knochentransplantaten Hinsichtlich der Postoperativen Präzision nach Umfangreicher Ablativer Chirurgie im Kopf-Hals-Bereich. [DOI] [Google Scholar]
- 113.Merta M., Kiukkonen A., Leikola J., Stoor P., Suojanen J. Skeletal stability after mandible bilateral sagittal split osteotomy—Comparison of patient-specific implant and mini-plate fixation: A retrospective study. J. Craniomaxillofac. Surg. 2024;52:93–100. doi: 10.1016/j.jcms.2023.11.010. [DOI] [PubMed] [Google Scholar]
- 114.Consorti G., Betti E., Catarzi L. Customized orbital implant versus 3D preformed titanium mesh for orbital fracture repair: A retrospective comparative analysis of orbital reconstruction accuracy. J. Craniomaxillofac. Surg. 2024;52:532–537. doi: 10.1016/j.jcms.2024.02.012. [DOI] [PubMed] [Google Scholar]
- 115.Krasovsky A., Hija A., Zeineh N., Capucha T., Haze D.A., Emodi O., Rachmiel A., Shilo D. Comparison of patient specific implant reconstruction vs conventional titanium mesh reconstruction of orbital fractures using a novel method. J. Craniomaxillofac. Surg. 2024;52:491–502. doi: 10.1016/j.jcms.2024.02.002. [DOI] [PubMed] [Google Scholar]
- 116.D’Alpaos D., Badiali G., Ceccariglia F., Tarsitano A. Delayed Orbital Floor Reconstruction Using Mirroring Technique and Patient-Specific Implants: Proof of Concept. J. Pers. Med. 2024;14:459. doi: 10.3390/jpm14050459. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117.Ettinger K.S., Mohamed A.K., Nathan J.M., Vierkant R.A., Morris J.M., Sears V.A., Arce K. Patient-specific Implants Improve Volumetric Surgical Accuracy Compared to Stock Reconstruction Plates in Modern Paradigm Virtual Surgical Planning of Fibular Free Flaps for Head and Neck Reconstruction. J. Oral Maxillofac. Surg. 2024;82:1311–1328. doi: 10.1016/j.joms.2024.06.166. [DOI] [PubMed] [Google Scholar]
- 118.Yildirim A., Hertach R., Yildirim V. Does Surgical Sequencing Still Mater in Fully Digital Bimaxillary Orthognathic Surgery? A Retrospective Accuracy Study of 46 PSI-Guided Cases. J. Med. Dent. 2026;2:37–43. doi: 10.64951/jmdnt.2026.1.04. [DOI] [Google Scholar]
- 119.Le P.H., Kim H., Son M.H., Cha I.H., Kim H.J., Nam W., Kim D. Comparison of Patient-Specific Plates and Pre-Bent Plates for Preserving Condylar Position in Mandible Reconstruction. J. Craniofac Surg. 2025;36:1109–1113. doi: 10.1097/SCS.0000000000010911. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.Mohamed A.K., Arce K., Nathan J.M., Vierkant R.A., Sears V.A., Morris J.M., Ettinger K.S. Patient-Specific Titanium Implants Reduce Hardware-Related Complications Compared to Stock Reconstruction Plates for Virtually Guided Fibular Free Flaps in Head and Neck Reconstruction. J. Oral Maxillofac. Surg. 2025;83:1157–1169. doi: 10.1016/j.joms.2025.05.008. [DOI] [PubMed] [Google Scholar]
- 121.Chen Y., Liu W. Clinical study on precision repair of orbital fractures using 3D-printed orbital models assisting with absorbable plates. BMC Ophthalmol. 2025;25:472. doi: 10.1186/s12886-025-04314-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Ren Z., Yu J., Sang Z., Xiong C., Gan P., Jin Q., Xu Q., Liao H. Efficacy and safety analysis of endoscopic navigation-guided 3D printing technology combined with absorbable materials in the treatment of orbital blowout fractures. Head Face Med. 2025;21:65. doi: 10.1186/s13005-025-00539-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Ismailov M.R.U., Mustafaev U.G.U., Tuerdi M., Khatamov E.B. Quantitative evaluation of surgical fidelity in vsp-guided mandibular reconstruction: A region-specific error analysis of complex defects. Zenodo. 2026;2:740–751. doi: 10.5281/zenodo.18811308. [DOI] [Google Scholar]
- 124.Zoabi A., Shhadeh A., Daoud S., Kablan F., Nahlieli O., Srouji S. Orbital Floor Reconstruction With Patient-Specific Implants: Comparable Outcomes Between Mirroring and Virtual Reduction Techniques. J. Oral Maxillofac. Surg. 2026;84:344–353. doi: 10.1016/j.joms.2025.11.006. [DOI] [PubMed] [Google Scholar]
- 125.Adell-Gómez N., Valls-Esteve A., Malet-Contreras A., Raurich-Leandro J., Ripoll-Palmés R., Valls-Ontañón A., Rubio-Palau J. Orthognathic surgical accuracy and cost-efficiency of patient-specific implants using intraoperative 3D CBCT superposition: A prospective cohort study with retrospective standard miniplates controls. J. Craniomaxillofac. Surg. 2026;54:104453. doi: 10.1016/j.jcms.2026.104453. [DOI] [PubMed] [Google Scholar]
- 126.Troise S., Calabria F., Spinelli R., De Masi S., Tramontano S., Abbate V., Bonavolontà P., Salzano G., De Fazio G.R., Romano A., et al. Virtual surgical planning and 3D custom-made implants in the management of orbital floor fractures: A case control study. J. Stomatol. Oral Maxillofac. Surg. 2026;127:102718. doi: 10.1016/j.jormas.2026.102718. [DOI] [PubMed] [Google Scholar]
- 127.Yang L.H., Merema B.B.J., Kraeima J., Boeve K., Schepman K.P., Huijing M.A., van der Beek E.S.J., Stenekes M.W., Vister J., de Visscher S.A.H.J., et al. Three-Dimensional Surgical Planning in Mandibular Cancer: A Decade of Clinical Experience and Outcomes. Cancers. 2026;18:271. doi: 10.3390/cancers18020271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.Mäntynen P., Koivisto J., Leikola J., Kaasalainen T., Sipos N., Heliövaara A., Horelli J., Wolff J., Lusila N., Suojanen J. Ultra-low-dose cone-beam CT for cleft and craniofacial deformity surgery: Radiation reduction and clinical applicability. Maxillofac. Plast. Reconstr. Surg. 2026;48:5. doi: 10.1186/s40902-026-00505-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Prikule I., Bagante I., Radzins O., Salms G. Impact of In-House 3D-Printed Models on Re-Operation Rates and Volumetric Precision in Orbital Floor Reconstruction: A Comparative Study. J. Clin. Med. 2026;15:2822. doi: 10.3390/jcm15082822. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Aguilar L., Buratovic J.P.V., Ibaceta V.M., Merchan F., Fuhrer A., Toledo X. Three-Dimensional Assessment of Maxillary Stability Using Customized Plates in Orthognathic Surgery: A Retrospective Cohort Study. Craniomaxillofac. Trauma. Reconstr. 2026;19:27. doi: 10.3390/cmtr19020027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 131.Rouhiainen V., Heliövaara A., Leikola J., Suojanen J., Juuri E. Patient-specific implants are more stable than conventional miniplates in Le Fort I osteotomy in patients with clefts. JPRAS Open. 2026;51:444–455. doi: 10.1016/j.jpra.2026.07.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132.de Oliveira Gorla L.F., Salvi S.M., Tweedale H., Lee N.J. PEEK patient-specific implants for orbital reconstruction: Accuracy and clinical outcomes. Br. J. Oral Maxillofac. Surg. 2026 doi: 10.1016/j.bjoms.2026.06.005. in press . [DOI] [PubMed] [Google Scholar]
- 133.Bertram J., Leonhardt H., Podmelle F., Teltzrow T., Franke A. Accuracy of PSI-based hybrid workflow using a temporary intermediate splint versus conventional splint-based maxillary positioning in orthognathic surgery: A retrospective cohort study. BMC Oral Health. 2026;26:1592. doi: 10.1186/s12903-026-09695-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 134.Desai J.B. Cost-Effective Technique of Fabrication of Polymethyl Methacrylate Based Cranial Implant Using Three-Dimensional Printed Moulds and Wax Elimination Technique. J. Craniofac Surg. 2019;30:1259–1263. doi: 10.1097/SCS.0000000000005539. [DOI] [PubMed] [Google Scholar]
- 135.Baldia M., Joseph M., Sharma S., Kumar D., Retnam A., Koshy S., Karuppusami R. Customized cost-effective polymethylmethacrylate cranioplasty: A cosmetic comparison with other low-cost methods of cranioplasty. Acta Neurochir. 2022;164:655–667. doi: 10.1007/s00701-022-05121-0. [DOI] [PubMed] [Google Scholar]
- 136.Kim Y.C., Lee S.J., Woo S.H., Yang S., Choi J.W. A Comparative Study of Titanium Cranioplasty for Extensive Calvarial Bone Defects: Three-Dimensionally Printed Titanium Implants Versus Premolded Titanium Mesh. Ann. Plast. Surg. 2023;91:446–455. doi: 10.1097/SAP.0000000000003663. [DOI] [PubMed] [Google Scholar]
- 137.Kim T.Y., Lee M.H. Surface Modification of Cranioplasty Implants to Reduce Postoperative Cerebrospinal Fluid Collection: A Preliminary Clinical Study. Korean J. Neurotrauma. 2026;22:36–41. doi: 10.13004/kjnt.2026.22.e10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Zhao D., Cheng L., Lu F., Zhang X., Ying J., Wei X., Cao F., Ran C., Zheng G., Liu G., et al. Design, fabrication and clinical characterization of additively manufactured tantalum hip joint prosthesis. Regen. Biomater. 2024;11:rbae057. doi: 10.1093/rb/rbae057. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139.Hu X., Lu M., Zhang Y., Wang Y., Min L., Tu C. A biomechanical comparison between cement packing combined with extra fixation and three-dimensional printed strut-type prosthetic reconstruction for giant cell tumor of bone in distal femur. J. Orthop. Surg. Res. 2022;17:151. doi: 10.1186/s13018-022-03039-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Liu X., Liu Y., Lu W., Liao S., Du Q., Deng Z., Lu W. Combined Application of Modified Three-Dimensional Printed Anatomic Templates and Customized Cutting Blocks in Pelvic Reconstruction After Pelvic Tumor Resection. J. Arthroplast. 2019;34:338–345.e1. doi: 10.1016/j.arth.2018.10.001. [DOI] [PubMed] [Google Scholar]
- 141.Pu F., Liu J., Shi D., Huang X., Zhang J., Wang B., Wu Q., Zhang Z., Shao Z. Reconstruction With 3D-Printed Prostheses After Sacroiliac Joint Tumor Resection: A Retrospective Case-Control Study. Front. Oncol. 2021;11:764938. doi: 10.3389/fonc.2021.764938. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 142.Hu X., Wen Y., Lu M., Luo Y., Zhou Y., Yang X., Tu C., Min L. Biomechanical and clinical outcomes of 3D-printed versus modular hemipelvic prostheses for limb-salvage reconstruction following periacetabular tumor resection: A mid-term retrospective cohort study. J. Orthop. Surg. Res. 2024;19:258. doi: 10.1186/s13018-024-04697-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 143.Zhang B.H., Fu J., Zhang G.Q., Zhou Y.G., Chen J.Y., Chai W. The reconstruction techniques and mid-term clinical outcomes of hip revision for acetabular bone defect after total hip arthroplasty. Zhonghua Wai Ke Za Zhi. 2024;62:836–846. doi: 10.3760/cma.j.cn112139-20240514-00243. [DOI] [PubMed] [Google Scholar]
- 144.Huang X., Huang D., Lin N., Yan X., Qu H., Ye Z. 3D-Printed Prosthesis with an Articular Interface for Anatomical Acetabular Reconstruction After Type I + II (+III) Internal Hemipelvectomy: Clinical Outcomes and Finite Element Analysis. J. Bone Jt. Surg. Am. 2025;107:184–195. doi: 10.2106/JBJS.23.01462. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 145.Guo Q., Li X., Tang Y., Huang Y., Luo L. Augmented reality and three-dimensional plate library-assisted posterior minimally invasive surgery for scapula fracture. Int. Orthop. 2022;46:875–882. doi: 10.1007/s00264-022-05303-6. [DOI] [PubMed] [Google Scholar]
- 146.Lv H., Xu M., Ai Q., Wang F., Zhang Z., Hou T. 3D-printed tantalum artificial vertebral body or titanium mesh in anterior column reconstruction post-spondylectomy for thoracic vertebral tumors: A retrospective comparative cohort study and finite element analysis. J. Orthop. Surg. Res. 2025;20:1059. doi: 10.1186/s13018-025-06351-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 147.van Doremalen R.F.M., van der Linde R.A., Kootstra J.J., van Helden S.H., Hekman E.E.G. Can 3D-printing avoid discomfort-related implant removal in midshaft clavicle fractures? A four-year follow-up. Arch. Orthop. Trauma Surg. 2021;141:1899–1907. doi: 10.1007/s00402-020-03654-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.Gong T., Luo Y., Lu M., Wang J., Zhang Y., Wang Y., Tang F., Li Z., Zhou Y., Min L., et al. The optimal strategy for 3D-printed uncemented endoprosthesis for the bone defect reconstruction of the distal radius, based on biomechanical analysis and retrospective cohort study. J. Surg. Oncol. 2023;127:1043–1053. doi: 10.1002/jso.27215. [DOI] [PubMed] [Google Scholar]
- 149.Velasquez Garcia A., Bukowiec L.G., Yang L., Nishikawa H., Fitzsimmons J.S., Larson A.N., Taunton M.J., Sanchez-Sotelo J., O’Driscoll S.W., Wyles C.C. Artificial intelligence-based three-dimensional templating for total joint arthroplasty planning: A scoping review. Int. Orthop. 2024;48:997–1010. doi: 10.1007/s00264-024-06088-6. [DOI] [PubMed] [Google Scholar]
- 150.Bramer W.M., Rethlefsen M.L., Kleijnen J., Franco O.H. Optimal database combinations for literature searches in systematic reviews: A prospective exploratory study. Syst. Rev. 2017;6:245. doi: 10.1186/s13643-017-0644-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Ewald H., Klerings I., Wagner G., Heise T.L., Stratil J.M., Lhachimi S.K., Hemkens L.G., Gartlehner G., Armijo-Olivo S., Nussbaumer-Streit B. Searching two or more databases decreased the risk of missing relevant studies: A metaresearch study. J. Clin. Epidemiol. 2022;149:154–164. doi: 10.1016/j.jclinepi.2022.05.022. [DOI] [PubMed] [Google Scholar]
- 152.Halladay C.W., Trikalinos T.A., Schmid I.T., Schmid C.H., Dahabreh I.J. Using data sources beyond PubMed has a modest impact on the results of systematic reviews of therapeutic interventions. J. Clin. Epidemiol. 2015;68:1076–1084. doi: 10.1016/j.jclinepi.2014.12.017. [DOI] [PubMed] [Google Scholar]
- 153.Shang Y. Subgraph Robustness of Complex Networks Under Attacks. IEEE Trans. Syst. Man Cybern. Syst. 2019;49:821–832. doi: 10.1109/tsmc.2017.2733545. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data presented in this study are available in the article and its Supplementary Materials. The search code, raw search responses, screening and charting outputs, instruction files, adjudicated corrections, analysis code, verification suite and the adversarial review reports are available on request from the corresponding author.