Automated Insulin Delivery for Patients with Type 2 Diabetes: A Narrative Review

Konstantinos Kitsios 1,*, Christina-Maria Trakatelli 1
PMCID: PMC13725152  PMID: 41788003

Abstract

Despite the availability of newer antidiabetic medications, many patients with type 2 diabetes mellitus still do not achieve their glycemic targets and require insulin treatment. Research from a limited number of randomized controlled trials and observational studies suggests that Automated Insulin Delivery (AID) systems are safe and effective for treating individuals with type 2 diabetes. With the implementation of a hybrid insulin pump, an average reduction in HbA1c of 0.6 to 0.9 percentage points is achieved, along with an increase in Time in Glucose Range (70-180 mg/dL) (3.9-10.0 mmol/L) of 7 to 21 percentage points. This improvement corresponds to an additional 1.7 to 5.1 hours of euglycemia per day, without increasing the risk of hypoglycemia or diabetic ketoacidosis, and with a slight increase in body weight, potentially reflecting improved glycemic control. Concomitant treatment with GLP-1RA and/or SGLT-2 inhibitors has minimal effect on the magnitude of glycemic improvement achieved after switching to AID therapy, although evidence remains limited. Both the effectiveness and device-related adverse reactions are similar to those observed in patients with type 1 diabetes. Although individuals with type 2 diabetes who require insulin treatment are generally older and may be less familiar with technology, training in the necessary skills and adherence to AID use in real-world settings appear feasible. Treatment with AID may slightly improve quality of life and reduce diabetes-related stress.

Keywords: Artificial pancreas, automated insulin delivery, fully closed-loop system, hybrid closed-loop system, hypoglycemia, type 2 diabetes

1. INTRODUCTION

Automated insulin delivery (AID) in the form of a hybrid insulin pump significantly improves glycemic control in patients with type 1 diabetes mellitus (T1DM) [1-3]. Despite the emergence of new classes of antidiabetic medications, a notable number of patients with type 2 diabetes mellitus (T2DM) still fail to achieve their glycemic targets and may require insulin therapy [4]. Patients using multiple daily injections (MDI) or insulin pumps could also benefit from hybrid AID systems; however, safety and efficacy data in T2DM remain limited and are derived mainly from hospitalized patients or small pilot studies conducted under controlled conditions [5-10]. Search strategy included PubMed using the following search string: ((closed-loop) OR (hybrid closed-loop)) OR (closed loop)) OR (hybrid closed loop)) OR (automated insulin delivery)) OR (omnipod)) OR (omnipod5)) OR (omnipod 5)) OR (Diabeloop)) OR (DBLHU)) OR (DBLG1)) OR (CamAPS Fx)) OR (control IQ)) OR (control-IQ)) OR (controlIQ)) OR (670G)) OR (Minimed 670g)) OR (780G)) OR (Minimed 780g)) AND ((Diabetes Mellitus) OR (diabet*)) and CENTRAL using the following search string: (diabet*) AND (Diabetes Mellitus) AND (Minimed 780g):ti,ab,kw OR (780G):ti,ab,kw OR (Minimed 670g):ti,ab,kw OR (670G):ti,ab,kw OR (controlIQ):ti,ab,kw OR (control-IQ):ti,ab,kw OR (control IQ):ti,ab,kw OR (CamAPSFx):ti,ab,kw OR (DBLG1):ti,ab,kw OR (DBLHU):ti,ab,kw OR (Diabeloop):ti,ab,kw OR (omnipod 5):ti,ab,kw OR (omnipod5):ti,ab,kw OR (omnipod):ti,ab,kw OR (automated insulin delivery):ti,ab,kw OR (hybrid closed loop):ti,ab,kw OR (closed loop):ti,ab,kw OR (hybrid closed-loop):ti,ab,kw OR (closed-loop):ti,ab,kw for relevant published trials from inception until September 26th, 2025. This review aims to critically evaluate the existing literature, focusing primarily on results from randomized clinical trials in real-world settings.

2. EVIDENCE FROM RANDOMIZED CLINICAL TRIALS

In the largest clinical trial conducted to date, 319 patients with a mean age of 59 years and a mean duration of T2DM of 18 years were randomly assigned in a 2:1 ratio to receive an AID system (t:slim X2 insulin pump with Control-IQ+ technology, Tandem, in conjunction with a Dexcom G6 sensor) or to continue their existing treatment regimen. The control group maintained their current approach, which involved multiple daily injections of both basal and mealtime insulin or the use of an insulin pump, alongside continuous real-time glucose monitoring [11].

The primary endpoint was the change in glycosylated hemoglobin (HbA1c). After 13 weeks of intervention, the AID group demonstrated a significant reduction in HbA1c, decreasing from 8.2 ± 1.4% (66.1 ± 15.3 mmol/mol) to 7.3 ± 0.9% (56.3 ± 9.8 mmol/mol), representing a drop of 0.9 percentage points (9.8 mmol/mol). In contrast, the control group experienced a more modest reduction of 0.3 percentage points (3.3 mmol/mol), from 8.1 ± 1.2% (65 ± 13.1 mmol/mol) to 7.8 ± 1.1% (61.7 ± 12.1 mmol/mol). The mean adjusted difference between the two groups was -0.6 percentage points (-6.5 mmol/mol), with a 95% confidence interval (CI) of -0.8 to -0.4 (-8.7 to -4.4 mmol/mol) (p < 0.001).

Additionally, the percentage of time that glucose levels were maintained within the therapeutic Target in Range (TIR) of 70-180 mg/dL (3.9-10.0 mmol/L) increased from 48 ± 24% to 64 ± 16% in the AID group, while the control group showed only a slight change from 51 ± 21% to 52 ± 21%. The mean difference in TIR between the two groups was 14 percentage points, with a 95% CI of 11 to 17 (p < 0.001). Notably, the occurrence of hypoglycemia was very low in both groups, with only one episode of severe hypoglycemia reported in a patient using the AID system.

Importantly, the mean total daily insulin dose decreased from 95 ± 47 units at baseline to 87 ± 46 units at 13 weeks in the AID group, whereas it increased from 102 ± 50 units to 104 ± 56 units in the control group. The mean change from baseline in weight was 2.4 ± 4.4 kg in the AID group and 0.9 ± 3.3 kg in the control group. Although the mechanism underlying the slight weight gain observed in the context of reduced total daily insulin delivery remains unclear, the investigators proposed decreased glycosuria-particularly in patients with higher baseline HbA1c levels-or increased calorie intake as possible explanations.

Most of the Patient-Reported Outcomes Surveys (PROs) did not differ between treatment groups. In a subgroup analysis, the treatment effect of AID did not differ between participants receiving Glucagon-Like Peptide-1 Receptor Agonists (GLP-1RA) or Sodium-Glucose Co-Transporter 2 inhibitors (SGLT-2inh) and those who were not taking these medications. However, patients treated with both SGLT-2 inhibitors and GLP-1RA exhibited a greater reduction in HbA1c compared to those not receiving these agents.

During a crossover clinical trial involving 17 patients with a mean age of 63 years, a mean HbA1c of 7.9% (62.8 mmol/mol), and a mean duration of type 2 diabetes of 18 years, participants on insulin pump therapy and Continuous Glucose Monitoring (CGM) were randomized to receive either a hybrid pump (DBLG1) or insulin pump therapy with CGM in an open-loop system over two 6-week periods [12]. The primary endpoint was Time in Range (TIR): 70-180 mg/dL (3.9-10.0 mmol/L). TIR increased to 76% with the closed-loop system compared with 61% in the open-loop phase (mean difference 15 percentage points, interquartile range [IQR] 8.0-22.0; p < 0.001).

The hybrid pump significantly reduced time spent in hyperglycemia, glucose coefficient of variation (CV), and glucose management index (GMI). Time in hypoglycemia was low and did not differ between the two interventions. In contrast, the daily insulin dose was significantly higher with the hybrid pump.

There was no change in weight from the beginning to the end of the study for either the entire population or for each intervention group separately. The mean sleep fragmentation index was lower with the closed-loop system than with conventional treatment (P = 0.047). However, the remaining PROs did not differ between interventions.

In a parallel-design randomized trial, the Tandem t:slim X2 Control-IQ technology hybrid insulin pump (Tandem Diabetes Care) was evaluated against multiple daily insulin injections (MDI) in a cohort of 30 patients with T2DM who had a mean age of 69 years and required residential care [13]. The primary endpoint was TIR: 70-180 mg/dL (3.9-10.0 mmol/L), while secondary endpoints included other parameters of the ambulatory glucose profile, HbA1c, and daily insulin doses. After 12 weeks of treatment, compared with MDI, the use of AID significantly increased TIR from 34.4% to 63% (adjusted mean difference 27.4%; 95% CI 15.0-39.8; p < 0.001), with no significant difference in time spent in hypoglycemia. In the AID group, HbA1c levels also showed a significant reduction, decreasing from 9% (74.9 mmol/mol) to 7.4% (57.4 mmol/mol) (adjusted mean difference -1.3%, 95% CI -1.9 to -0.7; −14.2 mmol/mol, 95% CI -20.8 to −7.7; p < 0.001).

There were no episodes of severe hypoglycemia or diabetic ketoacidosis (DKA) in either intervention group. Insulin use increased by 36.3 ± 65.5 U/day in the control group and by 13.8 ± 75.6 U/day in the intervention group (P = 0.52). Weight decreased by 0.76 ± 4.17 kg in the control group and increased by 1.62 ± 2.88 kg in the intervention group (p = 0.08). However, none of these differences were statistically significant. Satisfaction with hybrid pump therapy was high among both patients and caregivers, and by the end of the study, 77.8% of patients were willing to continue using the AID system.

The Cambridge group investigated the safety and efficacy of a fully automated closed-loop insulin delivery system through two crossover clinical trials. In the first trial, 26 adults with T2DM requiring dialysis, with a mean age of 68 years and a mean duration of diabetes of 20 years, were randomly assigned to two 20-day treatment periods using the CamAPS HX fully closed-loop app with ultra-rapid insulin aspart or conventional insulin therapy paired with a blinded CGM [14]. The primary endpoint was TIR: 70-180 mg/dL (3.9-10.0 mmol/L). TIR achieved with closed-loop insulin delivery was 52.8 ± 12.5%, compared with 37.7 ± 20.5% using conventional therapy, resulting in a mean difference of 15.1 percentage points (95% CI 8.0-22.2; p < 0.001).

Time spent in hypoglycemia (<70 mg/dL; 3.9 mmol/L) was significantly lower with the closed-loop system (0.1%, IQR 0.0-0.4%) compared with conventional therapy (0.2%, IQR 0.0-0.9%) (p = 0.04). Notably, one episode of severe hypoglycemia occurred during the closed-loop period, but it took place when the algorithm was not operational. Total daily insulin dose did not differ between treatments. Most participants reported spending less time managing their diabetes when using the closed-loop system and indicated they would recommend it to others.

In a second trial, 28 adults with T2DM, with a mean age of 59 ± 11 years and a mean HbA1c of 75 ± 15 mmol/mol (9.0% ± 1.4%), were randomly assigned to two 8-week treatment periods using the fully automated closed-loop insulin delivery system (CamAPS HX app) or conventional insulin therapy combined with a blinded CGM [15]. The primary endpoint was TIR: 70-180 mg/dL (3.9-10.0 mmol/L). Closed-loop therapy achieved a TIR of 66.3 ± 14.9%, compared with 32.3 ± 24.7% during conventional insulin therapy, yielding a mean difference of 35.3 percentage points (95% CI 28.0-42.6; p < 0.001).

Closed-loop therapy also resulted in significantly lower HbA1c values than conventional therapy, with levels of 57 ± 9 mmol/mol (7.3% ± 0.8%) versus 72 ± 13 mmol/mol (8.7% ± 1.2%), respectively. This corresponds to a mean difference of -15 mmol/mol (95% CI -11 to -20 mmol/mol), equivalent to -1.4% (95% CI -1.0 to -1.8%; p < 0.001).

Additionally, time spent in hypoglycemia (<70 mg/dL; 3.9 mmol/L) showed no significant difference between the two interventions, and no episodes of severe hypoglycemia were reported. There was only one serious treatment-related adverse event, an infusion-site abscess, during the closed-loop therapy period. Total daily insulin dose was higher during closed-loop therapy (median 0.90 U/kg/day; IQR 0.72-1.63 U/kg/day) compared with the control therapy period (median 0.71 U/kg/day; IQR 0.56-1.26 U/kg/day).

Hypoglycemia confidence scores and Problem Areas in Diabetes (PAID) scores were similar between interventions, while hypoglycemia worry scores were higher during closed-loop therapy. The baseline characteristics of participants included in the randomized controlled trials are summarized in Table 1. Tables 2A and B summarize the outcomes of the randomized controlled trials.

Table 1.

Baseline characteristics of the participants in randomized controlled trials.

Study ID Design AID Comparator Duration Sample size (N) Age (yrs) Diabetes duration (yrs) BMI
(Kg/m2)
HbA1c
(mmol/mol))
TIR
(%)
Insulin dose(u/d) GLP-1RA N(%) SGLT-2 inh
N(%)
Kudva et al. [11] RCT parallel Control-IQ MDI+CGM 13 weeks 319 59(12) 18 (11-26) 33 (29-40) 66.1(15.3) 48
(24)
95(47) 141
(44)
117
(37)
Borel et al. [12] RCT cross-over DBLG1 CSII + CGM 6 weeks 17 63(9) 24(9) 32(4) 62.8(9.9) 58.9
(20.1)
78.1(35.9) 7 (41) 5 (29)
Reznik et al. [13] RCT parallel Control-IQ MDI 12 weeks 29 69.3(6.7) 20.4
(12.3)
32.3 (4.3) 74.9(13.1) 34.4
(21.3)
62.8(37.2) 10 (34) 0
Boughton et al. [14] RCT
cross-
over
CamAPS HX fully closed-loop Standard insulin therapy 20 days 26 68.3(11.2) 20.0 (10.0) 30.4 (6.5) 55.2 (14.2)) 37.7
(20.5)
32.2(12.1-54.4) 0 0
Daly et al. [15] RCT cross-over CamAPS HX fully closed-loop Standard insulin therapy + masked CGM 8 weeks 26 59 (11) 17.5 (8.2) 35.3 (8.6) 74.9(15.3) 32.3
(24.7)
84 (54-129) 11 (42) 6 (23)

Data are reported as mean (SD), or median (IQR) unless otherwise indicated. AID: Automated Insulin Delivery. MDI: Multiple Daily Insulin injections. CSII: Continuous Subcutaneous Insulin Infusion. CGM: Continuous Glucose Monitoring. BMI: Body Mass Index. TIR: Time In Range. GLP-1RA: Glucagon-Like Peptide 1 Receptor Agonist. SGLT-2inh: Sodium-Glucose Co-Transporter 2 inhibitor. RCT: Randomized Controlled Trial.

Table 2A.

Treatment outcomes of the participants in randomized controlled trials.

Study ID BMI
(Kg/m2)
BMI-EOT (Kg/m2) BMI-control (Kg/m2) BMI-control EOT (Kg/m2) ΔBMI
(95% CI) (Kg/m2)
p HbA1c
(mmol/mol))
HbA1c-ΕΟΤ (mmol/mol) HbA1c-control (mmol/mol) HbA1c-control EOT(mmol/mol) ΔHbA1c (95% CI) p
Kudva et al. [11] 34.2(7.1) 35.1(7.5) 35.2(7.7) 35.5(7.6) 0.6(0.2 to 1.0) NA 66.1(15.3) 56.3 (9.8) 65(13.1) 60.7(12) -6.5(-8.7 to -4.4) <0.001
Borel et al. [12] 32(4) NA* 32(4) NA* NA* NA* 62.8(9.9) NA NA NA NA NA
Reznik et al. [13] 32.3 (4.3) NA 35.6(6.5) NA NA** NA 74.9(13.1) 57.4(6.5) 77.6(10.9) 72.7(10.9) -14.2(-20.8 to -7.7) <0.001
Boughton et al. [14] 30.4 (6.5) NA 30.4(6.5) NA NA NA 55.2 (14.2) NA 55.2(14.2) NA NA NA
Daly et al. [15] 35.3 (8.6) NA 35.3(8.6) NA NA NA 74.9(15.3) 57(9) 75(15) 72(13) -15(-20 to -11) <0.001

Data are reported as mean (SD), or median (IQR) unless otherwise indicated. BMI: Body Mass Index. EOT: End Of Treatment. NA: Not Applicable. CI: Confidence Interval.

*Body weight remained unchanged 97(16) Kg, p=0.64

**Δweight (Kg) 2.8 (-0.4 to 6.1), P=0.08.

Table 2B.

Treatment outcomes of the participants in randomized controlled trials.

Study ID TIR
(%)
TIR-EOT(%) TIR-control (%) TIR-control EOT(%) ΔTIR(95% CI) p Insulin dose(u/d) Insulin dose -ΕΟΤ(u/d) Insulin dose-control (u/d) Insulin dose-control EOT (u/d) Δinsulin (95% CI) p
Kudva et al. [11] 48
(24)
64(16) 51(21) 52(21) 14(11 to 17) <0.001 95(47) 87(46) 102(50) 104(56) -10(-20 to -0) NA
Borel et al. [12] 58.9
(20.1)
76(69-84) 58.9(20.1) 61(55-70) 15(8-22) <0.001 78.1(35.9) 103.7(76.1-122.8) 78.1(35.9) 78(59.3-95.3) 10 (3.3-30.6) 0.003
Reznik et al. [13] 34.4
(21.3)
63(9.4) 38.9(23.2) 36.6(22) 27.4(15 to 39.8) <0.001 62.8(37.2) 80.6(61.8) 81.8(38.2) 111.7(72.4) -18.5(-78.2 to 41.2) 0.52
Boughton et al. [14] 37.7
(20.5)
52.8(12.5) 37.7(20.5) 37.7(20.5) 15.1(8 to 22.2) <0.001 0.39* (0.25-0.59) 20.4(9.2-50.3) 0.39* (0.25-0.59) 32.2(12.1-54.4) NA 0.38
Daly et al. [15] NA 66.3(14.9) NA 32.3(24.7) 35.3(28 to 42.6) <0.001 NA 108(73-188) NA 84(54-129) NA NA

Data are reported as mean (SD), or median (IQR) unless otherwise indicated. BMI: Body Mass Index. TIR: Time In Range. EOT: End Of Treatment. NA: Not Applicable. CI: Confidence Interval.

* Reported as units/Kg/day.

In a recent meta-analysis of seven randomized controlled trials in patients with T2DM, a fully closed AID system, when compared with conventional insulin treatment, demonstrated a significant increase in time spent in euglycemia both overnight and during the day, with no difference in time spent in hypoglycemia [16]. However, five of the trials included in this meta-analysis were conducted in hospitalized patients [5-9].

3. NON-RANDOMIZED AND REAL-WORLD STUDIES

The safety and efficacy of AID systems in patients with T2DM have been investigated in a limited number of uncontrolled pilot and real-world studies. In the study by Davis et al., 24 patients with T2DM-mean age 61 years, mean HbA1c 9.4% (79.2 mmol/mol), and an average disease duration of 19 years-who were using basal insulin or an MDI regimen, with or without CGM, were transitioned to AID therapy with the Omnipod 5 hybrid system [17]. After 8 weeks of treatment, the time spent with glucose levels exceeding 250 mg/dL (13.9 mmol/L) decreased by 16.9 percentage points (p < 0.0001). This improvement was accompanied by a significant increase in TIR of 21.9 percentage points (p < 0.0001), while maintaining a low incidence of hypoglycemia, defined as glucose levels below 54 mg/dL (3.0 mmol/L). Additionally, HbA1c was reduced by 1.3 percentage points (14.2 mmol/mol) (p < 0.0001), with no significant changes in body weight. No episodes of severe hypoglycemia or DKA were recorded. Among participants previously treated with MDI, total daily insulin decreased by 32% (p = 0.003). Psychosocial outcomes assessed using validated questionnaires significantly favored AID treatment.

Twenty-two participants continued AID therapy during a 26-week extension phase, achieving a further decrease in time spent with glucose levels above 250 mg/dL (13.9 mmol/L) to 9.7% ± 9.2% [18]. The percentage of time spent < 54 mg/dL (3.0 mmol/L) remained low, while HbA1c decreased by 1.6% ± 1.2% (15.5 ± 13.1 mmol/mol, p < 0.0001), and TIR increased by 22.4% ± 19.2% (p < 0.0001) from standard therapy through the extension period. These improvements occurred without significant changes in body mass index and without an increase in total daily insulin requirements.

A recent study examined the safety and efficacy of the Omnipod 5 system in 305 patients with T2DM, with an average age of 57 years. These individuals were treated with various insulin regimens, including basal insulin alone, MDI, or an insulin pump, with or without CGM. Additionally, some participants used other antidiabetic agents such as GLP-1RA and SGLT-2 inhibitors [19]. After 13 weeks of AID treatment, HbA1c decreased from 8.2% (66.1 mmol/mol) to 7.4% (57.4 mmol/mol) (mean difference −0.8 percentage points [95% CI, −1.0 to −0.7], equivalent to −8.7 mmol/mol [95% CI, −10.9 to −7.6]; p < 0.001 for both noninferiority and superiority). TIR (70-180 mg/dL) (3.9-10.0 mmol/L) increased from 45% to 66% (mean difference 20 percentage points [95% CI, 18 to 22], p < 0.001). The proportion of time spent in hypoglycemia <70 mg/dL (3.9 mmol/L) and <54 mg/dL (3.0 mmol/L) did not differ significantly with AID therapy, although one episode of severe hypoglycemia was reported. The reduction in HbA1c was consistent regardless of concomitant treatment with SGLT-2 inhibitors or GLP-1RA. Body weight increased by 0.8 kg (p < 0.001), while total daily insulin significantly decreased from a mean (SD) of 0.80 (0.46) U/kg/day at baseline to 0.57 (0.29) U/kg/day with AID (p < 0.001). Most patient-reported outcomes assessed in the study favored AID treatment.

Bhargava et al. conducted a study to evaluate the safety and efficacy of the MiniMed 780G hybrid system in 95 patients with T2DM, who had a mean age of 60.3 years and a mean disease duration of 18.6 years [20]. These patients were treated with either MDI or an insulin pump, with or without CGM. After three months of treatment with the hybrid closed-loop insulin delivery system, HbA1c decreased from 7.9% (62.8 mmol/mol) to 7.2% (55.2 mmol/mol) (p < 0.001). TIR (70-180 mg/dL) (3.9-10.0 mmol/L) improved significantly, rising from 72.2% to 79.8% (mean difference 7.2%, p < 0.0001), while the time spent in level 1 and level 2 hypoglycemia remained stable. Total daily insulin requirements increased from 77.4 to 91.8 units (mean difference 13.9 units, p < 0.0001), with no change in body weight. Notably, no cases of severe hypoglycemia, DKA, or hyperosmolar hyperglycemic state were recorded. Psychometric questionnaire scores improved with AID use, including reductions in overall Diabetes Treatment Satisfaction Questionnaire (DTSQ) scores and in perceived frequencies of hyperglycemia and hypoglycemia, indicating enhanced quality of life.

In a real-world study, Breton and Kovatsev collected data over one year from 378 patients with T2DM using a hybrid insulin delivery system featuring the Control-IQ algorithm [21]. Compared with the previous treatment regimen, TIR significantly increased with the AID system, rising from 69.9% to 78.0% (p < 0.001). This improvement was accompanied by a reduction in the duration of hyperglycemia in the 180-250 mg/dL (10.0-13.9 mmol/L) range, which decreased from 23.9% to 19%, and a reduction in time spent with glucose levels exceeding 250 mg/dL (13.9 mmol/L), which fell from 3.6% to 2.3% (p < 0.001). Additionally, time in hypoglycemia below 54 mg/dL (3.0 mmol/L) increased slightly, from 0.00% to 0.04% (p < 0.001). However, time in hypoglycemia within the 54-70 mg/dL (3.0-3.9 mmol/L) range remained unchanged at 0.2%.

Table 3 summarizes the baseline characteristics of participants in non-randomized and real-world studies. The reported outcomes of these studies are presented in Table 4.

Table 3.

Baseline characteristics of the participants in non- randomized and real-world trials.

Study ID AID Duration Sample Size (N) Age (yrs) Diabetes duration (yrs) BMI (Kg/m2) HbA1c (mmol/mol) TIR
(%)
Insulin dose
(units/d)
GLP-1RA N(%) SGLT-2 inh N(%)
Davis et al. [17] Omnipod 5 8 weeks 24 60.6 (8.1) 19(9) 33.5(4.4) 79.2(9.9) 36.6(19.6) 61.5(46.4) 11(46) 6(25)
Davis et al. [18] Omnipod 5 26 weeks 22 61.2(7.7) 18.6(9.4) 33.9(4.4) 79.2(8.8) 36.6(19.6) 64.6(47.2) 9(41) 5(23)
Pasquel et al. [19] Omnipod 5 13 weeks 305 57 (11) 17(11-24) 35 (8) 66.1(14.2) 45 (25) 79.9(50) 168(55) 135(44)
Bhargava et al. [20] Minimed 780G 90 days 95 60.3(10.8) 18.6(8.6) 36(7.4) 62.8(11) 72.2(17.1) 77.4(38.5) 17(18) 20(21)
Breton et al. [21] Control-IQ 12 months 378 NA 20.76(10.3) NA NA 69.9(55.1-82.6) NA NA NA

Data are reported as mean (SD), or median (IQR) unless otherwise indicated. AID: Automated Insulin Delivery. BMI: Body Mass Index. TIR: Time In Range. GLP-1RA: Glucagon-Like Peptide 1 Receptor Agonist. SGLT-2inh: Sodium-Glucose Co-Transporter 2 inhibitor. NA: Not Applicable.

Table 4.

Treatment outcomes of the participants in non-randomized and real-world trials.

Study ID BMI (Kg/m2) BMI-EOT (Kg/m2) ΔBMI (95% CI)) p HbA1c (mmol/mol) HbA1c-EOT (mmol/mol) ΔHbA1c (95% CI) p TIR
(%)
TIR-EOT (%) ΔTIR(95% CI) p Insulin dose
(u/d)
Insulin dose-EOT(u/d) Δinsulin
(95% CI)
p
Davis et al. [17] 33.5(4.4) 33.9(4.3) 0.4*(1.3) 0.15 79(10) 65(8) -14*(8) <0.0001 36.6(19.6) 58.6(15.9) 21.9*(15.2) <0.0001 61.5(46.4) 52.6(34) -8.8(-26.3 to 3.9) 0.07
Davis et al. [18] 33.9(4.4) 31.9(3.9) -0.5*(2.2) 0.51 79(8.7) 62(7.7) -14.2*(7.7) <0.0001 36.6(19.6) 61.4(15.2) 22.4*(19.2) <0.0001 64.6(47.2) 68.7(28.5) 7.4*(20) 0.77
Pasquel et al. [19] 34.9 (7,5)) 35.1(7.6) 0.3(0.1 to 0.4) <0.001 66.1(14.2) 57.4(9.8) -8.7(-10.9 to -7.6) <0.001 45 (25) 66(17) 20(18 to 22) <0.001 79.9(50) 57.3(33) -22.7(-26.7 to-18.7) <0.001
Bhargava et al. [20] 36(7.4) 36.4(7.6) 0.3*(2.4) 0.371 62.8(11) 55.2(7.6) NA NA 72.2(17.1) 79.8(11.4) 7.2*(13.8) <0.0001 77.4(38.5) 91.8(49. 13.9*(28.7) <0.0001
Breton et al. [21] NA NA NA NA NA NA NA NA 69.9(55.1-82.6) 78(66.2-86.1) NA <0.001 NA NA NA NA

Data are reported as mean (SD), or median (IQR) unless otherwise indicated. BMI: Body Mass Index. TIR: Time In Range. NA: Not Applicable. EOT: End Of Treatment. CI: Confidence Interval.

*Reported as mean difference (SD).

4. DISCUSSION

Automated insulin delivery systems for treating insulin-treated patients with T2DM appear to be safe and effective. With the implementation of a hybrid insulin pump, an average reduction in HbA1c of 0.6 to 0.9 percentage points is achieved, along with an increase in TIR (70-180 mg/dL) of 7 to 21 percentage points, equivalent to 1.7 to 5.1 additional hours of euglycemia per day, without an increase in time spent in level 1 or level 2 hypoglycemia. Small absolute increases in time <54 mg/dL, although statistically significant, are clinically negligible and should be interpreted accordingly [21]. The effectiveness of AID treatment in T2DM patients is comparable to results observed in randomized clinical trials involving patients with T1DM [22, 23]. Furthermore, the improvement in glycemic control is achieved safely within the first weeks of treatment, without an increased risk of severe hypoglycemia or DKA. Adverse reactions associated with the devices and infusion systems are also comparable to those seen in patients with T1DM [24, 25].

Interestingly, the improvement in glycemic control appears to be greater in patients with HbA1c >9% (74.9 mmol/mol) when switching to AID [11, 17, 19]. In some studies, a slight increase in body weight has been observed in these patients despite a reduction in daily insulin requirements. This minimal increase may reflect normalization of glucosuria, especially among patients with higher baseline glycated hemoglobin levels, or increased caloric intake [11]. However, attributing this change to “improved control” is speculative, and more studies are needed to address concerns regarding clinically significant weight gain with AID use in patients with T2DM.

Following treatment guidelines, an increasing number of patients with T2DM and a history of cardiovascular disease are being treated with GLP-1RA and/or SGLT-2 inhibitors [26]. However, almost half of the T2DM participants in AID trials were not treated with these classes of cardioprotective anti-diabetic medications. In a randomized placebo-controlled trial in patients with T1DM, semaglutide as an adjunct to AID treatment resulted in significant increases in TIR and decreases in weight and daily insulin needs without increasing time spent in hypoglycemia [27]. In an observational prospective study, tirzepatide, a dual GLP-1/GIP agonist, used as adjunct therapy to AID reduced total daily insulin requirements and weight while increasing TIR in patients with T1DM [28]. In T2DM, concomitant use of GLP-1 RAs or SGLT2i has shown minimal additional benefit to AID efficacy. However, these findings are derived from subgroup analyses with potentially limited power to detect existing differences and therefore warrant further investigation in larger controlled studies [11, 19].

The effectiveness of hybrid pump therapy appears unaffected by prior treatment regimen or CGM use [11, 19]. Although patients with T2DM on insulin are generally older and may be less familiar with technology, structured training enables successful AID implementation in real-world practice [29-34]. Moreover, precise carbohydrate counting does not seem necessary for effective AID use in T2DM, as most trials relied on qualitative meal-size assessments [11, 19].

It is noteworthy that, unlike patients with T1DM-where the increase in TIR during nighttime is significantly greater than during the day-in patients with T2DM, the increase in euglycemia occurs equally during both night hours and daytime hours [35-37]. Patients generally comply well with the treatment, remaining in the system’s auto-mode for more than 90% of the total time.

Studies of AID therapies in patients with T1DM suggest increased treatment satisfaction, improved quality of life, and decreased diabetes-related distress, likely driven by the prevention of hypoglycemia [38-43]. Evidence regarding the impact of diabetes technologies on patient-reported outcome measures (PROMs) in individuals with T2DM remains limited [44]. Available data are mixed: while most studies report improved satisfaction and reduced treatment burden, some document persistent or increased hypoglycemia-related worry [15, 19, 21, 45-47]. This heterogeneity in PROs should be acknowledged, as it highlights the need for further research to better understand patient-specific factors influencing these outcomes.

In this review, we explored the efficacy and safety of AID treatment in patients with T2DM. We focused on clinical trials that enrolled patients already treated with insulin alone or in combination with other oral or injectable antidiabetic agents under real-life conditions. However, many relevant questions remain unanswered. Our findings do not apply to patients with T2DM who are not treated with insulin, as initiation of insulin therapy using AID has not yet been studied. Some trials enrolled special populations, such as patients on dialysis or those requiring residential assistance, and therefore their results may not be generalizable to the broader T2DM population. Adjunctive GLP-1RA or SGLT2i therapy may theoretically enhance outcomes such as weight reduction or lower insulin requirements, but supporting data in T2DM remain insufficient and warrant further investigation.

The cost of AID treatment is high and may pose a barrier to implementation in individuals with T2DM. Additionally, there is a lack of data regarding the cost-effectiveness of AID use in this population. Finally, a comprehensive analysis of the broader impact of diabetes technology on the treatment of patients with T2DM was beyond the scope of this review.

CONCLUSION

AID is an effective and safe treatment option for insulin-treated patients with T2DM who do not achieve their glycemic goals. It significantly decreases HbA1c and increases TIR, while time spent in hypoglycemia remains unchanged. The improvement in glycemic control is accompanied by a slight increase in weight and a modest improvement in quality of life. The need for training in specific technical skills and the associated cost may be potential barriers to AID treatment for many individuals with T2DM. Larger, long-term randomized clinical trials are needed to define the optimal application and cost-effectiveness of various AID algorithms in broader T2DM populations, as well as to identify which patient subgroups stand to benefit most from this technology.

ACKNOWLEDGEMENTS

Declared none.

LIST OF ABBREVIATIONS

AID

Automated Insulin Delivery

T1DM

Type 1 Diabetes Mellitus

T2DM

Type 2 Diabetes Mellitus

HbA1c

Glycosylated Hemoglobin A1c

CI

Confidence Interval

TIR

Time In Range

CV

Coefficient of Variation

GMI

Glucose Management Index

DKA

Diabetic Keto-Acidosis

IQR

Interquartile Range

GLP-1RA

Glucagon-Like Peptide Receptor Agonist

SGLT-2 inh

Sodium-Glucose Co-Transporter 2 Inhibitors

CGM

Continuous Glucose Monitoring

MDI

Multiple Daily Injections

AUTHORS’ CONTRIBUTIONS

The authors confirm their contributions to the paper as follows: K.K.: Study conception and design, and data collection; K.K., C.M.T.: Analysis and interpretation of results; K.K., C.M.T.: Draft manuscript; All authors reviewed the results and approved the final version of the manuscript.

CONSENT FOR PUBLICATION

Not applicable.

FUNDING

None.

CONFLICT OF INTEREST

The authors declare no conflict of interest, financial or otherwise.

REFERENCES

  • 1.Bekiari E., Kitsios K., Thabit H., Tauschmann M., Athanasiadou E., Karagiannis T., Haidich A.B., Hovorka R., Tsapas A. Artificial pancreas treatment for outpatients with type 1 diabetes: Systematic review and meta-analysis. BMJ. 2018;361:k1310. doi: 10.1136/bmj.k1310. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Weisman A., Bai J.W., Cardinez M., Kramer C.K., Perkins B.A. Effect of artificial pancreas systems on glycaemic control in patients with type 1 diabetes: A systematic review and meta-analysis of outpatient randomised controlled trials. Lancet Diabetes Endocrinol. 2017;5(7):501–512. doi: 10.1016/S2213-8587(17)30167-5. [DOI] [PubMed] [Google Scholar]
  • 3.Kitsios K., Trakatelli C.M., Sarigianni M. Hybrid closed-loop systems for the treatment of Type 1 Diabetes: Narrative Review. Endocrinol. Diabetes Metab. J. 2025;9(1):1–6. doi: 10.31038/EDMJ.2025913. [DOI] [Google Scholar]
  • 4.Nauck M.A., Wefers J., Meier J.J. Treatment of type 2 diabetes: Challenges, hopes, and anticipated successes. Lancet Diabetes Endocrinol. 2021;9(8):525–544. doi: 10.1016/S2213-8587(21)00113-3. [DOI] [PubMed] [Google Scholar]
  • 5.Thabit H., Hartnell S., Allen J.M., Lake A., Wilinska M.E., Ruan Y., Evans M.L., Coll A.P., Hovorka R. Closed-loop insulin delivery in inpatients with type 2 diabetes: A randomised, parallel-group trial. Lancet Diabetes Endocrinol. 2017;5(2):117–124. doi: 10.1016/S2213-8587(16)30280-7. [DOI] [PubMed] [Google Scholar]
  • 6.Bally L., Thabit H., Hartnell S., Andereggen E., Ruan Y., Wilinska M.E., Evans M.L., Wertli M.M., Coll A.P., Stettler C., Hovorka R. Closed-loop insulin delivery for glycemic control in noncritical care. N. Engl. J. Med. 2018;379(6):547–556. doi: 10.1056/NEJMoa1805233. [DOI] [PubMed] [Google Scholar]
  • 7.Boughton C.K., Bally L., Martignoni F., Hartnell S., Herzig D., Vogt A., Wertli M.M., Wilinska M.E., Evans M.L., Coll A.P., Stettler C., Hovorka R. Fully closed-loop insulin delivery in inpatients receiving nutritional support: A two-centre, open-label, randomised controlled trial. Lancet Diabetes Endocrinol. 2019;7(5):368–377. doi: 10.1016/S2213-8587(19)30061-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Herzig D., Suhner S., Roos J., Schürch D., Cecchini L., Nakas C.T., Weiss S., Kadner A., Kocher G.J., Guensch D.P., Wilinska M.E., Raabe A., Siebenrock K.A., Beldi G., Gloor B., Hovorka R., Vogt A.P., Bally L. Perioperative fully closed-loop insulin delivery in patients undergoing elective surgery: An open-label, randomised controlled trial. Diabetes Care. 2022;45(9):2076–2083. doi: 10.2337/dc22-0438. [DOI] [PubMed] [Google Scholar]
  • 9.Kumareswaran K., Thabit H., Leelarathna L., Caldwell K., Elleri D., Allen J.M., Nodale M., Wilinska M.E., Evans M.L., Hovorka R. Feasibility of closed-loop insulin delivery in type 2 diabetes: A randomized controlled study. Diabetes Care. 2014;37(5):1198–1203. doi: 10.2337/dc13-1030. [DOI] [PubMed] [Google Scholar]
  • 10.Taleb N., Carpentier A.C., Messier V., Ladouceur M., Haidar A., Rabasa-Lhoret R. Efficacy of artificial pancreas use in patients with type 2 diabetes using intensive insulin therapy: A randomised crossover pilot trial. Diabetes Care. 2019;42(7):e107–e109. doi: 10.2337/dc18-2406. [DOI] [PubMed] [Google Scholar]
  • 11.Kudva Y.C., Raghinaru D., Lum J.W., Graham T.E., Liljenquist D., Spanakis E.K., Pasquel F.J., Ahmann A., Ahn D.T., Aleppo G., Blevins T., Kruger D., Brown S.A., Levy C.J., Weinstock R.S., Steenkamp D.W., Spaic T., Hirsch I.B., Broyles F., Rickels M.R., Tsoukas M.A., Raskin P., Hatipoglu B., Desjardins D., Terry A.N., Singh L.G., Davis G.M., Schmid C., Kravarusic J., Coyne K., Casaubon L., Espinosa V., Jones J.K., Estrada K., Afreen S., Levister C., O’Malley G., Liu S.L., Marks S., Peleckis A.J., Pasqua M.R., Tardio V., Kurek C., Luker R.D., Churchill J., Tajrishi F.Z., Dean A., Dennis B., Fronczyk E., Perez J., Mukhashen S., Dhillon J., Ipek A., Bzdick S., Atakov Castillo A., Driscoll M., Averkiou X., Dalton-Bakes C.V., Moore A., Jordan L.F., Lesniak A., Pinsker J.E., Sasson-Katchalski R., Campos T., Spanbauer C., Kanapka L., Kollman C., Beck R.W. A randomised trial of Automated Insulin Delivery in type 2 diabetes. N. Engl. J. Med. 2025;392(18):1801–1812. doi: 10.1056/NEJMoa2415948. [DOI] [PubMed] [Google Scholar]
  • 12.Borel A.L., Lablanche S., Waterlot C., Joffray E., Barra C., Arnol N., Amougay H., Benhamou P.Y. Closed-Loop insulin therapy for people with type 2 diabetes treated with an insulin pump: A 12-week multicenter, open-label randomised, controlled, crossover trial. Diabetes Care. 2024;47(10):1778–1786. doi: 10.2337/dc24-0623. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Reznik Y., Carvalho M., Fendri S., Prevost G., Chaillous L., Riveline J.P., Hanaire H., Dubois S., Houéto P., Pasche H., Mianowska B., Renard E. Should people with type 2 diabetes treated by multiple daily insulin injections with home health care support be switched to hybrid closed-loop? The CLOSE AP + randomized controlled trial. Diabetes Obes. Metab. 2024;26(2):622–630. doi: 10.1111/dom.15351. [DOI] [PubMed] [Google Scholar]
  • 14.Boughton C.K., Tripyla A., Hartnell S., Daly A., Herzig D., Wilinska M.E., Czerlau C., Fry A., Bally L., Hovorka R. Fully automated closed-loop glucose control compared with standard insulin therapy in adults with type 2 diabetes requiring dialysis: An open-label, randomized crossover trial. Nat. Med. 2021;27(8):1471–1476. doi: 10.1038/s41591-021-01453-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Daly A.B., Boughton C.K., Nwokolo M., Hartnell S., Wilinska M.E., Cezar A., Evans M.L., Hovorka R. Fully automated closed-loop insulin delivery in adults with type 2 diabetes: An open-label, single-center, randomized crossover trial. Nat. Med. 2023;29(1):203–208. doi: 10.1038/s41591-022-02144-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Amer B.E., Yaqout Y.E., Abozaid A.M., Afifi E., Aboelkhier M.M. Does fully closed-loop automated insulin delivery improve glycaemic control in patients with type 2 diabetes? A meta-analysis of randomized controlled trials. Diabet. Med. 2024;41(1):15196. doi: 10.1111/dme.15196. [DOI] [PubMed] [Google Scholar]
  • 17.Davis G.M., Peters A.L., Bode B.W., Carlson A.L., Dumais B., Vienneau T.E., Huyett L.M., Ly T.T. Safety and efficacy of the omnipod 5 automated insulin delivery system in adults with type 2 diabetes: From injections to hybrid closed-loop therapy. Diabetes Care. 2023;46(4):742–750. doi: 10.2337/dc22-1915. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Davis G.M., Peters A.L., Bode B.W., Carlson A.L., Dumais B., Vienneau T.E., Huyett L.M., Ly T.T. Glycaemic outcomes in adults with type 2 diabetes over 34 weeks with the Omnipod® 5 Automated Insulin Delivery System. Diabetes Obes. Metab. 2025;27(1):143–154. doi: 10.1111/dom.15993. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Pasquel F.J., Davis G.M., Huffman D.M., Peters A.L., Parker J.C., Laffel L.M., Romeo G.R., Mathew J., Castorino K.N., Kruger D.F., Dungan K.M., Kipnes M., Jauch E.C., Oser T.K., Shah V.N., Horowitz B., Carlson A.L., Warren M.L., Deeb W., Buse J.B., Reed J.H., Berner J., Blevins T., Bajaj C., Kollman C., Raghinaru D., Ly T.T., Beck R.W., Guerrero-Arroyo L., Hughes-Palmer J., Walker M., Cannon K., Wagner A., Isganaitis E., Doolan J., Agarwal S., Zybert D., Zavala N., Church M.M., Gonzales K., Shuirman M., Estrada K., Jones J.K., Cushman T., Muhkahsen S., Faulds E., Aldrich L., Beltran S., Lane W., Wells C., Kaan D., Duncan R., Friedman S., Oser S.M., Kramer E.S., Huss K., Kaye W., Amole M., Diazgranados S., Bergenstal R.M., Martens T.W., Carlson M.J., Malaeb S., Lage H., Smith L., Dodis R., Hager M., Kirk E.A., Choate L., Brouillet C., Goldfaden R., Schaffner H.N., Niman S., Domingo D., Klein K., Young L., Ghorbani Rodriguez T., Dostou J., Diner J., Coviello A., Curlin E., Borkovich E.C., Tapia J., Korey L., Reinecke K., Caldwell S., Espinosa V., Casaubon L., Pandit K., Perez J., Tan A., Padilla C., Ruedy K.J., Dumais B., Namati J., Vienneau T., Miller K.M., Huyett L.M., Conroy L.R. Automated Insulin Delivery in adults with type 2 diabetes. A non randomised clinical trial. JAMA Netw. Open. 2025;8(2):2459348. doi: 10.1001/jamanetworkopen.2024.59348. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Bhargava A., Bergenstal R.M., Warren M.L., Thrasher J.R., Dempsey M.A., Bode B.W., LaRocque J., Carlson A.L., Keiter A., Ma H., Shin J.J., McVean J.J., Cordero T.L., Rhinehart A.S., Vigersky R.A. Safety and effectiveness of MiniMed 780G Advanced Hybrid Closed-Loop insulin intensification in adults with insulin-requiring type 2 diabetes. Diabetes Technol. Ther. 2025;27(5):366–375. doi: 10.1089/dia.2024.0586. [DOI] [PubMed] [Google Scholar]
  • 21.Breton M.D., Kovatchev B.P. One year real-world use of the control-iq advanced hybrid closed-loop technology. Diabetes Technol. Ther. 2021;23(9):601–608. doi: 10.1089/dia.2021.0097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Brown S.A., Kovatchev B.P., Raghinaru D., Lum J.W., Buckingham B.A., Kudva Y.C., Laffel L.M., Levy C.J., Pinsker J.E., Wadwa R.P., Dassau E., Doyle F.J., III, Anderson S.M., Church M.M., Dadlani V., Ekhlaspour L., Forlenza G.P., Isganaitis E., Lam D.W., Kollman C., Beck R.W. Six-month randomised, multicenter trial of closed-loop control in type 1 diabetes. N. Engl. J. Med. 2019;381(18):1707–1717. doi: 10.1056/NEJMoa1907863. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Russell S.J., Beck R.W., Damiano E.R., El-Khatib F.H., Ruedy K.J., Balliro C.A., Li Z., Calhoun P., Wadwa R.P., Buckingham B., Zhou K., Daniels M., Raskin P., White P.C., Lynch J., Pettus J., Hirsch I.B., Goland R., Buse J.B., Kruger D., Mauras N., Muir A., McGill J.B., Cogen F., Weissberg-Benchell J., Sherwood J.S., Castellanos L.E., Hillard M.A., Tuffaha M., Putman M.S., Sands M.Y., Forlenza G., Slover R., Messer L.H., Cobry E., Shah V.N., Polsky S., Lal R. Multicenter, randomized trial of a bionic pancreas in type 1 diabetes. N Engl J Med. 2022;387(13):1161–1172. doi: 10.1056/NEJMoa2205225. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Krouwer J.S. Adverse event data for years 2018 to 2020 for diabetes devices. J. Diab. Sci. Technol. 2022;16(5):1299–1302. doi: 10.1177/19322968211011688. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Krouwer J.S. An analysis of 2019 FDA adverse events for two insulin pumps and two continuous glucose monitors. J. Diabetes Sci. Technol. 2022;16(1):228–232. doi: 10.1177/1932296820951872. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.ElSayed N.A., McCoy R.G., Aleppo G., Bajaj M., Balapattabi K., Beverly E.A., Briggs Early K., Bruemmer D., Echouffo-Tcheugui J.B., Ekhlaspour L., Gaglia J.L., Garg R., Girotra M., Khunti K., Lal R., Lingvay I., Matfin G., Neumiller J.J., Pandya N., Pekas E.J., Pilla S.J., Polsky S., Segal A.R., Seley J.J., Stanton R.C., Bannuru R.R. 9. Pharmacologic approaches to glycemic treatment: Standards of care in diabetes—2025. Diabetes Care. 2025;48(1):S181–S206. doi: 10.2337/dc25-S009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Pasqua M.R., Tsoukas M.A., Kobayati A., Aboznadah W., Jafar A., Haidar A. Subcutaneous weekly semaglutide with automated insulin delivery in type 1 diabetes: A double-blind, randomized, crossover trial. Nat. Med. 2025;31(4):1239–1245. doi: 10.1038/s41591-024-03463-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Karakus K.E., Klein M.P., Akturk H.K., Shah V.N. Changes in basal and bolus insulin requirements with tirzepatide as an adjunctive therapy in adults with type 1 diabetes using tandem control-IQ. Diabetes Ther. 2024;15(7):1647–1655. doi: 10.1007/s13300-024-01592-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Liarakos A.L., Lim J.Z.M., Leelarathna L., Wilmot E.G. The use of technology in type 2 diabetes and prediabetes: A narrative review. Diabetologia. 2024;67(10):2059–2074. doi: 10.1007/s00125-024-06203-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Barchiesi M.A., Calabrese A., Costa R., Di Pillo F., D’Uffizi A., Tiburzi L., Zahid E. Continuous glucose monitoring in type 2 diabetes: A systematic review of barriers and opportunities for care improvement. Int. J. Qual. Health Care. 2025;37(3):mzaf046. doi: 10.1093/intqhc/mzaf046. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Jancev M., Vissers T.A.C.M., Visseren F.L.J., van Bon A.C., Serné E.H., DeVries J.H., de Valk H.W., van Sloten T.T. Continuous glucose monitoring in adults with type 2 diabetes: A systematic review and meta-analysis. Diabetologia. 2024;67(5):798–810. doi: 10.1007/s00125-024-06107-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Ajjan R.A., Battelino T., Cos X., Del Prato S., Philips J.C., Meyer L., Seufert J., Seidu S. Continuous glucose monitoring for the routine care of type 2 diabetes mellitus. Nat. Rev. Endocrinol. 2024;20(7):426–440. doi: 10.1038/s41574-024-00973-1. [DOI] [PubMed] [Google Scholar]
  • 33.Jackson M.A., Ahmann A., Shah V.N. Type 2 diabetes and the use of real-time continuous glucose monitoring. Diabetes Technol. Ther. 2021;23(S1):S-27–S-34. doi: 10.1089/dia.2021.0007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Daly A., Hovorka R. Technology in the management of type 2 diabetes: Present status and future prospects. Diabetes Obes. Metab. 2021;23(8):1722–1732. doi: 10.1111/dom.14418. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Thabit H., Lubina-Solomon A., Stadler M., Leelarathna L., Walkinshaw E., Pernet A., Allen J.M., Iqbal A., Choudhary P., Kumareswaran K., Nodale M., Nisbet C., Wilinska M.E., Barnard K.D., Dunger D.B., Heller S.R., Amiel S.A., Evans M.L., Hovorka R. Home use of closed-loop insulin delivery for overnight glucose control in adults with type 1 diabetes: A 4-week, multicentre, randomised crossover study. Lancet Diabetes Endocrinol. 2014;2(9):701–709. doi: 10.1016/S2213-8587(14)70114-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Collyns O.J., Meier R.A., Betts Z.L., Chan D.S.H., Frampton C., Frewen C.M., Hewapathirana N.M., Jones S.D., Roy A., Grosman B., Kurtz N., Shin J., Vigersky R.A., Wheeler B.J., de Bock M.I. Improved glycemic outcomes with Medtronic MiniMed advanced hybrid closed-loop delivery: Results from a randomized crossover trial comparing automated insulin delivery with predictive low glucose suspend in people with type 1 diabetes. Diabetes Care. 2021;44(4):969–975. doi: 10.2337/dc20-2250. [DOI] [PubMed] [Google Scholar]
  • 37.Carlson A.L., Sherr J.L., Shulman D.I., Garg S.K., Pop-Busui R., Bode B.W., Lilenquist D.R., Brazg R.L., Kaiserman K.B., Kipnes M.S., Thrasher J.R., Reed J.H.C., Slover R.H., Philis-Tsimikas A., Christiansen M., Grosman B., Roy A., Vella M., Jonkers R.A.M., Chen X., Shin J., Cordero T.L., Lee S.W., Rhinehart A.S., Vigersky R.A. Safety and glycemic outcomes during the MiniMed advanced hybrid closed-loop system pivotal trial in adolescents and adults with type 1 diabetes. Diabetes Technol. Ther. 2022;24(3):178–189. doi: 10.1089/dia.2021.0319. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Hermanns N., Kulzer B., Ehrmann D. Person-reported outcomes in diabetes care: What are they and why are they so important? Diabetes Obes. Metab. 2024;26(S1):30–45. doi: 10.1111/dom.15471. [DOI] [PubMed] [Google Scholar]
  • 39.Barnard-Kelly K., Marrero D., de Wit M., Pouwer F., Khunti K., Hermans N., Pierce J.S., Laffel L., Holt R.I.G., Battelino T., Naranjo D., Fosbury J., Fisher L., Polonsky W., Weissberg-Benchell J., Hood K.K., Schnell O., Messer L.H., Danne T., Nimri R., Skovlund S.E., Mader J.K., Sherr J.L., Schatz D., O’Neill S., Doble E., Town M., Lange K., de Beaufort C., Gonder-Frederick L., Jaser S.S., Liberman A., Klonoff D., ElSayed N.A., Bannuru R.R., Parkin C.G., Snoek F. Towards the standardisation of adult person-reported outcome domains in diabetes research: A Consensus Statement development panel. Diabet. Med. 2024;41(8):15332. doi: 10.1111/dme.15332. [DOI] [PubMed] [Google Scholar]
  • 40.Schneider-Utaka A.K., Hanes S., Boughton C.K., Hartnell S., Thabit H., Mubita W.M., Draxlbauer K., Poettler T., Hayes J., Wilinska M.E., Mader J.K., Narendran P., Leelarathna L., Evans M.L., Hovorka R., Hood K.K. Patient-reported outcomes for older adults on CamAPS FX closed loop system. Diabet. Med. 2023;40(9):15126. doi: 10.1111/dme.15126. [DOI] [PubMed] [Google Scholar]
  • 41.Polonsky W.H., Hood K.K., Levy C.J., MacLeish S.A., Hirsch I.B., Brown S.A., Bode B.W., Carlson A.L., Shah V.N., Weinstock R.S., Bhargava A., Jones T.C., Aleppo G., Mehta S.N., Laffel L.M., Forlenza G.P., Sherr J.L., Huyett L.M., Vienneau T.E., Ly T.T. How introduction of automated insulin delivery systems may influence psychosocial outcomes in adults with type 1 diabetes: Findings from the first investigation with the Omnipod® 5 System. Diabetes Res. Clin. Pract. 2022;190:109998. doi: 10.1016/j.diabres.2022.109998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Pinsker J.E., Müller L., Constantin A., Leas S., Manning M., McElwee Malloy M., Singh H., Habif S. Real-world patient-reported outcomes and glycemic results with initiation of control-IQ technology. Diabetes Technol. Ther. 2021;23(2):120–127. doi: 10.1089/dia.2020.0388. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Beato-Víbora P.I., Chico A., Moreno-Fernandez J., Bellido-Castañeda V., Nattero-Chávez L., Picón-César M.J., Martínez-Brocca M.A., Giménez-Álvarez M., Aguilera-Hurtado E., Climent-Biescas E., Azriel-Mir S., Rebollo-Román Á., Yoldi-Vergara C., Pazos-Couselo M., Alonso-Carril N., Quirós C. A multicenter prospective evaluation of the benefits of two advanced hybrid closed-loop systems in glucose control and patient-reported outcomes in a real-world setting. Diabetes Care. 2024;47(2):216–224. doi: 10.2337/dc23-1355. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Liarakos A.L., Crabtree T.S.J., Wilmot E.G. Patient-reported outcomes in studies of diabetes technology: What matters. Diabetes Obes. Metab. 2024;26(S7):59–73. doi: 10.1111/dom.15858. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Fisher L., Polonsky W.H., Perez-Nieves M., Desai U., Strycker L., Hessler D. A new perspective on diabetes distress using the type 2 diabetes distress assessment system (T2-DDAS): Prevalence and change over time. J. Diabetes Complications. 2022;36(8):108256. doi: 10.1016/j.jdiacomp.2022.108256. [DOI] [PubMed] [Google Scholar]
  • 46.Buysse D.J., Reynolds C.F., III, Monk T.H., Berman S.R., Kupfer D.J. The Pittsburgh sleep quality index: A new instrument for psychiatric practice and research. Psychiatry Res. 1989;28(2):193–213. doi: 10.1016/0165-1781(89)90047-4. [DOI] [PubMed] [Google Scholar]
  • 47.Polonsky W.H., Fisher L., Hessler D., Edelman S.V. Investigating hypoglycemic confidence in type 1 and type 2 diabetes. Diabetes Technol. Ther. 2017;19(2):131–136. doi: 10.1089/dia.2016.0366. [DOI] [PubMed] [Google Scholar]

Articles from Current Diabetes Reviews are provided here courtesy of Bentham Science Publishers