Abstract
The influence of endoscopic forceps variables (size, design, and make) on biopsy specimen weight, depth, and diagnostic adequacy has been studied in vitro on normal human stomach, and in vivo at endoscopy in dog and in patients. Greater pressure during the biopsy procedure yielded significantly heavier, deeper, and histologically more acceptable specimens. Standard sized forceps (diameter 2.4 mm) and those with ellipsoid cups consistently produced larger specimens than the smaller 'paediatric' forceps (1.8 mm) and those with round cups. Deeper specimens were also obtained using the standard forceps. In vitro and in vivo in the dog, the standard sized forceps also produced specimens of greater diagnostic adequacy than the paediatric forceps. At endoscopy in patients, however, we could not detect any influence of the size, shape, and presence of forceps spike or fenestration on the diagnostic adequacy of the specimens.
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Dekker W., Tytgat G. N. Diagnostic accuracy of fiberendoscopy in the detection of upper intestinal malignancy. A follow-up analysis. Gastroenterology. 1977 Oct;73(4 Pt 1):710–714. [PubMed] [Google Scholar]
- Graham D. Y., Schwartz J. T., Cain G. D., Gyorkey F. Prospective evaluation of biopsy number in the diagnosis of esophageal and gastric carcinoma. Gastroenterology. 1982 Feb;82(2):228–231. [PubMed] [Google Scholar]
- Misumi A., Mori K., Ikeda T., Misumi K., Ookubo F., Shimamoto M., Nakashima K., Fujishima K., Yagi Y., Takano S. Evaluation of fibergastroscopic biopsy in the diagnosis of gastric cancer: a study of 339 cases. Gastroenterol Jpn. 1978;13(4):255–263. doi: 10.1007/BF02774048. [DOI] [PubMed] [Google Scholar]
- Siegel M., Barkin J. S., Rogers A. I., Thomsen S., Clark R. Gastric biopsy: a comparison of biopsy forceps. Gastrointest Endosc. 1983 Feb;29(1):35–36. doi: 10.1016/s0016-5107(83)72498-3. [DOI] [PubMed] [Google Scholar]