Skip to main content
Annals of Surgery logoLink to Annals of Surgery
. 1987 Sep;206(3):335–339. doi: 10.1097/00000658-198709000-00011

Blood flow in colonic anastomoses. Effect of stapling and suturing.

R S Chung
PMCID: PMC1493166  PMID: 3307654

Abstract

Sutureline blood flow was measured with laser Doppler velocimetry in colonic anastomoses created with the stapler, manual suturing, or a combination. Blood flow was always reduced in the sutureline compared with normal mucosa. Of all the anastomoses studied, tight stapling reduced sutureline blood flow the most, whereas the two-layered manual anastomosis or stapling reinforced with sutures were somewhat less ischemic. Stapling adjusted to bowel wall thickness impaired flow only moderately. It was possible with practice to outperform the stapler by single-layered manual anastomosis using fine sutures. In humans, stapled ileocolostomy had a higher sutureline blood flow than the two-layered manual anastomosis. In view of the existing clinical experience of safe stapling without adjustment for bowel wall thickness, a low sutureline blood flow is probably tolerated to a considerable degree in humans. However, this study clearly shows that tight stapling can reduce sutureline blood flow unduly, whereas superior blood supply can be attained by attention to staple closure height relative to bowel wall thickness.

Full text

PDF
336

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Ahn H., Lindhagen J., Nilsson G. E., Salerud E. G., Jodal M., Lundgren O. Evaluation of laser Doppler flowmetry in the assessment of intestinal blood flow in cat. Gastroenterology. 1985 Apr;88(4):951–957. doi: 10.1016/s0016-5085(85)80013-5. [DOI] [PubMed] [Google Scholar]
  2. Brolin R. E., Ravitch M. M. Studies in intestinal healing VI. Effect of pharmacologically induced peristalsis on fresh intestinal anastomoses in dogs. Arch Surg. 1980 Mar;115(3):339–343. doi: 10.1001/archsurg.1980.01380030083020. [DOI] [PubMed] [Google Scholar]
  3. Bubrick M. P. Effect of technique on anastomotic dehiscence. Dis Colon Rectum. 1981 May-Jun;24(4):232–233. doi: 10.1007/BF02641865. [DOI] [PubMed] [Google Scholar]
  4. Chung R. S., Safaie-Shirazi S. The effect of secretin on pancreatic blood flow in the awake and anesthetized dog. Proc Soc Exp Biol Med. 1983 Sep;173(4):620–625. doi: 10.3181/00379727-173-41696. [DOI] [PubMed] [Google Scholar]
  5. Feld A. D., Fondacaro J. D., Holloway G. A., Jr, Jacobson E. D. Laser Doppler velocimetry: a new technique for the measurement of intestinal mucosal blood flow. Gastrointest Endosc. 1984 Aug;30(4):225–230. doi: 10.1016/s0016-5107(84)72391-1. [DOI] [PubMed] [Google Scholar]
  6. GOLDMAN A. AN EVALUATION OF AUTOMATIC SUTURE WITH UKL-60 AND UKL-40 DEVICES BY PULMONARY RESECTION. Dis Chest. 1964 Jul;46:29–36. doi: 10.1378/chest.46.1.29. [DOI] [PubMed] [Google Scholar]
  7. Marston A. Diagnosis and management of intestinal ischaemia. Ann R Coll Surg Engl. 1972 Jan;50(1):29–44. [PMC free article] [PubMed] [Google Scholar]
  8. Polglase A. L., Hughes E. S., McDermott F. T., Pihl E., Burke F. R. A comparison of end-to-end staple and suture colorectal anastomosis in the dog. Surg Gynecol Obstet. 1981 Jun;152(6):792–796. [PubMed] [Google Scholar]
  9. Ravitch M. M., Brolin R., Kolter J., Yap S. Studies in the healing of intestinal anastomoses. World J Surg. 1981 Jul;5(4):627–637. doi: 10.1007/BF01655019. [DOI] [PubMed] [Google Scholar]
  10. Shepherd A. P., Riedel G. L. Continuous measurement of intestinal mucosal blood flow by laser-Doppler velocimetry. Am J Physiol. 1982 Jun;242(6):G668–G672. doi: 10.1152/ajpgi.1982.242.6.G668. [DOI] [PubMed] [Google Scholar]
  11. VEST B., MARGULIS A. R. EXPERIMENTAL INFARCTION OF SMALL BOWEL IN DOGS. ROENTGENOGRAPHIC DEMONSTRATION WITH WATER-SOLUBLE IODINE CONTRAST MEDIUM. Am J Roentgenol Radium Ther Nucl Med. 1964 Nov;92:1080–1087. [PubMed] [Google Scholar]

Articles from Annals of Surgery are provided here courtesy of Lippincott, Williams, and Wilkins

RESOURCES