Skip to main content
Thorax logoLink to Thorax
. 1997 Dec;52(12):1068–1073. doi: 10.1136/thx.52.12.1068

Use of home sleep studies for diagnosis of the sleep apnoea/hypopnoea syndrome

A T Whittle, S P Finch, I L Mortimore, T W MacKay, N J Douglas
PMCID: PMC1758463  PMID: 9516901

Abstract

BACKGROUND: A study was undertaken to test the hypothesis that unsupervised domiciliary limited sleep studies do not impair the accuracy of diagnosis when used to investigate the sleep apnoea/hypopnoea syndrome (SAHS) and can be cheaper than laboratory polysomnography. METHODS: For validation, 23 subjects with suspected SAHS underwent laboratory polysomnography and a home study (EdenTec 3711) on successive nights. All subjects with > 15 apnoeas + hypopnoeas (A + H)/hour on polysomnography showed > 30 A + H/hour on their home study. Thereafter, in a prospective trial 150 subjects had a home study as the initial investigation and studies showing > 30 events/hour were regarded as diagnostic of SAHS. Those showing fewer events were investigated with polysomnography if necessary. Time to treatment, outcome, and costs of this protocol were compared with those of 75 patients investigated initially with polysomnography. RESULTS: Of the prospective trial subjects, 29% had > 30 A + H/hour and proceeded directly from home study to treatment; 15% without daytime sleepiness were not investigated further. Polysomnography was undertaken to establish a diagnosis in 56% of cases, including 18% whose home studies were unsuccessful. Compared with the 75 control patients, this protocol gave a diagnosis faster (median 18 (range 0-221) versus 47 (0-227) days, p < 0.001) and more cheaply (mean (SD) 164 pounds (104) versus 210 pounds (0), p < 0.001). The proportions offered CPAP (61% versus 67%) and subsequent objective CPAP usage (mean 4.7 (2.4) versus 5.0 (2.4) hours/night) were not different. CONCLUSIONS: Use of home sleep studies has benefits in time and cost. For diagnostic reliability a further sleep study was required in 56% of cases. 




Full Text

The Full Text of this article is available as a PDF (140.8 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Coppola M. P., Lawee M. Management of obstructive sleep apnea syndrome in the home. The role of portable sleep apnea recording. Chest. 1993 Jul;104(1):19–25. doi: 10.1378/chest.104.1.19. [DOI] [PubMed] [Google Scholar]
  2. Douglas N. J., Thomas S., Jan M. A. Clinical value of polysomnography. Lancet. 1992 Feb 8;339(8789):347–350. doi: 10.1016/0140-6736(92)91660-z. [DOI] [PubMed] [Google Scholar]
  3. Gould G. A., Whyte K. F., Rhind G. B., Airlie M. A., Catterall J. R., Shapiro C. M., Douglas N. J. The sleep hypopnea syndrome. Am Rev Respir Dis. 1988 Apr;137(4):895–898. doi: 10.1164/ajrccm/137.4.895. [DOI] [PubMed] [Google Scholar]
  4. Gyulay S., Gould D., Sawyer B., Pond D., Mant A., Saunders N. Evaluation of a microprocessor-based portable home monitoring system to measure breathing during sleep. Sleep. 1987 Apr;10(2):130–142. doi: 10.1093/sleep/10.2.130. [DOI] [PubMed] [Google Scholar]
  5. Hida W., Shindoh C., Miki H., Kikuchi Y., Okabe S., Taguchi O., Takishima T., Shirato K. Prevalence of sleep apnea among Japanese industrial workers determined by a portable sleep monitoring system. Respiration. 1993;60(6):332–337. doi: 10.1159/000196231. [DOI] [PubMed] [Google Scholar]
  6. Johns M. W. A new method for measuring daytime sleepiness: the Epworth sleepiness scale. Sleep. 1991 Dec;14(6):540–545. doi: 10.1093/sleep/14.6.540. [DOI] [PubMed] [Google Scholar]
  7. Man G. C., Kang B. V. Validation of a portable sleep apnea monitoring device. Chest. 1995 Aug;108(2):388–393. doi: 10.1378/chest.108.2.388. [DOI] [PubMed] [Google Scholar]
  8. Meyer T. J., Eveloff S. E., Kline L. R., Millman R. P. One negative polysomnogram does not exclude obstructive sleep apnea. Chest. 1993 Mar;103(3):756–760. doi: 10.1378/chest.103.3.756. [DOI] [PubMed] [Google Scholar]
  9. Redline S., Tosteson T., Boucher M. A., Millman R. P. Measurement of sleep-related breathing disturbances in epidemiologic studies. Assessment of the validity and reproducibility of a portable monitoring device. Chest. 1991 Nov;100(5):1281–1286. doi: 10.1378/chest.100.5.1281. [DOI] [PubMed] [Google Scholar]
  10. Stoohs R., Guilleminault C. MESAM 4: an ambulatory device for the detection of patients at risk for obstructive sleep apnea syndrome (OSAS). Chest. 1992 May;101(5):1221–1227. doi: 10.1378/chest.101.5.1221. [DOI] [PubMed] [Google Scholar]
  11. White D. P., Gibb T. J., Wall J. M., Westbrook P. R. Assessment of accuracy and analysis time of a novel device to monitor sleep and breathing in the home. Sleep. 1995 Feb;18(2):115–126. doi: 10.1093/sleep/18.2.115. [DOI] [PubMed] [Google Scholar]
  12. Wittig R. M., Romaker A., Zorick F. J., Roehrs T. A., Conway W. A., Roth T. Night-to-night consistency of apneas during sleep. Am Rev Respir Dis. 1984 Feb;129(2):244–246. [PubMed] [Google Scholar]
  13. Young T., Palta M., Dempsey J., Skatrud J., Weber S., Badr S. The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med. 1993 Apr 29;328(17):1230–1235. doi: 10.1056/NEJM199304293281704. [DOI] [PubMed] [Google Scholar]

Articles from Thorax are provided here courtesy of BMJ Publishing Group