Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2007 Oct 1.
Published in final edited form as: J Magn Reson. 2007 Jan 26;186(1):75–85. doi: 10.1016/j.jmr.2007.01.015

Artifacts in T1ρ-Weighted Imaging: Compensation for B1 and B0 Field Imperfections

Walter RT Witschey II 1, Arijitt Borthakur 1,2, Mark A Elliott 1,2, Eric Mellon 1,2, Sampreet Niyogi 2,3, Daniel J Wallman 2, Chenyang Wang 2,3, Ravinder Reddy 1,2
PMCID: PMC1995435  NIHMSID: NIHMS23183  PMID: 17291799

Abstract

The origin of spin locking image artifacts in the presence of B0 and B1 magnetic field imperfections is shown theoretically using the Bloch equations and experimentally at low (ω1 ≪ Δω0), intermediate (ω1 ~ Δω0) and high (ω1 ≫ Δω0) spin locking field strengths. At low spin locking fields, the magnetization is shown to oscillate about an effective field in the rotating frame causing signature banding artifacts in the image. At high spin lock fields, the effect of the resonance offset Δω0 is quenched, but imperfections in the flip angle cause oscillations about the ω1 field. A new pulse sequence is presented that consists of an integrated spin echo and spin lock experiment followed by magnetization storage along the -z-axis. It is shown that this sequence almost entirely eliminates banding artifacts from both types of field inhomogeneities at all spin locking field strengths. The sequence was used to obtain artifact free images of agarose in inhomogeneous B0 and B1 fields, off-resonance spins in fat and in vivo human brain images at 3T. The new pulse sequence can be used to probe very low frequency (0–400 Hz) dynamic and static interactions in tissues without contaminating B0 and B1 field artifacts.

Keywords: spin locking, T1ρ, relaxation, off-resonance T1ρ, inhomogeneous B0 and B1 fields

1. Introduction

Magnetic resonance (MR) tissue contrast depends on differences in tissue relaxation times T1 and T2, diffusion-weighting, magnetization transfer (MT) or perfusion effects to distinguish healthy and diseased tissues. In addition to these conventional contrast techniques, a powerful method to create tissue contrast is the spin-lattice relaxation time in the rotating frame (T1ρ) characterized first in spectroscopic experiments by Redfield [1].

T1ρ-weighted contrast is obtained by allowing magnetization to relax under the influence of an on-resonance, continuous wave (cw) radiofrequency (RF) pulse. T1ρ-weighted contrast is sensitive to both low frequency motional processes and static processes. Low frequency motional processes in tissues include proton exchange with hydroxyl or amide groups in proteins [24] while static processes include static dipolar interactions [5]. In particular, T1ρ-weighted contrast has distinguished early acute cerebral ischemia in rats [68], human gliomas [9] and tumor in breast tissues [10], tracked the early degeneration of cartilage in osteoarthritis [11] and the nucleus pulposus of lumbar intervertebral discs [12] and indirectly detected metabolic H217O in vivo [1316].

Variations of the preparatory pulse cluster used for T1ρ-weighted imaging are listed in Table 1. Two popular implementations involve either a rotary echo [17; 18] or adiabatic excitation [10] to compensate for B1 inhomogeneities. A degree of ΔB0 insensitivity is achieved using offset-independent adiabatic pulses, particularly those of the HSn family, however, there are restrictions on the minimum ω1max needed for a uniform flip angle across the sample [19]. Variations of the spin lock may also be used for low SAR acquisition by manipulating the T1ρ-weighting in k-space [20] or by spin-locking off-resonance [21; 22]. In addition, the spin locking pulse may be substituted with an adiabatic full passage pulse train to modulate T1ρ-weighted contrast dynamically during the preparation period [23].

Table 1.

Sources of artifacts in T1ρ-weighted imaging and their pulse sequence correction schemes. T1ρ clusters with adiabatic excitation and storage pulses are complementary to the four pulse sequences analyzed in this paper.

Spin Lock Sequence Reference B1 Insensitivity Flip Angle (α = 90°) ΔB0 Insensitivity Off-resonance
Conventional
Adiabatic (ω1max ≫ Δω0) [8] X
Rotary Echo (B1 Insensitive) [17] X X
Off-Resonance [21, 22] X X X
ΔB0 Insensitive [25] X X
B1 & ΔB0 Insensitive [this manuscript] X X X

Ideally, to achieve maximum T1ρ-weighted contrast, the spin locking amplitude (ω1 = γB1) should coincide with the T1ρ dispersion corresponding to the physical process involved, although, in practice, possible spin lock amplitudes are compromised by B0 field gradient artifacts and the high specific absorption rate (SAR) of radiation delivered to tissues. Rotary echoes or adiabatic excitation combined with a high amplitude spin lock (ω1 ≫ Δω0) remove artifacts from gradients in the B1 and B0 field [24], however, increasing ω1 to reduce artifacts and decreasing ω1 to reduce SAR is a vice, limiting the acceptable spin lock amplitude on 1.5 and 3 T clinical scanners to ω1 = 400–600 Hz to obtain a T1ρ-weighted image. Also scanners continue to increase in field strength to 7 T, there may be no range over which ω1 is acceptable. A technique to widen the acceptable spin lock range is necessary, particularly in the low-frequency regime where useful clinical contrast is generated by the scalar coupling between 1H and 17O in studies of brain metabolism or static dipole-dipole or 1H exchange dynamics in human cartilage.

Here, we examine the origin of ΔB0 and B1 spin locking artifacts using the Bloch equations and analyze a new pulse sequence which significantly corrects for these artifacts. This sequence allows spin lock amplitudes in the ω1 = 0–600 Hz range and is demonstrated on an agarose phantom, a water and fat phantom and in vivo in the human brain.

2. Theory

2.1 General Spin Lock

A general pulse sequence for spin locking is shown in Figure 1. In the following sections, we will analyze how the choice of phase or pulse composition in the above sequence can be used to eliminate artifacts from variations in the B0 or B1 fields.

Figure 1.

Figure 1

A generalized pulse sequence for T1ρ-weighted imaging. Each pulse is characterized by a flip angle α and phase θ. Spin locking pulses have both an amplitude ω1 and phase θ. Each of the four fixed amplitude spin lock pulse sequences in Sections 2.2–2.5 are special cases of this generalized sequence.

Using the Bloch equations, we can trace the path of the magnetization vector M(r; t) = [Mx (r; t), My (r; t), Mz (r; t)]T at the spatial location r = [x, y, z] and time t during the conventional spin locking pulse sequence (Fig. 1). Prior to excitation, the magnetization vector in the rotating frame is

M(r;0)=[0,0,M0(r;0)]T [1]

where M0 denotes the equilibrium magnetization and t = 0 is the time prior to excitation and T denotes the transpose operation. An instantaneous RF pulse may be represented compactly in matrix notation by terms like Rθ(α), where R denotes a matrix rotation, θ the pulse phase and α the pulse flip angle. Ideally, magnetization is excited by an instantaneous Rθ1(90°) (phase = θ1) pulse in the conventional spin locking sequence, however, variations in the B1 field may cause imperfect 90° flip angles across the sample, such that the flip angle α is instead

α(r)=γB1(r)τ [2]

where B1 is the actual RF field strength. While an ideal B1 field is uniform across the sample, e.g. B1 (r) = B1, the field may vary significantly in commercial systems with volume head coils.

After excitation, the magnetization is

M(0+)=Rθ1(α)M(0) [3]

where the spatial coordinate r is dropped for simplicity. Magnetization is now spin locked by a long duration RF pulse for a time τ. Ideally, nuclear spins are on-resonance during the spin lock, but B0 field gradients give a distribution of spins about resonance. Off-resonance magnetization is incorporated into the Bloch equations, neglecting relaxation

dM(t)dt=γM(t)x B(t) [4]

where the effective magnetic field in the rotating frame is

B(t)=[B1cosθ2,B1sinθ2,B0ωγ]T [5]

Here, ωis frequency of the rotating frame coordinate system chosen to coincide with the RF pulse carrier frequency ωRF, ω0 = γB0 is the spin Larmor frequency, θ2 is the phase of the spin locking pulse and we assume ω1 ≪ ω0. Of course, on-resonance ωRF = ω0, but we relax this assumption in Eq. [5] because of inhomogeneity in the B0 field. The solutions to Eq. [4] cause the net magnetization to nutate about the axis of the effective field z′, which makes an angle with the z-axis

ϕ=tan1(ω1Δω0) [6]

where Δω0 = ω0 −ωRF. In matrix notation M(t + τ) = Rz′eff τ)M(t), magnetization evolves during the spin locking pulse under the influence of the effective field

ωeff=ω12+Δω02 [7]

by the off-resonance pulse propagator Rz′eff τ). The nutation of the magnetization vector around the effective field is described by transformation to the tilted rotating frame

Rz(ωeffτ)=Rθ2(ϕ)Rz(ωeffτ)Rθ2(ϕ) [8]

Magnetization is now flipped by another instantaneous pulse Rθ32) after which it is spin locked by another cw pulse with phase θ4 and instantaneously flipped one last time. The final magnetization after this series of rotations and spin locks is

M(τ+)=Rθ5(α3)Rθ4(ϕ)Rz(ωeffτ/2)Rθ4(ϕ)Rθ3(α2)Rθ2(ϕ)Rz(ωeffτ/2)Rθ2(ϕ)Rθ1(α1)M(0) [9]

It is possible to further generalize Eq. [9] by making the first and second cw durations and amplitudes inequivalent, however, we will instead simplify to examine special cases of Eq. [9].

2.2 Conventional Spin Lock: 90xy-90−x

The conventional spin lock pulse cluster is sensitive to variations in both the B0 and B1 magnetic fields. While the conventional spin lock is in regular use in spectroscopy, it is replaced by adiabatic and rotary echo methods at spin lock amplitudes ω1 ≫ Δω0 in MR imaging.

For conventional spin locking, Eq.[9] reduces to

M(τ+)=Rx(α)Ry(ϕ)Rz(ωeffτ/2)Ry(ϕ)Rx(α)M(0) [10]

After excitation, the magnetization is

M(0)=[0M0sinαM0cosα] [11]

After the full spin locking duration τ, the magnetization is

M(τ)=[1cos(ϕ)sin(ϕ)sin(ϕ)cos(ϕ)][cos(ωeffτ)sin(ωeffτ)sin(ωeffτ)cos(ωeffτ)1][1cos(ϕ)sin(ϕ)sin(ϕ)cos(ϕ)][0M0sinαM0cosα] [12]

Finally, the T1ρ-weighted magnetization is stored longitudinally with a final α−x pulse, where, including inhomogeneity in the RF field

M(τ+)=[Mx(τ)My(τ)sinαMz(τ)cosα] [13]

A spoiler gradient eliminates the residual transverse magnetization, so the final longitudinal magnetization

Mz(τ+)=M0[cos(ωeffτ)sin(α+ϕ)sin(α+ϕ)+cos(α+ϕ)(cosα+ϕ)] [14]

whereα, ωeff and ϕ are functions of r. If the B1 field strength is much greater than the resonance offset (ω1 ≫ Δω), then the angle ϕ between z′ and z is nearly 90°. This amounts to a 90° phase shift of angular terms in Eq. [14], so

(ω1Δω0)Mz(r;τ+)=M0[cos(ω1τ)cos2(α)sin2(α)] [15]

When α is not 90°, there is a cosinusoidal ω1 τ dependence to the magnetization. These artifacts appear as oscillating regions of signal intensity arranged along the gradients of the B1 field. We show this special case in Figure 2A.

Figure 2.

Figure 2

Conventional (A) and rotary echo (B) composite pulses for T1ρ relaxation measurements and the magnetization path during on-resonance spin locking pulse. Magnetization flipped at an angle α with the y-axis (grey) nutates about the ω1 or y-axis and at τ/2 accumulates a phase ωeff τ/2 (black). While the traditional pulse sequence continues accumulating phase in the same direction (white), the rotary echo returns the magnetization back to its initial position (grey).

Conversely, if ω1 ≪ Δω0, then the angle ϕ between z′ and z is nearly 0° and

(ω1Δω0)Mz(r;τ+)=M0[cos(Δωτ)sin2(α)+cos2(α)] [16]

When α is 90°, T1ρ-weighted images are contaminated with artifacts due to gradients in the B0 field because of the first term in Eq. [16].

2.3 B1 Insensitive Spin Lock: 90x-τ/2y-τ/2−y-90−x

The B1 insensitive spin lock applies Solomon’s rotary echo to imaging pulse sequences [24]. Charagundla, et al. implemented the rotary echo to remove the signal dependence of the variation in the B1 field [17]. Instead of a single, long duration spin lock, the pulse is separated into two equal duration pulses with opposite phase ±y and Eq. [9] is reduced to

M(τ+)=Rx(α)Rx(ϕ)Rz(ωeffτ/2)Rx(ϕ)Rx(ϕ)Rz(ωeffτ/2)Rx(ϕ)Rx(α)M(0) [17]

During the second τ/2 spin locking pulse, the effective field angle

ϕ=tan1(ω1Δω) [18]

is rotated in the opposite sense about the z-axis.

The full expression for the longitudinal magnetization is extensive and is more easily implemented in matrix form as Eq. [17], however, in the limit ω1 ≫ Δω0, ϕ = 90° and Eq. [17] becomes

(ω1Δω0)M(τ+)=Rx(α)Ry(ωeffτ/2)Ry(ωeffτ/2)Rx(α)M(0) [19]

Which, because R−yRy = Inline graphic and Rx(α)Rx(−α) = Inline graphic may be simplified to

(ω1ω0)M(τ+)=M(0) [20]

where the direction of the effective field is entirely along +y for the first τ period and along −y for the second. The signal at M(τ+) no longer depends on cos(ω1τ) or α and so the image is free of banding artifacts inherent to the conventional spin lock. The B1 insensitive pulse cluster is ideal for situations where the B1 field greatly exceeds the resonance offset (see Figure 2B).

On the other hand, if ω1 ≪ Δω0, Eq. [17] may be reduced to

M(τ+)=Rx(α)Rz(Δωτ)Rx(α)M(0) [21]

and the T1ρ-weighted longitudinal component is the same as the conventional spin lock (Eq. [16]).

2.4 ΔB0 Insensitive Spin Lock: 90x-τ/2y-180y-τ/2−y-90−x

Avison, et al. introduced a spin locking pulse cluster which compensates for gradients in the B0 field, but does not eliminate artifacts from imperfect flip angles α ≠ 90° (Figure 2A) [25]. With a 2α (ideally, 2α = 180°) pulse between the spin locking pulse clusters Eq. [9] reduces to

M(τ+)=Rx(α)Rx(ϕ)Rz(ωeffτ/2)Rx(ϕ)Ry(2α)Rx(ϕ)Rz(ωeffτ/2)Rx(ϕ)Rx(α)M(0) [22]

If ω1 ≫ Δω0, ϕ = 90°, and Eq. [22] becomes

(ω1ω0)M(τ+)=Rx(α)Ry(ω1τ/2)Ry(ω1τ/2)Rx(α)M(0) [23]

If the initial flip angle α = 90°, then the magnetization is entirely along +y during both spin lock periods, Ry(2α) has no effect on the magnetization directed along +y and the result is Eq. [20]. Allowing for B1 imperfections, α ≠ 90° and the final longitudinal magnetization

(ω1Δω0)Mz(τ+)=M0[sin2(α)+cos2(α)cos(2α)] [24]

does not depend on the nutation frequency cos(ω1τ). Conversely, if ω1 ≪ Δω0, then ϕ = 0° and we rewrite Eq. [23] as

(ω1Δω0)M(τ+)=Rx(α)Rz(Δω0τ/2)Ry(2α)Rz(Δω0τ/2)Rx(α)M(0) [25]

Specifically, if Ry(2α) = Ry(180°), then Eq. [25] reduces to Eq. [20] and is independent of off-resonance effects Δω0. At intermediate field strengths (ω1 ~ Δω0), however, Eq. [22] requires both Rx(α) = Rx(90°) and Ry(2α) = Ry(180°) to remove terms like cos(ωeffτ) and this requirement is almost never satisfied across the sample.

2.5 ΔB0 and B1 Insensitive Spin Lock: 90x-τ/2y-180y-τ/2−y-90x

Alternating the phase of the last 90° pulse in the spin lock pulse cluster aligns the final magnetization along the −z-axis rather than along the +z-axis (Figure 2B). The alternation of the phase of the final 90° pulse from −x to +x in the cluster compensates for imperfect flip angles α ≠ 90° in an inhomogeneous B1 field. The expression for the full pulse propagator is the same as Eq. [22] except for the final phase shift

M(τ+)=Rx(α)Rx(ϕ)Rz(ωeffτ/2)Rx(ϕ)Ry(2α)Rx(ϕ)Rz(ωeffτ/2)Rx(ϕ)Rx(α)M(0) [26]

If ω1 ≫ Δω0, then ϕ = 90° and Eq. [26] reduces to

(ω1Δω0)M(τ+)=Rx(α)Ry(ω1τ/2)Ry(2α)Ry(ω1τ/2)Rx(α)M(0) [27]

and the final longitudinal magnetization is

(ω1Δω0)Mz(τ+)=M0[sin2(α)+cos2(α)cos(2α)] [28]

and is identical to Eq. [24] but with the first term inverted. The implication is that if α = 90°, the absolute magnetization in unaffected by the pulse phase shift. Instead, if ω1 ≪ Δω0 then magnetization is

(ω1Δω0)M(τ+)=Rx(α)Rz(Δω0τ/2)Ry(2α)Rz(Δω0τ/2)Rx(α)M(0) [29]

The key feature of Eq. [26] and [29] is that the final phase shift −x to +x no longer requires that Rx(α) = Rx(90°), however, to completely reduce Eq. [29] to Eq. [20] we still require Ry(2α) = Ry(180°). In this case Eq. [29] becomes

(ω1Δω0)Ry(2α)=Ry(180°)M(τ+)=M(0) [30]

Despite the inability to achieve a perfect 180° flip in practice, artifacts are less severe than in the ΔB0 insensitive spin lock. In addition, the rectangular 180° may be substituted with a composite 180° RF pulse to further reduce these artifacts [26].

3. Methods

Imaging was performed on a Siemens Trio 3T clinical imaging system equipped with a Bruker birdcage head coil. To maintain consistent B0 and B1 field maps throughout the experiment, automated single slice shim and pulse calibration was performed once and without any further adjustments. Volunteers were recruited to the study and scanned following a pre-approved protocol, described below, by the Institutional Review Board of the University of Pennsylvania. Agarose (3% w/v, 200 mM 23Na) or water/fat phantom (150 mL mineral oil/200 mL doped H2O) imaging was performed using a similar protocol (FOV = 15 cm2).

A B0 field map (Fig. 4, 68) was obtained from four complex gradient echo images with TE = 5, 10, 15 and 20 ms, TR = 700 ms, FOV = 23 cm2, slice thickness = 4 mm and BW = 130 Hz/pixel. Following phase unwrap, the accumulated pixel phase Δθ0 was related to the frequency offset by

Figure 4.

Figure 4

Simulated and actual spin lock artifacts at TSL = 30 ms in 3 different ω1 regimes: (1) Δω≫ ω11 = 0 Hz) (2) Δω~ ω11 = 25 Hz) and (3) Δω≪ ω11 = 400 Hz). B1 and B0 field maps were obtained after automatic pulse calibration and shimming protocols on a Siemens Trio 3T clinical imaging system. The B1 field map was scaled to an ideal π/2 flip angle with a 200 μs rectangular pulse (ω1 = 1250 Hz). The actual ω1 amplitude varies throughout the sample. To amplify image artifacts, the excitation and storage pulses were 80° instead of the nominal 90°.

Figure 6.

Figure 6

T1ρ-weighted images of the brain at 3T. Low spin lock amplitudes (ω1 = 25 Hz) induce Δω0 banding artifacts in both B1 compensation and B0 compensation sequence variants. These artifacts are greatly reduced at higher spin lock amplitudes or with both B1 and B0 compensation.

Figure 8.

Figure 8

The origin of relaxation-dependent artifacts in Fig. 6 and 7 is detailed. (A) The effective field ( ωeff=ω1+Δω0) varies spatially at ω1 = 25 Hz primarily because of variations in ΔB0 at 3T. (B) The angular difference between initial excitation and the orientation of the effective field (Δψ = α – ϕ). (C, D) T1ρ and T2ρ relaxation maps obtained from Eq. [9] using the transient solutions to the Bloch equations.

Δω0ΔTE=Δθ0 [31]

The final B0 field map was obtained by minimizing pixel by pixel the chi-square error statistic to Eq. [31] given the image echo times (TE) and pixel phases (Δθ0) using a linear least squares fitting algorithm in IDL (ITT Visual Information Solutions, Boulder, CO).

A B1 field map (Fig. 4, 68) was obtained using the following protocol. 4 images were obtained with varied rectangular pulse duration τ = 150, 200, 250 and 300 μs using a single rectangular pulse θ°x followed by a spoiler gradient and 2D single slice fast spin echo frequency and phase encoding sequence with the following imaging parameters: TEeff/TR = 13/2500 ms, 128×128 image matrix, FOV = 23 cm2, slice thickness = 4 mm, echo train length = 7, BW = 130 Hz/pixel. B1 field maps were generated using a function in IDL based on the Levenberg-Marquardt algorithm [27] to compute a non-linear least squares fit to the function

S(τ)=S(0)cos(ω1τ) [32]

where S(τ) denotes pixel signal in an image with rectangular pulse duration τ and ω1 is the B1 field amplitude.

Post-processing of both the ΔB0 and B1 field maps involved zeroing non-finite pixel values, 3×3 boxcar smoothing filter and a binary mask of linear fits with R2 < 0.995.

Four variations of a pulse cluster used for T1ρ-weighting (described each in Sections 2.2–2.5) were followed by a gradient to spoil residual transverse magnetization and a 2D single slice fast spin echo frequency and phase encoding sequence with spin lock duration (τ = TSL) = 40 ms, ω1 = 0, 25 or 400 Hz, echo train length = 7, BW = 130 Hz/pixel.

To verify the experimentally acquired T1ρ-weighted images (Fig. 4, 68), simulated images were created from experimental ΔB0 and B1 field maps using the Bloch equations with identical T1ρ-weighted imaging parameters: spin lock duration (τ = TSL) = 40 ms, ω1 = 0, 25 or 400 Hz. Rather than use explicit solutions to the Bloch equations for each pulse sequence (Eq.’s [14, 17, 22 and 26), it was fortuitous to use the generalized matrix notation of Eq. [9], however, Fig. 4 and 6 do not incorporate artifacts from relaxation processes.

The contribution to image artifacts from tissue relaxation and an imperfect refocusing pulse is examined in Fig. 7 and 8. Relaxation was modeled during the spin locking pulses using the transient solutions to the Bloch equations [28]. In the tilted rotating frame, the matrix formulation for both precession about the effective field and relaxation is

Figure 7.

Figure 7

Experimental and simulated T1ρ-weighted images of the brain using the pulse sequence described in Section 2.5 at a spin locking field strength ω1 = 25 Hz. Row 1: Experimental T1ρ-weighted images obtained for 5 spin lock durations 5–60 ms. Row 2: Simulated T1ρ-weighted images from B1 and ΔB0 field maps using transient solutions to the Bloch equations and a parametric fit of the unknown relaxation times T1ρ and T2ρ to the experimental images in Row 1. Row 3: Simulated T1ρ-weighted images from B1 and ΔB0 field maps to the Bloch equations, neglecting relaxation, but using the actual refocusing pulse Ry(2α ≠ 180°). Row 4: Simulated T1ρ-weighted images from B1 and ΔB0 field maps to the Bloch equations, neglecting relaxation and assuming a perfect refocusing pulse Ry(2α = 180°). Without modeling relaxation, there is consequently no relaxation-dependent contrast in Rows 3 and 4. Artifacts from relaxation effects are predominantly localized to the frontal cortex, suggesting relaxation artifacts are primarily dependent on regions of poor B0 field homogeneity, such as near nasal cavities. Artifacts from imperfect refocusing pulses localized to the brain periphery where B1 field inhomogeneity is poor.

Rz(ωeffτ)=[cos(ωeffτ)eτ/T2ρsin(ωeffτ)eτ/T2ρsin(ωeffτ)eτ/T2ρcos(ωeffτ)eτ/T2ρeτ/T1ρ] [33]

In general, both T1ρ and T2ρ are dependent on the B0 and B1 fields and may be written as T1ρ(Δω0 = 0, ω1) and T2ρ(Δω0 = 0, ω1) or, for arbitrary Δω0, T1ρoff(Δω0, ω1) and T2ρ off(Δω0, ω1). In addition, the steady-state solution to the Bloch equations requires an additional term

M(τ+)=Rz(ωeffτ)M(τ)+Mss(1eτ/T1ρ)z [34]

Where z″ denotes the unit vector in the direction of the effective field and Mss is the steady-state magnetization. The rotary echo further complicates an analysis of relaxation, since the magnetization approaches two distinct steady-states during each period.

Relaxation-dependent artifacts were examined by substituting Eq. [33] and [34] into Eq. [8] with T1ρ and T2ρ as unknowns. The steady-state magnetization was fixed (Mss = 0) for the simulation. T1ρ and T2ρ relaxation maps were calculated from Eq. [9] using a Levenberg-Marquardt algorithm [27] in IDL using initial estimates T1ρ = 50 ms and T2ρ = 140 ms.

The specific absorption rate (SAR) delivered during the T1ρ-weighted sequences was determined by estimating the SAR during an arbitrary RF pulse [29]:

SARτ/α=f(3msτ)2(α90°)2SAR3ms/90° [35]

where SAR3ms/90° is the average SAR delivered to the head during a 3 ms rectangular pulse with flip angle α = 90° using a quadrature birdcage coil, f is the pulse shape factor (rectangular pulse = 1.0, sinc pulse = 2.0), τ is the RF pulse duration and α is the RF pulse flip angle. For example, the average SAR3ms/90° in the brain (W/kg) at 3 T for a quadrature birdcage coil is between 0.242 W/kg (1.5 T) and 2.16 W/kg (4.1 T). For a generalized pulse sequence, the average SAR delivered is the sum of the energy absorbed by each RF pulse divided by the total time to acquire the image

SAR=Σn=1NSARτn/αnτnTT [36]

where SAR denotes the average delivered SAR over a total time period TT and τn is the nth RF pulse duration and αn the nth RF pulse flip angle. The FDA limits the delivered SAR to 3 W/kg averaged over the head during a 10 minute period and assuming continuous scanning during this period, 3 W/kg per TR. Eq. [36] estimates the average SAR delivered to the brain (Fig. 6) during the T1ρ-weighted sequences is approximately 0.5 W/kg/TR at ω1 = 400 Hz and 0.08 W/kg/TR at ω1 = 25 Hz. By comparison, the SAR delivered during a T1ρ-weighted sequence with TSL = 100 ms at ω1 = 800 Hz is 3.8 W/kg/TR, surpassing FDA regulations. The actual SAR may differ from this estimate because of coil and head geometry.

4. Results

We confirmed the theory in 3% agarose phantoms to demonstrate the effectiveness of the four different T1ρ pulse clusters to ΔB0 and B1 field gradients. While none of the four pulse clusters completely eliminated artifacts at all spin locking field strengths (ω1 = 0 Hz, 25 Hz and 400 Hz), we found the ΔB0 and B1 insensitive pulse cluster was the most robust. The remaining artifacts are attributed to an imperfect 180° pulses, but may be removed using either composite 180° or an adiabatic refocusing pulse.

T1ρ-weighted images were simulated from B1 and ΔB0 maps to verify that spin locking artifacts could be modeled using the Bloch equations. The simulated images were obtained using Eq.s [14], [17], [22] and [26] and are displayed alongside the actual T1ρ-weighted images in Figure 4. Actual images were collected using four spin locking composite pulse clusters and three different spin locking field strengths ω1: (1) Δω0 ≫ ω11 = 0 Hz) (2) Δω0 ~ ω11 = 25 Hz) and (3) Δω0 ≪ ω11 = 400 Hz).

At ω1 = 400 Hz and Δω0 ≪ ω1, conventional spin locking artifacts described by Eq. [15] are arranged along the gradients of the ω1 field. Banding artifacts form every ω1τ = 2π and while the B1 insensitive spin lock removes these artifacts, they reemerge in the ΔB0 insensitive sequence. Inverting the phase of the final 90° pulse +x removes these artifacts.

At ω1 = 0 Hz and Δω ≫ ω1, B0 field gradients create banding artifacts described by Eq. [16]. These artifacts are best known among fatty tissues or nasal cavities where the corresponding chemical shift or tissue susceptibility difference gives a large resonance offset. A sample in a Gaussian-like B0 field gradient forms bands for every Δω0τ = 2π and increasing either the resonance offset Δω0 or the spin lock duration τ increases the total number of bands in the image. The banding artifacts are identical in both the conventional and B1 insensitive spin locking sequences since, in the limit ω1 → 0, the two sequences are identical. Inserting a 180° pulse theoretically removes the dependence on the resonance offset (from Eq. 25), but banding artifacts remain from a combination of imperfect 180° (ΔB0 and B1 insensitive) and imperfect 90° pulses (ΔB0 insensitive).

To illustrate full ΔB0 insensitivity during spin locking, a series of images were collected in a fat and water phantom at varying spin lock durations τ (Fig. 5). At 3T and ω1 = 500 Hz, the effective field makes an angle ϕ ≈ 51° to the z-axis and produces severe banding artifacts in both conventional and B1 insensitive T1ρ-weighted imaging. The artifact is removed in ΔB0 or ΔB0 and B1 insensitive pulse clusters provided the hard pulse flip angles (90° or 180°) are conserved.

Figure 5.

Figure 5

T1ρ-weighted images of a water (bottom) and fat (top) phantom. Off-resonance fat protons produce artifacts in conventional and B1 insensitive T1ρ-weighted images, but are absent in ΔB0 insensitive methods. Contrast between water and fat varies in each of the four pulse sequences and is attributed to the off-resonance relaxation of fat spins.

T1ρ-weighted images were obtained of the human brain in vivo and show significant Δω0 banding artifacts at low spin lock amplitudes (ω1 ~ Δω0) in Figure 6. At higher spin lock amplitudes (ω1 ≫ Δω0), the magnetization for each of the pulse sequences is described by Eq.’s [20], [24] and [28], respectively, and the magnetization is independent of nutation about the effective field. By compensating for both B1 and B0 imperfections, artifacts are significantly reduced at ω1 = 25 Hz as well.

The high artifact suppression in the ΔB0 and B1 insensitive pulse sequence shows remarkable robustness to field inhomogeneities despite possible artifacts from relaxation processes. The pulse sequence used for ΔB0 and B1 correction, 90x-τ/2y-180y-τ/2−y-90x, is not a unique solution and an alternative such as 90x-τ/2y-180−x-τ/2y-90−x is equally robust (data not shown). Also, there exist 2 additional solutions for each excitation phase 90y, 90−x or 90−y

Additional artifacts emerge because of both tissue relaxation and an imperfect refocusing pulse; these additional artifacts are considered in Figure 7. Row 1 shows experimental images using the ΔB0 and B1 insensitive pulse sequence for 5 different spin locking durations (TSL = 5, 10, 20, 40, and 60 ms) at ω1 = 25 Hz. Except for the frontal cortex, a region of significant B0 field inhomogeneity, the experimental images are free of low ω1 band artifacts. The location of the artifacts can be reproduced in simulated images that model T1ρ and T2ρ relaxation (Row 2), but are not observed in simulations that do not model relaxation (Row 4). To some extent, an imperfect refocusing pulse produces artifacts around the periphery in regions of significant B1 field inhomogeneity (Row 3) and also in the frontal cortex.

The relationship between tissue relaxation, field inhomogeneity and the residual artifacts may be considered as follows: As magnetization nutates around the effective field, the component parallel to the effective field M|| will decay to the steady state magnetization at a rate 1/T1ρ and the component perpendicular to the effective field MT will decay at a rate 1/T2ρ. As the ratio MT/M|| changes during the spin lock, so does the nutation angle between the effective field and magnetization. If the nutation angle changes, the final spin locked magnetization (Eq. [30]) will not be stored parallel to the −z-axis and field dependent artifacts may emerge. Therefore, field inhomogeneity has two primary effects, to nutate the magnetization different angles because of variations in ωeff and also to cause relaxation-dependent artifacts. As Figures 6 and 7 demonstrate, small field inhomogeneities have a substantial effect on nutation angle banding artifacts, but don’t affect relaxation dependent artifacts nearly as much.

There is a threshold for field inhomogeneity beyond which low ω1 imaging becomes unacceptable. Relaxation-dependent artifacts are worse for larger spin lock durations and higher B0 field inhomogeneities and so the threshold is quantified in terms of τ and ΔB0. As stated previously, T2ρ relaxation will change the angle between the effective field and the initially excited magnetization, so artifacts are significant for when times τ ~ T2ρ and the difference angle between the initial excitation and the effective field

Δψ=αϕ [37]

is appreciable. There are three regimes for relaxation-dependent artifacts. For large Δψ, there is a significant amount of T2ρ relaxation at long τ durations. T2ρ relaxation can be easily quantified because a large component of the magnetization is perpendicular to the effective field. Conversely, for small Δψ, T1ρ relaxation can be easily quantified, because a large component of the magnetization is parallel to the effective field. Only gradients in Δψ will cause artifacts, because of significant differences in the relaxation times T1ρ and T2ρ. In the third regime, when Δψ ~ 45°, T1ρ and T2ρ can be quantified, but a convergent solution is less likely. Figure 8 demonstrates how Δψ affects quantification of T1ρ and T2ρ. While both T1ρ and T2ρ depend on ωeff (Fig. 8A), there is a clear spatial dependence of the relaxation times on Δψ (Fig 8B). For example, where Δψ is large, T2ρ takes on usual brain tissue values (T2ρ = 140 ms, Fig. 8D), but when Δψ is small, the magnetization lies parallel to the effective field and T2ρ quantification is not possible. Conversely, T1ρ is quantified when Δψ is small (Fig. 8C), but more difficult when Δψ is large, such as in the frontal cortex. T1ρ is nearly 5 ms lower in the midbrain white matter regions (Δψ = 45°) than in the white matter regions of the periphery (Δψ < 45°),

5. Discussion

Section 2.5 and Figure 4 suggest that the ΔB0 and B1 insensitive pulse sequence is the most robust against variations in B0 and B1 field gradients. Furthermore, the insensitivity of the sequence may be further improved by decomposing the central Ry(180°) into a composite 180° pulse or substitution with an adiabatic refocusing pulse.

While the foregoing theory entirely corrects for nutations about the effective magnetic field, magnetization dynamics are complicated by relaxation processes. Tissue relaxation during a cw pulse is a complex process that depends on ω1, Δω0 and the component of magnetization parallel (T1ρ) or perpendicular (T2ρ) to the applied RF field. In general, the system is driven by ω1 and damped by both T1 and T2 relaxation until it approaches a steady-state magnetization parallel to the effective field [28]. True monoexponential T1ρ relaxation is observed by flipping the magnetization parallel to ω1 on-resonance and is distorted by Δω0 and T2ρ. When ω1 ≪ Δω0, α ≠ 90° or both, these Δ effects compound to cause the magnetization to deviate from monoexponential T1ρ relaxation and cause additional image artifacts.

T1ρ relaxation measurements are a useful diagnostic tool in clinical imaging and any off-resonance spins interfere with T1ρ quantification. In particular, a T1ρ map displays the spatial variation in T1ρ across the sample. Often the T1ρ map contains artifacts, even if a single T1ρ-weighted image is artifact free, because the decay of spin magnetization is no longer monoexponential. The T1ρ map is improved by the ΔB0 and B1 insensitive pulse sequence and the possibility is now available for tissue studies of very low frequency T1ρ dispersion. Several T1ρ-weighted images collected at low frequency ω1 may be more sensitive to residual dipolar interactions (ωD ~ 200 Hz) in tissues and provide a useful form of dipolar contrast among tissues composed of oriented collagen, muscle fibers or myelinated axons. As in Figure 8, combined quantification of T1ρ and T2ρ is also possible, but must be interpreted in the context of a spatially varying effective field and usually both relaxation times will not be quantifiable simultaneously in a pixel.

Very low amplitude spin locking offers several interesting imaging capabilities. As Santyr, et al. initially suggested, it is possible to estimate the contribution of the local static field Bloc by measuring T1ρ dispersion in tissues [30]. In practice, at spin lock amplitudes ω1 ~ γBloc the contribution of the local static fields confounds a measurement of T1ρ because they induce oscillations of the magnetization about an effective field in the case of either off-resonance spins, chemical shift and secular J-coupling. And while this manuscript makes use of the Bloch equations, it can be shown using product operator theory that any static interaction that commutes with the spin operator Iz will be refocused. In this way, a low amplitude spin lock dispersion measurement can be used to probe dynamical interactions or spin diffusion independent of spin nutation about the effective field.

Low frequency T1ρ-weighted imaging has two potential applications in medicine where ω1 ~ γBloc. We suspect that it will generate useful contrast in indirectly detected H217O for studies of brain metabolism because of the low-frequency scalar coupling between 1H and 17O [13]. A bolus of enriched 17O2 may be given to a human subject and converted to metabolically produced H2O17. The magnetically active 17O nucleus interacts with water protons through scalar coupling and differences between low frequency (ω1 ~ JH–O) and high frequency (ω1 ≫ JH–O) T1ρ-weighted images are sensitive to 17O. This phenomenon has been observed in rat models, but is limited by B0 field inhomogeneities in the low frequency regime (ω1 ~ JH–O) [16]. In addition, static dipole-dipole interactions are known to exist in oriented tissues such as collagen layers of human cartilage[5; 31]. In the anisotropic environment of collagen tissues, 1H-1H static dipole-dipole interactions are partially averaged, such that ωD ~ 200 Hz. Low frequency T1ρ-weighted imaging may be sensitive to changes in this interaction, especially during diseases of the cartilage like osteoarthritis.

6. Conclusion

The origin of inhomogeneous Δω0 and ω1 field artifacts in T1ρ-weighted images are oscillations about the effective field at low ω1 and imperfect flip angles α ≠ 90° at all field strengths. We introduced a spin locking pulse sequence that almost entirely compensates for both types of artifacts at all field strengths. The sequence combines the familiar spin echo with a final pulse phase inversion to flip magnetization along the −z-axis. Consequently, several experimental images were shown on agarose gel and water fat phantoms confirming both Δω0 and ω1 insensitivity. These results were confirmed with the derived theory and simulation. The foregoing theory and sequence should be useful for spin locking in the low frequency (ω1 = 0–600 Hz) regime and generate useful contrast for T1ρ-weighted imaging applications.

Figure 3.

Figure 3

ΔB0 insensitive (A) and B1 and ΔB0 insensitive (B) composite pulses for T1ρ weighted imaging and the magnetization path during each sequence. In (A) magnetization is flipped along the y-axis (grey), where it nutates about the effective field (z′-axis) and at time τ/2 (black) is flipped 180° about the y-axis where it nutates around the effective field (z″-axis) back along the y-axis (grey). In (B) the magnetization follows a similar path, but with two differences: (1) the excitation flip angle does not need to be 90° and (2) B1 insensitivity is maintained by flipping the magnetization along the −z-axis. While (B) is more robust than (A), an imperfect 180° flip can still produce artifacts.

Acknowledgments

This study was performed at the Metabolic Magnetic Resonance Research and Computing Center, an NIH-supported resource center (NIH RR02305) supported by grants R01AR045404 and R01AR051041.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Redfield AG. Nuclear Magnetic Resonance Saturation and Rotary Saturation in Solids. Phys Rev. 1955;98:1787–1809. [Google Scholar]
  • 2.Duvvuri U, Goldberg AD, Kranz JK, Hoang L, Reddy R, Wehrli FW, Wand AJ, Englander SW, Leigh JS. Water magnetic relaxation dispersion in biological systems: The contribution of proton exchange and implications for the noninvasive detection of cartilage degradation. Proc Natl Acad Sci U S A. 2001;98:12479–12484. doi: 10.1073/pnas.221471898. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Makela HI, Grohn OHJ, Kettunen MI, Kauppinen RA. Proton exchange as a relaxation mechanism for T-1 in the rotating frame in native and immobilized protein solutions. Biochem Biophys Res Commun. 2001;289:813–818. doi: 10.1006/bbrc.2001.6058. [DOI] [PubMed] [Google Scholar]
  • 4.Knispel RR, Thompson RT, Pintar MM. Dispersion of Proton Spin-Lattice Relaxation in Tissues. J Magn Reson. 1974;14:44–51. [Google Scholar]
  • 5.Akella SV, Regatte RR, Wheaton AJ, Borthakur A, Reddy R. Reduction of residual dipolar interaction in cartilage by spin-lock technique. Magn Reson Med. 2004;52:1103–9. doi: 10.1002/mrm.20241. [DOI] [PubMed] [Google Scholar]
  • 6.Grohn OHJ, Lukkarinen JA, Silvennoinen MJ, Pitkanen A, van Zijl PCM, Kauppinen RA. Quantitative magnetic resonance imaging assessment of cerebral ischemia in rat using on-resonance T-1 in the rotating frame. Magn Reson Med. 1999;42:268–276. doi: 10.1002/(sici)1522-2594(199908)42:2<268::aid-mrm8>3.0.co;2-a. [DOI] [PubMed] [Google Scholar]
  • 7.Grohn OH, Makela HI, Lukkarinen JA, DelaBarre L, Lin J, Garwood M, Kauppinen RA. On- and off-resonance T1rho MRI in acute cerebral ischemia of the rat. Magn Reson Med. 2003;49:172–6. doi: 10.1002/mrm.10356. [DOI] [PubMed] [Google Scholar]
  • 8.Grohn OHJ, Kettunen MI, Makela HI, Penttonen M, Pitkanen A, Lukkarinen JA, Kauppinen RA. Early detection of irreversible cerebral ischemia in the rat using dispersion of the magnetic resonance imaging relaxation time, T1rho. J Cereb Blood Flow Metab. 2000;20:1457–66. doi: 10.1097/00004647-200010000-00007. [DOI] [PubMed] [Google Scholar]
  • 9.Aronen HJ, Ramadan UA, Peltonen TK, Markkola AT, Tanttu JI, Jaaskelainen J, Hakkinen AM, Sepponen R. 3D spin-lock imaging of human gliomas. Magn Reson Imaging. 1999;17:1001–10. doi: 10.1016/s0730-725x(99)00041-7. [DOI] [PubMed] [Google Scholar]
  • 10.Santyr GE, Henkelman RM, Bronskill MJ. Spin Locking for Magnetic-Resonance Imaging with Application to Human-Breast. Magn Reson Med. 1989;12:25–37. doi: 10.1002/mrm.1910120104. [DOI] [PubMed] [Google Scholar]
  • 11.Regatte RR, Akella SVS, Wheaton AJ, Lech G, Borthakur A, Kneeland JB, Reddy R. 3D-T-1 rho-relaxation mapping of articular cartilage: In vivo assessment of early degenerative changes in symptomatic osteoarthritic subjects. Acad Radiol. 2004;11:741–749. doi: 10.1016/j.acra.2004.03.051. [DOI] [PubMed] [Google Scholar]
  • 12.Auerbach JD, Johannessen W, Borthakur A, Wheaton AJ, Dolinskas CA, Balderston RA, Reddy R, Elliott DM. In vivo quantification of human lumbar disc degeneration using T-1 rho-weighted magnetic resonance imaging. Eur Spine J. 2006;15:S338–S344. doi: 10.1007/s00586-006-0083-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Reddy R, Stolpen AH, Leigh JS. Detection of O-17 by Proton T-1P Dispersion Imaging. J Magn Reson Series B. 1995;108:276–279. doi: 10.1006/jmrb.1995.1133. [DOI] [PubMed] [Google Scholar]
  • 14.Tailor DR, Poptani H, Glickson JD, Leigh JS, Reddy R. High-resolution assessment of blood flow in murine RIF-1 tumors by monitoring uptake of H217O with proton T1rho-weighted imaging. Magn Reson Med. 2003;49:1–6. doi: 10.1002/mrm.10375. [DOI] [PubMed] [Google Scholar]
  • 15.Tailor DR, Roy A, Regatte RR, Akella SVS, Leigh JS, Reddy R. Measuring regional blood flow in the rat brain using indirect 17O magnetic resonance imaging. Proc. Internatl. Soc. Magn. Reson. Med., ISMRM-ESMRB Joint Annual Meeting; Glasgow. 2001. p. 356. [Google Scholar]
  • 16.Tailor DR, Roy A, Regatte RR, Charagundla SR, McLaughlin AC, Leigh JS, Reddy R. Indirect 17O-magnetic resonance imaging of cerebral blood flow in the rat. Magn Reson Med. 2003;49:479–87. doi: 10.1002/mrm.10403. [DOI] [PubMed] [Google Scholar]
  • 17.Charagundla SR, Borthakur A, Leigh JS, Reddy R. Artifacts in T-1 rho-weighted imaging: correction with a self-compensating spin-locking pulse. J Magn Reson. 2003;162:113–121. doi: 10.1016/s1090-7807(02)00197-0. [DOI] [PubMed] [Google Scholar]
  • 18.Borthakur A, Wheaton A, Charagundla SR, Shapiro EM, Regatte RR, Akella SVS, Kneeland JB, Reddy R. Three-dimensional T1 rho-weighted MRI at 1.5 Tesla. J Magn Reson Im. 2003;17:730–736. doi: 10.1002/jmri.10296. [DOI] [PubMed] [Google Scholar]
  • 19.Garwood M, DelaBarre L. The return of the frequency sweep: Designing adiabatic pulses for contemporary NMR. J Magn Reson. 2001;153:155–177. doi: 10.1006/jmre.2001.2340. [DOI] [PubMed] [Google Scholar]
  • 20.Wheaton AJ, Borthakur A, Corbo M, Charagundla SR, Reddy R. Method for reduced SAR T1rho-weighted MRI. Magn Reson Med. 2004;51:1096–102. doi: 10.1002/mrm.20141. [DOI] [PubMed] [Google Scholar]
  • 21.Santyr GE, Fairbanks EJ, Kelcz F, Sorenson JA. Off-resonance spin locking for MR imaging. Magn Reson Med. 1994;32:43–51. doi: 10.1002/mrm.1910320107. [DOI] [PubMed] [Google Scholar]
  • 22.Fairbanks EJ, Santyr GE, Sorenson JA. One-Shot Measurement of Spin-Lattice Relaxation-Times in the Off-Resonance Rotating-Frame Using Mr-Imaging, with Application to Breast. J Magn Reson Series B. 1995;106:279–283. doi: 10.1006/jmrb.1995.1044. [DOI] [PubMed] [Google Scholar]
  • 23.Grohn HI, Michaeli S, Garwood M, Kauppinen RA, Grohn OHJ. Quantitative T-1p and adiabatic Carr-Purcell T-2 magnetic resonance imaging of human occipital lobe at 4 T. Magn Reson Med. 2005;54:14–19. doi: 10.1002/mrm.20536. [DOI] [PubMed] [Google Scholar]
  • 24.Solomon I. Rotary Spin Echoes. Phys Rev Lett. 1959;2:301–302. [Google Scholar]
  • 25.Avison M, Gochberg D, Gatenby C, Gore J, Zeng H, Zhao Y. A Composite Spin-Lock Pulse for deltaB0 + B1 Insensitive T1rho Measurement. Proc. Soc. Magn. Reson. Med., ISMRM Annual Meeting; Seattle, Washington. 2006. [Google Scholar]
  • 26.Levitt MH, Freeman R. Nmr Population-Inversion Using a Composite Pulse. J Magn Reson. 1979;33:473–476. doi: 10.1016/j.jmr.2011.08.016. [DOI] [PubMed] [Google Scholar]
  • 27.Marquardt DW. An Algorithm for Least-Squares Estimation of Nonlinear Parameters. J Soc Indus Appl Math. 1963;11:431–441. [Google Scholar]
  • 28.Torrey HC. Transient Nutations in Nuclear Magnetic Resonance. Phys Rev. 1949;76:1059–1068. [Google Scholar]
  • 29.Collins CM, Li SZ, Smith MB. SAR and B-1 field distributions in a heterogeneous human head model within a birdcage coil. Magn Reson Med. 1998;40:847–856. doi: 10.1002/mrm.1910400610. [DOI] [PubMed] [Google Scholar]
  • 30.Santyr GE, Henkelman RM, Bronskill MJ. Variation in Measured Transverse Relaxation in Tissue Resulting from Spin Locking with the Cpmg Sequence. J Magn Reson. 1988;79:28–44. [Google Scholar]
  • 31.Xia Y. Relaxation anisotropy in cartilage by NMR microscopy (μMRI) at 14-μm resolution. Magn Reson Med. 1998;39:941–9. doi: 10.1002/mrm.1910390612. [DOI] [PubMed] [Google Scholar]

RESOURCES