Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1995 Sep;33(9):2265–2270. doi: 10.1128/jcm.33.9.2265-2270.1995

Controlled evaluation of BacT/alert standard anaerobic and FAN anaerobic blood culture bottles for the detection of bacteremia and fungemia.

M L Wilson 1, M P Weinstein 1, S Mirrett 1, L G Reimer 1, R J Feldman 1, C R Chuard 1, L B Reller 1
PMCID: PMC228392  PMID: 7494013

Abstract

FAN medium was formulated to improve microbial recovery, particularly for fastidious microorganisms and for microorganisms causing sepsis in patients receiving antimicrobial therapy. In a controlled clinical evaluation performed at four university-affiliated hospitals, FAN anaerobic bottles were compared with standard anaerobic bottles for yield, speed of detection of microbial growth, and detection of septic episodes. A total of 10,431 blood culture sets were received; both anaerobic bottles of 7,694 blood culture sets were adequately filled with blood. Altogether, 925 isolates were recovered: 557 that were the cause of sepsis, 99 that were indeterminate as the cause of sepsis, and 269 contaminants. More Staphylococcus aureus (P < 0.001), coagulase-negative staphylococci (P < 0.001), Escherichia coli (P < 0.02), and all microorganisms combined (P < 0.005) were recovered from FAN bottles; more nonfermentative gram-negative bacilli (P < 0.05), Torulopsis glabrata (P < 0.001), and other yeasts (P < 0.01) were recovered from standard bottles. Growth of S. aureus (P < 0.001), coagulase-negative staphylococci (P < 0.001), Enterococcus faecalis (P < 0.025), streptococci other than Streptococcus pneumoniae (P < 0.01), and all microorganisms combined (P < 0.001) was detected earlier in standard bottles; growth of more isolates of E. coli (P < 0.05) and anaerobic bacteria (P < 0.01) was detected earlier in FAN bottles. The mean times to detection were 14.2 and 16.1 h for standard and FAN bottles, respectively.(ABSTRACT TRUNCATED AT 250 WORDS)

Full Text

The Full Text of this article is available as a PDF (345.5 KB).

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Appelbaum P. C., Beckwith D. G., Dipersio J. R., Dyke J. W., Salventi J. F., Stone L. L. Enhanced detection of bacteremia with a new BACTEC resin blood culture medium. J Clin Microbiol. 1983 Jan;17(1):48–51. doi: 10.1128/jcm.17.1.48-51.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Appleman M. D., Swinney R. S., Heseltine P. N. Evaluation of the Antibiotic Removal Device. J Clin Microbiol. 1982 Feb;15(2):278–281. doi: 10.1128/jcm.15.2.278-281.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Bille J., Stockman L., Roberts G. D. Detection of yeasts and filamentous fungi in blood cultures during a 10-year period (1972 to 1981). J Clin Microbiol. 1982 Nov;16(5):968–970. doi: 10.1128/jcm.16.5.968-970.1982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Doern G. V., Gantz N. M. Detection of bacteremia in patients receiving antimicrobial therapy: an evaluation of the antimicrobial removal device and 16B medium. J Clin Microbiol. 1983 Jul;18(1):43–48. doi: 10.1128/jcm.18.1.43-48.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Gantz N. M., Medeiros A. A., Swain J. L., O'Brien T. F. Vacuum blood-culture bottles inhibiting growth of Candida and fostering growth of Bacteroides. Lancet. 1974 Nov 16;2(7890):1174–1176. doi: 10.1016/s0140-6736(74)90813-7. [DOI] [PubMed] [Google Scholar]
  6. Hall M. M., Mueske C. A., Ilstrup D. M., Washington J. A., 2nd Evaluation of a biphasic medium for blood cultures. J Clin Microbiol. 1979 Nov;10(5):673–676. doi: 10.1128/jcm.10.5.673-676.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Hardy D. J., Hulbert B. B., Migneault P. C. Time to detection of positive BacT/Alert blood cultures and lack of need for routine subculture of 5- to 7-day negative cultures. J Clin Microbiol. 1992 Oct;30(10):2743–2745. doi: 10.1128/jcm.30.10.2743-2745.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Harkness J. L., Hall M., Ilstrup D. M., Washington J. A., 2nd Effects of atmosphere of incubation and of routine subcultures on detection of bacteremia in vacuum blood culture bottles. J Clin Microbiol. 1975 Oct;2(4):296–299. doi: 10.1128/jcm.2.4.296-299.1975. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Hopfer R. L., Moore D. G., Fainstein V., Watkins S., Wenglar M. Routine use of BACTEC 16B bottles to remove antibacterial and antitumor agents from blood cultures of cancer patients. J Clin Microbiol. 1983 Oct;18(4):759–764. doi: 10.1128/jcm.18.4.759-764.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Jessamine P. G., Hoban D. J., Forward K. R. Positive Bactec resin cultures do not influence antimicrobial selection. Diagn Microbiol Infect Dis. 1990 Jul-Aug;13(4):281–284. doi: 10.1016/0732-8893(90)90017-p. [DOI] [PubMed] [Google Scholar]
  11. Knepper J. G., Anthony B. F. Diminished growth of Pseudomonas aeruginosa in unvented blood-culture bottles. Lancet. 1973 Aug 11;2(7824):285–287. doi: 10.1016/s0140-6736(73)90792-7. [DOI] [PubMed] [Google Scholar]
  12. Moody J. A., Fasching C. E., Shanholtzer C. J., Gerding D. N., Peterson L. R. Evaluation of new blood culture processing systems. J Clin Microbiol. 1984 Sep;20(3):351–356. doi: 10.1128/jcm.20.3.351-356.1984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Peterson L. R., Shanholtzer C. J., Mohn M. L., Gerding D. N. Improved recovery of microorganisms from patients receiving antibiotics with the antimicrobial removal device. Am J Clin Pathol. 1983 Nov;80(5):692–696. doi: 10.1093/ajcp/80.5.692. [DOI] [PubMed] [Google Scholar]
  14. Ringertz S., Eriksson B., Malmborg A. S., Nyström S., Wretlind B. Use of the antimicrobial removal device prior to blood culture in patients on antibiotic therapy. Eur J Clin Microbiol. 1985 Dec;4(6):544–547. doi: 10.1007/BF02013391. [DOI] [PubMed] [Google Scholar]
  15. Strand C. L. Evaluation of the antimicrobial removal device when used with the BACTEC blood culture system. Am J Clin Pathol. 1982 Dec;78(6):853–856. doi: 10.1093/ajcp/78.6.853. [DOI] [PubMed] [Google Scholar]
  16. Thorpe T. C., Wilson M. L., Turner J. E., DiGuiseppi J. L., Willert M., Mirrett S., Reller L. B. BacT/Alert: an automated colorimetric microbial detection system. J Clin Microbiol. 1990 Jul;28(7):1608–1612. doi: 10.1128/jcm.28.7.1608-1612.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Weinstein M. P., Mirrett S., Reimer L. G., Wilson M. L., Smith-Elekes S., Chuard C. R., Joho K. L., Reller L. B. Controlled evaluation of BacT/Alert standard aerobic and FAN aerobic blood culture bottles for detection of bacteremia and fungemia. J Clin Microbiol. 1995 Apr;33(4):978–981. doi: 10.1128/jcm.33.4.978-981.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Weinstein M. P., Mirrett S., Wilson M. L., Harrell L. J., Stratton C. W., Reller L. B. Controlled evaluation of BACTEC Plus 26 and Roche Septi-Chek aerobic blood culture bottles. J Clin Microbiol. 1991 May;29(5):879–882. doi: 10.1128/jcm.29.5.879-882.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Weinstein M. P., Mirrett S., Wilson M. L., Reimer L. G., Reller L. B. Controlled evaluation of 5 versus 10 milliliters of blood cultured in aerobic BacT/Alert blood culture bottles. J Clin Microbiol. 1994 Sep;32(9):2103–2106. doi: 10.1128/jcm.32.9.2103-2106.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Weinstein M. P., Reller L. B., Murphy J. R., Lichtenstein K. A. The clinical significance of positive blood cultures: a comprehensive analysis of 500 episodes of bacteremia and fungemia in adults. I. Laboratory and epidemiologic observations. Rev Infect Dis. 1983 Jan-Feb;5(1):35–53. doi: 10.1093/clinids/5.1.35. [DOI] [PubMed] [Google Scholar]
  21. Wilson M. L., Harrell L. J., Mirrett S., Weinstein M. P., Stratton C. W., Reller L. B. Controlled evaluation of BACTEC PLUS 27 and Roche Septi-Chek anaerobic blood culture bottles. J Clin Microbiol. 1992 Jan;30(1):63–66. doi: 10.1128/jcm.30.1.63-66.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Wilson M. L., Mirrett S., Reller L. B., Weinstein M. P., Reimer L. G. Recovery of clinically important microorganisms from the BacT/Alert blood culture system does not require testing for seven days. Diagn Microbiol Infect Dis. 1993 Jan;16(1):31–34. doi: 10.1016/0732-8893(93)90127-s. [DOI] [PubMed] [Google Scholar]
  23. Wilson M. L., Weinstein M. P., Reimer L. G., Mirrett S., Reller L. B. Controlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems. J Clin Microbiol. 1992 Feb;30(2):323–329. doi: 10.1128/jcm.30.2.323-329.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Wright A. J., Thompson R. L., McLimans C. A., Wilson W. R., Washington J. A., 2nd The antimicrobial removal device. A microbiological and clinical evaluation. Am J Clin Pathol. 1982 Aug;78(2):173–177. doi: 10.1093/ajcp/78.2.173. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)