Abstract
We compared the BacT/Alert system using the aerobic FAN bottle with the ESP system using the 80A aerobic bottle for the detection of pediatric bloodstream pathogens at a children's hospital. From 6,636 blood culture sets complying with the inclusion criteria, 308 pathogens were detected, including 177 that were detected by both systems, 69 that were detected by BacT/Alert FAN only, and 62 that were detected by ESP 80A only (P = 0.6; not significant). BacT/Alert FAN detected more isolates of Staphylococcus aureus (47 versus 34; P = 0.02), while ESP 80A detected more episodes of streptococcal and enterococcal infection. BacT/Alert FAN detected more pathogens from patients receiving antibiotic therapy (107 versus 93; P = 0.04). Of 248 separate episodes of bacteremia or fungemia, 146 were detected by both systems, 56 were detected by ESP 80A only, and 46 were detected by BacT/Alert FAN only (P = 0.37; not significant). The median times to detection were 13.6 h for ESP 80A and 15.7 h for BacT/Alert FAN (P < 0.001). Both systems were considered easy to operate and were free from significant mechanical difficulties. False-positive or false-negative signals were rare or nonexistent with both systems. We conclude that both systems rapidly detect a broad range of pediatric bloodstream pathogens. BacT/Alert FAN provides better detection of Staphylococcus aureus, especially from patients receiving antibiotics. ESP 80A provides better detection of streptococci and enterococci.
Full Text
The Full Text of this article is available as a PDF (142.0 KB).
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Cockerill F. R., 3rd, Torgerson C. A., Reed G. S., Vetter E. A., Weaver A. L., Dale J. C., Roberts G. D., Henry N. K., Ilstrup D. M., Rosenblatt J. E. Clinical comparison of difco ESP, Wampole isolator, and Becton Dickinson Septi-Chek aerobic blood culturing systems. J Clin Microbiol. 1996 Jan;34(1):20–24. doi: 10.1128/jcm.34.1.20-24.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dunne W. M., Jr, Tillman J., Havens P. L. Assessing the need for anaerobic medium for the recovery of clinically significant blood culture isolates in children. Pediatr Infect Dis J. 1994 Mar;13(3):203–206. doi: 10.1097/00006454-199403000-00007. [DOI] [PubMed] [Google Scholar]
- Hellinger W. C., Cawley J. J., Alvarez S., Hogan S. F., Harmsen W. S., Ilstrup D. M., Cockerill F. R., 3rd Clinical comparison of the isolator and BacT/Alert aerobic blood culture systems. J Clin Microbiol. 1995 Jul;33(7):1787–1790. doi: 10.1128/jcm.33.7.1787-1790.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Krisher K. K., Whyburn D. R., Koepnick F. E. Comparison of the BacT/Alert pediatric blood culture system, Pedi-BacT, with conventional culture using the 20-milliliter Becton-Dickinson supplemented peptone broth tube. J Clin Microbiol. 1993 Apr;31(4):793–797. doi: 10.1128/jcm.31.4.793-797.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Morello J. A., Leitch C., Nitz S., Dyke J. W., Andruszewski M., Maier G., Landau W., Beard M. A. Detection of bacteremia by Difco ESP blood culture system. J Clin Microbiol. 1994 Mar;32(3):811–818. doi: 10.1128/jcm.32.3.811-818.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nolte F. S., Williams J. M., Jerris R. C., Morello J. A., Leitch C. D., Matushek S., Schwabe L. D., Dorigan F., Kocka F. E. Multicenter clinical evaluation of a continuous monitoring blood culture system using fluorescent-sensor technology (BACTEC 9240). J Clin Microbiol. 1993 Mar;31(3):552–557. doi: 10.1128/jcm.31.3.552-557.1993. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pohlman J. K., Kirkley B. A., Easley K. A., Basille B. A., Washington J. A. Controlled clinical evaluation of BACTEC Plus Aerobic/F and BacT/Alert Aerobic FAN bottles for detection of bloodstream infections. J Clin Microbiol. 1995 Nov;33(11):2856–2858. doi: 10.1128/jcm.33.11.2856-2858.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rohner P., Pepey B., Auckenthaler R. Comparative evaluation of BACTEC aerobic Plus/F and Septi-Chek Release blood culture media. J Clin Microbiol. 1996 Jan;34(1):126–129. doi: 10.1128/jcm.34.1.126-129.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schwabe L. D., Thomson R. B., Jr, Flint K. K., Koontz F. P. Evaluation of BACTEC 9240 blood culture system by using high-volume aerobic resin media. J Clin Microbiol. 1995 Sep;33(9):2451–2453. doi: 10.1128/jcm.33.9.2451-2453.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith J. A., Bryce E. A., Ngui-Yen J. H., Roberts F. J. Comparison of BACTEC 9240 and BacT/Alert blood culture systems in an adult hospital. J Clin Microbiol. 1995 Jul;33(7):1905–1908. doi: 10.1128/jcm.33.7.1905-1908.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Snyder J. W., Lude S. K. Comparison of BacT/Alert FAN medium with BACTEC NR660 Plus 26A medium. J Clin Microbiol. 1996 May;34(5):1327–1329. doi: 10.1128/jcm.34.5.1327-1329.1996. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Strand C. L., Wajsbort R. R., Sturmann K. Effect of iodophor vs iodine tincture skin preparation on blood culture contamination rate. JAMA. 1993 Feb 24;269(8):1004–1006. [PubMed] [Google Scholar]
- Thorpe T. C., Wilson M. L., Turner J. E., DiGuiseppi J. L., Willert M., Mirrett S., Reller L. B. BacT/Alert: an automated colorimetric microbial detection system. J Clin Microbiol. 1990 Jul;28(7):1608–1612. doi: 10.1128/jcm.28.7.1608-1612.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weinstein M. P., Mirrett S., Reimer L. G., Wilson M. L., Smith-Elekes S., Chuard C. R., Joho K. L., Reller L. B. Controlled evaluation of BacT/Alert standard aerobic and FAN aerobic blood culture bottles for detection of bacteremia and fungemia. J Clin Microbiol. 1995 Apr;33(4):978–981. doi: 10.1128/jcm.33.4.978-981.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weisse M. E., Bass J. W., Young L. M. Pediatric blood culture: comparison of yields using aerobic, anaerobic and hypertonic media. Pediatr Infect Dis J. 1992 Feb;11(2):123–125. [PubMed] [Google Scholar]
- Weisse M. E. Utility of anaerobic medium in pediatric blood cultures. Pediatr Infect Dis J. 1994 Sep;13(9):837–838. doi: 10.1097/00006454-199409000-00024. [DOI] [PubMed] [Google Scholar]
- Welby P. L., Keller D. S., Storch G. A. Comparison of automated Difco ESP blood culture system with biphasic BBL Septi-Chek system for detection of bloodstream infections in pediatric patients. J Clin Microbiol. 1995 May;33(5):1084–1088. doi: 10.1128/jcm.33.5.1084-1088.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wilson M. L., Weinstein M. P., Mirrett S., Reimer L. G., Feldman R. J., Chuard C. R., Reller L. B. Controlled evaluation of BacT/alert standard anaerobic and FAN anaerobic blood culture bottles for the detection of bacteremia and fungemia. J Clin Microbiol. 1995 Sep;33(9):2265–2270. doi: 10.1128/jcm.33.9.2265-2270.1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wilson M. L., Weinstein M. P., Reimer L. G., Mirrett S., Reller L. B. Controlled comparison of the BacT/Alert and BACTEC 660/730 nonradiometric blood culture systems. J Clin Microbiol. 1992 Feb;30(2):323–329. doi: 10.1128/jcm.30.2.323-329.1992. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zaidi A. K., Knaut A. L., Mirrett S., Reller L. B. Value of routine anaerobic blood cultures for pediatric patients. J Pediatr. 1995 Aug;127(2):263–268. doi: 10.1016/s0022-3476(95)70305-5. [DOI] [PubMed] [Google Scholar]
- Zwadyk P., Jr, Pierson C. L., Young C. Comparison of Difco ESP and Organon Teknika BacT/Alert continuous-monitoring blood culture systems. J Clin Microbiol. 1994 May;32(5):1273–1279. doi: 10.1128/jcm.32.5.1273-1279.1994. [DOI] [PMC free article] [PubMed] [Google Scholar]