Abstract
To our knowledge there have been no reports of late rupture of the posterior cruciate ligament (PCL) as a cause of instability in PCL-retaining total knee prostheses. In our experience of 150 total knee replacements using PCL-retaining prosthesis, three cases (2.0%) of late rupture of the posterior cruciate ligament have occurred, each leading to chronic instability, disabling pain, and revision arthroplasty. In each case rupture of the posterior cruciate ligament was confirmed at the time of revision arthroplasty. The use of a more constrained prosthesis led to a successful outcome in each case.
Full text
PDFImages in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Galinat B. J., Vernace J. V., Booth R. E., Jr, Rothman R. H. Dislocation of the posterior stabilized total knee arthroplasty. A report of two cases. J Arthroplasty. 1988;3(4):363–367. doi: 10.1016/s0883-5403(88)80039-1. [DOI] [PubMed] [Google Scholar]
- Insall J., Tria A. J., Scott W. N. The total condylar knee prosthesis: the first 5 years. Clin Orthop Relat Res. 1979 Nov-Dec;(145):68–77. [PubMed] [Google Scholar]
- Skolnick M. D., Coventry M. B., Ilstrup D. M. Geometric total knee arthroplasty. A two-year follow-up study. J Bone Joint Surg Am. 1976 Sep;58(6):749–753. [PubMed] [Google Scholar]