BioArgos: a fully automated blood culture system.

R J Courcol 1, M Duhamel 1, A Decoster 1, V M Lemaire 1, M L Rastorgoueff 1, D Ochin 1, G R Martin 1
PMCID: PMC265430  PMID: 1500505

Abstract

BioArgos (Sanofi Diagnostics Pasteur, Marnes-la-Coquette, France) is a fully automated blood culture system that detects carbon dioxide production by infrared spectroscopy through a glass bottle. This hands-off system was compared with the BACTEC NR-660 system (Becton Dickinson Diagnostic Instrument Systems, Towson, Md.). A total of 336 microorganisms belonging to 74 taxa were tested in simulated blood cultures by both systems. Experimental data showed no significant differences between the two systems. The inclusive detection times (+/- the standard deviations) were 33.2 +/- 28.7 and 35.0 +/- 30.6 h with BioArgos and BACTEC, respectively. Anaerobes were detected earlier with BioArgos, whereas detection of some organisms that need oxygen to grow was slightly delayed. In conclusion, BioArgos is as reliable and accurate as BACTEC NR-660 and shows better practicability owing to noninvasive detection, reduction of vial manipulation, and absence of daily maintenance.

Full text

PDF
1997

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Beckers B., Lang H. R. Bioluminescent measurement of ATP for the rapid detection of positive blood cultures. Naturwissenschaften. 1982 Mar;69(3):145–146. doi: 10.1007/BF00376723. [DOI] [PubMed] [Google Scholar]
  2. Courcol R. J., Fruchart A., Roussel-Delvallez M., Martin G. R. Routine evaluation of the nonradiometric BACTEC NR 660 system. J Clin Microbiol. 1986 Jul;24(1):26–29. doi: 10.1128/jcm.24.1.26-29.1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. DeBlanc H. J., Jr, DeLand F., Wagner H. N., Jr Automated radiometric detection of bacteria in 2,967 blood cultures. Appl Microbiol. 1971 Nov;22(5):846–849. doi: 10.1128/am.22.5.846-849.1971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. DeLand F. H., Wagner H. N., Jr Early detection of bacterial growth, with carbon-14-labeled glucose. Radiology. 1969 Jan;92(1):154–155. doi: 10.1148/92.1.154. [DOI] [PubMed] [Google Scholar]
  5. Finegold S. M., White M. L., Ziment I., Winn W. R. Rapid diagnosis of bacteremia. Appl Microbiol. 1969 Sep;18(3):458–463. doi: 10.1128/am.18.3.458-463.1969. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Graves M. H., Morello J. A., Kocka F. E. Sodium polyanethol sulfonate sensitivity of anaerobic cocci. Appl Microbiol. 1974 Jun;27(6):1131–1133. doi: 10.1128/am.27.6.1131-1133.1974. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Mantz J. M., Faller J. P., Le Faou A., Jaeger A., Minck R. Intérêt du comptage de germes pour l'interprétation d'une hémoculture positive. Nouv Presse Med. 1981 Jul 11;10(30):2489–2490. [PubMed] [Google Scholar]
  8. Thorpe T. C., Wilson M. L., Turner J. E., DiGuiseppi J. L., Willert M., Mirrett S., Reller L. B. BacT/Alert: an automated colorimetric microbial detection system. J Clin Microbiol. 1990 Jul;28(7):1608–1612. doi: 10.1128/jcm.28.7.1608-1612.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES