Abstract
The one-bottle Oxoid Signal blood culture system altered to provide a more aerobic bottle headspace was reassessed in a comparative study versus the two-bottle BACTEC radiometric system in 5,426 blood cultures. The BACTEC system detected more microorganisms (P less than 0.02), particularly anaerobes (P less than 0.05) and fungi (P less than 0.05).
Full text
PDFSelected References
These references are in PubMed. This may not be the complete list of references from this article.
- Murray P. R., Niles A. C., Heeren R. L., Curren M. M., James L. E., Hoppe-Bauer J. E. Comparative evaluation of the oxoid signal and Roche Septi-Chek blood culture systems. J Clin Microbiol. 1988 Dec;26(12):2526–2530. doi: 10.1128/jcm.26.12.2526-2530.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sawhney D., Hinder S., Swaine D., Bridson E. Y. Novel method for detecting micro-organisms in blood cultures. J Clin Pathol. 1986 Nov;39(11):1259–1263. doi: 10.1136/jcp.39.11.1259. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weinstein M. P., Mirrett S., Reimer L. G., Reller L. B. Effect of agitation and terminal subcultures on yield and speed of detection of the Oxoid Signal blood culture system versus the BACTEC radiometric system. J Clin Microbiol. 1989 Mar;27(3):427–430. doi: 10.1128/jcm.27.3.427-430.1989. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weinstein M. P., Mirrett S., Reller L. B. Comparative evaluation of Oxoid Signal and BACTEC radiometric blood culture systems for the detection of bacteremia and fungemia. J Clin Microbiol. 1988 May;26(5):962–964. doi: 10.1128/jcm.26.5.962-964.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weinstein M. P., Reller L. B., Murphy J. R., Lichtenstein K. A. The clinical significance of positive blood cultures: a comprehensive analysis of 500 episodes of bacteremia and fungemia in adults. I. Laboratory and epidemiologic observations. Rev Infect Dis. 1983 Jan-Feb;5(1):35–53. doi: 10.1093/clinids/5.1.35. [DOI] [PubMed] [Google Scholar]