Skip to main content
Journal of Clinical Microbiology logoLink to Journal of Clinical Microbiology
. 1992 Dec;30(12):3239–3242. doi: 10.1128/jcm.30.12.3239-3242.1992

Value of extended agitation and subculture of BACTEC NR 660 aerobic resin blood culture bottles for clinical yeast isolates.

E Prevost-Smith 1, N Hutton 1
PMCID: PMC270640  PMID: 1452708

Abstract

From 10,351 blood cultures, we prospectively studied 1,000 BACTEC NR 660 aerobic resin blood culture bottles (26+ and Peds Plus) for patients suspected of having yeast septicemia to determine whether extended agitation and subculturing would increase the recovery of yeasts. Aerobic bottles were agitated continuously for 144 h. On day 7, 1,000 culture-negative aerobic bottles which had fungal blood culture requests were agitated for an additional 14 days. During this time they were subcultured twice and read twice by BACTEC NR 660. ON days 1 to 7, 81 bottles were cultured positive for yeasts from 36 patients, which included 44 isolates of Candida albicans, averaging 1.4 days to detection, and 12 isolates of Cryptococcus neoformans, averaging 3.8 days to detection. The average detection time for all yeasts was 2.2 days. On days 7 to 21, no yeasts were detected by BACTEC or recovered from the subcultures. We conclude that when continuously agitated for at least 5 full days (120 h), the BACTEC NR 660 aerobic resin bottles reliably isolate yeasts, and it is unnecessary to subculture or hold these bottles beyond 5 days. It also eliminates the need for an additional blood culture system for yeast detection, thus saving (i) confusion in the collection process, (ii) patients' blood and money, and (iii) laboratory technologists' time.

Full text

PDF
3240

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Bille J., Stockman L., Roberts G. D., Horstmeier C. D., Ilstrup D. M. Evaluation of a lysis-centrifugation system for recovery of yeasts and filamentous fungi from blood. J Clin Microbiol. 1983 Sep;18(3):469–471. doi: 10.1128/jcm.18.3.469-471.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Body B. A., Pfaller M. A., Durrer J., Koontz F., Gröschel D. H. Comparison of the lysis centrifugation and radiometric blood culture systems for recovery of yeast. Eur J Clin Microbiol Infect Dis. 1988 Jun;7(3):417–420. doi: 10.1007/BF01962353. [DOI] [PubMed] [Google Scholar]
  3. Brannon P., Kiehn T. E. Large-scale clinical comparison of the lysis-centrifugation and radiometric systems for blood culture. J Clin Microbiol. 1985 Dec;22(6):951–954. doi: 10.1128/jcm.22.6.951-954.1985. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Huahua T., Rudy J., Kunin C. M. Effect of hydrogen peroxide on growth of Candida, Cryptococcus, and other yeasts in simulated blood culture bottles. J Clin Microbiol. 1991 Feb;29(2):328–332. doi: 10.1128/jcm.29.2.328-332.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Jungkind D., Millan J., Allen S., Dyke J., Hill E. Clinical comparison of a new automated infrared blood culture system with the BACTEC 460 system. J Clin Microbiol. 1986 Feb;23(2):262–266. doi: 10.1128/jcm.23.2.262-266.1986. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Kiehn T. E., Wong B., Edwards F. F., Armstrong D. Comparative recovery of bacteria and yeasts from lysis-centrifugation and a conventional blood culture system. J Clin Microbiol. 1983 Aug;18(2):300–304. doi: 10.1128/jcm.18.2.300-304.1983. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Koontz F. P., Flint K. K., Reynolds J. K., Allen S. D. Multicenter comparison of the high volume (10 ml) NR BACTEC PLUS system and the standard (5 ml) NR BACTEC system. Diagn Microbiol Infect Dis. 1991 Mar-Apr;14(2):111–118. doi: 10.1016/0732-8893(91)90044-g. [DOI] [PubMed] [Google Scholar]
  8. Murray P. R. Comparison of the lysis-centrifugation and agitated biphasic blood culture systems for detection of fungemia. J Clin Microbiol. 1991 Jan;29(1):96–98. doi: 10.1128/jcm.29.1.96-98.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Muñoz P., Bernaldo de Quirós J. C., Berenguer J., Rodríguez Créixems M., Picazo J. J., Bouza E. Impact of the BACTEC NR system in detecting Candida fungemia. J Clin Microbiol. 1990 Mar;28(3):639–641. doi: 10.1128/jcm.28.3.639-641.1990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Prevost E., Bannister E. Detection of yeast septicemia by biphasic and radiometric methods. J Clin Microbiol. 1981 Apr;13(4):655–660. doi: 10.1128/jcm.13.4.655-660.1981. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Robinson P. G., Sulita M. J., Matthews E. K., Warren J. R. Failure of the Bactec 460 radiometer to detect Cryptococcus neoformans fungemia in an AIDS patient. Am J Clin Pathol. 1987 Jun;87(6):783–786. doi: 10.1093/ajcp/87.6.783. [DOI] [PubMed] [Google Scholar]
  12. Tarrand J. J., Guillot C., Wenglar M., Jackson J., Lajeunesse J. D., Rolston K. V. Clinical comparison of the resin-containing BACTEC 26 Plus and the Isolator 10 blood culturing systems. J Clin Microbiol. 1991 Oct;29(10):2245–2249. doi: 10.1128/jcm.29.10.2245-2249.1991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  13. Telenti A., Roberts G. D. Fungal blood cultures. Eur J Clin Microbiol Infect Dis. 1989 Sep;8(9):825–831. doi: 10.1007/BF02185855. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)