Abstract

The automated and computerized AutoMicrobic system (AMS; Vitek Systems, Inc., subsidiary of McDonnell Douglas, Hazelwood, Mo.) was evaluated as a means of identifying the Enterobacteriaceae. The Micro-ID system (General Diagnostics, Morris Plains, N.J.) and, when necessary, conventional tubed media were used for comparison. Identification by AMS and Micro-ID differed in only 12 of 1,528 isolates (0.8%). Disagreements occurred primarily with Enterobacter spp. Precision testing of the AMS showed only 1 of 72 tests (1.4%) deviating from the expected. The AMS was found to be an accurate and precise method for the identification of Enterobacteriaceae.

Full text

PDF
333

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Aldridge C., Jones P. W., Gibson S., Lanham J., Meyer M., Vannest R., Charles R. Automated microbiological detection/identification system. J Clin Microbiol. 1977 Oct;6(4):406–413. doi: 10.1128/jcm.6.4.406-413.1977. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Barry A. L., Badal R. E. Rapid identification of Enterobacteriaceae with the micro-ID system versus API 20E and conventional media. J Clin Microbiol. 1979 Sep;10(3):293–298. doi: 10.1128/jcm.10.3.293-298.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Blazevic D. J., Mackay D. L., Warwood N. M. Comparison of Micro-ID and API 20E systems for identification of Enterobacteriaceae. J Clin Microbiol. 1979 May;9(5):605–608. doi: 10.1128/jcm.9.5.605-608.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Buesching W. J., Rhoden D. L., Esaias A. O., Smith P. B., Washington J. A., 2nd Evaluation of the modified Micro-ID system for identification of Enterobacteriaceae. J Clin Microbiol. 1979 Oct;10(4):454–458. doi: 10.1128/jcm.10.4.454-458.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Edberg S. C., Atkinson B., Chambers C., Moore M. H., Palumbo L., Zorzon C. F., Singer J. M. Clinical evaluation of the MICRO-ID, API 20E, and conventional media systems for identification of Enterobacteriacea. J Clin Microbiol. 1979 Aug;10(2):161–167. doi: 10.1128/jcm.10.2.161-167.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Isenberg H. D., Gavan T. L., Smith P. B., Sonnenwirth A., Taylor W., Martin W. J., Rhoden D., Balows A. Collaborative investigation of the AutoMicrobic System Enterobacteriaceae biochemical card. J Clin Microbiol. 1980 Jun;11(6):694–702. doi: 10.1128/jcm.11.6.694-702.1980. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Isenberg H. D., Gavan T. L., Sonnenwirth A., Taylor W. I., Washington J. A., 2nd Clinical laboratory evaluation of automated microbial detection/identification system in analysis of clinical urine specimens. J Clin Microbiol. 1979 Aug;10(2):226–230. doi: 10.1128/jcm.10.2.226-230.1979. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Sonnenwirth A. C. Preprototype of an automated microbial detection and identification system: a developmental investigation. J Clin Microbiol. 1977 Oct;6(4):400–405. doi: 10.1128/jcm.6.4.400-405.1977. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES