Abstract
A local estimation of pulse wave speed c, an important predictor of cardiovascular events, can be obtained at arterial locations where simultaneous measurements of blood pressure (P) and velocity (U), arterial diameter (D) and U, flow rate (Q) and cross-sectional area (A), or P and D are available, using the PU-loop, sum-of-squares (), lnDU-loop, QA-loop or new D2P-loop methods. Here, these methods were applied to estimate c from numerically generated P, U, D, Q and A waveforms using a visco-elastic one-dimensional model of the 55 larger human systemic arteries in normal conditions. Theoretical c were calculated from the parameters of the model. Estimates of c given by the loop methods were closer to theoretical values and more uniform within each arterial segment than those obtained using the . The smaller differences between estimates and theoretical values were obtained using the D2P-loop method, with root-mean-square errors (RMSE) smaller than , followed by averaging the two c given by the PU- and -loops (RMSE ). In general, the errors of the PU-, - and QA-loops decreased at locations where visco-elastic effects were small and nearby junctions were well-matched for forward-travelling waves. The performed better at proximal locations.
Keywords: Pulse wave propagation, Arterial stiffness, One-dimensional modelling, Voigt-type visco-elasticity, Well-matched bifurcations
1. Introduction
The changes in blood pressure and flow at the ascending aorta generated by the contraction of the left ventricle propagate throughout the systemic arterial network in the form of waves, referred to as pulse waves (Caro et al., 1978; Pedley, 1980). The speed at which pulse waves travel in the absence of any convective velocity, known as pulse wave speed c, is a measure of arterial stiffness, which has been identified as an important predictor of cardiovascular events (Blacher et al., 1999; Laurent et al., 2001; Meaume et al., 2001). Knowledge of c is also of paramount importance for pulse wave analysis; in particular to separate the pressure and flow waveforms measured at a single location into forward- and backward-travelling components (Parker and Jones, 1990).
Several methods have been proposed to estimate c from in vivo arterial waveforms. In a clinical setting, the most widely used method calculates c non-invasively as the ratio of the distance between two measuring sites (usually the carotid and femoral arteries) and the transit time between these sites (Laurent et al., 2006; Mancia et al. 2007). The recent Arteriograph method estimates c from the pressure waveform measured at the brachial artery with an occluding cuff (Horváth et al., 2010). By their nature, both methods provide some average estimation of c over a length of the arterial tree. Local estimations of c at a single measurement site are possible if simultaneous measurements of blood pressure (P) and velocity (U) (Khir et al., 2001; Davies et al., 2006), arterial diameter (D) and U (Feng and Khir, 2010), or flow rate (Q) and luminal area (A) (Rabben et al., 2004) are available. These methods are referred to as the PU-loop, sum-of-squares (), -loop and QA-loop, respectively, and will be described in Sections 2.3–2.6.
All these methods were developed in the time domain. Frequency-domain methods for the estimation of c have also been developed (O’Rourke and Taylor, 1967; Cox and Bagshaw, 1975; Milnor and Bertram, 1978; van Huis et al., 1987; Stergiopulos et al., 1999). However, they showed greater variation than the PU-loop and when applied to P and U measurements in the human carotids in normal conditions (Aguado-Sierra et al., 2006).
The accuracy of each method is difficult to assess in vivo, since the real value of c is usually unknown. Numerically simulated pulse waveforms allow us to compare estimates of c given by each method with theoretical values obtained from the parameters of the model. Pulse wave propagation in large systemic arteries can be accurately simulated at a reasonable computational cost using the one-dimensional (1-D) equations of blood flow in compliant vessels (Olufsen et al., 2000; Steele et al., 2003; Matthys et al., 2007; Bessems et al., 2008; Reymond et al., 2009). Trachet et al. (2010) applied these equations to compare theoretical values of c with estimates given by the carotid–femoral and Arteriograph methods, showing that both methods underestimate c.
The aim of this work is to test the performance of the PU-loop, , -loop and QA-loop methods using pulse waveforms generated in a 1-D model of the 55 larger human systemic arteries in normal conditions (Fig. 1). This model can capture the main features of in vivo pulse waveforms (Stergiopulos et al., 1992). In addition, a new method to estimate c from simultaneous P and D measurements (the D2P-loop) will be described (Section 2.7) and tested. The effect on the performance of each method of visco-elasticity and reflections at arterial junctions will be analysed, and the mechanisms underlying the discrepancies between theoretical and estimated c will be discussed.
2. Materials and methods
2.1. Numerical pulse waveforms
Pulse waveforms were simulated in the 55-artery network in Fig. 1 using the nonlinear 1-D equations of incompressible and axisymmetric flow in Voigt-type visco-elastic vessels
(1) |
where x is the axial coordinate along the vessel, t the time, A(x,t) the cross-sectional area of the lumen, U(x,t) the average axial velocity, P(x,t) the average internal pressure over the cross section, Kg m−3 the mass density of blood, f the friction force per unit length, and the blood viscosity. The arterial wall is assumed to be thin, homogeneous, incompressible and visco-elastic, with each cross section deforming axisymmetrically and independent of the others. The parameters h(x), E(x) and are the thickness, Young's modulus and Voigt-type viscous modulus of the arterial wall, respectively, and A0(x) is the luminal area at P=0. and are assumed to be constant and related to the elastic and visco-elastic properties of the arterial wall, respectively.
Eq. (1) were solved in the arterial network of Fig. 1 using a discontinuous Galerkin scheme with a spectralhp spatial discretisation and zero pressures and flows as initial conditions. The time step was for the elastic-wall models and for the visco-elastic-wall model. Arterial segments were divided into non-overlapping elements with a 2 cm length (when physically possible) and a polynomial and quadrature order of 3. Elements or segments shorter than 2 cm were given a polynomial and quadrature order of 2. Further details on the 1-D formulation are given in the supplementary material.
The physiological data of the models were taken from Alastruey (in press) and are shown in Tables 1–3 (supplementary material). A version of the 55-artery model with purely elastic vessels and modified radii so that junctions are well-matched for forward-travelling waves (i.e. waves are initially reflected at terminal branches) was also considered. This model will be referred to as the ‘well-matched model’.
The periodic flow rate shown in Fig. 1 was prescribed at the aortic root for the first 10 s, followed by zero flow to study the relaxation of the system. Terminal branches were coupled to matched three-element windkessel models, as described by Alastruey et al. (2008), with the parameters shown in Table 2 (supplementary material). Conservation of mass and continuity of the total pressure was enforced at the arterial junctions.
2.2. Theoretical pulse wave speed
System (1) is hyperbolic and admits wave-like solutions for P and U when . In that case, Riemann's method of characteristics shows that disturbances to the flow create changes in pressure and velocity across the wave fronts, which propagate forward (in the positive x-direction) with speed U+c and backward (in the negative x-direction) with speed U−c, where (Parker and Jones, 1990). For the tube law in (1), with and A0 constant, we have
(2) |
Moreover, and consist of a forward- (indicated with the subscript +) and a backward- (indicated with the subscript −) travelling component, which are related through the ‘water hammer’ equation
(3) |
In this work the theoretical value of the local c was calculated using Eq. (2) with A the mean cross-sectional area over the cardiac cycle.
2.3. PU-loop
This method assumes that there is a period in early systole when there are no reflected waves, so that and are zero. This leads to and , so that according to Eq. (3). Therefore c can be obtained as
(4) |
In the arteries simulated, the part of the PU-loop corresponding to early systole is approximately linear (Fig. 2a). Its slope is (Eq. (4)), which allows us to determine c using a reasonable approximation for (Khir et al., 2001). Later in the cardiac cycle, forward and backward waves are present and the PU-loop is no longer linear.
2.4. Sum of squares ()
This method was derived to estimate c in the coronary arteries, where the PU-loop does not show a period of time when pulse waves are unidirectional. Given simultaneous P and U measurements at a single location, c is obtained as
(5) |
where the sums extend over a cardiac cycle. Eq. (5) follows from Eq. (3) assuming to be small compared with . This is equivalent to finding the value of that minimises net wave energies (Davies et al., 2006).
2.5. -loop
Feng and Khir (2010) showed that c, and changes in the natural logarithm of the arterial diameter are related through
(6) |
during a reflection-free period of the cardiac cycle. In the arteries simulated, the part of the -loop corresponding to early systole is approximately linear (Fig. 2b). Its slope is 1/(2c) according to Eq. (6).
2.6. QA-loop
If there is a reflection-free period in the cardiac cycle, c can be estimated as the ratio between the change in flow rate and the change in cross-sectional area (Rabben et al., 2004); i.e.
(7) |
In the arteries simulated, the part of the QA-loop corresponding to early systole is approximately linear (Fig. 2c). Its slope is c according to Eq. (7).
2.7. D2P-loop
Assuming the arterial wall to be a Voigt-type visco-elastic material, the numerical results show that the visco-elastic term in the tube law in (1) becomes increasingly constant with time in diastole and the elastic term leads to an approximately linear relationship between D2 and P in late diastole (Fig. 2d). The slope can be calculated by differentiating the elastic term in the tube law, with and A0 assumed to be constant. This leads to
(8) |
at (P, A)=(0, A0). Eq. (2) with A=A0 was used for the second equality. Solving Eq. (8) for c with and yields
(9) |
Eq. (9) allows us to calculate c from the slope of the D2P-loop in late diastole using a reasonable value for and approximating D0 by the mean arterial diameter.
3. Results
The simulated pressure and velocity waveforms along the aorta and left iliac, femoral and tibial arteries of the purely elastic and visco-elastic 55-artery models are shown in Fig. 3 for the tenth cardiac cycle, when the pulse waveform has become periodic. At the aortic root, the systolic pressure is 15.6 kPa and the diastolic pressure 10.9 kPa for the purely elastic model. The corresponding values for the visco-elastic model are 16.6 and 10.6 kPa. The difference between these pressures (the pulse pressure) increases in the aorta toward distal locations. In both models, fluid viscosity decreases the mean pressure from 13.3 kPa at the aortic root to 13.2 kPa at the end of the aorta and 7.7 kPa at the end of the tibial arteries (Segments 49 and 55). The feet of the pressure waveforms show that the speed of pulse wave propagation increases distally; this is observed more clearly in the purely elastic model. In early diastole, the flow reverses in most of the aortic and iliac segments. Later in diastole pulse waveforms decrease approximately exponentially with similar time constants, especially for the pressure wave. This decay continues after 10 s, when a zero flow rate at the aortic root was prescribed.
Fig. 4 shows the theoretical wave speed c calculated using Eq. (2) along the aortic–iliac–femoral–tibial path in 1 cm increments, for the (a, b) purely elastic, (c, d) visco-elastic and (e, f) well-matched models. They are compared against the speeds determined using the PU-loop, , -loop, and QA-loop. Overall, the estimations of c given by these loop methods are closer to the theoretical value and more uniform within each segment than those obtained using the . The errors of these four methods increased in the visco-elastic model, especially at proximal locations, and decreased in the well-matched model. In those segments where the PU-loop overestimated c, the -loop underestimated it, and vice versa. This result suggests that c could be estimated closer to the theoretical value using the average of the two values obtained from the PU- and -loops. The c values estimated using the D2P-loop cannot be distinguished from the theoretical values in the scale of Fig. 4. Similar qualitative comparisons of c estimates and theoretical values were obtained in other arterial segments.
Table 1 shows the root-mean-square errors (RMSE) between theoretical and estimated c in the three arterial models. Absolute errors were calculated at 304 locations: the inlet, midpoint and outlet of each arterial segment, and at all the locations on the aortic–iliac–femoral–tibial path of Fig. 4. RMSE were then calculated for all the segments in the aorta and each generation of bifurcations from the aorta. The larger RMSE are mostly obtained in the 4th and 5th generation of bifurcations, which are terminal segments except for Segment 9 in the right arm.
Table 1.
Aorta | 1st gen. | 2nd gen. | 3rd gen. | 4th gen. | 5th gen. | |
---|---|---|---|---|---|---|
Samples | 70 | 39 | 50 | 80 | 59 | 6 |
PU-loop | 0.78 2.900.69 | 0.82 1.990.49 | 0.93 1.100.54 | 0.54 0.420.33 | 3.09 3.980.47 | 0.63 2.141.08 |
2.56 4.660.65 | 6.89 9.183.04 | 4.75 6.684.96 | 4.37 6.063.63 | 6.98 8.357.50 | 8.21 11.5123.45 | |
-loop | 0.76 3.580.64 | 0.69 0.650.43 | 1.90 2.400.50 | 0.57 0.360.28 | 2.41 2.850.50 | 0.68 1.971.11 |
QA-loop | 1.01 2.250.89 | 0.87 0.740.73 | 1.10 2.130.85 | 0.59 0.490.42 | 1.77 2.680.74 | 1.33 1.991.35 |
D2P-loop | 0.02 0.180.00 | 0.02 0.020.01 | 0.02 0.010.01 | 0.02 0.020.02 | 0.01 0.010.02 | 0.01 0.010.02 |
Average loops | 0.04 2.990.03 | 0.13 0.890.06 | 0.52 0.880.03 | 0.03 0.190.03 | 0.35 0.630.03 | 0.03 0.370.03 |
RMSE for the ‘loop’ methods decreased in most of the locations if bifurcations were well-matched for forward-travelling waves and the arterial wall was purely elastic. Estimation of c using the average of the two c from the PU- and -loops considerably reduced the errors in any group and model (with the exception of the segments of the aorta and first generation in the visco-elastic model), leading to RMSE smaller than 0.52 m s−1 in the purely elastic model, 2.99 m s−1 in the visco-elastic model, and 0.06 m s−1 in the well-matched model. RMSE for the decreased at all locations if visco-elastic effects were neglected, and mainly at proximal locations if bifurcations were well-matched. They increased considerably in the 5th generation of bifurcations of the well-matched model. Absolute errors smaller than 0.67 m s−1 were obtained using the D2P-loop at any location for any model.
In the visco-elastic model, the diastolic region of the pulse waveform used for the linear fitting in the D2P-loop was longer than the systolic region in the PU-, - and QA-loops. Fig. 2e compares these regions in an aortic pressure waveform.
The produced RMSE smaller than 1.4 m s−1 in the well-matched model if terminal boundary conditions were completely absorbent. If, in addition, the flow was inviscid (), then P, U, D, Q and A were approximately proportional at any location in the network and the absolute errors of any of the methods studied were less than 0.1 m s−1.
4. Discussion
The ability of the PU-loop, , -loop, QA-loop, and D2P-loop methods to estimate theoretical values of c was tested using pulse waveforms generated in a nonlinear 1-D model of the 55 larger systemic arteries in the human in normal conditions. Although these data are an approximation to in vivo pulse waveforms, they have the advantage of being free of measurement and alignment errors, and of providing theoretical values of c from the parameters of the model to compare with the estimates. The use of the 1-D formulation is justified because all the methods studied are based on this formulation.
Overall, the loop methods produced estimations of c closer to the theoretical value than did the . The D2P-loop led to the smallest RMSE, followed by averaging the estimations from the PU- and -loops. The latter method, however, requires simultaneous measurements of three waveforms (P, U and D), instead of the two required by the other methods studied. This limits its in vivo applicability.
The absolute errors in the estimates of c using the PU-, , and QA-loops increased in most of the arterial locations studied if visco-elasticity was modelled. They decreased if bifurcations were well-matched for forward-travelling waves. These observations are in agreement with the assumptions of purely elastic wall and absence of any reflected wave in early systole made in the derivation of these methods. Indeed, Alastruey et al. (2009) showed, for a 55-artery network with the same reflection coefficients as in the normal model used here, the arrival of waves reflected at nearby junctions in early systole, when the relationship between P and U, and U, or Q and A is still approximately linear (e.g. within the red regions in Fig. 2). These waves change the slope of the linear part of the PU-, - and QA-loops and, hence, are responsible for the errors observed in their estimates of c. Consequently, these loop methods are more useful for estimating the theoretical c at locations where the visco-elastic effects are small and nearby junctions are distant and close to well-matched for forward-travelling waves. Moreover, the difference between the estimates of c obtained from the PU- and -loops seem to correlate well with the intensity of reflected waves in the system; the greater the intensity is, the greater the difference is. The demonstration of the utility of this difference to estimate the intensity of reflected waves remains to be explored fully.
The new D2P-loop method does not have the problems discussed above, but it has only been tested using a Voigt-type visco-elastic model of the arterial wall. Although several studies have shown that this model provides a good approximation to the relationship between P and D observed in vivo (Armentano et al., 1995; Craiem et al., 2005; Čanić et al., 2006), more complex wall models have been proposed (e.g. Bessems et al., 2008; Devault et al., 2008; Valdez-Jasso et al., 2009) for which the D2P-loop was not tested. Moreover, the D2P-loop might not be linear in late diastole for short cardiac cycles. According to the simulations, it is linear when the pressure waveform becomes uniform in space. The PU-, - and QA-loops also have a linear part at the end of diastole, but their slopes seem not to be related to c.
The was originally derived for the coronary arteries, which were not considered in this study. Nevertheless, the errors obtained here should be similar or even greater in these arteries, since vessel movement due to myocardial contraction (which was not considered in the derivation of the ) is a potential additional source of error. Theoretically, the can provide flawed results when measurements are taken close to a reflection site with a large reflection coefficient, so that there is significant correlation between forward- and backward-travelling waves (Davies et al., 2006). According to Kolyva et al. (2008), however, the main source of error of the in the coronary arteries is their windkesselness nature rather than the existence of reflections. Indeed, they concluded that the is inappropriate for estimating c in coronary arteries. In the present study, the performance of the improved at proximal locations if bifurcations were well-matched for forward-travelling waves. This suggests that wave reflections should have some effect on the . Aguado-Sierra et al. (2006) showed, using in vitro data measured in a thin-walled latex tube, that the overestimated the foot-to-foot calculation of c by 3.5 m s−1 on average, with differences larger than 4.0 m s−1 at locations more than 10 cm away from the reflection sites. These differences are smaller than most of the RMSE reported in Table 1, but significant taking into account the relatively larger distance from the reflection sites of the experiment. The errors of the , however, do not seem to be critical when calculating the total wave energy carried during the heart cycle by forward- and backward-travelling waves in the coronary circulation (Siebes et al., 2009).
The PU-loop, and D2P-loop methods require knowledge of the blood density . The latter method, however, is less sensitive to changes in ; e.g. a 5% change in yields changes in the estimation of c smaller than 3% for the D2P-loop and of about 5% for the PU-loop and . Moreover, c varies nonlinearly with A (Eq. (2)) and, hence, P. The methods studied here, therefore, provide some average estimation of c over the cardiac cycle (Fig. 2f). Once the flow became periodic in the 55-artery models, changes in A within a cardiac cycle were smaller than 25% at aortic locations and 11% at the terminal branches of the head and limbs, leading to changes in the theoretical c smaller than 8% and 3%, respectively.
In vivo, the applicability and performance of the methods studied depends upon the feasibility of accurate measurements of the data they require. For the loop methods, a good accuracy is critical during the linear period, which is usually less than 50 m s (Khir et al., 2001; Rabben et al., 2004; Feng and Khir, 2010). Thus, a high sampling frequency is fundamental. According to this study, the D2P-loop is approximately linear for more than 50 ms (Fig. 2e), but this has not been tested in vivo. The existence of a time lag between P and U measurements is a potential additional source of error for the PU-loop, which can be corrected as described by Swalen and Khir (2009). The evolution of D can be measured non-invasively at some locations using ultrasound-based echo-tracking (Rabben et al., 2002). However, this technique, which is not applicable in the coronary arteries, may well introduce an additional error that limits the good performance of the D2P-loop method shown here.
5. Conclusions
According to the results presented here, the estimation of c closest to the theoretical value is obtained from simultaneous diameter and pressure measurements using the D2P-loop. If pressure and velocity measurements are available, the PU-loop produces estimates of c closer to the theoretical values than the , which performs better at proximal locations. The average c obtained from the PU- and -loops has a smaller error than the estimate provided by either method. Particular care should be taken when applying the PU-, - and QA-loops at locations with significant visco-elastic effects and reflected waves. These conclusions, however, still need to be assessed using in vivo data.
Conflict of interest statement
There is no conflict of interest between the author of this paper and other external researchers or organisations that could have inappropriately influenced this work.
Acknowledgments
The author would like to acknowledge Professor Peter Weinberg (Imperial College London) for his helpful comments on the paper. This work was funded by a British Heart Foundation Intermediate Basic Science Research Fellowship (FS/09/030/27812).
Footnotes
Supplementary data associated with this article can be found in the online version at doi:10.1016/j.jbiomech.2010.12.002.
Appendix A. Supplementary data
References
- Aguado-Sierra, J., Parker, K., Davies, J., Francis, D., Hughes, A., Mayet, J., 2006. Arterial pulse wave velocity in coronary arteries. In: 28th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, New York City, USA. Paper ThB09.2. [DOI] [PubMed]
- Alastruey J., Parker K., Peiró J., Sherwin S. Lumped parameter outflow models for 1-D blood flow simulations: effect on pulse waves and parameter estimation. Commun. Comput. Phys. 2008;4:317–336. [Google Scholar]
- Alastruey J., Parker K., Peiró J., Sherwin S. Analysing the pattern of pulse waves in arterial networks: a time-domain study. J. Eng. Math. 2009;64:331–351. [Google Scholar]
- Alastruey, J., On the mechanics underlying the reservoir—excess separation in systemic arteries and their implications for pulse wave analysis. Cardiov. Eng., in press, doi:10.1007/s10558-010-9109-9. [DOI] [PMC free article] [PubMed]
- Armentano R., Megnien J., Simon A., Bellenfant F., Barra J., Levenson J. Effects of hypertension on viscoelasticity of carotid and femoral arteries in humans. Hypertension. 1995;26:48–54. doi: 10.1161/01.hyp.26.1.48. [DOI] [PubMed] [Google Scholar]
- Bessems D., Giannopapa C., Rutten M., van de Vosse F. Experimental validation of a time-domain-based wave propagation model of blood flow in viscoelastic vessels. J. Biomech. 2008;41:284–291. doi: 10.1016/j.jbiomech.2007.09.014. [DOI] [PubMed] [Google Scholar]
- Blacher J., Asmar R., Djane S., London G., Safar M. Aortic pulse wave velocity as a marker of cardiovascular risk in hypertensive patients. Hypertension. 1999;33:1111–1117. doi: 10.1161/01.hyp.33.5.1111. [DOI] [PubMed] [Google Scholar]
- Čanić S., Tambača J., Guidoboni G., Mikelić A., Hartley C., Rosenstrauch D. Modeling viscoelastic behavior of arterial walls and their interaction with pulsatile blood flow. SIAM J. Appl. Math. 2006;67:164–193. [Google Scholar]
- Caro C., Pedley T., Schroter R., Seed W. Oxford University Press; 1978. The Mechanics of the Circulation. [Google Scholar]
- Cox R., Bagshaw R. Baroreceptor reflex control of arterial hemodynamics in the dog. Circ. Res. 1975;37:772–786. doi: 10.1161/01.res.37.6.772. [DOI] [PubMed] [Google Scholar]
- Craiem D., Graf S., Pessana F., Grignola J., Bia D., Gines F., Armentano R. Cardiovascular engineering: modelization of ventricular arterial interaction in systemic and pulmonary circulation. Latin Am. Appl. Res. 2005;35:111–114. [Google Scholar]
- Davies J., Whinnett Z., Francis D., Willson K., Foale R., Malik I., Hughes A., Parker K., Mayet J. Use of simultaneous pressure and velocity measurements to estimate arterial wave speed at a single site in humans. Am. J. Physiol. Heart Circ. Physiol. 2006;290:H878–H885. doi: 10.1152/ajpheart.00751.2005. [DOI] [PubMed] [Google Scholar]
- Devault K., Gremaud P., Novak V., Olufsen M., Vernières G., Zhao P. Blood flow in the circle of Willis: modeling and calibration. Multiscale Model. Simul. 2008;7:888–909. doi: 10.1137/07070231X. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Feng J., Khir A. Determination of wave speed and wave separation in the arteries using diameter and velocity. J. Biomech. 2010;43:455–462. doi: 10.1016/j.jbiomech.2009.09.046. [DOI] [PubMed] [Google Scholar]
- Horváth I., Németh A., Lenkey Z., Alessandri N., Tufano F., Kis P., Gaszner B., Cziráki A. Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity. J. Hypertension. 2010;28:2068–2075. doi: 10.1097/HJH.0b013e32833c8a1a. [DOI] [PubMed] [Google Scholar]
- Khir A., O’Brien A., Gibbs J., Parker K. Determination of wave speed and wave separation in the arteries. J. Biomech. 2001;34:1145–1155. doi: 10.1016/s0021-9290(01)00076-8. [DOI] [PubMed] [Google Scholar]
- Kolyva C., Spaan J., Piek J., Siebes M. Windkesselness of coronary arteries hampers assessment of human coronary wave speed by single-point technique. Am. J. Physiol. Heart Circ. Physiol. 2008;295:H482–H490. doi: 10.1152/ajpheart.00223.2008. [DOI] [PubMed] [Google Scholar]
- Laurent S., Boutouyrie P., Asmar R., Gautier I., Laloux B., Guize L., Ducimetiere P., Benetos A. Aortic stiffness is an independent predictor of all-cause and cardiovascular mortality in hypertensive patients. Hypertension. 2001;37:1236–1241. doi: 10.1161/01.hyp.37.5.1236. [DOI] [PubMed] [Google Scholar]
- Laurent S., Cockcroft J., van Bortel L., Boutouyrie P., Giannattasio C., Hayoz D., Pannier B., Vlachopoulos C., Wilkinson I. Expert consensus document on arterial stiffness: methodological issues and clinical applications. Eur. Heart J. 2006;27:2588–2605. doi: 10.1093/eurheartj/ehl254. [DOI] [PubMed] [Google Scholar]
- Mancia G., de Backer G., Dominiczak A., Cifkova R., Fagard R., Germano G., Grassi G., Heagerty A., Kjeldsen S., Laurent S., Narkiewicz K., Ruilope L., Rynkiewicz A., Schmieder R., Struijker Boudier H., Zanchetti A. 2007 guidelines for the management of arterial hypertension. J. Hypertension. 2007;25:1105–1187. doi: 10.1097/HJH.0b013e3282f0580f. [DOI] [PubMed] [Google Scholar]
- Matthys K., Alastruey J., Peiró J., Khir A., Segers P., Verdonck P., Parker K., Sherwin S. Pulse wave propagation in a model human arterial network: Assessment of 1-D numerical simulations against in vitromeasurements. J. Biomech. 2007;40:3476–3486. doi: 10.1016/j.jbiomech.2007.05.027. [DOI] [PubMed] [Google Scholar]
- Meaume S., Rudnichi A., Lynch A., Bussy C., Sebban C., Benetos A., Safar M. Aortic pulse wave velocity as a marker of cardiovascular disease in subjects over 70 years old. J. Hypertens. 2001;19:871–877. doi: 10.1097/00004872-200105000-00006. [DOI] [PubMed] [Google Scholar]
- Milnor W., Bertram C. The relation between arterial viscoelasticity and wave propagation in the canine femoral artery in vivo. Circ. Res. 1978;6:870–879. doi: 10.1161/01.res.43.6.870. [DOI] [PubMed] [Google Scholar]
- Olufsen M., Peskin C., Kim W., Pedersen E., Nadim A., Larsen J. Numerical simulation and experimental validation of blood flow in arteries with structured-tree outflow conditions. Ann. Biomed. Eng. 2000;28:1281–1299. doi: 10.1114/1.1326031. [DOI] [PubMed] [Google Scholar]
- O’Rourke M., Taylor M. Input impedance on the systemic circulation. Circ. Res. 1967;20:365–380. doi: 10.1161/01.res.20.4.365. [DOI] [PubMed] [Google Scholar]
- Parker K., Jones C. Forward and backward running waves in the arteries: analysis using the method of characteristics. J. Biomech. Eng. 1990;112:322–326. doi: 10.1115/1.2891191. [DOI] [PubMed] [Google Scholar]
- Pedley T. Cambridge University Press; 1980. The Fluid Mechanics of Large Blood Vessels. [Google Scholar]
- Rabben S., Bjærum S., Sørhus V., Torp H. Ultrasound-based vessel wall tracking: an auto-correlation technique with RF center frequency estimation. Ultrasound Med. Biol. 2002;28:507–517. doi: 10.1016/s0301-5629(02)00487-8. [DOI] [PubMed] [Google Scholar]
- Rabben S., Stergiopulos N., Hellevik L., Smiseth O., Slørdahl S., Urheim S., Angelsen B. An ultrasound-based method for determining pulse wave velocity in superficial arteries. J. Biomech. 2004;37:1615–1622. doi: 10.1016/j.jbiomech.2003.12.031. [DOI] [PubMed] [Google Scholar]
- Reymond P., Merenda F., Perren F., Rüfenacht D., Stergiopulos N. Validation of a one-dimensional model of the systemic arterial tree. Am. J. Physiol. Heart Circ. Physiol. 2009;297:H208–H222. doi: 10.1152/ajpheart.00037.2009. [DOI] [PubMed] [Google Scholar]
- Siebes M., Kolyva C., Verhoeff B.-J., Piek J., Spaan J. Potential and limitations of wave intensity analysis in coronary arteries. Med. Biol. Eng. Comput. 2009;47:233–239. doi: 10.1007/s11517-009-0448-x. [DOI] [PubMed] [Google Scholar]
- Steele B., Wan J., Ku J., Hughes T., Taylor C. In vivo validation of a one-dimensional finite-element method for predicting blood flow in cardiovascular bypass grafts. IEEE Trans. Biomed. Eng. 2003;50:649–656. doi: 10.1109/TBME.2003.812201. [DOI] [PubMed] [Google Scholar]
- Stergiopulos N., Young D., Rogge T. Computer simulation of arterial flow with applications to arterial and aortic stenoses. J. Biomech. 1992;25:1477–1488. doi: 10.1016/0021-9290(92)90060-e. [DOI] [PubMed] [Google Scholar]
- Stergiopulos N., Westerhof B., Westerhof N. Total arterial inertance as the fourth element of the windkessel model. Am. J. Physiol. 1999;276:H81–H88. doi: 10.1152/ajpheart.1999.276.1.H81. [DOI] [PubMed] [Google Scholar]
- Swalen M., Khir A. Resolving the time lag between pressure and flow for the determination of local wave speed in elastic tubes and arteries. J. Biomech. 2009;42:1574–1577. doi: 10.1016/j.jbiomech.2009.03.038. [DOI] [PubMed] [Google Scholar]
- Trachet B., Reymond P., Kips J., Swillens A., De Buyzere M., Suys B., Stergiopulos N., Segers P. Numerical validation of a new method to assess aortic pulse wave velocity from a single recording of a brachial artery waveform with an occluding cuff. Ann. Biomed. Eng. 2010;38:876–888. doi: 10.1007/s10439-010-9945-1. [DOI] [PubMed] [Google Scholar]
- Valdez-Jasso D., Haider M., Banks H., Santana D., German Y., Armentano R., Olufsen M. Analysis of viscoelastic wall properties in ovine arteries. IEEE Trans. Biomed. Eng. 2009;56:210–219. doi: 10.1109/TBME.2008.2003093. [DOI] [PubMed] [Google Scholar]
- van Huis G., Sipkema P., Westerhof N. Coronary input impedance during cardiac cycle as determined by impulse response method. Am. J. Physiol. Heart Circ. Physiol. 1987;22:317–324. doi: 10.1152/ajpheart.1987.253.2.H317. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.