Abstract
The magnetic susceptibility of tissue within and around an image voxel affects the magnetic field and thus the local frequency in that voxel. Recently, it has been shown that spatial maps of frequency can be used to quantify local susceptibility if the contributions of surrounding tissue can be deconvolved. Currently, such quantitative susceptibility mapping (QSM) methods employ gradient recalled echo (GRE) imaging to measure spatial differences in the signal phase evolution as a function of echo time, from which frequencies can be deduced. Analysis of these phase images, however, is complicated by phase wraps, despite the availability and usage of various phase unwrapping algorithms. In addition, lengthy high-resolution GRE scanning often heats the magnet bore, causing the magnetic field to drift over several Hertz, which is on the order of the frequency differences between tissues. Here, we explore the feasibility of applying the WAter Saturation Shift Referencing (WASSR) method for 3D whole brain susceptibility imaging. WASSR uses direct saturation of water protons as a function of frequency irradiation offset to generate frequency maps without phase wraps, which can be combined with any image or spectroscopy acquisition. By utilizing a series of fast short-echo-time direct saturation images with multiple radiofrequency offsets, a frequency correction for field drift can be applied based on the individual image phases. Regions of interest were delineated with an automated atlas-based method, and the average magnetic susceptibilities calculated from frequency maps obtained from WASSR correlated well with those from the phase-based multi-echo GRE approach at 3 Tesla.
Keywords: magnetic susceptibility, water saturation shift referencing (WASSR), field mapping, phase, direct saturation, quantitative susceptibility mapping (QSM)
Introduction
The recent availability of high magnetic fields (3T and 7T) has caused a surge in the technology development for quantitative susceptibility mapping (QSM) (de Rochefort et al., 2010; Haacke et al., 2010; Liu et al., 2009; Rauscher et al., 2008; Schweser et al., 2011b; Wharton and Bowtell, 2010; Wharton et al., 2010), as well as the determination of the susceptibility tensor in tissue (Li et al., 2012a; Li et al., 2012b; Liu, 2010; Liu et al., 2012). Currently, such QSM methods rely on phase maps generated from gradient-recalled echo (GRE) images to determine the voxel frequency (Duyn et al., 2007; Li et al., 2011; Liu, 2010; Liu et al., 2010a; Shmueli et al., 2009; Wharton and Bowtell, 2010). Although GRE is fast and available as a standard sequence on all human MRI scanners, this method has a few disadvantages. In particular, phase wraps complicate data interpretation at interfaces of structures with very different susceptibilities (e.g., near the sinuses), as well as at positions further from the magnetic field center that exhibit significantly different frequencies due to static magnetic field inhomogeneity. Problems also occur when spatial resolution is low with respect to the spatial rate of field change, causing intra-voxel inhomogeneity and phase measurement errors. Thus, for GRE imaging, there is a dilemma: increasing the echo time (TE) increases frequency contrast between tissues, but phase images acquired at longer TEs have an increasing number of phase wraps, while the magnitude images exhibit lower signal-to-noise ratio (SNR). However, SNR of the phase image is maximum at TE = T2* (Wu et al., 2012). Additionally, when high-resolution 3D imaging is performed, the GRE acquisition may take several minutes, thereby increasing sensitivity to motion artifacts and to static field drift, which may occur when gradient heating increases the bore temperature. Such drifts may range over several Hertz, which is on the order of magnitude of the measured frequency differences between tissues. When this occurs, the susceptibility measurements may not be reproducible.
The main parameter necessary for susceptibility quantification is the frequency shift in each voxel. Whole-brain resonance frequency maps without phase wraps can be obtained using the WAter Saturation Shift Referencing (WASSR) method, in which the resonance frequency per voxel is characterized by measuring direct water saturation as a function of saturation frequency offset (Kim et al., 2009; Smith et al., 2009). When using an RF pulse with low power and short duration, the frequency-dependent line shape is a Lorentzian that is not affected by inhomogeneous line broadening (Liu et al., 2010b; Mulkern and Williams, 1993) and has minimal contributions from conventional magnetization transfer (MT) and chemical exchange saturation transfer (CEST). The voxel resonance frequency can be determined from the minimum of the fitted Lorentzian curve, allowing the WASSR method to use only the signal magnitude for generating tissue frequency maps, which can then be processed to derive quantitative susceptibility maps with available technology. Here, we compare the use of WASSR and GRE for quantitative susceptibility mapping at 3 Tesla. This work has been presented partially in abstract form (Lim et al., 2011a; Lim et al., 2011b; Lim et al., 2012b).
Theory
Multi-Echo Gradient-Recalled Echo (GRE)
For each voxel, the rate of change of the signal phase as a function of echo time (TE) is related to Δ ωH2O = ωH2O − ωref, the radial resonance frequency of the voxel water shifted with respect to the scanner reference frequency (ωref):
[1] |
[2] |
where ϕ is the phase in radians, and ϕ0 corrects for any non-zero intercepts.
WAter Saturation Shift Referencing (WASSR)
Direct saturation (DS) of the water protons is achieved by applying a radiofrequency (RF) saturation pre-pulse of low power and short duration, minimizing contributions from magnetization transfer (MT) and chemical exchange saturation transfer (CEST) (Henkelman et al., 2001; Kim et al., 2009; Smith et al., 2009). By applying the saturation pulse across a range of RF frequencies, the resulting signal intensity in each voxel can be plotted as a function of RF offset frequency, resulting in a Z-spectrum for each voxel, as shown in Figure 1. The saturation spectrum can be described by an exact solution to the Bloch equations (Mulkern and Williams, 1993; Smith et al., 2009), in which the magnetization at steady state gives a Lorentzian lineshape as a function of the RF irradiation frequency (ωRF) with respect to the water frequency:
[3] |
where ΔωRF = ωRF − ωref, M0 is the initial signal without saturation in arbitrary units; R1 and R2 are the longitudinal and transverse relaxation rate constants in Hertz; ω1 = γB1 is the radial frequency of the saturation pulse field with amplitude B1
Several methods can be used to identify the water resonance frequency from the DS Z-spectrum, including polynomial fitting (Smith et al., 2009) and symmetry analysis with the Maximum Symmetry Algorithm with respect to the scanner reference frequency (Kim et al., 2009). The latter approach capitalizes on the fact that the shape of the direct saturation Z-spectrum is not affected by field inhomogeneities and is symmetric with respect to the center frequency (Kim et al., 2009). Dula et al. noted that the Maximum Symmetry Algorithm produced smoother frequency shift maps with less profound shift zones than seen in polynomial fitting (Dula et al., 2011; Dula et al., 2010), indicating that the Maximum Symmetry Algorithm produced a more robust map of the resonance frequency in a direct saturation experiment. However, the Maximum Symmetry Algorithm requires 16 to 33 points for accurate fitting (Kim et al., 2009).
To shorten the experiment and maintain a clinically viable scan time, we replaced Eqn. 3 with the equivalent general Lorentzian lineshape (Liu et al., 2010b; Sheth et al., 2011) to fit the direct saturation signal (Fig. 2a, b):
[4] |
where A is the signal amplitude in arbitrary units (Fig. 2d), C the signal remaining after saturation (Fig. 2e), and LW the linewidth at full-width-half-maximum (Fig. 2f) that depends on R1, R2, and ω1 = γB1 (Mulkern and Williams, 1993). After fitting this Lorentzian lineshape to the direct saturation spectrum in each voxel, we can map the water resonance frequency (Fig. 2g) with respect to the scanner reference frequency, similar to GRE imaging. This approach also allows one to map the signal amplitude A and the signal remaining after saturation C to measure the efficiency of the saturation pulse, as well as the linewidth LW to assess relaxation effects. For instance, the linewidth map allows a separate view into the effect of iron content (Smith et al., 2009).
As shown in Figure 1e, voxels close to the center of the brain usually have a resonance frequency that corresponds to the scanner reference frequency (referenced to 0Hz), which is automatically set by the scanner. At various other locations in the brain (Fig. 1a, c–d, f–g), background gradients and susceptibility effects from tissues can shift the resonance frequency of the voxel, resulting in a horizontal shift of the Z-spectra in these voxels. In the iron-rich globus pallidus (Fig. 1c), the T2 of the voxel is shortened, thereby changing the linewidth of the direct saturation spectra (Smith et al., 2009), so that the curve is not only shifted, but also broadened.
In direct saturation imaging, any B1 inhomogeneities affect only the linewidth (and thus the depth) of the Z-spectrum, not the lineshape or resonance frequency. The WASSR method should therefore provide robust measurements of raw resonance frequency even in the presence of B1 inhomogeneities.
Methods
Simulations
A Monte Carlo simulation was used to estimate the optimal parameters for WASSR imaging and to compare the robustness of finding the center frequency using a Lorentzian fit versus using the Maximum Symmetry Algorithm. The DS signal was simulated using Eqn. 3 (Mulkern and Williams, 1993), with added Rician noise to gauge the effect of SNR on the fitting. Variables included: the SNR of the volume acquired without a saturation prepulse, SNR(S0); the sweepwidth (range) across which the RF offset frequencies were prescribed; and the number of RF offset frequencies prescribed within the sweepwidth. The linewidth, or full-width half-maximum, of the normalized direct saturation spectrum (S(ωRF)/S0) plotted as a function of RF offset frequency can be calculated as (Kim et al., 2009):
[5] |
For gray matter at 3T, R1 and R2 are approximately 0.88Hz and 13.5Hz (Lu et al., 2005). We used B1 = 0.2μT so that ω1 = 53.5rad/sec. For these simulations, LW = 67.2Hz.
A set multiplier (between 1 and 5) was used to generate the sweepwidth based on this linewidth. For example, for a SW/LW ratio of 2, the sweepwidth would be 134Hz, with endpoints of +/− 67Hz. For each sweepwidth to linewidth ratio, a set number of RF offset frequencies (ranging from 5 to 61 offset frequencies) was equally spaced over the sweepwidth; this is needed because the field varies over a sample and the position of the DS curve differs between voxels. Experimentally, when acquiring a direct saturation dataset for five subjects and creating a histogram of the resonance frequency shifts, we found that the range of resonance frequency shifts across the brain had a standard deviation of approximately 21.6Hz. To simulate a resonance frequency shift, we therefore modeled the resonance frequencies as a zero-mean normal distribution, with a standard deviation equal to 21.6Hz.
We simulated the true direct saturation signal at infinite SNR, with the resonance frequency shifted as described above as well as at different SNR levels. We subsequently fit the signal to a Lorentzian lineshape, gauging the error for determining these resonance frequencies as a function of SNR and the number of offset frequencies over a range of sweepwidth to linewidth ratios. The mean absolute error was calculated from the simulated resonance frequency minus the fitted resonance frequency. This WASSR resonance frequency error was computed as a function of the number of offset frequencies for SNR values of 10:1, 20:1, 30:1, 40:1, 80:1, 100:1, 1000:1, and infinite over 1000 Monte Carlo iterations. All processing was performed with in-house software written with Java and MATLAB (The MathWorks, Natick, MA, USA).
Image Acquisition
To optimize image acquisition, direct saturation spectra were first acquired using several saturation pre-pulse powers and durations. While there is no strict rule, the general requirements are 1) sufficiently low power to avoid MT and CEST effects as much as possible, since they may shift the average water frequency; 2) sufficiently high power to generate at least 50% water saturation in each voxel to allow accurate Lorentzian fitting; 3) sufficiently long saturation pulse for it to be selective enough for minimizing the curve width; 4) sufficiently short saturation pulse for the sequence to be fast enough to be clinically viable. Because a short saturation pulse has an increased bandwidth and decreased selectivity of irradiation, we chose the shortest pulse at the lowest power necessary for selective saturation, in which the direct saturation spectra demonstrated the least saturation broadening around the resonance frequency. In our experimental setup at 3T with a body coil, proper saturation was achieved with a 0.2μT sinc-gauss pre-pulse applied for 70ms at 3T.
After IRB approval and written informed consent, five healthy male volunteers (aged 30 to 32 years old) were studied at 3T (Philips Healthcare, Best, Netherlands) using body-coil excitation and 32-channel head coil receive. A T1-weighted 3D MPRAGE scan was acquired for structural referencing (Turbo factor = 192, shot interval = 3000ms, SENSE = 2 × 1 × 2, TE = 3.5ms, TR = 7.5ms, α = 8°) with an acquired isotropic resolution of 1.1mm3 (FOV = 212mm × 212mm × 165mm, with a matrix reconstruction of 512). For all WASSR and GRE scans, the nominal resolution was 1.2mm isotropic, covering the entire brain (100 slices, FOV = 220mm × 220mm × 120mm, reconstructed resolution = 0.98mm × 0.98mm × 1.2mm). Fat suppression was accomplished using a water-selective ProSet 121 excitation pulse and one 60mm REST slab positioned inferior to the acquired volume. Phase images were acquired using a 3D ten-echo gradient-recalled echo sequence (SENSE = 2×1×2, TR = 70ms, TE1 = 6ms, ΔTE = 6ms, α = 20°, Scan Duration = 7:46min). For WASSR, each volume for the direct saturation spectra was acquired with a 3D GRE multi-shot EPI readout (EPI factor = 33, SENSE = 2×1×2, TR = 150ms, TE = 22ms, α = 20°, volume acquisition time = 24s) preceded by a saturation prepulse. The overall scan time depended on the number of offset frequencies utilized to sample the spectrum.
To provide a gold standard for WASSR, a highly sampled direct saturation dataset (57 RF offset frequencies sampled every 5Hz) was acquired over a large sweepwidth (−140Hz to 140Hz) to encompass the range of frequencies across a brain at 3T. One volume with no saturation prepulse and one volume with a prepulse at an RF offset of 80kHz were also acquired, leading to a total scan duration of 20:41minutes. A subset of these data was then fit to the Lorentzian lineshape to determine the minimum number of offsets necessary for generating a resonance frequency map of sufficient quality. Customized MATLAB software utilized a least squares curve-fitting algorithm written in Java for increased processing speed to calculate the resonance frequency maps, as well as the goodness-of-fit for fitting the Lorentzian lineshape.
With a saturation pre-pulse of 70ms at 0.2μT, the linewidth of the direct saturation spectra over the brain ranged from approximately 20 to 50Hz. The range of offset frequencies fell primarily between +/−80Hz. Therefore, for each subject, we applied the saturation pulse across a sweepwidth of +/−120Hz with respect to the scanner reference frequency, with a close sampling interval of 5Hz between +/−80Hz and larger sampling interval of 10Hz between 80Hz and 120Hz, in order to gather detail within the range of tissue frequencies and also encompass the full range of frequencies across the brain, respectively. Combined with the no-saturation and 80kHz offset volumes, this scan could be completed in a total time of 8:25min.
Image Analysis
The SNR within a region of interest of experimental data was calculated by acquiring two images with identical parameters, Image A and B, and using (Firbank et al., 1999):
[6] |
Frequency and Susceptibility Calculations
The data processing pipeline is illustrated by the flowchart in Figure 3. All processing was performed using custom-written scripts in MATLAB. It can be seen that the generation of the initial frequency maps is different between the two methods, as required by the different acquisition approaches, while the remainder of the processing is the same. However, because the phase-unwrapping procedure (Laplacian-based phase unwrapping) for the GRE QSM approach removes some of the large background gradients, the removal of these gradients is more demanding for WASSR processing.
Frequency mapping using gradient-recalled echo phase images was accomplished by first processing the phase signal from each echo with a Laplacian-based phase unwrapping method (Li et al., 2011), followed by linear fitting of the phase signal as a function of echo time for the first eight echoes (Eqn. 1) and calculation of the frequency map using Eqn. 2. The linear fitting of phase over time also gives an estimation of the initial phase (ϕoffset), which can be used as a threshold to exclude some voxels with unreliable phase measurement (Schweser et al., 2011b). Our threshold was π/8, meaning that voxels with an intercept larger than this threshold were considered unreliable, possibly due to turbulent flow, partial volume effects, or an unusually high frequency shift.
Frequency mapping using WASSR direct saturation spectra consisted of three steps, namely (i) correction for the field drift; (ii) coregistration of the image volumes for each irradiation frequency; (iii) fitting the direct saturation curves to a Lorentzian lineshape.
The field drift was calculated using the phase images acquired for the direct saturation volume at each RF frequency. Because all of these images were acquired with the same TE and TR, the theoretical phase signal in each image without field drift should be identical; therefore, any variations between the average phase of these volumes can be attributed to the field drifting. The slices with the most homogeneous phase were used for this correction. Slices near the top of the brain, with few phase wraps, were processed to select the largest continuous phase region over these slices across the duration of the direct saturation spectrum acquisition. The mean and standard deviation of the phase within this region were calculated as a function of dynamic scan time and fit to a fifth order polynomial function. The phase difference with respect to the first acquired image was converted to a frequency difference and subtracted from the prescribed offset frequency used to acquire that particular volume. No phase unwrapping was required.
After field drift compensation, the magnitude signal of each direct saturation volume was coregistered to the volume acquired with no saturation pre-pulse, using the Oxford FMRIb Software Library (FSL) Linear Image Registration Tool (FLIRT) (Jenkinson and Smith, 2001; Smith et al., 2004; Woolrich et al., 2009) with rigid body registration and a normalized mutual information algorithm. The signal per voxel (Fig. 2a) was then fit to a Lorentzian lineshape (Eqn. 4) as a function of the field-drift compensated RF offset frequency of the saturation prepulse (Fig. 2b–c) using a least-squares curve fit. Output parameters from this fit were maps of the amplitude of the initial signal (Fig. 2d), any residual signal from incomplete saturation (Fig. 2e), the linewidth at full-width-half-maximum (Fig. 2f), and the resonance frequency shift with respect to the scanner reference frequency (Fig. 2g).
To assess the quality of the Lorentzian fit, we calculated a map depicting the squared 2-norm of the difference between the experimental points and the fitted points per voxel across N offset frequencies:
[7] |
This residual norm map was combined with maps of the initial signal and residual signal from incomplete saturation to create a “goodness of fit” mask, indicating how well the Lorentzian lineshape fit the experimental data. For regions within the brain, we calculated a goodness of fit factor:
[8] |
with the square root of the residual norm map resnorm divided by the number of offset frequency points N, and A and C as defined with Eqn. 4 (Fig. 2). Because the signal remaining after saturation should generally be less than the initial signal without saturation, any negative differences in this “signal difference map” were set to very low values, using the “eps” function in MATLAB, which sets the selected voxels to the smallest values within a floating point precision of the software, approximately 2−52. The residual norm map divided by the signal difference map produced very high values in voxels with unreliable resonance frequency measurements, which could be due to incomplete saturation, noise, blood flow, or motion. The “goodness of fit” mask was created using all voxels having less than a ratio of 0.1.
To experimentally examine the effect of the number of collected points on the quality of fitting the direct saturation lineshape, we compared the highly-sampled dataset (57 offset frequencies) with datasets based on varying numbers of offset frequencies.
Susceptibility Calculation from the Field Map
The raw resonance frequency maps contain slowly-varying components from background gradient inhomogeneities that result in large field changes, as well as smaller local field changes from inhomogeneities due to the magnetic susceptibility of tissues. We used the dipole-fitting method (de Rochefort et al., 2010; Wharton and Bowtell, 2010) to model the background gradient field, first stripping the skull and dura and masking the brain, then solving a minimization problem to determine the susceptibility sources outside of the brain that would generate a background gradient field similar to that measured experimentally:
[9] |
where FT is the Fourier Transform operation and FT−1 is the inverse Fourier Transform operation, χ is the magnetic susceptibility distribution, is the unit vector in the direction of the main magnetic field in the subject frame, k̄subis the spatial frequency vector in the subject frame, and is the relative change in the magnetic field along the direction of the main magnetic field. W is a weighting matrix, chosen for the WASSR images to be the non-saturated volume from direct saturation imaging (TE = 22ms) multiplied by the goodness of fit mask. For the GRE images, W was chosen to be the corresponding fourth echo GRE magnitude image (acquired at TE = 24ms). M is a brain mask and β is a regularization parameter.
The mask of the brain region was created using the FSL Brain Extraction Tool (BET) (Smith, 2002; Smith et al., 2004; Woolrich et al., 2009), with areas outside the brain set at a value of zero, and areas inside the brain set at a value of one. This mask was then eroded by a disk element with a radius of three pixels, in order to eliminate extraneous signal from cerebrospinal fluid (CSF) surrounding the brain. The regularization parameter β was set to 1000, which makes the fitted susceptibility source inside the brain very small (less than 1 * 10−6 ppm), thereby emphasizing the susceptibility sources outside the brain as major contributions to field fluctuations. An iterative conjugate gradient-based solver was developed in MATLAB for solving the minimization problem, resulting in a map of the global magnetic field shift distribution from the fitted susceptibility sources. This global map was subtracted from the original resonance frequency map, which produced a map of the residual magnetic field shift.
The conversion from frequency to magnetic susceptibility is an ill-posed inverse problem (Wharton and Bowtell, 2010). We used the LSQR method (Li et al., 2011; Liu et al., 2009; Paige and Saunders, 1982; Wharton and Bowtell, 2010) to determine the magnetic susceptibilities. To calibrate the calculation for studies where we only have frequency maps for a single head orientation, we acquired a multi-orientation GRE and a multi-orientation WASSR dataset for one subject, then used the COSMOS method to find the solution χ to the following minimization problem (Liu et al., 2009; Wharton and Bowtell, 2010):
[10] |
where
[11] |
with ΔBn as the field shift produced by a susceptibility distribution (χ) in the nth orientation across a total of N orientations, χ̃ indicating the three-dimensional Fourier Transform of the magnetic susceptibility, Cn being the Fourier transform of the dipole convolution kernel linking susceptibility and field for that nth orientation from Eqn. 10, α as a regularization parameter set to 20, Mout as a mask that is zero inside the brain, and Wn as the chosen weighting matrix for the nth orientation. With α set to 20, the resulting susceptibility outside the brain is fairly small (less than 1 * 10−5ppm), thereby emphasizing susceptibility contributions from within the brain.
For a single orientation (N = 1), a relative residual threshold was set as the stopping criteria for the developed LSQR solver for the minimization problem in Eqn. 10. The weighting matrices for WASSR and GRE images were generated as described above for the multi-orientation data set.
Region of Interest Analysis
We parcellated our datasets into over sixty regions of interest through automated coregistration and segmentation with the Eve atlas using MRI Studio (https://www.mristudio.org) from the National Research Resource for Quantitative Functional MRI (http://www.mri-resource.kennedykrieger.org/software) (Jiang et al., 2006; Lim et al., 2013; Oishi et al., 2009). The Eve atlas from Johns Hopkins University is a single-subject female brain at 1mm3 isotropic resolution, put in standard Montreal Neurological Institute (MNI) coordinates (Mori et al., 2009; Mori et al., 2008; Oishi et al., 2009). Here, the Eve atlas templates used for coregistration were a skull-stripped GRE magnitude image at TE = 24ms (“EveGreMag”) and a QSM image (“EveQSM”), downloadable from the MRI Studio website as “JHU_MNI_SS_GreMag_ss” and “JHU_MNI_SS_QSM_ss,” with coregistration described in the methods section of (Lim et al., 2013). To coregister the outer surface of the brain, the skull-stripped magnitude image from the fourth echo of the GRE (TE = 24ms) from each subject was coregistered to the EveGreMag using Automated Image Registration (AIR) (Woods et al., 1998a; Woods et al., 1998b), with affine rigid body registration and trilinear interpolation. The resulting AIR transformation matrix was applied to the subject QSM. To facilitate processing, the images were transformed from float to byte format and intensity-corrected with automated histogram matching. To coregister the internal brain structures, the GRE magnitude image and QSM from each subject were coregistered using dual-channel Large Deformation Diffeomorphic Metric Mapping (Beg et al., 2005; Miller et al., 2002) to the EveGreMag and EveQSM. For the WASSR datasets, the skull-stripped non-saturated volume of the WASSR dataset (TE = 22ms) and the QSM from WASSR were coregistered using the above process to the EveGreMag and EveQSM, respectively.
After coregistration, the inverse transformation matrices from AIR and LDDMM were applied to the “Everything” Parcellation Map with deep gray matter ROIs delineated from QSM images and deep white matter ROIs from DTI FA images (downloadable as “JHU_MNI_SS_EvePM_V1.0”), transforming these brain regions into subject coordinates. These brain regions include gyral areas that contain both cortical gray matter and white matter tracts, which are difficult to align exactly per volunteer due to individual gyral variation. For the purposes of this paper, we included only deep gray matter and deep white matter ROIs in our analyses.
During the automated prescan, the scanner reference frequency (ωref) is placed at the frequency of the maximum water signal, which may be different for each scan depending on shimming and subject position in the coil. Therefore, a susceptibility reference standard is needed when quantifying susceptibilities. For each subject, deep white matter regions from the transformed EvePM were selected as a reference ROI in order to analyze all subjects on the same reference scale. The calculated susceptibility maps were then shifted so that the mean magnetic susceptibility value of white matter was −0.03ppm (Li et al., 2012b; Lim et al., 2013). This value was chosen because it led to a mean susceptibility of approximately 0ppm for cerebrospinal fluid (CSF), which is a common reference region used in QSM when assuming an isotropic susceptibility. Using the EvePM, we were able to measure the average susceptibility in various regions containing CSF, such as the frontal portion of the lateral ventricle and the body of the lateral ventricle, and saw a wide range of average values within these ROIs. Unlike the ventricles that may have partial volume effects with surrounding tissue, large white matter regions are more straightforward to select, and the mean magnetic susceptibility values within these white matter regions have smaller spatial variations (Li et al., 2012b; Lim et al., 2013; Wharton and Bowtell, 2010). We therefore used this indirect method to allocate a reference susceptibility of 0ppm to the CSF.
The transformed EvePM was applied to each subject’s referenced QSM, and an average susceptibility was determined for each region of interest. Values equal to exactly zero were excluded from the mean susceptibility calculation. The automated coregistration and segmentation process is described in detail in a recent paper (Lim et al., 2013).
To compare the agreement between the two methods for deep white matter and deep gray matter ROIs, we plotted the average susceptibilities for GRE vs. the average susceptibilities for WASSR alongside an identity line of y = x. We also created a Bland-Altman plot, with the average of the susceptibility values for GRE and WASSR plotted against the difference of the susceptibility values for GRE and WASSR for each of the regions of interest (Bland and Altman, 1986).
Results
Simulations
Monte Carlo simulations were used to compare the robustness of finding the resonance frequency with a Lorentzian fit or the Maximum Symmetry Algorithm. Figure 4a shows the mean absolute error in the resonance frequency as a function of the number of applied RF offset frequencies (N) for the optimal sweepwidth to linewidth ratio at different SNR levels of the non-saturated volume. As expected, using a higher number of points results in reduced error. Fitting with the Lorentzian lineshape consistently exhibited less error than with the Maximum Symmetry Algorithm. In the situation without noise, the mean absolute error for the Lorentzian fitting was about 10−8Hz (data smaller than axis limits of Fig. 4a), essentially the precision level of the software, whereas the Maximum Symmetry Algorithm reached an error limit of 10−6Hz.
Previous work on WASSR field mapping using the Maximum Symmetry Algorithm at 3T recommended acquiring 16 to 33 points (Kim et al., 2009). However, as seen in Figure 4a, with Lorentzian fitting (Liu et al., 2010b), a sub-Hertz error could be achieved using as few as nine offset frequencies with an SNR as low as 20:1. Experimentally, the SNR of the brain region for the non-saturated volume was determined to be 20 to 40 per slice, depending on the anatomical level of the slice. The pre-determined Lorentzian shape can be fitted quite well as long as there are some points without saturation and at least a few in the saturation dip. Figure 4b shows that using a sweepwidth to linewidth ratio greater than two does not improve the quality of the frequency determination but in fact reduces it slightly. As the accuracy reduces steeply below a ratio of two and linewidths vary, it is best to choose a conservative ratio between 2 and 4. That said, the shift range is also a factor, which generally requires one to increase the sweepwidth above optimal.
Determination of the Number of Points for WASSR Acquisition
Even though the simulations indicate that WASSR could measure frequency shifts with as few as nine data points for a sub-Hz error, the actual number of points (image volumes with different frequency offset) that is needed of course depends on the error required for determining accurate magnetic susceptibility differences from frequency differences on the order of a few Hz or less. The frequency shifts across a large organ like the brain fall across a fairly large range, which we experimentally determined to be approximately +/− 1ppm, or +/−128Hz at 3T. For these subjects, the linewidth (that is, the Full-Width Half Maximum of the fitted Lorentzian lineshape) generally ranged from 20Hz to 65Hz across the brain.
A 57-point direct saturation dataset was acquired with a saturation pre-pulse applied at an RF offset frequency every 5Hz between +/−140Hz. The SNR of the non-saturated signal ranged from 20 to 40, depending on the slice and structure in which the SNR was measured. Figure 5 shows a comparison between frequency maps calculated as a function of frequency resolution, ranging from 57-point 5Hz spacing (Fig. 5a) to 5-point 70Hz spacing (Fig. 5h). The highly-sampled dataset provided easy fitting of the frequency, but the total scan duration of 24 minutes is clearly not optimal for clinical studies. The sparse dataset with five points was insufficient for determining resonance frequency accurately (Fig. 5f). Acquiring fewer points shows a different profile from the gold standard, as seen in a graph of the line profiles (Fig. 5g inset) through the same slice of the resulting frequency maps.
A histogram of resonance frequencies across the brain showed that the majority of the resonance frequencies within the brain occur between +/− 80Hz, with fewer frequencies around +/− 120Hz located closer to the skull and CSF. Therefore, we decided to use 19 points (image volumes) spaced at intervals of 10Hz within +/− 80Hz (for finer detail within the brain, in accordance with our simulations) and then 20Hz between 80Hz and 120Hz (to capture the areas more affected by background gradients on the outer edges of the brain). This provided data comparable to the gold standard (Figs. 5e, g), while adhering to a clinically viable scan duration of approximately eight minutes.
This 19-volume non-linear scan does propagate more noise from regions with lower SNR (e.g., regions that contain flowing liquids like the blood and CSF with longer T1, which are less prone to complete saturation), but line profiles through resonance frequency maps from the 57-volume scan, 29-volume scan, and the 19-volume scan showed very comparable behavior (Fig. 5g, inset).
Field Drift Correction
Rapid GRE scanning often heats the magnet bore components, especially the passive shims. As a result, the magnet frequency may drift several Hertz over the total scan time, with the amount of drift depending on scan history. Figure 6 illustrates how a correction for this field drift can be obtained using WASSR. If the WASSR (or other GRE-based acquisition) experiment is conducted at the beginning of the overall study starting from a cold scanner, the magnet bore heats up slowly and continuously during the scan, causing the field to drift up to 10Hz over ten minutes (Fig. 6a). If a gradient-intensive sequence (e.g., DTI) is performed before the series of GRE acquisitions in the WASSR sequence, however, the magnet bore is closer to thermal equilibrium, and the field drifts much less (Fig. 6b). Because we cannot correct for field drift in the normal 3D single-volume GRE datasets, we usually ran the GRE sequence used for quantitative susceptibility comparison after the WASSR sequence, so that the field would drift less during the GRE image acquisition. The field drift can be measured using the WASSR frequency as a function of time (Fig. 6a, b), the result of which can be used for correction of the data by fitting the trend with a polynomial and shifting the prescribed offset frequencies used in Lorenztian fitting.
Calibration of Magnetic Susceptibilities
The COSMOS method, which utilizes data collected at multiple subject orientations, has the best inverse conditions compared to other methods (Liu et al., 2009). However, this is physically inconvenient for the subject and time consuming; therefore, most groups use only single orientation acquisitions. We used a COSMOS acquisition of one subject as the gold standard to calibrate the thresholds for stopping the iterative inverse calculation in the single orientation method. An example of this is shown in Figure 7 for both multi-echo GRE and WASSR. The LSQR iteration stopping criteria for the single-orientation susceptibility calculations were varied across a range of maximum relative residual thresholds (0.30, 0.20, 0.16, 0.12, 0.10, 0.08, 0.06, 0.04), as visualized in Fig. 7b. Figure 8 shows the average susceptibility (a, b) and standard deviation (c, d) within selected regions of interest for (a, c) GRE and (b, d) WASSR. The GRE single-orientation susceptibility maps corresponding to these relative residual thresholds converged at 28, 48, 67, 103, 136, 192, 299, and 514 iterations, respectively. The WASSR single-orientation susceptibility maps corresponding to these relative residual thresholds converged at 24, 46, 65, 108, 147, 215, 343, and 593 iterations, respectively. We determined the optimum relative residual threshold as the most consistent average susceptibility per ROI with the lowest standard deviation, which was approximately 0.06 for GRE and 0.06 − 0.08 for WASSR. We decided that the following parameters would produce appropriate susceptibility values for single-orientation acquisitions: α = 10, the maximum number of iterations = 200, and the tolerance threshold for the minimization = 0.08. For both acquisitions, the calculations generally converge within 80 to 200 iterations.
Frequency and Susceptibility Maps
The multi-echo gradient echo and WASSR direct saturation images produced the frequency maps shown in Figure 9 and the susceptibility maps shown in Figure 10. These maps were coregistered to GRE magnitude and QSM images from the Eve atlas (Lim et al., 2013) in order to display line profiles (Figs. 9g–i, 10g–i) through the same slices at several levels in the brain, which illustrate similarities and differences between the two methods.
Generally, the frequency and susceptibility maps for both GRE and WASSR show comparable trends; that is, the iron-rich deep gray matter structures such as the red nucleus and globus pallidus appear to be brighter, or more paramagnetic, than the reference of CSF, whereas the white matter structures such as the internal capsule of the corticospinal tract and the thalamic radiations appear darker, or more diamagnetic, than CSF.
Regions of Interest Analysis to compare WASSR and GRE
Figure 11 shows the regions of interest delineated via automated segmentation of QSM maps in our EveQSM reference atlas (Lim et al., 2013) for (a) GRE and (b) WASSR. Figure 11c displays the average referenced susceptibility and standard deviations for several deep white matter and deep gray matter regions of interest across our five subjects, showing comparable measurements for GRE and WASSR within the standard deviation for each ROI. Generally, the deep white matter regions exhibit an average susceptibility that is more negative than the deep gray matter regions. The globus pallidus exhibited the highest average susceptibility for all five volunteers. As shown in Figure 11c, the average susceptibility in two CSF-filled regions, the frontal portion and the body of the lateral ventricle, was approximately −0.03 to 0.04 ppm for WASSR and −0.03 to 0.02ppm for GRE, respectively, across the five volunteers. This typical large variation prompted the combined usage of several deep white matter bundles as a reference region, as described in the Methods section.
Figure 12a depicts the correlation between the GRE and WASSR methods for the mean susceptibilities across the white and gray matter regions of interest (R2 = 0.992) in Fig. 11c, plotted alongside a line of identity (y = x). For this particular set of volunteers, the correlation between the average susceptibility calculated using GRE versus WASSR follows the equation:
[12] |
Figure 12b contains a Bland-Altman bias plot that shows the degree of agreement for the ROIs displayed in Figure 11c between multi-echo GRE and WASSR (Bland and Altman, 1986) by comparing the difference between the mean susceptibilities of the two methods with their average. The mean difference between susceptibility measurements from WASSR and GRE was −0.0035, with a 95% confidence interval of [−0.012, 0.0055] ppm. Because the confidence interval is narrow and includes zero, the zero bias likely is the real bias, indicating agreement between the two methods.
Discussion
Gradient-recalled echo imaging is a convenient pulse sequence for frequency mapping that is widely available on all human MRI scanners, but exhibits a few disadvantages that may affect the accuracy of susceptibility maps. Firstly, phase images may have phase wraps at the interfaces of structures with largely different susceptibilities, thereby complicating data interpretation around these structures. Faulty phase wrap removal results in areas of discontinuous phase signal, which propagates to discontinuous susceptibility maps. Many methods have been developed to remove these phase wraps, including high pass filtering (Haacke et al., 2009; Haacke et al., 2004), region-based unwrapping methods (Jenkinson, 2003; Rauscher et al., 2008; Witoszynskyj et al., 2009), and Laplacian-based phase unwrapping (Li et al., 2011; Schofield and Zhu, 2003). The method of phase unwrapping may affect the resonance frequency map; for example, when using a high-pass filter to remove background gradients, the size of the filter and number of iterations applied will affect the final contrast in the resulting frequency map. Laplacian-based phase unwrapping, the method utilized here in our GRE method, removes part of the background gradients before dipole fitting, which results in a blurred resonance frequency map.
Secondly, phase contrast relies on proper choice of echo times, resulting in images with longer overall imaging times at lower fields; for example, commonly utilized TEs are approximately 40ms at 1.5T (Schweser et al., 2010) and 3T (Li et al., 2011), and about 15ms at 7T (Wharton and Bowtell, 2010). Phase images acquired at longer echo times will exhibit more phase wraps across the acquired volume, with maximum SNR at TE = T2* (Wu et al., 2012). Fast-imaging readouts, such as EPI, utilize longer echo times, which result in more phase wraps; therefore, phase imaging with GRE usually does not use EPI.
Thirdly, when acquiring the entire 3D GRE volume at high resolution, the subject is more likely to move, leading to image blurring. Examples from the literature show that, at 3T, the time needed to acquire a four-echo GRE at 2mm isotropic resolution was 26 minutes (Yao et al., 2009) and to acquire a dual-echo GRE at 0.6mm isotropic resolution was 19 minutes (Schweser et al., 2011b). Here, we utilized a 3D eight-echo GRE at 1.2mm isotropic resolution with a scan duration of approximately 8 minutes, which is still a long period of time for a subject to lie in the scanner without moving.
Fourthly, in conventional GRE, B1 inhomogeneities from RF transmission may affect the measured susceptibility (Schweser et al., 2011a). Fifthly, rapid GRE scanning often heats the magnet bore components, thereby changing the magnet frequency according to Curie’s Law, and the magnet frequency may drift over several Hertz during the acquisition.
Our results show that WASSR provides a comparable alternative, with several advantages over GRE imaging for QSM mapping. The WASSR method relies primarily on the signal magnitude, and no phase unwrapping is necessary. Because the generation of the resonance frequency map depends mainly on the direct saturation prepulse, any readout may be utilized; therefore, volumes may be acquired with a short TE (for higher SNR) and an EPI acquisition (approximately 20–30 seconds to acquire a whole-brain volume at 3T). A higher EPI factor may cause slight geometric distortions, but allows faster acquisition of each volume, which also decreases the likelihood of motion during acquisition. Even though structures may not correspond exactly between WASSR with EPI and 3D-GRE, which does not utilize EPI, the susceptibility measurements from WASSR were well reproducible and quite comparable to those from GRE. As shown by the Bland-Altman plot in Figure 12b, the susceptibility measurements using each of the two methods correlate well across regions of interest.
Acquiring separate volumes requires coregistration between the volumes, but also allows for quality control; if a subject moves while one volume is acquired at a particular RF offset frequency, that volume may be discarded without drastically affecting the entire experiment, because other volumes may still be fit to a Lorentzian lineshape. Therefore, the WASSR method is especially useful for acquiring susceptibility maps in subjects that are likely to move (e.g., children) or structures that are affected by involuntary motion. For example, acquiring direct saturation images and calculating frequency maps with WASSR has been used to calculate susceptibility maps in the cervical spinal cord (Lim et al., 2012a).
For our WASSR images at an SNR of 20 to 40 at 3T, while covering the entire brain from +/− 120Hz, we observed contributions from brain tissues within +/− 80Hz. Therefore, using simulations where the theoretical signal had a resonance frequency randomly selected from a distribution that had a mean of 0 and a standard deviation of 21.6Hz, we found that Lorentzian fitting had an error on the order of sub-Hertz even for low SNR conditions (Fig. 3b). The optimal sweepwidth to linewidth ratio was determined based on the number of volumes acquired; for 19 points, the optimal SW/LW was approximately 2 (Fig. 4b). A very low sweepwidth of acquired offset frequencies compared to the experimental linewidth (SW/LW ratio < 2) effectively undersamples the data, resulting in a high error for the Lorentzian fit. On the other hand, if the sweepwidth is too large (SW/LW > 4), the sampled points cannot adequately characterize the curve, also resulting in a slightly higher error. Experimentally, the sweepwidth should therefore encompass the distribution of frequency shifts across the brain, approximately +/− 1ppm from the scanner reference frequency. For subjects or structures across which the resonance frequency varies significantly, we recommend a longer scan to achieve higher offset frequency sampling across a larger sweepwidth.
With the WASSR method, any inhomogeneities from B1 transmission affect only the linewidth of the direct saturation lineshape and not the resonance frequency per voxel; that is, B1 inhomogeneities affect the breadth or width of the Lorentzian lineshape, but the Z-spectrum in each voxel will still be symmetric about and have a minimum signal occurring at the resonance frequency of that voxel. Because the calculated susceptibility is based on this resonance frequency, the WASSR method is expected to provide robust measurements of raw resonance frequency even in the presence of B1 inhomogeneities.
Using the WASSR method, the Z-spectrum of direct saturation images consists of a series of 3D brain volumes, each of which is acquired in less than half a minute with the same TE and TR. Even though the signal magnitude changes as a function of RF offset frequency, the signal phase depends mainly on the timing parameters (e.g., TE), and therefore should be the same for each volume. Any changes in this average phase with respect to the first volume correlate with changes in the main magnetic field. By acquiring a phase image for each 30-second WASSR volume, a frequency correction based on the individual volumes could be applied to determine field drift between acquired volumes.
The resonance frequency maps from GRE and WASSR do appear somewhat different. This could be due to either technical or mechanistic differences. From a technical point of view, the Laplacian-based phase unwrapping in the 3D-GRE removes a portion of the background gradient effects, resulting in a blurrier GRE resonance frequency map before background removal. For the WASSR method, background gradient removal for the resonance frequency maps relies solely on dipole fitting, which is more calculation intensive and may be less able to show small differences. Also, the GRE images did not utilize an EPI readout, leading to fewer geometric distortions but a longer acquisition time per volume. That said, EPI could theoretically be used to speed up the GRE acquisition, especially when using one TE for the phase information. The application of motion correction between the direct saturation images acquired at different frequencies may lead to reduced sharpness in the contrast of the WASSR images. Possible ways to overcome these limitations would be to restrain the subject’s motion further and to acquire the images with a lower EPI factor or other acquisition method. (Also see Supplementary Material.) The WASSR resonance frequency maps contained more extreme values, mainly due to differences in CSF saturation between acquired volumes, possible coregistration errors, geometric distortions from EPI, and larger contributions from background gradients that were not removed with preprocessing; however, these factors all contribute to a low “goodness of fit” factor when assessing how well the Lorentzian fit matched the experimental data. Therefore, any voxels with a low “goodness of fit” metric were given low weighting values in subsequent calculations, allowing the determination of meaningful corrected frequency and susceptibility maps. We used 0.08 as the relative residual threshold in susceptibility calculations for both GRE and WASSR in order to provide a comparable standard for the resulting quantitative susceptibility maps. An advantage of multi-echo GRE imaging is the ability to calculate an R2* map. The WASSR method allows for calculation of a linewidth map, which incorporates R1, R2, and B1. R2* maps (Haacke et al., 2005; Jensen et al., 2009) and saturation linewidth maps (Smith et al., 2009) have both been used to show iron distribution in the brain.
Interestingly, one feature that seems to differ between the methods is the appearance of some of the fiber bundles on the susceptibility images, for instance the optic radiations in Figures 9 and 10. In the future, we intend to perform higher resolution combined GRE and WASSR studies at 7T to see whether these differences can be used to study the mechanisms of contrast in susceptibility images. We hypothesize that these differences may be due to selection of different compartments (Sati et al., 2013) in the two methods, which provide an opportunity to learn more about the system. The WASSR approach is likely to measure different compartments from GRE for several reasons. First, with our current GRE approach, we calculated the resonance frequency maps as the average slope across eight echo times, ranging from TE1 = 6ms to TE8 = 48ms. The WASSR images were each acquired with a single TE of 24ms. Recent publications have suggested a difference in the frequency maps generated at different echo times, which could also contribute to the differences between these two images (Wharton and Bowtell, 2012). Using different echo times will affect the pools measured. At the longer TE used in the WASSR method, myelin water no longer contributes much to the water signal. Second, the water is saturated at a single frequency, after which this saturation can be transferred to other pools through multiple mechanisms of magnetization transfer, such as dipolar coupling (Henkelman et al., 2001; Wolff and Balaban, 1994), chemical exchange (van Zijl and Yadav, 2011), and physically through diffusion. Using a typical water diffusion constant in brain tissue of 1×10−9 m2/s, this can be as far as over a range of 29 μm. Of course, saturation is not instantaneous but follows T1, so the situation is more complex. The effect of magnetization exchange on susceptibility mapping (Luo et al., 2010; Shmueli et al., 2011; Zhong et al., 2008) is still not fully elucidated. More work is needed to extract the details of these mechanisms, but comparing these two methods provides an opportunity to study the complex biophysical system in more detail.
The accuracy of susceptibility maps obtained using the WASSR method depends mainly on sampling rate; more volumes acquired at closely-spaced RF offset frequencies will result in a better Lorentzian fit and a more clearly-defined susceptibility map. GRE is more conducive at higher magnetic fields, at which a shorter TE induces larger frequency shifts and therefore more contrast between gray and white matter structures. However, at lower fields with lower frequency shifts, GRE is limited by spatial SNR, longer echo times required for phase contrast, and resulting longer scan times. In WASSR, the sweepwidth needed to cover the distribution of frequency shifts across the brain is proportional to the field in Hz, but the same in ppm, allowing the same number of RF offset frequencies to be acquired at each field, and resulting in susceptibility maps that will be comparable between varying field strengths. The WASSR method may be most useful for centers that do not have access to high-field magnet systems.
Conclusions
The WASSR method utilizes direct saturation images to identify susceptibility shifts. The overall scan time for WASSR depends on the number of volumes acquired; acquiring more volumes result in susceptibility maps with a higher SNR, but also make WASSR somewhat slower than GRE MRI at high fields. That said, because the individual volumes are acquired quickly and are not significantly hampered by B1 or B0 field inhomogeneities, WASSR should be especially advantageous for low-field imaging with subjects that are likely to move (e.g., children), or across areas with large susceptibility differences that would produce many phase wraps.
Supplementary Material
Highlights.
We derived resonance frequency maps using WAter Saturation Shift Referencing (WASSR).
With WASSR, we can correct for field drift and motion, then calculate QSM images.
Quantitative Susceptibility Maps (QSMs) were also calculated from GRE signal phase.
We used MRI Studio with the Eve atlas for coregistration and automated segmentation.
QSM images and average QS per ROI were well correlated between WASSR and GRE methods.
Acknowledgments
The authors thank Mr. Carlos Renjifo, Dr. Brian Caffo, Dr. Andreia Faria, Mr. Joseph S. Gillen, Ms. Terri Brawner, Ms. Kathleen Kahl, Ms. Ivana Kusevic, Mr. Alan Huang, Dr. James Pekar, Dr. Seth Smith, Dr. Peter Barker, Dr. Doris Lin, and Dr. Raj Stewart for assistance with data acquisition and processing. This research was supported by NIH resource grant P41 EB051909 (formerly P41 RR015241), with educational funding from the NeuroEngineering Training Grant NIH T32EB003383, and an NIH grant for interdisciplinary training in psychiatry and neuroscience, 5 T32 MH015330. The contents of this publication are solely the responsibility of the authors and do not necessarily represent the official view of the NIH. Dr. Peter van Zijl is a paid lecturer for Philips Healthcare and is the inventor of technology that is licensed to Philips. Dr. Craig Jones is paid in part through a grant to Kennedy Krieger Institute from Philips Healthcare. These arrangements have been approved by The Johns Hopkins University in accordance with its Conflict of Interest policies.
Abbreviations
- aIC
Anterior limb of the Internal Capsule
- Am
Amygdala
- bC
Body of the Corpus callosum
- bLV
Body of the Lateral Ventricle
- CC
Corpus Callosum
- CI
Confidence Interval
- CN
Caudate Nucleus
- CSF
Cerebrospinal Fluid
- COSMOS
Calculation of Susceptibility through Multiple Orientation Sampling
- DS
Direct Saturation
- EC
External Capsule
- EvePM
“Everything” Parcellation Map from the Eve atlas at Johns Hopkins University
- fLV
Frontal portion of the Lateral Ventricle
- gC
Genu of the Corpus callosum
- GM
Gray Matter
- GP
Globus Pallidus
- GRE
Gradient Recalled Echo
- Hp
Hippocampus
- IC
Internal Capsule
- LSQR
Algorithm for sparse linear equations and sparse least squares
- MTR
Magnetization Transfer Ratio
- pIC
Posterior limb of the Internal Capsule
- Pt
Putamen
- QSM
Quantitative Susceptibility Mapping
- RN
Red Nucleus
- ROI
Region of Interest
- sC
Splenium of the Corpus callosum
- SN
Substantia Nigra
- Th
Thalamus
- TR
Thalamic Radiations
- WASSR
WAter Saturation Shift Referencing
- WM
White Matter
Footnotes
Author Contributions: IAL, XL, PCMvZ designed the research; IAL performed the research; IAL, XL, CKJ, JADF, DSV contributed new analytical tools; IAL, XL analyzed data; and IAL, XL, PCMvZ wrote the paper. CKJ, JADF, DSV edited the paper.
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
References
- Beg MF, Miller MI, Trouve A, Younes L. Computing Large Deformation Metric Mapping via Geodesic Flows of Diffeomorphisms. International Journal of Computer Vision. 2005;61:139–157. [Google Scholar]
- Bland JM, Altman DG. Statistical Methods for Assessing Agreement between Two Methods of Clinical Measurement. The Lancet. 1986;327:307–310. [PubMed] [Google Scholar]
- de Rochefort L, Liu T, Kressler B, Liu J, Spincemaille P, Lebon V, Wu J, Wang Y. Quantitative susceptibility map reconstruction from MR phase data using bayesian regularization: validation and application to brain imaging. Magnetic Resonance in Medicine. 2010;63:194–206. doi: 10.1002/mrm.22187. [DOI] [PubMed] [Google Scholar]
- Dula AN, Asche EM, Landman BA, Welch EB, Pawate S, Sriram S, Gore JC, Smith SA. Development of chemical exchange saturation transfer at 7 T. Magnetic Resonance in Medicine. 2011;66:831–838. doi: 10.1002/mrm.22862. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dula AN, Dortch RD, Landman BA, Welch EB, Gore JC, Smith SA. CEST Imaging at 7 Tesla: Comparison of the WASSR and Higher Order Polynomial Fit to Determine Center Frequency. International Society for Magnetic Resonance in Medicine; Stockholm, Sweden: 2010. [Google Scholar]
- Duyn JH, van Gelderen P, Li TQ, de Zwart JA, Koretsky AP, Fukunaga M. High-field MRI of brain cortical substructure based on signal phase. Proceedings of the National Academy of Sciences. 2007;104:11796–11801. doi: 10.1073/pnas.0610821104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Firbank M, Coulthard A, Harrison R, Williams E. A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol. 1999;44:N261–N264. doi: 10.1088/0031-9155/44/12/403. [DOI] [PubMed] [Google Scholar]
- Haacke EM, Cheng NYC, House MJ, Liu Q, Neelavalli J, Ogg RJ, Khan A, Ayaz M, Kirsch W, Obenaus A. Imaging iron stores in the brain using magnetic resonance imaging. Magn Reson Imaging. 2005;23:1–25. doi: 10.1016/j.mri.2004.10.001. [DOI] [PubMed] [Google Scholar]
- Haacke EM, Mittal S, Wu Z, Neelavalli J, Cheng YCN. Susceptibility-Weighted Imaging: Technical Aspects and Clinical Applications, Part 1. American Journal of Neuroradiology. 2009;30:19–30. doi: 10.3174/ajnr.A1400. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Haacke EM, Tang J, Neelavalli J, Cheng YCN. Susceptibility mapping as a means to visualize veins and quantify oxygen saturation. Journal of Magnetic Resonance Imaging. 2010;32:663–676. doi: 10.1002/jmri.22276. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Haacke EM, Xu Y, Cheng YCN, Reichenbach JR. Susceptibility Weighted Imaging (SWI) Magnetic Resonance in Medicine. 2004;52:612–618. doi: 10.1002/mrm.20198. [DOI] [PubMed] [Google Scholar]
- Henkelman RM, Stanisz GJ, Graham SJ. Magnetization transfer in MRI: a review. NMR Biomed. 2001;14:57–64. doi: 10.1002/nbm.683. [DOI] [PubMed] [Google Scholar]
- Jenkinson M. Fast, automated, N-dimensional phase-unwrapping algorithm. Magnetic Resonance in Medicine. 2003;49:193–197. doi: 10.1002/mrm.10354. [DOI] [PubMed] [Google Scholar]
- Jenkinson M, Smith S. A global optimisation method for robust affine registration of brain images. Med Image Anal. 2001;5:143–156. doi: 10.1016/s1361-8415(01)00036-6. [DOI] [PubMed] [Google Scholar]
- Jensen JH, Szulc K, Hu C, Ramani A, Lu H, Xuan L, Falangola MF, Chandra R, Knopp EA, Schenck J, Zimmerman EA, Helpern JA. Magnetic field correlation as a measure of iron-generated magnetic field inhomogeneities in the brain. Magn Reson Med. 2009;61:481–485. doi: 10.1002/mrm.21823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jiang H, van Zijl PCM, Kim J, Pearlson GD, Mori S. DtiStudio: resource program for diffusion tensor computation and fiber bundle tracking. Computer methods and programs in biomedicine. 2006;81:106–116. doi: 10.1016/j.cmpb.2005.08.004. [DOI] [PubMed] [Google Scholar]
- Kim M, Gillen JS, Landman BA, Zhou J, van Zijl PCM. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magnetic Resonance in Medicine. 2009;61:1441–1450. doi: 10.1002/mrm.21873. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li W, Wu B, Avram AV, Liu C. Magnetic susceptibility anisotropy of human brain in vivo and its molecular underpinnings. Neuroimage. 2012a;59:2088–2097. doi: 10.1016/j.neuroimage.2011.10.038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li W, Wu B, Liu C. Quantitative susceptibility mapping of human brain reflects spatial variation in tissue composition. Neuroimage. 2011;55:1645–1656. doi: 10.1016/j.neuroimage.2010.11.088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li X, Vikram DS, Lim IAL, Jones CK, Farrell JAD, van Zijl PCM. Mapping magnetic susceptibility anisotropies of white matter in vivo in the human brain at 7T. Neuroimage. 2012b;62:314–330. doi: 10.1016/j.neuroimage.2012.04.042. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lim IAL, Choe AS, Li X, Jones CK, van Zijl PCM. Frequency Mapping in the Spinal Cord with WASSR at 3 Tesla. International Society for Magnetic Resonance in Medicine; Melbourne, Victoria, Australia: 2012a. [Google Scholar]
- Lim IAL, Faria AV, Li X, Hsu JTC, Airan RD, Mori S, van Zijl PC. Human brain atlas for automated region of interest selection in quantitative susceptibility mapping: Application to determine iron content in deep gray matter structures. Neuroimage. 2013;82:449–469. doi: 10.1016/j.neuroimage.2013.05.127. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lim IAL, Farrell JAD, Jones CK, Vikram DS, Renjifo CA, Li X, van Zijl PCM. Frequency Mapping without Phase Wraps. ISMRM; Montreal, Quebec, Canada: 2011a. p. 4531. [Google Scholar]
- Lim IAL, Jones CK, Vikram DS, Renjifo CA, Li X, van Zijl PCM. Frequency Mapping without Phase Wraps. ISMRM Ultra-High Field Systems & Applications Workshop; Lake Louise, Alberta, Canada. 2011b. [Google Scholar]
- Lim IAL, Li X, Jones CK, Farrell JAD, Vikram DS, Renjifo CA, van Zijl PCM. Correlation of Brain Iron with Susceptibility: Comparison of Gradient Echo and WASSR Acquisition at 3 Tesla. International Society for Magnetic Resonance in Medicine; Melbourne, Victoria, Australia. 2012b. [Google Scholar]
- Liu C. Susceptibility tensor imaging. Magnetic Resonance in Medicine. 2010;63:1471–1477. doi: 10.1002/mrm.22482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu C, Jiang Y, Johnson GA. 3D tracking of magnetic pathways in white matter based on magnetic susceptibility anisotropy. Proceedings of the International Society of Magnetic Resonance in Medicine.2010a. [Google Scholar]
- Liu C, Li W, Wu B, Jiang Y, Johnson GA. 3D fiber tractography with susceptibility tensor imaging. Neuroimage. 2012;59:1290–1298. doi: 10.1016/j.neuroimage.2011.07.096. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu G, Gilad AA, Bulte JWM, van Zijl PCM, McMahon MT. High-throughput screening of chemical exchange saturation transfer MR contrast agents. Contrast media & molecular imaging. 2010b;5:162–170. doi: 10.1002/cmmi.383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu T, Spincemaille P, de Rochefort L, Kressler B, Wang Y. Calculation of susceptibility through multiple orientation sampling (COSMOS): a method for conditioning the inverse problem from measured magnetic field map to susceptibility source image in MRI. Magnetic Resonance in Medicine. 2009;61:196–204. doi: 10.1002/mrm.21828. [DOI] [PubMed] [Google Scholar]
- Lu H, Nagae-Poetscher LM, Golay X, Lin D, Pomper M, van Zijl PCM. Routine clinical brain MRI sequences for use at 3.0 Tesla. Journal of Magnetic Resonance Imaging. 2005;22:13–22. doi: 10.1002/jmri.20356. [DOI] [PubMed] [Google Scholar]
- Luo J, He X, d’Avignon DA, Ackerman JJH, Yablonskiy DA. Protein-induced water 1H MR frequency shifts: Contributions from magnetic susceptibility and exchange effects. J Magn Reson B. 2010;202:102–108. doi: 10.1016/j.jmr.2009.10.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miller MI, Trouve A, Younes L. On the Metrics and Euler-Lagrange Equations of Computational Anatomy. Annual review of biomedical engineering. 2002;4:375–405. doi: 10.1146/annurev.bioeng.4.092101.125733. [DOI] [PubMed] [Google Scholar]
- Mori S, Oishi K, Faria AV. White matter atlases based on diffusion tensor imaging. Current opinion in neurology. 2009;22:362–369. doi: 10.1097/WCO.0b013e32832d954b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mori S, Oishi K, Jiang H, Jiang L, Li X, Akhter K, Hua K, Faria AV, Mahmood A, Woods R, Toga AW, Pike GB, Neto R, Evans A, Zhang J, Huang H, Miller MI, Zijl PV, Mazziotta J. Stereotaxic white matter atlas based on diffusion tensor imaging in an ICBM template. Brain. 2008;40:570–582. doi: 10.1016/j.neuroimage.2007.12.035. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mulkern RV, Williams ML. The general solution to the Bloch equation with constant rf and relaxation terms: Application to saturation and slice selection. Med Phys. 1993;20:5–13. doi: 10.1118/1.597063. [DOI] [PubMed] [Google Scholar]
- Oishi K, Faria A, Jiang H, Li X, Akhter K, Zhang J, Hsu JT, Miller MI, van Zijl PCM, Albert M, Lyketsos CG, Woods R, Toga AW, Pike GB, Rosa-Neto P, Evans A, Mazziotta J, Mori S. Atlas-based whole brain white matter analysis using large deformation diffeomorphic metric mapping: Application to normal elderly and Alzheimer’s disease participants. Neuroimage. 2009;46:486–499. doi: 10.1016/j.neuroimage.2009.01.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Paige CC, Saunders MA. LSQR: An Algorithm for Sparse Linear Equations and Sparse Least Squares. ACM Transactions on Mathematical Software. 1982;8:43–71. doi: 10.1145/2527267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rauscher A, Barth M, Herrmann KH, Witoszynskyj S, Deistung A, Reichenbach JR. Improved elimination of phase effects from background field inhomogeneities for susceptibility weighted imaging at high magnetic field strengths. Magn Reson Imaging. 2008;26:1145–1151. doi: 10.1016/j.mri.2008.01.029. [DOI] [PubMed] [Google Scholar]
- Sati P, van Gelderen P, Silva AC, Reich DS, Merkle H, de Zwart JA, Duyn JH. Micro-compartment specific T2 relaxation in the brain. Neuroimage. 2013;77:268–278. doi: 10.1016/j.neuroimage.2013.03.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schofield MA, Zhu Y. Fast phase unwrapping algorithm for interferometric applications. Optics Letters. 2003;28:1194–1196. doi: 10.1364/ol.28.001194. [DOI] [PubMed] [Google Scholar]
- Schweser F, Atterbury M, Deistung A, Lehr BW, Sommer K, Reichenbach JR. International Society for Magnetic Resonance in Medicine. 2011a. Harmonic phase subtraction methods are prone to B1 background components. [Google Scholar]
- Schweser F, Deistung A, Lehr BW, Reichenbach JR. Quantitative imaging of intrinsic magnetic tissue properties using MRI signal phase: an approach to in vivo brain iron metabolism? Neuroimage. 2011b;54:2789–2807. doi: 10.1016/j.neuroimage.2010.10.070. [DOI] [PubMed] [Google Scholar]
- Schweser F, Deistung A, Lehr BW, Reichenbach R., Jr Differentiation between diamagnetic and paramagnetic cerebral lesions based on magnetic susceptibility mapping. Med Phys. 2010;37:5165–5178. doi: 10.1118/1.3481505. [DOI] [PubMed] [Google Scholar]
- Sheth VR, Liu G, Li Y, Pagel MD. Improved pH measurements with a single PARACEST MRI contrast agent. Contrast media & molecular imaging. 2011;7:26–34. doi: 10.1002/cmmi.460. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shmueli K, de Zwart JA, van Gelderen P, Li TQ, Dodd SJ, Duyn JH. Magnetic susceptibility mapping of brain tissue in vivo using MRI phase data. Magnetic Resonance in Medicine. 2009;62:1510–1522. doi: 10.1002/mrm.22135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shmueli K, Dodd SJ, Li TQ, Duyn JH. The Contribution of Chemical Exchange to MRI Frequency Shifts in Brain Tissue. Magnetic Resonance in Medicine. 2011;65:35–43. doi: 10.1002/mrm.22604. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith SA, Bulte JWM, van Zijl PCM. Direct saturation MRI: theory and application to imaging brain iron. Magnetic Resonance in Medicine. 2009;62:384–393. doi: 10.1002/mrm.21980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith SM. Fast robust automated brain extraction. Hum Brain Mapp. 2002;17:143–155. doi: 10.1002/hbm.10062. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Smith SM, Jenkinson M, Woolrich MW, Beckmann CF, Behrens TEJ, Johansen-Berg H, Bannister PR, De Luca M, Drobnjak I, Flitney DE, Niazy RK, Saunders J, Vickers J, Zhang Y, De Stefano N, Brady JM, Matthews PM. Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage. 2004;23:S208–S219. doi: 10.1016/j.neuroimage.2004.07.051. [DOI] [PubMed] [Google Scholar]
- van Zijl PCM, Yadav NN. Chemical Exchange Saturation Transfer (CEST): What is in a Name and What Isn’t? Magnetic Resonance in Medicine. 2011;65:927–948. doi: 10.1002/mrm.22761. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wharton S, Bowtell R. Whole-brain susceptibility mapping at high field: a comparison of multiple- and single-orientation methods. Neuroimage. 2010;53:515–525. doi: 10.1016/j.neuroimage.2010.06.070. [DOI] [PubMed] [Google Scholar]
- Wharton S, Bowtell R. Fiber orientation-dependent white matter contrast in gradient echo MRI. PNAS. 2012;109:18559–18564. doi: 10.1073/pnas.1211075109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wharton S, Schafer A, Bowtell R. Susceptibility Mapping in the Human Brain Using Threshold-Based k-Space Division. Magnetic Resonance in Medicine. 2010;63:1292–1304. doi: 10.1002/mrm.22334. [DOI] [PubMed] [Google Scholar]
- Witoszynskyj S, Rauscher A, Reichenbach JR, Barth M. Phase unwrapping of MR images using Phi UN--a fast and robust region growing algorithm. Med Image Anal. 2009;13:257–268. doi: 10.1016/j.media.2008.10.004. [DOI] [PubMed] [Google Scholar]
- Wolff SD, Balaban RS. Magnetization Transfer Imaging: Practical Aspects and Clinical Applications. Radiology. 1994;192:593–599. doi: 10.1148/radiology.192.3.8058919. [DOI] [PubMed] [Google Scholar]
- Woods RP, Grafton ST, Holmes CJ, Cherry SR, Mazziotta JC. Automated Image Registration: I. General Methods and Intrasubject, Intramodality Validation. J Comput Assist Tomogr. 1998a;22:139–152. doi: 10.1097/00004728-199801000-00027. [DOI] [PubMed] [Google Scholar]
- Woods RP, Grafton ST, Watson JDG, Sicotte NL, Mazziotta JC. Automated Image Registration: II. Intersubject Validation of Linear and Nonlinear Models. J Comput Assist Tomogr. 1998b;22:153–165. doi: 10.1097/00004728-199801000-00028. [DOI] [PubMed] [Google Scholar]
- Woolrich MW, Jbabdi S, Patenaude B, Chappell M, Makni S, Behrens T, Beckmann C, Jenkinson M, Smith SM. Bayesian analysis of neuroimaging data in FSL. Neuroimage. 2009;45:S173–S186. doi: 10.1016/j.neuroimage.2008.10.055. [DOI] [PubMed] [Google Scholar]
- Wu B, Li W, Avram AV, Gho SM, Liu C. Fast and tissue-optimized mapping of magnetic susceptibility and T2* with multi-echo and multi-shot spirals. Neuroimage. 2012;59:297–305. doi: 10.1016/j.neuroimage.2011.07.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yao B, Li TQ, van Gelderen P, Shmueli K, de Zwart JA, Duyn JH. Susceptibility contrast in high field MRI of human brain as a function of tissue iron content. Neuroimage. 2009;44:1259–1266. doi: 10.1016/j.neuroimage.2008.10.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhong K, Leupold J, von Elverfeldt D, Speck O. The molecular basis for gray and white matter contrast in phase imaging. Neuroimage. 2008;40:1561–1566. doi: 10.1016/j.neuroimage.2008.01.061. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.