Published in final edited form as: Lab Chip. 2011 Jun 27;11(14):2319–2324. doi: 10.1039/c1lc20042a

Three-dimensional continuous particle focusing in a microfluidic channel via standing surface acoustic waves (SSAW)

Jinjie Shi a,b, Shahrzad Yazdi a,c, Sz-Chin Steven Lin a, Xiaoyun Ding a, I-Kao Chiang a, Kendra Sharp d, Tony Jun Huang a,*
PMCID: PMC3997299  NIHMSID: NIHMS338058  PMID: 21709881

Abstract

Three-dimensional (3D) continuous microparticle focusing has been achieved in a single-layer polydimethylsiloxane (PDMS) microfluidic channel using a standing surface acoustic wave (SSAW). The SSAW was generated by the interference of two identical surface acoustic waves (SAWs) created by two parallel interdigital transducers (IDTs) on a piezoelectric substrate with a microchannel precisely bonded between them. To understand the working principle of the SSAW-based 3D focusing and investigate the position of the focal point, we computed longitudinal waves, generated by the SAWs and radiated into the fluid media from opposite sides of the microchannel, and the resultant pressure and velocity fields due to the interference and reflection of the longitudinal waves. Simulation results predict the existence of a focusing point which is in good agreement with our experimental observations. Compared with other 3D focusing techniques, this method is non-invasive, robust, energy-efficient, easy to implement, and applicable to nearly all types of microparticles.

Introduction

The ability to focus particles and cells into a confined region is important for many lab-on-a-chip applications such as flow cytometry,15 single molecule detection,69 protein folding,10 cell sorting11,12 and enzymatic kinetics.1315 Focusing particles to a narrow stream, however, is not a trivial task. Due to the laminar nature of microfluidic flows, particles suspended in a fluid medium tend to follow the fluidic streamlines unless a lateral force moves them from their original paths. Hydrodynamic forces have been commonly used to manipulate streamlines to guide particles to a confined region.1619 Sheath flows with high flow rates are often used to compress collections of particles for two-dimensional (2D) hydrodynamic focusing in planar microfluidic channels. However, for many applications, 2D hydrodynamic focusing is intrinsically problematic due to the lack of vertical focusing. For example, when using 2D hydrodynamic focusing for flow cytometry, cell velocity variation in the vertical direction causes difficulties for cell sorting, and two or more cells may enter optical interrogation region at the same time.20 In hydrodynamic focusing-based laminar mixers, flow velocity variations of the focused enzymes or chemical species in the vertical direction may result in different reaction times across the depth of the channel, which would make it extremely difficult to extract meaningful information about the reaction kinetics.14 The inherent limitations of 2D hydrodynamic focusing call for the development of three-dimensional (3D) focusing, in which the stream of particles is focused both horizontally and vertically.

To date, a few microfabricated 3D hydrodynamic focusing devices have been reported.1623 However, in many of these 3D focusing devices the introduction of excessive sheath solutions causes sample dilution and dispersion, which is not desirable for many applications. Moreover, most of these devices require the complicated construction of multi-layer channels, which are difficult to fabricate using conventional soft lithographic techniques. Sheathless 3D hydrodynamic focusing can also be achieved using inertial microfluidics. This technique requires relatively high flow rates and long channels in order to focus the particles.24

Alternative methods that have been utilized to demonstrate particle focusing in microfluidic channels are electroosmosis,2534 dielectrophoresis (DEP),3337 magnetophoresis,3840 and acoustics.4144 The acoustic-based methods appear to be ideal for on-chip microparticle manipulation because they are non-invasive, applicable to virtually all microparticles, and do not require sheath solutions.12,4552 Standing acoustic waves have been used for particle focusing via implementation of piezoelectric transducers. Particles exposed to standing acoustic waves tend to move towards nodal points. Although standing bulk acoustic waves (SBAWs) based techniques have shown promising results,53,54 their non-planar structures and complex fabrication processes make it challenging to integrate them with other planar lab-on-a-chip components. Previously, we introduced a 2D particle focusing technique using standing surface acoustic waves (SSAWs).47 A PDMS microchannel was aligned and bonded between two parallel interdigital transducers (IDTs). Upon the application of an RF signal to the IDTs, a SSAW was generated and the suspended particles were focused to pressure nodes located, by design, in the middle of the microchannel width. Thus, 2D particle focusing was observed.47

In this work, through experimental and computational studies, we demonstrate that utilizing the same SSAW method, 3D particle focusing in a continuously flowing microchannel can be achieved. In our 3D focusing device, longitudinal acoustic waves were radiated into the fluid media from the propagating surface acoustic waves (SAWs). The interference between these longitudinal waves and their reflections from the channel walls formed a standing acoustic wave field in the microchannel with a pressure node in the middle. Once the particles move towards the pressure nodal plane (2D focusing), they tend to aggregate at the maximum kinetic energy points within that plane (focusing in the vertical direction). Hence, suspended particles are focused in both the horizontal and vertical directions. This acoustic force-based 3D focusing technique is non-invasive, energy-efficient, applicable to virtually any microparticle, and easy to fabricate and implement. Utilizing this method, microparticles suspended in liquid media can be focused to a predesigned location, regardless of particle properties. Manipulation of various particle sizes and different cell types has been discussed in our previous studies.12,49 Unlike hydrodynamic focusing techniques, this SSAW focusing method does not require fluid inputs with high flow rates and is free from the particle dispersion problem. The high energy efficiency of our system stems from the fact that the acoustic energy is mainly confined to the surface of the substrate and dissipates weakly along the propagation line.55

Experimental study

Device fabrication and experimental setup

Fig. 1a shows the device used in our study. Two sets of IDTs with a period of 100 μm and electrode width of 25 μm were arranged parallel to each other on a 128° Y-cut LiNbO3 substrate. The IDTs were formed by e-beam evaporation of Ti (50 Å, adhesive layer) and Au (800 Å) followed by a lift-off process. A PDMS microchannel with 50 μm width and 100 μm height was fabricated by standard soft lithography and mold-replica techniques.56 The PDMS microchannel was then aligned with the IDTs and bonded on the LiNbO3 substrate such that a pressure node of the SSAW was located in the middle of the channel width. The bonding regions were treated with oxygen plasma to enhance the bonding between the PDMS channel and the LiNbO3 substrate. To generate continuous flow, a syringe pump (KDS 210, KD scientific, Holliston, MA) was used.

Fig. 1.

Fig. 1

(a) A photograph of the experimental apparatus is shown including a PDMS microchannel in between a pair of parallel IDTs. (b) A cross-section schematic of the microchannel and IDTs used to generate two SAWs travelling in the opposite direction. The leakage waves are radiated into the liquid phase under the Rayleigh angle (drawings are not in scale).

To visualize the focused microparticles when the SSAW was generated, the experimental apparatus was mounted on the stage of an inverted microscope (Nikon TE2000U). A fluorescent (dragon green) polystyrene bead (1.9 μm) solution was injected into the microchannel using the syringe pump. An RF signal generator (Agilent E4422B) produced an AC signal that was augmented by a power amplifier (Amplifier Research 100A250A). The amplified signal was equally partitioned into two separate signals, and applied to the two IDTs for SAW generation. The applied AC signal was set to 38.2 MHz, the excitation frequency for SAWs (Rayleigh mode) on LiNbO3, and the applied power was set to 24 dBm (~250 mW).

While particle behavior in the lateral (xz plane) direction can be conveniently visualized through the microscope, monitoring the microparticle distribution across the channel height (y-direction) presents a significant challenge. To address this challenge, a triangular 45 degree prism was placed adjacent to the sidewall of the microchannel to visualize the yz plane. As seen in Fig. 2a, the excitation light, coming from the bottom of the microscope stage into the triangular prism, was bent in the lateral direction toward the travelling particles in the microchannel. The excited microparticles emit fluorescent light that was reflected through the prism into the microscope to be recorded for analysis. Additionally, a CCD camera was connected to the microscope to capture images and real-time videos of our 3D focusing technique.

Fig. 2.

Fig. 2

(a) Schematic of the experimental setup for monitoring the particle behavior in the vertical direction (y-direction): a 45° prism adjacent to the channel was used to bend the excitation/fluorescent light for observation of particle migration in the y-direction. Recorded side view fluorescent images (b) before and (c) after the SSAW was generated.

Experimental results

As discussed in the previous section, in order to observe focusing in the vertical direction, a prism was positioned next to the PDMS channel sidewall. When the particles passed through the microchannel section adjacent to the prism at a flow rate of 7 μL min–1, their movement in the vertical direction (yz plane) was captured through the prism using a CCD camera. As shown in Fig. 2b, when the RF signal is off and no SAW is generated, the distribution of the fluorescent particles across the microchannel is random. When the RF is switched on, the particles experience the acoustic radiation forces from the SAW-induced longitudinal-mode leakage waves in the microchannel. As a result, they are focused to a balance line, located about 42 μm above the bottom of the channel (Fig. 2c). Further details on the focusing mechanism are discussed in the theoretical section of this article.

Alternatively, we can confirm the 3D focusing using the recorded top-view images (Fig. 3). Compared to the previous results in Fig. 2, the flow rates for the top-view imaging results were reduced by an order of magnitude to 0.7 μL min–1. The images were taken from a fixed camera position over a period of 2 s to track the redistribution of particles upon SSAW generation. At t = 0, the RF signal was off, and the distribution of the particles was random in both lateral and vertical directions. Note that the particles out of the focal plane appear blurry and large. At t = 0.1 s, the SSAW was generated and the particles started to redistribute in the lateral and vertical directions. The yellow dotted contours show the snapshots of three particles when they redistribute in the lateral direction (xz plane). As shown in different time snapshots, the acoustic radiation forces drove these particles to the pressure node of the SSAW located in the middle of the channel along the x-direction. The white dotted contours illustrate focusing in the vertical direction (y-direction): the particles out of the focal plane of the microscope, which appear large and blurry, were gradually driven to the focal plane of the microscope and aligned in the middle of the channel. The enlarged view of the selected particles in white contours is shown on the right-side column in Fig. 3. The duration of the entire focusing process is about 2 s, indicating the speed and efficiency of this focusing mechanism. It should be noted that a power less than 250 mW was used in this experiment in order to reduce the focusing speed to capture clear images. Reliable reversibility can be achieved by simply turning the SSAW on and off. After passing through the SSAW working region, the microparticles remain focused due to properties of laminar flow and large Peclet number.

Fig. 3.

Fig. 3

Top-view images of the fluorescent particles recorded at different time snapshots. Particles are randomly distributed in the channel before the RF power is on (t = 0). After the RF signal is applied and the SSAW is generated, particles are focused laterally and vertically. The yellow dotted contours show three particles that are driven to the center of the channel (pressure nodal plane). The white dotted contours illustrate focusing in the vertical direction; these contours are enlarged on the right-side column. The three particles are out of focus before the signal is applied. When SSAW is generated, they start to migrate to the focal plane of the optical apparatus, implying migration in the y-direction (vertical direction).

Theoretical and computational study

In order to understand the working principle of the SSAW-based 3D focusing and investigate the position of the focal point, we performed simulations using MATLAB. The propagating SAW (Rayleigh mode) from each IDT with amplitude normal to the substrate surface is partially absorbed by the fluid. Due to this energy absorption, a finite pressure difference is generated in the fluid between the peaks of the wave. The pressure difference above the surface of the substrate leads to the generation of a longitudinal sound wave in the liquid. This acoustic energy leakage into the liquid causes an exponential decay in the propagating SAW power. The leakage rate can be calculated as,57

α0=vRvSzwzSfscos(θR), (1)

where νR, νS, zw and zS are Rayleigh-wave velocity, shear-wave velocity, impedance of water, and shear-wave impedance of the substrate, respectively. fs is a dimensionless parameter of the substrate.57 Because of the difference between the velocity of sound in the fluid and substrate, longitudinal waves are radiated into the fluid at an angle θR, known as the Rayleigh angle, which is determined by the ratio of speed of sound in liquid to the speed of sound in solid given by the Snell's law,57,58

sin(θR)=vwvR, (2)

where νw is the speed of sound in liquid.

We assume that when the propagating SAW from each IDT reach the microchannel, each point on the substrate generates a plane longitudinal wave propagating along the Rayleigh angle into the water, as illustrated in Fig. 1b. As the longitudinal waves travel through the water, their energy attenuates with distance due to viscosity and thermal effects. Reflection from the channel walls and interference between the longitudinal waves also affect acoustic fields in the channel and is accounted for in the simulation. The attenuation and reflection coefficients used in the simulations are provided in the ESI†. Fig. 4a presents a time snapshot of the displacement field of a longitudinal wave in the microchannel radiated from a SAW propagating in the positive x-direction. The other IDT generates a SAW travelling in the opposite direction, i.e., negative x-direction (Fig. 4b). The arrows denote the local displacement magnitude and direction at a particular time. Both longitudinal waves enter the liquid under an angle of diffraction, θR, calculated as 22.5° according to eqn (2) with νw = 1490 m s–1 and νR = 3900 m s–1. Note that interference of these two oppositely propagating SAWs leads to the formation of a SSAW on the substrate.

Fig. 4.

Fig. 4

Time snapshots of displacement of the longitudinal-mode leakage wave from a SAW propagating in (a) the positive x-direction and (b) the negative x-direction. Resultant pressure field of the generated SSAW due to the penetration of two SAWs from opposite directions is shown in the cross-section of the channel at (c) t = t0 and (d) t = t0 + τ/2. Resultant pressure field distribution confirms the existence of a pressure nodal plane in the middle of the channel, as denoted by the dotted lines.

The resultant displacement field of these longitudinal waves generated by two oppositely propagating SAWs forms a standing wave along the x-direction in the cross-section of the microchannel. The corresponding pressure field of the combined displacement fields from Fig. 4a and b was simulated at two time snapshots with a time difference of half a period (τ/2). These simulation results are shown in Fig. 4c and d. Due to the interference of the radiated and reflected longitudinal waves, the resultant pressure field is not uniform. The strong discontinuity in the pressure field in Fig. 4c and d is due to the destructive interference of the radiated longitudinal waves and their reflections from the channel walls. During this time difference, the amplitude of the pressure in the yz plane of the channel at x = 25 μm remains zero (dashed line), while the pressure in the other analogous regions simply changes sign. These facts indicate that a pressure node exists in the middle plane of the channel (x = 25 μm), as denoted by the dashed lines in Fig. 4c and d.

The longitudinal waves induce pressure fluctuations in the liquid which, in turn, generate primary acoustic radiation forces. As a result of the radiation forces, suspended particles in the liquid aggregate at pressure nodes or anti-nodes depending on the relative density and compressibility of the particles and fluid media.59 As shown in Fig. 4c and d, a pressure node exists in the middle of the microchannel once the SSAW is generated. Eqn (3) gives the expression for the primary radiation forces Fax,

Fax=(πρ02Vcβw2λ)φ(β,ρ)sin(2kx), (3)

where P0 and Vc are the acoustic pressure and particle volume, respectively; λ is the wavelength; β is compressibility; ρ is density; and φ, the acoustic contrast factor, indicates whether particles aggregate at pressure nodes (positive φ) or pressure anti-nodes (negative φ):59

φ(β,ρ)=5ρc2ρw2ρc+ρwβcβw, (4)

where subscript c and w stand for particle and fluid, respectively. In our experiments, we used fluorescent polystyrene beads of a positive φ. Hence, the polystyrene particles in the microchannel move towards the pressure nodal plane. The constant values in eqn (3) and (4) can be found in ref. 60.

In our SSAW-based device, standing waves have a nonuniform acoustic field distribution in the vertical direction (i.e., y-direction), owing to a number of reasons such as near-field effects,61 source inhomogeneity,62 and energy density reduction towards the edges of localized axial field.63 This inhomogeneous acoustic energy distribution in the vertical direction generates the transverse primary radiation forces.55 Once the particles are moved, by the axial primary radiation forces, to the pressure nodal (yz) plane in the center of the microchannel, the transverse primary radiation forces move them to particular points with maximum acoustic kinetic energy within the nodal plane. These points are referred to by Schram as “acoustic hot spots”.64 Note that focusing of particles is due to the balance of axial and transverse primary radiation forces; however, the working mechanisms of these two forces are different. Axial primary forces are stronger, so they first push the particles to the pressure nodal plane. Then the weaker transverse force, which depends on acoustic energy gradient, forces the particles to maximum kinetic energy point. Please refer to ESI† for more details.

Because the time-averaged kinetic energy KE is proportional to the square of velocity (V2),60

KE=ρwV22, (5)

the distribution of the velocity field is required in order to find the maximum kinetic energy point(s) within the pressure nodal plane. Fig. 5a and b demonstrate the time-averaged velocity squared field in the x and y directions, respectively. Note that we are merely interested in the square of the velocity's magnitude within the pressure nodal plane. Fig. 5c shows the time-averaged distribution of velocity squared in x and y directions within the pressure nodal plane. The discontinuities near y = 20 μm and y = 60 μm are due to destructive interference of the longitudinal waves. As shown in Fig. 5a, the magnitude of the time-averaged velocity squared in the x-direction in the nodal plane is zero (black dotted contour in Fig. 5c). Hence, the magnitude of the time-averaged velocity squared is equal to the time-averaged velocity squared in the y-direction (Vy2), represented by the red solid curve. The local maximum of the Vy2 contour, which from eqn (5) is also the local maximum kinetic energy, is approximately at y = 46 μm. Thus, after the particles are focused to the middle of the microchannel due to the axial primary radiation forces, transverse primary radiation forces push them to a point around 46 μm above the bottom of the channel: the “acoustic hot spot.” It is worth mentioning that the position of the focal point is highly dependent on the channel design. The width of the channel is determined by the applied frequency. Therefore, after fixing the width, channel height can be used as a design parameter to change the position of the focused particles in the vertical direction (y-direction). Overall, increasing the channel height decreases the normalized focusing point. Please refer to ESI† for more details.

Fig. 5.

Fig. 5

Time-averaged velocity squared field in the (a) x-direction (Vx2) and (b) y-direction (Vy2) is shown. (c) Time-averaged velocity squared in the x-direction (Vx2) and y-direction (Vy2) within the middle plane (x = 25 μm) is plotted.

The minor discrepancy between experimental (y = 42 μm) and computational (y = 46 μm) results can be attributed to several factors. The main factor is the longitudinal waves penetrating the microchannel from the bulk acoustic waves (BAWs) generated from IDTs that travel within the bulk of the piezoelectric substrate.6568 Due to the complex nature of the transportation and attenuation of BAW energy and numerous interactions between the different travelling waves in the microchannel, this factor was not considered in our calculations. Another factor that could contribute to the discrepancy in the final focusing position is the secondary field forces.59,60 The resultant force is primarily attributed to the overall travelling forces, although the secondary forces do have minor, undetermined effects on the final balance point.

Conclusions

In conclusion, we demonstrate 3D continuous particle focusing in a microfluidic channel using SSAWs. Simulation results predict the existence and location of a focusing point, which is in good agreement with our experimental observations. Compared with many existing 3D particle focusing techniques, our SSAW-based method is non-invasive, reversible, energy-efficient, easy to implement, and applicable to almost any microparticle. With these advantages, this method is expected to be valuable in many lab-on-a-chip applications.

Supplementary Material

sd

Acknowledgements

The authors thank Brian Kiraly for the helpful discussion. This research was supported by NIH Director's New Innovator Award (1DP2OD007209-01), the National Science Foundation (0348149, 0824183, and 0801922), US Department of Agriculture (USDA/NRI), Air Force Office of Scientific Research (AFOSR), and the Penn State Center for Nanoscale Science (MRSEC). Components of this work were conducted at the Penn State node of the NSF-funded National Nanotechnology Infrastructure Network.

Footnotes

Electronic supplementary information (ESI) available: (1) Real-time video of fluorescent polystyrene beads focusing inside a microchannel; and (2) detailed information on device fabrication and simulation. See DOI: 10.1039/c1lc20042a

References

  • 1.Wang MM, Tu E, Raymond DE, Yang JM, Zhang H, Hagen N, Dees B, Mercer EM, Forster AH, Kariv I, Marchand PJ, Butler WF. Nat. Biotechnol. 2005;23:83–87. doi: 10.1038/nbt1050. [DOI] [PubMed] [Google Scholar]
  • 2.Mao X, Lin S-CS, Dong C, Huang TJ. Lab Chip. 2009;9:1583–1589. doi: 10.1039/b820138b. [DOI] [PubMed] [Google Scholar]
  • 3.Simonnet C, Groisman A. Anal. Chem. 2006;78:5653–5663. doi: 10.1021/ac060340o. [DOI] [PubMed] [Google Scholar]
  • 4.Yun H, Bang H, Min J, Chung C, Chang JK, Han D-C. Lab Chip. 2010;10:3243–3254. doi: 10.1039/c0lc00041h. [DOI] [PubMed] [Google Scholar]
  • 5.Hur SC, Kwong Tse HT, Di Carlo D. Lab Chip. 2010;10:274–280. doi: 10.1039/b919495a. [DOI] [PubMed] [Google Scholar]
  • 6.De Mello AJ, Edel JB. J. Appl. Phys. 2007;101:084903. [Google Scholar]
  • 7.Kim M, Hwang DJ, Jeon H, Hiromatsu K, Grigoropoulos CP. Lab Chip. 2009;9:311–318. doi: 10.1039/b808366e. [DOI] [PubMed] [Google Scholar]
  • 8.Wang TH, Peng Y, Zhang C, Wong PK, Ho CM. J. Am. Chem. Soc. 2005;127:5354–5359. doi: 10.1021/ja042642i. [DOI] [PubMed] [Google Scholar]
  • 9.Zhang C-Y, Chao S-Y, Wang TH. Analyst. 2005;130:483–488. doi: 10.1039/b415758c. [DOI] [PubMed] [Google Scholar]
  • 10.Gambin Y, Simonnet C, VanDelinder V, Deniz A, Groisman A. Lab Chip. 2010;10:598–609. doi: 10.1039/b914174j. [DOI] [PubMed] [Google Scholar]
  • 11.Franke T, Braunmüller S, Schmid L, Wixforth A, Weitz DA. Lab Chip. 2010;10:789–794. doi: 10.1039/b915522h. [DOI] [PubMed] [Google Scholar]
  • 12.Shi J, Hua H, Stratton Z, Huang Y, Huang TJ. Lab Chip. 2009;9:3354–3359. doi: 10.1039/b915113c. [DOI] [PubMed] [Google Scholar]
  • 13.Knight JB, Vishwanath A, Brody JP, Austin RH. Phys. Rev. Lett. 1998;80:3863–3866. [Google Scholar]
  • 14.Pollack L, Tate MW, Darnton NC, Knight JB, Gruner SM, Eaton WA, Austin RH. Proc. Natl. Acad. Sci. U. S. A. 1999;96:10115–10117. doi: 10.1073/pnas.96.18.10115. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Pabit SA, Hagen SJ. Biophys. J. 2002;83:2872–2878. doi: 10.1016/S0006-3495(02)75296-X. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Sundararajan N, Pio MS, Lee LP, Berlin AA. J. Microelectromech. Syst. 2004;13:559–567. [Google Scholar]
  • 17.Simonnet C, Groisman A. Appl. Phys. Lett. 2005;87:114104. [Google Scholar]
  • 18.Yang R, Feeback DL, Wang W. Sens. Actuators, A. 2005;118:259–267. [Google Scholar]
  • 19.Chang CC, Huang ZX, Yang RJ. J. Micromech. Microeng. 2007;17:1479–1486. [Google Scholar]
  • 20.Eyal S, Quake SR. Electrophoresis. 2002;23:2653–2657. doi: 10.1002/1522-2683(200208)23:16<2653::AID-ELPS2653>3.0.CO;2-H. [DOI] [PubMed] [Google Scholar]
  • 21.Mao X, Waldeisen JR, Huang TJ. Lab Chip. 2007;7:1260–1262. doi: 10.1039/b711155j. [DOI] [PubMed] [Google Scholar]
  • 22.Terray A, Hart SJ. Lab Chip. 2010;10:1729–1731. doi: 10.1039/b927244e. [DOI] [PubMed] [Google Scholar]
  • 23.Oakey J, Applegate RW, Arellano E, Di Carlo D, Graves SW, Toner M. Anal. Chem. 2010;82:3862–3867. doi: 10.1021/ac100387b. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Di Carlo D, Irimia D, Tompkins R, Toner M. Proc. Natl. Acad. Sci. U. S. A. 2007;104:18892–18897. doi: 10.1073/pnas.0704958104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Fu AY, Spence C, Scherer A, Arnold FH, Quake SR. Nat. Biotechnol. 1999;17:1109–1111. doi: 10.1038/15095. [DOI] [PubMed] [Google Scholar]
  • 26.Jacobson SC, Ramsey JM. Anal. Chem. 1997;69:3212–3217. doi: 10.1021/ac970192p. [DOI] [PubMed] [Google Scholar]
  • 27.Schrum DP, Culbertson CT, Jacobson SC, Ramsey JM. Anal. Chem. 1999;71:4173–4177. doi: 10.1021/ac990372u. [DOI] [PubMed] [Google Scholar]
  • 28.Wong PK, Wang TH, Deval JH, Ho CM. IEEE/ASME Trans. Mechatronics. 2004;9:366–376. [Google Scholar]
  • 29.Wong PK, Chen CY, Wang TH, Ho CH. Anal. Chem. 2004;76:6908–6914. doi: 10.1021/ac049479u. [DOI] [PubMed] [Google Scholar]
  • 30.Erickson D, Liu X, Venditti R, Li D, Krull UJ. Anal. Chem. 2005;77:4000–4007. doi: 10.1021/ac050236r. [DOI] [PubMed] [Google Scholar]
  • 31.Erickson D, Liu X, Krull U, Li D. Anal. Chem. 2004;76:7269–7277. doi: 10.1021/ac049396d. [DOI] [PubMed] [Google Scholar]
  • 32.Chung AJ, Kim D, Erickson D. Lab Chip. 2008;8:330–338. doi: 10.1039/b713325a. [DOI] [PubMed] [Google Scholar]
  • 33.Chu H, Doh I, Cho Y. Lab Chip. 2009;9:686–691. doi: 10.1039/b812213j. [DOI] [PubMed] [Google Scholar]
  • 34.Glawdel T, Ren CL. Encyclopaedia of Microfluidics and Nanofluidics, Springer-Verlag Heidelberg. 2008:516–516. [Google Scholar]
  • 35.Holmes D, Morgan H, Green N. Biosens. Bioelectron. 2006;21:1621–1630. doi: 10.1016/j.bios.2005.10.017. [DOI] [PubMed] [Google Scholar]
  • 36.Liang L, Qian S, Xuan X. J. Colloid Interface Sci. 2010;350:377–379. doi: 10.1016/j.jcis.2010.06.067. [DOI] [PubMed] [Google Scholar]
  • 37.Yu C, Vykoukal J, Vykoukal D, Schwartz J, Shi L, Gascoyne P. J. Microelectromech. Syst. 2005;14:480–487. [Google Scholar]
  • 38.Liu C, Stakenborg T, Peeters S, Lagae L. J. Appl. Phys. 2009;105:102014. [Google Scholar]
  • 39.Liu C, Lagae L, Borghs G. Appl. Phys. Lett. 2007;90:184109. [Google Scholar]
  • 40.James CD, McClain J, Pohl KR, Reuel N, Achyuthan KE, Bourdon CJ, Rahimian K, Galambos PC, Ludwig G, Derzon MS. J. Micromech. Microeng. 2010;20:045015. [Google Scholar]
  • 41.Nilsson A, Petersson F, Jönsson H, Laurell T. Lab Chip. 2004;4:131–135. doi: 10.1039/b313493h. [DOI] [PubMed] [Google Scholar]
  • 42.Goddard GR, Sanders CK, Martin JC, Kaduchak G, Graves SW. Anal. Chem. 2007;79:8740–8746. doi: 10.1021/ac071402t. [DOI] [PubMed] [Google Scholar]
  • 43.Petersson F, Nilsson A, Holm C, Jönsson H, Laurell T. Lab Chip. 2005;5:20–22. doi: 10.1039/b405748c. [DOI] [PubMed] [Google Scholar]
  • 44.Laurell T, Petersson F, Nilsson A. Chem. Soc. Rev. 2007;36:492–506. doi: 10.1039/b601326k. [DOI] [PubMed] [Google Scholar]
  • 45.Kuznetsova L, Coakley WT. J. Acoust. Soc. Am. 2004;116:1956–1966. [Google Scholar]
  • 46.Yasuda K, Umemura S, Takeda K. Jpn. J. Appl. Phys. 1996;1:3295–3299. [Google Scholar]
  • 47.Shi J, Mao X, Ahmed D, Colletti A, Huang TJ. Lab Chip. 2008;8:221–223. doi: 10.1039/b716321e. [DOI] [PubMed] [Google Scholar]
  • 48.Kulkarni KP, Ramarathinam SH, Friend J, Yeo L, Purcell AW, Perlmutter P. Lab Chip. 2010;10:1518–1520. doi: 10.1039/c001501f. [DOI] [PubMed] [Google Scholar]
  • 49.Shi J, Ahmed D, Mao X, Lin S-CS, Lawit A, Huang TJ. Lab Chip. 2009;9:2890–2895. doi: 10.1039/b910595f. [DOI] [PubMed] [Google Scholar]
  • 50.Rogers PR, Friend JR, Yeo LY. Lab Chip. 2010;10:2979–2985. doi: 10.1039/c004822d. [DOI] [PubMed] [Google Scholar]
  • 51.Ahmed D, Mao X, Shi J, Krishna Juluri B, Huang TJ. Lab Chip. 2009;9:2738–2741. doi: 10.1039/b903687c. [DOI] [PubMed] [Google Scholar]
  • 52.Ahmed D, Mao X, Krishna Juluri B, Huang TJ. Microfluid. Nanofluid. 2009;7:727–731. [Google Scholar]
  • 53.Petersson F, Aberg L, Sward-Nilsson AM, Laurell T. Anal. Chem. 2007;79:5117–5123. doi: 10.1021/ac070444e. [DOI] [PubMed] [Google Scholar]
  • 54.Goddard G, Martin J, Graves S, Kaduchak G. Cytometry, Part A. 2006;69:66–74. doi: 10.1002/cyto.a.20205. [DOI] [PubMed] [Google Scholar]
  • 55.Auld BA. Krieger Robert E., editor. Acoustic Fields and Waves in Solids. 1990;2:88–94. [Google Scholar]
  • 56.Xia Y, Whitesides M. Langmuir. 1997;13:2059–2067. [Google Scholar]
  • 57.Chou C-H, Khuri-Yakub BT, Kino GS. IEEE Trans. Ultrason. Ferroelectrics Freq. Contr. 1988;35:464–469. doi: 10.1109/58.4183. [DOI] [PubMed] [Google Scholar]
  • 58.Guttenberg Z, Rathgeber A, Keller S, Rädler JO, Wixforth A, Kostur M, Schindler M, Talkner P. Phys. Rev. E: Stat., Nonlinear, Soft Matter Phys. 2004;70:056311. doi: 10.1103/PhysRevE.70.056311. [DOI] [PubMed] [Google Scholar]
  • 59.Hill M, Harris NR. Microfluidic Technologies for Miniaturized Analysis Systems: Ultrasonic Particle Manipulation. Springer; 2007. pp. 357–392. [Google Scholar]
  • 60.Woodside SM, Bowen BD, Piret JM. AIChE J. 1997;43:1727–1736. [Google Scholar]
  • 61.Lilliehorn T, Simu U, Nilsson M, Almqvist M, Stepinski T, Laurell T, Nilsson J, Johansson S. Ultrasonics. 2005;43:293–303. doi: 10.1016/j.ultras.2004.11.001. [DOI] [PubMed] [Google Scholar]
  • 62.Groschl M. Acustica. 1998;84:432–447. [Google Scholar]
  • 63.Bazou D, Foster GA, Ralphs JR, Coakley WT. Mol. Membr. Biol. 2005;22:229–240. doi: 10.1080/09687860500093396. [DOI] [PubMed] [Google Scholar]
  • 64.Schram CJ. In: Advances in Sonochemistry: Manipulation of Particles in an Acoustic Field. Mason TJ, editor. Elsevier; 1991. pp. 293–322. [Google Scholar]
  • 65.Campbell CK. Surface Acoustic Wave Devices for Mobile and Wireless Communications. Academic Press; 1998. pp. 56–58. [Google Scholar]
  • 66.Shi J, Lin S-CS, Huang TJ. Appl. Phys. Lett. 2008;92:111901. [Google Scholar]
  • 67.Lin S-CS, Huang TJ. J. Appl. Phys. 2009;106:053529. [Google Scholar]
  • 68.Lin S-CS, Huang TJ, Sun J-H, Wu T-T. Phys. Rev. B: Condens. Matter Mater. Phys. 2009;79:094302. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

sd

RESOURCES