XeNA: An automated ‘open-source’ 129Xe hyperpolarizer for clinical use

Panayiotis Nikolaou a,b, Aaron M Coffey a,i, Laura L Walkup b,, Brogan M Gust b, Nicholas Whiting c,*, Hayley Newton c, Iga Muradyan e, Mikayel Dabaghyan e, Kaili Ranta d, Gregory D Moroz f, Matthew S Rosen g,h, Samuel Patz e, Michael J Barlow c, Eduard Y Chekmenev a,i,j, Boyd M Goodson b
PMCID: PMC4011489  NIHMSID: NIHMS568858  PMID: 24631715

Abstract

Here we provide a full report on the construction, components, and capabilities of our consortium’s “open-source” large-scale (~1 L/hr) 129Xe hyperpolarizer for clinical, pre-clinical, and materials NMR/MRI (Nikolaou et al., Proc. Natl. Acad. Sci. USA, 110, 14150 (2013)). The ‘hyperpolarizer’ is automated and built mostly of off-the-shelf components; moreover, it is designed to be cost-effective and installed in both research laboratories and clinical settings with materials costing less than $125,000. The device runs in the xenon-rich regime (up to 1800 Torr Xe in 0.5 L) in either stopped-flow or single-batch mode—making cryo-collection of the hyperpolarized gas unnecessary for many applications. In-cell 129Xe nuclear spin polarization values of ~30-90% have been measured for Xe loadings of ~300-1600 Torr. Typical 129Xe polarization build-up and T1 relaxation time constants were ~8.5 min and ~1.9 hr respectively under our SEOP conditions; such ratios, combined with near-unity Rb electron spin polarizations enabled by the high resonant laser power (up to ~200 W), permits such high PXe values to be achieved despite the high in-cell Xe densities. Importantly, most of the polarization is maintained during efficient HP gas transfer to other containers, and ultra-long 129Xe relaxation times (up to nearly 6 hr) were observed in Tedlar bags following transport to a clinical 3 T scanner for MR spectroscopy and imaging as a prelude to in vivo experiments. The device has received FDA IND approval for a clinical study of COPD subjects. The primary focus of this paper is on the technical / engineering development of the polarizer, with the explicit goals of facilitating the adaptation of design features and operative modes into other laboratories, and of spurring the further advancement of HP-gas MR applications in biomedicine.

Keywords: hyperpolarization, MRI, laser-polarized xenon, optical pumping, lung imaging

1. Introduction

The high sensitivity and biological compatibility of hyperpolarized xenon-129 (HP 129Xe) makes it attractive for a wide variety of potential biomedical applications (1-3)—particularly those involving pulmonary medicine (4-15). Xenon’s solubility (16) may be exploited to allow studies in bodily tissues (17) including blood (18,19), as well as those involving various biologically compatible liquids (20-22), and its wide chemical shift range and sensitive spin-relaxation response makes HP 129Xe a useful probe of the local molecular environment (1,23-25), even allowing local blood oxygenation maps as well as spectroscopic studies of flow (9,26-29). While much of the earlier efforts developing HP gases for MR applications concentrated on 3He, the world-wide shortage of this isotope (30) (a product of tritium decay) increases the urgency for the development of improved HP 129Xe approaches, with notable progress over the years (e.g., Refs. (31-39)). Despite these efforts, a major impediment towards wider-scale utilization of HP 129Xe in clinical applications has been the difficulty of reliably and inexpensively producing large quantities of hyperpolarized xenon with high 129Xe polarization (PXe).

In order to facilitate the implementation of HP 129Xe in biomedical and clinical applications, our collaboration has developed a 129Xe ‘hyperpolarizer’ for clinical, pre-clinical, and materials research (40). The device prepares HP 129Xe via the well-established technique of spin-exchange optical pumping (SEOP (41)), where unpaired electronic spins of an alkali metal vapor (typically Rb or Cs (42)) are polarized via depopulation optical pumping with circularly polarized laser light, and the polarization is subsequently transferred to xenon nuclear spins during gas-phase collisions via the hyperfine interaction. However, unlike most 129Xe polarization devices (including those from commercial sources), the presently described hyperpolarizer runs in batch / stopped-flow mode in the high-Xe-density regime (up to 1800 Torr in a 0.5 L cell). In many circumstances this mode of operation obviates, in part, the usual requirement to cryo-collect the HP 129Xe (e.g., Ref. (43))—a process that otherwise increases design complexity and can also lead to undesirable polarization losses during accumulation, storage, phase transitions, and transfer of HP 129Xe (44). Cryo-collection also helps ensure that residual Rb from the cell is not transferred to sample containers; thus our device includes a Teflon filter in the transfer line to getter any Rb or other particulate matter that might otherwise enter the sample space. Furthermore, the device is automated, modular, portable, and relatively easy to use, employing an “open-source” design comprised of mostly off-the-shelf components; as described below, many of the initially ‘custom’ components can now be readily purchased from commercial suppliers. Originating from our previous efforts studying batch-mode and stopped-flow (32) SEOP at high Xe densities and resonant laser fluxes (39,45,46), this new scaled-up design can achieve ~1 L/hr production rates, with 129Xe nuclear spin polarization values (PXe) of ~90%, ~57%, ~50%, and ~30% for Xe loadings of ~300, ~500, ~760, and ~1570 Torr, respectively (40). Furthermore, little polarization loss is suffered during cell cool-down and gas transfer from the device to sample containers. Long decay times for the HP 129Xe contrast agent (up to nearly 6 hr) were observed in Tedlar bags following transport to a clinical 3 T scanner. Importantly, XeNA is presently being used for a study involving COPD subjects at Brigham & Women’s Hospital (FDA IND #116,662).

The focus of this paper is on the technical / engineering development and capabilities of the polarizer (47,48). To briefly outline the subsequent sections: The Materials and Methods section focuses on describing the major components of the polarizer in great detail, comprising: (a) the framework design and chassis used to house and setup all the components, (b) the electromagnetic coils, (c) the laser, supporting optics, and optical path, (d) the optical pumping (OP) cell design and preparation including the filling procedure, (e) OP oven design and temperature control components, (f) gas manifold design, major components and vacuum system, (g) description of the (optional) cryo-collection setup and design, (h) microcontroller design, automation control and interface; and (i) in situ detection instrumentation. Commercial suppliers and part numbers for components are provided throughout in order to facilitate the construction of this ‘open-source’ device in other labs. The Results section describes the methods used for quality assurance (QA) of the PXe along with examples of 129Xe MRI from a human subject; for more in-depth analyses and theoretical considerations of the PXe results, please refer to Ref. (40). The Discussion section provides the practical advantages and limitations of the hyperpolarizer, as well as some possible design improvements to be investigated in the future. We briefly note here other 129Xe polarizers in the literature (e.g., Refs. (31-39,43,49-51)—as well as those from commercial sources (52-54)); those considering assembly or purchase of a polarizer are encouraged to review the designs and capabilities of many devices in light of their own applications, needs, and resources.

2. Materials and Methods

Our consortium’s “open-source” automated 129Xe polarizer (dubbed “XeNA” for XEnon polarizatioN Automated), is principally comprised of a mobile chassis, an electromagnetic coil assembly, a laser system, a vacuum / gas-handling manifold, on-board spectrometers (and other sensors) for QA and feedback, and a computer/automation system (Fig. 1A). The details of the design, components, assembly, and operation of XeNA are provided here and in the corresponding sections of the Supplemental Information documentation.

Figure 1.

Figure 1

The XeNA polarizer. (a) of the polarizer’s key components (the self-pressurized liquid N2 dewar that provides gas for heating and cooling of the oven and N2 gas cylinder used to operate pneumatic valves are not shown) (40). The optical path is represented by (“λ/4”) and is comprised of beam expanding optics, polarizing beam-splitter cube, quarter-wave plate, and heat sinks (see Fig. 2). For the gas cylinders, “N” and “E” designate xenon with naturally-abundant 129Xe and isotopically-enriched 129Xe, respectively (40). (b) Photograph of XeNA with open laser enclosure in its current location in a clinical MRI suite at Brigham and Womens’ Hospital, Boston, MA, USA.

2a. Frame

The lower portion of the framework of this polarizer design consists of two 19 in. rack-mountable cabinets used to house the required hardware (Figs. 1b & Fig. S1). The mobile device is supported by six 4 in. heavy-duty casters (McMaster-Carr, P/N #27075T72). The upper section of the polarizer is constructed from extruded aluminum slotted strut (45 mm × 45 mm) from Minitec Framing, which results in a strong and configurable framework—see Fig. S2 for additional details. The base support of the upper frame is bolted directly to the lower console to support the magnet coils and laser enclosure. The entire optics train and magnet assembly is contained in a light-tight enclosure with a top-mounted door built with Minitec slots rails with attached Alumalite panels (corrugated plastic core with 0.013 in. thick matte black painted aluminum skins on both sides).

2b. Magnets

A homogenous magnetic field is provided by a four-coil electromagnet assembly (P/N A650011, Acutran, Fombell, PA—23.6″ ID; 12 turns per layer, 17 layers, 12AWG) operating at either 5.26 mT (62.0 kHz 129Xe frequency) or 1.46 mT (62.0 kHz 1H frequency). The magnets are supported by the upper Minitec frame and held in place by 90° angle brackets (Minitec Framing). The four-coil geometry of the electromagnet is that of a Helmoltz pair series inspired by the designs Barker (55) and was optimized using BiotSavart software (Ripplon Software Inc, New Westminster, BC, Canada); a true Barker setup would require varying the number of turns, the current and/or the diameters of the inner versus outer coil pairs. The four-coil electromagnets are connected in series and are powered by a single power supply (Kenwood, P/N PDS60/12).

2c. Laser and Optical Path

A single 5.5 foot long optical breadboard (Thor Labs) is bolted to the center of the upper frame to mount the optics and serve as a platform for the laser and optical pumping oven. A custom translational platform (Minitec) with aluminum T-slotted rails is used to hold the laser in position (Fig. 2a). Each T-slotted rail’s position can be adjusted without disassembling the frame; this feature allows for simpler adjustment of the optical breadboard position so that the laser optical path and optical pumping cell may be better aligned with the axial isocenter of the four-coil electromagnet set-up. To further aid with the alignment procedures the laser is equipped with a <5 mW visible aiming beam (Fig. S3 Inset).

Figure 2.

Figure 2

(a) Laser diode array with translational mounting frame, and optical path assembly. (b) Schematic showing the principal elements of the optical path (40). The beam blocks are drawn separated from the polarizing beam splitter (PBS) housing for clarity.

The optical path (Fig. 2b) begins with a 200 W laser diode array (LDA, QPC Lasers, P/N 6507-0001 Brightlock Ultra-500) that produces near-IR laser output that is frequency-narrowed (nominal FWHM~0.27 nm) by ‘on-chip’ volume holographic gratings (VHGs) (46). The nominal output is resonant with the Rb D1 line at ~794.8 nm, and is tunable over >1 nm range by adjusting the LDA’s temperature, which is maintained by a 800 W water-chiller (K-O Concepts model LCR-8), and the laser driving current (Fig. S1b). This tunability allows the laser output to be placed on resonance with the Rb absorption center at different output powers (see Fig. S4b), as well as operate detuned for off-resonant SEOP (45,46,56). The laser beam is incident on a short (~6 in.) optical fiber that preserves most (>90%) of the linear polarization and then is expanded and collimated to 2 in. diameter by two lenses. Prior to entering a 2 in. polarizing beam-splitter (PBS) cube, the s-polarized beam component (<10% of incident power) is reflected 90° to the right and discarded into a beam block; the main (p-polarized) beam is transmitted forward into a quarter-wave (λ/4) plate, which renders the beam circularly-polarized.

The laser beam enters and exits the optical pumping oven through anti-reflection (AR)-coated optical flats (see below). Upon exiting the rear of the oven the beam is retro-reflected back through the cell via a 3 in. mirror (Thor Labs) and directed into the optics assembly, where the PBS reflects the beam into a second beam block. To provide additional flexibility for the polarizer, all optics were chosen to be sufficiently broad-band to also allow operation with Cs SEOP at 894.3 nm (e.g., Ref. (42)). Except for the PBS and λ/4 plate, all components of the optical path assembly attached to the laser were purchased through Thor Labs (See Table S1 and Fig. S3). Due to the high output power of the LDA, cooling fans are installed next to the beam blocks to direct air over their fins (Fig. S3).

2d. Optical Pumping Cell

The cell is a custom Pyrex (borosilicate glass) 2 in. diameter cylinder, with flat optical windows at the front and the rear (Mid-Rivers Glassblowing, Inc., St. Charles, MO, P/N MRG934-01A). Each OP-cell is 9.75 in. long with an internal volume of approximately 500 cc. The oven can accommodate variable cell lengths (6 in. – 12 in.). The OP-cells have one opening, sealed by a Chemglass Teflon stopcock that is pneumatically actuated using a Humphrey Rotary Actuator (P/N HRAPS5-360-S) and custom-designed Helical (P/N 9947-26mm-6mm, Item# 28571, Fig. 3d) closing off a ¼ in. O.D. vertical stem trapping the gas content inside the cell.

Figure 3.

Figure 3

(a) OP-Oven with OP-Cell (Mid-Rivers Glassblowing, Inc., St. Charles, MO, P/N MRG934-01A) mounted inside, with gas manifold components shown connected to the OP-Cell. (b) Glass (Pyrex) spiral storage condenser (Mid-Rivers Glassblowing, Inc., St. Charles, MO, P/N MRG927-01C) controlled via helical-rotary actuator assembly, residing in a strong (>500 Gauss) magnetic field produced by a pair of large 4 in. × 4 in. × 1 in. neodymium-iron-boride permanent magnets (Indigo Instruments). The magnet yoke was custom-machined from aluminum. (c) Two 400 W T-Type heaters (Omega Engineering, Stamford, CT) mounted inside aluminum enclosure. (d) Custom-made Helical connected and Humphrey Rotary Actuator assembly allowing automation of the Chem-Glass Teflon stopcocks. A video of the glass-valve actuation procedure can be viewed at:http://www.youtube.com/watch?v=w33xs9KHuB0.

Prior to installation in the polarizer, each OP-cell is pressure-tested and then prepared as follows. First, each one is soaked in a KOH/methanol base bath for 24 hours to remove impurities from the glass surface. The OP-Cell is then rinsed first with distilled water, and then with methanol. The cell is then placed in a bath ultrasonic cleaner filled with a mixture of methanol and distilled water and sonicated for about 1 hour at room temperature. The cell is then removed and rinsed again with distilled water followed by methanol, and then placed in an oven (110 °C) to dry. The dried OP-cell is coated with a siliconizing agent (SurfaSil, P/N PI-42800, Fisher Scientific) to slow 129Xe T1 relaxation due to interactions of 129Xe with paramagnetic centers in the glass (57-59); here, a 10 mL solution is prepared by diluting 1 mL of SurfaSil with hexane. Some of the solution is then pipetted into the cell and shaken to spread the solution over the entire inner surface. The cell is then emptied and washed with ~10 mL of pure hexane. This process is repeated 3 times, ending with a final hexane rinse. The cell is then placed in a warm oven (~60 °C) for at least 1 hour; afterwards, the cell is allowed to cool and its contents evacuated down to <1×10−3 Torr prior to loading the cell in the transfer chamber of a glove box. The cell is then loaded with ~250 mg of molten Rb metal via pipette in the glove box’s inert atmosphere. The OP-cell is then removed from the glove box and evacuated, and the Rb droplet is heated to its melting point to degas it prior to distribution as a thin film on the cell’s inner surface via local heating/cooling. Once Rb is distributed, the OP-cell can be placed in the OP-oven to be filled with a gas mixture of ultra-high purity 129Xe and N2. The N2 gas helps suppress unwanted re-emission of unpolarized light from the electronically excited Rb (31) (known as radiation trapping) as well as other undesirable energetic processes (60). Additionally, the N2 provides additional collision-broadening of the Rb absorption line, improving the absorption efficiency of the laser (61).

The OP-cell is attached to the polarizer’s gas manifold (Fig. 1A) via a ¼ in. PFA Swagelok union T fitting, and is loaded with a gas mixture of 129Xe and N2 automatically. This automation procedure involves evacuating the manifold lines prior to opening the OP-cell to the manifold to prevent oxidation of the Rb metal. The Edwards RV-5 rotary vane mechanical pump is used for achieving initial rough vacuum level (>10−2 Torr) before switching to an Edwards Turbo Pump Station (TS75W). Once a pressure of ~10−4-10−5 Torr is achieved, the OP-Cell may be closed. The automation sequence continues by performing three purge/vacuum cycles using ultra-high purity N2 gas in preparation to load the OP-cell. Once complete, the desired 129Xe/N2 mixture is loaded into the evacuated OP-cell, see Supporting Information for details. Once the OP-cell has been filled, then the optical pumping process can take place.

2e. Optical Pumping Oven

The oven is manufactured from non-magnetic materials: Teflon, nylon, and glass that can handle moderately high (>120 °C) operational temperatures (Figs. 3A & S4). The front and the rear walls of the oven contain 3 in. borosilicate (BK7) optical windows with anti-reflection coatings (CVI Melles-Griot P/N W2-PW1-3025-C-670-1064-0) appropriate for the near-IR wavelengths (640-1064 nm). The OP-cell and its contents are heated using a forced-air oven. The oven contains two inlets for heating or cooling air, and one exhaust port (Fig. S4). Since it is necessary to have the polarizer work in areas where compressed-air hook-ups may not be available, the gas supply of the temperature-control system is provided by a self-pressurizing liquid nitrogen dewar. A full dewar can supply sufficient gas for 12+ hours of continuous operation. The oven gas supply is routed through two 400 W heat pipes (Omega Engineering, Stamford, CT, P/N APH-5051) mounted inside an aluminum enclosure (Fig. 3C) and connected to the oven inlets via temperature-resistant silicone tubing (3/4 in. O.D. Silcon®, P/N 2802471, NewAge Industries). The heaters are controlled through a temperature controller (Omega Engineering, Stamford, CT, i-series CNi3244-C24) via feedback from thermocouple sensors inside the OP-Oven.

An opening in the side wall of the oven allows access to the cell’s stopcock, which is connected to the Helical-Rotary Actuator. The oven side walls include Pyrex window panels to allow monitoring of the SEOP process with optical probe beams and optimizing the laser beam alignment. The oven is mounted on lateral translational stages to permit fine adjustments to both the front and back of the oven with respect to the laser beam.

2f. Gas Manifold

This gas handling system consists of two distinct sections (Figs. 1a & S1a): the ‘upstream’ section contains unpolarized gas used to fill the OP-Cell, and the ‘downstream’ section is used to transfer hyperpolarized 129Xe to either a Tedlar bag or storage condensation coils. The ‘upstream’ system utilizes solenoid valves (green in Fig. 1a, McMaster-Carr P/N 5077T123 and Parker P/N 099-0167-900) and pneumatically actuated Teflon valves (red in Fig. 1a, Teqcom P/N M222CPFS-T). All valves are automated through the microcontroller box as follows. The valves allow gas flow regulation throughout the entire manifold. The loading ‘upstream’ section of the manifold is built using ¼ in. Swagelok pieces connecting stainless steel tubing and solenoid valves. The Edwards RV5 mechanical roughing pump and Turbo Pump Station are connected to a single vacuum line through a Swagelok T-connector that leads to both ‘upstream’ and ‘downstream’ sections of the gas handling system. It is extremely important to remove all O2 from the system as it not only reacts with the alkali metal in the OP-cell, but if present in the transfer lines will quickly de-polarize the hyperpolarized 129Xe gas.

The gas-handling system is equipped with three cylinders of ultra-high purity natural abundance 129Xe, enriched 129Xe, and N2 gas (all fitted with CG-580 two-stage regulators), thus allowing the option to use natural abundance or enriched 129Xe. After passing through oxygen filters/getters (Oxiclear / Labclear, P/N DGP-250-R1), these gases are delivered to the cell via a pulse-damping snubber (McMaster-Carr) that is used to restrict the flow rate of gases. The gas pressures are monitored by two different pressure gauges (for monitoring vacuum and high pressure respectively), which provide feedback to the microcontroller. The ‘fine’ gauge is a Teledyne Hastings HPM 2002, which is responsible for monitoring from ~800 Torr down to the 10−5 Torr vacuum range. The ‘coarse’ high-pressure gauge (rated to ~6000 Torr; Omega P/N PX309-100G5V) monitors pressure from 760 to 2000 Torr (total operational in OP-cell pressure).

The ‘downstream’ transfer section of the gas handling system is less complex than the ‘upstream’ loading section and contains fewer valves. Because HP 129Xe gas can easily be depolarized, all surfaces that may come into contact with the HP 129Xe must be non-magnetic. As a result, only Teflon components and SurfaSil-coated glass surfaces are used in the ‘downstream’ component of the gas-handling system. Furthermore, the HP 129Xe contrast agent ejection line is situated within the ambient field of the SEOP electromagnet to prevent any zero-field crossings, which could otherwise depolarize HP 129Xe. The transfer line is equipped with a gas filter (Entegris Wafergard GT-Plus; WGFG 21K P3; various suppliers, e.g. Venture Technologies Group, Farmington Hills, MI) to getter any residual Rb metal and other micro- and nanoparticles. The HP 129Xe cryo-collection stage (see below) is optional in the programmed GUI interface, and the condensation coil setup (Fig. 3b) may be installed in the transfer line. As mentioned above, the ‘downstream’ transfer section is also connected to the vacuum system, which evacuates the lines and the Tedlar bag while 129Xe is being optically polarized in the cell.

2g. Cryo-Collection Setup

This setup consists of a custom-made Pyrex tubular spiral storage vessel (Mid-Rivers Glassblowing, Inc., P/N MRG927-01C) that allows the HP 129Xe to be collected and stored prior to transfer to the Tedlar bag and administration to the subject or sample. The storage coil has Chemglass Teflon stopcock valves on each side of the coil. The spiral coil serves the functions of collecting, storing, and transferring the HP 129Xe, and its valves are each controlled via a helical-rotary actuator assembly. The coil resides in a strong (>500 Gauss) magnetic field produced by a pair of large permanent magnets (see above and Fig. 3b). The strong field prevents HP 129Xe depolarization in the solid state through coupling with neighboring (quadrupolar) 131Xe spins in the solid lattice (62). The magnets are held in place with custom-made aluminum holders with a mounting bar that can be attached directly to the main upper frame (Minitec Framing). The condenser may be cooled with liquid N2 (well below the freezing temperature of xenon, allowing N2 gas to pass through), and then, prior to the transfer to the Tedlar bag, warmed either with lukewarm water or with heated nitrogen gas. It is important that HP 129Xe is sublimated rapidly to suppress 129Xe relaxation that peaks as the xenon motion in the lattice increases, just below the triple point (44).

2h. Microcontroller Automation and Interface

The polarizer automation control box (Fig. 4) consists of numerous relays for turning alternating-current (AC) or direct-current (DC) powered devices on or off in a time-dependent sequence with sensor measurements allowing process control. An open-source microcontroller (Arduino Mega 2560 rev3, P/N 1050-1018-ND, Digikey, Thief River Falls, MN) orchestrates all operations. A dual-output power supply (CUI VF-D320-D1224A, P/N 102-2001-ND, Digikey) provides power for the automation control box, a gas valve manifold (Bürkert, Germany, P/N MP11) enabling pneumatic valve operation through 3/2-way solenoids valves (Bürkert, Germany, 6524 Type) of the manifold, and other devices used either in the internal pneumatic manifold or the external gas manifold. The microcontroller digital I/O lines provide activation of the AC or DC solid state relays (SSRs) (Grayhill I/O Relay Module, P/N GH3040-ND or GH-3038-ND, Digikey), thereby selectively powering other devices such as electronic solenoid valves, heaters, instrumentation, etc. Various controller outputs also include activating external devices via TTL level logic, such as shutting down the laser power supply unit (PSU) or triggering the low-field NMR spectrometer (Kea2. Magritek) to run a pulse program. Microcontroller inputs include status changes on digital input lines, and various sensors that can be interacted with via serial communications (see below). Additionally, the micro-controller interacts with a computer graphical user interface (GUI) through commands sent over the built-in USB-serial interface (Fig. S5). All programmed gas manifold load/transfer sequences are stored on the micro-controller board’s memory. The GUI itself was designed and programmed in open-source software (processing.org).

Figure 4.

Figure 4

(a) Schematic of all major components of the Microcontroller automation box. (b) Corresponding photograph.

Sensor inputs also initiate microcontroller responses. The automation control box features two safety interlocks. The first ensures that the water chiller is on before permitting device operation, whereas the second ensures that the upper frame door is closed during laser operation. The water chiller status is sensed by the water ‘low flow’ switch (P/N 2371K41, McMaster-Carr, Aurora, OH). The mechanical safety limit switch sensing the status of the upper frame is P/N 6352K41, McMaster-Carr. Activation of any of these sensors (5V → 0V) results in the following: (i) the laser PSU being sent to standby (laser off), (ii) the OP-cell being closed off to prevent contamination, and (iii) the microcontroller shutting down in a way that requires a physical reset. The pressure sensors described above are monitored by the microcontroller. Oven temperature monitoring and control are achieved by communication between the microcontroller and the oven’s temperature controller.

In general operation, the microcontroller runs a loop that obtains temperature and pressure measurements and transmits them to the computer GUI. Interrupt service routines (ISRs) monitor the water flow and door safety switches. Upon receipt of a command from the computer GUI during this general loop, the appropriate function is called after decoding. Such functions can include: oven temperature or manifold pressure level monitoring; timeouts on loops (as in the case of leaks preventing the gas-handling manifold from reaching specified vacuum levels); obtaining user input from the GUI (such as the requested xenon loading pressure); or reporting controller status information. GUI buttons launch automated procedures such as those necessary for replacing gas cylinders or for production of HP 129Xe.

2i. In-situ polarimetry

In situ detection is enabled by a low-field high-resolution NMR spectrometer (Kea2, Magritek) and an Ocean Optics HR2000+ high-resolution near-IR spectrometer. The IR probe detecting transmitted laser photons is fiber-optically connected behind the 3 in. retro-reflection mirror, whereas the RF coil (tuned to 62 kHz and matched to 50 Ohms) is located directly under the cell. The low-field NMR spectrometer is used for in situ QA of the SEOP process (detecting nuclear spin-polarization of 129Xe). In-situ NMR detection utilizes a small circular surface coil (~1.5 in. O.D.) with 342 turns using 34 AWG wire. This surface coil was tuned utilizing an Agilent E5071c ENA Series network analyzer for 129Xe detection at 5.26 mT and 1H at 1.46 mT respectively. The corresponding near-IR spectra provide detailed information regarding the status and output quality of the laser, and how well it is being absorbed by the Rb alkali metal in the cell. Additionally, the transmitted laser spectra can be used to infer the average electron spin polarization (40,45), see Supporting Information.

3. Results

Four independent methods have been used to determine nuclear and electron spin polarization (40). These methods include (i) in-situ (in OP-cell) field-cycled near-IR spectroscopy (45) to probe Rb electron spin polarization; (ii) in-situ (in OP-cell) 129Xe NMR at 5.26 mT; (iii) ex-situ HP 129Xe gas transferred to a “intermediate field” (47.5 mT) of a 122 mm bore magnet (Magritek); and (iv) ex-situ HP 129Xe gas transferred into a Tedlar bag and subsequent 129Xe NMR/MRI using a clinical 3 T scanner. Excellent agreement was observed among the different approaches used to quantify PXe (see Fig. 6 below). A detailed analysis of these results has been reported previously (40), and thus only a brief discussion is provided here.

Figure 6.

Figure 6

129Xe nuclear spin polarization values measured at 5.26 mT, 47.5 mT, and/or 3 T, plotted versus xenon partial pressure determined at loading (40). Labels ‘before transfer’ and ‘after transfer’ respectively refer to measurements obtained from Xe gas remaining within the SEOP cell before and after some of the gas was transferred to another container. Error bars are determined from the uncertainties in the spectral integral values obtained from the respective thermally-polarized reference samples. The value at 725 Torr was obtained with 82%-enriched 129Xe loaded with a different gas manifold.

The resulting in-situ HP 129Xe NMR spectra (e.g., Fig. 5a) were compared to the spectra of thermally polarized water at the same NMR frequency, allowing in-cell PXe values to be determined; in this example, PXe=49.5%±2.8% was achieved for a cell containing 761 Torr Xe. The thermal 1H reference signal (Fig. S9a,b) used to calculate PXe was obtained using an identical OP-cell filled with 5 mM CuSO4 doped water (63,64). Furthermore, in-situ NMR detection allows measurement of build-up and decay of HP 129Xe during and after SEOP. Examples of different PXe accumulation and 129Xe T1 relaxation studies are shown in Figs. 5b,c. The 129Xe polarization build-up and T1 decay time constants were 8.5±0.7 min and 1.9±0.6 hr, respectively.

Figure 5.

Figure 5

(a) Example of low-field in-situ HP 129Xe NMR used for QA, here obtained from a SEOP cell containing 761 Torr Xe and 1239 Torr N2. (b) Typical in-cell PXe build-up curve measured via in situ 129Xe NMR, here for a cell containing ~725 Torr Xe / ~1275 Torr N2. (c) Spin-lattice T1 decay curve for HP 129Xe NMR signals from Xe gas measured at 5.26 mT of a cell containing 495 Torr Xe / 1300 Torr N2 following SEOP and subsequent cell cool-down to 33 °C (by which point the Rb should be condensed—providing a more accurate measure of the intrinsic cell relaxation rate). In normal operation, HP 129Xe gas is typically transferred from the OP-cell to the sample at ~40 °C to help preserve more of the polarization.

Ex-situ detection at 47.5 mT allows for simultaneous evaluation of 129Xe polarization quality and agent-transfer efficiency. The polarized gas was transferred via a 2 meter long ¼ in. O.D. PTFE tubing into a 52 mL polypropylene spherical phantom located within the field of the 47.5 mT bore magnet system. HP 129Xe was detected at 559 kHz, whereas a 13C reference sample was detected at 508 kHz using a broad-band dual-channel RF probe (63). Comparison of the ex situ 47.5 mT results with the in-situ 5.26 mT 129Xe NMR polarimetry taken during the same experiment generally showed little loss of 129Xe polarization during transfer from the SEOP cell (40). Similar efficiencies were observed using ex-situ HP 129Xe spectroscopy at 3 T (40).

HP 129Xe gas expanded into 800 mL Tedlar bags was used to quantify the relaxation (i.e. life-time of the produced contrast agent) using a clinical 3 T MRI scanner. For most acquisitions, two different storage fields (0.0015 T and 3 T) were used to cover the dynamic range of FDA-approved magnetic fields for clinical MRI use. 129Xe T1 time constants were 38±12 min and 5.9±0.4 hr at 1.5 mT and 3 T respectively (40)—results that project positively for a variety of clinical and pre-clinical applications requiring gas-phase storage or accumulation of HP 129Xe. When polarizing 129Xe via batch/stopped-flow SEOP without cryo-collection, it is also important that residual Rb does not end up in sample containers. The combination of an efficient cell cool-down procedure (see Supplemental Information) and the use of the Teflon filter in the transfer line (described above) help to ensure that Rb (or other particulate matter) does not end up in the inhaled gas mixure. Following three separate SEOP runs, Tedlar bags were prepared and sent for elemental analysis (Element One, Wilmington, NC). In all cases, Rb content was found to be less than 5 ng / bag.

129Xe polarization values achieved with various in-cell Xe densities (i.e., with Xe partial pressures between ~300-1600 Torr), over a range of conditions including different magnetic field strengths, NMR acquisition methodologies, etc., are summarized in Fig. 6 (40). The data include values for in-cell PXe values of 90.9±5.2%, 57.1±3.3%, 50.1±2.9%, and 33.4±1.9% measured for Xe loadings of 300, 495, 765, and 1570 Torr, respectively; PXe values of 41±1.6% and 28±1.1% with ~760 and ~1545 Torr Xe loadings were obtained following transfer to Tedlar bags and subsequent transport to the 3 T scanner.

The polarizer has recently received FDA Investigational New Drug and IRB regulatory approval for use in a clinical study of COPD patients. Towards that end, the polarizer has already been used in several preliminary 129Xe MR imaging and spectroscopic experiments involving individual healthy human subjects; for example, images obtained in one experiment are shown in Fig. 7. For this experiment, a gas mixture comprised of 1300 Torr of 86%-enriched 129Xe gas and 700 Torr N2 gas was loaded and hyperpolarized in the OP-cell. Prior to use, PXe was measured to be 25% in the cell—somewhat lower than typical values (see Fig. 6), owing to the utilization of a OP-cell that was thermally regenerated following partial oxidation during installation. Some of the cell contents were then transferred to a 0.8 L Tedlar bag via expansion (thus, xenon comprised two-thirds of the bag contents). The bag was then transported to a Siemens 3 T clinical MRI scanner and the HP 129Xe agent was administered to a healthy human volunteer using the following inhalation protocol: the subject completed two full respiration cycles (total lung capacity (TLC) to functional residual capacity, (FRC), ×2), before being asked to inhale contents from the bag (followed by a small gulp of air to help push HP 129Xe out beyond the trachea). The demonstrated signal strength and image quality achieved are sufficient for the polarizer to support a number of potential pulmonary functional imaging/spectroscopic studies—the subject of future efforts.

Figure 7.

Figure 7

Selected (12 of 14) false-color 2D slices from a 3D 129Xe GRE chest image from a healthy subject following inhalation of HP 129Xe prepared using the XeNA polarizer (anterior to posterior, reading left to right, top to bottom). TE/TR 1.12/11 ms (SAR-limited), tipping angle α=6°, 80×80×14 matrix, acquisition time=4.5 s; FOV=320×320×196 mm3, data zero-filled to give 2×2×14 mm3 digital resolution (SNR~8-15).

4. Discussion

The 129Xe polarization values summarized here (40) represent by a significant margin the highest yet achieved at such high in-cell Xe densities—and establish the feasibility of attaining near-unity polarization in single (or multiple) batches with HP 129Xe quantities sufficient for most clinical uses (~1 L/hr). Furthermore, the hyperpolarizer is automated, modular, portable, and easy to maintain. XeNA is built mostly of commercially available components, and the complete material costs for construction were under $125,000.

The overall performance of the hyperpolarizer stems from a number of technical improvements of stopped-flow high-xenon-pressure SEOP process described here and fundamental improvements reported earlier (40). For instance, the VHG technology combined with high laser power (~170 W) provides high resonant photon flux through the cell. Next, to prevent any dark regions near the edges or longitudinal optical nodes of the OP-cell that could affect the SEOP process, the optical path setup was designed with care as to (i) reduce the loss of laser power through the optical train while expanding to a 2 in. diameter beam and (ii) to illuminate the OP-cell volume with near-unity circular polarization of the incident laser light. Furthermore, we have shown that retro-reflection of the beam back into the cell can provide a ~30% increase in HP 129Xe (45). Careful control of the cell temperature is key for SEOP optimization. Previously it was shown that there can be an inverse relationship between Xe density and the optimal temperature for SEOP (e.g., where lower cell temperatures are favored for higher Xe densities (39,46)—particularly when using spectrally-narrowed pump lasers). The steady-state 129Xe nuclear spin polarization is given by:

PXe(t=)=γSEγSE+ΓXe.PRb, (1)

where γSE is the Rb/Xe spin exchange rate (∞[Rb]), ΛXe is the 129Xe nuclear spin destruction rate (=1/T1), and 〈PRb〉 is the volume-averaged Rb electron spin polarization. For example, if one assumes that the intrinsic cell 129Xe T1 values are similar from cell to cell (and only mildly sensitive to the small differences in Xe loading or total pressure (65)), the time constants from the curves in Fig. 5 would correspond to ~12-fold larger γSE value compared to ΛXe, contributing to the efficiency of the 129Xe polarization process. Thus overall, the high in-cell 129Xe polarization values reported here are achieved primarily by: (i) maintaining a high “photon-to-Rb” ratio throughout the cell, thereby ‘forcing’ near-unity PRb despite the high Rb electron spin-destruction rates encountered at high Xe densities (66); and (ii) obtaining very slow 129Xe intrinsic relaxation rates, ensuring high γSEXe ratios despite the relatively modest Rb number densities expected at our mild operational oven temperatures (67).

High 129Xe polarization/collection efficiency is also promoted by the following: (i) the high quality and uniformity of the Rb coating within the cell; (ii) the presence of in situ real-time monitoring and feedback of the SEOP process of the Rb spectral absorption and 129Xe NMR signal; (iii) optimization of OP-cell cool-down procedure prior to transfer of the hyperpolarized contrast agent (discussed in detail in section ‘i’ of the Supporting Information); (iv) suppressing exposure to O2 and other paramagnetic materials throughout the gas lines; (v) rapid transfer of HP 129Xe contrast agent into a Tedlar transportation/delivery bag; and (vi) avoiding Xe phase transitions and other relaxation-susceptible portions of the Xe phase diagram (44).

Although Xe cryo-collection and subsequent sublimation prior to transfer to the sample is optional for XeNA, operation without cryo-collection does result in dilution of the HP 129Xe with N2 gas and leaves behind a significant fraction of the SEOP mixture in the OP-cell following the transfer process after contrast agent preparation. However, both of these issues can be mitigated by (i) using Xe-rich mixtures demonstrated here; (ii) the addition of a large automated gas piston or balloon (32,51), where the cell contents can be expanded into a much larger volume prior to transfer to the sample or transport vessel; and (iii) selecting the ‘topping-off’ procedure implemented in XeNA, which also slightly increases the device duty cycle. Indeed, when preparing multiple bags of HP 129Xe, this has become the standard mode of operation. Under these typical conditions, the time to produce each bag is ~35-40 min; ~15-20 min is spent polarizing the 129Xe, with the remaining time being used to purge/evacuate gas lines, transfer gases to and from the cell, and heat up / cool down the OP-cell before and after SEOP.

Future improvements may include (i) the re-designing of the OP-oven, because the current design makes installation and removal of the OP-cell somewhat difficult; (ii) the pressurized liquid N2 dewar can be replaced by a more suitable solution for gas heating and cooling to make the polarizer a true stand-alone device; (iii) The LDA and the optical train could be integrated into a single device for easier alignment with the OP-cell (see Supplemental Information); (iv) improving the thermal management of the cell to better mitigate any deleterious effects of high internal gas temperatures (68,69); (v) increased level of automation and components’ integration; (vi) utilization of pre-mixed Xe/N2 gas sources, allowing the duty cycle to be increased; and (vii) increasing the simplicity of the polarization procedures would be key for making a true “push-button” operated device. The current device increases the prospects of more widespread implementation of HP 129Xe MR technology, thus allowing feedback for future improvements.

5. Conclusion

We have described here a clinical-scale (~1 L/hr) hyperpolarizer that produces batches of spin-polarized 129Xe sufficient for clinical examination, with emphasis on specific design features and practical aspects of operation. The polarizer’s ‘open-source’ design and automation should facilitate implementation of HP 129Xe technology into clinical and pre-clinical settings. Four independent methods for characterizing the spin polarization have been utilized, and up to 90% 129Xe polarization is achieved. Efficient transfer of HP 129Xe from OP-cells into other sample containers was demonstrated, including Tedlar bags filled with up to 800 mL of polarized gas mixture for clinical imaging. The T1 of HP 129Xe was nearly 6 hr in a Tedlar bag used for the contrast agent delivery/transportation from the polarizer to a human volunteer. The ability to achieve high PXe values at high Xe densities, combined with stopped-flow operation and efficient Rb condensation / gettering, negates the usual requirement of Xe cryo-accumulation and storage, and also should provide improved polarization efficiency for quadrupolar isotopes (e.g. 83Kr and 131Xe (3,70-72)).

Supplementary Material

01

Acknowledgments

We thank B. Saam and G. Schrank for helpful conversations, and E. Koehnemann (Midrivers Glassblowing, Inc.) for expert glassblowing. We also thank John Gore for his support for polarizer construction, and S. Barcus for assistance during the early stages of polarizer construction. N.W. was supported by an NSF post-doctoral fellowship (OISE-0966393); B.M.G. and B.M.G. were supported in part by the NSF (DMR 0852004 & 1157058). This work was funded by NIH (1R01 HL096471) and SIU OSPA. M.J.B. is supported by the School of Medicine, U. Nottingham, UK. E.Y.C. thanks the support from NIH/NCI 5R00 CA134749-03 and DoD CDMRP Era of Hope Award W81XWH-12-1-0159/BC112431. A.M.C. thanks the support from training NIH grant R25 CA136440.

Abbreviations

AR

Anti-Reflection

COPD

Chronic Obstructive Pulmonary Disease

FOV

Field Of View

FRC

Functional Residual Capacity

FWHM

Full Width Half Max

GRE

Gradient Echo

GUI

Graphical User Interface

HP 129Xe

Hyperpolarized Xenon-129

I.D.

Inner Diameter

IR

Infra Red

ISRs

Input Service Routines

KOH

Potassium Hydroxide

LDA

Laser Diode Array

OP

Optical Pumping

O.D.

Outer Diameter

PBS

Polarizing Beam Splitter

PFA

Perfluoroalkoxy

PSU

Power Supply Unit

PRb

Rubidium Polarization

PXe

Xenon Polarization

QA

Quality Assurance

SAR

Specific Absorption Rate

SEOP

Spin-Exchange Optical Pumping

SNR

Signal-to-Noise Ratio

SSR

Solid-State Relay

TE

echo time

TLC

Total Lung Capacity

TR

repetition time

VHG

Volume Holographic Grating

XeNA

XEnon polarizatioN Automated

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • 1.Goodson BM. Nuclear magnetic resonance of laser-polarized noble gases in molecules, materials, and organisms. J Magn Reson. 2002;155(2):157–216. doi: 10.1006/jmre.2001.2341. [DOI] [PubMed] [Google Scholar]
  • 2.Bifone A, Cherubini A. Hyperpolarised Xenon in Biology. Prog Nucl Magn Reson Spectrosc. 2003;42(1-2):1–30. [Google Scholar]
  • 3.Lilburn D, Pavlovskaya G, Meersmann T. Perspectives of hyperpolarized noble gas MRI beyond He-3. J Magn Reson. 2013;229:173–186. doi: 10.1016/j.jmr.2012.11.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Mugler J, Altes T. Hyperpolarized 129Xe MRI of the human lung. J Magn Reson Imaging. 2013;37:313–331. doi: 10.1002/jmri.23844. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Albert M, Cates G, Driehuys B, Happer W, Saam B, Springer C, Wishnia A. Biological Magnetic Resonance Imaging Using Laser Polarized Xe-129. Nature. 1994;370:199–201. doi: 10.1038/370199a0. [DOI] [PubMed] [Google Scholar]
  • 6.Mugler JP, Driehuys B, Brookeman JR, Cates GD, Berr SS, Bryant RG, Daniel TM, deLange EE, Downs JH, Erickson CJ, Happer W, Hinton DP, Kassel NF, Maier T, Phillips CD, Saam BT, Sauer KL, Wagshul ME. MR imaging and spectroscopy using hyperpolarized Xe-129 gas: Preliminary human results. Magn Reson Med. 1997;37(6):809–815. doi: 10.1002/mrm.1910370602. [DOI] [PubMed] [Google Scholar]
  • 7.Leawoods JC, Yablonskiy DA, Saam B, Gierada DS, Conradi MS. Hyperpolarized He-3 gas production and MR imaging of the lung. Concepts Magn Reson. 2001;13(5):277–293. [Google Scholar]
  • 8.Moller HE, Chen XJ, Saam B, Hagspiel KD, Johnson GA, Altes TA, Lange EEd, Kauczor H-U. MRI of the Lungs Using Hyperpolarized Noble Gases. Magn Reson Med. 2002;47:1029–1051. doi: 10.1002/mrm.10173. [DOI] [PubMed] [Google Scholar]
  • 9.Patz S, Hersman FW, Muradian I, Hrovat MI, Ruset IC, Ketel S, Jacobson F, Topulos GP, Hatabu H, Butler JP. Hyperpolarized Xe-129 MRI: A viable functional lung imaging modality? Eur J Radiol. 2007;64(3):335–344. doi: 10.1016/j.ejrad.2007.08.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Dregely JPM, I, III, Ruset IC, Altes TA, Mata JF, Miller W, Ketel J, Ketel S, Distelbrink J, Hersman FW, Ruppert K. Hyperpolarized Xenon-129 Gas-Exchange Imaging of Lung Microstructure: First Case Studies in Subjects With Obstructive Lung Disease. J Magn Reson Imaging. 2011;33:1052–1062. doi: 10.1002/jmri.22533. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kaushik SS, Cleveland ZI, Cofer GP, Metz G, Beaver D, Nouls J, Kraft M, Auffermann W, Wolber J, McAdams HP, Driehuys B. Diffusion-Weighted Hyperpolarized 129Xe MRI in Healthy Volunteers and Subjects With Chronic Obstructive Pulmonary Disease. Magn Reson Med. 2011;65:1155–1165. doi: 10.1002/mrm.22697. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Matsuoka S, Patz S, Albert M, Sun Y, Rizi R, Gefter W, Hatabu H. Hyperpolarized Gas MR Imaging of the Lung: Current Status as a Research Tool. J Thorac Imaging. 2009;24:181–188. doi: 10.1097/RTI.0b013e3181b32bec. [DOI] [PubMed] [Google Scholar]
  • 13.Wild J, Marshall H, Xu X, Norquay G, Parnell S, Clemence M, Griffiths P, Parra-Robles J. Simultaneous Imaging of Lung Structure and Function with Triple-Nuclear Hybrid MR Imaging. Radiology. 2013;267:251–255. doi: 10.1148/radiol.12121153. [DOI] [PubMed] [Google Scholar]
  • 14.Lilburn DML, Hughes-Riley T, Six JS, Stupic KF, Shaw DE, Pavlovskaya GE, Meersmann T. Validating Excised Rodent Lungs for Functional Hyperpolarized Xenon-129 MRI. PLoS One. 2013;8(8):e73468. doi: 10.1371/journal.pone.0073468. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Kadlecek S, Mongkolwisetwara P, Xin Y, Ishii M, Profka H, Emami K, Rizi R. Regional determination of oxygen uptake in rodent lungs using hyperpolarized gas and an analytical treatment of intrapulmonary gas redistribution. NMR Biomed. 2011;24(10):1253–1263. doi: 10.1002/nbm.1685. [DOI] [PubMed] [Google Scholar]
  • 16.Chen RY, Fan FC, Kim S, Jan KM, Usami S, Chien S. Tissue-blood partition coefficient for xenon: temperature and hematocrit dependence. J Appl Physiology. 1980;49:178–183. doi: 10.1152/jappl.1980.49.2.178. [DOI] [PubMed] [Google Scholar]
  • 17.Swanson SD, Rosen MS, Coulter KP, Welsh RC, Chupp TE. Distribution and Dynamics of Laser-Polarized 129Xe Magnetization In Vivo. Magn Reson Med. 1999;42:1137–1145. doi: 10.1002/(sici)1522-2594(199912)42:6<1137::aid-mrm19>3.0.co;2-4. [DOI] [PubMed] [Google Scholar]
  • 18.Bifone A, Song YQ, Seydoux R, Taylor RE, Goodson BM, Pietrass T, Budinger TF, Navon G, Pines A. NMR of laser-polarized xenon in human blood. Proc Natl Acad Sci USA. 1996;93(23):12932–12936. doi: 10.1073/pnas.93.23.12932. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Albert M, Schepkin V, Budinger T. Measurement of Xe-129 T1 in blood to explore the feasibility of hyperpolarized Xe-129 MRI. J Comput Assist Tomog. 1995;19:975–978. doi: 10.1097/00004728-199511000-00025. [DOI] [PubMed] [Google Scholar]
  • 20.Acosta RH, Blümler P, Münnemann K, Spiess H-W. Mixture and dissolution of laser polarized noble gases: Spectroscopic and imaging applications. Prog Nucl Magn Reson Spectrosc. 2012;66:40–69. doi: 10.1016/j.pnmrs.2012.03.003. [DOI] [PubMed] [Google Scholar]
  • 21.Goodson BM. Using injectable carriers of laser-polarized noble gases for enhancing NMR and MRI. Concepts Magnetic Res. 1999;11(4):203–223. [Google Scholar]
  • 22.Venkatesh A, Zhao L, Balamore D, Jolesz F, Albert M. Evaluation of carrier agents for hyperpolarized xenon MRI. NMR Biomed. 2000;13:245–252. doi: 10.1002/1099-1492(200006)13:4<245::aid-nbm635>3.0.co;2-7. [DOI] [PubMed] [Google Scholar]
  • 23.Rubin SM, Spence MM, Goodson BM, Wemmer DE, Pines A. Evidence of nonspecific surface interactions between laser-polarized xenon and myoglobin in solution. Proc Natl Acad Sci USA. 2000;97:9472–9475. doi: 10.1073/pnas.170278897. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Jansch HJ, Gerhard P, Koch M. Xe-129 on Ir(111): NMR study of xenon on a metal single crystal surface. Proc Natl Acad Sci USA. 2004;101(38):13715–13719. doi: 10.1073/pnas.0405426101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Schröder L, Lowery TJ, Hilty C, Wemmer DE, Pines A. Molecular imaging using a targeted magnetic resonance hyperpolarized biosensor. Science. 2006;314(5798):446–449. doi: 10.1126/science.1131847. [DOI] [PubMed] [Google Scholar]
  • 26.Mazzanti M, Walvick R, Zhou X, Sun Y, Shah N, Mansour J, Gereige J, Albert M. Distribution of Hyperpolarized Xenon in the Brain Following Sensory Stimulation: Preliminary MRI Findings. PLoS One. 2011;6:e21607. doi: 10.1371/journal.pone.0021607. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Wolber J, Cherubini A, Leach M, Bifone A. Hyperpolarized Xe-129 NMR as a probe for blood oxygenation. Magn Reson Med. 2000;43:491–496. doi: 10.1002/(sici)1522-2594(200004)43:4<491::aid-mrm1>3.0.co;2-6. [DOI] [PubMed] [Google Scholar]
  • 28.Dabaghyan M, Dregely I, Muradian I, Hrovat M, Hatabu H, Patz S. Oxygen Partial Pressure and Uptake in the Lung with Hyperpolarized 129Xe MRI: Preliminary Results. Stockholm, Sweden. Proc. Intl. Soc. Magn. Reson. Med. 2010 [Google Scholar]
  • 29.Leung G, Norquay G, Wild JM. 129Xe as an in vivo probe for MR oximetry. Proc Intl Soc Mag Reson Med. 2013;21:0816. [Google Scholar]
  • 30.Shea DA, Morgan D. The Helium-3 Shortage: Supply, Demand, and Options for Congress. 2010. [Google Scholar]
  • 31.Driehuys B, Cates GD, Miron E, Sauer K, Walter DK, Happer W. High-volume production of laser-polarized Xe-129. Appl Phys Lett. 1996;69(12):1668–1670. [Google Scholar]
  • 32.Rosen MS, Chupp TE, Coulter KP, Welsh RC, Swanson SD. Polarized 129Xe optical pumping/spin exchange and delivery system for magnetic resonance spectroscopy and imaging studies. Rev Sci Instrum. 1999;70(2):1546–1552. [Google Scholar]
  • 33.Ruth U, Hof T, Schmidt J, Fick D, Jansch HJ. Production of nitrogen-free, hyperpolarized 129Xe gas. Appl Phys B. 1999;68(1):93–97. [Google Scholar]
  • 34.Zook AL, Adhyaru BB, Bowers CR. High capacity production of > 65% spin polarized xenon-129 for NMR spectroscopy and imaging. J Magn Reson. 2002;159(2):175–182. doi: 10.1016/s1090-7807(02)00030-7. [DOI] [PubMed] [Google Scholar]
  • 35.Mortuza MG, Anala S, Pavlovskaya GE, Dieken TJ, Meersmann T. Spin-exchange optical pumping of high-density xenon-129. J Chem Phys. 2003;118(4):1581–1584. [Google Scholar]
  • 36.Knagge K, Prange J, Raftery D. A continuously recirculating optical pumping apparatus for high xenon polarization and surface NMR studies. Chem Phys Lett. 2004;397:11–16. [Google Scholar]
  • 37.Ruset IC, Ketel S, Hersman FW. Optical Pumping System Design for Large Production of Hyperpolarized 129Xe. Phys Rev Lett. 2006;96(5):053002. doi: 10.1103/PhysRevLett.96.053002. [DOI] [PubMed] [Google Scholar]
  • 38.Schrank G, Ma Z, Schoeck A, Saam B. Characterization of a low-pressure high-capacity Xe-129 flow-through polarizer. Phys. Rev. A. 2009;80(6) [Google Scholar]
  • 39.Whiting N, Nikolaou P, Eschmann NA, Barlow MJ, Goodson BM. Interdependence of Xenon Density and Temperature on Rb/129Xe Optical Pumping Efficiency at High Xenon Densities. J Magn Reson. 2011;208(2):298–304. doi: 10.1016/j.jmr.2010.11.016. [DOI] [PubMed] [Google Scholar]
  • 40.Nikolaou P, Coffey A, Walkup L, Gust B, Whiting N, Newton H, Barcus S, Muradyan I, Moroz GD, Rosen M, Patz S, Barlow MJ, Chekmenev E, Goodson BM. Near-unity nuclear polarization with an open-source 129Xe hyperpolarizer for NMR and MRI. Proc Natl Acad Sci USA. 2013;110(35):14150–14155. doi: 10.1073/pnas.1306586110. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Walker T, Happer W. Spin-exchange optical pumping of noble-gas nuclei. Rev Mod Phys. 1997;69(2):629–642. [Google Scholar]
  • 42.Whiting N, Eschmann NA, Barlow MJ, Goodson BM. 129Xe/Cs (D1, D2) Versus 129Xe/Rb (D1) Spin-Exchange Optical Pumping at High Xenon Densities Using High-Power Laser Diode Arrays. Phys Rev A. 2011;83(5):053428. [Google Scholar]
  • 43.Six J, Hughes-Riley T, Stupic K, Pavlovskaya G, Meersmann T. Pathway to Cryogen Free Production of Hyperpolarized Krypton-83 and Xenon-129. PLoS One. 2012;7:e49927. doi: 10.1371/journal.pone.0049927. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Kuzma NN, Patton B, Raman K, Happer W. Fast nuclear spin relaxation in hyperpolarized solid Xe-129. Phys Rev Lett. 2002;88:147602. doi: 10.1103/PhysRevLett.88.147602. [DOI] [PubMed] [Google Scholar]
  • 45.Nikolaou P, Whiting N, Eschmann NA, Chaffee KE, Goodson BM, Barlow M. Generation of Laser-Polarized Xenon Using Fiber-Coupled Laser Diode Arrays Narrowed with Integrated Volume Holographic Gratings. J Magn Reson. 2009;197(2):249–254. doi: 10.1016/j.jmr.2008.12.015. [DOI] [PubMed] [Google Scholar]
  • 46.Whiting N, Nikolaou P, Eschmann NA, Goodson BM, Barlow MJ, Lammert R, Ungar J, Vaissie L. Using frequency-narrowed, tunable laser diode arrays with integrated volume holographic gratings for Rb/129Xe spin-exchange optical pumping at high resonant laser fluxes and high xenon densities. Appl Phys B. 2012;106(4):775–788. [Google Scholar]
  • 47.Nikolaou P, Coffey A, Walkup L, Gust B, Whiting N, Moroz GD, Rosen M, Chekmenev E, Barlow MJ, Patz S, Goodson BM. Towards the Development of an ‘Open-Source’ High Volume 129Xe/SEOP Apparatus for Potential Application in Human Lung Imaging and NMR/MRI Studies of Porous Materials. Frontiers in Biomedical Imaging. 2011;III [Google Scholar]
  • 48.Nikolaou P, Coffey A, Walkup L, Gust B, Whiting N, Newton H, Barcus S, Muradyan I, Moroz GD, Rosen M, Patz S, Barlow MJ, Chekmenev E, Goodson BM. An ‘Open-Source’ 129Xe Polarizer for Clinical Imaging, in vivo MRS/MRI, and NMR/MRI of Porous Materials. Miami, FL: 2012. [Google Scholar]
  • 49.Korchak S, Kilian W, Mitschang L. Configuration and Performance of a Mobile Xe-129 Polarizer. Appl Magn Reson. 2013;44:65–80. doi: 10.1007/s00723-012-0425-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Witte C, Kunth M, Döpfert J, Rossella F, Schröder L. Hyperpolarized Xenon for NMR and MRI Applications. J Vis Exp. 2012;67:e4268. doi: 10.3791/4268. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Hughes-Riley T, Six JS, Lilburn DML, Stupic KF, Dorkes AC, Shaw DE, Pavlovskaya GE, Meersmann T. Cryogenics free production of hyperpolarized 129Xe and 83Kr for biomedical MRI applications. J Magn Reson. 2013;237:23–33. doi: 10.1016/j.jmr.2013.09.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.XeMed LLC www.xemed.com.
  • 53.Polarean, Inc www.polarean.com.
  • 54.Toyoko Kagaku http://www.toyokokagaku.co.jp/english/product/02_02.html.
  • 55.Barker JR. New Coil Systems for the Production of Uniform Magnetic Fields. J Sci Instrum. 1949;26(8):273. [Google Scholar]
  • 56.Parnell SR, Deppe M, Parra-Robles J, Wild JM. Enhancement of 129Xe polarization by off-resonant spin exchange optical pumping. J Appl Phys. 2010;108:064908. [Google Scholar]
  • 57.Breeze SR, Lang S, Moudrakovski I, Ratcliffe CI, Ripmeester JA, Simard B, Santyr G. Coatings for optical pumping cells and extending the lifetime of hyperpolarized xenon. J Appl Phys. 1999;86(7):4040–4042. [Google Scholar]
  • 58.Breeze SR, Lang S, Moudrakovski I, Ratcliffe CI, Ripmeester JA, Santyr G, Simard B, Zuger I. Coatings for Optical Pumping Cells and Short-Term Storage of Hyperpolarized Xenon. J Appl Phys. 2000;87:8013–8017. [Google Scholar]
  • 59.Zeng X, Miron E, Van Wijngaarden WA, Schreiber D, Happer W. Wall relaxation of spin polarized 129Xe nuclei. Phys Lett A. 1983;96(4):191–194. [Google Scholar]
  • 60.Saha I, Nikolaou P, Whiting N, Goodson BM. Characterization of violet emission from Rb optical pumping cells used in laser-polarized xenon NMR experiments. Chem Phys Lett. 2006;428(4-6):268–276. [Google Scholar]
  • 61.Romalis MV, Miron E, Cates GD. Pressure broadening of Rb D-1 and D-2 lines by He-3, He-4, N-2, and Xe: Line cores and near wings. Phys Rev A. 1997;56(6):4569–4578. [Google Scholar]
  • 62.Cates GD, Benton DR, Gatzke M, Happer W, Hasson KC, Newbury NR. Laser production of large nuclear-spin polarization in frozen xenon. Phys Rev Lett. 1990;65:2591–2594. doi: 10.1103/PhysRevLett.65.2591. [DOI] [PubMed] [Google Scholar]
  • 63.Coffey AM, Shchepin RV, Wilkens K, Waddell KW, Chekmenev EY. A large volume double channel 1H-X RF probe for hyperpolarized magnetic resonance at 0.0475 T. J Magn Reson. 2012;220:94–101. doi: 10.1016/j.jmr.2012.04.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Waddell KW, Coffey AM, Chekmenev EY. In situ Detection of PHIP at 48 mT: Demonstration using a Centrally Controlled Polarizer. J Am Chem Soc. 2011;133(1):97–101. doi: 10.1021/ja108529m. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Anger BC, Schrank G, Schoeck A, Butler KA, Solum MS, Pugmire RJ, Saam B. Gas-phase spin relaxation of 129Xe. Phys Rev A. 2008;78(4):043406. [Google Scholar]
  • 66.Bouchiat MA, Brossel J, Pottier LC. Evidence for RbRareGas Molecules from the Relaxation of Polarized Rb Atoms in a Rare Gas. Experimental Results. J Chem Phys. 1972;56:3703–3714. [Google Scholar]
  • 67.Steck DA. Rb 85 & 87; Cs D-line Data. 20102008 [Google Scholar]
  • 68.Walter DK, Griffith WM, Happer W. Energy transport in high-density spin-exchange optical pumping cells. Phys Rev Lett. 2001;86:3264–3267. doi: 10.1103/PhysRevLett.86.3264. [DOI] [PubMed] [Google Scholar]
  • 69.Newton H, Walkup LL, Whiting N, West L, Carriere J, Havermeyer F, Ho L, Morris P, Goodson BM, Barlow MJ. Comparative study of in situ N2 rotational Raman spectroscopy methods for probing energy thermalisation processes during spin-exchange optical pumping. Appl Phys B. 2013 DOI 10.1007/s00340-013-5588-x. [Google Scholar]
  • 70.Cleveland ZI, Pavlovskaya GE, Stupic KF, LeNoir CF, Meersmann T. Exploring hyperpolarized 83Kr by remotely detected NMR relaxometry. J Chem Phys. 2006;124(4):044312. doi: 10.1063/1.2159493. [DOI] [PubMed] [Google Scholar]
  • 71.Stupic KF, Cleveland ZI, Pavlovskaya GE, Meersmann T. Quadrupolar Relaxation of Hyperpolarized Krypton-83 as a Probe for Surfaces. Solid State Nucl Magn Reson. 2006;29:79–84. doi: 10.1016/j.ssnmr.2005.08.008. [DOI] [PubMed] [Google Scholar]
  • 72.Stupic KF, Cleveland ZI, Pavlovskaya GE, Meersmann T. Hyperpolarized 131Xe NMR spectroscopy. J Magn Reson. 2011;208:58–69. doi: 10.1016/j.jmr.2010.10.004. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

01

RESOURCES