Abstract
PET imaging of non-human primates (NHP) requires correction of head motion if the subjects are scanned awake and their heads are unrestrained, as the NHPs move their heads faster and more frequently than human subjects. This work focuses on designing and validating two motion correction algorithms for awake NHP brain PET imaging.
Methods
Two motion correction methods were implemented for awake NHP brain PET imaging: multi-acquisition frame (MAF) and event-by-event (EBE). Motion data were acquired from an external motion tracking device. The MAF method divides scan data into short subframes, reconstructs each subframe individually and registers them to a reference orientation. This method suffers from residual intra-frame motion and data loss when motion is large as a minimum frame duration is often required. The EBE method, previously implemented for the human brain scanner HRRT and adapted for the microPET FOCUS-220 in this work, eliminates intra-frame motion and should have the best accuracy. We first evaluated the accuracy of both motion correction methods with moving phantom scans on the FOCUS-220. Both motion correction methods were then applied to awake NHP brain PET studies with a GABAA-benzodiazepine receptor ligand [11C]flumazenil, and the reconstructed images were compared with those from a motion-free anesthetized study.
Results
The phantom studies showed that EBE motion correction similarly recovers the contrast (within 3%) as the static study, whereas MAF motion correction using the standard algorithm setting showed 25% reduction in contrast from the static case. In awake NHP brain PET imaging, EBE motion correction better recovers the fine structures than the MAF method, as compared with anesthetized studies.
Conclusions
The large magnitude and frequency of NHP head motion suggests that EBE motion correction with accurate externally measured motion data can noticeably alleviate image blurring due to the intra-frame motion in the MAF motion correction method.
Keywords: PET, motion correction, awake, monkey
INTRODUCTION
PET imaging of non-human primates (NHP) is an important approach in understanding brain function and in evaluating the mechanisms of action of pharmacological agents (1, 2). The vast majority of NHP brain PET studies are conducted under anesthesia for obvious practical reasons including the elimination of head motion. With a high-resolution animal scanner, researchers can accurately measure the activity in small brain regions in NHP in the anesthetized state. However, anesthesia has been shown to have significant effects on PET metabolic and receptor binding measures (3, 4). If scanned awake and unrestrained, head motion would cause image blurring and inaccurate activity quantification. Alternatively, head fixation has been employed in a few centers (5, 6). This method is likely to create stress in the subjects (7) and requires extensive acclimation and training. Therefore, motion correction is essential to remove image blurring in awake NHP brain PET studies.
Various motion correction methods have been proposed and applied in PET. Picard and Thompson (8) proposed a multiple-acquisition frame (MAF) method, in which a new acquisition frame is started when the motion exceeds a threshold. This method has been implemented for small animal (9) and NHP PET imaging (10). However, the high speed and frequency of NHP head motion may introduce undesirable image blurring due to the intra-frame motion, uncorrected by the MAF method. In theory, event-by-event (EBE) motion correction has the potential for the highest accuracy, as each event is relocated to the LOR where it belongs. This method has been developed in the Motion-compensation OSEM List-mode Algorithm for Resolution-recovery reconstruction (MOLAR) (11) for the HRRT (12) and the biograph mCT (13).
The goal of this study was to develop and evaluate the accuracy of both MAF and EBE motion correction methods for awake NHP PET imaging on the microPET FOCUS-220 (Siemens Medical Solutions, Knoxville, USA). Motion data were acquired with the Vicra system (NDI, Waterloo, Canada) with a time synchronization technique to align the PET and the motion data. For the MAF method, the raw list-mode data are divided into subframes based on an intra-frame motion threshold (IFMT) and a minimum frame duration threshold (MFDT). For EBE motion correction, we adapted the MOLAR algorithm for the FOCUS-220. The modified version retains the theoretical framework of MOLAR, with major adjustments to accommodate the scanner geometry and normalization procedure. The accuracy of both motion correction algorithms was first validated with moving phantoms, and was then applied to awake NHP PET studies.
MATERIALS AND METHODS
The microPET FOCUS-220 Scanner
The microPET FOCUS-220 (14) is a high resolution PET scanner for imaging NHP and small animals. It consists of 168 detector blocks, each of which is divided into 12×12 lutetium oxyorthosilicate (LSO) crystals, which measure 1.51 mm × 1.51 mm × 10 mm. The axial field-of-view (FOV) is 7.6 cm, and transverse FOV is 19 cm. Scan data are acquired in list-mode as 48-bit binary data packets.
Motion Tracking
Subject motion was recorded with a Vicra optical tracking system controlled by a dedicated PC (Vicra PC). The Vicra employs infrared illuminators and stereo cameras to sense 3D positions of reflective spheres, which are mounted on a head tool. Multiple tools can be tracked simultaneously at 20 Hz. Tool position and orientation can be calculated relative to the Vicra’s internal coordinate system. To allow motion data to be independent of possible Vicra sensor vibration or movement, a reference tool is permanently attached to the scanner’s frame. A onetime calibration establishes the relationship between the scanner and reference tool coordinate systems (15). Using the measured reference tool to scanner relationship, post processing software calculates head tool motion relative to the scanner coordinate system. Ultimately, a set of time stamped affine transformation matrices are calculated and used to correct PET event data back to some strategically selected reference location.
Time Synchronization between PET Data and Motion Data
Precise synchronization of the motion data to the PET list-mode data is necessary, as awake NHP subjects exhibit frequent and rapid head movements. To synchronize the two data streams, a time injection algorithm was developed that periodically encodes and sends the absolute clock time of the Vicra PC into the PET list-mode data stream via a scanner physiological gating input. The algorithm and a validation scan are provided in the supplemental data.
MAF-based Motion Correction
The MAF method is summarized in Fig. 1. First, the list-mode data were divided into subframes based on the measured motion (step 1), using an IFMT of, e.g., 1 or 2 mm such that a new subframe begins when the motion magnitude, M, exceeds the IFMT. To calculate M, eight points were selected as the vertices of a virtual rectangular box (10 cm × 10 cm × 6 cm) centered in the scanner FOV, and the motion magnitude, m, for each point is calculated using Eq. 1.
Eq. 1 |
For each point, its transformed position is calculated using the motion transformation matrix at 20 Hz, and the standard deviations of the resulting x, y, and z coordinates are σx, σy, and σz. This definition of m was chosen because, if the subject stays at one location for the first half of the frame, and moves a distance, m, away for the second half, the motion magnitude within this frame is m. Within each subframe, the average motion of the eight points is M, which is continuously updated for each new motion data to determine if the IFMT has been exceeded.
In addition, an MFDT of, e.g., 2~4 seconds, is set to limit the computation demands of the algorithm. Specifically, when the intra-frame motion of a subframe exceeds the IFMT, the duration of the subframe is compared with the MFDT. The subframe is kept if the frame duration exceeds the MFDT. Otherwise, the subframe is discarded. This results in loss of data when the subject moves rapidly within short intervals.
Each subframe is reconstructed using FBP with a Ramp filter with cutoff at the Nyquist frequency (step 2). This first reconstruction includes normalization, randoms and decay correction, but ignores attenuation and scatter corrections, since an attenuation image at the orientation of this subframe is not available at this point. Next, the reconstructed image of each subframe is transformed to a reference orientation by the average motion data within that subframe (step 3). The average transformation matrix for each subframe is computed by averaging the quaternion components of all the transformation matrices with equal weight within the subframe. These images are summed together, and the transmission image of the same subject from an anesthetized study is mapped into the awake reference orientation by registering summed emission images of the anesthetized and awake studies to the MR image (10). The transmission image is then resliced to the orientation of each subframe with the inverse of the transformation matrices from step 2 (step 4). Each subframe is then re-reconstructed with attenuation and scatter corrections, and transformed to the reference orientation. Adjacent subframes are re-grouped into 5-min frames by summing the subframe images, weighted by their durations (step 5). The effective frame duration of each grouped frame is the sum of the subframe durations, and the effective frame mid-time is the weighted average of the subframes mid-times. Finally, all grouped images are decay corrected to the injection time for quantitative analyses.
Event-by-event Motion Correction
List-mode reconstruction with EBE motion correction was performed with MOLAR, an OP-OSEM algorithm:
Eq. 2 |
Each time frame of duration T (sec) is divided into nT sub-bins of duration Δt in seconds. k represents the index of each detected event, ik represents the LOR index of event k, tk represents the time stamp for event k. ci,t,j, the system matrix, represents the contribution of voxel j to LOR i in time bin t, accounting for geometry, resolution, solid angle, and motion effects. A spatially invariant Gaussian point-spread function (PSF) of 1.5mm FWHM was used. Li,t is the dimensionless product of decay factor at time t, livetime at time t, and positron branching fraction. Ai,t is the dimensionless attenuation factor. Ni is a sensitivity (normalization) factor, in units of (counts/sec)/(Bq/mL×mm), which converts the forward projection through the image grid λ (Bq/mL) to units of counts/sec. Normalization was performed with a component-based model that includes detector efficiencies, trans-axial and axial geometric effects, detector interference effects and parallax effects (16). Ri,t is the random coincidence rate in counts per second, estimated from the singles rates of the each detector pair. Si,t is the scattered coincidence rate in counts per second, estimated using the single scatter simulation model (17). When motion data are available, EBE motion correction transforms the locations of the endpoints of the LOR of each detected event to the reference position. Attenuation (A), scatter (S) and the system matrix (c) use the motion-corrected LOR, as denoted by the tilde sign (~) in Eq. 2. For example, Ãik, tk represents the attenuation factor for the motion-corrected event k detected on LOR ik at time tk. The reference position motion correction is defined as the orientation of the head during the transmission scan. Therefore, motion correction maps each LOR back to where the head was during the transmission scan, thus correctly aligning the attenuation map to the emission scan. In this way, no special processing of transmission data is required. Scatter is estimated iteratively during the iteration process based on the image at that iteration/subset. Every event is motion-corrected prior to the scatter estimation step. Therefore, scatter estimation uses all the motion-corrected events. A sensitivity image Q, in units of counts per second/(Bq/mL), is calculated uniquely for each frame. In principle, the summation in Eq. 2 occurs over all LORs i over all time periods t in the time frame. In MOLAR, Q is approximated by a random and unique sampling of LORs for each reconstruction, to avoid the calculation for all possible LORs, while accounting for the unique motion in each frame by repositioning the randomly sampled LORs (15). The final image sensitivity values are scaled to correct for the undersampling.
The list-mode data were divided into 30 subsets, with subsets defined based on the order of arrival of each event. Images were reconstructed for 4 iterations for the phantom studies, and 2 iterations for the awake NHP studies to control noise in the low-count NHP studies. The initial value of λ is uniform (8,500 Bq/mL) within the attenuating object and 0 outside it.
Moving Phantom Scan
In order to validate the motion correction algorithm, a moving mini-Derenzo phantom (Data Spectrum, Hillsborough, USA) was scanned and reconstructed with each motion correction algorithm. The diameters of the rod structures in the phantom are 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm. The phantom was filled with 37 MBq of [18F]FDG, and was positioned at 4 cm off the center of the transaxial FOV. A transmission scan was first performed to generate an attenuation map. A 5-min static scan was acquired, followed by a 5-min dynamic scan, in which the phantom was moved manually. Reflective markers were attached to the phantom and motion data were recorded by the Vicra system. The motion magnitude was calculated by dividing the scan data into 1-sec subframes, and calculating the displacement within each second. In this moving phantom scan, the average speed of motion was 1.4 mm/sec, and the motion magnitude in the entire scan was 48.2 mm, moderately larger than that of an NHP head motion in an awake study.
Awake NHP Brain Scans
Studies were performed in rhesus monkeys under a protocol approved by the Yale Institutional Animal Care and Use Committee. Awake rhesus monkeys were scanned in the FOCUS-220 fitted with a lifter-tilter (Agile Technologies, Knoxville, USA) (Supplemental Fig. 4), a mechanical device that lifts and rotate the scanner. The monkeys were trained to sit in a custom chair and tilted back ~35°, such that the long axis of their brain was centered in the scanner’s axial FOV when the scanner was tilted forward ~45°. Initial studies were performed using bolus/infusion administration of 252 MBq of the GABAA-benzodiazepine ligand [11C]flumazenil. List-mode data were acquired beginning 30 minutes post-injection for 30 minutes. The motion tracking tool was mounted onto a silicon rubber plug, which was adhered to the scalp via skin glue. The median speeds of motion for the awake NHP studies range from 1.0 to 2.4 mm/sec across studies. For list-mode reconstruction with MOLAR, one 30-min frame was reconstructed. The equivalent motion magnitude M during the 30-min scan was 34±6 mm for the five awake NHP studies. All PET images were registered to an NHP template via a linear transformation from the PET to the MR image space for the same monkey, and a nonlinear transformation from the MR space to the NHP template.
For a 30-min awake NHP study of 30M events, the MAF method involves reconstructing ~ 100 subframes twice, pre- and post-processing the subframes. The total computation time is ~ 10 hours (CPU speed = 3.0GHz). The EBE method based on the MOLAR platform takes ~ 6 hours (~1 hour for each 5-min frame) on 16 cluster nodes (CPU speed = 3200MHz).
RESULTS
Mini-Derenzo Phantom Scan
Fig. 2 shows the reconstructed images of the static and the moving mini-Derenzo phantom using both motion correction methods with time synchronization between the scan data and the motion data. All images are displayed in the transaxial view, and the image intensities are displayed to a global scale. In this implementation, with the phantom positioned 4-cm off center, the 1.6-mm rods were not clearly resolvable with OP-OSEM, even for static scans using 4 iterations and 30 subsets. These rods can be resolved by additional iterations (10) when the phantom is positioned at the center of the FOV (data not shown). MAF motion correction (Fig. 2B) with 1-mm IFMT and 3-sec MFDT corrects most of the motion, as compared to Fig. 2A. However, this frame division setting keeps only 31% of the events. Residual intra-frame motion causes slight blurring of the rod structures and reduced intensities in the rods (arrows), as compared with Fig. 2C, in which the same FBP reconstruction algorithm was performed on a static scan with the same number of events as Fig. 2B. EBE motion correction (Fig. 2E) used all of the counts, and gives comparable image quality as the static scan (Fig. 2F).
The contrast recovery coefficient (CRC), here defined as (Hot-Cold)/Hot, was used to examine the image resolution of the mini-Derenzo phantom. As shown in Fig. 3, ROIs were defined on Fig. 2C. The hot ROIs were defined on the 2.4 mm and the 3.2 mm diameter rods. For the 2.4 mm group, each hot rod contains 5 pixels of 0.95 mm × 0.95 mm. The elliptical cold ROIs, defined in the space between the hot ROIs, contain 4 pixels each. For the 3.2 mm diameter rods, each hot rod contains 9 pixels, and each cold ROI contains 8 pixels. The intensity of the hot and the cold region was the average intensity within the hot and the cold regions, respectively. The CRC was calculated from the images on Figs. 2B, C, E and F, and is shown in Table 1. For the 3.2 mm group, the CRC for the MAF method using 1-mm IMFT (Fig. 2B) was 0.77. For the static phantom (Fig. 2C), the CRC for the image reconstructed with FBP was 0.84, 9% higher than MAF using 1-mm IFMT, due to the intra-frame motion. The CRC for EBE motion correction (Fig. 2E) was 0.85, within 3% from the CRC for the static phantom reconstructed with MOLAR (0.87) (Fig. 2F). Fig. 2F gives higher CRC than Fig. 2C due to the resolution recovery PSF in MOLAR. The slight improvement in CRC between 2E and 2F is likely due to the uncertainty in the motion data (see Discussion). For 2-mm IFMT, the CRC for the 3.2 mm rods was 0.63, a 25% reduction from the static case. For 3-mm IFMT, the CRC dropped by 38% from the static case. Similar trend was observed for the 2.4 mm rods. Varying the MFDT did not affect the CRC values, as expected, since the MFDT affects image noise instead of resolution.
TABLE 1.
Rod diameter (mm) | Static (OSEM) | EBE | Static (FBP) | MAF (1mm IFMT) | MAF (2mm IFMT) | MAF (3mm IFMT) |
---|---|---|---|---|---|---|
2.4 | 0.74 | 0.72 | 0.63 | 0.57 | 0.55 | 0.50 |
3.2 | 0.87 | 0.85 | 0.84 | 0.77 | 0.63 | 0.52 |
Awake NHP Studies
Fig. 4 shows the reconstructed images of an awake (left and center columns) and an anesthetized NHP study (right column) in the same animal. The motion magnitude for the awake study was 1.3 mm/sec. The top row shows the images reconstructed with FBP with a Ramp filter at Nyquist frequency, and the bottom row shows the respective images reconstructed with MOLAR in the same study as the top row. Without motion correction (Figs. 4A and 4D), substantial blurring is observed. MAF motion correction (Fig. 4B) was able to correct for most of the motion, although some blurring is present in the fine cortical regions (arrows) when compared with an anesthetized scan of the same counts reconstructed with the same algorithm (Fig. 4C). MAF included 2-mm intra-frame motion with 3-sec MFDT, which retained 57% of the data. Using this setting, the subframe durations are 8.6 ± 8.8 sec for five awake NHP studies. EBE motion correction (Fig. 4E) gave comparable contrast as the anesthetized study (Fig. 4F) in the fine cortical regions indicated by the arrow. For MAF (Fig. 4B), some blurring is observed in the same cortical regions when compared with the anesthetized study (Fig. 4C). The noise level for EBE is improved over MAF as no counts are discarded for the EBE method.
DISCUSSION
In this study, we designed and evaluated the quality of MAF and EBE motion correction methods for awake NHP studies on the FOCUS-220. The image resolution of MAF depends on the residual motion allowed within each subframe. A low IFMT reduces residual motion, at the price of low statistics in each subframe. Also, the images from the subframes are transformed to a reference orientation, which may create additional image blurring due to interpolation of image pixels. In comparison, EBE motion correction uses all the counts without introducing image blurring caused by intra-frame motion. Compared to another EBE algorithm based on LOR rebinning (18), in MOLAR, the coordinates of each motion-corrected LOR are used in the reconstruction without binning into sinogram bins, which eliminates the loss of resolution due to the LOR rebinning.
The speed of motion is substantially higher in awake, unrestrained NHP head studies than in human brain imaging. The head motion of over 500 human brain studies (healthy controls and numerous patient populations), scanned on the human brain PET scanner HRRT, was also tracked with the Vicra camera. The fraction of instantaneous velocity of the head that is greater than 1 mm/s is ~ 0.8% in human studies, compared with ~ 16% in the five awake NHP brain studies. For instantaneous velocity greater than 5 mm/s, the fraction is ~ 0.02% in human studies, compared with ~ 2.3% in awake NHP brain imaging. In addition, the system resolution of the FOCUS-220 is 1.5 mm, better than the resolution of the HRRT (2.5 mm or higher). Therefore, while the MAF motion correction may be suitable for human brain PET imaging (8, 19, 20), EBE motion correction is superior in high resolution brain PET imaging of awake NHPs.
In MAF, the IFMT defines the permitted intra-frame motion and the MFDT controls the computation time by avoiding too many short subframes. The relationship between these two parameters and the fraction of data retained is shown in Table 2, based on five awake NHP studies. Little data are kept if the IFMT is less than 1 mm, due to the intrinsic uncertainty (~ 0.4 mm RMS) of the Vicra system. Reducing the MFDT retains more counts, but is computationally expensive. In a typical awake NHP study, reducing the MFDT from 4 sec to 2 sec increases the number of subframes from 76 to 133 for a 30-min acquisition. To balance the trade-off between resolution and noise, the default setting was 2 mm IFMT and 3 sec MFDT. Note that loss of counts may be acceptable in preclinical studies, if the injected dose can be increased.
TABLE 2.
Intraframe motion threshold (mm) | Minimum frame duration threshold (sec) | ||||
---|---|---|---|---|---|
1 | 2 | 3 | 4 | 5 | |
1 | 43% | 32% | 26% | 20% | 17% |
2 | 69% | 63% | 57% | 53% | 49% |
3 | 76% | 71% | 67% | 64% | 61% |
For MAF motion correction, the subframes were reconstructed with FBP to avoid possible low-count bias in the OSEM algorithm (21, 22). If no scan data were discarded, the MAF method would generate ~1,700 subframes for a 30-min study. The average subframe duration would be ~ 1 sec. In the awake NHP studies, the average count rate was ~ 40,000/sec. As such, 59% of the subframes would be shorter than 0.25 sec, and have less than 10,000 events, which is in the range of low-count bias in OSEM reconstructions (23). The choice of the reconstruction algorithm also affected the image resolution for MAF, in addition to intra-frame motion. This is because no spatial resolution recovery kernel is applied in the FBP reconstruction, whereas the MOLAR OSEM reconstruction uses a resolution recovering PSF. This can be seen in Fig. 2, in which the static phantom reconstructed with FBP (Fig. 2C) has lower CRC than MOLAR (Fig. 2F).
In our implementation of the MAF method, a new frame was started when the intra-frame motion exceeds a threshold M. More ideally, the criteria to start a new subframe might include a threshold for large step motion, which indicates the start of a new subframe. To calculate the motion transformation matrix for each subframe, we averaged the rotational components of the transformation by calculating the mean of the quaternion angles, and took the arithmetic mean of the translational components. An alternative procedure has been reported to estimate the average transformation matrix using Karcher’s weighted-mean transformation (24).
In the awake NHP scans, we affixed the reflective markers to a rigid tool, which was mounted onto a silicon rubber plug and adhered to the scalp via skin glue. We have observed cases where the reflective markers move on the loose skin of the NHP head. It is therefore necessary to avoid the forehead of the NHP when affixing the marker tool. Attaching the head tool to the rear of the NHP head may move the markers outside the FOV of the Vicra camera, resulting in missing motion data. In practice, using the lifter-tilter, we positioned the scanner such that the subject head was well centered in the scanner FOV, and adjusted the angle of the Vicra camera to include the largest possible range of detection. In this way, missing motion data were reduced to 0.1% in 4 studies.
Accurate time synchronization between the motion data and the scan data is necessary due to the high-speed head motion of NHP. Given the 1.5 mm intrinsic resolution of the FOCUS-220, 1 mm of motion may introduce noticeable image blurring. Our time synchronization method can reduce the temporal uncertainty in the motion data to 17 msec. To generate 1 mm of image blurring, the subject would need to move 60 mm/sec. In our awake NHP studies, the top speed (95th percentile) of motion ranged from 12 to 23 mm/sec, substantially below the 60 mm/sec threshold for noticeable resolution degradation. Therefore, an uncertainty of 17 msec in time should not affect the image resolution in this study.
Degradation in the spatial resolution was observed when the mini-Derenzo phantom was scanned at 4 cm off the center of the transaxial FOV in Fig. 2. The resolution degradation occurs primarily in the radial direction due to the uncertainty in the depth of interaction of the gamma photons with the detector crystals. In Fig. 2F, the largest (4.8 mm) rod became elliptical with the long axis in the radial direction. The effect of resolution degradation is more substantial in awake NHP PET imaging than human brain imaging, both because of larger motion and the reduced FOV of the FOCUS-220 compared to human brain PET scanners. This effect may be alleviated with a spatially dependent resolution model. A probability density function model that is particularly suited for list-mode reconstruction has been proposed by Yao et al (25). Initial implementation of the spatially dependent resolution model for the FOCUS-220 has been performed and additional validation is underway.
A quantitative assessment of the image quality would be helpful in comparing images to a reference image. In this work, two anesthetized images (FBP and OSEM) were used as reference images. However, the noise properties of FBP and OSEM images differ, making a numeric comparison difficult. A quantitative assessment to one reference image would enable a more definitive comparison, which will be our future work.
CONCLUSIONS
In this study, both MAF and EBE motion correction methods for awake NHP brain PET imaging were developed and evaluated. Moving phantom scans and awake NHP brain PET studies demonstrated that EBE motion correction can reduce image blurring caused by residual intra-frame motion in MAF. MAF discards a substantial amount of scan data, due to the rapid motion of the subject, based on the selection of minimum subframe duration. This leads to lower statistics than the EBE method. Future work will focus on the validation of a spatially variant resolution model to achieve uniform resolution throughout the FOV.
Supplementary Material
Acknowledgments
The authors thank the NHP team at the Yale-PET center and the laboratory of Drs. Graham Williams and Stacy Castner for performing the awake NHP studies that formed the basis of this work. The authors also thank Zhongdong Sun for programming support. This work was supported by Grant Number R01NS058360 from the National Institute of Neurological Disorders and Stroke. This publication was also made possible by CTSA Grant Number UL1 RR024139 from the National Center for Research Resources (NCRR) and the National Center for Advancing Translational Science (NCATS), components of the National Institutes of Health (NIH), and NIH roadmap for Medical Research. Its contents are solely the responsibility of the authors and do not necessarily represent the official view of NIH.
References
- 1.Meguro K, Blaizot X, Kondoh Y, Le Mestric C, Baron J, Chavoix C. Neocortical and hippocampal glucose hypometabolism following neurotoxic lesions of the entorhinal and perirhinal cortices in the non-human primate as shown by PET Implications for Alzheimer’s disease. Brain. 1999;122:1519–1531. doi: 10.1093/brain/122.8.1519. [DOI] [PubMed] [Google Scholar]
- 2.Nader MA, Morgan D, Gage HD, et al. PET imaging of dopamine D2 receptors during chronic cocaine self-administration in monkeys. Nat Neurosci. 2006;9:1050–1056. doi: 10.1038/nn1737. [DOI] [PubMed] [Google Scholar]
- 3.Bonhomme V, Fiset P, Meuret P, et al. Propofol anesthesia and cerebral blood flow changes elicited by vibrotactile stimulation: a positron emission tomography study. J Neurophysiol. 2001;85:1299–1308. doi: 10.1152/jn.2001.85.3.1299. [DOI] [PubMed] [Google Scholar]
- 4.Tsukada H, Nishiyama S, Kakiuchi T, et al. Isoflurane anesthesia enhances the inhibitory effects of cocaine and GBR12909 on dopamine transporter: PET studies in combination with microdialysis in the monkey brain. Brain Res. 1999;849:85–96. doi: 10.1016/s0006-8993(99)02018-1. [DOI] [PubMed] [Google Scholar]
- 5.Onoe H, Inoue O, Suzuki K, et al. Ketamine increases the striatal N-[11C]methylspiperone binding in vivo: positron emission tomography study using conscious rhesus monkey. Brain Res. 1994;663:191–198. doi: 10.1016/0006-8993(94)91263-7. [DOI] [PubMed] [Google Scholar]
- 6.Isoda M, Tsutsui K, Katsuyama N, et al. Design of a head fixation device for experiments in behaving monkeys. J Neurosci Methods. 2005;141:277–282. doi: 10.1016/j.jneumeth.2004.07.003. [DOI] [PubMed] [Google Scholar]
- 7.Arnsten AF. Stress impairs prefrontal cortical function in rats and monkeys: role of dopamine D1 and norepinephrine alpha-1 receptor mechanisms. Prog Brain Res. 2000;126:183–192. doi: 10.1016/S0079-6123(00)26014-7. [DOI] [PubMed] [Google Scholar]
- 8.Picard Y, Thompson CJ. Motion correction of PET images using multiple acquisition frames. IEEE Trans Med Imag. 1997;16:137–144. doi: 10.1109/42.563659. [DOI] [PubMed] [Google Scholar]
- 9.Kyme AZ, Zhou VW, Meikle SR, Fulton RR. Real-time 3D motion tracking for small animal brain PET. Phys Med Biol. 2008;53:2651–2666. doi: 10.1088/0031-9155/53/10/014. [DOI] [PubMed] [Google Scholar]
- 10.Sandiego CM, Jin X, Mulnix T, et al. Awake nonhuman primate brain PET imaging with minimal head restraint: Evaluation of GABAA Benzodiazepine binding with [11C]Flumazenil in awake and anesthetized animals. J Nucl Med. 2013 doi: 10.2967/jnumed.113.122077. In press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Carson RE, Barker W, Liow J-S, Adler S, Johnson C. Design of a motion-compensation OSEM List-mode Algorithm for Resolution-Recovery Reconstruction of the HRRT. IEEE Nucl Sci Symp Conf Rec. 2003:M16–6. [Google Scholar]
- 12.de Jong HW, van Velden FH, Kloet RW, Buijs FL, Boellaard R, Lammertsma AA. Performance evaluation of the ECAT HRRT: an LSO-LYSO double layer high resolution, high sensitivity scanner. Phys Med Biol. 2007;52:1505–1526. doi: 10.1088/0031-9155/52/5/019. [DOI] [PubMed] [Google Scholar]
- 13.Jin X, Chan C, Mulnix T, et al. List-mode reconstruction for the Biograph mCT with physics modeling and event-by-event motion correction. Physics in medicine and biology. 2013;58:5567–5591. doi: 10.1088/0031-9155/58/16/5567. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Tai YC, Ruangma A, Rowland D, et al. Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. J Nucl Med. 2005;46:455–463. [PubMed] [Google Scholar]
- 15.Jin X, Mulnix T, Gallezot J-D, Carson RE. Evaluation of motion correction methods in human brain PET imaging - A simulation study based on human motion data. Med Phys. 2013;40 doi: 10.1118/1.4819820. In press. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Casey ME, Gadagkar H, Newport D. A component based method for normalization in volume PET. In: Grangeat P, Amans J-L, editors. Third International Conference on Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine (Computational Imaging and Vision); New York, NY: Kluwer Academic Publisher; 1996. pp. 67–71. [Google Scholar]
- 17.Watson CC, Newport D, Casey ME. A single scatter simulation technique for scatter correction in 3D PET. In: PG, J-LA, editors. Computational Imaging and Vision. Dordrecht, The Netherlands: Kluwer Academic; 1996. pp. 255–268. [Google Scholar]
- 18.Fulton R, Nickel I, Tellmann L, Meikle SR, Pietrzyk U, Herzog H. Event-by-event motion compensation in 3D PET. IEEE NSS/MIC Conference Record. 2003;5:3286–3289. [Google Scholar]
- 19.Costes N, Dagher A, Larcher K, Evans AC, Collins DL, Reilhac A. Motion correction of multi-frame PET data in neuroreceptor mapping: simulation based validation. Neuroimage. 2009;47:1496–1505. doi: 10.1016/j.neuroimage.2009.05.052. [DOI] [PubMed] [Google Scholar]
- 20.Montgomery AJ, Thielemans K, Mehta MA, Turkheimer F, Mustafovic S, Grasby PM. Correction of head movement on PET studies: comparison of methods. J Nucl Med. 2006;47:1936–1944. [PubMed] [Google Scholar]
- 21.Reilhac A, Tomei S, Buvat I, Michel C, Keheren F, Costes N. Simulation-based evaluation of OSEM iterative reconstruction methods in dynamic brain PET studies. Neuroimage. 2008;39:359–368. doi: 10.1016/j.neuroimage.2007.07.038. [DOI] [PubMed] [Google Scholar]
- 22.van Velden FH, Kloet RW, van Berckel BN, Wolfensberger SP, Lammertsma AA, Boellaard R. Comparison of 3D-OP-OSEM and 3D-FBP reconstruction algorithms for High-Resolution Research Tomograph studies: effects of randoms estimation methods. Phys Med Biol. 2008;53:3217–3230. doi: 10.1088/0031-9155/53/12/010. [DOI] [PubMed] [Google Scholar]
- 23.Planeta-Wilson B, Yan J, Mulnix T, Carson RE. Quantitative accuracy of HRRT list-mode reconstructions: effect of low statistics. Paper presented at: Nuclear Science Symposium Conference Record, 2008. NSS’08. IEEE; 2008; [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Mohy-ud-Din H, Karakatsanis NA, Ay MR, Endres CJ, Wong DF, Rahmim A. Generalized Inter-Frame and Intra-Frame Motion Correction in PET Imaging - A Simulation Study. IEEE NSS/MIC Conference Record; 2011. [Google Scholar]
- 25.Yao R, Ramachandra RM, Mahajan N, et al. Assessment of a three-dimensional line-of-response probability density function system matrix for PET. Physics in medicine and biology. 2012;57:6827–6848. doi: 10.1088/0031-9155/57/21/6827. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.