Abstract
Background
An intriguing potential clinical use of cerebral oximeter measurements (SctO2) is the ability to noninvasively estimate jugular bulb venous oxygen saturation (SjvO2). Our purpose in this study was to determine the accuracy of the FORE-SIGHT® (CAS Medical Systems; Branford, CT), which is calibrated to a weighted average of 70% (SjvO2) and 30% arterial saturation, for Food and Drug Administration pre-market approval 510 (k) certification by adapting an industry standard protocol, ISO 9919:2005 [www.ISO.org] (used for pulse oximeters) and to evaluate the use of SctO2 and SpO2 measurements to noninvasively estimate jugular venous oxygen saturation (SnvO2).
Methods
Paired blood gas samples from the radial artery and the jugular venous bulb were collected from 20 healthy volunteers undergoing progressive oxygen desaturation from 100 to 70%. The blood sample pairs were analyzed via co-oximetry and used to calculate the approximate mixed vascular cerebral blood oxygen saturation, or reference SctO2 values (refSctO2), during increasing hypoxia. These reference values were compared to bilateral FORE-SIGHT SctO2 values recorded simultaneously with the blood gas draws to determine its accuracy. Bilateral SctO2 and SpO2 measurements were then used to calculate SnvO2 values which were compared to SjvO2.
Results
Two hundred forty-six arterial and 253 venous samples from 18 subjects were used in the analysis. The ipsilateral FORE-SIGHT SctO2 values showed a tolerance interval (TI) of [−10.72 10.90] Lin’s concordance correlation coefficient (CCC) with standard error (SE) of 0.83 ± 0.073 with the refSctO2 values calculated using arterial and venous blood gases. The combined data had a CCC of 0. 81 + 0.059 with TI of [−9.22 9.40] with overall bias was 0.09% and amplitude of the root mean square of error after it was corrected with random effects analysis was 2.92%. The bias and variability values between the ipsilateral and the contralateral FORE-SIGHT SctO2 measurements varied from person to person. The SnvO2 calculated from the ipsilateral SctO2 and SpO2 data showed a CCC + SE of 0.79 ± 0.088, TI = [−14.93 15.33], slope of 0.98, Y-Intercept of 1.14%) with SjvO2 values with a bias of 0.20% and an Arms of 4.08%. The SnvO2 values calculated independently from contralateral forehead FORE-SIGHT SctO2 values were not as correlated with the SjvO2 values (contralateral side CCC + SE = 0.72 ± 0.118, TI = [−14.86 15.20], slope of 0.66 and y-intercept of 20.36%).
Conclusions
The FORE-SIGHT cerebral oximeter was able to estimate oxygen saturation within the tissues of the frontal lobe under conditions of normocapnia and varying degrees of hypoxia (with 95% confidence interval of [−5.60 5.78] with ipsilateral blood ample data). These findings from healthy volunteers also suggest that the use of the calculated SnvO2 derived from SctO2 and SpO2 values may be a reasonable noninvasive method of estimating SjvO2 and therefore global cerebral oxygen consumption in the clinical setting. Further laboratory and clinical research is required to define the clinical utility of near-infrared spectroscopy determination of SctO2 and SnvO2 in the operating room setting.
Introduction
Cerebral oximetry using near-infrared spectroscopy (NIRS) is a continuous, noninvasive, optical-based method of measurement used to estimate cerebral tissue oxygen saturation (SctO2). NIRS devices, including both cerebral and conventional pulse oximeters, calculate oxy- and deoxyhemoglobin concentration by measuring the absorbance of light at specific wavelengths.1 However, cerebral oximeters do not preferentially measure oxygen saturation in pulsatile blood flow. Instead, cerebral NIRS devices estimate SctO2 by measuring oxy- and deoxyhemoglobin in arterioles, capillaries and venules in intracerebral tissue. The device interrogates a region of cerebral tissue approximately 1.5 cm below the sensor (half the distance between the transmitter and the receiver) and provides a weighted measure of hemoglobin changes in the arterial, capillary, and venous compartments, which makes this a regional cerebral tissue saturation monitor. As opposed to arterial saturation determined from conventional pulse oximetry (a measurement made from the changes in absorbance due to changes in concentration at the apex and the nadir of pulsatile flow between the transmitter and the receiver), cerebral oximetry may allow for more clinically nuanced information about cerebral oxygen supply and demand, which can be gleaned by examining the venous component of the cerebral blood flow (CBF) by mathematically manipulating SctO2 with SpO2.
The United States Food and Drug Administration requires cerebral oximeters meet parts of the International Organization for Standards (ISO) # 9919:2005 performance standards established for pulse oximeters when requesting pre-market 510(k) clearance. This ISO standard calls for a root-mean-square difference (measure of accuracy) of less than or equal to 4.0 % over a range of 70 to 100 % arterial blood oxygen saturation. While these standards are for pulse oximeters, they are generally adopted for the purpose of evaluating cerebral oximeters. The specifications also recommend stepped reductions in inspired oxygen concentration as a method to assess the accuracy of any type of oximeter used for patient monitoring.
There is no universally agreed upon “gold standard” measurement of SctO2 in healthy volunteers. However, comparison of NIRS generated SctO2-generated values with reference SctO2 (refSctO2) values derived from simultaneously sampling and measurement of radial artery oxygen saturation (SaO2) and jugular bulb venous oxygen saturation (SjvO2) via co-oximetry as performed by Henson et al2 as the standard.
The purpose of this study was to: 1) determine the accuracy of the FORE-SIGHT cerebral oximeter SctO2 values for Food and Drug Administration 510 (k) pre-market approval by comparing them to refSctO2 values derived from simultaneous radial artery and jugular bulb venous samples obtained from healthy subjects during room air breathing, mild hypoxemia and hyperoxemia in the presence of normocapnia; and 2) assess the accuracy of noninvasive venous oxygen saturation (SnvO2) derived from SpO2 and SctO2 values.
Methods
After IRB approval (August, 2004), 10 male and 10 female healthy volunteers aged 21 to 35 years with a body mass index of 18–30 kg·m−2 were enrolled into the study after signing a written consent form. Because of the relatively low systemic saturation targets, the IRB mandated that we report every 3 subjects to the Data Safety and Monitoring Board for review.
An 18G peripheral IV, 20G radial arterial and a 5F venous jugular bulb cannulae were inserted in each of the 20 subjects under 2% lidocaine HCl local anesthesia. The jugular bulb cannula3 was initially inserted under ultrasound guidance (Sonosite Titan, Sonosite, Seattle, WA) and advanced cephalad until the subject reported minor ear discomfort.4 A lateral skull radiograph confirmed cannula tip placement cranial to a line extending from the atlanto-occipital joint space and caudal to the lower margin of the orbit, thus insuring the absence of extracranial venous blood contamination.5 Bilateral oximeter sensors were applied to the subject’s forehead, symmetrically along the midline to avoid the sagittal sinus and as far above the orbital ridge to maximize interrogation of brain as the hairline allowed, and connected to a dual-channel FORE-SIGHT cerebral oximeter. Continuous venous jugular bulb (SjbO2) and cardiac output in liters per minute were monitored with a Vigilance hemodynamic monitor (Edwards LifeScience, Irvine, CA), and finger arterial oxygen saturations with a Nellcor pulse oximeter (Model N595, Coviden, Boulder CO) respectively. Periodic noninvasive arterial blood pressure, continuous arterial blood pressure, heart rate and electrocardiogram were monitored with a PROPAQ (Model CS, Welch Allyn, Beaverton, OR). Subjects were positioned supine with the head of the bed elevated 35°.
The subject inspired gas mixtures delivered to a facemask from a sequential gas delivery system6,7 (Respiract UDB 1000; Thornhill Research, Toronto, Canada) capable of producing and maintaining target end-tidal partial pressure of oxygen (PetO2) and carbon dioxide (PetCO2) values independent of one another. The system works by dividing the inspired breath into 2 components, fresh gas and the last end-tidal breath (which has already reached equilibrium with pulmonary venous gas pressures), such that only the small sample of fresh gas contributes to the gas exchange as long as the alveolar ventilation rate exceeds that of the fresh gas flow rate.8 End-tidal O2 and CO2 values were continuously measured using a Datex Capnomac monitor (Ultima I V90 EN, GE Healthcare, Waukesha, WI) via a sampling port located on the mask, with a sampling line of approximately 2 meters. Each subject initially inspired 21% O2 followed by 2 sequences of reduced oxygen breathing designed to produce a progressive and then an acute reduction in blood O2 saturation. After the period of reduced oxygen breathing, the inspired O2 concentration was returned to 21% and then increased to 50%. The subject was allowed a brief rest period between the 2 hypoxia sequences. The progressive sequence consisted of 10 steps of 4 minutes with inspired oxygen concentrations of 21, 12, 10, 9, 8, 9, 10, 12, 21, and 50%. The acute sequence consisted of 4 steps with inspired oxygen concentrations of 21, 8, 21, and 50%. The target PetCO2 during each step of the 2 sequences was 40 mmHg (Figure 1).
Figure 1.
Example of hemodynamic and respiratory data obtained during the progressive hypoxia sequence in a single subject.
Each step was maintained for 4 minutes and paired arterial and jugular venous blood samples were obtained during the last 2 minutes of the step. The jugular venous samples were drawn at a rate of 2 mL/min and the arterial samples were taken at the midpoint of the jugular venous draws (Figure 2).9,10
Figure 2.
Example of the SpO2, FORE-SIGHT™ SctO2, Vigilance SjbO2, and co-oximeter determined SjvO2 and SaO2 values during the progressive hypoxia sequence in a single subject.
If the subject did not reach the desired SpO2 minimum target value of 70% during either sequence, the O2 content in the delivered gas mixture was reduced to 5% FiO2 and additional blood samples were drawn at the new SpO2 plateau. Blood draws were omitted if the subject reached the desired SpO2 threshold of 70% in fewer steps or became excessively anxious. Blood samples were analyzed immediately by co-oximetry (Model IL282; Instrumentation Laboratories, Lexington, MA).
The NIRS-derived SctO2 values were calibrated against SaO2 and SjvO2 blood samples using a 3-compartment cerebral blood volume (CBV) model (refSctO2; equation 1) :11
| (1) |
where the distribution coefficients of arterial (Ka), venous (Kv) and capillary blood (Kc) oxygen saturations are set at 0.3, 0.7, and 0.0 for the FORE-SIGHT® (CAS Medical Systems; Branford, CT), respectively. The ratio of the volume of arterial blood (30%) to venous blood (70%) in the cerebral tissue is based on the work of Ito et al.36 The capillary blood compartment volume is considered negligible and so the capillary blood saturation (ScO2) contribution to CBV is considered nil (Kc = 0.0). The error term is a generic term that sums the various sources of error that may influence the SctO2 reading. Equation 1 was solved for jugular bulb venous oxygen saturation (SjvO2; equation 2).
| (2) |
Substituting pulse oximetry (SpO2) values for co-oximeter determined SaO2 and NIRS SctO2 values for co-oximeter determined refSctO2 produced a noninvasive estimate of venous oxygen saturation and global cerebral oxygen consumption (SnvO2; equation 3):
| (3) |
Furthermore, the relationship between the arterio-jugular oxygen difference (AVDO2), cerebral metabolic rate for O2 (CMRO2) and CBF as a function of arterial blood oxygen-carrying capacity (CaO2) and venous blood oxygen-carrying capacity (CjvO2)12 can be expressed in terms of SjvO2, SaO2, partial pressure of oxygen in the arteries (PaO2), and partial pressure of oxygen in the jugular vein (PjvO2), if the difference between the arterial and venous concentration of hemoglobin is negligible (equations 4 and 5).
| (4) |
| (5) |
Equation 6 shows the noninvasive version of equation 5 where SpO2 and SnvO2 have been substituted for SaO2 and SjvO2 respectively to determine the relationship between CMRO2 and CBF:
| (6) |
CMRO2 / CBF is approximately proportional to the difference between SnvO2 and SpO2 if the effects of dissolved oxygen (0.003 × (PaO2 – PjvO2)) are negligible. In the clinical setting, this approximation using noninvasive parameters can be important because the decrease in SctO2 can indicate either a decrease in CBF, or an increase in CMRO2, or both.
Data Analysis
A priori statistical power analysis determined that a minimum of 100 samples would be required to obtain a 99% confidence interval ((1 − □) of a width of ± 4% (each side) around the estimate of the mean with a standard deviation of 5. In the present study, we had 18 subjects with 2 sensors each, and each sensor yielded an average of 13 data points per sensor per subject, which netted an average of 27 data points total per subject. The ipsilateral and contralateral forehead SctO2 values, filtered within the device with a moving average filter with a window of 6 seconds, recorded at each step of the progressive and acute sequences at the midpoint of the predetermined 30 second jugular venous blood draw to avoid extracranial contamination, as reported by Matta and Lam,9 were compared to each other, and the refSctO2 value, and used to compute SnvO2 which was then compared to SjvO2. A paired Student’s t-test was used to compare the ipsilateral and contralateral sensor SctO2 measurements for each subject. The correlation and the Lin’s concordance correlation coefficient (CCC)13,14 (and the 95% confidence interval denoted as standard error (SE)15) between SctO2 and refSctO2 values were examined by constructing linear regression curves for each subject; for each forehead sensor independently and for both sensors combined. The agreement between ipsilateral and contralateral SctO2 and refSctO2 values across all subjects and all samples was explored by constructing the Bland and Altman16 plot and modified for random effect analysis for repeated measures by Myles and Cui.17 The degree of error, defined as the difference between the SctO2 and the refSctO2 values, was assessed in terms of parametric analysis of the errors, including bias (mean of the difference between FORE-SIGHT SctO2 and refSctO2), standard deviation (SD), confidence intervals (bias ±1.95 SD), tolerance intervals (for 99% of future measurements to be within 95% confidence), and first order linear regression,18 adjusted for random effects analysis as explained by Myles and Cui17 and adjustments for repeated measures as previously described.13 The analysis was repeated for the error between the calculated SnvO2 values and the co-oximeter-determined venous oxygen saturation values (SjvO2). Additional nonparametric analysis was performed in the form of a cumulative distribution function (CDF) because the error is not exactly normally distributed.19
Results
Of the 20 subjects enrolled in the study 18 were included in the analysis. The prototype FORE-SIGHT was unable to resolve NIRS measurements from 1 subject due to unexpected attenuation of the light through the tissue and blood gas equipment failure prevented blood gas analysis in another. The non-Gaussian distribution of subjects’ demographic information is summarized in Tables 1 and 2.
Table 1.
Demographics
| Age* | Weight* | BMI* | |
|---|---|---|---|
| Female | 26.0 [22.0, 33.4] |
65.4 [49.8, 75.6] |
22.2 [20.1, 26.5] |
| Male | 23.3 [21.4, 35.2] |
76.9 [63.4, 99.1] |
23.6 [21.8, 29.6] |
| Overall | 25.1 [21.4, 35.16] |
70.1 [49.8, 99.1] |
22.8 [20.1, 29.6] |
Median and Range
Table 2.
Demographics
| African America n |
Asian | Caucasia n |
Hispanic | Total | |
|---|---|---|---|---|---|
| Female | 0 | 2 | 7 | 0 | 9 |
| Male | 2 | 1 | 5 | 1 | 9 |
| Total | 2 | 3 | 12 | 1 | 18 |
Sixteen paired arterial and venous blood samples were collected during the progressive phase followed by the acute phase of the experiment from 2 subjects, 15 from 7 subjects, 14 from 6 subjects, 13 from 2 subjects and 7 from 1 subject. Variations in the number of samples collected were due to having to reduce or increase the inspired gas exposure duration to achieve the target SpO2 of 70% or the necessity to terminate the hypoxia exposure based on subject tolerance. Of the 254 total paired samples drawn, 8 arterial samples and 1 venous sample could not be analyzed due to clotting of the sample. Therefore 246 arterial and 253 venous samples were analyzed and included in the analysis. The lowest recorded SaO2 during the course of the study, across all subjects, was 64%. The PetCO2 was independently targeted to 40 mmHg for the entire experiment, which resulted in mean arterial PaCO2 during the 2 phase exposure (example in Figure 1) was 38.8 ± 2.2 mmHg (mean ± SD). The average of the ipsilateral and contralateral SctO2 values recorded at baseline during room air breathing was 74.6 ± 3.8% (mean ± SD) with a range of 66.6 to 83.3%.
The significance of the difference between ipsilateral and contralateral SctO2 varied from subject to subject, with overall significance of P = 0.61 (Table 3). The ipsilateral SctO2 correlated better with a CCC ± SE of 0.83 ± 0.073 and a tolerance interval (TI) of [−10.72 10.90] versus the contralateral side of 0.78 ± 0.098 with TI of [−10.59 10.74], with the combined data showing a CCC of 0. 81 ± 0.059 with TI of [−9.22 9.40] (Figure 3, Table 4).
Table 3.
Paired Student T-Test for the ipsilateral and contralateral values of SctO2 and SnvO2 per subject. P value below 0.05 is considered significant.
| Subject | Student T- Test P Value (SctO2) |
Student T- Test P Value (SnvO2) |
(SctO2) Ipsilateral / Contralateral dichotomy significant? |
(SnvO2) Ipsilateral / Contralateral dichotomy significant? |
|---|---|---|---|---|
| S02 | 2.86 e-9 | 2.86 e-9 | Yes | Yes |
| S03 | 4.85 e-6 | 4.85 e-6 | Yes | Yes |
| S04 | 2.45 e-9 | 2.45 e-9 | Yes | Yes |
| S05 | 2.08 e-8 | 2.08 e-8 | Yes | Yes |
| S06 | 1.70 e-4 | 1.70 e-4 | Yes | Yes |
| S07 | 9.64 e-1 | 9.00 e-1 | No | No |
| S08 | 3.67 e-1 | 3.64 e-1 | No | No |
| S09 | 6.78 e-2 | 1.18 e-1 | No | No |
| S10 | 4.46 e-3 | 1.88 e-3 | Yes | Yes |
| S11 | 1.51 e-1 | 1.99 e-1 | No | No |
| S13 | 3.42 e-12 | 3.42 e-12 | Yes | Yes |
| S14 | 1.69 e-1 | 2.98 e-1 | No | No |
| S15 | 8.33 e-3 | 7.17 e-3 | Yes | Yes |
| S16 | 1.33 e-2 | 2.46 e-2 | Yes | Yes |
| S17 | 3.99 e-5 | 9.10 e-7 | Yes | Yes |
| S18 | 1.29 e-2 | 1.78 e-2 | Yes | Yes |
| S19 | 1.68 e-1 | 1.68 e-1 | No | No |
| S20 | 3.98 e-1 | 8.98 e-1 | No | No |
| Overall | 7.23 e-1 | 6.93 e-1 | No | No |
Figure 3.
Plot of FORE-SIGHT™ SctO2 ipsilateral and contralateral combined sensor values versus calculated refSctO2 values.
Table 4.
Parametric analysis and Random Effects Analysis correction for the ipsilateral, contralateral and combined SctO2 values compared to refSctO2 values.
| Parametric Analysis with Random Effects Analysis Correction: |
SctO2 Ipsilateral |
SctO2 Contralateral |
SctO2 Combined |
|---|---|---|---|
| Bias | 0.09 | 0.08 | 0.09 |
| Standard Deviation | 2.92 | 2.88 | 2.86 |
| 95% Confidence Interval | [−5.60 5.78] | [−5.53 5.69] | [−5.48 5.57] |
| 95% Tolerance Interval | [−10.72 10.90] | [−10.59 10.74] | [−9.22 9.40] |
| Precision | 0.20 | 1.14 | 0.46 |
| Pearson Correlation | 0.85 | 0.78 | 0.81 |
|
Lin’s Concordance Correlation Coefficient ± Standard Error |
0.83 ± 0.073 | 0.78 ± 0.098 | 0.81 ± 0.059 |
| Slope | 1.05 | 0.74 | 0.89 |
| Intercept | −3.08 | 17.96 | 7.44 |
The complete parametric analysis of the errors for the ipsilateral, contralateral, and the combined data are in Table 4. The ipsilateral side was on the same side as the jugular bulb catheter, and the data reflect a more accurate measurement from the sensor located on the same side as the catheter.
The calculated proportional error for all ipsilateral and contralateral SctO2 values is shown on the random effect analysis-corrected Bland-Altman plot (Figure 4). There was a proportional error of less than 0.04% per decade (for every 10 % change in the (FORE-SIGHT SctO2 + refSctO2) / 2 value; X axis) and the error decreased as saturation values decreased.
Figure 4.
Bland-Altman plot of the difference between FORE-SIGHT SctO2 and refSctO2 values.
The relationship between blood gas-determined venous jugular bulb SjvO2 and the SnvO2 values calculated using the combined ipsilateral and contralateral FORESIGHT readings was examined using linear regression analysis (Figure 5) and a Bland-Altman plot (Figure 6)
Figure 5.
Plot of the calculated noninvasive venous oxygen saturation (SnvO2) versus the blood gas determined jugular bulb venous oxygen saturation (SjvO2) values.
Figure 6.
Bland-Altman plot of the calculated noninvasive venous oxygen saturation (SnvO2) versus the blood gas determined jugular bulb venous oxygen saturation (SjvO2) values.
The ipsilateral, contralateral and combined data for SnvO2 and the data obtained from the jugular bulb catheter (SjbO2) are compared to the SjvO2 values (Table 5). The data show that the SnvO2 calculated from the same side as the jugular bulb catheter (CCC ± SE = 0.79 ± 0.088, TI = [−14.93 15.33]) was more accurate than the contralateral sensor (CCC ± SE = 0.72 ± 0.118, TI = [−14.86 15.20]). The jugular bulb catheter (Vigilance) fared only slightly better with CCC + SE = 0.82 ± 0.078 with a TI of [−12.60 9.66].
Table 5.
Comparison of the calculated SnvO2 values derived from the ipsilateral, contralateral and combined measurements and the Vigilance (SjbO2) measurements to the venous jugular bulb oxygen saturation (SjvO2) values from blood samples.
| Parametric Analysis with Random Effects Analysis Correction: |
SnvO2 Ipsilateral |
SnvO2 Contralate ral |
SnvO2 Combined |
SjbO2 Vigilance |
|---|---|---|---|---|
| Bias | 0.20 | 0.17 | 0.19 | −1.47 |
| Standard Deviation | 4.08 | 4.06 | 4.01 | 3.00 |
| 95% Confidence Interval | [−7.80 8.20] | [−7.78 8.12] | [−7.68 8.05] | [−2.88 8.89] |
| 95% Tolerance Interval | [−14.93 15.33] |
[−14.86 15.20] |
[−12.89 13.26] |
[−12.60 9.66] |
| Precision | 0.17 | 1.92 | 0.99 | 0.34 |
| Pearson Correlation | 0.81 | 0.73 | 0.77 | 0.85 |
|
Lin’s Concordance Correlation Coefficient ± Standard Error |
0.79 ± 0.088 |
0.72 ± 0.118 |
0.76 ± 0.071 |
0.82 ± 0.078 |
| Slope | 0.98 | 0.66 | 0.82 | 0.93 |
| Intercept | 1.14 | 20.36 | 10.75 | 2.56 |
All values were compared against jugular bulb venous (SjvO2) blood sample values determined by co-oximetry.
The CDF plot of the errors in SctO2 (Figure 7 and Figure 8) clearly shows subject 2 to be an outlier, thus the parametric analysis will be partially biased away from the norm. It also shows that approximately 90% of the error is within 8.5% for the ipsilateral, and 10% for the contralateral. Similarly, the CDF plot of the SnvO2 shows once again that subject 2 is an outlier. It also shows that 90% of the SctO2 errors on the ipsilateral are within 12% and 14% on the contralateral.
Figure 7.
Cumulative Density Function Plot of the absolute value of the error between SctO2 and refSctO2.
Figure 8.
Cumulative Density Function Plot of the absolute value of the error between (SnvO2) and the blood gas determined jugular bulb venous oxygen saturation (SjvO2) values.
Discussion
We found a strong positive correlation between the ipsilateral forehead SctO2 measurements and the refSctO2 values calculated using arterial and jugular bulb oxygen saturations and the FORE-SIGHT SctO2 values. The ipsilateral sensor performed better than the contralateral, and the comparison between the 2 sides on a subject-by-subject basis showed that ipsilateral-contralateral dependency varied widely. Differences between the refSctO2 and FORE-SIGHT SctO2 values may reflect interindividual variability20 and/or differences in the regional blood flow of the cerebral tissue, as well as contamination from extracranial tissue21. Henson et al. sought to use SctO2 as a predictor of SjvO2, and found that the Somanetics 3100 cerebral oximeter device correlated well with venous oxygen saturation.2
The ± 10% tolerance interval of future readings may seem excessively large but given the sigmoid nature of the oxygen-hemoglobin dissociation curve22, this may not be clinically significant. As an example, a reading of 75% SctO2 may be between 65% or 85%, but a value within this range does not suggest an imminent clinical threat to the patient. An SctO2 of 55% indicates a venous saturation between 50% and 21%, which is the arbitrary threshold for neurocognitive impairment as set and published by Yao et al23, given that the arterial saturation is usually near 100%. Several studies have reported neurocognitive deficits and prolonged hospital stay when the INVOS™ (Covidien (formerly Somanetics), Mansfield, MA) rSO2 decreases below 50% for a prolonged period, 40% for more than 10 minutes, or nadir below 35% where the corresponding cerebral venous saturations decreased below 35%, 20%, and 13% accordingly.24,25,26 There is no clear SctO2 threshold for clinical intervention so the SctO2 may serve as a graded indicator of risk organized in decades. The normal range is considered 65% and above, elevated risk between 55% and 65%, and high risk at 55% and below (55% SctO2 corresponds to 35% venous saturation, assuming 100% SaO2). Because arterial blood oxygen saturations are very well managed in modern anesthesia practice, estimation of the noninvasive estimate of venous oxygen saturation (SnvO2) as a surrogate of SjvO2 may have some merit and indicate clinically relevant changes in the ratio between CMRO2 and CBF. A crude summary of physiologic conditions and their possible contributors are tabulated in Table 6.
Table 6.
Possible causes of inflections in the SnvO2 due to physiological state change in the cerebral tissue.
| Indication: | SaO2 | CMRO 2 |
CPP | CBF | Notes |
|---|---|---|---|---|---|
|
Low SnvO2 |
Low | Low | Low CPP may be due to Low MAP or high ICP, or both |
||
|
Low SnvO2 |
High | Some regions of the brain may be more metabolically active, which may indicate a low SnvO2 relative to systemic SvO2 |
|||
|
Low SnvO2 |
High | Low | Possibly vasoconstriction due to low PaCO2 and vasopressors |
||
|
Low SnvO2 |
Low | Low arterial blood saturation | |||
|
High SnvO2 |
High | Possibly vasodilatation due to high PaCO2, drugs that relax smooth muscle, and diuretics |
|||
|
High SnvO2 |
High | High | Possibly vasodilatation due to high PaCO2, vasopressors, high Cardiac Output, and / or high Mean Arterial Pressure (MAP) |
||
|
High SnvO2 |
High | Possibly vasodilatation due to high PaCO2 and vasopressors |
|||
|
High SnvO2 |
Low | Low brain activity |
All blank spaces denote normal ranges.
The Bland-Altman plot demonstrated a modest degree of proportional error, which declined as the blood oxygen saturation value decreased; a positive characteristic in the setting of clinical hypoxia. In addition, the data points were equally scattered above and below the zero line suggesting there is no consistent bias of 1 measurement approach (FORE-SIGHT SctO2) versus the refSctO2. SnvO2 calculated using the FORE-SIGHT SctO2 and SpO2 values was strongly correlated with blood gas determined SjvO2.
The wide variability of the average FORE-SIGHT SctO2 values recorded at baseline during room air breathing is generally consistent with what has been reported clinically.27,28,29 The significance of the cerebral oximeter can be further demonstrated by substituting equation 3 (assuming blood distribution to be 70% venous (Kv = 0.7), 30% arterial (Ka = 0.3), and negligible amount in the capillaries (Kc = 0) and ERROR assumed to be zero for the ideal case as suggested by Ito et al.11) into equation 6 and solving the ratio of CMRO2 to CBF as a function of SctO2 (equation 7).
| (7) |
Where K1 = (1.34 × [tHb]) /0.7
And
K2 = 0.003 × (PaO2 − PjbO2).
Solving equation 7 for SctO2 then becomes (equation 8):
| (8) |
From this equation, it can be deduced that the SctO2 has a proportional relationship to SpO2 and CMRO2 and inversely proportional with CBF. Thus, the variation in baseline SctO2 can either be directly due to interpersonal differences in brain activity and CBF. Any preexisting conditions affecting CBF such as sympathetic activation due to anxiety and reduced CBF due to vasoconstriction (abnormal PaCO2 due to hyperventilation, vasoconstrictors onboard, etc.), or atherosclerosis can affect the baseline SctO2.
In the current study, the effect of CO2 on CBF, and therefore oxygen delivery, was controlled by maintaining the PetCO2 at 40 mmHg with a scheme described by Slessarev et al.20 However, inspection of the individual subject’s linear regression plots show a slight inflection of the plots at lower saturations, which may indicate that the reference saturation and the NIRS saturation measurements may not be linear at lower saturation levels due to changing CMRO2 and other confounding effects. The end-tidal gas-forcing device software assumes constant oxygen consumption rates that may have introduced an effort-dependent error. Because Seifert and Secher have shown that sympathetic activity can possibly influence CBF and CMRO2 during cycling exercise,30 the increased respiratory effort with the diaphragm muscle during this study could have affected a change in CBF and CMRO2.
The morphology of the Bland-Altman plot [which resembles a “less than” symbol (<), with smaller spread in the 50% range than the 70% range], also suggests an error that is proportional to the magnitude of the SctO2 reading and corroborates the possibility that the SctO2 may be slightly nonlinear. In addition, hypoxemia produces changes in CBF31 and the full extent of the changes in the venous distribution and mixing of the various tributaries that drain into the jugular venous returns are not entirely known. Due to this uncertain venous mixing, the venous sample taken at the jugular bulb may not have reflected the true venous saturation at the interrogated tissue.
In this study refSctO2 values were calculated for the purposes of evaluating the accuracy of the FORE-SIGHT device using distribution coefficients (0.30*SaO2 + 0.70 * SjvO2) derived from Ito et al.32 However, Pollard et al.33 have estimated SctO2 using a slightly different CBV distribution (0.25*SaO2 + 0.75 * SjvO2) as has McCormick et al. (0.18*SaO2 + 0.82* SjvO2 and 0.28*SaO2 + 0.72* SjvO2).34 Not only does a device manufacturer’s choice of distribution coefficients affect the SctO2 reported value, but use of a fixed set of distribution coefficients can also be a source of error because the CBV distribution and CBF both change due to various activities and body positions.33 Functional magnetic resonance imaging studies by Ito et al35 have shown that an elevated CMRO2 during neural activation causes a decrease in venous blood oxygen saturation. Regional differences in cerebral activity may also cause cerebral blood distribution to change. Ito et al. used controlled visual stimulation to demonstrate significant differences in CBF and CBV during rest and periods of stimulation, which was consistent with positron emission tomography studies estimating CBF and CBV under hypercapnic conditions.36 However, Ito et al. did not precisely control PaCO2 and their findings indicate a wide range of possible distribution of the CBV under these conditions. They also found that CBV changed significantly owing to changes in PaCO2, and arterial compartment volume changed significantly in the direction of CBV changes. However, no significant change was observed in venous or capillary blood volume. This indicates that changes in CBV during hypercapnia and hypocapnia are produced by changes in arterial blood volume and not by changes in venous and capillary blood volume.37 The appropriate distribution coefficients for CBV within arterial and venous compartments remain a somewhat contentious issue but it is generally accepted that any increase in the concentration of CO2 within the arterial compartment and sympathetic activation due to mental activity can trigger vasodilatation and thus affect changes in the CBV.
Our results were obtained by controlling the PetCO2 in order to achieve a PaCO2 of approximately 40 mmHg (i.e., 38.8 ± 2.2 mmHg), which allowed a more consistent descent down the oxyhemoglobin dissociation curve. The dissociation curve from 20% to 80% is approximately linear and a small shift in the curve to the ipsilateral or contralateral may produce an unacceptable source of error when validating the calibration of a cerebral oximeter. Ito et al. found that the arterial fraction of human CBV during hypercapnia and hypocapnia measured by positron emission tomography changed such that the 25% arterial fraction assumed by the INVOS cerebral oximeter and the 30% arterial fraction assumed by the FORE-SIGHT cerebral oximeter for calculation of SctO2 were no longer valid under these conditions.37,38 The importance of maintaining normocapnia when evaluating the accuracy of a NIRS cerebral oximeter device which uses a fixed arterial to venous blood volume ratio is readily apparent from Ito et al.’s findings. Unfortunately, hypocapnia and hypercapnia in the clinical settings add to the uncertainty and error of these devices that have been calibrated under normocapnic conditions, regardless of whether a device is a “trend” or “absolute” monitor.
The gas delivery system required the gas blender to be calibrated at the beginning of the study to the subject’s rate of CO2 production (dVCO2/dt) and rate of O2 consumption (dVO2/dt). Any physical movement39 and / or any changes in the breathing effort alter the dVCO2/dt and dVO2/dt.40 The increased ventilatory drive at lower SaO2 increases respiratory effort thereby increasing both dVCO2/dt and dVO2/dt. The nature of the end-tidal-forcing device and technique used in this experiment made it impossible to adjust for these changes during the experiment and the initial set points for dVCO2/dt and dVO2/dt were no longer applicable at lower SaO2 and therefore resulted in some drift of the end-tidal target gas concentrations. Therefore vasodilatation and vasoconstriction due to hypercapnia and hypocapnia could have altered the intracranial blood volume distribution, and in turn the SctO2.
Aside from the sources of error noted, the NIRS device and its interaction with interindividual anatomical differences may also introduce error.41 The sensors are designed to be placed on the forehead equilaterally. If there is a positional bias towards one side or the other, the sagittal sinus, a prominent anatomical feature, may influence the accuracy of the sensor readings.42,43 Because the device is also designed to reject common mode noise, exposure to ambient light should not be an issue. However, an excessive amount of ambient light can saturate the sensors and introduce error in the reading. Individual skull thickness can also affect scattering of the NIRS optical signal and influence device readings.44,45 The present study included 2 African-Americans, 3 Asians, and 1 Hispanic male. The varying levels of skin pigmentation may have also manifested as artificially lower readings at lower saturations as Feiner et al. and Bickler et al. have reported.46,47
Conclusions
We found that the FORE-SIGHT NIRS device was able to estimate oxygen saturation (with 95% confidence interval of [−5.60 5.78] with ipsilateral blood sample data) within the tissues of the frontal lobe under conditions of normocapnia and varying degrees of hypoxia as suggested by the ISO 9919:2005 standard. The study findings also suggest that the use of the calculated venous oxygen saturation (SnvO2) may be a reasonable noninvasive method of estimating jugular bulb venous oxygen saturation and therefore global cerebral oxygen consumption in the clinical setting. Further laboratory and clinical research is required to define the value of NIRS determination of SctO2 and SnvO2 in the operating room and intensive care unit setting.
Acknowledgments
Funding: This study was funded by the CAS Medical Systems, Inc. under NIH SBIR grant NS045488
Footnotes
See Disclosures at end of article for Author Conflicts of Interest.
DISCLOSURES:
Name: Keita Ikeda, PhD
Contribution: This author helped with the data acquisition, data collection, and conduct of the study, data analysis, and manuscript preparation.
Attestation: Keita Ikeda attests to having approved the final manuscript.
Conflicts of Interest: The author has no conflicts of interest to declare.
Name: David B MacLeod, MBBS, FRCA
Contribution: study design, conduct of the study, data collection
Attestation: David MacLeod attests to having approved the final manuscript.
Conflicts of Interest: research funding from CAS Medical Systems, Inc.
Name: Hilary P. Grocott, MD, FRCPC, FASE
Contribution: Conduct of the study, manuscript writing
Attestation: Hilary Grocott attests to having approved the final manuscript
Conflicts of Interest: The author has no conflicts of interest to declare.
Name: Eugene W. Moretti, MD, MHSc
Contribution: Conduct of the study
Attestation: Dr. Eugene Moretti attests to having approved the final manuscript.
Conflicts of Interest: The author has no conflicts of interest to declare.
Name: Warwick Ames, MBBS, FRCA
Contribution: Conduct of the study
Attestation: Dr. Warwick Ames attests to having approved the final manuscript.
Conflicts of Interest: The author has no conflicts of interest to declare.
Name: Charles Vacchiano, PhD, CRNA
Contribution: This author helped with the manuscript preparation.
Attestation: Charles Vacchiano attests to having approved the final manuscript.
Conflicts of Interest: The author has no conflicts of interest to declare.
This manuscript was handled by: Gregory J. Crosby, MD
Contributor Information
Keita Ikeda, Department of Anesthesia, Duke University Medical Center, Durham, North Carolina.
David B MacLeod, Department of Anesthesia, Duke University Medical Center, Durham, North Carolina.
Hilary P. Grocott, Department of Anesthesia & Perioperative Medicine, University of Manitoba, Winnipeg, Manitoba, Canada.
Eugene W. Moretti, Department of Anesthesia, Duke University Medical Center, Durham, North Carolina.
Warwick Ames, Department of Anesthesia, Duke University Medical Center, Durham, North Carolina.
Charles Vacchiano, School of Nursing, Duke University, Durham, North Carolina.
References
- 1.Wray S, Cope M, Delpy DT, Wyatt JS, Reynolds EO. Characterization of the near infrared absorption spectra of cytochrome aa3 and haemoglobin for the non-invasive monitoring of cerebral oxygenation. Biochim Biophys Acta. 1988 Mar 30;933(1):184–192. doi: 10.1016/0005-2728(88)90069-2. [DOI] [PubMed] [Google Scholar]
- 2.Henson LC, Calalang C, Temp JA, Ward DS. Accuracy of a cerebral oximeter in healthy volunteers under conditions of isocapnic hypoxia. Anesthesiology. 1998 Jan;88(1):58–65. doi: 10.1097/00000542-199801000-00011. [DOI] [PubMed] [Google Scholar]
- 3.Chieregato A, Calzolari F, Trasforini G, Targa L, Latronico N. Normal jugular bulb oxygen saturation. J Neurol Neurosurg Psychiatry. 2003 Jun;74(6):784–786. doi: 10.1136/jnnp.74.6.784. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Matta B, Lam A. Retrograde cannulation of the jugular vein. Br J Anaesth. 1995 Sep;75(3):373. doi: 10.1093/bja/75.3.373. [DOI] [PubMed] [Google Scholar]
- 5.Kawahara F, Kadoi Y, Saito S, Goto F, Fujita N. Slow rewarming improves jugular venous oxygen saturation during rewarming. Acta Anaesthesiol Scand. 2003 Apr;47(4):419–424. doi: 10.1034/j.1399-6576.2003.00063.x. [DOI] [PubMed] [Google Scholar]
- 6.Slessarev M, Han J, Mardimae A, Prisman E, Preiss D, Volgyesi G, Ansel C, Duffin J, Fisher JA. Prospective targeting and control of end-tidal CO2 and O2 concentrations. J Physiol. 2007 Jun 15;581(Pt 3):1207–1219. doi: 10.1113/jphysiol.2007.129395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Somogyi RB, Vesely AE, Preiss D, Prisman E, Volgyesi G, Azami T, Iscoe S, Fisher JA, Sasano H. Precise control of end-tidal carbon dioxide levels using sequential rebreathing circuits. Anaesth Intensive Care. 2005 Dec;33(6):726–732. doi: 10.1177/0310057X0503300604. [DOI] [PubMed] [Google Scholar]
- 8.Mutch WA, Patel SR, Shahidi AM, Kulasekara SI, Fisher JA, Duffin J, Hudson C. Cerebral oxygen saturation: graded response to carbon dioxide with isoxia and graded response to oxygen with isocapnia. PLoS One. 2013;8(2):e57881. doi: 10.1371/journal.pone.0057881. Epub 2013 Feb 28. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Matta BF, Lam AM. The rate of blood withdrawal affects the accuracy of jugular venous bulb Oxygen saturation measurements. Anesthesiology. 1997 Apr;86(4):806–808. doi: 10.1097/00000542-199704000-00010. [DOI] [PubMed] [Google Scholar]
- 10.Matta BF, Lam AM, Mayberg TS. The influence of arterial oxygenation on cerebral venous oxygen saturation during hyperventilation. Can J Anaesth. 1994 Nov;41(11):1041–1046. doi: 10.1007/BF03015651. [DOI] [PubMed] [Google Scholar]
- 11.Ito H, Kanno I, Iida H, Hatazawa J, Shimosegawa E, Tamura H, Okudera T. Arterial fraction of cerebral blood volume in humans measured by positron emission tomography. Ann Nucl Med. 2001 Apr;15(2):111–116. doi: 10.1007/BF02988600. [DOI] [PubMed] [Google Scholar]
- 12.Valadka AB, Furuya Y, Hlatky R, Robertson CS. Global and regional techniques for monitoring cerebral oxidative metabolism after severe traumatic brain injury. Neurosurg Focus. 2000 Nov 15;9(5):e3. doi: 10.3171/foc.2000.9.5.3. Review. [DOI] [PubMed] [Google Scholar]
- 13.Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989 Mar;45(1):255–268. [PubMed] [Google Scholar]
- 14.Li R, Chow M. Evaluation of Reproducibility for Paired Functional Data. J Multivar Anal. 2005;93(1):81–101. doi: 10.1016/j.jmva.2004.01.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Lin LI. A Note on the Concordance Correlation Coefficient. Biometrics. 2000 Mar;56:324–325. [Google Scholar]
- 16.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310. [PubMed] [Google Scholar]
- 17.Myles PS, Cui J. Using the Bland-Altman method to measure agreement with repeated measures. Br J Anaesth. 2007 Sep;99(3):309–311. doi: 10.1093/bja/aem214. [DOI] [PubMed] [Google Scholar]
- 18.Statistical interpretation of data — Part 6: Determination of statistical tolerance intervals. ISO 16269-6. 2005:64. [Google Scholar]
- 19.Daniel Zwillinger, Stephen Kokoska. CRC Standard Probability and Statistics Tables and Formulae. CRC Press; 2010. p. 49. ISBN 978-1-58488-059-2. [Google Scholar]
- 20.Slessarev M, Mardimae A, Preiss D, Vesely A, Balaban DY, Greene R, Duffin J, Fisher JA. Differences in the control of breathing between Andean highlanders and lowlanders after 10 days acclimatization at 3850 m. J Physiol. 2010 May 1;588(Pt 9):1607–1621. doi: 10.1113/jphysiol.2009.186064. Epub 2010 Mar 15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Davie SN, Grocott HP. Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology. 2012 Apr;116(4):834–840. doi: 10.1097/ALN.0b013e31824c00d7. [DOI] [PubMed] [Google Scholar]
- 22.Balaban DY, Duffin J, Preiss D, Mardimae A, Vesely A, Slessarev M, Zubieta-Calleja GR, Greene ER, Macleod DB, Fisher JA. The in-vivo oxyhaemoglobin dissociation curve at sea level and high altitude. Respir Physiol Neurobiol. 2013 Mar 1;186(1):45–52. doi: 10.1016/j.resp.2012.12.011. Epub 2013 Jan 8. [DOI] [PubMed] [Google Scholar]
- 23.Yao FS, Tseng CC, Ho CY, Levin SK, Illner P. Cerebral oxygen desaturation is associated with early postoperative neuropsychological dysfunction in patients undergoing cardiac surgery. J Cardiothorac Vasc Anesth. 2004 Oct;18(5):552–558. doi: 10.1053/j.jvca.2004.07.007. [DOI] [PubMed] [Google Scholar]
- 24.Kazan R, Bracco D, Hemmerling TM. Reduced cerebral oxygen saturation measured by absolute cerebral oximetry during thoracic surgery correlates with postoperative complications. Br J Anaesth. 2009 Dec;103(6):811–816. doi: 10.1093/bja/aep309. [DOI] [PubMed] [Google Scholar]
- 25.de Tournay-Jetté E, Dupuis G, Bherer L, Deschamps A, Cartier R, Denault A. The relationship between cerebral oxygen saturation changes and postoperative cognitive dysfunction in elderly patients after coronary artery bypass graft surgery. J Cardiothorac Vasc Anesth. 2011 Feb;25(1):95–104. doi: 10.1053/j.jvca.2010.03.019. Epub 2010 Jul 22. [DOI] [PubMed] [Google Scholar]
- 26.Slater JP, Guarino T, Stack J, Vinod K, Bustami RT, Brown JM, 3rd, Rodriguez AL, Magovern CJ, Zaubler T, Freundlich K, Parr GV. Cerebral oxygen desaturation predicts cognitive decline and longer hospital stay after cardiac surgery. Ann Thorac Surg. 2009 Jan;87(1):36–44. doi: 10.1016/j.athoracsur.2008.08.070. discussion 44-5. [DOI] [PubMed] [Google Scholar]
- 27.MacLeod DB, Ikeda K, Vacchiano C, Lobbestael A, Wahr JA, Shaw AD. Development and validation of a cerebral oximeter capable of absolute accuracy. J Cardiothorac Vasc Anesth. 2012 Dec;26(6):1007–1014. doi: 10.1053/j.jvca.2012.06.010. Epub 2012 Aug 9. [DOI] [PubMed] [Google Scholar]
- 28.Kreeger RN, Ramamoorthy C, Nicolson SC, Ames WA, Hirsch R, Peng LF, Glatz AC, Hill KD, Hoffman J, Tomasson J, Kurth CD. Evaluation of pediatric near-infrared cerebral oximeter for cardiac disease. Ann Thorac Surg. 2012 Nov;94(5):1527–1533. doi: 10.1016/j.athoracsur.2012.05.096. Epub 2012 Aug 2. [DOI] [PubMed] [Google Scholar]
- 29.Rifai L, Winters J, Friedman E, Silver MA. Initial description of cerebral oximetry measurement in heart failure patients. Congest Heart Fail. 2012 Mar-Apr;18(2):85–90. doi: 10.1111/j.1751-7133.2011.00284.x. Epub 2012 Feb 16. [DOI] [PubMed] [Google Scholar]
- 30.Seifert T, Secher NH. Sympathetic influence on cerebral blood flow and metabolism during exercise in humans. Prog Neurobiol. 2011 Nov;95(3):406–426. doi: 10.1016/j.pneurobio.2011.09.008. Epub 2011 Sep 24. Review. [DOI] [PubMed] [Google Scholar]
- 31.Smith KJ, Wong LE, Eves ND, Koelwyn GJ, Smirl JD, Willie CK, Ainslie PN. Regional cerebral blood flow distribution during exercise: influence of oxygen. Respir Physiol Neurobiol. 2012 Oct 15;184(1):97–105. doi: 10.1016/j.resp.2012.07.014. Epub 2012 Aug 16. [DOI] [PubMed] [Google Scholar]
- 32.Ito H, Kanno I, Iida H, Hatazawa J, Shimosegawa E, Tamura H, Okudera T. Arterial fraction of cerebral blood volume in humans measured by positron emission tomography. Ann Nucl Med. 2001 Apr;15(2):111–116. doi: 10.1007/BF02988600. [DOI] [PubMed] [Google Scholar]
- 33.Pollard V, Prough DS, DeMelo AE, Deyo DJ, Uchida T, Widman R. The influence of carbon dioxide and body position on near-infrared spectroscopic assessment of cerebral hemoglobin oxygen saturation. Anesth Analg. 1996 Feb;82(2):278–287. doi: 10.1097/00000539-199602000-00011. [DOI] [PubMed] [Google Scholar]
- 34.McCormick PW, Stewart M, Goetting MG, Balakrishnan G. Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans. Stroke. 1991;22:596–602. doi: 10.1161/01.str.22.5.596. [DOI] [PubMed] [Google Scholar]
- 35.Ito H, Ibaraki M, Kanno I, Fukuda H, Miura S. Changes in cerebral blood flow and cerebral oxygen metabolism during neural activation measured by positron emission tomography: comparison with blood oxygenation level-dependent contrast measured by functional magnetic resonance imaging. J Cereb Blood Flow Metab. 2005 Mar;25(3):371–377. doi: 10.1038/sj.jcbfm.9600030. [DOI] [PubMed] [Google Scholar]
- 36.Ito H, Takahashi K, Hatazawa J, Kim SG, Kanno I. Changes in human regional cerebral blood flow and cerebral blood volume during visual stimulation measured by positron emission tomography. J Cereb Blood Flow Metab. 2001 May;21(5):608–612. doi: 10.1097/00004647-200105000-00015. [DOI] [PubMed] [Google Scholar]
- 37.Ito H, Ibaraki M, Kanno I, Fukuda H, Miura S. Changes in the arterial fraction of human cerebral blood volume during hypercapnia and hypocapnia measured by positron emission tomography. J Cereb Blood Flow Metab. 2005 Jul;25(7):852–857. doi: 10.1038/sj.jcbfm.9600076. [DOI] [PubMed] [Google Scholar]
- 38.Ito H, Kanno I, Fukuda H. Human cerebral circulation: positron emission tomography studies. Ann Nucl Med. 2005 Apr;19(2):65–74. doi: 10.1007/BF03027383. Review. [DOI] [PubMed] [Google Scholar]
- 39.Tachtsidis I, Elwell CE, Leung TS, Lee CW, Smith M, Delpy DT. Investigation of cerebral haemodynamics by near-infrared spectroscopy in young healthy volunteers reveals posture-dependent spontaneous oscillations. Physiol Meas. 2004 Apr;25(2):437–445. doi: 10.1088/0967-3334/25/2/003. [DOI] [PubMed] [Google Scholar]
- 40.Richardson DW, Kontos HA, Shapiro W, Patterson JL., Jr Role of hypocapnia in the circulatory responses to acute hypoxia in man. J Appl Physiol. 1966 Jan;21(1):22–26. doi: 10.1152/jappl.1966.21.1.22. [DOI] [PubMed] [Google Scholar]
- 41.Yoshitani K, Kawaguchi M, Miura N, Okuno T, Kanoda T, Ohnishi Y, Kuro M. Effects of hemoglobin concentration, skull thickness, and the area of the cerebrospinal fluid layer on near-infrared spectroscopy measurements. Anesthesiology. 2007 Mar;106(3):458–462. doi: 10.1097/00000542-200703000-00009. [DOI] [PubMed] [Google Scholar]
- 42.van der Zee P, Arridge SR, Cope M, Delpy DT. The effect of optode positioning on optical pathlength in near infrared spectroscopy of brain. Adv Exp Med Biol. 1990;277:79–84. doi: 10.1007/978-1-4684-8181-5_11. [DOI] [PubMed] [Google Scholar]
- 43.Dehghani H, Arridge SR, Schweiger M, Delpy DT. Optical tomography in the presence of void regions. J Opt Soc Am A Opt Image Sci Vis. 2000 Sep;17(9):1659–1670. doi: 10.1364/josaa.17.001659. [DOI] [PubMed] [Google Scholar]
- 44.Firbank M, Hiraoka M, Essenpreis M, Delpy DT. Measurement of the optical properties of the skull in the wavelength range 650–950 nm. Phys Med Biol. 1993 Apr;38(4):503–510. doi: 10.1088/0031-9155/38/4/002. [DOI] [PubMed] [Google Scholar]
- 45.Okada E, Delpy DT. Near-infrared light propagation in an adult head model. II. Effect of superficial tissue thickness on the sensitivity of the near-infrared spectroscopy signal. Appl Opt. 2003 Jun 1;42(16):2915–2922. doi: 10.1364/ao.42.002915. [DOI] [PubMed] [Google Scholar]
- 46.Feiner JR, Severinghaus JW, Bickler PE. Dark skin decreases the accuracy of pulse oximeters at low oxygen saturation: the effects of oximeter probe type and gender. Anesth Analg. 2007 Dec;105(6 Suppl):S18–S23. doi: 10.1213/01.ane.0000285988.35174.d9. tables of content. [DOI] [PubMed] [Google Scholar]
- 47.Bickler PE, Feiner JR, Severinghaus JW. Effects of skin pigmentation on pulse oximeter accuracy at low saturation. Anesthesiology. 2005 Apr;102(4):715–719. doi: 10.1097/00000542-200504000-00004. [DOI] [PubMed] [Google Scholar]








