Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2016 Jan 31.
Published in final edited form as: Nucl Med Commun. 2015 Feb;36(2):187–193. doi: 10.1097/MNM.0000000000000231

Image-derived input function with factor analysis and a priori information

Urban Simončič 1,2, Paolo Zanotti-Fregonara 3,4
PMCID: PMC4270921  NIHMSID: NIHMS629806  PMID: 25321159

Abstract

Background

Quantitative positron emission tomography (PET) studies often require the cumbersome and invasive procedure of arterial cannulation to measure the input function. This study sought to minimize the number of necessary blood samples by developing a factor-analysis based image-derived input function (IDIF) methodology for dynamic PET brain studies.

Methods

IDIF estimation was performed as follows: 1) carotid and background regions were segmented manually on early PET time frame; 2) blood-weighted and tissue-weighted time activity curves (TAC) were extracted with factor analysis; 3) factor analysis results were de-noised and scaled by using the voxels with the highest blood signal; 4) using population data and one blood sample at 40 minutes, the whole-blood TAC was estimated from post-processed factor analysis results; and 5) the parent concentration was finally estimated by correcting the whole blood curve with measured radiometabolite concentrations. The methodology was tested using data from 10 healthy subjects imaged with [11C](R)-rolipram. The accuracy of IDIFs was assessed against full arterial sampling by comparing the area under the curve (AUC) of the input functions and by calculating the total distribution volume (VT).

Results

The shape of the image-derived whole-blood TAC matched well the reference arterial curves, and the whole-blood AUCs was accurately estimated (mean error 1.0 ± 4.3%). The relative Logan-VT error was −4.1 ± 6.4%. Compartmental modeling and spectral analysis gave less accurate VT results than Logan.

Conclusion

A factor-analysis based IDIF for [11C](R)-rolipram brain PET studies that relies on single blood sample and population data can be used for accurate quantification of Logan-VT values.

Keywords: Image-derived input function, Factor analysis, [11C](R)-rolipram, PET

Introduction

An important logistical limitation of quantitative positron emission tomography (PET) scans is the need to measure, during image acquisition, the concentration over time of the parent radiotracer in plasma, i.e. the input function. This is generally done by collecting serial arterial blood samples, then separating the fraction of parent compound from its radiometabolites. The whole procedure is labor-intensive, time-consuming and uncomfortable for the patient, and alternative techniques are needed to reduce – or possibly eliminate – the need for blood sampling. Towards this end, one of the most often studied techniques is image-derived input function (IDIF). This approach relies on using physiological time activity curves (TAC) obtained from a blood pool visible on dynamic images. After opportune partial volume effect and radiometabolite correction, these TACs might replace serial arterial sampling as an input for kinetic modeling. While large blood pools, such as the heart or the aortic segments, might give IDIFs of good quality [1], getting an usable signal from smaller vessels, such as the carotids for brain studies, is much more challenging, due to partial volume effects [2]. Several strategies to obtain an IDIF from carotid arteries have been proposed. These include using accurate anatomical information from the co-registered MRI [3], correcting partial-volume effect during the image reconstruction procedure [4], blind source separation algorithms to derive the input function through the source signal mixing process [5], correcting spill-in by linear least-squares analysis with some blood samples [6] and many others [2].

Factor analysis is a statistical technique that can extract the physiological TACs from dynamic image sequences [78], by separating the dynamic PET image into its blood and tissue signal components over time. Factor analysis assumes that the observed signals are linearly transformed into a lower number of unobserved signals, called factors. Factor analysis can be implemented as either an algebraic manipulation of the observed signals [89] or as a least-squares-based problem with additional constraints such as non-negativity of factors and factor images [10].

The aim of the present study was to develop and validate an IDIF method for [11C](R)-rolipram brain PET studies based on factor analysis and relying on single blood sample and a priori information of radioligand’s kinetic behavior, vessel geometry, and scanner characteristics.

Materials and methods

Radiotracer preparation

[11C](R)-rolipram was synthesized as previously described [11] and according to Investigational New Drug Application #73,149, submitted to the US Food and Drug Administration. A copy of the application is available at: http://pdsp.med.unc.edu/snidd/nidpulldownPC.php. The radiotracer was obtained in high radiochemical purity (> 99%).

Image acquisition

Data from 10 consecutive healthy subjects who participated in a previous study were used [12]. All subjects gave written informed consent. The protocol was approved by the Ethics Committee of the National Institutes of Health.

Dynamic PET images were acquired for 90 minutes after a bolus injection of [11C](R)-rolipram (695 ± 152 MBq) with an Advance tomograph (GE Medical Systems, Waukesha, WI). The dynamic scan comprised 27 frames (six frames of 30 seconds each, then 3 × 60 s, 2 × 120 s, and 16 × 300 s). Images were reconstructed with filtered back-projection on a 128 × 128 matrix with a voxel size of 2.0 × 2.0 × 4.25 mm. An eight-minute 68Ge transmission scan was obtained before radiotracer injection for attenuation correction.

Input functions were obtained by drawing serial arterial samples from the radial artery: samples of 1 ml each were drawn from the radial artery at 15-second intervals until 150 seconds, followed by 3 ml samples at 3, 4, 6, 8, 10, 15, 20, 30, 40, and 50 minutes, and 4.5 ml at 60, 75, and 90 minutes. Decay-corrected whole-blood activity, the fraction of unchanged radiotracer in plasma, and the plasma/whole-blood ratio were calculated as described in previous studies [11, 13].

A 3D T1 weighted magnetic resonance imaging (MRI) scan was also obtained for each subject, using a 256 × 256 matrix. The machine used was either a 3-T Signa scanner (GE Medical Systems, Waukesha, WI; voxel size of 0.86 × 0.86 × 1.2 mm) or an Achieva 3-T MRI scanner (Philips Health Care, Andover, MA; voxel size of 1 × 0.94 × 0.94 mm).

Image analysis

The MR image from each subject was first coregistered to the average PET image using SPM5 (Wellcome Department of Imaging Neuroscience; University College London, UK). Then, the MR and PET images were normalized to a standard space (Montreal Neurological Institute space) using the transformation parameters from the MR images. Finally, a ROI template in the MNI space was used to derive the TACs of the regions used for kinetic modeling: thalamus (12.6 cm3), caudate (5.6 cm3), putamen (6.5 cm3), cerebellum (51.2 cm3), frontal (27.2 cm3), parietal (26.6 cm3), lateral temporal (25.0 cm3), occipital (31.2 cm3), anterior cingulate (7.5 cm3), and medial temporal (14.3 cm3) cortices.

Estimating the whole-blood TAC

Segmentation of carotid and background regions

Carotid and background regions of interest were manually drawn directly on the second PET frame (0.5 – 1 min) and then copied to all the frames (Figure 1). Carotid and background regions contained 110 ± 17 and 177 ± 31 voxels, respectively (i.e. the variation was slightly below 20%).

Figure 1.

Figure 1

Manually delineated carotid and background regions of interest.

Factor analysis

The factor analysis assumes that the dynamic image with n time frames and m voxels can be decomposed into a p number of time signals or factors (p<m), and TAC of each voxel can be represented with a linear combination of those p factors. In this particular case we assumed that two factors are sufficient – one for blood time signal and another for tissue time signal. Therefore, the image I of dimension n×m was formulated mathematically (Eq. 1) as the product of the factor matrix F of dimension n×2, which represents two factors, and the factor image matrix A of dimension 2×m, which describes the spatial distribution of the factors. Obviously, the radioactivity concentration cannot be negative and the contribution of blood or tissue signal to the voxels’ TAC cannot be negative. Therefore, the factor analysis results are physically viable only if the matrices F and A are non-negative.

I=FA+ε 1.
ΦFA(A,F)=W(IFA) 2.
Cb(t)={0;t<t1Qt1ti=13Aieδt(tτ)dτ;t1<t<t2Qt1t2i=13Aieδt(tτ)dτ;t2<t 3.
Cwb(t)=u×Cbf(t)v×Ctf(t) 4.
Σ=[Var(u)Cov(u,v)Cov(u,v)Var(v)] 5.
u=Cwb(ts)+v×Ctf(ts)Cbf(ts) 6.
ΦAP=[uumeanvvmean][Var(u)Cov(u,v)Cov(u,v)Var(v)]1[uumeanvvmean] 7.

Estimation of the factor images A and factors F was based on minimization of the least squares objective function ΦFA.

ΦFA(A,F)=W(IFA) 2.

The least squares objective function ΦFA is the second norm of the weighted difference between the matrix representation of the image I and the product of the factor matrix F and the factor image matrix A. Symbol ◦ denotes the entrywise product, where each element in the resulting matrix is product of the elements on the same position of the input matrices. Elements in weighting matrix W of Eq. 2 were aimed to be inversely proportional to the standard deviation of the corresponding element in the matrix representation of the image I. In this study we assumed spatially uniform noise with the standard deviation being inversely proportional to the square root of PET frame duration. The factor matrix F and the factor image matrix A were restricted to non-negative values to ensure physically viable results. In addition, the mean values of both factors were constrained to unity in order to avoid two additional free parameters for scaling.

Minimization of objective function ΦFA was done in Matlab (MathWorks, Natick, CA) using the routine fmincon that can find minimum of constrained nonlinear multivariable function [14]. The routine was set to use interior-point algorithm [1517] and the analytical gradient was provided. The factors (two columns) in matrix F were initialized with the normalized mean of the carotid and background TACs. Factor images in matrix A were initialized by solving the system of equations in Eq. 1, using the initialized matrix F. Non-negativity constraints and constrained mean values of both factors were specified in the routine optional arguments. If the blood-weighted factor appeared as the second factor in the resulting matrix F, the columns of the matrix were exchanged, along with the rows in the matrix A.

An intrinsic limitation of factor analysis is that the solution is not guaranteed to be unique [9]. In a preliminary version of this work we implemented an ambiguity correction procedure [18] which, in the present dataset, was found to have no impact on the shape of the input functions and was therefore removed from this work.

Removing the noise in blood-weighted factors

In order to minimize noise in the estimated TACs, blood-weighted factors were fitted with a three-exponential model. Although this step is not strictly necessary, it allows obtaining less noisy whole blood TACs. The model assumes that the injection of the radiotracer is of finite duration from t1 to t2. In this time-window, there is a constant flow (Q) of the radiotracer into the blood. After the peak, radiotracer concentration decreases following a three-exponential decay. The blood TAC can thus be calculated as the convolution between the square injection function and the three-exponential decay of radiotracer concentration in the blood.

Cb(t)={0;t<t1Qt1ti=13Aieδt(tτ)dτ;t1<t<t2Qt1t2i=13Aieδt(tτ)dτ;t2<t 3.

This model has eight free parameters that are set when fitting the model to the blood-weighted factor. First two parameters are time points t1 and t2. In addition, there are three pairs of parameters Ai and δi that determine exponential weight and decay constant. The value of Q is set to unity, because this parameter can be accounted for by adjusting the factors Ai in the three-exponential model. Having Q as an additional free parameter would cause the ambiguity of parameters.

Scaling factor analysis solutions

After de-noising by three-exponential model, the factors were scaled using the approach proposed by Sitek and colleagues [10, 19]. This step is necessary in order to account for the variability of TACs amplitudes, determined by the different partial volume effects among clinical subjects. Four voxels with the highest blood factor image values were selected within the carotid region. Factor analysis solutions were scaled so that the average factor image value among those four hottest voxels was equal to 1.

Generation of a priori input function data

The blood signal is reduced due to partial volume effects and contaminated by surrounding tissue activity. So, the actual concentration in blood Cwb(t) is:

Cwb(t)=u×Cbf(t)v×Ctf(t) 4.

where the Cbf(t) and Ctf(t) are the blood-weighted and tissue-weighted factors, respectively. If multiple measured values from blood-sampling are available, the coefficients u and v can be determined by linear least squares method, following the principle described by Chen and colleagues [6]. In the present study, the different ui and vi values from the various subjects were obtained by fitting the model in Eq. 4 to blood sampling data. Appropriate ui and vi values were averaged to obtain the average coefficients umean and vmean and the corresponding covariance matrix:

Σ=[Var(u)Cov(u,v)Cov(u,v)Var(v)] 5.

The average coefficients umean and vmean, their correlation and the corresponding covariance matrix, calculated on the data from all 10 subjects, are in Table 1. To validate the methodology, a priori input function data were generated from 9 subjects and tested on the remaining 1 – a leave-one-out approach.

Table 1.

Population data for coefficients u and v from the Eq. 4 in terms of mean, correlation between u and v, and covariance matrix. These figures are for all 10 subjects. When a leave-one-out procedure was used to validate the input function extraction, new figures, derived each time from 9 subjects, were used.

Coefficient Mean Correlation Covariance matrix

u v
u 2.40 0.36 0.037 0.017
v 0.66 0.017 0.061

To assess whether 9 subjects were sufficient to produce a stable estimate of the population parameters, we generated parameters using 3, 5 or 7 subjects, covering all possible combinations. The within-subject variations of the resulting AUC were 12.0 ± 3.8%, 4.5 ± 1.4% and 2.0 ± 0.8% when using 3, 5 and 7 subjects, respectively. This suggests that 9 subjects can provide an accurate estimate of the a priori input function data.

Generating whole-blood TACs

The relationship between coefficient u and v was determined using a single blood sample acquired at t=ts.

u=Cwb(ts)+v×Ctf(ts)Cbf(ts) 6.

As the coefficient v is unknown, it was determined by minimizing the objective function ΦAP that describes the deviation of coefficients u and v from the average coefficients umean and vmean, weighted with the corresponding inverse of the covariance matrix:

ΦAP=[uumeanvvmean][Var(u)Cov(u,v)Cov(u,v)Var(v)]1[uumeanvvmean] 7.

The time point ts was chosen at 40 minutes because it gave the highest Pearson’s correlation coefficient between the concentrations in plasma and the input function area under the curve in a previous study with [11C](R)-rolipram [20]. The final input functions were obtained by correcting for radiometabolites the whole blood TACs using four measured time-points, as previously described [2122].

Figures of merit

The accuracy of whole-blood TAC obtained with factor analysis and one blood sample was assessed 1) visually, 2) by calculating the relative error in AUCs for the whole-blood TACs at 5 min and 90 min post-injection, and 3) by comparing the IDIF-derived total distribution volume (VT) to the reference values obtained with full arterial input. VT was calculated in three different ways: with the Logan plot, with a two-tissue compartment model (2TCM) and with spectral analysis (SA). To our knowledge, this is the first time that SA is evaluated in association with IDIF. For 2TCM, blood volume was set at 5% and the curves were fitted for blood delay. For SA, the βj grid was defined by 100 components and a logarithmic distribution ranging from 0.01 to 1 (1/min). Kinetic analysis was performed with PMOD 3.1 (PMOD Technologies, Zurich, Switzerland) and the SAKE spectral analysis software [23].

Results

Visual analysis

The shape of the image-derived whole-blood TACs was consistently similar to the blood-sampling arterial whole-blood TAC. The peak was in most cases slightly underestimated, whereas the TACs at later times matched better to the reference arterial curves. shows the best and the worst example of image-derived whole-blood TACs obtained by factor analysis.

Area under the curve

Consistently with visual analysis, the IDIFs was less accurate in the estimation of the early part of the input function (AUC error over the first 5 minutes was −2.7 ± 8.7%), although the total AUC was more accurately estimated (mean error: 1.0 ± 4.3%).

To assess the robustness of the final TAC AUC with regards to operator's variability in drawing the ROIs, the manually-defined carotid and background regions were expanded or eroded by 20%. The resulting within-subject AUC variations were very small (0.67 ± 0.39%; highest variation: 1.6% in one subject).

Kinetic analysis

Logan-VT calculated with IDIF showed little bias and a moderate variability compared to three reference arterial Logan-VT values (mean error: −4.1 ± 6.4%; maximum error in one subject was 11%). When IDIF was used to calculate VT using 2TCM, the bias compared to 2TCM-VT obtained with the full arterial input was similar, but the variability substantially higher (mean error: −4.5 ± 9.6%, maximum error: 21%). Moreover, similarly to what we previously found [2425] individual rate constants showed important and unpredictable errors (Table 2). Finally, although SA-VT values were unbiased with regards to SA quantification with full input, they also displayed the highest variability (mean error: 0.4 ± 15.0%; maximum error: 28%).

Table 2.

Relative errors of kinetic parameters from compartmental modeling, Logan plot and spectral analysis.

2TCM Logan SA

Parameter K1 k2 k3 k4 VT VT VT
Mean error (SD) [%] 24.1 (34.8) 41.5 (71.7) 37.9 (9.4) 20.7 (69.1) −4.5 (9.6) −4.1 (6.4) 0.4 (15.0)

Discussion

The present study found that a factor analysis-derived IDIF and a single blood sample allows the estimation of an accurate input function, to be used for kinetic modeling with the Logan plot.

Typically, in brain PET studies, TACs from blood vessels cannot be directly used as a substitute for the arterial input function, because of the low spatial resolution of the machine. Due to this poor resolution, the blood signal in the vessel is not only underestimated, but it is also contaminated by the activity coming from the surrounding tissue, and each voxel would contain a mixture of the two signals in variable proportions. One possibility to recover the blood signal would be by using some blood measurements and a least squares analysis, as initially proposed by Chen and colleagues for [18F]-FDG [6] and subsequently by our laboratory for [11C](R)-rolipram [2021, 2526]. While only two samples are theoretically necessary to solve this system of equations, in practice more samples are necessary because of uncertainties in the sample measurements and the similarity between blood and tissue signals.

By using factor analysis we were able to obtain very reproducible blood-weighted and tissue-weighted TACs, which allowed us to generate the whole-blood TAC using a single blood sample and population data. Factor analysis does not aim to fully decompose the dynamic PET image into separate blood and tissue signals, but into a fixed combination, so that the amount of contamination of the blood-weighted factor from the tissue signal was constant, and therefore predictable. Notably, the amount of contamination of the blood-weighted factor with tissue signal depends on scanner resolution, vessel size, and degree of uptake of the radiotracer into the tissue. Therefore, the population-based approach to obtain whole-blood TAC from blood-weighted and tissue-weighted factors might be applicable only when using the scanner, the image reconstruction parameters, and the radiotracer from which the population parameters were derived.

In some cases, the technique might be simplified. For example, if the vessel was large enough or if the scanner resolution was higher, the partial volume correction factor could be set at one. Alternatively, if radiotracer tissue uptake was very low, the spill-over effect could be ignored. In these cases, the whole-blood TAC could be obtained directly from factor analysis results using only a single blood sample or only the population data. When a large blood pool is in the field-of-view, it might be already possible to estimate a fully non-invasive whole-blood TAC with the present technique. For brain PET studies, it would be possible after an improvement of current scanners resolution.

Although this technique minimizes the number of blood samples, it is not entirely blood-free. Therefore, an arterial line is still necessary. If the radioligand concentrations are the same in the artery and the vein, then a venous sample can effectively replace arterial blood. Unfortunately, [11C](R)-rolipram does not have a suitable artero-venous equilibrium [20], but some tracers do [6, 27]. Thus, we believe the methodology outlined in this work might have a practical interest for the quantification of tracers showing artero-venous equilibrium.

Using one blood sample allows reducing the workload to estimate each input function. This, however, comes with a cost, namely that the whole input function heavily depends on the accuracy of the radioactivity measurement of that single blood sample. Using the average of two or three samples, instead of one, would likely ensure an optimal estimation of the input function even if one of the samples happens to be wrongly measured. Similarly, in a previous study of 51 subjects scanned with [11C](R)-rolipram, three had inaccurate measurements in a single late blood sample. These errors were sufficient to cause a significant VT estimation error when using a population-based input function, but the impact of these wrong measurements was minimized by using the average of three blood samples [20].

Our technique has two limitations, which are however commonly shared by all IDIF techniques:

  1. Only the whole-blood TAC can be estimated from PET images. To obtain the parent plasma concentration (i.e. the actual input function), radiometabolite correction must be performed. In the present study, radiometabolite correction was performed using some blood samples. For some tracers, the use of a standardized metabolite curve might provide usable results [2830]. This approach however should be validated for each radioligand.

  2. The use of IDIF techniques essentially restricts the choice of kinetic modeling approaches to graphical analyses, such as the Logan or Patlak plots. In fact, in the present study, compartmental modeling and spectral analysis provided less accurate results than Logan. Although the average VT values showed little or no bias, the variability was substantially higher. These results echo those we previously obtained using different tracers and different machines [2425]. The relative robustness of the Logan plot is explained by the fact that this modeling technique relies on the total AUC of the input function, and is therefore less sensitive to the unavoidable errors in the estimation of the peak that are usually associated to the use of IDIFs.

Conclusions

An image-derived input function for [11C](R)-rolipram brain PET studies can be obtained with factor analysis, a priori information of radioligand’s kinetic behavior and PET scanner characteristics and a single blood sample.

Figure 2.

Figure 2

Two examples of image-derived whole-blood TACs obtained by factor analysis (solid line) compared to the reference arterial samples (dots). Inner plots magnify the first 10 minutes. The (A) plot is for the subject in whom factor analysis provided the most accurate estimation, while the (B) plot shows the worst case (AUC error of 7.7%). Please note that the peak estimation was better in the (B) curve. This underscores the fact that the peak contributes little to the estimation of the whole AUC for tracers like [11C](R)-rolipram, whose wash-out from the vascular compartment is very slow.

Acknowledgements

The authors thank Drs. Robert B. Innis, Masahiro Fujita, Sami S. Zoghbi, Victor W. Pike and the staff of the Molecular Imaging Branch of the NIMH for kindly providing the [11C](R)-rolipram images. Ioline Henter (NIMH) provided excellent editorial assistance.

Funding Source: This work was supported by the Slovenian Research Agency (ARRS); the Centre of Excellence for Biosensors, Instrumentation, and Process Control; the European Union; the European Regional Development Fund; the Ministry of Higher Education, Science and Technology of the Republic of Slovenia; and the Intramural Research Program of the National Institute of Mental Health, National Institutes of Health (IRP-NIMH-NIH).

Footnotes

Conflict of Interest Notification

There is no conflict of interests.

References

  • 1.de Geus-Oei LF, Visser EP, Krabbe PF, van Hoorn BA, Koenders EB, Willemsen AT, Pruim J, et al. Comparison of image-derived and arterial input functions for estimating the rate of glucose metabolism in therapy-monitoring 18F-FDG PET studies. J Nucl Med. 2006;47:945–949. [PubMed] [Google Scholar]
  • 2.Zanotti-Fregonara P, Chen K, Liow JS, Fujita M, Innis RB. Image-derived input function for brain PET studies: many challenges and few opportunities. J Cereb Blood Flow Metab. 2011;31:1986–1998. doi: 10.1038/jcbfm.2011.107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Fung EK, Carson RE. Cerebral blood flow with [15O]water PET studies using an image-derived input function and MR-defined carotid centerlines. Phys Med Biol. 2013;58:1903–1923. doi: 10.1088/0031-9155/58/6/1903. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Mourik JE, Lubberink M, Klumpers UM, Comans EF, Lammertsma AA, Boellaard R. Partial volume corrected image derived input functions for dynamic PET brain studies: methodology and validation for [11C]flumazenil. Neuroimage. 2008;39:1041–1050. doi: 10.1016/j.neuroimage.2007.10.022. [DOI] [PubMed] [Google Scholar]
  • 5.Naganawa M, Kimura Y, Ishii K, Oda K, Ishiwata K, Matani A. Extraction of a plasma time-activity curve from dynamic brain PET images based on independent component analysis. IEEE Trans Biomed Eng. 2005;52:201–210. doi: 10.1109/TBME.2004.840193. [DOI] [PubMed] [Google Scholar]
  • 6.Chen K, Bandy D, Reiman E, Huang SC, Lawson M, Feng D, Yun LS, et al. Noninvasive quantification of the cerebral metabolic rate for glucose using positron emission tomography, 18F-fluoro-2-deoxyglucose, the Patlak method, and an image-derived input function. J Cereb Blood Flow Metab. 1998;18:716–723. doi: 10.1097/00004647-199807000-00002. [DOI] [PubMed] [Google Scholar]
  • 7.Barber DC. The use of principal components in the quantitative analysis of gamma camera dynamic studies. Phys Med Biol. 1980;25:283–292. doi: 10.1088/0031-9155/25/2/008. [DOI] [PubMed] [Google Scholar]
  • 8.Di Paola R, Bazin JP, Aubry F, Aurengo A, Cavailloles F, Herry JY, Kahn E. Handling of Dynamic Sequences in Nuclear-Medicine. IEEE Trans Nucl Sci. 1982;29:1310–1321. [Google Scholar]
  • 9.Houston AS. The effect of apex-finding errors on factor images obtained from factor analysis and oblique transformation. Phys Med Biol. 1984;29:1109–1116. doi: 10.1088/0031-9155/29/9/007. [DOI] [PubMed] [Google Scholar]
  • 10.Sitek A, Gullberg GT, Huesman RH. Correction for ambiguous solutions in factor analysis using a penalized least squares objective. IEEE Trans Med Imaging. 2002;21:216–225. doi: 10.1109/42.996340. [DOI] [PubMed] [Google Scholar]
  • 11.Fujita M, Zoghbi SS, Crescenzo MS, Hong J, Musachio JL, Lu J-Q, Liow J-S, et al. Quantification of brain phosphodiesterase 4 in rat with (R)-[11C]rolipram-PET. NeuroImage. 2005;26:1201–1210. doi: 10.1016/j.neuroimage.2005.03.017. [DOI] [PubMed] [Google Scholar]
  • 12.Fujita M, Hines CS, Zoghbi SS, Mallinger AG, Dickstein LP, Liow JS, Zhang Y, et al. Downregulation of brain phosphodiesterase type IV measured with 11C-(R)-rolipram positron emission tomography in major depressive disorder. Biol Psychiatry. 2012;72:548–554. doi: 10.1016/j.biopsych.2012.04.030. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zoghbi SS, Shetty HU, Ichise M, Fujita M, Imaizumi M, Liow JS, Shah J, et al. PET imaging of the dopamine transporter with 18F-FECNT: a polar radiometabolite confounds brain radioligand measurements. J. Nucl. Med. 2006;47:520–527. [PubMed] [Google Scholar]
  • 14.Grace A MathWorks Inc. Optimization toolbox for use with MATLAB. Natick, Mass: MathWorks; 1992. [Google Scholar]
  • 15.Byrd RH, Gilbert JC, Nocedal J. A trust region method based on interior point techniques for nonlinear programming. Mathematical Programming. 2000;89:149–185. [Google Scholar]
  • 16.Byrd RH, Hribar ME, Nocedal J. An interior point algorithm for large-scale nonlinear programming. Siam Journal on Optimization. 1999;9:877–900. [Google Scholar]
  • 17.Waltz RA, Morales JL, Nocedal J, Orban D. An interior algorithm for nonlinear optimization that combines line search and trust region steps. Mathematical Programming. 2006;107:391–408. [Google Scholar]
  • 18.Simoncic U, Zanotti-Fregonara P. Image-derived input function with factor analysis and a priori information. J Nucl Med. 2014;55:2024. doi: 10.1097/MNM.0000000000000231. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Sitek A, Di Bella EV, Gullberg GT. Factor analysis with a priori knowledge--application in dynamic cardiac SPECT. Phys Med Biol. 2000;45:2619–2638. doi: 10.1088/0031-9155/45/9/314. [DOI] [PubMed] [Google Scholar]
  • 20.Zanotti-Fregonara P, Hines CS, Zoghbi SS, Liow JS, Zhang Y, Pike VW, Drevets WC, et al. Population-based input function and image-derived input function for [(1)(1)C](R)-rolipram PET imaging: methodology, validation and application to the study of major depressive disorder. Neuroimage. 2012;63:1532–1541. doi: 10.1016/j.neuroimage.2012.08.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Zanotti-Fregonara P, Zoghbi SS, Liow JS, Luong E, Boellaard R, Gladding RL, Pike VW, et al. Kinetic analysis in human brain of [11C](R)-rolipram, a positron emission tomographic radioligand to image phosphodiesterase 4: a retest study and use of an image-derived input function. Neuroimage. 2011;54:1903–1909. doi: 10.1016/j.neuroimage.2010.10.064. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Lyoo CH, Zanotti-Fregonara P, Zoghbi SS, Liow JS, Xu R, Pike VW, Zarate CA, Jr, et al. Image-derived input function derived from a supervised clustering algorithm: methodology and validation in a clinical protocol using [(11)c](R)-rolipram. PLoS One. 2014;9:e89101. doi: 10.1371/journal.pone.0089101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Veronese M, Rizzo G, Turkheimer FE, Bertoldo A. SAKE: A new quantification tool for positron emission tomography studies. Comput Methods Programs Biomed. 2013;111:199–213. doi: 10.1016/j.cmpb.2013.03.016. [DOI] [PubMed] [Google Scholar]
  • 24.Zanotti-Fregonara P, Fadaili el M, Maroy R, Comtat C, Souloumiac A, Jan S, Ribeiro MJ, et al. Comparison of eight methods for the estimation of the image-derived input function in dynamic [(18)F]-FDG PET human brain studies. J Cereb Blood Flow Metab. 2009;29:1825–1835. doi: 10.1038/jcbfm.2009.93. [DOI] [PubMed] [Google Scholar]
  • 25.Zanotti-Fregonara P, Liow JS, Fujita M, Dusch E, Zoghbi SS, Luong E, Boellaard R, et al. Image-Derived Input Function for Human Brain Using High Resolution PET Imaging with [11C](R)-rolipram and [11C]PBR28. Plos One. 2011;6:e17056. doi: 10.1371/journal.pone.0017056. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Zanotti-Fregonara P, Liow JS, Comtat C, Zoghbi SS, Zhang Y, Pike VW, Fujita M, et al. Image-derived input function in PET brain studies: blood-based methods are resistant to motion artifacts. Nucl Med Commun. 2012;33:982–989. doi: 10.1097/MNM.0b013e328356185c. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Hahn A, Nics L, Baldinger P, Ungersbock J, Dolliner P, Frey R, Birkfellner W, et al. Combining image-derived and venous input functions enables quantification of serotonin-1A receptors with [carbonyl-11C]WAY-100635 independent of arterial sampling. Neuroimage. 2012;62:199–206. doi: 10.1016/j.neuroimage.2012.04.047. [DOI] [PubMed] [Google Scholar]
  • 28.Lammertsma AA, Bench CJ, Hume SP, Osman S, Gunn K, Brooks DJ, Frackowiak RS. Comparison of methods for analysis of clinical [11C]raclopride studies. J Cereb Blood Flow Metab. 1996;16:42–52. doi: 10.1097/00004647-199601000-00005. [DOI] [PubMed] [Google Scholar]
  • 29.Mitkovski S, Villemagne VL, Novakovic KE, O'Keefe G, Tochon-Danguy H, Mulligan RS, Dickinson KL, et al. Simplified quantification of nicotinic receptors with 2[18F]F-A-85380 PET. Nucl Med Biol. 2005;32:585–591. doi: 10.1016/j.nucmedbio.2005.04.013. [DOI] [PubMed] [Google Scholar]
  • 30.Zanotti-Fregonara P, Maroy R, Peyronneau MA, Trebossen R, Bottlaender M. Minimally invasive input function for 2-(18)F-fluoro-A-85380 brain PET studies. Eur J Nucl Med Mol Imaging. 2012;39:651–659. doi: 10.1007/s00259-011-2004-9. [DOI] [PubMed] [Google Scholar]

RESOURCES