Abstract
Objective:
To investigate reproducibility of fluorine-18 fludeoxyglucose (18F-FDG) uptake on 18F-FDG positron emission tomography (PET)/CT and 18F-FDG PET/MR scans in patients with head and neck squamous cell carcinoma (HNSCC).
Methods:
30 patients with HNSCC were included in this prospective study. The patients were scanned twice before radiotherapy treatment with both PET/CT and PET/MR. Patients were scanned on the same scanners, 3 days apart and according to the same protocol. Metabolic tumour activity was measured by the maximum and peak standardized uptake value (SUVmax and SUVpeak, respectively), and total lesion glycolysis from the metabolic tumour volume defined from ≥50% SUVmax. Bland–Altman analysis with limits of agreement, coefficient of variation (CV) from the two modalities were performed in order to test the reproducibility. Furthermore, CVs from SUVmax and SUVpeak were compared. The area under the curve from cumulative SUV–volume histograms were measured and tested for reproducibility of the distribution of 18F-FDG uptake.
Results:
24 patients had two pre-treatment PET/CT scans and 21 patients had two pre-treatment PET/MR scans available for further analyses. Mean difference for SUVmax, peak and mean was approximately 4% for PET/CT and 3% for PET/MR, with 95% limits of agreement less than ±20%. CV was small (5–7%) for both modalities. There was no significant difference in CVs between PET/CT and PET/MR (p = 0.31). SUVmax was not more reproducible than SUVpeak (p = 0.09).
Conclusion:
18F-FDG uptake in PET/CT and PET/MR is highly reproducible and we found no difference in reproducibility between PET/CT and PET/MR.
Advances in knowledge:
This is the first report to test reproducibility of PET/CT and PET/MR.
Functional imaging with fluorine-18 fludeoxyglucose positron emission tomography combined with CT (18F-FDG PET/CT) has been shown to be useful for prognostication of head and neck squamous cell carcinoma (HNSCC),1–3 and the use of 18F-FDG PET/CT has also been shown to reduce interobserver variability in target delineation for radiotherapy.4,5 Furthermore, 18F-FDG PET/CT can identify regions of the tumour with a high risk of relapse, leading to the idea that 18F-FDG uptake might be a target for dose painting.6,7 Finally, 18F-FDG PET/CT may be used in response evaluation.8,9 Maximum standardized uptake value (SUVmax) has for many years been the main uptake measurement in prognostic studies for various malignancies. More recent studies have focused on demonstrating prognostic value of PET/CT-based volumetric parameters such as metabolic tumour volume (MTV) and total lesion glycolysis (TLG). MTV is the sum of the volume of voxels with standard uptake value (SUV) exceeding a certain threshold value in a tumour, and TLG is calculated by multiplying MTV and the mean standardized uptake value (SUVmean) of the MTV. These volume-based PET parameters have increasingly gained interest and have been reported to be significant prognostic factors for various malignancies including HNSCC.10–13 18F-FDG PET/CT is currently not routinely recommended as a diagnostic tool in HNSCC except in very specific situations,14 but reproducibility of the 18F-FDG signal is a prerequisite for a more widespread use of 18F-FDG PET for the above-mentioned indications. Yet, only a few studies of the reproducibility of 18F-FDG PET/CT exist8,15–21 and none of these studies includes patients with HNSCC.
MRI is gaining acceptance as an imaging modality for oncology as it offers superior soft-tissue contrast compared with CT alone, and it has been suggested that information from PET/CT and MR is complementary in head and neck cancer.22 The introduction of the integrated PET/MR scanner offers a unique opportunity to combine the high soft-tissue contrast of MR with the functional imaging from PET within a single imaging session. PET/MR is still in its infancy, but the combined modality imaging is potentially useful in the management of patients with HNSCC.22–28 However, the same criteria of reproducibility as with PET/CT should be upheld by this new modality. The purpose of this prospective test–retest study is to assess the reproducibility of both 18F-FDG PET/CT and 18F-FDG PET/MR in a homogenous cohort of patients with HNSCC.
METHODS AND MATERIALS
Patients
30 patients with HNSCC in the pharynx referred to radiotherapy were included in this prospective study. Patients with diabetes were excluded. All patients gave informed consent, and the study was approved by the local ethics committee, approval number H-3-2012-072. The patients were instructed to fast for at least 6 h before the examination. All patients were scanned prior to treatment with repeated scans on both modalities, exactly 3 days apart.
Fluorine-18 fludeoxyglucose positron emission tomography/CT
All PET/CT scans were performed on the same biograph mCT 64 (Siemens Medical Solutions, Malvern, PA), and patients were immobilized with an individually moulded thermoplastic mask in the supine position on a flat scanner couch. On Day 1, the scan was a whole-body PET/CT (apex of the skull to mid-thigh), and on Day 2, the scan covered the apex of the skull to apex of the lung. All CT images were acquired as diagnostic quality spiral CTs using 120 kVp and intravenous contrast agent, except for one patient who had renal insufficiency. A multibed PET scan with 2 min per bed position was performed after the CT, and the patients were scanned 60 min after 18F-FDG injection (4 MBq kg−1). In case of a delay on Day 1, a match of the actual time from 18F-FDG injection to scan was attempted on the second scanning day.
PET/CT images were reconstructed using an iterative three-dimensional ordinary poisson-ordered subset expectation maximization (OP-OSEM) algorithm using resolution modelling of the point spread function (PSF) and time of flight (TOF) with 2-mm gaussian post filter, which theoretically provides superior image reconstruction29–31 and is used in our clinical routine. However, on PET/MR, TOF is not available and PSF not comparable. To improve cross-modality comparison, an additional reconstruction of PET/CT data was carried out without PSF and TOF and with 4-mm gaussian post-filter. The two sets of PET data were denoted as “PET/CT (PSF/TOF)” and “PET/CT (OSEM)”, respectively. For both PET reconstructions, 3 iterations, 21 subsets and 400 × 400 matrices (2.0 × 2.0-mm voxels) were used.
Fluorine-18 fludeoxyglucose positron emission tomography/MR
The patients were scanned with an integrated PET/MR system (Siemens Biograph mMR) with a 3-T magnet using a head and neck coil. The MR sequences acquired were T2 weighted as well as a Dixon VIBE sequence for attenuation correction purposes. PET was performed as a single-bed, 20-min acquisition. The patients were planned to be scanned 100–120 min after 18F-FDG injection (after the PET/CT scan). A match of the 18F-FDG injection to PET/MR time interval on the second scanning day to the actual time interval on the first scanning day was attempted. PET data were reconstructed using OP-OSEM with 3 iterations, 24 subsets and 4-mm gaussian post-filter into 344 × 344 matrices to match the PET/CT reconstructions as close as possible. Resolution modelling (PSF) was not applied.
Imaging analysis and measurements of standardized uptake value
Lesions were counted from PET/CT (PSF/TOF) and PET/MR to make a simple test–retest measurement of the two modalities. The region of interest was defined as oropharynx, and in one case hypopharynx, including neck Levels I–V in case of node metastasis. An isocontour corresponding to 50% of SUVmax was generated and subsequently adapted by excluding regions with physiological 18F-FDG uptake by an experienced nuclear medicine physician (BMF). Only one MTV per patient, defined from SUVmax, was used since the main purpose was to investigate test–retest reproducibility and if the number of target lesions varies across patients, the correlation could be inflated.21
Metabolic activity (corrected for injected dose, decay and patient weight) was measured as SUVmax, peak standardized uptake value (SUVpeak), SUVmean, the volume encompassed by the 50% SUVmax isocontour (MTV50%) and TLG from this volume. The cumulative SUV–volume histograms (CSH)32 were plotted to test reproducibility of the distribution of 18F-FDG uptake (heterogeneity) in the tumour by calculating the area under the curve (AUC).33 The SUV and volume metrics were extracted from the software package Mirada XD3 (MIRADA Medical, Oxford, UK).
Statistical analysis
Statistical analysis was performed in SPSS® v. 19 (SPSS Inc., Chicago, IL) unless otherwise noted. Reproducibility of each measure was assessed from the difference between the two scans (Scan 1 − Scan 2), and coefficients of variation (CVs) as CVs = σ/μ (σ, spread; μ, mean) were assessed. For further analysis, visual assessment of P-P plots and Kolmogorov–Smirnov and Shapiro–Wilk tests were used to assess the distribution of SUV and volume measurements, and a log transformation was used where deemed necessary to match a normal distribution. Bland–Altman plots with limits of agreement from Bland–Altman analysis34 were generated. Variance component analysis was performed using the VARCOMP procedure in SPSS to compare variance between measurements to variance between patients (corresponding to signal-to-noise comparison). Wilcoxon signed rank test was used to test for difference in CVs between SUVmax and SUVpeak in PET/CT (PSF/TOF) and to test the difference in CVs of SUVmax from PET/CT (OSEM) and PET/MR.
RESULTS
Table 1 describes the number of scans that the 30 patients received. Figure 1 shows a flowchart of inclusion. Three did not receive the second scan; one patient was ill with flu, one dropped out and one patient forgot to come for the second scans. The remaining 27 had carcinoma in the oropharynx, but 3 patients had no pathological 18F-FDG uptake and were excluded from further analysis; Table 1 shows further characteristics of the remaining 24 patients. 21 of the 24 patients were also scanned twice with PET/MR, however, the data from 1 scan could not be reconstructed leaving 20 patients with 2 baseline PET/MR. All patients were scanned according to the same protocol, and the mean time difference between 18F-FDG injection and scanning (uptake time Scan 1 − uptake time Scan 2) was −1.7 min (range, −10 to 4 min) for PET/CT and −5.6 min (range, −49 to 8 min) minutes for PET/MR. The difference in uptake time was very small, but on average, the uptake time was a little longer on the second day because an unplanned delay on Day 1 could be matched by the same delay on Day 2, whereas it was not possible to compensate if an unplanned delay happened on the second day.
Table 1.
PET/CT | PET/MR | Number of patients (%) |
---|---|---|
Number of scans | ||
2 | 2 | 24 |
2 | 1 | 3 |
1 | 2 | 0 |
1 | 1 | 3 |
Patient characteristics of the 24 included patients | ||
Age (years) | ||
Median | 64 | |
Range | 44–74 | |
Sex | ||
Male | 20 (83.3) | |
Female | 4 (16.7) | |
T-stage | ||
1 | 0 (0) | |
2 | 17 (70.8) | |
3 | 7 (29.2) | |
4 | 0 (0) | |
N-stage | ||
0 | 9 (37.5) | |
1 | 2 (8.3) | |
2a | 1 (4.2) | |
2b | 5 (20.8) | |
2c | 6 (25.0) | |
3 | 1 (4.2) |
PET, positron emission tomography.
Table 2, shows the result for lesion counting, and Table 3 shows mean values for 18F-FDG measurements and volumes for both PET/CT and PET/MR. Mean difference in injected 18F-FDG activity between the scans was 0.01 MBq kg−1.
Table 2.
Number of lesions | (1) Scan |
(2) Scan |
||
---|---|---|---|---|
PET/CT (PSF) | PET/MR | PET/CT (PSF) | PET/MR | |
Lesions | 51 | 40 | 53 | 39 |
PSF, point spread function.
Fewer lymph nodes were noted in PET/MR owing to a smaller field of view.
Table 3.
18F-FDG uptake measurements | Mean PET/CT (PSF/TOF) | Mean PET/CT (OSEM) | Mean PET/MR |
---|---|---|---|
SUVmax | 20.4 | 15.0 | 16.1 |
SUVpeak | 14.0 | 12.1 | 13.3 |
SUVmean | 13.3 | 10.2 | 11.0 |
MTV50% | 5.0 | 7.1 | 6.6 |
Total lesion glycolysis | 74.8 | 80.9 | 81.5 |
MTV50%, metabolic tumour volume from threshold of ≥50% of SUVmax; OSEM, ordered subset expectation maximization; PSF, point spread function; SUVmax, maximum standardized uptake value; SUVmean, mean standardized uptake value; SUVpeak, peak standardized uptake value; TOF, time of flight.
MTV50% and TLG needed to be log transformed to follow normal distribution as assessed visually and from statistical tests (p < 0.03) using Kolmogorov–Smirnov and Shapiro–Wilk tests. After log transformation, there was no deviation from normality. We observed an increase in variation of absolute SUV difference with increasing SUV of the two pre-treatment PET/CT scans. Consequently, the difference in Bland–Altman analysis is reported in percent of SUV, which did not appear to depend on the pre-treatment SUV. Likewise, the Bland–Altman plots were generated with difference in percent to emphasize potential difference in variance, i.e. increasing variance with increasing SUVs. The Bland–Altman plots for SUVmax and SUVpeak for PET/CT (PSF/TOF, OSEM) and for PET/MR can be seen from Figure 2, and Bland–Altman plots for TLG and MTV50% can be seen from Figure 3. Results from Bland–Altman analysis are presented in Table 4, and Table 5 depicts CVs and the results from the variance component analysis from PET/CT (PSF/TOF, OSEM) and PET/MR. SUVmax, SUVmean and SUVpeak measures were generally very reproducible with limits of agreement within approximately ±20% for all measures. MTV and TLG measures were also very reproducible with limits of agreement within approximately ±25%.
Table 4.
18F-FDG uptake measurements | PET/CT (PSF/TOF) |
PET/CT (OSEM) |
PET/MR |
||||||
---|---|---|---|---|---|---|---|---|---|
Mean difference | SD | Limits of agreement (%) | Mean difference | SD | Limits of agreement (%) | Mean difference | SD | Limits of agreement (%) | |
SUVmax | −3.7 | 10.3 | (−24.4, 16.9) | −2.8 | 6.3 | (−15.5, 9.8) | 3.2 | 8.6 | (−14.0, 20.4) |
SUVpeak | −4.1 | 7.3 | (−18.7, 10.6) | −4.5 | 6.9 | (−18.2, 9.2) | 2.2 | 8.0 | (−13.9, 18.2) |
SUVmean50% | −3.5 | 9.5 | (−22.4, 15.5) | −4.2 | 7.1 | (−18.4, 10.0) | 3.1 | 7.8 | (−12.5, 18.7) |
MTV50% | N/A | N/A | (−29.7, 39.5) | N/A | N/A | (−30.9, 48.5) | N/A | N/A | (−27.7, 28.2) |
TLG50% | N/A | N/A | (−29.6, 29.7) | N/A | N/A | (−30.5, 36.3) | N/A | N/A | (−24.2, 30.1) |
AUC unnormalized | −3.5 | 9.6 | (−22.7, 15.6) | −4.2 | 7.1 | (−18.4, 10.0) | 3.1 | 7.9 | (−12.7, 18.9) |
AUC normalized | 0.2 | 2.7 | (−5.1, 5.5) | −1.4 | 2.5 | (−6.3, 3.6) | −0.1 | 1.5 | (−3.2, 2.9) |
AUC, area under the curve; MTV50%, metabolic tumour volume from threshold of ≥50% of SUVmax; OSEM, ordered subset expectation maximization; PET, positron emission tomography; PSF, point spread function; SD, standard deviation; SUVmax, maximum standard uptake value; SUVpeak, peak standard uptake value; TLG50%, total lesion glycolysis calculated as the product from MTV50% and SUVmean from MTV50%; TOF, time of flight.
The results are presented as percentages, except for MTV and TLG, where a log transformation was applied to fulfil the normal distribution assumption. The mean difference (SD) for lnMTV for PET/CT (PSF/TOF) was −0.01 (0.17), for PET/CT (OSEM) was 0.013 (0.19) and for PET/MR was −0.04 (0.14). The mean difference (SD) for lnTLG for PET/CT (PSF/TOF) was −0.01 to 0.05 (0.15), for PET/CT (OSEM) was −0.03 (0.17) and for PET/MR was −0.01 (0.13). Limits of agreement were assessed from mean difference ±2 SD. For MTV and TLG, the limits of agreement were transformed back using the exponential function.
Table 5.
18F-FDG uptake measurements | PET/CT (PSF/TOF) CV (%) | Variance component within subjects (%) | PET/CT (OSEM) CV (%) | Variance component within subjects (%) | PET/MR CV (%) | Variance component within subjects (%) |
---|---|---|---|---|---|---|
SUVmax | 7.6 | 3.4 | 4.8 | 1.5 | 6.4 | 3.6 |
SUVpeak | 5.8 | 1.7 | 5.7 | 1.6 | 5.7 | 2.6 |
SUVmean50% | 7.0 | 2.6 | 5.7 | 1.6 | 5.8 | 2.7 |
MTV50%a | 11.8 | 1.8 | 11.2 | 2.7 | 9.2 | 2.5 |
TLG50%a | 11.0 | 1.0 | 11.7 | 1.6 | 10.2 | 1.0 |
AUC unnormalized | 7.1 | 2.6 | 5.75 | 1.6 | 5.87 | 2.7 |
AUC normalized | 1.9 | 25.9 | 2.0 | 17.9 | 1.1 | 4.2 |
AUC, area under the curve; MTV50%, metabolic tumour volume from threshold of ≥50% of SUVmax; SUVmax, maximum standard uptake value; SUVpeak, peak standard uptake value; TLG50%, total lesion glycolysis calculated as the product from MTV50% and SUVmean from MTV50%; TOF, time of flight.
Variance of components within subjects was assessed from a variance component analysis and defined as total variance − variance between subjects.
Variance component was carried out after log transformation.
Figure 4 shows an example of a PET/CT (PSF) and a PET/MR scan from the patient with the largest difference in MTV (MTV50%) on PET/CT. Figure 5 shows an example of a patient with same tumour size, but with a larger difference in MTV50% on PET/MR than on PET/CT. The differences of 35% (Figure 4a,b) and of 13% (Figure 5c,d) are not easily distinguished from the two figures since the blue contour (MTV50%) obviously marks the same area in both PET/CT and PET/MR on both scan days.
Overall, the distribution of 18F-FDG uptake was very reproducible for both PET/CT and PET/MR reported as AUC from CSH, and the mean difference in AUC can be seen from Table 4. The small percentage variability for the AUC measures (Table 5) was not reflected in better variance components; 74% and 82% of the total variation in the data set for PET/CT was explained by interpatient variance. For all other measures, >95% of the total variance was owing to interpatient variance (Table 4), which illustrates small CVs for all SUV measurements (approximately 5–7%) and for MTV and TLG for both PET/CT and PET/MR.
There was no significant difference in CVs of SUVmax and SUVpeak in PET/CT (PSF/TOF) (p = 0.09); neither was there any significant difference in CVs of SUVmax from PET/CT (OSEM) and from PET/MR (p = 0.31).
DISCUSSION
This is, to our knowledge, the first study of reproducibility of 18F-FDG uptake in 18F-FDG PET/CT in patients with HNSCC and the very first to test reproducibility of 18F-FDG uptake values on PET/MR. There were no distinct difference, in lesions countered by the two PET/CT scans nor by the two PET/MR scans, except in one patient where lymph nodes were counted on the second scan, but not on the first scan, and this was only in PET/CT. In the remaining 20 patients, difference in lesions was because lymph nodes conglomerate was counted as two lesions in one scan and one lesion in the other scan. We observed no difference in primary tumours and no difference in lymph nodes side of the neck, i.e. ipsilateral and contralateral. The difference between lesions found in PET/CT and PET/MR were owing to the smaller field of view in PET/MR. 18F-FDG uptake measured as SUVs (SUVmax, SUVpeak, SUVmean) or as AUC from cumulative CSH is demonstrated to be very reproducible for both PET/CT and PET/MR.
Functional imaging with PET/CT, and in the future perhaps also with PET/MR, is increasingly used in the management of HNSCC for staging, radiotherapy planning and/or treatment response evaluation. 18F-FDG PET/CT is useful in patients with HNSCC for prognostication, target delineation, response assessment and possibly for dose painting. However, for appropriate implementation of all four applications, the reproducibility of 18F-FDG uptake must be known. In particular, if 18F-FDG PET images are used for dose painting with dose modification according to 18F-FDG uptake, the distribution of 18F-FDG uptake must be stable and reproducible before the start of radiotherapy.
Our study shows reasonably good reproducibility (i.e., a scan-to-scan variation of 20%) of the 18F-FDG uptake measurements on both PET/CT (PSF) and PET/MR. Of note, the present study represents the ideal scenario where the same scanners and protocols are used in the comparisons. If different scanners are used, additional margins for uncertainty may have to be included.35 Figure 4 depicts the worst-case scenario in difference in MTV50% in PET/CT (PSF), and such difference is not easy to see from the figure and the large relative difference is also owing to the actual tumour size, where a difference in 1 cm3 would yield a relative difference of 10%. Furthermore, the blue contour obviously marks the same area in both scans on both PET/CT (PSF) and PET/MR.
Most of the SUV metrics do not assess a change in tracer uptake within the entire tumour, and there is increasing awareness of intratumoral heterogeneity in 18F-FDG uptake.36 The distribution of 18F-FDG uptake assessed by CSH (with or without normalization to SUVmax) is very reproducible for both PET/CT and PET/MR. However, although the small CV for the normalized AUC seems very appealing, it should be noted from Table 5 that there is a relatively small interpatient variance component. This is because the good reproducibility of the normalized AUC (low noise) is outweighed by a narrow range of variation between patients (low signal).
The variability in the present reproducibility study is within the test–retest variability reported in the meta-analysis by de Langen et al.8 They reported that a difference should exceed both absolute and relative thresholds of 3.1% and 25% for SUVmax and 1.2% and 20% for SUVmean to be called a “true difference”. According to the proposed PERCIST response criteria,9 a change in ±30% in SULpeak (lean body mass corrected SUV) and an absolute change, at least 0.8 SUL, is classified as progression or remission. The worst-case scenario patient from Figure 4 had an increase in SUVmax from 14.7 to 19.2 (a change of 4.5 corresponding to 27%), however, the change in SUVpeak and SUVmean was <20%. Of note, the increase in SUVmax actually yields a decrease in MTV50%, illustrating that it is challenging to separate a metabolic response assessment from a structural response assessment, however, this discussion is beyond the scope of this article.
All patients in the present test–retest study would have been assessed as stable disease using the PERCIST criteria, although metabolic activity was assessed as SUV and not SUL. They would be expected to be similar since it is unlikely that a patient would change composition of body fat and lean tissue without a change in weight within 3 days. It must of course be noted that our patient cohort is very homogenous and only includes patients with HNSCC, and it is possible that a larger variation would be encountered in clinical practice with different tumour types. PERCIST proposes that no more than five lesions with a minimal 18F-FDG uptake of approximately 2.59 (calculated as 1.5 × the liver SUVmean + 2 standard deviations) be evaluated. In contrast to PERCIST, only one lesion per patient was assessed in the present test–retest study to avoid introducing bias in the evaluation of reproducibility if the number of lesions varied from patient to patient.
We observed an increase in absolute value variation with increasing SUVs. Consequently, reporting or using absolute threshold values for SUVs would result in too narrow limits at low SUV and too wide limits at high SUV, and this is not recommended in PERCIST criteria9 nor in the meta-analysis mentioned above.
Our scans were obtained 3 days apart, and therefore the fundamental assumption behind the study was that there should be no appreciable difference in tumour characteristics between the two scans, i.e. the true result is “stable disease”. Reassuringly, the values from PERCIST and the de Langen study would indeed classify all our patients as stable disease.
CONCLUSION
The 18F-FDG uptake in 18F-FDG PET/CT scans and 18F-FDG PET/MR scans are both very reproducible in patients with HNSCC. The uptake in PET/MR scans is as reproducible as in PET/CT scans. The analyses did not indicate that SUVpeak is more reproducible than SUVmax in neither PET/CT nor PET/MR. Pre-therapy 18F-FDG uptake is sufficiently stable in individual tumours to allow treatment modification on this basis.
FUNDING
The corresponding author has received a grant from the Arvid Nilssons Foundation. Supported by the Global Excellence in Health program of the Capital Region of Denmark.
Acknowledgments
ACKNOWLEDGMENTS
The authors thank Søren M Bentzen for suggesting the variance component analysis.
REFERENCES
- 1.Xie P, Li M, Zhao H, Sun X, Fu Z, Yu J. 18F-FDG PET or PET-CT to evaluate prognosis for head and neck cancer: a meta-analysis. J Cancer Res Clin Oncol 2011; 137: 1085–93. doi: 10.1007/s00432-010-0972-y [DOI] [PubMed] [Google Scholar]
- 2.Zhang B, Li X, Lu X. Standardized uptake value is of prognostic value for outcome in head and neck squamous cell carcinoma. Acta Otolaryngol 2010; 130: 756–62. doi: 10.3109/00016480903402981 [DOI] [PubMed] [Google Scholar]
- 3.Rasmussen JH, Vogelius IR, Fischer BM, Friborg J, Aznar MC, Persson GF, et al. Prognostic value of 18F-fludeoxyglucose uptake in 287 patients with head and neck squamous cell carcinoma. Head Neck May 2014. Epub ahead of print. doi: 10.1002/hed.23745 [DOI] [PubMed] [Google Scholar]
- 4.Troost EG, Schinagl DA, Bussink J, Oyen WJ, Kaanders JH. Clinical evidence on PET-CT for radiation therapy planning in head and neck tumours. Radiother Oncol 2010; 96: 328–34. doi: 10.1016/j.radonc.2010.07.017 [DOI] [PubMed] [Google Scholar]
- 5.Troost EG, Schinagl DA, Bussink J, Boerman OC, van der Kogel AJ, Oyen WJ, et al. Innovations in radiotherapy planning of head and neck cancers: role of PET. J Nucl Med 2010; 51: 66–76. doi: 10.2967/jnumed.108.061499 [DOI] [PubMed] [Google Scholar]
- 6.Madani I, Duprez F, Boterberg T, Van de Wiele C, Bonte K, Deron P, et al. Maximum tolerated dose in a phase I trial on adaptive dose painting by numbers for head and neck cancer. Radiother Oncol 2011; 101: 351–5. doi: 10.1016/j.radonc.2011.06.020 [DOI] [PubMed] [Google Scholar]
- 7.Vogelius IR, Håkansson K, Due AK, Aznar MC, Berthelsen AK, Kristensen CA, et al. Failure-probability driven dose painting. Med Phys 2013; 40: 081717. doi: 10.1118/1.4816308 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.de Langen AJ, Vincent A, Velasquez LM, van Tinteren H, Boellaard R, Shankar LK, et al. Repeatability of 18F-FDG uptake measurements in tumors: a metaanalysis. J Nucl Med 2012; 53: 701–8. doi: 10.2967/jnumed.111.095299 [DOI] [PubMed] [Google Scholar]
- 9.Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med 2009; 50(Suppl. 1): 122S–50S. doi: 10.2967/jnumed.108.057307 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Choi ES, Ha SG, Kim HS, Ha JH, Paeng JC, Han I. Total lesion glycolysis by 18F-FDG PET/CT is a reliable predictor of prognosis in soft-tissue sarcoma. Eur J Nucl Med Mol Imaging 2013; 40: 1836–42. doi: 10.1007/s00259-013-2511-y [DOI] [PubMed] [Google Scholar]
- 11.Moon SH, Choi JY, Lee HJ, Son YI, Baek CH, Ahn YC, et al. Prognostic value of 18F-FDG PET/CT in patients with squamous cell carcinoma of the tonsil: comparisons of volume-based metabolic parameters. Head Neck 2013; 35: 15–22. doi: 10.1002/hed.22904 [DOI] [PubMed] [Google Scholar]
- 12.Hyun SH, Ahn HK, Kim H, Ahn MJ, Park K, Ahn YC, et al. Volume-based assessment by (18)F-FDG PET/CT predicts survival in patients with stage III non-small-cell lung cancer. Eur J Nucl Med Mol Imaging 2014; 41: 50–8. doi: 10.1007/s00259-013-2530-8 [DOI] [PubMed] [Google Scholar]
- 13.Lee JW, Cho A, Lee JH, Yun M, Lee JD, Kim YT, et al. The role of metabolic tumor volume and total lesion glycolysis on 18F-FDG PET/CT in the prognosis of epithelial ovarian cancer. Eur J Nucl Med Mol Imaging 2014; 41: 1898–906. doi: 10.1007/s00259-014-2803-x [DOI] [PubMed] [Google Scholar]
- 14.Fletcher JW, Djulbegovic B, Soares HP, Siegel BA, Lowe VJ, Lyman GH, et al. Recommendations on the use of 18F-FDG PET in oncology. J Nucl Med 2008; 49: 480–508. doi: 10.2967/jnumed.107.047787 [DOI] [PubMed] [Google Scholar]
- 15.Hatt M, Cheze-Le Rest C, Aboagye EO, Kenny LM, Rosso L, Turkheimer FE, et al. Reproducibility of 18F-FDG and 3'-deoxy-3'-18F-fluorothymidine PET tumor volume measurements. J Nucl Med 2010; 51: 1368–76. doi: 10.2967/jnumed.110.078501 [DOI] [PubMed] [Google Scholar]
- 16.Lindholm H, Staaf J, Jacobsson H, Brolin F, Hatherly R, Sânchez-Crespo A. Repeatability of the maximum standard uptake value (SUVmax) in FDG PET. Mol Imaging Radionucl Ther 2014; 23: 16–20. doi: 10.4274/Mirt.76376 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Nahmias C, Wahl LM. Reproducibility of standardized uptake value measurements determined by 18F-FDG PET in malignant tumors. J Nucl Med 2008; 49: 1804–8. doi: 10.2967/jnumed.108.054239 [DOI] [PubMed] [Google Scholar]
- 18.Tixier F, Hatt M, Le Rest CC, Le Pogam A, Corcos L, Visvikis D. Reproducibility of tumor uptake heterogeneity characterization through textural feature analysis in 18F-FDG PET. J Nucl Med 2012; 53: 693–700. doi: 10.2967/jnumed.111.099127 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Velasquez LM, Boellaard R, Kollia G, Hayes W, Hoekstra OS, Lammertsma AA, et al. Repeatability of 18F-FDG PET in a multicenter phase I study of patients with advanced gastrointestinal malignancies. J Nucl Med 2009; 50: 1646–54. doi: 10.2967/jnumed.109.063347 [DOI] [PubMed] [Google Scholar]
- 20.Weber WA, Ziegler SI, Thödtmann R, Hanauske AR, Schwaiger M. Reproducibility of metabolic measurements in malignant tumors using FDG PET. J Nucl Med 1999; 40: 1771–7. [PubMed] [Google Scholar]
- 21.Rockall AG, Avril N, Lam R, Iannone R, Mozley PD, Parkinson C, et al. Repeatability of quantitative FDG-PET/CT and contrast-enhanced CT in recurrent ovarian carcinoma: test-retest measurements for tumor FDG uptake, diameter, and volume. Clin Cancer Res 2014; 20: 2751–60. doi: 10.1158/1078-0432.CCR-13-2634 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Becker M, Zaidi H. Imaging in head and neck squamous cell carcinoma: the potential role of PET/MRI. Br J Radiol 2014; 87: 20130677. doi: 10.1259/bjr.20130677 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Varoquaux A, Rager O, Poncet A, Delattre BM, Ratib O, Becker CD, et al. Detection and quantification of focal uptake in head and neck tumours: (18)F-FDG PET/MR versus PET/CT. Eur J Nucl Med Mol Imaging 2014; 41: 462–75. doi: 10.1007/s00259-013-2580-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Queiroz MA, Hüllner M, Kuhn F, Huber G, Meerwein C, Kollias S, et al. PET/MRI and PET/CT in follow-up of head and neck cancer patients. Eur J Nucl Med Mol Imaging 2014; 41: 1066–75. doi: 10.1007/s00259-014-2707-9 [DOI] [PubMed] [Google Scholar]
- 25.Platzek I, Beuthien-Baumann B, Schneider M, Gudziol V, Langner J, Schramm G, et al. PET/MRI in head and neck cancer: initial experience. Eur J Nucl Med Mol Imaging 2013; 40: 6–11. doi: 10.1007/s00259-012-2248-z [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Kubiessa K, Purz S, Gawlitza M, Kühn A, Fuchs J, Steinhoff KG, et al. Initial clinical results of simultaneous 18F-FDG PET/MRI in comparison to 18F-FDG PET/CT in patients with head and neck cancer. Eur J Nucl Med Mol Imaging 2014; 41: 639–48. doi: 10.1007/s00259-013-2633-2 [DOI] [PubMed] [Google Scholar]
- 27.Buchbender C, Heusner TA, Lauenstein TC, Bockisch A, Antoch G. Oncologic PET/MRI, part 1: tumors of the brain, head and neck, chest, abdomen, and pelvis. J Nucl Med 2012; 53: 928–38. doi: 10.2967/jnumed.112.105338 [DOI] [PubMed] [Google Scholar]
- 28.Boss A, Stegger L, Bisdas S, Kolb A, Schwenzer N, Pfister M, et al. Feasibility of simultaneous PET/MR imaging in the head and upper neck area. Eur Radiol 2011; 21: 1439–46. doi: 10.1007/s00330-011-2072-z [DOI] [PubMed] [Google Scholar]
- 29.Jakoby BW, Bercier Y, Conti M, Casey ME, Bendriem B, Townsend DW. Physical and clinical performance of the mCT time-of-flight PET/CT scanner. Phys Med Biol 2011; 56: 2375–89. doi: 10.1088/0031-9155/56/8/004 [DOI] [PubMed] [Google Scholar]
- 30.Panin VY, Kehren F, Michel C, Casey M. Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2006; 25: 907–21. [DOI] [PubMed] [Google Scholar]
- 31.Lois C, Jakoby BW, Long MJ, Hubner KF, Barker DW, Casey ME, et al. An assessment of the impact of incorporating time-of-flight information into clinical PET/CT imaging. J Nucl Med 2010; 51: 237–45. doi: 10.2967/jnumed.109.068098 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.El Naqa I, Grigsby P, Apte A, Kidd E, Donnelly E, Khullar D, et al. Exploring feature-based approaches in PET images for predicting cancer treatment outcomes. Pattern Recognit 2009; 42: 1162–71. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.van Velden FH, Cheebsumon P, Yaqub M, Smit EF, Hoekstra OS, Lammertsma AA, et al. Evaluation of a cumulative SUV-volume histogram method for parameterizing heterogeneous intratumoural FDG uptake in non-small cell lung cancer PET studies. Eur J Nucl Med Mol Imaging 2011; 38: 1636–47. doi: 10.1007/s00259-011-1845-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Sedgwick P. Limits of agreement (Bland-Altman method). BMJ 2013; 346: f1630. doi: 10.1136/bmj.f1630 [DOI] [PubMed] [Google Scholar]
- 35.Kumar V, Nath K, Berman CG, Kim J, Tanvetyanon T, Chiappori AA, et al. Variance of SUVs for FDG-PET/CT is greater in clinical practice than under ideal study settings. Clin Nucl Med 2013; 38: 175–82. doi: 10.1097/RLU.0b013e318279ffdf [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Vogelstein B, Papadopoulos N, Velculescu VE, Zhou S, Diaz LA, Jr, Kinzler KW. Cancer genome landscapes. Science 2013; 339: 1546–58. doi: 10.1126/science.1235122 [DOI] [PMC free article] [PubMed] [Google Scholar]