An evaluation of the Auto Analyzer SMA-4

T R J Lappin 1, A Lamont 1, M G Nelson 1
PMCID: PMC473995  PMID: 5784689

Abstract

The Technicon AutoAnalyzer SMA-4 has been systematically evaluated. Carry-over from the sample cups was found to be within an acceptable range. The precision and accuracy of the four parameters determined by the instrument have been investigated. The haemoglobin results were found to be accurate. There was good agreement with leucocyte counts performed on the model A Coulter electronic cell counter. After certain modifications had been made in the manifold, satisfactory degrees of accuracy were also obtained for the erythrocyte counts. Although the determination of the haematocrit by conductance is influenced by a variety of factors, the mean coefficient of variation was found to be 0·83% and 95% of the results agreed within ± 2% of those obtained by the microhaematocrit centrifugation method.

A logistic assessment of the SMA-4 when it was put into routine use indicated the need to select blood samples without excess anticoagulant, to calibrate the instrument before each run, to test for instrumental drift, to use a rapid method of correcting stoppages, and, if concurrent reporting is carried out, to employ the part-time services of a second operator. The presence of dust in the environment was found to have a deleterious effect on both the mechanical and electronic components of the system requiring preventive maintenance. It should be possible, by interfacing an analogue-digital converter and automatic punch, to produce data compatible with records.

Full text

PDF
13

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. BRECHER G., SCHNEIDERMAN M., WILLIAMS G. Z. Evaluation of electronic red blood cell counter. Am J Clin Pathol. 1956 Dec;26(12):1439–1449. doi: 10.1093/ajcp/26.12.1439. [DOI] [PubMed] [Google Scholar]
  2. KERNEN J. A., WURZEL H., OKADA R. New electronic method for measuring hematocrit: clinical evaluation. J Lab Clin Med. 1961 Apr;57:635–641. [PubMed] [Google Scholar]
  3. Lappin T. J., Lamont A. A cause of error in the determination of the haematocrit using the SMA-4. J Clin Pathol. 1968 Jan;21(1):114–114. doi: 10.1136/jcp.21.1.114-a. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. NELSON M. G., CARVILLE J. Blood cell counting: a comparison of the E.E.L. and Coulter machines. Ir J Med Sci. 1962 Oct;442:447–456. doi: 10.1007/BF02957182. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Pathology are provided here courtesy of BMJ Publishing Group

RESOURCES