Capillary zone electrophoresis as a tool for bottom-up protein analysis

Roza Wojcik 1,1, Guijie Zhu 1,1, Zhenbin Zhang 1,1, Xiaojing Yan 1,1, Yimeng Zhao 1,1, Liangliang Sun 1,1, Matthew M Champion 1,1, Norman J Dovichi 1,1,*
PMCID: PMC4763981  NIHMSID: NIHMS760060  PMID: 26651897

Background

A small community of researchers is reinvestigating capillary zone electrophoresis-MS (CZE-MS) for protein analysis. CZE provides a remarkably simple separation mechanism, very low cost instrumentation and very high resolution separations. Depending on its operational conditions, CZE can provide rapid separations for high-throughput analysis, or it can provide slow separations for deep bottom-up analysis of complex proteomes.

This effort has produced a remarkable improvement in performance. One metric is the number of detected peptides in a bottom-up single-shot analysis of complex proteomes, which has averaged roughly a tenfold improvement in each of the last 3 years. While pharmaceutical and clinical researchers may not require analysis of complex proteomes, the experience gained in proteomic analysis provides guidance for analysis of less complex samples.

This commentary highlights several challenges that have been faced and solutions demonstrated in CZE analysis of proteins.

Challenge 1: robust electrospray interface

CZE-electrospray interfaces present a serious design challenge. The interface must provide electrical contact at the distal end of the separation capillary in order to both complete the electrophoretic circuit and provide potential to drive the electrospray.

The first CZE electrospray interfaces were developed in the 1980s and were essentially modified HPLC interfaces [1]. These devices pump a sheath fluid at microliters per minute rates over the distal tip of the separation capillary and through a metal emitter, often with nebulizing gas to help stabilize the spray. The metal tip provides a convenient means of applying potential to the interface. The high sheath flow rate used in these early interfaces results in significant sample dilution.

Two interfaces were developed in the 2000s. In a sheathless interface, the distal tip of the separation capillary is etched to a few micrometer thickness [2]. The thin section is immersed in an electrolyte solution held at the electrospray potential, and the thin capillary wall provides sufficient conductivity to support electrospray. The absence of a sheath fluid eliminates dilution of the sample in the interface but requires that the background CZE electrolyte must support electrospray, which typically requires use of acidic solutions with volatile (e.g., methanol or acetonitrile) additives.

We have reported three generations of a nanospray interface based on electrokinetic pumping of a sheath liquid [3–5]. In these interfaces, a glass emitter is drawn to a 5–40 µm diameter tip using a micropipette puller. The separation capillary is threaded into this emitter and held in place using a plastic ‘T’ or cross. The sidearm of the T is connected to a sheath reservoir, and potential is applied to that reservoir to drive electrospray. Electro-osmosis at the interior surface of the glass emitter generates very stable flow of the sheath fluid in the nanoliters per minute range. The use of acidic sheath electrolytes with 10–50% (v/v) organic additive produces very stable spray without the need to add organics to the CZE background electrolyte. The latest generation of the interface is quite robust, and continuous operation over many days is routinely achieved.

Challenge 2: limited sample loading

To achieve the highest efficiency separation, CZE requires sample loadings of about 1 nl [6]. This minute sample volume can produce separations that approach a million theoretical plates. Unfortunately, this small sample volume contains a miniscule amount of analyte, severely stressing detection sensitivity.

Fortunately, several sample concentration methods have been developed that allow injection of very large sample volumes without serious degradation of separation efficiency. We highlight three: stacking, pH junction and online solid phase extraction (SPE).

In stacking, the sample is prepared in a low ionic strength electrolyte. The low ionic strength sample has correspondingly low conductivity that produces a high electric field across the sample plug, which concentrates the sample. A well-desalted tryptic digest will have ionic strength that is an order of magnitude lower than the background electrolyte, allowing injection of tens of nanoliters of sample without excessive band broadening. We have shown that stacking can be used to load approximately 400 ng of tryptic digest to identify over 10,000 peptides and 2100 proteins from the single-shot analysis of a HeLa tryptic digest in 110 min [7].

In the pH junction [8], tryptic peptides are prepared in a basic solution where the amphiprotic peptides are negatively charged. The sample is then pressure-injected into a capillary filled with an acidic background electrolyte. Upon application of an electric field, the negatively charged peptides migrate toward the positive, injection end of the capillary. Analytes are neutralized and concentrated when they reach the acidic background electrolyte. Concentration continues until the re-suspension buffer is depleted, at which point the peptides acquire a positive charge and undergo conventional migration and separation. This process requires no manipulation of the electrophoresis beyond preparation of the sample in a basic electrolyte. The pH junction can be used to inject hundreds of nanoliter samples, corresponding to over 30% of the capillary volume, with minimal loss of separation resolution [9].

Finally, a plug of stationary phase can be immobilized at the proximal tip of the separation capillary for SPE from very large volumes of sample. We typically employ a strong cation exchange solid-phase monolith for SPE and an acidic background electrolyte [10,11]. Tryptic digests are prepared in an acidic solution, where they carry a positive charge and are retained on the monolith. Components are eluted with a plug of a low ionic strength basic solution. Since the elution solution has lower ionic strength than the background electrolyte, stacking is produced upon application of an electric field. More importantly, the basic elution solution forms a pH junction, leading to additional band sharpening.

Challenge 3: control of separation speed

CZE is known for providing fast separations, which are valuable in high-throughput screens and other applications [12]. However, slow separations are desirable when performing deep proteomic analysis of complex samples; slower separations allow time to generate more tandem spectra for identification of large number of components. Separation speed can be controlled by adjusting the separation voltage, capillary length and extent of electro-osmosis. We find that the use of capillaries coated with a neutral polymer, such as linear polyacrylamide, coupled with a 1 M acetic acid background electrolyte provides very low electro-osmosis and advantageous separation conditions [7]. While there are commercially available coated capillaries, those capillaries tend to be expensive and can suffer from poor consistency. We have published a simple protocol for the preparation of the coated capillaries that results in a reproducible product [13].

CZE versus UPLC

Ultraperformance liquid chromatography (UPLC) has had many years of development, and is currently the separation mechanism of choice in protein and peptide analysis. CZE was first applied to a whole organism proteomic analysis 2 years ago, and is at a nascent stage of development. Nevertheless, there have been dramatic improvements in the performance in CZE in the past 2 years, and those improvements will undoubtedly continue as more experience is gained.

At this point in time (autumn 2015), we can make several observations. The loading capacity of CZE is roughly an order of magnitude lower than typical UPLC systems, and UPLC inevitably produces more peptide and protein identifications when large sample amounts are available. The state-of-the-art in the number of peptide and protein identifications produced in a single run is about three-times larger for UPLC than CZE [7]. CZE and UPLC produce similar numbers of protein and peptide identifications for approximately 50 ng samples, and CZE provides superior performance for smaller sample loadings [14–16]. It is clear that improved sample loading capacity will improve the performance of CZE, and the performance difference between CZE and UPLC will likely continue to shrink.

However, most protein analysis problems facing the clinical and pharmaceutical industries do not require a large numbers of protein and peptide identifications from a whole proteome. Instead, quality control for recombinant therapeutics and identification of biomarkers dominate routine analyses. In those cases, the reproducibility, resolution and simplicity of CZE become very attractive.

We conclude by mentioning three other applications of CZE. First, the technique has been applied to the top-down analysis of intact proteins [17–22]. Analysis of these proteins is challenging because of the requirement of a sophisticated mass spectrometer, complex data analysis and the presence of a range of proteoforms for each protein; these proteoforms are decorated with different sets of post-translational modifications that confound bottom-up analysis. Top-down analysis is ideal for characterizing the various modifications present on a protein because the proteoforms are not commingled as in bottom-up approaches. Typical applications are in the characterization of recombinant antibodies used in the biopharmaceutical industry. Second, CZE has proved to be a very powerful tool in the characterization of glycans [23–25]. Third, CZE is finding application in the clinical analysis of cerebral spinal fluids and other samples [26,27]. These and other applications of CZE will benefit from further improvements in sample loading, electrospray interface performance and control of separation speed.

Footnotes

Financial & competing interests disclosure

This work was supported by a grant from the NIH (R01GM096767). R Wojcik, L Sun, G Zhu and NJ Dovichi disclose that patents have been filed and that they receive royalties on the electrospray interfaces mentioned in this manuscript. The authors have no other relevant affiliations or financial involvement with any organization or entity with a financial interest in or financial conflict with the subject matter or materials discussed in the manuscript apart from those disclosed.

No writing assistance was utilized in the production of this manuscript.

References

  • 1.Smith RD, Barinaga CJ, Udseth HR. Improved electrospray ionization interface for capillary zone electrophoresis – mass-spectrometry. Anal. Chem. 1988;60(18):1948–1952. [Google Scholar]
  • 2.Moini M. Simplifying CE-MS operation. 2. Interfacing low-flow separation techniques to mass spectrometry using a porous tip. Anal. Chem. 2007;79(11):4241–4246. doi: 10.1021/ac0704560. [DOI] [PubMed] [Google Scholar]
  • 3.Wojcik R, Dada OO, Sadilek M, Dovichi NJ. Simplified capillary electrophoresis nanospray sheath-flow interface for high efficiency and sensitive peptide analysis. Rapid Commun. Mass Spectrom. 2010;24(17):2554–2560. doi: 10.1002/rcm.4672. [DOI] [PubMed] [Google Scholar]
  • 4.Sun L, Zhu G, Zhao Y, Yan X, Mou S, Dovichi NJ. Ultrasensitive and fast bottom-up analysis of femtogram amounts of complex proteome digests. Angew. Chem. Int. Ed. Engl. 2013;52(51):13661–13664. doi: 10.1002/anie.201308139. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Sun L, Zhu G, Zhang Z, Mou S, Dovichi NJ. Third-generation electrokinetically pumped sheath-flow nanospray interface with improved stability and sensitivity for automated capillary zone electrophoresis-mass spectrometry analysis of complex proteome digests. J. Proteome Res. 2015;14(5):2312–2321. doi: 10.1021/acs.jproteome.5b00100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Cheng YF, Wu SL, Chen DY, Dovichi NJ. Interaction of capillary zone electrophoresis with a sheath flow cuvette detector. Anal. Chem. 1990;62(5):496–503. [Google Scholar]
  • 7.Sun L, Hebert AS, Yan X, et al. Over 10,000 peptide identifications from the HeLa proteome by using single-shot capillary zone electrophoresis combined with tandem mass spectrometry. Angew. Chem. Int. Ed. Engl. 2014;53(50):13931–13933. doi: 10.1002/anie.201409075. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Britz-McKibbin P, Chen DD. Selective focusing of catecholamines and weakly acidic compounds by capillary electrophoresis using a dynamic pH junction. Anal. Chem. 2000;72(6):1242–1252. doi: 10.1021/ac990898e. [DOI] [PubMed] [Google Scholar]
  • 9.Zhu G, Sun L, Yan X, Dovichi NJ. Bottom-up proteomics of Escherichia coli using dynamic pH junction preconcentration and capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry. Anal. Chem. 2014;86(13):6331–6633. doi: 10.1021/ac5004486. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhang Z, Sun L, Zhu G, Yan X, Dovichi NJ. Integrated strong cation-exchange hybrid monolith coupled with capillary zone electrophoresis and simultaneous dynamic pH junction for large-volume proteomic analysis by mass spectrometry. Talanta. 2015;138:117–122. doi: 10.1016/j.talanta.2015.01.040. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Zhang Z, Yan X, Sun L, Zhu G, Dovichi NJ. Detachable strong cation exchange monolith, integrated with capillary zone electrophoresis and coupled with ph gradient elution, produces improved sensitivity and numbers of peptide identifications during bottom-up analysis of complex proteomes. Anal. Chem. 2015;87(8):4572–4577. doi: 10.1021/acs.analchem.5b00789. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Guetschow ED, Steyer DJ, Kennedy RT. Subsecond electrophoretic separations from droplet samples for screening of enzyme modulators. Anal. Chem. 2014;86(20):10373–10379. doi: 10.1021/ac502758h. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zhu G, Zun L, Dovichi NJ. Thermally-initiated free radical polymerization for reproducible production of stable linear polyacrylamide coated capillaries, and their application to proteomic analysis using capillary zone electrophoresis-mass spectrometry. Talanta. 2015 doi: 10.1016/j.talanta.2015.06.003. doi:10.1016/j.talanta.2015.06.003 In Press. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Faserl K, Sarg B, Kremser L, Lindner H. Optimization and evaluation of a sheathless capillary electrophoresis-electrospray ionization mass spectrometry platform for peptide analysis: comparison to liquid chromatography-electrospray ionization mass spectrometry. Anal. Chem. 2011;83(19):7297–7305. doi: 10.1021/ac2010372. [DOI] [PubMed] [Google Scholar]
  • 15.Li Y, Champion MM, Sun L, Champion PA, Wojcik R, Dovichi NJ. Capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry as an alternative proteomics platform to ultraperformance liquid chromatography-electrospray ionization-tandem mass spectrometry for samples of intermediate complexity. Anal. Chem. 2012;84(3):1617–1622. doi: 10.1021/ac202899p. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Wang Y, Fonslow BR, Wong CC, Nakorchevsky A, Yates JR., 3rd Improving the comprehensiveness and sensitivity of sheathless capillary electrophoresis-tandem mass spectrometry for proteomic analysis. Anal. Chem. 2012;84(20):8505–8513. doi: 10.1021/ac301091m. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Sun L, Knierman MD, Zhu G, Dovichi NJ. Fast top-down intact protein characterization with capillary zone electrophoresis-electrospray ionization tandem mass spectrometry. Anal. Chem. 2013;85(12):5989–5995. doi: 10.1021/ac4008122. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Zhao Y, Sun L, Champion MM, Knierman MD, Dovichi NJ. Capillary zone electrophoresis-electrospray ionization-tandem mass spectrometry for top-down characterization of the Mycobacterium marinum secretome. Anal. Chem. 2014;86(10):4873–4888. doi: 10.1021/ac500092q. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Li Y, Compton PD, Tran JC, Ntai I, Kelleher NL. Optimizing capillary electrophoresis for top-down proteomics of 30–80 kDa proteins. Proteomics. 2014;14(10):1158–1164. doi: 10.1002/pmic.201300381. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Zhao Y, Riley NM, Sun L, et al. Coupling capillary zone electrophoresis with electron transfer dissociation and activated ion electron transfer dissociation for top-down proteomics. Anal. Chem. 2015;87(10):5422–5429. doi: 10.1021/acs.analchem.5b00883. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Han X, Wang Y, Aslanian A, et al. In-line separation by capillary electrophoresis prior to analysis by top-down mass spectrometry enables sensitive characterization of protein complexes. J. Proteome Res. 2014;13(12):6078–6086. doi: 10.1021/pr500971h. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Han X, Wang Y, Aslanian A, Bern M, Lavallée-Adam M, Yates JR., 3rd Sheathless capillary electrophoresis-tandem mass spectrometry for top-down characterization of Pyrococcus furiosus proteins on a proteome scale. Anal. Chem. 2014;86(22):11006–11012. doi: 10.1021/ac503439n. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Balaguer E, Demelbauer U, Pelzing M, Sanz-Nebot V, Barbosa J, Neusüss C. Glycoform characterization of erythropoietin combining glycan and intact protein analysis by capillary electrophoresis – electrospray – time-of-flight mass spectrometry. Electrophoresis. 2006;27(13):2638–2650. doi: 10.1002/elps.200600075. [DOI] [PubMed] [Google Scholar]
  • 24.Zhuang Z, Starkey JA, Mechref Y, Novotny MV, Jacobson SC. Electrophoretic analysis of N-glycans on microfluidic devices. Anal. Chem. 2007;79(18):7170–7175. doi: 10.1021/ac071261v. [DOI] [PubMed] [Google Scholar]
  • 25.Jayo RG, Thaysen-Andersen M, Lindenburg PW, et al. Simple capillary electrophoresis-mass spectrometry method for complex glycan analysis using a flow-through microvial interface. Anal. Chem. 2014;86(13):6479–6486. doi: 10.1021/ac5010212. [DOI] [PubMed] [Google Scholar]
  • 26.Jahn H, Wittke S, Zürbig P, et al. Peptide fingerprinting of Alzheimer's disease in cerebrospinal fluid: identification and prospective evaluation of new synaptic biomarkers. PLoS ONE. 2011;6(10):e26540. doi: 10.1371/journal.pone.0026540. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Drube J, Schiffer E, Lau E, et al. Urinary proteome analysis to exclude severe vesicoureteral reflux. Pediatrics. 2012;129(2):e356–363. doi: 10.1542/peds.2010-3467. [DOI] [PubMed] [Google Scholar]

Articles from Bioanalysis are provided here courtesy of Taylor & Francis

RESOURCES