Abstract
Continuous wave Doppler echocardiography was used to study 41 adults with clinically suspected aortic stenosis undergoing cardiac catheterisation. Non-invasive assessment of the severity of stenosis was made before catheterisation using electrocardiograms, chest radiographs, and cross sectional echocardiography in addition to clinical examination and assessment modified, where appropriate, by the result of the Doppler examination. Catheterisation gradients were obtained in 33 patients and correlated well with those obtained by Doppler examination particularly when simultaneous recordings were obtained. All patients with surgically significant stenoses were identified by non-invasive assessment including Doppler examination and overestimation was not found in any patient with a less than significant stenosis. Thus surgery can be recommended in patients with aortic stenosis without the need for previous cardiac catheterisation.
Full text
PDFImages in this article
Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Angelsen B. A., Brubakk A. O. Transcutaneous measurement of blood flow velocity in the human aorta. Cardiovasc Res. 1976 May;10(3):368–379. doi: 10.1093/cvr/10.3.368. [DOI] [PubMed] [Google Scholar]
- Berger M., Berdoff R. L., Gallerstein P. E., Goldberg E. Evaluation of aortic stenosis by continuous wave Doppler ultrasound. J Am Coll Cardiol. 1984 Jan;3(1):150–156. doi: 10.1016/s0735-1097(84)80442-8. [DOI] [PubMed] [Google Scholar]
- DeMaria A. N., Bommer W., Joye J., Lee G., Bouteller J., Mason D. T. Value and limitations of cross-sectional echocardiography of the aortic valve in the diagnosis and quantification of valvular aortic stenosis. Circulation. 1980 Aug;62(2):304–312. doi: 10.1161/01.cir.62.2.304. [DOI] [PubMed] [Google Scholar]
- Gardin J. M., Burn C. S., Childs W. J., Henry W. L. Evaluation of blood flow velocity in the ascending aorta and main pulmonary artery of normal subjects by Doppler echocardiography. Am Heart J. 1984 Feb;107(2):310–319. doi: 10.1016/0002-8703(84)90380-6. [DOI] [PubMed] [Google Scholar]
- Godley R. W., Green D., Dillon J. C., Rogers E. W., Feigenbaum H., Weyman A. E. Reliability of two-dimensional echocardiography in assessing the severity of valvular aortic stenosis. Chest. 1981 Jun;79(6):657–662. doi: 10.1378/chest.79.6.657. [DOI] [PubMed] [Google Scholar]
- Hall R. J., Kadushi O. A., Evemy K. Need for cardiac catheterisation in assessment of patients for valve surgery. Br Heart J. 1983 Mar;49(3):268–275. doi: 10.1136/hrt.49.3.268. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hatle L., Angelsen B. A., Tromsdal A. Non-invasive assessment of aortic stenosis by Doppler ultrasound. Br Heart J. 1980 Mar;43(3):284–292. doi: 10.1136/hrt.43.3.284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hatle L., Brubakk A., Tromsdal A., Angelsen B. Noninvasive assessment of pressure drop in mitral stenosis by Doppler ultrasound. Br Heart J. 1978 Feb;40(2):131–140. doi: 10.1136/hrt.40.2.131. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martin R. P., Rakowski H., Kleiman J. H., Beaver W., London E., Popp R. L. Reliability and reproducibility of two dimensional echocardiograph measurement of the stenotic mitral valve orifice area. Am J Cardiol. 1979 Mar;43(3):560–568. doi: 10.1016/0002-9149(79)90014-6. [DOI] [PubMed] [Google Scholar]
- Stamm R. B., Martin R. P. Quantification of pressure gradients across stenotic valves by Doppler ultrasound. J Am Coll Cardiol. 1983 Oct;2(4):707–718. doi: 10.1016/s0735-1097(83)80311-8. [DOI] [PubMed] [Google Scholar]