Determination of pressure gradient in mitral stenosis with Doppler echocardiography.

J Holen, S Simonsen
PMCID: PMC482066  PMID: 465223

Abstract

The accuracy of a non-invasive ultrasound Doppler technique for the determination of the pressure gradient in mitral stenosis was evaluated in a study of 8 adult patients. Transseptal left atrial catheterisation and retrograde left ventricular catheterisation were performed. The same diastoles were used to compare the gradient constructed from the ultrasound data (delta PU) with that constructed from the manometric data (delta PM). In the 8 patients the difference betweent he mean diastolic values of delta PU and delta PM was - 0.54 +/- 1.0 (SD) mmHg. The corresponding figure for mid-diastole was 0.01 +/- 0.9 (SD) mmHg. The results indicate that the ultrasound technique is sufficiently accurate for diagnostic purposes.

Full text

PDF
531

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Holen J., Aaslid R., Landmark K., Simonsen S. Determination of pressure gradient in mitral stenosis with a non-invasive ultrasound Doppler technique. Acta Med Scand. 1976;199(6):455–460. doi: 10.1111/j.0954-6820.1976.tb06763.x. [DOI] [PubMed] [Google Scholar]
  2. Holen J., Aaslid R., Landmark K., Simonsen S., Ostrem T. Determination of effective orifice area in mitral stenosis from non-invasive ultrasound Doppler data and mitral flow rate. Acta Med Scand. 1977 Jan;201(1-2):83–88. doi: 10.1111/j.0954-6820.1977.tb15660.x. [DOI] [PubMed] [Google Scholar]

Articles from British Heart Journal are provided here courtesy of BMJ Publishing Group

RESOURCES