Abstract
A novel ion-imprinted electrochemical sensor based on AlGaN/GaN high electron mobility transistors (HEMTs) was developed to detect trace amounts of phosphate anion. This sensor combined the advantages of the ion sensitivity of AlGaN/GaN HEMTs and specific recognition of ion imprinted polymers. The current response showed that the fabricated sensor is highly sensitive and selective to phosphate anions. The current change exhibited approximate linear dependence for phosphate concentration from 0.02 mg L−1 to 2 mg L−1, the sensitivity and detection limit of the sensor is 3.191 μA/mg L−1 and 1.97 μg L−1, respectively. The results indicated that this AlGaN/GaN HEMT-based electrochemical sensor has the potential applications on phosphate anion detection.
Phosphate is one of the most important electrolytes and an essential component of all living organisms, as they take part in almost all metabolic processes1. A high phosphate level is detrimental to the environment and can lead to eutrophication which disrupts aquatic life cycles2,3,4,5,6,7,8,9. Phosphate levels is higher than 0.1 mg L−1, which may lead to many health concerns10,11, because moderate increase in phosphate concentration in natural water results in eutrophication and consequently lowers the dissolved oxygen concentration. Eventually, it will cause the death of certain fishes and other aquatic animals due to suffocation. While the normal phosphate level in the human body is higher, about 25–45 mg L−1 in blood serum12. It will be lead to the physiological and pathological changes of the human body when phosphate level is out of the normal range. So, the phosphate detection and recognition are of great significance for the controlling and guarding against the eutrophication and human disease occurrence.
Various detection strategies for phosphate have been developed including spectrophotometry13,14, fluorometry15,16, chromatography17. However, these methods are time consuming, laborious, require expensive instruments, well trained operators, and often employ potentially carcinogenic chemicals for the analysis5. The drawbacks of these techniques make them not easily to be used for on-line monitoring.
Recently AlGaN/GaN high electron mobility transistors (HEMTs) have been developed for favorable sensors, because the conducting two-dimensional electron gas (2DEG) channel is very close to the surface and extremely sensitive to adsorption of analytes18. In the past few years, development of ion selective monitoring such as potassium ions19, ammonium ions20, chloride ions21, mercury ions22,23,24,25 based on AlGaN/GaN HEMTs has attracted word-wide attentions. Due to excellent chemical and physical stability in water21,27, ion sensitive AlGaN/GaN HEMTs have showed great advantages in ion detection. Thus, it may be an excellent potential candidate for phosphate anion detection.
Molecular imprinting is a technology to create recognition sites in a macromolecular matrix using a template molecule28,29. Within molecularly imprinted polymers (MIPs), a large number of imprinted cavities designed for the template molecule are distributed, and these cavities are consistent to the template molecules in shape, size, and functional groups. So, MIPs not only have specific molecular recognition ability and high binding affinity for the template molecule, but also are described as artificial antibodies or receptors29,30,31. As the template is ion in the preparation process of MIPs, the synthetic products are called ion-imprinted polymers (IIPs)32,33,34,35,36. IIPs maintain all the advantages of MIPs, they can specially recognize the template ion. In the last few years, IIPs as selective sorbents for template ions have attracted much attention, and their applications in selective recognition, separation, and enrichment of ions as well as in the removal of toxic ions from aqueous medium for protecting the environment have been reported37,38,39,40.
In this work, we functionalized the ungated region of AlGaN/GaN HEMT sensor by phosphate anion imprinted polymers to detect and recognize phosphate anion. The current response showed that the fabricated sensor is very sensitive to phosphate anions, and meanwhile, it has no obvious response to other interfering anions, for example permanganate anion-sulfate anion and so on.
Results and Discussion
Figure 1 shows the structure diagram of the sensor. The current response was measured at room temperature using Cascade probe station. Figure 2 shows the voltage-current curves of the AlGaN/GaN HEMT-based sensor before and after dropping phosphate anion solution (PH = 5) onto the sensing region with a micropipette. We can observe that the source-drain current obviously decreases after dropping the phosphate anions. It is well known that the electrons in the 2DEG channel of undoped AlGaN/GaN HEMTs are induced by spontaneous and piezoelectric polarization charges. Any slight changes in the surface charge are transduced into a change in the concentration of the 2DEG in the AlGaN/GaN interface41. Therefore, the 2DEG concentration in the channel, which determines the source-drain resistance, is very sensitive to the change of surface charges of AlGaN/GaN HEMTs. When dropping the phosphate anion solution onto the functionalized surface of the AlGaN/GaN HEMT, phosphate anions with negative charges will occupy the imprinted caves in the sensing area due to the specific recognition and the electrostatic interaction between this template anion and the ion imprinted polymer. This process will decrease the surface potential (Vg) on the ungated area of the AlGaN/GaN HEMT sensor, then leads to a reduction in the concentration of 2DEG in the channel, as described in following equation42 (1), and finally causes a reduced source-drain current.
where εN is the permittivity of AlGaN, Vg is the gate voltage due to the surface states,
Voff is the threshold voltage, q is the electron charge, d is the total distance target ion to 2DEG, and V(x) is the channel potential.
However, when dropping permanganate and sulfate solution onto the sensing area, the source-drain current has almost no change as shown in Fig. 3, although permanganate and sulfate anions are of similar chemical structure to phosphate anion. This indicates that the sensor is of high selectivity for different anion solutions. This high selectivity can be attributed to the causes of specific recognition and electrostatic interaction between template ion and ion imprinted polymer. For the phosphate anion imprinted material, a great quantity of phosphate anion imprinted caves are distributed within the polymer layer on the sensing area of AlGaN/GaN HEMT. These caves are highly matched with phosphate anions in size and shape, which leads to the perfect recognition ability and strong binding action for phosphate anion43,44. In contrast, these imprinted caves are unmatched with MnO4− and SO42− ion in size and shape, leading to basically non-recognizing and non-binding for two contrast anions. The order magnitude of the ionic radii are successively rMnO4− > rPO43− > rSO42−, so the imprinted cavities are small for MnO4−, and are big for SO42−. Therefore, phosphate anion imprinted polymer exhibits a high selectivity to phosphate anion.
In addition, the temperature and pH value have effects on the current change of the sensors. The change of temperature can slightly affect the structural shrinkage and expansion of imprinted sites, and both tight and loose spatial structures of the imprinted sites will lead to the decrease in the recognition ability45. However, the change of temperature in a small range has little effect on current change according to our recorded experimental data. In our experiments, the current change was measured at room temperature which is consistent with the temperature that imprinted cavities formed, so the phosphate anions are matched with the imprinted cavities in size. Moreover, pH value of tested solution will affect the current change of the sensor. As pH < 4, phosphate predominantly exists as H3PO4 and H2PO4−, and H3PO4 species are hardly recognized by the ion imprinted polymers, so the current change of the sensor will decrease; As 7 < pH < 10, phosphate predominantly exists as HPO42− which is not matched with the imprinted cavities, and the sensor will be almost no response; As pH > 10, phosphate predominantly exists as PO43− and HPO42− which are also not matched with the imprinted cavities. As 4 < pH < 6, phosphate predominantly exists as H2PO4− which is consistent with the template ion, and hence the sensor has the best response in this pH value range43,46. Therefore, both the functionalization of GaN-based HEMT sensor and experimental test are carried out at pH value of 5 in this work, and pH value was adjusted with HCl solution.
Figure 4 shows the current response of the sensor to different concentration of phosphate anions at a constant voltage of 50 mV. Current is almost changeless when dropping water onto the sensitive area. In contrast to the water solution, when 0.02 mg L−1 of phosphate solution was dropped, an obvious current decrease was observed after the system reached a steady state. As expected, a gradual decrease in current was observed along with the increase of phosphate concentration in the measurement range from 0.02 mg L−1–20 mg L−1.
In order to study the performances of the sensor, we analyzed the relations between current change and phosphate concentration. As shown in Fig. 5, the current change of sensor is approximate linear dependence on the target phosphate concentration in the range from 0.02 mg L−1–2 mg L−1. With further increase of phosphate anion concentration, the decrease of the current will be gradually saturated. It can be easily understood because the number of cavities in the ion imprinted polymer is limited. When phosphate anion concentration reaches a certain value, these cavities will be occupied completely, and the current change will also reach saturation.
The sensitivity is an important performance of the sensor, which was defined as the slope of the calibration line according to IUPAC. It can be observed that change of source-drain current exhibits an approximate linear dependence with phosphate concentration in the range of 0.02 mg L−1–2 mg L−1, and the linear regression coefficient of the calibration line is 0.993. Therefore, the sensitivity of the sensor is 3.191 μA/mg L−1. The detection limit is obtained using 3σ method (IUPAC)47.
where σ is the standard deviation got from current method used in this experiment and m is the slope of calibration line. The detection limit is 1.97 μg L−1, which is better than the reported detection limits of 0.16 mg L−1 and 18.8 μg L−1 by Christopher48 and Gong49, respectively.
As the significant parameters of the sensor, the experimental reproducibility has to be investigated. Phosphate anions were detected for six independent experiments by using different AlGaN/GaN HEMT-based sensors. The maximum relative standard deviation (RSD) is 4.4% at the concentration level of 0.02 mg L−1, indicating acceptable reproducibility of the fabricated sensors. In addition, the fabricated sensor remained 98% of its initial current response after 15 days storage at room temperature, suggesting the satisfactory storage stability.
The AlGaN/GaN HEMT sensor functioned with ion imprinted polymer was fabricated successfully. It showed highly selectivity and recognition to phosphate anion and no response to other interfering anions, such as permanganate and sulfate anions. The sensitivity of sensor is 3.191 μA/mg L−1, and the detection limit is 1.97 μg L−1. Highly sensitivity and selectivity make the AlGaN/GaN HEMT-based sensor a promising application for phosphate anion on-line monitoring which may integrate easily the wireless microwave transmission function of HEMTs.
Methods
The growth process of AlGaN/GaN on SiC substrate and device preparation
Figure 1 shows the cross-sectional diagram of the sensor. The AlGaN/GaN heterostructure was grown by metal-organic chemical vapor deposition on a SiC substrate. A 3-μm-undoped GaN buffer layer was firstly grown on the substrate, followed by a 25 nm undoped Al0.3Ga0.7N layer was grown on the GaN layer to form the heterostructure. The two-dimensional electron gas (2DEG) is located at the interface between the undoped GaN layer and the AlGaN layer. Hall measurements showed a sheet carrier density of 1.85 × 1013 cm−2 and carrier mobility of 1650 cm2 v−1 s−1 at room temperature.
Mesa isolation was performed using an Inductively Coupled Plasma (ICP) etching system. Ohmic contacts were fabricated by depositing Ti/Al/Ni/Au (30/120/50/80 nm) layers using e-beam evaporation on the AlGaN layer and then annealing at 850 °C for 30 s in a flowing N2 atmosphere. The electrodes were protected to avoid direct contact to the solution by Si3N4 passivation layer (about 120 nm) deposited with plasma enhanced chemical vapor deposition. After that, photoresist (AZ5214) was used to encapsulate the source/drain regions, with only the ungated region open to allow the reactive ion etching to etch the surface. In order to reduce the surface damage induced by dry-etching, we adopted a method which combined virtues of dry-etching and wet-etching. Si3N4 passivation layer was etched 100 nm by dry-etching and the left was removed by wet-etching (HF:H2O = 1:10). At last, open ungated area was functioned with phosphate ion imprinted polymer.
The surface functionalization of sensor
To modify the ungated region of sensor with the phosphate ion imprinted polymer, the treated AlGaN surfaces were oxidized, silanized and ion-imprinted successively50. During oxidizing step, the sensing surface was oxidized 40 s with Inductively Coupled Plasma (ICP). After oxidation, the samples were boiled in deionized water for 30 min at 100 °C and then blown dry by a nitrogen gun. 10% 3-aminopropyltrimethoxy- silane (AMPS, KH-540) solution was used to treat the samples for 24 h at 50 °C. In the steps of phosphate ion imprinting, the phosphate anion surface-imprinting was conducted in an aqueous solution system. Deionized water 50 mL, 0.6985 g of Na3PO4∙12 H2O as template, 1.5 mL of monomer methacryloyloxyethyl trimethyl ammonium chloride (DMC), and 0.328 g of cross-linker N,N’-Methylenebisacrylami-De (MBA) were added to a beaker, silanized AlGaN/GaN HEMTs were put into solution and N2 was bubbled for 30 min to exclude air. The pH value of the solution was adjusted to 5. The substance was heated to 35 °C, followed by adding 0.02915 g of initiator ammonium persulfate. The graft/cross-linking polymerization reaction was performed at 35 °C for 14 h. To remove the template phosphate anion, the device was fully washed with 2 mol L−1 NaCl solution. After washing away the template ions, a large numbers of phosphate anion imprinted caves will remain within this thin polymer layer on the sensing surface. Through vacuum drying, the AlGaN/GaN sensors functionalized by the phosphate anion imprinted polymers were obtained. (The chemical reaction processes of ion imprinting see supplementary information).
Electrical measurements of the sensor
The current of source-drain was measured at room temperature using Cascade probe station (Instrument types: Tesla 200, Tungsten probe), Tungsten probes were pressed on the electrode of the source and drain during the measurements.
Additional Information
How to cite this article: Jia, X. et al. Highly selective and sensitive phosphate anion sensors based on AlGaN/GaN high electron mobility transistors functionalized by ion imprinted polymer. Sci. Rep. 6, 27728; doi: 10.1038/srep27728 (2016).
Supplementary Material
Acknowledgments
This work was supported by the National 973 project, China (2012CB619306), NSFC (Nos 61274075, 61474060), and the PAPD of Jiangsu Province, China.
Footnotes
Author Contributions X.J. performed experiments and data analysis. D.C. conceived and designed the research and experiments. B.L. fabricated the AlGaN/GaN heterostructure materials. H.L. contributed to data analysis. R.Z. and Y Z. supervised the project and discussed the results. X.J. and D.C. wrote the paper.
References
- Bard A. J. & Faulkner L. R. Electrochemical methods Fundamentals and Applications. 2nd ed. (eds Harris D. et al.) Ch. 2, Potentials and Thermodynamics of Cells, 44–86 (John Wiley and Sons Inc., 2001). [Google Scholar]
- Karthikeyan R. & Berchmans S. Inorganic-Organic Composite Matrix for the Enzymatic Detection of Phosphate in Food Samples. J Electrochem Soc 160, B73–B77 (2013). [Google Scholar]
- Camarero L. & Catalan J. Atmospheric phosphorus deposition may cause lakes to revert from phosphorus limitation back to nitrogen limitation. Nat Commun 3, (2012). [DOI] [PubMed] [Google Scholar]
- Knutsson J., Rauch S. & Morrison G. M. Performance of a passive sampler for the determination of time averaged concentrations of nitrate and phosphate in water. Environ Sci-Proc Imp 15, 955–962 (2013). [DOI] [PubMed] [Google Scholar]
- Kwan R. C. H., Leung H. F., Hon P. Y. T., Barford J. P. & Renneberg R. A screen-printed biosensor using pyruvate oxidase for rapid determination of phosphate in synthetic wastewater. Appl Microbiol Biot 66, 377–383 (2005). [DOI] [PubMed] [Google Scholar]
- Law A. T. & Adeloju S. B. Progress and recent advances in phosphate sensors: A review. Talanta 114, 191–203 (2013). [DOI] [PubMed] [Google Scholar]
- Penuelas J. et al. Human-induced nitrogen-phosphorus imbalances alter natural and managed ecosystems across the globe. Nat Commun 4, (2013). [DOI] [PubMed] [Google Scholar]
- Upadhyay L. S. B. & Verma N. Alkaline phosphatase inhibition based conductometric biosensor for phosphate estimation in biological fluids. Biosens Bioelectron 68, 611–616 (2015). [DOI] [PubMed] [Google Scholar]
- Warwick C., Guerreiro A. & Soares A. Sensing and analysis of soluble phosphates in environmental samples: A review. Biosens Bioelectron 41, 1–11 (2013). [DOI] [PubMed] [Google Scholar]
- Berchmans S., Issa T. B. & Singh P. Determination of inorganic phosphate by electroanalytical methods: A review. Anal Chim Acta 729, 7–20 (2012). [DOI] [PubMed] [Google Scholar]
- Neves M. S. A. C. et al. Spectrophotometric flow system using vanadomolybdophosphate detection chemistry and a liquid waveguide capillary cell for the determination of phosphate with improved sensitivity in surface and ground water samples. Talanta 77, 527–532 (2008). [Google Scholar]
- Tonelli M. et al. Relation between serum phosphate level and cardiovascular event rate in people with coronary disease. Circulation 112, 2627–2633 (2005). [DOI] [PubMed] [Google Scholar]
- Fujiwara K. et al. Determination of phosphorus at the parts per trillion level by laser-induced thermal lensing colorimetry. Analytical Chemistry 54, 2026–2029 (1982). [Google Scholar]
- Li C. M., Li Y. F., Wang J. & Huang C. Z. Optical investigations on ATP-induced aggregation of positive-charged gold nanoparticles. Talanta 81, 1339–1345 (2010). [DOI] [PubMed] [Google Scholar]
- Zhao H. X. et al. Highly selective detection of phosphate in very complicated matrixes with an off-on fluorescent probe of europium-adjusted carbon dots. Chem Commun 47, 2604–2606 (2011). [DOI] [PubMed] [Google Scholar]
- Bai J.-M., Zhang L., Liang R.-P. & Qiu J.-D. Graphene Quantum Dots Combined with Europium Ions as Photoluminescent Probes for Phosphate Sensing. Chem-Eur J 19, 3822–3826 (2013). [DOI] [PubMed] [Google Scholar]
- Quintana J. B., Rodil R. & Reemtsma T. Determination of phosphoric acid mono- and diesters in municipal wastewater by solid-phase extraction and ion-pair liquid chromatography–Tandem mass spectrometry. Anal Chem 78, 1644–1650 (2006). [DOI] [PubMed] [Google Scholar]
- Chu B. H. et al. Detection of an endocrine disrupter biomarker, vitellogenin, in largemouth bass serum using AlGaN/GaN high electron mobility transistors. Appl Phys Lett 96, (2010). [Google Scholar]
- Alifragis Y. et al. Potassium selective chemically modified field effect transistors based on AlGaN/GaN two-dimensional electron gas heterostructures. Biosens Bioelectron 22, 2796–2801 (2007). [DOI] [PubMed] [Google Scholar]
- Alifragis Y. et al. AlGaN/GaN high electron mobility transistor sensor sensitive to ammonium ions. Phys Status Solidi A 204, 2059–2063 (2007). [Google Scholar]
- Hung S. C. et al. Detection of chloride ions using an integrated Ag/AgCl electrode with AlGaN/GaN high electron mobility transistors. Appl Phys Lett 92, (2008). [Google Scholar]
- Chen K. H. et al. Low Hg(II) ion concentration electrical detection with AlGaN/GaN high electron mobility transistors. Sensor Actuat B-Chem 134, 386–389 (2008). [Google Scholar]
- Cheng J. et al. Ultrasensitive detection of Hg2+ using oligonucleotide-functionalized AlGaN/GaN high electron mobility transistor. Appl Phys Lett 105, (2014). [Google Scholar]
- Chu B. H. et al. AlGaN/GaN High Electron Mobility Transistor Based Sensors for Environmental and Bio-Applications. Nanosci Nanotech Let 2, 120–128 (2010). [Google Scholar]
- Wang H.-T. et al. Fast electrical detection of Hg(II) ions with AlGaN/GaN high electron mobility transistors. Appl Phys Lett 91, (2007). [Google Scholar]
- Khir F. L. M. et al. Synchrotron-based XPS studies of AlGaN and GaN surface chemistry and its relationship to ion sensor behaviour. Appl Surf Sci 314, 850–857 (2014). [Google Scholar]
- Ito T. et al. Self-assembled monolayers of alkylphosphonic acid on GaN substrates. Langmuir 24, 6630–6635 (2008). [DOI] [PubMed] [Google Scholar]
- Vlatakis G., Andersson L. I., Muller R. & Mosbach K. Drug assay using antibody mimics made by molecular imprinting. Nature 361, 645–647 (1993). [DOI] [PubMed] [Google Scholar]
- Wulff G. Molecular imprinting in cross-linked materials with the aid of molecular templates-a way towards artificial antibodies. Angew Chem Int Ed Engl 34, 1812–1832 (1995). [Google Scholar]
- Farid M. M., Goudini L., Piri F., Zamani A. & Saadati F. Molecular imprinting method for fabricating novel glucose sensor: Polyvinyl acetate electrode reinforced by MnO2/CuO loaded on graphene oxide nanoparticles. Food Chem 194, 61–67 (2016). [DOI] [PubMed] [Google Scholar]
- Shrivastav A. M. & Gupta B. D. SPR and Molecular Imprinting-Based Fiber-Optic Melamine Sensor With High Sensitivity and Low Limit of Detection. Ieee J Sel Top Quant 22, (2016). [Google Scholar]
- Preetha C. R., Gladis J. M., Rao T. P. & Venkateswaran G. Removal of toxic uranium from synthetic nuclear power reactor effluents using uranyl ion imprinted polymer particles. Environ Sci Technol 40, 3070–3074 (2006). [DOI] [PubMed] [Google Scholar]
- Singh D. K. & Mishra S. Synthesis, characterization and removal of Cd(II) using Cd(II)-ion imprinted polymer. J Hazard Mater 164, 1547–1551 (2009). [DOI] [PubMed] [Google Scholar]
- Shamsipur M. et al. Silver ion imprinted polymer nanobeads based on a aza-thioether crown containing a 1,10-phenanthroline subunit for solid phase extraction and for voltammetric and potentiometric silver sensors. Anal Chim Acta 852, 223–235 (2014). [DOI] [PubMed] [Google Scholar]
- Rajabi H. R., Shamsipur M. & Pourmortazavi S. M. Preparation of a novel potassium ion imprinted polymeric nanoparticles based on dicyclohexyl 18C6 for selective determination of K + ion in different water samples. Mat Sci Eng C-Mater 33, 3374–3381 (2013). [DOI] [PubMed] [Google Scholar]
- Rao T. P., Kala R. & Daniel S. Metal ion-imprinted polymers-Novel materials for selective recognition of inorganics. Anal Chim Acta 578, 105–116 (2006). [DOI] [PubMed] [Google Scholar]
- Xiang C., Pan J. M., Gao J., Yan Y. S. & Zhao G. Q. An ion-imprinted polymer supported by attapulgite with a chitosan incorporated sol-gel process for selective separation of Ce(III). Chinese Chem Lett 20, 985–989 (2009). [Google Scholar]
- Jo S.-H. et al. Continuous separation of copper ions from a mixture of heavy metal ions using a three-zone carousel process packed with metal ion-imprinted polymer. J Chromatogr A 1217, 7100–7108 (2010). [DOI] [PubMed] [Google Scholar]
- Wang X. et al. Enrichment and separation of silver from waste solutions by metal ion imprinted membrane. Hydrometallurgy 100, 82–86 (2009). [Google Scholar]
- Khajeh M., Heidari Z. S. & Sanchooli E. Synthesis, characterization and removal of lead from water samples using lead-ion imprinted polymer. Chem Eng J 166, 1158–1163 (2011). [Google Scholar]
- Talukdar A. et al. Piezotransistive transduction of femtoscale displacement for photoacoustic spectroscopy. Nat Commun 6, (2015). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Neaman D. A. Semiconductor Physics and Divices. 3rd ed. (eds Sayre D. et al.) Ch.13.5 High Electron Mobility Transistor, 604–605 (The McGraw-Hill Companies, Inc., 2010). [Google Scholar]
- Gao B., Du J. & Zhang Y. Preparation of Arsenate Anion Surface-Imprinted Material IIP-PDMC/SiO2 and Study on Its Ion Recognition Property. Ind Eng Chem Res 52, 7651–7659 (2013). [Google Scholar]
- Du J., Gao B., Huang X., Zhang Y. & Wang M. Preparing Surface Anion-imprinted Material Based on Ion Exchange and Surface-initiated Graft-polymerization and Studies on Its Recognition Character. Acta Chim Sinica 70, 1831–1838 (2012). [Google Scholar]
- Ma P. et al. Preparation and application of sulfadiazine surface molecularly imprinted polymers with temperature-responsive properties. J. Appl. Polym. Sci 132, (2015). [Google Scholar]
- Kugimiya A. & Takei H. Selectivity and recovery performance of phosphate-selective molecularly imprinted polymer. Anal. Chim. Acta 606, 252–256 (2008). [DOI] [PubMed] [Google Scholar]
- Mc Naught A. D. & Wilkinson A. IUPAC. Compendium of Chemical Terminology. Version.2.3.3. (eds Wurzbacher C. et al.) Ch.1, Alphabetical index, 1281–1282 (Blackwell Scientific Publications, Oxford, 2014). [Google Scholar]
- Warwick C. et al. A molecular imprinted polymer based sensor for measuring phosphate in wastewater samples. Water Sci Technol 69, 48–54 (2014). [DOI] [PubMed] [Google Scholar]
- Gong W. T., Na D., Fang L., Mehdi H. & Ning G. L. Preassembly-driven ratiometric sensing of H2PO4− anions in organic and aqueous environments. Org Biomol Chem 13, 1979–1982 (2015). [DOI] [PubMed] [Google Scholar]
- Gupta S. et al. Detection of clinically relevant levels of protein analyte under physiologic buffer using planar field effect transistors. Biosens Bioelectron 24, 505–511 (2008). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.