Stability of selected volatile breath constituents in Tedlar, Kynar and Flexfilm sampling bags

Paweł Mochalski a,b,*, Julian King a,c, Karl Unterkofler a, Anton Amann a,d
PMCID: PMC4909142  EMSID: EMS54753  PMID: 23323261

Abstract

The stability of 41 selected breath constituents in three types of polymer sampling bags, Tedlar, Kynar, and Flexfilm, was investigated using solid phase microextraction and gas chromatography mass spectrometry. The tested molecular species belong to different chemical classes (hydrocarbons, ketones, aldehydes, aromatics, sulphurs, esters, terpenes, etc.) and exhibit close-to-breath low ppb levels (3–12 ppb) with the exception of isoprene, acetone and acetonitrile (106 ppb, 760 ppb, 42 ppb respectively). Stability tests comprised the background emission of contaminants, recovery from dry samples, recovery from humid samples (RH 80% at 37 °C), influence of the bag’s filling degree, and reusability. Findings yield evidence of the superiority of Tedlar bags over remaining polymers in terms of background emission, species stability (up to 7 days for dry samples), and reusability. Recoveries of species under study suffered from the presence of high amounts of water (losses up to 10%). However, only heavier volatiles, with molecular masses higher than 90, exhibited more pronounced losses (20–40%). The sample size (the degree of bag filling) was found to be one of the most important factors affecting the sample integrity. To sum up, it is recommended to store breath samples in pre-conditioned Tedlar bags up to 6 hours at the maximum possible filling volume. Among the remaining films, Kynar can be considered as an alternative to Tedlar; however, higher losses of compounds should be expected even within the first hours of storage. Due to the high background emission Flexfilm is not suitable for sampling and storage of samples for analyses aiming at volatiles at a low ppb level.

1. Introduction

Preservation of the sample integrity during sampling and sample storage is probably one of the most demanding challenges in analytical chemistry. Different phenomena accompanying these phases of analysis like, e.g., background emission of pollutants, losses and interactions between sample constituents irreversibly modify the original sample composition and consequently distort the final results of analyses. This is particularly true in the case of exhaled breath analysis. Ultra-low concentrations of volatile organic breath constituents (from low ppb to low ppt), presence of highly reactive species and high humidity inducing wet chemistry make breath samples particularly vulnerable to all problems related to storage.16 Despite availability of real-time techniques such as proton transfer reaction mass spectrometry (PTR-MS),714 or selected ion flow tube mass spectrometry (SIFT-MS),15,16 gas chromatography mass spectrometry (GC-MS) remains the gold standard for the analysis of breath constituents.13,5,17 Since GC-MS analysis of exhaled air is usually coupled with a time-consuming sample pre-concentration method (e.g., solid phase microextraction (SPME), sorbent trapping, or needle traps (NTD)), sample storage is an inherent part of the analytical procedure. In this context the selection of the optimal storage conditions for breath samples is of particular importance.

Currently, Tedlar (PVF, polyvinyl fluoride) is one of the most popular and commonly accepted materials for collecting gaseous samples in general and breath gas samples in particular.1,3,1826 This is due to its moderate price, inertness, relatively good durability, and reusability. A number of studies have investigated the storage of breath constituents in polymer bags. Nevertheless, the majority of them focused on a limited number of species at levels much higher than the ones observed in breath, or dealt with a single analyte related to a specific disease or disorder.1921,27,28 For example, Groves and Zellers27 tested the influence of high humidity on the recovery of 6 breath-related compounds at the ppm level. The observed differences between dry and wet matrices were smaller than 10%. Steeghs et al.20 investigated the stabilities of 7 species (methanol, acetaldehyde, acetone, isoprene, benzene, toluene and styrene) at approximately 100 ppb level over a period of 72 hours in black-layered Tedlar. The results evidenced good recovery (better than 80%) of acetone, isoprene, acetaldehyde and benzene over this storage period. A more detailed study involving 12 breath species at levels of 70–85 ppb was performed by Beauchamp et al.19 After 10 h of storage, the observed losses were smaller than 20%.

The main goals of this study were the investigation and comparison of stabilities of selected breath constituents in three types of polymers bags, Tedlar, Kynar, and Flexfilm, as well as the identification of optimal storage conditions for breath samples. The 41 selected C3–C10 species represented different chemical classes (hydrocarbons, ketones, aldehydes, aromatics, sulphurs, esters, terpenes, heterocyclics, etc.) and exhibited close-to-breath low ppb concentrations. The tests comprised the background emission of pollutants, recovery from dry and humid (RH 100% at 32 °C) samples over a period of 7 days, and the influence of sample size (degree of bag filling) on sample stability. Finally, the effectiveness of the cleaning protocol was examined as a crucial factor for bag reusability. Gas chromatography with mass spectrometric detection coupled with solid phase microextraction (SPME) as the pre-concentration method was selected as the analytical tool during all experiments. In the framework of the present study SPME exhibits some advantages over alternative pre-concentration methods (e.g., sorbent trapping) like ease of operation, good sensitivity, excellent reproducibility, full automation and relatively small extraction dependence on humidity.29 Finally, this pre-concentration method requires relatively small amounts of sample volume (10–20 ml) to perform extraction. The latter feature was particularly beneficial during this study, as the initial volumes of the samples in the bags remained almost unaffected during experiments. Alternative techniques (e.g., SPE) are more time- and effort-consuming and usually require much larger sample volumes.

2. Experimental

2.1. Sampling bags

Within this study, three types of sampling bags were compared with respect to the stability of breath constituents:

  • 3 l in volume transparent Tedlar (PVF – polyvinyl fluoride) bags (SKC Inc., USA) equipped with a single polypropylene valve (dimensions when deflated: 26 cm × 24.5 cm, film thickness: 50 μm).

  • 3 l in volume SamplePro Flexfilm bags (unknown polymer – trade secret of SKC Inc., USA) equipped with a single polypropylene valve (dimensions when deflated: 21 cm × 41.5 cm, film thickness: 76 μm).

  • 3 l in volume Kynar (PVDF – polyvinylidene difluoride) bags (SKC Inc., USA) equipped with a single polypropylene valve (dimensions when deflated: 26 cm × 28.5 cm, film thickness: 50.8 μm).

All bags were new and flushed five times with high-purity nitrogen (type 6.0 – 99.9999%) directly before their use.

2.2. Chemicals and standards

Multi-compound test gas mixtures as well as calibration mixtures were prepared from pure liquid or gaseous substances. The majority of them were purchased from Sigma-Aldrich (Vienna, Austria): n-butane (99%), n-pentane (99.8%), n-hexane (99%), n-octane (99.8%), n-decane (99%), isobutane (99%), 3-methyl pentane (99%), 2-butene E and Z (99%), 2-pentene E and Z (99%), 1-hexene (97%), methylcyclopentane (97%), α-pinene (98%), (+)-3-carene (98.5%), p-cymene (99%), D-limonene (99%), eucalyptol (99%), benzene (99.8%), toluene (99.8%), p-xylene (99%), o-xylene (99%), acetone (99.8%), 2-butanone (99.5%), 2-pentanone (99%), 4-heptanone (97%), 2-butenone (99%), propanal (97%), 2-methyl propanal (99.5%), butanal (99%), hexanal (98%), octanal (99%), 2-methyl-2-propenal (95%), furan (99%), 2-methyl furan (99%), 2,5-dimethyl furan (99%), thiophene (99%), 3-methyl thiophene (98%), methyl acetate (99.5%), ethyl acetate (99.9%), n-propyl acetate (98%), methyl methacrylate (99%), dimethyl selenide (99%), ethyl ether (99.7%), pyrimidine (99%) and acetonitrile (99.8%). Moreover, 2-methyl pentane (99.5%), 4-methyl heptane (97%), isoprene (99%), ethylbenzene (99.8%), dimethyl sulfide (99%), 2-methyl-1-pentene (99.5%) and n-butyl acetate (99.7%) were obtained from Fluka (Switzerland), whereas, 2,4-dimethyl heptane (95%), 2,4-dimethyl-1-heptene (94%) and 4-methyl octane (97.5%) were provided by Chemsampco (USA). 3-Methyl furan (98%) was purchased from Acros Organic (Belgium) and methyl propyl sulfide (98%) from SAFC (USA).

The standard mixtures were prepared in two steps. Firstly, multi-compound primary standards were prepared in 1 l glass bulbs (Supelco, Canada). Prior to the use, each bulb was thoroughly cleaned with methanol and dried at 70 °C for at least 12 h. Then, the bulb was evacuated using a vacuum membrane pump and approximately 1 μl of a liquid (or 0.5 ml of gaseous) analyte was injected through a rubber septum. Next, the bulb was heated to 60 °C for 30 min to ensure complete evaporation and subsequently balanced to ambient pressure with high-purity nitrogen (6.0 – 99.9999%). The final calibration or test mixtures were prepared by transferring appropriate volumes of primary standard with Hamilton syringes into sampling bags filled in advance with predefined amounts of high-purity nitrogen. Calibration curves were obtained on the basis of triplicate analyses of 7 mixtures. Humid test mixtures were prepared in an analogous way as dry samples; however, during the last step polymer bags were filled with humid zero-air produced by means of a generator GasLab (Breitfuss Messtechnik, Germany). The GasLab unit comprises an integrated zero-air generator and a humidification module enabling the preparation of gas mixtures at predefined humidity levels. To avoid water condensation and to mimic conditions during breath sampling, the transfer line and polymer bags were maintained at 37 °C during the filling procedure. In all cases the gas volumes in the polymer bags were measured using an EL-FLOW F201CV digital mass flow controller (Bronkhorst hightech B.V., Netherlands).

A great majority of human breath constituents exhibit very low concentration levels ranging from ppt to several ppb.14,22 Consequently, an effort has been made to investigate stabilities of breath compounds at levels close to the ones observed in real samples. Effectively, the multi-compound test mixture contained analytes with concentrations falling within the range of 3–12 ppb. The three exceptions were acetone (720 ppb) and isoprene (106 ppb) exhibiting higher physiological levels in human breath,23,24 as well as acetonitrile (42 ppb) showing higher LOD for the applied analytical method. The range of volume fractions used during calibration and validation of the analytical method as well as the compounds’ concentration levels in the multi-compound test mixture are presented in Table 1.

Table 1.

Retention times Rt (min), LODs (ppb), RSDs (%), coefficients of variation (R2), linear ranges (ppb) of compounds under study and levels of species in the multicompound test mixture. Compounds are ordered with respect to retention time

VOC CAS Rt [min] Test mixture level [ppb] RSD [%] LOD [ppb] R2 Linear range [ppb] Quantifier ion
Isobutane 75-28-5 10.90 6   6.7 0.32 0.998 1–30   43
2-Butene, (E) 624-64-6 11.01   1.8 0.3 0.995 1–17   56
2-Butene, (Z) 590-18-1 11.11   3 0.3 0.994 1–22   56
Acetonitrile 75-05-8 11.44 42   7.4 4 0.999 12–62   41
n-Butane 106-97-8 11.85 6.2   4.3 0.19 0.987 0.63–25   43
Furan 110-00-9 13.39 12   2.1 0.22 0.999 0.6–22   68
Propanal 123-38-6 13.51 22   2.6 0.6 0.997 2–45   58
Acetone 67-64-1 13.65 720   2.6 0.74 0.999 3–1000   58
Dimethyl sulfide 75-18-3 14.33 10   1.5 0.1 0.999 0.3–30   62
Methyl acetate 79-20-9 15.06 12   2.2 0.14 0.999 0.4–25   43
Ethyl ether 60-29-7 15.94 8   1.2 0.29 0.999 1–20   74
Isoprene 78-79-5 16.10 106   1.3 0.1 0.999 0.5–175   67
2-Pentene, (E) 646-04-8 16.32 8   1.6 0.1 0.999 0.4–10   55
2-Pentene, (Z) 627-20-3 16.48 5   2.5 0.14 0.998 0.3–6   55
n-Pentane 109-66-0 16.57 6.2   1.6 0.11 0.996 0.4–25   43
Dimethyl selenide 593-79-3 16.76 10   3 0.23 0.998 0.6–12.2   95
2-Propenal, 2-methyl- 78-85-3 16.98   1 0.11 0.998 0.4–29   70
Propanal, 2-methyl- 78-84-2 17.25   6.5 0.26 0.997 0.8–15.7   43
3-Buten-2-one 78-94-4 17.59   5 0.19 0.998 0.6–23   55
Butanal 123-72-8 18.03   3 0.4 0.988 1.2–12   72
Furan, 2-methyl- 534-22-5 18.10 7   2 0.1 0.998 0.3–18   82
2-Butanone 78-93-3 18.20 9   7 0.13 0.997 0.4–36   43
Furan, 3-methyl- 930-27-8 18.39   3 0.15 0.997 0.4–20   82
Ethyl acetate 141-78-6 18.96 8   2.2 0.13 0.996 0.4–17   43
Thiophene 110-02-1 19.93 9   3.3 0.15 0.999 0.45–21   84
1-Pentene, 2-methyl- 763-29-1 19.95   3.2 0.1 0.999 0.3–15   56
Pentane, 2-methyl- 107-83-5 20 9.5   1 0.18 0.999 0.55–11   43
Pentane, 3-methyl- 96-14-0 20.19   1.5 0.1 0.999 0.4–12   57
1-Hexene 592-41-6 20.22 9   1.7 0.2 0.999 0.6–10   56
Benzene 71-43-2 20.38 12   3.8 0.3 0.998 1–36   78
Cyclopentane, methyl- 96-37-7 20.45   9 0.1 0.991 0.3–11   56
n-Hexane 110-54-3 20.70 6.2   1.6 0.12 0.995 0.4–25   57
Pyrimidine 289-95-2 21.70 10   9 0.1 0.972 0.4–28   80
2-Pentanone 107-87-9 21.98 8   2.2 0.1 0.998 0.4–24   43
Furan, 2,5-dimethyl- 625-86-5 22.04 7.5   1.4 0.08 0.999 0.3–15   96
Methyl methacrylate 80-62-6 22.11   1.6 0.11 0.999 0.4–18   69
n-Propyl acetate 109-60-4 22.71   1.3 0.15 0.999 0.5–17   43
Methyl propyl sulfide 3877-15-4 22.73 6   2.1 0.04 0.996 0.2–30   61
Thiophene, 3-methyl- 616-44-4 24.00 6.5   4.2 0.1 0.996 0.3–22   97
Toluene 108-88-3 24.30 12   2.9 0.1 0.993 0.3–30   91
Hexanal 66-25-1 25.76 5   9 0.4 0.996 1.2–10   56
n-Butyl acetate 123-86-4 26.21 6   2.1 0.4 0.995 1.2–12   56
Heptane, 4-methyl- 589-53-7 26.76   2.8 0.24 0.989 0.6–11.6   43
Ethylbenzene 100-41-4 27.45   7 0.25 0.989 0.75–13   91
n-Octane 111-65-9 27.60 8   2.8 0.1 0.998 0.3–14   85
p-Xylene 106-42-3 27.72 8.5   8 0.07 0.986 0.3–18   91
o-Xylene 95-47-6 28.03   6 0.15 0.991 0.4–15   91
4-Heptanone 123-19-3 28.36 5   6.3 0.06 0.978 0.2–17   71
Heptane, 2,4-dimethyl- 2213-23-2 28.98   5.9 0.1 0.987 0.3–8.8   85
1-Heptene, 2,4-dimethyl- 19549-87-2 29.05   6 0.12 0.986 0.4–9   83
Octane, 4-methyl- 2216-34-4 29.76 7   3.4 0.2 0.995 0.6–11   43
α-Pinene 80-56-8 30.81 6   8 0.46 0.985 1.4–19   93
Octanal 124-13-0 31.87 3 11 0.3 0.974 1–17   84
(+)-3-Carene 498-15-7 32.14 4   5.3 0.61 0.954 1.8–12   93
p-Cymene 99-87-6 32.67 7   5.7 0.1 0.973 0.4–21 119
D-Limonene 5989-27-5 32.88 9   6 0.45 0.954 1.4–18   68
n-Decane 124-18-5 33.21 7   9 0.4 0.978 1.2–17   57
Eucalyptol 470-82-6 33.46 8.5   6.1 1 0.986 3–25   43
n-Dodecane 112-40-3 36.22   8 0.5 0.964 1.5–15   71

2.3. SPME procedure and chromatographic analysis

The test gas samples were taken using a 20 ml gas-tight glass syringe (Roth, Germany) equipped with a replaceable needle. Sampling was achieved manually by drawing a volume of 18 ml from the sampling bag and subsequent injection of this volume into an evacuated SPME vial (20 ml in volume, Gerstel, Germany) sealed with a 1.3 mm butyl/PTFE septum (Macherey-Nagel, Germany). To avoid loss of analytes during the sample storage in SPME vials an effort was made to analyze samples within 3 h after vial filling. The SPME procedure was carried out automatically using a multipurpose sampler MPS (Gerstel, Germany). SPME was achieved by inserting a 75 μm carboxen-polydimethylsiloxane (CAR–PDMS) fiber (Supelco, Canada) into the vial and exposing it to its content for 10 minutes at 37 °C. Immediately after extraction, the fiber was introduced into the inlet of the gas chromatograph where the sorbed VOCs were thermally desorbed at 290 °C. The fiber was conditioned at 290 °C for 5 minutes prior to each analysis.

The GC-MS analyses were performed using an Agilent 7890A/5975C GC-MS system (Agilent, USA). During the fiber desorption, the split/splitless inlet operated in a splitless mode (1 min), followed by the split mode at a ratio of 1 : 20. The analytes under study were separated using a PoraBond Q column (25 m × 0.32 mm, film thickness 5 μm, Varian, USA) working in a constant flow mode of helium at 1.4 ml min−1. The column temperature program was as follows: 40 °C for 2 min, increase to 260 °C at a rate of 7 °C min−1, held at 260 °C for 7 min. The mass spectrometer worked in a SCAN mode with an associated m/z range set from 20 to 200. The quadrupole, ion source and transfer line were kept at 150 °C, 230 °C and 280 °C, respectively.

The identification of compounds was performed in two steps. First, the peak spectrum was checked against the NIST mass spectral library. Next, the NIST identification was confirmed by the retention times obtained on the basis of standards prepared from pure compounds. Peak integration was based on extracted ion chromatograms. The retention times of the investigated compounds for the applied chromatographic parameters as well as the ions used for the integration are presented in Table 1. An exemplary chromatogram from a test mixture analysis is presented in Fig. 1.

Fig. 1.

Fig. 1

An exemplary chromatogram from a test mixture analysis.

2.4. Sampling bags tests

2.4.1. Background test

To identify contaminants emitted by the polymer films 3 new bags of each type were filled with 2000 ml of high-purity nitrogen (corresponding to approximately 67% of their nominal volume), stored at room temperature for 24 hours and analyzed after certain time periods. The time intervals of the measurements were defined as follows: the first sampling was carried out immediately after filling the bag, subsequent ones after 6, 12 and 24 h of storage. Additionally, blank (nitrogen) and laboratory air measurements were performed.

Next, an additional cleaning procedure was applied to check if it is possible to reduce contaminant emission from the bags under study. For this purpose, after five-fold flushing bags were filled with nitrogen and heated overnight (approximately 12 h) in an oven at 50 °C to induce potential contaminant desorption from the polymer film or from the valves. Next, bags were again flushed five times with nitrogen and the aforementioned background test procedure was repeated; however, in this case only 3 samplings were performed, immediately after filling, and after 6 and 24 hours of storage.

2.4.2. Dry standard stability test

To investigate the stability of breath species, a 41-component test mixture was prepared using the aforementioned procedure and injected into the tested polymer bags. The nominal levels of all compounds in the test mixture are presented in Table 1. To study the influence of different (film) surface-to-(sample) volume ratios (SA : V) on the sample integrity, three bags of each type of film were filled with different volumes of standard mixture: 2.4, 1.2 and 0.6 l (i.e., 80, 40, and 20% of the maximum capacity). Due to some differences in the film dimensions these volumes corresponded to SA : V ratio values of 53, 106, and 212 m−1 for Tedlar bags, 73, 145, and 291 m−1 for Flexfilm bags, and 62, 124, 247 m−1 for Kynar bags respectively. All bags were pre-conditioned and filled at the same time with the same test mixture and were stored at room temperature (24 °C) exposed to daylight. The stability of the test gas was monitored over a period of 7 days with the time instants for drawing the samples defined as follows: the first sample was taken approximately 10 minutes after the bag filling and the next ones after 6, 24, 48, 72, 126 and 168 hours of storage.

To confirm repeatability, the stability test was repeated for the test gas mixture volume of 2.4 l with the same sampling protocol, however, with three bags of each type being involved.

2.4.3. Humid standard stability test

The stability of compounds under study in humid matrices was investigated using a test mixture having water content similar to breath leaving the upper airways, i.e., RH of 100% at 32 °C.30 Three new bags from each type of film were filled with 2.4 l of humid test mixture and sampled immediately after filling and after 6, 24 and 48 h of storage. To avoid condensation and to mimic the sampling of real breath samples (having body temperature) during filling all bags were heated to 37 °C. However, during the experiment they were stored at room temperature. The duration of the experiment was restricted to 2 days as water vapor permeates relatively easily through all tested polymer films, and after a few hours sample humidity reaches ambient levels.19,31,32 Additionally, one bag of each type was sampled after 2 and 4 hours to study the evolution of the VOC concentrations during the first hours of storage, when the humidity still remains elevated.

2.4.4. Reusability test

The reusability test was focused on studying the effectiveness of the bag cleaning protocol developed during one of our previous studies.21 Polymer bags involved in the dry standard stability test (i.e., containing the test mixture for 7 days) were used during the test. Firstly, bags were flushed five times with high purity nitrogen to remove remainings of the test gas. Next, all bags were filled with 2 l of nitrogen and conditioned at 50 °C for approximately 12 h to remove volatiles of interest from the bags’ material (film, valve etc.). After heating, the bags were again rinsed five times, filled with 2 l of high-purity nitrogen and stored at room temperature for 24 hours. The effectiveness of the applied cleaning protocol was checked by comparing the levels of test mixture VOCs before and after this time of storage.

3. Results and discussion

3.1. Method validation

Limits of detection (LODs) were calculated using the mean value of the blank responses and their standard deviations obtained on the basis of 10 blank measurements33 and are presented in Table 1. The limit of quantification (LOQ) was defined as three times the LOD. The relative standard deviations (RSDs) were calculated on the basis of five consecutive analyses of standard mixtures. The calculated RSDs varied from 1% to 9% and were recognised as satisfactory for the aims of this study. The system response was found to be linear within the investigated concentration ranges, as shown in Table 1, with coefficients of variation ranging from 0.954 to 0.999.

3.2. Background test

All volatiles found to be emitted by the investigated polymer sampling bags are summarised in Tables 24. The presented concentrations are the mean values of VOC levels in three bags. The emission rates were calculated for unconditioned bags assuming that the contaminants are emitted by the polymer film.

Table 2.

Contaminants emitted by Flexfilm bags [ppb]. Compounds are ordered with respect to increasing retention time. ʺ—ʺ denotes that the VOC was not detected, whereas ʺ<LOQʺ stands for VOC level below LOQ

New bag sampling time [h]
Preconditioned bag sampling time [h]
VOC CAS 0 6 12 24 0 6 24 VOC emission × 10−12 [g × h−1 × cm−2]
Carbonyl sulfide (COS) 463-58-1 Not quantified
Acetaldehyde 75-07-0 Not quantified
2-Butene, (E) 624-64-6 1.6 2.0 2.3 0.6 0.39
2-Butene, (Z) 590-18-1 3.6 6.7 12.2  0.8  3.6 1.53
n-Butane 106-97-8 1.6 6.2 11.1 19.1 <LOQ  3.7  8.1 2.62
Propanal 123-38-6 <LOQ 1.4 2.1 2.4 0.48
Acetone 67-64-1 14 75 102 140  9.9  19 25.7
Carbon disulfide (CS2) 75-15-0 Not quantified
Methyl acetate 79-20-9 <LOQ <LOQ 0.3 0.4 0.1
n-Pentane 109-66-0 0.4 2.0 3.4 5.5 0.4  1.7  3.4 0.99
2-Propenal, 2-methyl- 78-85-3 <LOQ <LOQ 0.3 0.4 <LOQ 0.1
3-Buten-2-one 78-94-4 <LOQ <LOQ 0.6 0.6 <LOQ <LOQ 0.19
Butanal 123-72-8 <LOQ 2.2 3.6 4.5  1.3  1.3 1.01
2-Butanone 78-93-3 0.5 1.8 2.9 4.1  0.6  1.0 0.85
Pentane, 2-methyl- 107-83-5 <LOQ 0.7 1.1 1.8 <LOQ  1.4  3.0 0.41
1-Hexene 592-41-6 <LOQ 0.9 1.4 2.2 <LOQ  0.8  1.6 0.49
Pentane, 3-methyl- 96-14-0 <LOQ 1.1 2.0 3.1 0.4  1.7  3.8 0.68
Cyclopentane, methyl- 96-37-7 1.7 6.6 11.1 17.3 2.0  9.2  18.7 3.78
n-Hexane 110-54-3 3.8 17.1 28.6 44.8 4.8  22.7  45.6 10
Methyl methacrylate 80-62-6 1.6 6.7 10.1 14.6 0.4  1.4  2.6 4.22
Toluene 108-88-3 <LOQ 1.2 1.8 2.4 <LOQ  0.3  0.5 0.67
Heptane, 4-methyl- 589-53-7 <LOQ 1.2 2.0  1.2  2.4 0.53
Heptane, 2,4-dimethyl- 2213-23-2 1.9 8.9 14.8 23.2 3.4  14.8  30.4 7.76
1-Heptene, 2,4-dimethyl- 19549-87-2 <LOQ 0.6 0.8 <LOQ  0.5  1.1 0.26
Octane, 4-methyl- 2216-34-4 <LOQ 2.8 4.5 6.8 1.0  4.1  8.6 2.37
Caprolactam 105-60-2 Not quantified
Dodecane 112-40-3 <LOQ 3.3 5.3 6.4 2.0  3.3  5.2 3.51

Table 4.

Contaminants emitted by Tedlar bags [ppb]. Compounds are ordered with respect to the increasing retention time. ʺ—ʺ denotes that the VOC was not detected, whereas ʺ<LOQʺ stands for VOC level below LOQ

New bag sampling time [h]
Preconditioned bag sampling time [h]
VOC CAS 0 6 12 24 0 6 24 VOC emission × 10−12 [g × h−1 × cm−2]
Carbonyl sulfide (COS) 463-58-1 Not quantified
Acetonitrile 75-05-8 <LOQ 14.2 17.3 18.9 4.23
Carbon disulfide (CS2) 75-15-0 Not quantified
n-Hexane 110-54-3 0.5 0.15
Acetamide, N,N-dimethyl- 127-19-5 Not quantified
2-Propyl acetate, 1-methoxy- 108-65-6 Not quantified
Heptane, 2,4-dimethyl- 2213-23-2 0.5 0.8 1.3 0.3 0.4 0.5
Octane, 4-methyl- 2216-34-4 <LOQ 0.8 1.0 1.5 <LOQ <LOQ 0.82
Phenol 108-95-2 Not quantified

A total of 27 compounds were emitted by Flexfilm bags. The most dominant chemical classes were hydrocarbons with 14 and aldehydes with four species, respectively. Amongst the remaining compounds, there were three ketones, two esters, two volatile sulphur compounds, one aromatic and one amide. Sulphur compounds (COS and CS2) were found to be produced by all three types of bags and seem to be emitted by rubber parts of the sampling valves.21 As the emission of these two species was investigated thoroughly in our previous paper21 they were not quantified within this study. Acetone was the most abundant compound with the concentration level reaching 140 ppb after 24 hours of storage. Apart from acetone, high concentrations were observed for some hydrocarbons (2-butene, n-butane, methylcyclopentane, n-hexane, 2,4-dimethyl heptane). 75% of the contaminants were detected shortly after filling the Flexfilm bag. Six hours later all of them were present in the bags at levels of several ppb. In the context of breath gas analysis this emission can be considered as significant. Pre-conditioning of Flexfilm bags reduced the emission of aldehydes, ketones and esters by 50–80%; however, the emission rates of hydrocarbons remained unchanged (with the exception of 2-butene and toluene). Acetone background was particularly improved with levels spreading around 20 ppb after conditioning and 1 day of storage. Nevertheless, despite conditioning considerable concentrations of contaminants could be found after several hours of storage. Repeating the pre-conditioning step (data not shown) further improved the background of all species apart from hydrocarbons.

Kynar bags released 21 species. The predominant chemical class was aromatics with four compounds. Apart from them, there were two hydrocarbons, two volatile sulphur compounds (COS and CS2), three aldehydes, three ketones and three esters, one CFC, one halide and one nitro compound. In the case of unconditioned Kynar bags the great majority of contaminants could be detected in small amounts (usually below 1 ppb) after 6 hours of storage. At the end of the experiment the highest levels were noted for acetone and toluene, 33 and 9.4 ppb respectively. The applied pre-conditioning method was found to be very efficient in the case of Kynar bags. After cleaning only five species (2-butanone, toluene, p-xylene, COS and CS2) could be detected in the Kynar bag samples after 6 hours of storage. After 1 day several additional contaminants were found in the Kynar bags in detectable amounts, however, their levels were below the LOQs of the analytical method. Taking into account the good results of the pre-conditioning protocol it can be surmised that additional cleaning/s could further reduce the contaminants emission to the levels acceptable for breath analysis. Conversely, conditioning promoted the emission of sulphur species – COS and CS2 – which is consistent with the findings of our previous paper21 indicating rubber parts of polymer bags (o-ring, septum) as potential sources of these species.

Only 9 compounds were found to be emitted by the Tedlar film: three hydrocarbons (n-hexane, 2,4-dimethylheptane and 4 methyl octane), two volatile sulphur compounds (COS and CS2), N,N-dimethylacetamide, phenol, acetonitrile and 1-methoxy-2-propyl acetate. N,N-Dimethylacetamide and phenol are commonly known and well documented contaminants in Tedlar bags.19,34 Amongst the quantified species the highest levels were noted for acetonitrile (19 ppb). 2,4-Dimethylheptane and 4-methyl octane could be detected within few hours of storage, however, n-hexane was found only at the end of the experiment. Like in the case of Kynar bags pre-conditioning considerably improved the background emission. However, small amounts of 2,4-dimethylheptane and 4-methyl octane could still be detected after 6 hours of storage.

Several compounds identified as contaminants in the tested bags were found also in room air at low ppb levels. Hence, permeation from room air might be a possible source of pollution, e.g., for acetone, 2-butanone, n-butane, n-pentane, methyl acetate, toluene, and p-xylene. However, the fact that several other species having similar physicochemical properties were detected in room air but not in bag air favours emissions from the polymer film as the main source of contamination. For example, 2-methyl butane – a hydrocarbon very similar to n-pentane – was present in room air at levels of several ppb but was not detected in the bag content during the background test.

3.3. Dry standard stability test

The stability of test mixture compounds in tested polymer bags over a period of one week is presented in Table 5. A compound’s concentration was considered stable when its level was higher than 80% of its initial value.

Table 5.

Stability of selected breath constituents in Tedlar, Kynar and Flexfilm sampling bags for the dry matrix. Vf–filling volume in %. The recoveries exceeding 100% reflect the emission of pollutants

Recovery in Flexfilm bag [%]
Recovery in Kynar bag [%]
Recovery in Tedlar bag [%]
VOC Vf[%]  6 h  24 h  48 h  72 h  126 h  168 h  6 h  24 h  48 h  72 h  126 h  168 h  6 h  24 h  48 h  72 h  126 h  168 h
n-Butane 20 180 424 714 1020 1621 1978 113 107 109 131 130 95 94 92 90 88 85
40 159 296 454 607 963 0 93 94 89 85 86 84 93 95 94 91 94 81
80 142 238 352 471 693 840 112 113 103 106 104 104 103 102 97 93 94 93
n-Pentane 20 120 176 210 296 412 486 94 91 91 84 82 92 93 89 85 79 76
40 111 132 162 193 260 0 96 92 84 82 78 77 99 93 88 81 85 80
80 98 113 137 154 170 205 112 106 105 98 96 90 102 100 91 86 86 81
n-Hexane 20 215 472 721 992 1448 1723 103 101 100 129 125 99 97 95 90 82 87
40 179 347 528 673 1019 0 92 91 88 81 83 75 94 96 90 90 93 82
80 152 263 387 495 729 855 111 114 107 103 103 107 106 104 95 94 96 92
n-Octane 20 95 89 81 78 69 67 91 86 83 74 75 99 91 89 86 82 78
40 97 91 85 85 81 0 93 92 88 84 81 78 98 95 88 85 86 80
80 93 92 89 84 80 80 108 107 102 102 98 84 98 91 91 87 86 82
n-Decane 20 82 71 59 50 56 40 95 78 73 64 57 86 78 69 57 50 50
40 96 92 69 59 55 0 98 83 78 71 63 49 92 85 76 71 62 57
80 106 92 83 70 62 56 102 91 94 87 85 76 100 92 89 85 84 70
Isobutane 20 96 88 88 95 96 99 90 93 79 92 86 107 100 86 83 86 84
40 99 93 86 69 80 0 99 96 79 77 74 80 101 90 88 74 77 74
80 89 88 69 72 75 74 100 85 85 82 76 89 98 99 82 84 87 80
Pentane, 2-methyl- 20 102 132 138 149 187 203 102 100 89 92 88 94 81 84 83 83 81
40 91 92 98 95 104 0 101 98 93 84 88 83 89 92 82 85 85 72
80 87 86 87 85 89 97 110 102 101 95 98 88 99 97 86 88 89 93
Octane, 4-methyl- 20 124 178 231 263 332 358 100 99 104 105 104 105 125 123 147 148 179
40 110 140 162 166 222 0 98 103 93 105 103 108 91 101 112 110 125 107
80 110 107 140 131 170 169 99 104 102 103 103 89 102 103 100 104 111 112
2-Pentene, (E) 20 94 93 83 80 82 81 88 75 63 48 37 89 93 88 84 83 81
40 93 81 85 80 80 0 93 87 73 63 44 30 97 94 90 87 86 82
80 89 84 87 85 85 82 109 105 97 88 72 58 98 95 89 88 87 78
2-Pentene, (Z) 20 103 94 85 96 82 80 85 77 64 53 41 91 93 90 84 82 80
40 113 105 104 99 104 93 90 75 66 50 36 96 92 91 85 84 83
80 94 88 93 88 88 85 109 105 95 89 76 65 99 93 89 89 87 82
1-Hexene 20 96 127 154 172 253 285 99 88 77 78 75 96 92 89 85 83 80
40 99 112 129 138 173 94 88 85 76 74 68 92 91 84 87 88 80
80 97 103 111 115 135 145 111 107 101 96 93 88 97 94 83 85 81 84
Isoprene 20 93 87 80 70 70 65 85 77 69 57 47 88 90 86 82 79 77
40 95 83 83 77 77 93 88 76 71 60 50 97 92 89 86 84 81
80 91 86 87 83 80 77 109 105 98 92 83 75 101 102 93 94 96 89
Benzene 20 85 80 69 65 58 53 90 77 63 54 54 92 85 79 74 68 65
40 93 86 82 71 71 88 80 71 65 61 56 88 86 79 77 74 69
80 93 87 83 73 76 70 105 99 91 84 81 70 97 91 82 79 81 79
Toluene 20 85 78 67 61 56 51 81 66 60 52 46 90 81 73 64 59 55
40 87 84 79 70 66 88 79 71 63 56 52 90 84 78 72 66 62
80 92 88 82 77 73 67 100 94 85 78 70 64 95 91 85 82 78 73
p-Xylene 20 80 72 55 49 42 40 71 55 48 40 39 84 69 64 52 45 43
40 89 78 74 63 63 0 76 69 56 51 48 38 87 79 68 62 59 53
80 86 83 76 70 64 52 94 84 73 68 58 49 91 88 82 76 72 67
Acetone 20 95 95 92 94 92 90 71 58 53 39 37 89 79 71 65 55 52
40 90 93 90 89 92 73 59 47 43 32 26 90 82 77 72 67 61
80 94 91 91 88 86 86 97 82 70 62 51 39 103 97 92 88 84 81
2-Butanone 20 123 133 158 210 217 228 87 80 80 72 71 89 79 76 69 55 56
40 121 139 155 152 226 82 72 60 60 56 52 102 96 89 83 77 65
80 105 122 131 123 140 163 103 83 81 74 67 62 99 92 86 85 79 76
2-Pentanone 20 86 74 65 60 55 47 64 42 34 24 20 89 78 65 61 55 47
40 89 86 79 75 70 71 53 40 33 23 19 91 82 72 70 65 59
80 94 92 81 82 76 70 96 78 59 54 43 36 100 93 84 78 81 77
4-Heptanone 20 86 65 51 42 29 26 42 27 22 10 12 82 64 55 51 38 31
40 87 78 66 60 51 59 38 26 20 16 13 89 73 69 60 55 49
80 92 87 85 68 60 53 82 60 44 38 24 21 100 83 86 80 76 70
Propanal 20 132 174 208 233 278 299 84 75 75 71 67 96 86 83 76 70 67
40 118 154 175 194 239 85 78 76 71 66 66 94 89 83 80 76 71
80 108 121 134 142 156 166 109 98 90 91 85 75 99 93 87 85 81 81
Hexanal 20 104 86 121 139 125 145 72 130 115 98 92 101 100 95 94 91 85
40 103 101 135 114 129 92 78 78 73 72 72 103 101 98 95 93 91
80 100 107 123 98 76 126 155 140 115 126 110 125 99 98 99 103 88 95
Octanal 20 101 95 96 80 65 82 72 63 54 45 61 98 66 69 45 50 51
40 113 85 87 109 80 79 45 63 52 40 51 83 87 85 61 55 50
80 132 99 95 90 85 89 110 93 86 80 53 53 101 88 77 64 64 52
Furan 20 93 97 97 100 106 107 85 70 60 46 37 86 86 79 75 70 64
40 90 97 99 102 111 86 81 74 63 54 47 93 87 84 79 76 71
80 98 94 92 92 92 92 107 101 93 87 78 67 97 93 88 86 84 81
Furan, 2-methyl- 20 88 76 70 65 58 53 77 59 49 34 27 93 82 76 71 66 62
40 92 88 83 80 78 87 74 65 56 42 34 103 96 90 83 81 76
80 95 89 84 79 76 73 106 89 87 76 65 56 97 87 78 84 81 78
Furan, 2,5-dimethyl- 20 89 74 57 44 29 22 66 42 26 12 7 91 80 67 62 54 47
40 86 81 68 58 46 86 68 47 29 13 6 96 86 75 72 66 59
80 99 91 82 77 67 59 106 93 72 59 37 22 101 92 82 77 79 72
Methyl acetate 20 89 86 79 68 73 71 65 53 47 42 44 81 79 68 64 57 53
40 94 88 83 79 81 71 61 50 41 29 22 92 85 79 74 67 63
80 93 88 86 84 80 72 99 85 72 63 47 38 99 92 87 84 80 77
Ethyl acetate 20 91 80 67 66 54 48 70 54 47 34 29 89 80 73 67 58 53
40 89 76 74 71 69 79 57 54 46 37 32 100 94 87 79 74 67
80 93 84 87 78 77 71 100 87 76 67 58 52 98 93 83 84 80 78
n-Butyl acetate 20 84 82 66 64 52 59 71 59 51 45 38 86 78 69 57 50 50
40 92 86 81 75 64 82 69 60 52 43 39 92 85 76 71 62 57
80 93 89 82 78 72 69 97 83 69 64 53 41 101 93 86 82 79 73
Dimethyl sulfide 20 87 73 59 50 36 30 69 45 30 16 9 90 86 80 73 67 61
40 88 76 63 55 46 80 75 63 50 39 31 89 86 80 77 71 69
80 93 83 78 73 65 61 104 95 85 76 63 55 97 89 85 84 82 79
Methyl propyl sulfide 20 89 76 64 56 44 37 73 47 31 14 7 93 83 79 73 65 59
40 86 84 75 66 58 85 74 59 48 35 25 91 88 82 76 73 69
80 95 89 77 77 68 66 104 96 79 73 60 52 95 91 85 84 81 78
Thiophene 20 77 70 58 54 45 40 75 61 52 39 31 86 76 65 59 50 47
40 82 79 73 65 60 83 72 63 55 46 40 87 80 74 68 62 51
80 86 84 78 74 68 62 101 93 84 77 68 62 99 100 82 84 79 80
Thiophene, 3-methyl- 20 73 56 43 38 29 24 64 45 37 27 20 83 66 55 48 39 34
40 83 71 60 53 42 77 61 50 42 33 27 86 75 65 59 52 46
80 89 78 71 64 57 51 96 83 71 63 52 46 99 93 85 80 74 69
(+)-3-Carene 20 85 71 61 54 42 37 88 84 58 48 42 83 84 80 71 57 60
40 97 89 88 76 67 92 80 79 64 46 36 94 91 95 85 77 62
80 104 98 93 80 68 66 91 92 91 84 76 69 93 82 88 83 76 67
α-Pinene 20 93 82 78 68 58 60 101 95 75 65 57 99 87 84 85 81 88
40 86 89 77 76 70 99 90 86 80 67 56 93 95 92 87 84 78
80 99 81 92 82 86 81 115 103 107 100 83 75 101 99 90 94 89 88
p-Cymene 20 77 59 45 38 29 27 67 50 40 34 26 83 75 63 56 44 43
40 85 77 67 58 50 79 65 54 48 39 30 93 87 75 67 63 51
80 98 87 78 68 59 54 89 76 69 63 52 46 98 93 85 81 78 63
D-Limonene 20 82 67 57 51 39 41 79 60 50 36 30 81 82 75 70 54 53
40 87 83 75 71 58 86 77 63 54 39 28 96 92 75 80 71 59
80 107 98 90 82 68 66 99 79 82 76 54 41 103 100 95 92 86 69
Eucalyptol 20 81 86 70 33 29 36 80 69 63 57 58 96 125 138 142 123 122
40 74 82 71 65 63 91 86 80 83 75 87 106 108 113 111 109 88
80 112 107 95 101 90 84 105 96 94 102 87 82 98 97 102 101 102 85
Ethyl ether 20 94 89 82 70 73 69 86 82 79 72 67 88 90 88 85 82 78
40 96 86 83 80 78 94 90 84 83 76 73 96 92 89 85 87 81
80 88 86 88 84 80 79 106 103 98 93 91 87 94 90 89 87 86 79
Acetonitrile 20 47 31 27 34 22 17 57 34 31 12 7 57 38 27 24 16 11
40 69 42 27 32 24 45 26 18 12 8 2 64 47 38 31 23 17
80 74 50 43 36 27 28 66 39 22 12 4 6 87 65 56 49 38 33
Dimethyl selenide 20 79 65 46 44 30 26 37 12 4 2 1 89 77 65 51 40 31
40 82 64 54 45 37 78 53 29 17 7 3 96 86 76 64 61 48
80 87 81 73 71 64 59 96 74 55 40 23 15 96 89 84 78 73 66
Pyrimidine 20 45 26 18 18 12 10 47 27 25 14 13 53 31 21 16 11 10
40 52 40 30 24 17 36 19 13 11 6 5 65 42 29 25 17 14
80 80 58 42 38 30 25 56 33 17 17 10 8 82 66 50 42 32 28

For all bags the stability of the compounds of interest was strongly correlated with the volume of the test sample filled into the bag. Regardless of the chemical class of a compound, its recovery was significantly better when the sampling bag was filled up to 80% of its nominal volume. For the majority of all species, SA : V ratios below 100 m−1 provided good recoveries even after 7 days of storage. Samples with the highest surface-to-volume ratios (above 200 m−1) were stable only for several hours. This finding is not surprising, as the area of the bag materials (polymer film, valve, etc.) having contact with the sample for all SA : V ratios remained the same and its potential for interactions with the sample constituents was comparable. Consequently, large samples containing higher masses of the investigated species were more resistant to losses during storage. Additionally, samples stored at lower SA : V ratios were less susceptible to the emission of contaminants, as can be seen in Table 5. For example, in Flexfilm bags after 24 hours of storage the concentration of n-butane remained stable when the bag was filled up to 80% of its nominal volume, whereas in the bags filled up to only 20% of its nominal volume its concentration increased two-fold. Thus, it is strongly recommended to collect the largest possible volume of the sample in order to provide the optimal conditions for the preservation of its integrity.

Considerable differences were found for the stabilities of compounds in different polymer bags. Since the superiority of lower SA : V ratios of stored samples is undeniable, further discussion of the compound recoveries will refer to bags filled up to 80% (2.4 l) with the test mixture, unless otherwise stated.

3.3.1. Aliphatic hydrocarbons

The stabilities of hydrocarbons tested within this study suffered significantly from the background emission in Flexfilm and Kynar bags. Despite pre-conditioning, levels of numerous species tended to increase rapidly (even within 6 hours of storage). This phenomenon was particularly pronounced for Flexfilm bags, confirming the finding of the background tests. For HCs found not to be emitted by bag materials, good recoveries were noted even after 3 days of storage. Due to the much lower background emission Kynar bags provided better stability of HCs. The majority of species from this chemical class remained stable up to 3 days of storage (when filled up to 80% of maximum volume). Nevertheless, the risk of contamination considerably limits the applicability of Kynar and Flexfilm bags during breath studies aiming at hydrocarbons at low ppb levels. Hydrocarbons stored in Tedlar bags exhibited excellent recoveries over the whole investigated storage period, even for higher SA : V ratios. Only heavier hydrocarbons (e.g., n-decane) showed higher losses. Specifically, unsaturated hydrocarbons were much better preserved in Tedlar bags than in the other ones. Interestingly, in Kynar and Flexfilm bags the drop of isoprene levels (initial value of 106 ppb) was accompanied by the increase of 2-methyl-2-propenal, 3-buten-2-one and 3-methylfuran – species known to be the products of isoprene degradation in the atmosphere.35 For example, after 7 days concentrations of these species in Kynar bags filled with 2.4 l of test mixture were 12, 11, and 1.8 ppb, respectively. In Tedlar bags this effect was much less evident. In the case of this film only 3-buten-2-one was found to be produced (1 ppb after 1 day, 3 ppb after 7 days). Perhaps the presence of Kynar and Flexfilm films promotes the degradation of isoprene. As a result, due to the good background and excellent recoveries Tedlar bags seem to be the best choice for sampling and storage of breath hydrocarbons.

3.3.2. Aromatic hydrocarbons

In Kynar and Flexfilm bags the stability of the studied aromatics was relatively poor. In general, acceptable recoveries were observed only up to 24 hours of storage. In samples having higher SA : V ratios losses were pronounced even within the first hours of storage. In Tedlar bags recovery of species from this class was over 80% at the end of the investigated period, however, only in bags filled up to 80% of the maximum capacity. In all cases the values of recovery tended to decrease with increasing molecular mass of a compound.

3.3.3. Ketones

Recovery of ketones in Kynar bags was unsatisfactory. Their levels rapidly dropped below the arbitrarily chosen threshold of 80%. Even acetone having an initial concentration of 720 ppb followed this pattern. The apparently better stability of 2-butanone can easily be explained by its background emission from the Kynar film. Consequently, ketones stored in Kynar bags should be analysed within several hours after sampling. Much better recoveries were observed in Flexfilm bags, with characteristic drops related to the molecular mass of the compound. Once more Tedlar bags provided the best storage conditions for the discussed species. Apart from 4-heptanone, all ketones were stable for up to 7 days of storage. These results and the fact that Tedlar bags do not exhibit ketone release render this material optimal for the storage of species from this chemical category.

3.3.4. Aldehydes

Flexfilm bags were found to be inappropriate for the storage of aldehydes. Background emission significantly affected their initial concentrations. The 7 day monitoring period even revealed the emission of additional aldehydes (e.g., n-heptanal) not being detected during the 24 hour background test. Nevertheless, n-octanal stored in Flexfilm bags exhibited the best stability. In Kynar and Tedlar bags the stability of all tested aldehydes was comparable (up to 3 days).

For all remaining compounds (e.g., sulphurs, esters, terpenes) the superiority of storage in Tedlar bags is undisputed. For species with molecular mass up to 90 Tedlar bags provided good stability for up to 7 days of storage. Recoveries of heavier species were better than 80% only within 3–4 days. In Kynar and Flexfilm bags losses of these analytes were more evident and usually exceeded 20% in samples stored longer than one day. The stability of pyrimidine, acetonitrile and dimethyl selenide was especially poor. In all cases acetonitrile concentrations rapidly dropped even within the first several hours of storage. This finding is consistent with previous studies evidencing huge losses of this compound during storage due to the permeation through the polymer film.19 Its slightly better recovery in Tedlar bags can be explained by a background emission. A similar progression of stability was noted for pyrimidine with losses being acceptable only for Tedlar bags within 6 hours of storage. Dimethyl selenide showed good recoveries only in Tedlar bags.

3.4. Humid standard stability test

The comparison of recoveries of volatiles for dry and humid test mixtures is presented in Table 6. It must be stressed here that water permeates relatively easily through all tested materials and consequently, sample humidity remains elevated only for several hours of storage19,31,32 and subsequently reaches an ambient level. The contrary holds true for dry samples that exhibit ambient levels of water vapour after a few hours of storage. Consequently, the humid standard stability test was restricted to a period of 48 h only.

Table 6.

Comparison of recoveries of volatiles under study for dry and humid test mixtures in Tedlar, Kynar and Flexfilm bags

Recovery from Flexfilm bag [%]
Recovery from Kynar bag [%]
Recovery from Tedlar bag [%]
VOC   2 h   4 h   6 h   24 h   48 h   2 h   4 h   6 h   24 h 48 h   2 h   4 h   6 h   24 h   48 h
n-Butane Humid 105 106 108 115 123 105 99 99 95 91 105 106 68 99 95
Dry 115 130 131 202 214 107 101 110 103 85 100 103 103 101 88
n-Pentane Humid 103 104 105 109 114 105 104 103 89 85 103 104 73 99 100
Dry 103 105 109 128 153 101 100 100 78 77 96 98 99 96 95
n-Hexane Humid 103 104 104 111 117 99 101 88 79 76 103 102 81 98 100
Dry 126 151 188 412 583 103 103 99 80 81 102 98 99 102 97
n-Octane Humid 90 92 93 92 91 101 101 100 90 86 90 92 93 92 91
Dry 99 101 99 101 94 103 98 98 87 73 95 98 98 99 95
n-Decane Humid 61 61 61 60 57 82 89 82 82 66 71 61 60 60 57
Dry 84 101 92 90 79 103 97 97 68 52 100 116 104 114 104
Isobutane Humid 102 100 89 73 78 101 105 100 99 83 100 98 89 71 76
Dry 100 98 96 87 94 102 100 94 78 84 94 88 89 86 76
Pentane, 2-methyl- Humid 80 79 79 92 84 104 104 102 92 89 80 79 70 92 84
Dry 100 91 101 115 124 101 98 100 82 75 100 98 97 100 95
Octane, 4-methyl- Humid 96 98 99 106 104 98 98 95 87 90 96 98 79 106 104
Dry 93 108 105 158 189 104 98 96 84 81 103 98 100 103 110
2-Pentene, (E) Humid 91 91 90 90 85 104 101 97 84 42 91 91 92 90 89
Dry 98 98 98 93 83 101 99 99 60 44 96 99 98 98 96
2-Pentene, (Z) Humid 90 102 92 92 93 106 102 97 88 49 90 100 98 92 93
Dry 99 100 98 93 100 94 105 103 57 53 87 86 88 84 87
1-Hexene Humid 99 100 100 104 109 104 102 99 90 78 99 100 73 76 79
Dry 102 98 109 145 168 101 98 98 72 65 101 98 99 98 97
Isoprene Humid 104 104 105 102 99 104 102 99 86 64 98 99 105 97 94
Dry 100 99 98 93 96 100 97 96 60 50 97 98 99 99 95
Benzene Humid 87 86 85 83 81 98 96 94 86 76 87 86 88 83 81
Dry 96 88 95 92 85 98 95 95 74 63 90 97 96 95 92
Toluene Humid 90 89 89 87 84 97 96 93 94 87 90 89 88 87 84
Dry 96 93 95 86 85 97 92 92 80 64 93 96 96 93 84
p-Xylene Humid 86 88 89 82 79 94 90 89 85 74 86 88 89 82 79
Dry 97 96 94 86 79 98 93 91 66 54 80 95 95 90 80
Acetone Humid 99 100 100 98 99 97 92 89 80 66 99 100 97 98 99
Dry 99 97 97 91 88 92 86 84 61 57 98 96 95 93 81
2-Butanone Humid 96 98 97 99 102 106 94 100 100 76 96 98 84 95 97
Dry 115 129 131 140 151 96 93 95 66 71 101 105 89 85 97
2-Pentanone Humid 91 93 95 85 83 88 83 78 72 43 91 93 95 85 83
Dry 99 91 97 92 84 94 84 84 52 37 100 99 96 96 88
4-Heptanone Humid 87 87 89 64 67 87 87 89 64 67 87 87 89 64 67
Dry 90 100 99 93 86 90 100 99 93 86 90 100 99 93 86
Propanal Humid 97 98 98 97 96 104 104 101 97 97 97 98 93 97 96
Dry 101 102 101 95 89 98 96 96 76 76 100 104 97 95 84
Hexanal Humid 101 95 96 96 109 101 89 95 103 92 100 94 97 96 95
Dry 102 104 102 98 85 116 119 113 85 58 89 100 103 98 90
Octanal Humid 69 92 69 93 76 83 53 50 72 49 81 92 59 93 76
Dry 94 102 101 99 71 105 106 130 86 58 102 98 89 81 73
Furan Humid 90 90 90 88 88 103 103 99 90 77 90 90 93 88 88
Dry 99 97 97 94 88 98 96 95 68 62 99 97 95 97 87
Furan, 2-methyl- Humid 90 91 86 85 81 101 95 93 85 58 90 91 90 85 81
Dry 96 97 95 93 89 99 96 93 60 49 98 96 96 97 92
Furan, 2,5-dimethyl- Humid 93 91 90 79 65 98 89 81 68 23 93 91 92 79 65
Dry 99 96 98 90 83 99 92 94 39 18 98 98 97 95 88
Methyl acetate Humid 90 90 90 86 84 96 93 89 82 68 90 90 92 86 84
Dry 97 97 95 90 88 90 88 86 55 54 97 96 96 93 87
Ethyl acetate Humid 92 92 91 87 86 98 95 93 87 76 92 92 93 87 86
Dry 95 95 95 92 90 96 92 90 64 59 99 97 99 93 89
n-Butyl acetate Humid 87 92 93 73 78 89 85 81 74 55 87 92 91 85 78
Dry 102 103 98 92 94 97 94 91 53 52 98 98 101 95 82
Dimethyl sulfide Humid 92 92 91 86 80 100 101 104 94 84 92 92 96 86 80
Dry 96 94 95 91 91 94 96 94 63 74 98 97 97 97 90
Methyl propyl sulfide Humid 94 95 93 87 81 104 103 98 90 69 94 95 98 87 81
Dry 96 92 97 92 83 98 93 94 85 75 99 96 95 97 92
Thiophene Humid 89 89 88 83 79 100 97 96 86 77 89 89 88 83 79
Dry 95 88 90 85 81 96 93 91 69 63 100 97 96 92 87
Thiophene, 3-methyl- Humid 87 86 85 77 70 94 90 87 79 63 87 86 87 77 70
Dry 94 87 89 78 73 94 88 86 58 49 94 94 94 88 81
(+)-3-Carene Humid 82 82 83 76 69 96 92 90 86 58 82 82 80 76 69
Dry 96 96 91 90 83 99 93 93 72 46 101 96 96 104 98
α-Pinene Humid 91 93 95 85 89 92 105 99 92 68 91 93 87 85 89
Dry 94 97 83 94 90 110 100 102 90 57 101 108 104 100 97
p-Cymene Humid 68 69 69 64 59 88 85 84 79 68 68 69 65 64 59
Dry 90 94 88 86 73 96 85 86 72 58 109 105 104 112 98
d-Limonene Humid 68 72 75 66 63 88 85 81 76 36 68 72 72 66 63
Dry 95 97 91 88 80 102 90 91 48 32 100 95 96 100 90
Eucalyptol Humid 88 84 84 45 57 88 84 84 45 57 88 84 84 45 57
Dry 93 107 101 105 96 93 107 101 105 96 93 107 101 105 96
Ethyl ether Humid 90 90 92 89 87 106 106 102 94 88 90 90 96 89 87
Dry 98 96 97 91 93 100 97 96 68 72 96 98 98 99 96
Acetonitrile Humid 72 63 64 59 42 73 62 56 54 30 72 63 69 59 42
Dry 84 80 77 62 43 69 56 52 26 15 87 81 78 64 50
Dimethyl selenide Humid 90 91 90 80 70 107 99 89 80 25 90 91 94 80 70
Dry 94 99 96 93 88 102 97 94 37 18 92 98 98 96 84

For the majority of compounds the difference between recoveries in dry samples and humid samples was smaller than 10%, which is in good agreement with the results obtained by Groves and Zellers.27 Nevertheless, species in humid samples exhibited usually slightly poorer stability. In general, recovery differences tended to increase with increasing molecular masses of the compounds. For the heaviest species investigated within this study (n-decane, eucalyptol, D-limonene, p-cymene, α-pinene), they amounted to 20–40%, thus significantly reducing the safe storage time. Interestingly, in Flexfilm and Kynar bags the presence of large amounts of water reduced the emission of contaminants. Most probably water condensing and permeating through polymer films forms a kind of barrier protecting samples from background emission of pollutants. The same water layer seems to induce higher losses of less volatile and more soluble species tending to go into the liquid phase. Consequently, a rapid drop in the concentrations of hydrophilic compounds is observed during the first hours of their storage. Amongst the remaining volatiles acetonitrile was especially sensitive to the presence of water with losses of 30% already after 2 hours of storage. The humidity influence is relatively similar for all bag materials tested. To sum up, high humidity is a crucial factor considerably reducing safe storage time of breath constituents. Since the recoveries of compounds heavier than 90 drop significantly during the first hours of storage, it is recommended to analyse breath samples within six hours of storage.

3.5. Reusability test

The applied cleaning protocol was found to be efficient. In Tedlar bags after 24 hours of storage of pure nitrogen only 4 compounds from the tested ones were detected: 2-butanone, 3-methylthiophene, hexanal and p-xylene. However, their levels were below the LOQs of the applied method. The same number of species was found in Kynar bags: acetonitrile, 3-methyl-thiophene, n-octane and octanal. Amongst them acetonitrile exhibited quantified levels spreading around 4 ppb. Cleaning of Flexfilm bags was more difficult. Excluding species known to be released, six artifacts from the test mixture were detected in these bags after storage for one day: acetonitrile, acetone, pyrimidine, n-octane, p-xylene, and octanal. Acetone showed an average concentration of 18 ppb, whereas acetonitrile levels reached a mean value of 6 ppb. Additional cleaning cycles might be necessary to further remove remainings of the previous sample.

4. Conclusions

In general, several valuable pieces of information on the storage of breath gas samples (as well as other samples containing species at the ppb level) in polymer bags can be extracted from the results of this study.

Firstly, the background emission of pollutants is one of the most important factors when selecting the optimal polymer. High contaminants release distorts the original sample composition already during sampling (bag filling). In the context of breath research aiming at VOCs at low ppb or even ppt levels, Tedlar bags with only nine identified contaminants seem to be the best choice. However, two-fold pre-conditioning of bags before usage is highly recommended. On the other hand, Kynar and particularly Flexfilm were found to emit numerous pollutants (mainly hydrocarbons) detectable immediately, or after few hours of storage at ppt to ppb levels. Preconditioning, even when repeated for several times was not efficient in the case of Flexfilm bags; consequently, this type of material is only suitable for studies aiming at much higher levels of VOCs (e.g., at the ppm level). Due to the quite effective cleaning Kynar could be considered as an alternative to Tedlar, however, it must be remembered that even repeated conditioning does not guarantee the reduction of contaminant emission to a safe level. Finally, it must be underlined that within this study due to the chromatographic limitations only C3–C12 contaminants were monitored, thus the emission of heavier pollutants cannot be excluded.

Secondly, in the case of all tested materials the recovery of volatiles strongly depends on the degree of bag filling (i.e., on the polymer surface-to-(sample) volume ratio (SA : V)). The recoveries of the investigated species in bags with low SA : V values (below 100 m−1) were satisfactory up to 7 days of storage. The increase of the SA : V ratio values above 200 decreases the storage time with acceptable recovery (>80%) by a factor of 3–6. This finding is not surprising as the VOC levels in smaller samples (containing smaller masses of species) are more vulnerable to losses related to sorption or permeation. Additionally, samples in bags filled up to 80% of their maximum volume were less affected by the background emission of contaminants. Consequently, if breath samples are to be stored in polymer bags it is strongly recommended to collect a sample volume as large as possible.

Stability comparisons of the analytes under study in the three polymer bags demonstrated the supremacy of Tedlar over remaining films. In the case of a dry test mixture, recoveries from Tedlar bags (when filled up to 80% of nominal volume) were excellent even after one week of storage. Nevertheless, this safe storage time decreases with the increase of the VOC’s molecular mass. For Kynar the storage times for which an acceptable stability of the tested VOCs can be expected are generally shorter. For this type of bag low ppb VOCs should be analyzed within one day of storage. Nevertheless, Kynar is not suitable for storage of some classes of compounds like ketones (poor recovery), or hydrocarbons (high background emission, or polymer dependent decomposition as it may be the case for isoprene). The suitability of Flexfilm bags for storing breath C3–C10 species is very limited. Although the losses of compounds under study in this polymer were lower than in Kynar bags, the levels of pollutants (hydrocarbons, aldehydes, ketones) were especially high, thereby considerably affecting sample integrity.

High humidity also affects the species’ recoveries. For the majority of compounds stabilities in humid air were up to 10% lower than in a dry matrix. Higher losses (20–40%) detectable even shortly after the bags filling were observed for volatiles with molecular mass above 110. Consequently, in order to reduce losses of heavier species it is advised to analyze breath samples within 6 hours after sampling.

All tested polymers can be reused. The applied cleaning protocol was found to be quite efficient for the removal of artifacts from a previous sample. Nonetheless, to provide maximum security the cleaning procedure proposed here should be repeated at least two times.

In the context of reusability one important factor was not investigated within this study, namely the ageing effect of the polymer film. It was demonstrated in our previous paper21 that used polymer bags with a scratched film exhibit poorer recoveries for sulphur compounds; a similar effect is expected for other classes of species. Therefore, an effort must be made to protect the polymer film and control its quality during usage.

Finally, it must be stressed that due to limitations of the analytical method applied, only C3–C10 volatiles were tested within the study. For heavier or for more reactive species problems arising from sample storage can be much more apparent and demand additional studies.

Table 3.

Contaminants emitted by Kynar bags [ppb]. Compounds are ordered with respect to increasing retention time. ʺ—ʺ denotes that the VOC was not detected, whereas ʺ<LOQʺ stands for VOC level below LOQ

New bag sampling time [h]
Pre-conditioned bag sampling time [h]
VOC CAS 0 6 12 24 0 6 24 VOC emission × 10−12 [g × h−1 × cm−2]
Carbonyl sulfide (COS) 463-58-1 Not quantified
Acetone 67-64-1 2.7 41.4 29.4 32.8 11
Carbon disulfide (CS2) 75-15-0 Not quantified
Trimethylsilyl fluoride 420-56-4 Not quantified
Methyl acetate 79-20-9 <LOQ 0.4 0.5   0.14
n-Pentane 109-66-0 0.5 <LOQ   0.08
2-Propenal, 2-methyl- 78-85-3 <LOQ 0.4 0.4   0.12
Propanal, 2-methyl- 78-84-2 <LOQ 1.3 1.6 1.9 <LOQ   0.59
2-Butanone 78-93-3 0.5 2.3 3.6 4.3 <LOQ 0.5 0.7   1.19
Ethyl acetate 141-78-6 <LOQ 1.0 1.4 1.8 <LOQ   0.59
1,2-Dichlorohexafluoropropane 661-97-2 Not quantified
n-Hexane 110-54-3 <LOQ <LOQ 0.4 <LOQ   0.12
Propane, 2-nitro- 79-46-9 Not quantified
2-Pentanone 107-87-9 <LOQ <LOQ <LOQ   0.09
n-Propyl acetate 109-60-4 <LOQ 0.6 0.8 1.0   0.43
Toluene 108-88-3 1.2 5.6 7.4 9.4 0.3 0.4   3.44
Hexanal 66-25-1 1.1 1.1   0.34
1,3-Dioxane, 4,4-dimethyl- 766-15-4 Not quantified
Ethylbenzene 100-41-4 <LOQ 1.0 1.4   0.51
p-Xylene 106-42-3 0.5 1.0 1.5 0.3 0.4   0.48
o-Xylene 95-47-6 <LOQ 0.4 0.5   0.22

Acknowledgements

P.M., J.K., and K.U. gratefully acknowledge support from the Austrian Science Fund (FWF) under grant no. Y330 and P24736-B23. We appreciate funding from the Austrian Federal Ministry for Transport, Innovation and Technology (BMVIT/BMWA, project 836308, KIRAS). We greatly appreciate the generous support of the government of Vorarlberg, Austria.

References

  • 1.Amann A, Corradi M, Mazzone P, Mutti A. Expert Rev Mol Diagn. 2011;11:207–217. doi: 10.1586/erm.10.112. [DOI] [PubMed] [Google Scholar]
  • 2.Amann A, Smith D. Breath Analysis for Clinical Diagnosis and Therapeutic Monitoring. World Scientific; New Jersey: 2005. [Google Scholar]
  • 3.Bajtarevic A, Ager C, Pienz M, Klieber M, Schwarz K, Ligor M, Ligor T, Filipiak W, Denz H, Fiegl M, Hilbe W, et al. BMC Cancer. 2009;9:348. doi: 10.1186/1471-2407-9-348. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Miekisch W, Schubert JK, Noeldge-Schomburg GF. Clin Chim Acta. 2004;347:25–39. doi: 10.1016/j.cccn.2004.04.023. [DOI] [PubMed] [Google Scholar]
  • 5.Phillips M, Altorki N, Austin JH, Cameron RB, Cataneo RN, Kloss R, Maxfield RA, Munawar MI, Pass HI, Rashid A, Rom WN, et al. Clin Chim Acta. 2008;393:76–84. doi: 10.1016/j.cca.2008.02.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Poli D, Carbognani P, Corradi M, Goldoni M, Acampa O, Balbi B, Bianchi L, Rusca M, Mutti A. Respir Res. 2005;6:71. doi: 10.1186/1465-9921-6-71. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.King J, Kupferthaler A, Frauscher B, Hackner H, Unterkofler K, Teschl G, Hinterhuber H, Amann A, Hogl B. Physiol Meas. 2012;33:413–428. doi: 10.1088/0967-3334/33/3/413. [DOI] [PubMed] [Google Scholar]
  • 8.King J, Mochalski P, Unterkofler K, Teschl G, Klieber M, Stein M, Amann A, Baumann M. Biochem Biophys Res Commun. 2012;423:526–530. doi: 10.1016/j.bbrc.2012.05.159. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Koc H, King J, Teschl G, Unterkofler K, Teschl S, Mochalski P, Hinterhuber H, Amann A. J Breath Res. 2011;5:037102. doi: 10.1088/1752-7155/5/3/037102. [DOI] [PubMed] [Google Scholar]
  • 10.King J, Unterkofler K, Teschl G, Teschl S, Koc H, Hinterhuber H, Amann A. J Math Biol. 2011;63:959–999. doi: 10.1007/s00285-010-0398-9. [DOI] [PubMed] [Google Scholar]
  • 11.King J, Koc H, Unterkofler K, Mochalski P, Kupferthaler A, Teschl G, Teschl S, Hinterhuber H, Amann A. J Theor Biol. 2010;267:626–637. doi: 10.1016/j.jtbi.2010.09.028. [DOI] [PubMed] [Google Scholar]
  • 12.King J, Kupferthaler A, Unterkofler K, Koc H, Teschl S, Teschl G, Miekisch W, Schubert J, Hinterhuber H, Amann A. J Breath Res. 2009;3:027006. doi: 10.1088/1752-7155/3/2/027006. [DOI] [PubMed] [Google Scholar]
  • 13.King J, Mochalski P, Kupferthaler A, Unterkofler K, Koc H, Filipiak W, Teschl S, Hinterhuber H, Amann A. Physiol Meas. 2010;31:1169–1184. doi: 10.1088/0967-3334/31/9/008. [DOI] [PubMed] [Google Scholar]
  • 14.King J, Unterkofler K, Teschl G, Teschl S, Mochalski P, Koc H, Hinterhuber H, Amann A. J Breath Res. 2012;6:016005. doi: 10.1088/1752-7155/6/1/016005. [DOI] [PubMed] [Google Scholar]
  • 15.Endre ZH, Pickering JW, Storer MK, Hu WP, Moorhead KT, Allardyce R, McGregor DO, Scotter JM. Physiol Meas. 2011;32:115–130. doi: 10.1088/0967-3334/32/1/008. [DOI] [PubMed] [Google Scholar]
  • 16.Spanel P, Smith D. Mass Spectrom Rev. 2011;30:236–267. doi: 10.1002/mas.20303. [DOI] [PubMed] [Google Scholar]
  • 17.Ligor M, Ligor T, Bajtarevic A, Ager C, Pienz M, Klieber M, Denz H, Fiegl M, Hilbe W, Weiss W, Lukas P, et al. Clin Chem Lab Med. 2009;47:550–560. doi: 10.1515/CCLM.2009.133. [DOI] [PubMed] [Google Scholar]
  • 18.Buszewski B, Ulanowska A, Ligor T, Denderz N, Amann A. Biomed Chromatogr. 2008;23:551–556. doi: 10.1002/bmc.1141. [DOI] [PubMed] [Google Scholar]
  • 19.Beauchamp J, Herbig J, Gutmann R, Hansel A. J Breath Res. 2008;2:046001. doi: 10.1088/1752-7155/2/4/046001. [DOI] [PubMed] [Google Scholar]
  • 20.Steeghs MM, Cristescu SM, Harren FJ. Physiol Meas. 2007;28:73–84. doi: 10.1088/0967-3334/28/1/007. [DOI] [PubMed] [Google Scholar]
  • 21.Mochalski P, Wzorek B, Sliwka I, Amann A. J Chromatogr B: Anal Technol Biomed Life Sci. 2009;877:189–196. doi: 10.1016/j.jchromb.2008.12.003. [DOI] [PubMed] [Google Scholar]
  • 22.Ligor T, Ligor M, Amann A, Ager C, Bachler M, Dzien A, Buszewski B. J Breath Res. 2008;2:046006. doi: 10.1088/1752-7155/2/4/046006. [DOI] [PubMed] [Google Scholar]
  • 23.Kushch I, Arendacka B, Stolc S, Mochalski P, Filipiak W, Schwarz K, Schwentner L, Schmid A, Dzien A, Lechleitner M, Witkovsky V, et al. Clin Chem Lab Med. 2008;46:1011–1018. doi: 10.1515/CCLM.2008.181. [DOI] [PubMed] [Google Scholar]
  • 24.Schwarz K, Pizzini A, Arendacka B, Zerlauth K, Filipiak W, Schmid A, Dzien A, Neuner S, Lechleitner M, Scholl-Burgi S, Miekisch W, et al. J Breath Res. 2009;3:027003. doi: 10.1088/1752-7155/3/2/027003. [DOI] [PubMed] [Google Scholar]
  • 25.Schwarz K, Filipiak W, Amann A. J Breath Res. 2009;3:027002. doi: 10.1088/1752-7155/3/2/027002. [DOI] [PubMed] [Google Scholar]
  • 26.Erhart S, Amann A, Haberlandt E, Edlinger G, Schmid A, Filipiak W, Schwarz K, Mochalski P, Rostasy K, Karall D, Scholl-Burgi S. J Breath Res. 2009;3:016004. doi: 10.1088/1752-7155/3/1/016004. [DOI] [PubMed] [Google Scholar]
  • 27.Groves WA, Zellers ET. Am Ind Hyg Assoc J. 1996;57:257–263. doi: 10.1080/15428119691014981. [DOI] [PubMed] [Google Scholar]
  • 28.Gilchrist FJ, Razavi C, Webb AK, Jones AM, Spanel P, Smith D, Lenney W. J Breath Res. 2012;6:036004. doi: 10.1088/1752-7155/6/3/036004. [DOI] [PubMed] [Google Scholar]
  • 29.Pawliszyn J. Solid Phase Microextraction: Theory and Practice. Wiley-VCH, Inc; New York: 1997. [Google Scholar]
  • 30.McFadden ER, Jr, Pichurko BM, Bowman HF, Ingenito E, Burns S, Dowling N, Solway J. J Appl Physiol. 1985;58:564–570. doi: 10.1152/jappl.1985.58.2.564. [DOI] [PubMed] [Google Scholar]
  • 31.Cariou S, Guillot JM. Anal Bioanal Chem. 2006;384:468–474. doi: 10.1007/s00216-005-0177-4. [DOI] [PubMed] [Google Scholar]
  • 32.Beghi S, Guillot JM. J Chromatogr A. 2006;1127:1–5. doi: 10.1016/j.chroma.2006.05.102. [DOI] [PubMed] [Google Scholar]
  • 33.Huber W. Accredit Qual Assur. 2003;8:213–217. [Google Scholar]
  • 34.Trabue SL, Anhalt JC, Zahn JA. J Environ Qual. 2006;35:1668–1677. doi: 10.2134/jeq2005.0370. [DOI] [PubMed] [Google Scholar]
  • 35.Dibble TS. J Phys Chem A. 1999;103:8559–8565. [Google Scholar]

RESOURCES