Abstract
Colorimetric enzyme-linked immunosorbent assay utilizing 3′-3-5′-5-tetramethylbenzidine(TMB) as the chromogenic substrate has been widely used in the hospital for the detection of all kinds of disease biomarkers. Herein, we demonstrate a strategy to change this single-color display into dual-color responses to improve the accuracy of visual inspection. Our investigation firstly reveals that oxidation state of 3′-3-5′-5-tetramethylbenzidine (TMB2+) can quantitatively etch gold nanoparticles. Therefore, the incorporation of gold nanoparticles into a commercial TMB-based ELISA kit could generate dual-color responses: the solution color varied gradually from wine red (absorption peak located at ~530 nm) to colorless, and then from colorless to yellow (absorption peak located at ~450 nm) with the increase amount of targets. These dual-color responses effectively improved the sensitivity as well as the accuracy of visual inspection. For example, the proposed dual-color plasmonic ELISA is demonstrated for the detection of prostate-specific antigen (PSA) in human serum with a visual limit of detection (LOD) as low as 0.0093 ng/mL.
Colorimetric enzyme-linked immunosorbent assay (ELISA) is one of the most popular techniques for clinical diagnosis of diverse diseases1,2,3. Most of commercially available colorimetric ELISAs employ horseradish peroxidase (HRP) as the enzyme label and 3,3′,5,5′-tetramethylbenzidine (TMB) as the chromogenic substrate4,5. In the presence of targets, HRP catalyzes the oxidation of TMB (colorless) into TMB2+ (yellow)6, and the concentration of targets is quantified on the basis of the optical density of the yellow solution7,8. Naked-eye inspection of the targets is inaccurate because human eyes are insensitive to optical density variations of the same color. Normally, the optical density of colorimetric ELISA is detected with a sophisticated readout (e.g. an automatic microplate reader) in the hospital9,10. However, in some cases, the use of a costly and bulky readout is not applicable. For example, portable and affordable devices are greatly welcome for in-home personal healthcare11,12,13,14,15,16. Therefore, the development of colorimetric ELISA with improved accuracy for naked-eye inspection could be greatly desired.
In the past two decades, the unique optical properties of noble metal nanoparticles have attracted intense attention to both the academic and industrial communities17,18,19,20,21,22. For instance, Stevens and coworkers incorporated the optical responses of noble metal nanoparticles into conventional ELISA to develop a novel immunoassay strategy called plasmonic ELISA14,23,24,25. This new ELISA method was based on enzyme-controlled growth of gold nanoparticles (AuNPs). In the absence of analyte, the reduction of gold ions was guided to form quasi-spherical, non-aggregated gold nanoparticles, and the solution color was red; in the presence of analyte, the reduction of gold ions was guided to form an ill-defined morphology comprising aggregated nanoparticles, and the solution color was blue. The color transformation from red to blue was easily distinguishable with the naked eyes24. Thus the proposed plasmonic ELISA showed an ultralow limit of detection for prostate specific antigen (PSA) and HIV-1 capsid antigen p24 (1 × 10−18 g/mL). Jiang and coworkers26 developed a strategy to conduct plasmonic ELISA based on click chemistry induced formation of AuNP aggregates. This work can well accommodate conventional immunoassays by employing alkaline phosphatase (ALP) as the enzyme label. Note that all above mentioned plasmonic ELISA methods present color variations from red to blue in response with different concentrations of targets. The red and blue colors are easily to distinguish with the naked eyes; however, the discrimination of colors in-between red and blue are still a big challenge with the naked eyes. In addition, color variations between red and blue result in continuous changes of the absorption spectral peaks, which is inconvenient to be monitored with a microplate reader for the quantification of the analytes.
In this work, we demonstrate another alternative to conduct plasmonic ELISA based on enzyme mediated etching of gold nanoparticles. Our approach displays two distinct colors in responses of different concentrations of analytes, namely, the solution gradually changed from wine red to colorless, and then from colorless to yellow with an increase amount of analytes. Therefore, during the whole detection process, the absorption spectra of the solution only shows two peaks; one located at ~530 nm, which corresponds to the wine red solution; the other located at ~450 nm, which corresponds to the yellow solution. These two distinct absorption peaks are easily to be detected with the microplate reader. Comparing with the conventional ELISA, this dual color plasmonic ELISA can effectively improve the accuracy of visual inspection.
Results and Discussion
Principle of the dual-color ELISA
Figure 1 depicts the principle of the dual-color ELISA. Generally, our detection scheme can be divided into two steps. The first step was a standard ELISA process, which was directly adopted from a commercial available ELISA Kit. The target molecule (antigen) was immobilized on the microplate via antibody-antigen interaction; then a second antibody conjugated with HRP was immobilized onto the antigen. After the addition of enzymatic reaction substrate containing peroxide and TMB, HRP would catalyze the oxidation of the colorless TMB into blue TMB+. The blue TMB+ would turn into yellow TMB2+ after the addition of a stop solution containing hydrochloric acid. The concentration of TMB2+ was proportional to the concentration of antigen in the assay samples, which was the basis of the commercial ELISA for quantitative determination of the concentration of proteins. The second step was the colorimetric reaction process, and this step was the main difference between our method and the conventional ELISA. An appropriate amount of CTAB-capped AuNPs was added to the yellow solution containing TMB2+ which was produced in the first step. The solution color turned into red after the addition of AuNPs (Fig. 1). The optical density of the red solution decreased with the increase of the analyte concentration, and the solution was then turned into colorless when further increasing the analyte concentration. The colorless solution indicated that there was neither AuNPs nor TMB2+ in the solution. Higher concentration of analytes would turn the solution into yellow color. This yellow solution indicated the presence of excess TMB2+.
Proof and mechanism of TMB2+ induced etching of AuNPs
To show the experimental evidence of TMB2+ induced etching of AuNPs, control studies were performed to investigate the effect of TMB2+, Br−, and CTA+ on etching process of AuNPs. As shown in Fig. 2A, the absorption value (450 or 530 nm) seldom changed when only Br− or CTA+ existed (sample a and b); they showed an orange red color (the tertiary color of red and yellow) which was the same as the control (only AuNPs and TMB2+ existed, sample e). On the contrary, when CTAB or CTA+ and Br− existed (sample c and d), a sharp decrease appeared and the color of the mixture correspondingly changed from red to pink. These results demonstrate that (1) neither the mixture of TMB2+ and Br− nor the mixture of TMB2+ and CTAC can induce significant etching of AuNPs; and (2) either the presence of the mixture of TMB2+ and CTAB or the mixture of TMB2+, CTAC, and Br− can generate significant color variations.
It has been reported that H2O2 can be hydrolyzed and therefore generates some powerful oxidant which can efficiently etch bulk Au27,28,29. However, the TMB2+ solution used in this work was produced by HRP-H2O2-TMB system, namely, the TMB2+ solution may contain H2O2 which may influence the process of TMB2+ etching. In this section, another comparative study was conducted to demonstrate whether H2O2 could influence the etching effect of TMB2+. As presented in Fig. 2A, compared with the blank (AuNPs with water), the mixture of H2O2, CTAB, and AuNPs shows seldom change in color and absorption value (530 nm). The absorption variations of AuNPs with different reaction time were compared between TMB2+ and H2O2 (Fig. 2B). The result shows that the etching time for TMB2+ and H2O2 was almost the same (within 1 min), that is, the residual H2O2 in TMB2+ solution has no influence on TMB2+ etching process. Many studies have reported the etching mechanism by various oxidizing agents in the past years. It has reported that the potential of AuBr2−-CTA+/Au is no more than 0.2 V (versus normal hydrogen electrode, NHE)30. Therefore, in the presence of CTA+ and Br−, many reagents can induce the etching of AuNPs. Zhang et al.31 reported that copper (II) can effectively etch gold nanorods in the presence of CTAB. Zhu et al.32 demonstrated that bromine anions were oxidized by hydrogen peroxide to form tribromide, which would then etch gold nanorods to Au+. Nevertheless, TMB2+ may be a stronger oxidant than hydrogen peroxide and can efficiently etch AuNPs in the presence of CTAB in this work.
The reaction process of TMB2+ and AuNPs was also studied by UV-vis spectrum. The maximum absorption peak of TMB is at 285 nm6. Figure 1C presents that the chromogenic substrate (TMB + H2O2) and TMB2+ solution both exhibited the characteristic absorption peak, and the absorption intensity of pure substrate was nearly three times than that of TMB2+. When TMB2+ mixed with AuNPs, the absorption intensity at 285 nm rose, but the absorption intensity at 450 nm decreased. This phenomenon suggests the TMB generated after mixing TMB2+ with AuNPs. The reaction relationship may follow the equation (1) shown below:
Figure 3A,B display the transmission electron microscopy (TEM) images of CTAB-capped AuNPs in the presence and absence of TMB2+. In the absence of TMB2+, the initial size of CATB-capped AuNPs is ~45 nm, and they are quasi-spherical; the solution color is deep wine red (Fig. 3A). After the addition of TMB2+, the size is ~4 nm (Fig. 3B). Figure 3C displays their corresponding UV-vis spectra, in which a significant decrease was observed at ~530 nm in the presence of TMB2+. These results also indicate that TMB2+ can effectively etch AuNPs.
Optimization of experimental conditions
To obtain the best etching effect, we investigated some experimental conditions. The colorimetric assay is based on the reaction between TMB2+ and AuNPs, so the production of TMB2+ is built on HRP-H2O2-TMB system. Hence, the concentrations of H2O2 and TMB were firstly evaluated. As shown in Fig. 4A,B, Δλ530 increased as the H2O2/TMB concentration increased and tended to stable at 1.2 mM (H2O2) and 0.75 mM (TMB). Thus the optimum concentrations of H2O2 and TMB were 1.2 mM and 0.75 mM, respectively. The CTAB concentration is another key factor for TMB2+ etching process. Figure 4C shows Δλ530 increased with the increment of CTAB concentration and reached a platform at 0.15 M. Therefore, 0.15 M CTAB was used for the following study.
HRP concentration dependent etching of AuNPs
HRP is used as a label to indicate the presence of analyte in conventional ELISA, the concentration of analyte is therefore proportional to the concentration of HRP adsorbed on the substrate via antibody-antigen interaction. Hence firstly we assessed the relationship between the concentration of HRP and the color display of the solution. Figure 5A shows the UV-vis spectra of the system in response to different concentrations of HRP. The UV-vis spectrum of AuNPs initially exhibits a peak at ~530 nm. Upon the addition of HRP, the intensity significantly decreased. When the HRP concentration was 20 mU/mL, no obvious absorption peak was observed in the visible region (e.g. 400–800 nm) and the solution was colorless. The result indicates that no AuNPs exists in the solution. Figure 5B correspondingly displays the color variations from 1 to 12, and the intensity change at 530 nm (ΔA530) increased linearly with the HRP concentration ranging from 4 to 20 mU/mL (Fig. 5C). After the solution turned into colorless, further increase of HRP concentration made the solution turn into yellow. This yellow solution indicates the presence of access TMB2+. In this case, the optical density at 450 nm was used for the quantification of HRP (Fig. 5D), which is similar to conventional ELISA.
Visual detection of PSA in human serum
The feasibility of the proposed method was evaluated for visual detection of PSA in human serum. A commercially available human PSA ELISA kit was adopted for the detection of PSA. All operation procedures were followed by the manual of the kit with the only difference in the color display step. The conventional ELISA was added a stop solution to turn the TMB+ into yellow TMB2+ (Fig. 6A), while we added an additional AuNPs solution into the yellow solution to display the dual colors (Fig. 6B). In the absence of PSA, the solution color was deep wine red, and the solution showed an absorption peak at ~530 nm. After the addition of PSA, the solution colors varied from wine red to colorless in the concentration range from 0.3 to 3 ng/mL; and then it displayed yellow when the concentration was more than 3 ng/mL (Fig. 6C). Compared with the conventional ELISA, this dual-color method shows at least two advantages. First, the conventional ELISA only shows one color (yellow) while our method shows red and yellow in response with different concentrations of PSA. This vivid color display could greatly improve the accuracy of visual inspection. Second, for conventional ELISA, the absorption value of 450 nm exceeded the acceptable range of microplate reader when the PSA concentration was more than 2.5 ng/mL (Fig. 6D); whereas the absorption value significantly decreased when AuNPs were added and then was signaled by the microplate reader. By record the decrease of absorption intensity at the wavelength of ~530 nm or absorption intensity increase at the wavelength of ~450 nm, two working curve which corresponding to the red region or yellow region were obtained (Fig. E,F). Other than conventional ELISA, this dual-color ELISA could detect PSA above 2.5 ng/mL, owing to the reason that part of the TMB2+ was consumed by AuNPs. These results indicate that this dual-color ELISA has a wider detection dynamic range than traditional ELISA. The limit of detection was 0.0093 ng/mL (3σ/slope), which is similar to traditional ELISA.
Conclusion
In summary, we have developed a dual-color plasmonic ELISA for the sensitive detection of disease biomarkers with the naked eyes. Our investigation reveals that TMB2+ can quantitatively etch AuNPs, thus the addition of AuNPs to the solution of conventional HRP-TMB based ELISA can turn the single color (e.g. yellow) into dual color (e.g. red and yellow) display. This vivid color display improves the accuracy of visual inspection with the naked eyes. The dual-color plasmonic ELISA might have a potential promising market for the detection of many disease biomarkers that are presently detected with the conventional ELISA.
Methods
Reagents and materials
Gold chloride solution (HAuCl4) and TMB liquid substrate system were purchased from Aladdin (Shanghai, China). The sodium citrate, ascorbic acid, and sodium bromide were obtained from Fu Chen Chemistry (Tianjin, China), and CTAB was purchased from J & K Chemical Technology (Beijing, China). HRP and human total PSA ELISA kit were purchased from Sigma-Aldrich (USA). All other reagents were used as received. UV-vis absorption spectra were recorded with Tianmei UV 2310II (Shanghai, China) or Multiskan spectrum microplate spectrophotometer (Thermo, USA). Ultrapure water was from Direct-Q3 UV system (Millipore, 18.2 MΩ·cm), and all photos were taken by Canon EOS 600D digital camera.
Preparation of the CTAB capped AuNPs
Au seeds were prepared according to previous method33. Briefly, 100 mL of HAuCl4 aqueous solution (2.5 × 10−4 M) was refluxed in the oil bath at 120 °C with vigorous stirring for 30 min, followed by adding 10 mL of sodium citrate (1%). The mixture solution was continued refluxing for 20 min. Then the gold seed solution was cooled down to the room temperature. CTAB-capped AuNPs were synthesized with a seed mediated method reported previously34. Briefly, CTAB (0.21862 g) was dissolved in HAuCl4 (60 mL, 2.5 × 10−4 M) and the mixture solution was incubated in water bath (30 °C). Then 3 mL of 0.1 M freshly prepared ascorbic acid was added followed by gentle stirring for 2 min. The color of the liquid turned into colorless when the ascorbic acid was added. Finally, 30 mL of Au seed solution was added and the mixtures were kept at 30 °C in a water bath for at least 6 h. The absorption peak of CTAB-capped AuNPs was ~530 nm, and the average diameter was ~45 nm calculated by the UV-vis adsorption35 (the detail can be seen in supporting information 2).
Preparation of TMB2+ stock solution
The TMB2+ stock solution was prepared by HRP-H2O2-TMB system. Firstly, 100 μL of HRP (0.15 U/mL) was added to a test tube, followed by adding H2O2 (1 mL) and TMB (1 mL). The mixture turned into light blue quickly and then blue color deepened as time extended. When the enzymatic reaction was about 10 min, HCl (1 mL, 2 M) was added to the mixture to end the reaction. The color changed to yellow immediately, indicating that the TMB2+ is produced. According to previous literature6, the extinction coefficient of the TMB2+ is 5.9 × 104 M−1 cm−1.
Visual detection of HRP
To verify the effectiveness of this method, we investigated the relationship between HRP concentration and the color. Different concentrations of TMB2+ (150 μL) mixed with 100 μL of CTAB-capped AuNPs (~4.182 nM, the concentration was calculated by the UV-vis adsorption36, and the detail can be seen in supporting information 1). The color turned from red to colorless and then yellow, indicating the concentration of HRP is from low to high. Then we diluted HRP stock solution to different concentration solution. HRP (20 μL) mixed with H2O2 (100 μL) and TMB (100 μL), followed by adding HCl (80 μL, 2 M) after 10 min. The absorption value was measured at 450 nm.
Detection of PSA in human serum
A human total PSA ELISA kit was adopted for the detection of PSA. Firstly, different concentrations of human PSA standard solution (100 μL) were added to well, followed by incubation for 2.5 h at room temperature. Aspirate and wash the wells three times with wash buffer. Then, 100 μL of biotinylated detection antibody was added to each well, followed by incubation for 1 h at room temperature with gentle shaking. After that, add 100 μL of HRP-Streptavidin conjugate to each well and incubate 45 min at room temperature with gentle shaking. Aspirate and wash the wells three times with wash buffer. Next, 100 μL of ELISA colorimetric TMB reagent was added to each well and incubate for 30 min at room temperature with gentle shaking. Then 50 μL of stop solution was added. Finally, 100 μL of AuNPs was added to each well and incubated for 5 min with gentle shaking. The absorption intensity was obtained with a microplate reader, and the photos were took in photo studio.
Additional Information
How to cite this article: Guo, L. et al. Dual-color plasmonic enzyme-linked immunosorbent assay based on enzyme-mediated etching of Au nanoparticles. Sci. Rep. 6, 32755; doi: 10.1038/srep32755 (2016).
Supplementary Material
Acknowledgments
The authors gratefully acknowledge the financial support of the National Natural Science Foundation of China (21277025, 21575027, 21275031, 21375021).
Footnotes
The authors declare no competing financial interests.
Author Contributions S.X., X.M. and L.G. designed the research; S.X. and X.M. performed the experiments; S.X., X.M. and L.G. analyzed the data; Y.L. synthesized the AuNPs; B.Q., Z.L. and G.C. contributed to the reagents/materials/ analysis tools; S.X., X.M. and L.G. wrote the paper; all authors approved the final manuscript.
References
- Berg B. et al. Cellphone-Based Hand-Held Microplate Reader for Point-of-Care Testing of Enzyme-Linked Immunosorbent Assays. Acs Nano 9, 7857–7866, doi: 10.1021/acsnano.5b03203 (2015). [DOI] [PubMed] [Google Scholar]
- Liang J. J. et al. Aggregated Silver Nanoparticles Based Surface-Enhanced Raman Scattering Enzyme-Linked Immunosorbent Assay for Ultrasensitive Detection of Protein Biomarkers and Small Molecules. Anal. Chem. 87, 5790–5796, doi: 10.1021/acs.analchem.5b01011 (2015). [DOI] [PubMed] [Google Scholar]
- Lai W. Q., Tang D. P., Zhuang J. Y., Chen G. N. & Yang H. H. Magnetic Bead-Based Enzyme-Chromogenic Substrate System for Ultrasensitive Colorimetric Immunoassay Accompanying Cascade Reaction for Enzymatic Formation of Squaric Acid-Iron(III) Chelate. Anal. Chem. 86, 5061–5068, doi: 10.1021/ac500738a (2014). [DOI] [PubMed] [Google Scholar]
- Zhang S. Y., Garcia-D’Angeli A., Brennan J. P. & Huo Q. Predicting detection limits of enzyme-linked immunosorbent assay (ELISA) and bioanalytical techniques in general. Analyst 139, 439–445, doi: 10.1039/c3an01835k (2014). [DOI] [PubMed] [Google Scholar]
- Penezina O. et al. Performance of a Redesigned HIV Selectest Enzyme-Linked Immunosorbent Assay Optimized To Minimize Vaccine-Induced Seropositivity in HIV Vaccine Trial Participants. Clin. Vaccine Immunol. 21, 391–398, doi: 10.1128/Cvi.00748-13 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Josephy P. D. E. T. & Mason R. P. The horseradish peroxidase-catalyzed oxidation of 3,5, 3′,5′-tetramethylbenzidine. Free radical and charge-transfer complex intermediates. The Journal of Biological Chemistry 257, 3669–3675 (1982). [PubMed] [Google Scholar]
- Druilhe P., Moreno A., Blanc C., Brasseur P. H. & Jacquier P. A colorimetric in vitro drug sensitivity assay for Plasmodium falciparum based on a highly sensitive double-site lactate dehydrogenase antigen-capture enzyme-linked immunosorbent assay. Am. J. Trop. Med. Hyg. 64, 233–241 (2001). [DOI] [PubMed] [Google Scholar]
- Cho I. H., Paek E. H., Kim Y. K., Kim J. H. & Paek S. H. Chemiluminometric enzyme-linked immunosorbent assays (ELISA)-on-a-chip biosensor based on cross-flow chromatography. Anal. Chim. Acta 632, 247–255, doi: 10.1016/j.aca.2008.11.019 (2009). [DOI] [PubMed] [Google Scholar]
- Gao Z. Q., Hou L., Xu M. D. & Tang D. P. Enhanced Colorimetric Immunoassay Accompanying with Enzyme Cascade Amplification Strategy for Ultrasensitive Detection of Low-Abundance Protein. Sci. Rep. 4, doi: 10.1038/Srep03966 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhang P. H. et al. A Highly Sensitive Europium Nanoparticle-Based Immunoassay for Detection of Influenza A/B Virus Antigen in Clinical Specimens. J. Clin. Microbiol. 52, 4385–4387, doi: 10.1128/Jcm.02635-14 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhu Z. et al. Au@Pt Nanoparticle Encapsulated Target-Responsive Hydrogel with Volumetric Bar-Chart Chip Readout for Quantitative Point-of-Care Testing. Angew. Chem. Int. Ed. 53, 12503–12507, doi: 10.1002/anie.201405995 (2014). [DOI] [PubMed] [Google Scholar]
- Zhu Z. et al. Translating Molecular Recognition into a Pressure Signal to enable Rapid, Sensitive, and Portable Biomedical Analysis. Angew. Chem. Int. Ed. 54, 10448–10453 (2015). [DOI] [PubMed] [Google Scholar]
- Tokel O., Inci F. & Demirci U. Advances in Plasmonic Technologies for Point of Care Applications. Chem. Rev. 114, 5728–5752, doi: 10.1021/cr4000623 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Song Y. J. et al. Point-of-care technologies for molecular diagnostics using a drop of blood. Trends Biotechnol. 32, 132–139, doi: 10.1016/j.tibtech.2014.01.003 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jang J. S. et al. Thin-walled SnO2 nanotubes functionalized with Pt and Au catalysts via the protein templating route and their selective detection of acetone and hydrogen sulfide molecules. Nanoscale 7, 16417–16426, doi: 10.1039/c5nr04487a (2015). [DOI] [PubMed] [Google Scholar]
- Chen Y. F. et al. Optofluidic opportunities in global health, food, water and energy. Nanoscale 4, 4839–4857, doi: 10.1039/c2nr30859b (2012). [DOI] [PubMed] [Google Scholar]
- O’Brien M. N., Jones M. R., Brown K. A. & Mirkin C. A. Universal Noble Metal Nanoparticle Seeds Realized Through Iterative Reductive Growth and Oxidative Dissolution Reactions. J. Am. Chem. Soc. 136, 7603–7606, doi: 10.1021/ja503509k (2014). [DOI] [PubMed] [Google Scholar]
- Anker J. N. et al. Biosensing with plasmonic nanosensors. Nat. Mater. 7, 442–453, doi: 10.1038/nmat2162 (2008). [DOI] [PubMed] [Google Scholar]
- Arvizo R. R. et al. Intrinsic therapeutic applications of noble metal nanoparticles: past, present and future. Chem. Soc. Rev. 41, 2943–2970, doi: 10.1039/c2cs15355f (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guo L. H. et al. Distance-Mediated Plasmonic Dimers for Reusable Colorimetric Switches: A Measurable Peak Shift of More than 60 nm. Small 9, 234–240, doi: 10.1002/smll.201201061 (2013). [DOI] [PubMed] [Google Scholar]
- Guo L. H., Ferhan A. R., Lee K. & Kim D. H. Nanoarray-Based Biomolecular Detection Using Individual Au Nanoparticles with Minimized Localized Surface Plasmon Resonance Variations. Anal. Chem. 83, 2605–2612, doi: 10.1021/ac200432c (2011). [DOI] [PubMed] [Google Scholar]
- Guo L. H., Xu Y., Ferhan A. R., Chen G. N. & Kim D. H. Oriented Gold Nanoparticle Aggregation for Colorimetric Sensors with Surprisingly High Analytical Figures of Merit. J. Am. Chem. Soc. 135, 12338–12345, doi: 10.1021/ja405371g (2013). [DOI] [PubMed] [Google Scholar]
- Howes P. D., Rana S. & Stevens M. M. Plasmonic nanomaterials for biodiagnostics. Chem. Soc. Rev. 43, 3835–3853, doi: 10.1039/c3cs60346f (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- de la Rica R. & Stevens M. M. Plasmonic ELISA for the ultrasensitive detection of disease biomarkers with the naked eye. Nat. Nanotechnol. 7, 821–824, doi: 10.1038/Nnano.2012.186 (2012). [DOI] [PubMed] [Google Scholar]
- de la Rica R. & Stevens M. M. Plasmonic ELISA for the detection of analytes at ultralow concentrations with the naked eye. Nat. Protoc. 8, 1759–1764, doi: 10.1038/nprot.2013.085 (2013). [DOI] [PubMed] [Google Scholar]
- Xianyu Y. L., Wang Z. & Jiang X. Y. A Plasmonic Nanosensor for Immunoassay via Enzyme-Triggered Click Chemistry. Acs Nano 8, 12741–12747, doi: 10.1021/nn505857g (2014). [DOI] [PubMed] [Google Scholar]
- Liu X. et al. A plasmonic blood glucose monitor based on enzymatic etching of gold nanorods. Chem. Commun. 49, 1856–1858 (2013). [DOI] [PubMed] [Google Scholar]
- Zhang Z. & Chen Z., Pan, D. & Chen, L. Fenton-like Reaction-Mediated Etching of Gold Nanorods for Visual Detection of Co2+. Langmuir 31, 643–650 (2014). [DOI] [PubMed] [Google Scholar]
- Takahashi Y., Miyahara N. & Yamada S. Gold nanorods embedded in titanium oxide film for sensing applications. Anal. Sci. 29, 101–105 (2013). [DOI] [PubMed] [Google Scholar]
- Rodriguez-Fernandez J., Perez-Juste J., Mulvaney P. & Liz-Marzan L. M. Spatially-directed oxidation of gold nanoparticles by Au(III)-CTAB complexes. J. Phys. Chem. B 109, 14257–14261 (2005). [DOI] [PubMed] [Google Scholar]
- Zhang Z. Y., Chen Z. P., Qu C. L. & Chen L. X. Highly Sensitive Visual Detection of Copper Ions Based on the Shape-Dependent LSPR Spectroscopy of Gold Nanorods. Langmuir 30, 3625–3630, doi: 10.1021/la500106a (2014). [DOI] [PubMed] [Google Scholar]
- Zhu Q., Wu J., Zhao J. & Ni W. Role of bromide in hydrogen peroxide oxidation of CTAB-stabilized gold nanorods in aqueous solutions. Langmuir 31, 4072–4077, doi: 10.1021/acs.langmuir.5b00137 (2015). [DOI] [PubMed] [Google Scholar]
- FRENS G. Controlled Nucleation for the Regulation of the Particle Size in Monodisperse Gold Suspensions. Nature Phys. Sci. 241, 20–22 (1973). [Google Scholar]
- Huang Y. J. & Kim D. H. Synthesis and Self-Assembly of Highly Monodispersed Quasispherical Gold Nanoparticles. Langmuir 27, 13861–13867, doi: 10.1021/la203143k (2011). [DOI] [PubMed] [Google Scholar]
- Khlebtsov N. G. Determination of size and concentration of gold nanoparticles from extinction spectra. Anal. Chem. 80, 6620–6625, doi: 10.1021/ac800834n (2008). [DOI] [PubMed] [Google Scholar]
- Haiss W., Thanh N. T., Aveyard J. & Fernig D. G. Determination of size and concentration of gold nanoparticles from UV-vis spectra. Anal. Chem. 79, 4215–4221 (2007). [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.