Skip to main content
The British Journal of Ophthalmology logoLink to The British Journal of Ophthalmology
. 1991 Jan;75(1):42–44. doi: 10.1136/bjo.75.1.42

Clinical evaluation of automated refraction in anterior chamber pseudophakia.

P Sunder Raj 1, J R Villada 1, K Myint 1, A E Lewis 1, T Akingbehin 1
PMCID: PMC504105  PMID: 1991086

Abstract

Automated refraction with the Canon RK-1 Autoref keratometer was evaluated in 85 eyes of 85 patients with an anterior chamber intraocular lens implant and a best corrected visual acuity of at least 6/12. Autorefraction readings were obtained in 80 (94%) of these eyes. The agreement between autorefraction and clinical refraction data was 89% for spherical equivalence less than 0.51 dioptres (D), 91% for sphere power less than 0.51 D, 82% for cylinder power less than 0.51 D, and 91% for cylinder axis less than 11 degrees. Autorefraction provides acceptably accurate postoperative refraction values in anterior chamber pseudophakia.

Full text

PDF
44

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. Dang M. S., Raj P. P. SRK II formula in the calculation of intraocular lens power. Br J Ophthalmol. 1989 Oct;73(10):823–826. doi: 10.1136/bjo.73.10.823. [DOI] [PMC free article] [PubMed] [Google Scholar]
  2. Dufier J. L., Abitbol M., Pigamo F., Prete T., Paris J. P., Poitrenaud O. Evaluation des performances d'un réfractomètre objectif automatique sur 3,618 yeux. J Fr Ophtalmol. 1987;10(4):301–308. [PubMed] [Google Scholar]
  3. Ghose S., Nayak B. K., Singh J. P. Critical evaluation of the NR-1000F Auto Refractometer. Br J Ophthalmol. 1986 Mar;70(3):221–226. doi: 10.1136/bjo.70.3.221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Guillon M. Automated refraction in aphakia--II. Its repeatability and accuracy compared to conventional techniques. Ophthalmic Physiol Opt. 1986;6(1):85–89. doi: 10.1016/0275-5408(86)90123-7. [DOI] [PubMed] [Google Scholar]
  5. Guyton D. L. Automated refractors-1983. Ophthalmology. 1983 Sep;Suppl:36–44. doi: 10.1016/s0161-6420(83)34441-9. [DOI] [PubMed] [Google Scholar]
  6. Hunold W., Auffarth G., Effert R. Klinische Anwendung des Canon Autorefraktometers R 10 zur Refraktionsbestimmung bei Schielkindern. Klin Monbl Augenheilkd. 1988 Jan;192(1):58–65. doi: 10.1055/s-2008-1050075. [DOI] [PubMed] [Google Scholar]
  7. Infra-red autorefractors--here to stay? Ophthalmic Physiol Opt. 1985;5(3):237–239. [PubMed] [Google Scholar]
  8. Klein B. E., Davis M. D., Segal P., Long J. A., Harris W. A., Haug G. A., Magli Y. L., Syrjala S. Diabetic retinopathy. Assessment of severity and progression. Ophthalmology. 1984 Jan;91(1):10–17. doi: 10.1016/s0161-6420(84)34374-3. [DOI] [PubMed] [Google Scholar]
  9. Nakazawa M., Ohtsuki K. Apparent accommodation in pseudophakic eyes after implantation of posterior chamber intraocular lenses. Am J Ophthalmol. 1983 Oct;96(4):435–438. doi: 10.1016/s0002-9394(14)77905-x. [DOI] [PubMed] [Google Scholar]
  10. Nayak B. K., Ghose S., Singh J. P. A comparison of cycloplegic and manifest refractions on the NR-1000F (an objective Auto Refractometer). Br J Ophthalmol. 1987 Jan;71(1):73–75. doi: 10.1136/bjo.71.1.73. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Papers presented at the meeting of the Society for Experimental Optometry. Birmingham, 24-25 July 1989. Abstracts. Ophthalmic Physiol Opt. 1990 Jan;10(1):97–101. [PubMed] [Google Scholar]
  12. Reddy N. S., Agarwal S. Clinical evaluation of Topcon auto refractometer. Indian J Ophthalmol. 1987;35(5-6):407–410. [PubMed] [Google Scholar]
  13. Wesemann W., Rassow B. Automatic infrared refractors--a comparative study. Am J Optom Physiol Opt. 1987 Aug;64(8):627–638. doi: 10.1097/00006324-198708000-00011. [DOI] [PubMed] [Google Scholar]
  14. Wong E. K., Jr, Patella V. M., Pratt M. V., Myers S. W., Gaster R. N., Leopold I. H. Clinical evaluation of the Humphrey automatic refractor. Arch Ophthalmol. 1984 Jun;102(6):870–875. doi: 10.1001/archopht.1984.01040030690020. [DOI] [PubMed] [Google Scholar]
  15. Wood I. C. A review of autorefractors. Eye (Lond) 1987;1(Pt 4):529–535. doi: 10.1038/eye.1987.80. [DOI] [PubMed] [Google Scholar]
  16. Wood I. C., Papas E., Burghardt D., Hardwick G. A clinical evaluation of the Nidek Autorefractor. Ophthalmic Physiol Opt. 1984;4(2):169–178. [PubMed] [Google Scholar]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group

RESOURCES