Abstract
A strategy of suppressing the fast degradation behaviour of Mg-based biomaterials by the introduction of one of Mg degradation products Mg(OH)2 was proposed according to the following degradation mechanism, Mg + 2H2O ⇋ Mg(OH)2 + H2↑. Specifically, Mg(OH)2 submicron particles were mixed into poly (L-lactic acid) (PLLA) to synthesize a composite coating onto hydrofluoric acid-pretreated Mg-Nd-Zn-Zr alloy. The in vitro degradation investigations showed that the addition of Mg(OH)2 particles not only slowed down the corrosion of Mg matrix, but also retarded the formation of gas pockets underneath the polymer coating. Correspondingly, cytocompatibility results exhibited significant improvement of proliferation of endothelial cells, and further insights was gained into the mechanisms how the introduction of Mg(OH)2 particles into PLLA coating affected the magnesium alloy degradation and cytocompatibility. The present study provided a promising surface modification strategy to tailor the degradation behaviour of Mg-based biomaterials.
Magnesium and magnesium alloys have been considered as a promising candidate for the next-generation biodegradable cardiovascular stent as it shows the possibility to degrade or be absorbed once fulfilling the aim of strutting a narrowed arterial vessel1,2. Their application is likely to overcome the limitations of permanent stents in growing young aged, particularly in newborns, considering the issues of growth3. Recently, the PROGRESS-AMS and DREAMS clinical trials sponsored by BIOTRONIK AG have shown promising results4,5,6. However, the too rapid corrosion/degradation of magnesium matrix, consequently resulting in the early loss of structural support way before the recovery of vascular function, has posed a huge challenge for the clinical use of magnesium alloy implants2,6,7. Thus, it is essential for magnesium and magnesium alloys to acquire adequate surface modifications/treatments to enhance corrosion resistance.
Among the varieties of surface modification techniques7,8,9,10,11,12,13,14,15,16, polymer coating presents as a facile and effective method for protecting magnesium alloy from rapid corrosion. An ideal polymer coating for biodegradable magnesium alloy implants is considered to own suitable mechanical properties, excellent biocompatibility and controlled degradation rate, apart from preventing magnesium substrate from too rapid corrosion, thus prolonging its life expectancy. Poly(L-lactic acid) (PLLA) is one of the several biodegradable polymers preferred for biomedical applications, of which the hydrolysis product lactic acid is found naturally in the body10. Several studies have reported that the corrosion rates of PLLA-coated Mg or Mg alloys decreased to certain extent compared with those of uncoated samples in DMEM or SBF, and exhibited enhanced cytocompatibility10,17,18. Nevertheless, the too rapid release of hydrogen from the corroding Mg substrate still often leads to the formation of gas pockets under the coating19, which may cause the coating spalling and peeling off the substrate, resulting in deleterious consequences for clinical use.
Essentially, the degradation process of magnesium alloy is a corrosion reaction of magnesium with water20:
insoluble Mg(OH)2 is formed once the metal reaches direct contact with water, but it is not stable in the presence of anions like Cl−, SO42− or under acidic conditions21:
Considering the above mechanisms, several measures can be taken to impede the dissolution of the magnesium alloy: (1) to control the amount of the aggressive anions in the environment; (2) to increase the content of the reaction products, thus changing the reaction equilibrium and retarding the progress of the forward reaction; (3) providing a slightly alkaline environment to delay the forward reaction. Based on the above analyses (2) and (3), it can be hypothesized that the introduction of some Mg(OH)2 particles, whose chemical composition is identical to the degradation product of Mg matrix, into the PLLA coating as an additive, would be beneficial for enhancing the protection of coating against corrosion reaction.
In this paper, a Mg-Nd-Zn-Zr alloy (noted as JDBM) was used as the substrate material, whose biodegradable behaviours have been studied in previous works in our group22,23,24, showing a nanophasic uniform and slower corrosion behaviour compared to commercial magnesium alloys. Here, a composite PLLA coating mixed with Mg(OH)2 submicron particles was prepared on JDBM covered with a conversion layer generated by the pre-treatment with hydrofluoric acid (noted as HF) to provide short-term protection. To our knowledge, this is the first time reported to prepare a composite coating consisting of PLLA and Mg(OH)2 on magnesium alloy-based implant materials and to achieve a significant protective effect and good biocompatibility with this coating. The influence of the particles on the in vitro protective effect of the polymer coating as well as the corrosion mechanism was investigated and discussed. The effect on the cytocompatibility of the composite coating was quantitatively evaluated with endothelial cell line model.
Results
Physio-chemical characterizations of coatings
Figure 1 depicts the SEM surface morphologies and cross-section view of HF samples23, the PLLA-coated samples (noted as PLLA) and Mg(OH)2/PLLA-coated samples (noted as HM/PLLA). On the HF sample (Fig. 1a and b), the MgF2 layer, with a thickness of ~1.5 μm, was generally flat except for some parallel lines, which are scratches resulting from the polishing process. The white particles distributing uniformly on the surface were second-phase particles mainly composed of Mg12Nd25. PLLA coating displayed uniform and smooth surface morphology as shown in Fig. 1c, while Mg(OH)2 particles distributed discretely in the HM/PLLA coating (white particles in Fig. 1e. The thickness of the both of coatings was about 6~7 μm (Fig. 1d and f), and no statistic difference in the thickness could be observed. As the mean diameter of the Mg(OH)2 particles was 750~850 nm (Supplementary Fig. S1), Mg(OH)2 particles confirmed by SEM-EDS with size of less than 1 μm inside the composite coating (in black circles, Fig. 1d) indicated that the particles distributed in the coating with few apparent aggregates.
The contact angle tests showed that the freshly prepared HF sample was super-hydrophilic with a small contact angle of less than 5°. The PLLA coating exhibited hydrophobicity with a contact angle of 78.7 ± 2.1°, while the addition of Mg(OH)2 particles reduced surface hydrophobicity, as the contact angle of HM/PLLA dropped to 65.2 ± 3.6°.
In vitro degradation tests
Electrochemical results of JDBM samples with various coatings are shown in Fig. 2. The natural corrosion current (Icorr) data calculated using Tafel fitting from the polarization curves (Fig. 2a) of HF, PLLA and HM/PLLA samples are 4.16 ± 0.48 μA/cm2, 1.25 ± 0.08 μA/cm2 and 0.52 ± 0.12 μA/cm2, respectively. The PLLA coatings slowed down the degradation of the magnesium substrate, as Icorr values dropped by ~70%, compared to the results of HF samples. The HM/PLLA coating further lowered the natural corrosion current by ~58%.
The EIS spectra (Fig. 2b and c) illustrated that PLLA and HF/PLLA specimens presented different impedance responses from those of the HF-treated specimens. The method of electronic circuit models in elucidating the corrosion mechanism of coated specimens in details has been used by many researchers9,26,27,28, and in the present work, equivalent circuits were utilized to fit the EIS results. It should be noted that the equivalent circuits are theoretical models resulting from mathematic deduction without considering the practical dispersion effect. However, dispersion effect cannot be ignored in experimental spectra. Therefore, constant phase element CPE components (with impedance defined as Z = 1/[Y (I*ω)n] were used instead of pure capacitances in the equivalent circuits29. The Nyquist plots of the HF samples exhibited two constants while the samples with PLLA or composite coatings displayed three. As a result, different equivalent circuits (Fig. 3) were applied to fit the experimental spectra of different coating samples to obtain more detailed analyses and information of the protective mechanism of the coatings.
Figure 3(a–c) corresponds to the equivalent circuit of HF, PLLA and HM/PLLA samples. Regarding the HM/PLLA samples, as the particles distributed inside the coating to be an integral coating, the electrochemical properties of the composite coating could be characterized using the same electrical element as PLLA coating. Here Rs was the solution resistance; Rf and CPEf were applied to describe the properties of the polymer coatings; CPEb and Rb characterized the capacitance performance and the resistance property of the barrier MgF2 layer, respectively. Due to the existence of defects in the polymer and the MgF2 layer, the solution can infiltrate through the coatings, and once the solution reached and interacted with the substrate, the charge transfer process occurred30 and an electric double-layer formed. The characterizations of the double layer were described by a CPEdl and a Rct in parallel with it. In addition, in the initial stage of immersion, the infiltration process of the electrolyte into the substrate through the polymer coating and the MgF2 layer successively was a slow diffusion reaction due to the relatively compact structure of the coatings. This diffusion characteristics could be described using a W element.
The fitting results are listed in Table 1. The total resistance of HF samples (Rb + Rct) mainly came from the barrier MgF2 layer, while the protective effect of polymer-coated groups was mostly attributed to the resistance of the polymer coatings, and their total resistances (Rf + Rb + Rct) were significantly greater than those of HF samples. It was noteworthy that the addition of Mg(OH)2 imposed a great influence on the electrochemical properties of PLLA coating-the resistance improved from 3.38E4 Ω cm2 to 8.23E4 Ω cm2. Furthermore, the Rct value also increased from 1249 Ω cm2 to 3592 Ω cm2, indicating a retarded charge transfer process, most probably due to the existence of Mg(OH)2 particles.
Table 1. The fitting results of EIS spectra of different coating samples.
Rs (Ω cm2) | Yf (Ω−1 cm−2 sn) | n | Rf (Ω cm2) | Yb (Ω−1 cm−2 sn) | n | Rb (Ω cm2) | Ydl (Ω−1 cm−2 sn) | n | Rct (Ω cm2) | |
---|---|---|---|---|---|---|---|---|---|---|
HF | 94.1 | — | — | — | 1.02E-5 | 0.88 | 8245 | 1.51E-4 | 1 | 1522 |
PLLA | 100.1 | 6.67E-9 | 0.82 | 3.38E4 | 1.45E-6 | 0.71 | 5015 | 1.37E-5 | 0.91 | 1249 |
HM/PLLA | 103.3 | 1.86E-8 | 0.72 | 8.23E4 | 6.94E-7 | 0.73 | 4055 | 7.89E-6 | 0.80 | 3592 |
The results of corrosion rate measured from weight loss during static immersion process is displayed in Fig. 4a, and hydrogen evolution results in Fig. 4b. It can be seen from Fig. 4a that the PLLA coating reduced the corrosion rate of HF samples significantly, from about 0.77 ± 0.01 mg/cm2/day to about 0.47 ± 0.05 mg/cm2/day. The addition of Mg(OH)2 further decreased the weight loss rate to 0.36 ± 0.02 mg/cm2/day. These results were in accordance with those obtained from hydrogen evolution test (Fig. 4b). During the 10 days of immersion, the volume of evolved H2 from polymer-coated specimens, especially the composite coating group, was remarkably less than that of HF samples, further demonstrating the improved corrosion resistance by the composite PLLA/Mg(OH)2 particles coating.
Figure 5 shows the optical images revealing surface morphologies of HF, PLLA and HM/PLLA-coated samples after immersion for 1 day, 4 days and 30 days in cell culture medium (DMEM with 10% FBS), respectively. The surface of the HF samples kept flat during the entire immersion process, unless some white spots observed at day 30. For PLLA-coated specimen, after 1 day of immersion, there were a multitude of small bubbles distributing uniformly on the surface, and after immersion for 4 days, quite a few of the tiny bubbles grew larger and merged to form gas pockets, thus delaminating the coating from the matrix. As a result, at the end of the immersion (30 days), most part of the coating peeled off the magnesium alloy substrate, while in the case of HM/PLLA, the condition was significantly alleviated. At day 1, bubbles were barely observed and the coating appeared intact. After 4 days of immersion, only a few small bubbles began to emerge, with both smaller size and lower quantity than those of PLLA-coated specimens. At day 30, the entire coating still adhered onto the substrate tightly, with only a small region at the edge detached from the substrate.
Direct cell adhesion and proliferation
Direct response of human umbilical vein endothelium cells (ECs) to PLLA and HM/PLLA coatings was evaluated by culturing ECs on various samples for 1 day and 4 days (Fig. 6a), with live cell density quantitatively and statistically analyzed (Fig. 6b). While the highest cell density of the NC group were observed, only a few live ECs adhered onto HF samples, and on samples coated with PLLA and HM/PLLA coatings, the adhesive cell density increased distinctly with no statistic difference between the two groups. When the culture period was extended to 4 days, ECs proliferation was recorded on all the samples, and the proliferation rate differed. Cells of the NC group exhibited the highest proliferation rate, from 120 ± 19 cells/mm2 at day 1 to 864 ± 15 cells/mm2. On HF samples, the cell density increased from 28 ± 9 cells/mm2 to 78 ± 22 cells/mm2, while the value increased significantly from 49 ± 18 cells/mm2 to 238 ± 15 cells/mm2 and from 67 ± 14 cells/mm2 to 415 ± 51 cells/mm2 on substrates with PLLA and HM/PLLA coatings, respectively. It is noteworthy that the proliferation rate on HM/PLLA samples was only slightly lower than that of the NC group, displaying greatly improved cell adhesion and proliferation in comparison with PLLA coating. Furthermore, the live cell ratio (Fig. 6c) on HM/PLLA coating were higher than that on PLLA coating at both day 1 and day 4. Specifically, at day 1, the live cell ratios of PLLA and HM/PLLA were (61.8 ± 12.8)% and (83.5 ± 4.4)%, and at day 4, the ratios were (88.7 ± 0.1)% and (96.5 ± 1.56)%, respectively. These results confirmed the HM/PLLA coating represented significantly more desirable interfacial conditions for the viability and proliferation of ECs, demonstrating superior cytocompatibility with the addition of Mg(OH)2 particles into PLLA coating.
Discussion
Hydrofluoric acid treatment is a frequently used method to improve the corrosion resistance and biocompatibility of magnesium-based implants by in-situ forming MgF2 conversion layers, however, it fails to produce long-term stable coatings in ionic solutions21,31. Therefore, it is usually applied as a pre-treatment method and further surface modifications like polymer deposition coatings are required for long-term application. As a commonly used polymer coating for protecting magnesium and magnesium alloys, PLLA exhibits improved corrosion resistance and biocompatibility, nevertheless, the interaction between the polymer and Mg alloy substrate, as well as the hydrogen gas generated from the degradation process, often undermined the protection effect of the coating. Thus, new strategies to modify the polymer coating are essential to be developed for clinical implant applications.
In this study, Mg(OH)2 submicron particles were added into PLLA to modify polymer coating properties, and the influence of the Mg(OH)2 particles on the corrosion protection and the cytocompatibility were systematically investigated. The electrochemical results (Fig. 2) demonstrated the superiority of the HM/PLLA coating in protecting the Mg substrates. The HM/PLLA-coated samples exhibited the lowest natural corrosion current and the highest total resistance (Table 1, Rf + Rb + Rct for polymer-coated samples, and Rb + Rct for HF-treated ones), indicating the greatest resistance against corrosion among the three different surface modifications. More specifically, compared with PLLA coating, the introduction of Mg(OH)2 particles increased the resistance of the polymer Rf as well as the value of Rct, implying the charge transfer process was impeded and the corrosion reaction occurred at the interface of MgF2/Mg was slower than that of the PLLA-coated substrate. The results of in vitro degradation tests (Fig. 4) confirmed that the addition of Mg(OH)2 particles further reduced the corrosion rate of the HF samples as compared with PLLA coating, as both the corrosion rate measured from weight loss (Fig. 4a) and the volume of H2 evolution (Fig. 4b) of HM/PLLA samples were lower than those of PLLA and HF samples, respectively.
As shown in Fig. 7, a schematic illustration was proposed to depict the influence of the addition of Mg(OH)2 particles on the degradation process. For the PLLA samples, the degradation process was supposed to involve the following steps:
In the initial stage of immersion, the electrolyte penetrated or diffused through the PLLA coating and the MgF2 layer and reached the interface of MgF2/Mg, thus the Mg substrate reacted with water according to the eq. (1).
Due to the existence of Cl−, some of Mg(OH)2 formed at the interface of MgF2/Mg was gradually dissolved, leading to the release of OH− to the vicinity, resulting in local alkaline environment, as shown in (eq. 2).
In addition, though the MgF2 layer is generally considered insoluble in water, it can be gradually dissolved in Cl− containing solution23 according to the following equation:
Mg2+ released from this reaction disrupted the dissolution equilibrium of eq. (2), so the reaction goes on reversely to form Mg(OH)2 in the alkaline vicinity:
Combining the eqs (3) and (4), that is
The eq. (5) indicates conversion of the MgF2 layer to Mg(OH)2.
Due to the barrier layer of the dense and hydrophobic PLLA coating, the diffusion of the water molecules and the ions were retarded, thus the above corrosion reaction rates were considered slower than that without polymer coating. Therefore, the polymer-coated Mg substrate was better protected and degraded slower than HF samples. However, the PLLA coating could also hinder the fast generated H2 from immediate diffusing to the surrounding medium, thus once the generating rate of H2 was remarkably higher than the diffusion rate, the H2 molecules tended to accumulate at the interface of PLLA/MgF2, leading to some scattered gas bubbles as shown in Fig. 7a, and with the growing of H2 volume, the gas bubbles integrated to form big gas pockets, resulting in the shedding or peeling-off of the coating from the substrate as shown in Fig. 7b. This interpreted the peeling-off phenomenon of PLLA coating in Fig. 5.
The addition of 2% Mg(OH)2 particles into PLLA coating (Fig. 7c) decreased the corrosion rate of the alloy as was described above. Compared to the PLLA-coated samples, the main distinction of the degradation of specimens with HM/PLLA coating was that the Mg(OH)2 particles participated in the overall process. Mg(OH)2 were attacked by penetrated Cl− in the electrolyte and could be gradually dissolved according to eq. (2), releasing OH− and Mg2+ to the surrounding environment as shown in Fig. 7d, increasing pH in the vicinity of the substrate. The alkaline microenvironment decelerated the dissolution of the product Mg(OH)2 deposited on the substrate and also that converted from MgF2 layer, thus the forward reaction of the substrate with water (eq. (1)) was also retarded and the Mg-based substrate degraded at a slower rate.
Further research (Supplementary Table S1) showed that the protective effect with Mg(OH)2 particles was content-dependent. Addition of 0.5% Mg(OH)2 particles into the PLLA coating accelerated the corrosion (from 0.47 ± 0.05 to 0.57 ± 0.03 mg/cm2/day), while with the increase of the amount of the Mg(OH)2 particles, the corrosion protection effect of the coating was enhanced gradually. The introduction of Mg(OH)2 particles might slightly destroy the compactness of the polymer coating, which is adverse for the protection of the PLLA coating. Nevertheless, with respect to corrosion resistance, this deterioration effect can be compensated by the retard of the forward corrosion reaction of the substrate with water according to eq. (1). Therefore, it is expected that there exists equilibrium between the breach of coating compactness and the hindrance of forward corrosion reaction, thus the amount of particles added to the coating plays a crucial role.
From the images of surface morphologies after immersion in cell culture medium (Fig. 5), it appeared no significant peelings or gas pockets of the HM/PLLA composite coating. Two important factors are considered to be contributed to this phenomenon: the generation rate of H2 was relatively slower due to the slower substrate corrosion, and the diffusion rate through the composite coating was faster than that through the PLLA coating, as the compactness of the composite coating was somewhat deteriorated by the addition of the inorganic particles. The addition of Mg(OH)2 particles reduced the hydrophobicity and increased the water uptake of the PLLA. The infiltrating water caused swelling of polymer, leading to a number of defects/channels inside the polymer coating due to the expansion of the coating volume during the swelling. Therefore, H2 could diffuse through the composite coating more easily than through the compact PLLA coating. Thus, H2 did not gather to form gas pockets underneath the coating, preventing the coating from peeling off the substrate. No detachment or peeling-off of the coatings from the substrate is a most fundamental requirement in the application of medical implants.
In addition, the introduction of magnesium hydroxide particles further enhanced the cell adhesion and proliferation as compared to the bare PLLA coating. As the cytotoxicity assay with the extracts of samples with polymer or composite coating (shown in Supplementary Fig. S2) showed similar cell viability of grade 1 (>75%), the higher cell proliferation on HM/PLLA coating most probably resulted from more favourable surface properties, in our study, e.g. the flatter surface during degradation, where fewer gas bubbles formed due to the relatively lower corrosion rate as well as higher H2 diffusion rate. On the contrary, with the occurrence of many bubbles under the PLLA coating, some part of the coating delaminated from the substrate, thus it became difficult for the cells to attach and grow on the surface. Furthermore, in the case of the composite coating, the local pH and concentration of Mg2+ would be more close to the physiological condition due to slower Mg degradation with the protection of HM/PLLA coating, which is also likely to benefit cell adhesion and proliferation.
As reported in previous works32,33,34, in the blood vessel healing process, the neighbouring ECs migration or the circulating endothelial cells (CEC) recruiting process are assumed to be beneficial for the fast endothelialisation of the vascular. The composite coating modified the magnesium alloy to be a physio-chemical condition which attracted more ECs and enhanced their proliferation, thus accelerating the regrowth of the endothelial layers. In a word, the addition of Mg(OH)2 particles into the PLLA coating did not only protect the substrate from rapid corrosion, but also provide a much more desirable interfacial environment for adhesion and proliferation of ECs. This study provides a new strategy to further enhance the corrosion resistance and biocompatibility of magnesium alloy for medical applications.
Conclusion
A composite PLLA coating with Mg(OH)2 particles was prepared on Mg alloy substrate, with the Mg(OH)2 particles uniformly distributing in the polymer coating. The in vitro degradation tests showed the protective effect of PLLA was improved with the introduction of Mg(OH)2 particles, as the degradation rate significantly decreased. A model was proposed to understand the degradation process and mechanisms of corrosion protection of the HM/PLLA composite coating. The analyses indicated that the improvement of the protection most probably resulted from Mg(OH)2 particles in the coating retarding the dissolution of the corrosion product Mg(OH)2 on the Mg alloy substrate and impeding the forward corrosion reaction of Mg with water, i.e. the degradation of the matrix. Additionally, the introduction of Mg(OH)2 particles reduced the damaging effect of the bubbling phenomena occurred with bare polymer coating, improving the adhesion between the coating and the magnesium alloy substrate. Furthermore, the HM/PLLA coating, where fewer gas bubbles formed due to slower corrosion rate and higher H2 diffusion rate, presents a more favorable surface for ECs adhesion and proliferation, showing excellent cytocompatibility. The addition of Mg(OH)2 particles into PLLA coating on Mg-based substrate appears to be a promising strategy to modify the biodegaradable magnesium-based medical devices for enhanced corrosion resistance and improved biocompatibility.
Materials and Methods
Sample preparation
The composition and the processing of Mg-Nd-Zn-Zr alloy (JDBM) alloy could be found in ref. 23. Φ12 × 3 mm, Φ15 × 3 mm and Φ19 × 3 mm disks were used in this study for different measurements. All the specimens were polished with SiC paper progressively up to 3000 grits, ultrasonically cleaned in acetone and ethanol for 10 min, respectively, and dried with warm air. Thereafter, the specimens were pre-treated by 40 wt.% hydrofluoric acid for 24 hours at room temperature to form a compact and protective MgF2 layer of ~1.5 μm thickness23 on the surface.
PLLA (-(CHCH3COO)n-) with an weight-average molecular weight of ~100, 000 g/mol was used as the coating material. A PLLA solution was prepared by dissolving the polymer in dichloromethane (analytical grade) at a concentration of 2 wt.%. Prior to use, magnesium hydroxide particles (purchased from Aladdin, China) with an average particle diameter of 750~850 nm (Supplement Fig. S1) were ultrasonically dispersed in DCM for at least 1 hours to obtain uniformly dispersed solution. Then a small amount of the dispersed solution was added to the PLLA solution to obtain a dispersion solution with a Mg(OH)2/PLLA ratio of ~2 wt.%. The PLLA and Mg(OH)2/PLLA coatings with magnesium hydroxide particles were prepared by a dip-coating method, simply dipping the samples into the solutions for 120 s followed by pulling them out at a constant rate to eliminate the residual solvent at room temperature and the dipping process was repeated 3 times. Then the samples were transferred to a vacuum oven for further drying at 30 °C for at least 48 h.
Characterization of the polymer coatings
The morphologies and the thickness of the polymer coatings in this study were observed by a scanning electron microscopy (SEM, Hitachi S4800, Japan) equipped with energy dispersive X-ray spectroscopy (EDS). The cross-sections of the coated samples were prepared by imbedding the samples in PMMA and polishing with SiC paper progressively up to 7000 grits. Prior to SEM observation, the samples were coated with a thin layer of gold for better conductivity with a sputter coater (Hitachi E-1045, Japan).
Static contact angle (CA) test was performed to evaluate the surface wettability of the HF, PLLA and HM/PLLA samples using the sessile drop method with an optical contact angle system (OCA20, Germany). A minimum of 6 values of contact angle were collected per group of samples.
In vitro degradation tests
Electrochemical corrosion measurements, static immersion tests and hydrogen evolution experiments were performed in conventional simulated body fluid (c-SBF)35 at 37 °C, pH 7.40 for evaluating the influence of the introduction of Mg(OH)2 particles on the protective effect of PLLA coating for magnesium matrix.
Electrochemical impedance spectroscopy (EIS) and potentiodynamic polarization curves were measured in the c-SBF with an advanced electrochemical system (Princeton PARSTAT 2273, USA), equipped with a three-electrode cell featuring a Pt counter electrode and a saturated calomel reference electrode (SCE). A surface area of 0.5 cm2 for each sample was exposed to the solution during the tests. After immersion in the SBF at open circuit potential (OCP) for 1 h to reach a steady state, EIS tests were conducted at OCP from 100 mHz up to 105 Hz, with an amplitude of ±10 mV. The impedance date fitting was performed using a Zsimpwin software. Potentiodynamic polarization experiments were carried out from −0.3 V vs. OCP at a scan rate of 1 mV/s, and the measurement ended once the polarization current reached 1 × 10−3 A.
Immersion tests were carried out in accordance with ASTM G31-7236 (the ratio of surface area to solution volume was 1 cm2:30 mL). The initial weight of the samples was recorded prior to immersion. After immersion in c-SBF for 10 days, the samples were cleaned with boiling chromate acid (200 g/L CrO3 + 10 g/L AgNO3) for 1 min to remove the surface corrosion products followed by drying with warm air. The dried samples were weighed and the corrosion rates were calculated according to ASTM G1-03.
The hydrogen evolution test was carried out as depicted in ref. 37. The samples were suspended in beakers containing c-SBF with a funnel placed over the specimens to collect hydrogen from the specimen surface. A burette was mounted over the funnel and filled with solution. In this way, the volume of the evolved hydrogen was measured.
In order to better simulate the in vivo environment containing proteins and some other organic molecules, Φ19 × 3 mm disks were immersed in Dulbecco’s modified eagle medium (DMEM, Gibco, USA) containing 10% fetal bovine serum (FBS, Gibco, USA) in a cell incubator and the ratio of medium volume to sample surface area was 1 mL/1.25 cm3, according to ISO 10993-12. The medium was replaced periodically. After immersion for 1, 4 and 30 days, the coating morphologies were observed with a stereo microscope.
Direct cell adhesion assay
The human umbilical vein endothelium cell line EA.hy926, was purchased from the Cell Bank, Chinese Academy of Sciences. The cells were cultured in DMEM supplemented with 10% FBS in a cell incubator (37 °C, 5% CO2 and 95% humidity). Cells were subcultured every 2–3 days. Disk samples were placed in a 24-well plate and 1 mL of 3 × 104/mL cell suspension was added to each well. Cells seeded in a blank well served as negative control (NC). After 1 d and 4 d of incubation, Live/Dead staining (LIVE/DEAD® Viability kit, Thermo Fisher Scientific, US) was conducted according to the manufacturer’s protocol, and the stained cells were observed by an inverted fluorescence microscopy (IX 71, Olympus). The adhesion assays were carried out in triplicate and repeated at least twice.
Statistical analysis
The results were expressed as the means ± standard deviations. Difference between two groups were tested using the Student’s test. For variance between different groups, one-way ANOVA was performed and statistical significance was defined as p < 0.05.
Additional Information
How to cite this article: Shi, Y.-j. et al. Enhanced corrosion resistance and cytocompatibility of biodegradable Mg alloys by introduction of Mg(OH)2 particles into poly (L-lactic acid) coating. Sci. Rep. 7, 41796; doi: 10.1038/srep41796 (2017).
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Material
Acknowledgments
This work was supported by the Science and Technology Commission of Shanghai Municipality (No. 14DZ1940800), the National Key Technology R&D Program of the Ministry of Science and Technology (2012BAI18B01) and the National Natural Science Foundation of China (No.: 51571143; 51501115).
Footnotes
The authors declare no competing financial interests.
Author Contributions G.-y.Y. and S.-r.G. designed the study. Y.-j.S. and J.P. performed the study and contributed equally to this work. Z.-h Li and J.-l.N. helped prepare the samples. Y.-j. Shi drafted the manuscript. J.P., J.Z., H.Z. and G.-y.Y. revised the manuscript. All authors reviewed the manuscript.
References
- Heublein B. et al. Biocorrosion of magnesium alloys: a new principle in cardiovascular implant technology? Heart 89, 651–6 (2003). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zeng R., Dietzel W., Witte F., Hort N. & Blawert C. Progress and challenge for magnesium alloys as biomaterials. Adv. Eng. Mater. 10, B3–B14 (2008). [Google Scholar]
- Schranz D., Zartner P., Michel-Behnke I. & Akinturk H. Bioabsorbable metal stents for percutaneous treatment of critical recoarctation of the aorta in a newborn. Catheter. Cardiovasc. Interv. 67, 671–3 (2006). [DOI] [PubMed] [Google Scholar]
- Erbel R. et al. Temporary scaffolding of coronary arteries with bioabsorbable magnesium stents: a prospective, non-randomised multicentre trial. The Lancet 369, 1869–75 (2007). [DOI] [PubMed] [Google Scholar]
- Haude M. et al. Safety and performance of the drug-eluting absorbable metal scaffold (DREAMS) in patients with de-novo coronary lesions: 12 month results of the prospective, multicentre, first-in-man BIOSOLVE-I trial. The Lancet 381, 836–44 (2013). [DOI] [PubMed] [Google Scholar]
- Waksman R. et al. Early-and long-term intravascular ultrasound and angiographic findings after bioabsorbable magnesium stent implantation in human coronary arteries. JACC Cardiovasc. Interv. 2, 312–20 (2009). [DOI] [PubMed] [Google Scholar]
- Hornberger H., Virtanen S. & Boccaccini A. R. Biomedical coatings on magnesium alloys – A review. Acta Biomater. 8, 2442–55 (2012). [DOI] [PubMed] [Google Scholar]
- Alabbasi A., Mehjabeen A., Kannan M. B., Ye Q. & Blawert C. Biodegradable polymer for sealing porous PEO layer on pure magnesium: An in vitro degradation study. Appl. Surf. Sci. 301, 463–7 (2014). [Google Scholar]
- Conceicao T. F., Scharnagl N., Blawert C., Dietzel W. & Kainer K. U. Corrosion protection of magnesium alloy AZ31 sheets by spin coating process with poly(ether imide) [PEI]. Corros. Sci. 52, 2066–79 (2010). [Google Scholar]
- Gray‐Munro J. E., Seguin C. & Strong M. Influence of surface modification on the in vitro corrosion rate of magnesium alloy AZ31. J. Biomed. Mater. Res. A 91, 221–30 (2009). [DOI] [PubMed] [Google Scholar]
- Hu R.-G., Zhang S., Bu J.-F., Lin C.-J. & Song G.-L. Recent progress in corrosion protection of magnesium alloys by organic coatings. Prog. Org. Coat. 73, 129–41 (2012). [Google Scholar]
- Li J., Cao P., Zhang X., Zhang S. & He Y. In vitro degradation and cell attachment of a PLGA coated biodegradable Mg–6Zn based alloy. Journal of materials science 45, 6038–45 (2010). [Google Scholar]
- , Wang J. et al. A surface-eroding poly(1,3-trimethylene carbonate) coating for fully biodegradable magnesium-based stent applications: Toward better biofunction, biodegradation and biocompatibility. Acta Biomater. 9, 8678–89 (2013). [DOI] [PubMed] [Google Scholar]
- , Zomorodian A. et al. Corrosion resistance of a composite polymeric coating applied on biodegradable AZ31 magnesium alloy. Acta Biomater. 9, 8660–70 (2013). [DOI] [PubMed] [Google Scholar]
- Supplit R., Koch T. & Schubert U. Evaluation of the anti-corrosive effect of acid pickling and sol–gel coating on magnesium AZ31 alloy. Corros. Sci. 49, 3015–23 (2007). [Google Scholar]
- Barchiche C.-E., Rocca E., Juers C., Hazan J. & Steinmetz J. Corrosion resistance of plasma-anodized AZ91D magnesium alloy by electrochemical methods. Electrochim. Acta 53, 417–25 (2007). [Google Scholar]
- Xu L. & Yamamoto A. In vitro degradation of biodegradable polymer-coated magnesium under cell culture condition. Appl. Surf. Sci. 258, 6353–8 (2012). [Google Scholar]
- Xu L. & Yamamoto A. Characteristics and cytocompatibility of biodegradable polymer film on magnesium by spin coating. Colloids Surf. B. Biointerfaces 93, 67–74 (2012). [DOI] [PubMed] [Google Scholar]
- Chen Y. et al. Interaction between a high purity magnesium surface and PCL and PLA coatings during dynamic degradation. Biomed Mater 6, 025005 (2011). [DOI] [PubMed] [Google Scholar]
- Witte F. et al. Degradable biomaterials based on magnesium corrosion. Curr. Opin. Solid State Mater. Sci. 12, 63–72 (2008). [Google Scholar]
- da Conceicao T. F., Scharnagl N., Blawert C., Dietzel W. & Kainer K. U. Surface modification of magnesium alloy AZ31 by hydrofluoric acid treatment and its effect on the corrosion behaviour. Thin Solid Films 518, 5209–18 (2010). [Google Scholar]
- Mao L. et al. Nanophasic biodegradation enhances the durability and biocompatibility of magnesium alloys for the next-generation vascular stents. Nanoscale 5, 9517–22 (2013). [DOI] [PubMed] [Google Scholar]
- Mao L., Yuan G., Niu J., Zong Y. & Ding W. In vitro degradation behavior and biocompatibility of Mg–Nd–Zn–Zr alloy by hydrofluoric acid treatment. Mater. Sci. Eng., C 33, 242–50 (2013). [DOI] [PubMed] [Google Scholar]
- Zhang J. et al. Influence of proteins and cells on in vitro corrosion of Mg–Nd–Zn–Zr alloy. Corros. Sci. 85, 477–81 (2014). [Google Scholar]
- Zhang X., Yuan G., Niu J., Fu P. & Ding W. Microstructure, mechanical properties, biocorrosion behavior, and cytotoxicity of as-extruded Mg–Nd–Zn–Zr alloy with different extrusion ratios. Journal of the Mechanical Behavior of Biomedical Materials 9, 153–62 (2012). [DOI] [PubMed] [Google Scholar]
- Grundmeier G., Schmidt W. & Stratmann M. Corrosion protection by organic coatings: electrochemical mechanism and novel methods of investigation. Electrochim. Acta 45, 2515–33 (2000). [Google Scholar]
- , da Conceição T. F., Scharnagl N., Dietzel W. & Kainer K. U. Controlled degradation of a magnesium alloy in simulated body fluid using hydrofluoric acid treatment followed by polyacrylonitrile coating. Corros. Sci. 62, 83–9 (2012). [Google Scholar]
- Zhang J.-T., Hu J.-M., Zhang J.-Q. & Cao C.-N. Studies of water transport behavior and impedance models of epoxy-coated metals in NaCl solution by EIS. Prog. Org. Coat. 51, 145–51 (2004). [Google Scholar]
- Song G.-L. & Xu Z. Crystal orientation and electrochemical corrosion of polycrystalline Mg. Corros. Sci. 63, 100–12 (2012). [Google Scholar]
- Duan H., Du K., Yan C. & Wang F. Electrochemical corrosion behavior of composite coatings of sealed MAO film on magnesium alloy AZ91D. Electrochim. Acta 51, 2898–908 (2006). [Google Scholar]
- Zhang J. et al. Influence of fluoride treatment on surface properties, biodegradation and cytocompatibility of Mg–Nd–Zn–Zr alloy. J. Mater. Sci. Mater. Med. 25, 791–9 (2013). [DOI] [PubMed] [Google Scholar]
- Rotmans J. I. et al. In vivo cell seeding with anti-CD34 antibodies successfully accelerates endothelialization but stimulates intimal hyperplasia in porcine arteriovenous expanded polytetrafluoroethylene grafts. Circulation 112, 12–8 (2005). [DOI] [PubMed] [Google Scholar]
- Avci-Adali M., Paul A., Ziemer G. & Wendel H. P. New strategies for in vivo tissue engineering by mimicry of homing factors for self-endothelialisation of blood contacting materials. Biomaterials 29, 3936–45 (2008). [DOI] [PubMed] [Google Scholar]
- Brewster L. P., Bufallino D., Ucuzian A. & Greisler H. P. Growing a living blood vessel: insights for the second hundred years. Biomaterials 28, 5028–32 (2007). [DOI] [PMC free article] [PubMed] [Google Scholar]
- Oyane A. et al. Preparation and assessment of revised simulated body fluids. J. Biomed. Mater. Res. A 65, 188–95 (2003). [DOI] [PubMed] [Google Scholar]
- ASTM G31-72: standard practice for laboratory immersion corrosion testing of metals. Published: American Society for Testing and Materials (Place published, 2004).
- Zong Y. et al. Comparison of biodegradable behaviors of AZ31 and Mg–Nd–Zn–Zr alloys in Hank’s physiological solution. Mater. Sci. Eng., B 177, 395–401 (2012). [Google Scholar]