Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2018 Feb 1.
Published in final edited form as: IEEE Trans Med Imaging. 2016 Oct 4;36(2):467–477. doi: 10.1109/TMI.2016.2614967

Optimizing MR Scan Design for Model-Based T1, T2 Estimation from Steady-State Sequences

Gopal Nataraj 1,, Jon-Fredrik Nielsen 2, Jeffrey A Fessler 3
PMCID: PMC5378699  NIHMSID: NIHMS824206  PMID: 27893386

Abstract

Rapid, reliable quantification of MR relaxation parameters T1 and T2 is desirable for many clinical applications. Steady-state sequences such as Spoiled Gradient-Recalled Echo (SPGR) and Dual-Echo Steady-State (DESS) are fast and well-suited for relaxometry because the signals they produce are quite sensitive to T1 and T2 variation. However, T1, T2 estimation with these sequences typically requires multiple scans with varied sets of acquisition parameters. This paper describes a systematic framework for selecting scan types (e.g., combinations of SPGR and DESS scans) and optimizing their respective parameters (e.g., flip angles and repetition times). The method is based on a Cramér-Rao Bound (CRB)-inspired min-max optimization that finds scan parameter combinations that robustly enable precise object parameter estimation. We apply this technique to optimize combinations of SPGR and DESS scans for T1, T2 relaxometry in white matter (WM) and grey matter (GM) regions of the human brain at 3T field strength. Phantom accuracy experiments show that SPGR/DESS scan combinations are in excellent agreement with reference measurements. Phantom precision experiments show that trends in T1, T2 pooled sample standard deviations reflect CRB-based predictions. In vivo experiments show that in WM and GM, T1 and T2 estimates from a pair of optimized DESS scans exhibit precision (but not necessarily accuracy) comparable to that of optimized combinations of SPGR and DESS scans. To our knowledge, T1 maps from DESS acquisitions alone are new. This example application illustrates that scan optimization may help reveal new parameter mapping techniques from combinations of established pulse sequences.

Index Terms: T1, T2 relaxometry; optimal experimental design; Cramér-Rao Bound; magnetic resonance imaging

I. Introduction

Fast, accurate quantification of spin-lattice and spin-spin relaxation parameters T1 and T2 has been of longstanding interest in MRI. Many researchers have suggested that T1, T2 maps may serve as biomarkers for monitoring the progression of various disorders [1]. Neurological applications include: lesion classification in multiple sclerosis [2]; tumor characterization [3, 4]; and symptom onset prediction in stroke [5, 6]. In addition, T1, T2 have shown promise for detecting hip and knee cartilage degeneration [7, 8] and for assessing cardiac dysfunction due to iron overload [9] or edema [10]. Motivated by this broad interest in T1, T2 mapping, this paper describes a systematic method to guide MR scan design.

Classical pulse sequences such as inversion/saturation recovery (IR/SR) or (single) spin echo (SE) yield relatively simple methods for T1 or T2 estimation, respectively; however, these methods require several scans, each with long repetition time TR, leading to undesirably long acquisitions. Numerous modifications such as the Look-Locker method [11], multi-SE trains [12], or fast k-space trajectories [1315] have been proposed to accelerate T1 [1619] and T2 [2023] relaxometry with these classical sequences. These techniques are more sensitive to model non-idealities [2426], and are still speed-limited by the long TR required for (near)-complete T1 recovery.

Steady-state (SS) pulse sequences [27, 28] permit short TR, and are thus inherently much faster than classical counterparts. SS techniques are well-suited for relaxometry because the signals produced are highly sensitive to T1 and T2 variation. However, short TR times also cause SS signals to be complex functions of both desired and undesired (nuisance) parameters, complicating quantification. Furthermore, some such methods [29, 30] still require scan repetition, though individual scans are now considerably shorter. Despite these difficulties, the potential for rapid scanning with high T1, T2 sensitivity has motivated numerous SS relaxometry studies [2938].

The dual-echo steady-state (DESS) sequence [39] was recently proposed as a promising SS imaging technique for T2 estimation [35]. Because it produces two distinct signals per excitation, the DESS sequence can reduce scan repetition requirements by recording twice as much data per scan. As with other SS methods, the resulting signals [40, 41] are complicated functions of T1, T2, and other parameters. Prior works have isolated T2 dependencies using either algebraic manipulations of the first- and second-echo signals [35, 36] or separate scans to first estimate nuisance parameters [42]. Although DESS concurrently encodes rich T1 and T2 information, these methods have shied away from using DESS for T1 estimation, either through bias-inducing approximations, or noise-propagating sequential estimation, respectively.

Whether it be with DESS, other sequences, or even combinations thereof, it is generally unclear how to best assemble a scan profile (i.e., a collection of scans) for a fixed amount of scan time. Furthermore, for a given scan profile, it is typically not obvious how to best select acquisition parameters (e.g., flip angles, repetition times, etc.) for relaxometry. In this paper, the term scan design refers to the related problems of scan profile selection and scan parameter optimization.

Historically, scan design for relaxometry has predominantly been explored using figures of merit related to estimator precision. In particular, several studies have used the Cramér-Rao Bound (CRB), a statistical measure that bounds the minimum variance of an unbiased estimator. Earlier works have used the CRB and variations to select inversion times for recovery experiments [43, 44], flip angles for spoiled gradient-recalled echo (SPGR) sequences [45], and echo times for SE experiments [46]. More recent studies have considered additional scan design challenges, including scan time constraints [47], multiple latent parameters [48], multiple scan parameter types [49], and latent parameter spatial variation [50, 51].

The aforementioned studies consider scan parameter optimization for profiles consisting of only one pulse sequence. In contrast, this paper introduces a general framework for robust, application-specific scan design for parameter estimation from combinations of pulse sequences. The framework first finds multiple sets of scan parameters that achieve precise estimation within a tight, application-specific range of object parameters (e.g., T1, T2, etc.). The framework then chooses the one scan parameter set most robust to estimator precision degradation over a broader range of object parameters. As a detailed example, we optimize three combinations of DESS and SPGR sequences for T1, T2 mapping. For a fixed total scan time, we find that well-chosen DESS scans alone can be used to estimate both T1 and T2 with precision and robustness comparable to combinations of DESS and SPGR. This example illustrates that, with careful scan profile design, well-established pulse sequences can find use in new estimation problems.

This paper is organized as follows. Section II uses a general signal model to describe a CRB-inspired min-max optimization problem for robust, application-specific scan optimization. Section III adapts the DESS signal models to our framework and optimizes three practical DESS/SPGR combinations to show that, even in the presence of radiofrequency (RF) field inhomogeneity, DESS is a promising option for T1, T2 relaxometry. Section IV describes simulation, phantom, and in vivo experiments and discusses corresponding results. Section V discusses practical challenges and suggests future directions. Section VI summarizes key contributions.

II. A CRB-Inspired Scan Selection Method

A. A Generalized Signal Model

A broad class of MR pulse sequences useful for parameter mapping produce, after reconstruction, a set of noisy images yd(·) that can be described with the following general model:

yd(r)=fd(x(r);ν(r),pd)+εd(r),d=1,,D, (1)

where x(r) ∈ ℂL collects L latent object parameters, at position r; ν(r) ∈ ℂK gathers K known object parameters at r; pd ∈ ℝP denotes a set of P scan parameters chosen for the dth dataset; fd : ℂL×ℂK ×ℝP ↦ ℂ is a (pulse-sequence dependent) function that models the noiseless signal obtained from the dth dataset; and εd~N(0,σd2) is assumed, for sake of simplicity, to be independent, complex Gaussian noise1 [54, 55]. (As a concrete example, for T2(·) mapping from single SE datasets, x(·) could collect spin density and T2(·); ν(·) could collect known off-resonance and RF field inhomogeneities; and pd could assign the dth echo time, chosen to yield image yd(·).)

A scan profile consists of D datasets from a combination of pulse sequences. Let y(r) := [y1(r),, yD(r)] ∈ ℂD collect the noisy signals at r from all datasets, P := [p1,, pD] ∈ ℝP×D gather the corresponding scan parameters, and vector function f : ℂL × ℂK × ℝP×D ↦ ℂD naturally extend scalar function f, where (·) denotes vector transpose. Then the log-likelihood function is (to within a constant independent of x(r)):

lnL(x(r))=-12y(r)-f(x(r);ν(r),P)-1/22, (2)

where covariance matrix :=diag(σ12,,σD2) is diagonal due to the assumption of independence between scans.

Under suitable regularity conditions2, the Fisher information matrix I(x(r); ν(r),P) ∈ ℂL×L [56] is a measure useful for characterizing the precision of unbiased estimates of x(r) from y(r), given ν(r) and P:

I(x(r);ν(r),P):=E([xlnL(x(r))][xlnL(x(r))])=[xf(x(r);ν(r),P)]-1[xf(x(r);ν(r),P)], (3)

where 𝔼(·) denotes expectation; ∇x denotes a row gradient with respect to x at fixed r; and (·) denotes Hermitian transpose. In particular, the matrix Cramér-Rao Bound (CRB) [57] ensures that any unbiased estimator (r) satisfies

cov(x^(r);ν(r),P)_I-1(x(r);ν(r),P), (4)

where for arbitrary, equally sized matrices A and B, matrix inequality AB means AB is positive semi-definite. In the following, we design an optimization problem based on the CRB to guide MR scan design for relaxometry.

B. Min-max Optimization Problem for Scan Design

Following [58], we focus on minimizing a weighted average of the variances in each of the L latent object parameter estimates. A reasonable objective function for overall estimator precision is therefore given by

Ψ(x(r);ν(r),P)=tr(WI-1(x(r);ν(r),P)W), (5)

where W ∈ ℝL×L is a diagonal, application-specific matrix of weights, preselected to control the relative importance of precisely estimating the L latent object parameters; and tr (·) denotes the matrix trace operation. For scan design, we would like to minimize (5) with respect to scan parameters P.

The CRB depends not only on P but also on the spatially varying object parameters x(·) and ν(·). Thus, one cannot perform scan design by “simply” minimizing Ψ over the scan parameters P. Instead, we pose a practical and application-dependent min-max optimization problem for scan design: we seek candidate scan parameters over a search space ℘ that minimize the worst-case, i.e., maximum cost Ψ̃t, as viewed over “tight” parameter ranges 𝒳t and 𝒩t:

PargminPPΨt(P),where (6)
Ψt(P)=maxxXtνNtΨ(x;ν,P). (7)

Here, we select latent object parameter set 𝒳t based on the application and known parameter set 𝒩t based on the spatial variation typically observed in the known parameters ν(·). Min-max approach (6) should ensure good estimation precision over a range of parameter values.

Since Ψ is in general non-convex with respect to P, it may have multiple global minimizers as well as other scan parameters that are nearly global minimizers. To improve robustness to object parameter variations, we form an expanded set of candidate scan parameters by also including scan parameters that yield costs to within a tolerance δ ≪ 1 of the optimum. Mathematically, we define this expanded set of candidate scan parameter combinations (for a given scan profile) as

S:={P:Ψt(P)-Ψt(P)δΨt(P)}. (8)

To select amongst these candidate scan parameters, we employ a robustness criterion: we select the single scan parameter P* that degrades the least when the worst-case cost is viewed over widened object parameter sets 𝒳b ⊇ 𝒳t and 𝒩b ⊇ 𝒩t:

P=argminPSΨb(P),where (9)
Ψb(P)=maxxXbνNbΨ(x;ν,P). (10)

To compare different scan profiles, we select corresponding search spaces ℘ to satisfy acquisition constraints (e.g., total scan time), but otherwise hold optimization parameters W, δ, 𝒳t,b, and 𝒩t,b fixed. Since Ψ is data-independent, we can solve (6) and (9) offline for each scan profile. The result of each profile’s min-max optimization process (9) is a corresponding optimized scan parameter matrix P* that is suitable for the range of latent x and known ν object parameters specified in 𝒳t and 𝒩t, and is robust to variations in those parameters over broader sets 𝒳b and 𝒩b, respectively.

III. Application: Optimizing SS Sequences for T1, T2 Relaxometry in the Brain

This section applies the methods of Section II.B to the problem of scan design for joint T1, T2 estimation from combinations of SS sequences. Section III.A presents a brief overview of the DESS signals, formulating models to permit estimation of as few nuisance parameters as possible. Section III.B details how we use optimization problems (6) and (9) to tailor three combinations of DESS and SPGR scans for precise T1, T2 estimation in white matter (WM) and grey matter (GM) regions of the human brain. Section III.C compares the predicted performance of the three optimized scan profiles.

A. The DESS Signal Model

The DESS sequence interlaces RF excitations with unbalanced spoiler gradients of fixed area [39] to produce two distinct signals per excitation. If the gradient lobe area is carefully chosen to dominate through-voxel field inhomogeneity gradients, yet not introduce significant diffusive effects [5961], the bulk steady-state signal sD (from a voxel centered at position r) immediately before (t ≈ 0) and after (t ≈ 0+) an RF excitation centered at time t = 0 can be written as

sD(r,0-)=-iM0(r)tanα(r)2[1-η(r,TR)]and (11)
sD(r,0+)=+iM0(r)tanα(r)2[1-η(r,TR)ξ(r,TR)];where (12)
η(r,t):=1-E22(r,t)1-E22(r,t)/ξ2(r,t),and (13)
ξ(r,t):=1-E1(r,t)cosα(r)E1(r,t)-cosα(r). (14)

Here, Eu (r, t) := exp (−t/Tu(r)) for u ∈ {1, 2} describes longitudinal or transverse relaxation at time t, respectively; M0(r) ∈ ℂ is proportional to spin density; flip angle α(r) = α0κ(r) is decomposed as a nominal (prescribed) value α0 with spatial variation κ(r) ≈ 1 due to RF field inhomogeneities; TR is repetition time; and i:=-1. Signal models (11) and (12) neglect relaxation and off-resonance effects during each (short) RF pulse3, and assume RF rotations about the î-axis.

We model each voxel’s macroscopic broadening distribution to be Cauchy (ω̄(r), R2(r)), where ω̄(r) denotes median off-resonance frequency and R2(r) is the broadening bandwidth. If we time readout gradients to form echoes symmetrically centered at echo time TE before and after RF excitation, then the noiseless DESS signals are well approximated as

sD(r,-TE)sD(r,0-)E2-1(r,TE)e-(R2(r)+iω¯(r))TE (15)
sD(r,+TE)sD(r,0+)E2(r,TE)e-(R2(r)-iω¯(r))TE. (16)

In (15) and (16), nuisance parameters M0(r), R2(r), κ(r), and ω̄(r) complicate estimation of T1(r) and T2(r). For simplicity, we take κ(r) to be known from a separately acquired RF transmit field B1+(r) scan. To avoid (separate or joint) ω̄(r) estimation, we choose to use magnitude DESS data, at the expense of slight model mismatch (studied with simulations in Section IV.A) in (1) due to Rician noise. By fixing TE across acquisitions and defining

ME(r):=M0(r)E2(r,TE)e-R2(r)TE, (17)

we can rewrite magnitude signals as functions of L = 3 latent and K = 1 known object parameters, collected as x(r) := [ME(r), T1(r), T2(r)] and ν(r) := κ(r), respectively:

f1(x(r);ν(r),p):=sD(r,-TE)=ME(r)tanα(r)2E2-2(r,TE)[1-η(r,TR)] (18)
f2(x(r);ν(r),p):=sD(r,+TE)=ME(r)tanα(r)2[1-η(r,TR)ξ(r,TR)]; (19)

where p := [α0, TR] collects the P = 2 scan parameters we are free to optimize. From here, we use the scan design method described in Section II to select an optimized p corresponding to each DESS scan within a particular scan profile.

B. Scan Design Details

There are numerous candidate scan profiles involving DESS and/or other pulse sequences that may be useful for fast, accurate T1, T2 mapping. In this work, we consider combinations of DESS and SPGR scans [63]. With proper RF phase cycling and gradient spoiling, the SPGR signal sS can be expressed without any explicit T2(r) dependence:

sS(r,TE)=iME(r)sinα(r)(1-E1(r,TR))1-E1(r,TR)cosα(r). (20)

SPGR’s reduced dependence on spatially varying unknowns is reason for its use for T1 mapping [3032] and subsequent T2 mapping from other sequences [29, 42]. In a similar spirit, we examine scan profiles containing SPGR over other SS sequences because we predict that the SPGR sequence’s T2-independence may help estimators disentangle T2 from other unknown sources of DESS signal contrast.

As written in (20), each SPGR scan also leaves p = [α0, TR] as P = 2 scan parameters available to optimize. A given scan profile consisting of CSPGR SPGR and CDESS DESS scans yields D = CSPGR + 2CDESS datasets. We optimize such a scan profile by solving (9) over a dimension-PD = 2(CSPGR + 2CDESS) space of scan parameters.

We select constraints on search space ℘ based on hardware limitations and desired scan profile properties. Since each pair of DESS signals must share the same choice of p, the search space is reduced to A0,SPGRCSPGR×A0,DESSCDESS×TR,SPGRCSPGR×TR,DESSCDESS (superscripts denote Cartesian powers). We assign flip angle ranges 𝒜0,SPGR = 𝒜0,DESS ← [5, 90]° to restrict RF energy deposition. We set feasible TR solution sets 𝒯R,SPGR ← [12.2,+∞)ms and 𝒯R,DESS ← [17.5,+∞)ms based on pulse sequence designs that control for other scan parameters. These control parameters are described in further detail in Section IV, and are held fixed in all subsequent SPGR and DESS experiments. To equitably compare optima from different scan profiles, we require TR := [TR,1,, TR,CSPGR, TR,CSPGR+1,, TR,CSPGR+CDESS] to satisfy a total time constraint, ||TR||1TR,max. For a scan profile consisting of CSPGR SPGR and CDESS DESS scans, these constraints collectively reduce the search space dimension from PD to 2(CSPGR + CDESS) − 1.

Prior works have considered T1 or T2 estimation from as few as 2 SPGR [29, 45] or 1 DESS [35] scan(s), respectively. We likewise elect to optimize the (CSPGR,CDESS) ← (2, 1) scan profile as a benchmark. We choose TR,max ← 2(12.2)+ 1(17.5) = 41.9ms and select other scan profiles capable of meeting this time constraint. Requiring that candidate profiles contain CDESS ≥ 1 DESS scans for T2 contrast and satisfy DL(= 3) for well-conditioned estimation, we note that (1, 1) and (0, 2) are the only other eligible profiles.

In the ensuing experiments, we focus on precise T1, T2 estimation in the brain and design latent object parameter ranges 𝒳t = ℳE,t × 𝒯1,t × 𝒯2,t and 𝒩t = 𝒦t accordingly. Noting that T1 ~ 10T2, we choose W ← diag (0, 0.1, 1) in (5) to place roughly equal importance on precise T1 vs. T2 estimation. Since W places zero weight on ME estimation (obviating the need for complex differentiation in (3)), it is easily shown that Ψ depends on ME only through a constant scale factor; thus it suffices to consider ℳE,t ← 1. We select 𝒯1,t ← [800, 1400]ms and 𝒯2,t ← [50, 120]ms to correspond with WM and GM regions of interest (ROIs) at 3T [64, 65].We take 𝒦t ← [0.9, 1.1] to account for 10% spatial variation in flip angle. Broadened ranges 𝒳b ← 1×[400, 2000]ms×[40, 200]ms and 𝒦b ← [0.5, 2] are constructed to encourage solutions robust to a wide range of object parameters. We assume constant noise variance σ12==σD2:=σ2, where σ2 ← 1.49 × 10−7 is selected to reflect measurements from normalized phantom datasets (cf. Sections IV.B.1 and S.V.A-S. V.B for acquisition details). Lastly, we set δ ← 0.01 to select a robust scan parameter P* with associated worst-case cost Ψ̃t(P*) within 1% of global optimum Ψ̃t().

C. Scan Profile Comparisons

We solve (6) and (9) via grid search to allow illustration (§S.I in Supplement4) of Ψ̃t(P) as well as worst-case T1, T2 standard deviations σT1t(P) and σT2t(P), each defined as

σT1t(P):=maxxXbνNbσT1(x;ν,P)and (21)
σT2t(P):=maxxXbνNbσT2(x;ν,P), (22)

where σT1 (x; ν, P) and σT2 (x; ν, P) are corresponding diagonal elements of inverse Fisher matrix I−1(x; ν, P). Grid searches for the (2, 1), (1, 1), and (0, 2) profiles each took about 4, 43, and 28 minutes, respectively. All experiments described hereafter were carried out using MATLAB® R2013a on a 3.5GHz desktop with 32GB RAM.

Table I compares optimized scan parameters for profiles consisting of (2, 1), (1, 1), and (0, 2) SPGR and DESS scans, respectively. In addition to σT1t(P) and σT2t(P), Table I presents analogous worst-case standard deviations σT1b(P) and σT2b(P) over 𝒳b × 𝒦b to show how each estimator degrades over the broadened object parameter range. When viewed over tight range 𝒳t × 𝒦t, the (0, 2) profile provides a 11.5% reduction in worst-case cost over the other choices. Extending to broadened range 𝒳b×𝒦b, this reduction grows dramatically to 31.4%. We thus observe that while the different optimized profiles afford similar estimator precision over a narrow range of interest, the (0, 2) profile may be preferable due to its robustness to a wide range of object parameters.

TABLE I.

Performance summary of different scan profiles, optimized by solving (9) subject to scan time constraint TR,max = 41.9ms. The first column defines each profile. The next four columns describe P*. The latter three pairs of columns show how worst-case σ̃T1 σ̃T2, and Ψ̃ values degrade from tight to broad ranges. Flip angles are in degrees; all other values are in milliseconds.

Scan
α^0spgr
α^0dess
T^Rspgr
T^Rdees
σT1t(P)
σT1b(P)
σT2t(P)
σT2b(P)
Ψ̃t(P*) Ψ̃b(P*)
(2, 1) (15, 5)° 30° (12.2, 12.2) 17.5 28 154 1.3 9.1 4.0 17.7
(1, 1) 15° 10° 13.9 28.0 27 169 2.8 8.8 4.9 17.9
(0, 2) (35, 10)° (24.4, 17.5) 21 113 1.5 6.0 3.5 12.2

As the DESS sequence has already found success for T2 mapping from even one scan [35], it is reassuring but unsurprising that our analysis finds two DESS scans to yield the most precise T2 estimates. More interestingly, our methods suggest that, with a minimum CDESS = 2 scans, DESS can be used to simultaneously estimate T1 as well. In fact, for certain choices of parameter ranges, a second DESS scan is predicted to afford 1 precision comparable to two SPGR scans.

IV. Experimental Validation and Results

To test our approach to optimized scan design (described in Section II.B), we next estimate T1 and T2 maps (using maximum-likelihood (ML) and regularized least squares (RLS) methods detailed in Section S.II) from datasets collected using the scan profiles optimized in Section III. In Section IV.A, we study estimator statistics from simulated data. In Sections IV.B–IV.C, we progress to phantom and in vivo datasets to evaluate scan profile performance and estimator robustness under increasingly complex settings. For the latter experiments, we use reference parameter maps from classical (long) pulse sequences, in lieu of ground truth maps.

A. Numerical Simulations

We select T1 and T2 WM and GM values based on previously reported measurements at 3T [64, 65] and extrapolate other unimportant latent object parameters M0 and T2 from measurements at 1.5T [66]. We assign these parameter values to the discrete anatomy of the BrainWeb digital phantom [66, 67] to create ground truth M0, T1, T2, and T2 maps. We then choose acquisition parameters based on Table I (with fixed TE = 4.67ms) and apply models (20) and (15)–(16) to the 81st slices of these true maps to compute noiseless 217×181 SPGR and DESS image-domain data, respectively.

For each scan profile, we corrupt the corresponding (complex) noiseless dataset F with additive complex Gaussian noise, whose variance σ2 ← 1.49 × 10−7 is set to match CRB calculations. This yields realistically noisy datasets Y ranging from 105–122 SNR, where SNR is defined here as

SNR(F,Y):=FFY-FF. (23)

We use each profile’s noisy magnitude dataset |Y| to compute estimates E, 1, and 2 (images and histograms in Section S.III). We then evaluate estimator bias and variance from latent ground truth T1 and T2 maps.

In these simulations, we intentionally neglect to model a number of physically realistic effects because their inclusion would complicate study of estimator statistics. First and foremost, we assume knowledge of a uniform transmit field, to avoid confounding B1+ and T1, T2 estimation errors. For a similar reason, spatial variation in the sensitivity of a single receive coil is also not considered. We omit modeling partial volume effects to ensure deterministic knowledge of WM and GM ROIs. We will explore the influence of these (and other) nuisance effects on scan design in later subsections.

To isolate bias due to estimator nonlinearity from regularization bias, we minimize the ML initialization cost (S.1) only, and do not proceed to solve RLS problem (S.2). This permits consideration of T1, T2 estimation from each of the 7733 WM or 9384 GM data points as voxel-wise independent realizations of the same estimation problem. To minimize quantization bias, we optimize (S.1) using a very finely spaced dictionary of signal vectors from 1000 T1 and T2 values logarithmically spaced between [102, 103.5] and [101, 102.5], respectively. Using 106 dictionary elements, solving (S.1) took less than 7 minutes for each tested scan design P*.

Table II5 verifies that, despite model nonlinearity and Rician noise, estimation bias in WM- and GM-like voxels is negligible. Sample standard deviations are consistent with σT1t and σT2t (cf. Table I). In WM and GM, we observe that the (1, 1) and (0, 2) profiles afford high T^1ML precision, while the (2, 1) and (0, 2) scans afford high T^2ML precision. In agreement with the predictions of Ψ̃t and Ψ̃b, these simulation studies suggest that at these SNR levels, an optimized profile containing 2 DESS scans can permit T1 and T2 estimation precision in WM and GM comparable to optimized profiles containing SPGR/DESS combinations.

TABLE II.

Sample means ± sample standard deviations of T1 and T2 ML estimates in WM and GM ROIs of simulated data, compared across different optimized scan profiles. Sample means exhibit insignificant bias, and sample standard deviations are consistent with worst-case standard deviations σT1t and σT2t reported in Table I. All values are reported in milliseconds.

Scan (2, 1) (1, 1) (0, 2) Truth
WM T^1ML 830 ± 17 830 ± 15 830 ± 14 832
GM T^1ML 1330 ± 30. 1330 ± 24 1330 ± 24 1331

WM T^2ML 80. ± 1.0 80. ± 2.1 79.6 ± 0.94 79.6
GM T^2ML 110. ± 1.4 110. ± 3.0 110. ± 1.6 110

B. Phantom Experiments

This subsection describes two experiments. In the first experiment, we compare the SPGR/DESS scan profiles described in Table I (as well as a reference profile consisting of IR and SE scans) against nuclear magnetic resonance (NMR) measurements from the National Institute for Standards and Technology (NIST) [68]. These measurements provide information about ROI sample means and ROI sample standard deviations (Fig. 1), which we define as first- and second-order statistics computed across voxels within an ROI. In the second experiment, we repeat the SPGR/DESS scan profiles 10 times and compute sample standard deviation maps across repetitions (not shown). Taking ROI sample means of these maps gives pooled sample standard deviations (Table III), which indicate relative scan profile precision.

Fig. 1.

Fig. 1

Phantom within-ROI sample statistics of T1 and T2 ML estimates from optimized SPGR/DESS and reference IR/SE scan profiles, vs. NIST NMR measurements [68]. Markers and error bars indicate ROI sample means and ROI sample standard deviations within the 14 labeled and color-coded vials in Fig. S.7. Tight 𝒳t and broad 𝒳b latent parameter ranges are highlighted in orange and yellow, respectively. Fig. S.8 provides analogous plots for RLS estimates. Table S.2 replicates sample statistics within Vials 5–8. Our MR measurements are at 293K, while NIST NMR measurements are at 293.00K. Within the designed parameter ranges, estimates from different acquisitions are in reasonable agreement with NIST measurements.

TABLE III.

Phantom pooled sample standard deviations ± pooled standard errors of sample standard deviations, from optimized SPGR/DESS scan profiles. Each entry is a measure of uncertainty of a typical voxel’s T1 or T2 ML estimate. For sake of brevity, sample statistics corresponding only to phantom vials within (or nearly within) tight design range 𝒳t (color-coded orange in Fig. S.7) are reported. ‘V#’ abbreviates vial numbers. All values are reported in milliseconds.

(2SP,1DE) (1SP,1DE) (0SP,2DE)
V5 σ^T^1ML(P) 50 ± 12 40 ± 10. 39 ± 9.4
V6 σ^T^1ML(P) 70 ± 18 60 ± 15 70 ± 16
V7 σ^T^1ML(P) 60 ± 13 50 ± 13 50 ± 13
V8 σ^T^1ML(P) 23 ± 5.4 20. ± 4.7 18 ± 4.3
(2SP,1DE) (1SP,1DE) (0SP,2DE)
V5 σ^T^2ML(P) 2.6 ± 0.63 6 ± 1.4 3.5 ± 0.84
V6 σ^T^2ML(P) 1.9 ± 0.46 5 ± 1.1 2.3 ± 0.54
V7 σ^T^2ML(P) 1.4 ± 0.34 3.4 ± 0.80 1.5 ± 0.35
V8 σ^T^2ML(P) 1.1 ± 0.26 3.5 ± 0.84 1.4 ± 0.33

1) Within-ROI Statistics

We acquire combinations of (2, 1), (1, 1), and (0, 2) SPGR and DESS coronal scans of a High Precision Devices® MR system phantom T2 array. For each scan profile, we prescribe the optimized flip angles α̂0 and repetition times R listed in Table I, and hold all other scan parameters fixed. We achieve the desired nominal flip angles by scaling a 20mm slab-selective Shinnar-Le Roux excitation [69], of duration 1.28ms and time-bandwidth product 4. For each DESS (SPGR) scan, we apply 2 (10) spoiling phase cycles over a 5mm slice thickness. We acquire all steady-state phantom and in vivo datasets with a 256 × 256 × 8 matrix over a 240 × 240 × 30 mm3 field of view (FOV). Using a 31.25kHz readout bandwidth, we acquire all data at minimum TE ← 4.67ms before or after RF excitations. To avoid slice-profile effects, we sample k-space over a 3D Cartesian grid. After Fourier transform of the raw datasets, only one of the excited image slices is used for subsequent parameter mapping. Including time to reach steady-state, each steady-state scan profile requires 1m37s scan time.

To validate a reference scan profile for use in in vivo experiments, we also collect 4 IR and 4 SE scans. For (phase-sensitive, SE) IR, we hold (TR, TE) ← (1400, 14)ms fixed and vary (adiabatic) inversion time TI ∈ {50, 150, 450, 1350}ms across scans. For SE, we similarly hold TR ← 1000ms fixed and vary echo time TE ∈ {10, 30, 60, 150}ms across scans. We prescribe these scan parameters to acquire 256×256 datasets over the same 240 × 240 × 5 mm3 slice processed from the SPGR/DESS datasets. Each IR and SE scan requires 5m58s and 4m16s, for a total 40m58s scan time.

We additionally collect a pair of Bloch-Siegert shifted 3D SPGR scans for separate B1+ estimation [70]. We insert a 9ms Fermi pulse (peak amplitude B1pk0.075G) at ±8 kHz off-resonance into an SPGR sequence immediately following on-resonant excitation. We estimate regularized B^1+ maps [71] from the resulting pair of datasets. We then estimate flip angle variation κ̂ as B^1+/B1pk, calibrate κ̂ (via separate measurements described in Section S.IV), and thereafter take κ as known. For consistency, we account for flip angle variation when estimating T1 and T2 from both the candidate (SPGR/DESS) and reference (IR/SE) aforementioned scan profiles. With a repetition time of 21.7ms, this B1+ mapping acquisition requires 1m40s total scan time.

Fig. 1 plots sample means and sample standard deviations computed within circular ROIs of phantom T1 and T2 ML estimates (reconstruction details, analogous plots for RLS estimates, and images in Sections S.V.A–S.V.C). The highlighted orange and yellow parameter spaces correspond to design ranges 𝒳t and 𝒳b. T1 estimates from both the candidate (2, 1), (1, 1), and (0, 2) (SPGR, DESS) and reference (4, 4) (IR, SE) profiles are in reasonable agreement with NIST estimates [68] across the vial range. T2 estimates from all profiles are also in good agreement with NIST for vials within 𝒳b. SPGR/DESS profiles likely underestimate large T2 values (≥200ms) due to greater influence of diffusion in DESS [5961]. SPGR/DESS profiles possibly overestimate and the IR/SE profile likely underestimates short (≤30ms) and very short (≤15ms) T2 values, respectively, due to poorly conditioned estimation.

2) Across-Repetition Statistics

In a second study, we repeat the (2, 1), (1, 1), and (0, 2) scan profiles 10 times each and separately estimate T1 and T2 for each repetition of each scan profile. We then estimate the standard deviation across repetitions on a per-voxel basis, to produce sample standard deviation maps for each profile. Each ROI voxel of the sample standard deviation map is a better estimate of the population standard deviation (which the CRB characterizes) than the ROI sample standard deviation from a single repetition, because the latter estimate is contaminated with slight spatial variation of voxel population means (due to imaging non-idealities such as Gibbs ringing due to k-space truncation).

Table III reports pooled sample standard deviations and pooled standard errors of the sample standard deviations (computed via expressions in [72]) for phantom vials within (or nearly within) tight design range 𝒳t (marked orange in Fig. S.7). Due to error propagation from coil combination and κ̂ estimation, pooled ML sample standard deviations cannot be compared in magnitude to worst-case predicted standard deviations (Table I); however, trends of empirical and theoretical standard deviations are overall similar. In particular, the optimized (0, 2) DESS-only scan profile affords T1 ML estimation precision (in vials whose T1, T2 is similar to that of WM/GM) comparable to optimized (2, 1) and (1, 1) mixed (SPGR, DESS) profiles. Also in agreement with predictions, the optimized (2, 1) and (0, 2) profiles afford greater T2 ML estimation precision than the optimized (1, 1) profile.

C. In Vivo Experiments

In a single long study of a healthy volunteer, we acquire the same optimized scan profiles containing (2, 1), (1, 1), and (0, 2) SPGR and DESS scans (cf. Table I), as well as the reference profile containing (4, 4) IR and SE scans. We obtain axial slices from a 32-channel Nova Medical® receive head array. To address bulk motion between acquisitions and to compare within-ROI statistics, we rigidly register each coil-combined image to an IR image (details in Section S.V.D) prior to parameter mapping. All acquisition (cf. Section IV.B.1) and reconstruction (cf. Sections S.V.A–S.V.B) details are otherwise the same as in phantom experiments.

Fig. 2 compares brain T1 and T2 ML estimates from optimized scan profiles (Fig. S.9 and Fig. S.10 provide corresponding colorized and grayscale RLS estimates, respectively). Though in-plane motion is largely compensated via registration, through-plane motion and non-bulk motion likely persist, and will influence ROI statistics. Due to motion (and scan duration) considerations, we examine within-ROI statistics from a single repetition as in Section IV.B.1, and do not attempt across-repetition statistics as in Section IV.B.2.

Fig. 2.

Fig. 2

Left: WM and GM ROIs, overlaid on a representative anatomical (coil-combined IR) image. Separate WM ROIs are distinguished with anterior/posterior (A/P) and right/left (R/L) directions. Four small anterior cortical GM polygons are pooled into a single ROI (cyan). Right: Colorized T1 and T2 ML estimates from the brain of a healthy volunteer. Columns correspond to profiles consisting of (2 SPGR, 1 DESS), (1 SPGR, 1 DESS), (0 SPGR, 2 DESS), and (4 IR, 4 SE) acquisitions. Parameter maps are cropped in post-processing for the purpose of display. Figs. S.9 (colorized) and S.10 (grayscale) provide analogous full-FOV maps estimated via both ML and RLS estimators. Colorbar ranges are in milliseconds.

Visually, 1 maps from steady-state profiles exhibit similar levels of contrast in WM/GM regions well away from cerebrospinal fluid (CSF) as that seen in the reference 1 estimate. Since we did not optimize any scan profiles for estimation in high-T1 regions, it is expected that greater differences may emerge in voxels containing or nearby CSF. In particular, T1 is significantly underestimated within and near CSF by the (0, 2) DESS-only profile. This suggests that with the signal models used in this work, including at least one SPGR scan in an optimized profile may offer greater protection against estimation bias in high-T1 regions.

Table IV summarizes within-ROI sample means and sample standard deviations computed6 over four separate WM ROIs containing 96, 69, 224, and 148 voxels and one pooled cortical GM ROI containing 156 voxels (cf. Fig. 2). Within-ROI 1 sample standard deviations are comparable across steady-state profiles. In agreement with Table I, T2 estimates from the optimized (1, 1) scan profile exhibit higher within-ROI sample variation than corresponding (2, 1) and (0, 2) 2 maps.

TABLE IV.

Within-ROI sample means ± within-ROI sample standard deviations of T1 and T2 ML estimates from the brain of a healthy volunteer. Sample statistics are computed within ROIs indicated in Fig. 2. All values are reported in milliseconds.

ROI (color) (2SP,1DE) (1SP,1DE) (0SP,2DE) (4IR,4SE)
T^1ML
anterior right WM (yellow) 840 ± 32 770 ± 31 840 ± 43 780 ± 22
anterior left WM (magenta) 740 ± 61 660 ± 45 740 ± 55 760 ± 24
posterior right WM (green) 890 ± 88 860 ± 72 960 ± 84 810 ± 26
posterior left WM (blue) 860 ± 70. 850 ± 61 880 ± 79 820 ± 37
anterior GM (cyan) 1200 ± 210 1200 ± 230 1300 ± 230 1300 ± 180

T^2ML
anterior right WM (yellow) 40. ± 1.3 54 ± 3.8 46 ± 1.5 55 ± 1.9
anterior left WM (magenta) 40. ± 1.7 50. ± 4.5 44 ± 1.7 53 ± 1.8
posterior right WM (green) 43 ± 2.7 60. ± 6.9 51 ± 3.6 59 ± 2.1
posterior left WM (blue) 43 ± 1.8 57 ± 4.9 49 ± 2.5 57 ± 1.8
anterior GM (cyan) 50 ± 12 60 ± 15 60 ± 11 59 ± 6.0

In most cases, 1 within-ROI sample means from optimized SPGR/DESS scan profiles do not deviate substantially from each other or from reference IR/SE measurements. Two notable exceptions are T^1ML in anterior left and posterior right WM from (1, 1) and (0, 2) profiles: these estimates are significantly lower and higher than analogous estimates from other profiles, respectively. Results thus suggest that the optimized (2, 1) scan profile yields WM T^1ML estimates that are more consistently similar to IR WM T^1ML estimates than other optimized SPGR/DESS profiles.

Systematic differences in 2 sample means are evident across scan profiles, particularly within WM ROIs. Curiously, the (1, 1) profile agrees most consistently (in WM/GM T^2ML within-ROI sample mean) with reference estimates, though with relatively high sample variation. The (2, 1) and (0, 2) SPGR/DESS profiles produce consistently lower WM T^2ML than the reference IR/SE profile, though the (0, 2) profile is in reasonable agreement with other steady-state estimates [73].

These discrepancies may due to differences in sensitivity to multi-compartmental relaxation [74]. Specifically, different signal models with different scan parameter choices might be more or less sensitive to the model mismatch incurred by neglecting to distinguish the multiple T2 components within each voxel. Section S.VI studies T2 estimation bias due to multi-compartmental relaxation in more detail.

V. Discussion and Future Work

Phantom experiments show that optimized scan profiles consisting of (2, 1), (1, 1), and (0, 2) (SPGR, DESS) scans yield accurate WM/GM T1, T2 estimates, and that empirical precision trends across profiles agree reasonably with CRB-based predictions. However, in vivo experiments reveal that even with scan optimization, it may be challenging to achieve clinically viable levels of precision from the aforementioned steady-state profiles, at least at 3T. At the expense of greater scan time, it is of course possible that optimized profiles containing greater numbers of SPGR, DESS, and/or other steady-state scans can provide clinically acceptable precision levels. For these and other more complicated scan profiles, estimator dependence on scan parameters becomes even less intuitive, increasing the need for scan design.

The proposed scan design framework addresses spatial variation in object parameters through a min-max design criterion. The min-max criterion guarantees an upper bound on a weighted sum of variances and assumes no prior knowledge of distributions. However, in general it is non-differentiable in P, precluding gradient-based optimization. Furthermore, it is conservative by nature, and often selects scan parameters based on corner cases of the object parameter space. To reduce the influence of corner cases, it may be desirable to instead construct a cost function related to the coefficient of variation as in [44, 4648], perhaps by setting parameter weights W−1 ← diag (x) for x ≠ 0 in (5).

As a less conservative alternative to min-max design, other recent works [50, 51] have addressed object parameter spatial variation by instead constructing cost functions related to the Bayesian CRB [75], which characterizes the expected precision with respect to a prior distribution on object parameters. Bayesian cost functions are usually differentiable and can also, with appropriate priors, penalize object parameter coefficients of variation instead of variances, as in [50]. However, prior distributions are generally unknown, and may need to be estimated from data, as in [51].

Careful calibration of flip angle scaling κ is essential for accurate T1,T2 estimation from SPGR/DESS scan profiles. In this work, we estimate κ from separate acquisitions and adjust nominal flip angles prior to reconstruction, but acknowledge that non-idealities in those separate acquisitions may themselves cause resultant B^1+ errors to propagate into our T1,T2 estimates. To reduce error propagation, it may be desirable to instead design scan profiles to permit joint estimation of κ, in addition to other latent object parameters. Unfortunately, we find that optimizing the (2, 1) or (0, 2) profile to allow for four-parameter x(r) := [ME(r), T1(r), T2(r), κ(r)] estimation results in unacceptably high amplification of the worst-case T1 standard deviation. (Incidentally, precise T2 ML and RLS estimation alone from the (2, 1) or (0, 2) profile is possible [42].) It remains an open scan design question as to whether time spent collecting Bloch-Siegert data for separate B1+ mapping could instead be better spent collecting additional SPGR, DESS, and/or other data for joint estimation.

By working with closed-form signal expressions, we neglect to model several higher-order effects. However, it is apparent that the nonlinear estimation procedures required for many mapping problems can amplify the influence of these secondary effects, often inducing substantial bias. Since the CRB (as described) applies only to unbiased estimators, it is thus desirable to use signal models that are as complete as possible for CRB-based scan design. In theory, scan optimization approach (9) is even compatible with acquisitions where a closed-form model relating data to latent and scan parameters is unknown, as in [22, 76]. In practice, difficulties arise in efficient computation of signal gradients required in (3), which may demand more specialized techniques, as in [77]. Designing scan profiles involving such complex signal models would likely necessitate optimization techniques more involved than the simple grid searches used in this work.

VI. Conclusion

We have introduced a CRB-inspired min-max optimization approach to aid robust, application-specific MR scan selection and optimization for precise parameter estimation. As a detailed example, we have optimized combinations of fast SPGR and DESS scans for T1, T2 relaxometry in WM and GM regions of the human brain at 3T. Numerical simulations show that at typical noise levels and with accurate flip angle prior knowledge, WM- and GM-like T1, T2 ML estimates from optimized scans are nearly unbiased, and so worst-case CRB predictions yield reliable bounds on ROI sample variances. Phantom accuracy experiments show that optimized combinations of (2, 1), (1, 1), or (0, 2) (SPGR, DESS) scans are in excellent agreement with NIST and IR/SE measurements over the designed latent object parameter range of interest. Phantom precision experiments show that these SPGR/DESS combinations exhibit trends in pooled sample standard deviations that reasonably reflect CRB predictions. In vivo experiments suggest that with optimization, the (0, 2) profile can yield comparable 1, 2 precision to the more conventional (2, 1) [42] scan profile in well-isolated WM/GM ROIs; however, the (0, 2) T1 estimates are unreliable within and near the CSF and do not agree with IR measurements in WM as consistently as the (2, 1) profile. This and other disagreements across profiles in vivo may be attributable to differences in signal model sensitivities to neglected higher-order effects. Nevertheless, this simple example application illustrates that scan optimization may enable new parameter mapping techniques from established pulse sequences.

Supplementary Material

supplement

Acknowledgments

This work was supported in part by the National Institutes of Health (NIH) under Grant P01 CA87634, and in part by the University of Michigan’s MCubed funding program.

The authors thank Dr. Kathryn Keenan and Dr. Stephen Russek at NIST for generously lending a prototype [78] (used in initial experiments) of the High Precision Devices® MR system phantom, and for sharing an early preprint of [68]. The authors also thank the reviewers for their helpful suggestions.

Footnotes

1

Though the noise distribution of k-space raw data is usually well-modeled as complex white Gaussian, the noise distribution of the dth reconstructed image yd(·) depends both on the acquisition and reconstruction. If single receive channel k-space data is fully-sampled on a Cartesian grid, each dataset yd(·) is recoverable via separate Fourier transform, and is thus complex Gaussian and independent across datasets. However, if k-space data is multi-channel, undersampled, and/or non-Cartesian, it may be preferable that yd(·) be estimated by more sophisticated techniques, e.g., [52, 53]. In such cases, reconstructed image noise is unlikely to be Gaussian-distributed.

2

In particular, f must be analytic in complex components of x at each r.

3

Finite-duration RF effects influence SS signals more strongly for shorter TR, larger α0, and smaller T2/T1 [62]. In this work, RF pulse durations do not exceed 11% of repetition times and do not excite nominal flip angles greater than 35°. Phantom results (cf. Fig. 1) show that for these scan parameters, the influence of finite-duration RF effects on T1, T2 estimates is small even for T2/T1 ≈ 0.1 as in WM/GM.

4

Supplementary material is available in the /media tab on IEEEXplore.

5

Each sample statistic presented hereafter is rounded off to the highest place value of its corresponding uncertainty measure. For simplicity, each uncertainty measure is itself endowed one extra significant figure. Decimal points indicate the significance of trailing zeros.

6

We have taken effort to try and select ROIs that reflect expected anatomy in all coil-combined and registered images, including adjacent slices in images from 3D acquisitions. However, we acknowledge the possibility of some contamination across tissue boundaries, especially WM and/or CSF contamination into cortical GM.

Color versions of one or more of the figures in this paper are available online at http://ieeexplore.ieee.org.

Contributor Information

Gopal Nataraj, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI, USA.

Jon-Fredrik Nielsen, Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI, USA.

Jeffrey A. Fessler, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, MI, USA.

References

  • 1.Cheng HLM, Stikov N, Ghugre NR, Wright GA. Practical medical applications of quantitative MR relaxometry. J Mag Res Im. 2012 Oct;36(4):805–24. doi: 10.1002/jmri.23718. [DOI] [PubMed] [Google Scholar]
  • 2.Larsson HBW, Frederiksen J, Kjaer L, Henriksen O, Olesen J. In vivo determination of T1 and T2 in the brain of patients with severe but stable multiple sclerosis. Mag Res Med. 1988 May;7(1):43–55. doi: 10.1002/mrm.1910070106. [DOI] [PubMed] [Google Scholar]
  • 3.Kurki T, Lundbom N, Komu M, Kormano M. Tissue characterization of inter cranial tumors by magnetization transfer and spin-lattice relaxation parameters in vivo. J Mag Res Im. 1996 Aug;6(4):573–9. doi: 10.1002/jmri.1880060403. [DOI] [PubMed] [Google Scholar]
  • 4.Englund E, Brun A, Gyorffy-Wagner Z, Larsson E, Persson B. Relaxation times in relation to grade of malignancy and tissue necrosis in astrocytic gliomas. Mag Res Im. 1986;4(5):425–9. [Google Scholar]
  • 5.Siemonsen S, Mouridsen K, Holst B, Ries T, Finsterbusch J, Thomalia G, Ostergaard L, Fiehler J. Quantitative T2 values predict time from symptom onset in acute stroke patients. Stroke. 2009 May;40(5):1612–6. doi: 10.1161/STROKEAHA.108.542548. [DOI] [PubMed] [Google Scholar]
  • 6.DeWitt LD, Kistler JP, Miller DC, Richardson EP, Buonanno FS. NMR-neuropathologic correlation in stroke. Stroke. 1987;18(2):342–51. doi: 10.1161/01.str.18.2.342. [DOI] [PubMed] [Google Scholar]
  • 7.Matzat SJ, Tiel JV, Gold GE, Oei EHG. Quantitative MRI techniques of cartilage composition. Quant Imaging Med Surg. 2013 Jun;3(3):162–74. doi: 10.3978/j.issn.2223-4292.2013.06.04. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Mosher TJ, Dardzinski BJ. Cartilage MRI T2 relaxation time mapping: overview and applications. Semin Musculoskelet Radiol. 2004;8(4):355–68. doi: 10.1055/s-2004-861764. [DOI] [PubMed] [Google Scholar]
  • 9.Guo H, Au W-Y, Cheung JS, Kim D, Jensen JH, Khong P-L, Chan Q, Chan KC, Tosti C, Tang H, Brown TR, Lam WWM, Ha S-Y, Brittenham GM, Wu EX. Myocardial T2 quantification in patients with iron overload at 3 Tesla. J Mag Res Im. 2009 Aug;30(2):394–400. doi: 10.1002/jmri.21851. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Giri S, Chung YC, Merchant A, Mihai G, Rajagopalan S, Raman SV, Simonetti OP. T2 quantification for improved detection of myocardial edema. Cardiovasc Magn Reson. 2009;11(1):56–68. doi: 10.1186/1532-429X-11-56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Look DC, Locker DR. Time saving in measurement of NMR and EPR relaxation times. Rev Sci Instrum. 1970 Feb;41(2):250–1. [Google Scholar]
  • 12.Carr HY, Purcell EM. Effects of diffusion on free precession in nuclear magnetic resonance experiments. Phys Rev. 1954 May;94(3):630–8. [Google Scholar]
  • 13.Stehling MK, Turner R, Mansfield P. Echo-planar imaging: magnetic resonance imaging in a fraction of a second. Science. 1991 Oct;254(5028):43–50. doi: 10.1126/science.1925560. [DOI] [PubMed] [Google Scholar]
  • 14.Ahn CB, Kim JH, Cho ZH. High-speed spiral-scan scho planar NMR imaging - I. IEEE Trans Med Imag. 1986 Mar;5(1):2–7. doi: 10.1109/TMI.1986.4307732. [DOI] [PubMed] [Google Scholar]
  • 15.Meyer CH, Hu BS, Nishimura DG, Macovski A. Fast spiral coronary artery imaging. Mag Res Med. 1992 Dec;28(2):202–13. doi: 10.1002/mrm.1910280204. [DOI] [PubMed] [Google Scholar]
  • 16.Kay I, Henkelman RM. Practical Implementation and Optimization of One-shot T1 imaging. Mag Res Med. 1991 Dec;22(2):414–24. doi: 10.1002/mrm.1910220249. [DOI] [PubMed] [Google Scholar]
  • 17.Gowland PA, Leach MO. Fast and accurate measurements of T1 using a multi-readout single inversion-recovery sequence. Mag Res Med. 1992 Jul;26(1):79–88. doi: 10.1002/mrm.1910260109. [DOI] [PubMed] [Google Scholar]
  • 18.Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery (MOLLI) for high-resolution T1 mapping of the heart. Mag Res Med. 2004 Jul;52(1):141–6. doi: 10.1002/mrm.20110. [DOI] [PubMed] [Google Scholar]
  • 19.Stehling MK, Ordidge RJ, Coxon R, Mansfield P. Inversion-recovery Echo-planar imaging (IR-EPI) at 0.5T. Mag Res Med. 1990 Mar;13(3):514–7. doi: 10.1002/mrm.1910130320. [DOI] [PubMed] [Google Scholar]
  • 20.Bonny JM, Zanca M, Boire JY, Veyre A. T2 maximum likelihood estimation from multiple spin-echo magnitude images. Mag Res Med. 1996 Aug;36(2):287–93. doi: 10.1002/mrm.1910360216. [DOI] [PubMed] [Google Scholar]
  • 21.Kumar D, Nguyen TD, Gauthier SA, Raj A. Bayesian algorithm using spatial priors for multiexponential T2 relaxometry from multiecho spin echo MRI. Mag Res Med. 2012 Nov;68(5):1536–43. doi: 10.1002/mrm.24170. [DOI] [PubMed] [Google Scholar]
  • 22.Ben-Eliezer N, Sodickson DK, Block KT. Rapid and accurate T2 mapping from multi–spin-echo data using Bloch-simulation-based reconstruction. Mag Res Med. 2015 Feb;73(2):809–17. doi: 10.1002/mrm.25156. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Nguyen TD, Wisnieff C, Cooper MA, Kumar D, Raj A, Spincemaille P, Wang Y, Vartanian T, Gauthier SA. T2prep three-dimensional spiral imaging with efficient whole brain coverage for myelin water quantification at 1.5 tesla. Mag Res Med. 2012 Mar;67(3):614–21. doi: 10.1002/mrm.24128. [DOI] [PubMed] [Google Scholar]
  • 24.Majumdar S, Orphanoudakis SC, Gmitro A, O’Donnell M, Gore JC. Error in the measurements of T2 using multiple-echo MRI techniques: 1. Effect of radiofrequency pulse imperfections. Mag Res Med. 1986 Jun;3(3):397–417. doi: 10.1002/mrm.1910030305. [DOI] [PubMed] [Google Scholar]
  • 25.Majumdar S, Orphanoudakis SC, Gmitro A, O’Donnell M, Gore JC. Error in the measurements of T2 using multiple-echo MRI techniques: 2. Effects of static field inhomogeneity. Mag Res Med. 1986 Aug;3(4):562–74. doi: 10.1002/mrm.1910030410. [DOI] [PubMed] [Google Scholar]
  • 26.Farzaneh F, Riederer SJ, Pelc NJ. Analysis of T2 limitations and off-resonance effects on spatial resolution and artifacts in echo-planar imaging. Mag Res Med. 1990 Apr;14(1):123–39. doi: 10.1002/mrm.1910140112. [DOI] [PubMed] [Google Scholar]
  • 27.Hinshaw WS. Image formation by nuclear magnetic resonance: The sensitive point method. J Appl Phys. 1976 Aug;47(8):3709. [Google Scholar]
  • 28.Scheffler K. A pictorial description of steady-states in rapid magnetic resonance imaging. Concepts in Magnetic Resonance. 1999;11(5):291–304. [Google Scholar]
  • 29.Deoni SCL, Rutt BK, Peters TM. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Mag Res Med. 2003 Mar;49(3):515–26. doi: 10.1002/mrm.10407. [DOI] [PubMed] [Google Scholar]
  • 30.Chang LC, Koay CG, Basser PJ, Pierpaoli C. Linear least-squares method for unbiased estimation of T1 from SPGR signals. Mag Res Med. 2008 Aug;60(2):496–501. doi: 10.1002/mrm.21669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Fram EK, Herfkens RJ, Johnson GA, Glover GH, Kaaris JP, Shimakawa A, Perkins TG, Pelc NJ. Rapid calculation of T1 using variable flip angle gradient refocused imaging. Mag Res Im. 1987;5(3):201–8. doi: 10.1016/0730-725x(87)90021-x. [DOI] [PubMed] [Google Scholar]
  • 32.Wang H, Cao Y. Spatially regularized T1 estimation from variable flip angles MRI. Med Phys. 2012 Jul;39(7):4139–48. doi: 10.1118/1.4722747. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Deoni SCL, Ward HA, Peters TM, Rutt BK. Rapid T2 estimation with phase-cycled variable nutation steady-state free precession. Mag Res Med. 2004 Aug;52(2):435–9. doi: 10.1002/mrm.20159. [DOI] [PubMed] [Google Scholar]
  • 34.Deoni SCL. Transverse relaxation time (T2) mapping in the brain with off-resonance correction using phase-cycled steady-state free precession imaging. J Mag Res Im. 2009 Aug;30(2):411–7. doi: 10.1002/jmri.21849. [DOI] [PubMed] [Google Scholar]
  • 35.Welsch GH, Scheffler K, Mamisch TC, Hughes T, Millington S, Deimling M, Trattnig S. Rapid estimation of cartilage T2 based on double echo at steady state (DESS) with 3 Tesla. Mag Res Med. 2009 Aug;62(2):544–9. doi: 10.1002/mrm.22036. [DOI] [PubMed] [Google Scholar]
  • 36.Heule R, Ganter C, Bieri O. Rapid estimation of cartilage T2 with reduced T1 sensitivity using double echo steady state imaging. Mag Res Med. 2014 Mar;71(3):1137–43. doi: 10.1002/mrm.24748. [DOI] [PubMed] [Google Scholar]
  • 37.Stöcker T, Keil F, Vahedipour K, Brenner D, Pracht E, Shah NJ. MR parameter quantification with magnetization-prepared double echo steady-state (MP-DESS) Mag Res Med. 2014 Jul;72(1):103–11. doi: 10.1002/mrm.24901. [DOI] [PubMed] [Google Scholar]
  • 38.Heule R, Ganter C, Bieri O. Triple echo steady-state (TESS) relaxometry. Mag Res Med. 2014 Jan;71(1):230–7. doi: 10.1002/mrm.24659. [DOI] [PubMed] [Google Scholar]
  • 39.Bruder H, Fischer H, Graumann R, Deimling M. A new steady-state imaging sequence for simultaneous acquisition of two MR images with clearly different contrasts. Mag Res Med. 1988 May;7(1):35–42. doi: 10.1002/mrm.1910070105. [DOI] [PubMed] [Google Scholar]
  • 40.Gyngell ML. The steady-state signals in short-repetition-time sequences. J Mag Res. 1989 Feb;81(3):474–83. [Google Scholar]
  • 41.Hänicke W, Vogel HU. An analytical solution for the SSFP signal in MRI. Mag Res Med. 2003 Apr;49(4):771–5. doi: 10.1002/mrm.10410. [DOI] [PubMed] [Google Scholar]
  • 42.Nataraj G, Nielsen J-F, Fessler JA. Model-based estimation of T2 maps with dual-echo steady-state MR imaging. Proc IEEE Intl Conf on Image Processing. 2014:1877–81. [Google Scholar]
  • 43.Weiss GH, Gupta RK, Ferretti JA, Becker ED. The choice of optimal parameters for measurement of spin-lattice relaxation times. I. Mathematical formulation. J Mag Res. 1980 Feb;37(3):369–79. [Google Scholar]
  • 44.Zhang Y, Yeung HN, O’Donnell M, Carson PL. Determination of sample time for T1 measurement. J Mag Res Im. 1998 May;8(3):675–81. doi: 10.1002/jmri.1880080324. [DOI] [PubMed] [Google Scholar]
  • 45.Wang HZ, Riederer SJ, Lee JN. Optimizing the precision in T1 relaxation estimation using limited flip angles. Mag Res Med. 1987 Nov;5(5):399–416. doi: 10.1002/mrm.1910050502. [DOI] [PubMed] [Google Scholar]
  • 46.Jones JA, Hodgkinson P, Barker AL, Hore PJ. Optimal sampling strategies for the measurement of spin-spin relaxation times. J Mag Res B. 1996 Oct;113(1):25–34. [Google Scholar]
  • 47.Imran J, Langevin François, Saint-Jalmes Hervé. Two-point method for T1 estimation with optimized gradient-echo sequence. Mag Res Im. 1999 Nov;17(9):1347–56. doi: 10.1016/s0730-725x(99)00092-2. [DOI] [PubMed] [Google Scholar]
  • 48.Deoni SCL, Peters TM, Rutt BK. Determination of optimal angles for variable nutation proton magnetic spin-lattice, T1, and spin-spin, T2, relaxation times measurement. Mag Res Med. 2004 Jan;51(1):194–9. doi: 10.1002/mrm.10661. [DOI] [PubMed] [Google Scholar]
  • 49.Fleysher L, Fleysher R, Liu S, Zaaraoui W, Gonen O. Optimizing the precision-per-unit-time of quantitative MR metrics: Examples for T1, T2, and DTI. Mag Res Med. 2007 Feb;57(2):380–7. doi: 10.1002/mrm.21144. [DOI] [PubMed] [Google Scholar]
  • 50.Akçakaya M, Weingärtner S, Roujol Sébastien, Nezafat R. On the selection of sampling points for myocardial T1 mapping. Mag Res Med. 2015 May;73(5):1741–53. doi: 10.1002/mrm.25285. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Lewis CM, Hurley SA, Meyerand ME, Koay CG. Data-driven optimized flip angle selection for T1 estimation from spoiled gradient echo acquisitions. Mag Res Med. 2016 doi: 10.1002/mrm.25920. to appear. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Fessler JA, Sutton BP. Nonuniform fast Fourier transforms using min-max interpolation. IEEE Trans Sig Proc. 2003 Feb;51(2):560–74. [Google Scholar]
  • 53.Muckley MJ, Noll DC, Fessler JA. Fast parallel MR image reconstruction via B1-based, adaptive restart, iterative soft thresholding algorithms (BARISTA) IEEE Trans Med Imag. 2015 Feb;34(2):578–88. doi: 10.1109/TMI.2014.2363034. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Macovski A. Noise in MRI. Mag Res Med. 1996 Sep;36(3):494–7. doi: 10.1002/mrm.1910360327. [DOI] [PubMed] [Google Scholar]
  • 55.Lei T. Statistics of MR signals: revisited. Proc SPIE 6510 Medical Imaging 2007: Phys Med Im. 2007:651052. [Google Scholar]
  • 56.Fisher RA. Theory of statistical estimation. Proc Cambridge Philo- sophical Society. 1925 Jul;22(5):700–25. [Google Scholar]
  • 57.Cramér H. Mathematical methods of statistics. Princeton: Princeton Univ. Press; 1946. [Google Scholar]
  • 58.Chernoff H. Locally optimal designs for estimating parameters. Ann Math Stat. 1953 Dec;24(4):586–602. [Online]. Available: http://www.jstor.org/stable/2236782. [Google Scholar]
  • 59.Carney CE, Wong STS, Patz S. Analytical solution and verification of diffusion effect in SSFP. Mag Res Med. 1991 Jun;19(2):240–6. doi: 10.1002/mrm.1910190209. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Wu EX, Buxton RB. Effect of diffusion on the steady-state magnetization with pulsed field gradients. J Mag Res. 1990 Nov;90(2):243–53. [Google Scholar]
  • 61.Kaiser R, Bartholdi E, Ernst RR. Diffusion and field-gradient effects in NMR Fourier spectroscopy. J Chem Phys. 1974 Apr;60(8):2966–79. [Google Scholar]
  • 62.Bieri O, Scheffler K. SSFP signal with finite RF pulses. Mag Res Med. 2009 Nov;62(5):1232–41. doi: 10.1002/mrm.22116. [DOI] [PubMed] [Google Scholar]
  • 63.Zur Y, Wood ML, Neuringer LJ. Spoiling of transverse magnetization in steady-state sequences. Mag Res Med. 1991 Oct;21(2):251–63. doi: 10.1002/mrm.1910210210. [DOI] [PubMed] [Google Scholar]
  • 64.Wansapura JP, Holland SK, Dunn RS, Ball WS. NMR relaxation times in the human brain at 3.0 Tesla. J Mag Res. 1999 Apr;9(4):531–8. doi: 10.1002/(sici)1522-2586(199904)9:4<531::aid-jmri4>3.0.co;2-l. [DOI] [PubMed] [Google Scholar]
  • 65.Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM. T1, T2 relaxation and magnetization transfer in tissue at 3T. Mag Res Med. 2005 Sep;54(3):507–12. doi: 10.1002/mrm.20605. [DOI] [PubMed] [Google Scholar]
  • 66.Kwan RKS, Evans AC, Pike GB. MRI simulation-based evaluation of image-processing and classification methods. IEEE Trans Med Imag. 1999 Nov;18(11):1085–97. doi: 10.1109/42.816072. [DOI] [PubMed] [Google Scholar]
  • 67.Collins DL, Zijdenbos AP, Kollokian V, Sled JG, Kabani NJ, Holmes CJ, Evans AC. Design and construction of a realistic digital brain phantom. IEEE Trans Med Imag. 1998 Jun;17(3):463–8. doi: 10.1109/42.712135. [DOI] [PubMed] [Google Scholar]
  • 68.Keenan KE, Stupic KF, Boss MA, Russek SE, Chenevert TL, Prasad PV, Reddick WE, Cecil KM, Zheng J, Hu P, Jackson EF. Multi-site, multi-vendor comparison of T1 measurement using ISMRM/NIST system phantom. Proc Intl Soc Mag Res Med. 2016:3290. [Google Scholar]
  • 69.Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm. IEEE Trans Med Imag. 1991 Mar;10(1):53–65. doi: 10.1109/42.75611. [DOI] [PubMed] [Google Scholar]
  • 70.Sacolick LI, Wiesinger F, Hancu I, Vogel MW. B1 mapping by Bloch-Siegert shift. Mag Res Med. 2010 May;63(5):1315–22. doi: 10.1002/mrm.22357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Sun H, Grissom WA, Fessler JA. Regularized estimation of Bloch-Siegert B1+ Maps in MRI. Proc IEEE Intl Conf on Image Processing. 2014:3646–50. [Google Scholar]
  • 72.Ahn S, Fessler JA. Tech Rep 413. Comm. and Sign. Proc. Lab., Dept. of EECS, Univ. of Michigan; Ann Arbor, MI, 48109-2122: Jul, 2003. Standard errors of mean, variance, and standard deviation estimators. [Online]. Available: http://web.eecs.umich.edu/~fessler/papers/lists/files/tr/stderr.pdf. [Google Scholar]
  • 73.Heule R, Bär P, Mirkes C, Scheffler K, Trattnig S, Bieri O. Triple-echo steady-state T2 relaxometry of the human brain at high to ultra-high fields. NMR in Biomedicine. 2014 Sep;27(9):1037–45. doi: 10.1002/nbm.3152. [DOI] [PubMed] [Google Scholar]
  • 74.Mackay A, Whittall K, Adler J, Li D, Paty D, Graeb D. In vivo visualization of myelin water in brain by magnetic resonance. Mag Res Med. 1994 Jun;31(6):673–7. doi: 10.1002/mrm.1910310614. [DOI] [PubMed] [Google Scholar]
  • 75.Gill RD, Levit BY. Applications of the van Trees inequality: A Bayesian Cramér-rao bound. Bernoulli. 1995;1(1/2):59–79. [Online]. Available: http://www.jstor.org/stable/3318681. [Google Scholar]
  • 76.Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA. Magnetic resonance fingerprinting. Nature. 2013 Mar;495:187–93. doi: 10.1038/nature11971. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 77.Zhao B, Haldar J, Setsompop K, Wald LL. Optimal experiment design for magnetic resonance fingerprinting. embc. 2016 doi: 10.1109/EMBC.2016.7590737. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.Russek SE, Boss M, Jackson EF, Jennings DL, Evelhoch JL, Gunter JL, Sorensen AG. Characterization of NIST/ISMRM MRI system phantom. Proc Intl Soc Mag Res Med. 2012:2456. [Online]. Available: http://cds.ismrm.org/protected/12MProceedings/files/2456.pdf.

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

supplement

RESOURCES