Air Filters for Germ-free Isolators

Charles E Yale 1, Alapakkam R Vivek 1
PMCID: PMC547734  PMID: 4881953

Abstract

A germ-free isolator must have a perfect bacterial filter. This paper describes a new, relatively inexpensive, stainless-steel filter frame, which is easily and quickly assembled and protects the enclosed filter material at all times. Resistance to the flow of air was less than 4 inches of water at an airflow of 30 ft3/min through the filter frame with 204 inches2 of surface area and four, one-half inch thick pieces of fiberglass filter material. This filter performed satisfactorily in our gnotobiotic laboratory and was found to be consistently 100% efficient in removing an aerosol containing Serratia marcescens from an air stream under a variety of operating conditions.

Full text

PDF
1653

Images in this article

Selected References

These references are in PubMed. This may not be the complete list of references from this article.

  1. HUMPHREY A. E. Air sterilization. Adv Appl Microbiol. 1960;2:301–311. doi: 10.1016/s0065-2164(08)70132-1. [DOI] [PubMed] [Google Scholar]
  2. MAY K. R., HARPER G. J. The efficiency of various liquid impinger samplers in bacterial aerosols. Br J Ind Med. 1957 Oct;14(4):287–297. doi: 10.1136/oem.14.4.287. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. TREXLER P. C., REYNOLDS L. I. Flexible film apparatus for the rearing and use of germfree animals. Appl Microbiol. 1957 Nov;5(6):406–412. doi: 10.1128/am.5.6.406-412.1957. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Applied Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES