. 2017 Jul 6;2017:bcr2016218757. doi: 10.1136/bcr-2016-218757

Substantial harm associated with failure of chronic paediatric central venous access devices

Amanda J Ullman 1,2,3, Tricia Kleidon 3,4, Marie Cooke 1,3, Claire M Rickard 1,3
PMCID: PMC5534914  PMID: 28687683

Abstract

Central venous access devices (CVADs) form an important component of modern paediatric healthcare, especially for children with chronic health conditions such as cancer or gastrointestinal disorders. However device failure and complications rates are high.

Over 2½ years, a child requiring parenteral nutrition and associated vascular access dependency due to ‘short gut syndrome’ (intestinal failure secondary to gastroschisis and resultant significant bowel resection) had ten CVADs inserted, with ninesubsequently failing. This resulted in multiple anaesthetics, invasive procedures, injuries, vascular depletion, interrupted nutrition, delayed treatment and substantial healthcare costs. A conservative estimate of the institutional costs for each insertion, or rewiring, of her tunnelled CVAD was $A10 253 (2016 Australian dollars).

These complications and device failures had significant negative impact on the child and her family. Considering the commonality of conditions requiring prolonged vascular access, these failures also have a significant impact on international health service costs.

Keywords: Congenital Disorders, Intravenous / Infusion Nursing, Infection Control In Hospital

Background

Central venous access devices (CVADs) are an essential component of short-term and long-term care provision for ~50 000 children in Australia and millions of children internationally.1 2 Despite their necessity, one in four of these devices is associated with a severe complication and fails prior to completion of treatment.3 These rates of CVAD failure, in addition to higher rates of complications that resolve after interventions, are unacceptable and place an enormous burden on the healthcare system, children and families.

For children with chronic health conditions, and dependence on vascular access for nutrition and treatment, device failures substantially reduce morbidity, quality of life and longevity.4 However seminal literature and international guidelines5–7 are principally focused on the prevention of only infectious complications, despite the higher prevalence of other types of CVAD dysfunction.3 Mechanical and vascular injuries associated with CVADs, such as venous thrombosis, are critical. Each result in treatment disruption, additional resources and difficulty associated with replacement CVAD insertion, morbidity and long-term vessel depletion.3 8 Vessel depletion further seriously impacts on the health and well-being of children who have a lifelong dependency on vascular access and existing chronic conditions such as immunocompromised or malabsorption nutritional disorders.4

A demonstration of the burden and harm associated with CVADs for children with chronic, vascular access dependent conditions is not evident in the literature. With consent from the legal guardian, a case study has been used to illustrate the current complex and harmful experience of paediatric CVAD dependency and failure. The case study also explains the local resources wasted due to potentially preventable adverse events associated with ineffective paediatric CVAD practices.

Case presentation

The child was born in an Australian tertiary referral hospital in June 2014 with gastroschisis. After early surgical repair, she required significant bowel resection (15 cm, including the ileocecal valve), resulting in short bowel syndrome (or intestinal failure) and a long-term dependency on parenteral nutrition (PN). While the child’s treatment nutritional goal is enteral autonomy, she is currently heavily dependent on vascular access for the safe administration of PN and is likely to remain so for the foreseeable future.

Outcome and follow-up

By 2½ years of age, she has had ten CVADs inserted, with ninesubsequently failing: five tunnelled CVADs; three peripherally inserted central catheters (PICCs) and one non-tunnelled CVAD, with one tunnelled CVAD still in situ. The timeline and sequelae of each CVAD is described in table 1. This equates to the child having a CVAD inserted every 100 days, each requiring a general anaesthetic and surgical procedure. Most complications resulting in CVAD failure were related to mechanical causes, most frequently dislodgement and fracture. In addition, over the last 2 years, the child’s central vasculature has developed significant thrombotic occlusion, limiting future CVAD placements. Each failed CVAD restricted her PN, required the insertion of temporary peripheral intravenous devices and/or necessitated antimicrobial therapy.

Table 1.

Central vascular access device timeline

Date inserted CVAD type inserted and site Complications (not resulting in CVAD removal) during CVAD dwell Date CVAD removed Complication and reason for CVAD removal Time period CVAD successful
18 June 2014 PICC;
left brachial
6 August 2014 Complete occlusion 49 days
7 August 2014 PICC;
right brachial
18 November 2014 Dislodged; tip no longer in central position 103 days
18 November 2014 Tunnelled, cuffed CVAD;
left internal jugular
6 March 2015: occluded right subclavian vein
7 March 2015: tip position in right atrium, traversing tricuspid valve.
Reposition required
2 June 2015 CVAD blocked, right chest swelling.
Internal catheter fracture, confirmed by contrast study
196 days
7 June 2015 Tunnelled, cuffed CVAD;
right internal jugular
5 June 2015: occluded true right subclavian vein; multiple collaterals formed
9 June 2015: partial CVAD dislodgement (from SVC/RA junction to proximal SVC)
29 August 2015 Dislodged; tip no longer in central position 83 days
29 August 2015 Tunnelled, cuffed CVAD;
right internal jugular (rewire)
14 October 2015 CVAD-associated bloodstream infection (Klebsiella pneumoniae) and catheter fracture 46 days
25 October 2015 PICC;
right brachial
26 March 2016 Dislodged; tip no longer in central position
Unsuccessful PICC replacement
153 days
28 March 2016 Non-tunnelled CVAD;
right internal jugular
30 March 2016 Elective replacement of an inappropriate device 2 days
30 March 2016 Tunnelled, cuffed CVAD;
right internal jugular
3 April 2016 Catheter fracture; repair unsuccessful
Required prophylactic intravenous antibiotics, PIVC insertion, partial PN only
4 days
4 April 2016 Tunnelled, cuffed CVAD;
right internal jugular (rewire)
6 June 2016: Catheter fracture; repaired successfully
Required prophylactic intravenous antibiotics, PIVC insertion, partial PN only
6 October 2016 Catheter fracture; repair unsuccessful
Required prophylactic intravenous antibiotics, PIVC insertion, partial PN only
185 days
7 October 2016 Tunnelled, cuffed CVAD;
right internal jugular (rewire)
Currently in situ

CVAD, central venous access device; PICC, peripherally inserted central catheter; PIVC, peripheral intravenous catheter; PN, parenteral nutrition; RA, right atrium; SVC, superior vena cava.

In 2015, the child developed a CVAD-associated bloodstream infection caused by Klebsiella pneumoniae. This infection resulted in a 22-day admission to a tertiary paediatric hospital; 14 days of intravenous antibiotics (ceftriaxone) and 3 days of gentamicin locks; hypokalaemia (K+ 2.7 mmol/L) requiring overnight potassium chloride infusion; CVAD fracture during hospital stay that was unable to be repaired resulting in CVAD removal; insertion of three peripheral intravenous catheters (one necessitating ultrasound guidance) until PICC insertion and 7 days of missed PN.

The local, institutional resources used when inserting each of these CVADs is substantial. An example of the immediate financial costs associated with inserting, or rewiring, a tunnelled CVAD at the case study facility is described in table 2. This does not include additional costs associated with treating, or attempting to treat, the underlying complication. A cost of $A10 253 (2016 dollars) per device is a major cost to paediatric hospitals, and the child has undergone six tunnelled CVAD insertion, or rewiring, in her short life time.

Table 2.

Local resources used to insert, or rewire, a tunnelled CVAD in 2016 Australian dollars

Resources Cost ($A)
Staff24 25
Consultant paediatric surgeon ($A199.16/hour): 2 hours 398.32
Training (registrar, resident medical officer) paediatric surgeon ($A128.98/hour): 2 hours 257.55
Registered nurses or equivalent (operating specialty)($A42.4711/hour): two nurses, 2½ hours per nurse 212.36
Consultant anaesthetist ($A199.16/hour): 2½ hours 497.90
Training (registrar, resident medical officer) paediatric anaesthetist ($A128.98/hour): 2½ hours 322.45
Registered nurse or equivalent (anaesthetics and recovery specialty) ($A42.4711/hour): 3 hours 127.42
Location26
Dedicated theatre suite (including equipment such as ultrasonography, imaging) ($A80/minute): 1 ½ hours 7200.00
Anaesthetic recovery: at least 30 min 89.98
Hospital bed: at least 8 hours 803.95
Consumables26 27
Central venous access device and insertion equipment: frequently >1 required 275.00
Sterile personal protective equipment ($A5.00 each): four at least 20.00
Anaesthetic medications (including inhaled, intravenous): multiple 34.38
Device dressing and security 14.00
Total 10 253.31

CVAD, central venous access device.

Discussion

The child’s vascular access progression may seem unremarkable to paediatric clinicians, because such complications are common and in-depth data on repercussions are lacking. Yet these experiences were extraordinarily negative for the child, her family and the healthcare system. Children with chronic health conditions requiring prolonged vascular access are common place.9 The potentially preventable harm associated with their CVADs places a large burden on the healthcare system.

The costs and events described by the case study are in accordance with recent case-control studies demonstrating that paediatric CVAD-associated bloodstream infections cost healthcare systems around US$55 646 (2011 US dollars) and 19 additional days in hospital,10 with even higher costs (U$69 332; 2011 US dollars; 21.2 additional days in hospital) for the haematology and oncology paediatric population.11 Delays to treatment and the insertion of replacement CVADs due to other (non-infectious) types of CVAD failure and complication are also expensive for healthcare systems, and significantly reduce the quality of life, morbidity and mortality of paediatric patients.12

As demonstrated in the case study, the immediate interruption to necessary treatment results in an inability to receive prescribed fluids, nutrition, antibiotics and other necessary medicines. In addition to the immediate costs associated with the insertion of a replacement device, the insertion of new CVADs can result in serious complications such as pneumothorax and arterial puncture.3 13 Overall, the treatment of the complication, the interruption to necessary treatment and complications associated with the new CVAD insertion are associated with an increased length of hospital stay, ICU stay and mortality.14 15 This case study has not explored the further financial and other costs borne by the patient and family such as lost time in paid employment and opportunity costs in attendance at playgroup or family occasions.

It is timely for a new focus on long-term vessel health and preservation in paediatrics. Children with CVADs are frequently managed by multiple medical and health professionals. ‘Siloed’ healthcare commonly sees decisions regarding device selection, placement and management made in isolation by individual clinicians from varying backgrounds (eg, oncologist, surgeons, anaesthetists, nurse consultants). A paediatric vascular access continuity of care model, which has been advocated across many other areas of healthcare,16 17 has not been applied, and this is resulting in harm to the patients, their families and the healthcare institutions.

It is also time for high-quality evidence to be generated to improve vascular access outcomes for paediatric patients. A recent focus on preventing harm associated with CVAD insertion and immediate infection has been highly successful in intensive care units;7 18 19 however, evidence is weaker on how to maintain CVAD performance during device dwell and in chronic use settings. The child’s case study, in agreement with previous literature,2 3 has demonstrated that the harm associated with CVADs frequently occurs during the later stages of CVAD dwell, not on the early days after insertion. Paediatric CVAD maintenance procedures such as flushing,20 dressing,21 administration set changes,22 and hub decontamination23 are not well supported by evidence, likely perpetuating these preventable complications for patients.

Learning points.

  • Children with vascular access dependent chronic health conditions are at great risk for harm associated with vascular access complications.

  • Paediatric central venous access device failure and complication rates are high and have a substantial impact on the healthcare system, patient and family.

  • High-quality evidence to improve vascular access outcomes for paediatric patients is urgently needed.

Footnotes

Contributors: All authors have made substantial contributions to the paper and are in agreement with the content. AJU and TK analysed and interpreted the patient data regarding the underlying vascular progression. AJU was the major contributor in writing the manuscript. MC and CRM contributed to the conceptual design and manuscript preparation. All authors read and approved the final manuscript.

Competing interests: Griffith University has received unrestricted investigator initiated research grants to support research projects undertaken by AJU from product manufacturers (3M, Adhezion, Angiodynamics, B. Braun, Centurion Medical Products). Griffith University has received consultancy payments on AJU’s behalf from manufacturers (3M, BD). Griffith University has received unrestricted investigator initiated research and education grants to support projects undertaken by TK from product manufacturers (3M, Adhezion, Angiodynamics, Baxter, Centurion Medical Products). Griffith University has received consultancy payments on TK’s behalf from manufacturers (BD and Angiodynamics). Griffith University has received unrestricted investigator initiated research or educational grants on CRM’s behalf from product manufacturers (3M, Adhezion, Angiodynamics; B. Braun, Becton Dickinson; Carefusion; Centurion Medical Products; Entrotech, Smiths Medical, Teleflex). Griffith University has received consultancy payments on CRM’s behalf from manufacturers (3M, B. Braun, BD, ResQDevices, Smiths Medical). Griffith University has received an unrestricted investigator initiated research or educational grant on MC’s behalf from manufacturers (Baxter, BD).

Patient consent: Obtained from Guardian.

Provenance and peer review: Not commissioned; externally peer reviewed.

References

  • 1. Australian Institute of Health and Welfare. Admitted patient care 2013-14: Australian hospital statistics Australian Institute of Health and Welfare. Canberra: Australian Government, 2015. [Google Scholar]
  • 2. Ullman AJ, Cooke M, Kleidon T, et al. Road map for improvement: point prevalence audit and survey of central venous access devices in paediatric acute care. J Paediatr Child Health 2017;53:123–30. 10.1111/jpc.13347 [DOI] [PubMed] [Google Scholar]
  • 3. Ullman AJ, Marsh N, Mihala G, et al. Complications of Central Venous access Devices: a systematic review. Pediatrics 2015;136:e1331–e1344. 10.1542/peds.2015-1507 [DOI] [PubMed] [Google Scholar]
  • 4. Modi BP, Jaksic T. Pediatric intestinal failure and vascular access. Surg Clin North Am 2012;92:729–43. 10.1016/j.suc.2012.03.012 [DOI] [PubMed] [Google Scholar]
  • 5. World Health Organization. Preventing bloodstream infections from central line venous catheters 2012. (accessed 12 nov 2012). [Google Scholar]
  • 6. O'Grady NP, Alexander M, Burns LA, et al. Guidelines for the prevention of intravascular catheter-related infections. Clin Infect Dis 2011;52:e162–93. 10.1093/cid/cir257 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Pronovost P, Needham D, Berenholtz S, et al. An intervention to decrease catheter-related bloodstream infections in the ICU. N Engl J Med 2006;355:2725–32. 10.1056/NEJMoa061115 [DOI] [PubMed] [Google Scholar]
  • 8. Chopra V, Anand S, Hickner A, et al. Risk of venous thromboembolism associated with peripherally inserted central catheters: a systematic review and meta-analysis. Lancet 2013;382:311–25. 10.1016/S0140-6736(13)60592-9 [DOI] [PubMed] [Google Scholar]
  • 9. Barczykowska E, Szwed-Kolińska M, Wróbel-Bania A, et al. The use of central venous lines in the treatment of chronically ill children. Adv Clin Exp Med 2014;23:1001–9. 10.17219/acem/37359 [DOI] [PubMed] [Google Scholar]
  • 10. Goudie A, Dynan L, Brady PW, et al. Attributable cost and length of stay for central line-associated bloodstream infections. Pediatrics 2014;133:e1525–32. 10.1542/peds.2013-3795 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Wilson MZ, Rafferty C, Deeter D, et al. Attributable costs of central line-associated bloodstream infections in a pediatric hematology/oncology population. Am J Infect Control 2014;42:1157–60. 10.1016/j.ajic.2014.07.025 [DOI] [PubMed] [Google Scholar]
  • 12. van Miert C, Hill R, Jones L. Interventions for restoring patency of occluded central venous catheter lumens. Cochrane Database Syst Rev 2012;4:Cd007119 10.1002/14651858.CD007119.pub2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Rey C, Alvarez F, De La Rua V, et al. Mechanical complications during central venous cannulations in pediatric patients. Intensive Care Med 2009;35:1438–43. 10.1007/s00134-009-1534-0 [DOI] [PubMed] [Google Scholar]
  • 14. Schwebel C, Lucet JC, Vesin A, et al. Economic evaluation of chlorhexidine-impregnated sponges for preventing catheter-related infections in critically ill adults in the Dressing Study. Crit Care Med 2012;40:11–17. 10.1097/CCM.0b013e31822f0604 [DOI] [PubMed] [Google Scholar]
  • 15. Soufir L, Timsit JF, Mahe C, et al. Attributable morbidity and mortality of catheter-related septicemia in critically ill patients: a matched, risk-adjusted, cohort study. Infect Control Hosp Epidemiol 1999;20:396–401. 10.1086/501639 [DOI] [PubMed] [Google Scholar]
  • 16. Chopra V, Flanders SA, Saint S, et al. The Michigan Appropriateness Guide for intravenous catheters (MAGIC): Results from a Multispecialty Panel using the RAND/UCLA Appropriateness Method. Ann Intern Med 2015;163(Suppl 6):S1–40. 10.7326/M15-0744 [DOI] [PubMed] [Google Scholar]
  • 17. Moureau NL, Trick N, Nifong T, et al. Vessel health and preservation (Part 1): a new evidence-based approach to vascular access selection and management. J Vasc Access 2012;13:351–6. 10.5301/jva.5000042 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Lau CS, Chamberlain RS. Ultrasound-guided central venous catheter placement increases success rates in pediatric patients: a meta-analysis. Pediatr Res 2016;80:178–84. 10.1038/pr.2016.74 [DOI] [PubMed] [Google Scholar]
  • 19. Gilbert RE, Mok Q, Dwan K, et al. Impregnated central venous catheters for prevention of bloodstream infection in children (the CATCH trial): a randomised controlled trial. Lancet 2016;387:1732–42. 10.1016/S0140-6736(16)00340-8 [DOI] [PubMed] [Google Scholar]
  • 20. Bradford NK, Edwards RM, Chan RJ. Heparin versus 0.9% sodium chloride intermittent flushing for the prevention of occlusion in long term central venous catheters in infants and children. Cochrane Database Syst Rev 2015;11:Cd010996. [DOI] [PubMed] [Google Scholar]
  • 21. Ullman AJ, Cooke ML, Mitchell M, et al. Dressings and securement devices for central venous catheters (CVC). Cochrane Database Syst Rev 2015;9:Cd010367. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Ullman AJ, Cooke ML, Gillies D, et al. Optimal timing for intravascular administration set replacement. Cochrane Database Syst Rev 2013;9:CD003588 10.1002/14651858.CD003588.pub3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Moureau NL, Flynn J. Disinfection of Needleless Connector Hubs: clinical evidence systematic review. Nurs Res Pract 2015;2015:1–20. 10.1155/2015/796762 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Queensland Health and Department of Education and Training. Nurses and Midwives Certified Agreement (EB9) Training QHaDoEa. Queensland: Queensland Government, 2016. [Google Scholar]
  • 25. Queensland Health and Department of Education and Training. Medical Officers' (Queensland Health) Certified Agreement (No.4) Queensland Health and Department of Education and training. Brisbane: Queensland Government, 2015. [Google Scholar]
  • 26. Health Q. Queensland Health Tender costings In: Health Q, ed Brisbane, Australia: queensland Government, 2016. [Google Scholar]
  • 27. Rickard CM, Edwards M, Spooner AJ, et al. A 4-arm randomized controlled pilot trial of innovative solutions for jugular central venous access device securement in 221 cardiac surgical patients. J Crit Care 2016;36:35–42. 10.1016/j.jcrc.2016.06.006 [DOI] [PubMed] [Google Scholar]

Articles from BMJ Case Reports are provided here courtesy of BMJ Publishing Group