Abstract
Reflection mode Terahertz (THz) imaging of corneal tissue water content (CTWC) is a proposed method for early, accurate detection and study of corneal diseases. Despite promising results from ex vivo and in vivo cornea studies, interpretation of the reflectivity data is confounded by the contact between corneal tissue and dielectric windows used to flatten the imaging field. Herein, we present an optical design for non-contact THz imaging of cornea. A beam scanning methodology performs angular, normal incidence sweeps of a focused beam over the corneal surface while keeping the source, detector, and patient stationary. A quasioptical analysis method is developed to analyze the theoretical resolution and imaging field intensity profile. These results are compared to the electric field distribution computed with a physical optics analysis code. Imaging experiments validate the optical theories behind the design and suggest that quasioptical methods are sufficient for designing of THz corneal imaging systems. Successful imaging operations support the feasibility of non-contact in vivo imaging. We believe that this optical system design will enable the first, clinically relevant, in vivo exploration of CTWC using THz technology.
Index Terms: Biological and medical imaging, THz imaging of cornea, medical diagnostics, clinical instruments
I. INTRODUCTION
Corneal disorders, such as Fuchs’ endothelial dystrophy [1], keratoconus [2], pseudophakic bullous keratopathy and graft rejection [3, 4], are characterized by increased corneal tissue water content (CTWC) and subsequent swelling of the cornea, leading to chronic vision impairment and often requiring surgical intervention. Corneal disorders affect large populations worldwide especially that of elderly [5]. It is believed that abnormal CTWC is a key clinical manifestation of endothelial malfunctions and corneal dystrophies [6–10]. Because abnormal CTWC is an important diagnostic target for assessing the extent of tissue damage in vivo [6–10], quantifying and tracking CTWC can (1) provide a better understanding of the formation, development, and progression of these disorders; and (2) become a clinically useful method for early diagnosis and assist in the choice and timing of interventions. However, accurate and non-invasive in vivo measurement of CTWC remains elusive.
Terahertz (THz) imaging is a promising method for the sensing and imaging of CTWC due to the homogeneity of bulk corneal tissue and relative lack in physiologic variations in corneal topography compared to other structures in the body. However, despite the acquisition of clinically relevant data from ex vivo and in vivo cornea, interpretation of the results has been confounded by contact between cornea and field flattening dielectric windows that are required in standard flat-field THz imaging methods [11, 12].
In this work, we describe reflective terahertz (THz) imaging optical principles that use a combination of planar and off-axis parabolic (OAP) mirrors to scan a beam at the normal incidence across the surface of cornea while keeping the source, detector, and patient stationary. This method acquires an image of a spherical surface with an arbitrary radius of curvature by an orthographic projection of the spherical surface to Cartesian coordinates. The design enables non-contact imaging of corneal reflectivity and avoids deformations of the cornea, which was a critical confounder in our previous work [11, 12].
The following sections introduce the design and simulation of scanning optics for non-contact CTWC imaging of human cornea. Imaging principles, quasioptical modeling, and physical optics modeling are presented. The quasioptical modeling treats the OAP mirror as an ensemble of thin lenses of varying effective focal lengths. The accuracy of this quasioptical technique is assessed with a physical optics simulation code. A prototype imaging system is presented and was used to evaluate the utility of the presented optical simulations.
II. Motivation for current work
In previous studies [11–13], we investigated the utility of THz and millimeter wave imaging and sensing to track changes in CTWC. A rabbit model study (n=5) was designed to mimic disease-relevant CTWC perturbations. A dhydration protocol was applied to one cornea of each animal and the subsequent, expected rehydration was observed over a 90-minute period. During this time window, central corneal thickness (CCT) measurements were acquired with an ultrasound pachymeter (current clinical standard), and the corresponding THz reflectivities were acquired with a 100 GHz (narrowband) reflectometer and ~525 GHz (broadband) imaging system. The protocol included the application of a 12μm thick Mylar window to gently flatten the cornea and provide a planar surface for both systems.
The experiment revealed a strong positive correlation between increasing CCT and increasing 100 GHz reflectivity; both consistent with the intended increase in CTWC. However, the 525 GHz imaging data did not demonstrate a statistically significant increase or decrease in reflectivity and produced a limited correlation with the millimeter wave and the CCT measurement [11, 12]. An electromagnetic modeling of this problem indicated that in vivo cornea is a lossy etalon at millimeter wave and THz frequencies and, thus, its spectral properties are a complex coupling of CTWC and CCT [11, 12]. Model based analysis suggested that the protocol most likely modified the thickness of the cornea while leaving the aggregate CTWC relatively unperturbed, thus resulting in a resolvable “etalon effect” at 100 GHz and an apparent absence of etalon at 525 GHz. It is likely that the contact pressure of the window altered the CCT while the stream of air produced negligible variation in CTWC.
These experiments confirmed weaknesses in the exploration of THz imaging for CWTC diagnostics with current techniques. First, active imaging of the corneal surface is difficult with simple x–y scanning techniques. The corneal surface is only ~10 mm in diameter en face and nearly spherical with a mean radius of curvature of ~ 8 mm [14]. This geometry constrains measurements to the central apex of the cornea when using conventional, planar raster-scanning [11]. Second, all current clinically accepted techniques of determining CTWC in vivo are based on thickness measurements that extrapolate CTWC from CCT [15]. Application of a dielectric field flattening window confounds CCT-to-CTWC mapping and perturb measured THz reflectivity. Considering these shortfalls, it is apparent that THz imaging of CTWC requires an improved THz imaging method that can acquire reflectivity maps of the cornea in vivo without contact.
To address these needs, we have designed, constructed, and characterized a system that acquires reflectivity maps of the cornea without contact. The system architecture was motivated by two key observations: (1) The cornea is nearly a perfect hypo hemisphere with respect to a THz wavelength and the person-to-person variations of corneal curvature are relatively small; (2) The en face diameter of the cornea is also relatively consistent across the adult population. The target curvature and field of view can be assumed a priori with a high degree of confidence. Therefore, corneal imaging is unique amongst all THz medical imaging applications and allows the system design to accommodate only a very limited set of use cases.
III. Spherical properties of the cornea
The anterior cornea (surface) is aspherical where the deviation from an ideal sphere increases toward the periphery [16]. Human cornea is also astigmatic, demonstrating different curvature and, hence, optical power as a function of cross-sectional meridian angles [17]. Characterization of corneal topology is important for interventions such as Laser-Assisted in situ Keratomileusis (LASIK) and Photorefractive Keratectomy (PRK). Topological maps can be obtained with Videokeratography [18] or Scheimpflug photography [19] techniques which interpret the surface height measurements in refractive power at each local surface. In particular, Videokeratography [20] can be used to construct dioptric power maps from corneal surface height data, which can be further fitted to a parametric surface to analyze higher-order surface features from disease or refractive surgery [21]. However, despite the maturity of topology mapping technology, there is no standardized method for analyzing topographic information [22]. Therefore, the following analysis is performed to characterize physiologic variations of anterior corneal surface.
Corneal surface data is decomposed to into a set of concentric rings centered at the corneal apex. Fourier decomposition analysis was utilized to explore the deviations between average human cornea topology and an ideal sphere [22, 23]. The fundamental mode (first term in the Fourier series expansion) is attributed to de-centration (Figure 1(a)), and the second term is attributed to “regular” astigmatism (Figure 1(b)). The RoC deviation data in Figure 2 represents the expected geometric variation from an ideal sphere. Higher order Fourier terms are grouped together as “higher order surface irregularities” and are shown to contribute much less to the overall corneal shape than either decentration or astigmatism individually. In general, the Fourier coefficients increase for equal height contour lines further from the corneal apex.
Figure 1.

Corneal anatomy. The anterior and posterior surfaces are represented by the curved solid and dotted lines respectively. (a) Decentration: local surfaces may have centers of curvature (CoC), that do not lie on the central optical axis. and (b) Astigmatism: Radial asymmetry results in and focal points from different axes lying at different depths along the optical axis.
Figure 2.

Average corneal RoC variation (Δr, equation (1)) for healthy cornea (left), Keratoconus (middle), and Grafted cornea (right). The contribution to overall RoC variation is decomposed into decentration, regular Astigmatism, and higher order irregularities [22]. The maximum RoC variation is less than 1λ @ 650 GHz.
In [22], corneal surface curvature measurements are reported as diopters (D) across the meridian cross-section as a function of radial distance from the center of the cornea and the meridian angle. The diopter values are averaged and converted to an expected RoC deviation using equation (1) where Ptypical is the focusing power of a cornea representative of typical population (Ptypical = 44D), Pmeasured is the data from [22], n is the refractive index of the cornea (n = 1.376), and Δr is the resulting standard deviation in the RoC.
| (1) |
The average RoC deviation from the first term in the series (decentration), the second term in the series (astigmatism) and the higher order terms are displayed in Figure 2 for healthy eyes (N=25), Keratoconic eyes (N=13), and grafted cornea (N=20) [22]. The format represents the expected geometric variation from an ideal sphere. Data for endothelial dystrophies, such as Fuchs, was not available during the preparation of this manuscript. However the expected deviation arising from dystrophies is expected to much less that Keratoconus thus the pathologies in Figure 2 serve as sufficient upper bound.
For normal, healthy eyes, decentration produces a mean RoC displacement deviation of ~0.05 mm, regular astigmatism is slightly higher at ~0.08 mm, and the sum total of higher modes contributes a negligible ~ 0.015 mm. Keratoconus, a condition where the cornea thickens and the CCT increases at a rate faster than the periphery [2, 24, 25], exhibits mean RoC displacement deviations up to ~0.4 mm. ~ Finally, for grafted cornea, where a patient’s diseased cornea has been removed and replaced with a donor cornea [4], the mean RoC displacement deviation arising for decentration, astigmatism, and higher order irregularities, are ~0.3 mm, ~0.35 mm, and ~0.08 mm, respectively.
It is instructive to view this RoC curvature variation with respect to a free space wavelength of 0.462 mm (650 GHz), the center wavelength of the system in section X (Figure 10). When normalized to illumination wavelength, the corneal radius of curvature is ~ 8 mm/0.462 mm = 17.32λ. Assume that the variations described by the decentration, astigmatism, and higher order irregularities are uncorrelated and that total standard deviation can be computed as the square root of the sum of squares of each individual standard deviation. Then, the resulting RoC variations are following: Normal, healthy cornea = 17.32λ ± 0.21λ, Keratoconus 17.32λ ± 1.16λ, and grafted cornea = 17.32λ ± 1.01λ, corresponding to coefficients of variance (100*σ/μ) of 1.21%, 6.69% and 5.83% respectively. Hence, with respect to the 650 GHz illumination, the corneal surface can be considered an ideal sphere with maximum expected deviations approximately one free space wavelength.
Figure 10.

Block diagram of the prototype system.
IV. Spherical Surface Imaging
A. Spherical Surface Scanning Principle
Imaging of a spherical surface is performed by positioning the CoC of the target (cornea) coincident with the focal point of an OAP mirror, and then transmitting a collimated illumination beam into the mirror clear aperture (CA), parallel to the CA normal (Figure 3).
Figure 3.

Mapping of transverse translation in the mirror CA plane to angular translation on the corneal surface. In (a) and (b), transverse locations R1 – R5 are mapped to angular scan locations θ1 – θ5. In (c) and (d), transverse scan locations d1 – d5 are mapped to angular scan locations ϕ1 – ϕ5. Note that the focused beam is intercepted prior to the focal point and that all beams are normal to the corneal surface.
The focused radiation is normal to the spherical surface and, in the limit of geometric optics, has a phase front curvature equal to the spherical surface RoC. The reflected, diverging beam is re-collimated by the OAP mirror and arrives coincident with the transmitted beam path. Modulating the transverse location of the collimated beam while maintaining a path parallel to the plane of mirror’s clear aperture sweeps the location of the illumination spot on the spherical surface, and an image can be constructed (Figure 3). The retro-directive nature of this arrangement is compatible with any transceiver design that can multiplex/demultiplex the input and output beams using, e.g., a wire grid, thin film, or polarizing beam splitter.
This beam scanning technique accomplishes spherical surface (θ,ϕ) scanning by the geometrical projection of the target’s spherical surface into the planar coordinate system (x,y) in the mirror clear aperture. This is a restatement of the Fourier-Transform property of an optical focusing element as demonstrated with a 90° off-axis parabolic mirror (OAP) in Figure 3 where transverse locations Rk and dk are mapped to angular locations θk and ϕk respectively. Unlike the Mercator projection which transforms the surface from spherical coordinates via a cylindrical projection to rectilinear coordinates [26], this method performs conformal mapping from spherical coordinates via projection by a paraboloid surface into a rectilinear coordinate system. Although this exact mapping operation is uncommon, it shares resemblance to the conic orthographic mapping projections in cartography, such as the Lambert Conformal Conic Projection [27].
B. Effective Imaging Optic f/# and offset
The mirror scanning solid angle can be characterized by the mirror f/# = fe/A = 2f0 A where fe is the effective focal length (EFL), f0 is the parent focal length (PFL, fe = 2f0 for a 90° OAP) and A is the CA diameter (Figure 4(a)). A human cornea spans ~60° (±30°) about its apex, thus an imaging optic should approach fe/A = 2tan(30°) ~ 1.15 to span ±30° in both azimuthal (Figure 3(c)) and elevation (Figure 3(a)) about the corneal apex.
Figure 4.

Half angles of 90° OAP mirrors: ϕ δ θ+ and θ-. (a) side view. (b) Clear aperture view. (c) Design space bordered by offset fe - A/2 = Rc and fe = 1.15A. Standard, commercially available mirrors indicated by [○] 25.4 mm, [□] 50.8 mm, and [Δ] 76.2 mm apertures.
Table I lists the azimuthal and elevation angles subtended by a 90° off-axis parabolic reflector, according to commonly available f/#s from vendors. Because the focusing geometry (Figure 4) is asymmetric, the azimuthal coverage angle (θ− + θ+ (Figure 4(a)) is not bisected by the 90° ray (i.e. θ− > θ+, ∀ fe, A). The 90° geometry results in the relation θ+ < δ < ϕ < θ− therefore θ+ forms the lower bound on the OAP f/#. The results in Table 1 and Figure 4 that a 25.4 mm PFL, 76.2 mm CA OAP (2·25.4/76.2 = f/0.66) mirror can scan the entire angular extent of the cornea.
Table I.
Spherical scanning angles (deg.) by imaging reflector f/#
| f/# (fe/A) | 0.66 | 1 | 2 | 3 |
|---|---|---|---|---|
| θ− | 62.8 | 36.9 | 16.3 | 10.4 |
| θ+ | 30.7 | 22.6 | 12.7 | 8.8 |
| ϕ | 44.5 | 29.0 | 14.4 | 9.6 |
| δ | 37.2 | 26.6 | 14.0 | 9.5 |
f/# is scale invariant and does not consider the size of the target relative to the size of the focusing objective. Practical considerations restrict combinations of focal length and clear aperture to pairs that avoids positioning the apex of the cornea inside the mirror. In other words, the “flange distance” should be greater than the corneal radius of curvature: fe – A/2 > Rc. The flange distance and f/# are displayed in Figure 4(c), where all combinations to the right of the fe – A/2 = Rc line yield sufficiently large offsets and all combinations to the left of the fe = 1.15A yield f/#s sufficient angular coverage. Markers representing standard, commercially available pairs of effective focal lengths and apertures are superimposed on this space with the 76.2 mm diameter mirrors indicated by the dotted contour. This work utilized a 76.2 mm diameter f/0.66 OAP.
V. Quasioptical Analysis
A method to implement the beam scanning described in Figure 3 utilizes a set of plane mirrors, labeled x-scan and y-scan, that translate a collimated input beam in two orthogonal directions parallel to the clear aperture plane of the OAP mirror. This concept is displayed in Figure 5 where the input and output beams are multiplexed/demultiplexed by a beam splitter that optically collocates the source and detector.
Figure 5.

Imaging system design. The raypath is plotted starting from the source, then traced to the scanning mirrors (only x- scanning shown), target, and back to the detector. If the target is spherical, all rays focus to a point on the detector, invariant to scan mirror position.
The optics are laid out in three groups which transmit and receive collimated beams: (1) Transceiver subsystem, (2) Scanning subsystem, and (3) Imaging mirror. The corneal CoC is coincident with the OAP mirror’s focal point and all rays of the focused beam are orthogonal to the spherical surface. Thus, all reflected rays, independent of scan mirror position will arrive at the detector feedhorn with identical convergence angle, transverse location, and extent as demonstrated with the three-superimposed beam paths in Figure 5 (ASAP, Breault Inc.).
A. Quasioptical Analysis Setup and modeling
The imaging mirror in Figure 5 is much larger than the transverse extent of the collimated input beam (A≫ω0). At any scan location, the beam only sees a local surface of the imaging OAP mirror, from here on referred to as sub-reflector. Therefore, the transformation of the THz beam, as a function of scan location can be approximated as the transformation by an optic whose properties represent the sub-reflector; mirror surface subtended by the beam waist. Given a reasonable input collimated beam size, the segments span moderate changes in curvature and focal distance and substantially reduced asymmetry about the beam centroid. These points suggest that the sub-reflector can be accurately modeled as a thin lens whose focal length is equal to the central beam path length from the mirror surface to the focus. This setup enables the use of quasioptical analysis with Gaussian optics and ABCD matrices to compute the resulting beam transformations as a function of scan location and, thus, the image transfer function of the imaging OAP mirror [28].
The Gaussian beam setup is displayed in Figure 6(a), where the shaded areas represents the cross sections of a cornea (8 mm RoC) and a 76.2 mm clear aperture, 25.4 mm PFL, 90° OAP mirror. Three parallel ray bundles are normal to the clear aperture plane and the corresponding sub-reflectors are represented by the darker shaded areas. A Gaussian beam can be traced through each sub-reflector using (1) the distance of free space travel between a reference plane and the mirror surface height defined by the intersection of the parabolic profile and collimated beam centroid, and (2) the effective focal length of the sub reflector defined by the line segment connecting the intersection in (1) and the effective focal point of the parent OAP mirror. This concept yields a collimated path length, focused path length, and effective focal length that are all dependent on the lateral distance (Rs) between the mirror focal point and collimated beam centroid (horizontal axis in Figure 6(a)). The problem is recast to that described by Figure 6(c). To complete the problem description, a reference plane is defined at a distance d0 from the upper tip of the OAP represented by the dotted line in Figure 6(a). This reference plane is where the collimated beam (plane coincident with the beam waist) was launched. The collimated beam travels a distance d(Rs) + d0, is focused by a thin lens of fe(Rs) and then travels fe(Rs) – Rc to a convex reflector of radius Rc (cornea). The mirror and corneal geometry ensure normal incidence for all Rs within the clear aperture of the mirror, resulting in the beam retracing its path back through the mirror collinear with the incidence path.
Figure 6.

Raypath diagram of the beam scanning. (a) parabolic mirror segmentation location, (b) definition of reference plane tangent to the corneal surface at the intersection of beam centroid and cornea. The focused spot size was characterized on these reference planes. This is where spot size on the surface of the cornea. (c) Thin lens equivalent of the overall beam path.
| (2) |
| (3) |
Expressions for the radially dependent focal length and beam clear aperture path length are defined in equations (2) and (3). Note that these expressions are defined for the domain RS ε 2fo ± A/2 and are explicitly for a 90° OAP.
| (4) |
| (5) |
| (6) |
| (7) |
| (8) |
| (9) |
M1 = MP→C (equations (4), (7)) is the transfer matrix describing the propagation of the beam from the reference plane (P) to the cornea (C). M2 = MCMP→C (Equations (6), (8)) describes the transformation of the beam by M1 and the subsequent reflection from the spherical reflector. The overall retrodirective transmission through the optical system is described by M3 = MC→PMCMP→C (equations (6), (9)), where the beam starts and stops at the reference plane.
| (10) |
| (11) |
| (12) |
The standard complex beam parameter relations listed in equations (10) – (12) were used to compute the spot size on target (M1), the beam radius of curvature immediately prior to (M1) and following (M2) reflection from the cornea, and the coupling coefficient between the input and output beams at the reference plane (M3). These metrics were calculated for all relevant Rs with input complex beam parameter q0 parameterized by ω0 and a constant R0 → ∞.
B. Spot radius on the corneal surface
The spot size on the cornea was explored as a function of input beam waist size at the reference plane and the input, collimated beam location described by its radial distance (Rs) from the focal point of the OAP mirror using equations (7) and (10) – (13). These equations characterize the spot size on a plane tangent to the corneal surface and normal to the beam centroid as depicted in Figure 6(b). Due the relatively small extent of the focused beam with respect to the corneal RoC, the spot size computed on a plane was considered a close representation of projecting the beam on to the corneal surface and assessing the extent of the intersection contour.
| (13) |
The input Gaussian beam radius (ω0) at the reference plane was varied from 4 mm to 12 mm and the resulting output spot radius as a function of input radius and mirror scan radius (Rs) are superimposed on the shaded outline profile of a 76.2 mm CA, 25.4 mm PFL, 90° OAP mirror length for reference. Note the dotted line style at the extreme ends of each curve. The exterior points on both ends of the curves represent the edge of the mirror CA. The interior points, located closer to the mirror axis, represent one input beam radius (ω0) from the edge of the mirror. In practice, diffractive effects may lead to a divergence between quasioptical analysis and realized performance for beam centroid locations outside the solid line intervals.
The 4 mm input spot size demonstrates a decreasing focused spot size ω1 for decreasing scam radius from Rs = 85.05 mm down to Rs = 35.02 mm corresponding to a beam radii of 3.36 mm 1.96 mm respectively. The focused beam radii then increase monotonically to 2.09 mm at the near edge of the mirror (Rs = 15.78 mm). For small radii input beams, the combination of the sub-reflector focal length and center wavelength, results in a beam waist that occurs “outside” of the cornea for large Rs, thus the beam is diverging prior to incidence on the corneal surface. As the beam f# is decreased (decreasing Rs) and the focused path length decreases (decreasing Rs), the beam waist decreases and its location converges to the surface of the cornea. Further decreases in Rs move the beam waist to “inside” the cornea and is concomitant with an increasing convergence angle. While the focused waist continues to decrease in size, the corneal surface intercepts the beam at a distance that is increasing from the location of the waist.
This behavior is somewhat reversed for the 12 mm input radius. The spot size on the corneal surface reaches its maximum at the shortest scan radius (shortest effective focal length), Rs = 24.25. mm, ω1 = 3.1 mm., and its minimum at the largest scan radius, Rs = 76.58. mm, ω1 = 1.5 mm. Unlike the 4 mm input beam, the 12 mm input beam produces a waist that is “inside” the cornea for all Rs, and the offset between the sub-reflector focal point and waist location is small with respect to radius of the cornea (Rc) for nearly all Rs. Since the input beam is large, the effective beam f# is small and leads to large convergence angles and thus large spot sizes on the corneal surface which is located a large number of Rayleigh lengths from the focal point (i.e. 8 mm >> zR).
The intermediate input spot sizes demonstrate varying dependencies on the aforementioned factors, and it is clear by inspection of Figure 7(a) that an intermediary balances the following three competing factors:
1. Decreasing Rs decreases the effective focal length of the sub-reflector and, thus, decreases the size of the focused beam waist.
2. Decreasing Rs increases the convergence angle of the beam and can increase the beam spot size on the corneal surface depending on beam input parameters.
3. Decreasing the input beam waist decreases the convergence angle while increasing the offset between sub-reflector focal point (corneal center of curvature) and focused beam waist. This can result in an increase or decrease in the spot size on the corneal surface depending on the f/# of the OAP, the input diameter of the beam, and the wavelength.
Figure 7.

Quasioptical and physical optics computations of relevant beam parameters. The physical optics results are discussed in section VI. (a) Focused spot radius on the analysis plane (Figure 6(b)) as a function of the input beam location (Rs) over the parabolic profile of the imaging OAP mirror. Each curve represents different collimated beam radii (ω0) ranging from 4 mm – 12 mm. A scaled overlay of parabolic cross section is shown to depict the location of the beam. (b) Coupling coefficient between the input and output beam as a function of input beam location over the parabolic profile of the imaging OAP mirror. (c) Differential between quasioptics and physical optics computations for ω1. (d) Differential between quasioptics and physical optics computations for K3. Radius of curvature of the beam immediately on the target surface (e) and immediately after reflection (f), as a function of the input beam radial location over the parabolic profile of the imaging OAP mirror. Strong agreement was observed between the quasioptical analysis and physical optics computations. Maximum deviation occurs for larger input diameters near the edge of the mirror indicating diffraction.
The results in Figure 7 demonstrate a tradeoff relationship between input (ω0) and focused (ω1) beam radii and scan radius (Rs) which arises from characterizing the focused beam at a distance Rc>0 from the geometric focal point; a factor unique to corneal imaging.
C. Beam Coupling efficiency
The 2D beam coupling coefficient beam is defined in equation (15) where the relationship between q0 and q3 is defined in equation (14) and constructed with M3 from equation (9). The coupling coefficient was calculated at the reference plane and is written explicitly in terms of the input and output spot sizes and radii of curvature, which together define their respective complex beam parameters. Note that equation (15) assumes infinite reflector transverse extent and does not account for diffraction.
| (14) |
| (15) |
The coupling coefficient as a function of mirror position and input beam diameter are displayed in Figure 7(b), which demonstrate the expected relation between input and output beam matching. As the beam input radius increases and the sub-reflector effective focal length decreases (decreasing Rs) the RoC matching between the focused beam and cornea improves, resulting in a reduced perturbation of the illumination beam and increased matching at the reference plane. The 12 mm beam ranges from 0.57 – 0.95 and the 4 mm beam ranges from 0.30 – 0.03. These results confirm a monotonically decreasing coupling efficiency for increasing input radius, independent of the input beam radius.
The source and detector mirrors in Figure 5 were oriented, with respect to the beam splitter, to maintain an “ortho” configuration [29, 30]. Similarly, the beam bath through the focused mirror was retro directive thus its beam transformation can be described as the output of two identical OAP mirrors also oriented in the “orthro” configuration. Finally, the beam patterns of the source and detector were considered identical to reflect the 26 dB feedhorns used in Section X. Due to the optical system symmetry, the similarly of the transmit and received beams at the beam splitter (reference plane) was deemed a sufficient estimate of the coupling efficiency between the detector pattern and the transformed source pattern.
D. Beam RoC and Coupling
The RoCs for pre (R1) and post (R2) corneal reflection are displayed in Figure 7(e) and (f) respectively and were computed with equation (16). The pre-reflection RoC are consistent with the observed behavior in the spot size characterization shown in Figure 7(a).
| (16) |
The RoCs resulting from the smaller input spot sizes are asymptotic and rapidly diverge to −∞ for larger mirror Rs and then should approach, from +∞, +1/Rs. Note that in Figure 7(e) only the negative RoCs are plotted and asymptotes/asymptotic behavior were omitted for clarity. As the input spot size is increased the beam waist decreases and its location converges to the corneal CoC. While most of the focused beam RoCs appear to be asymptotic, only the RoCs computed from the 4 mm, 5 mm and 6 mm beams flip sign. These occur at Rs = 68.17 mm, Rs = 75.68 mm, and Rs = 82.26 mm for the ω0 = 4mm, ω0 = 5mm, and ω0 = 6 mm respectively. All other considered beam radii maintain focused waist location “inside the cornea” for all Rs.
The reflected beam RoC demonstrates significantly less variation than the pre-reflection RoC as a function of mirror radius Rs and exhibit positive radii (divergence) for all input beam radii and mirror scan locations. Further, the high, empirical correlation of the trends with the computed coupling coefficients suggest that the post reflection RoC is the primary determinant of the coupling efficiency in the presented optics configuration.
E. Observations
The quasioptical design and analysis space for corneal imaging is unique because (1) the target is spherical, leading to increased divergence in reflection compared to the canonical flat target and (2) the target surface is NOT located at the focal point defined by the optics or beam calculations, leading to complex beam dynamics on target.
The results suggest that for a fixed input beam radius, and optimal alignment, the signal from the near edge (small Rs) of the mirror will always be higher than the far edge of the mirror (large Rs). These observations also suggest that the imaging field may be homogenized in terms of beam radius and coupling efficiency if the input beam radius is allowed to vary as a function of Rs. Finally, the results indicate that the minimum spot size on target does not coincide with the colocation of beam waist and corneal surface for the range of parameters examined.
Note that the analysis method used in this section can be extended to remove thin lens approximation treatment of the mirror surface segment. Any propagated beam can be decomposed to include higher-order Hermite-Gaussian or Laguerre-Gaussian modes, and an augmented ray-transfer method can be applied to beam propagation, thereby accounting for the asymmetric geometry of the mirror segment [31–33].
VI. Physical Optics Analysis
The quasioptical analysis of beam propagation was compared to computations by GRASP (Ticra Inc., Copenhagen, Denmark), a physical optics code typically used in reflector antenna design [34–36]. The cornea was modeled as dielectric half sphere with a refractive index of n = 1.376 and a radius of 8 mm. The quasi-optical system along with the sphere are simulated by a Gaussian beam source polarized along x. The source is located at the input beam plane with a vertical distance of 74.6 mm above the center of the sphere.
A. Spot size
The focused electric field distribution was sampled in a transverse/tangent plane, located at the intersection of the beam centroid and corneal surface, for every input collimated beam radius (ω0) and location (Rs) evaluated in Figure 7. These planes are described by the x’, and z’ axes in the Figure 8(a),(f) and are defined by the rotation of the reference frame (x-y-z) about the y-axis. The input beam was TM polarized (E-field aligned with the x-z or x’-z’ planes) to match the source in section X.
Figure 8.

(a)Layout for ω0 = 4 mm and R = 76.9 mm. (b) cross section of beam in the x’-z’ plane. section (c) beam profile on the y’-z’ plane tangent to the cornea and coincident with the intersection of the focused beam centroid and corneal surface. Electric field magnitude (d) and phase (e) of the beam at the reference plane (x-y plane) following reflection from the cornea and recollimation from the OAP. (f) Layout for ω0 = 12 mm and R = 24.7 mm. (g)-(j) same as (b)-(e). Note that in configuration (a) the fields are nearly Gaussian while in configuration (f) they are not due to the offset reflector configuration and possible diffraction. Blue correspond to lower E-field amplitude and yellow corresponds to higher.
An example of the focused beam for the input parameters ω0 = 4 mm, Rs = 76.9 mm is displayed in Figure 8(a) – (c) in dB scale with accompanying scale bars indicating FOV dimensions. The E-field distribution in the x’-z’ plane confirms that the focused fields is nearly Gaussian. This distribution shape was also observed on the tangent target plane (y’-z’) which demonstrates vanishingly small radial asymmetry (ellipticity ~ 1) and an apparent lack of cross polarization. The predicted short Rayleigh length at ω0 = 4 mm was observed and inspection of Figure 8 (b) confirms the focused beam waist is located prior to the surface of the cornea. The nearly Gaussian field distribution was attributed to the negligible effect of reflector geometry, as the local curvature of the OAP segment, subtended by a 4 mm radius beam, for large Rs, was nearly symmetric about the beam centroid.
A complementary example of a large input beam radius (ω0 = 12 mm) paired with a short scan radius (Rs = 24.7 mm) is displayed in Figure 8(f)-(h). The increased beam radius and mirror curvature result in the focused beam waist lying nearly coincident with the corneal CoC/OAP focal point which is visible in the x’-z’ cut in Figure 8(g). Increased diffraction and the effects of increased mirror asymmetry about the beam centroid are also apparent and manifest as multiple local extremum in the field distributions of Figure 8(g)-(h) and significant radial asymmetry in Figure 8(h).
The average spot size on target (x’-y’ tangent plane) was computed numerically by obtaining the modulus of the E-field, finding the location of the peak amplitude, computing the 1/e closed contour of the modulus, and then computing the average radius of the 1/e contour. These results are superimposed with a square marker (□) in Figure 7(a) and demonstrate good agreement between quasioptical analysis and physical optics. The level of fit is further explored in Figure 7(b) which reports the difference between quasioptical radius (ω1,q) and physical optics average radius (ω1,p): ω 1,q – ω 1,p. The spot size differential further substantiates the general correlation between decreasing spot size radius and improved goodness of fit. Additionally, the graph elucidates the increased effects of diffraction and offset reflector configuration for larger input radii (ω0) and smaller scan radii (Rs).
B. Coupling
The electric field of the input and return beams at the reference plane were sampled for every pair of ω0 and Rs. The magnitude of the reflected beam for the ω0 = 4 mm, Rs = 76.9 mm pair demonstrates a significant increase in main lobe extent compared to the initial 4 mm. This corroborates with the results obtained with quasioptical analysis that indicate substantial beam divergence over the optical path which manifests in a broad re-collimated beam radius (Figure 8(c,h)). Additionally, the multiple extremum in the wrapped phase plot (Figure 8 (e),(j)) suggests that the reference plane is located approximately one Rayleigh length from the mirror surface corroborating a curved phased front and thus appreciable beam divergence.
The converse is true for the ω0 = 12 mm, R = 24.7 mm beam which demonstrates a transverse extent at the reference plane only slight larger than the initial beam. The phase plot is also more uniform with extrema spaced farther apart. These plots indicate that the reference plane is likely well within one collimated beam Rayleigh length and strongly corroborate with the quasioptical analysis.
| (17) |
The coefficient quantifies the field correlation between the incident electric field at the input beam plane, , and the field reflected from the cornea through the OAP up to the same input plane, This coefficient was computed with equation (17) which references the square of the inner product of the vector fields with the product of the total energy in each field. The complex vector inner product accounts for mismatch between amplitude, phase, and polarization. The coupling coefficient computations were superimposed with a square marker (□) in Figure 7(b) and again demonstrate good agreement between quasitopical methods and physical optics. The differential between the quasioptical coupling coefficient (K3,p) and the physical optics coupling coefficient (K3,q): K3,p – K3,q is plotted in Figure 7(d) and reveals a maximum deviation of ~ 6.5 % at ω0 ~ 8 mm, R ~ 76.9 mm. The differentials for larger input beam radii are not monotonic across the scan range and this variation is likely the result of edge diffraction and cross polarization.
IX. Beam Radius Optimization
Inspection of Figure 7(a) and the tradeoffs listed in section V.B motivate the optimization of input spot size subject to a specific resolution criterion. Two candidate metrics were considered in the following equations where ω0 is the input collimated beam radius and ω1 is the focused beam radius on the cornea. Rs,1 = 2f0 – A/2 + ω0 is the “unclipped” scan radius corresponding to the near edge of the OAP, and Rs,2 = 2f0 + A/2 – ω0 is the “unclipped” scan radius corresponding to the far edge of the OAP mirror.
| (18) |
| (19) |
Equation (18) chooses input beam diameter ω0 such that the resulting spot sizes on the target at the extremums of the scan range [Rs,1, Rs,2] are equal. The solution to equation (18) corresponds to an approximate colocation of the smallest focused spot size with the apex of the cornea, while trying to minimize the asymmetry of the spot size about the corneal apex. Equation (19) chooses a collimated beam radius where the average focused spot size at the corneal surface is minimized. The parameter space and solutions to these equations are displayed in Figure 9 using the quasioptical along with physical optics results of Figure 7(a). The cross over point of the two curves that satisfy equation (18) is denoted with the gray circle (○) marker and occurs at a collimated input waist of 6.25 mm.
Figure 9.

Optimization space for equations (18) and (19). The intersection of the near (red) and far (blue) edge traces is the solution to equation (18). The black curve represents the argument of the derivative in equation (19) and the black circular (○) marker is the solution of equation (19). The square (□) markers were computed from the physical optics simulations.
The input spot size dependent behavior of the solution of the differential in equation (19) is denoted with the solid black line in Figure 9 and its minimum, which satisfies (19), is denoted by the circular marker (○) at ω0 = 7.6 mm. The physical optics derived solution is plotted with the square (□) marker. The results demonstrated good agreement and identified an optimal input radius of ω0 ~ 7.0 mm.
For current clinical practice, the corneal center provides most utility when identifying disease processes, thus justifying the metric in equation (18) [37]. However, certain corneal pathologies, e.g. corneal graft rejection, can create “edematous fronts” that can start from anywhere in the periphery and migrate across the extent of the corneal surface [4]. Detection of these features supports minimizing the spatial bias of imaging system, thus justifying the metric in (19). Note that with the optimized ω0 = 7.6 mm the focused beam waist is never coincident with the corneal surface.
X. Experimental Validation of Optical Theory
A prototype system was assembled to explore the accuracy of the quasioptical theory and physical optics simulations. The coupling coefficient was chosen as a representative test because it is dependent on four parameters: (1) input radius of curvature, (2) output radius of curvature, (3) input beam radius, and (4) output beam radius.
The matching between theoretical and experimental coupling coefficient was explored by imaging a 7.93 mm radius brass sphere. Given the lack of contrast in the target, any spatial variation in signal was attributed to the predicted, spatially varying coupling efficiency.
A. System
A prototype system, based on the optics of Figure 5, was constructed with a solid-state frequency-modulated continuous wave THz source (Amplifier-multiplier chain, Virginia Diodes, Virginia) centered at 650 GHz. The detector was a WR1.5 waveguide mounted Schottky diode detector (ZBD) (Virginia Diodes, VA) with a 500GHz-700GHz detection bandwidth. Both the source and detector were coupled to diagonal feedhorn antennas with 26 dB of gain and aperture dimensions of 2.4 mm × 2.4 mm. To mitigate standing waves, the output was frequency modulated over ~2 GHz at a rate of 100 kHz, producing an RF bandwidth that exceeded the expected full etalon period of the optical path. The source was also amplitude modulated at ~900 Hz, and the rectified signal from the ZBD was detected with a Lock-in amplifier (Stanford Research Systems, CA) using an integration time of 3 milliseconds (ms). These parameters are summarized in the block diagram of Figure 10.
The optical layout of the system (Figure 5) used 25.4 mm PFL, 76.2 mm CA, 90° OAP mirrors to collimate radiation from the multiplier chain and focus reflected radiation into the detector aperture. The combination of feedhorn directivity and OAP PFL yielded a collimated spot 1/e field radius of ~ 10 mm as measured with a knife edge target. The imaging mirror was a 76.2 mm CA, 25.4 mm PFL OAP. Beam scanning was performed with two 50.8 mm diameter gold coated plane mirrors.
B. Image and Coupling Coefficient Fits
A THz image of the brass sphere is displayed in the inset of Figure 11 and was acquired by translating the plane mirrors with 5 mm steps. Note that this data is a mapping of the reflectivity on the sphere to the transverse coordinates of the clear aperture and thus the data was directly compared to the simulation results in Figure 7 and Figure 8. The image was masked with a circle of radius A/2-ω0 to exclude aberrations due to beam clipping.
Figure 11.

Coupling calibration target with data fit to equation (15).
The inset image in Figure 11 shows a general trend of decreasing signal starting from the near edge of the mirror aperture and continuing, monotonically, to the far edge (decreasing signal for increasing scan radius). A localized increase in signal is apparent in the bottom left corner of the image, which is likely due to beam diffraction. A contour line was superimposed on the aperture diameter. The arrows on the contour line corresponds to increasing scan radius and the pixels sampled by this profile are plotted with a circular line style (○) in Figure 11.
The quasioptical (equation (15)) and physical optics simulations for an input beam radius of 10 mm are superimposed on the data in a solid line style (–) and square marker (□) respectively (obtained directly from Figure 7). The experimental data and simulation results demonstrate good agreement across the center of the masked aperture of the mirror. The minimum deviation occurs at the mirror center (Rs = 50.8 mm) while the maximum deviation occurs towards the mirror near edge (Rs = 28 mm). This indicates that diffraction may contribute to observed contrast beyond A/2-ω0 and supports the exploration of smaller mask radii.
XI. Conclusions
This work presents a novel imaging principle and analysis method for normal incidence, non-contact THz imaging of spherical targets with applications to corneal diagnostics. An analysis of the variation in the geometric properties of the cornea was presented. It was shown that when referenced to a THz wavelength the cornea can be considered an ideal sphere; even under perturbations from corneal diseases and the effects of surgical interventions. This analysis motivated the design of a corneal imaging system that places the corneal CoC coincident with the focal point of a low f/# OAP mirror. Transverse sweeping of a collimated beam in the clear aperture plane produces angular scanning of a focused beam along the surface of the cornea.
Quasioptical techniques were introduced that segment parabolic mirrors into an ensemble of thin lenses where the effective focal length and free space beam paths are proportional to the distance between the parent focal point and mirror clear aperture centroid. The spot size on target, radius of curvature pre and post reflection, and coupling coefficient between transmit and received beams were simulated for a single OAP reflector.
The spot size on target and quasioptical coupling coefficient were also computed with full wave physical optics and good agreement between the physical optics results and the quasioptical results was demonstrated. The imaging mirror was over dimensioned (A>> ω0) therefore diffractive effects were limited for most of the considered input parameter space. The concurrence of the results strongly supports the use of quasioptical techniques as a design and analysis tool.
An imaging system based on Figure 5 was constructed and used to acquired images of a brass ball target with an RoC matching that of cornea. A profiles of intensity variation along the vertical diameter of the image was compared with the quasioptical and physical optics theory and good agreement was observed between image contrast and predicted coupling coefficients.
The limited variation of the shape and size of cornea among adult population is unique amongst all structures in the body and, when referenced to the typical center wavelength of a THz imaging system, even more negligible. We believe that the presented systems and analysis techniques provide a good first step towards the in vivo translation of this technology to human subjects.
Acknowledgments
This work is supported by the National Eye Institute (NEI) Grant# 5R01EY021590. The authors would like to thank Mr. Lu Han at Taylor and Francis for facilitating the discussions that lead to this research.
This work was supported in part by the National Eye institute under Grant Number 5R01EY021590.
Contributor Information
Shijun Sung, UCLA Dept. of Electrical Engineering, Los Angeles, CA 90095.
Shahab Dabironezare, Center for Wireless Systems and Technology, TU Delft, Netherlands.
Nuria Llombart, Center for Wireless Systems and Technology, TU Delft, Netherlands.
Skyler Selvin, UCLA Dept. of Electrical Engineering, Los Angeles, CA 90095.
Neha Bajwa, UCLA Dept. of Bioengineering, Los Angeles, CA 90095 USA.
Somporn Chantra, UCLA Dept. of Ophthalmology, Los Angeles, CA 90095 USA.
Bryan Nowroozi, UCLA Dept. of Bioengineering, Los Angeles, CA 90095; Mimeo Labs Inc, Santa Monica, CA 90404 USA.
James Garritano, UCLA Dept. of Bioengineering, Los Angeles, CA 90095 USA.
Jacob Goell, UCLA Dept. of Bioengineering, Los Angeles, CA 90095 USA.
Alex Li, UCLA Dept. of Bioengineering, Los Angeles, CA 90095 USA.
Sophie X. Deng, UCLA Dept. of Ophthalmology, Los Angeles, CA 90095 USA
Elliott Brown, Wright State University Dept. of Electrical Engineering, Dayton, OH 45435 USA.
Warren S. Grundfest, UCLA Dept. of Electrical Engineering, Los Angeles, CA 90095 UCLA Dept. of Bioengineering, Los Angeles, CA 90095 USA.
Zachary D. Taylor, UCLA Dept. of Electrical Engineering, Los Angeles, CA 90095 UCLA Dept. of Bioengineering, Los Angeles, CA 90095 USA.
References
- 1.Adamis AP, Filatov V, Tripathi BJ, Tripathi RAMC. Spotlight on ExhibitorsFuchs’ endothelial dystrophy of the cornea. Survey of Ophthalmology. 1993;38:149–168. doi: 10.1016/0039-6257(93)90099-s. 1993/09/01. [DOI] [PubMed] [Google Scholar]
- 2.Rabinowitz YS. Keratoconus. Survey of Ophthalmology. 1998;42:297–319. doi: 10.1016/s0039-6257(97)00119-7. 1// [DOI] [PubMed] [Google Scholar]
- 3.Taylor DM, Atlas BF, Romanchuk KG, Stern AL. Pseudophakic Bullous Keratopathy. Ophthalmology. 1983;90:19–24. doi: 10.1016/s0161-6420(83)34607-8. 1// [DOI] [PubMed] [Google Scholar]
- 4.Panda A, Vanathi M, Kumar A, Dash Y, Priya S. Corneal Graft Rejection. Survey of Ophthalmology. 2007;52:375–396. doi: 10.1016/j.survophthal.2007.04.008. 7// [DOI] [PubMed] [Google Scholar]
- 5.Whitcher JP, Srinivasan M, Upadhyay MP. Corneal blindness: a global perspective. Bulletin of the World Health Organization. 2001;79:214–221. [PMC free article] [PubMed] [Google Scholar]
- 6.Waring GO, III, Stulting R, Street D. Penetrating keratoplasty for pseudophakic corneal edema with exchange of intraocular lenses. Archives of Ophthalmology. 1987;105:58–62. doi: 10.1001/archopht.1987.01060010064032. [DOI] [PubMed] [Google Scholar]
- 7.Ytteborg J, Dohlman CH. Corneal edema and intraocular pressure: Ii. clinical results. Archives of Ophthalmology. 1965;74:477–484. doi: 10.1001/archopht.1965.00970040479008. [DOI] [PubMed] [Google Scholar]
- 8.Waring GO, Rodrigues MM, Laibson PR. Corneal dystrophies. II. Endothelial dystrophies. Survey of Ophthalmology. 1978;23:147–168. doi: 10.1016/0039-6257(78)90151-0. 1978/11/01. [DOI] [PubMed] [Google Scholar]
- 9.Rodrigues MM, Krachmer JH, Hackett J, Gaskins R, Halkias A. Fuchs’ Corneal Dystrophy. Ophthalmology. 1986;93:789–796. 1986/06/01. [PubMed] [Google Scholar]
- 10.Krachmer JH, Purcell JJ, Jr, Young CW, Bucher KD. Corneal endothelial dystrophy: A study of 64 families. Archives of Ophthalmology. 1978;96:2036–2039. doi: 10.1001/archopht.1978.03910060424004. [DOI] [PubMed] [Google Scholar]
- 11.Taylor ZD, Garritano J, Shijun S, Bajwa N, Bennett DB, Nowroozi B, Tewari P, Sayre JW, Hubschman JP, Deng SX, Brown ER, Grundfest WS. THz and mm-Wave Sensing of Corneal Tissue Water Content: In Vivo Sensing and Imaging Results. Terahertz Science and Technology, IEEE Transactions on. 2015;5:184–196. doi: 10.1109/TTHZ.2015.2392628. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Taylor ZD, Garritano J, Sung S, Bajwa N, Bennett DB, Nowroozi B, Tewari P, Sayre J, Hubschman JP, Deng S, Brown ER, Grundfest WS. THz and mm-Wave Sensing of Corneal Tissue Water Content: Electromagnetic Modeling and Analysis. Terahertz Science and Technology, IEEE Transactions on. 2015;5:170–183. doi: 10.1109/TTHZ.2015.2392619. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Bennett DB, Taylor ZD, Sassoon D, Tewari P, Johnson RD, Singh R, Culjat MO, Hubschman J-P, Grundfest WS. Terahertz sensing in corneal tissues. J Biomed Opt. 16:8. doi: 10.1117/1.3575168. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Liu Z, Huang AJ, Pflugfelder SC. Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system. British Journal of Ophthalmology. 1999;83:774–778. doi: 10.1136/bjo.83.7.774. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Li Y, Shekhar R, Huang D. Corneal Pachymetry Mapping with High-speed Optical Coherence Tomography. Ophthalmology. 2006;113:792–799.e2. doi: 10.1016/j.ophtha.2006.01.048. 5// [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Gullstrand A. Appendix. In: Helmholtz HV, editor. Handbuch der physiologischen Optik. Vol. 1. New York, Dover: pp. 351–352. [Google Scholar]
- 17.Srivannaboon S, Reinstein DZ, Sutton HFS, Holland SP. Accuracy of Orbscan total optical power maps in detecting refractive change after myopic laser in situ keratomileusis. Journal of Cataract & Refractive Surgery. 1999;25:1596–1599. doi: 10.1016/s0886-3350(99)00286-2. 12// [DOI] [PubMed] [Google Scholar]
- 18.Bogan SJ, Waring GO, III, Ibrahim O, Drews C, Curtis L. Classification of normal corneal topography based on computer-assisted videokeratography. Archives of Ophthalmology. 1990;108:945–949. doi: 10.1001/archopht.1990.01070090047037. [DOI] [PubMed] [Google Scholar]
- 19.Hockwin O, Weigelin E, Laser H, Dragomirescu V. Biometry of the Anterior Eye Segment by Scheimpflug Photography. Ophthalmic Research. 1983;15:102–108. doi: 10.1159/000265242. [DOI] [PubMed] [Google Scholar]
- 20.Applegate RA, Howland HC. Noninvasive measurement of corneal topography. IEEE Engineering in Medicine and Biology Magazine. 1995;14:30–42. [Google Scholar]
- 21.SCHWIEGERLING J, GREIVENKAMP JE. Using Corneal Height Maps and Polynomial Decomposition to Determine Corneal Aberrations. Optometry & Vision Science. 1997;74:906–916. doi: 10.1097/00006324-199711000-00024. [DOI] [PubMed] [Google Scholar]
- 22.Hjortdal JØ, Erdmann L, Bek T. Fourier analysis of video-keratographic data. A tool for separation of spherical, regular astigmatic and irregular astigmatic corneal power components. Ophthalmic and Physiological Optics. 1995;15:171–185. [PubMed] [Google Scholar]
- 23.RAASCH TW. Corneal Topography and Irregular Astigmatism. Optometry & Vision Science. 1995;72:809–815. doi: 10.1097/00006324-199511000-00006. [DOI] [PubMed] [Google Scholar]
- 24.Krachmer JH, Feder RS, Belin MW. Keratoconus and related noninflammatory corneal thinning disorders. Survey of Ophthalmology. 1984;28:293–322. doi: 10.1016/0039-6257(84)90094-8. 1// [DOI] [PubMed] [Google Scholar]
- 25.GROMACKI SJ, BARR JT. Central and Peripheral Corneal Thickness in Keratoconus and Normal Patient Groups. Optometry & Vision Science. 1994;71:437–441. doi: 10.1097/00006324-199407000-00003. [DOI] [PubMed] [Google Scholar]
- 26.Lee LP. THE TRANSVERSE MERCATOR PROJECTION OF THE ENTIRE SPHEROID. Empire Survey Review. 1962;16:208–217. 1962/01/01. [Google Scholar]
- 27.Snyder JP. Map projections used by the U.S. Geological Survey. USGS Numbered Series. 1982;1532:101. [Google Scholar]
- 28.Goldsmith P. Gaussian Beam Quasioptical Propogation and Applications Piscataway. NJ: IEEE Press; 1998. [Google Scholar]
- 29.Murphy JA. Distortion of a simple Gaussian beam on reflection from off-axis ellipsoidal mirrors. International Journal of Infrared and Millimeter Waves. 1987;8:1165–1187. 1987/09/01. [Google Scholar]
- 30.Malone RM, Becker SA, Dolan DH, Hacking RG, Hickman RJ, Kaufman MI, Stevens GD, Turley WD. Design of a thermal imaging diagnostic using 90-degree off-axis parabolic mirrors. 2006:62880Z–62880Z-9. [Google Scholar]
- 31.Zauderer E. Complex argument Hermite–Gaussian and Laguerre– Gaussian beams. Journal of the Optical Society of America A. 1986;3:465–469. 1986/04/01. [Google Scholar]
- 32.Siegman AE. Hermite–gaussian functions of complex argument as optical-beam eigenfunctions. Journal of the Optical Society of America. 1973;63:1093–1094. 1973/09/01. [Google Scholar]
- 33.Pampaloni F, Enderlein J. Gaussian, hermite-gaussian, and laguerre-gaussian beams: A primer. arXiv preprint physics/0410021. 2004 [Google Scholar]
- 34.Llombart N, Cooper KB, Dengler RJ, Bryllert T, Siegel PH. Confocal Ellipsoidal Reflector System for a Mechanically Scanned Active Terahertz Imager. IEEE Transactions on Antennas and Propagation. 2010;58:1834–1841. [Google Scholar]
- 35.Cooper KB, Dengler RJ, Llombart N, Bryllert T, Chattopadhyay G, Mehdi I, Siegel PH. An Approach for Sub-Second Imaging of Concealed Objects Using Terahertz (THz) Radar. Journal of Infrared, Millimeter, and Terahertz Waves. 2009;30:1297–1307. [Google Scholar]
- 36.Cooper KB, Dengler RJ, Llombart N, Thomas B, Chattopadhyay G, Siegel PH. THz Imaging Radar for Standoff Personnel Screening. IEEE Transactions on Terahertz Science and Technology. 2011;1:169–182. [Google Scholar]
- 37.LACKNER B, SCHMIDINGER G, PIEH S, FUNOVICS MA, SKORPIK C. Repeatability and Reproducibility of Central Corneal Thickness Measurement With Pentacam, Orbscan, and Ultrasound. Optometry and Vision Science. 2005;82:892–899. doi: 10.1097/01.opx.0000180817.46312.0a. [DOI] [PubMed] [Google Scholar]