Abstract
Purpose
The Regularly Incremented Phase Encoding – Magnetic Resonance Fingerprinting (RIPE-MRF) method is introduced to limit the sensitivity of preclinical MRF assessments to pulsatile and respiratory motion artifacts.
Methods
As compared to previously reported standard Cartesian MRF methods (SC-MRF), the proposed RIPE-MRF method uses a modified Cartesian trajectory that varies the acquired phase encoding line within each dynamic MRF dataset. Phantoms and mice were scanned without gating or triggering on a 7T preclinical MRI scanner using the RIPE-MRF and SC-MRF methods. In vitro phantom T1 and T2 measurements as well as in vivo liver assessments of artifact-to-noise ratio (ANR) and MRF-based T1 and T2 mean and standard deviation were compared between the two methods (n=5).
Results
RIPE-MRF showed significant ANR reductions in regions of pulsatility (P<0.005) and respiratory motion (P<0.0005). RIPE-MRF also exhibited improved precision in T1 and T2 measurements in comparison to the SC-MRF method (P< 0.05). The RIPE-MRF and SC-MRF methods displayed similar mean T1 and T2 estimates (difference in mean values <10%).
Conclusion
These results show that the RIPE-MRF method can provide effective motion artifact suppression with minimal impact on T1 and T2 accuracy for in vivo small animal MRI studies.
Keywords: Magnetic Resonance Fingerprinting, motion artifacts, artifact suppression, view ordering, Cartesian trajectory
INTRODUCTION
Quantitative preclinical MRI methods can be used to provide objective assessments of animal models of human disease for comparison with gold standard histological assessments (1–6). Unfortunately, quantitative imaging in small animals can be challenging due to anesthetized respiration rates of 40-80 breaths per minute and heart rates of >200 beats per minute causing blurring and ghosting artifacts that can result in quantification errors (4,5,7–11). Unfortunately, prospective respiratory/cardiac gating to reduce the motion artifacts (9,12–15) may not be practical because they can substantially extend acquisition times (16). Intubation and significant motion restriction (11,17) can also help reduce motion artifacts, but may cause increased mortality in animal models with advanced disease. As a result, preclinical respiratory and cardiovascular motion remains a significant challenge for accurate quantification in preclinical MRI studies.
Magnetic Resonance Fingerprinting (MRF) (18,19) is a quantitative MRI framework that has demonstrated unique properties including robustness to bulk patient motion and preclinical respiratory motion (18,20). However, this resistance to respiratory motion was only observed for low to moderate respiration rates. The same preclinical MRF study also showed that Cartesian MRF is susceptible to artifacts due to cardiac pulsatility. In this work, we propose the Regularly Incremented Phase Encoding (RIPE)-MRF methodology (21) for use on high field preclinical MRI scanners to suppress both respiratory and cardiac pulsatility motion artifacts. This new approach utilizes an incremented view ordering of the k-space phase encoding, within a fully-sampled Cartesian MRF trajectory, to reduce the impact of motion on the MRF-based T1 and T2 maps. The intention of the RIPE-MRF view ordering is to add temporal incoherence to the motion artifacts using altered k-space trajectories similar to the temporal incoherence instituted in highly undersampled non-Cartesian MRF (22–25). Herein, we compare the quantitative and motion suppression capabilities of this new RIPE-MRF technique to the previously described in vivo preclinical MRF methodology (20).
METHODS
RIPE-MRF Encoding
For this study, two Cartesian MRF methods were implemented on a Bruker Biospec 7T MRI Scanner (Bruker Inc., Billerica, MA). The previously described preclinical MRF method (20) will be referred to as Standard Cartesian MRF (SC-MRF). The RIPE-MRF method is derived from the SC-MRF method but has modified the acquisition order of the phase encoding lines as shown in Figure 1. The SC-MRF method (Fig. 1a) acquires the same phase encoding line during each variable repetition time for an entire set of dynamic MRF images. After acquiring the entire set of MRF data for this single phase encoding line, the phase encoding line is incremented and another set of dynamic MRF data is acquired. This process is repeated for all phase encoding lines, starting with the edge line of k-space and continuing in a sequential fashion. In contrast, the RIPE-MRF method (Fig. 1b) linearly increments the acquired k-space line for each repetition time throughout the entire set of dynamic MRF images. The result of this phase encoding variation is a shuffled set of fully sampled Cartesian k-space time points that can be reordered to enable image reconstruction and quantification.
FIG. 1.

Schematic representation of two different MRF acquisition schemes. (a) Standard Cartesian MRF (SC-MRF) and (b) RIPE-MRF phase encoding schema showing four (out of 1024) MRF k-space time points. The red line in each time point dataset is acquired during the first set of dynamic MRF images (1024 total k-space lines). The blue line is the phase encoding line acquired during the second set of dynamic MRF images (second set of 1024 k-space lines). The green line is the third set of acquired MRF phase encoding lines. SC-MRF acquires the same phase encoding line during each set of dynamic MRF images, while the RIPE-MRF strategy increments the acquired phase encoding line during each set of dynamic MRF images to provide temporal incoherence for respiratory and pulsatile motion artifacts.
MRF Design and Quantification
Both the SC-MRF and RIPE-MRF acquisitions used a Fast Imaging with Steady-state Free Precession (FISP) imaging kernel (19,20) with patterns of flip angles and repetition times based on the original MRF method ((18) and Supporting Fig. S1). Following the MRF acquisitions, all imaging data were exported to MATLAB (The Mathworks, Natick, MA) for analysis. MRF quantification was performed by matching the acquired profiles on a pixel by pixel basis to a dictionary of simulated profiles from all logical combinations of T1 and T2. For each pixel the maximum inner product between the acquired profile and the individual dictionary entry yielded the matched T1 and T2 values (26). M0 was also estimated as a scale factor between the acquired data and simulated dictionary profile (18). In vitro MRF experiments were also performed to compare the accuracy of the T1 and T2 estimates from SC-MRF and RIPE-MRF with conventional spin echo MRI assessments (Supplemental Methods).
In vivo RIPE-MRF
All studies were conducted according to protocols approved by the CWRU Institutional Animal Care and Use Committee (IACUC). In vivo, single slice axial liver scans of wild type female C57BL/6 mice (8–12 weeks of age, n=5) were acquired sequentially using both SC-MRF and RIPE-MRF methods resulting in a set of T1, T2, and M0 maps for each method. The order of the RIPE-MRF and SC-MRF scans was alternated between the animals to avoid potential bias due to the acquisition order (Supporting Fig. S2). Each mouse was scanned in two separate imaging sessions, using two different levels of anesthesia to evaluate the impact of respiration and heart rate on the two MRF methods. For the first imaging session, respiration was maintained at 45-60 breaths per minute by adjusting the depth of isoflurane anesthesia (high anesthesia state). The animals underwent a second imaging session with the respiration rate maintained at 80-100 breaths per minute (low anesthesia state). No gating or triggering was utilized so as to assess each method’s baseline sensitivity to motion artifacts. Axial liver images were chosen to provide an MRF dataset with significant respiratory and pulsatile motion artifacts. Imaging parameters for the in vivo MRF studies were: 3×3 cm FOV, 128 × 128 matrix, and 1 mm slice thickness. The total imaging time for each mouse imaging session was 1.5 hours (45 minutes each for SC-MRF and RIPE-MRF).
We analyzed the SC-MRF and RIPE-MRF datasets to assess the effect of motion on the MRF signal evolution profiles. For this qualitative analysis, we utilized the magnitude of the signal intensity at the center of k-space for the MRF datasets which demonstrates both sensitivity to motion artifacts as well as limited noise levels in comparison to individual image voxels. An ideal dictionary profile was plotted with each acquired profile to illustrate the appearance of deviations due to motion. In addition, composite MRF images were generated by taking the complex sum of the reconstructed dynamic MRF images across the entire time domain to visualize the temporal coherence of the motion artifacts for the RIPE-MRF and SC-MRF methods (Supporting Fig. S3). In these composite images incoherent noise should add destructively while coherent signals will add constructively highlighting any motion artifact coherence through time.
Coherence of the motion artifacts was quantified by manually selecting three ROIs in the background region of the composite images: 1) a region of pulsatility artifacts; 2) a region of respiratory motion artifacts; and 3) an overall artifact ROI with respiration and pulsatility artifacts. To ensure consistency, ROIs were selected to cover the entire background in the phase encoding direction for the desired region (Supporting Fig. S4). To measure the relative magnitude of motion artifacts (27), mean values for each of the artifact ROIs were divided by the mean of four separate ROIs taken in background regions with minimal motion artifacts to calculate an artifact-to-noise ratio (ANR). The ANR results from the RIPE-MRF and SC-MRF acquisitions were then compared using unpaired, two-tailed Student’s t-tests.
A similar ROI analysis was also performed on the in vivo MRF-based T1 and T2 maps. Two ROIs were selected within the liver to compare the impact of the motion artifacts on the mean and standard deviation of the measured T1 and T2 values: 1) an area impacted by pulsatility artifact; and 2) an area impacted by respiratory motion artifact. To ensure consistency of the ROIs selection, the ROI for pulsatile motion artifact was chosen in the liver directly anterior to the aorta, while the ROI for respiratory motion artifact was chosen to be lateral to the pulsatile motion area (Supporting Fig. S4). The mean T1 and T2 value as well as the standard deviation for these ROIs as a percent of the mean value within the ROI were compared between corresponding areas from both methods using unpaired, Student’s t-tests. MRF-based M0 maps were analyzed separately in this study (Supporting Fig. S5) as M0 is a scale factor while T1 and T2 are matched parameters. For all experiments, statistical significance was set at P<0.05.
RESULTS
In Vivo Studies
Representative MRF profiles from SC-MRF and RIPE-MRF along with the dictionary profiles are shown in Figure 2 for both the high and low anesthesia states. The SC-MRF profiles exhibit regular coherent deviations from the matched dictionary profiles. For the RIPE-MRF profiles, these deviations are more distributed over the MRF profile and manifest as noise-like spikes. The temporal distribution of these motion-induced variations is consistent between the two anesthesia states, but more frequent in the low anesthesia state as expected.
FIG. 2.

Comparison of MRF signal profiles. Signal intensity profiles from the center of k-space for the MRF datasets are shown with the corresponding dictionary entry overlaid. Profiles from SC-MRF and RIPE-MRF for both the (a) high and (b) low anesthesia states highlight differences in the appearance of motion between the methods. SC-MRF profiles show coherent deviations due to motion, while the RIPE-MRF profiles exhibit “noise-like” deviations distributed more evenly throughout the profile. The frequency of these deviations is increased in the high anesthesia state (b) as expected.
Artifact-to-Noise Ratio (ANR) measurements from the ROI analysis of the in vivo composite MRF images (Supporting Fig. S3) are shown in Figure 3. ANR from the RIPE-MRF scans was significantly reduced in all ROIs for both anesthesia states in comparison to SC-MRF (*P< 0.005). The ANR was also reduced for the low anesthesia state compared to the high anesthesia state with significant differences seen between anesthesia states for SC-MRF in the respiration and overall motion artifact regions (*P< 0.005) and in the overall motion artifact region for RIPE-MRF (**P< 0.0005).
FIG. 3.

Artifact-to-noise ratios (ANR) for SC-MRF and RIPE-MRF. An ROI analysis was used on composite MRF images to measure mean ANRs for each MRF acquisition. Three regions corresponding to predominantly pulsatile motion artifacts, predominantly respiratory motion artifacts, and an overall motion artifact were analyzed. Noise ROIs were selected from motion-free regions of the image background. The RIPE-MRF method exhibited significantly reduced ANR for both the high and low anesthesia states in the regions of pulsatility (**P< 0.0005, *P<.005, respectively), respiration (**P<0.0005, **P<0.0005, respectively), and overall (***P<0.00005, **P<0.0005, respectively) motion artifacts in comparison to SC-MRF. The high and low anesthesia states showed significant differences within a method in the area of overall motion artifact for both SC-MRF and RIPE-MRF (*P< 0.005, **P<.0005, respectively) and in SC-MRF in the area of respiration artifact (*P< 0.005).
Representative in vivo MRF-based T1 and T2 maps are shown in Figure 4 for both the high and low anesthesia states with parameter ranges chosen for display purposes. Consistent with ANR results, the in vivo SC-MRF maps exhibited visible motion artifacts in the phase encoding direction (blue and green arrows, Fig. 4). In contrast, the in vivo RIPE-MRF maps show visibly reduced artifacts both within the liver and in the background regions of the MRF maps. The ROI analysis of the MRF maps (Fig. 5) demonstrated that the mean parametric estimates were similar for both methods. The only significant difference was in mean T1 values in the region of respiration artifact between anesthesia levels for RIPE-MRF (Fig. 5a, *P<0.05). The spatial standard deviation in the parametric values as a percentage of the mean was generally reduced in the RIPE-MRF maps in comparison to SC-MRF suggesting improved uniformity of the RIPE-MRF maps. In particular, RIPE-MRF exhibited significantly reduced variation in three out of four regions with pulsatility artifacts (Fig. 5c,d, *P<0.05). The corresponding M0 maps are presented in Supporting Figure S5 and these demonstrate similar artifact patterns as the T1 and T2 maps.
FIG. 4.

Representative T1 (a) and T2 (b) relaxation time maps of a healthy mouse liver. Example axial in vivo SC-MRF and RIPE-MRF maps, acquired sequentially in the same imaging session, are shown for both the high and low anesthesia states. Pervasive motion artifacts are seen in the phase encoding direction of the SC-MRF maps. The green arrow highlights an area of predominantly pulsatile motion artifacts in the SC-MRF scan while the blue arrow shows an area impacted by respiratory motion artifact. These artifacts are minimized in the RIPE-MRF maps and the areas appear more homogeneous.
FIG. 5.

Liver ROI T1 and T2 measurement analysis. Mean and standard deviation as a percentage of mean value for in vivo liver T1 (a,c) and T2 (b,d) values obtained from the SC-MRF and RIPE-MRF maps in regions of predominantly pulsatile and respiratory motion artifact (green and blue arrows respectively in Fig. 4) for both high and low anesthesia states. Statistically significant differences in T1 relaxation time were seen between high and low anesthesia states in the area of respiration motion for RIPE-MRF (*P<.05). RIPE-MRF demonstrates more precise estimates of T1 and T2 as measured by the reduction in the spatial standard deviation of the measurements. For the RIPE-MRF assessments, the standard deviation was significantly reduced in three out of four T1 comparisons to SC-MRF (*P<0.05) and one of the four comparisons in T2 (*P<0.05). The low anesthesia state resulted in increased variation for all comparisons, but only the area of respiration artifact in T2 was statistically significant for both SC-MRF and RIPE-MRF (***P<0.0005, **P<0.005, respectively).
Results of phantom studies showed significant differences between spin echo and both MRF methods. The MRF methods over-estimated the mean phantom T1 values in comparison to conventional spin echo methods. In addition, the RIPE-MRF method showed small but significant reduction in mean T2 values in comparison to both SC-MRF and the spin echo methods (Supporting Fig. S6).
DISCUSSION
Herein, we describe the Regularly Incremented Phase Encoding (RIPE) – MRF scheme to suppress respiratory and pulsatile motion artifacts in preclinical MRF applications. The RIPE-MRF approach attempts to decrease the temporal coherence of motion artifacts by varying the acquired k-space line during the MRF acquisition (Fig. 1). Linear increments in the phase encoding line were added during the dynamic MRF acquisition resulting in significant reductions in the coherence of motion artifacts (Figs. 2-3, Supporting Fig. S3). The application of modified view ordering to minimize the effect of motion artifacts has been previously used to improve image quality (28–31). The goal of this work was to utilize the concept of view ordering within the MRF framework to limit the impact of motion artifacts on the resulting quantitative maps. Subsequently, we show that this new RIPE-MRF trajectory provides a significant reduction in the coherence of motion artifacts resulting in more precise in vivo T1 and T2 measurements (Figs. 4, 5) compared to previously reported preclinical MRF methods (20).
In vivo application of the RIPE-MRF methodology altered the temporal distribution of motion artifacts in the MRF profiles (Fig. 2). While the periodic, motion-induced deviations in the MRF profiles are coherent for SC-MRF, RIPE-MRF deviations are “noise-like” displacements that are more evenly distributed across the entire MRF signal evolution profile. ANR measurements (Fig. 3) and composite images (Supporting Fig. S3) show these differences in the image domain. The reordering of the k-space datasets with RIPE-MRF limited the temporal coherence of the motion artifacts and resulted in significantly reduced ANR. Importantly, these ANR reductions were consistent for both pulsatility and respiratory motion artifacts and for both high and low anesthesia states.
Motion-induced artifacts in the MRF signal evolution profiles also resulted in corresponding alterations in the MRF-based T1 and T2 maps. In vivo RIPE-MRF maps demonstrated a visible reduction in the appearance of motion artifacts in the liver compared to SC-MRF maps (Fig. 4). The impact of adding temporal incoherence to the motion artifacts on MRF-based quantification was manifest through reduced variation in the hepatic T1 and T2 values measured by RIPE-MRF in both regions of respiration and pulsatility artifacts (Fig. 5). Further, RIPE-MRF resulted in improved precision for both levels of anesthesia indicating it is effective in improving quantification over a range of physiological states. Both MRF methods resulted in mean hepatic T1 estimates consistent with previous reports (4,17). Liver T2 estimates from both MRF methods were reduced in comparison to reported literature values (11,17). This T2 underestimation was also observed for the MRF phantom results in comparison to conventional MRI techniques (Supporting Fig. S6). T2 underestimation may be partially ameliorated through corrections to B0 and B1 heterogeneities but were not further explored in this initial study.
The primary advantage of the RIPE-MRF method is eliminating the need for gating/triggering to get artifact free quantitative maps in small animal imaging. Prior preclinical studies have performed self-navigation and/or motion correction (32–35), retrospectively identified corrupted data using monitoring systems (7,36), or triggered the acquisition during the quiescent period of motion (16,37). These methods reduce motion artifacts but also significantly increase scan time and/or require complex post-processing. In contrast, RIPE-MRF provides a straightforward quantification process utilizing all available data with no restrictions. Additionally, the method is robust across a range of physiological motion rates reducing the need for precise physiological controls (i.e., intubation). Higher variation in the MRF maps in the regions of high frequency pulsatility artifacts indicate that the frequency of motion artifacts is an important factor in MRF trajectory design. For example, further reductions in motion artifacts may be needed to provide reliable MRF maps in regions with high-frequency / rapid motion (e.g., pulmonary / cardiac imaging). Therefore, future studies will be needed to investigate alternative view ordering methods (e.g., random phase encoding) and / or gated MRF approaches to provide artifact-free MRF-based T1 and T2 maps (23).
In this initial study, non-Cartesian trajectories were not explored as a potential form of motion suppression. Prior preclinical studies incorporating non-Cartesian trajectories, such as PROPELLER (2,38), spiral (16,37,39,40), and radial (41–43) k-space sampling, show promising motion artifact suppression. However, the resistance of non-Cartesian MRF methods to motion must be balanced with additional error sources such as eddy current - induced trajectory errors (44) and off-resonance artifacts prevalent on high field preclinical MRI scanners (16). From a practical perspective, non-Cartesian trajectories may be difficult or even impossible to implement on some preclinical MRI scanners. Further, this Cartesian RIPE-MRF approach may be useful for both preclinical and clinical MRF applications in combination with established parallel imaging strategies (45). Regardless, future studies will be needed to thoroughly compare the relative motion artifact resistance of the Cartesian RIPE-MRF approach with non-Cartesian MRF trajectories.
CONCLUSIONS
In conclusion, we have developed the motion artifact-resistant RIPE-MRF method to provide reliable multi-parametric quantification for preclinical MRF applications. In this initial implementation, incrementing the phase encoding line during the dynamic MRF acquisition resulted in suppression of both pulsatility and respiratory motion artifacts in the in vivo MRF-based T1 and T2 relaxation time maps of mouse abdomens. This improved resistance to motion artifacts was achieved with no change in the MRF acquisition time as well as minimal impact on the accuracy of the T1 and T2 estimates. The RIPE-MRF method represents a shift in the preclinical quantitative imaging paradigm, from attempting to reduce or eliminate motion artifacts in the underlying images to accepting the presence of motion artifacts but manipulating them so they are suppressed during quantification. Thus, the RIPE-MRF method serves as a foundation for free-breathing Cartesian MRF in rodents.
Supplementary Material
SUPPORTING FIG. S1. Schematic of MRF pulse sequence with TR and FA patterns. The top panel demonstrates the inversion preparation combined with variable flip angles and repetition times played out until FAn and TRn (n=1024). The slice gradient is unbalanced resulting in the dephasing needed to perform FISP imaging. This pattern is repeated for all lines of k-space changing the acquired phase encoding line based on the method implemented (SC-MRF or RIPE-MRF; Fig. 1). For each TR the phase encoding is perfectly balanced allowing for an arbitrary phase encoding order to be applied.
SUPPORTING FIG. S2. Workflow of in vivo MRF experiments. Shown is the experimental work flow for two animals (out of 5 total) representing the two possible experimental designs. Three mice were imaged using the protocol demonstrated in Mouse 1 and two mice were imaged using the protocol for Mouse 2. The high anesthesia session was performed first to evaluate motion suppression in the presence of lower respiration and heart rates. Within this session the scan order was alternated to average out any physiological changes over the course of the experiment. Then 3-4 weeks later the low anesthesia state was imaged to analyze higher respiration and heart rates with the same alternation of the scan order. Total imaging time for each MRF acquisition was 45 minutes with the total scanning session being 1.5 hours (RIPE-MRF and SC-MRF).
SUPPORTING FIG. S3. Composite MRF images for both SC-MRF and RIPE-MRF. Top and bottom rows are the same images; the top row is windowed to show anatomy, the bottom row is windowed to show artifacts. These composite MRF images were generated by calculating the magnitude of the complex sum of the dynamic MRF images. For these images, temporally coherent signals add constructively resulting in high signal magnitude while temporally incoherent signals will add destructively giving lower magnitude. SC-MRF shows high signal magnitude from artifact in the image background compared to RIPE-MRF indicating an element of incoherence being added to the artifact in the time domain when using the RIPE-MRF method. Some residual pulsatility artifacts are seen in the RIPE-MRF. Additionally, RIPE-MRF images appear to have less blurring.
SUPPORTING FIG. S4. Shown are representative composite MRF images and MRF-based T1 and T2 maps to illustrate how ROIs were selected. For ANR measurements on the composite images (left column) ROIs were chosen to cover the entire phase encoding direction for the respiration artifact (blue), pulsatility artifact (green), and overall artifact (red) to analyze a similar number of pixels for each animal. T1 and T2 ROIs (middle and right column, respectively) show the presence of ROIs for respiration artifact (blue) and pulsatility artifact (green). Pulsatility was chosen anterior to the aorta and respiration was chosen laterally from the area of pulsatility to get consistently selected ROIs between animals.
SUPPORTING FIG. S5. Representative M0 maps from in vivo SC-MRF and RIPE-MRF acquisitions. These maps correspond to the T1 and T2 maps seen in Figure 4. In this study, the MRF-based M0 maps are estimated as a scale factor and are not matched like T1 and T2. RIPE-MRF maps show distinct reductions in motion artifacts in comparison to SC-MRF similar to the T1 and T2 maps shown in Figure 4.
SUPPORTING FIG. S6. Phantom results from SE, SC-MRF, and RIPE-MRF. T1 and T2 maps show similar quantitative values between both MRF methods and reasonable agreement with SE. Error bars are shown as the standard deviation of the 5 mean T1 and T2 estimates obtained for each phantom in the in vitro repeatability study. MRF-based T1 estimates are significantly higher than conventional MRI estimates while MRF-based T2 estimates are under-estimated. The phantom results are more consistent between the two MRF methods. (*P<0.05, **P<0.01, ***P<0.005, ****P<0.0005)
Acknowledgments
The authors would like to acknowledge the support of the Interdisciplinary Biomedical Imaging Training Program, NIH T32EB007509 administered by the Department of Biomedical Engineering, Case Western Reserve University; NIH/NHLBI F30 HL136190; NIH/NHLBI R21 HL130839; the Case Comprehensive Cancer Center (NIH/NCI P30 CA43703); the Clinical and Translation Science Collaborative of Cleveland (NIH/NCATS UL1 TR000439); the Cystic Fibrosis Foundation; and the Polycystic Kidney Disease Foundation.
References
- 1.Chouhan M, Mookerjee R, Bainbridge A, Walker-Samuel S, Davies N, Halligan S, Lythgoe MF, Taylor SA. Use of Caval Subtraction 2D Phase-Contrast MR Imaging to Measure Total Liver and Hepatic Arterial Blood Flow : Preclinical Validation and Initial Clinical Translation. Radiology. 2016;280(3):916–923. doi: 10.1148/radiol.2016151832. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Yamaguchi M, Mitsuda M, Ezawa K, Nakagami R, Furuta T, Sekine N, Niitsu M, Fujii H. Artifact-reduced simultaneous MRI of multiple rats with liver cancer using PROPELLER. J Magn Reson Imaging. 2013;38(1):225–230. doi: 10.1002/jmri.23969. [DOI] [PubMed] [Google Scholar]
- 3.Pan F, Den J, Zhang C, Wang H, Cheng J, Wu W, Hong N, Wang Y. The Therapeutic Response of Gastrointestinal Stromal Tumors to Imatinib Treatment Assessed by Intravoxel Incoherent Motion Diffusion-Weighted Magnetic Resonance Imaging with Histopathological Correlation. PLoS One. 2016;11(12) doi: 10.1371/journal.pone.0167720. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Gao Y, Erokwu BO, DeSantis DA, Croniger CM, Schur RM, Lu L, Mariappuram J, Dell KM, Flask CA. Initial evaluation of hepatic T1 relaxation time as an imaging marker of liver disease associated with autosomal recessive polycystic kidney disease (ARPKD) NMR Biomed. 2016;29(1):84–89. doi: 10.1002/nbm.3442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Chen W, Zhang B, Xia R, Zhang R, Xu Z, Chen Y, Wang C, Wang L, Zheng J, Gao F. T2 mapping at 7T MRI can quantitatively assess intramyocardial hemorrhage in rats with acute reperfused myocardial infarction in vivo. J Magn Reson Imaging. 2016;44(1):194–203. doi: 10.1002/jmri.25145. [DOI] [PubMed] [Google Scholar]
- 6.Lemasson B, Pannetier N, Coquery N, Boisserand LSB, Collomb N, Schuff N, Moseley M, Zaharchuk G, Barbier EL, Christen T. MR Vascular Fingerprinting in Stroke and Brain Tumors Models. Sci Rep. 2016 Nov;6:37071. doi: 10.1038/srep37071. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Ramasawmy R, Campbell-Washburn AE, Wells JA, Johnson SP, Pedley RB, Walker-Samuel S, Lythgoe MF. Hepatic arterial spin labelling MRI: an initial evaluation in mice. NMR Biomed. 2015;28(2):272–280. doi: 10.1002/nbm.3251. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Yin T, Peeters R, Feng Y, Liu Y, Yu J, Dymarkowski S, Himmelreich U, Oyen R, Ni Y. Characterization of a rat orthotopic pancreatic head tumor model using three-dimensional and quantitative multi-parametric MRI. NMR Biomed. 2017;30(2):e3676. doi: 10.1002/nbm.3676. [DOI] [PubMed] [Google Scholar]
- 9.Li W, Griswold M, Yu X. Rapid T1 mapping of mouse myocardium with saturation recovery look-locker method. Magn Reson Med. 2010;64(5):1296–1303. doi: 10.1002/mrm.22544. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Naresh NK, Chen X, Roy RJ, Antkowiak PF, Annex BH, Epstein FH. Accelerated dual-contrast first-pass perfusion MRI of the mouse heart: Development and application to diet-induced obese mice. Magn Reson Med. 2015;73(3):1237–1245. doi: 10.1002/mrm.25238. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Zhang H, Yang Q, Yu T, Chen X, Huang J, Tan C, Liang B, Guo H. Comparison of T2, T1rho, and diffusion metrics in assessment of liver fibrosis in rats. J Magn Reson Imaging. 2017;45(3):741–750. doi: 10.1002/jmri.25424. [DOI] [PubMed] [Google Scholar]
- 12.Parzy E, Miraux S, Franconi JM, Thiaudière E. In vivo quantification of blood velocity in mouse carotid and pulmonary arteries by ECG-triggered 3D time-resolved magnetic resonance angiography. NMR Biomed. 2009;22(5):532–537. doi: 10.1002/nbm.1365. [DOI] [PubMed] [Google Scholar]
- 13.Fries P, Seidel R, Müller A, Matthes K, Denda G, Massmann A, Menger MD, Sperling J, Morelli JN, Altmeyer K, et al. Comparison of self-gated and prospectively triggered fast low angle shot (FLASH) sequences for contrast-enhanced magnetic resonance imaging of the liver at 9.4 T in a rat model of colorectal cancer metastases. Invest Radiol. 2013;48(10):738–744. doi: 10.1097/RLI.0b013e318294dd0e. [DOI] [PubMed] [Google Scholar]
- 14.Zhang H, Ye Q, Zheng J, Schelbert EB, Hitchens TK, Ho C. Improve myocardial T1 measurement in rats with a new regression model: Application to myocardial infarction and beyond. Magn Reson Med. 2013;72(3):737–748. doi: 10.1002/mrm.24988. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Herrmann KH, Pfeiffer N, Krumbein I, Herrmann L, Reichenbach JR. MRI compatible small animal monitoring and trigger system for whole body scanners. Z Med Phys. 2014;24(1):55–64. doi: 10.1016/j.zemedi.2013.07.005. [DOI] [PubMed] [Google Scholar]
- 16.Castets CR, Ribot EJ, Lefrancois W, Trotier AJ, Thiaudiere E, Franconi JM, Miraux S. Fast and robust 3D T1 mapping using spiral encoding and steady RF excitation at 7T: Application to cardiac manganese enhanced MRI (MEMRI) in mice. NMR Biomed. 2015;28(7):881–889. doi: 10.1002/nbm.3327. [DOI] [PubMed] [Google Scholar]
- 17.Chow AM, Gao DS, Fan SJ, Qiao Z, Lee FY, Yang J, Man K, Wu EX. Measurement of liver T1 and T2 relaxation times in an experimental mouse model of liver fibrosis. J Magn Reson Imaging. 2012;36(1):152–158. doi: 10.1002/jmri.23606. [DOI] [PubMed] [Google Scholar]
- 18.Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA. Magnetic resonance fingerprinting. Nature. 2013;495(7440):187–192. doi: 10.1038/nature11971. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Jiang Y, Ma D, Seiberlich N, Gulani V, Griswold MA. MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout. Magn Reson Med. 2015;74(6):1621–1631. doi: 10.1002/mrm.25559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Gao Y, Chen Y, Ma D, Jiang Y, Herrmann KA, Vincent JA, Dell KM, Drumm ML, Brady-Kalnay SM, Griswold MA, et al. Preclinical MR fingerprinting (MRF) at 7 T: effective quantitative imaging for rodent disease models. NMR Biomed. 2015;28(3):384–394. doi: 10.1002/nbm.3262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Anderson C, Wang C, Gu Y, Griswold M, Yu X, Flask C. Regularly incremented phase encoding magnetic resonance fingerprinting (RIPE-MRF) for enhanced motion suppression in cartesian MRF. Proc. 25th Annu. Meet; ISMRMH, onolulu, Hawaii, USA. 2017; p. 5307. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Cloos MA, Knoll F, Zhao T, Block K, Bruno M, Wiggins C, Sodickson D. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun. 2016;53(9):1689–1699. doi: 10.1038/ncomms12445. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Hamilton JI, Jiang Y, Chen Y, Ma D, Lo W-C, Griswold M, Seiberlich N. MR fingerprinting for rapid quantification of myocardial T1, T2, and proton spin density. Magn Reson Med. 2016;77(4):1446–1459. doi: 10.1002/mrm.26216. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Cauley SF, Setsompop K, Ma D, Jiang Y, Ye H, Adalsteinsson E, Griswold MA, Wald LL. Fast group matching for MR fingerprinting reconstruction. Magn Reson Med. 2014;74(2):523–528. doi: 10.1002/mrm.25439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Ye H, Ma D, Jiang Y, Cauley SF, Du Y, Wald LL, Griswold MA, Setsompop K. Accelerating magnetic resonance fingerprinting (MRF) using t-blipped simultaneous multislice (SMS) acquisition. Magn Reson Med. 2016;75(5):2078–2085. doi: 10.1002/mrm.25799. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Rieger B, Zimmer F, Zapp J, Weingärtner S, Schad LR. Magnetic resonance fingerprinting using echo-planar imaging: Joint quantification of T1 and T2* relaxation times. Magn Reson Med. 2016 doi: 10.1002/mrm.26561. [DOI] [PubMed] [Google Scholar]
- 27.Epstein FH, Arai AE. Optimization of fast cardiac imaging using an echo-train readout. J Magn Reson Imaging. 2000;11(2):75–80. doi: 10.1002/(sici)1522-2586(200002)11:2<75::aid-jmri1>3.0.co;2-p. [DOI] [PubMed] [Google Scholar]
- 28.Nguyen TD, Ding G, Watts R, Wang Y. Optimization of view ordering for motion artifact suppression. Magn Reson Imaging. 2001;19(7):951–957. doi: 10.1016/s0730-725x(01)00426-x. [DOI] [PubMed] [Google Scholar]
- 29.Maclaren JR, Bones PJ, Millane RP, Watts R. A modified view ordering for artifact reduction in MRI. Annu Int Conf IEEE Eng Med Biol - Proc. 2007;(2):2053–2056. doi: 10.1109/IEMBS.2007.4352724. [DOI] [PubMed] [Google Scholar]
- 30.Wilman AH, Riederer SJ. Performance of an elliptical centric view order for signal enhancement and motion artifact suppression in breath-hold three-dimensional gradient echo imaging. Magn Reson Med. 1997;38(5):793–802. doi: 10.1002/mrm.1910380516. [DOI] [PubMed] [Google Scholar]
- 31.Haacke EM, Patrick JL. Reducing motion artifacts in two-dimensional Fourier transform imaging. Magn Reson Imaging. 1986;4(4):359–376. doi: 10.1016/0730-725x(86)91046-5. [DOI] [PubMed] [Google Scholar]
- 32.Winter P, Kampf T, Helluy X, Gutjahr FT, Meyer CB, Bauer WR, Jakob PM, Herold V. Self-navigation under non-steady-state conditions: cardiac and respiratory self-gating of inversion recovery snapshot FLASH acquisitions in mice. Magn Reson Med. 2016;76(6):1887–1894. doi: 10.1002/mrm.26068. [DOI] [PubMed] [Google Scholar]
- 33.Yang HJ, Sharif B, Pang J, Kali A, Bi X, Cokic I, Li D, Dharmakumar R. Free-breathing, motion-corrected, highly efficient whole heart T2 mapping at 3T with hybrid radial-cartesian trajectory. Magn Reson Med. 2016;75(1):126–136. doi: 10.1002/mrm.25576. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Tibiletti M, Bianchi A, Kjorstad A, Wundrak S, Stiller D, Rasche V. Respiratory self-gated 3D UTE for lung imaging in small animal MRI. Magn Reson Med. 2016 doi: 10.1002/mrm.26463. [DOI] [PubMed] [Google Scholar]
- 35.Ribot EJ, Duriez TJ, Trotier AJ, Thiaudiere E, Franconi JM, Miraux S. Self-gated bSSFP sequences to detect iron-labeled cancer cells and/or metastases in vivo in mouse liver at 7 Tesla. J Magn Reson Imaging. 2015;41(5):1413–1421. doi: 10.1002/jmri.24688. [DOI] [PubMed] [Google Scholar]
- 36.Campbell-Washburn AE, Price AN, Wells JA, Thomas DL, Ordidge RJ, Lythgoe MF. Cardiac arterial spin labeling using segmented ECG-gated Look-Locker FAIR: Variability and repeatability in preclinical studies. Magn Reson Med. 2013;69(1):238–247. doi: 10.1002/mrm.24243. [DOI] [PubMed] [Google Scholar]
- 37.Vandsburger MH, Janiczek RL, Xu Y, French BA, Meyer CH, Kramer CM, Epstein FH. Improved arterial spin labeling after myocardial infarction in mice using cardiac and respiratory gated look-locker imaging with fuzzy C-means clustering. Magn Reson Med. 2010;63(3):648–657. doi: 10.1002/mrm.22280. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Pandit P, Qi Y, Story J, King KF, Johnson GA. Multishot PROPELLER for high-field preclinical MRI. Magn Reson Med. 2010;64(1):47–53. doi: 10.1002/mrm.22376. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.van de Looij Y, Mauconduit F, Beaumont M, Valable S, Farion R, Francony G, Payen JF, Lahrech H. Diffusion tensor imaging of diffuse axonal injury in a rat brain trauma model. NMR Biomed. 2012;25(1):93–103. doi: 10.1002/nbm.1721. [DOI] [PubMed] [Google Scholar]
- 40.Castets CR, Lefrancois W, Wecker D, Ribot EJ, Trotier AJ, Thiaudiere E, Franconi JM, Miraux S. Fast 3D ultrashort echo-time spiral projection imaging using golden-angle: A flexible protocol for in vivo mouse imaging at high magnetic field. Magn Reson Med. 2016;77(5):1831–1840. doi: 10.1002/mrm.26263. [DOI] [PubMed] [Google Scholar]
- 41.Zurek M, Johansson E, Risse F, Alamidi D, Olsson LE, Hockings PD. Accurate T1 mapping for oxygen-enhanced MRI in the mouse lung using a segmented inversion-recovery ultrashort echo-time sequence. Magn Reson Med. 2014;71(6):2180–2185. doi: 10.1002/mrm.24876. [DOI] [PubMed] [Google Scholar]
- 42.Messroghli DR, Nordmeyer S, Buehrer M, Kozerke S, Dietrich T, Kaschina E, Becher PM, Hucko T, Berger F, Klein C, et al. Small Animal Look-Locker Inversion Recovery (SALLI) for Simultaneous Generation of Cardiac T1 Maps and Cine and Inversion Recovery-prepared Images at High Heart Rates: Initial Experience. Radiology. 2011;261(1):258–265. doi: 10.1148/radiol.11101943. [DOI] [PubMed] [Google Scholar]
- 43.Vautier J, Heilmann M, Walczak C, Mispelter J, Volk A. 2D and 3D radial multi-gradient-echo DCE MRI in murine tumor models with dynamic R2*-corrected R1 mapping. Magn Reson Med. 2010;64(1):313–318. doi: 10.1002/mrm.22396. [DOI] [PubMed] [Google Scholar]
- 44.Beaumont M, Lamalle L, Segebarth C, Barbier EL. Improved k-space trajectory measurement with signal shifting. Magn Reson Med. 2007;58(1):200–205. doi: 10.1002/mrm.21254. [DOI] [PubMed] [Google Scholar]
- 45.Davies M, Puy G, Vandergheynst P, Wiaux Y. A compressed sensing framework for magnetic resonance fingerprinting. SIAM J Imaging Sci. 2014;7(4):2623–2656. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
SUPPORTING FIG. S1. Schematic of MRF pulse sequence with TR and FA patterns. The top panel demonstrates the inversion preparation combined with variable flip angles and repetition times played out until FAn and TRn (n=1024). The slice gradient is unbalanced resulting in the dephasing needed to perform FISP imaging. This pattern is repeated for all lines of k-space changing the acquired phase encoding line based on the method implemented (SC-MRF or RIPE-MRF; Fig. 1). For each TR the phase encoding is perfectly balanced allowing for an arbitrary phase encoding order to be applied.
SUPPORTING FIG. S2. Workflow of in vivo MRF experiments. Shown is the experimental work flow for two animals (out of 5 total) representing the two possible experimental designs. Three mice were imaged using the protocol demonstrated in Mouse 1 and two mice were imaged using the protocol for Mouse 2. The high anesthesia session was performed first to evaluate motion suppression in the presence of lower respiration and heart rates. Within this session the scan order was alternated to average out any physiological changes over the course of the experiment. Then 3-4 weeks later the low anesthesia state was imaged to analyze higher respiration and heart rates with the same alternation of the scan order. Total imaging time for each MRF acquisition was 45 minutes with the total scanning session being 1.5 hours (RIPE-MRF and SC-MRF).
SUPPORTING FIG. S3. Composite MRF images for both SC-MRF and RIPE-MRF. Top and bottom rows are the same images; the top row is windowed to show anatomy, the bottom row is windowed to show artifacts. These composite MRF images were generated by calculating the magnitude of the complex sum of the dynamic MRF images. For these images, temporally coherent signals add constructively resulting in high signal magnitude while temporally incoherent signals will add destructively giving lower magnitude. SC-MRF shows high signal magnitude from artifact in the image background compared to RIPE-MRF indicating an element of incoherence being added to the artifact in the time domain when using the RIPE-MRF method. Some residual pulsatility artifacts are seen in the RIPE-MRF. Additionally, RIPE-MRF images appear to have less blurring.
SUPPORTING FIG. S4. Shown are representative composite MRF images and MRF-based T1 and T2 maps to illustrate how ROIs were selected. For ANR measurements on the composite images (left column) ROIs were chosen to cover the entire phase encoding direction for the respiration artifact (blue), pulsatility artifact (green), and overall artifact (red) to analyze a similar number of pixels for each animal. T1 and T2 ROIs (middle and right column, respectively) show the presence of ROIs for respiration artifact (blue) and pulsatility artifact (green). Pulsatility was chosen anterior to the aorta and respiration was chosen laterally from the area of pulsatility to get consistently selected ROIs between animals.
SUPPORTING FIG. S5. Representative M0 maps from in vivo SC-MRF and RIPE-MRF acquisitions. These maps correspond to the T1 and T2 maps seen in Figure 4. In this study, the MRF-based M0 maps are estimated as a scale factor and are not matched like T1 and T2. RIPE-MRF maps show distinct reductions in motion artifacts in comparison to SC-MRF similar to the T1 and T2 maps shown in Figure 4.
SUPPORTING FIG. S6. Phantom results from SE, SC-MRF, and RIPE-MRF. T1 and T2 maps show similar quantitative values between both MRF methods and reasonable agreement with SE. Error bars are shown as the standard deviation of the 5 mean T1 and T2 estimates obtained for each phantom in the in vitro repeatability study. MRF-based T1 estimates are significantly higher than conventional MRI estimates while MRF-based T2 estimates are under-estimated. The phantom results are more consistent between the two MRF methods. (*P<0.05, **P<0.01, ***P<0.005, ****P<0.0005)
