Performance of the Xpert HIV-1 Viral Load Assay: a Systematic Review and Meta-analysis

Madlen Nash a,b, Sophie Huddart a,b, Sayema Badar c, Shrikala Baliga d,e, Kavitha Saravu e,f, Madhukar Pai a,b,
Editor: Yi-Wei Tangg
PMCID: PMC5869835  PMID: 29386266

ABSTRACT

Viral load (VL) is the preferred treatment-monitoring approach for HIV-positive patients. However, more rapid, near-patient, and low-complexity assays are needed to scale up VL testing. The Xpert HIV-1 VL assay (Cepheid, Sunnyvale, CA) is a new, automated molecular test, and it can leverage the GeneXpert systems that are being used widely for tuberculosis diagnosis. We systematically reviewed the evidence on the performance of this new tool in comparison to established reference standards. A total of 12 articles (13 studies) in which HIV patient VLs were compared between Xpert HIV VL assay and a reference standard VL assay were identified. Study quality was generally high, but substantial variability was observed in the number and type of agreement measures reported. Correlation coefficients between Xpert and reference assays were high, with a pooled Pearson correlation (n = 8) of 0.94 (95% confidence interval [CI], 0.89, 0.97) and Spearman correlation (n = 3) of 0.96 (95% CI, 0.86, 0.99). Bland-Altman metrics (n = 11) all were within 0.35 log copies/ml of perfect agreement. Overall, Xpert HIV-1 VL performed well compared to current reference tests. The minimal training and infrastructure requirements for the Xpert HIV-1 VL assay make it attractive for use in resource-constrained settings, where point-of-care VL testing is most needed.

KEYWORDS: HIV, Xpert HIV-1, accuracy, diagnosis, viral load

INTRODUCTION

Despite the recommendations put forward by the World Health Organization (WHO), viral load (VL) monitoring of antiretroviral therapy (ART) is not routinely performed in many low-resource countries, and treatment failure is diagnosed on the basis of clinical or immunological criteria. Currently used VL assays demand sophisticated facilities, expensive equipment, and skilled technicians, making them unaffordable and largely impractical for scale-up in resource-limited settings (1). To expand the use of targeted and routine VL monitoring, inexpensive, low-complexity assays are needed, preferably for point-of-care use.

The Xpert HIV-1 VL assay (Cepheid, Sunnyvale, CA, USA), performed on the GeneXpert instrument system, is an in vitro diagnostic test designed for the rapid quantification of HIV-1 in human plasma from individuals with an active HIV infection. It uses real-time quantitative reverse transcription PCR technology and targets HIV-1 group M subtypes A, B, C, D, AE, F, G, H, AB, AG, J, and K and groups N and O. It has a limit of quantitation of 40 copies/ml and can detect HIV-1 RNA over a linear range of 40 to 10,000,000 copies/ml. The GeneXpert platform allows on-demand molecular testing in one fully integrated closed cartridge and provides results in 90 min (2). The Xpert HIV-1 VL assay runs on the same GeneXpert platform as the WHO-endorsed Xpert MTB/RIF cartridge, used for diagnosis of tuberculosis (TB).

The Xpert MTB/RIF assay is a major advance in tuberculosis diagnostics, and more than 23 million cartridges have been used around the world (3, 4). Given the high coprevalence of TB and HIV in many settings (5) and the need for greater integration of TB-HIV services, leveraging the existing GeneXpert network for HIV may substantially increase access to VL testing. Similar to the Xpert MTB/RIF assay, the Xpert HIV-1 VL assay demands minimal training and modest infrastructure requirements while providing rapid results. If proven to be as accurate as current, established reference standard VL tests, this monitoring tool has the potential for rapid scale-up in countries already using the GeneXpert platform for TB. Scaling up routine HIV-1 VL testing is essential to meet the UNAIDS 90-90-90 targets (6).

In July of 2017, the Xpert HIV-1 VL was accepted for the WHO list of prequalified in vitro diagnostics (7). Country-level validation studies on the accuracy of this assay have now been conducted in a variety of settings. We conducted a systematic review to synthesize evidence from these validation studies and estimate the overall agreement between the Xpert HIV-1 VL assay and current reference standard assays.

MATERIALS AND METHODS

We conducted a systematic review of the literature according to PRISMA guidelines (8). A written protocol for the systematic review was prepared a priori. A modification to the pooling criteria was made after the authors were made aware of the tendency of I2, a measure of heterogeneity among studies, to approach 100% as sample sizes of studies increase (9). Thus, for the pooling of correlations, where internal study variability was very small and the resulting I2 was very high, the decision to pool was based on clinical heterogeneity.

Search strategy and selection criteria.

Using Ovid, we systematically searched Medline (1946 to 2017), Embase (1947 to 2017), and Global Health (1973 to 2017) for studies evaluating the performance of the Xpert HIV-1 VL assay in HIV-positive patients (see Appendix S1 in the supplemental material for complete search strategy). No language or geographic restriction was applied. The study search was conducted on 25 September 2017.

All studies that enrolled participants with a known HIV-positive status were eligible for inclusion.

We restricted inclusion to studies of patients of known HIV status, because VL testing is not recommended as an HIV diagnostic but as a treatment-monitoring tool. The included studies were required to compare Cepheid's Xpert HIV-1 VL assay to another established PCR-based VL assay and report at least one comparison measure between Xpert VLs and gold standard VLs (i.e., Pearson or Spearman correlation, Bland-Altman agreement, etc.). Conference abstracts and posters were excluded, as their quality could not be comprehensively assessed (Fig. 1).

FIG 1.

FIG 1

PRISMA flow chart.

Two reviewers (S. Huddart and M. Nash) screened the titles and abstracts of citations retrieved from all sources. Duplicates were removed, and studies which met the inclusion criteria were flagged for further review. A full-text screen of relevant studies was then performed by the same independent reviewers.

Data extraction.

Two reviewers (S. Huddart and M. Nash) independently extracted data from all eligible studies on patient demographics, correlation coefficients, and results of Bland-Altman analyses comparing Xpert HIV-1 VL assays and the reference test and study quality (see Appendix S1).

Extracted data were adjudicated by the two reviewers, with discrepancies resolved by a third reviewer (S. Badar). Study authors were contacted regarding information not reported in the included manuscripts, but no responses were received.

Risk of bias and assessment of study quality.

A modified version of the QUADAS-2 criteria for diagnostic tests was used to assess study quality (10). Because of the numerical output of both the index and reference tests, biased interpretation of either assay was unlikely; thus, we did not evaluate blinding of the index test or reference test as sources of bias. We assessed patient selection and patient flow according to QUADAS-2 guidelines. Overall risk of bias for patient selection and flow was summarized as low, high, or unclear.

Statistical analysis.

Correlation coefficients, both Spearman and Pearson, were extracted from studies reporting these measures. Because of the theoretical heterogeneity between parametric and nonparametric measures, meta-analysis was only considered within strata of correlation type. The decision to pool was based on the authors' assessment of clinical heterogeneity.

The results of Bland-Altman analyses, a summary of the mean differences between VL measures, were also extracted. Twelve Bland-Altman analyses were performed among the included studies, but one (11) did not report the corresponding standard deviation, confidence interval (CI), or limit of agreement. Without a measure of variance, it is difficult to appropriately interpret this Bland-Altman analysis, thus the study's Bland-Altman analysis was not included in the quantitative summary. The decision to pool for Bland-Altman was based on I2, as internal study variability was moderate and I2 appropriately reflected heterogeneity.

All extracted studies performed their correlation and Bland-Altman analyses on log-transformed viral loads. For studies on patient samples, samples were independent, with one sample per patient included in the agreement analyses. For one included study using laboratory-quality assessment samples, the sampling procedure was not described (reference 12, referring to the laboratory samples).

Meta-analysis was performed using a DerSimonian and Laird random effects model (13).

The number of studies was insufficient to conduct metaregression. Traditional methods to assess publication bias, such as Egger's test or funnel plots, are severely underpowered, especially with small numbers of studies, thus publication bias was not formally assessed but is assumed to exist to some degree in all systematic reviews.

All statistical analyses were performed in R.

RESULTS

Figure 1 shows the study selection flow chart. Twelve articles covering 13 studies were identified during the systematic search of the literature; these studies enrolled more than 3,300 individual patients, 2,011 of whom had VLs quantifiable on Xpert and the reference test (Table 1) (11, 12, 1423). For most studies, patient blood samples were collected as part of routine clinical practice either within a larger HIV/AIDS study or through health care institutions. One paper included data on the quality assessment samples (12). Three studies enrolled HIV patients in India (20, 22, 23), and three studies enrolled HIV patients in South Africa (references 12 [referring to both the patient and the laboratory samples] and 14). Overall, the quality of reporting of patient demographics was poor. Only 3 of 13 studies reported a complete set of demographic covariates (average patient age, gender distribution, and the proportion of participants receiving ART) (12, 14, 20). Of those studies that reported ART coverage, treatment rates ranged from 0 to 100%. The reference standards used included PCR-based VL assays (Abbott RealTime HIV-1 m2000rt, Abbott Molecular; COBAS AmpliPrep/COBAS TaqMan, Roche Diagnostics; NucliSENS EasyQ HIV-1 v2.0, bioMérieux; and Versant HIV-1 RNA 1.5 assay, Siemens HealthCare Diagnostics).

TABLE 1.

Study characteristics

Study Yr No. enrolled No. quantifiable on index and reference test Location Patient population Age % Female % on ART Test and manufacturera
Avidor et al. (21) 2017 383 254 Israel HIV+, routine care Roche TaqMan
Bruzzone et al. (19) 2017 50 41 Italy HIV+ laboratory samples Versant
Ceffa et al. (11) 2015 300 274 Malawi Pediatric and adult HIV+, routine care 100 Abbott
Garrett et al. (14) 2016 42 42 South Africa HIV+ women in CAPRISA 002 study 33 100 66 Roche TaqMan
Gous et al., laboratory samples (12) 2016 42 20 South Africa Quality assessment HIV+ laboratory samples Roche TaqMan
Gous et al., patient samples (12) 2016 158 53 South Africa HIV+, routine care 42 54 100 Roche TaqMan
Gueudin et al. (15) 2016 295 162 France HIV+, routine care 20 Abbott
Jordan et al. (16) 2016 764 390 Europe, USA Adult HIV+, routine care 45 28 Abbott
Kulkarni et al. (22) 2017 219 167 India HIV+, routine care 38 52 Abbott
Mor et al. (17) 2015 404 146 Israel HIV+, routine care NucliSens v2.0
Moyo et al. (18) 2016 302 277 Botswana HIV+, ART naive, rural patients 0 Abbott
Nash et al. (20) 2017 246 89 India Adult HIV+, routine care 41 40 70 Roche TaqMan
Swathirajan et al. (23) 2017 103 96 India HIV+, routine care Abbott
a

Full names and manufacturers of reference tests are the following: Abbot, Abbott RealTime HIV-1 m2000rt, Abbott Molecular; Roche TaqMan, COBAS AmpliPrep/COBAS TaqMan, Roche Diagnostics; NucliSENS, NucliSENS EasyQ HIV-1 v2.0, bioMérieux; Versant, HIV-1 RNA 1.5 assay, Siemens HealthCare Diagnostics.

Overall, when pertinent information was reported, study quality was high (Table 2). Validation studies require cross-sectional design, which minimizes the risk of potential biases. Risk of selection bias, nongeneralizability, or risk of flow bias was generally low. Some studies did not clearly state aspects related to study design (i.e., sequential or random enrollment and reasons for exclusion from analysis), which limits the ability to draw conclusions about study quality.

TABLE 2.

Study quality assessment using modified QUADAS-2a

Study Risk of selection bias Risk of nongeneralizability Risk of flow bias
Avidor et al. (21) Unclear Unclear Low
Bruzzone et al. (19) Unclear Low Low
Ceffa et al. (11) Low Low Low
Garrett et al. (14) Low Low Low
Gous et al. (12) Unclear Unclear Unclear
Gous et al. (12) Unclear Low Unclear
Gueudin et al. (15) Unclear Low Low
Jordan et al. (16) Low Low High
Kulkarni et al. (22) Unclear Unclear Low
Mor et al. (17) Unclear Unclear Low
Moyo et al. (18) Unclear Unclear Unclear
Nash et al. (20) Low Low Low
Swathirajan et al. (23) Unclear Unclear Low
a

Risk of biases assessed using modified QUADAS-2 (10). Unclear risk denotes there was insufficient information provided in the paper to assess the particular bias.

All reported correlations between the Xpert HIV-1 VL assay and gold standard tests were very high for both Pearson and Spearman correlation coefficients (Fig. 2). As all correlations within strata were very close, ranging from 0.81 to 0.98, studies were pooled within strata of correlation type. The pooled correlations were 0.94 (95% CI, 0.89, 0.97) and 0.96 (95% CI, 0.86, 0.99) for Pearson and Spearman correlations, respectively. These values indicate a very high degree of agreement between Xpert VL and reference standard VL values.

FIG 2.

FIG 2

Forest plot for Pearson and Spearman correlation coefficients from comparison of VL values by Xpert and a reference test for VL.

Bland-Altman results are normally distributed, thus I2 is an appropriate metric for heterogeneity. The I2 for the Bland-Altman results was 96%, which suggests enough heterogeneity to preclude pooling. However, all studies reported a mean difference within 0.34 U of zero (the ideal value) (Fig. 3). We stratified the correlation and Bland-Altman values by ART status and found no major difference in studies with patients on ART versus studies with ART-naive or mixed-ART-status patients (see Fig. S1 and S2 in the supplemental material).

FIG 3.

FIG 3

Forest plot for Bland-Altman (BA) correlation coefficients from comparison of VL values by Xpert and a reference test for VL.

There was a sufficient number of Bland-Altman studies to stratify by type of reference test, although heterogeneity remained too high to provide pooled estimates (Fig. S3). No major differences in Bland-Altman results by reference test were identified.

We observed substantial variability in the number and type of agreement measures reported in the included studies (Table 3). Correlation, either Pearson or Spearman, and Bland-Altman measures were the most commonly reported, but several other measures were reported. Additionally, one study (11) reported a mean log difference as part of its Bland-Altman analysis but did not provide standard deviations, confidence intervals, or limits of agreement. This prevented some studies from being included in the Bland-Altman meta-analysis.

TABLE 3.

Diversity of measures of agreement reported in Xpert HIV evaluations

Measure Result by studya
Avidor et al. (21) Bruzzone et al. (19) Ceffa et al. (11) Garrett et al. (14) Gous et al., laboratory samples (12) Gous et al., patient samples (12) Gueudin et al. (15) Jordan et al. (16) Kulkarni et al. (22) Mor et al. (17) Moyo et al. (18) Nash et al. (20) Swathirajan et al. (23)
Pearson correlation *
Spearman correlation
Bland-Altman ** ***
Passing Bablok regression
Deming regression
Agreement, binary threshold (%)
Agreement, 3+ categories (%)
Similarity (%)
Similarity coefficient of variation (%)
Kappa Statistic
Concordance Correlation
a

*, Study estimated Pearson correlation as a secondary metric, so the Spearman correlation was used for this analysis; **, study performed a Bland-Altman analysis but did not report a measure of variance and thus was excluded from the relevant analyses; ***, study provided a Bland-Altman plot but did not report the numerical mean log difference, which we deemed necessary to be classified as a Bland-Altman analysis. Shaded cells indicate that the measure was reported in the cited study. Blank cells indicate that the measure was not reported.

DISCUSSION

Overall, our systematic review showed Xpert HIV-1 VL performs well compared with current established reference standard VL assays, both when measured by correlation and by Bland-Altman analysis. These findings might help inform policy guidance on this new assay, along with data on costs, feasibility, clinical impact, and cost-effectiveness.

Our systematic review has several potential limitations. First, as the search was conducted shortly after Xpert HIV-1 VL received WHO in vitro diagnostic prequalification, few validation studies had been conducted and published at the time of the search. Consequently, only 12 articles meeting our inclusion criteria were identified and included. We hope to update our meta-analysis in the future to account for new studies that will emerge on the Xpert VL assay. Second, the decision to exclude conference abstracts resulted in the exclusion of multiple relevant studies. Abstracts were excluded because study quality could not be effectively assessed, but as Xpert HIV-1 VL is a new tool, many studies on the assay have yet to be published as full-length articles. However, due to poor reporting, study quality was challenging to assess even in full-length articles. Where sufficient information was available, quality was high.

The final limitation pertains to generalizability. Many studies failed to report important demographic and clinical information, such as gender distribution, age of study participants, and patient ART status. This information is critical to understand the clinical relevance and generalizability of the reported data. Authors of studies with missing information were contacted; however, no responses were received. There was also substantial variability in which measures of agreement were calculated. To improve the ability to systematically review and pool the literature, a minimum set of standard measures of agreement should be agreed upon and used. Correlation and Bland-Altman could provide such a minimum set, as they measure both general agreement and magnitude and direction of bias between assays. Many studies reported percent agreement analyses where Xpert HIV-1 VL and the reference test were used to classify patients above or below certain thresholds. However, the specific thresholds used varied considerably, preventing a meaningful comparison. The use of a standard clinical threshold in future evaluations of Xpert HIV-1 VL would allow for direct comparisons in systematic reviews. The generalizability of our study is also limited due to the geographical distribution of the primary studies included; they were exclusively performed in countries where the predominant HIV subtype is C (e.g., India and Africa) or B (e.g., Israel, Europe, and the United States) (24). As more validation studies continue to be published on the Xpert assay, the results of this meta-analysis can be updated to include data from different countries where other HIV-1 subtypes predominate.

The purpose of this study was to assess VL agreement between Xpert and other gold standard assays as a measure of accuracy. However, a final recommendation for the use of Xpert VL will need to consider not just accuracy but also country-specific implementation research that addresses the feasibility of Xpert VL and its impact on patient outcomes.

The low-complexity nature of the Xpert HIV-1 VL assay, coupled with its ability to deliver same-day test results, make this technology particularly well suited for use in resource-constrained settings, where point-of-care VL testing is most needed (6). Deployment of such a tool, as well as utilization of preexisting Gene Xpert systems that are used in TB diagnostics, has the potential to increase access to VL testing, which will be necessary to achieve the 90-90-90 global targets for HIV/AIDS. Further research is needed to assess the impact of this VL assay on important patient outcomes and to establish its cost-effectiveness compared to VL assays currently used in centralized laboratories.

Supplementary Material

Supplemental material

ACKNOWLEDGMENTS

This study was supported by a TMA Pai Endowment Chair to M.P. at Manipal University, India, and the Manipal McGill Centre for Infectious Diseases (MAC ID). M.P. is also supported by a Canada Research Chair award. S.H. has a fellowship from the FRQS.

M.N. and S.H. conceptualized the article, wrote the protocol, and screened and extracted the papers. S.B. resolved disagreements in inclusion and extraction. S.H. performed the analysis. M.N. and S.H. drafted the manuscript. S.B., K.S., and M.P. critically reviewed the protocol and manuscript.

Footnotes

Supplemental material for this article may be found at https://doi.org/10.1128/JCM.01673-17.

REFERENCES

  • 1.UNAIDS. 2014. The gap report. UNAIDS, Geneva, Switzerland. [Google Scholar]
  • 2.Cepheid Technical Support. 2016. Xpert HIV-1 viral load training. Cepheid, Sunnyvale, CA: http://www.cepheid.com/en/cepheid-solutions/clinical-ivd-tests/virology/xpert-hiv-1-viral-load Accessed 11 December 2017. [Google Scholar]
  • 3.World Health Organization. 2017. WHO monitoring of Xpert MTB/RIF roll-out. World Health Organization, Geneva, Switzerland. http://www.who.int/tb/areas-of-work/laboratory/mtb-rif-rollout/en/ Accessed 9 February 2017.
  • 4.Cazabon D, Suresh A, Oghor C, Qin ZZ, Kik SV, Denkinger CM, Pai M. 2017. Implementation of Xpert MTB/RIF in 22 high tuberculosis burden countries: are we making progress? Eur Respir J 50:1700918. doi: 10.1183/13993003.00918-2017. [DOI] [PubMed] [Google Scholar]
  • 5.World Health Organization. 2016. Global tuberculosis report. World Health Organization, Geneva, Switzerland. [Google Scholar]
  • 6.Roberts T, Cohn J, Bonner K, Hargreaves S. 2016. Scale-up of routine viral load testing in resource-poor settings: current and future implementation challenges. Clin Infect Dis 62:1043–1048. doi: 10.1093/cid/ciw001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.World Health Organization. 2017. WHO prequalification of in vitro diagnostics public report. World Health Organization, Geneva, Switzerland. [Google Scholar]
  • 8.Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gøtzsche PC, Ioannidis JPA, Clarke M, Devereaux PJ, Kleijnen J, Moher D. 2009. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. PLoS Med 6:e1000100. doi: 10.1371/journal.pmed.1000100. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Rucker G, Schwarzer G, Carpenter JR, Schumacher M. 2008. Undue reliance on I(2) in assessing heterogeneity may mislead. BMC Med Res Methodol 8:79. doi: 10.1186/1471-2288-8-79. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Whiting PF, Rutjes AW, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, Leeflang MM, Sterne JA, Bossuyt PM. 2011. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med 155:529–536. doi: 10.7326/0003-4819-155-8-201110180-00009. [DOI] [PubMed] [Google Scholar]
  • 11.Ceffa S, Luhanga R, Andreotti M, Brambilla D, Erba F, Jere H, Mancinelli S, Giuliano M, Palombi L, Marazzi MC. 2016. Comparison of the Cepheid GeneXpert and Abbott M2000 HIV-1 real time molecular assays for monitoring HIV-1 viral load and detecting HIV-1 infection. J Virol Methods 229:35–39. doi: 10.1016/j.jviromet.2015.12.007. [DOI] [PubMed] [Google Scholar]
  • 12.Gous N, Scott L, Berrie L, Stevens W. 2016. Options to expand HIV viral load testing in South Africa: evaluation of the GeneXpert HIV-1 viral load assay. PLoS One 11:e0168244. doi: 10.1371/journal.pone.0168244. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.DerSimonian R, Laird N. 1986. Meta-analysis in clinical trials. Control Clin Trials 7:177–188. doi: 10.1016/0197-2456(86)90046-2. [DOI] [PubMed] [Google Scholar]
  • 14.Garrett NJ, Drain PK, Werner L, Samsunder N, Abdool Karim SS. 2016. Diagnostic accuracy of the point-of-care xpert HIV-1 viral load assay in a South African HIV clinic. J Acquir Immune Defic Syndr 72:e45–e48. doi: 10.1097/QAI.0000000000000978. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Gueudin M, Baron A, Alessandri-Gradt E, Lemee V, Mourez T, Etienne M, Plantier JC. 2016. Performance evaluation of the new HIV-1 quantification assay, Xpert HIV-1 viral load, on a wide panel of HIV-1 variants. J Acquir Immune Defic Syndr 72:521–526. [DOI] [PubMed] [Google Scholar]
  • 16.Jordan JA, Plantier JC, Templeton K, Wu AHB. 2016. Multi-site clinical evaluation of the Xpert HIV-1 viral load assay. J Clin Virol 80:27–32. doi: 10.1016/j.jcv.2016.04.014. [DOI] [PubMed] [Google Scholar]
  • 17.Mor O, Gozlan Y, Wax M, Mileguir F, Rakovsky A, Noy B, Mendelson E, Levy I. 2015. Evaluation of the realtime HIV-1, Xpert HIV-1, and Aptima HIV-1 Quant Dx assays in comparison to the NucliSens EasyQ HIV-1 v20 assay for quantification of HIV-1 viral load. J Clin Microbiol 53:3458–3465. doi: 10.1128/JCM.01806-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Moyo S, Mohammed T, Wirth KE, Prague M, Bennett K, Holme MP, Mupfumi L, Sebogodi P, Moraka NO, Boleo C, Maphorisa CN, Seraise B, Gaseitsiwe S, Musonda RM, van Widenfelt E, Powis KM, Gaolathe T, Tchetgen Tchetgen EJ, Makhema JM, Essex M, Lockman S, Novitsky V. 2016. Point-of-care Cepheid Xpert HIV-1 viral load test in rural African communities is feasible and reliable. J Clin Microbiol 54:3050–3055. doi: 10.1128/JCM.01594-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Bruzzone B, Caligiuri P, Nigro N, Arcuri C, Delucis S, Di Biagio A, Icardi G. 2017. Xpert HIV-1 viral load assay and versant HIV-1 RNA 1.5 assay: a performance comparison. J Acquired Immune Defic Syndr 74:e86–e88. doi: 10.1097/QAI.0000000000001227. [DOI] [PubMed] [Google Scholar]
  • 20.Nash M, Ramapuram J, Kaiya R, Huddart S, Pai M, Baliga S. 2017. Use of the GeneXpert tuberculosis system for HIV viral load testing in India. Lancet Glob Health 5:e754–e755. doi: 10.1016/S2214-109X(17)30247-4. [DOI] [PubMed] [Google Scholar]
  • 21.Avidor B, Matus N, Girshengorn S, Achsanov S, Gielman S, Zeldis I, Schweitzer I, Adler A, Turner D. 2017. Comparison between Roche and Xpert in HIV-1 RNA quantitation: a high concordance between the two techniques except for a CRF02_AG subtype variant with high viral load titters detected by Roche but undetected by Xpert. J Clin Virol 93:15–19. doi: 10.1016/j.jcv.2017.05.014. [DOI] [PubMed] [Google Scholar]
  • 22.Kulkarni S, Jadhav S, Khopkar P, Sane S, Londhe R, Chimanpure V, Dhilpe V, Ghate M, Yelagate R, Panchal N, Rahane G, Kadam D, Gaikwad N, Rewari B, Gangakhedkar R. 2017. GeneXpert HIV-1 quant assay, a new tool for scale up of viral load monitoring in the success of ART programme in India. BMC Infect Dis 17:506. doi: 10.1186/s12879-017-2604-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Swathirajan CR, Vignesh R, Boobalan J, Solomon SS, Saravanan S, Balakrishnan P. 2017. Performance of point-of-care Xpert HIV-1 plasma viral load assay at a tertiary HIV care centre in Southern India. J Med Microbiol doi: 10.1099/jmm.0.000514. [DOI] [PubMed] [Google Scholar]
  • 24.Hemelaar J, Gouws E, Ghys PD, Osmanov S. 2006. Global and regional distribution of HIV-1 genetic subtypes and recombinants in 2004. AIDS 20:W13–W23. doi: 10.1097/01.aids.0000247564.73009.bc. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplemental material

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)

RESOURCES