Skip to main content
HHS Author Manuscripts logoLink to HHS Author Manuscripts
. Author manuscript; available in PMC: 2018 Aug 22.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2018 Feb;65(2):137–148. doi: 10.1109/TUFFC.2017.2778566

Pressure Pulse Distortion by Needle and Fiber Optic Hydrophones due to Nonuniform Sensitivity

Keith A Wear 1., Yunbo Liu 1., Gerald R Harris 1.
PMCID: PMC6103641  NIHMSID: NIHMS984310  PMID: 29389648

Abstract

Needle and fiber optic hydrophones have frequency-dependent sensitivity, which can result in substantial distortion of nonlinear or broadband pressure pulses. A rigid cylinder model for needle and fiber optic hydrophones was used to predict this distortion. The model was compared with measurements of complex sensitivity for a fiber optic hydrophone and 3 needle hydrophones with sensitive element sizes (d) of 100, 200, 400, and 600 μm. Theoretical and experimental sensitivities agreed to within 12 ± 3% (RMS normalized magnitude ratio) and 8 ± 3 degrees (RMS phase difference) for the four hydrophones over the range from 1 – 10 MHz. The model predicts that distortions in peak positive pressure can exceed 20% when d0 < 0.5 and SI > 7% and can exceed 40% when d0 < 0.5 and SI > 14%, where λ0 is the wavelength of the fundamental component and SI (spectral index) is the fraction of power spectral density contained in harmonics. The model predicts that distortions in peak negative pressure can exceed 15% when d0 < 1. Measurements of pulse distortion using a 2.25 MHz source and needle hydrophones with d = 200, 400 and 600 μm agreed with the model to within a few percent on the average for SI values up to 14%. This work 1) identifies conditions for which needle and fiber optic hydrophones produce substantial distortions in acoustic pressure pulse measurements and 2) offers a practical deconvolution method to suppress these distortions.

Keywords: hydrophone, needle, fiber-optic, acoustic output measurement

I. Introduction

Accurate measurement of pressure fields is essential to ensure safety and effectiveness for diagnostic and therapeutic ultrasound devices [1]. Hydrophones are the most common devices for measurement of acoustic pressure fields. Membrane hydrophones are often preferred because they tend to exhibit less pressure pulse distortion than needle and fiber-optic hydrophones due to their relatively uniform (frequency-independent) sensitivity. However, membrane hydrophones are bulky and therefore impractical for many applications. In addition, membrane hydrophones, with their big planar surfaces, can create reflections that interfere with measurements.

Needle hydrophones are often more economical than membrane hydrophones. Needle and fiber-optic hydrophones are compact and can be inserted into small spaces inaccessible to membrane hydrophones. Compactness makes needle hydrophones preferable for in vitro testing of transcranial systems [24]. In addition, some needle [5] and fiber-optic [614] hydrophones are more robust for high-intensity therapeutic ultrasound (HITU) measurements than most membrane hydrophones, although robust membrane hydrophone designs also have been proposed [15, 16]. Fiber–optic hydrophones have been used extensively for calibration of HITU [1723] and lithotripsy [24] systems. Some fiber optic hydrophones have exceptional spatial resolution on the order of 100 μm or less, far superior to most membrane hydrophones. One challenge with needle and fiber optic hydrophones is that they exhibit non-uniform (frequency-dependent) sensitivity at low frequencies. However, pressure pulse distortion due to nonuniform sensitivity can be suppressed in post processing by applying deconvolution to hydrophone measurements [2530].

Many needle and fiber-optic hydrophones may be modeled as rigid, semi-infinite rods. This model is straightforward for needle hydrophones. This model also applies to many fiber-optic hydrophones for the following reasons. A common fiber-optic hydrophone design involves propagating an optical beam along a fiber core and measuring the optical reflection from the fiber tip / fluid interface. Pressure variations in the fluid induce density variations near the fiber tip / fluid interface. These density variations result in variations in the fluid index of refraction, which result in variations in the reflection coefficient encountered by the optical beam travelling down the fiber core and reflecting back from the tip. Since the fiber-optic hydrophone core typically has far higher characteristic impedance than the surrounding fluid, it may be accurately modeled as a rigid disk [31]. This model applies to many fiber-optic hydrophones, such as those commonly used in HITU [17, 18, 2123] and lithotripsy [24]. Other fiber-optic hydrophones, such as those based on multi-layer [32], distributed Bragg reflector [33], Fabry-Perot interferometer [3436] and other [37] designs, have a more complex response than the rigid, semi-infinite rods model predicts.

Jones studied the frequency response of a rigid, semi-infinite rod to a normally-incident plane wave and showed that the spatial-average (over the rod surface) pressure amplitude has a maximum response at fmax=1.2 c / (πa), where f = frequency, c = sound speed and a = cylinder radius, that is approximately double the response in the limit f << fmax [38]. An equivalent approach that uses vibrating planar piston theory [31] is discussed in Sec. II. The model does not simply average the free (hydrophone absent) field over a surface corresponding to the hydrophone sensitive element but also accounts for the fact that the presence of the hydrophone perturbs the field by imposing a boundary condition of zero normal component of particle velocity on the hydrophone sensitive element surface (as a consequence of the assumption of rigidity). This frequency-dependent response can introduce substantial pressure pulse distortion for broadband signals measured with needle and fiber-optic hydrophones. In addition, Jones obtained the phase response, which is also frequency-dependent and therefore contributes to pulse distortion. Measured frequency responses for typical commercial needle and fiber-optic hydrophones are consistent with this model at low frequencies, as will be shown in this paper. Some hydrophones incorporate features such as pre-amplification that cause them to deviate from the model. However, if the frequency dependence of these features is known, then the model can be adapted accordingly.

Figure 1 illustrates a potential problem of nonuniform response in the context of medical ultrasound signals. The top panel shows the spectrum of a nonlinear signal (blue) and the hydrophone frequency-dependent sensitivity (red). The bottom panel shows an unfiltered waveform (blue) and a waveform distorted by the hydrophone (red dash). The hydrophone boosts harmonics relative to the fundamental, resulting in a waveform with exaggerated high-frequency content, which is manifested as sharper, taller, compressional peaks.

Fig. 1.

Fig. 1.

a) spectrum of a nonlinear signal (blue) and the hydrophone frequency-dependent sensitivity (red). b) unfiltered waveform (blue) and a waveform distorted by the hydrophone (red dash). The hydrophone boosts harmonics relative to the fundamental, resulting in a waveform with reduced rarefactional pressure and exaggerated high-frequency content, which is manifested as sharper, taller, compressional peaks.

In this paper, the model is validated with measurements of complex sensitivity performed using time delay spectrometry (TDS) for a fiber-optic hydrophone and 3 needle hydrophones with sensitive element sizes (d) of 100, 200, 400, and 600 μm. A simulation based on the model is then validated with experimental measurements and used to predict distortions in measurements of peak compressional pressure (p+) peak rarefactional pressure (p-), and pulse intensity integral (pii) due to non-uniform sensitivity over a wide range of experimental parameters (sensitive element size, fundamental frequency, bandwidth, and nonlinearity). The likelihood for mechanical bioeffects is thought to be related to p- while the likelihood for thermal bioeffects is thought to be related to PII [39]. In histotripsy applications, in which mechanical fractionation of the tissue is accomplished using repetitive high intensity ultrasound pulses, p+ is important for distinguishing regimes of cavitation histotripsy and boiling histotripsy [20].

There are many sources of uncertainty in hydrophone measurements, including calibration uncertainty, spatial averaging, variation of hydrophone sensitivity with temperature, and hydrophone sensitivity drift over time. These affect all hydrophones, not just needle and fiber-optic hydrophones. The goal of the present study was to isolate the effects of nonuniform sensitivity of needle and fiber-optic hydrophones on estimates of p+, p-, and pii.

II. Model

A. Frequency-dependent Hydrophone Response

Needle and fiber optic response was modeled by spatially averaging acoustic pressure over the end of a rigid, semi-infinite, cylindrical rod [31, 38, 40]. One approach involves a solution for scattering of a plane harmonic scalar wave by a semi-infinite circular cylindrical rod of diameter 2a when the boundary condition is u=0 or ∂u/∂v=0, where u is the scalar field and v is the normal to the rod [38]. A related approach models a fiber optic hydrophone as a vibrating planar piston (fiber core) with radius a2 in an infinite rigid baffle (fiber cladding) with inner radius a2 and outer radius a1 [31]. The two approaches should give similar results when a = a1 = a2. The transfer function for the fiber optic hydrophone model is [31]

H(f)=21πa2202π0a2ej2πfb(r,θ)/crdrdθ

where f = frequency, c = sound speed, and

b(r,θ)=r2cos2θ+(a12r2)rcosθ.

In the present paper, both models [31, 38] were tested by comparison with our experimental measurements of hydrophone sensitivity based on time delay spectrometry (TDS). Pulse distortions for a wide variety of pulse shapes were estimated by subjecting simulated nonlinear diagnostic pulses to theoretical frequency-dependent complex sensitivities. Another model for needle hydrophone sensitivity accounts not only for diffraction but also for multiple reflections within the layered structure and includes properties of polyvinylidene fluoride (PVDF) such as attenuation [41]. This model, however, requires detailed knowledge of the hydrophone structural characteristics, which is not always available.

B. Nonlinear Pressure Pulse

The pressure pulses used in simulations were based on the model of Ayme and Carstensen for nonlinear medical ultrasound signals [42]. Their formula represents the signal as a sum of n sinusoids, each with frequency nfc (where fc is the center frequency for the fundamental component) and a π/4 phase shift, with amplitudes proportional to 1 / n. Zeqiri and Bond [43] modified this formula by changing the harmonic amplitude, 1 / n, to the form derived by Blackstock for plane periodic waves in a semi-infinite fluid, Bn(σ) [44]. With this modification, the acoustic pressure pulse, pi(t), has a continuously adjustable shock parameter, σ, to modulate the level of nonlinearity.

pi(t)=m(t)[n=1100Bn(σ)sin(2πnfct+π4)]

where t is time and

Bn(σ)=2nπVb+2nπσΦminπcosn(ΦσsinΦ)dΦ

For 0 ≤ σ < 1, the shock amplitude Vb = 0, Φmin = 0, and Bn(σ) = (2/nσ)Jn(nσ) where Jn is the ordinary Bessel function of order n. For σ > 1, Φmin = σVb and Vb is the first nonzero root to the transcendental equation Vb = sin(σVb) [43]. As pointed out by Zeqiri and Bond, Bacon has proposed a method for estimating σ from experimentally measured quantities [45, 46].

For the present study, an envelope function, m(t), was included to simulate the pulsed nature of medical ultrasound signals. m(t) had three phases: a rising portion proportional to [1 – exp(-t / t1)] followed by a constant middle portion followed by a decaying portion proportional to exp(-t / t2). The durations of the three phases of the envelope could be adjusted in order to control the fractional bandwidth of the simulated signals. One hundred harmonics were summed in order to result in nonlinear properties observed in medical ultrasound: compressional peaking, rarefactional rounding, and shock front formation. Signals were sampled at 2 GHz using Matlab (Natick, MA).

The degree of nonlinearity in the signal was quantified using the “spectral index,” SI, originally proposed by Duck [47].

SI=faP(f)df0P(f)df

where P(f) is the power spectrum and fa is a spectral boundary, chosen here to be equal to 1.5 times the center frequency of the fundamental spectral component in order to fall half way between the fundamental and the first harmonic. Another alternative would be to set fa to the frequency at which the spectral value first falls to a level 6 dB below its value at the center frequency [47], but this approach can be problematic for highly nonlinear, very wideband signals for which the -6 dB point can lie far outside the fundamental lobe. The spectral index reflects the fraction of power contained in frequencies above the fundamental.

Figure 2 shows the 16 signals tested in the simulation. Nonlinearity (SI) increases from left to right, and fractional bandwidth (-6 dB) of the fundamental lobe (BW0) increases from top to bottom. The center frequencies of the fundamental spectral lobes were chosen to avoid a simple monotonic relationship with either SI or BW0. It can be seen that as nonlinearity (SI) increases, compressional peaks become sharper while rarefactional peaks become more rounded. Figure 3 shows the corresponding spectra.

Fig. 2.

Fig. 2.

Signals tested in the simulation. SI = spectral index. BW0 = fractional bandwidth of the fundamental lobe. The center frequencies of the fundamental lobes are given in the right column of figures.

Fig. 3.

Fig. 3.

Spectra of signals tested in the simulation. SI = spectral index. BW0 = fractional bandwidth (-6dB) of the fundamental lobe. The center frequencies of the fundamental lobes are given in the right column of figures.

Although the relative amplitudes of the harmonics in the simulations were based on theory for plane waves [44] and the constant phase shift of π/4 for all harmonics in the simulations [42] may seem like an oversimplification, Figure 2 suggests that the signal model proposed here produces signals that are consistent with signals produced by focused transducers encountered in medical ultrasound [48].

Several authors have explored many metrics for quantification of nonlinearity [47, 4952]. Some metrics are convenient because they may be computed directly from simulated or measured signals, without knowledge of the source transducer or beam properties. These include the following. As mentioned above, the spectral index, SI, is the fraction of power contained in frequencies above the fundamental. The second harmonic ratio is the ratio of the amplitudes of the second to first harmonics in the spectrum. The asymmetric ratio, p+ / p- is the ratio of the peak compressional pressure to the peak rarefactional pressure. The Ostrovskii / Sutin propagation parameter, σs, is the ratio of the difference between p+ and p- to the sum of p+ and p-. The acoustic pulse crest factor is defined as p+/prms, where prms is the root-mean-square value of the pressure. Other proposed metrics such as the acoustic propagation parameter, σm, and the local distortion parameter, σq, have merits but require empirical measurements of system properties such as focal gain and local area factor (square root of the ratio of the source aperture to the beam area) [52].

The relationships among nonlinearity metrics depend on the detailed structure of the acoustic beam, including relative amplitudes and phases of the harmonics in the spectrum. For example, the time domain metrics are thought to be particularly sensitive to diffraction (that is, interference among harmonics) [52]. In addition, time-domain metrics would be expected to exhibit more dependence on hydrophone phase response (which is often unknown) than SI. Factors such as these lead to the concern that predictions of the model could depend on the particular choice of nonlinearity metric (in this case, SI). In order to investigate this possibility, SI, second harmonic ratio, p+ / p-, σs, and p+/prms were computed and compared for all simulated and experimental signals.

III. Experimental Methods

A. Sensitivity Measurements

Theoretical models were compared with measurements of complex sensitivities for four hydrophones without preamplifiers (see Table 1). The complex sensitivity of the fiber optic hydrophone was obtained from the Fourier transform of the impulse response provided by the manufacturer. Complex sensitivities for the three needle hydrophones were measured using a digital time-delay spectrometry (TDS) system previously described [27, 53].(There are many methods for measuring sensitivity magnitude but fewer methods for measuring sensitivity phase [5357], with some TDS methods extending to frequencies as high as 40 MHz [27, 56] and a nonlinear method providing phase measurements at harmonic frequencies up to 100 MHz [55] Some methods have shown that common hydrophone measurement systems can be accurately modeled as minimum phase systems [27, 53, 56].) A PVDF membrane hydrophone (HMB-200, Onda Corp., Sunnyvale, CA) was used for a reference in substitution experiments. The complex sensitivity of the reference membrane hydrophone was measured from 1 to 40 MHz by an independent laboratory (National Physical Laboratory, U.K.). TDS sensitivity measurements were performed using four different broadband source transducers (approximate center frequencies: 2 MHz, 5 MHz, 30 MHz, and 50 MHz) in order to obtain hydrophone magnitude and phase response over four frequencies bands (roughly 1–3 MHz, 3–6 MHz, 6–25 MHz, 25–40 MHz) that collectively spanned the range from 1 to 40 MHz. The transducers included (in order of ascending center frequency) 1) a custom bi-concave transducer [58], 2) a 1.27 cm diameter, 5 MHz transducer (IS0504HR, Valpey Fisher, Hopkinton, MA), 3) a 0.635 cm, 30 MHz transducer (ZF3001, Panametrics, Waltham, MA), and 4) a 0.635 cm, 50 MHz transducer (P150-2-R2.00, Olympus, Waltham, MA). For more detail, see [27].

Table I.

Hydrophones Used For Experimental Measurements

Manufacturer Model Type Sensitive Element Size d(μm)
Nominal Fit
RP Acoustics (Leutenbach, DE) FOPH 2000 Fiber Optic 100 128
Onda (Sunnyvale, CA) HNC-200 Ceramic Needle 200 176
Onda HNA-400 Ceramic Needle 400 308
Force Technology Inst. (Brondby, DK). Custom PVDF Needle 600 584

Model functions were fit to experimental data by minimizing the mean square difference between the model and the experimental sensitivity magnitude over a range of frequencies. The adjustable parameters in the fit were a multiplicative constant and sensitive element size d. The upper limit of the frequency range of the fit could be limited by two factors. First, one model only provided predictions up to a maximum frequency corresponding to ka = 10 [38] where k=2π/λ and λf=c. Second, as will be seen in the Results section, two needle hydrophones had design complexities that caused their frequency-dependent sensitivities to deviate from the simple cylinder model at frequencies much higher than the maximum response frequency, fmax = 2.4c/(πd). In these cases, the upper limit was set near the frequency at which the deviations began.

B. Pulse Distortion Measurements

The effects of nonuniform sensitivity on measurements of p+ and p- were investigated by comparing measurements of ultrasound fields in a water tank performed using the HMB-200 membrane hydrophone (with a relatively flat frequency response from 1 – 20 MHz) with measurements performed using the needle and fiber optic hydrophones. The source was a KB K12401 Aerotech (Lewistown, PA) single-element transducer with center frequency = 2.25 MHz, diameter = 1.91 cm, and focal distance = 10.2 cm. A Tektronix (Beaverton, OR) AFG 3102 dual channel arbitrary function generator and an Amplifier Research (Souderton, PA) Model 150A 100B 150W power amplifier were used to generate the electric signal that was fed into the source transducer. Voltage outputs from hydrophones were digitized at 1 GHz using a Tektronix 3012C Digital Phosphor Oscilloscope. The transducer was driven by 8-cycle tone bursts. This burst duration was much shorter than durations for HITU pulses but was long enough to achieve an approximately continuous wave condition in the center of the pulse, allowing measurements of pressure distortion relevant for HITU. In addition to driving the transducer at its resonance frequency of 2.25 MHz, measurements were performed at 1.75 MHz and 3 MHz in order to investigate different values for d / λ0 where λ0 is the wavelength corresponding to the fundamental frequency, fc. For each driving frequency and each hydrophone, measurements were performed at 6 different driving levels in order to span a range of signal nonlinearity (SI) but avoiding levels that could produce sufficiently high pressures to damage the hydrophones. The function generator peak-to-peak voltages were 31, 62, 93, 124, 156, 190 mV (1.75 MHz), and 20, 40, 60, 80, 100, 120 mV (2.25 and 3 MHz).

Voltage output signals were analyzed off-line using Matlab. Signals were low-pass filtered with a Gaussian global low-pass filter exp(-f2/2Σ2) with Σ = 40 MHz. Signals were normalized to their spectral magnitude at the driving frequency. This eliminated pressure discrepancies due to uncertainties in sensitivity magnitude, which has been reported to be on the order of ten percent [59], so that pressure discrepancies exclusively due to frequency-dependent sensitivity could be elucidated.

IV. Results

A. Hydrophone Sensitivity Measurements

Figure 4 shows experimental measurements and theoretical fits for sensitivities of the four hydrophones. Since the complex sensitivity of the fiber optic hydrophone was obtained from the Fourier transform of the impulse response provided by the manufacturer, the experimental values in the top row of Figure 4 go all the way down to zero frequency. The experimental values for the three needle hydrophones do not include values below 1 MHz because of experimental signal-to-noise ratio limitations. Similar high-pass filter behavior for low frequencies (f < fmax) has been measured for needle [5, 26, 27, 29, 40, 41, 60] and fiber optic [8, 14] hydrophones previously.

Fig. 4.

Fig. 4.

Theoretical and experimental forms for hydrophone sensitivities for the fiber optic hydrophone (top row) and 3 needle hydrophones. The two models are “Model 1” [38] and “Model 2” [31]. The arrows show the maximum response at ka = 2.4 or, equivalently, f = 2.4 c / (πd).

The two models, “Model 1” [38] and “Model 2” [31], gave nearly identical results. Theoretical and experimental sensitivities agreed to within 12 ± 3% (mean ± standard deviation of RMS magnitude difference) and 8 ± 3 degrees (mean ± standard deviation of RMS phase difference) for the four hydrophones. The effective sensitive element diameter was an adjustable parameter in the fits and is shown in Table I. The effective sensitive element diameter was always within 28% of the nominal value. The maximum sensitivity is expected at fmax = 2.4c/(πd) [38]. The optimal estimates of sensitive element size based on the minimum least-squares model fits to experimental sensitivity gave values of d of 128, 176, 308, and 584 μm for the four hydrophones, which corresponds to fmax =8.8, 6.4, 3.7, and 1.9 MHz, consistent with Fig. 4. Needle #3 has a peak between 1 and 2 MHz, which is consistent with simulation based on the multilayered structure model [41]. In order to capture the high-pass filter behavior for rigid cylindrical hydrophones with sensitive elements larger than 600 μm, sensitivity should be measured at frequencies below 1 MHz [61].

As can be seen in Fig. 1a, the cylinder model predicts an approximate doubling of sensitivity magnitude between very low frequencies (f << fmax) and the maximum response frequency, fmax, followed by relatively constant sensitivity magnitude for frequencies above fmax. Fig. 4 shows that Needle #1 and Needle #2 are consistent with the rigid cylinder model for frequencies up to 10 MHz, but show diminished sensitivity for frequencies above 10 MHz compared with the nearly-constant sensitivity predicted by the cylinder model. This disparity is presumably due to design features that deviate from the simple cylinder model. In these cases, the upper limits of the displayed model values were set near 10 MHz. As can be seen in Fig. 3, most of the spectral energy in the test signals was contained in frequencies below 10 MHz.

Figure 5 shows relationships, obtained from simulation and experiments, between the spectral index, SI, and four other metrics to describe nonlinearity. All metrics appear to obey simple monotonic relationships with SI. Although relationships among nonlinearity metrics in principle depend on the structure of an acoustic beam, including relative amplitudes and phases of the harmonics in the spectrum, the simulated and experimental values seem very consistent with each other, supporting the relevance of the simulation signal model. Although the relative amplitudes of the harmonics in the simulations were based on theory for plane waves [44], the signal model used here seems consistent with experimental data acquired using a focused transducer [48]. Although the constant phase shift of π/4 for all harmonics in the simulations [42] may seem like an oversimplification, the signal model shows nonlinearity metrics (both time-domain and frequency-domain) similar to experimentally measured values.

Fig. 5.

Fig. 5.

Metrics to quantify nonlinearity for the signals studied in this paper, obtained from simulation and experiment. Power law fits to simulated data (y = A SIn + B; solid lines) are shown where A and n were fitting parameters and the intercept value B was forced to the theoretical value for a purely linear signal (0, 1, 0, and 2 respectively). Cubic polynomial fits (y = a3 SI3 + a2 SI2 + a1 SI + B; dashed lines) are shown where a3, a2, and a1 were fitting parameters.

B. Acoustic Output Measurements

Figure 6 shows distortions in simulated peak compressional pressure before and after application of the model sensitivity function. The results are shown as functions of d / λ0 and SI where λ0 is the wavelength corresponding to the fundamental frequency, fc. Experimental needle hydrophone data points are plotted in the panel in Figure 6 with SI value (3%, 7%, 14%, or 22%) closest to that obtained from companion membrane hydrophone measurements, provided the experimental SI value was less than 3% above and no more than 1.5% below the simulated value (3%, 7%, 14%, or 22%). It can be seen that the distortion in peak compressional pressure caused by the non-uniform hydrophone sensitivity is primarily due to two factors: d / λ0 and SI. Center frequency and fractional bandwidths of the signals were far less influential in distorting peak compressional pressure. Experimental measurements were consistent with the simulations. Data were not acquired at SI = 22% because the required intensities were feared to potentially damage hydrophones.

Fig. 6.

Fig. 6.

Percent changes in simulated peak compressional pressure before and after application of the model sensitivity function. Experimental data points correspond to measurements acquired with the Onda HNC-200 (circles), Onda HNA-400 (asterisks), and Danish Institute of Biomedical Engineering (squares) needle hydrophones respectively.

Figure 7 shows distortions in simulated peak rarefactional pressure before and after application of the model sensitivity function. As with peak compressional pressure, d / λ0 is an important factor in determining distortion of peak rarefactional pressure. In contrast with peak compressional pressure, fractional bandwidth, BW0, is a more important determinant of distortion of peak rarefactional pressure than SI. Therefore, Fig. 7 is drawn with four panels corresponding to different values for BW0 instead of SI (as in Fig. 6). Experimental data are not shown in Fig. 7 because they would obscure simulation results but are shown on a zoomed scale in Fig. 8 instead.

Fig. 7.

Fig. 7.

Percent changes in simulated peak rarefactional pressure before and after application of the model sensitivity function.

Fig. 8.

Fig. 8.

Percent changes in simulated peak rarefactional pressure before and after application of the model sensitivity function for BW0 = 17%. Experimental data points correspond to measurements acquired with the Onda HNC-200 (circles), Onda HNA-400 (asterisks), and Danish Institute of Biomedical Engineering (squares) needle hydrophones respectively.

Figure 8 shows distortions in simulated peak rarefactional pressure before and after application of the model sensitivity function for BW0=17%, along with experimental data. The biggest discrepancies occur for 0.2 < d / λ0 < 0.6.

Fig. 9 shows percent changes in simulated pulse intensity integral before and after application of the model sensitivity function. As with peak compressional pressure, the distortion in pulse intensity integral caused by the non-uniform hydrophone sensitivity is primarily due to d / λ0 and SI Comparison of Fig.s 6 and 9 show similar trends of distortions of p+ and pii as functions of SI. This may be explained by the fact that as SI increases, the time-domain waveform becomes increasingly asymmetric with respect to the time axis (see Fig. 2) and therefore p+ has an increasing relative influence (compared with p-) on pii.

Figure 9.

Figure 9.

Percent changes in simulated pressure pulse intensity integral before and after application of the model sensitivity function.

V. Discussion

Needle and fiber optic hydrophones are desirable for many applications in medical ultrasound but can distort acoustic pressure pulses. The rigid cylindrical rod model presented here is accurate for predicting responses of the needle and fiber-optic hydrophones investigated for frequencies up to at least 1.5 fmax. This high-pass filter behavior has been previously shown for needle [5, 26, 27, 29, 40, 41, 60] and fiber optic [8, 14] hydrophones. This work identifies conditions for which nonuniform sensitivity of needle and fiber optic hydrophones results in substantial distortions in acoustic pressure pulse measurements. Knowledge of these conditions can help inform the appropriate choice for a hydrophone for a particular measurement.

Comparison of Fig.s 1a and 4 shows that the cylinder model was consistent with experimental sensitivities for all four hydrophones investigated at frequencies up to and beyond the maximum sensitivity magnitude frequency, fmax= 2.4c/(xd). However, two needle hydrophones showed sensitivity magnitude reductions compared to the model for frequencies above 10 MHz. As can be seen from Fig. 3, the test signals used in this investigation contained relatively little energy above 10 MHz so the differences between theoretical and experimental sensitivities were not consequential for this investigation. However, for signals with substantial fractions of energy above 10 MHz (say, over 25%), a modified model or the empirical sensitivity might improve accuracy of predictions of output pressure pulses. Still, the cylinder model faithfully modelled sensitivities for frequencies up to at least 1.5 fmax for the needle hydrophones investigated. Moreover, for HITU signals, fiber optic hydrophones are generally preferred due to their superior robustness to high pressures (e.g. > 10 MPa) [1724], and the cylinder model performs very well for frequencies up to 100 MHz for the fiber optic hydrophone investigated, as may be seen in Fig. 10 (maximum relative magnitude error = 23% and maximum phase error = 13 degrees).

Fig. 10.

Fig. 10.

Experimental and theoretical sensitivity for the RP Acoustics FOPH 2000 fiber-optic hydrophone from 0 to 100 MHz. The experimental sensitivity of the fiber-optic hydrophone was obtained from the Fourier transform of the impulse response provided by the manufacturer.

Distortions in measurements of peak positive pressure, Δp+, due to nonuniform sensitivity are more sensitive to d0 and SI then BW0. They are highest when d0 < 0.5 and increase with increasing SI. Distortions in measurements of peak negative pressure, Δp-, due to nonuniform sensitivity tend to be smaller than Δp+ and are highest when d0 < 1. Figure 11 illustrates why Δp+ is much more sensitive to nonlinearity than Δp-. When a nonlinear medical ultrasound signal, which typically has sharp compressional peaks and rounded rarefactional segments, is decomposed into compressional and rarefactional components, spectral analysis shows that the nonlinear content of the signal is carried disproportionately by the compressional component.

Fig. 11.

Fig. 11.

A nonlinear tone burst decomposed into compressional and rarefactional components (left). Spectra (right) show that the nonlinearity is carried disproportionately by the compressional component.

It is commonly thought that it is desirable to use a hydrophone with sensitive element as small as possible (provided sufficient sensitivity is achieved) in order to minimize spatial averaging artifacts and to allow finely-spaced contiguous measurements in a pressure field scan. However, Fig.s 69 show that reductions in sensitive element size (d) can actually increase signal distortion for needle and fiber optic hydrophones.

Common robust fiber-optic hydrophones have sensitive element sizes around d = 100 μm [10, 17, 21, 23, 62]. Highly nonlinear waveforms in combination with values of d0 near 0.1–0.4 can result in considerable potential distortion in peak compressional pressure (see Fig. 6, lower right panel). Such values are achievable in well-validated, extremely promising approaches including histotripsy (e.g., d0 ≈ 0.1 at 1.5 MHz [23] and d0 ≈ 0.07 at 1 MHz [62]) and other HITU applications (e.g., d0 ≈ 0.14 at 2 MHz [17] and d0 ≈ 0.08 at 1.2 MHz [21]).

When distortions reach unacceptable levels, they can be suppressed by applying deconvolution of the frequency-dependent hydrophone sensitivity. The value of deconvolution in the context of hydrophone measurements is appreciated in the ultrasound metrology community [17, 19, 21, 2530, 63, 64] and to some extent outside that community [65, 66]. Generally speaking, in order to perform a deconvolution, the magnitude and the phase of the hydrophone sensitivity must be known over all frequencies present in the signal spectrum. In the event that the hydrophone manufacturer provides only limited sensitivity information (e.g. magnitude from 1 – 20 MHz but not phase information), the model presented in this paper provides a method to estimate sensitivity phase and to extrapolate magnitude to unspecified frequencies (e.g. 0 – 1 MHz, and above 20 MHz).

Limitations of this work include the following. First, this analysis only considers uncertainties (or pressure discrepancies) due to nonuniform sensitivity. It does not consider discrepancies due to uncertainties in the magnitude of sensitivity, which are on the order of 10 percent [59]. Second, this analysis does not consider spatial averaging effects, which can be a problem for highly focused nonlinear beams.

Previous investigations of pressure pulse distortion by membrane hydrophones due to nonuniform response [67, 68] played an influential role in acoustic output measurement standards [6971]. The present paper extends these ideas to needle and fiber-optic hydrophones. While low-frequency (e.g., f < 1 MHz) response of membrane hydrophones is dominated by capacitance of the hydrophone and associated electronics [70, 72, 73], low-frequency (f < fmax) responses of needle and fiber optic hydrophones are dominated by diffraction effects related to the dimensions of the needle or fiber.

VI. Conclusion

Sensitivities for many needle and fiber-optic hydrophones commonly used in medical ultrasound can be accurately predicted using a rigid rod model for frequencies up to 1.5 fmax and beyond. Distortions in measurements of peak positive pressure due to nonuniform sensitivity can exceed 20% when d/x0 < 0.5 and SI > 7% and can exceed 40% when d0 < 0.5 and SI > 14%, where d is the sensitive element size, λ0 is the wavelength of the fundamental component, and SI (spectral index) is the fraction of power spectral density contained in harmonics. Distortions in peak negative pressure due to nonuinform sensitivity can exceed 15% when d0 < 1 and are particularly high for very broadband signals. When these experimental circumstances cannot be avoided, deconvolution of hydrophone sensitivity from measurements can be used to suppress these distortions. The rigid rod model provides a method to extrapolate sensitivity magnitude data and to generate sensitivity phase data to facilitate complex deconvolution.

Acknowledgment

The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. The authors are grateful for funding support from the FDA Office of Women’s Health.

Biographies

graphic file with name nihms-984310-b0001.gif

Keith A. Wear received his B.A. in Applied Physics from the University of California at San Diego in 1980. He received his M.S. and Ph.D. in Applied Physics with a Ph.D. minor in Electrical Engineering from Stanford University in 1982 and 1987. He was a post-doctoral research fellow with the Physics department at Washington University, St. Louis from 1987–1989. He has been a research physicist specializing in biomedical ultrasound at the FDA since 1989. He is an Associate Editor of 3 journals: Journal of the Acoustical Society of America; IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control; and Ultrasonic Imaging. He was the Technical Program Chair of the 2008 IEEE International Ultrasonics Symposium in Beijing, China. He was the General Program Chair of the 2017 IEEE International Ultrasonics Symposium in Washington, DC. He was chair of the American Institute of Ultrasound in Medicine (AIUM) Technical Standards Committee from 2014–2016. He is a Fellow of the Acoustical Society of America, the American Institute for Medical and Biological Engineering, and the AIUM. He is a senior member of IEEE.

graphic file with name nihms-984310-b0002.gif

Yunbo Liu received B.S. in Applied Mechanics from the Peking University, Beijing, China in 2001 and Ph.D. in Mechanical Engineering from Duke University, Durham, North Carolina in 2006. His research activity was focused on therapeutic ultrasound for cancer treatment. Since 2006 he has been employed as a research engineer by the Center for Devices and Radiological Health of Food and Drug Administration, Silver Spring, MD. His main research area is experimental evaluation of biomedical ultrasound exposimetry and safety.

graphic file with name nihms-984310-b0003.gif

Gerald R. Harris (M’72-S’76-M’79-SM’82-F’94-LF’11) was born in Jacksonville, NC, on November 22, 1945. He received the bachelor’s degree in electrical engineering in 1967 from the Georgia Institute of Technology, Atlanta, the M.S. degree in biological engineering in 1971 from the Rose-Hulman Institute of Technology, Terre Haute, IN, and the Ph.D. degree in electrical engineering in 1982 from the Catholic University of America, Washington, DC.

He is retired from the Food and Drug Administration’s Center for Devices and Radiological Health, Silver Spring, MD, where his main activities comprised the experimental and theoretical evaluation of medical ultrasound transducers and systems.

Dr. Harris is a member of Sigma Xi and has been elected to the rank of Fellow of the Acoustical Society of America, the American Institute of Ultrasound in Medicine, the American Institute for Medical and Biological Engineering, and the Institute of Electrical and Electronics Engineers (IEEE Life Fellow).

References

  • [1].Harris GR, “Progress in medical ultrasound exposimetry,” IEEE Trans. Ultrason. Ferroelectr. Freq. Control vol. 52, no. 5, pp. 717–736. Sept. 2005. [DOI] [PubMed] [Google Scholar]
  • [2].White J, Clement GT, and Hynynen K, “Transcranial ultrasound focus reconstruction with phase and amplitude correction,” IEEE Trans. IEEE Trans. Ultrason., Ferroelect., Freq. Contr, vol. 52, no. 9, pp. 1518–1522, Sept. 2005. [DOI] [PubMed] [Google Scholar]
  • [3].Samiotaki G, Vlachos F, Tung Y, and Konofagou EE, “A quantitative pressure and microbubble-size dependence study of focused ultrasound-induced blood-brain barrier opening reversibility in vivo using MRI,” Mag. Res. Med, vol. 67, pp. 769–777, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Robertson J, Martin E, Cox B, and Treeby BE, “Sensitivity of simulated transcranial ultrasound fields to acoustic medium property maps,” Phys. Med. & Biol In press. [DOI] [PubMed] [Google Scholar]
  • [5].Zanelli CI and Howard SM, “A robust hydrophone for HIFU metrology,” AIP Conf. Proc vol. 829, 618, 2006, DOI: 10.1063/1.2205548 [DOI] [Google Scholar]
  • [6].Staudenraus J and Eisenmenger W, “Fiber-optic probe hydrophone for ultrasonic and shock-wave measurements in water,” Ultrasonics, vol. 31, no. 4, pp. 267–273, July, 1993. [Google Scholar]
  • [7].Koch C, Molenstruck W, and Reibold R, “Shock-wave measurement using a calibrated interferometric fiber-tip sensor,” Ultrasound Med. & Biol, vol. 23, no. 8, pp. 1259–1266, 1997. [DOI] [PubMed] [Google Scholar]
  • [8].Wang ZQ, Lauxmann P, Wurster C, Kohler M, Gompf B, and Eisenmenger W, “Impulse response of a fiber-optic probe hydrophone determined with shock waves in water,” J. Appl. Phys, vol. 85, no. 5, pp. 2514–2516, March, 1999, DOI: 10.1063/1.369569. [DOI] [Google Scholar]
  • [9].Lewin PA, Mu C, Umchid S, Daryoush A, and El-Sherif M, “Acousto-optic, point receiver hydrophone probe for operation up to 100 MHz,” Ultrasonics vol. 43, pp. 815–821, 2005. [DOI] [PubMed] [Google Scholar]
  • [10].Parsons JE, Cain CA, and Fowlkes JB, “Cost-effective assembly of a basic fiber-optic hydrophone for measurement of high-amplitude therapeutic ultrasound fields,” J. Acoust. Soc. Am, vol. 119, no. 3, pp. 1432–1440, March, 2006, DOI: 10.1121/1.2166708 [DOI] [PubMed] [Google Scholar]
  • [11].Zhou Y, Zhai L, Simmons R, and Zhong P, “Measurement of high intensity focused ultrasound fields by a fiber optic probe hydrophone,” J. Acoust. Soc. Am, vol. 120, no. 2, 676–685, 2006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Umchid S, Gopinath R, Srinivasan K, Lewin PA, Daryoush AS, Bansal L, and El-Sherif M, “Development of calibration techniques for ultrasonic hydrophone probes in the frequency range from 1 to 100 MHz.” Ultrasonics, vol. 49, pp. 306–311. 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Haller J, Wilkens V, Jenderka K, and Koch C, “Characterization of a fiber-optic displacement sensor for measurement in high-intensity focused ultrasound fields,” J. Acoust. Soc. Am, vol. 129, no. 6, pp. 3676–3681, June, 2011. [DOI] [PubMed] [Google Scholar]
  • [14].Howard SM, “Calibration of reflectance-based fiber-optic hydrophones,” IEEE Symp. Proc. 2016, 978-1-4673-9897-8/16/
  • [15].Bessonova OV and Wilkens V, “Membrane hydrophone measurement and numerical simulation of HIFU fields up to developed shock regimes,” IEEE Trans. Ultrason., Ferroelect., Freq. Contr, vol. 60, no. 2, pp. 290–300, Feb., 2013. [DOI] [PubMed] [Google Scholar]
  • [16].Wilkens V, Sonntag S, and Georg O, “Robust spot-poled membrane hydrophones for measurement of large amplitude pressure waveforms generated by high intensity therapeutic ultrasonic transducers,” J. Acoust. Soc. Am, vol. 139, no. 3, pp. 1319–1332, March, 2016. [DOI] [PubMed] [Google Scholar]
  • [17].Canney MS, Bailey MR, Crum LA, Khokhlova VA, and Sapozhnikov OA, “Acoustic characterization of high intensity focused ultrasound fields: a combined measurement and modeling approach,” J. Acoust. Soc. Am, vol. 124, no. 4, pp. 2406–2420, Oct., 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Maxwell AD, Wang T, Cain CA, Fowlkes JB, Sapozhnilov OA, Bailey MR, and Xu Z, “Cavitation clouds created by shock scattering from bubbles during histotripsy,” J. Acoust. Soc. Am, vol. 130, no. 4, pp. 1888–1898, Oct. 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Haller J, Jenderka K, Durando G, and Shaw A, “A comparative evaluation of three hydrophones and a numerical model in high intensity focused ultrasound fields,” J. Acoust. Soc. Am, vol. 131, no. 2, pp. 1121–1130, Feb., 2012. [DOI] [PubMed] [Google Scholar]
  • [20].Maxwell A, Sapozhnikov O, Bailey M, Crum L, Xu Z, Fowlkes B, Cain C, and Khokhlova V, “Disintegration of tissue using high intensity focused ultrasound: two approaches that utilize shock waves,” Acoustics Today, vol. 8, no. 4, pp. 24–37, Oct. 2012. [Google Scholar]
  • [21].Kreider W, Yuldashev PV, Sapozhnikov OA, Farr N, Partanen A, Bailey BR, and Khokhlova VA, Characterization of a multi-element clinical HIFU system using acoustic holography and nonlinear modeling,” IEEE Trans. Ultrason., Ferroelect., Freq. Contr, vol. 60, no. 8, pp. 1683–1698, Aug., 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Jing Y, Cannata J, and Wang T, “Experimental verification of transient nonlinear acoustical holography,” J. Acoust. Soc. Am, vol. 133, no. 5, pp. 2533–2540, May, 2013. [DOI] [PubMed] [Google Scholar]
  • [23].Vlaisavljevich E, Lin K, Maxwell A, Warnez MT, Mancia L, Singh R, Putnam AJ, Fowlkes B, Jonsen E, Cain C, and Xu Z, “Effects of ultrasound frequency and tissue stiffness on the histotripsy intrinsic threshold for cavitation,” Ultrasound. Med. & Biol, vol. 41, no. 6, pp. 1651–1667, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Maxwell AD, Cunitz BW, Kreider W, Sapozhnikov OA, His RS, Harper JD, Bailey MR, and Sorensen MD, “Fragmentation of urinary calculi in vitro by burst wave lithotripsy,” J. Urology, vol. 193, pp. 338–344, Jan., 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Hurrell A, “Voltage to pressure conversion: are you getting ‘phased’ by the problem?” J. Phys. conf. Series, Adv. Metrol., Ultrasound Med, pp. 57–62, Apr. 27–28 2004. [Google Scholar]
  • [26].Wilkens V and Koch C “Amplitude and phase calibration of hydrophones up to 70 MHz using broadband pulse excitation and an optical reference hydrophone.” J. Acoust. Soc. Am vol. 115, no. 6, pp. 2892–2903, June 2004. [Google Scholar]
  • [27].Wear KA, Gammell PM, Maruvada S, Liu Y, and Harris Gerald, “Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity,” IEEE Trans. Ultrason., Ferro., Freq. Contr vol. 61, pp. 62–75, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28].Wear KA, Liu Y, Gammell PM, Maruvada S, and Harris GR, “Correction for frequency-dependent hydrophone response to nonlinear pressure waves using complex deconvolution and rarefactional filtering: application with fiber optic hydrophones,” IEEE Trans. Ultrason., Ferro., Freq. Contr vol. 62, no. 1, pp. 152–164, Jan. 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Hurrell AM and Rajagopal S, “The practicalities of obtaining and using hydrophone calibration data to derive pressure waveforms,” IEEE Trans. Ultrason., Ferro., Freq. Contr, vol. 64, no. 1, pp. 126–140, Jan., 2017. [DOI] [PubMed] [Google Scholar]
  • [30].Eichstadt S, Wilkens V, Dienstfrey A, Hale P, Hughes B, and Jarvis C, “On challenges in the uncertainty evaluation for time-dependent measurements,” Metrologia, vol. 53, pp. S125–S135, 2016. [Google Scholar]
  • [31].Krucker JF, Eisenberg A, Krix M, Lotsch R, Pessel M, and Trier H, “Rigid piston approximation for computing the transfer function and angular response of a fiber-optic hydrophone,” J. Acoust. Soc. Am, vol. 107, no. 4, pp. 1994–2003, April, 2000. [DOI] [PubMed] [Google Scholar]
  • [32].Weise W, Wilkens V, and Koch C, “Frequency response of fiber-optic multilayer hydrophones: experimental investigation and finite element simulation,” IEEE Trans. Ultrason., Ferro., Freq. Contr, vol. 49, no. 7, pp. 937–946, July, 2002. [DOI] [PubMed] [Google Scholar]
  • [33].Guan B, Tam H, Lau S, and Chan HLW, “Ultrasonic hydrophone based on distributed Bragg reflector fiber laser,” IEEE Photonics Tech. Let, vol. 16, no. 1, pp. 169–171, Jan., 2005. [Google Scholar]
  • [34].Morris P, Hurrell A, Shaw A, Zhang E, and Beard P, “A Fabry-Perot fiber-optic ultrasonic hydrophone for the simultaneous measurement of temperature and acoustic pressure,” J. Acoust. Soc. Am, vol. 125, no. 6, pp. 3611–3622, June 2009. [DOI] [PubMed] [Google Scholar]
  • [35].Bull V, Civale J, Rivens I, and ter Haar G, “A comparison of acoustic cavitation detection thresholds measured with piezo-electric and fiber-optic hydrophone sensors,” Ultrasound in Med. & Biol, vol. 39, no. 12, pp. 2406–2421, 2013. [DOI] [PubMed] [Google Scholar]
  • [36].Kim K, Mizuno Y, Nakamura K, “Fiber-optic ultrasonic hydrophone using short Fabry-Perot cavity with multilayer reflectors deposited on small stub,” Ultrasonics, vol. 54, pp. 1047–1051, 2014. [DOI] [PubMed] [Google Scholar]
  • [37].Vannacci E, Granchi S, Belsito L, Roncaglia A, and Biagi E, “Wide bandwidth fiber-optic ultrasound probe in MOMS technology: preliminary signal processing results, Ultrasonics, vo. 75, pp. 164–173, 2017 [DOI] [PubMed] [Google Scholar]
  • [38].Jones DS, “The scattering of a scalar wave by a semi-infinite rod of circular cross section,” Phil. Trans. Royal Soc. London, Series A, Math. And Phys. Sciences, vol. 247, no. 934, pp. 499–528, April, 1955. [Google Scholar]
  • [39].International Electrotechnical Commission, “Medical electrical equipment Part 2–37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment,” Ed. 2.1, Geneva, 2015. [Google Scholar]
  • [40].Harris GR and Gammell PM, “Sensitivity measurements of piezoelectric polymer hydrophones from 0.2 – 2 MHz using a broadband-pulse technique,” J. Acoust. Soc. Am, vol. 105, no. 2, Feb., pp. 725–731, 1999. [Google Scholar]
  • [41].Fay B, Ludwig G, Lankjaer C, and Lewin PA, “Frequency response of PVDF needle-type hydrophones,” Ultrasound in Med. & Biol, vol. 20, no. 4, pp. 361–366, 1994. [DOI] [PubMed] [Google Scholar]
  • [42].Ayme EJ and Carstensen EL, “Cavitation induced by asymmetric distorted pulses of ultrasound: Theoretical predictions.” lEEE Trans. Ultrason. Ferroelec. Freq. Contr, vol. 36, no. 1, pp. 32–40, Jan. 1989. [DOI] [PubMed] [Google Scholar]
  • [43].Zeqiri B and Bond AD, “The influence of waveform distortion on hydrophone spatial-averaging corrections—Theory and measurement,” J. Acoust. Soc. Am, vol. 92, no. 4, pp. 1809–1821, Oct. 1992. [Google Scholar]
  • [44].Blackstock DT, “Connection between the Fay and Fubini solutions for plane sound waves of finite amplitude,” J. Acoust. Soc. Am, vol. 39, no. 6, pp. 1019–1026, 1966. [Google Scholar]
  • [45].Bacon DR, “Finite amplitude distortion of the pulsed fields used in diagnostic ultrasound,” Ultrasound Med. & Biol, vol. 10, no. 2, pp. 189–195, 1984. [DOI] [PubMed] [Google Scholar]
  • [46].Bacon DR, “Prediction of in situ exposure to ultrasound: An improved method,” Ultrasound Med. & Biol, vol. 15, no. 4, pp. 355–361, 1989. [DOI] [PubMed] [Google Scholar]
  • [47].Duck FA, “Estimating in situ exposure in the presence of acoustic nonlinearity,” J. Ultrasound Med, vol. 18, pp. 43–53, 1999. [DOI] [PubMed] [Google Scholar]
  • [48].Szabo TL, Diagnostic Ultrasound Imaging: Inside Out, Second Ed., Academic Press, USA, 2014, ch. 12. [Google Scholar]
  • [49].Szabo TL, Clougherty F, and Grossman C, “Effects of nonlinearity on the estimation of in situ values of acoustic output parameters,” J. Ultrasound Med, vol. 18, pp. 33–41, 1999. [DOI] [PubMed] [Google Scholar]
  • [50].Bigelow TA, and O’Brien WD Jr., “Experimental evaluation of indicators of nonlinearity for use in ultrasound transducer characterizations,” Ultrasound Med. & Biol, vol. 28, no. 11/12, pp. 1509–1520, 2002. [DOI] [PubMed] [Google Scholar]
  • [51].Duncan T, Humphrey VF, and Duck FA, “A numerical comparison of nonlinear indicators for diagnostic ultrasound fields,” Proc. World Cong. Ultrason, Sept. 7–10, Paris, 2003. [Google Scholar]
  • [52].International Electrotechnical Commission Technical Specification IEC/TS 61949, “Ultrasonics – Field characterization – In situ exposure estimation in finite-amplitude ultrasonic beams,” IEC; Geneva, Switzerland, 2007. [Google Scholar]
  • [53].Wear KA, Gammell PM, Maruvada S, Liu Y, and Harris GR, “Time-delay-spectrometry measurement of magnitude and phase of hydrophone response,” IEEE Trans. Ultrason., Ferro., Freq. Cont, vol. 58, no. 11, pp. 2325–2333, Nov. 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [54].Cooling MP and Humphrey VF, “A nonlinear propagation model-based phase calibration technique for membrane hydrophones,” IEEE Trans. Ultrason., Ferro., Freq. Cont, vol. 55, no. 1, pp. 84–93, Jan., 2008. [DOI] [PubMed] [Google Scholar]
  • [55].Bloomfield PE, Gandhi G, and Lewin PA, “Membrane hydrophone phase characteristics through nonlinear acoustics measurements,” IEEE Trans. Ultrason., Ferro., Freq. Cont, vol. 58, no. 11, pp. 2418–2437, Nov., 2011. [DOI] [PubMed] [Google Scholar]
  • [56].Bloomfield PE and Lewin PA, “Determination of ultrasound hydrophone phase from Fourier-Hilbert transformed 1 to 40 MHz time delay spectrometry amplitude,” IEEE Trans. Ultrason., Ferro., Freq. Cont, vol. 61, no. 4, pp. 662–672, April, 2014. [Google Scholar]
  • [57].Olveira EG, Costa-Felix RPB, and Machado JC, “Primary reciprocity-based method for calibration of hydrophone magnitude and phase sensitivity: complete tests at frequencies from 1 to 7 MHz,” Ultrasonics, vol. 58, pp. 87–95, 2015. [DOI] [PubMed] [Google Scholar]
  • [58].Harris GR and Gammell PM, “1–3 piezoelectric composite transducers for swept-frequency calibration of hydrophones from 100 kHz to 2 MHz,” J. Acoust. Soc. Am, vol. 115, no. 6, pp. 2914–2918, 2004. [DOI] [PubMed] [Google Scholar]
  • [59].Preston RC, Bacon DR, and Smith RA, “Calibration of medical ultrasonic equipment—procedures and accuracy assessment,” IEEE Trans. Ultrason., Ferroelec., Freq. Contr, vol. 35, no. 2, pp. 110–121, March, 1988. [DOI] [PubMed] [Google Scholar]
  • [60].Lewin PA, Bautista R, and Devaraju V, “Voltage sensitivity response of ultrasonic hydrophones in the frequency range 0.25 – 2.5 MHz,” Ultrasound Med. & Biol, vol. 25, no. 7, pp. 1131–1137, 1999. [DOI] [PubMed] [Google Scholar]
  • [61].Devaraju V, Lewin PA, and Bleeker H, “Determination of sensitivity versus frequency characteristics of miniature ultrasonic hydrophones below 1 MHz using planar scanning technique,” J. Ultrasound Med, vol. 21, pp. 261–268, 2002. [DOI] [PubMed] [Google Scholar]
  • [62].Wang T, Xu Z, Hall TL, Fowlkes JB, and Cain CA, “An efficient treatment strategy for histotripsy by removing cavitation memory,” Ultrasound in Med. & Biol, vol. 38, no. 5, pp. 753–766, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [63].Eichstadt S and Wilkens V, “Evaluation of uncertainty for regularized deconvolution: A case study in hydrophone measurements,” J. Acoust. Soc. Am, vol. 141, no. 6, pp. 4155–4167, 2017. [DOI] [PubMed] [Google Scholar]
  • [64].Liu Y, Wear KA, and Harris GR, “Variation of high intensity therapeutic ultrasound (HITU) pressure field characterization: effects of hydrophone choice, nonlinearity, spatial averaging, and complex deconvolution,” Ultrasound in Med. & Biol, vol. 43, no. 10, pp. 2329–2342, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [65].Deng Y, Palmeri ML, Rouze NC, Rosenzweig SJ, Abdelmalek MF, and Nightingale KR, “Analyzing the impact of increasing mechanical index and energy deposition on shear wave speed reconstruction in human liver,” Ultrasound in Med. & Biol, vol. 41, no. 7, pp. 1948–1957, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [66].Deng Y, Palmeri ML, Rouze NC, Trahey GE, Haystead CM, and Nightingale KR, “Quantifying image quality improvement using elevated acoustic output in B-mode harmonic imaging,” Ultrasound in Med. & Biol, vol. 43, no. 10, pp. 2416–2425, 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [67].Harris GR, “Are current hydrophone low frequency response standards acceptable for measuring mechanical/cavitation indices?” Ultrasonics, vol. 34, pp. 649–654, 1996. [DOI] [PubMed] [Google Scholar]
  • [68].Harris GR, “Pressure pulse distortion by hydrophones due to diminished low frequency response,” IEEE Trans. Ultrason., Ferroelec., Freq. Contr, vol. 42, no. 6, pp. 989–992, Nov. 1995. [Google Scholar]
  • [69].AIUM/NEMA, “Acoustic Output Measurement Standard for Dianostic Ultrasound Equipment, Revision 3: NEMA Standards Publication UD 2–2004, American Institute of Ultrasound in Medicine, Laurel, MD,; National Electrical Manufacturers Association, Rosslyn, VA, 2004. [Google Scholar]
  • [70].International Electrotechnical Commission IEC 62127–1. Ultrasonics – Hydrophones – Part 1: Measurement and characterization of ultrasonic fields up to 40 MHz (IEC, Geneva, Switzerland: ), 2007 [Google Scholar]
  • [71].International Electrotechnical Commission IEC 62127–2. Ultrasonics – Hydrophones – Part 2: Calibration for ultrasonic fields up to 40 MHz (IEC, Geneva, Switzerland: ). Annex I 2007. [Google Scholar]
  • [72].Lewin PA, Schafer ME, and Chivers RC, “Integrated preamplifiers for ultrasound transducers,” Proc. 1985 IEEE Ultrasonics Symposium, 503–506, 0090–5607/85/0000–0503
  • [73].Lewin PA, Schafer ME, and Chivers RC, “Factors affecting the choice of preamplfication for ultrasonic hydrophone probes,” Ultrasound Med. & Biol, vol. 13, no. 3, pp. 141–148, 1987. [DOI] [PubMed] [Google Scholar]

RESOURCES