A Comprehensive Comparative Study on Inductive and Ultrasonic Wireless Power Transmission to Biomedical Implants

Ahmed Ibrahim 1, Miao Meng 2, Mehdi Kiani 3
PMCID: PMC6192045  NIHMSID: NIHMS959182  PMID: 30344453

Abstract

This paper presents a comprehensive comparison between inductive coupling and ultrasound for wireless power transmission (WPT) to biomedical implants. Several sets of inductive and ultrasonic links for different powering distances (d12) and receiver dimensions have been optimized, and their key parameters, including power transmission efficiency (PTE) and power delivered to the load (PDL) within safety constraints, have been compared to find out which method is optimal for any given condition. Two design procedures have been presented for maximizing the PTE of inductive and ultrasonic links by finding the optimal geometry for the transmitter (Tx) and receiver (Rx) coils and ultrasonic transducers as well as the optimal operation frequency (fp). Our simulation and measurement results showed that the ultrasonic link transcends the inductive link in PTE and somewhat in PDL for a small Rx of 1.1 mm3 (diameter of 1.2 mm), particularly when the Rx was deeply implanted inside the tissue (d12 ≥ 10 mm). However, for a larger 20 mm3 Rx (diameter of 5 mm), the inductive link achieved higher PTE and PDL, particularly at shorter distances (d12 < 30 mm). The optimal loading condition is shown to be quite different in inductive and ultrasonic links. Despite higher performance for small Rx and large d12, the ultrasonic link is more sensitive to Rx misalignments and orientations. This led us to propose a new design procedure based on the worst-case misalignment scenario. The simulation results have been validated by measurements. The inductive and ultrasonic links, operating at 30 MHz and 1.1 MHz, achieved measured PTEs of 0.05% and 0.65% for the 1.1 mm3 Rx located 30 mm inside tissue and oil environments with optimal load resistances of 295 Ω and 3.8 kΩ, respectively.

Index Terms: Wireless power transmission, inductive links, ultrasound, biomedical implants, efficiency, design methodology

I. Introduction

WIRELESS power transmission (WPT) is widely used in many applications with different power requirements to eliminate power cords [1]-[8]. WPT to biomedical implants can eliminate the need for bulky batteries with a limited lifetime, and reduce the implant size [9]. WPT is vital in today’s implantable medical devices (IMDs) that interface with the central and peripheral nervous systems (CNS and PNS) for recording and stimulating neuronal activity using different modalities such as electricity, light, and ultrasound [10]-[15].

IMDs can potentially be utilized in a wide range of applications such as cochlear, retinal, cortical, and PNS implants [16]-[19]. The required power range for IMDs varies significantly due to widely varying system requirements and constraints. For example, IMDs are used with various media—such as air, bone, and tissue—and with different implantation depths from several millimeters to centimeters, as well as different functions, such as monitoring and/or modulation of neural activity with different numbers of channels and modalities (e.g. electrical, optical, and ultrasound) [20]. Cochlear and retinal implants require tens of mW transferred across sub-cm to cm distances [16], [17]. The power consumption of cortical implants ranges from sub-mW to tens of mW, which highly depends on the number of recording/stimulation channels and their function (recording and/or stimulation) [18]. State-of-the-art recording front ends consume several μW of power, however, the data transmitter often dominates the total power consumption in IMDs, particularly if the number of channels is high [21]. PNS implants for restoring the dysfunctional motor and sensory functions in the limbs require transmitted power levels of tens to hundreds of mW across several millimeters due to the large stimulation current [22]. There are also cardiac and gastric slow-wave recording implants that only require hundreds of μW or less, but across several centimeters [15], [19]. Achieving robust, efficient, and safe WPT to small implants located deep inside the body is of great importance in all applications, regardless of the challenges posed by the wide range of system requirements.

Fig. 1 shows the generic block diagram for a wirelessly powered implant with emphasis on its main blocks for WPT. In the transmitter side (Tx), it includes an external energy source that supplies an efficient power amplifier (PA) to drive the primary element, which, in this paper, is either a series resonant LC-tank (L1C1) or an ultrasonic transducer (U1). The energy source drives the primary element at the power carrier frequency (fp). In the receiver side (Rx), it includes a secondary element (parallel resonant L2C2-tank or U2) that receives the AC power carrier, and a power management block to convert the incoming AC signal to a constant DC supply for the implant core. In this paper for simplicity, the power management and implant core are modeled as an AC load, RL in Fig. 1.

Fig. 1.

Fig. 1

Generic WPT link across the skin for powering biomedical implants using a pair of primary and secondary elements. In the inductive and ultrasonic links, these elements are LC-tanks (series L1C1 and parallel L2C2) and ultrasonic transducers (U1-U2), respectively. For simplicity, the power management and implant core have been modeled as an AC load, RL. Depending on the application, the propagation medium can include one or more layers of air, skin, bone, soft tissue, fat, muscle, etc. It should be noted that the ultrasonic link is not suitable for air and bone media.

Until recently, inductive coupling was the conventional method for WPT to biomedical implants, particularly for those with centimeter (cm) dimensions. This is due to its high power transmission efficiency (PTE) defined as the delivered power to RL divided by the Tx element’s (L1C1 or U1) input power as well its minimal dissipation of the magnetic field inside the tissue, and safety considerations [23]-[26]. Recently, several efforts have been made towards the miniaturization of implants to millimeter (mm) and even sub-mm dimensions in order to reduce the tissue damage and inflammation, and improve the implant’s longevity [27]-[33]. For improving PTE of mm-sized coils, a common theme has been the significant increase of fp to hundreds of MHz.

In [28]-[30], it has been suggested that the optimal fp to deliver power to an implant coil with an outer diameter (Do2) of 2 mm is within the GHz/sub-GHz range. An fp of 120 MHz has been proposed in [31] for powering an on-chip coil with Do2 of 1 mm. An optimal fp around 300 MHz has been used to deliver power to a single-loop square coil with Do2 of 1 mm and 5 mm in [21],[32]. The geometry and fp of the inductive link have been optimized in [33] for Do2 of 1 mm to maximize PTE for a fixed RL of 5 kΩ, resulting in the optimal fp = 200 MHz. We have also presented the design and optimization of inductive links for powering mm-sized implants by proposing a new figure-of-merit (FoM) to achieve not only high PTE but also maximum delivered power to the load (PDL) within safety constraints [34], [35]. At the powering distance (d12) of 12 mm with Do2 = 1 mm, we have demonstrated two inductive links at fps of 20 MHz and 100 MHz with the PTE and maximum PDL of 1.4% and 3.3%, and 2.2 mW and 0.72 mW, respectively.

Although these new designs and optimizations have partially improved the PTE and PDL of inductive/RF-based links, small PTE and stringent safety constraints have recently led to the emergence of ultrasound-based WPT links for mm-sized implants [36]-[41]. The ultrasonic link in [36] has achieved a PTE of 0.002% in delivering power to an ultrasonic Rx with a size of 0.127 mm3 at d12 = 3 cm using a commercial Tx transducer (U1) operating at fp = 5 MHz. An ultrasonically powered implant has been reported in [37], delivering a regulated power of 100 μW to an ultrasonic Rx with a diameter (Dou2) of 1 mm at d12 = 3 cm using a 1 MHz commercial U1. For optimal WPT to mm-sized implants, we have also presented the design and optimization of ultrasonic links, based on which an ultrasonic link at fp = 1.8 MHz has been demonstrated with the measured PTE of 0.66% at a d12 of 3 cm for Dou2 = 1.2 mm [39].

A comprehensive study of inductive and ultrasonic links for different conditions in terms of implant size, depth, misalignment, and orientation can help the designers of WPT for biomedical implants to choose the optimal modality. In [40], simulated PTEs of inductive and ultrasonic WPT links for Do2 of 2–10 mm have been compared for different d12’s, however, the conclusions have been made based on non-optimized inductive and ultrasonic links, which can be misleading. In [41], the measured and simulated PTEs of an ultrasonic link has been compared with the simulated PTEs of a comparably sized inductive link with identical Tx and Rx diameters of 4.4 mm. Again, both inductive and ultrasonic links have not been optimized in this work. In addition, the maximum PDLs of inductive and ultrasonic links under safety constraints have not been compared in the past.

Recently, we have also briefly compared the simulated PTEs of inductive and ultrasonic links for small and large implants [42]. However, optimal link parameters for different d12’s were not reported, and no comparisons were provided for different d12’s or for the rotation of both inductive and ultrasonic implants, and maximum PDL under safety constraints was not considered. In addition, no detailed optimization procedure, measurement results, or design guidelines for ultrasonic links based on their misalignment were provided in [42]. Therefore, the literature lacks a comprehensive, accurate, and validated (through measurements) comparison between key parameters (both PTE and PDL) of inductive and ultrasonic links for WPT to biomedical implants with different size, depth, alignment, orientation, and power requirements in order to clarify which WPT method is more efficient and robust for a given application.

In this paper, we present a performance comparison in terms of PTE and PDL between several sets of inductive and ultrasonic links. The links have optimal geometries and fp for two different Rx sizes of 1.1 mm3 (diameter = 1.2 mm) and 20 mm3 (diameter = 5 mm) at various d12’s of 5–50 mm. The PTE and PDL of these links vs. Rx misalignment and orientation will also be discussed. A design guideline for ultrasonic links will also be presented based on worst-case misalignment scenarios. This paper focuses on providing design guidelines and clarity for the designers of IMDs as to which modality is more suitable for any given application. For applications that involve bone (e.g. cortical implants) or air (e.g. retinal implants), ultrasound by itself is not suitable due to the large ultrasound loss and reflection. However, we have recently demonstrated a hybrid inductive-ultrasonic link, which can be used for such applications [43]. To demonstrate the proposed links in a real-world application, we will also provide a design example for a 1-channel gastric slow-wave recording implant with 10 μW of power consumption at an implantation depth of 5 cm [15].

Section II summarizes the design parameters and procedures for both inductive and ultrasonic links. The comparative simulation and measurement results will be discussed in Sections III and IV, respectively. Finally, discussions and concluding remarks will be presented in Sections V and VI, respectively.

II. Design and Optimization of Inductive and Ultrasonic WPT Links

A. Inductive WPT Link

The WPT link in Fig. 1 can be converted to a conventional 2-coil inductive link by replacing the primary and secondary elements with two resonant LC-tanks, i.e., series L1C1 and parallel L2C2, respectively. The PTE of the inductive link, as its model is shown in Fig. 2, can be found from,

η=k122Q1Q2L1+k122Q1Q2L·Q2LQL, (1)

where Q1 = ωL1/R1, Q2 = ωL2/R2, ω =2π fp = (L1C1)−0.5 = (L2C2)−0.5, Q2L = Q2QL/(Q2 + QL), and QL = RL/ωL2 is referred to as the load quality factor [44]. The term Q2L/QL represents the loss inside Rx and any RL mismatch. In order to find the optimal RL that maximizes PTE, i.e., RL,PTE, the optimal QL,PTE = RL,PTE/ωL2 can be found from

QL,PT E=Q21+k122Q1Q2, (2)

which can be further simplified to Q2 for loosely coupled coils with small k12, particularly in WPT to mm-sized implants [34]. In such small k12 and optimal-loading conditions, PTE in (1) can be simplified to k122Q1Q2/4 [34]. Therefore, for optimal load condition in (2), k12, Q1, and Q2 should all be maximized to optimize PTE. As we have discussed in detail in [34], any given RL smaller than the optimal RL,PTE can be transformed to RL,PTE by adding one off-chip capacitor, but matching any given RL larger than RL,PTE requires an inductor, which is not feasible in mm-sized implants.

Fig. 2.

Fig. 2

(a) The 2-coil inductive link model in HFSS to find the optimal geometries of L1 and L2, i.e., Do1,2, w1,2, n1,2, s1,2, as well as the optimal fp based on our design procedure in [34]. L2 is located inside muscle tissue. The total powering distance (d12) is d + ds.

The IEEE standard for maximum human exposure to RF electromagnetic fields states that the tissue specific absorption rate (SAR) should not exceed a certain value for safe operation. The peak SAR limit, averaged over 1 g of tissue, for RF exposure to the human head is 1.6 W/kg, which will be considered in the rest of this paper without the loss of generality [35]. The maximum allowable PDL under SAR constraints (P L) can be calculated by multiplying the power provided by the PA with the PTE from (1),

PL=0.5(R1+Rref)I1,SAR2×η=0.5R1(1+k122Q1Q2L)I1,SAR2×η (3)

where I1,SAR is the maximum Tx coil current under the SAR constraint and Rref is the reflected resistance to L1 [44]. For small k12, PL in (3) can be simplified to 0.5R1I1,SAR2×η [34].

In order to optimize the inductive links we proposed a detailed design procedure for WPT to mm-sized inductive links, made with wire-wound coils (WWCs), in [34]. But that design procedure optimizes a figure of merit for mm-sized implants as opposed to PTE. Fig. 3 shows our modified design procedure to maximize PTE with a matched load condition for cm- and mm-sized implants. We will provide a summary of this procedure in this paper, but we would like to refer the reader to [34] for a more detailed procedure as well as 3D surfaces of optimized parameters. We choose WWCs because they can achieve higher Q and, consequently, higher PTE [45]. The optimization should be done in a full-wave electromagnetic field simulator, such as HFSS (Ansoft, Pittsburgh, PA), as shown in Fig. 2. The design procedure is as follows. Step-1: design constraints are imposed by the application and fabrication technology, including the maximum values for the implant size, i.e., outer diameter (Do2) and height of L2, as well as L1 and L2 wire widths (w1, max, w2, max), the nominal value for the implant depth inside the tissue (d in Fig. 2), and finally the minimum value for the wire spacing between turns (s,min) as twice the thickness of the wire insulation. Step-2: the initial values for L1 and L2 geometries, such as w1, the number of turns (n1 and n2), w2, and s1, and L1 spacing from tissue (ds in Fig. 2) are chosen. Step-3: Q2 is maximized in an iterative process by sweeping n2 and w2 for different fps using s2,min. This step finds the optimal L2 geometry. Step-4: PTE in (1) will be maximized in an iterative process by sweeping n1, Do1, w1, and s1 for different fps. This step leads to optimal L1 and fp. Step-5: the optimal ds that increases Q1 and consequently maximizes PTE in (1) is found. This leads to a slight increase in the powering distance (d12) to d + ds. Finally, the design is further validated and fine-tuned through measurements in step 6 [34].

Fig. 3.

Fig. 3

Iterative optimization flowchart for designing efficient inductive WPT links. The optimal values for L1/L2 geometries, fp, and RL that maximize PTE can be found.

It should be noted that Q1 also depends on the medium surrounding L1 (particularly its conductance), which affects PTE and PDL in (1) and (3). As shown in Fig. 2, separation of L1 from the skin (ds) can help to improve Q1 at the cost of lowering k12. Therefore, to strike a balance between Q1 and k12 in (1), optimal ds will be in the range of several millimeters, suggesting that L1 should always be carried by the subject.

B. Ultrasonic WPT Link

An ultrasonic WPT link can be realized in Fig. 1 by replacing the primary and secondary elements with two disc-shaped ultrasonic transducers (U1 and U2); its model is shown in Fig. 4. In this paper, symmetric disk-shaped piezoelectric transducers, made with lead zirconate titanate (PZT), will be used to achieve high electromechanical coupling and less complexity in 3D modeling and computation in finite-element method (FEM) simulation tools [39]. It should be noted that similar design methodologies can be generalized to other types of ultrasonic transducers. As shown in Fig. 4, the ultrasonic transmitter (U1) should be in perfect contact with the skin to eliminate any acoustic reflections at this interface. In practice, some ultrasonic gels can also be applied at the interface. We have already discussed in detail the theory behind ultrasonic WPT using disk-shaped piezoelectric transducers in [39]. The PTE of such links can be found from,

PT E=Io×A2×η2Pin, (4)

where η2 is the mechanical-electrical power conversion efficiency of U2 including any RL mismatch within the Rx, Io is the acoustic intensity of the U1 beam, A2 is the effective cross-section area of U2, and Pin is the delivered power to U1 from the energy source. Any impedance mismatch at the Tx-transducer/skin interface can significantly reduce the PTE due to acoustic reflections that drastically reduce Io in (4). As discussed in [39], disk-shaped unfocused transducers have a natural focal point where the beam is the narrowest and Io is maximum. The focal point distance of U1 depends on the outer diameter of U1 (Dou1) as well as fp. The thickness-mode resonance frequencies of U1 and U2 should be matched to fp to increase their mechanical-electrical power conversion efficiencies. In addition, A2 in (4) is a function of Dou2. Finally, for maximum power transfer in the Rx, any given RL should be matched to the U2 impedance at resonance, depending on Dou2 and t2. Therefore, U1 and U2 geometries, including Dou1,2 and t1,2 in Fig. 4 as well as fp, need to be co-optimized in a matched-load condition to achieve the highest PTE in (4).

Fig. 4.

Fig. 4

The ultrasonic WPT link model in COMSOL to find the optimal geometries for U1 and U2, i.e., Dou1,2, t1,2, as well as the optimal fp. Matching layers are added on both U1 and U2. The soft tissue is mimicked by the castor oil with similar acoustic attenuation of 0.8 dB/cm/MHz. A perfect matching layer (PML) has been considered at the boundaries of the medium to avoid acoustic reflections.

It should be noted that the optimal geometries of the ultrasonic transducers can be accurately predicted if the aspect ratio (Dou/t) is large [39]. However, for most applications involving mm-sized implants, the aspect ratio is restricted due to the small Rx dimension requirement. Therefore, FEM simulation tools such as COMSOL Multiphysics (COMSOL, Burlington, MA), need to be used in simulating and optimizing ultrasonic links for WPT to mm-sized implants. Fig. 4 shows the ultrasonic WPT link model in COMSOL, in which U1 is air-backed to improve its conversion efficiency and U2 is mounted on a silicon substrate, integrating the power management and implant-core circuitry, and is located inside castor oil with an attenuation coefficient of 0.8 dB/cm/MHz to mimic the soft tissue. In ultrasonic WPT, since U1 needs to be in contact with the skin, d12 and d are the same (ds = 0). In order to minimize reflections at U1/U2 interfaces due to a mismatch in acoustic impedance of two media, matching layers with two materials are used [39]. A perfect matching layer (PML) has also been considered at the boundaries of the medium to avoid acoustic reflections.

For ultrasonic WPT links, the maximum achievable PDL under FDA safety constraints (PL) is limited by the maximum spatial-peak temporal-averaged intensity (ISPTA) of 720 mW/cm2 [46]. In order to find PL in COMSOL, the link is simulated in 2D Axial Symmetric configuration, and the intensity magnitude, which is equivalent to ISPTA, is plotted. Then the input power is increased by increasing the AC input voltage until ISPTA of 720 mW/cm2 is observed, at which PL can be calculated from the AC voltage across RL.

Fig. 5 shows the simulated sound-wave intensity of the optimized 1.1 mm3 ultrasonic link (shown later in Table III) at d = 10 mm, when the maximum ISPTA of 720 mW/cm2 was achieved at the vicinity of U1, i.e., in the near field region of ~1 mm. Although matching layers were added to Rx, there are still some reflecting waves from the Rx, which can partially either reinforce or cancel out the transmitted waves depending on the Rx location (d) and fp. As shown in Fig. 5, the ultrasound intensity within the whole medium (not necessarily at the implant location) needs to be found for safety constraints.

Fig. 5.

Fig. 5

Simulated sound-wave intensity of the 1.1 mm3 ultrasonic link optimized at d = 10 mm, where the maximum spatial-peak temporal time-averaged intensity (ISPTA) of 720 mW/cm2 was achieved at d = 1 mm.

TABLE III.

Key electrical and Acoustic Parameters of the Human Abdominal Layers

Layer Sound Speed (m/s) Density (kg/m3) Attenuation (dB/cm/MHz) Acoustic Impedance (MRayls) Conductivity [S/m] Relative Permittivity
Skin 1730 1150 1.8 1.99 0.0004 4
Muscle 1575 1065 1 1.68 0.2 4
Fat 1450 950 0.6 1.38 0.035 2.5
Soft Tissue 1474 956 0.8 1.41 0.02 4

Our design procedure in [39] can only maximize ultrasonic link PTE for a fixed RL. A revised design procedure is presented in Fig. 6 that maximizes ultrasonic link PTE for matched-load condition by optimizing the U1 and U2 geometries, fp, and RL. While a summary of the design procedure is presented here, we would like to refer the reader to our previous work in [39] for more details as well as 3D surfaces of optimized parameters. The optimization flowchart starts with the design constraints imposed by the application, including the maximum implant size, i.e., outer diameter (Dou2) and overall thickness of U2 (t2ov = t2 + tm2, where tm2 is the matching-layer thickness), as well as the nominal value for d12 in step 1.

Fig. 6.

Fig. 6

Iterative optimization flowchart for designing efficient ultrasonic WPT links in COMSOL. The optimal values for U1/U2 geometries, fp, and RL that maximize PTE can be found.

In step 2, initial values for fp, Dou1, the thicknesses of U1 piezoelectric transducer and its matching layer (t1 and tm1), tm2, and RL (several kΩ) are chosen. The detailed guidelines for selecting initial values of fp, Dou1, t1, tm1, t2, and tm2 can be found in our previous work [39]. Rx geometry is optimized in step 3 to maximize PTE by sweeping both t2 and Dou2, resulting in a 3D surface as shown in [39], followed by sweeping RL in an iterative process until RL changes less than 1%. This leads to the optimal U2 geometry with optimal RL (RL,PTE) for the given U1 geometry.

In step 4, U1 geometry is optimized by sweeping both t1 and Dou1 to maximize PTE, using the values for t2, Dou2, and RL from step 3. Steps 3 and 4 are repeated iteratively until t1 and Dou1 change less than 1%. This leads to optimal U1/U2 geometries and RL,PTE for the selected fp. In step 5, fp is changed slightly and steps 1-4 are repeated for the new fp in an iterative process to find the optimal fp that leads to the maximum PTE. The procedure in Fig. 6 should be repeated for several fps until the new PTE values for higher fps are at least two times smaller than the peak PTE. Step 5.1 determines the optimal U1/U2 geometries, RL, and fp that achieve the highest PTE which are further validated and fine-tuned through measurements.

C. Design Examples

In order to compare the performance of inductive and ultrasonic WPT links for different conditions including implant size and depth, different sets of inductive and ultrasonic links were optimized for PTE considering the same design constraints: 1) implant size was limited to 1.1 mm3 and 20 mm3 by limiting its thickness to 1 mm and outer diameter to 1.2 mm and 5 mm, respectively. This covers small and relatively large implant sizes. 2) In order to compare these links at different implant depth, d was set to 5 mm, 10 mm, 30 mm, and 50 mm. 3) In optimizations, optimal an load condition was considered for all links. In our optimization setups as shown in Fig. 4, U2 was mounted on and L2 was wound around a silicon substrate with 0.3 mm thickness, mimicking the implant core circuitry. Finally, coils and ultrasonic transducers were made of single conductor copper wire and lead zirconate titanate (PZT), respectively.

III. Inductive and Ultrasonic WPT Links Performance Comparison

Tables I and II summarize the optimal coils and transducers geometries and fp for 16 sets of inductive and ultrasonic links for the Rx size of 1.1 mm3 (Do2 = Dou2 = 1.2 mm) and 20 mm3 (Do2 = Dou2 = 5 mm) at four different d within 5-50 mm range, based on our design procedures in Sections II.A and II.B, respectively. For inductive links in Table I, the optimal fp of the small 1.1 mm3 Rx was increased to 100 MHz at d of 5 mm and 10 mm to increase Q2 and consequently PTE. At d ≥ 30 mm, fp was reduced to 30 MHz and 20 MHz, because Do1 was increased by ~2 times to improve k12 at large distances, and larger Do1 reduced the self-resonance frequency of L1. Therefore, fp was reduced to increase Q1 and PTE. For the larger 20 mm3 Rx, optimal fp of 20 MHz was achieved for all d, because a large Q2 could be achieved at lower fp, considering large Do2 of 5 mm. In all inductive links, an optimal ds of 2 mm further improved Q1 and PTE.

TABLE I.

Optimized Coil Geometries and Operation Frequencies for 8 Design Examples

Parameters Symbols 1.1-mm3 Rx 20-mm3 Rx
L1 Inductance (nH) L1 359 409 3816 4426 912 2396 3152 6285
Outer Diameter (mm) Do1 18 20 44 43 32 38 46 50
Number of Turns n1 5 5 13 12 7 9 8 12
Wire Width (μm) w1 644
Wire Spacing (μm) s1 644
Quality Factor Q1 236 136 214 282 248 186 147
L2 Inductance (nH) L2 42 42 147 210 298
Outer Diameter (mm) Do2 1.2 5
Number of Turns n2 7 6
Wire Width (μm) w2 100 100
Wire Spacing (μm) s2 30 30
Quality Factor Q2 55 36 30.7 59.3
Optimal Operation Frequency (MHz) fp 100 100 30 20 20
Implant Depth Inside Tissue (mm) d 5 10 30 50 5 10 30 50
Optimal L1 and Tissue Spacing (mm) ds 2 2 2 2 2
Optimal Load (Ω) RL,PTE 1550 1550 295 167 2168
PTE (%) η 13.5 3.65 0.078 0.0064 67.4 47.8 4.71 0.45
Max PDL Within Safety Limits (mW) PL 4.4 0.88 0.068 0.0056 346.4 131.8 5 0.55

Grayed cells indicate the design constraints.

TABLE II.

Optimized Ultrasonic Transducer Geometries and Operation Frequencies for 8 Design Examples

Parameters Symbols 1.1-mm3 Rx 20-mm3 Rx
U1 Diameter (mm) Dou1 5.8 6.4 10.8 13.3 5.7 5.2 7 8.9
Thickness (mm) t1 1.05 1.15 1.05 1.13 0.6 0.64 0.63 0.65
Backing Air
lst + 2nd Matching Layers Thickness (mm) tm1 0.25 + 0.08 0.26 + 0.08 0.30 + 0.09 0.16 + 0.05 0.16 + 0.05 0.17 + 0.05
U2 Diameter (mm) Dou2 1.1 1.2 1.2 1.2 4.7 4.9 4.8 5
Thickness (mm) t2 0.32 0.31 0.3 0.3 0.45 0.47 0.48 0.47
Backing Silicon
lst + 2nd Matching Layers Thickness (mm) tm2 0.25 + 0.08 0.26 + 0.08 0.30 + 0.09 0.16 + 0.05 0.16 + 0.05 0.17 + 0.05
Optimal Operation Frequency (MHz) fp 1.9 1.8 1.8 1.6 3 2.9 2.9 2.8
Implant Depth Inside Tissue (mm) d 5 10 30 50 5 10 30 50
Optimal Load (Ω) RL,PTE 4700 4500 4800 2000 140 130 140 100
PTE (%) η 19.4 10.6 2.3 0.7 40.4 28 8.5 2.5
Max PDL Within Safety Limits (mW) PL 1.5 2 2.1 1 8.1 7.4 2.2 2

Grayed cells indicate the design constraints.

All ultrasonic links in Table II were optimized in castor oil medium with the mass density of 956 kg/m3 and sound velocity of 1474 m/s [47]. Schott glass and araldite glue were used as the first and second layers for acoustic matching in both U1 and U2 [48]. The optimal fp for the ultrasonic links were close to each other, within 1.6-3 MHz range for both small and large Rx’s. As d was increased to 50 mm, Dou1 was also increased to extend the focal zone and improve PTE, particularly for the 1.1 mm3 Rx, in which achieving a narrow focal zone was key [39]. Table II also lists the thicknesses of each two-material matching layer (tm1 and tm2), which are added to t1 and t2. Since U2 overall thickness should be limited to 1 mm and U2 is mounted on a silicon die with a thickness of 0.3 mm, t2 + tm2 was limited to 0.7 mm in our design procedure.

Figs. 7a and 7b compare the simulated PTEs and maximum PDLs within safety limits (PL) vs. Rx implant depth (d) for optimal inductive and ultrasonic links in Tables I and II for the Rx sizes of 20 mm3 and 1.1 mm3, respectively. For the 20 mm3 Rx, as shown in Fig. 7a the inductive link PTE and PL were significantly higher at shorter d of 5 mm and 10 mm. At d = 10 mm, the inductive link achieved 1.7 times more PTE (47.8% vs. 28%) and 17.8 times more PL (131.8 mW vs. 7.4 mW) compared to the ultrasonic link. At d = 30 mm, although the ultrasonic link achieved higher PTE (8.5% vs. 4.7%), the inductive link PL was still significantly higher (5 mW vs. 2.2 mW). At larger d = 50 mm, the ultrasonic link achieved higher PTE and PL of 2.5% and 2 mW compared to the inductive link PTE and PL of 0.45% and 0.55 mW, respectively.

Fig. 7.

Fig. 7

Comparison between simulated values for PTE and maximum PDL within safety limits (PL) vs. d for the optimal inductive and ultrasonic links with the Rx sizes of (a) 20 mm3 and (b) 1.1 mm3. The links specifications have been listed in Tables I and II.

For the smaller 1.1 mm3 Rx, as shown in Fig. 7b the ultrasonic link outperformed the inductive link with significantly higher PTEs at all distances. The PTE for the ultrasonic link was 1.4 times higher than the inductive link (19.4% vs. 13.5%) at small d = 5 mm, and 109 times higher (0.7% vs. 6.4×10−3%) at larger d of 50 mm. Although the inductive link PL at short d = 5 mm was higher (4.4 mW vs. 1.5 mW), the ultrasonic link achieved much higher PL at d ≥ 10 mm, particularly at large d of 50 mm, in which the PL for the ultrasonic link was 178.6 times higher than the inductive link (1 mW vs. 5.6×10−3 mW). Several lessons can be learned from Fig. 7. 1) For large Rx sizes, the inductive link PL is much larger than that of the ultrasonic link because magnetic fields at MHz frequencies are quite safe [35]. 2) At large d, particularly when d is significantly larger than Do2 or Dou2, the ultrasonic link achieves higher PTE and PL. 3) For small mm-sized Rx’s, the ultrasonic link can achieve much higher PTE and PL, particularly at larger distances. 4) The ultrasonic link PL within the safety constraint is not significantly higher than that of the inductive link. As shown in Fig. 7, there are conditions for which the ultrasonic link achieves higher or comparable PTE but less PL. Indeed, the main reason for 178.6 times higher PL in the ultrasonic link for 1.1 mm3 Rx at d = 50 mm is the 109 times improvement in PTE.

It should be noted that the ultrasonic link PL for 1.1 mm3 Rx in Fig. 7b peaked at d of 30 mm because at larger distances Dou1 was increased by the optimization procedure to provide a more distant focal zone, resulting in less acoustic intensity for the same input power. As shown in [39], Dou1 mainly determines the beam diameter. In addition, the overlapping sound waves generated by the reflections from Rx are more dominant at shorter distances. The ultrasonic link PL decreased at d > 30 mm due to a PTE drop at larger distances.

Fig. 8 shows the optimal loadings (RL,PTE) that were found in the optimization of inductive and ultrasonic links for the 1.1 mm3 and 20 mm3 Rx’s at different d. For the large 20 mm3 Rx, the inductive link RL,PTE is within kΩ-range (PDL of several mW), while the ultrasonic links require much smaller RL,PTE of hundreds of ohms (PDL of tens of mW). Therefore, for small mW-range, the PDL ultrasonic links suffer from mismatch to the Rx, further decreasing their PTE. If more PDL is needed, large RL,PTE in inductive links can easily be realized by adding one capacitor or employing multi-coil links, since the Rx size is quite large [34], [43]. Therefore, inductive links are superior for large Rx’s. For a small 1.1 mm3 Rx, the inductive link RL,PTE is smaller, particularly at large distances. Since less power (mW-range and below) can practically be delivered to mm-sized implants due to safety issues, ultrasonic links are superior, particularly for deeply-implanted mm-sized devices.

Fig. 8.

Fig. 8

The optimal loading (RL,PTE) vs. d for maximizing PTE in optimized inductive and ultrasonic links listed in Tables I and II.

Figs. 9a and 9b show the simulated values of PTE and PL of optimized inductive and ultrasonic links for a 1.1 mm3 Rx at d = 10 mm vs. Rx lateral misalignment in the Y direction, as shown in Fig. 2, and angular rotation (θ), respectively. Although the ultrasonic link achieved higher PTE (10.6% vs. 3.65%) when Tx and Rx were perfectly aligned, it is drastically prone to Rx misalignment and rotation due to its focused nature. As shown in Fig. 9a, the ultrasonic link PTE and PL reduced by 88 and 95 times from 10.6% to 0.12% and 2 mW to 210 μW for only 3 mm of misalignment, respectively. For the same condition, the inductive link PTE and PL slightly reduced by 1.3 times from 3.65% to 2.9% and 0.88 mW to 0.6 mW, respectively.

Fig. 9.

Fig. 9

Simulated PTE and PL of optimized inductive and ultrasonic links for 1.1 mm3 Rx at d = 10 mm vs. (a) Rx lateral misalignment in Y direction as shown in Figs. 2 and 4, and (b) Rx angular rotation (θ).

A similar trend can be seen in Fig. 9b for Rx rotation around the Y-axis. For a small θ of 30°, the ultrasonic link PTE and PL reduced by 29 and 32 times to 0.35% and 62 μW, respectively. However, the inductive link PTE and PL only dropped by ~1.3 times to 2.8% and 0.6 mW for θ = 30°, respectively. Fig. 9 clearly shows that inductive coupling is a more robust technique for WPT to implants that involve misalignment and rotations compared to ultrasound because both PTE and PL of the inductive link are significantly higher for > 1 mm of misalignment and > 20° of rotation. Therefore, ultrasonic links need to be designed such that they become less prone to misalignment and rotation.

IV. Measurement Results

The accuracy of our HFSS and COMSOL simulations for inductive and ultrasonic links, respectively, were validated through measurements particularly at large d and small Do2/Dou2. To reduce the fabrication complexity, we optimized an ultrasonic link without matching layers at d = 30 mm and Dou2 = 1.2 mm and used it in measurements. Our optimization resulted in optimal Dou1 = 15.9 mm, t1 = 1.95 mm, t2 = 0.95 mm, fp = 1.1 MHz, and RL,PTE = 3.8 kΩ. Figs. 10a and 10b show the experimental setups for measuring the PTE for the optimized inductive and ultrasonic links, respectively, for a 1.1 mm3 Rx at d = 30 mm. We first measured S-parameters using a network analyzer and then calculated PTEs using the equations in our previous works [34], [39].

Fig. 10.

Fig. 10

Measurement setup for measuring S-parameters of (a) inductive link and (b) ultrasonic link using a network analyzer to calculate PTE for 1.1 mm3 Rx at d = 30 mm.

Inside the inductive link measurement setup in Fig. 10a, the L1 and L2 coils with the specifications in Table I (Do2 = 1.2 mm, d = 30 mm) were connected to a pair of SMA connectors, and held in parallel and perfectly aligned using a Plexiglas/plastic frame. In order to mimic the lossy tissue environment, L2 was wrapped in a 3-cm thick layer of chicken breast. Similar to Fig. 2, L1 was spaced 2 mm from the tissue. Our proposed de-embedding method in [34] was used to improve the accuracy of measured Q1, Q2, and k12 from the S-parameters.

Inside the ultrasonic link measurement setup in Fig. 10b, U1 and U2 (without matching layers) were connected to a pair of SMA connectors and held perfectly aligned inside an oil tank, mimicking soft tissue. U2 was mounted on a small printed circuit board (PCB) and held above the bottom of the tank using a 3D-printed stand which was fixed at the center of the oil tank. U1 was glued on a 3D-printed holder which could move in the Y and Z directions as shown in Fig. 10b.

Fig. 11a shows the simulated and measured PTEs of the optimized inductive links vs. fp for the Rx’s size of 1.1 mm3 (Do2 = 1.2 mm) at d = 30 mm and ds = 2 mm, i.e., d12 = 32 mm. The measured PTE at the optimal fp of 30 MHz was 0.053%, which was slightly smaller than the simulated PTE of 0.078%. This discrepancy could be due to the geometry variations of the hand-fabricated L2, and any possible L2 misalignments in measurements. Nonetheless, the measured PTEs were close to the simulated ones, particularly at smaller fp < 20 MHz.

Fig. 11.

Fig. 11

Comparison between simulated and measured PTEs of optimized inductive link for 1.1 mm3 Rx and d = 30 mm vs. (a) fp and (b) lateral misalignment in the Y direction.

Fig. 11b shows the simulated and measured PTEs of the same inductive link vs. Rx lateral misalignment in the Y direction (Fig. 2). The simulated and measured PTEs at the Rx lateral misalignment of 10 mm were 0.047% and 0.024%, respectively. The measured results in Fig. 11b further confirms that the inductive link is quite robust to misalignment.

Figs. 12a and 12b show the simulated and measured PTEs for the optimized ultrasonic link for a 1.1 mm3 Rx with a matched load of 3.8 kΩ vs. (a) fp at d = 30 mm, and (b) dat the optimal fp of 1.1 MHz. Close simulated and measured PTEs of 0.69% and 0.62% were achieved at d = 30 mm and fp = 1.1 MHz, respectively, validating the accuracy of our simulations.

Fig. 12.

Fig. 12

Comparison between simulated and measured PTEs of optimized ultrasonic link (without matching layers) for 1.1 mm3 Rx vs. (a) fp at d = 30 mm, and (b) d at the optimal fp of 1.1 MHz.

V. Discussion

Our measurement setups in Fig. 10 use a homogenous medium for WPT. To verify the effects of the multilayer propagation medium, which is more practical, on the PTE of the inductive and ultrasonic WPT, we provide a real-word example of PNS interfacing in the following. Our current active research is on recording gastric slow-wave signals by implanting an array of mm-sized implants inside the sub-mucosa space of the stomach to study the electrophysiology behind stomach dismotility [15]. Fig. 13 shows the abdominal multilayer structure of a human for such an application, in which a 1.1 mm3 inductive/ultrasonic Rx, optimized for d = 50 mm as listed in Tables I and II, is located at d = 50 mm. Table III shows the key electrical and acoustic parameters of these layers [49]-[53].

Fig. 13.

Fig. 13

Human abdominal multilayer structure for a mm-sized inductive/ultrasonic gastric implant, located at d = 50 mm. The key parameters of these layers for simulations have been listed in Table III.

The inductive and ultrasonic links were simulated in both a homogeneous medium (muscle for inductive link and soft tissue for ultrasonic link) as well as the multilayer structure in Fig. 13. Figs. 14a and 14b compare the simulated PTEs of inductive and ultrasonic links, respectively, in single-layer and multilayer media. As shown in Fig. 14a, the inductive link PTE was slightly increased in the multilayer medium because the conductivity of the skin layer is less than that of muscle (as shown in Table III) and, therefore, the quality factor of the Tx coil, which faces the skin and muscle in multilayer and single-layer media, respectively, was increased in the multilayer medium. Fig. 14b also shows that the ultrasonic PTE changes were negligible, because the calculated reflections at the interface of layers in the multilayer media were as small as 3.28% at the skin-fat interface, 0.96% at the fat-muscle interface, and 0.76% at the muscle-soft tissue interface.

Fig. 14.

Fig. 14

Comparison between simulated PTEs of optimized WPT links for 1.1 mm3 Rx and d = 50 mm in single-layer medium and multilayer medium in Fig. 12 for (a) inductive link and (b) ultrasonic link, showing close PTE results.

Although ultrasonic links achieve high PTE and PL, particularly for WPT to deeply-implanted mm-sized devices, they are drastically prone to any Rx misalignment due to their focused nature, as shown in Fig. 9a. This may cause ultra-low PTE and even malfunction in powering implants that involve dynamic movements, notably PNS interfaces for recording from or stimulating actively moving organs, such as the stomach and heart [15], [54]. Using ultrasonic beamforming to steer the ultrasound beam towards the implant can partially mitigate the misalignment issue [55]. However, this requires knowledge about the implant’s location at any moment as well as complex and power-consuming external hardware. In addition, beam-forming can be a suitable choice when a network of implants is wirelessly powered. Therefore, we believe that the ultrasonic WPT link for a single implant needs to be optimized for the worst-case scenario of the misalignment.

As shown in [39], the ultrasound beam is narrowest in the focal zone, which is an optimal location for the Rx when it is perfectly aligned. However, by placing the Rx inside the deep far-field region, PTE can be improved for a wider range of misalignments at the cost of PTE reduction in perfectly aligned conditions. This can be done by reducing Dou1, which decreases the focal length [39].

Due to the limited aspect ratio of the Tx transducer, t1 should also be optimized along with Dou1 to maintain U1 resonance at fp. It should be noted that due to the side beams, increasing Dou1 might also improve PTE for a certain misalignment condition, as shown in Fig. 5. However, this approach places Rx in the near-field region with several maxima and minima in the sound intensity, which can lead to drastic PTE reduction if the Rx moves slightly [39].

Our proposed optimization of ultrasonic links for worst-case misalignment scenarios is as follows. Step 1: the design procedure in Fig. 6 with fully-aligned Rx will be followed first to find optimal U1 and U2 geometries as well as fp. Step 2: the optimal Dou1 in step 1 is reduced by 0.1 mm and the corresponding t1 for resonating U1 at fp is found by sweeping t1 around its optimal value from step 1. Step 3: using the new Dou1 and t1 values, PTE will be found for all practical misalignments. It is clear that for the new Dou1 and t1, PTE for the fully-aligned condition will be reduced. However, steps 2 and 3 will be repeated in a recursive process until the PTE at the worst-case misalignment keeps decreasing. Finally, among the U1 geometries that lead to the maximum or close-to-maximum PTEs at the worst-case misalignment, the one with relatively higher PTEs for lower misalignments, including the aligned condition, will be chosen.

For a worst-case misalignment of 3 mm, we re-optimized the ultrasonic link with a 1.1 mm3 Rx at d = 10 mm by changing Dou1 and t1. The specifications of the ultrasonic link, including U1 and U2 geometries, for the perfectly-aligned condition, are listed in Table II. Fig. 15a shows the simulated 3-D surface of PTE vs. t1 and Dou1 for the perfectly aligned condition. It can be seen that the maximum PTE of 10.6% was achieved at Dou1 = 6.4 mm and t1 = 1.15 mm (similar to the results in Table II), which validates our design procedure in Fig. 6 for fully aligned implants. In Fig. 15a, for each Dou1 there is an optimal t1 that corresponds to U1 resonance at an optimal fp of 1.8 MHz. As Dou1 was decreased from 6.4 mm to 2 mm, the peak PTE also decreased from 10.6% to 1.45% due to widening the ultrasound beam.

Fig. 15.

Fig. 15

Simulated 3D PTE surfaces of the ultrasonic link for 1.1 mm3 Rx at d = 10 mm vs. (a) different t1and Dou1 values in perfectly aligned condition, and (b) Rx lateral misalignment in Y direction (as shown in Fig. 2) and Dou1, demonstrating that reducing Dou1 helps to improve PTE for misaligned links.

Fig. 15b shows the simulated 3-D surface of PTE vs. Dou1 and Rx misalignment in the Y direction, as shown in Fig. 4, using the optimal t1 in Fig. 15a for each Dou1. It can be seen that at 3 mm of misalignment, the highest PTE of 0.36% was achieved with a relatively small Dou1 of 2.4 mm. At similar misalignment of 3 mm, the links with Dou1 of 3 mm and 2.2 mm achieved slightly lower PTEs of 0.32% and 0.34%, respectively. We prefer Dou1 of 3 mm, because 1) at 3 mm of misalignment, its PTE is very close to the maximum achievable PTE, 2) in fully-aligned condition, it achieves a higher PTE of 4% compared with the PTEs of 2.85% and 2.58% for Dou1 of 2.4 mm and 2.2 mm, respectively, and 3) as shown in Fig. 15b it achieves relatively higher PTE for a wide range of misalignments. As shown in Figs. 15a and 15b, for the misalignment of 0 to 3 mm, the link with the relatively small Dou1 of 3 mm shows only 12.5 times reduction in PTE from 4% to 0.32%, as opposed to 88 times decrease in PTE for the larger Dou1 of 6.4 mm from 10.6% to 0.12%. In other words, by reducing Dou1 from 6.4 mm to 3 mm, PTE was increased by 2.7 times (0.32% vs. 0.12%) for 3 mm of misalignment at the cost of 2.6 times less PTE without misalignment (4% vs. 10.6%). Such smaller variation in PTE, i.e., flatter PTE surface, for a smaller Dou1 as shown in Fig. 15b can also highly benefit the external power driver and battery by reducing their output power range [56].

Fig. 16 compares the measured PTEs of two ultrasonic links with different U1 geometries and identical 1 mm3 Rx’s for d = 30 mm and Rx misalignment of 0–5 mm in the Y direction as shown in Fig. 10b. The “large U1” in Fig. 16 refers to the optimal U1 geometry without a matching layer (Dou1 = 15.9 mm, t1 = 1.95 mm) for the aligned condition, while the “small U1” refers to a transducer with a reduced Dou1 of 9.5 mm (t1 = 1.95 mm) to improve robustness against misalignment. It can be seen in Fig. 16 that the PTE for the large U1 is 3.1 times higher than the PTE for the small U1 (0.62% vs. 0.2%) when U2 is perfectly aligned. However, for 5 mm misalignment in the Y direction, the link with the small U1 outperformed the link with the large U1 by 2.6 times (0.12% vs. 0.047%) due to the wider ultrasonic beam for the small U1 at d = 30 mm. When U2 was misaligned by 5 mm, the PTE for the link with large U1 was significantly reduced by 13 times, while the PTE for the link with small U1 only reduced by 1.7 times,

Fig. 16.

Fig. 16

Measured values of PTE at d = 30 mm vs. lateral misalignment in Y direction for two different U1 transducers with Dou1 of 15.9 mm and 9.5 mm.

This paper mainly focused on the performance (PTE and PDL) comparison between inductive and ultrasonic links. There have also been works that have used GHz/sub-GHz-range RF links for WPT to mm-sized implants [28]-[33]. Table III compares the specifications and performance of our inductive and ultrasonic links with the state-of-the-art ultrasonic and RF links for WPT to deeply implanted mm-sized devices. Since the helix Rx coil in [30] has a Do2 of 2 mm and height of 3 mm, i.e., a total volume of 9.4 mm3, for a fair comparison we have also optimized an inductive link for similar Rx volume at d = 50 mm. The specifications of this inductive link are as follows: Do1 = 60 mm, w1 = 700 μm, n1 = 12, s1 = 350 μm, Do2 = 2 mm, w2 = 100 μm, n2 = 23, s2 = 30 μm, ds = 2 mm, optimal fp = 11 MHz, and RL,PTE = 3.6 kΩ.

As shown in Table IV, the ultrasonic link with the small Rx size of 1 mm3 has still achieved higher PTE of 0.7% at d12 = 50 mm compared with its inductive and RF counterparts, which have an even larger Rx size of 9.4 mm3, with the PTEs of 0.13% and 0.04%, respectively. Despite smaller Rx size, the ultrasonic link PL is ~1 mW, which is also higher than the RF link in [30] with PL = 0.2 mW under the SAR limit of 2.5 W/kg averaged over 10 g of tissue. This clearly shows that ultrasonic links are superior for powering deeply implanted mm-sized devices if they are properly optimized for worst-case misalignment scenarios.

TABLE IV.

Benchmarking of Recent WPT Links for Deeply-Implanted mm-Sized Devices with Different Modalities

Parameters This work 2014, [30] RF 2015, [36] Ultrasound 2015, [38] Ultrasound
Inductive *Inductive Ultrasound
Tx Outer Diameter (mm) 43 60 13.3 60 6.35 20 30
Rx Outer Diameter/Cross-Section Area (mm/mm2) 1.2/1.1 2/3.14 1.2/1.1 2/3.14 0.127×0.127 1×5 2×2 2×4
Volume (mm3) 1.1 9.4 1.1 9.4 0.002 5 8 16
Powering Distance (mm) 52 52 50 50 30 200
Optimal Operation Frequency (MHz) 20 11 1.6 1600 5 2.3 1.15
PTE (%) 0.0064 0.13 0.7 0.04 0.002 0.4 1.7 2.7
Max PDL Within Safety Limit (mW) 0.0056 1.4 1 +0.2
*

Link specs: w1= 700 μm, n1 = 12, s1 = 350 μm, w2 = 100 μm, n2 = 23, s2 = 30 μm, RL,PTE = 3.6 kΩ.

+

For peak SAR limit of 2.5 W/kg averaged over 10 g of tissue.

Since the PTE and PL of inductive and ultrasonic links vary significantly with d12 and implant size (as shown in Fig. 7), a designer can utilize the FoM formulas, which we have proposed in [34] and [57], to strike a balance between these two key parameters. Our design procedures in Section II can easily be extended to optimization of FoM by replacing the PTE equation with the FoM equation in each step.

The changes in the implant’s position over time highly depend on its size and application. For instance, cochlear implants utilize a pair of magnets to fully align the Tx and Rx coils and fix the implant’s position. For cm-sized implants, this can be a practical solution. For mm-sized implants, particularly in the PNS with active movements, the implant is supposed to be light enough to freely float inside the tissue, reducing tissue damage [27]. Therefore, the implant will move, which is the main reason we proposed a revised design procedure in this section that compensates for the implant’s movements. For our design example of a gastric implant similar to [58], the implant will be endoscopically placed into the submucosa space of the stomach by making small submucosal pockets.

The above-mentioned IMD applications require the delivery of tens of mW of power or below, which requires only several Watts in the worst-case scenarios (with low PTE). The power rating of electronic and acoustic components is well above this range. Thus the inductive and ultrasonic WPT link reliability against the exposure of electronic and acoustic components to high power levels is not an issue here. Regarding the reliability of the WPT links against distance, misalignment, orientation, and loading variations, we have extensively discussed them in both Sections III and V.

VI. Conclusion

The performances of different scenarios for inductive and ultrasonic WPT to biomedical implants for different Rx sizes and implantation depths were compared to clarify which modality is more suitable for any given biomedical application. Both PTE and maximum PDL under safety constraints were compared, suggesting that 1) for short distances and large Rx size, inductive coupling is the suitable choice, 2) for mm-sized devices, particularly those deeply implanted (several cm) inside the body, ultrasonic links achieve higher performance, 3) optimal load condition is very different for inductive and ultrasonic links and plays an important role in their performances, 4) ultrasonic links cannot necessarily deliver much higher power due to high sound intensity close to the Tx transducer, and 5) if the application involves implant movements, ultrasonic links are drastically prone to misalignments and should be designed for the worst-case misalignment scenarios. These conclusions were made based on our study on several sets of optimized inductive and ultrasonic links with the Rx volumes of 1.1 mm3 and 20 mm3 at different powering distances of 5-50 mm. Our simulation results have been validated by measurements.

Acknowledgments

This work was supported in part by the National Institutes of Health under Grant NIBIB-1U18EB021789-01.

Biographies

graphic file with name nihms959182b1.gif

Ahmed Ibrahim (S’18) received the B.Sc. and M.Sc. degrees in electrical engineering from Ain-Shams University, Cairo, Egypt, in 2008 and 2014, respectively. During his Master’s studies, he designed inverter-based low-voltage continuous-time sigma delta ADCs. From 2008 to 2011, he was working as an RF Design Engineer at Sysdsoft, Cairo, Egypt, where he designed ADCs. From 2011 to 2014, he worked as an Analog Design Engineer at Si-Ware Systems, Cairo, Egypt, where he designed sensor interfaces circuits for consumer MEMS based accelerometers, gyroscopes, and spectrometer including clocking and phase-locked loop circuits. In August 2014, he joined the Penn State University Integrated Circuits and Systems Lab (ICSL), where he is working toward the Ph.D. degree. His research interests are integrated circuit design for bio and wireless applications and wireless power transfer. He serves as a reviewer for the IEEE SENSORS JOURNAL, IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I & II.

graphic file with name nihms959182b2.gif

Miao Meng received his B.S. in Electrical and Computer Engineering from University of Wisconsin Madison, Wisconsin, in May 2011, and M.S. in Electrical Engineering from Columbia University, New York, NY in December 2012. He interned as an IC Designer and worked on level shifter optimization for I/O interface at IBM, China. In Aug. 2014, he joined the integrated circuits and systems lab (ICSL) as a PhD student. His research interest is integrated circuits for bio applications and ultrasound-based wireless power transfer and data communication to/with miniature implantable devices.

graphic file with name nihms959182b3.gif

Mehdi Kiani (S’09-M’14) received the B.S. degree from Shiraz University, Shiraz, Iran, and the M.S. degree from Sharif University of Technology, Tehran, Iran, in 2005 and 2008, respectively. He received his M.S. and Ph.D. degrees in Electrical and Computer Engineering from the Georgia Institute of Technology in 2012 and 2013, respectively. He is currently an assistant professor in the department of Electrical Engineering at the Pennsylvania State University. His research interest is integrated circuits and systems design for bio application. He was the recipient of the Georgia Tech Sigma Xi Best PhD Thesis Award, and Georgia Tech Chih Foundation Research Award for excellent research in the fields of engineering and health sciences. He is an Associate Editor of the IEEE Transactions on Biomedical Circuits and Systems and has served on the subcommittees of the IEEE Sensors and IEEE Custom Integrated Circuits Conference.

Contributor Information

Ahmed Ibrahim, Electrical Engineering Department at the Pennsylvania State University, University Park, PA 16802, USA.

Miao Meng, Electrical Engineering Department at the Pennsylvania State University, University Park, PA 16802, USA.

Mehdi Kiani, Electrical Engineering Department at the Pennsylvania State University, University Park, PA 16802, USA.

References

  • 1.Finkenzeller K. RFID-Handbook. 2nd. Hoboken, NJ: Wiley; 2003. [Google Scholar]
  • 2.Kim C, Seo D, You J, Park J, Cho B. Design of a contactless battery charger for cellular phone. IEEE Trans Ind Electron. 2001 Dec;48(6):1238–1247. [Google Scholar]
  • 3.Hatanaka K, Sato F, Matsuki H, Kikuchi S, Murakami J, Kawase M, Satoh T. Power transmission of a desk with a cord-free power supply. IEEE Trans Magn. 2002 Sep;38(5):3329–3331. [Google Scholar]
  • 4.Lee B, Kiani M, Ghovanloo M. A smart wirelessly powered homecage for long-term high-throughput behavioral experiments. IEEE Sensors Journal. 2015 Sep;15(9):4905–4916. doi: 10.1109/JSEN.2015.2430859. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Lee S, Lee B, Kiani M, Mahmoudi B, Gross R, Ghovanloo M. An inductively-powered wireless neural recording system with a charge sampling analog front-end. IEEE Sensors Journal. 2016 Jan 15;16(2):475–484. doi: 10.1109/JSEN.2015.2483747. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Kilinc EG, Conus G, Weber C, Kawkabani B, Maloberti F, Dehollain C. A system for wireless power transfer of micro-systems in-vivo implantable in freely moving animals. IEEE Sensors Journal. 2014 Feb;14(2):522–531. [Google Scholar]
  • 7.Yilmaz G, Atasoy O, Dehollain C. Wireless energy and data transfer for in-vivo epileptic focus localization. IEEE Sensors Journal. 2013 Nov;13(11):4172–4179. [Google Scholar]
  • 8.Badr B, Somogyi-Gsizmazia R, Delaney K, Dechev N. Wireless power transfer for telemetric devices with variable orientation, for small rodent behavior monitoring. IEEE Sensors Journal. 2015 Apr;15(4):2144–2156. [Google Scholar]
  • 9.Zhou D, Greenbaum E. Implantable neural prostheses 1. New York, NY: Springer; 2009. [Google Scholar]
  • 10.Chen K, Yang Z, Hoang L, Weiland J, Humayun M, Liu W. An integrated 256-channel epiretinal prosthesis. IEEE J Solid State Cir. 2010 Sep;45:1946–1956. [Google Scholar]
  • 11.Yin M, Borton D, Aceros J, Patterson W, Nurmikko A. A 100-channel hermetically sealed implantable device for chronic wireless neurosensing applications. IEEE Trans Biomed Cir Syst. 2013 Apr;7:115–128. doi: 10.1109/TBCAS.2013.2255874. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Nicolelis M. Actions from thoughts. Nature. 2001 doi: 10.1038/35053191. [DOI] [PubMed] [Google Scholar]
  • 13.Lee H, Kwon K, Li W, Ghovanloo M. A power-efficient switched capacitor stimulating system for electrical/optical deep brain stimulation. IEEE Int Solid-State Cir Conf. 2014 Feb;:414–416. [Google Scholar]
  • 14.Deisseroth K. Optogenetics. Nature. 2011 Jan;8 doi: 10.1038/nmeth.f.324. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Ibrahim A, Farajidavar A, Kiani M. A 64-channel wireless implantable system-on-chip for gastric electrical-wave recording. IEEE Sensors Conf. 2016 Oct; [Google Scholar]
  • 16.Zeng F-G, Rebscher S, Harrison W, Sun X, Feng H. Cochlear implants: System design, integration, and evaluation. IEEE Rev Biomed Eng. 2008;1:115–142. doi: 10.1109/RBME.2008.2008250. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Weiland J, Humayun M. Visual prosthesis. Proc IEEE. 2008 Jul;96(7):1076–1084. [Google Scholar]
  • 18.Lebedev MA, Nicolelis MA. Brain–machine interfaces: Past, present and future. Trends Neurosci. 2006;29(9):536–546. doi: 10.1016/j.tins.2006.07.004. [DOI] [PubMed] [Google Scholar]
  • 19.DiMarco J. Implantable cardioverter-defibrillators. New England J Med. 2003;349(19):1836–1847. doi: 10.1056/NEJMra035432. [DOI] [PubMed] [Google Scholar]
  • 20.Agarwal K, Jegadeesan R, Guo Y, Thakor N. Wireless power transfer strategies for implantable bioelectronics. IEEE Rev Biomed Eng. 2017 Dec;10:136–161. doi: 10.1109/RBME.2017.2683520. [DOI] [PubMed] [Google Scholar]
  • 21.Muller R, Le H, Li W, Ledochowitsch P, Gambini S, Bjorninen T, Koralek A, Carmena J, Maharbiz M, Alon E, Rabaey J. A minimally invasive 64-channel wireless μECoG implant. IEEE J Solid State Cir. 2015 Jan;50:344–359. [Google Scholar]
  • 22.Borton DA, Yin M, Aceros J, Nurmikko A. An implantable wireless neural interface for recording cortical circuit dynamics in moving primates. J Neural Eng. 2013 Apr;10(2) doi: 10.1088/1741-2560/10/2/026010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Harrison R. Designing efficient inductive power links for implantable devices. IEEE Int Symp Cir Syst. 2007 May;:2080–2083. [Google Scholar]
  • 24.Agarwal K, Jegadeesan R, Guo YX, Thakor NV. Wireless power transfer strategies for implantable bioelectronics. IEEE Reviews in Biomedical Engineering. 2017;10:136–161. doi: 10.1109/RBME.2017.2683520. [DOI] [PubMed] [Google Scholar]
  • 25.Baker MW, Sarpeshkar R. Feedback analysis and design of RF power links for low-power bionic systems. IEEE Trans Biomed Cir Syst. 2007 Mar;1(1):28–38. doi: 10.1109/TBCAS.2007.893180. [DOI] [PubMed] [Google Scholar]
  • 26.Jow UM, Ghovanloo M. Design and optimization of printed spiral coils for efficient transcutaneous inductive power transmission. IEEE Trans Biomed Cir Syst. 2007 Sep;1:193–202. doi: 10.1109/TBCAS.2007.913130. [DOI] [PubMed] [Google Scholar]
  • 27.McConnell G, Rees H, Levey A, Gutekunst C, Gross R, Bellamkonda R. Implanted neural electrodes cause chronic, local inflammation that is correlated with local neurodegeneration. J Neural Eng. 2009 Oct;6 doi: 10.1088/1741-2560/6/5/056003. [DOI] [PubMed] [Google Scholar]
  • 28.Pivonka D, Yakovlev A, Poon A, Meng T. A mm-sized wirelessly powered and remotely controlled locomotive implant. IEEE Trans Biomed Cir Syst. 2012 Dec;6(6):523–532. doi: 10.1109/TBCAS.2012.2232665. [DOI] [PubMed] [Google Scholar]
  • 29.Kim S, Ho J, Poon A. Wireless power transfer to miniature implants: transmitter optimization. IEEE Trans Ant Propag. 2012 Oct;60(10):4838–4845. [Google Scholar]
  • 30.Ho J, Yeh A, Neofytou E, Kim S, Tanabe Y, Patlolla B, Beygui R, Poon A. Wireless power transfer to deep-tissue microimplants. Proc Natl Acad Sci. 2014 Jun;111:7974–7979. doi: 10.1073/pnas.1403002111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Zargham M, Gulak P. Maximum achievable efficiency in near-field coupled power-transfer systems. IEEE Trans Biomed Cir Syst. 2012 Jun;6:228–245. doi: 10.1109/TBCAS.2011.2174794. [DOI] [PubMed] [Google Scholar]
  • 32.Mark M, Bjorninen T, Ukkonen L, Sydanheimo L, Rabaey J. SAR reduction and link optimization for mm-size remotely powered wireless implants using segmented loop antennas. Biomed Wireless Techn Networks Sensing Syst (BioWireleSS) 2011 Jan;:7–10. [Google Scholar]
  • 33.Ahn D, Ghovanloo M. Optimal design of wireless power transmission links for millimeter-sized biomedical implants. IEEE Trans Biomed Cir Syst. 2016 Feb;10(1):125–137. doi: 10.1109/TBCAS.2014.2370794. [DOI] [PubMed] [Google Scholar]
  • 34.Ibrahim A, Kiani M. A figure-of-merit for design and optimization of inductive power transmission links for millimeter-sized biomedical implants. IEEE Trans Biomed Cir Syst. 2016 Dec;10(6):1100–1111. doi: 10.1109/TBCAS.2016.2515541. [DOI] [PubMed] [Google Scholar]
  • 35. (IEEE Standard C95.1).IEEE standard for the safety levels with respect to human exposure to radiofrequency electromagnetic fields, 3 KHz to 300 GHz. 2006 [Google Scholar]
  • 36.Seo D, Carmena J, Rabaey J, Maharbiz M, Alon E. Model validation of untethered, ultrasonic neural dust motes for cortical recording. J Neurosci Methods. 2015 Apr;244:114–122. doi: 10.1016/j.jneumeth.2014.07.025. [DOI] [PubMed] [Google Scholar]
  • 37.Charthad J, Weber M, Chang T, Arbabian A. A mm-sized implantable medical device (IMD) with ultrasonic power transfer and a hybrid bi-directional data link. IEEE J Solid-State Cir. 2015 Aug;50:1–13. [Google Scholar]
  • 38.Song S, Kim A, Ziaie B. Omni-directional ultrasonic powering for millimeter-scale implantable devices. IEEE Trans Biomed Eng. 2015 Nov;62(11):2717–2723. doi: 10.1109/TBME.2015.2444854. [DOI] [PubMed] [Google Scholar]
  • 39.Meng M, Kiani M. Design and optimization of ultrasonic wireless power transmission links for millimeter-sized biomedical implants. IEEE Trans Biomed Cir Syst. 2017 Feb;11(1):98–107. doi: 10.1109/TBCAS.2016.2583783. [DOI] [PubMed] [Google Scholar]
  • 40.Denisov A, Yeatman E. Ultrasonic vs. inductive power delivery for miniature biomedical implants. International Conf Body Sensor Networks. 2010:84–89. [Google Scholar]
  • 41.Chou T, Subramanian R, Park J, Mercier P. A miniaturized ultrasonic power delivery system. IEEE Biomed Cir Syst Conf. 2014 Oct;:440–443. M. [Google Scholar]
  • 42.Ibrahim A, Meng M, Kiani M. Inductive and ultrasonic wireless power transmission to biomedical implants. Intern Symp Cir Syst. 2017 May; doi: 10.1109/JSEN.2018.2812420. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Meng M, Kiani M. A hybrid inductive-ultrasonic link for wireless power transmission to millimeter-sized biomedical implants. IEEE Trans Cir Syst II. 2017 Oct;64(10):1137–1141. doi: 10.1109/TBCAS.2016.2583783. [DOI] [PubMed] [Google Scholar]
  • 44.Kiani M, Jow U, Ghovanloo M. Design and optimization of a 3-coil inductive link for efficient wireless power transmission. IEEE Trans Biomed Cir Syst. 2011 Dec;5:579–591. doi: 10.1109/TBCAS.2011.2158431. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Ibrahim A, Kiani M. Inductive power transmission to millimeter-sized biomedical implants using printed spiral coils. IEEE 38th Eng Med Biol Conf. 2016 Aug; doi: 10.1109/EMBC.2016.7591801. [DOI] [PubMed] [Google Scholar]
  • 46.Food and Drug Administration (FDA) Information for manufacturers seeking marketing clearance of diagnostic ultrasound systems and transducers. Guidance for Industry and FDA Staff. 2008 Sep 9; [Google Scholar]
  • 47.Fyke F, Greenleaf J, Johnson S. Continuous wave measurements of acoustic attenuation in an oil/polymer mixture. Ultrasound Med Biol. 1978;5:87–90. doi: 10.1016/0301-5629(79)90132-7. [DOI] [PubMed] [Google Scholar]
  • 48.Callens D, Bruneel C, Assaad J. Matching ultrasonic transducer using two matching layers where one of them is glue. NDT&E International. 2004 Dec;37:591–596. [Google Scholar]
  • 49.Kim J, Lim H, Lee S, Kim Y. Thickness of rectus abdominis muscle and abdominal subcutaneous fat tissue in adult women: correlation with age, pregnancy, laparotomy, and body mass index. Archives of Plastic Surgery. 2012;39(5):528. doi: 10.5999/aps.2012.39.5.528. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Akkus O. Evaluation of skin and subcutaneous adipose tissue thickness for optimal insulin injection. Journal of Diabetes & Metabolism. 2012;03(8) [Google Scholar]
  • 51.Edelman S. ESP Ultrasound. 4. Jul, 2012. Understanding ultrasound physics. [Google Scholar]
  • 52.Azhari H. Basics of biomedical ultrasound for engineers. John Wiley & Sons; Apr, 2010. [Google Scholar]
  • 53.Gabriel S, Lau R, Gabriel C. The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues. Phys Med Biol. 1996 Nov;41(11):2271–2293. doi: 10.1088/0031-9155/41/11/003. [DOI] [PubMed] [Google Scholar]
  • 54.Timko B, Cohen-Karni T, Yu G, Qing Q, Tian B, Lieber C. Electrical recording from hearts with flexible nanowire device arrays. Nano letters. 2009 Jan;9:914–918. doi: 10.1021/nl900096z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Seo D, Tang H, Carmena J, Rabaey J, Alon E, Boser B, Maharbiz M. Ultrasonic beamforming system for interrogating multiple implantable sensors. IEEE 37th Eng Med Biol Conf. 2015 Aug; doi: 10.1109/EMBC.2015.7318942. [DOI] [PubMed] [Google Scholar]
  • 56.Kiani M, Ghovanloo M. An RFID-based closed loop wireless power transmission system for biomedical applications. IEEE Trans Cir Syst-II. 2010 Apr;57(4):260–264. doi: 10.1109/TCSII.2010.2043470. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Kiani M, Ghovanloo M. A figure-of-merit for designing high performance inductive power transmission links. IEEE Trans Indus Elect. 2013 Nov;60:5292–5305. doi: 10.1109/TIE.2012.2227914. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Paskaranandavadivel N, Wang R, Sathar S, O’Grady G, Cheng L, Farajidavar A. A wireless multichannel system for in-vivo monitoring of gastric activity propagation. Neurogastroenterology & Motility. 2015 Apr;27:580–585. doi: 10.1111/nmo.12515. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES