Abstract
Background
The concept of ‘patient experience’ has become central to how to improve healthcare. Remote communication with patients is today a frequent practice in healthcare services, showing similar outcomes to standard outpatient care while enabling cost reduction in both formal and informal care. The purpose of this study was to analyse the experiences of people with telemonitoring pacemakers.
Methods
Patients were randomly allocated to either the telemonitoring or hospital monitoring follow-ups. Using the ‘Generic Short Patient Experiences Questionnaire’ (GS-PEQ), as well as an ad-hoc survey from the ‘telehealth patient satisfaction survey’ and ‘costs survey’, patients’ experiences were measured six months after the pacemaker implant in a cohort of 50 consecutive patients. The mean age was 74.8 (± 11.75) years and 26 (52%) patients were male of which 1 was lost in follow-up. Finally, 24 patients were followed up with standard hospital monitoring, while 25 used the telemonitoring system. Differences in baseline characteristics between groups were not found.
Results
Findings showed overall positive and similar experiences in patients living with telemonitoring and hospital monitoring pacemakers. Significant differences were found in GS-PEQ concerning how telemonitoring patients received less information about their diagnosis/afflictions (p = 0.046). We did not find significant differences in other items such as ‘confidence in the clinicians’ professional skills’, ‘treatment perception adapted to their situation’, ‘involvement in decisions regarding the treatment’, ‘perception of hospital organisation’, ‘waiting before admission’, ‘satisfaction of help and treatment received’, ‘benefit received’, and ‘incorrect treatment’.
Conclusions
The remote communication of pacemakers was met with positive levels of patients’ experiences similarly to patients in the hospital monitoring follow-up. However, telemonitoring patients received less information. Thus, improving the quality and timing of information is required in telemonitoring patients in the planning and organisation of future remote communication healthcare services for people living with a pacemaker implant.
1. Introduction
Cardiovascular diseases are one of the most frequent reasons of disease-associated mortality [1]. In Norway, cardiovascular diseases are one of the main cause-of-death disease groups, with myocardial infarction being one of the five most important specific causes of death [2]. Worldwide around 3 million people have a pacemaker and around 600,000 pacemakers are implanted every year [3]. According to clinical guidelines, patients with an implanted PM need to be followed-up every 3–12 months [4, 5]. Every consultation frequently involves an assessment of the device’s function, cardiac events, and the patient’s clinical status and, if needed, the pacemaker is reprogrammed or pharmacotherapy is changed [4, 6]. In our current ageing population, there are increasing indications with respect to persons carrying an implant. The follow-up is today a substantial load for national health services [7] as well as for patients and their relatives [8, 9]. In fact, it has been shown that people with cardiovascular conditions demand a large amount of follow-ups and hospital admissions in Norway [10]. In this regard, remote communication in cardiology could mitigate the increasing workload of follow-ups of pacemakers [11].
In the last Norwegian Coordination Reform in the healthcare sector [12], remote communication and the use of telehealth strategies were emphasised. Remote communication between healthcare providers and patients is considered today a tool that decreases outpatient consultations and costs [13]. In fact, long-distance communication with implanted devices is now a reality. Recent studies have shown that using telemonitoring in patients with pacemakers can result in similar clinical outcomes to standard outpatient care while allowing more flexible services organisation and greater cost reduction in both formal and informal care [9, 14, 15]. Moreover, remote monitoring (RM) or telemonitoring (TM) systems have potential advantages such as early detection of cardiovascular events and early response to technical problems in the device or alterations in the patient’s clinical condition [16, 17]. TM could represent a possible solution in helping to reduce the number of consultations and travels to hospital, thereby optimising healthcare resources [15].
Pacemakers are devices that help to monitor abnormal heart rhythms and regulate heartbeats being placed in the chest [1,18]. A pacemaker implant is a very important life event, so these patients may have different and particular experiences of living with such a device [1]. In this regard, it is important to analyse their experiences in order to develop high-quality healthcare services [19]. Indeed, the experience of patients is more and more being considered clinically important and the concept of ‘patient experience’ has become relevant and central when it comes to increase the quality of healthcare services [20]. In telehealth, both the technology and users’ experiences are key determinants of the acceptance and accomplishment of these implementations [21].
Although many economic evaluations and outcome studies have been conducted, few have inquired as to how individuals experience living with remote monitoring pacemakers. We found one study analysing patients’ experiences with standard on-clinic monitored pacemakers [22]. A previous literature review [23] found three studies comparing the experiences of people with pacemakers and implantable cardiac defibrillators (ICD) [24–26]. In the field of telemonitoring cardiac devices, one study analysed diverse aspects of remote monitored ICDs, such as ease of use of the system, acceptance, and satisfaction of patients and clinicians [21]. A recent study [20] enrolled 14 patients with an insertable cardiac monitor (ICM) so as to explore patients’ experiences, and claimed the need for studies on other cardiac telemonitoring technologies. We have not found previous studies specifically on patients’ experiences with remote monitoring pacemakers. Therefore, the aim of this study was to explore the experiences of people living with telemonitoring pacemakers. The rationale was to produce pertinent and translatable knowledge for future opportunities in healthcare contexts of these patients and to direct future research.
2. Materials and methods
This paper is part of a larger project in Norway, the NORDLAND study (2014–2017), wherein a team has collaborated which includes chronic heart patients with a pacemaker, their relatives, cardiologists, nurses, psychologists, and health communication experts. This study is based on a randomised, non-masked observational design where participants were assigned to either follow-up consultations in the hospital or follow-up by remote communication technologies. Participants were recruited in Nordland Hospital, Bodø, Norway. This hospital with a pacemaker centre covers 170,000 inhabitants and conducts around 80–90 pacemaker implants per year [27].
The following protocol has been described in detail previously [27]. Every patient who was scheduled for a pacemaker implant between August 2014 and October 2015 and met all of the inclusion criteria: aged 18 years or older, capacity to provide informed consent and to manoeuvre the home monitor system, and life expectancy >1 year, as well as none of the exclusion criteria: scheduled for an implantable cardioverter-defibrillator (ICD) or cardiac resynchronisation therapy (CRT) and involvement in other studies at the same time, was invited to participate. A total of 76 patients were screened and 50 patients were included and randomised to either telemonitoring (TM, n = 25) or hospital monitoring (HM, n = 25), before being implanted with the pacemaker. The randomisation process was performed as follows: a person unrelated to the study prepared a total of 50 sealed envelopes, 25 of which included a note reading "tele-monitoring" and 25 a note reading "hospital monitoring". The envelopes were thoroughly mixed and numbered from 1 to 50. When a patient accepted the invitation to participate in the trial and signed the informed consent he or she received a consecutive number and was allocated to follow-up in accordance with the specification included in the corresponding envelope. Thus, the investigators had no knowledge of or influence on the randomisation result prior to inclusion.
Depending on their diagnosis, patients were implanted with either a single (VVIR) or a dual chamber (DDDR) pacemaker. For further characteristics and descriptions of the devices that were received by each group of participants please see the previously published article [27].
Data collection was performed 6 months after surgery, by a phone call from one of the research team members. In total, 24 HM patients and 25 TM patients participated, with each participant answering 20 questions. To assess the experiences of the users with PM, an ad-hoc questionnaire was created by merging validated questionnaires assessing users’ experiences. The items included comprised the full version of the Generic Short Patient Experiences Questionnaire (GS-PEQ), adding some questions from the telehealth patient satisfaction survey and a costs survey. The Generic Short Patient Experiences Questionnaire (GS-PEQ) is a short set of questions on user experiences with specialist healthcare that covers certain relevant topics. It was created and has been validated in Norway [28]. The questionnaire includes 10 questions (see questions 1–10 in the S1 File) about the following topics: outcome (2), clinician services (2), user involvement (2), incorrect treatment (1), information (1), organisation (1), and accessibility (1). Additionally, to evaluate other aspects of the telehealth experience by patients who have been implanted with home-monitored pacemakers, we used an adapted version of the telehealth patient satisfaction survey [29] and the costs survey [30]. These surveys are composed of closed- and open-ended questions exploring the patients’ experiences with the home monitoring technology, as well as some specific data on costs that they might have with regard to the pacemaker monitoring (see questions 11–20 in the S1 File).
The protocol was approved by the Regional Ethics Committee–REK Nord (Tromsø, Norway), with the committee´s reference number being as follows: 2014/383/REK Nord. The study was developed in accordance with the precepts of the Declaration of Helsinki. All patients signed the corresponding informed consent prior to their enrolment and appropriate measures were taken to ensure data privacy.
With regard to statistical analyses, following a similar approach of a previous study [9], first patient baseline characteristics and potential differences between groups were compared using a difference in means test for continuous variables and a difference in proportions test (binomial method) or chi-square test (replaced by the Fisher exact test for cells with n<5 cases) for qualitative variables. Secondly, results from the questionnaire were presented on a single question basis with comparison between the two groups, telemonitoring and hospital monitoring, using the Mann–Whitney U test for ordinal data and the chi-square test for nominal data. Analyses were carried out with SPSS 24th edition (SPSS Institute, Inc., Chicago, IL, USA) statistical software. Fig 1 shows the flow (CONSORT) diagram of the study. This figure has been previously published [27].
3. Results
Baseline characteristics are presented in Table 1. This table has been previously published [27]. There were no major differences between the telemonitoring and the hospital monitoring groups with regard to age, gender, pacing indication, and other clinical characteristics. Differences were only found in those items that are directly influenced by the type of monitoring: ‘transmissions from patient’s home’ and ‘calls/letters sent to the patients’.
Table 1. Selected patient baseline characteristics by intervention status.
All (n = 50) | Groups | P-value | ||
---|---|---|---|---|
Telemonitoring | Hospital monitoring | |||
Age | 74.84 (± 11.75) | 73.68 (± 14.22) | 76.00 (± 8.77) | 0.676 |
Men | 26 (52.0) | 13 (52.0) | 13 (52.0) | 1.00 |
Pacing indication N (%) | ||||
Sick sinus syndrome | 24 (48.0) | 12 (48.0) | 12 (48.0) | 0.648 |
Atrioventricular block | 20 (40.0) | 11 (44.0) | 9 (36.0) | |
Chronic AF with bradycardia | 6 (12.0) | 2 (8.0) | 4 (16.0) | |
Disease manifestations N (%) | ||||
Syncope | 14 (28.0) | 8 (32.0) | 6 (24.0) | 0.812 |
Dizziness | 25 (50.0) | 12 (48.0) | 13 (52.0) | |
Dyspnoea | 11 (22.0) | 5 (20.0) | 6 (24.0) | |
Service derived N (%) | ||||
Emergency dept. | 3 (6.0) | 1 (4.0) | 2 (8.0) | 0.505 |
Cardiology ward | 14 (28.0) | 5 (20.0) | 9 (36.0) | |
Primary healthcare | 4 (8.0) | 2 (8.0) | 2 (8.0) | |
Other hospitals | 29 (58.0) | 17 (68.0) | 12 (48.0) | |
Stimulation N (%) | ||||
DDDR | 44 (88.0) | 23 (92.0) | 21 (84.0) | 0.334 |
VVIR | 6 (12.0) | 2 (8.0) | 4 (16.0) | |
Comorbidities N (%) | ||||
Dislipidemia | 27 (54.0) | 13 (52.0) | 14 (56.0) | 0.500 |
Obesity (BMI >30) | 1 (2.0) | 0 (0.0) | 1 (4.0) | 0.500 |
Tachyarrhythmia | 18 (36.0) | 7 (28.0) | 11 (44.0) | 0.189 |
Hypertension | 32 (64.0) | 17 (68.0) | 15 (60.0) | 0.384 |
Other comorbidities N (%) | ||||
None | 18 (36.0) | 11 (44.0) | 7 (28.0) | 0.388 |
Others | 10 (20.0) | 6 (24.0) | 4 (16.0) | |
Coronary heart diseases | 22 (44.0) | 8 (32.0) | 14 (56.0) | |
Pharmaceutical treatment N (%) | ||||
Antiaggregants | 18 (36.0) | 8 (32.0) | 10 (40.0) | 0.384 |
Anticoagulants | 25 (50.0) | 10 (40.0) | 15 (60.0) | 0.129 |
Antiarrhythmics | 18 (36.0) | 7 (28.0) | 11 (44.0) | 0.189 |
Antihypertensives | 32 (64.0) | 18 (72.0) | 14 (56.0) | 0.189 |
n = 50 (Telemonitoring group: 25; Hospital monitoring group: 25). Values are expressed as means or proportions. SD: Standard deviation; 95 CI: 95% confidence interval of means; DDDR: Bicameral pacemaker with two electrodes placed in the atrium and in the ventricle; VVIR: Unicameral pacemaker with an electrode in the ventricle with the ability to modulate frequency of stimulation; BMI: Body mass index. Note: this table has been previously published in a previous article [27].
In relation to the results derived from the questionnaire, overall experience with both types of follow-ups was positive and, as presented in Table 2, there were few differences between the home monitoring and the hospital monitoring groups with respect to the individual questions. Using the Mann–Whitney U test, χ2 test and Fisher exact test, we only found significant p values in two questions (Q3 and Q12). Q3 asked whether patients obtained sufficient information about their diagnosis/afflictions, and Q12 asked for the time that it takes patients to attend a cardiology consultation. In general, responses showed positive experiences, and although these were also positive, the lowest scoring was for Q4 and Q6, for both the intervention and control group patients. These questions concern treatment and organisational perceptions by patients (see Table 3 and full data set in the S1 File).
Table 2. Follow-up information at 6 months.
All (n = 49) | Groups | P-value | ||
---|---|---|---|---|
Telemonitoring | Hospital monitoring | |||
Number of transmissions from hospital N (%) | ||||
0 | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0.26 |
1 | 41 (83.7) | 21 (84.0) | 20 (83.3) | |
2 | 6 (12.2) | 2 (8.0) | 4 (16.7) | |
3 | 2 (4.1) | 2 (8.0) | 0 (0.0) | |
Number of transmissions from patient’s home N (%) | ||||
0 | 29 (59.2) | 5 (20.0) | 24 (100) | <0.001 |
3–5 | 15 (30.6) | 15 (60.0) | 0 (0.0) | |
6–8 | 5 (10.2) | 5 (20.0) | 0 (0.0) | |
Extra transmissions from patient’s home N (%) | ||||
0 | 45 (91.8) | 21 (84.0) | 24 (100) | 0.12 |
1 | 1 (2.0) | 1 (4.0) | 0 (0.0) | |
3 | 3 (6.2) | 3 (12.0) | 0 (0.0) | |
Cardiovascular events N (%) | ||||
None | 46 (93.9) | 23 (92.0) | 23 (95.8) | 0.40 |
PCI | 1 (2.0) | 1 (4.0) | 0 (0.0) | |
Angina | 1 (2.0) | 0 (0.0) | 1 (4.2) | |
Lead dislodgement x 2 | 1 (2.0) | 1 (2.0) | 0 (0.0) | |
Calls/letters sent to the patients N (%) | ||||
0 | 27 (55.1) | 4 (16.0) | 23 (95.8) | <0.001 |
1 | 21 (42.9) | 20 (80.0) | 1 (4.2) | |
3 | 1 (2.0) | 1 (4.0) | 0 (0.0) | |
Changes of medication N (%) | ||||
0 | 33 (67.3) | 17 (68.0) | 16 (66.7) | 0.11 |
1 | 7 (14.3) | 5 (20.0) | 2 (8.3) | |
2 | 3 (6.1) | 1 (4.0) | 2 (8.3) | |
3 | 4 (8.2) | 0 (0.0) | 4 (16.7) | |
4 | 2 (4.1) | 2 (8.0) | 0 (0.0) | |
Changes of pacemaker’s reprogramming N (%) | ||||
0 | 34 (69.4) | 16 (64.0) | 18 (75.0) | 0.34 |
1 | 13 (26.5) | 7 (28.0) | 6 (25.0) | |
2 | 2 (4.1) | 2 (8.0) | 0 (0.0) | |
Number of hospitalisations (related or not to pacemaker’s implant) N (%) | ||||
0 | 30 (61.2) | 14 (56.0) | 16 (66.7) | 0.55 |
1 | 14 (28.6) | 7 (28.0) | 7 (29.2) | |
2 | 4 (8.2) | 3 (12.0) | 1 (4.2) | |
5 | 1 (2.0) | 1 (2.0) | 0 (0.0) | |
Number of hospitalisation days (related or not to pacemaker’s implant) N (%) | ||||
0 | 30 (61.2) | 14 (56.0) | 16 (66.7) | 0.54 |
1–5 | 12 (24.5) | 6 (24.0) | 6 (25.1) | |
6–10 | 4 (8.1) | 2 (8.0) | 2 (8.4) | |
+10 | 3 (6.0) | 3 (12.0) | 0 (0.0) | |
Reasons for hospitalisation N (%) | ||||
None | 30 (61.2) | 14 (56.0) | 16 (66.7) | 0.37 |
Others | 6 (12.3) | 3 (12.0) | 3 (12.5) | |
Cancer | 1 (2.0) | 1 (4.0) | 0 (0.0) | |
Coronary problems | 9 (18.4) | 4 (16.0) | 5 (20.8) | |
Pacemaker dysfunction | 3 (6.1) | 3 (12.0) | 0 (0.0) |
n = 50 (Telemonitoring group: 25; Hospital monitoring group: 24). Values are expressed as means or proportions. SD: Standard deviation; 95 CI: 95% confidence interval of means; PCI: Percutaneous coronary intervention. Note: this table has been previously published in a previous article [27].
Table 3. Results derived from the Generic Short Patient Experiences Questionnaire (GS-PEQ).
Question | Answering categories* | Telemonitoring group (n = 23) |
Hospital monitoring group (n = 23) |
p-value |
---|---|---|---|---|
Question 1† | 1 = Not at all; 2 = To a small extent; 3 = To a moderate extent; 4 = To a large extent; 5 = To a very large extent | 4 (1, 5) | 5 (2, 5) | 0.214 |
Question 2† | 5 (4, 5) | 5 (3, 5) | 0.326 | |
Question 3† | 4 (2, 5) | 5 (1, 5) | *0.046 | |
Question 4† | 4 (3, 5) | 5 (3, 5) | 0.241 | |
Question 5† | 3 (1, 5) | 3 (1, 5) | 0.091 | |
Question 6† | 4 (3, 5) | 4 (3, 5) | 0.712 | |
Question 7† | 5 (3, 5) | 5 (3, 5) | 0.362 | |
Question 8† | 1 (1, 3) | 1 (1, 2) | 0.613 | |
Question 9† | 1 (1, 2) | 1 (1, 5) | 0.492 | |
Question 10† | 4 (1, 5) | 5 (3, 5) | 0.404 | |
Question 11† | Number of kilometres | 65 (1, 500) | 40 (1, 300) | .183 |
Question 12† | 1 = <1 hour; 2 = 1–2 hours; 3 = 2–3 hours; 4 = 3–4 hours; 5 = >4 hours | >4 (<1, >4) | 2–3 (<1, >4) | *.041 |
Question 13‡ | 1 = Public transport | 8 (34.8) | 3 (13.0) | .323 |
2 = Own car | 0 | 1 (4.3) | ||
3 = Ambulance | 10 (43.5) | 10 (43.5) | ||
4 = Taxi | 1 (4.3) | 3 (13.0) | ||
5 = Other | 4 (17.4) | 6 (26.0) | ||
Question 14‡ | 1 = Working | 1 (4.3) | 2 (8.7) | .229 |
2 = Unemployed | 0 | 0 | ||
3 = Pensionist | 12 (52.2) | 8 (34.8) | ||
4 = Sick leave | 8 (34.8) | 13 (56.5) | ||
5 = Other | 2 (8.7) | 0 | ||
Question 15‡ | Yes | 3 (13.0) | 6 (26.1) | .265 |
No | 20 (87.0) | 17 (73.9) | ||
Question 16‡ | 1 = Working | 2 (50.0) | 4 (57.1) | .368 |
2 = Unemployed | 0 | 0 | ||
3 = Pensioner | 1 (25.0) | 3 (42.9) | ||
4 = Sick leave | 1 (25.0) | 0 | ||
5 = Other | 0 | 0 | ||
Question 17‡ | Yes | 17 (73.9) | 15 (65.2) | .522 |
No | 6 (26.1) | 8 (34.8) | ||
Question 18† | 36.55 (0–33.25) | 22.10 (0–110.75) | .611 | |
Question 19‡ | 1 = None | 19 (82.6) | 22 (95.6) | .200 |
2 = Once | 3 (13.0) | 0 | ||
3 = Twice | 0 | 0 | ||
4 = More than 2 | 1 (4.4) | 1 (4.4) | ||
Question 20‡ | 1 = None | 20 (87.0) | 22 (95.6) | .351 |
2 = Once | 2 (8.7) | 0 | ||
3 = Twice | 0 | 0 | ||
4 = More than 2 | 1 (4.3) | 1 (4.4) |
*For questions 1–10, the following scoring was used: 1, not at all; 2, to a small extent; 3, to some extent; 4, to a large extent; and 5, to a very large extent.
†Data presented as median (min., max.).
‡Data presented as total number (percentage).
4. Discussion
The NORDLAND study explored the patients’ experiences with living with telemonitoring pacemakers. To the best of our knowledge, this is the first study to explore patients’ experiences in respect of living with telemonitoring pacemakers. The findings revealed: i) overall positive experiences in patients living with telemonitoring pacemakers; ii) significant differences in the GS-PEQ between both groups concerning telemonitoring patients receiving less information about their diagnosis/afflictions than those ones in hospital monitoring; iii) significant differences in how TM patients take more time to attend a cardiology consultation at hospital than HM patients; and iv) that no significant differences between groups were found in the rest of the items, such as ‘confidence in the clinicians’ professional skills’, ‘treatment perception adapted to their situation’, ‘involvement in decisions regarding the treatment’, ‘perception of hospital organisation’, ‘waiting before admission’, ‘satisfaction of help and treatment received’, ‘benefit received’, and ‘incorrect treatment’.
The overall positive experience found in our study is confirmed and well aligned with previous studies analysing experiences of telehealth devices. In a previous study evaluating telehealth in primary care [31], patients found that this new remote communication with healthcare providers was a flexible, convenient, easy-to-use and acceptable means of their jointly managing their condition with a responsible health professional. Another study [21] found that the ease of use, satisfaction, and acceptance of remote monitoring of implantable defibrillators appeared to be elevated both for patients and for clinicians. However, in spite of these positive results, we need to be cautious, since the implementation of telehealth in regular healthcare practice is difficult [21]. Indeed, 15–20 years ago, it was reported that most telehealth initiatives had become a failure in daily practice, since poor technical feasibility often results in distrust from users and low levels of satisfaction [32, 33]. Nowadays, we can see that technical problems are very much minimised; thus, users’ experiences are more positive, as our findings have revealed.
Besides the overall positive experiences, one relevant outcome found in our study was that telemonitoring patients received less information about their diagnosis/afflictions than those ones in hospital monitoring. This is an important finding concerning the expectations of the information that are normally provided by conversations with the health professionals and the delivery of pamphlets. We believe that due to the reduction of face-to-face consultations, these TM patients had a lesser chance of posing questions or clarifying issues in relation to their diagnosis with the health professionals. Being well informed is a key ingredient in patients’ subjective experiences [34]. A recent study [20] also found that a characteristic part of the diagnostic process in ICM remote monitoring is that patients experience the feeling of ‘not knowing’ or ‘being uninformed’. According to the authors, this might be a result of “no news is good news” in the home monitoring environment. The monitoring systems only inform the health professionals when something goes wrong. According to the authors, this collides with the needs of patients. Frequent contact with clinicians is needed so as to receive information on the health status, especially for patients who are asymptomatic (as this contact is the only way in which to know about their progress). Other previous studies [24, 25] carried out on patients with hospital monitoring cardiac implants also support the recommendations of providing more information such as preceding to implantation. Among the patients, these studies found more fear of death, of device malfunction, greater concerns about not being able to work, and more worry with regard to having sex and driving. These studies recommend that as these patients are mentally ill from depression or anxiety, they suggest that health professionals should be aware of significant symptoms, meeting the patients’ needs. Although we did not find similar studies on telemonitoring patients with pacemakers, we believe that these patients might have higher information needs than hospital monitoring, since their face-to-face or phone contact with healthcare professionals is lower and, therefore, they might have fewer opportunities to obtain information. This may lead us to an Albert Einstein quote: “It has become appallingly obvious that our technology has exceeded our humanity.” In this regard, we agree with Boriani et al. [11] in suggesting that the evolving capabilities of implanted devices to monitor patients’ cardiac status (heart rhythm, fluid overload, right ventricular pressure, oximetry, etc.) may imply a shift from strictly device-centred follow-up to perspectives centred on the patient (and patient–device interactions). A patient-centred approach could provide improvements in healthcare delivery and clinical outcomes in telemonitoring patients.
Finally, we should mention that we found another difference between groups in Q12, which asked for the time that it takes patients to attend a cardiology consultation. However, we believe that this significant difference was a result of chance in spite of the p-value, as all participants were randomly allocated to one of both groups, either telemonitoring or hospital monitoring follow-up.
Despite the relevant results obtained, the study has certain limitations. Firstly, some issues in relation to measuring patient experience include the feedback biases by answering intentionally to the questionnaires in order to accomplish positive outcomes. Some patients are also confused because they consider the questions to be on their experience of their health condition [35]. Secondly, this is an open study wherein clinicians, researchers and patients knew the follow-up category for all participants. Nevertheless, we believe that our study presents some important strong points since the NORDLAND study is a randomised study in a field where it is not frequent to conduct such method design. This ensures a major evidence level, a lesser chance of bias due to random selection of the groups, and it might be repeatable and comparable with other studies [27].
5. Conclusions
Our results showed that patients living with telemonitoring pacemakers have overall positive experiences, similar to those with hospital monitoring pacemakers. Areas for improvement should focus on improving the quality and timing of information during the entire therapeutic process. To the best of our knowledge, this is the first publication to show patients’ experiences of living with telemonitoring pacemakers.
Our study also includes some important practical implications. We believe that by achieving a better understanding of the experiential dimensions of these patients we can inform communication practices between healthcare professionals and patients in the follow-up process of a pacemaker implant. In fact, such little research into patient experiences of living with telemonitoring pacemakers could also be a result of the poor educational preparation of health professionals responsible for these patients. Therefore, our findings will support the development of this important research field [36, 37]. Further organisation and management of healthcare services should ensure the delivery of quality and timely information to patients during the entire process, from pre-implantation to the follow-up phases.
Supporting information
Acknowledgments
We would like to thank the patients who took part in the study and acknowledge the invaluable support of the NORDLAND research team. Furthermore, special thanks go to Mr. Terje Enebakk and Ms. Hilde Thunhaug, who gathered the information from patients.
Data Availability
All relevant data are within the paper and its Supporting Information files.
Funding Statement
This study has received financial support from the European Economic Area Grants – Science and Sustainability Program, ref. 008/ABELCM/2014A, received by DCM, and the Consejería de Salud, Junta de Andalucía, ref. PI/0256/2017, received by ALV. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
References
- 1.Ghojazadeh M, Azami-Aghdash S, Sohrab-Navi Z, Kolahdouzan K. Cardiovascular patients’ experiences of living with pacemaker: Qualitative study. ARYA Atheroscler. 2015;11:281–8. [PMC free article] [PubMed] [Google Scholar]
- 2.Norwegian Institute of Public Health. Disease Burden in Norway 2015 Results from the Global Burden of Diseases, Injuries, and Risk Factors Study 2015. Oslo, Norway: Norwegian Institute for Public Health; 2017. https://www.fhi.no/globalassets/dokumenterfiler/rapporter/2015/sykdomsbyrde_i_norge_2015.pdf. Accessed 12 Apr 2018. [Google Scholar]
- 3.Lippeveld T, Sauerborn R, Bodart C, World Health Organization, editors. Design and implementation of health information systems. Geneva: World Health Organization; 2000. [Google Scholar]
- 4.Epstein AE, DiMarco JP, Ellenbogen KA, Estes NAM, Freedman RA, Gettes LS, et al. ACC/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the ACC/AHA/NASPE 2002 Guideline Update for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices) developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. J Am Coll Cardiol. 2008;51:e1–62. 10.1016/j.jacc.2008.02.032 [DOI] [PubMed] [Google Scholar]
- 5.Wilkoff BL, Auricchio A, Brugada J, Cowie M, Ellenbogen KA, Gillis AM, et al. HRS/EHRA expert consensus on the monitoring of cardiovascular implantable electronic devices (CIEDs): description of techniques, indications, personnel, frequency and ethical considerations. Heart Rhythm. 2008;5:907–25. [DOI] [PubMed] [Google Scholar]
- 6.Cronin EM, Varma N. Remote monitoring of cardiovascular implanted electronic devices: a paradigm shift for the 21st century. Expert Rev Med Devices. 2012;9:367–76. 10.1586/erd.12.18 [DOI] [PubMed] [Google Scholar]
- 7.Folino AF, Breda R, Calzavara P, Migliore F, Iliceto S, Buja G. In-home controls of pacemakers in debilitated elderly patients. Geriatr Gerontol Int. 2012;12:30–5. 10.1111/j.1447-0594.2011.00723.x [DOI] [PubMed] [Google Scholar]
- 8.Folino AF, Breda R, Calzavara P, Borghetti F, Comisso J, Iliceto S, et al. Remote follow-up of pacemakers in a selected population of debilitated elderly patients. Eur Eur Pacing Arrhythm Card Electrophysiol J Work Groups Card Pacing Arrhythm Card Cell Electrophysiol Eur Soc Cardiol. 2013;15:382–7. [DOI] [PubMed] [Google Scholar]
- 9.López-Villegas A, Catalán-Matamoros D, Robles-Musso E, Peiró S. Workload, time and costs of the informal cares in patients with tele-monitoring of pacemakers: the PONIENTE study. Clin Res Cardiol. 2016;105:307–13. 10.1007/s00392-015-0921-5 [DOI] [PubMed] [Google Scholar]
- 10.Sakshaug S, Selmer R-M, Graff-Iversen S. Cardiovascular disease in Norway. Oslo, Norway: Norwegian Institute for Public Health; 2016. https://www.fhi.no/en/op/hin/health—disease/cardiovascular-disease-in-norway—/#about-cardiovascular-disease. Accessed 12 Mar 2018. [Google Scholar]
- 11.Boriani G, Diemberger I, Martignani C, Biffi M, Valzania C, Bertini M, et al. Telecardiology and Remote Monitoring of Implanted Electrical Devices: The Potential for Fresh Clinical Care Perspectives. J Gen Intern Med. 2008;23:73–7. 10.1007/s11606-007-0355-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Iversen MM, Espehaug B, Hausken MF, Graue M, Østbye T, Skeie S, et al. Telemedicine Versus Standard Follow-Up Care for Diabetes-Related Foot Ulcers: Protocol for a Cluster Randomized Controlled Noninferiority Trial (DiaFOTo). JMIR Res Protoc. 2016;5:e148 10.2196/resprot.5646 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Daniel Catalán-Matamoros, Antonio López-Villegas. Telehealth and the current society: challenges and opportunities. Rev Esp Comun En Salud. 2016;7:336–45. [Google Scholar]
- 14.Lopez-Villegas A, Catalan-Matamoros D, Robles-Musso E, Bautista-Mesa R, Peiro S. Cost-utility analysis on telemonitoring of users with pacemakers: The PONIENTE study. J Telemed Telecare. 2018;:1357633X1876718. [DOI] [PubMed] [Google Scholar]
- 15.Lopez-Villegas A, Catalan-Matamoros D, Robles-Musso E, Peiro S. Effectiveness of pacemaker tele-monitoring on quality of life, functional capacity, event detection and workload: The PONIENTE trial: Effectiveness of pacemaker tele-monitoring. Geriatr Gerontol Int. 2016;16:1188–95. 10.1111/ggi.12612 [DOI] [PubMed] [Google Scholar]
- 16.López-Villegas A, Catalán-Matamoros D, Robles-Musso E, Peiró S. [Comparative Effectiveness of Remote Monitoring of People with Cardiac Pacemaker versus Conventional: quality of Life at the 6 Months]. Rev Esp Salud Publica. 2015;89:149–58. 10.4321/S1135-57272015000200004 [DOI] [PubMed] [Google Scholar]
- 17.Varma N, Ricci RP. Telemedicine and cardiac implants: what is the benefit? Eur Heart J. 2013;34:1885–95. 10.1093/eurheartj/ehs388 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Dreifus LS, Cohen D. Editorial: Implanted pacemakers: Medicolegal implications. Am J Cardiol. 1975;36:266–7. [DOI] [PubMed] [Google Scholar]
- 19.Schiffer AA, Denollet J, Pedersen SS, Broers H, Widdershoven JW. Health Status in Patients Treated with Cardiac Resynchronization Therapy: Modulating Effects of Personality: PERSONALITY AND CRT. Pacing Clin Electrophysiol. 2007;31:28–37. [DOI] [PubMed] [Google Scholar]
- 20.Andersen TO, Andersen PRD, Kornum AC, Larsen TM. Understanding patient experience: a deployment study in cardiac remote monitoring. ACM Press; 2017. p. 221–30. 10.1145/3154862.3154868 [DOI] [Google Scholar]
- 21.Marzegalli M, Lunati M, Landolina M, Perego GB, Ricci RP, Guenzati G, et al. Remote Monitoring of CRT-ICD: The Multicenter Italian CareLink Evaluation-Ease of Use, Acceptance, and Organizational Implications. Pacing Clin Electrophysiol. 2008;31:1259–64. 10.1111/j.1540-8159.2008.01175.x [DOI] [PubMed] [Google Scholar]
- 22.Malm D, Hallberg LR-M. Patients’ experiences of daily living with a pacemaker: a grounded theory study. J Health Psychol. 2006;11:787–98. 10.1177/1359105306066642 [DOI] [PubMed] [Google Scholar]
- 23.Tagney J. A literature review comparing the experiences and emergent needs of adult patients with permanent pacemakers (PPMs) and implantable cardioverter defibrillators (ICDs): Patients with ICDs and PPMs: a literature review. J Clin Nurs. 2010;:no-no. [DOI] [PubMed] [Google Scholar]
- 24.Duru F, Büchi S, Klaghofer R, Mattmann H, Sensky T, Buddeberg C, et al. How different from pacemaker patients are recipients of implantable cardioverter-defibrillators with respect to psychosocial adaptation, affective disorders, and quality of life? Heart Br Card Soc. 2001;85:375–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Leosdottir M, Sigurdsson E, Reimarsdottir G, Gottskalksson G, Torfason B, Vigfusdottir M, et al. Health-related quality of life of patients with implantable cardioverter defibrillators compared with that of pacemaker recipients. Eur Eur Pacing Arrhythm Card Electrophysiol J Work Groups Card Pacing Arrhythm Card Cell Electrophysiol Eur Soc Cardiol. 2006;8:168–74. [DOI] [PubMed] [Google Scholar]
- 26.Newall EG, Lever NA, Prasad S, Hornabrook C, Larsen PD. Psychological implications of ICD implantation in a New Zealand population. Eur Eur Pacing Arrhythm Card Electrophysiol J Work Groups Card Pacing Arrhythm Card Cell Electrophysiol Eur Soc Cardiol. 2007;9:20–4. [DOI] [PubMed] [Google Scholar]
- 27.Lopez-Villegas A, Catalan-Matamoros D, Lopez-Liria R, Enebakk T, Thunhaug H, Lappegård KT. Health-related quality of life on tele-monitoring for users with pacemakers 6 months after implant: the NORDLAND study, a randomized trial. BMC Geriatr. 2018;18 10.1186/s12877-018-0911-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Sjetne IS, Bjertnaes OA, Olsen RV, Iversen HH, Bukholm G. The Generic Short Patient Experiences Questionnaire (GS-PEQ): identification of core items from a survey in Norway. BMC Health Serv Res. 2011;11 10.1186/1472-6963-11-88 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Abrams DJ, Geier MR. A comparison of patient satisfaction with telehealth and on-site consultations: a pilot study for prenatal genetic counseling. J Genet Couns. 2006;15:199–205. 10.1007/s10897-006-9020-0 [DOI] [PubMed] [Google Scholar]
- 30.Bas Villaobos M. Propuesta de un sistema de monitorización remota de dispositivos implantados en pacientes cardiológicos: Consideraciones económicas, organizativas y de calidad percibida. Tesis Doctoral. Universidad Politécnica de Madrid; 2010. http://oa.upm.es/5681/. Accessed 12 Mar 2018. [Google Scholar]
- 31.Cottrell E, McMillan K, Chambers R. A cross-sectional survey and service evaluation of simple telehealth in primary care: what do patients think? BMJ Open. 2012;2:e001392 10.1136/bmjopen-2012-001392 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Berg M. Patient care information systems and health care work: a sociotechnical approach. Int J Med Inf. 1999;55:87–101. [DOI] [PubMed] [Google Scholar]
- 33.Brebner JA, Brebner EM, Ruddick-Bracken H. Experience-based guidelines for the implementation of telemedicine services. J Telemed Telecare. 2005;11 Suppl 1:3–5. 10.1177/1357633X0501100102 [DOI] [PubMed] [Google Scholar]
- 34.Menkes DB, Davison MP, Costello SA, Jaye C. Stereotactic radiosurgery: the patient’s experience. Soc Sci Med. 2005;60:2561–73. 10.1016/j.socscimed.2004.11.021 [DOI] [PubMed] [Google Scholar]
- 35.Manary MP, Boulding W, Staelin R, Glickman SW. The patient experience and health outcomes. N Engl J Med. 2013;368:201–3. 10.1056/NEJMp1211775 [DOI] [PubMed] [Google Scholar]
- 36.Catalan-Matamoros D. El gran avance del ámbito científico y académico de la comunicación en salud = The great development of the scientific and academic field of health communicationn salud. Rev Esp Comun En Salud. 2017;8:114–7. [Google Scholar]
- 37.Catalán Matamoros DJ, Muñoz-Cruzado y Barba M, Fuentes Hervías MT. Técnicas de comunicación para la prevención y el control de enfermedades (Communication techniques for disease prevention and control). Rev Esp Comun En Salud. 2010;1:50–65. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper and its Supporting Information files.