. 2018 Mar 27;26(6):1869–1874. doi: 10.1007/s12350-018-1261-4

Saline-push improves rubidium-82 PET image quality

Jennifer M Renaud 1, Kai Yi Wu 1, Kimberly Gardner 1, May Aung 1, Rob S B Beanlands 1, Robert A deKemp 1,
PMCID: PMC6908549  PMID: 29589206

Abstract

Introduction

Rubidium-82 (82Rb) PET is used widely for myocardial perfusion imaging. The purpose of this study was to investigate if an additional saline-push following the 82Rb elution improves PET image quality.

Methods

82Rb PET scans were acquired with and without 26 mL saline-push in six patients as part of a clinical quality improvement program. Dynamic images were analyzed to measure the total activity delivered to the superior vena cava (SVC) and retained in the left ventricle (LV) myocardium. Tracer uptake images were used to measure blood background coefficient-of-variation (COV), myocardium-to-blood signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) to assess image quality.

Results

Similar eluted activity was measured with/without the saline-push (830 vs 795 MBq; P = 0.24). The activity delivered to the heart and retained in the myocardium was consistently increased more than twofold (SVC: + 114% and LV: + 104%; P < 0.001) with the saline-push. Image quality was improved in all patients, with lower background noise (COV: − 19%), higher SNR (+ 24%) and CNR (+ 27%) (all P ≤ 0.01).

Conclusions

The saline-push used to flush 82Rb activity out of the infuser tubing, patient injection and intravenous access lines consistently increased the activity delivered to the heart by twofold. This technique is recommended to maximize image quality with 82Rb PET.

Electronic supplementary material

The online version of this article (10.1007/s12350-018-1261-4) contains supplementary material, which is available to authorized users.

Keywords: Rubidium-82, myocardial perfusion imaging, continuous quality improvement, positron emission tomography

Introduction

Rubidium-82 (82Rb) is the most commonly used PET tracer for stress myocardial perfusion imaging (MPI) and is increasingly used for quantification of myocardial blood flow and stress/rest flow reserve.1 The ultra-short half-life of 76 seconds requires careful timing of the tracer infusion with the start of PET imaging for optimal and reproducible image quality. However, this can be challenging in practice as the parent Strontium-82 (82Sr) activity in the 82Sr/Rb generator decays over several weeks, causing the volume of saline eluate and the corresponding time for elution to increase for the same prescribed activity. This results in additional decay of the 82Rb activity during the tracer elution, which can reduce image quality particularly near the end of the generator shelf-life. Recently, a new 82Rb elution system has become available which allows consistent administration of 82Rb activity regardless of the generator age.24 This system normally includes an automated post-elution flush (saline-push) of the infuser tubing and patient injection line (6 mL) to help deliver all the eluted activity to the patient intravenous (IV) site, as well as an additional (+ 20 mL) saline volume to help decrease the transit-time from the IV access site to the superior vena cava and right atrium of the heart (Figure 1). The purpose of this study was to evaluate the impact of this added saline-push on 82Rb PET image quality.

Figure 1.

Figure 1

Three-way valves are used to ‘bypass’ the 82Rb generator, allowing saline flush (blue) of the infuser tubing from the generator to the patient infusion and intravenous (IV) access lines (orange), and to push the eluted 82Rb activity through the peripheral and central veins to the heart

Methods

As part of the routine clinical quality improvement program at our institution, six scans (4 rest + 2 stress) were identified during a 4-month period (Sep to Dec 2017) with premature termination of the rubidium injection, before the start of the normal saline-push.4 In these scans (−push), the flow of saline was stopped (aborted) in error, immediately after the prescribed total activity was eluted from the 82Rb generator, and without administration of the standard saline-push volume. As such, the infuser tubing from the generator to the patient infusion line and IV access site were not flushed normally, and tracer transit to the heart was reliant only on the venous return blood flow. Due to clinical concern about the potential impact on image quality with these aborted injections, a repeat injection and scan (+push) was performed immediately following the first scan; the same activity was prescribed and then delivered normally including the standard saline-push volume (26 mL) to flush the infuser tubing, patient infusion, and IV lines as well as the cephalic and saphenous veins. The imaging protocol is illustrated in Figure S1 (online supplement). Patients provided written informed consent as part of the REST-PET trial (NCT02117284).

List-mode acquisitions were started at the time of tracer injection (30 s ‘square-wave’ at constant-activity-rate of 0.3 MBq/kg/s = 9 MBq/kg) at rest or following dipyridamole stress (140 μg/min/kg), according to current guidelines and local practice.5,6 Dynamic PET images (9 × 10 s, 3 × 30 s, 1 × 60 s = 4 min total) were acquired in rapid succession (∆t = 6 ± 1 min) first without and then with the saline-push on a Discovery 690 PET-VCT scanner (GEHC, Waukesha, WI). Dynamic images were reconstructed using the vendor iterative method (VuePoint HD) with 8-mm post-reconstruction Hann filter. Total counts (prompt − random coincidences) were recorded during the tracer uptake phase (2-4 min) and used to reconstruct static (ungated) images, then smoothed with 12-mm Hann post-filter. The aborted elution (− push) scans were shortened compared to our normal practice (4 vs 8 min) to minimize inter-scan delay, particularly during stress to ensure that the repeated clinical scan was acquired near peak hyperemia. The repeated elution (+ push) scans were analyzed using the same 4-min data to avoid any bias in image quality that would arise from a longer scan.

As illustrated in Figure S2A (online supplement), arterial blood time-activity curves (TAC) were measured using a volume-of-interest (VOI) placed in the superior vena cava (SVC), to compare the delivered 82Rb activity profiles with and without the saline-push. The SVC curve was also integrated over time, to assess the total activity (MBq·min) delivered to the heart over the course of the 4-minute scan. TAC values in the left ventricle (LV) myocardium were measured using FlowQuant® automated processing as previously described.6 Background noise was measured as the coefficient-of-variation (COV = SD/mean) of activity values in a spherical VOI placed in the left atrium (LA) blood pool as shown in Figure S2B (online supplement). The myocardium (LV) and blood background (COV) values measured during the uptake phase were used to calculate 82Rb image signal-to-noise ratios (SNR) and contrast-to-noise ratios (CNR) for each scan according to Equations 1 and 2 below.

SNR=LV÷COV 1
CNR=(LV-LA)÷COV. 2

Results

Patient demographics and 82Rb activity elution data are summarized in Table 1. The mean patient age (65 years) and weight (94 kg) are typical of those referred for 82Rb PET MPI at our institution. The total injected saline volume was 29 ± 3 mL higher for elutions with vs without the saline-push, although the activity eluted from the generator was not significantly different between the repeat injections (830 vs 795 MBq; P = 0.24). The differences in volume injected with vs without the saline-push were slightly higher than the 26 mL expected in most patients, likely due to incomplete recharging of the generator activity in the 5- to 6-min time interval between elutions.

Table 1.

Patient and rubidium elution data

Num Sex Scan type Patient age (years) Patient weight (kg) LVEF (%) Activity delivered (− push) Activity delivered (+ push) Volume delivered (− push) Volume delivered (+ push) Time interval (∆min)
1 M Rest 59 138 25 980 1150 20 55 6
2 F Rest 75 104 66 937 942 11 39 6
3 M Stress 75 70 29 633 648 11 39 5
4 M Rest 65 73 48 652 663 11 39 6
5 M Stress 54 80 60 709 718 12 39 5
6 M Rest 64 95 65 856 860 10 36 9
Mean ± Std Dev 65 ± 8 94 ± 25 49 ± 18 795 ± 150 830 ± 195 13 ± 4 41 ± 7 6 ± 1

The total coincidence counts recorded during the uptake phase of the PET scans were twofold higher using the saline-push (31.4 ± 5.5 vs 15.6 ± 2.1 million counts; P < 0.001). Despite no measured difference in the eluted activity (MBq) from the generator, the integral activity delivered to the heart (SVC) increased more than 2-fold using the saline-push (P < 0.001), as shown in Figure 2. The average 82Rb activity retained by the heart (LV myocardium) was also increased by 2-fold (P < 0.001), as shown individually for all 6 patients in Figure S3 (online supplement).

Figure 2.

Figure 2

Effects on 82Rb image quality with vs without saline-push. Total 82Rb activity (MBq) eluted from the generator was unchanged, but the activity delivered to the heart (SVC) and retained in the LV myocardium increased by more than 2-fold. LV-to-blood SNR and CNR increased by 24% and 27%, with concordant decrease in the blood background COV by 19% (all P ≤ 0.01). Error bars are ± 95% confidence intervals of the mean

The average image SNR and CNR were both improved in the LV myocardium by approximately 25% using the saline-push (P = 0.002 and 0.01, respectively) as shown in Figure 2, and for all patients individually in Figure S4 (online supplement). The improvement in image quality was also associated with a similar change in the LA blood background noise (COV) which decreased by approximately 20% (P = 0.01). The effects of the saline-push on 82Rb image quality are shown qualitatively for one patient in Figure 3. This patient had the lowest left ventricle ejection fraction (LVEF) as shown in Table 1. Analysis of the changes in CNR, SNR, and background noise COV with LVEF (Figure S5) demonstrated a significant effect of LVEF on improved SNR (P = 0.034) and a trend for reduced COV (P = 0.078), suggesting that the benefits of the saline-push may be greater in patients with lower LVEF.

Figure 3.

Figure 3

82Rb PET perfusion images at rest in patient #1 (59-year-old male, 138 kg, BMI = 48, LVEF = 25%) obtained with (A) and without (B) the additional saline-push following elution of the prescribed activity from the generator. Image quality and myocardium-to-blood contrast are improved markedly using the saline-push

Discussion

As one of the two FDA-approved myocardial perfusion tracers, 82Rb PET is a robust tool for diagnosis, risk stratification, and management of patients with suspected or known coronary artery disease (CAD).7 The useful shelf-life of the 82Rb generator is typically determined by the minimum activity available for patient injection, which decreases as the parent 82Sr isotope decays over time (25.5-day half-life). PET image quality (SNR) is directly related to the number of recorded counts during the uptake phase.8 In this small study, we have demonstrated the consistent and highly significant increase in recorded counts and image quality obtained using ~ 25 mL saline-push without increasing the 82Rb activity eluted from the generator. This volume of saline is similar to that used routinely to optimize image quality during contrast administration for x-ray CT coronary angiography.9 The 82Rb activity delivered to the heart and retained in the myocardium was consistently increased more than 2-fold, which provides tremendous opportunity to increase image quality or to extend the useful shelf-life of the 82Rb generator. This additional activity is a result of 2 separate processes: i. flushing the eluate in the tubing volume between the generator and the patient IV access site, and ii. decreasing the 82Rb isotope decay-time during venous transit from the IV access site to the heart.

The measured increase in SNR of ~ 25% is consistent with the ~ 20% decrease in image noise (i.e., 1.25 = 1/0.8) but is somewhat less than the theoretical improvement (√2) expected from a 2-fold increase in counts.8 This is likely due to the use of statistical iterative reconstruction and image smoothing, which also have independent effects to improve image quality, compared to filtered-backprojection without the use of an apodizing window of the ramp-filter. The improvements in image quality were measured using a relatively short (4 min) scan time compared to the typical 8-min dynamic scan duration used clinically at our institution. The 4-min period used for analysis was the longest common time available to compare the properties of the aborted and repeated scans. While the exact numerical results may differ slightly, the benefits of the saline-push should translate directly to the longer scan times, since 70% of the counts are acquired in the first 2 min of the uptake phase, i.e., 2-4 min after the scan starts.

In the present study, an integrated ‘bypass line’ was used to administer the saline-push automatically following elution of the 82Rb generator.4 However, similar results may be expected with other generator systems using a manual flush, which could be administered by inserting a 3-way valve and flush-syringe at the junction between the generator tubing and the patient infusion line. If the saline-push is administered manually, adequate shielding should be employed to minimize radiation exposure to the imaging technologist. The patient benefits from improvement in image quality should be balanced against the risk of increased occupational exposure.

While the current data were acquired using the ‘constant-activity-rate’ mode of infusion,4 the effects of the saline-push should apply generally to other infusion modes, e.g., constant-flow-rate used with this or other rubidium generator systems and may help overcome the practical difficulties of weight-based dosing reported for 82Rb PET.10 The saline flow-rate is not changed during any of the patient infusions, even in the constant-activity-rate mode, therefore the time-delay to flush activity from the generator tubing and patient infusion lines will be similar for a given flow-rate regardless of the mode of tracer infusion. For other PET perfusion tracers such as 13N-ammonia or 15O-water, a dual-syringe pump could be used to automatically deliver the saline-push following tracer administration, again similar to methods used routinely for coronary CT angiography.9

Conclusion

An additional 25-30 mL saline-push is recommended following elution of the prescribed activity from a 82Rb generator to flush the infuser tubing, patient infusion, and intravenous access lines, and to increase venous return to the heart. This simple technique increases myocardial uptake by a factor of 2, and image SNR and CNR by 25%.

New knowledge gained

82Rb PET myocardial perfusion image quality is significantly improved by flushing all of the eluted activity out of the infuser tubing and patient IV lines, followed by an additional 20 mL saline-push to accelerate the tracer transit to the heart. The benefits to improve image quality may be highest in patients with low cardiac output, where delayed clearance from the blood pool is often observed.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Acknowledgments

Disclosure

This study was supported by a grant from the Ontario Research Fund (ORF-RE7-021) and an unrestricted research grant from Jubilant DraxImage for the REST-PET trial. JMR is a consultant for Jubilant DraxImage and receives royalties from the sale of FlowQuant software. RSB has received research grants and honoraria from Lantheus Medical Imaging, GE Healthcare and Jubilant DraxImage. RdK receives royalty revenues from rubidium-82 generator technologies, FlowQuant software licenses, and is a consultant for Jubilant DraxImage. KYW, KG, MA have no conflicts of interest related to this work.

Abbreviations

CNR

Contrast-to-noise ratio

COV

Coefficient-of-variation

LA

Left atrium

LV

Left ventricle

LVEF

Left ventricle ejection fraction

MBq

Megabecquerel

MPI

Myocardial perfusion imaging

PET

Positron emission tomography

82Rb

Rubidium-82

SNR

Signal-to-noise ratio

SVC

Superior vena cava

Footnotes

The authors of this article have provided a PowerPoint file, available for download at SpringerLink, which summarises the contents of the paper and is free for re-use at meetings and presentations. Search for the article DOI on SpringerLink.com.

References

  • 1.Murthy VL, Bateman TM, Beanlands RS, Berman DS, Borges-Neto S, Chareonthaitawee P, Cerqueira MD, deKemp RA, DePuey EG, Dilsizian V, Dorbala S, Ficaro EP, Garcia EV, Gewirtz H, Heller GV, Lewin HC, Malhotra S, Mann A, Ruddy TD, Schindler TH, Schwartz RG, Slomka PJ, Soman P, Di Carli MF, Einstein A, Russell R, Corbett JR. Clinical quantification of myocardial blood flow using PET: Joint Position Paper of the SNMMI Cardiovascular Council and the ASNC. J Nucl Cardiol. 2017;25:269–297. doi: 10.1007/s12350-017-1110-x. [DOI] [PubMed] [Google Scholar]
  • 2.Klein R, Adler A, Beanlands RS, deKemp RA. Precision-controlled elution of a 82Sr/82Rb generator for cardiac perfusion imaging with positron emission tomography. Phys Med Biol. 2007;52(3):659–673. doi: 10.1088/0031-9155/52/3/009. [DOI] [PubMed] [Google Scholar]
  • 3.Klein R, Ocneanu A, Renaud JM, Ziadi MC, Beanlands RS, deKemp RA. Consistent tracer administration profile improves test-retest repeatability of myocardial blood flow quantification with 82Rb dynamic PET imaging. J Nucl Cardiol [DOI] [PMC free article] [PubMed]
  • 4.RUBY-FILL rubidium chloride rb-82 injection, Jubilant DraxImage Inc. product insert. 2016. http://www.draximage.com/wp-content/uploads/2017/01/RUBY-FILL_Insert-USA.pdf. Accessed 21 Dec 2017.
  • 5.Dilsizian V, Bacharach SL, Beanlands RS, Bergmann SR, Delbeke D, Dorbala S, Gropler RJ, Knuuti J, Schelbert HR, Travin MI. ASNC imaging guidelines/SNMMI procedure standard for positron emission tomography (PET) nuclear cardiology procedures. https://www.asnc.org/Files/Guidelines and Quality/PET GuidelineASNC SNMMI2016.pdf. Accessed 21 Dec 2017.
  • 6.Klein R, Renaud JM, Ziadi MC, Thorn SL, Adler A, Beanlands RS, deKemp RA. Intra- and inter-operator repeatability of myocardial blood flow and myocardial flow reserve measurements using rubidium-82 pet and a highly automated analysis program. J Nucl Cardiol. 2010;17(4):600–616. doi: 10.1007/s12350-010-9225-3. [DOI] [PubMed] [Google Scholar]
  • 7.Ziadi MC. Myocardial flow reserve (MFR) with positron emission tomography (PET)/computed tomography (CT): clinical impact in diagnosis and prognosis. Cardiovasc Diagn Ther. 2017;7(2):206–218. doi: 10.21037/cdt.2017.04.10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Strother SC, Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating countrates to image signal-to-noise ratios using noise equivalents counts. IEEE Trans Nucl Sci. 1990;37(2):783–788. doi: 10.1109/23.106715. [DOI] [Google Scholar]
  • 9.Bae KT. Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology. 2010;256(1):32–61. doi: 10.1148/radiol.10090908. [DOI] [PubMed] [Google Scholar]
  • 10.Moody JB, Hiller KM, Lee BC, Corbett JR, Ficaro EP, Murthy VL. Limitations of Rb-82 weight-adjusted dosing accuracy at low doses. J Nucl Cardiol. 2017;24(4):1395–1401. doi: 10.1007/s12350-016-0531-2. [DOI] [PubMed] [Google Scholar]