Skip to main content
HHS Author Manuscripts logoLink to HHS Author Manuscripts
. Author manuscript; available in PMC: 2019 Dec 29.
Published in final edited form as: IEEE Trans Ultrason Ferroelectr Freq Control. 2018 Oct 1;65(12):2343–2348. doi: 10.1109/TUFFC.2018.2872840

Directivity and Frequency-Dependent Effective Sensitive Element Size of a Reflectance-Based Fiber Optic Hydrophone: Predictions from Theoretical Models Compared with Measurements

Keith A Wear 1, Samuel M Howard 2
PMCID: PMC6935507  NIHMSID: NIHMS1063893  PMID: 30281445

Abstract

The goal of this work was to measure directivity of a reflectance-based fiber-optic hydrophone at multiple frequencies and to compare it to four theoretical models: Rigid Baffle (RB), Rigid Piston (RP), Unbaffled (UB), and Soft Baffle (SB). The fiber had a nominal 105 μm diameter core and a 125 μm overall diameter (core + cladding). Directivity measurements were performed at 2.25, 3.5, 5, 7.5, 10, and 15 MHz from ±90° in two orthogonal planes. Effective hydrophone sensitive element radius was estimated by least-squares fitting the four models to directivity measurements using the sensitive element radius as an adjustable parameter. Over the range from 2.25 to 15 MHz, the average magnitudes of differences between the effective and nominal sensitive element radii were 59 ± 49% (RB), 10 ± 5% (RP), 46 ± 38% (UB), and 71 ± 19% (SB). Therefore, directivity of a reflectance-based fiber-optic hydrophone may be best estimated by the RP model.

Keywords: fiber optic hydrophone, directivity

I. Introduction

Quantitative measurement of acoustic pressure waveforms is essential to ensure safety and effectiveness for medical ultrasound devices. Pressure wavefonns are usually measured using hydrophones. For diagnostic medical ultrasound applications, there is a variety of choices of hydrophones, including membrane, needle, capsule, and fiber-optic designs. However, for therapeutic applications, including high intensity therapeutic ultrasound (HITU), the choices are much more limited because the output pressure from HITU devices is often too great to measure with most coimnon hydrophones without causing damage or cavitation-induced measurement unreliability.

Reflectance-based fiber optic hydrophones can withstand very high pressures (e.g., up to 70 MPa and beyond [14]) and have been used for HITU [18] and lithotripsy [9] applications. The most common design of a reflectance-based fiber optic hydrophone propagates laser light along a fiber and measures the signal reflected from the tip immersed in a fluid. Changes in pressure due to an acoustic wave incident on the tip cause changes in index of refraction in the fluid, which result in changes in the reflection coefficient at the fiber tip[1, 1013].

Fiber optic hydrophones have exceptional spatial resolution, often approximately 100 μm or less, making them appropriate to capture frequencies up to 100 MHz [1, 1013]. The use of fiber tips tapered out to diameters down to 7 μm has been proposed to enhance spatial resolution [12, 13]. A membrane hydrophone with a steel foil front protection layer has been demonstrated to withstand peak compressional pressures up to 75 MPa and has a nominal geometrical sensitive element diameter of approximately 200 microns [14, 15]. A robust needle hydrophone has been developed for HITU applications and has a nominal geometrical sensitive element diameter of approximately 400 microns [16]. Fiber optic and needle hydrophones have smaller cross sections than membrane hydrophones, which reduces the magnitude of undesired reflections in a water tank due to planar interfaces. Fiber optic hydrophones that operate on the basis of Fabry-Perot interferometry[17, 18] have exceptional spatial resolution (10 μm) but have much lower maximum pressure tolerances than reflectance-based fiber-optic hydrophones[19].

Reflectance-based fiber optic hydrophones have drawbacks. First, they have relatively high minimum detectable (e.g. noise equivalent) pressure amplitudes, on the order of 0.5 MPa, making them unsuitable for low pressure measurements. While this is not a severe limitation for measuring maximum pressures on the order of tens of MPa from HITU systems, it can be a limitation for mapping the entire HITU field, including side lobes. Second, flexing of fiber optic probes due to radiation forces at clinical HITU levels can cause signal and positioning distortions.

Complete hydrophone characterization involves specification of complex sensitivity, directivity, and frequency-dependent effective sensitive element size. Sensitivity corresponds to voltage output for normally-incident, quasi-planar pressure waves as a function of frequency. Frequency-dependent sensitivity functions are useful for deconvolving the hydrophone impulse response from the hydrophone voltage output, resulting in improved accuracy in pressure measurements[18, 2027]. For example, in one study of eight hydrophones (4 membrane, 2 needle, 1 capsule, and one fiber optic) measuring the same reference pressure signal, deconvolution reduced the coefficient of variation (ratio of standard deviation to mean across all 8 hydrophones) from 29% to 8% (peak compressional pressure), 39% to 13% (peak rarefactional pressure), and 58% to 10% (pulse intensity integral) [23].

While sensitivity of reflectance-based fiber-optic hydrophones has been considered in previous publications[1, 11, 28, 29], directivity has received less attention. Directivity corresponds to voltage output for quasi-planar pressure waves as a function of angle of incidence. Directivity is important because 1) its measurement can be used to determine the effective size of the sensitive element, 2) it describes how critical the angular orientation of the hydrophone must be to faithfully measure an incoming pressure wave, 3) it describes the potential for distortion of highly-focused beams with wide angular spectra, and 4) it describes how well the hydrophone can reject undesired off-axis waves (e.g., reflections in a water tank).

Hydrophone directivity is often modeled by integrating the complex free-field pressure incident on a surface corresponding to the hydrophone sensitive element. This leads to the rigid baffle (RB) model for hydrophone directivity [30] (p. 381),

DRB(k,θ)=2J1(kasinθ)kasinθ

where θ = angle of incidence, k = 2π / λ, λ = wavelength, and a = the radius of the hydrophone sensitive element [31]. This model is commonly used to infer a value for “effective” sensitive element radius, aeff, from measurements of directivity (IEC). Effective radius aeff may be distinguished from nominal or geometrical radius ag. One limitation of the RB model is that it does not necessarily fully account for the effect that the presence of the hydrophone can have on the field that it is used to measure. Other models for hydrophone directivity have been proposed, including the unbaffled (UB) model [32]

DUB(k,θ)=DRB(k,θ)(1+cosθ2),

and the soft baffle (SB) model [33],

DSB(k,θ)=DRB(k,θ)cosθ.

The RB, UB, and SB approaches have previously been used to model membrane and needle hydrophones[31].

Krucker el al. modeled a reflectance-based fiber optic hydrophone as a rigid piston (RP) [34]. The RP model imposes a boundary condition that the component of acoustic velocity perpendicular to the sensitive element must be zero [34]. Although the RP model was primarily motivated by reflectance-based fiber optic hydrophones, Krucker et al. suggested that the RP model could also be appropriate for needle hydrophones. Their theoretical directivity (Eq. 19) is in the form of an integral that must be evaluated numerically. For the sake of brevity, the reader is referred to their paper [34]. Krucker et al. did not provide experimental validation for the RP model for sensitivity or directivity of fiber-optic or needle hydrophones. However, Wear et al. subsequently demonstrated good agreement between the RP model and experimental measurements of sensitivity of needle and reflectance-based fiber optic hydrophones (12% ± 3% root-mean-square difference in normalized sensitivity [29]) and directivity of needle hydrophones (7% ± 5% accuracy in measurement of geometrical sensitive element radius [35]).

The goal of the present paper is to report measurements of directivity and frequency-dependent effective sensitive element size of a reflectance-based fiber optic hydrophone at multiple frequencies and compare them to predictions based on RB, RP, UB, and SB models.

II. Methods

An Onda (Sunnyvale, CA) HFO-690 fiber optic hydrophone was used for directivity measurements. The multi-mode fiber, operated at a wavelength of 690 nm, has a nominal 105 μm diameter core and a 125 μm overall diameter (core + cladding). The fiber-tip was cleaved prior to the measurements according to the manufacturer’s instructions, in order to present a flat tip to an incoming pressure wave.

The hydrophone was inserted into a tank custom-designed for directvity measurements. The water temperature was 23.5 ± 0.5°, as measured with a calibrated mercury-in-glass thermometer (VWR Scientific Products Corporation, East Hills, NY, USA).

Directivity measurements were performed at 2.25, 3.5, 5 MHz, 7.5, 10 MHz, and 15 MHz, which meant that the wavelength ranged from approximately 1 to 6.5 times the nominal fiber core diameter. The acquisition method followed IEC 62127-2, Section 12 (IEC). The hydrophone was aligned to the axis of rotation by comparing the time of flight of the ultrasound signal at three or more angles of incidence and adjusting the hydrophone position to minimize any differences.

Source transducers were driven with tone-bursts from a Hewlett-Packard (Santa Clara, CA) 3314A function generator through an ENI (Rochester, NY) 3100LA power amplifier (nominal 55 dB gain). The function generator voltage of 600 mV was chosen to be comfortably within the power amplifiers input limit of 1V input. Table 1 lists transducer properties and experimental parameters. For measurements at 2.25 and 3.5 MHz, the sources were Olympus (Waltham, MA) nondestructive evaluation transducers. For measurements at 5 MHz and above, the source was an Onda BBS, which is a broadband, mildly-focused (F/10) transducer made with PVDF film [36]. When paired with the amplifier, the BBS has a center frequency of about 10 MHz and a −6 dB bandwidth of approximately 16 MHz. Waveforms were acquired through a Tektronix (Beaverton, OR) TDS 540 oscilloscope. Waveform acquisition was delayed to acquire five (Olympus) or ten (BBS) microseconds of signal, which had pulse duration of ten (Olympus) or fifteen (BBS) microseconds. To minimize noise and harmonic distortion, the oscilloscope’s 20 MHz bandwidth filter was used. Below 10 MHz, an additional Mini-Circuits (Brooklyn, NY) 10 MHz filter was used. 64 waveforms were averaged for data collected with the BBS (i.e., at 5 MHz and above). 256 wavefonns were averaged for data collected with the Olympus transducers (i.e. at 2.25 and 3.5 MHz).

TABLE I.

Source Transducers

Model Frequency (MHz) Diameter (cm) Focal Properties Distance between source and hydrophone (cm)
Center Drive
Olympus A304-SU 2.25 2.25 2.54 Planar 12.7
Olympus V381 3.5 3.5 1.91 Planar 12.7
Onda BBS 10 5 2.54 F/10 25.4
Onda BBS 10 7.5 2.54 F/10 25.4
Onda BBS 10 10 2.54 F/10 25.4
Onda BBS 10 15 2.54 F/10 25.4

Directivity measurements were performed in two orthogonal planes by rotating a fiber holder (which gripped the fiber) around its axis of symmetry. Two hundred measurements were performed in each plane for angles ranging from −90° to 90° with angular interval between measurements of 0.9°. Approximately 2 cm of bare fiber stuck out of the fiber holder. This distance provided sufficient fiber rigidity, while delaying the potential arrival of spurious reflections until after the signal was acquired. In all cases a duty factor of less than 1% was utilized in the measurements, as a precaution to avoid flexing of the fiber tip due to acoustic radiation pressure. Temporal average intensities were less than 10 mW/cm2, which corresponds to a radiation pressure of 0.06 Pa (based on the standard plane wave approximation of calculating radiation pressure by dividing intensity by speed of sound).

Alternative methods for directivity measurement include a pulsed near-field method [37], a harmonic-based approach [35, 38] a time-delay-spectrometry (TDS) based method [39], and a method based on using a photoacoustic source consisting of a blackened planar surface illuminated by a laser [40, 41].

In order to estimate uncertainty of measurements, the standard deviations of paired directivity measurements taken at the same angle from both orthogonal planes were computed for each frequency. These standard deviations were then averaged over all angles for each frequency. These uncertainty estimates included random uncertainties, temperature variations (±0.5°C), source stability (±2%), digitizer error (±2%), positional repeatability (this effect varies with frequency from ±2.5% at 1 MHz to 5% at 20 MHz), and potential imperfect symmetry of the fiber optic hydrophone.

“Effective” sensitive element radii, aeff, were estimated for all four models at each of the six frequencies. The hydrophone sensitive element radius a was used as a free parameter in functional fits of theoretical directivity models to experimental data. For each model and frequency, the value of a that minimized the mean square difference between the theoretical form and experimental data was designated as aeff. Effective radius aeff may be distinguished from nominal or geometrical radius ag. Note that while IEC 62127-3 defines effective radius by comparing measured directivity data to the RB model (IEC 62127-3), the present paper defines multiple effective radii obtained by fitting measured directivity data to multiple models. IEC 62127-3 recommends computation of aeff from directivity values at −3 dB and –6 dB. However, because of the small sensitive element size of the fiber-optic hydrophone, and the corresponding wide directivity patterns, directivity measurements did not always fall to −3 dB or −6 dB over the range of angles investigated (−90° to 90°), especially at low frequencies. Therefore, the fitting method was used instead. A similar method was described by Wilkens and Molkenstruck for estimating aeff from directivity measurements for a membrane hydrophone [42].

III. Results

Figure 1 shows directivity measurements for six frequencies as functions of angle of incidence θ. As frequency increases, the directivity functions get narrower. This is expected because wavelength is inversely related to frequency, and at a given angle of incidence, decreasing wavelength is associated with increasing phase cancellation across the hydrophone sensitive element. Figure 1 shows that the lower frequency measurements (2.25 and 3.5 MHz) appear to be noisier than the higher frequency measurements. This is likely due to lower acoustic pressures produced by the lower frequency transducers, which were unfocussed, and also the −6 dB loss in hydrophone response at lower frequencies [11, 29]. (RP hydrophone sensitivity roughly doubles between zero frequency and fmax = 1.2c/(πag) where c = speed of sound [29]. For the present reflectance-based fiber optic hydrophone, fmax = 11 MHz. See Fig. 2 in [11]. Similar behavior has been observed for needle hydrophone sensitivity [29, 43, 44].)

Fig. 1.

Fig. 1.

Directivity measurements at 6 frequencies.

Fig. 2.

Fig. 2.

Directivity measurements at 6 frequencies compared with 4 theoretical models: Rigid Baffle (RB), Rigid Piston (RP), Unbaffled (UB), and Soft Baffle (SB). Model functions are based on nominal or geometrical sensitive element size, ag, for the fiber optic hydrophone.

The standard deviations of directivity measurements, averaged over all angles were 0.03 (2.25 MHz), 0.03 (3.5 MHz), 0.02 (5 MHz), 0.01 (7.5 MHz), 0.01 (10 MHz), and 0.02 (15 MHz). These were small (≤ 3%) compared with the maximum value of directivity, which is 1 by definition.

Figure 2 shows theoretical and experimental directivity plots at all six frequencies. The theoretical models were computed based on the nominal (geometrical) sensitive element radius ag (52.5 μm). The RP model appears to conform to the data more consistently across all six frequencies than the other three models. For frequencies 5 MHz and above (kag > 1.1 where k = 2π / λ and λ is the wavelength), the RP and UB models are very close to each other and appear to model the experimental directivity better than the common RB model until kag exceeds 3 and the three models become comparable.

At 15 MHz (kag = 3.3), all four models approach each other but noticeably underestimate the experimental directivity at |θ| > 45°. This discrepancy is likely due to acoustic pressure from the 15 MHz transducer falling to low values at the hydrophone sensitive element for |θ| > 45° and becoming comparable to the noise, thus making the directivity measurements less reliable. The fiber-optic hydrophone has a noise equivalent pressure ranging from 0.25 MPa at 2 MHz bandwidth to 0.5 MPa at 100 MHz bandwidth, defined by the manufacturer as the rms signal measured in the absence of sonication. The averaging described in Section II further improved the SNR, which nevertheless varied with frequency because of variations in the source and hydrophone responses. At normal incidences, it is estimated that SNR varied from 11 at 10 MHz to 4 at 2.25 MHz, 3.5 MHz and 15 MHz.

Root-mean-square differences (RMSDs) between models (based on ag) and measurements (averaged over frequencies from 2.25 to 10 MHz over the range from −90° to 90°) were 17% (RB), 6% (RP), 8% (UB), and 31% (SB). (The 15 MHz data were excluded because of the unreliable measurements for |θ| > 45°).

Figure 3 shows effective radius, aeff, plotted vs. kag for all 6 frequencies and all 4 models. The RP model conforms most consistently to the nominal (geometrical) effective radius, ag. The discrepancies between effective and nominal radii decrease with kag for all four models. Over the range from 2.25 to 15 MHz, the average magnitudes of differences between the effective and nominal sensitive element radii were 59 ± 49% (RB), 10 ± 5% (RP), 46 ± 38% (UB), and 71 ± 19% (SB). The decline of aeff with frequency is consistent with reported measurements for needle [35, 39, 45], fiber optic [45], and membrane [42] hydrophones. (Recall that k is directly proportional to frequency.)

Fig. 3.

Fig. 3.

Effective sensitive element radii based on best fit model functions with sensitive element size as an adjustable parameter. The nominal or geometric sensitive element radius, ag, is shown in the horizontal black dotted line. The average standard deviation, obtained by comparing estimates of aeff derived from directivity measurements in two orthogonal planes, is 2 μm.

Figure 4 shows the relative differences between effective and geometrical sensitive element radii for the four models. A fit to the RB model of the form C / kag is also shown. The value of C that yielded the minimum squared difference between model and data was 0.67 with a 95% confidence interval of (0.56, 0.79).

Fig. 4.

Fig. 4.

Relative difference between effective and geometrical sensitive element sizes as a function of kag. A fit to the RB model is also shown. The average standard deviation, obtained by comparing estimates of (aeffag) / ag derived from directivity measurements in two orthogonal planes, is 4%.

IV. Conclusion

The commonly-used rigid baffle (RB) model for characterizing hydrophone directivity was compared to three alternative models—soft baffle (SB), unbaffled (UB), and rigid piston (RP)—for a reflectance-based fiber-optic hydrophone over the range from 2.25 to 15 MHz (0.5 < kag < 3.3). The discrepancies among the four models are considerable for low kag (e.g., 0 < kag < 1) and decrease as kag increases. The RP model proved superior to the other three models in terms of similarity of nominal or geometric sensitive element radius, ag, and effective sensitive element radius, aeff, derived by fitting model directivity functions to measured directivity data. The UB model is much easier to compute than the RP model (because it does not require numerical analysis of an integral) and is nearly identical to the RP model for ka > 1.3. The superior performance of the RP model, which is similar to previous findings with needle hydrophones, may be attributable to its more realistic representation of the boundary condition on the surface of the needle hydrophone sensitive element.

Acknowledgment

The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. The authors are grateful for funding support from the FDA Office of Women’s Health.

This work was supported in part by a grant from the FDA Office of Women’s Health.

Biography

graphic file with name nihms-1063893-b0005.gif

Keith A. Wear received his B.A. in Applied Physics from the University of California at San Diego. He received his M.S. and Ph.D. in Applied Physics with a Ph.D. minor in Electrical Engineering from Stanford University. He was a post-doctoral research fellow with the Physics department at Washington University, St. Louis. He is the FDA Acoustics Laboratory Leader. He is an Associate Editor of 3 journals: Journal of the Acoustical Society of America, IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control; and Ultrasonic Imaging. He was the Technical Program Chair of the 2008 IEEE International Ultrasonics Symposium in Beijing, China. He was the General Program Chair of the 2017 IEEE International Ultrasonics Symposium in Washington, DC. He was chair of the American Institute of Ultrasound in Medicine (AIUM) Technical Standards Committee from 2014-2016. He is a Fellow of the Acoustical Society of America, the American Institute for Medical and Biological Engineering, and the AIUM. He is a senior member of IEEE.

graphic file with name nihms-1063893-b0006.gif

Samuel M. Howard received his A.B. in Engineering Sciences from Harvard University. He received his M.S. and Ph.D in Theoretical and Applied Mechanics from Cornell University with minors in Applied Mathematics and Physics. From 1989 to 2001 he worked for Acuson (later Siemens) in transducer R&D. From 2001 to the present he has been at Onda Corporation, where he is the CTO and directs the Acoustics Laboratory. His current interests focus on ultrasound metrology for medical and industrial applications.

Contributor Information

Keith A. Wear, US Food and Drag Administration, Silver Spring, MD, 20852.

Samuel M. Howard, Onda Corporation, Sunnyvale, CA, 94089

REFERENCES

  • [1].Staudenraus J and Eisemnenger W, “Fiberoptic Probe Hydrophone for Ultrasonic and Shock-Wave Measurements in Water,” (in English), Ultrasonics, vol. 31, no. 4, pp. 267–273, July 1993. [Google Scholar]
  • [2].Canney MS, Bailey MR, Crum LA, Khokhlova VA, and Sapozhnikov OA, “Acoustic characterization of high intensity focused ultrasound fields: A combined measurement and modeling approach,” (in English), J. Acoust. Soc. Am, vol. 124, no. 4, pp. 2406–2420. October 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [3].Maxwell AD et al. , “Cavitation clouds created by shock scattering from bubbles during histotripsy,” (in English), J. Acoust. Soc. Am, vol. 130, no. 4, pp. 1888–1898, October 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Kreider W et al. , “Characterization of a Multi-Element Clinical HIFU System Using Acoustic Holography and Nonlinear Modeling,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 60, no. 8, pp. 1683–1698, August 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Parsons JE, Cain CA, and Fowlkes JB, “Cost-effective assembly of a basic fiber-optic hydrophone for measurement of high-amplitude therapeutic ultrasound fields,” (in English), J. Acoust. Soc. Am, vol. 119, no. 3, pp. 1432–1440, March 2006. [DOI] [PubMed] [Google Scholar]
  • [6].Haller J, Jenderka KV, Durando G, and Shaw A, “A comparative evaluation of three hydrophones and a numerical model in high intensity focused ultrasound fields,” (in English), J. Acoust. Soc. Am, vol. 131, no. 2, pp. 1121–1130, February 2012. [DOI] [PubMed] [Google Scholar]
  • [7].Jing Y, Cannata J, and Wang T, “Experimental verification of transient nonlinear acoustical holography,” J Acoust Soc Am. vol. 133 no. 5 pp. 2533–40. May 2013.” [DOI] [PubMed] [Google Scholar]
  • [8].Vlaisavljevich E et al. , “Effects of ultrasound frequency and tissue stiffness on the histotripsy intrinsic threshold for cavitation,” Ultrasound Med Biol, vol. 41, no. 6, pp. 1651–67, June 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Maxwell AD et al. , “Fragmentation of urinary calculi in vitro by burst wave lithotripsy,” J Urol, vol. 193, no. 1, pp. 338–44, January 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [10].Arvengas A, Davitt K, and Caupin F, “Fiber optic probe hydrophone for the study of acoustic cavitation in water,” (in English), Rev Sci Instrum, vol. 82, no. 3, March 2011. [DOI] [PubMed] [Google Scholar]
  • [11].Howard SM, “Calibration of Reflectance-Based Fiber-Optic Hydrophones,” (in English), 2016. IEEE Inti. Ultrasonics Symp,2016. [Google Scholar]
  • [12].Umchid S et al. , “Development of calibration techniques for ultrasonic hydrophone probes in the frequency range from 1 to 100 MHz,” (in English), Ultrasonics, vol. 49, no. 3, pp. 306–311, March 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Lewin PA, Mu C, Umchid S, Daryoush A, and El-Sherif M, “Acousto-optic, point receiver hydrophone probe for operation up to 100 MHz,” (in English), Ultrasonics, vol. 43, no. 10, pp. 815–821. December 2005. [DOI] [PubMed] [Google Scholar]
  • [14].Bessonova ΟV and Wilkens V, “Membrane Hydrophone Measurement and Numerical Simulation of HIFU Fields up to Developed Shock Regimes,” (in English), IEEE Trans. Ultrason. Ferroelectr., Freq. Control, vol. 60, no. 2, pp. 290–300, February 2013. [DOI] [PubMed] [Google Scholar]
  • [15].Wilkens V, Sonntag S, and Georg O, “Robust spot-poled membrane hydrophones for measurement of large amplitude pressure waveforms generated by high intensity therapeutic ultrasonic transducers,” (in English), J. Acoust. Soc. Am, vol. 139, no. 3 pp. 1319–1332. March 2016. [DOI] [PubMed] [Google Scholar]
  • [16].Zanelli CI and Howard SM, “A robust hydrophone for HIFU metrology,” (in English), Therapeutic Ultrasound, vol. 829, pp. 618–+ 2006. [Google Scholar]
  • [17].Morris P, Hurrell A, Shaw A, Zhang E, and Beard P, “A Fabry-Perot fiber-optic ultrasonic hydrophone for the simultaneous measurement of temperature and acoustic pressure,” (in English), J. Acoust. Soc. Am, vol. 125, no. 6, pp. 3611–3622, June 2009. [DOI] [PubMed] [Google Scholar]
  • [18].Wear KA. Liu YB. Gammell PM. Maruvada S. and Harris GR, “Correction for Frequency-Dependent Hydrophone Response to Nonlinear Pressure Waves Using Complex Deconvolution and Rarefactional Filtering: Application With Fiber Optic Hydrophones,” (in English), IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 62, no. 1, pp. 152–164, January 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Liu Y, Wear KA, and Harris GR, “Variation of High-Intensity Therapeutic Ultrasound (HITU) Pressure Field Characterization: Effects of Hydrophone Choice, Nonlinearity, Spatial Averaging and Complex Deconvolution,” Ultrasound Med Biol. vol. 43, no. 10 pp. 2329–2342. October 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Wilkens V and Koch C, “Amplitude and phase calibration of hydrophones up to 70 MHz using broadband pulse excitation and an optical reference hydrophone,” (in English), J. Acoust. Soc. Am, vol. 115, no. 6, pp. 2892–2903, June 2004. [Google Scholar]
  • [21].Hurrell A, “Voltage to Pressure Conversion: Are you getting “phased” by the problem,” (in English), AMUM 2004: Advanced Metrology for Ultrasound in Medicine 2004, vol. 1, pp. 57–62, 2004. [Google Scholar]
  • [22].Wear KA, Gammell PM, Maruvada S, Liu Y, and Harris GR, “Time-delay spectrometry measurement of magnitude and phase of hydrophone response,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 58, no. 11, pp. 2325–33, November 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [23].Wear KA, Gammell PM, Maruvada S, Liu Y, and Harris GR, “Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 61, no. 1, pp. 62–75, January 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Eichstadt S, Wilkens V, Dienstfrey A, Hale P, Hughes B, and Jarvis C, “On challenges in the uncertainty evaluation for time-dependent measurements,” (in English), Metrologia, vol. 53, no. 4, pp. S125–S135, August 2016. [Google Scholar]
  • [25].Eichstadt S and Wilkens V, “GUM2DFT-a software tool for uncertainty evaluation of transient signals in the frequency domain,” (in English), Meas. Sci. Tech, vol. 27, no. 5, May 2016. [Google Scholar]
  • [26].Eichstadt S and Wilkens V, “Evaluation of uncertainty for regularized deconvolution: A case study in hydrophone measurements,” (in English), J. Acoust. Soc. Am, vol. 141, no. 6, pp. 4155–4167, June 2017. [DOI] [PubMed] [Google Scholar]
  • [27].Hurrell AM and Rajagopal S, “The Practicalities of Obtaining and Using Hydrophone Calibration Data to Derive Pressure Waveforms,” (in English), IEEE Trans. Ultrason., Ferroelectr. Freq. Control, vol. 64, no. 1, pp. 126–140, January 2017. [DOI] [PubMed] [Google Scholar]
  • [28].Wang ZQ, Lauxmann P, Wurster C, Kohler M, Gompf B, and Eisenmenger W, “Impulse response of a fiber optic probe hydrophone determined with shock waves in water,” (in English), J. Appl. Phys, vol. 85, no. 5, pp. 2514–2516, March 1 1999. [Google Scholar]
  • [29].Wear KA, Liu Y, and Harris GR, “Pressure pulse distortion by needle and fiber-optic hydrophones due to nonuniform sensitivity,” IEEE Trans Ultrason Ferroelectr. Freq Control, vol. 65, no. 2, pp. 137–148, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30].Morse PM and Ingard KU, Theoretical Acoustics. Princeton, NJ: Princeton University Press, 1968. [Google Scholar]
  • [31].Shombert DG, Smith SW, and Harris GR, “Angular Response of Miniature Ultrasonic Hydrophones at Medical Diagnostic Frequencies,” (in English), Ultrasonic Imaging, vol. 3, no. 2, pp. 194–195, 1981. [DOI] [PubMed] [Google Scholar]
  • [32].Delannoy B, Lasota H, Bruneel C, Torguet R, and Bridoux E, “Infinite Planar Baffles Problem in Acoustic Radiation and Its Experimental-Verification,” (in English), J. Appl. Phys, vol. 50, no. 8, pp. 5189–5195, 1979. [Google Scholar]
  • [33].Selfridge AR, Kino GS, and Khuriyakub BT, “A Theory for the Radiation-Pattern of a Narrow-Strip Acoustic Transducer,” (in English), Appl. Phys. Let, vol. 37, no. 1, pp. 35–36, 1980. [Google Scholar]
  • [34].Krucker JF, Eisenberg A, Krix M, Lotsch R, Pessel M, and Trier HG, “Rigid piston approximation for computing the transfer function and angular response of a fiber-optic hydrophone,” (in English), J. Acoust. Soc. Am, vol. 107, no. 4, pp. 1994–2003, April 2000. [DOI] [PubMed] [Google Scholar]
  • [35].Wear KA, Baker C, and Miloro P, “Directivity of Needle Hydrophones: Predictions from Four Theoretical Forms Compared with Experimental Measurements,” IEEE Trans Ultrason Ferroelectr Freq Control, p. in press, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [36].Selfridge A and Lewin PA, “Wideband spherically focused PVDF acoustic sources for calibration of ultrasound hydrophone probes,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 47, no. 6, pp. 1372–1376, November 2000. [DOI] [PubMed] [Google Scholar]
  • [37].Harris GR and Shombert DG, “A Pulsed near-Field Technique for Measuring the Directional Characteristics of Acoustic Receivers,” (in English), IEEE Trans Son. Ultrason, vol. 32, no. 6, pp. 802–808, 1985. [Google Scholar]
  • [38].Smith RA and Bacon DR, “A Multiple-Frequency Hydrophone Calibration Technique,” (in English), J. Acoust. Soc. Am, vol. 87, no. 5, pp. 2231–2243, May 1990. [DOI] [PubMed] [Google Scholar]
  • [39].Radulescu EG, Lewin PA, Nowicki A, and Berger WA, “Hydrophones’ effective diameter measurements as a quasi-continuous function of frequency,” (in English), Ultrasonics, vol. 41, no. 8, pp. 635–641, November 2003. [DOI] [PubMed] [Google Scholar]
  • [40].Conjusteau A, Ermilov SA, Su R, Brecht HP, Fronheiser MP, and Oraevsky AA, “Measurement of the spectral directivity of optoacoustic and ultrasonic transducers with a laser ultrasonic source,” Rev Sci Instrum, vol. 80, no. 9, p. 093708, September 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [41].Guggenheim JA, Zhang EZ, and Beard PC, “A Method for Measuring the Directional Response of Ultrasound Receivers in the Range 0.3–80 MHz Using a Laser-Generated Ultrasound Source,” (in English), IEEE Trans. Ultrason. Ferroelectr., Freq. Control, vol. 64, no. 12, pp. 1857–1863, December 2017. [DOI] [PubMed] [Google Scholar]
  • [42].Wilkens V and Molkenstruck W, “Broadband PVDF membrane hydrophone for comparisons of hydrophone calibration methods up to 140 MHz,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 54, no. 9, pp. 1784–1791, September 2007. [DOI] [PubMed] [Google Scholar]
  • [43].Fay B, Ludwig G, Lankjaer C, and Lewin PA, “Frequency-Response of Pvdf Needle-Type Hydrophones,” (in English), Ultrasound in Med. & Biol, vol. 20, no. 4, pp. 361–366, 1994. [DOI] [PubMed] [Google Scholar]
  • [44].Harris GR and Gammell PM, “Sensitivity measurements of piezoelectric polymer hydrophones from 0.2–2 MHz using a broadband-pulse technique,” (in English), Journal of the Acoustical Society of America, vol. 105, no. 2, pp. 725–731, February 1999. [Google Scholar]
  • [45].Beard PC, Hurrell AM, and Mills TN, “Characterization of a polymer film optical fiber hydrophone for use in the range 1 to 20 MHz: A comparison with PVDF needle and membrane hydrophones,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 47, no. 1, pp. 256–264, January 2000. [DOI] [PubMed] [Google Scholar]

RESOURCES