Abstract
High-intensity therapeutic ultrasound (HITU) pressure is often measured using a hydrophone. HITU pressure waves typically contain multiple harmonics due to nonlinear propagation. As harmonic frequency increases, harmonic beam width decreases. For sufficiently high harmonic frequency, beam width may become comparable to the hydrophone effective sensitive element diameter, resulting in signal reduction due to spatial averaging. An analytic formula for a hydrophone spatial averaging filter for beams with Gaussian harmonic radial profiles was tested on HITU pressure signals generated by three transducers (1.45 MHz, F/1; 1.53 MHz, F/1.5; 3.91 MHz, F/1) with focal pressures up to 48 MPa. The HITU signals were measured using fiber-optic and needle hydrophones (nominal geometrical sensitive element diameters: 100 μm and 400 μm). Harmonic radial profiles were measured with transverse scans in the focal plane using the fiber-optic hydrophone. Harmonic radial profiles were accurately approximated by Gaussian functions with root-mean-square (RMS) differences between transverse scans and Gaussian fits less than 9% for frequencies up to approximately 50 MHz. The Gaussian harmonic beam width parameter σn varied with harmonic number n according to a power law, σn = σ1/nq where 0.5 < q < 0.6. RMS differences between experimental and theoretical spatial averaging filters were 11% ± 1% (1.45 MHz), 8% ± 1% (1.53 MHz), and 4% ± 1% (3.91 MHz). For the two more highly focused (F/1) transducers, the effect of spatial averaging was to underestimate peak compressional pressure (pcp), peak rarefactional pressure (prp), and pulse intensity integral (pii) by (mean ± standard deviation) (pcp: 4.9% ± 0.5%, prp: 0.4% ± 0.2% pii: 2.9% ± 1.0%) and (pcp: 28.3% ± 9.6% prp: 6.0% ± 2.4% pii: 24.3% ± 6.7%) for the 100 μm and 400 μm diameter hydrophones respectively. These errors can be suppressed by application of the inverse spatial averaging filter.
Keywords: Acoustic output measurement, high intensity therapeutic ultrasound, hydrophone, needle, fiber optic, spatial averaging, spatiotemporal transfer function
I. Introduction
High intensity therapeutic ultrasound (HITU) is a minimally-invasive technology for treatment of cancer [1–6], uterine fibroids [7], brain disorders [3, 8–10], and other conditions [11–19]. It is essential to accurately measure acoustic output of HITU devices in order to ensure that they deliver appropriate dose to targeted tissues. Characterization of HITU systems is challenging because standardized methods for HITU dose measurements do not exist yet and high intensities of ultrasound can damage common measurement devices or render them ineffective [20, 21]. To address this challenge, the International Electrotechnical Commission recently initiated an effort to draft a technical specification on measurement of ultrasound at high pressure therapeutic levels (IEC/TS 62937) [22].
Acoustic dose is often characterized by acoustic power, which is typically measured using a radiation force balance [23, 24], and acoustic pressure, which is typically measured using a hydrophone [25, 26]. Reflectance-based fiber-optic hydrophones are often used to characterize HITU and lithotripsy pressure fields because they have high spatial resolution (on the order of 100 μm or less) and can withstand very high pressures (up to 70 MPa and beyond) [12, 27–37]. Robust needle [38] and membrane [20, 39] hydrophones have also been developed for HITU applications.
Hydrophones can distort pressure signals due to 1) frequency-dependent sensitivity and 2) spatial averaging across the finite sensitive element. These forms of distortion are relevant to HITU signals, which can be highly nonlinear and can therefore have very broad bandwidths due to the presence of multiple harmonics [28, 29, 40–43]. Nonlinearity affects spatial averaging because as harmonic frequency increases, harmonic beam width decreases [40, 44–49] and the potential for harmonic spatial averaging increases [50, 51].
The distorting effects of frequency-dependent sensitivity can be suppressed by performing sensitivity deconvolution of hydrophone measurements [27–30, 52–61]. Recently, an inverse-filter method for correction for spatial averaging artifacts in nonlinear ultrasound signals was derived and validated for signals in the diagnostic pressure range [50, 51]. The objective of the present paper is to extend that validation to HITU pressure signals.
II. Methods
A. Hydrophone Response Model
The hydrophone spatiotemporal response, Hp(f), is modeled as the product of spatial and temporal factors.
| (1) |
where f is frequency, ML(f) is the loaded sensitivity of the hydrophone, defined as the ratio of the frequency-domain voltage output response U(f) to a normally-incident input plane pressure wave PNP(f)
| (2) |
and Sp(f) is a filter that accounts for spatial averaging. The subscript p is a reminder that the function depends on the particular pressure field distribution incident upon the hydrophone. The separability assumption was previously shown to be valid for a wide variety of transducers and hydrophones [50, 51].
B. Spatial Averaging Filter
The frequency-dependent voltage output of a hydrophone is proportional to the integral of the complex pressure field incident upon its sensitive element. The ratio of this integral to the area of the sensitive element quantifies the effects of spatial averaging [62–66]. The numerator and denominator of the ratio should be calculated based on the hydrophone effective sensitive element radius aeff(f), which is a function of frequency and can differ from the geometrical radius, ag [50]. Functional forms for aeff(f) have been reported for needle and fiber optic hydrophones [50, 67, 68]. The spatial averaging filter may be written as [50]
| (3) |
where r is the radial coordinate and p(r, f) is the pressure distribution normalized to its maximum (axial) value. (Previous publications neglected to specify that p(r, f) in this formula should be normalized [50, 51], although this should have been apparent to most readers anyway. Otherwise the formula would predict the average pressure across the sensitive element instead of the dimensionless spatial averaging filter.) In this investigation, harmonic components of ultrasound beams are assumed to have Gaussian radial profiles. (This assumption has been validated previously for low-intensity nonlinear pressure fields [50] and will be tested experimentally for high-intensity pressure fields in the present paper.) For Gaussian harmonic beams, it can be shown that the spatial averaging filter is given by [50],
| (4) |
where
| (5) |
f = nf1, f1 is the fundamental frequency, n is the harmonic number [50] and the radial profile of each harmonic component of the nonlinear beam is given by
| (6) |
The coefficient gn is related to the rate of phase change with the radial coordinate. The full width half maximum (FWHM) of each harmonic component is .
The spatial averaging filter requires knowledge of the harmonic beam width parameter σn for all harmonics of significant amplitude in the beam. Ideally, these could be measured from a high-resolution lateral hydrophone scan of the pressure field in the focal plane. If the data are analyzed in frequency domain, then σn may be measured for each individual harmonic.
If lateral scan data are not available, σ1 may sometimes be modelled based on simulation [41] or diffraction theory. If the diffraction pattern for the fundamental frequency is the same as the low-amplitude case (so-called “classic-jinc-fundamental” or CJF condition [50]), then σ1 for a circular focused piston transducer in the focal plane may be found by minimizing the mean-square difference between the magnitudes of (6) and the classic theoretical diffraction pattern over the half width half maximum (HWHM) of the beam [69], which results in σ1 = 1.93 D / (k1as), where D = focal length. k1 = 2π / λ1, λ1 is the wavelength of the fundamental component of the beam, and as is the radius of the source [50]. The CJF approximation will be tested in the present investigation of HITU signals.
The dependence of σn on σ1 and n may sometimes be accurately approximated as a power law, σn=σ1/nq [40, 44, 45, 47, 48, 50]. Simulation analysis has suggested that the power law approximation is accurate for transducers with center frequencies between 1 MHz and 5 MHz and f-numbers between 1.4 and 2.5 provided that σm (the nonlinear propagation parameter [70], not to be confused with σn) < 2.4, which corresponds approximately and conservatively to σq (the local distortion parameter [71]) < 3.0 or the spectral index SI (the fraction of signal power in frequencies above the fundamental lobe [72]) < 0.4 [50]. This power law approximation will be tested in the present investigation of HITU signals.
A formula for the frequency-dependent effective sensitive element size [50], which is required for (5), can be derived from a rigid piston model [73, 74] that has been previously validated for sensitivity [75, 76] and directivity [67] of needle hydrophones and sensitivity [76] and directivity [68] of fiber optic hydrophones.
| (7) |
The coefficients A and B depend weakly on the effective range of angle θ in the angular spectrum of the beam [50]. A varies from 1.81 (|θ| < 10°) to 2.13 (|θ| < 90°). B varies from 1.07 (|θ| < 10°) to 0.94 (|θ| < 90°) [50]. The values chosen for the present investigation were A = 1.85 and B = 1.05, corresponding to |θ| < 30°. This formula applies to needle hydrophones and reflectance-based fiber optic hydrophones that measure changes in a fluid refractive index caused by pressure changes [27, 30, 31, 68, 77, 78]. It does not apply to fiber optic displacement sensors [79], Fabry-Perot interferometric fiber optic hydrophones [57, 80–83], other Fabry-Perot sensors [84, 85], or other fiber optic designs [86–88].
C. Substitution Experiment
If the pressure field incident upon a hydrophone has a spectrum Pp(f), then the hydrophone output voltage may be expressed as
| (8) |
The same spectrum Pp(f) may be measured using two different hydrophones (here labeled 1 and 2). If (8) is rewritten for each hydrophone by adding a subscript 1 or 2 to Up(f), ML(f), and Sp(f) and then the ratio of the two equations is taken, the pressure spectrum Pp(f) (same for both measurements) will cancel out and the following substitution formula may be obtained [51],
| (9) |
The quotients of functions in the numerator and denominator, Upi(f) / MLi(f), may be regarded as voltage spectra that have been deconvolved for hydrophone sensitivities [56].
In order to isolate the effects of spatial averaging, each frequency-dependent spatial averaging filter function was scaled so that its average value was equal to the average value of the theoretical spatial averaging filter over the range of harmonic frequencies measured. This eliminated discrepancies due to uncertainties in sensitivity magnitude, which have been reported to be on the order of ten percent [62]. There are many sources of uncertainty in hydrophone sensitivity including initial calibration uncertainty, dependence of sensitivity with temperature, and sensitivity drift over time [61].
D. Experimental Methods
Table I lists the hydrophones used. Hydrophone 1 (the reference hydrophone) was a fiber optic hydrophone (HFO, Onda, Sunnyvale, CA), with a nominal geometrical sensitive element diameter of 100 μm [30]. Hydrophone 2 was a robust needle hydrophone (HNA-0400, Onda) designed to withstand and measure high pressures, with a nominal geometrical sensitive element diameter of 400 μm [38].
TABLE I.
Hydrophones
| Hydrophone | Type | Nominal Geometrical Sensitive Element Diameter dg (μm) |
|---|---|---|
| Onda HFO | Fiber optic | 100 |
| Onda HNA-0400 | Ceramic | 400 |
Table II lists the transducers used. Sources were driven in tone bursts with a 100 Hz repetition rate (10 msec repetition period) through an Agilent (Santa Clara, CA) HP3314A function generator into an ENI (Rochester, NY) 240L amplifier. Waveforms were acquired using an Agilent 5012A digital storage oscilloscope after ringing-up to steady-state conditions. Data acquisition times were 5 - 10 μs.
TABLE II.
Transducer and Wave Properties
| Frequency (MHz) | 1.45 | 1.53 | 3.91 | |
| Diameter (mm) | Geometrical | 100 | 100 | 25 |
| Effective | 100 ± 3 | 83 ± 1 | 17 ± 1 | |
| Focal Length (mm) | 100 | 150 | 25 | |
| Spectral Index SI | 0.42 ± 0.01 | 0.34 ± 0.01 | 0.41 ± 0.01 | |
| Nonlinear propagation parameter σm | 2.45 ± 0.06 | 2.30 ± 0.03 | 2.44 ± 0.07 | |
| Local distortion parameter σq | 3.79 ± 0.06 | 2.97 ± 0.01 | 2.89 ± 0.05 | |
| Exponent q (σn = σ1 / nq) | n =1-5 | 0.72 ± 0.01 | 0.64 ± 0.03 | 0.59 ± 0.01 |
| 0-40 MHz | 0.60 ± 0.01 | 0.50 ± 0.03 | 0.53 ± 0.01 | |
| Focal pressure (MPa) | 48.7 ± 0.3 | 16.2 ± 0.4 | 29.4 ± 0.4 | |
After alignment of the beam, the depth of the maximum pulse-pressure-squared was found, and transverse scans were conducted in horizontal and vertical directions. Radiofrequency (RF) data were stored at every scan point and post-processed afterward. HFO signals were averaged in order to improve signal-to-noise ratio (SNR). The numbers of averages were 64 (1.45 MHz), 128 (1.53 MHz), and 32 (3.91 MHz). For the HNA-0400 measurements, an Onda AH2020 (1.45 MHz) or Onda AH2010 (1.53 MHz and 3.91 MHz) preamplifier was used.
The HNA-0400 spatial averaging filter Sp2(f) was evaluated from (9) as follows. The HFO was used to measure voltage response Up1(t) from a nonlinear tone burst pressure wave at the focal point of the transducer. Voltage response Up2(t) was measured by replacing the HFO with the HNA-0400 and repeating the measurement. Windowed steady-state signals (excluding ring-up and ring-down portions) were analyzed. Spectra of time-domain voltage responses Up1(f) and Up2(f) were estimated using the Fast Fourier Transform (FFT). Sensitivities ML1(f) and ML2(f) in (9) were provided by the manufacturer. Sensitivities for needle [75, 76] and fiber optic [76] hydrophones have been shown to be accurately predicted by the rigid piston model [73, 74]. The sensitivity deconvolution bandwidth was limited by the maximum frequency for which the HNA-0400 sensitivity was calibrated, which was 60 MHz. Spatial averaging filters were computed up to a maximum frequency that varied between 43 MHz and 55 MHz among the three source transducers, depending on SNR.
For the HFO, the sensitivity was deconvolved as previously reported [30]. Briefly, a static (no sound on) measurement of HFO output was used to determine the small-disturbance relationship between HFO output and pressure (with knowledge of known properties of the water and fiber). Next, the slight nonlinearity in the dependence of the HFO output on pressure (on the order of 10% at 50 MPa and due to the nonlinear properties of water) was corrected for. This dependence was based on standard tables for water, assuming no significant frequency dependence. Finally, the frequency-dependent diffraction of the sound wave at the tip was corrected via deconvolution, using a diffraction kernel derived via Finite Element Analysis [30], calculated over the electronic bandwidth of the HFO, which was 150 MHz. When the Nyquist frequency (i.e., half of the sampling frequency) exceeded 150 MHz, the deconvolution kernel was extended by assuming it is constant between 150 MHz and the Nyquist frequency.
Harmonic beam width parameters σn were measured from FFTs of transverse scans using the HFO fiber optic hydrophone. The exponent q was estimated by fitting harmonic FWHM to a power law function of harmonic number n that was proportional to 1/nq as described previously [50].
E. Data Analysis
Effective radii of transducers were determined by fitting the focal plane diffraction pattern (measured with the HFO) for the first harmonic to a theoretical 2J1(x) / x function where J1() is a Bessel function of the first kind, x = k1 as r / D, k1 = 2π/λ1, λ1 = fundamental wavelength, as = effective radius of transducer (adjustable fitting parameter), r = radial coordinate, and D = focal length [69]. More sophisticated transducer geometric modeling is possible if an axial scan [28, 42, 89] (instead of or in addition to a lateral scan) or a two-dimensional planar scan [29, 34, 90, 91] is available. While other methods evaluate effective radius in the linear regime, the present method evaluated effective radius in the nonlinear regime. Therefore, effective radius in the present context may incorporate effects of nonlinear propagation.
Nonlinearity of axial signals was characterized by the nonlinear propagation parameter σm [70], the local distortion parameter σq [71], and the spectral index SI [72]. Note that σm is approximately equal to σq when the local area factor Fa (square root of the ratio of the source aperture area to beam area) is between 2 and 12. Numerical analysis suggests that σm and σq have high correlations with energy transfer from the fundamental to higher harmonics [71]. SI is the fraction of the power spectrum contained in frequencies above the fundamental frequency [72, 92].
The effects of spatial averaging corrections on time-domain signals were investigated as follows. The spatial averaging correction was achieved by 1) applying an FFT to the voltage signals, 2) multiplying by the inverse of the spatial averaging filter (4), and 3) applying an inverse FFT. The HNA-0400 inverse spatial averaging filter was multiplied by a tenth-order Butterworth low-pass filter [93] in order to prevent excessive amplification of frequencies with low SNR that could result from inverse filtering. Low-pass regularization is a common component of inverse filtering. Regularization may induce systematic errors that can be addressed by uncertainty contributions using methods described in [94]. Pressure signals were deconvolved for hydrophone sensitivity. Noise in the rarefactional component of sensitivity-deconvolved signals was suppressed with a rarefactional filter [57].
III. Results
A. Spectra
Fig. 1 shows spectra of HFO and HNA hydrophon measurements from the three transducers. Spectra were deconvolved for hydrophone frequency-dependent sensitivities and normalized to their values at their fundamental frequencies. The nonlinear nature of the signals results in spectral peaks at integer multiples of the fundamental frequency. HNA data decrease with frequency more rapidly than HFO data due to greater spatial averaging effects resulting from the larger geometrical sensitive element diameter (400 μm vs. 100 μm).
Fig. 1.
Spectra of HFO and HNA hydrophone measurements from three transducers. Transducer center frequencies are given in the upper right corner of each plot. Spectra are deconvolved for hydrophone frequency-dependent sensitivities and normalized to their values at their fundamental frequencies. HNA data decrease with frequency more rapidly than HFO data due to greater spatial averaging effects resulting from the larger nominal geometrical sensitive element diameter (400 μm vs. 100 μm).
B. Nonlinearity Indexes
Table II shows measurements of σm, σq, and SI for the axial signals produced by the three transducers. The approximate factor of 6 between σm and SI is similar to that found by 1) Duck for a pulsed single-element 3.38 MHz transducer with a focal depth of 95 mm (see Fig. 7 in [72]), 2) Duncan et al. for many transducers when energy transferred from the fundamental frequency to higher harmonics was less than 20 percent (see Fig. 5 in [71]), and 3) Wear for 6 transducers for SI < 0.4, σm < 2.4 and σq < 3 [50]. Values for σq are well in excess of the IEC 62127-1 threshold for “considerable nonlinear distortion” of σq > 1.5 [95].
Table II indicates that the average value for σm for the pressure fields produced by the three transducers was approximately 2.4. The exponent q was approximately 0.5 - 0.6 (power law fits from 0 MHz to 40 MHz) or 0.6 - 0.7 (power law fits over the first 5 harmonics). The latter range is very consistent with previously-reported simulations based on 6 transducer geometries for σm ≈ 2.4 (See Fig. 9 in [50] which shows least-squares linear fit of q = 0.79 – 0.042σm implying that q = 0.69 when σm = 2.4).
C. Beam Properties
Fig.s 2–4 show harmonic beam profile plots at three different zoom levels for nonlinear pressure waves produced by the three transducers. The harmonic number is denoted by n. Gaussian fits are shown in dotted lines. Error bars denote plus and minus one standard deviation, obtained from transverse scans obtained horizontally and vertically. The black vertical lines show the spatial extent of the HNA-0400 geometrical sensitive element. The beam plots support the Gaussian model (6) for radial profiles throughout the spatial extent of the HNA-0400 sensitive element.
Fig. 2.
Harmonic beam profile plots at three different zoom levels for nonlinear pressure waves produced by the 1.45 MHz transducer. The harmonic number is denoted by n. Gaussian fits are shown in dotted lines. Error bars denote plus and minus one standard deviation, obtained from transverse scans obtained horizontally and vertically. The black vertical lines show the spatial extent of the HNA-0400 sensitive element, within which Gaussian shape is assumed by (4). The beam plots support the Gaussian model for radial profiles across the HNA-0400 sensitive element.
Fig. 4.
Harmonic beam profile plots at three different zoom levels for nonlinear pressure waves produced by the 3.91 MHz transducer. The harmonic number is denoted by n. Gaussian fits are shown in dotted lines. Error bars denote plus and minus one standard deviation, obtained from transverse scans obtained horizontally and vertically. The black vertical lines show the spatial extent of the HNA-0400 sensitive element, within which Gaussian shape is assumed by (4). The beam plots support the Gaussian model for radial profiles across the HNA-0400 sensitive element.
Fig. 5 shows RMS differences (RMSD) between experimental beam radial profiles and Gaussian fits over the HNA-0400 geometrical sensitive element (± 200 μm) as functions of frequency. RMSD remained below 9% for all three transducers over the range of frequencies shown. Error bars are not shown (to reduce clutter), but the average standard deviations of RMSD for the three transducers over the ranges of frequency shown were 6% (1.45 MHz), 3% (1.53 MHz), and 1% (3.91 MHz). These relatively low values for RMSD support the Gaussian model for radial profiles across the HNA-0400 sensitive element.
Fig. 5.
RMS differences (RMSD) between experimental beam radial profiles and Gaussian fits as functions of frequency. Error bars are not shown (to reduce clutter), but the average standard deviations of RMSD values for the three transducers over the ranges of frequency shown were 6% (1.45 MHz), 3% (1.53 MHz), and 1% (3.91 MHz).
Fig. 6 shows full width half maxima (FWHM) of harmonic radial profiles vs. harmonic number for the three transducers. Least-squares power law fits are shown in the continuous curves. The RMSDs between experimental and power-law-fit FWHM for data shown in Fig. 5 were 11% (1.45 MHz), 10% (1.53 MHz), and 7% (3.91 MHz). It is conceivable that the measured FWHM values are larger than the true FWHM values for higher harmonic numbers due to convolution with the HFO transverse spatial response. However, it can be seen that the power law fits are quite good for those harmonics for which FWHM > 0.4 mm (four times the HFO sensitive element diameter) and therefore relatively unaffected by the transverse convolution. Therefore, the power law fits appear to be reasonable approximations. If q is estimated based on just the first 5 harmonics rather than all the harmonics shown in Fig. 6, estimates of q increase by approximately 0.1 (see Table II).
Fig. 6.
Full width half maxima (FWHM) of harmonic radial profiles vs. harmonic number for the three transducers. Least-squares power law fits are shown in the continuous curves.
D. Spatial Averaging Filters
Fig. 7 shows spatial averaging filters computed for the HNA hydrophone using 1) the substitution formula (9), 2) σ1 estimated from classic diffraction theory and the effective transducer radius (see Section II.E), 3) values of exponent q shown in Fig. 6 and Table II, and 4) the frequency-dependent effective hydrophone sensitive element radius (7). The spatial averaging filters for these signals are monotonically decreasing functions of harmonic frequency. The RMSDs between experimental and theoretical spatial averaging filters for data shown in Fig. 7 are 11% ± 1% (1.45 MHz), 8% ± 1% (1.53 MHz), and 4% ± 1% (3.91 MHz).
Fig. 7.
Spatial averaging filters for the three transducers: 1.45 MHz (top), 1.53 MHz (middle), and 3.91 MHz (bottom). Error bars denote plus and minus one standard deviation, obtained from transverse scans obtained horizontally and vertically.
Fig. 8 shows averaged hydrophone pressure tone burst measurements from the 1.45 MHz transducer (windowed steady-state signals excluding ring-up and ring-down portions) before (top) and after (bottom) correction for spatial averaging effects. Spatial averaging affects compressional segments more than rarefactional segments because spatial averaging is a low-pass filter and compressional segments have far more high-frequency content than rarefactional segments (see Fig. 3 in [57]). After spatial averaging correction, the waveforms acquired with the two hydrophones were much more consistent with each other. After spatial averaging correction, the HNA signal is somewhat noisier than before, which is a common result of inverse filtering.
Fig. 8.
Hydrophone pressure measurements from the 1.45 MHz transducer before (a) and after (b) correction for spatial averaging. All waveforms in (a) and (b) were deconvolved for hydrophone sensitivities.
E. Effect of Spatial Averaging on Pressure Measurements
The effects of spatial averaging were greater for the two F/1 (nominally) transducers than for the F/1.5 transducer because of their tighter focusing. For the two F/1 transducers, the effect of spatial averaging was to underestimate peak compressional pressure by 4.9% ± 0.5% (HFO) and 28.3% ± 9.6% (HNA) (mean ± standard deviation; results from the two transducers are averaged). The first value is similar to that reported by Canney et al. [28] (2%) obtained by simulation for a 2 MHz, F/1 (nominally) transducer measured using a reflectance-based fiber-optic hydrophone with a similar sensitive element diameter (100 μm). The present investigation supports Canney et al.’s conclusion that the difference that they observed between measured and modeled values for peak compressional pressure at high drive levels was not primarily attributable to spatial averaging effects. Canney et al. also reported “significant distortion” when the simulated hydrophone diameter was changed from 100 μm to 500 μm. Similarly, in the present investigation, distortion significantly increased when the 100 μm diameter hydrophone was replaced by the 400 μm diameter hydrophone.
For the two F/1 transducers, the effect of spatial averaging was to underestimate peak rarefactional pressure by 0.4% ± 0.2% (HFO) and 6.0% ± 2.4% (HNA) (results from the two transducers are averaged). These effects are much smaller than those measured for peak compressional pressure because (again) spatial averaging is a low-pass filter and compressional segments have far more high-frequency content than rarefactional segments (see Fig. 3 in [57]).
For the two F/1 transducers, the effect of spatial averaging was to underestimate pulse intensity integral by 2.9% ±1.0% (HFO) and 24.3% ± 6.7% (HNA) (results from the two transducers are averaged).
IV. Discussion
Many investigations have considered corrections for spatial averaging across a hydrophone sensitive element [50, 62–66, 96–101], but only one, to our knowledge, derives an analytic formula for spatial averaging of a nonlinear field [50]. The formula expresses the spatial averaging filter as a function of frequency-dependent hydrophone sensitive element size and frequency-dependent beam width [50]. Spatial averaging is especially important for nonlinear fields with multiple harmonics, which are characterized by harmonic beam widths decreasing as harmonic number increases [40, 44–49]. The spatial averaging filter formula was previously validated for nonlinear focused pressure waves with maximum pressures of a few MPa generated using source transducers with center frequencies in the range from 1 MHz to 5 MHz and f-numbers in the range from 1.4 to 2 [51]. This previous validation supported a graphical guide to aid investigators in choosing an appropriate hydrophone sensitive element size for their measurements [50]. In the present paper, the validation was extended to HITU pressure beams with maximum pressures up to 48 MPa generated using source transducers with center frequencies in the range of 1.5 MHz – 4 MHz and f-numbers in the range of 1 – 1.5.
The exponent q determines how rapidly harmonic beam width decreases with harmonic number n and therefore can play an influential role in determining the extent of spatial averaging across the hydrophone sensitive element. For the pressure waves investigated here, q fell approximately in the range from 0.6 to 0.7 if based on the first five harmonics and from 0.5 to 0.6 if based on all harmonics up to about 43-55 MHz (see Table II). The values of q (based on the first five harmonics) found in the present investigation are consistent with previous simulation analysis that suggested that q (based on the first five harmonics) would be expected to be approximately 0.69 when σm ≈ 2.4 (see Fig. 9 in [50]). While a power law representation simplifies computation of the spatial averaging filter (4), it is not required. The spatial averaging filter formula (4) can accommodate any functional form for the dependence of the beam width parameter σn on harmonic number n.
The spatial averaging correction method used in this paper has at least two important advantages over the method found in IEC 62127-1 Annex E [95]. First, it accommodates nonlinear signals. Second, it provides an analytic expression for the spatial averaging filter in terms of beam and hydrophone dimensions.
The concordance between theory and experiment in Fig. 7 supports some assumptions underlying the spatial averaging filter formula [50] for the HITU conditions considered here. First, hydrophone spatiotemporal response is separable into the product of frequency-dependent sensitivity and a spatial averaging filter. Second, spatial averaging effects can be accurately modeled by integrating the magnitude of the acoustic pressure over a frequency-dependent effective sensitive element area.
The sensitivity deconvolutions performed here required sensitivity calibration for frequencies up to at least 40 MHz, which can be achieved using some reported methods [54, 56, 85, 102–106]. While the experiments were performed with rigid piston (RP) hydrophones (i.e., needle and fiber-optic hydrophones), the spatial averaging filter formula (4) can be applied to non-RP hydrophones provided spectra that are accurately deconvolved for frequency-dependent sensitivity and the appropriate frequency-dependent hydrophone sensitive element size is used. The theory assumes uniform response across the active area of the hydrophone. The theory may be less accurate at frequencies where uniformity is compromised (for example, if a hydrophone has a strong higher order radial resonance).
The distinction between effective and geometrical source transducer diameter can be important [28]. As shown in Table II, effective source diameter was essentially equal to geometrical source diameter for the 1.45 MHz transducer but considerably less than the geometrical source diameter for the 1.53 MHz and 3.91 MHz transducers. The mean reductions in spatial averaging filter (over the ranges of frequencies in Fig. 7) produced by using geometrical instead of effective source transducer diameter in (4) were 1% (1.45 MHz), 5% (1.53 MHz), and 18% (3.91 MHz). In the first two cases, the difference is not very consequential. In the third case, the difference is greater but still provides a reasonable approximation to the spatial averaging filter. In cases in which the effective source transducer diameter is not known, the geometrical source diameter can be used to provide a useful lower bound for the spatial averaging filter (and therefore an upper bound for spatial averaging reduction).
V. Conclusion
This paper provides experimental validation for a theoretical model for spatial averaging effects for nonlinear HITU beams with focal pressures up to 48 MPa measured with hydrophones. Spatial averaging effects are characterized by a spatial averaging filter, which is a low-pass filter. Spatial averaging artifacts may be suppressed by applying the inverse of the spatial averaging filter to measurements.
The spatial averaging filter requires knowledge of the frequency-dependent hydrophone effective sensitive element radius, which can obtained from a rigid piston model for needle and fiber optic hydrophones [50] or from empirical data for membrane hydrophones [107]. The spatial averaging filter also requires knowledge of frequency-dependent beam width. For the transducers and nonlinearity indexes considered here (center frequency: 1.5 MHz – 4 MHz; f-number: 1 – 1.5; σm < 2.4; σq < 3.8; SI < 0.4), the focal plane harmonic beams were accurately approximated by Gaussian radial functions. The width of the fundamental component, σ1, was approximately equal to the width of the classic linear diffraction pattern based on the effective transducer diameter. Beam width varied with harmonic number approximately as a power law, σn=σ1/nq where q = 0.5 – 0.6 (see Table II). For conditions outside the range of frequencies, f-numbers, and pressure levels considered here and previously [50, 51], nonlinear simulation tools [20, 28, 39, 41] may offer a useful alternative to estimating harmonic beam properties to be used as inputs to the spatial averaging filter formula (4). If upper and lower bounds for frequency-dependent effective hydrophone sensitive element size and harmonic beam width can be estimated (from theory, simulation, or experiment) then upper and lower bounds for the spatial averaging filter can be determined.
The spatial averaging filter is useful for 1) correction of hydrophone-based measurements for spatial averaging artifacts, 2) retrospective analysis of HITU acoustic output measurements, 3) informing choice of an appropriate hydrophone for measuring HITU system acoustic output, and 4) informing design of HITU systems in conjunction with hydrophone test procedures. In some cases, this approach may allow relatively expensive fiber-optic hydrophones to be replaced by relatively economical needle hydrophones (provided they are sufficiently robust to withstand HITU pressures) for characterization of HITU beams.
Fig. 3.
Harmonic beam profile plots at three different zoom levels for nonlinear pressure waves produced by the 1.53 MHz transducer. The harmonic number is denoted by n. Gaussian fits are shown in dotted lines. Error bars denote plus and minus one standard deviation, obtained from transverse scans obtained horizontally and vertically. The black vertical lines show the spatial extent of the HNA-0400 sensitive element, within which Gaussian shape is assumed by (4). The beam plots support the Gaussian model for radial profiles across the HNA-0400 sensitive element.
Acknowledgements
The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services. The authors are grateful for funding support from the FDA Office of Women’s Health.
Biographies
Keith A. Wear received his B.A. in Applied Physics from the University of California at San Diego. He received his M.S. and Ph.D. in Applied Physics with a Ph.D. minor in Electrical Engineering from Stanford University. He was a post-doctoral research fellow with the Physics department at Washington University, St. Louis. He is the FDA Acoustics Laboratory Leader. He is an Associate Editor-in-Chief for IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. He is an Associate Editor of 3 journals: J. Acoust. Soc. Am.; IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control; and Ultrasonic Imaging. He was the Technical Program Chair of the 2008 IEEE International Ultrasonics Symposium in Beijing, China. He was the General Program Chair of the 2017 IEEE International Ultrasonics Symposium in Washington, DC. He was elected to serve as vice chair (2012-2014) and chair (2014-2016) of the American Institute of Ultrasound in Medicine (AIUM) Technical Standards Committee. He was elected to serve as vice-chair of the AIUM Bioeffects Committee (2019-2021). He received the 2019 AIUM Joseph H. Holmes Basic Science Pioneer Award. He is a Fellow of the Acoustical Society of America, the American Institute for Medical and Biological Engineering, and the AIUM. He is a senior member of IEEE.

Samuel M. Howard received his A.B. in Engineering Sciences from Harvard University. He received his M.S. and Ph.D in Theoretical and Applied Mechanics from Cornell University with minors in Applied Mathematics and Physics. From 1989 to 2001 he worked for Acuson (later Siemens) in transducer R&D. From 2001 to the present he has been at Onda Corporation, where he is the CTO and directs the Acoustics Laboratory. His current interests focus on ultrasound metrology for medical and industrial applications.
Contributor Information
Keith A. Wear, Center for Devices and Radiological Health, Food and Drug Administration, Silver Spring, MD, 20993 USA.
Samuel M. Howard, Onda Corporation, Sunnyvale, CA 94089
References
- [1].McDannold K Hynynen D Wolf G. Wolf, and Jolesz F, “MRI evaluation of thermal ablation of tumors with focused ultrasound,” J Magn Reson Imaging, vol. 8, no. 1, pp. 91–100, Jan-Feb 1998. [DOI] [PubMed] [Google Scholar]
- [2].Palussiere J et al. , “Feasibility of MR-guided focused ultrasound with real-time temperature mapping and continuous sonication for ablation of VX2 carcinoma in rabbit thigh,” Magn Reson Med. vol. 49, no. 1, pp. 89–98, January 2003. [DOI] [PubMed] [Google Scholar]
- [3].McDannold N, Clement GT, Black P, Jolesz F, and Hynynen K, “Transcranial magnetic resonance imaging- guided focused ultrasound surgery of brain tumors: initial findings in 3 patients,” Neurosurgery, vol. 66, no. 2, pp. 323–32; discussion 332, February 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [4].Payne A et al. , “In vivo evaluation of a breast-specific magnetic resonance guided focused ultrasound system in a goat udder model,” Med Phys. vol. 40, no. 7, p. 073302, July 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Minalga E et al. , “An 11-channel radio frequency phased array coil for magnetic resonance guided high-intensity focused ultrasound of the breast,” Magn Reson Med, vol. 69, no. 1, pp. 295–302, January 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Sommer G, Bouley D, Gill H, Daniel B, Pauly KB, and Diederich C, “Focal ablation of prostate cancer: four roles for magnetic resonance imaging guidance,” Can J Urol, vol. 20, no. 2, pp. 6672–81, April 2013. [PMC free article] [PubMed] [Google Scholar]
- [7].Hindley J et al. , “MRI guidance of focused ultrasound therapy of uterine fibroids: early results,” AJR Am J Roentgenol, vol. 183, no. 6, pp. 1713–9, December 2004. [DOI] [PubMed] [Google Scholar]
- [8].Hynynen K et al. , “500-element ultrasound phased array system for noninvasive focal surgery of the brain: a preliminary rabbit study with ex vivo human skulls,” Magn Reson Med, vol. 52, no. 1, pp. 100–7, July 2004. [DOI] [PubMed] [Google Scholar]
- [9].Elias WJ et al. , “A pilot study of focused ultrasound thalamotomy for essential tremor,” N Engl J Med, vol. 369, no. 7, pp. 640–8, August 15 2013. [DOI] [PubMed] [Google Scholar]
- [10].Chauvet D et al. , “Targeting accuracy of transcranial magnetic resonance-guided high-intensity focused ultrasound brain therapy: a fresh cadaver model,” J Neurosurg, vol. 118, no. 5, pp. 1046–52, May 2013. [DOI] [PubMed] [Google Scholar]
- [11].Hynynen K, “MRI-guided focused ultrasound treatments,” Ultrasonics, vol. 50, no. 2, pp. 221–9, February 2010. [DOI] [PubMed] [Google Scholar]
- [12].Maxwell AD et al. , “Cavitation clouds created by shock scattering from bubbles during histotripsy,” (in English), J. Acoust. Soc. Am, vol. 130, no. 4, pp. 1888–1898, October 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [13].Aptel F et al. , “Miniaturized high-intensity focused ultrasound device in patients with glaucoma: a clinical pilot study,” Invest Ophthalmol Vis Sci, vol. 52, no. 12, pp. 8747–53, November 11 2011. [DOI] [PubMed] [Google Scholar]
- [14].Wang TY, Hall TL, Xu Z, Fowlkes JB, and Cain CA, “Imaging feedback of histotripsy treatments using ultrasound shear wave elastography,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 59, no. 6, pp. 1167–81, June 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [15].Maxwell A et al. , “Disintegration of tissue using high intensity focused ultrasound: two approaches that utilize shock waves,” Acoustics Today, pp. 24–37, 2012. [Google Scholar]
- [16].ter Haar G, “Heat and sound: focused ultrasound in the clinic,” Int J Hyperthermia, vol. 31, no. 3, pp. 223–4, May 2015. [DOI] [PubMed] [Google Scholar]
- [17].Khokhlova VA et al. , “Histotripsy methods in mechanical disintegration of tissue: towards clinical applications,” Int J Hyperthermia, vol. 31, no. 2, pp. 145–62, March 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [18].Bader KB et al. , “Efficacy of histotripsy combined with rt-PA in vitro,” Phys Med Biol, vol. 61, no. 14, pp. 5253–74, July 21 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [19].Bader KB, Crowe MJ, Raymond JL, and Holland CK, “Effect of Frequency-Dependent Attenuation on Predicted Histotripsy Waveforms in Tissue-Mimicking Phantoms,” Ultrasound Med Biol, vol. 42, no. 7, pp. 1701–5, July 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [20].Wilkens V, Sonntag S, and Georg O, “Robust spot-poled membrane hydrophones for measurement of large amplitude pressure waveforms generated by high intensity therapeutic ultrasonic transducers,” (in English), J. Acoust. Soc. Am, vol. 139, no. 3, pp. 1319–1332, March 2016. [DOI] [PubMed] [Google Scholar]
- [21].Shaw A, ter Haar G, Haller J, and Wilkens V, “Towards a dosimetric framework for therapeutic ultrasound,” Int J Hyperthermia, vol. 31, no. 2, pp. 182–92, March 2015. [DOI] [PubMed] [Google Scholar]
- [22].Szabo TL, “New international standards for high pressure therapeutic ultrasound fields,” J. Acoust. Soc. Am, abstract vol. 144, no. 3, Pt. 2, p. 1700, 2018. [Google Scholar]
- [23].Maruvada S, Harris GR, Herman BA, and King RL, “Acoustic power calibration of high-intensity focused ultrasound transducers using a radiation force technique,” J Acoust Soc Am, vol. 121, no. 3, pp. 1434–9, March 2007. [DOI] [PubMed] [Google Scholar]
- [24].Haller J and Wilkens V, “Derivation of continuous wave mode output power from burst mode measurements in high-intensity ultrasound applications,” J Acoust Soc Am, vol. 135, no. 3, pp. EL123–7, March 2014. [DOI] [PubMed] [Google Scholar]
- [25].Harris GR, “Progress in medical ultrasound exposimetry,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 52, no. 5, pp. 717–36, May 2005. [DOI] [PubMed] [Google Scholar]
- [26].Szabo TL, Diagnostic Ultrasound Imaging: Inside Out, Second ed. Oxford, UK: Elsevier, 2014. [Google Scholar]
- [27].Staudenraus J and Eisenmenger W, “Fiberoptic Probe Hydrophone for Ultrasonic and Shock-Wave Measurements in Water,” (in English), Ultrasonics, vol. 31, no. 4, pp. 267–273, July 1993. [Google Scholar]
- [28].Canney MS, Bailey MR, Crum LA, Khokhlova VA, and Sapozhnikov OA, “Acoustic characterization of high intensity focused ultrasound fields: A combined measurement and modeling approach,” (in English), J. Acoust. Soc. Am, vol. 124, no. 4, pp. 2406–2420, October 2008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [29].Kreider W et al. , “Characterization of a Multi-Element Clinical HIFU System Using Acoustic Holography and Nonlinear Modeling,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 60, no. 8, pp. 1683–1698, August 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [30].Howard SM, “Calibration of Reflectance-Based Fiber-Optic Hydrophones,” (in English), 2016 IEEE Intl. Ultrasonics Symp., 2016. [Google Scholar]
- [31].Parsons JE, Cain CA, and Fowlkes JB, “Cost-effective assembly of a basic fiber-optic hydrophone for measurement of high-amplitude therapeutic ultrasound fields,” (in English), J. Acoust. Soc. Am, vol. 119, no. 3, pp. 1432–1440, March 2006. [DOI] [PubMed] [Google Scholar]
- [32].Haller J, Jenderka KV, Durando G, and Shaw A, “A comparative evaluation of three hydrophones and a numerical model in high intensity focused ultrasound fields,” (in English), J. Acoust. Soc. Am, vol. 131, no. 2, pp. 1121–1130, February 2012. [DOI] [PubMed] [Google Scholar]
- [33].Umchid S et al. , “Development of calibration techniques for ultrasonic hydrophone probes in the frequency range from 1 to 100 MHz,” (in English), Ultrasonics, vol. 49, no. 3, pp. 306–311, March 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [34].Jing Y, Cannata J, and Wang T, “Experimental verification of transient nonlinear acoustical holography,” J Acoust Soc Am, vol. 133, no. 5, pp. 2533–40, May 2013. [DOI] [PubMed] [Google Scholar]
- [35].Vlaisavljevich E et al. , “Effects of ultrasound frequency and tissue stiffness on the histotripsy intrinsic threshold for cavitation,” Ultrasound Med Biol, vol. 41, no. 6, pp. 1651–67, June 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Maxwell AD et al. , “Fragmentation of urinary calculi in vitro by burst wave lithotripsy,” J Urol, vol. 193, no. 1, pp. 338–44, January 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [37].Kothapalli S et al. , “Acoustic field characterization of a clinical magnetic resonance-guided high-intensity focused ultrasound system inside the magnet bore,” Med Phys, vol. 44, no. 9, pp. 4890–4899, September 2017. [DOI] [PubMed] [Google Scholar]
- [38].Zanelli CI and Howard SM, “A robust hydrophone for HIFU metrology,” (in English), Therapeutic Ultrasound, vol. 829, pp. 618-+, 2006. [Google Scholar]
- [39].Bessonova OV and Wilkens V, “Membrane hydrophone measurement and numerical simulation of HIFU fields up to developed shock regimes,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 60, no. 2, pp. 290–300, February 2013. [DOI] [PubMed] [Google Scholar]
- [40].Soneson JE and Myers MR, “Gaussian representation of high-intensity focused ultrasound beams,” J Acoust Soc Am, vol. 122, no. 5, pp. 2526–31, November 2007. [DOI] [PubMed] [Google Scholar]
- [41].Soneson JE, “A user-friendly software package for HIFU simulation,” AIP Conference Proceedings, vol. 1113, pp. 165–169, 2009. [Google Scholar]
- [42].Rosnitskiy PB, Yuldashev PV, Vysokanov BA, and Khokhlova VA, “Setting Boundary Conditions on the Khokhlov–Zabolotskaya Equation for Modeling Ultrasound Fields Generated by Strongly Focused Transducers,” Acoustical Physics, vol. 62, no. 2, pp. 151–159, 2016. [Google Scholar]
- [43].Soneson E, “Extending the Utility of the Parabolic Approximation in Medical Ultrasound Using Wide-Angle Diffraction Modeling,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 64, no. 4, pp. 679–687, April 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [44].Du G and Breazeale MA, “Harmonic distortion of a finite amplitude Gaussian beam in a fluid,” J Acoust Soc Am, vol. 80, no. 1, pp. 212–216, 1986. [Google Scholar]
- [45].Reilly CR and Parker KJ, “Finite-amplitude effects on ultrasound beam patterns in attenuating media,” J Acoust Soc Am, vol. 86, no. 6, pp. 2339–48, December 1989. [DOI] [PubMed] [Google Scholar]
- [46].Averkiou MA and Hamilton MF, “Measurements of harmonic generation in a focused finite-amplitude sound beam,” J Acoust Soc Am, vol. 98, no. 6, pp. 3439–42, December 1995. [DOI] [PubMed] [Google Scholar]
- [47].Ward B, Baker AC, and Humphrey VF, “Nonlinear propagation applied to the improvement of resolution in diagnostic medical ultrasound,” J Acoust Soc Am, vol. 101, no. 1, pp. 143–54, January 1997. [DOI] [PubMed] [Google Scholar]
- [48].Hamilton MF and Blackstock DT, Nonlinear Acoustics. San Diego, CA: Academic Press, 1998. [Google Scholar]
- [49].Cherin EW, Poulsen JK, van der Steen AF, Lum P, and Foster FS, “Experimental characterization of fundamental and second harmonic beams for a high-frequency ultrasound transducer,” Ultrasound Med Biol, vol. 28, no. 5, pp. 635–46, May 2002. [DOI] [PubMed] [Google Scholar]
- [50].Wear KA, “Considerations for choosing sensitive element size for needle and fiber-optic hydrophones I: Theory and graphical guide,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 66, no. 2, pp. 318–339, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].Wear KA and Liu Y, “Considerations for choosing sensitive element size for needle and fiber-optic hydrophones II: Experimental validation of spatial averaging model,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 66, no. 2, pp. 340–347, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].Liu Y, Wear KA, and Harris GR, “Variation of High-Intensity Therapeutic Ultrasound (HITU) Pressure Field Characterization: Effects of Hydrophone Choice, Nonlinearity, Spatial Averaging and Complex Deconvolution,” Ultrasound Med Biol, vol. 43, no. 10, pp. 2329–2342, October 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [53].Lewin PA, Mu C, Umchid S, Daryoush A, and El-Sherif M, “Acousto-optic, point receiver hydrophone probe for operation up to 100 MHz,” (in English), Ultrasonics, vol. 43, no. 10, pp. 815–821, December 2005. [DOI] [PubMed] [Google Scholar]
- [54].Wilkens V and Koch C, “Amplitude and phase calibration of hydrophones up to 70 MHz using broadband pulse excitation and an optical reference hydrophone,” (in English), J. Acoust. Soc. Am, vol. 115, no. 6, pp. 2892–2903, June 2004. [Google Scholar]
- [55].Hurrell A, “Voltage to Pressure Conversion: Are you getting “phased” by the problem,” (in English), AMUM 2004: Advanced Metrology for Ultrasound in Medicine 2004, vol. 1, pp. 57–62, 2004. [Google Scholar]
- [56].Wear KA, Gammell PM, Maruvada S, Liu Y, and Harris GR, “Improved measurement of acoustic output using complex deconvolution of hydrophone sensitivity,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 61, no. 1, pp. 62–75, January 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [57].Wear KA, Liu YB, Gammell PM, Maruvada S, and Harris GR, “Correction for Frequency-Dependent Hydrophone Response to Nonlinear Pressure Waves Using Complex Deconvolution and Rarefactional Filtering: Application With Fiber Optic Hydrophones,” (in English), IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 62, no. 1, pp. 152–164, January 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [58].Eichstadt S and Wilkens V, “GUM2DFT-a software tool for uncertainty evaluation of transient signals in the frequency domain,” (in English),Meas. Sci. Tech, vol. 27, no. 5, May 2016. [Google Scholar]
- [59].Eichstadt S, Wilkens V, Dienstfrey A, Hale P, Hughes B, and Jarvis C, “On challenges in the uncertainty evaluation for time-dependent measurements,” (in English), Metrologia, vol. 53, no. 4, pp. S125–S135, August 2016. [Google Scholar]
- [60].Hurrell AM and Rajagopal S, “The Practicalities of Obtaining and Using Hydrophone Calibration Data to Derive Pressure Waveforms,” (in English), IEEE Trans. Ultrason., Ferroelectr. Freq. Control, vol. 64, no. 1, pp. 126–140, January 2017. [DOI] [PubMed] [Google Scholar]
- [61].Martin E and Treeby B, “Investigation of the repeatability and reproducibility of hydrophone measurements of medical ultrasound fields,” J. Acoust. Soc. Am, vol. 145, no. 3, pp. 1270–1282, 2019. [DOI] [PubMed] [Google Scholar]
- [62].Preston RC, Bacon DR, and Smith RA, “Calibration of medical ultrasonic equipment-procedures and accuracy assessment,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 35, no. 2, pp. 110–21, 1988. [DOI] [PubMed] [Google Scholar]
- [63].Zeqiri B and Bond AD, “The Influence of Wave-Form Distortion on Hydrophone Spatial-Averaging Corrections - Theory and Measurement,” (in English), J. Acoust. Soc. Am, vol. 92, no. 4, pp. 1809–1821, October 1992. [Google Scholar]
- [64].Radulescu EG, Lewin PA, Goldstein A, and Nowicki A, “Hydrophone spatial averaging corrections from 1 to 40 MHz,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 48, no. 6, pp. 1575–1580, November 2001. [DOI] [PubMed] [Google Scholar]
- [65].Radulescu EG, Lewin PA, and Nowicki A, “1-60 MHz measurements in focused acoustic fields using spatial averaging corrections,” Ultrasonics, vol. 40, no. 1-8, pp. 497–501, May 2002. [DOI] [PubMed] [Google Scholar]
- [66].Cooling MP, Humphrey VF, and Wilkens V, “Hydrophone area-averaging correction factors in nonlinearly generated ultrasonic beams,” J. Phys. Conf. Series, Adv. Metro for Ultrasound in Med, vol. 279, pp. 1–6, 2011. [Google Scholar]
- [67].Wear KA, Baker C, and Miloro P, “Directivity and Frequency-Dependent Effective Sensitive Element Size of Needle Hydrophones: Predictions from Four Theoretical Forms Compared with Measurements,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 65, no. 10, pp. 1781–1788, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [68].Wear KA and Howard SM, “Directivity and frequency-dependent effective sensitive element size of a reflectance-based fiber optic hydrophone: predictions from theoretical models compared with measurements,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 65, no. 12, pp. 2343–2348, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [69].Lucas BG and Muir TG, “The field of a focusing source,” J. Acoust. Soc. Am, vol. 72, no. 4, pp. 1289–1296, 1982. [Google Scholar]
- [70].Bacon DR, “Finite-Amplitude Distortion of the Pulsed Fields Used in Diagnostic Ultrasound,” (in English), Ultrasound in Medicine and Biology, vol. 10, no. 2, pp. 189–195, 1984. [DOI] [PubMed] [Google Scholar]
- [71].Duncan T, Humphrey VF, and Duck FA, “A numerical comparison of nonlinear indicators for diagnostic ultrasound fields,” Proceedings World Congress of Ultrasonics, pp. 85–88, 2003. [Google Scholar]
- [72].Duck FA, “Estimating in situ exposure in the presence of acoustic nonlinearity,” J Ultrasound Med, vol. 18, no. 1, pp. 43–53, January 1999. [DOI] [PubMed] [Google Scholar]
- [73].Jones DS, “The Scattering of a Scalar Wave by a Semi-Infinite Rod of Circular Cross Section,” (in English), Philosophical Transactions of the Royal Society of London Series a-Mathematical and Physical Sciences, vol. 247, no. 934, pp. 499–528, 1955. [Google Scholar]
- [74].Krucker JF, Eisenberg A, Krix M, Lotsch R, Pessel M, and Trier HG, “Rigid piston approximation for computing the transfer function and angular response of a fiber-optic hydrophone,” (in English), J. Acoust. Soc. Am, vol. 107, no. 4, pp. 1994–2003, April 2000. [DOI] [PubMed] [Google Scholar]
- [75].Harris GR and Gammell PM, “Sensitivity measurements of piezoelectric polymer hydrophones from 0.2-2 MHz using a broadband-pulse technique,” (in English), Journal of the Acoustical Society of America, vol. 105, no. 2, pp. 725–731, February 1999. [Google Scholar]
- [76].Wear KA, Liu Y, and Harris GR, “Pressure pulse distortion by needle and fiber-optic hydrophones due to nonuniform sensitivity,” IEEE Trans Ultrason Ferroelectr. Freq Control, vol. 65, no. 2, pp. 137–148, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [77].Wang ZQ, Lauxmann P, Wurster C, Kohler M, Gompf B, and Eisenmenger W, “Impulse response of a fiber optic probe hydrophone determined with shock waves in water,” (in English), J. Appl. Phys, vol. 85, no. 5, pp. 2514–2516, March 1 1999. [Google Scholar]
- [78].Arvengas A, Davitt K, and Caupin F, “Fiber optic probe hydrophone for the study of acoustic cavitation in water,” (in English), Rev Sci Instrum, vol. 82, no. 3, March 2011. [DOI] [PubMed] [Google Scholar]
- [79].Haller J, Wilkens V, Jenderka KV, and Koch C, “Characterization of a fiber-optic displacement sensor for measurements in high-intensity focused ultrasound fields,” (in English), J. Acoust. Soc. Am, vol. 129, no. 6, pp. 3676–3681, June 2011. [DOI] [PubMed] [Google Scholar]
- [80].Weise W, Wilkens V, and Koch C, “Frequency response of fiber-optic multilayer hydrophones: Experimental investigation and finite element simulation,” (in English), IEEE Trans. Ultrason. Ferroelectr. Freq. Control, vol. 49, no. 7, pp. 937–946, July 2002. [DOI] [PubMed] [Google Scholar]
- [81].Cox BT, Zhang EZ, Laufer JG, and Beard PC, “Fabry Perot polymer film fibre-optic hydrophones and arrays for ultrasound field characterisation,” (in English), AMUM 2004: Advanced Metrology for Ultrasound in Medicine 2004, vol. 1, pp. 32–37, 2004. [Google Scholar]
- [82].Morris P, Hurrell A, Shaw A, Zhang E, and Beard P, “A Fabry-Perot fiber-optic ultrasonic hydrophone for the simultaneous measurement of temperature and acoustic pressure,” (in English), J. Acoust. Soc. Am, vol. 125, no. 6, pp. 3611–3622, June 2009. [DOI] [PubMed] [Google Scholar]
- [83].Kim KS, Mizuno Y, and Nakamura K, “Fiber-optic ultrasonic hydrophone using short Fabry-Perot cavity with multilayer reflectors deposited on small stub,” (in English), Ultrasonics, vol. 54, no. 4, pp. 1047–1051, April 2014. [DOI] [PubMed] [Google Scholar]
- [84].Martin E, Zhang EZ, Guggenheim JA, Beard PC, and Treeby BE, “Rapid Spatial Mapping of Focused Ultrasound Fields Using a Planar Fabry-Perot Sensor,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 64, no. 11, pp. 1711–1722, November 2017. [DOI] [PubMed] [Google Scholar]
- [85].Wilkens V, “Characterization of an optical multilayer hydrophone with constant frequency response in the range from 1 to 75 MHz,” (in English), Journal of the Acoustical Society of America, vol. 113, no. 3, pp. 1431–1438, March 2003. [DOI] [PubMed] [Google Scholar]
- [86].Vannacci E, Granchi S, Belsito L, Roncaglia A, and Biagi E, “Wide bandwidth fiber-optic ultrasound probe in MOMS technology: Preliminary signal processing results,” (in English), Ultrasonics, vol. 75, pp. 164–173, March 2017. [DOI] [PubMed] [Google Scholar]
- [87].Guan BO, Tam HY, Lau ST, and Chan HLW, “Ultrasonic hydrophone based on distributed Bragg reflector fiber laser,” (in English), Ieee Photonics Technology Letters, vol. 17, no. 1, pp. 169–171, January 2005. [Google Scholar]
- [88].Tseng YH and Wang JS, “Single-crystalline tellurite optical fiber hydrophone,” Opt Lett, vol. 41, no. 5, pp. 970–3, March 1 2016. [DOI] [PubMed] [Google Scholar]
- [89].Rosnitskiy PB et al. , “Design of HIFU Transducers for Generating Specified Nonlinear Ultrasound Fields,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 64, no. 2, pp. 374–390, February 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [90].Sapozhnikov OA, Pishchal’nilov YA, and Morozov AV, “Reconstruction of the Normal Velocity Distribution on the Surface of an Ultrasonic Transducer from the Acoustic Pressure Measured on a Reference Surface,” Acoustical Physics, vol. 49, no. 3, pp. 416–424, 2003. [Google Scholar]
- [91].Sapozhnikov OA, Tsysar SA, Khokhlova VA, and Kreider W, “Acoustic holography as a metrological tool for characterizing medical ultrasound sources and fields,” J Acoust Soc Am, vol. 138, no. 3, pp. 1515–32, September 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [92].Bigelow TA and O’Brien WD, “Experimental evaluation of indicators of nonlinearity for use in ultrasound transducer characterizations,” (in English), Ultrasound in Medicine and Biology, vol. 28, no. 11-12, pp. 1509–1520, Nov-Dec 2002. [DOI] [PubMed] [Google Scholar]
- [93].Oppenheim AV and Schafer RW, Digital Signal Processing. Englewood Cliffs, NJ: Prentice-Hall, Inc., 1975. [Google Scholar]
- [94].Eichstadt S and Wilkens V, “Evaluation of uncertainty for regularized deconvolution: A case study in hydrophone measurements,” (in English), J. Acoust. Soc. Am, vol. 141, no. 6, pp. 4155–4167, June 2017. [DOI] [PubMed] [Google Scholar]
- [95].IEC 62127-1 Ultrasonics – Hydrophones – Part 1: Measurement and characterization of medical ultrasonic fields up to 40 MHz, 2013.
- [96].Beissner K, “Maximum hydrophone size in ultrasonic field measurements,” Acustica, vol. 59, pp. 61–66, 1985. [Google Scholar]
- [97].Smith RA, “Are hydrophones of diameter 0.5 mm small enough to characterise diagnostic ultrasound equipment?,” Phys Med Biol, vol. 34, no. 11, pp. 1593–607, November 1989. [DOI] [PubMed] [Google Scholar]
- [98].Boutkedjirt T and Reibold R, “Improvement of the lateral resolution of finite-size hydrophones by deconvolution,” Ultrasonics, vol. 38, no. 1-8, pp. 745–8, March 2000. [DOI] [PubMed] [Google Scholar]
- [99].Boutkedjirt T and Reibold R, “Reconstruction of ultrasonic fields by deconvolving the hydrophone aperture effects. I. Theory and simulation,” Ultrasonics, vol. 39, no. 9, pp. 631–9, August 2002. [DOI] [PubMed] [Google Scholar]
- [100].Boutkedjirt T and Reibold R, “Reconstruction of ultrasonic fields by deconvolving the hydrophone aperture effects. II. Experiment,” Ultrasonics, vol. 39, no. 9, pp. 641–8, August 2002. [DOI] [PubMed] [Google Scholar]
- [101].Radulescu EG, Lewin PA, Nowicki A, and Berger WA, “Hydrophones’ effective diameter measurements as a quasi-continuous function of frequency,” (in English), Ultrasonics, vol. 41, no. 8, pp. 635–641, November 2003. [DOI] [PubMed] [Google Scholar]
- [102].Esward TJ and Robinson SP, “Extending the frequency range of the National Physical Laboratory primary standard laser interferometer for hydrophone calibrations to 80 MHz,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. 46, no. 3, pp. 737–44, 1999. [DOI] [PubMed] [Google Scholar]
- [103].Selfridge A and Lewin PA, “Wideband spherically focused PVDF acoustic sources for calibration of ultrasound hydrophone probes,” (in English), IEEE Trans Ultrason Ferroelectr Freq Control, vol. 47, no. 6, pp. 1372–1376, November 2000. [DOI] [PubMed] [Google Scholar]
- [104].Radulescu EG, Lewin PA, Wojcik J, and Nowicki A, “Calibration of ultrasonic hydrophone probes up to 100 MHz using time gating frequency analysis and finite amplitude waves,” (in English), Ultrasonics, vol. 41, no. 4, pp. 247–254, June 2003. [DOI] [PubMed] [Google Scholar]
- [105].Bloomfield PE and Lewin PA, “Determination of Ultrasound Hydrophone Phase From Fourier-Hilbert Transformed 1 to 40 MHz Time Delay Spectrometry Amplitude,” (in English), IEEE Trans. Ultrason. Ferroelectr., Freq. Control, vol. 61, no. 4, pp. 662–672, April 2014. [Google Scholar]
- [106].Weber M and Wilkens V, “Using a heterodyne vibrometer in combination with pulse excitation for primary calibration of ultrasonic hydrophones in amplitude and phase,” (in English), Metrologia, vol. 54, no. 4, pp. 432–444, August 2017. [Google Scholar]
- [107].Wear KA, Baker C, and Miloro P, “Directivity and Frequency-Dependent Effective Sensitive Element Size of Membrane Hydrophones: Theory vs. Experiment,” IEEE Trans Ultrason Ferroelectr Freq Control, vol. submitted, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]








