Abstract
Multispectral photoacoustic imaging (MPAI) is a promising emerging diagnostic technology, but fluence artifacts can degrade device performance. Our goal was to develop well-validated phantom-based test methods for evaluating and comparing MPAI fluence correction algorithms, including a heuristic diffusion approximation, Monte Carlo simulations, and an algorithm we developed based on novel application of the diffusion dipole model (DDM). Phantoms simulated a range of breast-mimicking optical properties and contained channels filled with chromophore solutions (ink, hemoglobin, or copper sulfate) or connected to a previously developed blood flow circuit providing tunable oxygen saturation (SO2). The DDM algorithm achieved similar spectral recovery and SO2 measurement accuracy to Monte Carlo-based corrections with lower computational cost, potentially providing an accurate, real-time correction approach. Algorithms were sensitive to optical property uncertainty, but error was minimized by matching phantom albedo. The developed test methods may provide a foundation for standardized assessment of MPAI fluence correction algorithm performance.
Keywords: spectral coloring, oximetry, oxygen saturation, tissue phantoms, diffusion theory, Monte Carlo
1. Introduction
Multispectral Photoacoustic Imaging (MPAI) is an emerging hybrid biomedical imaging modality that combines the high functional contrast of optics with the deep penetration of ultrasound [1]. By employing multiple optical wavelengths, absorption spectra can be measured, enabling mapping of chromophore concentrations including oxyhemoglobin (HbO2) and deoxyhemoglobin (Hb) as well as contrast agents (e.g. dyes, nanoparticles) [2]. Additionally, HbO2 and Hb can be used to compute blood oxygen saturation (SO2) maps. Many preclinical and clinical applications of MPAI have been reported, including tissue oximetry [[3], [4], [5]], cancer detection and diagnosis [[6], [7], [8], [9]], cerebrovascular imaging [[10], [11], [12]], surgical guidance [13], and tumor margining [14].
One critical technical challenge in MPAI is the corruption of measured absorption spectra due to “spectral coloring” artifacts caused by both spatial and spectral variations in local fluence within the tissue [[15], [16], [17]]. Photoacoustic signal is proportional to absorption coefficient as well as local fluence as evident from the well-known photoacoustic equation:
(1) |
where is photoacoustic pressure amplitude, is 3D spatial position, is wavelength, is absorption coefficient, is reduced scattering coefficient, and is local fluence. Fluence generally decreases with depth in tissue and thus degrades image uniformity, but spectral variations in tissue optical properties also cause significant wavelength-dependent differences in fluence distribution that corrupt measured spectra. Absorption spectra are generally assumed to be linear combinations of signals from several chromophores, and spectral unmixing algorithms are used to solve for chromophore concentrations at each pixel, often using least-squares methods [18,19]. Clearly, if measured MPAI spectra differ from target absorption spectra (a performance characteristic we denote here as spectral recovery), MPAI measurement accuracy of spectrum-derived biomarkers such as SO2 or contrast agent concentration may be significantly degraded. Fluence correction can be performed on photoacoustic images by dividing by , thus yielding a more accurate estimate of . As subsurface fluence is difficult to measure, computational modeling represents an appealing technique for estimating fluence distributions.
Numerous methods of modeling light propagation in tissue have been proposed for performing fluence correction in MPAI systems. Because the Radiative Transfer Equation (RTE) is difficult to solve analytically, a common alternative is to use the diffusion approximation (equivalent to Pn approximation, where n = 1, 2 or 3 depending on truncation of spherical harmonic expansion) when the assumptions of isotropic photon scattering and time invariance are valid [20,21]. The P1 approximation is widely used in biophotonics, offering reasonable accuracy in the diffusive regime, and the delta-P1 approximation can further improve accuracy in the ballistic regime [[22], [23], [24]]. Both P1 and delta-P1 approximations can be solved numerically (e.g., using finite element methods) to obtain fluence, which is especially useful for studying complex heterogeneous media but carries high computational cost [[25], [26], [27]]. Analytical solutions of these models are possible for homogeneous media [24,28], and one-dimensional solutions of the P1 approximation have been used for MPAI fluence correction due to their simple implementation [29,30]. However, these models can potentially be inaccurate for MPAI if assumptions regarding source geometry are not satisfied.
An alternative approximation is the diffusion dipole model (DDM), which simulates the fluence distribution produced by a collimated pencil beam in a semi-infinite turbid medium as the response to a pair of point sources, with a source at a particular tissue depth and an imaginary source above the tissue surface [31]. This model is well studied and yields good accuracy far from boundaries and sources, but to our knowledge has not been previously applied as an MPAI fluence correction technique. Since a large, finite-area beam may be approximated as a sum of pencil beams, the DDM may be considered a Green’s function or impulse response that can be convolved with the incident beam geometry to compute the fluence produced by large beams used in MPAI systems. Another advantage is that DDM can easily produce fluence values at arbitrary points, most notably in planes offset from the beam center, which is relevant to clinical MPAI devices that use offset illumination-detection geometries [32]. Monte Carlo (MC) simulation has also been applied for photoacoustic fluence corrections [[33], [34], [35]], and may be considered the ‘gold standard’ method for simulating photon propagation and computing fluence in turbid media as it offers high accuracy in both ballistic and diffusive regimes and supports arbitrary tissue optical properties and geometries. However, MC-based fluence corrections can carry high computational cost. MC runtimes can be reduced through various methods such as GPU parallelization and scaling methods [36,37], although it is unclear whether such methods could be made fast enough to enable real-time or adaptive MPAI fluence corrections.
While these models can estimate fluence distributions in tissue, a priori knowledge of tissue optical properties and their spatial distribution is required, which is difficult to ascertain for real tissue. MPAI can be combined with independent tissue optical property measurements fed back into fluence models, such as diffuse optical tomography [38,39] or acousto-optics [40,41], but incorporating these methods increases system cost. Development of fluence correction methods that reduce or eliminate a priori specification of tissue optical properties remains an active research area and has included use of multiple illumination geometries [42,43], eigenspectral analysis [44], machine-learning based fluence quantification [45], monitoring dynamic changes in target absorption [46], differential contrast agent methods [47], and SNR-regularization [48]. Additionally, several models have been used in iterative minimization schemes to move towards adaptive fluence corrections that do not require a priori information, including finite element models [49], direct solutions of the RTE [50], Bayesian approaches [51,52], and adjoint radiance MC [53,54].
Fluence correction algorithms are a key design aspect of MPAI devices, and the wide variation in approaches implies potentially significant differences in performance. Fluence model accuracy, appropriateness for particular device illumination/detection geometries, and robustness to uncertainty in optical properties are not well understood. The availability of well-validated test methods for assessing algorithm performance would greatly facilitate MPAI device development and optimization, and as a result, improve diagnostic interpretation of preclinical and clinical MPAI data and support regulatory evaluation. Our group has previously developed test methods for image quality evaluation and SO2 measurement accuracy [[55], [56], [57]], but fluence correction performance was not considered. The purpose of this work was to develop best practices for evaluating and comparing performance of fluence correction algorithms. Towards this overall goal, our study objectives were to (1) develop and validate a novel, computationally-efficient DDM-based fluence correction, (2) evaluate spectral recovery and SO2 measurement accuracy of several fluence correction algorithms in tissue phantoms, and (3) evaluate and compare fluence correction robustness to uncertainty in tissue optical properties.
2. Methods
2.1. Fluence Correction Models
In this study, we developed and compared performance of three fluence correction algorithms: (1) a simple, 1D diffusion approximation (1D-DA), (2) a novel approach based on the diffusion-dipole model (DDM), and (3) Monte Carlo simulations including modeling of the ultrasound transducer in tissue contact (MC-UST). These three methods illustrate algorithms of varying complexity, ease of implementation, and computational cost. The 1D-DA algorithm may be considered a fast but very simple, even heuristic model, the DDM algorithm represents a more complex closed-form model tailored to our specific MPAI system configuration, and the MC-UST algorithm represents a highly detailed approach based on a gold-standard method.
2.1.1. 1D Homogeneous Diffusion Approximation (1D-DA)
For an infinite uniform planar beam in a semi-infinite homogeneous medium, assuming conservation of energy at the tissue surface and that fluence vanishes far from sources, the delta-P1 or delta-Eddington approximation can be used to solve for fluence as a function of depth, :
(2) |
(3) |
where is incident radiant exposure, is surface specular reflectivity, is scattering coefficient, is anisotropy factor, is effective attenuation coefficient, and and are lumped parameters computed from tissue optical properties and surface reflectance conditions [24]. Fluence dependence on wavelength is left implicit here and in subsequent sections for clarity. The first term in brackets in equation 2 describes the collimated fluence while the second term describes diffuse fluence. The collimated term decays rapidly with depth such that its contribution to is less than 0.1% for depths greater than 4 mm, assuming typical tissue optical properties. Thus, a simple, heuristic one-dimensional diffusion approximation (1D-DA) can be derived as reported by others [15,29]:
(4) |
where is the incident radiant exposure after specular reflection losses. While this heuristic model is quite simple and accurately models fluence directly beneath a large beam, its underlying assumptions are likely not well suited to predicting fluence distributions for our specific MPAI system configuration, which uses an offset beam-transducer geometry (see Section 2.2). However, a device designer could choose to use this approach due to its ease of implementation and very low computational cost. This algorithm is also fast enough to support real-time computations for adaptive or manually adjustable fluence compensation.
2.1.2. Diffusion Dipole Model (DDM)
In this section, we briefly summarize our implementation of the DDM and refer the reader to Wang and Wu’s full description of this model [31]. Assuming a semi-infinite medium and using an extrapolated boundary condition [58], the fluence produced by a pencil beam, , can be written in terms of fluence due to a point source within the tissue, , and an imaginary point source above the tissue, [28,31]:
(6) |
(7) |
(8) |
where and are the distances from the real point source and the imaginary point source to the observation point , is the transport albedo, is the diffusion coefficient, and is the location of extrapolated boundaries where fluence equals zero:
where is the ratio of the tissue refractive index to the top medium (e.g., air). We set the real point source position at .
Rather than pencil beams or point sources, however, most MPAI systems use large beams on the order of centimeters in length or diameter, including our custom system (see Section 2.2.1). In our approach, we convolve the above Green’s function with a uniform, normally-incident elliptical beam with minor radius , major radius , and radiant exposure [59] (Fig. 1):
(9) |
(10) |
where describes the boundary of the elliptical beam. Because is zero outside the ellipse, the integral bounds for equation (4) can be chosen as and . The integral (equation 9) was computed using a 0.2 mm step size for and , as convergence analysis yielded maximum residual error of < 1.3% near point sources between step sizes of 0.2 mm and 0.1 mm (data not shown). By fixing equal to the elevational offset between the centers of the optical beam and ultrasound transducer and choosing and corresponding to MPAI pixel coordinates, the required 2D fluence correction map can be generated (Fig. 1). DDM fluence outputs were verified against MC simulations with = 0.1 cm−1, = 10 cm−1, g = 0.9, and n = 1.4, for four beam geometries with spatially uniform energy: 1) a 0.1 mm diameter point source, 2) a 4 cm diameter circular beam, 3) an elliptical beam with = 0.5 cm, = 2 cm, and 4) a 0.1 mm x 4 cm line source. Fluence error was spatially averaged over three regions in depth: shallow zone (0-0.5 cm), middle zone (0.5-3.0 cm), and deep zone (3.0-4.5 cm). Additional verification was performed for the case of a uniform elliptical beam ( = 1.75 cm, = 0.25 cm) with various sets of tissue optical properties, specifically all combinations of 0.02, 0.05, 0.1, or 0.2 cm−1, 5, 10, 15, or 20 cm−1, and 1.4, 1.45, or 1.5. All MC simulations used for DDM verification assumed a normal incidence beam angle, matching the assumed condition of DDM and thus providing a true reference solution for comparison.
To investigate the potential to further reduce computational cost, we evaluated a one-dimensional DDM approximation (1D-DDM), where fluence vs. depth was computed at = 9 mm, = 0 mm (Fig. 1) and this 1D fluence distribution in depth was applied as a fluence correction to each column of the image data. 1D-DDM outputs were also compared against MC simulations.
2.1.3. MC Simulations
A previously validated 3D voxel-based MC model was used to verify DDM algorithm accuracy and model the presence of the ultrasound transducer (denoted as MC-UST corrections) [60]. Based on prior convergence analysis, 8 × 107 photons were launched into a 300 × 300 × 300 cubic grid with voxel dimensions of 0.02 × 0.02 x 0.02 cm. Unlike MC simulations used for DDM verification, MC-UST simulations assumed a 45° beam angle matching our MPAI system used in phantom experiments. MC-UST models used for fluence correction assumed an air boundary at the tissue surface ( = 1.0), except for a rectangular area corresponding to the transducer face (12 mm x 40 mm) in contact with the tissue and positioned at a 9 mm offset from a 5 × 35 mm elliptical beam, mimicking our imaging configuration (Fig. 1). As our system’s transducer was covered with a gel-coupled aluminum foil layer to reduce surface clutter, we modeled photon interactions with the transducer using the Fresnel equations and a wavelength-dependent complex refractive index of aluminum [61]. Non-reflected photons are considered absorbed by the foil. Accurately modeling the transducer is a key consideration as transducer reflections have been shown to alter the fluence distribution in tissue [62] and transducer optical absorption can produce undesirable image clutter [32,56].
2.2. Phantom Experiments
2.2.1. Custom PAI System
Phantom imaging experiments were performed using a custom MPAI system described previously [57]. In brief, the system uses a near-infrared optical parametric oscillator (OPO) delivering 5 ns pulses at a 10 Hz pulse repetition rate over a tunable wavelength range of 690-950 nm (Phocus Mobile, Opotek, Inc., Carlsbad, CA) and a 128-channel ultrasound acquisition system (Vantage 128, Verasonics, Inc., Kirkland, WA) with a 128-element, 7 MHz linear array transducer (L11-4v, Verasonics). Light is coupled into a fiber bundle and engineered diffuser to produce a 5 mm x 35 mm uniform elliptical beam positioned with a 9 mm offset relative to the transducer’s elevational center (i.e., 9 mm from the image plane). The transducer was gel-coupled with aluminum foil to reduce image clutter due to light absorption at the transducer surface. Multispectral imaging scans were performed from 700 to 898 nm in 2 nm steps, acquiring ten scans for averaging purposes, with wavelength-dependent radiant exposure of 8-10 mJ/cm2. Photoacoustic images were reconstructed in real-time using Verasonics’ proprietary pixel-based algorithm. MATLAB (Mathworks, Inc., Natick, MA) was used for all data acquisition and post-processing. MPAI data was fluence corrected using 1D-DA, 1D-DDM, or MC-UST algorithms, then least squares spectral unmixing using the pseudo-inverse approach [63] was applied to compute SO2 per pixel. Images acquired at eleven wavelengths were used for unmixing calculations (700, 720, 740, 760, 780, 800, 820, 840, 860, 880, 898 nm).
2.2.2. Spectral Recovery
Phantom experiments were used to validate fluence models and evaluate spectral recovery using different fluence correction algorithms. Phantoms were constructed from acrylic molds with polytetrafluoroethylene (PTFE) tubes suspended at various positions and filled with various chromophore solutions. To characterize spectral recovery as a function of depth, a “penetration” phantom was constructed containing a diagonal array of PTFE tubes (0.56 mm inner diameter, STT-24, Component Supply Company, Fort Meade, FL) at depths of 5 to 35 mm in 5 mm steps with 5 mm horizontal spacing (Fig. 2). To improve validation as well as study robustness to uncertainty in tissue optical properties, three different Intralipid-ink solutions were used outside the tubes to represent low, average, and high optical attenuation of breast tissue at 800 nm: () = (0.02 cm−1, 5 cm−1), (0.05 cm−1, 10 cm−1), and (0.1 cm−1, 15 cm−1), respectively. Phantom channels were filled with an India ink solution (Super Black India ink, Speedball Art Products Co., LLC, Statesville, NC) with 4.5 cm−1 at 800 nm. Total transmittance and diffuse reflectance were measured over 400-1000 nm by integrating sphere spectrophotometry (Lambda 1050, PerkinElmer Inc., Waltham, MA) and the optical absorption coefficient and reduced scattering coefficient were computed using the inverse adding-doubling algorithm [64](Fig. 3).
To demonstrate adequate spectral recovery for a wide variety of target chromophore spectra, a second “chromophore” phantom was fabricated containing a row of seven tubes at a depth of 15 mm with 5 mm lateral spacing (Fig. 2). Four tubes were filled with either India ink, non-turbid hemoglobin solution (Multi-4 CO-Oximeter Control, Level 2, Instrumentation Laboratory, Inc., Bedford, MA), turbid bovine whole blood (Quad Five, Inc., Ryegate, MT), or copper sulfate (209198, Sigma-Aldrich Inc., St. Louis, MO). Solution concentrations were selected to yield absorption coefficients within 1.5 – 8 cm−1 over 700 to 900 nm, and blood was diluted to a total hemoglobin of 13.5 ± 0.1 g/dL. Reference extinction coefficient spectra of each solution were either measured by spectrophotometry or taken from the literature [65]. This phantom was filled with an Intralipid-ink solution representing the average attenuation case described above. Root mean square difference (RMSD) between the MPAI spectrum, , and reference spectrum, , over 11 wavelengths was computed for each target tube as:
(12) |
Depth-averaged RMSD was also computed for penetration phantom data by averaging RMSD over all detectable target fluid channels. To quantify measurement variability, average spectral coefficient of variation (COV) was computed on uncorrected spectra for each target tube as:
(13) |
where and are the standard deviation and mean of target intensity over ten scans, respectively, for the -th scan wavelength.
2.2.3. SO2 Measurement Accuracy and Robustness
To evaluate SO2 measurement accuracy of fluence correction models, we used our previously validated tunable blood oxygenation phantom approach [57]. Briefly, a penetration phantom was constructed with an array of tubes connected to a flow circuit that included a membrane oxygenator. By delivering adjustable concentrations of nitrogen and oxygen gas to the oxygenator, different blood SO2 levels can be achieved stably and reproducibly. Whole bovine blood was centrifuged to prepare red blood cell suspensions in phosphate buffered saline, and MPAI SO2 measurements were taken at setpoints of 99%, 80%, 60%, and 40% SO2. At each SO2 setpoint, a 0.2 mL blood sample was drawn from the circuit and measured using a clinical whole blood CO-oximeter (Avoximeter 4000, Accriva Diagnostics, Inc., San Diego, CA) to provide ground truth SO2 measurements both before and after MPAI measurements.
SO2 measurement accuracy was quantified using several performance metrics, including those described in an international performance standard for pulse oximeters [66] and in cerebral oximeter phantom test methods [67]. Sensitivity was defined as the slope of a linear fit to the SO2 data, and mean bias was computed as the mean difference between MPAI SO2 values, , and CO-oximeter SO2 values, , over SO2 setpoints as:
(14) |
RMSD is a more comprehensive metric that characterizes overall SO2 measurement performance and was computed for each target depth as:
(15) |
Again, depth-averaged was computed to summarize performance over all target depths.
To evaluate the impact of uncertainty in tissue optical properties on SO2 measurement accuracy, MPAI data acquired in a penetration phantom with () = (0.05 cm−1, 10 cm−1) and corrected using the 1D-DDM and 1D-DA algorithms were re-analyzed by varying input optical properties by = −0.03 to +0.05 cm−1 in steps of 0.01 cm−1 and = −3.0 to +5.0 cm−1 in steps of 1.0 cm−1. Depth-averaged was computed for all combinations of and .
3. Results and Discussion
3.1. DDM Verification
DDM was verified against MC simulations with four commonly used beam geometries (Fig. 4). DDM and MC fluence values were generally in good agreement, and the lowest percent error was observed in the middle zone (Table 1). Greater error in the shallow zone is expected based on the limitations of diffusion theory near boundaries and sources, while errors in the deep zone are likely due to insufficient photon penetration and resultant noise in the MC outputs. Error in the shallow zone was generally higher for point and line beams, suggesting that the DDM spatial integration step value (optimized for the elliptical beam case, which is most relevant to our system design) should be decreased for these geometries.
Table 1.
Source Geometry | Shallow Zone (0 – 0.5 cm) |
Middle Zone (0.5 – 3 cm) |
Deep Zone (3 – 4.5 cm) |
---|---|---|---|
Point ( = = 0.05 cm) |
47.1% | 5.4% | 34.5% |
Circle ( = = 2 cm) |
6.0% | 6.2% | 21.1% |
Ellipse ( = 0.5 cm, = 2 cm) |
4.5% | 4.9% | 24.5% |
Line ( = 0.05 cm, = 2 cm) |
11.8% | 5.2% | 31.9% |
DDM was also in good agreement with MC simulations for the elliptical beam case, both at beam center and at offset planes useful for fluence correction in our MPAI system (Table 2). While the DDM and MC models can compute fluence in different offset planes, the 1D-DA model only produces a single fluence depth profile, indicating a lack of flexibility for modeling offset illumination-transducer geometries. As shown in Fig. 5, as tissue optical attenuation increases, the absolute error between 1D-DA and DDM/MC fluence increases significantly in the shallow zone. Similar results were found for the condition () = (0.02 cm−1, 15 cm−1), which corresponds to the lowest absorption and highest reduced scattering coefficients used in verification studies. For a beam offset of 1 cm and for the high attenuation case, 1D-DA actually agreed better with DDM and MC results. However, results under other scenarios clearly indicate that this is not always the case. These results show that DDM is consistent with MC simulations and can predict fluence distributions in photoacoustic image planes offset from an elliptical laser beam. Comparisons of fluence between 1D-DDM and MC models showed that the regions of greatest error between 1D-DDM and MC fluence distributions are near the lateral edges of the field of view, with maximum errors exceeding 50% (Fig. 6). However, all but the deepest phantom targets were outside of these high-error zones as shown by the overlaid photoacoustic image in Fig. 6.
Table 2.
Y Axis |
X Axis |
|||||
---|---|---|---|---|---|---|
Shallow Zone (0 – 0.5 cm) |
Middle Zone (0.5 – 3 cm) |
Deep Zone (3 – 4.5 cm) |
Shallow Zone (0 – 0.5 cm) |
Middle Zone (0.5 – 3 cm) |
Deep Zone (3 – 4.5 cm) |
|
Percent Error (%) | 1.0 – 16.9 | 1.3 – 9.3 | 0 – 16.5 | 1.2 – 24.0 | 1.1 – 13.8 | 0 – 21.3 |
As shown in Fig. 7, the 1D-DA algorithm predicts a significantly different log-linear slope of fluence versus depth than other models, with average percent error of 37% vs. both 1D-DDM and MC-UST. Agreement between 1D-DA and other algorithms improved somewhat at a 9 mm elevational offset corresponding to our system’s image plane, but average percent error remained large (31% and 22% vs. 1D-DDM and MC-UST, respectively). This improvement should be considered coincidental because the 1D-DA algorithm is only expected to be valid beneath the beam. At the beam center, the 1D-DDM and MC-UST fluence depth profiles in close agreement except from 0-2 mm (near the boundary and source). In the offset image plane, MC-UST predicts lower fluence than 1D-DDM for shallow depths (0-1 cm). This may be due to light absorption by the aluminum foil, which was included in the MC-UST model but not the 1D-DDM model.
3.2. Spectral Recovery
Mean-normalized, fluence-corrected photoacoustic spectra acquired using each fluence correction algorithm were compared with the raw, uncorrected spectra (Fig. 8). Spectral RMSD and COV (equations 12 and 13) were analyzed and compared. All fluence correction methods improved recovery of the target spectrum, although spectral recovery generally decreased with depth. For different phantom optical property cases, the number of visible targets decreases with increasing attenuation as expected, but RMSD per target was comparable across algorithms provided measurements presented low enough COV. Qualitatively, the three fluence corrections provided comparable performance for all phantom optical property cases, but depth-averaged RMSD shows some slight difference between 1D-DA, 1D-DDM and MC-UST. The low attenuation case had the greatest depth-averaged RMSD since all seven targets were detectable and included in RMSD computations. Algorithm performance ranking in terms of RMSD was not consistent over all optical property cases, although the MC-UST algorithm’s RMSD was least variant with respect to phantom optical property cases (i.e., MC-UST was more robust). Small differences in RMSD between algorithms, especially for the high attenuation case, may be due to noise and may not indicate significantly different levels of performance. As shown in Fig. 9, fluence corrections accurately recovered spectra of various target chromophores. RMSD was low for all correction methods and similar across all chromophores imaged. This illustrates that fluence correction algorithm performance, in terms of spectral recovery, is generally independent of target absorption spectra presenting different magnitude and spectral features.
3.3. SO2 Measurement Accuracy and Robustness
MPAI SO2 measurement accuracy varied significantly among fluence correction methods (Fig. 10), and MPAI measurements showed negative bias relative to CO-oximetry. SO2 sensitivity decreased with depth regardless of fluence correction algorithm, and the 1D-DA algorithm provided the most improvement in sensitivity (Fig. 11). However, the 1D-DA algorithm also produced the largest mean bias, particularly for shallow targets, which is consistent with fluence profiles shown in Fig. 5. If considering sensitivity and bias results alone, it is unclear how to rank algorithms by performance. We propose that (equation 15) is a suitable metric for summarizing SO2 measurement accuracy, as applied in our previous work [57]. All algorithms improved over performing no correction, but the 1D-DA algorithm had a strong depth-dependency due to inaccurate prediction of fluence attenuation rate (Fig. 11). The 1D-DDM algorithm achieved a depth-averaged of < 4%, which is the acceptable performance threshold defined by ISO 80601-2-61:2017 for pulse oximeter measurement accuracy. 1D-DDM was slightly lower than that achieved by the MC-UST algorithm, but this small difference may not be clinically significant. However, this result indicates that the simpler 1D-DDM model can achieve similar performance as the gold-standard MC approach with lower computational cost and ease of implementation.
The requirement for a priori tissue property information is a significant limitation of many fluence correction algorithms for in vivo preclinical or clinical imaging, including those studied here. Results indicated that MPAI can achieve high SO2 measurement accuracy when the tissue optical properties are known a priori, but algorithm performance was highly sensitive to uncertainty in optical properties, with increases in up to 15% (Fig. 12). Interestingly, we observed a region of the optical property tuning space that showed minimal changes in . This zone represents near-constant albedo, as confirmed by plotting depth-averaged versus corresponding albedo for each pair of and values (Fig. 12). We also observed that our “true” optical property values measured by spectrophotometry were near this minimum, with small discrepancies attributable to measurement uncertainty. These results suggest that for models requiring a priori specification of optical properties, the design problem may be simplified to matching the tissue albedo, rather than matching both and . This observation may inform design of future iterative or adaptive fluence correction strategies by reducing dimensionality of the fitting problem. These algorithms also assume a homogeneous medium, which may not be well-matched to in vivo conditions. Variations in tissue morphology and heterogeneity of optical properties are likely to affect accuracy and robustness to uncertainty of fluence correction algorithms, especially those investigated in this study. Additionally, variations in device design (beam position and angle, transducer face optical properties) may affect algorithm performance. These effects will be addressed in future work, for instance using multi-domain computational modeling [68].
4. Conclusions
We developed a novel, fast fluence correction based on the DDM and tailored for an offset illumination-detection geometry seen in clinical MPAI devices. This model was verified against Monte Carlo simulations and validated through phantom experiments. The 1D-DDM algorithm achieved similar spectral recovery and SO2 measurement accuracy to Monte Carlo-based corrections, but the 1D-DDM method offers lower computational cost and may enable real-time or adaptive fluence corrections. Assumption of a homogeneous medium and requirements for a priori assumption or measurement of tissue optical properties remain significant limitations of fluence correction algorithms evaluated in this study. While algorithm performance was sensitive to uncertainty in tissue optical properties, error was small when the simulated albedo was close to the true phantom albedo. This observation may aid design of future real-time, adaptive fluence correction strategies. Future work will include parametric study of algorithm robustness to uncertainty in device design parameters as well as tissue properties and morphology. The performance test methods and metrics proposed in this work may facilitate standardization of best practices for evaluating fluence correction algorithm performance in MPAI devices.
Transparency document
Declaration of Competing Interest
The authors declare that there are no conflicts of interest.
Acknowledgements
The authors gratefully acknowledge funding support from FDA Office of Women’s Health, FDA Medical Countermeasures Initiative, and the ORISE fellowship program through Oak Ridge Associated Universities. We also thank Dr. Quanzeng Wang (CDRH/FDA) for assistance with Monte Carlo software as well as the FDA High-Performance Computing team. The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services.
Biographies
Xuewen Zhou received her BS degree in chemistry from Illinois Wesleyan University, and her MS degree in chemical engineering from University of Rochester in 2017. She joined FDA in 2017 as an ORISE fellow and was trained on developing phantom-based test methods for photoacoustic imaging systems. She was accepted as a PhD student by Fischell Department of Bioengineering in University of Maryland in 2018 and her research interests include biomedical imaging, photoacoustic imaging, and Brillouin microscopy.
Nima Akhlaghi received his BS and MS degrees in physical and biomedical electronics from Kyiv Polytechnic Institute, Ukraine, and Sumy State University, Ukraine, in 2009 and 2010. He received his PhD in electrical and computer engineering from George Mason University in 2017. He joined the FDA in 2017 as a visiting scientist, where he is conducting ultrasound imaging research in the Office of Science and Engineering Laboratories (OSEL). His research interests include ultrasound and photoacoustic imaging, modeling, and elastography.
Keith A. Wear received his BA degree in applied physics from the University of California, San Diego and his MS and PhD degrees in applied physics from Stanford University. He is a research physicist at the FDA. He is an associate editor of the Journal of the Acoustical Society of America, Ultrasonic Imaging, and IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control. He is a fellow of the Acoustical Society of America, the American Institute for Medical and Biological Engineering, and the American Institute of Ultrasound in Medicine.
Brian S. Garra received his MD from the University of Washington followed by residency in diagnostic radiology at the University of Utah, becoming board certified in 1980. He is a professor of radiology at the University of Vermont and at George Washington University. A fellow of the American Institute of Ultrasound in Medicine and the Society of Radiologists in Ultrasound, he does device evaluation and research in elastography and complex phantom development at the FDA. He is an associate editor of Ultrasonic Imaging and reviews for numerous journals including Radiology and Ultrasound in Medicine and Biology. He spends half his time practicing clinical radiology focusing on ultrasound and spinal MRI, including an active ultrasound outreach program in rural Peru.
T. Joshua Pfefer received his PhD in biomedical engineering from the University of Texas at Austin and trained as a research fellow at the Wellman Laboratories of Photomedicine. In 2000, he joined the FDA, where he is the leader of the Optical Diagnostic Devices Laboratory. His research focuses on elucidating light-tissue interactions of emerging technologies and developing new methods to improve and standardize testing. In 2018, he was named a fellow of SPIE for contributions to biophotonics.
William C. Vogt received his BS degree in mechanical engineering from the University of Massachusetts Amherst in 2009 and his PhD in biomedical engineering from Virginia Polytechnic Institute and State University in 2013. Since joining the FDA in 2013, he has been conducting regulatory science to develop tools and test methods for evaluating the safety and effectiveness of photoacoustic imaging devices. His research interests include photoacoustics, tissue phantoms, standardization, and biophotonic medical device characterization and evaluation.
References
- 1.Beard P. Biomedical photoacoustic imaging. Interface Focus. 2011;1(4):602–631. doi: 10.1098/rsfs.2011.0028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Weber J., Beard P.C., Bohndiek S.E. Contrast agents for molecular photoacoustic imaging. Nat. Methods. 2016;13(8):639–650. doi: 10.1038/nmeth.3929. [DOI] [PubMed] [Google Scholar]
- 3.Petrova I., Petrov Y., Esenaliev R., Deyo D., Cicenaite I., Prough D. Noninvasive monitoring of cerebral blood oxygenation in ovine superior sagittal sinus with novel multi-wavelength optoacoustic system. Opt. Express. 2009;17(9):7285–7294. doi: 10.1364/oe.17.007285. [DOI] [PubMed] [Google Scholar]
- 4.Hu S., Rao B., Maslov K., Wang L.V. Label-free photoacoustic ophthalmic angiography. Opt. Lett. 2010;35(1):1–3. doi: 10.1364/OL.35.000001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Neuschler E.I., Butler R., Young C.A., Barke L.D., Bertrand M.L., Böhm-Vélez M., Destounis S., Donlan P., Grobmyer S.R., Katzen J., Kist K.A., Lavin P.T., Makariou E.V., Parris T.M., Schilling K.J., Tucker F.L., Dogan B.E. A Pivotal Study of Optoacoustic Imaging to Diagnose Benign and Malignant Breast Masses: A New Evaluation Tool for Radiologists. Radiology. 2018;287(2):398–412. doi: 10.1148/radiol.2017172228. [DOI] [PubMed] [Google Scholar]
- 6.Mallidi S., Luke G.P., Emelianov S. Photoacoustic imaging in cancer detection, diagnosis, and treatment guidance. Trends Biotechnol. 2011;29(5):213–221. doi: 10.1016/j.tibtech.2011.01.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Heijblom M., Piras D., Xia W., van Hespen J.C.G., Klaase J.M., van den Engh F.M., van Leeuwen T.G., Steenbergen W., Manohar S. Visualizing breast cancer using the Twente photoacoustic mammoscope: What do we learn from twelve new patient measurements? Opt. Express. 2012;20(11):11582–11597. doi: 10.1364/OE.20.011582. [DOI] [PubMed] [Google Scholar]
- 8.Lin L., Hu P., Shi J., Appleton C.M., Maslov K., Li L., Zhang R., Wang L.V. Single-breath-hold photoacoustic computed tomography of the breast. Nat. Commun. 2018;9 doi: 10.1038/s41467-018-04576-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Valluru K.S., Willmann J.K. Clinical photoacoustic imaging of cancer. Ultrasonography. 2016;35(4):267–280. doi: 10.14366/usg.16035. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Burton N.C., Patel M., Morscher S., Driessen W.H., Claussen J., Beziere N., Jetzfellner T., Taruttis A., Razansky D., Bednar B., Ntziachristos V. Multispectral opto-acoustic tomography (MSOT) of the brain and glioblastoma characterization. Neuroimage. 2013;65:522–528. doi: 10.1016/j.neuroimage.2012.09.053. [DOI] [PubMed] [Google Scholar]
- 11.Xiang L., Ji L., Zhang T., Wang B., Yang J., Zhang Q., Jiang M.S., Zhou J., Carney P.R., Jiang H. Noninvasive real time tomographic imaging of epileptic foci and networks. Neuroimage. 2013;66:240–248. doi: 10.1016/j.neuroimage.2012.10.077. [DOI] [PubMed] [Google Scholar]
- 12.Yao J., Xia J., Maslov K.I., Nasiriavanaki M., Tsytsarev V., Demchenko A.V., Wang L.V. Noninvasive photoacoustic computed tomography of mouse brain metabolism in vivo. Neuroimage. 2013;64:257–266. doi: 10.1016/j.neuroimage.2012.08.054. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Gandhi N., Allard M., Kim S., Kazanzides P., Bell M.A.L. Photoacoustic-based approach to surgical guidance performed with and without a da Vinci robot. J. Biomed. Opt. 2017;22(12) [Google Scholar]
- 14.Li R., Wang P., Lan L., Lloyd F.P., Goergen C.J., Chen S., Cheng J.-X. Assessing breast tumor margin by multispectral photoacoustic tomography. Biomed. Opt. Express. 2015;6(4):1273–1281. doi: 10.1364/BOE.6.001273. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Cox B., Laufer J.G., Arridge S.R., Beard P.C. Quantitative spectroscopic photoacoustic imaging: a review. J. Biomed. Opt. 2012;17(6) doi: 10.1117/1.JBO.17.6.061202. [DOI] [PubMed] [Google Scholar]
- 16.Hu S., Wang L.V. Photoacoustic imaging and characterization of the microvasculature. J. Biomed. Opt. 2010;15(1) doi: 10.1117/1.3281673. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Xia J., Yao J., Wang L.V. Photoacoustic tomography: principles and advances. Prog. Electromagn. Res. 2014;147:1–22. doi: 10.2528/pier14032303. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Chatni M.R., Xia J., Maslov K.I., Guo Z., Wang K., Anastasio M.A., Wang L.V., Sohn R., Arbeit J.M., Zhang Y. Tumor glucose metabolism imaged in vivo in small animals with whole-body photoacoustic computed tomography. J. Biomed. Opt. 2012;17(7) doi: 10.1117/1.JBO.17.7.076012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Leavesley S.J., Britain A.L., Cichon L.K., Nikolaev V.O., Rich T.C. Assessing FRET using spectral techniques. Cytom. A. 2013;83(10):898–912. doi: 10.1002/cyto.a.22340. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Star W.M. Comparing the P3-approximation with diffusion theory and with Monte Carlo calculations of light propagation in a slab geometry; Proc. SPIE 10305; 1989. [Google Scholar]
- 21.Wright S., Schweiger M., Arridge S. Reconstruction in optical tomography using the PN approximations. Meas. Sci. Technol. 2006;18(1):79. [Google Scholar]
- 22.Laufer J., Elwell C., Delpy D., Beard P. In vitro measurements of absolute blood oxygen saturation using pulsed near-infrared photoacoustic spectroscopy: accuracy and resolution. Phys. Med. Biol. 2005;50(18):4409. doi: 10.1088/0031-9155/50/18/011. [DOI] [PubMed] [Google Scholar]
- 23.Joseph J.H., Wiscombe W.J., Weinman J.A. The delta-Eddington approximation for radiative flux transfer. J. Atmos. Sci. 1976;33(12):2452–2459. [Google Scholar]
- 24.Carp S.A., Prahl S.A., Venugopalan V. Radiative transport in the delta-P1 approximation: accuracy of fluence rate and optical penetration depth predictions in turbid semi-infinite media. J. Biomed. Opt. 2004;9(3):632–648. doi: 10.1117/1.1695412. [DOI] [PubMed] [Google Scholar]
- 25.Brochu F.M., Brunker J., Joseph J., Tomaszewski M.R., Morscher S., Bohndiek S.E. Towards quantitative evaluation of tissue absorption coefficients using light fluence correction in optoacoustic tomography. IEEE Trans. Med. Imaging. 2016;36(1):322–331. doi: 10.1109/TMI.2016.2607199. [DOI] [PubMed] [Google Scholar]
- 26.Cox B.T., Arridge S.R., Köstli K.P., Beard P.C. Two-dimensional quantitative photoacoustic image reconstruction of absorption distributions in scattering media by use of a simple iterative method. Appl. Opt. 2006;45(8):1866–1875. doi: 10.1364/ao.45.001866. [DOI] [PubMed] [Google Scholar]
- 27.Tarvainen T., Cox B.T., Kaipio J., Arridge S.R. Reconstructing absorption and scattering distributions in quantitative photoacoustic tomography. Inverse Probl. 2012;28(8) [Google Scholar]
- 28.Seo I., Hayakawa C.K., Venugopalan V. Radiative transport in the delta‐approximation for semi‐infinite turbid media. Med. Phys. 2008;35(2):681–693. doi: 10.1118/1.2828184. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Maslov K., Zhang H.F., Wang L.V. Effects of wavelength-dependent fluence attenuation on the noninvasive photoacoustic imaging of hemoglobin oxygen saturation in subcutaneous vasculature in vivo. Inverse Probl. 2007;23(6):S113–S122. [Google Scholar]
- 30.Xia W., Nikitichev D.I., Mari J.M., West S.J., Pratt R., David A.L., Ourselin S., Beard P.C., Desjardins A.E. Performance characteristics of an interventional multispectral photoacoustic imaging system for guiding minimally invasive procedures. J. Biomed. Opt. 2015;20(8) doi: 10.1117/1.JBO.20.8.086005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Wang L.V., Wu H.-i. John Wiley & Sons; 2007. Biomedical Optics: Principles and Imaging. [Google Scholar]
- 32.Oraevsky A., Clingman B., Zalev J., Stavros A., Yang W., Parikh J.R. Clinical optoacoustic imaging combined with ultrasound for coregistered functional and anatomical mapping of breast tumors. Photoacoustics. 2018;12:30–45. doi: 10.1016/j.pacs.2018.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Liu Y., Jiang H., Yuan Z. Two schemes for quantitative photoacoustic tomography based on Monte Carlo simulation. Med. Phys. 2016;43(7):3987–3997. doi: 10.1118/1.4953185. [DOI] [PubMed] [Google Scholar]
- 34.Jacques S.L. Coupling 3D Monte Carlo light transport in optically heterogeneous tissues to photoacoustic signal generation. Photoacoustics. 2014;2(4):137–142. doi: 10.1016/j.pacs.2014.09.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Hochuli R., Powell S., Arridge S., Cox B. Quantitative photoacoustic tomography using forward and adjoint Monte Carlo models of radiance. J Biomed Opt. 2016;21(12) doi: 10.1117/1.JBO.21.12.126004. [DOI] [PubMed] [Google Scholar]
- 36.Zhu C., Liu Q. Review of Monte Carlo modeling of light transport in tissues. J. Biomed. Opt. 2013;18(5) doi: 10.1117/1.JBO.18.5.050902. [DOI] [PubMed] [Google Scholar]
- 37.Graaff R., Koelink M.H., de Mul F.F., Zijistra W.G., Dassel A.C., Aarnoudse J.G. Condensed Monte Carlo simulations for the description of light transport. Appl. Opt. 1993;32(4):426–434. doi: 10.1364/AO.32.000426. [DOI] [PubMed] [Google Scholar]
- 38.Kumavor P.D., Xu C., Aguirre A., Gamelin J.K., Ardeshirpour Y., Tavakoli B., Zanganeh S., Alqasemi U.S., Yang Y., Zhu Q. Target detection and quantification using a hybrid hand-held diffuse optical tomography and photoacoustic tomography system. Journal of Biomedical Optics. 2011;16(4):1–13. doi: 10.1117/1.3563534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Bauer A.Q., Nothdurft R.E., Culver J.P., Erpelding T.N., Wang L.V. Quantitative photoacoustic imaging: correcting for heterogeneous light fluence distributions using diffuse optical tomography. J. Biomed. Opt. 2011;16(9):1–8. doi: 10.1117/1.3626212. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Daoudi K., Hussain A., Hondebrink E., Steenbergen W. Correcting photoacoustic signals for fluence variations using acousto-optic modulation. Opt. Express. 2012;20(13):14117–14129. doi: 10.1364/OE.20.014117. [DOI] [PubMed] [Google Scholar]
- 41.Hussain A., Petersen W., Staley J., Hondebrink E., Steenbergen W. Quantitative blood oxygen saturation imaging using combined photoacoustics and acousto-optics. Opt. Lett. 2016;41(8):1720–1723. doi: 10.1364/OL.41.001720. [DOI] [PubMed] [Google Scholar]
- 42.Zemp R.J. Quantitative photoacoustic tomography with multiple optical sources. Appl. Opt. 2010;49(18):3566–3572. doi: 10.1364/AO.49.003566. [DOI] [PubMed] [Google Scholar]
- 43.Held K.G., Jaeger M., Rička J., Frenz M., Akarçay H.G. Multiple irradiation sensing of the optical effective attenuation coefficient for spectral correction in handheld OA imaging. Photoacoustics. 2016;4(2):70–80. doi: 10.1016/j.pacs.2016.05.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Tzoumas S., Nunes A., Olefir I., Stangl S., Symvoulidis P., Glasl S., Bayer C., Multhoff G., Ntziachristos V. Eigenspectra optoacoustic tomography achieves quantitative blood oxygenation imaging deep in tissues. Nat. Commun. 2016;7:12121. doi: 10.1038/ncomms12121. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Kirchner T., Gröhl J., Maier-Hein L. Context encoding enables machine learning-based quantitative photoacoustics. J. Biomed. Opt. 2018;23(5):1–9. doi: 10.1117/1.JBO.23.5.056008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Xia J., Danielli A., Liu Y., Wang L., Maslov K., Wang L.V. Calibration-free quantification of absolute oxygen saturation based on the dynamics of photoacoustic signals. Opt. Lett. 2013;38(15):2800–2803. doi: 10.1364/OL.38.002800. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Rajian J.R., Carson P.L., Wang X. Quantitative photoacoustic measurement of tissue optical absorption spectrum aided by an optical contrast agent. Opt. Express. 2009;17(6):4879–4889. doi: 10.1364/oe.17.004879. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Naser M.A., Sampaio D.R.T., Munoz N.M., Wood C.A., Mitcham T.M., Stefan W., Sokolov K.V., Pavan T.Z., Avritscher R., Bouchard R.R. Improved Photoacoustic-Based Oxygen Saturation Estimation With SNR-Regularized Local Fluence Correction. IEEE Trans. Med. Imaging. 2019;38(2):561–571. doi: 10.1109/TMI.2018.2867602. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Laufer J., Cox B., Zhang E., Beard P. Quantitative determination of chromophore concentrations from 2D photoacoustic images using a nonlinear model-based inversion scheme. Appl. Opt. 2010;49(8):1219–1233. doi: 10.1364/AO.49.001219. [DOI] [PubMed] [Google Scholar]
- 50.Saratoon T., Tarvainen T., Cox B.T., Arridge S.R. A gradient-based method for quantitative photoacoustic tomography using the radiative transfer equation. Inverse Probl. 2013;29(7) [Google Scholar]
- 51.Pulkkinen A., Cox B.T., Arridge S.R., Kaipio J.P., Tarvainen T. A Bayesian approach to spectral quantitative photoacoustic tomography. Inverse Probl. 2014;30(6) [Google Scholar]
- 52.Pulkkinen A., Cox B.T., Arridge S.R., Goh H., Kaipio J.P., Tarvainen T. Direct Estimation of Optical Parameters From Photoacoustic Time Series in Quantitative Photoacoustic Tomography. IEEE Trans. Med. Imaging. 2016;35(11):2497–2508. doi: 10.1109/TMI.2016.2581211. [DOI] [PubMed] [Google Scholar]
- 53.Buchmann J., Kaplan B.A., Powell S., Prohaska S., Laufer J. Three-dimensional quantitative photoacoustic tomography using an adjoint radiance Monte Carlo model and gradient descent. J. Biomed. Opt. 2019;24(6) doi: 10.1117/1.JBO.24.6.066001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Buchmann J., Bernhard K., Powell S., Prohaska S., Laufer J. Quantitative PA tomography of high resolution 3-D images: Experimental validation in a tissue phantom. Photoacoustics. 2020;17 doi: 10.1016/j.pacs.2019.100157. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Vogt W.C., Jia C., Wear K.A., Garra B.S., Pfefer T.J. Biologically relevant photoacoustic imaging phantoms with tunable optical and acoustic properties. J. Biomed. Opt. 2016;21(10) doi: 10.1117/1.JBO.21.10.101405. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Vogt W.C., Jia C., Wear K.A., Garra B.S., Pfefer T.J. Phantom-based image quality test methods for photoacoustic imaging systems. J. Biomed. Opt. 2017;22(9):1–14. doi: 10.1117/1.JBO.22.9.095002. [DOI] [PubMed] [Google Scholar]
- 57.Vogt W.C., Zhou X., Andriani R., Wear K.A., Pfefer T.J., Garra B.S. Photoacoustic oximetry imaging performance evaluation using dynamic blood flow phantoms with tunable oxygen saturation. Biomed. Opt. Express. 2019;10(2):449–464. doi: 10.1364/BOE.10.000449. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Haskell R.C., Svaasand L.O., Tsay T.-T., Feng T.-C., McAdams M.S., Tromberg B.J. Boundary conditions for the diffusion equation in radiative transfer. J. Opt. Soc. Am. A. 1994;11(10):2727–2741. doi: 10.1364/josaa.11.002727. [DOI] [PubMed] [Google Scholar]
- 59.Wang L., Jacques S.L., Zheng L. CONV—convolution for responses to a finite diameter photon beam incident on multi-layered tissues. Comput. Methods Programs Biomed. 1997;54(3):141–150. doi: 10.1016/s0169-2607(97)00021-7. [DOI] [PubMed] [Google Scholar]
- 60.Pfefer T.J., Barton J.K., Chan E.K., Ducros M.G., Sorg B.S., Milner T.E., Nelson J.S., Welch A.J. A three-dimensional modular adaptable grid numerical model for light propagation during laser irradiation of skin tissue. IEEE J. Sel. Top. Quantum Electron. 1996;2(4):934–942. [Google Scholar]
- 61.Rakic A.D. Algorithm for the determination of intrinsic optical constants of metal films: application to aluminum. Appl. Opt. 1995;34(22):4755–4767. doi: 10.1364/AO.34.004755. [DOI] [PubMed] [Google Scholar]
- 62.Tavakoli B., Kumavor P.D., Aguirre A., Zhu Q. Effect of ultrasound transducer face reflectivity on the light fluence inside a turbid medium in photoacoustic imaging. J. Biomed. Opt. 2010;15(4) doi: 10.1117/1.3462930. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Laufer J., Elwell C., Delpy D., Beard P. In vitro measurements of absolute blood oxygen saturation using pulsed near-infrared photoacoustic spectroscopy: accuracy and resolution. Phys. Med. Biol. 2005;50(18):4409–4428. doi: 10.1088/0031-9155/50/18/011. [DOI] [PubMed] [Google Scholar]
- 64.Prahl S.A., van Gemert M.J., Welch A.J. Determining the optical properties of turbid media by using the adding–doubling method. Appl. Opt. 1993;32(4):559–568. doi: 10.1364/AO.32.000559. [DOI] [PubMed] [Google Scholar]
- 65.Bosschaart N., Edelman G.J., Aalders M.C., van Leeuwen T.G., Faber D.J. A literature review and novel theoretical approach on the optical properties of whole blood. Lasers Med. Sci. 2014;29(2):453–479. doi: 10.1007/s10103-013-1446-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.ISO 80601-2-61:2017 . 2017. Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment. [Google Scholar]
- 67.Kleiser S., Nasseri N., Andresen B., Greisen G., Wolf M. Comparison of tissue oximeters on a liquid phantom with adjustable optical properties. Biomed. Opt. Express. 2016;7(8):2973–2992. doi: 10.1364/BOE.7.002973. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Akhlaghi N., Pfefer T.J., Wear K.A., Garra B.S., Vogt W.C. Multidomain computational modeling of photoacoustic imaging: verification, validation, and image quality prediction. J. Biomed. Opt. 2019;24(12) doi: 10.1117/1.JBO.24.12.121910. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.