Abstract
Magnetic resonance spectroscopic imaging (MRSI) offers considerable promise for monitoring metabolic alterations associated with disease or injury; however, to date, these methods have not had a significant impact on clinical care, and their use remains largely confined to the research community and a limited number of clinical sites. The MRSI methods currently implemented on clinical MRI instruments have remained essentially unchanged for two decades, with only incremental improvements in sequence implementation. During this time, a number of technological developments have taken place that have already greatly benefited the quality of MRSI measurements within the research community and which promise to bring advanced MRSI studies to the point where the technique becomes a true imaging modality, while making the traditional review of individual spectra a secondary requirement. Furthermore, the increasing use of biomedical MR spectroscopy studies has indicated clinical areas where advanced MRSI methods can provide valuable information for clinical care. In light of this rapidly changing technological environment and growing understanding of the value of MRSI studies for biomedical studies, this article presents a consensus from a group of experts in the field that reviews the state-of-the-art for clinical proton MRSI studies of the human brain, recommends minimal standards for further development of vendor-provided MRSI implementations, and identifies areas which need further technical development.
Keywords: brain, magnetic resonance spectroscopic imaging, review
1 ∣. INTRODUCTION
The combination of magnetic resonance spectroscopy with spatial encoding methods enables spectral information to be mapped in a noninvasive manner. Proton MR spectroscopic imaging (MRSI) has been of particular value for in vivo measurements in the brain, for both clinical and biomedical research studies.1,2 Although proposed over 30 years ago3-5 and widely implemented on commercial MRI systems, the clinical adoption of MRSI remains limited for reasons that include long acquisition times, low spatial resolutions, variable quality, inadequate analysis software and a lack of appreciation in the clinical community of the potential impact on clinical care. However, recent technological developments promise to transform MRSI into a more reliable and higher throughput modality through the increased availability of higher magnetic fields (≥3 T), multichannel detector systems, new encoding methods and new approaches for obtaining metabolite maps over large brain volumes. These developments promise to bring the quality of MRSI to the point where the technique becomes a true imaging modality and the more traditional analysis of individual spectra becomes a secondary requirement. Complementary to these technological developments has been increased experience within the research community of the potential value of MRSI-detected metabolic biomarkers for clinical studies. With this background of an evolving technology and increased understanding of the potential of MRSI, this report aims to: (1) summarize the state-of-the-art acquisition and analysis methods and review the status of clinical applications; (2) make specific recommendations for minimum implementation standards that apply to all MRSI studies; (3) recommend particular areas where further development is needed to bring advanced MRSI methods into the clinical setting; and (4) propose future directions where advanced MRSI methods should be implemented as part of research and clinical imaging protocols.
Both MRI and MRS technologies have numerous options in the choice of acquisition methods and parameters, which can result in multiple tradeoffs between information content and implementation considerations. Together with an ongoing and steady rate of development, these characteristics present challenges to the development of a set of implementation recommendations. Instead, this report aims to identify broad categories of clinical applications and emerging translational research areas for which different spatial or spectral sampling capabilities are recommended. For example, some applications can make effective use of relatively simple spectroscopic acquisitions to generate a map of the choline to N-acetylaspartate ratio, suggesting that acquisition design considerations can focus on optimizing spatial resolution and acquisition times. By contrast, studies that aim to measure neurotransmitters or tumor-specific molecular markers require optimum spectral discrimination and sensitivity, while spatial sampling becomes a secondary consideration. Therefore, an additional aim of this report is to summarize these application-specific considerations in the choice of MRSI methods selection.
Limitations of this report include that it considers methods that have already been demonstrated for in vivo human studies and for which there is reasonable expectation that these can be translated into clinical practice. Technologies such as hyperpolarized MRS and multinuclear detection are not considered. Considerations for 1H MRSI studies at ultrahigh field (>7 T) are discussed; however, this also remains an area of active development.
2 ∣. CLINICAL UTILITY OF MRSI OF THE BRAIN
2.1 ∣. Clinical applications
Despite the clear diagnostic potential of 1H MRS and decades of effort demonstrating that it provides complementary information relative to MRI, it largely remains an investigational tool that is not recommended for reimbursement in several countries. As an “imaging” method, MRSI offers greater promise for routine clinical use than single voxel spectroscopy (SVS), although at the expense of increased measurement complexity and typically longer acquisition times. While a direct comparison of the clinical value and reliability of MRSI relative to SVS may not be possible, it is known that the limited spatial sampling of SVS methods can impact diagnostic value.6,7 The typically smaller voxel volumes of MRSI have also been shown to enable sampling of brain regions where SVS results were of poor quality,8 and while motion artifacts may be visible in MRSI, these can equally affect SVS measurements but without being apparent in the result.9
The spectral sampling in most MRSI studies is comparable with that used for SVS, so in principle MRSI can provide the same spectroscopic information as SVS, although acquisitions are typically designed to sample smaller voxel volumes, which reduces detection sensitivity10 (signal-to-noise ratio [SNR] per unit time). As a result, most clinical studies using MRSI have focused on detection of prominent spectral components, namely, N-acetylaspartate (NAA), creatine (Cr) and choline (Cho) compounds, followed by a smaller number of studies that also include lactate (Lac), myo-inositol (mI) and the combined signal from glutamate and glutamine (Glx). Lower spatial resolution MRSI studies have also been demonstrated for detection of compounds that can be challenging even for SVS measurements, such as γ-aminobutyric acid (GABA)11,12 and glutathione (GSH).13
An obvious benefit of MRSI over SVS for clinical applications is that spectra are obtained from multiple regions in a single study, which particularly benefits studies of heterogeneous pathologies. The following sections review clinical applications that particularly benefit from the multi-voxel sampling provided by MRSI.
2.1.1 ∣. Brain tumors
Many brain tumors exhibit considerable heterogeneity and may have multi-focal lesions, making MRSI useful for identifying regions of infiltrative tumor14-18 and guiding biopsy localization.19-24 There is long-standing interest in the use of MRSI for treatment planning,25-31 for which the widely reported metabolic changes that occur outside of MRI-observed abnormalities may have a significant impact on outcomes.26,32-34 Additional applications include assessment of treatment response,35-39 identifying recurrent tumor,40,41 and stratifying patients into subgroups for outcome assessments.36 These applications benefit from having metabolic information over a large region and volumetric measurements are particularly beneficial.41-45 Most studies are based upon the observation of increased Cho and decreased NAA in tumor tissue (Figure 1),46 with improved contrast between normal and tumor tissue at longer TEs.47 Meta-analyses provide evidence that Cho/NAA is effective for distinguishing high and low grade glioma and for separating regions with high grade pathology from areas of necrosis and less malignant tumor, although large variations in these findings have also been reported.16-18 There is also increasing interest in making use of the entire metabolic profile, including changes of lipid,48 glutamate,13 glutamine, alanine,49 GSH,50 mI51 (Figure 1) and D-2-hydroxglutarate (2HG), which is a marker of oncogenic IDH mutation status.52,53 Since oncogenic IDH mutations are only found within the tumor there is no background 2HG signal, presenting considerable potential for mapping tumor distributions to guide and monitor treatment.13,48,50,51,54-56 Implementation of 2HG measurement for volumetric acquisitions has also been demonstrated (Figure 2).54,57,58
2.1.2 ∣. Epilepsy
Presurgical evaluation of epilepsy remains challenging, typically requiring co-localizing information from multiple diagnostic tests, including structural and diffusion MRI, PET, EEG and MEG. MRSI has been shown to provide complementary information via detection of altered tissue metabolism in the local neighborhood of a seizure focus, primarily from the reduction of NAA. Long echo time studies have indicated reduced NAA compared with healthy controls, as well as abnormal metabolism in the limbic and subcortical regions.59 Detection of neocortical epilepsy particularly benefits from the availability of whole-volume mapping due to the cortical location and limitations of prior localizing information to direct placement of a smaller imaging volume.8,60 This application may also particularly benefit from increased spatial resolution provided at higher magnetic fields. A study at 7 T showed that a positive outcome was associated with the extent of resection of the region of abnormal NAA/Cr.61 Other metabolites of interest are GABA and glutamate,62 although diagnostic value for MRSI remains to be shown.
2.1.3 ∣. Traumatic brain injury
Traumatic brain injury and chronic traumatic encephalopathy are characterized by widespread metabolic alterations that can occur remote from regions indicated by structural MRI-observed lesions or altered DTI measures,63,64 therefore, given the limitations of other localizing information, wide FOV MRSI measurements are best suited for studies of TBI. These metabolic alterations vary with severity and time after injury,65 with dominant findings being decreased NAA and increased Cho in the subacute phase that may persist for many years.66,67 Several studies have demonstrated an association of MRS biomarkers with cognitive assessments,63,68 although this association is not strong within the mild injury group, for which 15% to 30% of subjects will experience longer term postconcussion symptoms.69 Longer TE measurements, with sampling of NAA and Cho, appear to be suitable for studies of TBI; however, further studies are needed to determine if other MRS biomarkers are of value and whether there are associations of the spatial distributions of altered metabolites with cognitive outcomes.
2.1.4 ∣. Multiple sclerosis
Multiple sclerosis is associated with demyelination, remyelination, gliosis and axonal loss over multiple brain regions. While focal lesions are observed on structural MRI, it has been shown that there are also widespread metabolic alterations in normal-appearing brain tissue that are relevant for assessing disease progression.70 More specifically decreased NAA and increased Cho, mI and Glu have been found in normal-appearing white matter (NAWM) and, inside the acute lesions, reduced NAA, GABA and Glu and elevated mI have also been detected in gray matter.71 MRSI is thus able to localize changes in both lesions and normal-appearing brain tissue72 and is valuable for serial evaluation of disease progression and response to therapy.73,74 With the more widespread availability of ultrahigh field scanners, it is also possible to evaluate changes of GABA and GSH.74,75
2.1.5 ∣. Mitochondrial disorders
Mitochondrial disorders are a clinically, morphologically and biochemically diverse group of energy metabolism disorders with a variety of causes, including genetic, physiological and environmental.76 Diagnosis is complicated by varied etiopathophysiology, novel mitochondrial DNA mutations, and heterogeneity of the genotypes and phenotypes. Existing screening tests are effective for only a limited number of variants, are invasive, and/or may not provide the required information. Because metabolic changes associated with dysfunctional mitochondria are generally distinctive and brain tissue- or region-specific, MRSI measurements can aid in diagnosis and provide indices of disease progression and therapeutic response. Particularly sensitive markers of mitochondrial dysfunction that are of clinical utility are changes in NAA77,78 and parenchymal and/or CSF lactate,77-79 which is upregulated in mitochondrial dysfunction.
2.1.6 ∣. Other potential applications
MRS has been widely used for studies of psychiatric disorders, pain disorders and neurodegenerative diseases; however, definitive clinical applications outside of the research arena, and the value of spatial information on metabolite distributions, remain elusive.80-82 Continued interest remains in the extension to ultrahigh field (7 T) studies for measurements of glutamate, glutamine, GABA11,12 and GSH,83 for example for response assessment84 and to provide objective criteria for studies of psychopathological and therapeutic mechanisms. Another topic of interest is the search for biomarkers that discriminate disease subtypes with different disease progression or therapy response.
Brain infection is a rare application where 1H MRSI has already proven clinical impact. An additional succinate peak reflects bacterial metabolism and is paralleled by NAA decrease and lactate increase.
2.2 ∣. Status of clinical acceptance
MRSI sequences provided by MRI systems manufacturers are largely based on older technologies and as a result many research groups have developed their own methods. These implementations have not been widely distributed within the clinical MR community and this has led to a lack of standardization and poor integration into the clinical workflow. Additional barriers to more widespread clinical adoption include a requirement for specialized expertise for quality control and interpretation, and difficulties associated with obtaining reimbursement.
Several technical limitations must be addressed by the equipment manufacturers before advanced MRSI methods can be widely accepted within the clinical community. The most challenging is that some brain regions cannot be reliably sampled due to local B0 inhomogeneities, which cannot be corrected by manufacturer-implemented B0 shim system designs, a limitation that has become of greater importance as target volumes have increased. However, for full clinical acceptance there needs to be an expectation that pathologies within any brain region can be reliably sampled. A second requirement is that MRSI reconstruction, postprocessing and quantitative analysis will have to be fully automated and integrated into the clinical workflow, including robust quality control (see section 5.1). A further limitation remains relatively lengthy acquisition times, which can present challenges for clinical workflows. Additional effort is also required to ensure quantitative equivalence between scanners and to demonstrate reproducibility to support multi-center studies.
3 ∣. DATA ACQUISITION METHODS
3.1 ∣. Spectral acquisition methods
The selection of MRSI acquisition methods must first consider the required spatial and spectral information requirements and relative difficulties for acquiring that information. Intermediate to long TE (>50 ms) acquisitions with volume or slice selection provide spectra that are simple to interpret85 and less susceptible to lipid contamination and baseline distortions, and therefore are preferred for whole-slice or whole-brain studies.86 Short TE sequences are generally better suited for mapping J-coupled metabolites but must account for stronger lipid and macromolecular contributions.87 PRESS or sLASER excitation sequences with sampling of the second half of a spin-echo is most widely implemented, although pulse-acquire FID acquisitions are being increasingly used for measurements at ≥7 T due to lower SAR requirements, while also increasing SNR due to minimal T2-relaxation and J-evolution,88,89 having negligible chemical shift displacement errors (CSDEs) and lower sensitivity to B1 inhomogeneities.90-92 Since FID-excitation requires a full-slice or whole-brain selection, this also requires effective control of extracranial lipid signal during image reconstruction88,89,93-95 and careful handling of macromolecular signals.96
Sampling of metabolites at the lower limits of MRS detection (≤2 mM) or with strong spectral overlap (e.g. GABA, GSH) typically requires spectral editing and large voxels,97 although implementations at high field strengths are changing the volume considerations.98,99 The main obstacle for spectrally edited MRSI is the sensitivity to B0 inhomogeneities and temporal frequency drifts, although three-dimensional (3D) MRSI implementations have been reported.11,12,99
3.2 ∣. Spatial acquisition methods
3.2.1 ∣. Volume selection and lipid removal
Primary acquisition considerations for brain MRSI include reduction of signal contamination from extracranial lipids, efficient sampling of spatial k-space information to obtain the target spatial resolution and SNR within acceptable acquisition times,100 and an ability to achieve the necessary B0 homogeneity over the volume of interest (VOI). VOI excitation by PRESS, STEAM, semi-LASER or LASER101 is commonly used to avoid extracranial regions,102,103 although it has limitations of requiring positioning of the VOI and poor sampling of cortical regions. In addition, PRESS-based volume selection at higher fields leads to an unacceptably large CSDE.103
Full-slice or whole-brain excitations, as well as multi-slice MRSI with carefully tailored outer volume saturation (OVS),104 improve the ability to investigate cortical regions and have the significant advantage that knowledge about the location of pathological abnormalities is not required. However, they face additional challenges with B0 and B1 field inhomogeneity that impact spatial or spectrally selective excitation or saturation methods. Whole-brain excitation has chiefly been combined with intermediate or long echo times to mitigate the lipid contamination and baseline problems, but this limits the number of metabolites and the achievable spatial resolution. Whole-brain coverage at medium and short TEs requires additional considerations for reduction of lipid signals. This may include OVS105,106 or lipid inversion-nulling107; however, OVS requires automated placement procedures for clinical applications,108,109 is SAR-demanding and difficult to implement for larger field of views (FOVs), and inversion-nulling incurs a loss of sensitivity, suppresses diagnostically relevant lipids, and still benefits from selective saturation of the orbits and fronto-temporal regions. Additional approaches include the use of specialized gradient coils for dephasing signals near the scalp94,110 and relying entirely on higher spatial resolutions and optimization of the spatial response function (SRF), which can be done using acquisition,79 reconstruction,111-115 and postprocessing (see section 4) methods. For large VOI acquisitions, additional considerations include that a wide range of spectral quality is inevitably obtained, placing additional demands on quality control and spectral analysis. With the use of multichannel detection and increasing resolutions the sizes of the sampled volumetric raw data can become significant (~20-100 Gb), making rapid processing on current scanners challenging.
Typical spatial resolutions for 3 T measurements of the major singlet resonances and high abundance multiplets range from 0.3 to 1.0 cc with scan times of the order of 10 to 18 minutes.91,113 With FID acquisition and constrained reconstruction methods, voxel sizes as small as 0.04 cc nominal voxel volume have been reported.116,117 At 7 T, nominal voxel sizes of 0.023 cc have been reported using a FID acquisition method and scan time of 17 minutes.118
3.2.2 ∣. MRSI encoding
Cartesian phase-encoded MRSI, often combined with elliptical k-space distribution, is the most commonly implemented acquisition strategy; however, for higher spatial resolutions this leads to long scan times. Several k-space undersampling methods such as parallel imaging, compressed sensing or multi-band encoding119-121 have been demonstrated, with acceptable tradeoffs between data quality, scan time and spatial resolution, although with increased motion sensitivity and lipid fold-in artifacts for higher accelerations. Increased sampling efficiency can be obtained using spatial-spectral encoding (SSE) methods that combine sampling of spectral information with simultaneous sampling along one or two k-space dimensions. Initially implemented using echo planar readout (EPSI),87,122 several non-Cartesian trajectories have also been used.112,123,124 These can also be combined with k-space undersampling and multi-band encoding to further increase sampling efficiency.125-130 When implemented at higher field strengths these advanced MRSI methods can provide metabolite maps of relatively high spatial resolution within clinically acceptable acquisition times (Figure 3).131 Limitations of these methods include increased noise, gradient heating leading to frequency drifts, more complex image reconstruction, and maintaining sufficient spectral bandwidth for ≥7 T, although temporal interleaves can be used to increase spectral sweepwidth.132,133
3.3 ∣. Implementation considerations
3.3.1 ∣. B1 and B0 inhomogeneity
Both intra-voxel and global B0 homogeneity must be considered for MRSI. Intra-voxel inhomogeneity is dominated by local magnetic susceptibility variations, with strong effects in regions such as the temporal-frontal brain due to the tissue-air interface, or near hemorrhage, calcifications, or surgical cavities. Smaller voxels or spatial oversampling can be used to increase the volume over which suitable quality spectra can be obtained, although with a tradeoff with SNR.134-136 For whole-brain studies at 3 T using spatial oversampling with a nominal voxel volume of 0.31 cc, the effect of intra-voxel inhomogeneity limits sampling to ~ 75% of the brain volume137 although there is considerable variability between subjects and instruments. Global B0 inhomogeneity, i.e. over the whole FOV, results in spatially dependent frequency shifts that can degrade frequency-selective water or lipid suppression, or spectral editing. Achieving an acceptable global B0 homogeneity becomes increasingly difficult for larger FOVs and for whole-brain studies there is a high likelihood of large unsuppressed water or extracranial lipid signals, and metabolite signals close to water or lipid may be impacted. With current B0 shim capabilities, frequency-selective spectral editing is therefore only recommended for centrally located volumes where global B0 homogeneity can meet the strict requirements.
Image-based B0 shim algorithms with robust convergence, spatial constraints and reliable calibration are required, together with suitable B0 shim strengths.138 Shim array coils,139,140 high order shim systems141 and dynamic B0 shimming93 promise to provide considerably improved performance; however, these are currently not supported by MRI manufacturers.
B1 inhomogeneity leads to spatially dependent signal losses in metabolite images. These effects can be minimized by using sequences with B1-insensitive or low flip angle pulses,102,103 and can be corrected for by applying signal normalization (see section 5.2).
3.3.2 ∣. Temporal instabilities
Subject motion results in image artifacts, decreased SNR and degraded SRFs. In addition, B0 homogeneity is altered, meaning that a full correction of these effects requires real-time updates of the scanner frequency, gradients and B0 shims. B0 field drifts can be caused by gradient heating, particularly with the SSE methods,142 or following gradient-intensive MRI sequences such as diffusion or fMRI.143 If not corrected, these lead to degraded water suppression and cause problems with acquisition methods that include temporal interleaves or signal averaging.
3.3.3 ∣. Reference measurements
Several of the artifacts mentioned in the previous sections can be addressed with one or more reference measurements. Motion correction can use methods described for MRI,144 and real-time frequency correction can be implemented using a frequency navigator145 or a full B0 shim correction measurement,146 which may increase the minimum TR. Implementation of both frequency drift and motion correction during the acquisition is highly recommended for clinical studies.
Separate water MRSI or MRI measurements can be used to support several processing functions. A water MRSI reference is the most convenient method for applying B0 and lineshape corrections and signal normalization, similar to the approach widely used for SVS, and is strongly recommended. This can be acquired as a separate measurement performed with reduced TR and reduced k-space sampling to minimize the additional scan time,147-149 while assuming the same head position, or as an interleaved measurement following the water-suppressed acquisition,142 which lengthens the minimum TR time.
4 ∣. MRSI PROCESSING METHODS
4.1 ∣. Spatial reconstruction and spectral processing
Multichannel detection is routinely used to optimize sensitivity and enable parallel-imaging acceleration methods. The processing methods used for MRSI are similar to those widely implemented for MRI,120,128,150-155 although they additionally require information on the phase of the detection sensitivity distributions.
Spectral processing for MRSI includes, as a minimum, the typical steps used in SVS, including removal of residual water, frequency/lineshape/phase correction (ECC) and Fourier transform. The spatial dimensions increase the data storage requirements, complexity and computation time relative to SVS processing, requiring 10s to 100s of minutes to process volumetric MRSI that may contain several thousand voxels.86,87
The limitations of detection sensitivity and spatial encoding efficiency (section 3.2) may result in low spatial resolutions and signal contamination due to the broad SRF, which can notably result in the propagation of extracranial lipid signals. These effects can be reduced using spatial smoothing, at the expense of reducing the effective spatial resolution, or weighted k-space acquisition.131 Alternatively, image reconstruction approaches that control the SRF by incorporating a priori spatial information of the lipid regions,156-158 or signal removal based on lipid spectral patterns,157,159 can be applied.
Several noise reduction methods can be used to improve visual interpretation of metabolite images and spectra. Spatial smoothing and image interpolation can improve the appearance of metabolite images and noise reductions can be obtained from constrained spatial reconstructions.117,160 When reporting acquisition parameters, the effective spatial resolution, ie the final value after smoothing, should be stated. Spectral filtering or denoising161,162 can also be used to improve the appearance of spectra, although apodization-smoothing should be kept to small values (e.g. ≤2 Hz at 3 T) so as not to impact the accuracy of the spectral analysis. However, in this regard, the contrast in MRSI metabolite maps is commonly of greater importance than quantitation accuracy.
4.2 ∣. Spectral fitting
Given the large number of spectra in MRSI data, fully automated spectral analysis methods are essential, for which methods based on iterative parametric modeling and incorporation of a priori metabolite spectral information have been widely used.163-165 Newer algorithmic approaches using machine learning are also anticipated to play an increasing role.166,167 Since MRSI spectra may include varying lineshapes and large lipid and unsuppressed water signals, analysis methods must include robust handling of these features. In consideration of this requirement and the typical SNR levels, the selection of simpler parametric spectral models is generally recommended; for example, while the use of a Lorentz-Gauss or variable lineshape model may be preferred for analysis of SVS data, a Gaussian lineshape may be more robust for MRSI. Similarly, inclusion of metabolite basis functions for signals with amplitudes below the detection threshold is not recommended as this will likely result in overfitting.
Spectral analysis for MRSI can benefit from application of frequency and phase corrections prior to fitting,168,169 and from inclusion of a priori spatial information, which may include removal of local outliers or enforcing spatial smoothness of parameters. Spatial constraints can be included in the penalty function used in the optimization170-172 or by modifying starting values for repeated applications of the spectral fitting.164
4.3 ∣. Scanner integration
Considerations for integrating advanced MRSI methods into the clinical workflow include efficient processing of large datasets and combination with information derived from other imaging modalities as may be needed for several processing and analysis steps. Processing for multichannel and high-resolution volumetric MRSI acquisitions has large memory requirements, which may not be available in standard instrument configurations, and lengthy processing times, which could interfere with subsequent imaging protocols, therefore the use of automated data transfers to dedicated servers and application-specific processing programs may be required.166,167
5 ∣. DATA ANALYSIS METHODS
5.1 ∣. Quality control
Quality control is essential to reduce both false positive and false negative results. The traditional quality assessment metric, Cramer-Rao lower bounds (CRLB), needs to be used with caution, as it is only valid if the spectrum is free of artifacts and accurately described by the model used for calculation. Other measures include the fitted metabolite linewidth or the SNR; however, all of these measures would incorrectly identify as poor quality a spectrum with low metabolite content within a lesion.173 For this situation, the linewidth measured from a coregistered water spectrum can be used, which would, for example, appropriately label necrotic regions that contain water but no measurable metabolite signals. Data-driven analysis tools can also be used to spatially segregate spectral features such as lipid contamination and baseline distortion,174 and machine learning approaches are also anticipated to play an increasing role.175,176
Specific criteria for quality evaluation have not been established and can depend on the application. The commonly applied criteria for SVS, of CRLB ≤20% and linewidth ≤0.1 ppm (13 Hz at 3 T) are equally applicable to MRSI studies,9 although these may be loosened for MRSI measurements of singlet resonances where data review will be based on visual review of metabolite images (e.g. to 15 Hz), or tightened to provide a more rigid criteria for quantitative measurements (e.g. to 11 Hz). The SNR can also be used (for nonlesional volumes only), and for quantitative measurements a minimum SNR (defined as NAA peak amplitude to RMS noise) of >10 for measurements of the primary singlet resonances or >20 for measurements of multiplet resonance groups is recommended. Depending on the disease, false negative results might be of more concern than false positive results (i.e. cancer detection), or vice versa, and quality criteria could be set to be more or less conservative with either type I or type II errors. If metabolite maps are to be used for clinical procedures, for example, neurosurgical or radiation treatment planning, it is recommended that voxels identified as being of inadequate quality be set to zero to avoid possible misinterpretation (but leaving the spectrum unchanged and available for visual review). However, for other applications it may be possible to retain all the image data, but to also display a “Quality Map” together with the metabolite maps, as illustrated in Figure 4.135 In either situation, visual review of spectra in critical areas is still recommended.
Motion and outer volume contamination can cause artifacts that can be difficult to recognize in individual spectra or reconstructed metabolite maps, although may be better recognized as out-of-object image artifacts in maps created by spectral integration over the lipid signals. Chemical shift displacement artifacts can change spectral patterns at the edges of a volume-selected region, although this can be mitigated by improved excitation methods,103,177 and these edge voxels should be excluded from analysis.
5.2 ∣. Signal normalization and quantification
For comparative analyses of metabolite maps between subjects or across multiple studies, a signal normalization is required. Similar to methods used for SVS,1,178 these may include taking metabolite ratios, ratios to tissue water, or using an external quantitation reference. Metabolite ratio maps conveniently account for bias-field intensity variations and CSF partial volume effects, and may take advantage of the complementary changes of individual metabolites, e.g. as occurs for Cho and NAA with a number of pathologies; however, precautions must be taken to prevent outliers caused by small denominator values.179 Individual metabolite maps may also be normalized by taking the ratio to an internal reference region, e.g. from NAWM.180 This can be done using the same metabolite, e.g. NAA/NAANAWM, or by using another metabolite as a reference, e.g. NAA/CrNAWM. These latter two approaches provide a “self-normalization” that takes into account within-subject variations, such as due to age, or changes in normal tissue associated with disease, such as brain cancer.181,182
For semi-quantitative single metabolite maps, both a bias-field correction (to account for transmit and receive sensitivity) and signal normalization (to account for acquisition variables) must be applied, which can be done using a coregistered proton-density MRI,183 or a water reference MRSI.86,184 The use of a tissue water MRSI reference has advantages of convenience, ensuring exact coregistration and SRF, and can be used for several additional processing steps (section 3.3.3). The tissue volume fraction in each voxel can be obtained following segmentation of a high-resolution MRI and downsampled to the SRF of the MRSI study.184,185 For truly quantitative measurements, the reference should be corrected for water content, partial volume effects, and water and metabolite relaxation times; however, these requirements, such as relaxation correction, are impractical for clinical diagnostics and results are commonly reported as institutional units to indicate that they only apply to data obtained with the identical acquisition and processing.
5.3 ∣. Visualization and multi-modal integration
Visual inspection of individual spectra in MRSI data can be informative and frequently necessary. While system vendors provide display functionality on the MRI systems, support for interactive viewing of spectra on the PACS system is currently unavailable, which is critical for clinical integration. Support for the DICOM MRS standard186 is also variable, with some systems using proprietary file formats or nonstandard formulations of DICOM SOP (service-object pair) classes that are incompatible with PACS. As an alternative, static DICOM secondary capture reports can be generated that show selected images and spectra, which can be sent to PACS.
Analysis of individual spectra is time-consuming and subjective, and given that advanced MRSI methods can provide metabolite maps of relatively good spatial resolution, an image review format offers clear benefits. While metabolite maps can be sent to PACS as DICOM images, these may not show sufficiently detailed anatomical structure and there is frequently missing spatial information due to FOV selection or inadequate spectral quality; therefore, additional co-localizing information from a coregistered MRI is needed. One approach is to use color-coded metabolite images overlaid on a high-resolution MRI (e.g. Figures 3 and 4), although with a caution to limit transparency to avoid overemphasizing the background MRI. A disadvantage of this approach is that the images are sent using a true color format, which does not allow further image contrast manipulation. The selection of color tables can also be problematic, particularly for metabolite ratio maps that have a very large dynamic range.
Several stand-alone MRSI display systems have been developed at research sites that provide features such as interactive spectral selection, display of spatially registered MRIs, image overlay functions and quality map information, and support for the DICOM spectroscopy standard is provided in TARQUIN,187,188 SIVIC189,190 and jMRUI.191 The extension of these features into commercial medical image display packages is strongly encouraged, together with further support for the DICOM MRS standard. The development of standardized signal normalization and display methods is also recommended to increase clinical acceptance.
5.4 ∣. Analysis methods
The combined spatial and multiparametric information of MRSI presents opportunities for novel quantitative analyses. Spatial averaging over anatomically defined regions of interest (ROIs) can be applied, either using the individual voxel fit results or by averaging the spectra and fitting that result.192 Voxels not meeting the quality evaluation criteria should be excluded, and for spectral summation the phase and frequency correction must be performed prior to averaging. Both methods can be extended to incorporate information on the tissue volume fraction in each voxel to separate multiple contributions such as gray and white matter.193,194 Other examples include measurements over specific neuronal tracts,195,196 smaller tissue regions with volume contributions from neighboring regions,194 and different tumor regions.197,198 ROIs can also be automatically defined using atlas registration methods, which greatly benefit from having fully 3D information to support nonlinear registration.86,192
The image format of MRSI naturally supports the use of voxel-based analysis (VBA) methods, which may be preferred when prior localization information is not available. An alternative to a simple calculation of a metabolite ratio map is to detect outlying values relative to a regression line generated for two metabolites, using values selected from normal tissue regions in the same subject. For example, a Cho and NAA index (CNI) has been used to highlight tumor regions (Figure 5A).199 Comparisons between datasets from different subjects or subject groups can be performed following nonlinear spatial transform to register multiple images to a standard frame of reference where statistical tests can be carried out. This can be done for single subject data by comparing metabolite maps against mean values from a control group using a z-score analysis.65,200 VBA methods account for normal regional variations of metabolite concentrations and any spatial variability in the reproducibility of the data; however, additional steps are needed to account for the relative tissue volume contributions,65 and control values must be derived from a subject group that matches known covariates, notably age, while sex and body weight may also be considered.181 The display of z-score maps greatly facilitates interpretation by directly identifying regions that are statistically different from control values (Figure 5B); however, the need for standardized acquisition methods and availability of the normative reference data make this approach difficult to translate to routine clinical studies.
The multiparametric nature of MRSI allows for multiple metabolite maps to be analyzed together, using techniques for pattern recognition analysis, as has been investigated for SVS.201 This has primarily been done as additional image contrasts to a MRI-based tissue classification, to provide metabolic information to improve tissue classification for brain tumors.202,203 Results indicate considerable potential for these analysis approaches; however, these methods have not been widely implemented and the requirement for consistency of imaging protocols may mean that these would be difficult to translate to clinical practice.
6 ∣. RECOMMENDATIONS
6.1 ∣. Acquisition methods: application-dependent selection
As described in section 3, MRSI acquisition methods offer multiple tradeoffs in terms of ease of use, acquisition times, spatial coverage and achievable spectral information. In the selection of which technique should be used, the primary considerations are the metabolic information that is needed, over what volume, and at what spatial resolution that information can be reliably obtained. The target location will also impact B0, B1 and lipid contamination considerations; for example, large volume excitations would be suitable for measurements in the occipital and parietal lobes; studies of temporal lobes, where B0 inhomogeneity is more of concern, may benefit from slice-selective excitation; and studies of the hippocampi, subcortical gray matter, or brain stem would be better sampled with 3D VOI localization methods. Considerations for acquisition variables that impact data acquisition time depend primarily on the subject group, with, for example, studies of pediatrics, advanced brain cancer, or neurodegenerative diseases benefiting from the shortest possible acquisitions (<10 minutes), whereas studies of relatively cooperative patients, as may be the case for subjects with epilepsy or migraine, may tolerate a longer examination (e.g. 12 to 20 minutes). Other considerations include available expertise, ranging from fully vendor-supported implementations that can be routinely implemented by MRI technologists to methods that require specialized expertise to ensure quality at the time of the acquisition and in the data analysis and interpretation.
Listed in Table 1 are summarized broad categories of metabolite targets, brain volumes and field strengths that correspond to different acquisition approaches, together with examples of potential clinical applications. These range from acquisition of only the primary singlet resonances, which can be reliably done at longer TEs (e.g. >50 ms), to obtaining full spectral information using short TEs with analyses of the strongly overlapping multiplet resonance patterns, and finally to the case of using specialized acquisition methods to resolve overlapping spectral compounds such as for GABA and GSH. In general, the tradeoffs of acquisition time, spatial resolution and B0 inhomogeneity mean that smaller VOIs are more suitable if detailed spectral information is to be obtained. Therefore, the singlet resonances can be readily mapped using longer TE 3D or multi-slice MRSI over large brain volumes, without the use of spatial volume selection, whereas applications that have stricter requirements for B0 homogeneity and robust water and subcutaneous lipid suppression, are currently best suited to volume-selected acquisitions. This last recommendation is made with consideration of the current shimming hardware and software provided by scanner manufacturers; however, following future improvements of the shim hardware/software systems, large brain volumes should also be possible for mapping of metabolites such as GABA and GSH.
TABLE 1.
Target metabolites | Methods | Field strength | Example clinical applications |
---|---|---|---|
NAA, Cr, Cho | Long TE, large FOV 3D volume | 1.5 and 3 T |
|
NAA, Cr, Cho, Lac | Long TE, 3D volume with suppression of subcutaneous lipids | 1.5 and 3 T |
|
NAA, Cr, Cho, Glu, Gln, mI, Lac | Short TE, volume-selected 2D or 3D | All |
|
GABA, GSH, 2HG, etc. | Spectral editing with volume-selected 2D | 3 and 7 T |
|
NAA, Cr, Cho, Glu, mI, Gln*, NAAG*, GSH*, GABA* | FID, 2D (multi) slice or 3D volume with high spatial resolution | 3, 7 and 9.4 T |
|
The feasibility of implementing these MRSI acquisition approaches, for each of these broad categories of metabolite target groups, are summarized in Figure 6, which range from being widely available and fully integrated into available MRI systems, to requiring specialized pulse sequences and experienced personnel.
6.2 ∣. Recommendations for minimum standards
6.2.1 ∣. Acquisition
B0 shimming
The biggest limiting factor and critical requirement for MRSI quality and reliability is excellent B0 shimming. A frequency dispersion of a maximum of 20 Hz across the entire brain at 3 T should be achievable. Automated and robust B0 shim algorithms are required138 and further development of advanced shim hardware is strongly recommended (section 3.3.1).
Encoding and acceleration methods
Cartesian encoding methods will continue to be used for 2D or multi-slice studies but should include standard k-space undersampling options. The implementation of multi-slice and 3D SSE methods is strongly encouraged.
Motion
Incorporation of prospective motion and frequency drift correction into acquisition sequences is also highly recommended, with automatic re-shimming in case of substantial motion. Given the typical acquisition times of MRSI sequences, additional attention should be given to restricting head motion and to patient comfort.
Spatial selection
To support multiple scan protocols, several MRSI implementations are required and should include volume-selection and slice- and volume-selective localization methods, using both FID and spin-echo excitation. For volume selection the use of adiabatic excitation (e.g. semi-LASER and LASER) is recommended for ≥3 T to minimize intensity variations across the FOV204 and chemical shift displacement.103 For studies ≥7 T, STEAM volume selection or slice-selected FID excitation is preferred to reduce B1 requirements and CSDE.
Water suppression
Water suppression schemes such as VAPOR or WET205,206 that are robust against variations in transmit B1 field intensity and T1 relaxation times are recommended.
Extracranial lipid signals
Multiple approaches for reducing contamination from extracranial lipids should be available, including use of higher spatial resolutions and postprocessing methods. For outer volume suppression, automatic placement of multiple bands and broadband saturation pulses should be implemented.
Scanner performance
MRI systems can vary in terms of eddy currents, B0 shimming and temporal B0 field stability, and regular maintenance and documentation of these performance parameters is recommended.
RF performance
A minimum of 24 μT transmit field strength at 3 T is recommended for adequate performance of spectroscopy localization pulses, to limit chemical shift displacement artefacts, and to yield short TEs and adiabaticity in semi-LASER. For studies at ≥7 T, multichannel transmit is recommended and the use of spatially selective RF pulses (ie high flip angle refocusing) should be avoided to minimize CSDEs and the effects of B1 inhomogeneities.
6.2.2 ∣. Processing and analysis
Clinical review
Comprehensive methods for review of MRSI results on clinical PACS systems are currently not available and further commercial support for these systems is essential for clinical acceptance. Current options include transfer of metabolite maps or color-coded metabolite overlay images using standard DICOM protocols, and formation of technologist-generated static reports showing sample spectra and metabolite maps.
Image formation
A primary benefit of advanced MRSI methods is that metabolite maps can be generated with a high enough spatial resolution that image review becomes the preferred analysis method. Relative metabolite concentrations should be derived by spectral fitting and all processing steps should be fully automated.
Individual metabolite maps should be corrected for bias-field intensity variations and signal-normalized. Current practice does not indicate a recommended normalization method and the choice is application-dependent, although preferred options include taking metabolite ratios to Cr, normalization by a reference brain region (e.g. mean value in NAWM), or as a ratio to a tissue water reference for pathologies where no change of water content may be anticipated.
Data processing
Standard spectral preprocessing methods should be used and relative metabolite concentrations should be derived by spectral fitting. Attention should be paid to maintaining performance for fitting of spectra with large baseline variations and low SNR, particularly for acquisitions involving larger FOVs. For whole-slice or whole-brain acquisitions, the use of postprocessing or specialized image reconstruction methods for reduction of extracranial lipid contamination is strongly recommended.
Quality control
Regions of poor spectral quality must be identified to avoid misinterpretation. Given the large data sizes of advanced MRSI methods, automated methods for quality control are essential; however, recognition of image artifacts can be difficult and visual review of selected voxels, particularly if indicating potential pathologies, is encouraged. It is recommended that voxels not meeting the quality criteria be set to zero value in the metabolite maps prior to clinical review, although for experienced users with access to appropriate display methods it may be preferable to keep the original metabolite images and display these together with quality maps.
Data analysis
Calculation of both metabolite ratio and individual metabolite maps is recommended, with the data used for image review being application-dependent.
For studies of brain tumors, the use of image analysis methods such as CNI maps or threshold detection of Cho/NAA maps can be used. Further development and evaluation of statistical image analysis methods for specific clinical applications are recommended.
Scanner integration
For clinical acceptance, it is essential that metabolite maps be made available shortly after acquisition. However, the combination of large data sizes and the algorithmic requirements for many image reconstruction and spectral analysis steps mean that on-scanner reconstruction for advanced MRSI methods may, in the short term, be difficult to support by the available scanner hardware, therefore dedicated computation servers may be required. Given that innovative sampling and reconstruction methods continue to be investigated, improved support for integrating custom-built reconstruction pipelines into clinical workflows needs to be provided by the system manufacturers.
Minimum MRSI display requirements must include full 3D image review capabilities with spatial coregistration with other image series, and interactive selection and display of spectra. Integration of MRSI display with interactive spectral selection into PACS systems is identified as an area where further development will be required.
7 ∣. FUTURE DEVELOPMENT
Several ongoing technological developments will benefit future implementations of advanced MRSI methods. Continued improvements of detection sensitivity, including optimization of ultrahigh field MRI systems and improved B1 detection and transmit array technologies, will further extend the trend to detection of an increasing number of metabolites (Figure 7) and higher spatial resolutions. This in turn leads to decreased intra-voxel B0 inhomogeneity and a corresponding increase of the volume of the brain that can be sampled. The use of advanced shimming methods (section 3.3.1) promises to provide improved B0 homogeneity, which together with motion correction will contribute to both greater reliability and spatial coverage. Several studies have already demonstrated that the use of higher spatial resolution acquisitions and improved reconstruction and postprocessing methods will additionally benefit sampling of cortical surface regions.98,116,118 Further developments of under-sampled image encoding and reconstruction strategies will increase the efficiency of MRSI data collection.116,207 With current performance providing voxel dimensions of the order of ~ 3 mm isotropic at ≥7 T for multi-slice studies (Figure 7),98,118 when combined with high-performance volumetric spatial sampling methods, it is anticipated that comparable performance can be obtained within acquisition times from 5 to 15 minutes, depending on the FOV and resolution.
With the development of more robust B0 shimming methods and MRSI methods that can sample a wide brain volume, the scan prescription for MRSI studies will become equivalent to any MRI method, making the acquisition methods feasible to implement in a routine clinical setting. There are also no technical barriers to providing rapid metabolite image reconstruction to provide reconstructed metabolite maps shortly after the end of the data acquisition, as is done for other clinical MRI protocols. It is anticipated that once vendor implementations of MRSI provide acquisition methods that are as easy to use as any MRI sequence, together with a full integration of the processing, then MRSI will play a greater role in clinical studies. One remaining barrier requirement will be the introduction of standardized approaches for image analysis and quality evaluation. These requirements include the development of automated and robust approaches to control for spectral quality and standardization of acquisition protocols that are matched with normal reference values to enable automated image analysis.
8 ∣. CONCLUSION
Historically, variability in data quality, concerns with reproducibility, limitations of detection sensitivity, inefficient sampling of spatial and spectral dimensions, and limitations of review and analysis software, have held back the dissemination of MRSI. As major technological advances of MRI technologies were introduced, such as multichannel detection and transmission and higher magnetic field strengths, these have provided improvements in MRSI data quality that in turn have led to an increasing interest in clinical applications. These ongoing hardware improvements are being combined with novel spatial-spectral sampling and processing methods that provide the level of performance needed to incorporate MRSI into standardized imaging protocols. Further developments are still needed to fully integrate advanced MRSI methods into clinical studies, including robust and automated acquisition methods, efficient processing of the large 4D datasets, and integration with PACS; however, these have largely been implemented in the research setting and there are no technical barriers to these developments. Once implemented, MRSI will become a robust image-based modality suitable for routine use. The clinical success of MRI has benefited from the high degree of automation that has been achieved and immediate presentation of the images for interpretation. Similarly, once MRSI is fully integrated as an “imaging” method it will become more widely used for routine clinical studies.
As with MRI sequences, variants of MRSI implementations are anticipated to address differing clinical requirements, with the available selection not being restricted by instrumentation performance. For this reason, this report has recommended a range of MRSI implementations that reflect the multiple tradeoffs between information content and ease of implementation. In part, these recommendations rely on continued commitment from manufacturers in implementing newer hardware and software that have been demonstrated to improve data quality.
There is ample literature demonstrating the clinical utility of MRSI, which also supports the need to translate advanced MRSI technologies into commercial products. While most of these reports of clinical value have to date been developed by the research community, a more widespread distribution of advanced MRSI technologies is now needed to support future multisite studies that will provide stronger evidence of clinical efficacy.
ACKNOWLEDGEMENTS
To the memory of Dr. Sarah Nelson, an esteemed colleague and tireless investigator of MRSI.
AAM is supported by NIH grants R01EB016064 and RO1CA172210. OCA is supported by NIH grants R01CA211080 and P50CA165962. WB is supported by the Austrian Science Fund (FWF): KLI-718, and P30701. AH is supported by European Research Council grant 679927, SYNAPLAST MR, Horizon 2020 project grant 634541, CDS-QUAMRI and CPRIT Established Researcher Recruitment grant RR180056. SJN is supported by NIH grants R01CA127612, P50CA097257 and P01CA118816. SP is supported by NIH grants 1P30GM122734 and 1R21CA241714. BJS is supported by NIH grant 1R01CA200808.
Funding information
Austrian Science Fund (FWF), Grant/Award Numbers: KLI-718, P30701; CDS-QUAMRI and CPRIT, Grant/Award Number: RR180056; European Research Council, Grant/Award Number: 679927; National Institutes of Health, Grant/Award Numbers: 1P30GM122734, 1R01CA200808, 1R21CA241714, P01CA118816, P50CA097257, P50CA165962, R01CA127612, R01CA211080, R01EB016064, RO1CA172210; SYNAPLAST MR, Horizon 2020 project grant, Grant/Award Number: 634541
Abbreviations used:
- 2HG
D-2-hydroxyglutarate
- CRLB
Cramer-Rao lower bounds
- CSDE
chemical shift displacement error
- CSF
cerebrospinal fluid
- DICOM
digital imaging and communications in medicine
- ECC
eddy current correction
- FOV
field of view
- IDH
isocitrate dehydrogenase
- LASER
localization by adiabatic selective refocusing
- NAWM
normal-appearing white matter
- OVS
outer volume suppression
- PACS
picture archiving and communication system
- PRESS
Point-REsolved SpectoScopy
- ROI
region of interest
- SIVIC
spectroscopic imaging, visualization and computing
- SNR
signal-to-noise ratio
- SOP
service-object pair
- SRF
spatial response function
- SSE
spatial-spectral encoding
- STEAM
STimulated Echo Acquisition Mode
- SVS
single voxel spectroscopy
- VBA
voxel-based analysis
- VOI
volume of interest
REFERENCES
- 1.Oz G, Alger JR, Barker PB, et al. Clinical proton MR spectroscopy in central nervous system disorders. Radiology. 2014;270:658–679. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Barker PB, Bizzi A, De Stefano N, Gullapalli R, Lin DDM. Clinical MR spectroscopy: Techniques and applications. Cambridge, UK: Cambridge University Press; 2009. [Google Scholar]
- 3.Brown TR, Kincaid BM, Ugurbil K. NMR chemical shift imaging in three dimensions. PNAS. 1982;79:3523–3526. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Maudsley AA, Hilal SK, Perman WH, Simon HE. Spatially resolved high-resolution spectroscopy by "four-dimensional" NMR. J Magn Reson. 1983;51: 147–152. [Google Scholar]
- 5.Mansfield P Spatial-mapping of the chemical-shift in NMR. J Phys D Appl Phys. 1983;16:L235–L238. [DOI] [PubMed] [Google Scholar]
- 6.Ricci PE, Pitt A, Keller PJ, Coons SW, Heiserman JE. Effect of voxel position on single-voxel MR spectroscopy findings. Am J Neuroradiol. 2000;21: 367–374. [PMC free article] [PubMed] [Google Scholar]
- 7.Steffen-Smith EA, Venzon DJ, Bent RS, Hipp SJ, Warren KE. Single- and multivoxel proton spectroscopy in pediatric patients with diffuse intrinsic pontine glioma. Int J Radiat Oncol Biol Phys. 2012;84:774–779. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Maudsley AA, Domenig C, Ramsay RE, Bowen BC. Application of volumetric MR spectroscopic imaging for localization of neocortical epilepsy. Epilepsy Res. 2010;88:127–138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kreis R Issues of spectral quality in clinical 1H-magnetic resonance spectroscopy and a gallery of artifacts. NMR Biomed. 2004;17:361–381. [DOI] [PubMed] [Google Scholar]
- 10.Edelstein WA, Glover GH, Hardy CJ, Redington RW. The intrinsic signal-to-noise ratio in NMR imaging. Magn Reson Med. 1986;3:604–618. [DOI] [PubMed] [Google Scholar]
- 11.Bogner W, Gagoski B, Hess AT, et al. 3D GABA imaging with real-time motion correction, shim update and reacquisition of adiabatic spiral MRSI. Neuroimage. 2014;103:290–302. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Zhu H, Edden RA, Ouwerkerk R, Barker PB. High resolution spectroscopic imaging of GABA at 3 Tesla. Magn Reson Med. 2011;65:603–609. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Li Y, Larson P, Chen AP, et al. Short-echo three-dimensional H-1 MR spectroscopic imaging of patients with glioma at 7 Tesla for characterization of differences in metabolite levels. J Magn Reson Imaging. 2015;41:1332–1341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Li Y, Park I, Nelson SJ. Imaging tumor metabolism using in vivo magnetic resonance spectroscopy. Cancer J. 2015;21:123–128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Bendini M, Marton E, Feletti A, et al. Primary and metastatic intraaxial brain tumors: prospective comparison of multivoxel 2D chemical-shift imaging (CSI) proton MR spectroscopy, perfusion MRI, and histopathological findings in a group of 159 patients. Acta Neurochir. 2011;153:403–412. [DOI] [PubMed] [Google Scholar]
- 16.Wang Q, Zhang H, Zhang J, et al. The diagnostic performance of magnetic resonance spectroscopy in differentiating high-from low-grade gliomas: A systematic review and meta-analysis. Eur Radiol. 2016;26:2670–2684. [DOI] [PubMed] [Google Scholar]
- 17.Wang W, Hu Y, Lu P, et al. Evaluation of the diagnostic performance of magnetic resonance spectroscopy in brain tumors: a systematic review and meta-analysis. PLoS ONE. 2014;9:e112577. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Usinskiene J, Ulyte A, Bjornerud A, et al. Optimal differentiation of high- and low-grade glioma and metastasis: a meta-analysis of perfusion, diffusion, and spectroscopy metrics. Neuroradiology. 2016;58:339–350. [DOI] [PubMed] [Google Scholar]
- 19.Dowling C, Bollen AW, Noworolski SM, et al. Preoperative proton MR spectroscopic imaging of brain tumors: correlation with histopathologic analysis of resection specimens. Am J Neuroradiol. 2001;22:604–612. [PMC free article] [PubMed] [Google Scholar]
- 20.McKnight TR, von dem Bussche MH, Vigneron DB, et al. Histopathological validation of a three-dimensional magnetic resonance spectroscopy index as a predictor of tumor presence. J Neurosurg. 2002;97:794–802. [DOI] [PubMed] [Google Scholar]
- 21.Chang SM, Nelson S, Vandenberg S, et al. Integration of preoperative anatomic and metabolic physiologic imaging of newly diagnosed glioma. J Neurooncol. 2009;92:401–415. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Price SJ, Gillard JH. Imaging biomarkers of brain tumour margin and tumour invasion. Br J Radiol. 2011;84:S159–S167. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Guo J, Yao C, Chen H, et al. The relationship between Cho/NAA and glioma metabolism: implementation for margin delineation of cerebral gliomas. Acta Neurochir. 2012;154:1361–1370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.He T, Qiu T, Wang X, et al. Multivoxel magnetic resonance spectroscopy identifies enriched foci of cancer stem-like cells in high-grade gliomas. Onco Targets Ther. 2017;10:195–203. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Chan AA, Lau A, Pirzkall A, et al. Proton magnetic resonance spectroscopy imaging in the evaluation of patients undergoing gamma knife surgery for Grade IV glioma. J Neurosurg. 2004;101:467–475. [DOI] [PubMed] [Google Scholar]
- 26.Pirzkall A, Li X, Oh J, et al. 3D MRSI for resected high-grade gliomas before RT: tumor extent according to metabolic activity in relation to MRI. Int J Radiat Oncol Biol Phys. 2004;59:126–137. [DOI] [PubMed] [Google Scholar]
- 27.Ballangrud AM, Lymberis S, Thakur SB, et al. Magnetic resonance spectroscopy imaging in radiotherapy planning for recurrent glioma. Med Phys. 2011;38:2724–2730. [DOI] [PubMed] [Google Scholar]
- 28.Einstein DB, Wessels B, Bangert B, et al. Phase II trial of radiosurgery to magnetic resonance spectroscopy-defined high-risk tumor volumes in patients with glioblastoma multiforme. Int J Radiat Oncol Biol Phys. 2012;84:668–674. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Anwar M, Molinaro AM, Morin O, et al. Identifying voxels at risk for progression in glioblastoma based on dosimetry, physiologic and metabolic MRI. Radiat Res. 2017;188:303–313. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Muruganandham M, Clerkin PP, Smith BJ, et al. 3-Dimensional magnetic resonance spectroscopic imaging at 3 Tesla for early response assessment of glioblastoma patients during external beam radiation therapy. Int J Radiat Oncol Biol Phys. 2014;90:181–189. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Ken S, Vieillevigne L, Franceries X, et al. Integration method of 3D MR spectroscopy into treatment planning system for glioblastoma IMRT dose painting with integrated simultaneous boost. Radiat Oncol. 2013;8:1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Pirzkall A, McKnight TR, Graves EE, et al. MR-spectroscopy guided target delineation for high-grade gliomas. Int J Radiat Oncol Biol Phys. 2001;50: 915–928. [DOI] [PubMed] [Google Scholar]
- 33.Stadlbauer A, Moser E, Gruber S, et al. Improved delineation of brain tumors: an automated method for segmentation based on pathologic changes of 1H-MRSI metabolites in gliomas. Neuroimage. 2004;23:454–461. [DOI] [PubMed] [Google Scholar]
- 34.Parra NA, Maudsley AA, Gupta RK, et al. Volumetric spectroscopic imaging of glioblastoma multiforme radiation treatment volumes. Int J Radiat Oncol Biol Phys. 2014;90:376–384. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Park I, Tamai G, Lee MC, et al. Patterns of recurrence analysis in newly diagnosed glioblastoma multiforme after three-dimensional conformal radiation therapy with respect to pre-radiation therapy magnetic resonance spectroscopic findings. Int J Radiat Oncol Biol Phys. 2007;69:381–389. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Nelson SJ, Kadambi AK, Park I, et al. Association of early changes in 1H MRSI parameters with survival for patients with newly diagnosed glioblastoma receiving a multimodality treatment regimen. Neuro Oncol. 2017;19:430–439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Nelson SJ, Li Y, Lupo JM, et al. Serial analysis of 3D H-1 MRSI for patients with newly diagnosed GBM treated with combination therapy that includes bevacizumab. J Neurooncol. 2016;130:171–179. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.van Dijken BRJ, van Laar PJ, Holtman GA, van der Hoorn A. Diagnostic accuracy of magnetic resonance imaging techniques for treatment response evaluation in patients with high-grade glioma, a systematic review and meta-analysis. Eur Radiol. 2017;27:4129–4144. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Yang I, Huh NG, Smith ZA, Han SJ, Parsa AT. Distinguishing glioma recurrence from treatment effect after radiochemotherapy and immunotherapy. Neurosurg Clin N Am. 2010;21:181–186. [DOI] [PubMed] [Google Scholar]
- 40.Matsusue E, Fink JR, Rockhill JK, Ogawa T, Maravilla KR. Distinction between glioma progression and post-radiation change by combined physiologic MR imaging. Neuroradiology. 2010;52:297–306. [DOI] [PubMed] [Google Scholar]
- 41.Sawlani V, Taylor R, Rowley K, Redfern R, Martin J, Poptani H. Magnetic resonance spectroscopy for differentiating pseudo-progression from true progression in GBM on concurrent chemoradiotherapy. Neuroradiol J. 2012;25:575–586. [DOI] [PubMed] [Google Scholar]
- 42.Li Y, Lupo JM, Parvataneni R, et al. Survival analysis in patients with newly diagnosed glioblastoma using pre- and postradiotherapy MR spectroscopic imaging. Neuro Oncol. 2013;15:607–617. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Quon H, Brunet B, Alexander A, et al. Changes in serial magnetic resonance spectroscopy predict outcome in high-grade glioma during and after postoperative radiotherapy. Anticancer Res. 2011;31:3559–3565. [PubMed] [Google Scholar]
- 44.Ratai EM, Zhang Z, Snyder BS, et al. Magnetic resonance spectroscopy as an early indicator of response to anti-angiogenic therapy in patients with recurrent glioblastoma: RTOG 0625/ACRIN 6677. Neuro Oncol. 2013;15:936–944. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Cordova JS, Shu HK, Liang Z, et al. Whole-brain spectroscopic MRI biomarkers identify infiltrating margins in glioblastoma patients. Neuro Oncol. 2016;18:1180–1189. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Trattnig S, Springer E, Bogner W, et al. Key clinical benefits of neuroimaging at 7T. Neuroimage. 2018;168:477–489. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Li Y, Lafontaine M, Chang S, Nelson SJ. Comparison between short and long echo time magnetic resonance spectroscopic imaging at 3T and 7T for evaluating brain metabolites in patients with glioma. ACS Chem Neurosci. 2018;9:130–137. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Li X, Vigneron DB, Cha S, et al. Relationship of MR-derived lactate, mobile lipids, and relative blood volume for gliomas in vivo. Am J Neuroradiol. 2005;26:760–769. [PMC free article] [PubMed] [Google Scholar]
- 49.Crisi G 1H MR spectroscopy of meningiomas at 3.0T: the role of glutamate-glutamine complex and glutathione. Neuroradiol J. 2011;24:846–853. [DOI] [PubMed] [Google Scholar]
- 50.Thelwall PE, Yemin AY, Gillian TL, et al. Noninvasive in vivo detection of glutathione metabolism in tumors. Cancer Res. 2005;65:10149–10153. [DOI] [PubMed] [Google Scholar]
- 51.Hattingen E, Raab P, Franz K, Zanella FE, Lanfermann H, Pilatus U. Myo-inositol: a marker of reactive astrogliosis in glial tumors? NMR Biomed. 2008; 21:233–241. [DOI] [PubMed] [Google Scholar]
- 52.Andronesi OC, Kim GS, Gerstner E, et al. Detection of 2-hydroxyglutarate in IDH-mutated glioma patients by in vivo spectral-editing and 2D correlation magnetic resonance spectroscopy. Sci Transl Med. 2012;4:116ra4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Choi C, Raisanen JM, Ganji SK, et al. Prospective longitudinal analysis of 2-hydroxyglutarate magnetic resonance spectroscopy identifies broad clinical utility for the management of patients with IDH-mutant glioma. J Clin Oncol. 2016;34:4030–4039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Jafari-Khouzani K, Loebel F, Bogner W, et al. Volumetric relationship between 2-hydroxyglutarate and FLAIR hyperintensity has potential implications for radiotherapy planning of mutant IDH glioma patients. Neuro Oncol. 2016;18:1569–1578. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Andronesi OC, Loebel F, Bogner W, et al. Treatment response assessment in IDH-mutant glioma patients by noninvasive 3D functional spectroscopic mapping of 2-hydroxyglutarate. Clin Cancer Res. 2016;22:1632–1641. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Choi C, Ganji SK, DeBerardinis RJ, et al. 2-hydroxyglutarate detection by magnetic resonance spectroscopy in IDH-mutated patients with gliomas. Nat Med. 2012;18:624–629. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Strasser B, Gagoski B, Bijaya T, et al. Whole-brain high resolution 3D MRSI for measuring 2HG and tumor metabolism in mutant IDH glioma patients. International Society for Magnetic Resonance in Medicine, Montreal. 2019;7635. [Google Scholar]
- 58.An Z, Tiwari V, Baxter J, et al. 3D high-resolution imaging of 2-hydroxyglutarate in glioma patients using DRAG-EPSI at 3T in vivo. Magn Reson Med. 2019;81:795–802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Pan JW, Kuzniecky RI. Utility of magnetic resonance spectroscopic imaging for human epilepsy. Quant Imaging Med Surg. 2015;5:313–322. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Mueller SG, Laxer KD, Barakos JA, et al. Metabolic characteristics of cortical malformations causing epilepsy. J Neurol. 2005;252:1082–1092. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Pan JW, Duckrow RB, Gerrard J, et al. 7T MR spectroscopic imaging in the localization of surgical epilepsy. Epilepsia. 2013;54:1668–1678. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Hattingen E, Luckerath C, Pellikan S, et al. Frontal and thalamic changes of GABA concentration indicate dysfunction of thalamofrontal networks in juvenile myoclonic epilepsy. Epilepsia. 2014;55:1030–1037. [DOI] [PubMed] [Google Scholar]
- 63.Govind V, Gold S, Kaliannan K, et al. Whole-brain proton MR spectroscopic imaging of mild-to-moderate traumatic brain injury and correlation with neuropsychological deficits. J Neurotrauma. 2010;27:483–496. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Maudsley AA, Govind V, Levin B, Saigal G, Harris L, Sheriff S. Distributions of magnetic resonance diffusion and spectroscopy measures with traumatic brain injury. J Neurotrauma. 2015;32:1056–1063. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Maudsley AA, Govind V, Saigal G, Gold SG, Harris L, Sheriff S. Longitudinal MR spectroscopy shows altered metabolism in traumatic brain injury. J Neuroimaging. 2017;27:562–569. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Brown M, Baradaran H, Christos PJ, Wright D, Gupta A, Tsiouris AJ. Magnetic resonance spectroscopy abnormalities in traumatic brain injury: A meta-analysis. J Neuroradiol. 2018;45:123–129. [DOI] [PubMed] [Google Scholar]
- 67.Cohen BA, Inglese M, Rusinek H, Babb JS, Grossman RI, Gonen O. Proton MR spectroscopy and MRI-volumetry in mild traumatic brain injury. Am J Neuroradiol. 2007;28:907–913. [PMC free article] [PubMed] [Google Scholar]
- 68.Yeo RA, Phillips JP, Jung RE, Brown AJ, Campbell RC, Brooks WM. Magnetic resonance spectroscopy detects brain injury and predicts cognitive functioning in children with brain injuries. J Neurotrauma. 2006;23:1427–1435. [DOI] [PubMed] [Google Scholar]
- 69.McKee AC, Daneshvar DH. The neuropathology of traumatic brain injury. Handb Clin Neurol. 2015;127:45–66. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Fleischer V, Kolb R, Groppa S, Zipp F, Klose U, Gröger A. Metabolic patterns in chronic multiple sclerosis lesions and normal-appearing white matter: intra-individual comparison by using 2D MR spectroscopic imaging. Radiology. 2016;281:536–543. [DOI] [PubMed] [Google Scholar]
- 71.Moccia M, Ciccarelli O. Molecular and metabolic imaging in multiple sclerosis. Neuroimaging Clin N Am. 2017;27:343–356. [DOI] [PubMed] [Google Scholar]
- 72.Heckova E, Strasser B, Hangel GJ, et al. 7 T magnetic resonance spectroscopic imaging in multiple sclerosis: How does spatial resolution affect the detectability of metabolic changes in brain lesions? Invest Radiol. 2019;54:247–254. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73.Rovira A, Alonso J. 1H magnetic resonance spectroscopy in multiple sclerosis and related disorders. Neuroimaging Clin N Am. 2013;23:459–474. [DOI] [PubMed] [Google Scholar]
- 74.Prinsen H, de Graaf RA, Mason GF, Pelletier D, Juchem C. Reproducibility measurement of glutathione, GABA, and glutamate: Towards in vivo neurochemical profiling of multiple sclerosis with MR spectroscopy at 7T. J Magn Reson Imaging. 2017;45:187–198. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 75.Srinivasan R, Ratiney H, Hammond-Rosenbluth KE, Pelletier D, Nelson SJ. MR spectroscopic imaging of glutathione in the white and gray matter at 7 T with an application to multiple sclerosis. Magn Reson Imaging. 2010;28:163–170. [DOI] [PubMed] [Google Scholar]
- 76.Craven L, Alston CL, Taylor RW, Turnbull DM. Recent advances in mitochondrial disease. Annu Rev Genomics Hum Genet. 2017;18:257–275. [DOI] [PubMed] [Google Scholar]
- 77.Lin DD, Crawford TO, Barker PB. Proton MR spectroscopy in the diagnostic evaluation of suspected mitochondrial disease. Am J Neuroradiol. 2003; 24:33–41. [PMC free article] [PubMed] [Google Scholar]
- 78.Weiduschat N, Kaufmann P, Mao X, et al. Cerebral metabolic abnormalities in A3243G mitochondrial DNA mutation carriers. Neurology. 2014;82:798–805. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 79.Kaufmann P, Shungu DC, Sano MC, et al. Cerebral lactic acidosis correlates with neurological impairment in MELAS. Neurology. 2004;62:1297–1302. [DOI] [PubMed] [Google Scholar]
- 80.Quadrelli S, Mountford C, Ramadan S. Systematic review of in-vivo neuro magnetic resonance spectroscopy for the assessment of posttraumatic stress disorder. Psychiatry Res Neuroimaging. 2018;282:110–125. [DOI] [PubMed] [Google Scholar]
- 81.Moriguchi S, Takamiya A, Noda Y, et al. Glutamatergic neurometabolite levels in major depressive disorder: a systematic review and meta-analysis of proton magnetic resonance spectroscopy studies. Mol Psychiatry. 2018;24:952–964. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Chandra A, Dervenoulas G, Politis M. Magnetic resonance imaging in Alzheimer's disease and mild cognitive impairment. J Neurol. 2019;266:1293–1302. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Wang AM, Pradhan S, Coughlin JM, et al. Assessing brain metabolism with 7-T proton magnetic resonance spectroscopy in patients with first-episode psychosis. JAMA Psychiat. 2019;76:314–323. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Li Y, Jakary A, Gillung E, et al. Evaluating metabolites in patients with major depressive disorder who received mindfulness-based cognitive therapy and healthy controls using short echo MRSI at 7 Tesla. Magn Reson Mater Phy. 2016;29:523–533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Xin L, Gambarota G, Mlynarik V, Gruetter R. Proton T-2 relaxation time of J-coupled cerebral metabolites in rat brain at 9.4T. NMR Biomed. 2008;21:396–401. [DOI] [PubMed] [Google Scholar]
- 86.Maudsley AA, Domenig C, Govind V, et al. Mapping of brain metabolite distributions by volumetric proton MR spectroscopic imaging (MRSI). Magn Reson Med. 2009;61:548–559. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Posse S, DeCarli C, Le Bihan D. Three-dimensional echo-planar MR spectroscopic imaging at short echo times in the human brain. Radiology. 1994;192:733–738. [DOI] [PubMed] [Google Scholar]
- 88.Henning A, Fuchs A, Murdoch JB, Boesiger P. Slice-selective FID acquisition, localized by outer volume suppression (FIDLOVS) for (1)H-MRSI of the human brain at 7 T with minimal signal loss. NMR Biomed. 2009;22:683–696. [DOI] [PubMed] [Google Scholar]
- 89.Gruber S, Heckova E, Strasser B, et al. Mapping an extended neurochemical profile at 3 and 7 T using accelerated high-resolution proton magnetic resonance spectroscopic imaging. Invest Radiol. 2017;52:631–639. [DOI] [PubMed] [Google Scholar]
- 90.Boer VO, Siero JC, Hoogduin H, et al. High-field MRS of the human brain at short TE and TR. NMR Biomed. 2011;24:1081–1088. [DOI] [PubMed] [Google Scholar]
- 91.Chang P, Nassirpour S, Avdievitch N, Henning A. Non-water-suppressed H-1 FID-MRSI at 3T and 9.4T. Magn Reson Med. 2018;80:442–451. [DOI] [PubMed] [Google Scholar]
- 92.Bogner W, Gruber S, Trattnig S, Chmelik M. High-resolution mapping of human brain metabolites by free induction decay 1H MRSI at 7 T. NMR Biomed. 2012;25:873–882. [DOI] [PubMed] [Google Scholar]
- 93.Boer VO, Klomp DW, Juchem C, Luijten PR, de Graaf RA. Multislice 1H MRSI of the human brain at 7 T using dynamic B0 and B1 shimming. Magn Reson Med. 2012;68:662–670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Boer VO, van de Lindt T, Luijten PR, Klomp DW. Lipid suppression for brain MRI and MRSI by means of a dedicated crusher coil. Magn Reson Med. 2015;73:2062–2068. [DOI] [PubMed] [Google Scholar]
- 95.Nassirpour S, Chang P, Kirchner T. Henning A. Over-discretized SENSE reconstruction and B0 correction for accelerated non-lipid-suppressed 1H FID MRSI of the human brain at 9.4 T. NMR Biomed. 2018;31:e4014. [DOI] [PubMed] [Google Scholar]
- 96.Povazan M, Hangel G, Strasser B, et al. Mapping of brain macromolecules and their use for spectral processing of H-1-MRSI data with an ultra-short acquisition delay at 7 T. Neuroimage. 2015;121:126–135. [DOI] [PubMed] [Google Scholar]
- 97.Mullins PG, McGonigle DJ, O'Gorman RL, et al. Current practice in the use of MEGA-PRESS spectroscopy for the detection of GABA. Neuroimage. 2014;86:43–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Nassirpour S, Chang P, Henning A. High and ultra-high resolution metabolite mapping of the human brain using 1H FID MRSI at 9.4T. Neuroimage. 2018;168:211–221. [DOI] [PubMed] [Google Scholar]
- 99.Moser P, Hingerl L, Strasser B, et al. Whole-slice mapping of GABA and GABA+ at 7T via adiabatic MEGA-editing, real-time instability correction, and concentric circle readout. Neuroimage. 2019;184:475–489. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Pohmann R, von Kienlin M, Haase A. Theoretical evaluation and comparison of fast chemical shift imaging methods. J Magn Reson. 1997;129: 145–160. [DOI] [PubMed] [Google Scholar]
- 101.Oz G, Tkac I. Short-echo, single-shot, full-intensity proton magnetic resonance spectroscopy for neurochemical profiling at 4 T: validation in the cerebellum and brainstem. Magn Reson Med. 2011;65:901–910. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102.Andronesi OC, Ramadan S, Ratai EM, Jennings D, Mountford CE, Sorensen AG. Spectroscopic imaging with improved gradient modulated constant adiabaticity pulses on high-field clinical scanners. J Magn Reson. 2010;203:283–293. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Scheenen TW, Klomp DW, Wijnen JP, Heerschap A. Short echo time 1H-MRSI of the human brain at 3T with minimal chemical shift displacement errors using adiabatic refocusing pulses. Magn Reson Med. 2008;59:1–6. [DOI] [PubMed] [Google Scholar]
- 104.Duyn JH, Gillen J, Sobering G, van Zijl PC, Moonen CT. Multisection proton MR spectroscopic imaging of the brain. Radiology. 1993;188:277–282. [DOI] [PubMed] [Google Scholar]
- 105.Osorio JA, Xu D, Cunningham CH, et al. Design of cosine modulated very selective suppression pulses for MR spectroscopic imaging at 3T. Magn Reson Med. 2009;61:533–540. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Henning A, Schar M, Schulte RF, et al. SELOVS: brain MRSI localization based on highly selective T1- and B1- insensitive outer-volume suppression at 3T. Magn Reson Med. 2008;59:40–51. [DOI] [PubMed] [Google Scholar]
- 107.Spielman DM, Pauly JM, Macovski A, Glover GH, Enzmann DR. Lipid-suppressed single- and multisection proton spectroscopic imaging of the human brain. J Magn Reson Imaging. 1992;2:253–262. [DOI] [PubMed] [Google Scholar]
- 108.Martinez-Ramon M, Gallardo-Antolin A, Cid-Sueiro J, et al. Automatic placement of outer volume suppression slices in MR spectroscopic imaging of the human brain. Magn Reson Med. 2010;63:592–600. [DOI] [PubMed] [Google Scholar]
- 109.Ozhinsky E, Vigneron DB, Nelson SJ. Improved spatial coverage for brain 3D PRESS MRSI by automatic placement of outer-volume suppression saturation bands. J Magn Reson Imaging. 2011;33:792–802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.de Graaf RA, Brown PB, De Feyter HM, McIntyre S, Nixon TW. Elliptical localization with pulsed second-order fields (ECLIPSE) for robust lipid suppression in proton MRSI. NMR Biomed. 2018;31:e3949. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 111.Sarkar S, Heberlein K, Hu X. Truncation artifact reduction in spectroscopic imaging using a dual-density spiral k-space trajectory. Magn Reson Imaging. 2002;20:743–757. [DOI] [PubMed] [Google Scholar]
- 112.Adalsteinsson E, Star-Lack J, Meyer CH, Spielman DM. Reduced spatial side lobes in chemical-shift imaging. Magn Reson Med. 1999;42:314–323. [DOI] [PubMed] [Google Scholar]
- 113.Zhang SH, Maier SE, Panych LP. Improved spatial localization in magnetic resonance spectroscopic imaging with two-dimensional PSF-Choice encoding. J Magn Reson. 2018;290:18–28. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 114.Panych LP, Roebuck JR, Chen NK, et al. Investigation of the PSF-choice method for reduced lipid contamination in prostate MR spectroscopic imaging. Magn Reson Med. 2012;68:1376–1382. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Kirchner T, Fillmer A, Tsao J, Pruessmann KP, Henning A. Reduction of voxel bleeding in highly accelerated parallel 1H MRSI by direct control of the spatial response function. Magn Reson Med. 2015;73:469–480. [DOI] [PubMed] [Google Scholar]
- 116.Ma C, Lam F, Ning Q, Johnson CL, Liang ZP. High-resolution 1H-MRSI of the brain using short-TE SPICE. Magn Reson Med. 2017;77:467–479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117.Klauser A, Courvoisier S, Kasten J, et al. Fast high-resolution brain metabolite mapping on a clinical 3T MRI by accelerated 1 H-FID-MRSI and low-rank constrained reconstruction. Magn Reson Med. 2019;81:2841–2857. [DOI] [PubMed] [Google Scholar]
- 118.Hangel G, Strasser B, Povazan M, et al. Ultra-high resolution brain metabolite mapping at 7 T by short-TR Hadamard-encoded FID-MRSI. Neuroimage. 2018;168:199–210. [DOI] [PubMed] [Google Scholar]
- 119.Nassirpour S, Chang P, Avdievitch N, Henning A. Compressed sensing for high-resolution nonlipid suppressed 1H FID MRSI of the human brain at 9.4T. Magn Reson Med. 2018;80:2311–2325. [DOI] [PubMed] [Google Scholar]
- 120.Dydak U, Weiger M, Pruessmann KP, Meier D, Boesiger P. Sensitivity-encoded spectroscopic imaging. Magn Reson Med. 2001;46:713–722. [DOI] [PubMed] [Google Scholar]
- 121.Strasser B, Povazan M, Hangel G, et al. (2+1)D-CAIPIRINHA accelerated MR spectroscopic imaging of the brain at 7T. Magn Reson Med. 2017;78:429–440. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 122.Matsui S, Sekihara K, Kohno H. High-speed spatially resolved NMR spectroscopy using phase-modulated spin-echo trains. Expansion of the spectral bandwidth by combined use of delayed spin-echo trains. J Magn Reson. 1985;64:167–171. [Google Scholar]
- 123.Schirda CV, Zhao T, Andronesi OC, et al. In vivo brain rosette spectroscopic imaging (RSI) with LASER excitation, constant gradient strength readout, and automated LCModel quantification for all voxels. Magn Reson Med. 2016;76:380–390. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 124.Furuyama JK, Wilson NE, Thomas MA. Spectroscopic imaging using concentrically circular echo-planar trajectories in vivo. Magn Reson Med. 2012; 67:1515–1522. [DOI] [PubMed] [Google Scholar]
- 125.Lin FH, Tsai SY, Otazo R, et al. Sensitivity-encoded (SENSE) proton echo-planar spectroscopic imaging (PEPSI) in the human brain. Magn Reson Med. 2007;57:249–257. [DOI] [PubMed] [Google Scholar]
- 126.Larson PEZ, Bok R, Kerr AB, et al. Investigation of tumor hyperpolarized [1-C-13]-pyruvate dynamics using time-resolved multiband RF excitation echo-planar MRSI. Magn Reson Med. 2010;63:582–591. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 127.Zhu X, Ebel A, Ji JX, Schuff N. Spectral phase-corrected GRAPPA reconstruction of three-dimensional echo-planar spectroscopic imaging (3D-EPSI). Magn Reson Med. 2007;57:815–820. [DOI] [PubMed] [Google Scholar]
- 128.Tsai SY, Otazo R, Posse S, et al. Accelerated proton echo planar spectroscopic imaging (PEPSI) using GRAPPA with a 32-channel phased-array coil. Magn Reson Med. 2008;59:989–998. [DOI] [PubMed] [Google Scholar]
- 129.Mayer D, Kim DH, Spielman DM, Bammer R. Fast parallel spiral chemical shift imaging at 3T using iterative SENSE reconstruction. Magn Reson Med. 2008;59:891–897. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 130.Shankar RV, Chang JC, Hu HCH, Kodibagkar VD. Fast data acquisition techniques in magnetic resonance spectroscopic imaging. NMR Biomed. 2019;32:e4046. [DOI] [PubMed] [Google Scholar]
- 131.Hingerl L, Bogner W, Moser P, et al. Density-weighted concentric circle trajectories for high resolution brain magnetic resonance spectroscopic imaging at 7T. Magn Reson Med. 2018;79:2874–2885. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132.Adalsteinsson E, Irarrazabal P, Topp S, Meyer C, Macovski A, Spielman DM. Volumetric spectroscopic imaging with spiral-based k-space trajectories. Magn Reson Med. 1998;39:889–898. [DOI] [PubMed] [Google Scholar]
- 133.Andronesi OC, Gagoski BA, Sorensen AG. Neurologic 3D MR spectroscopic imaging with low-power adiabatic pulses and fast spiral acquisition. Radiology. 2012;262:647–661. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 134.Li BS, Regal J, Gonen O. SNR versus resolution in 3D 1H MRS of the human brain at high magnetic fields. Magn Reson Med. 2001;46:1049–1053. [DOI] [PubMed] [Google Scholar]
- 135.Ebel A, Maudsley AA. Improved spectral quality for 3D MR spectroscopic imaging using a high spatial resolution acquisition strategy. Magn Reson Imaging. 2003;21:113–120. [DOI] [PubMed] [Google Scholar]
- 136.Gruber S, Mlynarik V, Moser E. High-resolution 3D proton spectroscopic imaging of the human brain at 3 T: SNR issues and application for anatomy-matched voxel sizes. Magn Reson Med. 2003;49:299–306. [DOI] [PubMed] [Google Scholar]
- 137.Sabati M, Sheriff S, Gu M, et al. Multivendor implementation and comparison of volumetric whole-brain echo-planar MR spectroscopic imaging. Magn Reson Med. 2015;74:1209–1220. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Nassirpour S, Chang P, Fillmer A, Henning A. A comparison of optimization algorithms for localized in vivo B0 shimming. Magn Reson Med. 2018;79:1145–1156. [DOI] [PubMed] [Google Scholar]
- 139.Han H, Song AW, Truong TK. Integrated parallel reception, excitation, and shimming (iPRES). Magn Reson Med. 2013;70:241–247. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 140.Stockmann JP, Witzel T, Keil B, et al. A 32-channel combined RF and B0 shim array for 3T brain imaging. Magn Reson Med. 2016;75:441–451. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 141.Pan JW, Lo KM, Hetherington HP. Role of very high order and degree B0 shimming for spectroscopic imaging of the human brain at 7 Tesla. Magn Reson Med. 2012;68:1007–1017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 142.Ebel A, Maudsley AA. Detection and correction of frequency instabilities for volumetric 1H echo-planar spectroscopic imaging. Magn Reson Med. 2005;53:465–469. [DOI] [PubMed] [Google Scholar]
- 143.Lange T, Zaitsev M, Buechert M. Correction of frequency drifts induced by gradient heating in 1H spectra using interleaved reference spectroscopy. J Magn Reson Imaging. 2011;33:748–754. [DOI] [PubMed] [Google Scholar]
- 144.Godenschweger F, Kagebein U, Stucht D, et al. Motion correction in MRI of the brain. Phys Med Biol. 2016;61:R32–R56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 145.Posse S, Cuenod CA, Le Bihan D. Motion artifact compensation in 1H spectroscopic imaging by signal tracking. J Magn Reson B. 1993;102:222–227. [Google Scholar]
- 146.Hess AT, Andronesi OC, Tisdall MD, et al. Real-time motion and B0 correction for localized adiabatic selective refocusing (LASER) MRSI using echo planar imaging volumetric navigators. NMR Biomed. 2012;25:347–358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 147.Posse S, Otazo R, Caprihan A, et al. Proton echo-planar spectroscopic imaging of J-coupled resonances in human brain at 3 and 4 Tesla. Magn Reson Med. 2007;58:236–244. [DOI] [PubMed] [Google Scholar]
- 148.Otazo R, Tsai SY, Lin FH, Posse S. Accelerated short-TE 3D proton echo-planar spectroscopic imaging using 2D-SENSE with a 32-channel array coil. Magn Reson Med. 2007;58:1107–1116. [DOI] [PubMed] [Google Scholar]
- 149.Zhao X, Prost RW, Li Z, Li SJ. Reduction of artifacts by optimization of the sensitivity map in sensitivity-encoded spectroscopic imaging. Magn Reson Med. 2005;53:30–34. [DOI] [PubMed] [Google Scholar]
- 150.Strasser B, Chmelik M, Robinson SD, et al. Coil combination of multichannel MRSI data at 7 T: MUSICAL. NMR Biomed. 2013;26:1796–1805. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Dong Z, Peterson B. The rapid and automatic combination of proton MRSI data using multi-channel coils without water suppression. Magn Reson Imaging. 2007;25:1148–1154. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 152.Maril N, Lenkinski RE. An automated algorithm for combining multivoxel MRS data acquired with phased-array coils. J Magn Reson Imaging. 2005;21:317–322. [DOI] [PubMed] [Google Scholar]
- 153.Rodgers CT, Robson MD. Receive array magnetic resonance spectroscopy: Whitened singular value decomposition (WSVD) gives optimal Bayesian solution. Magn Reson Med. 2010;63:881–891. [DOI] [PubMed] [Google Scholar]
- 154.Brown MA. Time-domain combination of MR spectroscopy data acquired using phased-array coils. Magn Reson Med. 2004;52:1207–1213. [DOI] [PubMed] [Google Scholar]
- 155.Abdoli A, Maudsley AA. Phased-array combination for MR spectroscopic imaging using a water reference. Magn Reson Med. 2016;76:733–741. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 156.Haupt CI, Schuff N, Weiner MW, Maudsley AA. Removal of lipid artifacts in 1H spectroscopic imaging by data extrapolation. Magn Reson Med. 1996;35:678–687. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 157.Tsai SY, Lin YR, Lin HY, Lin FH. Reduction of lipid contamination in MR spectroscopy imaging using signal space projection. Magn Reson Med. 2019;81:1486–1498. [DOI] [PubMed] [Google Scholar]
- 158.Bilgic B, Chatnuntawech I, Fan AP, et al. Fast image reconstruction with L2-regularization. J Magn Reson Imaging. 2014;40:181–191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 159.Laudadio T, Mastronardi N, Vanhamme L, Van Hecke P, Van Huffel S. Improved Lanczos algorithms for blackbox MRS data quantitation. J Magn Reson. 2002;157:292–297. [DOI] [PubMed] [Google Scholar]
- 160.Liu Y, Ma C, Clifford BA, et al. Improved low-rank filtering of magnetic resonance spectroscopic imaging data corrupted by noise and B0 field inhomogeneity. IEEE Trans Biomed Eng. 2016;63:841–849. [DOI] [PubMed] [Google Scholar]
- 161.Zhu XP, Du AT, Jahng GH, et al. Magnetic resonance spectroscopic imaging reconstruction with deformable shape-intensity models. Magn Reson Med. 2003;50:474–482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 162.Abdoli A, Stoyanova R, Maudsley AA. Denoising of MR spectroscopic imaging data using statistical selection of principal components. Magn Reson Mater Phy. 2016;29:811–822. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 163.Provencher SW. Estimation of metabolite concentrations from localized in vivo proton NMR spectra. Magn Reson Med. 1993;30:672–679. [DOI] [PubMed] [Google Scholar]
- 164.Soher BJ, Young K, Govindaraju V, Maudsley AA. Automated spectral analysis III: Application to in vivo proton MR spectroscopy and spectroscopic imaging. Magn Reson Med. 1998;40:822–831. [DOI] [PubMed] [Google Scholar]
- 165.Poullet JB, Sima DM, Van Huffel S. MRS signal quantitation: a review of time- and frequency-domain methods. J Magn Reson. 2008;195:134–144. [DOI] [PubMed] [Google Scholar]
- 166.Bhat H, Sajja BR, Narayana PA. Fast quantification of proton magnetic resonance spectroscopic imaging with artificial neural networks. J Magn Reson. 2006;183:110–122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167.Gurbani SS, Sheriff S, Maudsley AA, Shim H, Cooper LAD. Incorporation of a spectral model in a convolutional neural network for accelerated spectral fitting. Magn Reson Med. 2019;81:3346–3357. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 168.Ebel A, Soher BJ, Maudsley AA. Assessment of 3D 1H NMR echo-planar spectroscopic imaging using automated spectral analysis. Magn Reson Med. 2001;46:1072–1078. [DOI] [PubMed] [Google Scholar]
- 169.Soher BJ, Maudsley AA. Evaluation of variable line-shape models and prior information in automated 1H spectroscopic imaging analysis. Magn Reson Med. 2004;52:1246–1254. [DOI] [PubMed] [Google Scholar]
- 170.Kornak J, Young K, Soher BJ, Maudsley AA. Bayesian k-space-time reconstruction of MR spectroscopic imaging for enhanced resolution. IEEE Trans Med Imaging. 2010;29:1333–1350. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 171.Laruelo A, Chaari L, Tourneret JY, et al. Spatio-spectral regularization to improve magnetic resonance spectroscopic imaging quantification. NMR Biomed. 2016;29:918–931. [DOI] [PubMed] [Google Scholar]
- 172.Kelm BM, Kaster FO, Henning A, et al. Using spatial prior knowledge in the spectral fitting of MRS images. NMR Biomed. 2012;25:1–13. [DOI] [PubMed] [Google Scholar]
- 173.Kreis R The trouble with quality filtering based on relative Cramer-Rao lower bounds. Magn Reson Med. 2016;75:15–18. [DOI] [PubMed] [Google Scholar]
- 174.Ladroue C, Howe FA, Griffiths JR, Tate AR. Independent component analysis for automated decomposition of in vivo magnetic resonance spectra. Magn Reson Med. 2003;50:697–703. [DOI] [PubMed] [Google Scholar]
- 175.Kyathanahally SP, Doring A, Kreis R. Deep learning approaches for detection and removal of ghosting artifacts in MR spectroscopy. Magn Reson Med. 2018;80:851–863. [DOI] [PubMed] [Google Scholar]
- 176.Gurbani SS, Schreibmann E, Maudsley AA, et al. A convolutional neural network to filter artifacts in spectroscopic MRI. Magn Reson Med. 2018;80:1765–1775. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 177.McNab JA, Bartha R. Quantitative short echo-time 1H LASER-CSI in human brain at 4 T. NMR Biomed. 2006;19:999–1009. [DOI] [PubMed] [Google Scholar]
- 178.Jansen JF, Backes WH, Nicolay K, Kooi ME. 1H MR spectroscopy of the brain: absolute quantification of metabolites. Radiology. 2006;240:318–332. [DOI] [PubMed] [Google Scholar]
- 179.Maudsley AA. Lesion segmentation for MR spectroscopic imaging using the convolution difference method. Magn Reson Med. 2019;81:1499–1510. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180.Elias AE, Carlos RC, Smith EA, et al. MR spectroscopy using normalized and non-normalized metabolite ratios for differentiating recurrent brain tumor from radiation injury. Acad Radiol. 2011;18:1101–1108. [DOI] [PubMed] [Google Scholar]
- 181.Maudsley AA, Govind V, Arheart KL. Associations of age, gender and body mass with 1H MR-observed brain metabolites and tissue distributions. NMR Biomed. 2012;25:580–593. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Maudsley AA, Roy B, Gupta RK, et al. Association of metabolite concentrations and water diffusivity in normal appearing brain tissue with glioma grade. J Neuroimaging. 2014;24:585–589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 183.Gasparovic C, Neeb H, Feis DL, et al. Quantitative spectroscopic imaging with in situ measurements of tissue water T1, T2, and density. Magn Reson Med. 2009;62:583–590. [DOI] [PubMed] [Google Scholar]
- 184.Gasparovic C, Song T, Devier D, et al. Use of tissue water as a concentration reference for proton spectroscopic imaging. Magn Reson Med. 2006;55:1219–1226. [DOI] [PubMed] [Google Scholar]
- 185.Pfefferbaum A, Adalsteinsson E, Spielman D, Sullivan EV, Lim KO. In vivo spectroscopic quantification of the N-acetyl moiety, creatine, and choline from large volumes of brain gray and white matter: Effects of normal aging. Magn Reson Med. 1999;41:276–284. [DOI] [PubMed] [Google Scholar]
- 186.NEMA. http://dicom.nema.org/medical/dicom/current/output/html/part03.html. Accessed August 29, 2018.
- 187.Wilson M, Reynolds G, Kauppinen RA, Arvanitis TN, Peet AC. A constrained least-squares approach to the automated quantitation of in vivo (1)H magnetic resonance spectroscopy data. Magn Reson Med. 2011;65:1–12. [DOI] [PubMed] [Google Scholar]
- 188.TARQUIN. http://tarquin.sourceforge.net. Accessed August 29, 2018.
- 189.Crane JC, Olson MP, Nelson SJ. SIVIC: open-source, standards-based software for DICOM MR spectroscopy workflows. Int J Biomed Imaging. 2013; 2013:169526. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 190.SIVIC. https://sourceforge.net/projects/sivic. Accessed August 29, 2018.
- 191.JMRUI. http://www.jmrui.eu. Accessed August 29, 2018.
- 192.Goryawala MZ, Sheriff S, Maudsley AA. Regional distributions of brain glutamate and glutamine in normal subjects. NMR Biomed. 2016;29:1108–1116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 193.Hetherington HP, Pan JW, Mason GF, et al. Quantitative 1H spectroscopic imaging of human brain at 4.1 T using image segmentation. Magn Reson Med. 1996;36:21–29. [DOI] [PubMed] [Google Scholar]
- 194.Goryawala MZ, Sheriff S, Stoyanova R, Maudsley AA. Spectral decomposition for resolving partial volume effects in MRSI. Magn Reson Med. 2018;79:2886–2895. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 195.Mandl RC, van den Heuvel MP, Klomp DW, et al. Tract-based magnetic resonance spectroscopy of the cingulum bundles at 7 T. Hum Brain Mapp. 2012;33:1503–1511. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 196.Govind V, Sharma KR, Maudsley AA, Arheart KL, Saigal G, Sheriff S. Comprehensive evaluation of corticospinal tract metabolites in amyotrophic lateral sclerosis using whole-brain 1H MR spectroscopy. PLoS ONE. 2012;7:e35607. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 197.Su Y, Thakur SB, Karimi S, et al. Spectrum separation resolves partial-volume effect of MRSI as demonstrated on brain tumor scans. NMR Biomed. 2008;21:1030–1042. [DOI] [PubMed] [Google Scholar]
- 198.Szabo de Edelenyi F, Simonetti AW, Postma G, Huo R, Buydens LMC. Application of independent component analysis to 1H MR spectroscopic imaging exams of brain tumours. Anal Chim Acta. 2005;544:36–46. [Google Scholar]
- 199.McKnight TR, Noworolski SM, Vigneron DB, Nelson SJ. An automated technique for the quantitative assessment of 3D-MRSI data from patients with glioma. J Magn Reson Imaging. 2001;13:167–177. [DOI] [PubMed] [Google Scholar]
- 200.Signorini M, Paulesu E, Friston K, et al. Rapid assessment of regional cerebral metabolic abnormalities in single subjects with quantitative and non-quantitative [18F]FDG PET: A clinical validation of statistical parametric mapping. Neuroimage. 1999;9:63–80. [DOI] [PubMed] [Google Scholar]
- 201.Perez-Ruiz A, Julia-Sape M, Mercadal G, et al. The INTERPRET decision-support system version 3.0 for evaluation of magnetic resonance spectroscopy data from human brain tumours and other abnormal brain masses. BMC Bioinformatics. 2011;11:581. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 202.De Edelenyi FS, Rubin C, Estève F, et al. A new approach for analyzing proton magnetic resonance spectroscopic images of brain tumors: nosologic images. NatMed. 2000;6:1287–1289. [DOI] [PubMed] [Google Scholar]
- 203.Raschke F, Barrick TR, Jones TL, Yang G, Ye X, Howe FA. Tissue-type mapping of gliomas. Neuroimage Clin. 2019;21:101648. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 204.Wilson M, Andronesi OC, Barker PB, et al. A methodological consensus on clinical proton MR spectroscopy of the brain: review and recommendations. Magn Reson Med. 2019;82:527–550. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 205.Tkac I, Starcuk Z, Choi IY, Gruetter R. In vivo 1H NMR spectroscopy of rat brain at 1 ms echo time. Magn Reson Med. 1999;41:649–656. [DOI] [PubMed] [Google Scholar]
- 206.Ogg RJ, Kingsley PB, Taylor JS. WET, a T1-and B1-intensive water-suppression method for in vivo localized 1H NMR spectroscopy. J Magn Reson. 1994;104:1–10. [DOI] [PubMed] [Google Scholar]
- 207.Cao P, Shin PJ, Park I, et al. Accelerated high-bandwidth MR spectroscopic imaging using compressed sensing. Magn Reson Med. 2016;76:369–379. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 208.Hingerl L, Strasser B, Moser P, et al. Towards full-brain FID-MRSI at 7T with 3D concentric circle readout trajectories. International Society for Magnetic Resonance in Medicine, Toronto, 2019; 618. [Google Scholar]
- 209.Magnusson PO, Boer VO, Marsman A, et al. Gamma-aminobutyric acid edited echo-planar spectroscopic imaging (EPSI) with MEGA-sLASER at 7T. Magn Reson Med. 2018;81:773–780. [DOI] [PMC free article] [PubMed] [Google Scholar]