This guideline summarizes updated safety data (2017–2025) and provides expert recommendations on the use of low intensity transcranial electrical stimulation (tES) in humans. tES encompasses several techniques including transcranial direct current stimulation (tDCS), oscillatory transcranial direct current stimulation (otDCS), transcranial alternating current stimulation (tACS), transcranial random noise stimulation (tRNS), transcranial temporal interference stimulation (tTIS), and their combinations or variations. Across over 300,000 sessions involving healthy individuals, patients with neuropsychiatric conditions, and other clinical populations, no tES-related serious adverse events (AEs) have been reported. Moderate AEs are rare and limited to a small range of specific applications. Mild AEs are common and include transient symptoms such as localized sensations (e.g., tingling or burning), headaches, and fatigue. Similar mild AEs are also reported by individuals receiving placebo stimulation. The frequency, magnitude, and type of AEs are comparable across healthy, clinical, and vulnerable groups, including children, elderly, or pregnant women. Combined interventions (e.g., co-application with EEG, TMS, or neuroimaging) have not shown increased safety risks. Safety is well-established for both bipolar and multichannel tES when applied up to 4 mA and up to 60 min per day. Higher intensities and longer stimulation durations may also be safe. Nevertheless, the number of studies using intensities above 4 mA or stimulating longer than 60 min is low. Home-based use of treatments is growing rapidly, leveraging remote supervision to provide patients with greater access and enable repeated, sustained dosing paradigms. We recommend using screening and AE questionnaires in future controlled studies, in particular when planning to extend the stimulation parameters applied. We discuss recent regulatory and ethical issues.
Keywords: tDCS, tACS, tES, Safety, Adverse events, Ethics, Regulation, Training
1. Introduction
1.1. Overview of the update
Low intensity transcranial electrical stimulation (tES) encompasses a group of non-invasive brain stimulation techniques—including transcranial direct current stimulation (tDCS), transcranial alternating current stimulation (tACS), and transcranial random noise stimulation (tRNS)—that apply low intensity electrical currents to the scalp to modulate cortical excitability and neural oscillations. The applications of tES have grown exponentially since its development in the early 2000s. The safety, ethical, legal, regulatory and application guidelines published in 2017 (Antal et al., 2017) provide a comprehensive overview of the state of the art including the available data up to 2016. With the field rapidly progressing, we provide here an update of the safety, regulatory, and application guidelines based on the available evidence published until the middle of 2025. The essentials of the present manuscript were agreed upon at a two-day conference held in Munich, Germany on 12th and 13th June 2025. Participants included research and clinical experts from neurophysiology, neurology, cognitive neuroscience, and psychiatry, as well as ethicists. Industry representatives (i.e., tES equipment manufacturers) also contributed to safety and regulatory issues.
To ensure a comprehensive update, we conducted a systematic search of the literature: the PubMed database using a combination of tES and safety-related keywords (see Appendix for search syntax and output) published between 01/01/2017 and 31/12/2024. The search identified 937 records, which were narrowed down to 340 records after screening titles and abstracts. The records were then categorized based on their type (case reports, qualitative reviews, systematic reviews, meta-analyses, randomized controlled trials, guidelines, and opinion articles) and the topics they addressed (healthy participants, neurology, psychiatry, cognition, pediatrics, elderly populations, pregnancy, ethics, animal studies, modeling and theoretical works, accelerated protocols, and home-based stimulation). Each category was reviewed by experts in the respective research areas, who were encouraged to ensure the inclusion of all relevant works by conducting additional topic-focused searches in different databases (e.g., see the section on aging population), as well as including any gray-literature sources, and then provided a narrative summary of the relevant findings.
We relied on summarizing and interpreting data regarding (1) available animal studies, (2) computational modeling, and (3) human trials, including reports on healthy participants, patients, and vulnerable groups, such as children, elderly, and pregnant women. With regard to animal data, the main effort was devoted to understanding the translation of findings to human applications (e.g., the relationship of dose of stimulation and safety).
Special stimulation conditions that are increasingly used during the last years, e.g., combination of tES with other methods, such as pharmacological and non-pharmacological treatments, stimulating patients with intracranial implants, combination of tES with transcranial magnetic stimulation (TMS) or functional magnetic resonance imaging (fMRI), and EEG were also considered, to assess if they change the riskprofile of the tES application. Furthermore, other stimulation protocols, such as temporal interference stimulation – tTIS and transorbital stimulation) and other settings than ‘transcranial’, in which recent safety data are available, were also integrated (e.g., transcutaneous spinal direct current stimulation (tsDCS).
In this paper, we first provide an overview of the terminology of side effects (SE) and adverse events (AE), technical parameters and basic principles of low intensity tES used alone or combined with other methods, and safety aspects of the stimulation with a summary of the published AEs in healthy participants and different patient populations. The presumed mechanisms and the efficacy of tES in eliciting desired outcomes are not within the scope of this review except for instances, in which they inform about safety. Other stimulation methods such as cranial electrical stimulation (CES) or electroconvulsive therapy (ECT) are also not incorporated here. We also present recent regulatory issues across continents and recommend standards for reporting in research and clinical practice, and finally we summarize existing data and provide recommendations for training and certification and future safety monitoring.
Consensus with regard to the definitions, recommendations, among others were reached after extensive discussions and by voting. The experts first summarized safety data related to their fields and answered questions in several rounds. The key results were presented and discussed at the meeting in Munich. After that the experts were encouraged to support or revise their earlier answers in light of the replies and critique of other members of the panel. The meeting was organized by the Brain Stimulation Special Interest Group of the International Federation of Clinical Neurophysiology (IFCN) and the European Society for Brain Stimulation (ESBS). The IFCN and ESBS have endorsed the guideline.
1.2. Basic aspects: nomenclature and explanations
This document adopts the definitions used in the previous edition of this guideline (Antal et al., 2017) which align with earlier proposals (e. g., (Bikson et al., 2016; Woods et al., 2016), and the nomenclature consensus (Bikson et al., 2019). Additional key terms are clarified below.
Low intensity tES is defined as electric current stimulation with total intensities of ≤4 mA (for AC waveforms ≤4 mA peak-to-peak), a total stimulation duration of up to 60 min per session, and using electrode sizes between 1 cm2 and 100 cm2 with frequencies between 0 and 10,000 Hz. For multi-channel stimulation the total intensity is the sum across all electrodes of the same polarity. For multiple sessions per day the total duration is the sum of all session durations. The current can be of constant intensity and polarity (tDCS) or can be of variable intensities, polarities, or both, alternating between polarities (either at a steady sinusoidal frequency, as in tACS, or a randomly changing frequency, as in tRNS) or within the same polarity (as in oscillatory tDCS (otDCS)) (Guleyupoglu et al., 2013). The intensity of tDCS is always defined as baseline-to-peak, while with tACS and tRNS baseline-to-peak or peak-to-peak intensities can be used and should be explicitly specified. The 4 mA and 60 min parameter limits are based on convention both for ‘low’ and ‘limited’ (Antal et al., 2017; Bikson et al., 2016, 2018, 2019, 2023) and do not represent formal safety limits. Table 1 provides an overview of tES parameters and compatibility features for which the strongest evidence is currently available. Nevertheless, these parameters do not represent absolute safety or application’s limits, and they are not fixed thresholds.
Table 1.
Overview of most commonly used low-intensity transcranial electrical stimulation (tES) parameters and compatibility features. Note: These parameters reflect commonly accepted definitions and applications for low-intensity tES used in experimental and clinical research and are consistent with safety guidelines for human applications. Nevertheless, these parameters do not represent absolute safety or application’s limits, they are not fixed thresholds but indicate the ranges for which the strongest evidence (highest amount of data) supporting safety is currently available.
| Parameter | Specification |
|---|---|
| Current intensity | ≤4 mA total (for alternating currents: ≤4 mA peak-to-peak) |
| Stimulation duration | up to 60 min per session |
| Session frequency | 1–5 sessions per day |
| Inter-session interval | from several days down to 10–20 min |
| Electrode size | 1–100 cm2 |
| Frequency range | 0–10,000 Hz |
| Electrode configuration | 2–9 electrodes |
| Multimodal compatibility | Can be combined with MRI/fMRI (using compatible stimulators), EEG, and other brain stimulation modalities (TMS, rTMS, TBS, ECT) |
| Concurrent interventions | Can be administered concurrently with behavioral, cognitive, motor and pharmacological interventions |
Throughout this review, statements are specific to low intensity tES (as herewith defined) unless otherwise indicated, and studies with evidence for safety which exceed 4 mA or 60 min total per day are also described. The use of intensity ramp-ups and ramps-downs is universal in tES studies and so the claims made in this report (e.g., regarding tolerability) imply the use of ramp up/down. Generally, the ramp up/down durations are 30–60 s and are not calculated in the stimulation duration (Bikson et al., 2019). Sham typically includes brief ramp-up and ramp-down phases, one at the onset and one at the end of the stimulation session, each lasting no more than one minute in total.
Safety is best considered under medical device and drug regulations, which generally define a procedure as safe if the expected benefits outweigh potential risks. An intervention or procedure, such as tES is considered safe if it does not pose an unreasonable risk when implemented as intended, according to defined procedures and administered by trained personnel, and when its benefits justify any known, expected or remaining risks. According to the EU-MDR, safety is determined through a benefit–risk analysis (Article 2), and risks must be mitigated as far as possible, in line with current technology and knowledge.
The term side effect (SE) implies a known causal relationship with the intervention and refers to an effect different from the targeted or intended one. An SE is not always harmful or unpleasant, it may also be neutral or even beneficial. Following the recommendations of the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH; https://www.ich.org) (Baber, 1994; ICH Expert Working Group, 1994) and the Common Terminology Criteria for Adverse Events (CTCAE) v5 (U.S. Department of Health and Human Services, 2017), the term SE will be used in this document when necessary for historical or other legitimate reasons (e.g., communications with participants/patients, or discussions on physiological mechanisms).
Instead, we follow ICH and CTCAE guidelines and use the term adverse event (AE), defined as any unfavorable and unintended sign, symptom, or disease temporally associated with the procedure, regardless of proven causality. AEs are graded according to severity: Grade 1 are mild AEs, defined as involving mild symptoms for which no medical treatment is necessary (e.g., skin redness or tingling during tDCS). Grade 2 are moderate AEs with the need of local or noninvasive treatment (e.g., in the case of tES, the local application of a skin cream for a skin irritation). Serious AEs (SAE) (Grade 3) are severe or medically significant but not immediately life-threatening events, which include the requirement for inpatient hospitalisation or prolongation of hospitalisation. Grade 4 includes life-threatening AEs, where urgent intervention is often required, while Grade 5 means death related to AE. An AE is generally considered unexpected if it is not listed in the information brochure or is not listed at the specificity or severity level that has been observed, or it is not consistent with the risk information described in the investigational plan (FDA regulations, 21CFR312.32, safety reporting).
Following the ISO 14155:2020 classification and CTCAE grading, all unintended effects fall in one of two categories: (1) Non-serious AEs: Grade 1–2: benign, short-lasting effects that may affect comfort or perceived burden, but posing no safety risk. These correspond to commonly reported SEs (for details see Section 6) (2) Serious AEs: Grade 3–5: events of safety concern, such as those requiring hospitalization or causing lasting harm. Therefore, for safety considerations, only serious AEs are directly relevant, while non-serious AEs and known SE of tES are more appropriately evaluated in terms of tolerability and participant burden.
1.3. History of tES and state of the art
The origins of tES can be traced back to electrotherapy practices in the 18th and 19th centuries, when rudimentary electrical devices were used to relieve melancholia (Aldini, 1804), headaches, and induce electronarcosis and electrosleep (dArsonval, 1892; Leduc, 1902; Robinovitch, 1906). However, their effectiveness and mechanisms of action were largely inconclusive, and not extensively explored (Lolas, 1977). Scientific foundations for modern tES applications were laid much later in the 20th century, through early experimental studies in humans and animal models on cerebral excitability, activity, and neuroplasticity (Guleyupoglu et al., 2013). Modern applications of tES began to gain momentum in the early 2000s, when systematic investigations demonstrated that tDCS could induce polarity-dependent changes in cortical excitability in a safe and reproducible manner (Nitsche and Paulus, 2000; Priori et al., 1998), catalyzing a new wave of cognitive and clinical neuroscience research. What followed was a rapid diversification of techniques. The stimulation targets expanded to also include the cerebellum and spinal cord (Priori et al., 2014). Oscillatory stimulation protocols, such as tACS and tRNS, offered a unique approach to interacting with brain networks (Antal et al., 2007, 2017). tACS in particular introduced the ability to interact with endogenous neural oscillations, thus offering a tool for modulation and investigation of specific brain rhythms underpinning cognition and behavior (Wischnewski et al., 2023). These developments prompted a growing interest in network-based, state-dependent, and frequency-specific stimulation paradigms tailored to a specific clinical population, brain mechanism, or personal brain dynamics.
The simplicity, portability, and low cost of tES devices, combined with their favorable safety profile, have contributed to widespread adoption in both basic and clinical neuroscience research. In recent years, the field has expanded significantly, with tES being investigated across a variety of applications including neurorehabilitation, psychiatry, aging, epilepsy, and cognitive enhancement. One direction of progress toward clinical application is the emergence of multi-session and multi-day trials (Couture et al., 2025; Schwippel et al., 2024). This progress has been accompanied by increased focus on improving the spatial, functional, and temporal precision of tES. Specifically, introducing individualized stimulation protocols improves functional precision of the intervention. Such protocols leverage computational modeling for tailored stimulation dose (Puonti et al., 2020a), optimized electrode locations adapted to individual anatomy (Ruffini et al., 2014), physiology-inspired stimulation frequencies, polarities, and waveforms (Zaehle et al., 2011), and connectivity inspired manipulation with phases of periodic currents (Alekseichuk et al., 2019). Multimodal integration with EEG and neuroimaging enables individual targeting based on structural and functional idiosyncrasies and closed-loop designs, which promise more robust effects. Additionally, large-scale computational models have promoted our understanding of current distribution and prompted new approaches to precision and network targeting. Modeling was initially used for post hoc explanations of observed study results and has subsequently been increasingly exploited for ad-hoc planning tools and meta-analytic frameworks. The second notable source of modern understanding of tES mechanisms comes from translational animal research (Johnson et al., 2019; Krause et al., 2019; Vöröslakos et al., 2018), improving the comprehension of dosing and brain state selectivity. However, the rapid adoption of tES has outpaced the development of unified methodological, ethical, legal, and regulatory standards. As the field evolves, it becomes increasingly important to reassess safety guidelines, ethical considerations and regulatory frameworks especially in light of emerging use in vulnerable populations, home-based applications, expanded clinical indications, and multimodal approaches.
2. Principle-based verification of tES
2.1. Mechanism of action – neuronal polarization and other mechanisms
The low intensity tES modulates brain activity primarily through subthreshold neuronal polarization. tES induces electric fields that shift the membrane potentials without directly causing action potentials, making neurons more or less likely to fire in response to ongoing synaptic input. Such modulation can influence synaptic efficacy, alter neuronal excitability, and shape network-level oscillatory dynamics. The mechanisms of low-intensity tES vary, depending on the current properties and electrode placement. For instance, tDCS which always has at least one anode and at least one cathode, is explained by a sustained polarity-specificity change in membrane potential (Buch et al., 2017; Jackson et al., 2017). Direct effects of tES that follow neuronal polarization by tES-generated brain EFs have been extensively studied (Bikson et al., 2010; Froehlich and McCormick, 2010; A. Liu et al., 2018). tACS applies sinusoidal current through two or more electrodes, thus there is no fixed anode or cathode given the oscillatory nature of the current, and the proposed mechanism action is linked to modulation of brain oscillations (Huang et al., 2021; Reato et al., 2013).
Still, there are additional parallel supplementary mechanisms to be considered. As summarized previously (Antal et al., 2017), these include galvanotaxis and electroporation – both requiring higher intensities than those applied in tES. Furthermore, the effects of tES on glial cells have been demonstrated in animal studies, including implications for tDCS (Cancel et al., 2022; Monai et al., 2016; Ruohonen and Karhu, 2012; Tsui et al., 2022). Next, effects of tES, especially tDCS, on brain clearance mechanisms (Khadka et al., 2023; Wang and Monai, 2024; Xia et al., 2020), as well as on vascular and metabolic function (Bahr-Hosseini and Bikson, 2021; Gellner et al., 2023; Khadka et al., 2018; Luo et al., 2022; Sprugnoli et al., 2019) have also been documented, including changes in blood brain barrier/endothelial cell activation (Cancel et al., 2018; Xia et al., 2021). These mechanisms are not independent of each other or of direct neuronal stimulation (e.g. stimulation of the blood brain barrier will impact both neurons and glia, leading to brain transport/clearance changes, and so impacting neuronal function/metabolism). It remains important to understand if/how a given form of tES activates one of these parallel pathways as this may indicate unique therapeutic mechanistic pathways, as well as better explain secondary (following) neuronal changes.
Depending on dose, tES may also induce or attenuate neuro-inflammatory processes in the brain. In rodents, anodal tDCS (~100 kC/m2) decreases the number of activated microglia (Pikhovych et al., 2016), while higher charge densities have pro-inflammatory effects (Rueger et al., 2012). Effects vary by species, disease model, timing, and stimulation parameters. For example, anodal tDCS (4 kC/m2) down-regulates hippocampal inflammatory mediators in a chronic pain model (Spezia Adachi et al., 2012), but upregulates them by both cathodal and anodal tDCS of the same charge density in cortical tissue in a rat epilepsy model (Regner et al., 2020). In post-stroke rodent model, cathodal tDCS (~66 kC/m2) reduces microglial activation and immune cell invasion (Peruzzotti-Jametti et al., 2013) and induces a shift in microglia polarization (Braun et al., 2016), while dual-tDCS down-regulates inflammatory mediators in cortex and hippocampus (Huang et al., 2021). In contrast, anodal tDCS (~57 kC/m2) applied in the acute phase of stroke increases microglia activity throughout the rat brain (Fritsch et al., 2024). In animal models of multiple sclerosis (MS) it was shown that cathodal tDCS attenuates inflammation and axonal loss (Marenna et al., 2022) and that anodal tDCS enhances remyelination and reduces microglial activity (Rossi et al., 2024).
Conclusions: Neuronal function is directly modulated by low-intensity tES through membrane polarization. There remains the possibility of parallel mechanisms of action such as glia, vascular, or brain clearance stimulation. While effects can be demonstrated in isolated tissue models, it is difficult to isolate direct vs indirect effects precisely in vivo because neuronal, glial, vascular, and transport factors are tightly linked. Beyond direct neuronal effects, tES has been shown to modulate neuroinflammatory processes in a context-dependent manner—either attenuating or exacerbating inflammation depending on dose, timing, and disease model. Given the importance of these mechanisms in the etiology and the treatment of various brain disorders, the secondary effects of tES are an important consideration, but they do not raise specific safety concerns in respect to existing practices.
2.2. Assumptions regarding dose–response relationship
tES dose is defined by all of the parameters of the stimulation device that affect the generated electric field (EF) in the body with units of V/m (or, equivalently, mV/mm) (Peterchev et al., 2012). This includes the relevant parameters of the electrode montage (skin contact area), and the intensity, waveform, polarity (if asymmetric), and the duration the current is applied for. For all waveforms except DC this includes stimulation frequency. Electrode size (area, shape) refers to the interface between the electrolyte (e.g. gel, saline) and the skin – that is specifically the surface when current flows in/out the body. Electrode current density is the current amplitude divided by the electrode size. Note this is distinct from current density in the body / brain. Electrode charge density is the total charge applied during a session divided by the electrode size. For tDCS, electrode charge density is the current density times the session duration. As with current density, care should be taken in distinguishing electrode charge density and charge densities in the body / brain. Electrode current / charge density are metrics that can be uniquely calculated given dose parameters, while the converse is not true (e.g. given only a current density one cannot calculate a unique current or electrode size).
These “extrinsic” parameters delivered by the stimulation equipment are well defined and reproducible, while factors influencing current flow patterns and physiological responses are not controllable (e.g., individual tissue properties and anatomy, age, gender, baseline neuro-transmitter concentrations, genetics, dynamic state of the brain before and during stimulation). These “intrinsic” parameters shape the physiological responses to the stimulation and should therefore be considered along with the dose selection (Bikson et al., 2013a; Li et al., 2019; Wei et al., 2024). This presents a challenge to researchers and clinicians when finding the ‘optimal’ dose for a given application. At present, in most studies, the dose (including all parameters: intensity, waveform, electrode size, stimulation duration) is chosen based on previously published data, prior research and clinical experience, individual measures like thresholds, computational models, summary metrics such as the electrode charge density calculated from electrode area, current, and time and related safety considerations based on human and animal experimental data.
tES produces an EF in the brain, which then governs all physiological effects. However, the relation between EF and outcome, including its accumulation over time, is not trivial and may be non-linear (Esmaeilpour et al., 2018; Miranda et al., 2009; Ruffini et al., 2014). On the other hand, even though tDCS induces significant polarization changes, there is a rapid physiological “accommodation” of the axonal membrane potential to a prolonged depolarizing or hyperpolarizing current, mainly involving potassium channels, and probably limiting these changes (Lefaucheur and Wendling, 2019). Thus, given the differences in the biophysical mechanisms of varied tES strategies, dose–response estimates are not directly comparable between methods and the only major factor controlling stimulation efficacy. Actually, optimal dosing depends on the intended neuromodulatory effects.
With regards to informing safety standards based on animal models where brain injury was accessed, it is generally assumed that there is a monotonic dose response (i.e. risk increases with intensity). Factors for interpretation are considered below but we start by noting there is no evidence for brain injury from low intensity tES (tDCS) in humans (Bikson et al., 2016; Nitsche et al., 2004).
Skin effects depend heavily on electrode design (see Electrode section) and best practices (Khadka et al., 2018; Pilloni et al., 2022). Skin is exposed to electrode electrochemical products (Minhas et al., 2010) and higher current density than the brain (Bikson et al., 2018). Standards for skin tolerance should be developed separately and validated in human subjects through empirical testing (Minhas et al., 2010; Woods et al., 2016).
For safety, one approach to dose–response is to 1) adopt the lowest reported brain injury threshold across all animal studies and 2) rely on electrode current density in the animal models as an upper limit for brain current density in human standards. This: a) avoids assumptions about dose–response curves (Bikson et al., 2016); b) neglects variable results (Zhang et al., 2019); including c) methodological issues (e.g. electrode materials that exaggerate sensitivity in some animal studies. This is thus an overall conservative approach. Note limits developed according to this approach for brain current density do not apply for electrode current density limits. Studies using invasive electrodes (Bikson et al., 2017) are excluded due to different injury mechanisms.
The lowest threshold needed to cause a lesion from either anodal or cathodal tDCS in rodent models can be used. Rodent data suggests higher sensitivity under anodal stimulation (Jackson et al., 2017), though studies indicate cathodal stimulation lesion threshold may be higher than previously reported (Zhang et al., 2019). In rodent models, the lowest reported current density threshold for detectable histological tissue damage thresholds for tDCS in animal models were 142.9 A/m2 for cathodal (Liebetanz et al., 2009), and 20 A/m2 for anodal for tDCS (Jackson et al., 2017). Current flow models of tDCS in humans predict the ratio of electrode/skin to brain current density greater than 10:1, and in cases greater than 100:1 (Bikson et al., 2018). In rodent models because of the lower volume (~1000×) and electrode-to-brain distance, as well as use of epicranial electrodes (which prevents skin shunting) the ratio of electrode to brain current density is lower than in humans. To relate safety threshold from rodent studies to human, translation computational models are used (Bikson et al., 2016) and would be montage specific (See modeling section).
The tDCS safety threshold for any montage/subject can be set using two factors. 1) The lowest reported current density causing brain injury in rodents: 20 A/m2 (Jackson et al., 2017). Noting the majority of animal trials indicate even higher electrode current densities are safe (Blaschke et al., 2023; Braun et al., 2016; Callai et al., 2022; Fritsch et al., 2024; Milighetti et al., 2020; Sánchez-León et al., 2021; Zhang et al., 2019). 2) An additional limit based on charge density, which factors in both current density and time (specifically charge density is current density × stimulation duration). The classical “zero lesion size” threshold 5240 mC/cm2 (Liebetanz et al., 2009), which is widely referenced (Antal et al., 2017; Liebetanz et al., 2009) and is lower than observed injury thresholds (Fritsch et al., 2024; Jackson et al., 2017; Zhang et al., 2019). We emphasize both these limits were determined based on electrode/scalp levels (not brain levels), though are conservatively applied here to brain levels in determining thresholds.
Conclusion and recommendations: Defining precise dose–response relationships in tES remains a significant challenge, given the complex interplay between standardized extrinsic stimulation parameters and individual-specific intrinsic factors such as anatomy, physiology, and baseline brain state. Although safety thresholds derived from animal studies serve as valuable benchmarks, their translation to humans must be approached with rigor, due to interspecies variability and methodological heterogeneity. In this context, we propose a precautionary strategy whereby 1) the lowest electrode current density / electrode charge density shown to be injurious in any rodent model is determined; 2) these limits are applied to set the maximum brain current density / brain charge density in human trials; 3) the electrode current density / electrode charge density limits for human trials are then derived, which may be electrode montage specific (e.g. using computational models). Low intensity tES results in brain current/charge-density well below animal safety thresholds. For example, if 1) accepting 20 A/m2 and 5240 mC/cm2 as current density and charger density limits – based on conservative estimates in rodent experiments at electrode level; 2) applying these limits to the human brain; 3) for a conventional bifrontal 5 × 5 cm2 electrode montage the predicted brain current density is 0.22 A/m2 per 1 mA applied (Bikson et al., 2018). Therefore, 2 mA for 30 min yields 0.44 A/m2 and 79.2 mC/cm2; this is 45× below the current density threshold and 68× below the charger density threshold. 6 mA for 120 min yields 1.32 A/m2 and 950.4 mC/cm2; this is 15× below the current density threshold and 5.5× below the charger density threshold. However, our goal here is not to suggest absolute safety limits but to show that existing low-intensity approaches are well below even conservatively estimated safety limits.
2.3. Modeling (heating, induced voltages)
Computational modeling in tES provides a subject-specific estimation of current flow distribution within the head by integrating anatomical imaging (typically magnetic resonance imaging (MRI)) data with finite element models (Fig. 1). These simulations allow investigation of the spatial distribution and intensity of the EF induced by surface-applied currents (i.e., tDCS, tACS), providing information into the physical dose received by neurons in different cortical and subcortical regions. The EF magnitude and direction is the cardinal agent of tES techniques (Galan-Gadea et al., 2023; Radman et al., 2009) and it is molded by the details of individual anatomy (Mosayebi-Samani et al., 2021; Opitz et al., 2015), complex arrangement of cortical folding, and by the electrode montage (electrode positions, currents, and their geometry and composition (Miranda et al., 2013; Saturnino et al., 2015)). While the total applied current at the scalp is typically fixed across study participants (e.g., 1–4 mA), the resulting intracerebral EF can vary substantially between groups and individuals (Laakso et al., 2015), potentially contributing to variability in stimulation outcomes. Computational models are needed precisely because the relation between dose and brain EF is complex (i.e. it cannot be predicted by rules of thumb).
Fig. 1. Electric field modeling pipeline.
Computational modeling in tES studies provides a subject-specific estimation of current flow distribution within the head by integrating anatomical imaging (typically using MR data with finite element models). Using hypothetical electrode positions on the individually reconstructed brain/scalp, the spatial distribution and intensity of the electric field (EF) can be estimated.
Therefore, a primary function of computational models is to guide the development of new electrode montages and dosing strategies for targeting specific brain regions (Dmochowski et al., 2011; Ruffini et al., 2014; Saturnino et al., 2019), as well as to characterize which regions are affected by existing configurations (Splittgerber et al., 2020a). Accordingly, one of the most common uses of computational modeling is the comparison of different montages in terms of their capacity to target desired brain areas.
In this context, computational modeling is a valuable tool for investigating EF distributions across diverse anatomical profiles, including pediatric populations and individuals with neurological conditions (Kasten et al., 2019; Laakso et al., 2019; Van Hoornweder et al., 2022). For example, children, who typically have thinner skulls, may experience significantly stronger cortical EF compared to adults (Ciechanski et al., 2018; Schaper et al., 2023). Individualized EF distributions can also be estimated in the presence of anatomical abnormalities, such as skull defects (Opitz et al., 2018), as well as in stroke patients with lesions filled with cerebrospinal fluid (Minjoli et al., 2017; Yoon et al., 2024) and patients with implants (Mercadal et al., 2022). In such cases, simulations can identify localized field enhancements in the brain or surrounding tissues, including the scalp – an important consideration for safety and risk assessment.
The precision of these predictions, however, depends on the accuracy of the model, not necessarily the complexity (Bikson and Datta, 2012). Since the introduction of MRI-derived models (Datta et al., 2009b), it has been recognized that the CSF compartment should be segmented to avoid direct brain-skull contact and that cortical gyration may generate hotspots (i.e., strong EFs) between, rather than directly beneath, electrodes.
Typical models include six tissue compartments: scalp, skull, CSF, grey matter, white matter, and air (Guidetti et al., 2022). However, these compartments encompass multiple tissue types, and assumptions must be made either through explicit compartmentalization or by assigning aggregate conductivities. For instance, the space between brain and skull is not pure CSF but includes meninges; using a reduced conductivity of 0.8 S/m instead of 1.65 S/m for this compartment has been proposed for greater accuracy (Jiang et al., 2020; Weise et al., 2022). Similarly, the scalp is typically modeled as a single compartment, with an aggregate conductivity of 0.2–0.465 S/m, even though it includes layers such as fat. Such simplifications are often necessary but introduce potential sources of deviation from real-world outcomes.
While invasive measurements in patients have addressed model accuracy (Guidetti et al., 2022; Huang et al., 2017; Opitz et al., 2016; Puonti et al., 2020b), discrepancies remain, especially at the individual level. Key challenges include uncertainties in tissue resistivity and permittivity, particularly at low frequencies (Saturnino et al., 2019), as well as poorly understood conductivity profiles of pathological and perilesional tissues (Datta et al., 2010; Sprugnoli et al., 2019). Precise segmentation of lesions may require multimodal imaging beyond standard T1-weighted MRI. Similarly, a consensus has yet to be reached on the best way to model implants, such as titanium plates and SEEG leads (Alonso et al., 2023; Karimi et al., 2025; Mercadal et al., 2022). Some evidence suggests metals may behave as non-conductors at low frequencies due to insulating tissue-metal interfaces (Mercadal et al., 2022).
Underlying these modeling efforts is a key conceptual simplification: the assumption that the EF in a region of interest predicts the degree of modulation. This is known as the quasi-uniform assumption (Bikson et al., 2013b), which posits that at the scale of individual neurons, the EF can be treated as spatially uniform. The effects of (uniform) EFs on neuronal polarization depend strongly on both the magnitude and direction of the EF relative to a neuron’s morphology and orientation (Bikson et al., 2004). Therefore, modeling the EF vector, not just its magnitude, is critical (Galan-Gadea et al., 2023). This assumption remains implicit in most large-scale tES simulations, including when modeling effects on neurons (Aberra et al., 2023; Rathour and Kaphzan, 2024) or other cell types (Khadka and Bikson, 2022), where each cell is treated as being exposed to a locally uniform field.
The EF strength and its spatial distribution in tACS are expected to be similar to that observed with tDCS (Huang et al., 2017; Opitz et al., 2016). It remains unclear whether the high electric permittivity of brain tissues can significantly affect the strength of the EF in the brain and shift the phase of the sinusoidal waves, in particular with higher frequencies (Gaugain et al., 2023).
Models focused on skin current flow have informed electrode design, as tolerability of tES is often related to skin effects (Khadka et al., 2018; Kronberg and Bikson, 2012; Saturnino et al., 2015). The skin is, in fact, a complex multilayer/structures barrier, and efforts have been made to model current flow through these structures explicitly (Gomez-Tames et al., 2016; Khadka and Bikson, 2020). Practical factors affecting tolerability (such as electrodes drying or leaking saline/gel) may not be addressed by idealized computational models.
Notwithstanding the noted limitations, EF simulations can support risk analyses. For example, it is reasonable to assume that AE scales with the local electric field strength. When simulations predict EF levels within previously validated safety margins, this supports confidence in a given approach.
Extending simulations beyond the EF to include tissue heating can help to improve their utility for risk assessment. Heating of the brain during tES, scales with the strength and duration of the EF, and can be predicted by bioheat models. For conventional tES protocols predicted heating is insignificant (not hazardous) (Datta et al., 2009a) including in the presence of implants, (Karimi et al., 2025), but might be re-evaluated for higher doses, atypical anatomy, or specific implants. Heating simulations at the MRI radiofrequency, complemented by measurements of skin temperature increases, have been important for the safety evaluation of tES electrodes and cables used during magnetic resonance scanning (Gregersen et al., 2021).
Models of other mechanisms of AEs, such as electrochemical reactions at the skin-electrode interface, are not well established, limiting the possibilities offered by simulations for risk assessments of those aspects. Computational models are also a key tool in scaling rodent to human dosimetry and developing safety standards (Bikson et al., 2016).
Previous findings on anatomical differences in head and cortical bone structure between males and females (Russell et al., 2014), suggest that females tend to have thicker cranial bone structures, which may reduce the intensity of electrical currents reaching brain tissue. However, such differences may be less impactful than general anatomical (size) differences across individuals. These findings along with the reciprocal dynamics between fluctuating sex and reproductive hormones and cortical excitability (Finocchi and Ferrari, 2011; Aftanas et al., 2022), remain underexplored.
Conclusions and recommendations: Models are an important part of risk analyses, especially in relating dose to EF and for identifying possible worst-case scenarios. No significant / hazardous brain heating is predicted during conventional protocols. However, given the model uncertainties, the need for further validation, and the incomplete understanding of the link between EF and AEs, it is recommended that EF modeling not be used on its own to assess safety. Instead, it should be complemented by empirical safety data and established risk assessment procedures.
3. Electrode design for tES
Electrode design and placement govern brain current flow and thus, the spatial effects of stimulation (for a detailed review, see (Solomons and Shanmugasundaram, 2020)). While the shape and materials of the electrode have moderate influence on brain current flow, the primary importance of electrode design is tolerable effects on the skin. We summarize electrode design impact on brain current flow and on skin effects.
Conventional tDCS uses two “pad” electrodes (one anode, one cathode) of comparable size (25–35 cm2). The term “return” electrodes does not mean an electrode is inert but rather reflects the hypothesis that it plays a less critical role in the targeted outcome. Notably, the return electrode can be either the anode or cathode, depending on the montage and the intended direction of current flow. Similarly, designations like “anodal tDCS” or “cathodal tDCS” highlight the presumed functional importance of the anode or cathode, respectively, within a given protocol (Bikson et al., 2019). Increasing the size of one electrode relative to the other has been proposed as a method to reduce the impact of the larger electrodes (Nitsche et al., 2007), though computational models suggest this has moderate effect (Bikson et al., 2010). Positioning the “return” electrode in an extracephalic position does not make it inert but produces current through the deep brain (Noetscher et al., 2014; Sadeghihassanabadi et al., 2022).
Scalp electrodes used for HD-tDCS® are smaller (~2.5 cm2) (Minhas et al., 2010). The use of smaller electrodes in itself provides only a moderate benefit in focality (Miranda et al., 2009). However, multiple HD electrodes can be arranged that can alter current flow. In the 4×1 Laplacian montage (Datta et al., 2009b), the center electrode determines the active polarity and the radius of the four electrode ring controls focality (Alam et al., 2016). Overall, the 4×1 montage enables a more focal cortical current flow (Edwards et al., 2013). It has been applied to multiple cortical targets including the motor cortex and the DLPFC (Cacciamani et al., 2024; Caparelli-Daquer et al., 2012; Hemmerich et al., 2024; Johari et al., 2022; Kuo et al., 2013; Müller et al., 2022). The 4 × 1 montage was also used to deliver tACS (Gebodh et al., 2024; Grover et al., 2022; Zhang et al., 2022a) or otDCS (Manojlović et al., 2025).
Alternatively, multiple electrodes distributed across several sites can be used to optimize the EF strength in the target brain region or to deliver network-targeted stimulation (Fisher et al., 2017; Gregoret et al., 2023). Another approach using four electrodes, tTIS, was recently shown to be well tolerated (Cassarà et al., 2025; Vassiliadis et al., 2024). There is a tradeoff between maximum EF intensity in the brain and focality (Alam et al., 2016). With multi electrode montages, it is also possible to provide higher total current by splitting current to individual electrodes (Harrie et al., 2023).
Electrode design plays a fundamental role in tolerability. While for any given current, decreasing electrode size will proportionally increase current density; electrode materials have a more important role in tolerability. For example, well-designed HD electrodes produce higher current density than conventional tDCS pad electrodes without necessarily enhancing skin sensation (Reckow et al., 2018; Turski et al., 2017). Electrode shape appears to have minimal effect on the resulting sensation (Ambrus et al., 2011; Minhas et al., 2011). The resistivity of the electrolyte (sponge or hydrogel) may impact tolerability (Dundas et al., 2007; Khadka et al., 2018). Skin lesions are not an expected AE of tES when the best practices are followed and appropriate equipment used; essentially every case-report of a skin lesion during tDCS was correlated to poor technique or technology (Pilloni et al., 2021). Impedance measurements are a valued quality-control step to identify potentially non-ideal electrode set-up (e.g. electrolyte dries up). However, low impedance is neither a necessary or sufficient condition to ensure tolerated and reliable stimulation. It is possible for an approach to show low impedance but not be tolerated (Woods et al., 2016). For home-based techniques (see Section 10.4), and use in the MRI (see Section 9.1.), special considerations apply.
Finally, given the central role of electrode design in sensation, it is also a relevant factor in the blinding reliability (e.g. electrode designs that produce less sensation in the active arm will produce more effective blinding). This is an important point when considering the reliability of sham across studies that may not use the same electrode designs.
Conclusions and recommendations: A multitude of possible electrode shapes, materials, and placements are possible. Electrode montage primarily shapes current flow through the brain while electrode design primarily impacts tolerability. Skin preparation (cleaning and drying) is recommended, while abrasion is not valued or recommended. Skin lesions are not an expected AE of tES using verified devices/electrodes and standard protocols (e.g. over intact skin). Electrodes verified for specific tES dose should be used; electrodes not designed for stimulation (recording electrodes) or for distinct stimulation applications (eg. TENS electrodes) are not a priori verified for tES. For sponge electrodes, using desalinated (tap) water is not recommended. Appropriate protocols for electrode preparation and placement are required. Protocols for electrode re-use or single-use should be guided by the manufacturer’s recommendations. Tolerability is always a function of both electrode design and the waveform applied to the electrode (intensity, frequency). Therefore, tolerability should be evaluated for each dose.
4. Safety considerations of different tES techniques
4.1. Conventional (bipolar) and 4 × 1 montages
The modern development of tDCS (Nitsche and Paulus, 2000, 2001; Priori, 2003), and then related low-intensity tES techniques such as tACS (Antal et al., 2008), used an electrolyte (saline) saturated sponge “pocket” around a conductive rubber electrode, which was in turn connected by a lead to the stimulator (DaSilva et al., 2011). Conventionally, two electrodes were used, one cathode and one anode. Variations in this approach include changing sponge electrode size and shape (Nitsche et al., 2007). Approaches using more than two electrodes were considered but the size of electrodes limited their number and proximity on the scalp (Angius et al., 2018).
For a given current applied to an electrode, the current density increases proportionally with decreased electrode size, and for this reason, it was assumed that stimulation with small electrodes would not be tolerated. However, in 2008, this was overturned by delivering 2 mA tDCS with ~4 cm2 electrodes, with the key design factor being the material and shape rather than the surface area of the electrode (Minhas et al., 2010). For example, the HD-tDCS® system uses specialized hydrogel. With smaller electrodes, it is possible to create montages with electrode arrays, including one electrode surrounded by a ring of electrodes of the opposite polarity (Laplacian montage as used in HD-tDCS® (Datta et al., 2009b)) as all bipolar montages (et al., 2019; Rawji et al., 2018). HD-tDCS® is not less tolerated than conventional bipolar tDCS and high currents have also been tested for HD-tDCS® (Reckow et al., 2018), as well as various tES waveforms, such as tACS (Grover et al., 2022; Helfrich et al., 2014) or otDCS (Manojlović et al., 2025). Therapeutic neuromodulation trials with HD-tDCS® also followed (Caparelli-Daquer et al., 2012; Kuo et al., 2013).
Modeling software can be used to optimize the number of electrodes (multi-channel application), their size and placement, and the current per electrode (Dmochowski et al., 2011; Nishimoto et al., 2024; Zhang et al., 2022a) for both superficial and deep brain targets (Antonakakis et al., 2024; Huang and Parra, 2019). Some approaches like tTIS depend on the use of precisely placed smaller electrodes. Smaller electrodes also support the use of head-gear combining tES and EEG.
Conclusions and recommendations: The 4 × 1 montage (e.g., used for HD-tDCS®) has the same safety profile as conventional bipolar tES. The selection of sponge pad or gel-based electrodes depend on the trial hypothesis, brain targeting constraints, and practical factors.
4.2. tACS
tACS continues to demonstrate a favorable safety profile when used within typical experimental and clinical parameters. As previously reported, sensations under the electrodes are generally less intense than during tDCS (Fertonani et al., 2015). This may be due to reduced electrochemical effects and the frequency-filtering properties of neuronal membranes (Deans et al., 2007). Phosphenes and cutaneous sensations are most pronounced at frequencies between 10 and 30 Hz, with a peak around 20 Hz, and tend to diminish at both lower and higher frequencies (Turi et al., 2014). Phosphenes are more prominent with frontal montages, while skin sensations increase with central montages. Both effects scale with intensity (Raco et al., 2014).
Since the last review, multiple studies have confirmed the absence of SAEs in both healthy and clinical populations across a wide range of standard tACS protocols (Evans et al., 2019; Indahlastari et al., 2018; Matsumoto and Ugawa, 2017). Structured assessments consistently report SEs such as transient sensations such as tingling, itching, phosphenes and only mild AEs (Grade 1), including mild headaches (Daughters et al., 2020; Kvašňák, 2019; Zeng et al., 2019). These effects are generally short-lived and often reported at comparable rates in sham conditions (Bland and Sale, 2019). Studies using intensities of up to 2 mA baseline to peak and durations of 20 to 40 min per session, administered across multiple days, have demonstrated consistent tolerability in both experimental and clinical settings, including in patients with depression (Alexander et al., 2019) and substance use disorders (Daughters et al., 2020).
Conclusion and recommendation: Current evidence shows that tACS is safe and well tolerated when applied within standard parameters. Across healthy and clinical populations, only mild AEs (Grade 1) and transient SE typically dependent on stimulation frequency and montage, have been reported, with incidence rates often comparable to sham, and did not require medical care nor necessitate early termination.
4.3. Oscillatory tDCS
Oscillatory transcranial direct current stimulation (otDCS) is a form of tES that combines features of both tDCS and tACS. It involves the application of low-intensity sinusoidally modulated currents with a direct current offset. This means that the current oscillates at a specific frequency while for each electrode remaining entirely within either the positive or negative polarity range (i.e., either anodal or cathodal) and intensity within the standard range (i.e., between 1 and 2 mA). Given that otDCS incorporates elements of both tDCS and tACS, its safety profile may be expected to be similar to those of these established techniques. A recent study compared the subjective experience (tolerability) and SEs of tDCS (1.5 mA), tACS (0 ± 1 mA, i.e., 2 mA peak-to-peak), and otDCS (1.5 ± 0.5 mA, i.e., between 1 and 2 mA) in healthy young adults (Bjekić et al., 2024). The findings indicated that otDCS induced discomfort levels comparable to tDCS while producing similarly mild but somewhat less frequent AEs, including scalp itching (31.0 % post-tDCS, 26.2 % post-otDCS), tingling sensations (42.9 % post-tDCS, 23.8 % post-otDCS), and mild scalp irritation (28.6 % post-tDCS, 16.7 % post-otDCS). SEs typically associated with alternating currents, such as phosphenes, were reported by three participants during tACS and one during otDCS, while shaking of the visual field was reported during tACS only. Specific to otDCS, a negligible-to-mild burning sensation was reported by 14.3 % of participants, while 19.1 % experienced increased tiredness following stimulation. This sham-controlled study delivered otDCS as well as tACS and tDCS via electrodes placed on the left parietal cortex (P3) and contralateral cheek while the frequency was in theta range (4–8 Hz) for otDCS and tACS. The same profile of AEs was reported in other studies that applied theta-band otDCS (Vulić et al., 2021; Živanović et al., 2022), while there are no available reports on protocols in other frequency bands.
Conclusions and recommendations: Current evidence suggests that otDCS, like tDCS and tACS, is safe and associated with minimal AE (Grade 1) when applied within standard parameters. However, research on otDCS remains limited, and several factors—including electrode positioning, current intensity, and stimulation frequency—may influence its tolerability. Additionally, the long-term effects of repeated otDCS sessions, particularly in individuals with neurological conditions, remain unexplored.
4.4. tRNS
Transcranial random noise stimulation (Bikson et al., 2019) is mostly used with frequencies between 100–640 Hz (Terney et al, 2007), has been recognized for its good safety profile, with AEs comparable to sham. tRNS includes symmetric biphasic waveforms (with zero DC) and asymmetric waveforms including monophasic or DC offset (Mondino et al., 2022; Murphy et al., 2020; Tokikuni et al., 2024).
Sheffield et al. (2022) analyzed AE data from 1019 healthy adults who underwent 9–11 daily sessions of tRNS, tDCS, or tACS using small circular electrodes (3.14 cm2) centred on four different cortical target sites. Using Bayesian statistics, they found support for the null hypothesis, indicating that AEs were similar between active and sham tRNS. Additionally, blinding data revealed excellent blinding efficacy for tRNS. These findings align with previous studies using fewer sessions and larger sponge electrodes, which also reported no significant differences in AEs between tRNS and sham, as well as advantages such as higher cutaneous perception thresholds and lower AE reporting rates compared to tDCS (Ambrus et al., 2010). A notable finding is that AEs (Grade 1) varied by stimulation and collection site, suggesting the potential influence of training differences or population characteristics on reported AEs (Sheffield et al., 2022). However, no longer-term data are provided.
Large-scale studies on clinical populations remain limited, although the reported AEs (Grade 1) are similar to those ones reported in healthy adults. tRNS has been used in multiple studies on adults with tinnitus, with a recent review reporting only mild AEs (Alashram, 2024). However, AE variability across studies mirrors the SE observed by Sheffield et al. (2022), which may be attributed to different experimenters’ expertise or types of population.
For individuals with brain damage, preliminary evidence suggests no reported AEs with tRNS. In a study investigating vision recovery in patients with cortical blindness due to V1 damage, the patients did not report any AEs (Herpich et al., 2019).
Most paediatric studies have focused on atypically developing children. A recent study examined 1032 tES sessions in children and adolescents with neuropsychiatric and neurodevelopmental disorders (Battisti et al., 2025), confirming effective blinding, consistent with earlier paediatric research (Dakwar-Kawar et al., 2022, 2023; Looi et al., 2017; Splittgerber et al., 2020b). Although AEs were similar between active and sham tRNS, an overall higher AE (Grade 1) reporting rate for tRNS compared to tDCS (including both sham and active conditions) was noted. This discrepancy is unlikely to stem from tRNS, and is likely to be due to differences in participant populations or experimenter effects. In animal models, juvenile mice underwent the same protocol as children with atypical development (nine twice-weekly sessions of 0.75 mA (peak-to-peak) over one month) (Looi et al., 2017), with no detectable macroscopic tissue damage (Sánchez-León et al., 2021).
Research on home-use tRNS remains scarce. The only known study compared 20 min AI-driven personalised current intensity tRNS, 1.5 mA (peak-to-peak) tRNS, and sham tRNS in healthy adults at home. No AEs (Grade 1) differences between active tRNS and sham were found, and the participants tolerated the intervention well (Cohen Kadosh et al., In press).
Conclusion and recommendation: In summary, the existing evidence supports the safety and tolerability of tRNS in the lab and at home. However, most studies have used relatively lower current intensities (0.03 mA/cm2 in children, and mostly <0.318 mA/cm2 in adults). It is recommended to carefully test higher intensities, because it remains an open question whether higher intensities would maintain similar AE rates and blinding efficacy.
4.5. Network-targeted tES
Interest has grown in tES to modulate large-scale brain networks typically using multi-channel2 tES– i.e., simultaneous stimulation of functionally or structurally connected regions that exhibit correlated activity across time (Ruffini et al., 2018). These networks are typically identified through fMRI (task-based or resting-state BOLD) (Fox and Greicius, 2010), (s)EEG (Bartolomei et al., 2018; Centeno and Carmichael, 2014), magnetoencephalography (MEG) (Cho et al., 2024, p. 202), and/or diffusion-weighted MRI (Babaeeghazvini et al., 2021). Examples include canonical networks such as the default mode network, salience network, and dorsal attention network (Zhou et al., 2018). Alternatively, networks may be defined as sets of brain parcels, based on standardized parcellations, supporting cognitive or sensorimotor functions (Lawrence et al., 2021; Schaefer et al., 2018; Wang et al., 2021).
Many studies end up stimulating networks by default due to the spatial scale of the electric field generated by typical electrode montages. However, this section specifically addresses network stimulation studies that have generated empirical data by explicitly targeting networks defined through functional or structural imaging modalities (Table S1). It excludes stimulation protocols characterized by broad, non-specific cortical coverage (Splittgerber et al., 2020a; Sprugnoli et al., 2019), as well as approaches primarily aimed at suppressing activity within single epileptogenic zones (Daoud et al., 2022; Kaye et al., 2021; Simula et al., 2024). Epileptogenic zones can be conceptualized as networks due to the pathological synchrony observed across different regions and characteristic seizure propagation patterns (Daoud et al., 2025), and current efforts are investigating the potential of stimulating epileptogenic networks using multichannel tDCS (Ruffini et al., 2018). However, clinical data regarding safety and efficacy are not yet available.
The studies with network-based targets (Abellaneda-Pérez et al., 2021; Adhia et al., 2023; Dagan et al., 2018; Ester-Nacke et al., 2024, 2025; Fischer et al., 2017; Goede et al., 2024; Mencarelli et al., 2020; Salehinejad et al., 2025; Smeele et al., 2023; Wei et al., 2024, 2024; Zhou et al., 2022) share protocol characteristics with potential safety implications: they often use smaller electrodes (contact areas < 8 cm2); they employ intensities commonly exceeding 2 mA and going as high as 4 mA (defined as the sum of the currents delivered through the anodes); and they generate diffuse EF distributions often optimized using algorithmic approaches to achieve network level targeting (Ruffini et al., 2014, 2018). Still, no SAEs have been reported, and only common AEs (Grade 1) are reported (e.g., tingling, itching, transient redness under the electrodes, and mild headaches). Studies comparing network-targeted protocols with conventional montages, including tDCS and tRNS, report no significant differences in AEs and SE (Fischer et al., 2017; Mencarelli et al., 2020). One under-examined aspect is state dependence: the idea that stimulation effects may vary depending on ongoing brain activity (Luo et al., 2025). While not anticipated to raise safety concerns, state dependence may affect response consistency, including of pathological networks (Daoud et al., 2025; Schaper et al., 2023).
Conclusions and recommendations: Network-targeted tES demonstrates minimal AEs (Grade 1) within the conventional parameters including intensities under 2.0 mA per-electrode and electrode contact areas higher than 3.14 cm2. Montage optimization algorithms are recommended to optimize the EF distribution, especially in complex networks. There is no evidence for new potential safety issues with network stimulation that necessitate special standards.
4.6. Transcranial temporal interference stimulation (tTIS)
Transcranial temporal interference stimulation (tTIS) is designed to target deeper brain structures in humans with good depth-focality trade-off (e.g.,(Beanato et al., 2024; Hummel and Wessel, 2024; Kurtin et al., 2025; Vassiliadis et al., 2024; Violante et al., 2023; Wessel et al., 2023)), demonstrating a favorable safety profile when used within conventional parameters. Studies have consistently reported minimal AEs, while no SAEs have been documented.
Typically, stimulation parameters for tTIS include carrier frequencies ranging between 1000–20,000 Hz, beating envelope frequencies between 5–130 Hz and currents usually up to approximately 4 mA peak to peak with session durations commonly around seconds to minutes. Most frequent AEs, expected from tES, were perceived sensations such as tingling, burning, itching ((Vassiliadis et al., 2024) for review (Demchenko et al., 2025)). The perception of stimulation-induced sensations has been shown to depend directly on stimulation intensity, becoming more noticeable as the current is approaching higher thresholds, e.g., around 2 mA (Vassiliadis et al. 2024). It has been demonstrated that the perceived sensations are comparable between the temporal interference condition and control conditions, including a high-frequency control (with no difference between carrier frequencies) and conventional sham (involving fade-in and fade-out stimulation) (Vassiliadis et al., 2024). This renders tTIS as a method with excellent blinding opportunities well suited for double-blinded and placebo-controlled trials. A recent meta-analysis (Demchenko et al., 2025) suggests that perceived mild sensations like tingling, itching, though rather rare, might be more frequently reported in verum tTIS (11 %) compared to control (5 %), an important aspect that must be further explored. Rarely, subjects describe headache, sleepiness/drowsiness, nausea or pain. However, these AEs are comparably frequent during tTIS and control stimulation (Demchenko et al., 2025).
Studies that directly assessed the safety profile of tTIS, such as the detailed examination by Piao et al. (2022) and Wang et al. (2024), have reinforced its general tolerability and absence of significant AEs. Piao et al. (2022) and Wang et al. (2024) investigated 38 and 88 healthy subjects, respectively, monitoring neuropsychological, electrophysiological, and behavioral parameters, including serum NSE levels and EEG, and lent additional support for tTIS’s favorable safety profile.
First applications in patient cohorts confirmed a favorable safety and AEs profile of tTIS (e.g., (Lamoš et al., 2025; Missey et al., 2025; Ploumitsakou et al., 2025; Vassiliadis et al., 2024; Yang et al., 2024)).
However, the long-term safety profile and the effects of repeated or prolonged applications remain incompletely evaluated and have to be addressed in upcoming studies.
Conclusion and recommendation: tTIS using carrier frequencies of 1000–20,000 Hz and currents ≤4 mA peak-to-peak (single studies up to 30 mA) is currently regarded as safe for use in both basic neuroscience and early clinical-translational research, with only mild AEs reported (Grade 1). Initial clinical applications support this favorable safety profile. Notably, tTIS provides strong potential for effective participant blinding, a key advantage in experimental design. Continued, rigorous monitoring of AEs remains essential—particularly as stimulation protocols are refined and new ones developed, patient populations are included, and long-term safety is assessed and needs to be demonstrated. Overall, the safety profile of tTIS is comparable to that of other established NIBS techniques, such as tACS.
4.7. Transcutaneous spinal direct current stimulation (tsDCS)
During transcutaneous spinal DCS (tsDCS) (Cogiamanian et al., 2008), electrical current is delivered through an electrode positioned on the skin over the spinal cord with the return electrode placed over various regions according to different protocols (mainly the shoulder, the anterior aspect of the trunk, or along the spine). tsDCS appears to influence ascending and descending spinal pathways and to modify the excitability of various spinal reflexes in humans and animals (Priori et al., 2014). In general, anodal tsDCS tends to suppress conduction along spinal pathways and to facilitate reflexes, while cathodal tsDCS tends to enhance responses mediated by spinal ascending pathways and inhibit reflexes.
In several clinical trials including patients with neuropathic, musculoskeletal, or visceral pain the analgesic effect of tsDCS was proven (thoracic anode placement; 2.5 mA, 20 min, at least 5 days stimulation) (Guidetti et al., 2021; Hodaj et al., 2023). The mechanisms underlying the analgesic effects of tsDCS remain largely unknown. However, in healthy subjects, the technique likely modulates synaptic efficacy at the segmental level, influencing the local processing and/or transmission of nociceptive inputs in the dorsal horn (Lenoir et al., 2018). Interestingly, anodal tsDCS reduced reactive oxygen species (ROS) levels and increased antioxidant capacity markers such as glutathione (GSH) and total antioxidant capacity (TAC), suggesting improved mitochondrial efficiency in multiple sclerosis (MS) (Mrakic-Sposta et al., 2025).
Clinical experience and the available reports suggest that tsDCS does not induce SAEs. Computational simulations support this insight (Guidetti et al., 2023), also in presence of metallic spinal implants (Kuck et al., 2019), as induced electric field values are reported to be over a thousand times lower than the safety thresholds for tissue injury (Bikson et al., 2009). While one study reported no blood biomarkers indicative of neuronal damage following stimulation (Cogiamanian et al., 2008), a preliminary account (H. Zhao et al., 2023) indicates that the most frequently AEs are mild sensations of burning, tingling, and itching below the stimulating electrodes. Skin redness may occur, however, this tends to be transient and self-resolving. Thus, tsDCS does not induce changes in bodily state, at least as measured by autonomic nervous system metrics.
Conclusions and recommendations: Carefully inspect the spine for vertebral malformations (e.g., spina bifida, myelomeningocele) as these can alter the electrical environment of the spinal cord, affecting current flow during tsDCS. Syringomyelia can also significantly alter the pattern and effectiveness of current flow during tsDCS due to the presence of a fluid-filled cavity in the spinal cord, which changes local conductivity and may shunt current away. Careful consideration of individual anatomy and possible adjustment of stimulation protocols are necessary in patients with any kind of spine abnormalities. No safety data exist for pregnant women and for patients with implanted active devices in the torso (injection pumps, stimulators, cardiac pacemakers, etc). For these cases, no specific safety recommendation is made but study-specific risk analysis is warranted.
4.8. Optic nerve/retina stimulation
During repetitive transorbital alternating current stimulation (rtACS) a weak electrical current is delivered to the eye through a pair or multiple electrodes attached to the skin around the eye. Animal studies using crush and transection models of the optic nerve indicate that the electrical stimulation induces structural and functional neurorestoration (axonal regeneration, normalised visual evoked potentials), and has a neuroprotective effect (better survival of ganglion cells) (Miyake et al., 2007; Morimoto et al., 2005; Tagami et al., 2009; Yin et al., 2016). These processes are assumed to be mediated by the release of neurotrophic factors and increased chorioretinal blood flow (Fu et al., 2015).
In the previous version of these guidelines (Antal et al., 2017), we reported on the safety and tolerability outcomes from studies involving 760 patients with optic neuropathies, such as those following stroke or with post-chiasmatic lesions, who were treated in various clinical trials using this technology (Fedorov et al., 2011; Gall et al., 2013, 2010, 2016, 2011, 2015; Sabel et al., 2011; Schmidt et al., 2013). The most common AEs were mild skin sensations and irritation, headache, drowsiness, and sleep disturbances. No stimulation-related SAEs were reported.
Since 2017, the number of clinical trials using rtACS has been limited. In one study (Räty et al., 2021), 56 patients with occipital stroke were stimulated using 5–15 Hz rtACS, or with the combination of occipital tDCS and rtACS. No serious AEs occurred. The most common AEs were mild skin irritation under stimulation electrodes during the treatment, tiredness, ‘other’ skin sensations (about 30–100 % of the cases, depending on the experimental setup), mild headache, and phospheneslike visual phenomena (about 30 %). A low number of patients (maximum 2) reported metallic taste, myokymia, dizziness, sleeping difficulties (only in the sham group), and lack of concentration (only in the sham group). No difference in the frequency of any of the reported AEs, nor in their composite number per subject was detected between the different modalities.
A recent review of 10 RCTs and 15 observational studies, including transcorneal electrical stimulation, transpalpebral electrical stimulation, transdermal electrical stimulation, and rtACS, reported no SAEs (Liu et al., 2021). Minor complications included foreign body sensation, dry eye, transient superficial keratitis, cutaneous sensation at electrode sites, and mild headache in sporadic cases.
An interventional, randomized study on 51 patients using trans-corneal electrical stimulation reported mild AEs, including predominantly dry eye symptoms, but also a few cases of ocular discomfort, ocular pain, transient visual disturbance, increased tear production, and itching (Stett et al., 2023).
Two studies using rTACS to target phosphene induction in 15 (Sabel et al., 2021) and 61 (Hunold et al., 2025) healthy participants, respectively, reported that the stimulation was well tolerated and no AEs occurred.
An alternative terminology for transorbital stimulation is retinofugal alternating current stimulation (rACS) targeting α-oscillations in the visual cortex via periorbital electrodes (Haberbosch et al., 2019). In this study, multichannel stimulation was applied at 10 Hz, with an amplitude of 120 % phosphene threshold (resulting in 351.69 μA peak-to-peak). Topographical modeling revealed the highest current densities in the anterior visual pathway, the retina, and the optic nerve. No serious AEs occurred in 20 healthy subjects, and the stimulation was well tolerated.
An ongoing multicenter clinical study in Germany is investigating the efficacy of the rtACS in open-angle glaucoma (Schittkowski et al., 2025). As of the publication date of this paper, no safety-related concerns have been reported.
Another stimulation form, using a rectangular waveform, is also applied to treat glaucoma. This type of stimulation is described as optic nerve stimulation (ONS) (Erb et al., 2022). In a recent study, seventy glaucoma patients with progressive vision loss underwent electrical ONS (5–35 Hz), with stimulus intensities up to 1.2 mA, for 10 days, about 80 min of stimulation/day. In this paper, no AEs or SAEs were reported.
Conclusions and recommendations: Based on the published data, the likelihood of detrimental effects using rtACS is extremely low and not different from other protocols using tES. It is recommended that personalised protocols (e.g. slightly below phosphene threshold intensities) are used at least at the beginning of the treatments. Furthermore, RCTs are needed to establish the optimal protocols for the application to different kinds of visual disorders.
4.9. Cerebellar stimulation
Cerebellar tES, including tDCS, tACS and tRNS, has emerged as a non-invasive technique targeting cerebellar networks implicated in motor control, cognition, and affective regulation. While its therapeutic and experimental applications are expanding, the safety profile of cerebellar tES remains an important topic for clinical and research deployment.
Systematic reviews and individual randomized controlled trials indicate that cerebellar tES administered within established current intensity ranges from 1 to 2 mA and ≤30 min per session is safe and well tolerated in healthy adults and patient populations (Manto et al., 2022; Warthen et al., 2024).
In a recent meta-analysis of 14 studies involving cerebellar tDCS in patients with degenerative ataxia (N = 406), no serious AEs were reported; only mild, transient SEs occurred in two studies (Gong et al., 2023).
Commonly observed AEs, as in cortical tDCS, include scalp tingling, itching, mild headache, skin redness, fatigue, and dizziness, these symptoms typically resolve shortly after stimulation and rarely necessitate treatment discontinuation (Antal et al., 2017; Herzog et al., 2022).
Several trials combining cerebellar tES with motor rehabilitation in stroke survivors or individuals with balance disorders have similarly reported no serious complications. For example, in a controlled study applying cerebellar tDCS during virtual balance training in stroke patients, mild headache was the most frequently noted symptom (Qurat-Ul-Ain et al., 2023). Studies employing cerebellar tACS or tRNS also demonstrate good tolerability, with minor cutaneous or vestibular sensations but no severe neurological or systemic effects (Tavakoli and Yun, 2017).
Although, data remain limited; preliminary trials indicate that cerebellar tES is generally well tolerated in adolescents with neuro-developmental disorders (D’Urso et al., 2021): the anatomical proximity of the cerebellum to the brainstem warrants conservative dosing and individualized current modeling to avoid off-target stimulation. Ethical aspects and regulatory oversight are essential when applying tES in pediatric populations (Salehinejad et al., 2022).
There is no single standard for cerebellar stimulation electrode placement, which is usually guided by previous research (Ferrucci et al., 2015). Stimulation sites vary by function, for example in stroke rehabilitation: superior-lateral cerebellum for motor tasks, pharyngeal area for dysphagia, and lobule VII for aphasia, contralesional stimulation is often more effective due to crossed cerebro-cerebellar pathways, and bilateral stimulation has shown greater enhancement of swallowing function compared to unilateral, likely due to increased input to the cerebral swallowing cortex (Liu et al., 2024).
Conclusion and recommendation: cerebellar tES—when delivered within conventional parameters—is a safe and well-tolerated neuromodulatory intervention in adult populations. No serious AEs have been documented to date. Mild, transient SEs are common and manageable. However, further studies are required to assess the long-term safety of repeated or chronic cerebellar stimulation, particularly in vulnerable populations such as children or individuals with neurological comorbidities. From a procedural perspective, it is recommended to use montages that place the active electrode over the cerebellar hemisphere (whole or lateral) — typically located 1 cm below and 2 cm lateral to the inion—and the reference electrode at an extracephalic site, such as the ipsilateral shoulder. This configuration is proposed to reduce current spread to brainstem structures (Parazzini et al., 2013).
5. Safety considerations of repeated tES application
5.1. Repeated/multiple sessions across days
The rationale for delivering tDCS across multiple days is to enhance its efficacy and promote longer lasting neuroplastic after-effects. In line with this objective, numerous studies over the past two decades have investigated the efficacy and safety of multi-session tDCS protocols.
A key meta-analysis conducted by Bikson et al. (2016) assessed the safety of repeated tDCS interventions, defined as more than four sessions per week, drawing on data from over 1000 participants. The analysis confirmed the safety and tolerability of repeated tDCS, reporting no SAEs. Subsequent studies applying up to 10 tDCS sessions per subject in controlled clinical or research settings, and administered by trained personnel, have consistently replicated this safety profile (Antonenko et al., 2022, 2024; Choy et al., 2023; Juras et al., 2025; Ke et al., 2023; Talsma et al., 2017; Teixeira-Santos et al., 2022). A comprehensive meta-analysis of 158 studies (n = 4130) further confirmed that AE rates did not increase with cumulative tDCS exposure—defined as two or more sessions spaced no more than a day apart—nor did they differ across diagnostic groups, reinforcing the safety of repeated tDCS protocols (Nikolin et al., 2018).
The use of home-based tDCS has expanded the feasibility of long-term interventions, enabling daily stimulation protocols extending over several weeks or months (Grønli et al., 2022; Im et al., 2019; Madhavan et al., 2025a; Martens et al., 2018). For instance, Martens et al. (2018) administered a 4-week daily tDCS intervention (2 mA, 20 min/session, 5 sessions/week; 20 sessions total; ~400 min per subject) targeting the left DLPFC in 27 patients with chronic minimally conscious state. No SAEs were observed in the active tDCS group. One SAE (epileptic seizure) occurred in the sham group on day 4, though this event was deemed unrelated to stimulation, as the sham protocol involved only 5 s of active stimulation (5 s ramp-up, 5 s stimulation, 5 s ramp-down).
In another study, Im et al. (2019) evaluated the efficacy of a 6-month home-based tDCS protocol in 20 patients with AD, with 12 receiving active and 8 receiving sham stimulation (2 mA, 30 min/session; ~182 sessions; ~5460 min total per subject; bilateral DLPFC montage: anode over F3, cathode over F4). No information on AEs or SEs was reported. Nonetheless, 18 of the 20 participants completed the study; the two drop-outs were attributed to refusal or caregiver-related issues rather than stimulation-related AEs.
Further home-based studies have supported the safety and tolerability of long-term tDCS. In amyotrophic lateral sclerosis (ALS) patients, Madhavan et al. (2025) administered 72 sessions over 24 weeks (2 mA, 20 min/session, 3 sessions/week; 1440 min total per subject) without SAEs. Similarly, Grønli et al. (2022) applied daily tDCS for 18 weeks in AD patients (2 mA, 30 min/session, 7 sessions/week; ~122 sessions; ~3660 min per subject), with no SAEs reported. Ghazi-Noori et al. (2024) conducted a 6-week study in patients with bipolar disorder (2 mA, 30 min/session; 21 sessions; 630 min total per subject), reporting that 90.6 % of AEs were rated as mild, 9 % as moderate, and 0.4 % as severe. The most common side effects included tingling (83.5 %), skin redness (40.6 %), itching (29.3 %), and burning sensations (26.5 %). The most severe AEs comprised one case each of tingling and burning sensations, and two cases each of itching and skin redness.
Conclusions and recommendations: Multi-session tDCS delivered at conventional intensities (≤4 mA) by trained professionals in clinical or research settings remains well-tolerated, with no SAEs related to tDCS reported. Emerging data from home-based protocols, extending up to 5000 min of total stimulation time, suggest initial safety under real-world conditions. However, to substantiate these findings, rigorously controlled trials with larger sample sizes and systematic AE monitoring are essential.
5.2. Safety considerations of repeated/multiple-sessions per day
Applying multiple tES sessions per day (also often referred to as an “accelerated protocol”, “intensive application” or “spaced tES” in the literature) builds on conventional protocols, which typically deliver stimulation once per day. Inspired by findings from accelerated repetitive transcranial magnetic stimulation (rTMS) studies, where multiple daily sessions have been shown to augment treatment effects, these multi-session-per-day tES protocols aim to enhance efficacy by increasing the frequency of sessions within a single day.
A potential concern with this approach is the increased risk of AEs due to the cumulative effects of stimulation. However, there is little evidence to suggest that intensifying the application schedule (to 2–5 sessions per day) and reducing the inter-session intervals (for several hours to 10–20 min) elevates the risk of serious AE. While participants receiving active tES appear to have a higher frequency of mild AEs associated with tES (such as burning sensations or itching) compared to sham, similar effects have been observed in conventional once daily protocols (Nikolin et al., 2018).
The literature search identified 70 studies ranging from case reports to randomized controlled trials, covering both healthy and pathological conditions. In total, these studies involved over 1900 individuals and more than 14,000 active sessions.
Although few studies have directly compared the safety of once-versus multiple-daily tES protocols, available evidence from neurophysiological studies suggests that multi-session protocols do not substantially elevate the risk of serious AEs. These include consecutive tDCS sessions over the motor cortex spaced by short (3–20 mins) or long intervals (3–24 h) compared to a single session (Monte-Silva et al., 2013), thrice-daily 20 min tRNS sessions spaced 30 mins apart (Brevet-Aeby et al., 2019), and twice-daily vs once-daily tDCS (20 min, 5 h apart) in patients with vascular depression (Zanardi et al., 2020).
The majority of studies retrieved in the literature search involve patients with neuropsychiatric conditions and compare multiple-daily tES protocols to sham. One of the earliest applications of a multi-sessions tES protocol was in patients with schizophrenia (Brunelin et al., 2012). These received twice-daily tDCS sessions, spaced 2 to 3 h apart, with the anode placed over the left prefrontal cortex and the cathode over the left temporoparietal junction (ten 20-min sessions, 2 mA). While this study primarily investigated therapeutic outcomes, it also reported no significant AEs and preserved participant blinding. Subsequently, a meta-analysis of eight studies (329 participants) with schizophrenia showed no difference in all-cause discontinuation between twice-daily tDCS and sham groups (Jiang et al., 2022). Safe use of twice-daily protocols has also been observed in numerous randomized sham-controlled tDCS trials across several other clinical populations using twice-daily protocols, including subacute stroke (Hsu et al., 2023), stroke (Pinto et al., 2021), Parkinson disease (Aksu et al., 2022; Dashtelei et al., 2024), tobacco use disorder (Mondino et al., 2022), obsessive compulsive disorder (Bation et al., 2019), major depression (Moirand et al., 2022), bipolar depression with suicidal ideation (Wang et al., 2024), social anxiety disorder (Jafari et al., 2021), epilepsy (Kaufmann et al., 2021), and acute respiratory distress syndrome (Andrade et al., 2022). These twice-daily session studies employed varying electrode montages, session durations, and stimulation intensities (ranging from 9 to 30 min at 1–3 mA), as well as differing inter-session intervals (from 10 min to 6 h).
Some recent studies have also explored delivering more than two tES sessions per day in clinical populations, with some protocols including up to five daily sessions. This intensified approach has been applied in patients with major depression (Couture et al., 2025), schizophrenia (Mondino et al., 2021), and catatonia (Bouaziz et al., 2023). Importantly, while most studies report only mild, transient effects (e.g., tingling, itching, redness), comparable to those observed with oncedaily protocols, a trial of 29 patients with depression receiving 50 tDCS sessions (five per day, 20 min duration at 2 mA, 20 min inter session interval with left anodal/right cathodal DLPFC stimulation) found that 18 (62 %) developed mild, persistent erythematous-squamous plaques under the anode, consistent with irritative contact dermatitis (Miron et al., 2023). Additionally, they reported high rates of redness at electrodes sites (100 %), headaches (67.9 %) and fatigue (57.1 %). No participant required specific treatments (e.g., topical corticosteroids). Notably, the authors reported that tDCS parameters, including current intensity and density, were within established safety guidelines, used saline pre-saturated electrode sponges (‘SNAPpads’), which are high-quality, individually packaged electrodes and maintained impedance levels below 30 kΩ. It is also unlikely that the symptoms were caused by the accumulation of toxic by-products. These details suggest that poor technique or non-compliance with safety recommendations cannot explain the observed AEs. In addition to recommending the application of a moisturizing cream to the exposed scalp area after each treatment day in affected participants, all subjects were advised to gently cleanse the exposed scalp area with a mild soap before each session. Conclusions and recommendations: Direct comparisons between multiple and once daily tES protocols regarding safety remain limited. Delivering up to five 20-mins sessions per day (at 2 mA, separated by 20 min) is tolerable, with a safety profile comparable to oncedaily tES protocols. Particular attention should be paid to potential skin reactions, and special care, including the use of optimized equipment, should be taken to minimize these effects. Given their safety profile, multi-sessions-per-day protocols may be particularly suitable for users who are unable or unwilling to attend multiple visits, as well as for those with severe cognitive or clinical symptoms who may benefit from more intensive treatment.
6. Summary of side effects and adverse events associated with low intensity tES in human studies
6.1. Perceptual side effects
At the perceptual level, commonly reported SEs remain scalp discomfort due to cutaneous sensations (i.e., somatosensory perceptions), such as tingling, itching, and burning. Recent evidence from a large body of tES literature, including studies using tDCS, tACS, tRNS, and HD-tDCS, supports that perceptual SEs are generally mild, transient, and well tolerated in both healthy individuals and clinical populations for intensities up to 2 mA. Some HD-tDCS studies have explored intensities up to 4 mA, where effects remain tolerable but perceptual sensations tend to be more pronounced. However, it should be noted that a considerable proportion of people receiving sham stimulation report similar perceptions too, warranting caution in attributing every occurrence to a tES effect.
Tingling is one of the most commonly reported sensations during tES and the sensation is typically localized at the site of stimulation (Bjekić et al., 2024; Chang et al., 2021; Fertonani et al., 2015). This sensation is generally mild and short-lived (Gairola et al., 2024; Hawas et al., 2025; Latrèche et al., 2024; Ruffini et al., 2024; Upadhyay et al., 2025; Wang et al., 2024). Numerous studies have reported tingling sensations in both active and sham conditions, with some studies noting a higher incidence in sham groups. However, the underlying reasons for this observation remain unclear (Ul-Ain et al., 2024). Delicado-Miralles et al. (2024) also showed that tingling sensations were most prominent during the initial sessions and declined in frequency and intensity over time, highlighting habituation across repeated exposures. A large-scale analysis of 6779 sessions of remotely supervised at-home tDCS (RS-tDCS) in clinical trial contexts reported no serious SEs and a very low session abort rate (0.04 %), with the most common SE being mild tingling, itching, and warmth, equally distributed across active and sham conditions. Notably, no participants discontinued due to tolerability concerns (Pilloni et al., 2022). The prevalence of tingling in pediatric populations was also shown by Buchanan et al. (2021), noting it was typically mild and not associated with dropouts.
Thus, mild local pain or discomfort at the stimulation site is also among the reported SE associated with tES, particularly during tDCS sessions (Bjekić et al., 2024; Pérez-Borrego et al., 2024). This discomfort is typically characterized by sensations such as burning, itching, or pressure beneath the electrodes and is usually transient, subsiding shortly after the end of the stimulation. Several studies report that these sensations are more prevalent at the beginning of stimulation sessions and tend to decrease with repeated exposure (Bjekić et al., 2024; Delicado-Miralles et al., 2024). The intensity and frequency of local pain and discomfort are influenced by stimulation parameters, electrode type, and individual sensitivity (Bjekić et al., 2024). All these aspects can be addressed through strategies such as the use of ramp-up and -down protocols avoiding abrupt current onset, adequate electrode preparation, consistent saline/gel application (Fertonani et al., 2015).
Phosphene and “vibration” perceptions are mainly induced during tACS, particularly at frequencies in the alpha and beta range and current intensities ≥1 mA, remains an expected SE as consequence of visual (Schwiedrzik, 2009) or tactile systems stimulation (Zeng et al., 2019), and is not considered harmful, despite its impact on perception. Several studies have confirmed that phosphenes, while occasionally interfering with tasks, are non-distressing and mitigated through protocol adjustments (e.g., montage design).
These results are consistent with earlier findings (Antal et al., 2017) and may vary with electrode positioning and whether current ramping is applied for tDCS. While in general, effects like phosphenes or tingling may impact task performance or engagement temporarily, they are not sufficient grounds to halt or contraindicate stimulation protocols, unless requested by the participant.
6.2. Adverse events
Redness of the skin at the electrode site is an anticipated, though relatively infrequent non-serious event. This erythema is typically mild and often resolves within hours. Woodham et al. (2024) reported that 63.5 % of participants in the active tDCS group experienced skin redness, compared to 18.5 % in the sham group. Zhou et al. (2024) and Park et al. (2024) also found skin redness among the most frequent mild AEs, with no long-term consequences.
In home-based multichannel tDCS, similar effects such as scalp burning and skin irritation have been noted (e.g., Park et al., 2024), although they are infrequent and typically resolve spontaneously (Ruffini et al., 2024). Importantly, these events rarely result in dropout, and adherence remains high across clinical and home-based trials when appropriate safety measures are implemented (Pérez-Borrego et al., 2024). In pediatric and adolescent populations, Ciechanski et al. (2017) and Buchanan et al. (2021) emphasized that such events were generally well tolerated and did not lead to increased dropout, supporting the safety of tES across age groups. These AEs have been associated with non-conventional protocols/equipment, reinforcing the importance of adequate device design and user training (for details see Section 10). Provided appropriate protocols, equipment, electrodes and skin preparation are used, skin injury is not an expected AE of low-intensity tES (Pilloni et al., 2021; Woods et al., 2016). Monitoring for skin integrity can help minimize local irritation and enhance tolerability (Ruffini et al., 2024; Woodham et al., 2024).
Fatigue is a less consistently reported AE of tES. It has been described in some studies, though often without a clear association with active stimulation. Sleepiness and general tiredness have been reported in both active and sham conditions, suggesting non-specific origins potentially related to session timing, individual baseline state, or expectations (Latréche et al., 2024; Woodham et al., 2024). In fibromyalgia and schizophrenia studies, fatigue was sometimes reported but did not significantly differ between active and sham conditions (Lyu et al., 2024; Yang et al., 2024).
In one experiment, intended alterations, including dizziness, motion perception shifts, and spatial disorientation, have been reported with tACS at low frequencies (1–2 Hz, 2.5 mA) in healthy individuals. These effects appear to be frequency-dependent and due to the modulation of the vestibular system function (Rossi et al., 2023).
Headaches are another reported AE across both tDCS and tACS protocols. These are generally mild, often occur during or shortly after stimulation, and tend to be self-limiting (Chen et al., 2024). Studies such as those of Woodham et al. (2024), Wang et al. (2024), and Zhou et al. (2024) observed similar rates of headaches in both active and sham groups, suggesting that while common, headaches may not be directly induced by stimulation. Similarly, in a double-blind pilot trial of athome remotely supervised tDCS (RS-tDCS) for veterans with persistent post-traumatic headache, no AEs were reported, and the intervention resulted in significant reductions in headache days and severity, further supporting the safety and potential of RS-tDCS in populations with high symptom burden (Charvet et al., 2023b). Similarly, Park et al. (2024) found that headaches were reported in the active group and in the sham group, with no statistical significance, reinforcing the notion that these effects are often nonspecific and mild.
Regarding cognitive AEs, there is no evidence that tES causes long-term or significant cognitive impairments. However, a few studies have reported short-lived mild AEs, affecting non-targeted functions, such as reduced attention, impaired concentration, and increased mental fatigue (e.g., (Yuan et al., 2022)). These effects were primarily self-reported and listed in specific populations, such as older adults and individuals with medical conditions undergoing prolonged or repeated stimulation protocols.
On the other hand, when tES is applied to modulate cognition by down- or up-regulating cortical states, the intended effects arise from the modulation of cortical excitability, which can interfere with non-targeted functions, such as practice effects or learning. This suggests that tES may, in some cases, influence cognitive engagement indirectly rather than directly impairing it (e.g., (Leaver, 2025)). Reports of cognitive AEs remain rare, typically self-reported and not performance-limiting. However, while most cognitive outcomes in studies are still tied to the primary stimulation target and interpreted as functional neuromodulation rather than AEs, the challenge of capturing off-target or secondary cognitive effects remains.
Conclusion and recommendations: tES is generally safe and well tolerated, with no evidence of serious or persistent SEs or AEs in healthy or sensitive populations when standard safety protocols are followed. The most common events are typically mild, short-lived, and manageable (AEs grade 1). Their occurrence can be reduced through proper electrode application, technical adjustments based on individual modelling of the current flow and user education. Routine monitoring and reporting of these effects are recommended to refine individual tolerability profiles and assess any potential trade-offs in cognitive function. These effects do not pose a barrier to experimental, clinical or home-based use of tES. The summary of SE/AE is presented in Table 2.
Table 2.
Summary of SE/AE across different tES techniques, with severity, frequency of occurrence and respective SE/AE classification i.e. grading in line with definitions outlined in Section 2.1.
| Effect | Affected Area | Severity | Frequency | Stimulation Types | SE/AE grade | Notes |
|---|---|---|---|---|---|---|
| Tingling/Itching | Scalp | Mild | Very Common | tDCS, tACS, otDCS, tsDCS | SE/AE Grade 1 | Transient, higher with tDCS and with higher intensities; reduced with longer ramping; present in sham. |
| Burning sensation | Scalp | Mild to Moderate sensation | Common | tDCS, otDCS, tsDCS | SE/AE Grade 1 | Linked to electrode contact; more common with poor electrode preparation or repeated sessions. |
| Skin redness | Scalp | Mild | Common | tDCS, HD-tDCS, tsDCS, tACS | SE/AE Grade 1 | Temporary; persistent with accelerated protocols; irritation or dermatitis. |
| Phosphenes | Perceptual Visual Field | Mild | Common with tACS | tACS, otDCS | SE/AE Grade 1 | Frequency dependent; stronger when electrodes are closer to the eyes; usually not distressing. |
| Visual flickering/Distortion | Perceptual Visual Field | Mild | Occasional | tACS | SE/AE Grade 1 | Frequency dependent effect, during stimulation, not harmful. |
| Headache | General | Mild | Common | tDCS, tACS | AE Grade 1 | Self-resolving, timely, linked to stimulation sessions. |
| Fatigue/Sleepiness | Systemic | Mild | Common | tDCS, otDCS, tACS | AE Grade 1 | Reported more in older adults and in postpartum women. |
| Dizziness/Vertigo | Vestibular | Mild | Uncommon | tACS | AE Grade 1 | Low frequency vestibular effects; short duration, reversible. |
| Local pain or discomfort | Scalp | Mild | Common | All | AE Grade 1 | Usually due to electrode placement, wrong electrode preparation or stimulation start. |
| Difficulty concentrating | Cognitive | Mild | Rare | tDCS | AE Grade 1 | Mostly self-reported, short-lived. |
| Contact dermatitis | Skin | Moderate | Occasional | Accelerated tDCS | AE Grade 2 | Observed in prolonged/accelerated protocols. |
| Mood changes (e.g. Hypomania) | Mood | Moderate | Rare | tDCS in MDD | AE Grade 2 | In patients with depression; causal link unclear. |
6.3. The association between ANS and tES
The autonomic nervous system (ANS) regulates involuntary functions such as heart rate, blood pressure, and respiration. It is closely connected to the brain via structures like the nucleus of the solitary tract, which integrates afferent signals — many of which travel through the vagus nerve—and coordinates responses through cortical and subcortical circuits. Given this brain–ANS–vagus integration, it is essential to consider how tES may modulate autonomic function (Makovac et al., 2017) and whether this may result in potential AEs. Based on the reports, tES (tACS, tDCS, tRNS) is safe for ANS function, with no reported AEs or impairments in cardiovascular or other ANS parameters. For example, tACS did not modulate pain-related autonomic responses, which was confirmed by Bayesian statistics (May et al., 2021; Prim et al., 2019). Also an extracephalic (arm/shoulder) electrode positioning in tDCS, one targeting M1 (Santarnecchi et al., 2014) and the other one targeting the frontal midline (Vandermeeren et al., 2010) suggest null or mild effects on autonomic functions. Li et al. reported a transient increase in automatic arousal at the end of the tDCS session, an effect that disappeared afterwards (Li et al., 2024). These results highlight that stimulating the brain through tES has no AEs at the level of ANS functioning.
In fact, only after applying tDCS, several studies suggest potential benefits on ANS functioning, such as transient improvements in heart rate variability (HRV), which reflects better autonomic regulation (Ko et al., 2024). Razza et al. (2024) recently showed a dose-dependent effect of tDCS for HRV, with stronger effects on HRV increases after 3 mA as compared to 1.5 mA. This ANS functioning has increasingly been employed as a supportive physiological marker in tES research, complementing behavioral and cognitive outcomes. For example, the stimulation of the DLPFC with tDCS has been shown to modulate ANS processing without reliably altering self-reported mood (Mondino et al., 2015), suggesting subtle, yet meaningful physiological shifts. Similarly, Schroeder et al. (2015) reported reduced autonomic arousal during emotional tasks following anodal tDCS over the left prefrontal cortex, reductions linked to distraction and mild behavioral improvements. Recently, Allaert et al. (2022) also demonstrated that active (vs. sham) tDCS over the left DLPFC was associated with decreased skin conductance responses, indicative of lower emotional arousal during emotion regulation. These findings suggest that HRV and other ANS indices may not only reflect the physiological effects of tES but also serve as potential biomarkers of treatment response or even therapeutic targets, particularly in conditions marked by dysregulated autonomic function.
Conclusion: The tES induced effects on ANS appear to be reversible, within normal physiological bounds, and in line with expected neuropsycho-physiological patterns—raising no safety concerns. Long term SEs have not been reported. However, research on the autonomic effects of other NIBS modalities, such as tACS and tRNS, remains limited and warrants further investigation.
7. Safety considerations in neurotypical adults and special populations
7.1. Absolute and relative exclusion criteria for tES in non-clinical settings
In research involving healthy participants, particularly in basic and translational neuroscience research, it is essential to adhere to clearly defined inclusion and exclusion criteria in order to on one hand minimize potential risks and ensure participant safety and, on the other, ensure methodological rigor and exclude potential confounds.
By definition, the inclusion criteria for studies in healthy adults are the absence of neurological, psychiatric, or neurodevelopmental conditions and legal adulthood status (age of majority is ≥18 years in most countries, for detailed consideration of tES application in children and older adults see Section 7.3. and 7.4.).
Exclusion criteria are typically categorized as absolute or relative, depending on the nature and severity of the associated risks. Absolute exclusion criteria refer to conditions that categorically disqualify a participant from taking part in a study due to the known potential for harm or significantly increased risk. The absolute exclusion criteria (i.e. contraindication) for tES application are open wounds, preexisting skin abnormalities (e.g. nevi, angiomas or previous burns/scars, neurodermitis, etc.), or active infection at the stimulation site, which could be conducive to or lead to increased risk of skin damage, or secondary infection associated with tES and/or the use of conductive gel/paste/solution.
Other exclusion criteria may apply depending on the study aims, available resources and overall cost-benefit analysis. These are relative exclusion criteria that do not have documented safety risks but are applied based on the principal investigator’s decision either for precautionary or methodological reasons. For a conservative approach the following relative exclusion criteria can be applied: self-reported history of epilepsy or uncontrolled seizures and/or seizure activity in the past year, given the theoretical risk that modulation of cortical excitability may lower the seizure threshold – nevertheless, based on the available evidence, seizures have not been reported as an AE associated with tES; self-reported history of head or spine injury and/or surgery, which may alter brain structure or excitability, thereby increasing unpredictability of response to tES; as well as implanted electronic or metallic devices (e. g. cochlear implants or metal plates in the head) that can alter current flow or induce heating (see Section 9.4 for detailed consideration); current use of medications or drugs such as antidepressants, antipsychotics, or stimulants; self-reported history of persistent SE/AE that could be associated with tES, e.g. frequent headaches and migraines; tattoo at the stimulation site, as it would require consideration of the ink used (Maas et al., 2021); excessive scalp sensitivity and discomfort during electrode placement; possible pregnancy (for detailed consideration see Section 7.2.); low tolerability of previous NIBS applications, etc. Further common relative exclusion criteria in studies that aim to explore basic neuromodulatory mechanisms and effects of tES in neurotypical adults include factors that may act as confounds and add to the variability of outcomes, such as: excessive caffeine intake; smoking; alcohol, or recreational drug use; sleep deprivation before or severe fatigue on the day of the session; high scores on screening questionnaires for anxiety, depression, or distress, which may influence placebo/nocebo responses or AE reporting. Other relative exclusion criteria may be introduced depending on the specific study methods, aims and objectives.
7.2. tES in pregnant and breastfeeding or not-breastfeeding postpartum women
So far, nine studies have been published about tES during pregnancy, including tDCS (Kurzeck et al., 2021; Laurin et al., 2022; Shenoy et al., 2015; Sreeraj et al., 2016; Strube et al., 2016; Vigod et al., 2019; Q. Zhao et al., 2022), TNS (Trevizol et al., 2015), and tACS (Wilkening et al., 2019). Among these, only two RCTs were published: Vigod et al. (2019) observed the effect of 10 tDCS sessions delivered between 14–32 weeks of gestation and Zhao et al. (2022) observed the effect of one tDCS session pre-c-section. The remaining evidence comes from case studies encompassing 85 women receiving active stimulation, 83 of whom underwent tDCS.
Based on the current evidence, tES – particularly tDCS – appears to be well-tolerated during pregnancy, with minimal AEs reported. tDCS was used as an add-on treatment to Iloperidone (12-mg/day; Shenoy et al., 2015), to lamotrigine (100-mg/day; case report 1 from Laurin et al. 2022) and to Venlafaxine (75-mg/day, case report 2 from Laurin et al. 2022). Zhao et al. (2022) also reported tDCS concomitant with lumbar anesthesia (Bupivacaine only [n = 52] or Bupivacaine with Lidocaine [n = 10]) with no AEs due to the combination between tES and medication being reported.
As in the general population, studies present only transient mild AEs (tingling, itching and mild burning sensations at the application site, transitory experience of phosphenes, mild headache during and right after stimulation, dizziness and insomnia, difficulty in concentrating,) describing symptoms that are common in pregnancy therefore they cannot be directly attributed to tES, (Kurzeck et al., 2021; Laurin et al., 2022; Shenoy et al., 2015; Sreeraj et al., 2016; Wilkening et al., 2019; Zhao et al., 2022). One preterm birth was reported (Vigod et al., 2019), but no direct association with tDCS was established. No other neonatal complications were noted.
These findings suggest that tES has a benign safety profile during pregnancy and that the risk of fetal exposure to electric current is negligible. In a comparative perspective, modeling studies in ECT, which produces electric field densities approximately 100 times higher than tES, suggest that fetal exposure remains well within the safety limits set by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) (Bikson et al., 2023; Kibret et al., 2018). Consequently, it is reasonable to infer that tES, with its significantly lower electric field intensity, poses no additional fetal safety concerns. Furthermore, no major safety concerns for maternal health, fetal exposure, or newborn development have been reported at follow-up (between 24 h after the intervention and up to 3 years after intervention), even when tES was used as an add-on treatment to medication (see Table S2).
In the postpartum period, evidence is limited. There is one case report of a depressed breastfeeding woman (Laurin et al., 2022) and an open-arm study (Griffiths et al., 2025) with 20 postpartum depressed women. In both studies tDCS was used as an adjunct treatment to antidepressants and the F3-anode, F4-cathode montage at 2 mA intensity was used. In both studies the acute treatment included 15 daily sessions across 4 weeks, followed by a maintenance treatment of three to four sessions a week for three to four weeks. Side and AEs reported were minor and transient, like those expected in the general population (mild fatigue, paresthesia of the scalp, mild headache). Hence, as tES does not enter systemic circulation, no risk of transmission through breast milk could be thought of. However, fatigue after tDCS sessions (an uncommon AEs) should be monitored in the postpartum, given that it could in extremis indirectly affect newborn care routines.
Overall, critical knowledge gaps persist on the reciprocal interactions between tES and reproductive/polypeptide hormones. Hormones such as progesterone, estrogen, luteinizing hormone, and follicle-stimulating hormone undergo significant fluctuations during pregnancy and the postpartum period, influencing cortical excitability and potentially modulating the physiological effects of tES. These hormonal shifts may hypothetically increase the vulnerability of women to seizures (Finocchi and Ferrari, 2011). Conversely, like rTMS, tES may also influence hormonal regulation, impacting neuroendocrine functions (e.g., (Aftanas et al., 2022)). Although fluctuating sex hormones during pregnancy or across the menstrual cycle may affect cortical excitability warrants further investigation, so far there has been no evidence that women are at increased risk when using tES in comparison to men.
Recommendations: tES, particularly tDCS may be considered a non-pharmacological treatment during pregnancy and postpartum, particularly for women who prefer to avoid, cannot use, or do not respond to medication, with informed consent addressing limited evidence. In both clinical and research contexts, pregnant and postpartum women, including those breastfeeding, should not be excluded by default if the benefit–risk balance is favorable. Protocols should monitor maternal health, fetal signals, and infant outcomes, if possible. Future studies should collect reproductive, hormonal, and developmental data to better understand sex-specific effects of tES.
Precautions that should be considered with pregnant women: Avoid stimulation with an extracephalic electrode as there is no evidence in the literature about its effects. Do not stimulate directly over the abdominal/lumbar area (not part of standard tES protocols). After the 24th week of gestation, avoid stimulating in the supine position for theoretical concerns about dizziness and blood pressure drops due to compressing major blood vessels. For pregnant women either in clinical or research contexts, implement monitoring protocols that include frequent supervision by a specialized maternal health care team (especially when using home-based devices). Avoid stimulation when holding a baby/infant.
7.3. tES in pediatric populations
Since 2017, the number of tES studies in pediatric populations has significantly increased. In 2020, Zewdie et al. reported on a large-scale non-invasive brain stimulation study involving 384 participants who underwent at least one stimulation procedure. In total, 612 tDCS sessions were administered, including sessions with 92 children (12.9± 3.4 yrs). The intervention was well tolerated, with mild itching or tingling reported in 37 % of participants. Other observed AEs included headache, neck pain, presyncope, tingling, combined itching and burning sensations, and nausea. Major reported AEs and the corresponding stimulation protocols in pediatric populations since Antal et al. (2017) are summarized in Table S3.
Recent reviews have demonstrated that tES is generally safe, with no AEs reported when established protocols are properly followed (Buchanan et al., 2021; Salehinejad et al., 2020, 2022). A more recent review by Salehinejad and Siniatchkin (2024) provides a comprehensive overview of the latest evidence on NIBS in pediatric populations. Consistent with findings from the past decade, data from the last two years (2022–2023) further confirm the safety of NIBS in children and adolescents when established protocols are strictly followed. Both tES and TMS are generally well tolerated, provided that safety guidelines, including exclusion criteria, are followed.
Computational models have indicated that the smaller cranial anatomy in pediatric populations will result in a higher brain electric field per applied current intensity (mA) (Kessler et al., 2013; Minhas et al., 2012). Predicted current densities were not higher at the skin, consistent with comparable tolerability. Differences in brain electric fields will depend on the montage and specific subject anatomy (e.g. a minor with a relatively large head may be less sensitive than an adult with a relatively small head). Differences in electric field sensitivity are consistent with differences in dose response measured by changes in the amplitude of TMS evoked potentials, with some suggesting the use of reduced currents in pediatrics (Moliadze et al., 2015, 2018). This is not evidently based on safety. 2 mA currents have been used in pediatric populations, and the increased sensitivity remains far from levels considered injurious.
In 2018, Meiron et al. (2018), successfully applied 10 sessions of 4×1 HD-tDCS (intensities up to 1 mA) in a case report of a 30-month-old child suffering from early onset epileptic encephalopathy. There were no SAEs or SEs related to the HD-tDCS intervention. Battist et al (2025) extensively evaluated the tolerability, safety, and blinding of tDCS and tRNS in paediatric clinical populations, composed of 92 children and adolescents (54 females, age range: 8–17 years), involving 1032 sessions across neuropsychiatric (i.e., anorexia nervosa) and neuro-developmental (i.e., attention deficit and hyperactivity disorder, developmental dyscalculia) conditions. It compared AEs occurrence between active and sham tES conditions (i.e., 528 active vs. 504 sham sessions) as well as tDCS and tRNS (i.e., 772 tDCS sessions vs. 260 tRNS sessions), while considering demographic and emotional-behavioural factors. Results demonstrate the safety of tES, with no moderate or severe AEs reported. Approximately 77 % of sessions were free of any AEs, supporting the use of tES in these populations. Although active sessions showed a higher likelihood of AEs compared to sham, these events were predominantly mild, mostly limited to itching, indicating a generally benign safety profile. However, the study did not assess the effect of AEs occurrence separately for tDCS and tRNS.
In 2020, Splittberger et al. (2020) reported the case of a healthy 13-year-old girl who experienced her first generalized tonic-clonic seizure within a week after undergoing a single tDCS session. In the present observation, no direct causal relationship between the tDCS and the seizure can be established, especially since the seizure occurred five days after stimulation. Rather, this case highlights that previously undiagnosed conditions can emerge by chance and underscores the need to take appropriate measures to minimize such risks as much as possible.
Conclusions and recommendations: Studies conducted in pediatric populations have not reported a significantly higher incidence of AEs compared to those observed in adults. In terms of tolerability, there are no major concerns regarding the use of tES in pediatric populations. There is no evidence for long-term aftereffects, including in children; ongoing studies on long-term effects are useful. In research involving children and adolescents, particular attention should be paid to the language used during informed consent and when assessing eligibility. To ensure participant safety, tES studies in this population should implement age-appropriate screening procedures and systematically monitor AEs after each stimulation session. Furthermore, a rigorous ethical framework is needed, including greater transparency and consistent reporting of unusual or unexpected cases (Sierawska et al., 2019, 2020; Splittgerber et al., 2020a; Splittgerber et al., 2020c).
7.4. tES in aging populations
Building upon Antal et al. (2017), we focused on studies published since 2017 that explicitly report SEs or AEs in their title or abstract, selecting those involving healthy (Menassa et al., 2023) older adults (≥60 years) (World Health Organization, 2021). While not exhaustive of all NIBS research in this population, our analysis includes both experimental and review studies on tDCS, and tACS. A summary of the selected articles is presented in Table S4.
Consistent with previous findings, most studies continue to use tDCS with a current intensity of 1.5–2 mA and 35 cm2 electrodes, targeting the frontal or parietal cortex for 15–30 min. Several studies investigated safety outcomes in DLPFC stimulation. Antonenko et al (2022) applied 1 mA tDCS with electrodes of 5 cm diameter and found no significant differences in reported sensations between active and sham conditions; similar to Yu et al. (2020) that applied 1.5 mA tDCS with 35 cm2 electrodes. Manor et al. (2018) used 2 mA and observed only mild, transient SEs, primarily sensations under the electrode (active: 65 %, sham: 70 %), skin redness (32 % vs. 16 %), and sleepiness (30 % vs. 14 %), and Assecondi et al. (2022), reported only mild AEs over 5 sessions of 2 mA with 3.14 cm2 electrodes over the DLPFC. Similarly, two studies analyzing data from the “Augmenting Cognitive Training in Older Adults” study (Woods et al., 2018) found that, across 20 sessions of 2 mA tDCS over the bilateral DLPFC, participants reported mild sensations (mostly tingling and burning), with no other AEs (Hausman et al., 2023, 2024). A higher number of individuals in the active group reported high blood pressure; however, this has not been linked to the effects of stimulation (Hausman et al., 2023). Self-administered at-home application of 1 mA combined with cognitive training demonstrated not only similar sensations between active and sham conditions, but also excellent feasibility and acceptability (Rocke et al., 2024). Other studies targeted different cortical areas. Dumel et al. (2018) applied 2 mA anodal tDCS over the left M1 hand area (42 cm2 sponge electrodes) but did not report AEs. Luckey et al. (2020) used 1.5 mA tDCS over the C2 dermatomes for 12.5 min (35 cm2 sponge electrodes), with only mild sensations of itching and tingling and no significant differences between active and sham groups. One study also investigated tDCS safety in older adults (70–90 yrs) with pacemakers, applying 2 mA over the DLPFC with no reported AEs (Roncero et al., 2020).
For HD-tDCS, safety considerations were discussed in three studies, including two reviews analyzing clinical trials in healthy and clinical populations, with partial data overlap (El Jamal et al., 2023; Reckow et al., 2018). These reviews examined SEs across different intensities (2 and 3 mA) and stimulation targets (parietal, temporal, or prefrontal cortex), reporting no unexpected or severe AEs. The most frequently reported sensations were tingling, burning, and itching, occurring in 15–59 % of sessions, with most effects rated as mild. Severe sensations were reported in fewer than 4 % of cases, and no serious AEs were observed. One experimental study by Lo et al. (2023) investigated HD-tDCS safety in six participants undergoing 1.5 mA anodal stimulation over the DLPFC for 20 min. Sensations under the electrodes were reported in 12.1 % of sessions, and skin redness in 2.4 %, with no severe effects.
For tACS, no studies in older adults are explicitly assessing safety. However, two reviews (Al Qasem et al., 2022; Wu et al., 2024) discussed its use in modulating cognitive functions in healthy older adults without reporting safety concerns. An ad-hoc search identified 16 tACS studies employing current intensities of up to 2 mA, with stimulation commonly targeting the DLPFC (Diedrich et al., 2024a; Diedrich et al., 2024b; Draaisma et al., 2022; Grover et al., 2022; Krebs et al., 2021), motor cortex (Gamage et al., 2025; Guerra et al., 2021; Rumpf et al., 2019), posterior parietal cortex (Draaisma et al., 2022; Misselhorn et al., 2020), and temporal regions (Reinhart and Nguyen, 2019). These studies utilised a range of frequencies, including theta, alpha, beta, gamma, the-ta–gamma coupling, and individualized frequency protocols (Draaisma et al., 2022; Fresnoza et al., 2018). Overall, only mild AEs were reported, with no significant differences observed between active and sham stimulation groups. Notably, phosphenes were reported in only three studies (Diedrich et al., 2024; Rumpf et al., 2019), and one study using alpha and beta frequency tACS reported ineffective blinding (Rumpf et al., 2019).
In studies employing high-frequency tRNS (100–640 Hz) only mild AEs were reported, although one participant dropped out from the active group due to headache (Brambilla et al., 2021). Studies targeted various brain regions, including the dorsolateral prefrontal cortex (DLPFC) and broader prefrontal areas (Brambilla et al., 2021; Maltezou-Papastylianou et al., 2022), as well as the occipital and fronto-occipital cortex (Esposito et al., 2021; Fertonani et al., 2019).
Conclusions and recommendations: The safety recommendations outlined by Antal et al. (2017) for tDCS remain valid, as no evidence contradicts them. Overall, reported AEs across tES modalities were mild and transient, with no significant differences between active and sham conditions, highlighting the effectiveness of standard sham protocols. However, a lack of use of standardized tools and requirements for AE assessment and reporting remains a limitation, particularly for older adults. Further research is needed, particularly on tACS, and tRNS, and special aging populations such as individuals with pacemakers.
8. Safety for tES as therapeutic intervention
8.1. Note on inclusion and exclusion criteria in therapeutic interventions
The exclusion criteria in therapeutic interventions are generally the same as for healthy people, (see Section 7.1), with the exception of the pathology they are intended to treat. For example, tES interventions for epilepsy will have all exclusion and inclusion criteria as studies in healthy people, but they will include (not exclude) people with a history of epilepsy or epileptic seizures. Furthermore, the possible add-on effects of different kinds of medications should also be considered.
8.2. Psychiatric conditions
8.2.1. tDCS in major depressive disorder and bipolar depression
In major depressive disorder (MDD), large randomized trials applying 2 mA over the DLPFC for 20–30 min per session have reported excellent tolerability, with AEs limited to transient scalp discomfort and no serious AE (Brunoni et al., 2011, 2012). Treatment-emergent mania or hypomania is exceedingly rare, with isolated case reports in bipolar II patients (Chao et al., 2018; Gálvez et al., 2011) and systematic reviews indicating incidence rates below 1 %, comparable to those seen with pharmacotherapy (Dondé et al., 2018). No patients have been reported to discontinue tDCS due to adverse psychiatric or neurological symptoms, underscoring its safety in mood disorders. These findings are reinforced by two recent studies demonstrating the feasibility and safety of at-home tDCS for major depression. In an observational cohort, Charvet et al. (2023) showed significant symptom reductions by week 2 of a 6-week home-based protocol, with continued improvement during a taper phase (Charvet et al., 2023b). In a large, fully remote, multisite randomized controlled trial, Woodham et al. (2025) further demonstrated the efficacy and acceptability of 10 weeks of home-based tDCS, with greater reductions in depressive symptoms compared to sham and no differences in discontinuation rates and no induced manic/hypomanic states (Woodham et al., 2025).
8.2.2. tDCS in schizophrenia
tDCS protocols targeting hallucinations, negative symptoms or cognitive deficits in schizophrenia also reveal no safety issues In childhood-onset schizophrenia, bilateral 2 mA tDCS over the DLPFC and superior temporal gyrus was well tolerated with no serious AEs in 10 sessions (Mattai et al., 2011). Similarly, the STARTS randomized clinical trial of 100 adults with predominant negative symptoms found no significant difference in overall AE rates between active and sham groups (10 sessions, 2 per day), aside from a higher rate of a mild scalp burning sensation in the active arm (active 43.8 % vs sham 14.3 %) and no participants withdrew due to AEs (Valiengo et al., 2020). A recent meta-analysis compiling data from multiple randomized controlled trials (RCTs) reported that tDCS using the same electrode montage was safely administered to over 320 patients with treatment-resistant auditory hallucinations, without any reported serious AEs, no exacerbation of positive psychotic symptoms, and no discontinuation (Jiang et al., 2022).
8.2.3. tDCS in other psychiatric conditions
In drug-resistant obsessive compulsive disorder (OCD), add-on tDCS trials have reported only mild, self-limited itching, tingling, and transient headache, without serious AEs or treatment discontinuations ((da Silva et al., 2019), for both once- and twice-daily sessions: (Bation et al., 2019)).
In posttraumatic stress disorder (PTSD), 10 sessions of tDCS (2 mA) delivered over the DLPFC documented no serious AE, with mild Aes confined to brief scalp sensations and transient headaches (Ahmadizadeh et al., 2019).
In anorexia nervosa, a double-blind, sham-controlled tDCS trial in 43 inpatients reported no serious device-related AEs; AE rates were comparable between active and sham groups, with two withdrawals due to headache and one transient mood elevation suggestive of hypomania (Baumann et al., 2021).
8.2.4. tACS in psychiatric conditions
A limited number of controlled trials have investigated the efficacy of tACS in psychiatric conditions (Frohlich and Riddle, 2021; Gholamali Nezhad et al., 2024; Perera et al., 2023). In this study population, additional awareness about potential risks is required, including elevated baseline suicide rates when compared to healthy populations without psychiatric conditions. Three RCTs have examined tACS for depression. Two of them used a proprietary waveform that the authors refer to as tACS despite both the waveform and montage differs from tACS. An RCT of this stimulation paradigm in the first depressive episode in MDD(20 daily sessions, 40 min,77.5 Hz,15 mA) reported tinnitus, discomfort, headache and itches as the most frequent AEs with no statistically significant difference between groups (Wang et al., 2022). No serious AEs occurred, and importantly, no transitions to (hypo-)manic symptoms or suicide attempts were reported. The same stimulation paradigm was then studied in patients who were receiving SSRI medication treatment (Zhou et al., 2024). Similarly, no serious AEs and no statistically significant difference in AEs (headaches, drowsiness, dizziness) was reported. An RCT of alpha- and gamma-frequency tACS demonstrated safety for lower frequency stimulation that directly targets cortical oscillations associated with depression symptoms (Alexander et al., 2019). This study reported worsening of suicidal ideation in four participants, all of which were in the sham group. No participant developed (hypo-)mania and no suicide attempt occurred in the study. The only stimulation experience that differed between groups was phosphenes, which do not represent a safety concern but rather a nuisance for effective study blinding. Together, these studies support the safety of tACS in depression.
RCTs of tACS in psychotic disorders (Chang et al., 2021; Mellin et al., 2018; Zhang et al., 2022b) further support this conclusion. An additional study investigated the use of individualized alpha tACS in OCD (Perera et al., 2023). No SAEs were reported, and only minor infrequent SEs and AEs (headache, tingling, phosphenes, itching). Overall, the current literature supports the safety of tACS in psychiatric populations given the lack of serious AEs and no statistically significant differences in stimulation-associated sensory perception and SEs. However, the heterogeneity of stimulation protocols and study populations together with the small sample size make general conclusions challenging.
Conclusions and recommendations: Overall, evidence across a broad spectrum of psychiatric conditions attests to a favorable safety and tolerability profile of tDCS, characterized by only mild, transient AEs and an absence of serious or lasting adverse outcomes. Regarding tACS, none of the studies reported any serious AEs. Furthermore, most stimulation related SEs were not statistically different between active and sham groups. Of particular relevance, no attempted or completed suicide attempts have been reported in any of the published tACS trials in psychiatric disorders. Together, the current literature supports the safety of tACS in psychiatric disorders. Nevertheless, given the vulnerability of this population, assessment of suicide risk at enrollment together with monitoring of changes to suicide risk throughout the study are recommended.
8.3. Chronic pain and migraine
There is evidence demonstrating the general tolerability, favorable safety profile, and low incidence of AEs in individuals with chronic pain or migraine. The most commonly reported SEs and AEs are: (i) mild and transient sensory disturbances at the site of stimulation, including itching, tingling, burning sensations, or light headaches; (ii) general symptoms, including sleepiness, fatigue, dizziness, nausea, mood change and trouble to concentrate; (iii) local AEs, such as skin redness (erythema) (Ahdab et al., 2019; Hervik et al., 2024; Jobin et al., 2025; Mendes et al., 2024). In sham-controlled studies, the occurrence of such SEs and AEs, except for local erythema under the electrodes, did not differ between the active and sham conditions (Alwardat et al., 2020; Cerrahoğlu Şirin et al., 2021; Hong et al., 2022). However, it remains essential to implement systematic monitoring and documentation of AEs throughout the intervention period to ensure participant safety.
Notably, no serious AEs of tDCS treatment have been reported across a range of clinical populations, including individuals with fibromyalgia (Yang et al., 2024), chronic low back pain (Alwardat et al., 2020), spinal cord injury (Portaro et al., 2025), chronic cancer pain (Grenouillet et al., 2025), episodic migraine (Ahdab et al., 2019; Hong et al., 2022), abdominal pain (Bayer et al., 2019), chronic headache (Hervik et al., 2024), and comorbid psychiatric conditions such as depression, anxiety (Wen et al., 2022), and bipolar disorder (Mastria et al., 2021). These findings support the safe application of tDCS in both clinical and investigational contexts in the pain domain.
Home-based tDCS has also been explored for chronic pain treatment and appears to be generally safe, well-tolerated, and feasible for self-administration (Antonioni et al., 2024). This is further supported by a pilot randomized controlled trial in veterans with persistent post-traumatic headache, where 20 sessions of home-based, remotely supervised tDCS significantly reduced moderate-to-severe headache days and total headache days, with high adherence and no significant safety concerns reported (Charvet et al., 2023). In one study of 12 patients with chronic pain syndrome treated by tDCS at home for several weeks (Garcia-Larrea et al., 2019), AEs were further detailed: two patients experienced acute AEs, including nausea, headache, or digestive disorders. It was concluded that these events were unrelated to the tDCS procedure. The treatment was not interrupted and was effective in reducing pain; two other patients reported superficial skin burns beneath the electrodes due to incorrect electrode application. This led to changing the type of electrodes (from Ag/AgCl electrodes of 3.14 cm2 contact area to larger sponge electrodes of 8 cm2 contact area) without further problems for the rest of the treatment.
tACS has been proposed to restore pathologically altered rhythmic brain activity associated with pain. There are only a few tACS trials performed in chronic pain patients, in the context of migraine (Antal et al., 2020) or chronic low-back pain (Ahdab et al., 2019; Ahn et al., 2019; Prim et al., 2019). Neither study reported any serious AEs. For the migraine study (Antal et al., 2020), tACS was performed at home (single session, 15 min at 0.4 mA at 140 Hz, Oz-Cz montage) and a few participants reported pain under electrodes, tingling, itching, nervousness, fatigue, and unpleasantness. However, the occurrence rate was low and no statistically significant differences were reported compared to sham procedure. Similarly, a single session of alpha-frequency tACS (20 min at 1 mA, 10 Hz) applied in a cross-over design with an active sham comparator in individuals with chronic low back pain was not associated with a statistically significant difference in stimulation-related SEs (Ahn et al., 2019; Prim et al., 2019). Overall, tACS appears well tolerated and safe. However, the small number of studies and the low number of participants do not allow for any strong conclusions. However, it is noteworthy that, in a sham-controlled study on patients with fibromyalgia, one patient in the active group attempted suicide by overdose. In this study, an unusual theta-burst protocol was delivered (monophasic square wave with a pulse width of 0.5 ms and intensity of 1 mA at 50 Hz, repeated at duty cycle with on time of 2 s and off time of at 1 mA). The psychiatric history of the patient was not provided, and caution is advised in determining whether this event was causally related to the study intervention. In any case, it is recommended that in participants with complex medical or psychiatric histories, a suicide risk screening is included in the study protocol.
Recommendation: While no significant risks have been identified for the use of tDCS to treat chronic pain, even for several weeks at home, the long-term safety of extended tDCS use remains to be investigated to refine safety protocols for prolonged therapeutic use. Regarding tACS, it appears well tolerated and safe. However, the small number of studies and the low number of participants do not allow for any strong conclusions.
8.4. Stroke neurorehabilitation
tDCS has been applied in stroke rehabilitation across 177 published studies, involving a total of 6,509 patients. Depending on the rehabilitation target, tDCS targeted the ipsi- or contralesional motor cortex in most studies, but also other areas such as nodes of the fronto-parietal language network. Stimulation intensities typically ranged between 1–2 mA, though safe administration at intensities up to 4.0 mA has been reported (Chhatbar et al., 2017; Schlaug et al., 2025). The typical session duration in stroke recovery trials ranged from 20 to 30 min, with some extending up to 40 min. tDCS was well-tolerated, and AEs were both mild and transient. The most commonly reported AEs during active stimulation included tingling (n = 108 vs. 45 for sham), itching (n = 56 vs. 34), skin redness (n = 36 vs. 8), headache (n = 17 vs. 9), mild burning (n = 16 vs. 11), and fatigue (n = 16 vs. 2). Serious AEs, such as stroke, myocardial infarction, or death, were rare and occurred at similar rates in both the active (n = 12) and sham (n = 13) groups. Notably, the study authors consistently judged these serious events to be unrelated to the tDCS intervention and more likely attributable to the patients’ pre-existing medical conditions. Home use of tDCS has also been validated for patients recovering from aphasia (Richardson et al., 2023).
TDCS in the hyperacute phase of stroke has been less well studied. The first two RCTs applying conventional tDCS and HD-tDCS in hyper-acute stroke were conducted in France (STICA) and the USA (TESSERACT), (Bahr-Hosseini and Bikson, 2021; Pruvost-Robieux et al., 2021). These pilot studies were underpowered for efficacy evaluation. However, the results suggested a benefit upon penumbral salvage. The stimulation was safe and well tolerated in emergency settings.
Anodal tDCS was applied to the swallowing sensorimotor cortex or the pharyngeal motor cortex to improve swallowing in patients with post-stroke dysphagia, showing a strong safety profile. In an RCT with 42 patients receiving either up to two daily 20-min sessions of anodal tDCS targeting the unaffected swallowing motor cortex at 2 mA for 10 days or sham tDCS, no decline in global neurological, motor or swallowing functions or seizures were reported in any of the intervention arms (Kumar et al., 2022). Unanticipated serious AEs reported were adjudged as not being related to the stimulation (see Section 13.1 for details). Anodal tDCS of the contralesional swallowing sensorimotor cortex at 1.6 mA for 20 min daily for 6 days over 8 weeks in patients with brainstem stroke was without any AEs (Mao et al., 2022). No AEs were also reported in other trials with dysphagic stroke patients applying either four anodal tDCS of the contralesional swallowing motor cortex 20-min sessions at 1 mA (Suntrup-Krueger et al., 2018) or bilateral anodal tDCS of the pharyngeal motor cortices at 1 mA for 20 min daily for 10 days (Ahn et al., 2017).
tACS was reported in 11 studies involving 156 subacute or chronic stroke patients that were treated with a broad variety of tACS protocols. Protocols varied widely in intensity (0.4–3.0 mA), frequency (6–120 Hz), and session duration (up to 120 min), with treatment plans including up to 20 sessions. Only 5 articles reported on perceptual SEs, AEs and/or tolerability (Grigutsch et al., 2024; Kitatani et al., 2020; Middag-van Spanje et al., 2024; Omae et al., 2024; Schuhmann et al., 2022). No SEs or AEs were noted, and tolerability was high. Five patients discontinued study participation, four with reasons not related to the tACS intervention, and one due to a mismatch of expectations with the therapeutic program.
tRNS has demonstrated a favorable safety profile in stroke, with SEs predominantly mild and transient. Current evidence indicates that tRNS applied at 2 mA (100–500 Hz) for 30 min is generally associated with minimal sensations such as tingling (Sethi et al., 2023). When combined with movement-triggered stimulation (Hayward et al., 2017) or integrated into motor rehabilitation programs such as the graded repetitive arm supplementary program (Arnao et al., 2019), tRNS has not been associated with AEs. Across the three published studies (n = 19 patients treated with tRNS), no serious AEs have been reported, and participant adherence has been consistently very high, with no dropouts observed.
Recommendation: While there is no evidence of serious AEs associated with tDCS, tACS, and tRNS in stroke patients, ongoing monitoring for mild AEs is recommended. Adjustment of stimulation parameters (intensity, duration, electrode montage) may further improve comfort and tolerability. The integration of tDCS, tACS or tRNS with adjunctive therapies (e.g., functional electrical stimulation or motor training) appears feasible and well tolerated, though standardization and individualization of protocols are recommended to ensure safety and optimize outcomes. The safety evidence is strongest for tDCS, with over 6500 S patients studied. In contrast, evidence for tACS and tRNS remains limited, highlighting the need for larger, well-controlled trials.
8.5. Multiple sclerosis (MS)
tDCS has been extensively studied for its safety and tolerability in individuals with multiple sclerosis (MS), particularly as a home-based intervention delivered through remotely supervised tDCS (RS-tDCS). Fatigue was among the earliest symptoms targeted, as this debilitating condition lacks consistently effective treatments in MS. In an initial randomized, sham-controlled trial using home-based RS-tDCS targeting the left DLPFC, 20 sessions were well-tolerated and associated with reductions in fatigue (Charvet et al., 2018). A subsequent larger multisite trial confirmed the safety and feasibility of the same protocol (Charvet et al., 2023a). There were no serious AEs, across all participants in both conditions, 9 AEs were rated >7 on the VAS; in all instances, the session resumed. One participant discontinued one session due to discomfort at the stimulation midway check-in point. Otherwise, all AEs were reported as mild in intensity and resolved after the session ended. Concerning feasibility, in the 117 individuals randomized, an average of 26 ± 9 sessions were completed; only four patients withdrew before completing any sessions and seven withdrew who completed fewer than 10 sessions. Attrition remained minimal, demonstrating a high level of fidelity to treatment, with 92.0 % of participants completing at least 25 of the 30 sessions by the end of treatment.
Importantly, tDCS has demonstrated a favorable safety profile in individuals with MS over extended treatment intervals (Pilloni et al., 2022). Two systematic reviews report that adverse events such as headache, insomnia, and mild sensory symptoms occurred at similar rates between active and sham stimulation, with no SAEs reported (Ashrafi et al., 2020). Applications of RS-tDCS for cognitive symptoms have also consistently demonstrated strong safety and tolerability. Multiple randomized trials have shown no SAEs while reporting benefits in complex attention, response variability, and overall cognitive performance, particularly in participants with greater baseline disability (Charvet et al., 2018, 2025; Riemann et al., 2025). Applications for motor symptoms have been similarly well-tolerated. A trial combining 10 sessions of M1-targeted tDCS (2.5 mA) with aerobic exercise demonstrated excellent tolerability along with improved gait speed and endurance (Pilloni et al., 2020a). Only perceptual effects of itching, tingling, and head pain were reported with no higher intensity level of >7 (rated on a 0- to 10-point scale) for any participant, and all of them were resolved at the end of the stimulation period. In progressive MS, home-based RS-tDCS paired with dexterity training over 20 sessions showed positive outcomes in hand function and quality of life, with no serious safety concerns (Pilloni et al., 2024). 1077 at-home RS-tDCS sessions were completed, no session was discontinued due to tolerability, and only during six sessions (n = 3 active and n = 3 sham tDCS sessions) did participants report discomfort >7 on the VAS. Only one participant reported itching with an intensity higher than 7 in two separate sessions. In all cases, all participants opted to continue treatment and the session was resumed without further tolerability issues. Across sessions, tingling was the most frequently reported sensation (reported by 30 % active versus 35 % sham participants) followed by itching (reported by 18 % vs 7 %) and warmth sensations (reported by 15 % vs 8 %). tDCS has also been explored for comorbid behavioral conditions. In a randomized trial of women with MS and cannabis use disorder, RS-tDCS paired with mindfulness meditation was safely administered and associated with reduced cannabis use, decreased withdrawal symptoms, and improved cognitive function (Pilloni et al., 2025). 45 participants initiated the intervention (n = 30 active tDCS, n = 15 sham tDCS), with 83 % completing at least 14 of the 20 planned sessions. There were no dropouts related to the intervention, the most commonly reported sensations during the sessions were tingling, itching, and warmth, in agreement with our previous reports.
Beyond tDCS, other noninvasive stimulation modalities have also been evaluated in MS. A study of bifrontal 6 Hz tACS reported only mild to moderate AEs across 180 sessions, with no differences between active and sham groups, and notably, no phosphenes reported by any participant (Hsu et al., 2023). Similarly, tRNS was studied for effects on affect, pain, and attention; AEs such as insomnia, nausea, and headache occurred at comparable rates in active and sham groups, and only one sham-treated participant reported phosphenes (Palm et al., 2018). Discomfort ratings did not differ by stimulation condition, further supporting the tolerability of these interventions.
Conclusions and recommendation: While no significant risks have been identified for the use of tDCS to treat fatigue and cognitive symptoms in MS even for several weeks at home, the long-term safety of extended tDCS use (e.g, years) remains to be investigated to refine safety protocols, for prolonged therapeutic use.
8.6. Dementia and mild cognitive impairment (MCI)
A network meta-analysis of randomised controlled trials on the beneficial effects of NIBS on cognition in dementia (Tseng et al., 2023), which included 7 tDCS studies, found that none of the NIBS methods were associated with different rates of any AEs compared with sham controls. A study of 11 patients with mild cognitive impairment (MCI), who received 2 mA tDCS for 20 min per day over 5 consecutive days (55 sessions in total), reported a pricking sensation (n = 30, 55 %) and a mild burning sensation (n = 8, 14 %) as the most common AEs; no AE was reported in 15 sessions (31 %), while all reported AEs were of mild intensity (Murugaraja et al., 2017). Another study that assessed 46 patients with subjective or objective cognitive impairment after 9 sessions of combined cognitive training with either anodal tDCS (1 mA) or sham tDCS over the DLPFC found 15 AEs in 4 participants in the target group, and 16 AEs in 7 participants in the control group, without significant difference between groups; no SAEs occurred in this study (Antonenko et al., 2024). In terms of efficacy, no significant difference emerged in the primary outcome measure of trained task performance in the letter updating task, but a significant difference was found for the N-back working memory tasks (secondary outcome).In individuals with both MCI and depression, a six-week randomized sham-controlled trial demonstrated that prefrontal tDCS was ‘well tolerated’ and associated with electrophysiological markers of enhanced neuroplasticity, although behavioral improvements did not reach significance (Kim et al., 2024). In the paper no AEs or perceptual effects of the stimulation are reported. A larger multisite randomized clinical trial combining prefrontal tDCS with cognitive remediation in older adults with MCI and/or remitted major depressive disorder (rMDD) also confirmed the safety and acceptability with no serious AEs reported (Rajji et al., 2025). Another double-blind randomized controlled trial in amnestic MCI similarly confirmed the feasibility and tolerability of tDCS delivered with cognitive training; there were no significant differences between conditions in the frequency of the following reported AEs: redness, tingling, mild burning, pain, nausea, light headedness, headache, blurred vision, fatigue (Martin et al., 2019) For the ITT sample, the most common AEs reported across all sessions were: tingling (25.9 %), redness (16.2 %), mild burning (11 %), and itching (10.5 %) (Martin et al., 2019). Another trial similarly confirmed the feasibility and tolerability of tDCS delivered with cognitive training, although it did not find an additive cognitive benefit over sham (Hausman et al., 2023). Throughout the trial, the sham group reported more body-related AEs, while the active group reported more cardiac-related AEs. In particular, there was a greater number of individuals reporting high blood pressure in the active group (15 in the active vs 6 in the sham group), which has not been shown to be related to effects of stimulation. In the domain of language disorders, a recent study in individuals with primary progressive aphasia found that home-based RS-tDCS paired with speech-language therapy was feasible, well tolerated, (George et al., 2025). In this observational study only 10 patients participated. All participants successfully completed all scheduled intervention sessions, resulting in a 100 % completion rate.. Consistent with prior studies, there were no serious AEs reported during the intervention. Mild, self-limited sensations such as tingling at the electrode site were reported by two participants but did not disrupt adherence or require further intervention. Conclusions: No specific safety concerns related to tES are detected specific to this patient group.
8.7. Amyotrophic lateral sclerosis (ALS)
A recent study of home-based tDCS in amyotrophic lateral sclerosis (ALS) (Madhavan et al., 2025b), following 72 tele-tDCS sessions over 24 weeks in 16 patients, reported only usual mild to moderate AEs (with itching and tingling being the most pronounced), without significant difference between the intervention and delayed-onset (control) groups.
Conclusions: No specific safety concerns related to tES are detected specific to this patient group.
8.8. Parkinson’s disease (PD)
No AEs were reported in studies including patients with Parkinson’s disease (PD), as shown in a network meta-analysis (Qiu et al., 2023). Moreover, RS-tDCS has been found feasible for home use in PD, yielding benefits in fatigue and cognitive processing speed with high tolerability; tingling (43 %) and burning sensation (29 %) were common but represent mild AEs (Dobbs et al., 2018). Regarding cerebellar tDCS, in a meta-analysis including 12 studies on movement disorders, one reported local skin erythema, five reported no AEs, but the other six did not mention relevant information(França et al., 2018)). Other tES methods have been tested as a treatment for PD as well. Two studies, based on gamma-frequency tACS (Guerra et al., 2022) or tRNS (Monastero et al., 2020) delivered over M1 did not report significant safety concerns. Conclusions: No specific safety concerns related to tES are detected specific to this patient group.
8.9. Epilepsy
Low-intensity tES was used in patients with epilepsy for various purposes, much of it using cathodal tDCS. Across all these use cases, there is no evidence of an increased safety risk to epileptic patients (Bikson et al., 2016; Nitsche and Paulus, 2009; Simula et al., 2022; Sudbrack-Oliveira et al., 2021). However, the available evidence remains limited, and while hyper-excitable brain networks may respond differently to electric fields, this does not constitute evidence of a safety risk. The common exclusion of history of seizures from tES protocols is based on precautions using rTMS.
Conclusions: No specific safety concerns related to tES are detected specific to this patient group.
8.10. Other clinical conditions
8.10.1. Viral infections
In patients with Human Immunodeficiency Virus (HIV), 10 HD-tDCS sessions targeting the cingulate cortices at 1.5 mA over two to three weeks to improve brain function was well-tolerated, with no significant difference in self-reported discomfort ratings between sham and active conditions (Jiang et al., 2022). Ten 20-min sessions of anodal tDCS of the right inferior frontal cortex at 2 mA combined with cognitive training to improve cognitive functioning in 33 older adults with HIV did not lead to any AEs (Fazeli et al., 2019). Cody et al. (2020) tested the same protocol as Fazeli et al. (2019) comparing its impact on sleep quality and processing speed in the elderly with versus without HIV. One participant reported a burning sensation similar to a sunburn during the stimulation and the tDCS intensity was reduced to a level she could tolerate. Nevertheless, it was not specified whether the participant was in the HIV or non-HIV sample.
The safety profile of tDCS was also explored in different subgroups of COVID-19 patients. The combination of HD-tDCS targeting the left diaphragmatic M1 (twice per day at 3 mA for 30 min) and respiratory rehabilitation for up to 10 days in 56 critically ill patients was well-tolerated and safe, with similar patterns in AEs in both sham and active groups (Andrade et al., 2022). The only AE associated with active tDCS was mild and transient redness of the scalp. In a recent trial testing the effects of 10 sessions of HD-tDCS of the left M1 at 3 mA for 20 min paired with rehabilitation on fatigue in 70 patients with post-acute sequelae of COVID-19, the occurrence of no other AEs, except skin redness, differed significantly between sham and active arms (Santana et al., 2023, p. 20). In addition, no serious AEs were reported in this study. A single session of bilateral prefrontal tDCS over the DLPFC at 2 mA for 30 min in 40 COVID-19 inpatients was safe, with no serious AEs for any patient up to 1 h post intervention and no differences in occurrence and intensity of AEs between sham and active groups (Pinto et al., 2023). Headaches and sleepiness were the most frequently reported consequences in the sham group while tingling was the most reported SE in the active group.
Currently, a large multi-center trial is evaluating a 50-session RS-tDCS protocol paired with cognitive training to treat cognitive impairment in individuals with post-acute sequelae of SARS-CoV-2 infection (Knopman et al., 2024).
8.10.2. Cancer
In a recent scoping review by Capetti et al. (2024) including 10 trials testing the effects of tDCS in the reduction of pain and psychocognitive symptoms in cancer patients, only Gaynor et al. (2020) reported AEs. Two sessions of anodal tDCS at 1 mA over the left DLPFC in an F3-F4 montage to improve sustained attention in 16 breast cancer survivors was well-tolerated, with burning sensation on the scalp and itching being the most frequent AEs rated as mild or moderate. No sessions needed to be stopped because of AEs (Gaynor et al., 2020).
8.10.3. Obesity
tDCS protocols targeting weight loss in obesity demonstrated a favorable safety profile in patients with obesity. A systematic review including 7 randomized clinical trials with 170 patients using mainly anodal tDCS of the DLPFC (unilateral or bilateral) at 2 mA for a duration of 20 to 40 min documented no severe AEs (Gouveia et al., 2021). Only transient mild AEs including headache, neck and scalp pain, scalp burn, and skin redness were reported in both active and sham intervention groups.
8.10.4. Type 1 diabetes mellitus
In type 1 diabetes mellitus, one 20-min session of HD-tDCS targeting the bilateral DLPFCs with total current delivered of 1.5 mA for improving performance in cyclists was well-tolerated and no SE or AE was reported during or after the stimulation (Filipas et al., 2022).
8.10.5. Systemic lupus erythematosus
Only one study investigated the impact of tDCS in patients with systemic lupus erythematosus (de Andrade et al., 2024). The combination of five sessions of anodal tDCS over the M1 at 2 mA and low-intensity aerobic exercise was well tolerated, with the presence of only mild and transient AEs and the absence of any disease flares.
8.10.6. End-stage renal disease
The safety of 10 20-min daily sessions of anodal tDCS of the M1 at 2 mA in 30 patients with end-stage renal disease undergoing hemodialysis was evaluated in an exploratory study through monitoring the participants’ tolerance and subjective sensations to the stimulation (Pegado et al., 2024). Commonly reported SEs and AEs were headache, nausea and tingling, with no serious AE. The stimulation was well tolerated by the participants.
8.10.7. Tinnitus
A variety of tDCS, tACS, tRNS protocols have all been investigated for the treatment of tinnitus, (Chen et al., 2020). When AEs were reported, there were either no AEs in either group or transient events such as tingling, dizziness, headaches, facial numbness, itching, scalp pain, burning, pinching, fatigue, headaches or skin irritation, which were typically observed in both active and sham stimulation groups (Kreuzer et al., 2019; Pal et al., 2015; Smeele et al., 2023; Yang et al., 2021). None of the investigated tES protocols were associated with significantly different dropout rates when compared with the sham control (Chen et al., 2020). Occasionally, transient increases in tinnitus were reported, sometimes resulting in patients dropping out of the studies. However, these increases ceased spontaneously. No SAEs have been reported in any study.
Conclusions: This section reviews the safety of tES in less commonly studied disorders, mainly focusing on tDCS. No specific safety concerns were identified for these clinical indications. However, many trials did not assess SE/AE, thus more comprehensive data recording is needed.
9. Combining tES with other methods and interventions
9.1. tES combined with MRI/fMRI
Low-intensity tES during MRI has become a well-established protocol for functional MRI and MR spectroscopy investigating tES-induced brain effects, and electrical tissue property imaging. The first edition of the “Safety, Ethical, Legal, Regulatory, and Application Guidelines,” published in 2017, included a dedicated chapter addressing tES during MRI (Antal et al., 2017). It concluded that low-intensity tES can be safely used during MRI if specific precautions are strictly followed. This section synthesizes three new studies that specifically addressed safety aspects of concurrent tES-MRI applications and concludes with an updated version of the original recommendations from 2017.
During MR scanning, the TES electrodes and leads inside the scanner bore are exposed to strong radiofrequency (RF) excitation fields (reaching 10–30 kW), requiring special safety considerations to prevent ohmic heating and skin burns. Heating due to low-frequency currents induced by MR gradient switching is usually not of concern. A dedicated tES-MRI safety study has evaluated a new setup for low-intensity tES during MRI using low-conductivity silicon rubber leads to reduce RF coupling and heating (Gregersen et al., 2021). If MRI-compatible tES stimulators use highly conductive copper leads (σ = 5.8 × 107 S/m), 5 kΩ resistors are added to avoid antenna effects in the copper wire at ¼ RF wavelength. High RF electric fields may occur near the safety resistors, so they should be well-separated from the skin. To reduce RF-induced heating and antenna effects, leads with distributed resistance and lower total impedance are preferred. Carbon fiber leads, common in EEG-MRI, lower SAR compared to copper. For tES-fMRI applications, even safer performance is achieved using low conductivity silicone rubber leads (σ = 29.4 S/m). Simulations and temperature tests confirmed improved safety, minimal SAR increase, and no overheating. At the same time, the ohmic resistance of the silicone rubber leads could be kept below 5 kΩ, enabling higher tES current intensities compared to copper leads with added safety resistors.
A simulation study developed a workflow to assess RF-induced heating during MRI with a tES setup, showing that temperature rise is highly sensitive to skin conductivity, electrode and gel geometry, and phantom materials. These findings underscore the importance of precise modeling and setup design to ensure accurate safety evaluations and prevent overheating risks during combined tES-MRI procedures (Kozlov et al., 2020).
More than 100 peer-reviewed studies have used low-intensity tES concurrently during MR scanning for functional MRI, MR-spectroscopy, or electrical tissue property imaging. These studies have not reported severe AEs, but explicit reporting of AEs is poor in the existing literature. To address this issue, a checklist for assessing the methodological quality of concurrent tES-fMRI studies, the ContES checklist, has been developed through expert consensus using a delphi procedure (Ekhtiari et al., 2022). In addition to key methodological and technological factors such as electrode setup and stimulation parameters, key safety-relevant factors were identified and grouped as a separate category of items. These items emphasize the need to report MR conditionality of the tES setup, results of safety and artifact testing, any participant intolerance, and impedance measurements to ensure safe and interpretable tES-fMRI integration. Reporting these items was considered to be extremely or highly important. This contrasts with a relatively low reporting rate of approximately 30 % for most items that emerged when three independent raters evaluated 57 concurrent tES-fMRI studies. Reporting rate of MR conditionality specifications, such as safety tests for tES-fMRI setting or safety was very low (5 %) in these 57 studies, and thus standardized reporting of presence and absence of safety aspects, including AEs, is needed in future tES-MRI studies.
Conclusions and recommendations: Specific safety precautions apply for concurrent tES-MRI. The study protocol must always comply with the safety standards for both tES and MRI. The stimulator (containing ferromagnetic parts) must remain outside the MR room, and the use of MR-compatible cables, electrodes and electrode leads is strictly required. The use of RF filters is strongly encouraged to prevent noise interference on the MR data. Filters are also effective in preventing the cables transmit RF energy of the MR scanner to the stimulator. The electrode leads close to and inside the scanner bore must be arranged without loops and their path lengths inside the bore should be kept as short as possible. Low conductivity silicone rubber electrodes in conjunction with a thick conductive paste are recommended to minimize the risk of burns during prolonged imaging sessions. Saline-soaked sponges are contraindicated due to their tendency to desiccate during extended scans. MR conditionality of tES equipment refers to specific application scenarios (such as scanner field strengths, MR sequences and sequence parameters) approved by the manufacturer. Concurrent tES-MRI studies outside these scenarios require additional clearance by the manufacturer and/or safety evaluations by qualified MR experts.
9.2. tES simultaneously combined with TMS/rTMS/TBS/ECT
Since 2017, there have been only a few clinical trials in which some form of tES has been applied simultaneously with plasticity-inducing forms of rTMS or theta burst stimulation (TBS) for therapeutic purposes: tDCS with rTMS in moderate Alzheimer’s Disease (Hu et al., 2022), or tDCS with TBS in MDD (Wivatvongvana et al., 2024), without noting any AEs. In addition, the concomitant use of tDCS over the DLPFC (anode left/cathode right, 2 mA, 30 min), followed by ultrabrief right unilateral electroconvulsive therapy (ECT) administered 5–10 min later, appears to be both feasible and safe (Mayur et al., 2018).
tACS at different frequencies has been used in combination with TBS to investigate physiological effects of brain stimulation, including plasticity, in healthy subjects, without noting any AEs (Guerra et al., 2018; Maiella et al., 2022; Nakazono et al., 2021; Ogata et al., 2021; Takahashi et al., 2024; Tremblay et al., 2017).
All the remaining studies were performed in healthy subjects, again for physiological purposes, mainly in order to study relationships between the effects of EEG activity, tACS frequencies and TMS pulse(s) on cortical excitability, either when TMS pulse was phase-locked or not with the underlying oscillatory EEG activity or with the applied tACS (Fehér et al., 2017, 2022; Feurra et al., 2019; Geffen et al., 2021; Guerra et al., 2016, 2020; Pozdniakov et al., 2021; Raco et al., 2016, 2017; Schilberg et al., 2018; Shpektor et al., 2017). No AEs were noted.
Conclusions: when tES simultaneously applied with TMS/rTMS/TBS there is a possibility that changes in neuronal excitability produced by electrical polarisation can directly affect the magnetic stimulation induced changes. However, there is no evidence that the addition of single pulses of TMS to tES increases the incidence of AEs over that of tES alone.
The above findings confirm the conclusion from our previous review (Antal et al., 2017) that there is currently no evidence that simultaneous tES and TMS, when established protocols are followed, introduces a safety risk. Factors such as tolerability (TMS induction of current in tES wires) or equipment interaction should be considered, but are not known to impact safety.
9.3. tES combined with EEG/MEG
To demonstrate on-line and/or off-line effects of tES, it is desirable to record EEG or MEG during and/or immediately before and after tES experiments. Recording EEG during and/or immediately before and after tES requires having EEG electrodes positioned on the scalp and often connected to an EEG amplifier during the stimulation. While applied current might hypothetically divert into EEG electrodes which might then heat up, this is not in fact expected or observed. The reason is the very high input impedance of modern EEG amplifiers (typically in the range of megaohms – MΩ), thus limiting currents to the range of microamps, which cannot result in relevant heating. In addition, the EEG electrode itself presents an electrochemical impedance. Moreover, even if an EEG amplifier could hypothetically represent a short-circuit at the end of the EEG wires, the tES stimulator, being a constant current source, would still limit its current to the selected strength, thus eliminating any potential for tissue damage. Another hypothetical concern is changes in current flow pattern by the presence of the EEG electrodes on the scalp. However, no heating of EEG electrodes was reported in numerous published tES-EEG studies. Moreover, those studies have not reported AEs other than those seen in tES without additional electro-physiological monitoring (Antal et al., 2008; Helfrich et al., 2014; Zaehle et al., 2010) or tDCS (Antal et al., 2004; Cunillera et al., 2016; Faria et al., 2012; Luft et al., 2014; Mancini et al., 2016; Marshall et al., 2006).
Simultaneously recording MEG during tES experiments poses no risks to the participants, since the MEG sensors do not touch the skin and thus cannot conduct tES currents. Various authors have demonstrated combined MEG/tES studies without safety issues (Neuling et al., 2015; Ruhnau et al., 2016). It should be mentioned that, meanwhile, even intracranial recordings have been conducted during tES without safety concerns (Guidetti et al., 2022; Huang et al., 2017; Lamoš et al., 2025; Louviot et al., 2022).
When tES is combined with EEG, if a stimulation electrode is short-circuited to another stimulation electrode, current will most likely become ineffective. If a stimulation electrode is short-circuited to an EEG electrode, there will be an exceptionally strong artefact in the EEG. If two EEG electrodes are short-circuited, no EEG signal will be observable. Therefore conductive fluids between electrodes must be prevented to avoid short-circuiting adjacent electrodes. Electrode gel is preferable to saline solution (Fehér and Morishima, 2016).
When EEG is recorded during tES, artefacts contaminate the EEG recordings. Artefacts are not a straightforward safety concern, but a recording loaded with artefacts is useless and conducting it would represent an unnecessary burden to participants in experiments and/or treatment procedures, thus causing ethical concerns. Although no device produced so far is able to simultaneously stimulate and record EEG, using the same electrodes, the development of multichannel stimulation/recording devices has allowed EEG recordings from surrounding electrodes even during the stimulation. To acquire meaningful EEG data that leads to interpretable results, the artefacts of tES need to be avoided, removed or minimized ((Helfrich et al., 2014; Neuling et al., 2015), for a review, see (Kasten et al., 2019)). Some authors, however, were not able to remove the artefacts and claimed that they are inherent to recording EEG during tACS and might be impossible to completely remove (Gebodh et al., 2019; Noury et al., 2016). It should be noted that the artefact of tES can exceed amplitudes of 100 mV and bring some EEG amplifiers into saturation (Fehér and Morishima, 2016). Thus, it is advisable to use EEG amplifiers with 24 bit or more and to set the analog/digital (AD) range to well cover the size of the artefact (von Conta et al., 2022). Importantly, EEG amplifiers already show small non-linearities with the specified AD range, which can result in artifactual EEG oscillations at the frequency of interest due to nonlinear distortion of tACS signals (Kasten et al., 2018). Furthermore, the way the sine wave is generated in the tES device (e.g., for tACS) can be responsible for whether artifacts around the stimulation frequency can be seen in the EEG/MEG. Thus, tES devices that work with a high digital to analog conversion quantization (minimum 16 bit) and high sampling rate (e.g. >10 kHz) to synthesize a sine wave signal are advantageous here.
Conclusion: Combining tES with EEG/MEG does not pose additional safety risk. In the simultaneous recording of brain activity by either EEG or MEG during/before/after tES, no additional sensations or other AEs, beyond ones generally encountered with tES, have been reported.
9.4. Special considerations for intracranial electrodes and implants
Measurements using intracranial recordings to study the mechanisms of tES in humans have become increasingly popular (Guidetti et al., 2022). Early studies using stereo EEG (sEEG) recordings found that tES produces electric fields up to 0.4 mV/mm per 1 mA (Huang et al., 2017; Opitz et al., 2016). No AEs were reported for stimulation intensities 0.5–2 mA and tACS frequencies up to 100 Hz. Measured electric field values are in line with predictions from electric field models (Datta et al., 2009a; Miranda et al., 2013; Opitz et al., 2018). Since 2017, various labs have performed intracranial recordings in humans during tES without any reported AEs. The location of tES electrodes were different, many times over the same hemisphere/region as the implanted electrodes but not always, guided by the scientific question. Recording modalities involved sEEG (Lafon et al., 2017; Louviot et al., 2022; Shan et al., 2023; Simula et al., 2024; Wang et al., 2022), DBS leads (Chhatbar et al., 2018; Esmaeilpour et al., 2017; Ruhnau et al., 2018) and grid electrodes (Salimpour et al., 2017). Invasive recordings were also performed in an individual with brain lesions without reported AEs (Jiang et al., 2022). A recent computational study suggests that the implants could possibly enhance the tES electric field but the effects stay within a safe range (Karimi et al., 2025). However, direct electric field measurements of AC currents in a phantom next to a DBS lead showed only small effects of the implant (10 % reduction) (Lamoš et al., 2025). This is in line with another modeling study which predicts only moderate effects of implants on the EF (Mercadal et al., 2022). Local electric field perturbations can affect the stimulation dose and thus complicate the study of tES mechanisms using invasive recordings. This effect may be amplified by the lower resistance of the burr hole if the transcranial electrode is in close vicinity (Datta et al., 2010). For more details regarding the safety of various configurations and the arrangement of tES electrodes and implants we refer to Karimi et al. (2025).
Recommendation: tES appeared to be safe in subjects with implanted DBS or sEEG electrodes within stimulation parameters typically applied in healthy participants. TES on humans carrying any implants in the brain or in the skull should only be performed in well-supervised and controlled studies. Special attention is required for adaptive implanted brain stimulation which may sense the electric field artifacts by tES.
9.5. Pharmacological interventions combined with tES
The interaction between tES, particularly tDCS, and pharmacological treatments remains a topic of clinical relevance and ongoing investigation. The 2017 guidelines noted that some pharmacological agents may modulate the effects of tES, potentially enhancing or reducing its efficacy, and raised the possibility of safety concerns when combining the two interventions. Since then, several additional studies have investigated the safety, feasibility, and effectiveness of combined tES and pharmacotherapy, particularly in remote or home-based treatment settings.
Recent evidence, primarily from feasibility studies, pilot trials, and case series, supports the general safety and tolerability of tES in patients receiving stable pharmacotherapy across various clinical populations, including MDD, bipolar depression, Parkinson’s disease, multiple sclerosis, fibromyalgia, schizophrenia, and temporal lobe epilepsy. SEs and AEs reported in these studies were predominantly mild and transient, such as tingling, itching, or transient discomfort at the electrode site, with no evidence of increased risk attributable to the pharmacological status of the participants (Cappon et al., 2022; Flynn et al., 2024; Ghazi-Noori et al., 2024; Pilloni et al., 2022).
Some studies, including large-scale fully remote RCTs, have provided initial evidence of clinical benefit of at-home tES when administered alongside stable pharmacological regimens (Ruffini et al., 2024; Woodham et al., 2024). Nevertheless, most of these studies did not systematically analyze drug-tES interactions, and participants were often required to maintain stable medication regimens or exclude drugs known to lower seizure threshold (e.g., bupropion, clozapine). Moreover, no study so far has reported data about medications prescribed for unrelated conditions that may nonetheless interact with brain stimulation. For example, beta-blockers, sodium and calcium channel blockers—used in antihypertensive, antiarrhythmic and antiepileptic therapies—can diminish the modulatory effects of tDCS by stabilizing membrane excitability. Though statins, hypoglycemic agents, steroid or hormone therapies have not been studied in this context, their wide-spread use and their indirect effects on the CNS suggest they could also influence tDCS outcomes. Furthermore, drugs potentially leading to cutaneous AEs, e.g., immunosuppressants and immunomodulators, could change the tolerability profile of tES at the site of stimulation.
However, although long-term follow-up data are sparse, available findings suggest that sustained use of home-based tDCS with pharmacotherapy is feasible and generally well tolerated, provided appropriate training, supervision, and adherence to safety protocols are ensured (Dragon et al., 2024; Lee et al., 2022; Lench et al., 2024).
While the safety profile of combined tES and pharmacotherapy appears reassuring, the clinical rationale for their combination should be carefully considered. Specifically, concurrent use should be pursued primarily when a synergistic or additive therapeutic effect is anticipated. Recent evidence suggests that this may not always be the case: a large multicenter RCT found no additional benefit of adding tDCS to sertraline for MDD compared to sertraline alone (Burkhardt et al., 2023). On the other hand, potential risk associated with tES in depression is treatment-emergent mania (Brunoni et al., 2017), for which co-administration of mood stabilizers has been recommended to mitigate this risk (Tatti et al., 2022).
Conclusions and recommendations: While no new evidence has emerged that clearly indicates an increased risk or clinically significant adverse interaction between tES and pharmacotherapy, current findings are largely based on feasibility and small-scale studies, often without dedicated analysis of pharmacodynamic interactions. Therefore, until high-quality evidence becomes available, we recommend that tES be used in combination with neuropsychiatric and non-neuropsychiatric pharmacotherapy only under conditions of careful clinical supervision.
Patients should be screened for medications known to directly or indirectly act on CNS and skin, and any changes in pharmacotherapy during tES treatment should be avoided whenever possible. Moreover, clinicians should evaluate whether a combined approach is likely to confer added clinical value on a case-by-case basis. Further studies are warranted to explore synergistic effects, pharmacological enhancement of tES efficacy, and safety across diverse drug classes.
9.6. TES combined with other non-pharmacological interventions
The administration of a concurrent non-pharmaceutical intervention can alter the brain state in turn modulating the effect and response of NIBS (Bradley et al., 2022; Opitz et al., 2015). The term “Non-pharmacological intervention” includes a variety of treatments. For the sake of this review, we followed the taxonomy proposed by Ninot et al., (Castellano-Tejedor, 2022; Ninot et al., 2017), who identifies five main groups, which are addressed below (for overview see Table S5).
9.6.1. Psychological health interventions
Psychological health interventions range from prevention programs to psychotherapy interventions. Studies combined tDCS with fear conditioning paradigms, which are commonly used to model anxiety-related and broader emotional processes. Four studies used standard tDCS with different montages, all targeting the PFC (Dittert et al., 2018; Ganho-Ávila et al., 2019, 2022; Ma et al., 2024), and the inferior frontal gyrus (Ma et al., 2024) while one applied frontopolar HD-tDCS (Adams et al., 2023) to the DLPFC. All were single-session trials with young adults, conducted online or offline, using currents up to 1.5 mA. The combination was safe, with tDCS producing only mild AEs, such as itching, tingling, headache, skin redness, and tiredness. Blinding, when reported, was effective.
In a meta-analysis of 17 studies, 13 of which employed tDCS, NIBS combined with psychosocial interventions was found to be a safe approach for alleviating depressive symptoms (He et al., 2022). Commonly reported SEs included headache, scalp discomfort, skin redness, and itching, occurring in both active and sham stimulation groups. Most studies concluded that NIBS paired with psychosocial intervention is well-tolerated, with only mild and transient AEs, such as post-stimulation headache.
The safety and tolerability of NIBS, including tES combined with mindfulness-based interventions were evaluated in a recent meta-analysis by Demina et al., (2024). The analysis reported good overall tolerability, no serious AEs, and generally high levels of treatment satisfaction.
One study combined tACS (delta-beta waveform) with behavioral activation, a third-wave evidence-based cognitive behavioral therapy in patients with depression. No safety concerns were noted (Carlton et al., 2025).
9.6.2. Physical health interventions
Five studies combined tDCS with physical training interventions in healthy young adults. Thomas et al., (2021) applied tDCS during a combined aerobic and cognitive task but did not report collecting or analyzing AE data. In contrast, Jung et al. (2024), Lo et al. (2023), and (da Silva et al., 2019) implemented multi-session offline tDCS protocols (1.5 or 2 mA) during balance training, physical therapy, and time to exhaustion, respectively. All studies reported only mild AE and successful blinding. One study combined tRNS to the DLPFC and V1 with exergame training, with participants reporting no AEs (Moret et al., 2021). When reported, blinding was effective. Dumel et al. (2018) did not report collecting or analyzing AEs data, while the other studies reported only mild AEs
In clinical populations, motor training combined with RS-tDCS has also shown efficacy. In individuals with impaired ambulation due to multiple sclerosis, 10 sessions of M1-targeted tDCS (2.5 mA) paired with aerobic exercise significantly improved gait speed and walking endurance, with benefits maintained four weeks post-intervention (Pilloni et al., 2020). A more recent trial demonstrated that 20 sessions of home-based, remotely supervised M1-SO tDCS paired with manual dexterity training improved hand function and MS-related quality of life in participants with impairment due to progressive MS, many of whom were also of older age (Pilloni et al., 2024).
9.6.3. Nutritional health interventions
Span from the use of dietary supplements to structured nutritional therapy and diet-based interventions. tDCS has been explored in combination with hypocaloric diets in individuals with excess weight. Natividade et al., (2021) reported only mild AEs, such as itching, tingling, and fatigue, with no significant differences in the frequency of AEs between active and sham groups. Similarly, Heinitz et al., (2017) found a higher incidence of skin redness in the active tDCS group, while other mild AEs occurred at comparable rates across active and sham conditions
9.6.4. Digital health
Interventions, encompassing wearable and handheld eHealth devices, therapeutic games, and virtual reality therapy, have been integrated with NIBS in training paradigms in recent studies. Specifically, four studies combined tDCS with various cognitive tasks (Duffy et al., 2024; Holczer et al., 2023; Yu et al., 2020; Zhu et al., 2025), three studies combined tACS with working memory tasks (Diedrich et al., 2024a; Diedrich et al., 2024b; Krebs et al., 2021b), and two used tRNS (Brambilla et al., 2021; Chenot et al., 2022). All were conducted during stimulation, mostly using multi-session designs in young and older adults, with variable electrodes’ montages, targeting the DLPFC. tDCS studies reported mixed cognitive findings and only mild sensations, similarly in active and sham groups (Ruf et al., 2017). Interestingly, while AEs remain mild, higher intensities generally cause more noticeable sensations (Weller et al., 2020). Two analyses from the “Augmenting cognitive training in older Adults” study (Woods et al., 2018) found that 20 sessions of 2 mA tDCS over bilateral DLPFC were well tolerated, with only mild sensations (e.g., tingling, burning) and no other adverse events(Hausman et al., 2023, 2024). While more participants in the active group reported high blood pressure, this was not attributed to stimulation (Hausman et al., 2023), see also Section 7.4
In the tACS studies, two participants experienced phosphenes (Diedrich et al., 2024a; Diedrich et al., 2024b). Two tRNS studies reported only mild sensations, but one participant dropped out because of a headache in the tACS group (Brambilla et al., 2021). Notably, Sheffield et al. (2022) conducted a multi-site review on blinding efficacy and AEs in studies combining cognitive training with NIBS. They found that AEs were more strongly associated with contextual factors such as testing site and session number, rather than the type of transcranial electrical stimulation itself. Blinding was less effective for tDCS and tACS compared to tRNS, particularly in conditions producing stronger cutaneous sensations. While smaller electrodes and repeated sessions did not increase AEs, the findings underscore the need for improved blinding protocols, especially in studies using higher-intensity stimulation.
In clinical populations, RS-tDCS paired with cognitive training has shown feasibility and cognitive benefits. In MS, studies report improvements in complex attention, reduced response variability, and enhanced cognitive outcomes after 30 sessions, particularly in those with greater disability (L. Charvet et al., 2018, 2025; Charvet et al., 2023b). In Parkinson’s disease, two open-label studies demonstrated that RS-tDCS was safe, well-tolerated, and feasible for home-based delivery resulting in modest gains in processing speed, fatigue, and motor functions (Agarwal et al., 2018; Dobbs et al., 2018). Across both trials, tolerability was high, and adverse events were minimal and mild.
Two studies combined tDCS with activity performed in a virtual reality environment (Bulteau et al., 2017; Ciechanski et al., 2017), simulating height exposure or surgical resection, respectively, over one or two sessions, with a current intensity of 1 mA, reporting itching and tingling as the most common complaints, with no serious AEs (Ciechanski et al., 2017). Neri et al. (2025) combined tRNS to the sensori-motor and visual networks with five days of a VR first-person shooter training, reporting only mild AEs and no differences between the active and sham groups.
No other health interventions (eg., phytotherapy, aromatherapy, etc.,) were found combined with noninvasive brain stimulation.
Conclusions and recommendations: The combination of tES with non-pharmacological interventions appears to be safe in adults across age groups. Reported AEs were mild and consistent with standalone tES applications. No significant safety concerns emerged across the reviewed studies. Despite these reassuring findings, differences in stimulation parameters and outcome measures limit the ability to draw definitive conclusions about safety in all contexts. Care must be taken when higher intensities are used to achieve successful blinding. In particular, a few studies did not systematically report or analyze AEs, showing the need for more rigorous and standardized safety monitoring. Given the wide range of potential interventions that can be combined with NIBS, it is essential that researchers implement rigorous monitoring and transparent reporting of AEs, particularly when evaluating novel combinations.
10. Ethical, legal and regulatory aspects
10.1. Ethics (doping, neuroenhancements, Neurorights)
In navigating the ethical landscape of tES, four core principles emerge: 1) Rigorous safety, 2) Equitable access, 3) Strict oversight against misuse, and 4) Protection of neurorights, all underpinned by comprehensive informed consent.
Firstly, adherence to established parameters, proper electrode preparation, and continuous monitoring is essential to minimise mild AEs (e.g., skin tingling, headache) and eliminate harms such as burns (Bolognini et al., 2017). Secondly, tES therapies must be made available through regulated clinical channels, ensuring that socio-economic barriers do not exclude patients with a genuine therapeutic need (Voarino et al., 2016). The compliance of the health care personnel and the medical organisations to these technologies is essential in the successful implementation of the treatments (e.g. clinics buy tES devices but then lack the resources to keep them functioning). In terms of research and clinical use, the sensitivity to differences in populations in producing research data and implementing protocols to patients and participants need to be present. Equitable access, prevention of misuse, and protection of neurorights all concern wider frameworks than merely focus on the mental and physiological health of the patient. The implementation of treatment needs to acknowledge that values inherent in the technologies or values in the research data do not necessarily transfer from one society to another, from a given health care system to another and patients should not be ignored. One of the challenges for evidence based tES treatments is the ever changing (developing) devices: many treatment devices do not stay long enough to gather a good amount of evidence. Nevertheless, our aim is to comment on the tES protocols rather than on device-specific considerations.
Third, home-use systems require built-in lockouts, remote monitoring, and mandatory user training to prevent unsupervised off-label use and cognitive or sport “doping” in healthy individuals – a practice that not supported outside of approved research to ensure fair competition in academic, athletic, and professional settings (Voarino et al., 2016), although it is still unclear whether the benefit is a proven fact or a myth (Lefaucheur, 2019). Neuroenhancement using tES in healthy people is a frequently discussed issue, without official consensus (Antal et al., 2022). Finally, special consideration for vulnerable populations – particularly minors (See Section 7.3.) – requires age-appropriate consent processes and developmental risk–benefit analyses that respect emerging neurorights, including the right to mental privacy and cognitive integrity (Auvichayapat and Auvichayapat, 2022; Buchanan et al., 2022). Together, these principles ensure that tES remains a safe, patient-centred modality for therapy and rehabilitation, while minimising the risks of abuse, injustice, and exploitation.
10.2. Regulation of tES across the world
10.2.1. Europe
In the EU, in May 2021 the Regulation (EU) 2017/745 (https://eur-lex.europa.eu/eli/reg/2017/745/oj/eng), hereafter referred to as Medical Device Regulation (MDR), has superseded the previous medical device directive (MDD), and now regulates all aspects concerning “the placing on the market, making available on the market or putting into service of medical devices for human use and accessories for such devices in the Union” as well as “clinical investigations concerning such medical devices and accessories conducted in the Union.” (Art. 1(1)). Importantly, according to Art. 1(2) the MDR also applies to “groups of products without an intended medical purpose that are listed in Annex XVI”, where it explicitly lists under point 6. “Equipment intended for brain stimulation that apply electrical currents or magnetic or electro-magnetic fields that penetrate the cranium to modify neuronal activity in the brain.”, and Art. 1(3) details that “Devices with both a medical and a non-medical intended purpose shall cumulatively fulfil the requirements applicable to devices with an intended medical purpose and those applicable to devices without an intended medical purpose.”
MDR Art. 2(1) rather widely defines a ‘medical device’ as “any instrument, apparatus, appliance, software, implant, reagent, material or other article intended by the manufacturer to be used, alone or in combination, for human beings for one or more of the following specific medical purposes: (1) diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease; (2) diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability; (3) investigation, replacement or modification of the anatomy or of a physiological or pathological process or state; (4) providing information by means of in vitro examination of specimens derived from the human body, including organ, blood and tissue donations; and (5) which does not achieve its principal intended action by pharmacological, immunological or metabolic means, in or on the human body, but which may be assisted in its function by such means.”
Medical purposes relevant for tES devices are in particular the diagnosis, prevention, and treatment or alleviation of a disease (i.e., therapeutic applications), but also the investigation or modification of a physiological or pathological process or state (i.e., research applications), whereas for TMS also diagnosis and monitoring purposes are relevant (e.g., via TMS evoked MEPs, TEPs, etc). It should be noted that there is a fundamental difference between the intended medical purpose (e.g., non-invasive stimulation of the human brain) and a specific clinical indication (e.g., major depressive disorder or chronic pain).
According to Art. 51(1), “devices shall be divided into classes I, IIa, IIb and III, taking into account the intended purpose of the devices and their inherent risks” and “classification shall be carried out in accordance with Annex VIII”. Following Rule 9 of the Annex VIII, “all active therapeutic devices intended to administer or exchange energy are classified as class IIa unless their characteristics are such that they may administer energy to or exchange energy with the human body in a potentially hazardous way, taking account of the nature, the density and site of application of the energy, in which case they are classified as class IIb”. Accordingly, tES devices, which qualify as active devices and have a comparatively low risk profile as outlined in this paper, fall into class IIa if they have a therapeutic medical purpose.
Notably, Rule 10 refers to “Active devices intended for diagnosis and monitoring. […]” (not applicable to tES) and Rule 13 states that “all other active devices are classified as class I”. Unlike in the original definition in Art. 2(1), a rule explicitly referring to the medical purpose of “investigation” (as in academic or clinical research) is missing in Annex VIII, so that tES devices without a therapeutic purpose were automatically classified as class I (instead of the appropriate class IIa). Following a letter by six EU member states, based on an inaccurate risk assessment without proper review of scientific evidence and following a flawed line of argument (Antal et al., 2025), in 2022 the Regulation (EU) 2022/2347 (https://eur-lex.europa.eu/eli/reg_impl/2022/2347/oj/eng) re-classified all non-invasive brain stimulation devices without a medical purpose as class III, i.e., the highest risk category normally reserved for e.g., implantable devices, as it was reasoned that “while such equipment is not surgically invasive, the electrical currents or magnetic or electromagnetic fields do penetrate the cranium to modify neuronal activity in the brain.” Energy delivery to the body is, however, part of the definition of active therapeutic devices, which explicitly belong to class II, and, given the actual risk profile of tES devices provided in this paper, they more specifically belong to class IIa (see also the explanations provided by the Medical Device Coordination Group (MDCG) in document MDCG 2021–24, “Guidance on classification of medical devices”, (Directorate-General for Health and Food Safety, 2021). Therefore, both the evidence provided for an elevated risk of tES and TMS devices was incorrect (Antal et al., 2025) and the classification rationale was flawed. Unfortunately, this incorrect reclassification caused enormous irritations in the field and misinterpretations by regulatory bodies and regional ethic committees due to the unclear and ambiguous status of tES and TMS devices used for investigational research as medical purpose (cf. Art. 2(1)), rather than therapeutic or diagnostic applications (which are explicitly mentioned also in Rule 9 and 10 of Annex XVI). This confusion led to unjustified interruptions of research activities in many academic and clinical institutions. Researchers, clinicians, academic societies, manufacturers, and patient organizations have repeatedly expressed their concerns (Antal et al., 2024). Following a formal complaint to the EU Ombudsman and their respective recommendation in 2024 (https://www.ombudsman.europa.eu/en/decision/en/185531), the European Commission has asked the Scientific Committee on Health, Environmental and Emerging Risks (SCHEER) to issue a new evidence-based scientific opinion on the risk that is posed by using brain stimulation devices for non-medical purposes, which is expected to be published in 2026. Based on this opinion, the European Commission will hopefully re-reclassify non-invasive brain stimulation devices without medical purpose to the same class as those with a medical purpose, namely class IIa.
Since the MDR came into force in May 2021, manufacturers are required to conduct the conformity assessment procedure to obtain a CE-certification as a medical device for their product according to the MDR. However, since the notified bodies allowed to monitor the certification procedure and issue the CE-certificates simply cannot process the enormous numbers of re-certifications required, transition periods have been introduced and extended, e.g., to Dec 31, 2028 for class II medical devices that previously did not require involvement of a notified body under the MDD. The CE-certification process requires the manufacturer to explicitly state the intended medical purpose as well as one or more clinical indications and to provide clinical evidence for the safety and effectiveness of the device. At the time of submission of this consensus paper, the first tES devices have received CE-certification as a medical device under the MDR with major depressive disorder as indication, and more are likely soon to follow.
Finally, there is a lot of irritation in the field regarding when research studies using a medical device, specifically a tES device, have to be conducted as ‘clinical investigation’ under the MDR (requiring registration with the national authorities, full compliance with ISO 14155 GCP, as well as costly and time consuming documentation, monitoring, and auditing activities) and when they may be conducted as mere ‘clinical study’ outside the MDR (i.e., only requiring a positive vote by the responsible ethics committee) (Rethwilm et al., 2024). An important tipping point in the decision tree is whether the study investigates the safety and/or performance of a specific investigational device (-> clinical investigation) or merely uses an existing medical device (which could in principle be replaced by another device with comparable features) as an auxiliary tool to investigate a fundamental neuroscientific or clinical research question (-> clinical study). For the latter, the medical device does not even need to be used for the medical purpose or the clinical indication originally indicated by the manufacturer. Unfortunately, while the MDR applies in all EU countries, the specific legal implementations of the regulation at the national level differ across EU member states, and even within each state, the local editorial committees often have their own interpretation regarding this issue, currently creating very unequal challenges for tES research across labs and countries within the EU.
10.2.2. Asia
Based on our research, no country in Asia has officially approved tES devices as medical devices, except for Singapore and South Korea. In 2017, Singapore’s Health Sciences Authority granted Soterix Medical approval to market tDCS devices. These devices are approved for treating major depression and fibromyalgia (Soterix Medical Inc, 2025). Similarly, South Korea’s Ministry of Food and Drug Safety (MFDS) approved the home-based tDCS treatment device MINDD STIM+ (Ybran Inc, n.d.) in 2021 for therapeutic use (Bikson et al., 2023).
In Japan, medical devices are categorized into four classes based on risk levels by the Pharmaceuticals and Medical Devices Agency (PMDA): Class I (extremely low risk), Class II (low risk), Class III (medium risk), and Class IV (high risk) (Medical Devices Agency, n.d.). For example, TMS devices are classified as Class II. Although tES devices would likely fall into a similar category, no such devices have been approved for medical use in Japan to date. The use of tES devices for basic or scientific research must comply with the Ethical Guidelines for Medical and Biological Research Involving Human Subjects, while clinical trials are governed by separate regulations. The Japanese Society for Clinical Neurophysiology issued safety guidelines for the research use of tES in 2019, which remain a critical resource for investigators using tES devices for research purposes (https://www.jstage.jst.go.jp/article/jscn/49/2/49_109/_article/-char/ja). However, the distinction between basic research and clinical trials can be ambiguous, raising the question of “where to draw the line between medical and non-medical use” (Antal et al., 2025).
Other Asian countries employ comparable risk-based classification systems. For instance, South Korea: The Ministry of Food and Drug Safety classifies medical devices into Classes I–IV based on risk (Food and Drug Administration, 2018); India: The Central Drugs Standard Control Organization uses a classification system ranging from Class A (lowest risk) to Class D (highest risk), as outlined in the Medical Devices Rules of 2017 (Central Drugs Standard Control Organization, n.d.); China: The National Medical Products Administration categorizes medical devices into three classes (I–III) based on increasing risk, as defined in the Rules for Classification of Medical Devices (The National Medical Products Administration, n.d.).
10.2.3. Australia
In Australia, medical devices are regulated through the Therapeutics Goods Administration and the Australian Register of Therapeutic Goods. Several devices are listed on the ARTG allowing for use, typically under clinical supervision. The Sooma tDCS/Transcranial electrical stimulation system was first listed in 2016 under an intended purpose of the treatment of “unipolar depression in the adult population under the supervision of a qualified health practitioner”. The neuroCare Group tES system was listed in 2017 but without a specific detailed therapeutic claim (“tDCS may represent an effective treatment option for patients presenting with major depressive episodes”). Finally, the Soterix Medical tES system has also been listed since 2017 with a lengthy description of application including “The device is intended to treat different neurological and psychiatric disorders. In particular, tDCS increases spontaneous brain activity and metabolism in areas found to be hypoactive in patients suffering from major depressive disorder.”
10.2.4. America (USA & Canada)
In the USA, the framework comprises a system of regulations and recommendations issued by the Good Practices in Clinical Research, Code of Federal Regulations, and/or the Food and Drug Administration (FDA). The FDA defines medical devices as products that are “intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease in man… or intended to affect the structure or any function of the body….…” The FDA’s classification for medical devices are based on risk level (Bikson et al., 2023; Food and Drug Administration, 2018). Class I devices are low-risk devices, such as band-aids, which are subject to general controls, such as registration of facilities and compliance with good manufacturing practices. Class II devices are moderate risk devices that are subject to additional special controls, such as performance standards and special labeling. Class III devices are those that pose a high risk of illness or injury. Most devices that provide a low level of electrical stimulation to the body for medical purposes, such as transcutaneous electrical nerve stimulation devices and powered muscle stimulation devices, are considered Class II devices (Fregni et al., 2015). However, tES devices without an intended medical purpose may not fall under the scope of FDA regulation (Antal et al., 2022; Vasquez and Fregni, 2021; Wexler, 2015).
Medical devices not cleared or approved by the FDA in the US are required to follow the Investigational Device Exemptions regulation (21 CFR Part 812). This regulation describes three types of device studies: significant risk (SR), non-significant risk (NSR), and exempt studies. Under 21 CFR 812.3(m), a SR device study is defined as a clinical investigation using a device that is intended as an implant, is represented to be for a use in supporting human life, is for a use of substantial importance in mitigating and treating disease or presents a potential for serious risk to the health, safety or welfare of a subject. A NSR device study is one that does not meet the definition for an SR device study. Certain studies are exempt from the requirements of 21 CFR Part 812, for example: studies of an already cleared medical device in which the device is used or investigated in accordance with the indications in the cleared labeling. Clinical trials using tDCS and tACS devices in the US have been generally classified as NSR. Sponsors of investigational SR device studies are required to get an approved IDE from the FDA before starting their study. In addition, in accordance with the regulations at Part 812, the study may not start until both FDA and the Institutional Review Board of clinical setting have given their approval.
Clinical Example – NYU Langone Health tDCS Telehealth Program: To meet growing clinical demand and align with regulatory standards, NYU Langone Health launched a telehealth-based RS-tDCS clinical service in 2019. This program operates under NSR designation and IRB oversight, offering home-based tDCS as innovative care for conditions such as multiple sclerosis, Parkinson’s disease, mood disorders, and post-stroke recovery. The protocol includes medical clearance, HIPAA-compliant remote supervision via video, and individualized treatment plans specifying electrode placement, current intensity, and dosing schedules. As of 2025, over 230 patients across 26 U.S. states have participated, completing more than 16,000 sessions. The program has demonstrated high engagement, safety, and feasibility, and serves as a model for large-scale implementation of RS-tDCS within U.S. healthcare systems.
10.2.5. America (South America)
In Brazil, tDCS devices are regulated as medical equipment requiring formal registration by the National Health Surveillance Agency (Fregni et al., 2015). Only NeuroConn’s DC-Stimulator is specifically ANVISA-approved for tDCS use (ANVISA certification is analogous to a CE mark). Clinical research must be approved by local ethics committees with possible review by the national regulatory body (Fregni et al., 2015).
Argentina’s National Administration of Drugs, Foods and Medical Devices (ANMAT) oversees medical device regulation and aligns with international health standards. tDCS devices would require ANMAT clearance; none are currently registered (https://www.argentina.gob.ar/anmat/anmat-en/). All clinical studies need institutional ethics committee approval.
Chile’s Public Health Institute (ISP) manages medical device approval under national health law. tDCS equipment must comply with ISP standards and be registered. Research or clinical use of tDCS requires review by local ethics committees under Ministry of Health regulations.
Other South American nations follow similar models. For example, Colombia’s INVIMA and Peru’s DIGEMID regard tDCS equipment as medical devices, and Uruguay’s health authority takes a similar stance; in all cases trials require ethics/IRB clearance.
10.3. Where should/could tES be performed and by whom?
The tES can be applied in both clinical and non-clinical settings, serving different purposes such as treating patients, conducting clinical research, or advancing basic neuroscience by studying neurotypical participants in experimental brain research. In clinical contexts, tES may be administered in hospitals or private practices for evidence-based treatment of neurological and psychiatric conditions or as part of clinical trials involving patient populations. In non-clinical settings, tES is commonly used in research laboratories to explore brain function in neurotypical healthy participants. Additionally, tES can be delivered outside a lab/clinic i.e. in home-based environments, either as part of remotely supervised research protocols or as prescribed clinical interventions. Given the diverse applications of tES, it is important to distinguish its use along three key dimensions: 1) Participants: tES either being used in neurotypical adults, clinical populations or populations requiring special considerations (e.g. children); 2) Settings: tES either being applied in research laboratories, clinical environments (e. g., hospitals or private practices), or home-based contexts; 3) Personnel involved: tES either being administered by trained researchers, by healthcare professionals, or self-administered by participants, patients or non-professionals under or without remote supervision (for detailed consideration regarding home-based use see Section 10.4. and 10.5).
A primary consideration in determining where and by whom tES should be applied is the assessment of risk. Risk evaluation must account not only for the characteristics of the technique itself but also for individual participant factors such as potential vulnerability or the possibility of interactions with concurrent oral medications. In research contexts (e.g., academic institutions), tES is typically administered by trained research personnel from diverse disciplinary backgrounds, most commonly biology, psychology, medicine and engineering. These individuals are required to have appropriate training, in order to independently perform tES (see Section 11). Ethical approval for such research applications must be obtained from the relevant institutional review board (IRB) or local ethics committee. In contrast, clinical applications of tES are subject to national regulatory frameworks, and formal approval from competent regulatory authorities may be required, where applicable (see Section 10.2).
Conclusions and recommendations: When conducted in accordance with the safety standards and parameters outlined in previous sections, tES can be safely administered in clinical settings under the supervision of a healthcare professional or clinician, as well as in research environments by trained researchers under the oversight of a principal investigator with a medical or non-medical degree. Remotely supervised home-based tES (RS-tES) is also considered safe and offers the added benefits of increased accessibility and the potential to scale both research and clinical applications.
10.4. Remotely delivered tES – home-based use under medical prescription or in research settings (Remote-Supervised tES)
The use of tES—particularly tDCS—in home settings has expanded rapidly, driven by the need to accelerate clinical trials, deliver scalable and accessible treatment options, and reduce the burden on healthcare systems. This approach is rapidly advancing research on optimal clinical use, safety, tolerability—especially over longer timeframes—and standardized implementation guidelines.
This successful implementation of guided home-based use has been facilitated by adherence to equipment and protocols that ensure reliable stimulation. The Remotely Supervised tDCS (RS-tDCS) guidance (Charvet et al., 2015; Charvet et al., 2020) outlines 8 conditions, including staff training, to satisfy for home-based tES, which have been adopted in over 60 human trials (Charvet et al., 2018, Charvet et al., 2025; Charvet et al., 2023a; Charvet et al., 2023b; Dobbs et al., 2018; George et al., 2025; Knopman et al., 2024; Pilloni et al., 2022, 2024, 2025; Richardson et al., 2023). Adopting these guidelines, a single center provided almost 7000 sessions in adults without significant AEs (Pilloni et al., 2022) and a similar outcome was achieved by a study in pediatric populations (Simpson et al., 2022) with associated controls.
The eight RS-tDCS guidelines define minimum training requirements for researchers, clinicians, and staff, as well as criteria for equipment use and patient monitoring, to ensure the safe and consistent implementation of RS-tDCS protocols.
This section’s conclusions are specified only for RS-tES (home based under medical prescription/research setting and following indicated guidelines). Digitalized tES is the integration of home-based tES with mobile-Health technologies (Brunoni et al., 2022), and when following RS-tES is within this section scope. So called “DIY-tDCS” (Wexler and Reiner, 2018) cannot be considered RS-tES due to the lack of ongoing supervision by a trained and experienced tES practitioner and provider. RS-tES interventions enhanced by feedback loops for personalization (e. g., adjusting stimulation parameters and session frequency based on symptom change or tolerability), or those integrating AI-based or other rule-based decision-support tools (e.g., IoT-enabled systems), should still be embedded within a broader connected care framework that includes active human supervision (Cohen Kadosh et al., In press). Securing human-factors by maintaining a clinician or researcher in the loop is essential to ensure safety, contextual interpretation of data, and clinical accountability.
RS-tES rules balance rigor with the need for flexibility. For example, the requirement for televisits during tES will vary between RS-tES addressing different indications and patient groups, ranging from at every session to only during orientation and certification. In contrast to self-directed home use, RS-tES requires ongoing professional supervision. For this reason, devices for RS-tES cannot continue to operate if professional supervision is discontinued (e.g., they are remotely deactivated).
Several additional studies support the safety and feasibility of RS-tDCS. Ruffini et al. (2024) showed that 34 of 35 patients successfully completed an 8-week program of left DLPFC stimulation for depression, including 4 weeks of daily treatment, with a 73 % response rate and no AEs. Park et al. (2024) studied 19 patients for improvement in MCI after two weeks of tDCS therapy to the dorsal frontal cortex, with 3 patients experiencing minor burns that resolved over time. Ghazi-Noori et al. (2024) studied 44 persons with bipolar depression over 6 weeks with moderate success and mild AEs—mainly skin reactions such as redness (40.6 %) or burning (26.5 %).
One publication summarized six clinical trials in which home-stimulation was performed. Pilloni et al. (2022) amalgamated 308 subjects covering various disorders and 6779 RS-tDCS sessions. Participants tolerated treatment well, with only minimal mild AEs such as tingling (68 %), itching (41 %), and warmth (42 %), equally reported in active and sham conditions. Class I RCTs demonstrated rapid enrollment, high adherence, and preserved blinding. Another review by Palm et al. (2017) amalgamated 22 clinical studies, again confirming minimal AEs.
Global feasibility of RS-tDCS is also being demonstrated. Silva-Filho et al. (2022) discussed successful availability of RS-tDCS in middle-income countries, such as Brazil. Vogelmann et al. (2024) outlined a roadmap to developing at-home tDCS into a scalable tool for depression treatment.
As above, while tDCS remains under investigation status in the United States, a clinical care model for home-based tDCS has been successfully implemented at NYU Langone Health and other institutions offer both onsite and remotely accessible services under models of innovative care and outside of insurance coverage. Similarly, two studies were successfully conducted within the UK NHS using a RS-tDCS system complemented by digital psychological intervention, to treat major depressive disorder (and RCT with 174 participants (Woodham et al., 2024)) and perinatal depression (an open-label study with 25 participants (Griffiths et al., 2025)). Mild transient AEs occurred mostly in the active arm of the RCT and no serious adverse events were reported. This allows for the future of a prescription-based model for tDCS and tES treatment at home, typically paired with digital therapeutics and/or other rehabilitative exercises.
Recommendation: We recommend the adoption of the RS-tES guidelines, including provisions for staff and patient training and specialized equipment, in all clinical and research applications of RS-tES conducted under medical prescription or within research settings. This recommendation is supported by evidence from over 60 trials and thousands of RS-tDCS sessions across adult and pediatric populations with varying levels of disability, demonstrating that adherence to the eight RS-tES guidelines (detailed below) ensures safety and feasibility without compromising scalability or accessibility.
RS-TES Guidelines:
-
(1)
General and study/treatment-specific tES training and certification of staff.
-
(2)
Criteria and procedures for screening and selection of patients/participants and/or caregivers, concerning their capability to participate in RS-tES-based interventions (include cognitive, psychological, physical and contextual prerequisites).
-
(3)
Training procedures and materials for participants and/or caregivers, including instructional materials, intervention administration and assessment of training effectiveness. Minimum competencies should be assessed by the research team/healthcare professionals, confirming suitability of the user. Consent forms for RS-tES, should include i) information importance of using equipment as directed, ii) data privacy, iii) information about remote monitoring and opt-out procedures.
-
(4)
Techniques for electrode preparation and placement to minimize risk of user error, including consistent montage for reproducibility
-
(5)
Equipment and procedures for reliable control of the prescribed stimulation dosage per session and across the entire intervention protocol. This includes all aspects of tES dose (Peterchev et al., 2012) (e.g. current intensity, waveform/time)
-
(6)
Routine monitoring strategies should be in place (with a schedule appropriate to the study objects and populations) to ensure adherence to the protocol, including consistent device setup, electrode saturation and placement, stimulation parameters and predefined corrective steps in case of deviations. When software is used to automate and manage dosing, procedures for staff oversight (review and changes) must be in place.
-
(7)
Methods for detecting, recording, managing and reporting any intervention-emergent AEs during or after tES sessions. RS-tES interventions should include procedures for access to a trained clinician or researcher, either in-person or via teleconference. Protocols should include standardized troubleshooting instructions and emergency response measures to manage potential AEs or protocol deviations.
-
(8)
Criteria for when individual sessions are discontinued such as due to discomfort, technical issues, or personal request, while remaining in the intervention, and criteria for when the intervention is discontinued (for instance, missing more than 20 % of sessions (e.g., >4 of 20) or ≥ 3 consecutive sessions). All communications between patients, staff, and devices should protect guidelines for confidentiality and cybersecurity.
10.5. Safety and ethical aspects of freely available (direct-to-consumer) brain stimulation
Direct-to-consumer devices for brain stimulation are widely available for purchase online (Wexler, 2015, 2020), at least in some societies. To date, there have been three studies of users of direct-to-consumer tES devices (Jwa, 2015; Wexler, 2016; Wexler and Reiner, 2018). All found that users generally adhered to the current levels (1–2 mA) and typical length of stimulation session (20 min) used in scientific studies at the time. Home users did, however, depart from conventional protocols with regard to number of sessions: Wexler (2018) found that 40 % of respondents had self-administered over 21 sessions of tDCS. Among them, 8 % were “super-users” who reported over 100 sessions of tDCS. Approximately one-third of respondents reported skin irritation and 3 % of users reported significant skin burns (Wexler and Reiner, 2018). Thus, while to some extent, home users adhere to the current levels employed in scientific studies, they tend to experiment with the conventional duration and number of sessions. This of course raises questions of safety, effectiveness and unintended effects on the users. Even if the use is typically well tolerated during a session, this does not preclude the possibility of long term AEs arising due to frequent and/or repeated stimulation, particularly in the case of “super-users” who experience a large cumulative ‘dose’ (see also (Riggall et al., 2015)).
Aside from self-reported studies of users, there is little peer-reviewed scholarship examining the safety or effectiveness of direct-to-consumer tES devices. One study found that stimulation with the Foc.us v1 device caused subjects to perform worse on the accuracy component of a working memory task than subjects who received sham (Steenbergen et al., 2016); other studies have assessed the efficacy of Halo Sport for enhancing aspects of athletic performance (Jiang et al., 2022; Lu et al., 2021) or the Thync device (Paneri et al., 2016). However, none of these devices are currently available on the market.
One company, Flow Neuroscience, exemplifies how the lines between medical use and the freely available direct-to-consumer devices are increasingly blurred. The company markets a CE-certified tDCS device, which has shown mixed results for depression in clinical studies (Borrione et al., 2024; Woodham et al., 2024). Flow is sold directly to consumers in the EU, UK, Norway, and Hong Kong, accompanied by a smartphone app offering behavioral therapy. The device includes built-in usage restrictions: during the first three weeks, users can administer up to five sessions per week, followed by a limit of two sessions per week thereafter.
On a general level, the debate about treatment versus enhancement is still ongoing and use of these direct-to-consumer devices for enhancement have their own ethical aspects (Voarino et al., 2016). This debate also bears relevance to health care systems and there are arguments for promoting access to technologies that increase well-being regardless of whether they are for treatment or enhancement purposes (Zohny, 2014). Engineering (manufacturers) standards have been proposed that apply to all limited-output tES devices including direct-to-consumer products (LOTES), which suggest medical-grade risk management and precise labeling (Bikson et al., 2018, 2023).
Conclusions and recommendation: More data is needed regarding the safety and efficacy of direct-to-consumer brain stimulation devices marketed to the general public. In some tES “super-users,” engagement with direct-to-consumer devices may escalate to patterns characteristic of behavioral addiction. We recommend distinguishing between direct-to-consumer devices manufactured respecting medical-device standards and those with no evidence for quality control.
10.6. Blinding efficacy – placebo stimulation
Establishing adequate sham conditions and maintaining blinding integrity are critical for ensuring the validity of safety data in randomized controlled trials using tES. Inadequate blinding may not only compromise the interpretability of efficacy results, but also obscure whether reported SEs or AEs are attributable to the stimulation itself or to psychological or other nonspecific factors. Despite this importance, the evaluation of blinding success has historically been neglected in clinical research in general; only 2 % of randomized trials across medical disciplines tested blinding success as of 2001.
In tES studies, participants may perceive physical sensations such as scalp tingling or itching, potentially enabling them to correctly guess whether they are receiving active stimulation. Neuromodulation studies, including those using tES, are susceptible to this issue (Knotkova et al., 2021). A review of screened papers consisting of 35 randomized controlled trials published after the previous tES safety guideline examined whether blinding integrity was assessed. Of these, 16 studies (45.7 %) included assessments of blinding success, typically by asking participants whether they believed they received real or sham stimulation (Cherney et al., 2021; daSilva Morgan et al., 2022; Flynn et al., 2024; Gehrman et al., 2024; Lee et al., 2022; Leffa et al., 2022; Manor et al., 2018; Nikolin et al., 2020; Pereira et al., 2021; Sampaio-Junior et al., 2018; Shaw et al., 2020; Smeele et al., 2023; Valiengo et al., 2017, 2020; Wandrey et al., 2023; H. Wang et al., 2019). The remaining 19 did not report such assessment.
While most of these studies reported blinding success, notable exceptions exist. For example, Leffa et al. (2022) observed that participants in the active tDCS group guessed their condition more accurately than expected by chance. Cherney et al. (2021) reported that all four participants in the sham group mistakenly believed they received active stimulation. Nikolin et al. (2020) found statistically significant differences in participant guesses between sham and active tRNS groups, indicating inadequate blinding despite limited impact on outcome measures.
The effectiveness of a sham will depend on several factors: 1) the goals of the trial and how sham is assessed (e.g. subject’s guess on condition), 2) the equipment (electrodes) and other protocol details (e.g. participant instructions, distractors), and 3) the sham protocol itself. Given this variability, sham designs may reasonably vary across trial contexts. Typically, sham stimulation is implemented using a “fade-in, short-stimulation, fade-out” approach, in which current is ramped up and down as in active tES, but stimulation at target intensity is delivered only briefly (e.g., less than one minute) (Turi et al., 2019). Comparisons between active and sham tES are therefore relative: AEs observed under the sham condition may reflect effects of brief current delivery and/or nonspecific factors, rather than the active protocol’s specific parameters.
However, the literature indicates that, when participants are explicitly asked, physical sensations during tDCS and sham stimulation are often different, potentially leading to decrease blinding efficacy (Bjekić et al., 2024; Greinacher et al., 2019; Turi et al., 2019). This highlights the need for context-specific validation of sham protocols and transparent reporting of blinding integrity.
Conclusion: Sham protocols used in the field do not raise safety concerns per se. However, proper sham and blinding procedures are essential for evaluating the safety of tES. Building consensus toward the harmonization of sham methodology remains an important goal to support valid and ethical safety evaluations.
11. Training and certification programs: assuring safety in the future
The favorable safety profile and ease of use of tES make it appealing for clinical applications and research. Still, proper training is necessary to ensure the correct use of the device, knowing how to prevent possible AEs and how to deal with them should they arise. While clinical treatments should obviously be supervised by respective health care professionals there is no requirement for a specific occupation for persons doing the stimulation or medical supervision in a research setting if proper training has been ensured.
The proper training ensures the acquisition of theoretical and practical knowledge about tES. For example, researchers need to know the principles of tES and the physiology of its wanted and unwanted effects. Furthermore, researchers, technicians and even research participants/patients themselves, in the setting of home use, would need to know how to set up the equipment, how to place the electrodes correctly, and how to assure the prescribed dose of stimulation. After a period of instruction, individuals should be assessed to make sure they can do all the procedures correctly. It might help compliance in certain settings to have the instructor and the trainee sign the same document confirming the satisfactory completion of the training. Specialized training in tES can be formally recognized trough certification programs.
It is of note that improper protocols may be associated with subjective reports of atypical pain, so protocols may include a session termination rule based on a pain level. Though the perception of heat is not associated with a risky increase in skin temperature (Khadka et al., 2018), proper training is necessary to ensure adequate reaction. In certain circumstances, it would be appropriate to know how to deal with cognitive or emotional changes that also may occur, such as participants/patients feeling uneasy or anxious during the procedure. Persons performing tES should follow SOPs, that prescribes notification paths and procedures for reporting AEs.
12. Statement from tES manufacturers/manufacturers perspective
“THE HEALTH AND WELL-BEING OF MY PATIENT will be my first consideration” This is an excerpt of the physician’s pledge from the Declaration of Geneva (“Declaration of Geneva,” 2025), which describes the many ethical implications to which physicians are bound. These ethical standards apply to the development of devices for our field of low-intensity tES. Most of our companies work already according to ISO 13485 standard, which outlines a Quality Management System (QMS) for the medical device industry. This standard emphasizes patient safety by maintaining consistent quality throughout the entire lifecycle of medical devices and it is recognised globally by various organisations and regulatory bodies (International Organization for Standardization, 2016). We believe that a well-regulated market is essential for the sustainable growth of industry, and we welcome regulation based on principles like the EU’s Better regulation framework (Antal et al., 2024). However, any legislation must be based on real, solid scientific evidence or comprehensive data from all market participants and not on arbitrary, quickly compiled and superficial data. Everyone must be able to rely on official authorisations as a reliable indicator of the safety, efficacy and reliability of products.
Safety of patients and users is of paramount importance (for us) when developing devices for medical, non-medical and research use. This involves rigorous testing, clinical and safety trials and continuous post-marketing surveillance of devices.
Developers from our industry must ensure that devices do not pose unnecessary risks and that potential risks are clearly communicated to providers, users and patients. Therefore, we as the manufacturers of those devices must identify the hazards associated with the medical device and devices without an intended medical purpose, to estimate and evaluate the associated risks, to control these risks to the maximum technical possible, and to monitor the effectiveness of the controls. The requirements of the risk analysis are applicable to all phases of the life cycle of a medical device. The process described in the standard applies to risks associated with a medical device, such as risks related to biocompatibility, data and systems security, electricity, moving parts, radiation, and usability. A risk management process is therefore mandatory for every tES device brought to the market, ensuring safety and performance (International Organization for Standardization, 2019). Further, manufacturers of low-intensity tES devices already provide their users with instructions for safe handling of the technology, the proper labeling as well as the duration and strength of the applications as well as information on SEs and contraindications.
Our field of low-intensity tES and software for Digital Health will come closer together. The medical device industry will be undergoing a transformation in the future. With artificial intelligence, wearables, digital diagnostics, connected health platforms, and robotics, devices in our field are becoming smarter, more precise, and deeply integrated into patient care. This brings new ethical and legal challenges to the development of low-intensity tES in the future. However, these technologies will offer significant benefits, such as personalized treatments, but they also raise concerns about algorithmic bias, transparency, and accountability.
Post market surveillance is a tool to monitor safety, performance and effectiveness of any device placed on the market and to install a proactive and systematic process as required in ISO 13485 / MDR / FDA in order to derive the necessary corrective and preventive action (CAPA). Companies in our sector often have large data sets on the real-world use of NIBS devices, which could provide insights into their safety and performance to regulators monitoring them, but also to researchers and the public (International Organization for Standardization, 2016).
Recommendations: Compliance with established regulatory and quality management standards (e.g. FDA, MDR or ISO 13485) demonstrates the ability of manufacturers to design, develop, manufacture, market, install and maintain safe devices for low-intensity tES. According to international understanding, the classification of medical devices is a risk-based system that takes into account the vulnerability of the human body and the potential risks associated with the devices (Antal et al., 2024). By following the regulation and state of the art, there is no a priori difference in risk between medical, NIBS devices without an intended use including research devices. Different classifications on safety within a regulatory framework are therefore unnecessary since the risk level is the same. Technical standards on low-intensity tES are necessary to address specific aspects of safety and performance for various types of such devices (Bikson et al., 2023). Post Market Surveillance data sets from manufacturers of medical devices and devices without an intended medical purpose for our field of low-intensity tES should be included in publications on safety since they include basic information on quality and safety and maybe also on performance. Open cooperation with all stakeholders must be sought in order to find an efficient set of rules for all parties involved in the ongoing digitalization of our field. The FDA’s very pragmatic approach in the US, which already allows the local ethics committee to classify low-intensity tES studies as non-significant risk studies (Food and Drug Administration, 2018), will continue to make it easier for local scientists to expand the parameter range of low-intensity tES devices as in (Cassarà et al., 2025) or to study their interactions with other modalities when a NIBS device is classified as a non-significant risk device. In this way, preclinical, safety and efficacy data can continue to be obtained quickly and efficiently, which is essential for successful collaboration between industry and academia to enrich the overall research process.
13. Monitoring procedures for tES
13.1. Note on serious and unexpected side and adverse events
As with any research or clinical intervention, participants may experience AEs that occur during the course of the study or treatment but are unrelated to the intervention itself. This may include physical accidents, psychological distress, decreased well-being, disease onset, or any other unfortunate life events that coincide temporally with tES application. In this regard, Kumar et al. (2022), who combined 2 mA anodal tDCS daily over 5 days with swallowing exercises in post-stroke dysphagia reported that three patients did not complete the study on dysphagia due to unrelated deaths, which were appropriately documented as part of good clinical reporting. A self-described case report (Boccard-Binet and Sen, 2020) summarizes the emergence of lucid dreams, deja vu, anxiety, and panic attacks 3 weeks after a cognitive tES experiment. This person later developed neurological symptoms and was diagnosed with epilepsy, but no link between symptoms and tES has been established. Regardless of the severity, the procedures for monitoring and reporting of AEs must be rigorously implemented. On this note, Vogelmann et al. (2025) reported the premature termination of a clinical trial of home-use tES in depression treatment due to an accumulation of AEs. Several patients experienced skin lesions and safety monitoring was insufficient to detect or prevent AEs. Skin burns can occur if electrodes are improperly designed or employed and so are not an expected AE in conventionally designed tES interventions (Woods et al., 2016).
In conclusion, while serious or unexpected AEs can occur in any clinical or research context, particularly among older or medically complex populations, there is no evidence to suggest these are caused by tES itself. Still, even if unrelated, these rare incidents should be reported in line with good research and clinical practice and following national and institutional regulatory standards.
13.2. Participant/Patient screening
Prior to including a participant in a study or a patient for treatment using tES, investigators or clinicians should assess the exclusion criteria using a standard questionnaire. Consensus was reached for the following self-report questionnaire (Table 3). The current screening questionnaire is an updated version of the one proposed in Antal et al. (2017) and is recommended for use, when applying tES in research and clinical settings. An affirmative answer to the first question (i.e., skin) constitutes an absolute contraindication to tES and the rest of the questions listed refer to relative exclusion criteria depending on the study population and the objectives or the clinical goals (see Section 8). In line with the Declaration of Geneva, it is the responsibility of the principal investigator or the attending physician to assess each participant/patient’s health and well-being as their primary concern.
Table 3. Screening questionnaire for transcranial electrical stimulation.
| Screening Questions | Yes | No | |
|---|---|---|---|
| Absolute exclusion criterion | |||
| 1 | Do you have open wound(s), preexisting skin disease (e.g., nevi, angiomas or previous burns/scars, neurodermitis, etc.), or active infection at [specify the stimulation sites e.g. on the head]? If yes, please specify the issue and location: |
||
| Note: Assess if it can lead to increased risk of skin irritation, damage, or secondary infection associated with tES and/or the use of conductive gel/paste/solution; if yes – exclude. | |||
| Relative exclusion criteria | |||
| 1 | Do you have any metal or electronic implants (e.g. splinters, fragments, clips, cochlear implants, deep brain stimulation, metallic tattoo, etc.) in the brain/skull. If yes, please specify the type of metal and the location. | ||
| 2 | Note: Tooth fillings and braces are no exclusion criteria Do you have metal or any electronic device at other sites in the head, neck, or chest area, such as a cardiac pacemaker or traumatic metallic residual fragments? If yes, please specify the location or device | ||
| 3 | Have you ever had unexplained or repeated loss of consciousness? If yes, please describe: | ||
| 4 | Have you ever had seizures and/or seizure activity in the past year? If yes, please describe: | ||
| 5 | Have you ever had a head trauma followed by impairment of consciousness? | ||
| 6 | Do you have a history of migraines or frequent headaches? | ||
| 7 | Do you have known neurological or psychiatric disorders (including alcohol, medication, and drug dependence or abuse). If yes, please specify: | ||
| 8 | Are you currently pregnant or is there a chance that you might be? | ||
| 9 | Are you breastfeeding? | ||
| 10 | Do you have a history of surgical procedures involving your head or spinal cord? If yes, please specify the locations: | ||
| 11 | Did you have major surgical procedures in the past 2 months. If yes, please specify: | ||
| 12 | Do you take medication? If yes, please specify: | ||
| 13 | Did you have low tolerability of previous tES applications? If you have never had tES before answer NA | ||
The questions 2 through 13 are proposed as a question pool that can be expanded or reduced depending on the specific context and purpose of the tES application. An affirmative answer to one or more of these questions does not constitute an absolute contraindication to tES. In line with the Declaration of Geneva, it is the responsibility of the Principal Investigator or the attending physician to assess each participant’s health and well-being as their primary concern.
13.3. Side effect and adverse event reporting
One of the crucial measures in ensuring safety and tolerability of the stimulation protocols in a study using tES is the active query of SEs and AEs [refer to Section 2.1 for further details on understanding SEs and AEs]. Standardized questionnaires implementing numerical scales to assess and rate the severity of AEs have been previously recommended (Antal et al., 2017; Brunoni et al., 2011; Fertonani and Miniussi, 2017; Poreisz et al., 2007). However, the internationally adopted standardized approach for SE/AE recording and reporting is yet to be established.
It is important to address the fact that in the vast majority of studies, AEs are assessed after stimulation only, and no baseline assessment of potential AEs as already existing symptoms is done. This is why, when only “after-tES” assessments are done, many times it is difficult to decide whether an AE is related to tES or simply this symptom (e.g., tiredness) has already existed before the session. This could be one of the explanations for the discrepancies in the AE frequencies reported across studies.
Recommendation: Given the reported SE/AE, researchers and clinicians are strongly advised to monitor and evaluate the following symptoms before, during and after tES application: skin sensations (tingling, itching), headache/scalp pain, pressure under electrode, fatigue/sleepiness/trouble concentrating, warming sensation/heat, negative mood, neck pain/stiffness, nausea, lightheadedness, flickering lights, vibration perception under the electrodes. Furthermore, it is advisable to monitor and record not only potentially unpleasant but also possible positive SEs such as alertness, calmness, positive mood, sense of wellbeing, increased concentration. The protocols should be established so that if negative SEs or AEs persist beyond the immediate post-stimulation period, participants can notify the investigator, who will then assess the severity and decide whether to continue the stimulation or take another appropriate action.
13.4. Acceptability of tES
Studying patient’s acceptability of an intervention is important in health innovation, as higher acceptability by patients and health professionals increases the adherence to prescribed recommendations, including safety procedures by patients/careers and clinicians/prescribers/technical assistants, consequently improving overall results (Borrelli et al., 2005; Hommel et al., 2013; Peters et al., 2013; Proctor et al., 2009). Importantly, acceptability assessments in tES need to go beyond drop-out rates and limitations to accessibility. The Theoretical Framework of Acceptability (Sekhon et al., 2017) offers a broader definition of acceptability that include seven components: the affective attitude, burden, perceived effectiveness, ethicality, intervention coherence, opportunity costs and self-efficacy.
Moreover, as tES in health innovation moves increasingly toward digital health, with growing reliance on remote clinical oversight with AI-based systems for diagnosis and personalization of treatment, new challenges emerge that should be integrated in acceptability measures. Key factors related with hybrid and digital interventions are outlined in the Digital Therapeutic Alliance framework (Malouin-Lachance et al., 2025) and include empathy and trust, personalization, therapeutic agreement, treatment goals, and ethical concerns.
Recommendation: Future clinical studies should include well-grounded and comprehensive assessments of acceptability. Theoretical models such as the Theoretical Framework of Acceptability and the Digital Therapeutic Alliance framework (for studies involving digital components) provide reliable foundations for developing such measures.
14. Summary and final conclusions
In conclusion, this review provides an updated and comprehensive synthesis of the safety evidence for low-intensity tES, based on studies published between 2017 and 2024. It integrates data from animal studies, computational modeling, and human trials in both healthy and clinical populations, including vulnerable groups (children, older adults, pregnant women) and special stimulation conditions (e.g., home-based use). Overall, common mild AEs associated with low-intensity tES include skin redness (erythema) and paraesthesias (e.g., tingling, itching, burning), whilst other symptoms reported in the literature (e.g. skin sensations, headache, fatigue, etc.) are typically same as those seen with sham protocols. Before stimulation starts, carefully inspecting the skin for diseases or focal lesions in the area of stimulation, is advised.
Current evidence supports the continued classification of tES as a low-risk intervention when used within established safety parameters. As such tES can be used in research and clinical settings by trained non-medical and medical professionals. Still, the current safety guidelines highlight several domains where the available evidence remains inconclusive and additional systematic data are required. First, the interaction of tES with pharmacotherapy across different clinical populations lacks robust, large-scale clinical trials; existing studies are typically small and feasibility-oriented, precluding definitive conclusions regarding pharmacodynamic interactions or synergistic effects. Evidence in specific populations, such as pregnant and postpartum women and children, also remains insufficient. Furthermore, the longterm and cumulative effects of repeated or multiple daily stimulation sessions need more systematic data. Additional uncertainty concerns the safety of emerging digital, AI-assisted, and home-based stimulation systems, for which real-world post-market surveillance data remain scarce. Finally, the absence of universally standardized frameworks for AE monitoring and reporting across studies limits comparability and meta-analytic synthesis. Collectively, these gaps delineate priority areas where coordinated, large-scale, and methodologically harmonized research is essential to inform future guideline updates.
As the field expands and its applications diversify, a key challenge is to maintain an appropriate balance between regulatory oversight and scientific advancement. Regulatory frameworks must ensure human protection and minimize the risk of misuse, while simultaneously enabling and facilitating basic and translational research. Overly stringent or misapplied regulations, particularly in non-clinical research settings involving healthy volunteers, may impede methodological innovation and hinder the development of the field, including the advancement of therapeutic applications of tES. In contrast, inadequate oversight and regulation may lead to misuse and expose people, particularly those with mental health problems, to physical, psychological, and financial harm. Accordingly, this review highlights the need for clear, nuanced guidelines that distinguish between clinical and non-clinical use, and take into account contextual factors such as setting, population characteristics and stimulation protocols. In addition, the implementation of standardized SE and AE monitoring and reporting practices, harmonized regulatory procedures, and structured training requirements are essential to ensure the responsible, safe, and effective use of tES in both research and clinical settings.
Supplementary Material
Appendix A. Supplementary data
Supplementary data to this article can be found online at https://doi.org/10.1016/j.clinph.2025.2111436.
Acknowledgements
This guideline was financially supported by the German Research Society (Project number: 561490349) and by the International Federation of Clinical Neurophysiology. We thank Yasmin Lopes Vieira for administrative support during the workshop and preparation of this manuscript.
AA is supported by the German Research Foundation (DFG AN 687/9-1), EU-Horizon 2020 (PAINLESS, No. 101057367), BMBF (NEURO VR-Plus 16SV9248), DAAD-PPP (57702891).
AF is supported by the German Research Foundation (ResearchU-nit5429/1 [467143400]; CRC1315-B03 [327654276]; FL379/22-1; FL379/26-1; FL379/34-1; FL379/35-1; FL379/37-1).
AGA is supported by the Portuguese Foundation for Science and Technology (FCT) Grants 2020.02059. CEECIND (Doi: 10. 54499/2020.02059.CEECIND/CP1609/CT0015) and the research center CIN-EICC is funded by national funds through FCT – Fundação para a Ciência e a Tecnologia, I.P., under the project/support UID/00730 - Center for Research in Neuropsychology and Cognitive Behavioral Intervention – CINEICC.
AP acknowledges the support provided by Matteo Guidetti in preparation of the manuscript.
AT was supported by Innovation Fund Denmark (grant 9068-00025B), the Lundbeck Foundation (grant R313-2019-622), the National Institute of Health (grant R01MH128422), and by the German Research Foundation (project grants: TH 1330/6-1, TH 1330/7-1, part of Research Unit FOR 5429 “MeMoSLAP”).
CSH was funded by the German Research Foundation (DFG) under Germany’s Excellence Strategy (EXC 2177/1 - project ID 390895286) and the research training group on Neuromodulation of Motor and Cognitive Function in Brain Health and Disease (RTG 2783), as well as the Bundesministerium fur Bildung and Forschung (BMBF, Germany 13GW0665D).
HRS has received funding as principal investigator for the project “Precision Brain-Circuit Therapy - Precision-BCT” from Innovation Funds Denmark (grant nr. 9068-00025B) and the project “ADAptive and Precise Targeting of cortex-basal ganglia circuits in Parkinsońs Disease - ADAPT-PD” from Lundbeckfonden (collaborative project grant, grant nr. R336-2020-1035).
JB receives institutional support from the Ministry of Science, Technological Development and Innovation of the Republic of Serbia (451-03-136/2025-03/200015).
JB, SRF, HRS, CM and WP are supported by EU-funded HORIZON Collaboration and Support Action TWINNIBS (101059369).
JOS is a Sir Henry Dale Fellow funded by the Royal Society and the Wellcome Trust (215451/Z/19/Z).
LBR receives funds from FWO-Flanders (12AZO24N).
LD is supported by the German Federal Ministry of Research, Technology and Space (BMFTR) (project: NeuroVR-Plus; 16SV9248).
MAN is supported by the Horizon 2020 research and innovation program of the European Union under grant agreement No 101017716 (Neurotwin); Deutsche Forschungsgemeinschaft (German Research Foundation) (DFG) Research Unit 5429/1 (467143400), NI 683/17-1; DFG Coordinated Research Center, project number 316803389 - SFB 1280, subproject A06, DFG NI683-13-1, and by the German Center for Mental Health (DZPG), Center Bochum.
MAV received funding from FWO-Flanders for research projects for fundamental research (Grant Numbers: G044222N; G044019N).
MH received support from the NINDS intramural program.
PBF is supported by a National Health and Medical Research Council of Australia Investigator grant (1193596).
PR is supported by EU-Horizon 2020 (PAINLESS, No. 101057367).
RF was supported by ‘Aldo Ravelli Research Center’ University of Milan.
RHH receives grants from National Institutes of Health (R01AG075111) and Chan Zuckerberg Initiative (2022-309349) and receives research support from the Allison and Jed Hart Fund for Cognitive Neuroscience.
SU is supported by the University of Oulu and the Research Council of Finland Profi 7 352788.
UZ receives grants from the European Research Council (ERC Synergy Grant Agreement No 810377), German Ministry of Education and Research (BMBF, 01KG2125), German Research Foundation (DFG, ZI 542/13-1), Else-Kröner Fresenius Foundation (2022_EKTP22).
WP received support from JPND (REMOPD) and Horizon EUROPE (TWINNIBS).
Abbreviations
- AC
alternating current
- AD
Alzheimer’s Disease
- AE
adverse event
- ALS
amyotrophic lateral sclerosis
- ANS
autonomic nervous system
- CES
cranial electrical stimulation
- CNS
central nervous system
- CTCAE
Common Terminology Criteria for Adverse Events
- DBS
deep brain stimulation
- DC
direct current
- DLPFC
dorsolateral prefrontal cortex
- ECT
electroconvulsive therapy
- EEG / sEEG
electroencephalography / stereo EEG
- EF
electric field
- ESBS
European Society for Brain Stimulation
- HIV
human immunodeficiency virus
- ICH
International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use
- IFCN
International Federation of Clinical Neurophysiology
- IRB
institutional review board
- LOTES
limited-output tES
- MCI
mild cognitive impairment
- MDD
major depressive disorder
- MDR
Medical Device Regulation
- MEG
magnetoencephalography
- MR / MRI / fMRI
magnetic resonance / MR imaging / functional MRI
- MS
multiple sclerosis
- NIBS
noninvasive brain stimulation
- OCD
obsessive compulsive disorder
- otDCS
oscillatory transcranial direct current stimulation
- ONS
optic nerve stimulation
- PD
Parkinson’s disease
- PFC
prefrontal cortex
- PMDA
Pharmaceuticals and Medical Devices Agency
- PTSD
posttraumatic stress disorder
- RCT
randomized control trial
- RS-tES /RS-tDCS
remotely supervised tES/tDCS
- rtACS
repetitive transorbital alternating current stimulation
- SAE
serious adverse event
- SE
side effect
- SSRI
selective serotonin reuptake inhibitors
- TBS
theta burst stimulation
- tACS
transcranial alternating current stimulation
- tDCS
transcranial direct current stimulation
- tES
transcranial electrical stimulation
- tsDCS
transcutaneous spinal direct current stimulation
- tTIS
transcranial temporal interference stimulation
- tRNS
transcranial random noise stimulation
- TMS /rTMS
transcranial magnetic stimulation / repetitive TMS
- VAS
visual-analogue scale
Footnotes
Multi-channel is a misnomer referring to the use of three or more channels.
Declaration of competing interest
The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: AA is the vice president of the European Society for Brain Stimulation, Member-at-Large at the EMEAC – IFCN, she serves as a paid consultant at NeuroConn, Ilmenau, and she is a paid advisor at Electromedical Products International (Pulvinar), USA. Member of the advisory board at PlatoScience, has non-financial support from Sooma Medical, she is Paid Editor by Elsevier. AGA has nonfinancial support for research from Flow Neuroscience and Sooma Medical. ATS is Chief Scientific Advisor of PlatoScience, Scientfic Advisor of Alphasys, Funder and Dierector of the Academy of Brain Stimulation and the AoBS Foundation as well as the International Clinical TMS Certification Course, receiving equpment support from MagVenture, MagStim, Deymed, Yingchi, Brainsway. CSH holds a patent on transcranial brain stimulation. FF is a paid consultant and shareholder of Electromedical Products International (EPI) and is the inventor of non-invasive brain stimulation technology for which he receives royalty payments from UNC. FP is a member of the European Scientific Advisory Board of BrainsWay Inc. (Jerusalem, Israel), and the International Scientific Advisory Board of Sooma (Helsinki, Finland); he has received speaker honoraria from Mag&More GmbH, and the neuroCare Group (Munich, Germany); his lab has received support with equipment from neuroConn GmbH (Ilmenau, Germany), Mag&More GmbH, and BrainsWay Inc. HRS has received honoraria as speaker and consultant from Lundbeck AS, Denmark, and as editor (Neuroimage Clinical) from Elsevier Publishers, Amsterdam, The Netherlands. He has received royalties as book edi ni mi tor from Springer Publishers, Stuttgart, Germany, Oxford University Press, Oxford, UK, and from Gyldendal Publishers, Copenhagen, Denmark. KS is an employee and shareholder of Neurocare group AG. LC has received speaker fees from Soterix Medical and serves as a consultant for Syntropic Medical and Ceragem. LD has received nonfinancial support for research from NeuroCare. He has received honorarium from NeuroCare for a training course. MAN is a member of the Scientific Advisory Boards of Neuroelectrics and Precisis. MSG serves as a consultant to Abbott (DBS), Livanova (VNS) and Neuralief (trigeminal). He has ongoing grants from Livanova and Abbott. In the last 3 years PBF has received equipment for research from Neurosoft and Nexstim. He has served on a scientific advisory board for Magstim and received speaker fees from Otsuka. He has also acted as a founder and board member for TMS Clinics Australia and Resonance Therapeutics. PR has received nonfinancial support for research from Sooma Medical™, NeuroCare, and QuantalX Neuroscience. He has received honorarium from NeuroCare for a training course. RCK serves on the scientific advisory boards of Neuroelectrics Inc., Innosphere Inc., is the founder and shareholder of Cognite Neurotechnology Ltd. RF and AP are shareholder of Newronika SpA, Italy. SA is named inventor on a patent application (WO/2022/106850) jointly submitted by the University of Birmingham (UK), and Dalhousie University (CA), titled “Improving cognitive function”. SR is member of the advisory board of Sinaptica therapeutics. UZ serves on the scientific advisory board of CorTec GmbH (Freiburg, Germany). Other authors have no conflict of interest to declare.
Annexes
Literature search syntax.
PubMed.
syntax: ((((((safety[Title/Abstract]) OR (side effect*[Title/Abstract])) OR (adverse effect*[Title/Abstract])) OR (tolerability[Title/Abstract])) AND ((((((((tES[Title/Abstract]) OR (transcranial electric* stimulation[Title/Abstract])) OR (tDCS[Title/Abstract])) OR (transcranial direct current stimulation[Title/Abstract])) OR (tACS[Title/Abstract])) OR (transcranial alternating current stimulation[Title/Abstract])) OR (transcranial random noise stimulation[Title/Abstract])) OR (tRNS[Title/Abstract]))) AND (2017/1/1:3000/12/12 [pdat]))
Examples of topic specific searches:
Ageing:
The following search string was used: ((((((safety[Title/Abstract]) OR (side effect*[Title/Abstract])) OR (adverse effect*[Title/Abstract])) OR (tolerability[Title/Abstract])) AND ((((((((tES[Title/Abstract]) OR (transcranial electric* stimulation[Title/Abstract])) OR (tDCS[Title/Abstract])) OR (transcranial direct current stimulation[Title/Abstract])) OR (tACS[Title/Abstract])) OR (transcranial alternating current stimulation[Title/Abstract])) OR (transcranial random noise stimulation [Title/Abstract])) OR (tRNS[Title/Abstract]))) AND (2017/1/1:3000/12/12[pdat]) AND ((Healthy aging) OR (healthy ageing) OR (older adults) OR (elderly))
Combination with nonpharmacological treatments:
(((((safety[Title/Abstract]) OR (side effect*[Title/Abstract])) OR (adverse effect*[Title/Abstract])) OR (tolerability[Title/Abstract])) AND ((((((((tES[Title/Abstract]) OR (transcranial electric* stimulation [Title/Abstract])) OR (tDCS[Title/Abstract])) OR (transcranial direct current stimulation[Title/Abstract])) OR (tACS[Title/Abstract])) OR (transcranial alternating current stimulation[Title/Abstract])) OR (transcranial random noise stimulation[Title/Abstract])) OR (tRNS [Title/Abstract]))) AND (2017/1/1:3000/12/12[pdat]) AND ((training) or (exercise) or (diet) OR (repetition) OR (practice) OR (neuroenhancement) OR (neuro-enhancement) OR (enhancement) OR (combination)).
References
- Abellaneda-Pérez K, Vaqué-Alcázar L, Perellón-Alfonso R, Solé-Padullés C, Bargalló N, Salvador R, Ruffini G, Nitsche MA, Pascual-Leone A, Bartrés-Faz D. Multifocal transcranial direct current stimulation modulates resting-state functional connectivity in older adults depending on the induced current density. Front Aging Neurosci. 2021;13 doi: 10.3389/fnagi.2021.725013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Aberra AS, Wang R, Grill WM, Peterchev AV. Multi-scale model of axonal and dendritic polarization by transcranial direct current stimulation in realistic head geometry. Brain Stimul. 2023;16(6):1776–1791. doi: 10.1016/j.brs.2023.11.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Adams TG, Kelmendi B, George JR, Forte J, Hubert TJJ, Wild H, Rippey CS, Pittenger C. Frontopolar multifocal transcranial direct current stimulation reduces conditioned fear reactivity during extinction training: A pilot randomized controlled trial. Neurobiol Learn Mem. 2023;205:107825. doi: 10.1016/j.nlm.2023.107825. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Adhia DB, Mani R, Mathew J, O’Leary F, Smith M, Vanneste S, De Ridder D. Exploring electroencephalographic infraslow neurofeedback treatment for chronic low back pain: A double-blinded safety and feasibility randomized placebo-controlled trial. Sci Rep. 2023;13(1):1177. doi: 10.1038/s41598-023-28344-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Aftanas LI, Zhanaeva SY, Al’perina EL, Idova GV, Gevorgyan MM, Dzemidovich SS, Kulikova KI. Effect of Repetitive Transcranial Magnetic Stimulation on Serum Levels of Steroid Adrenal Hormones in Parkinson’s Disease: Sex Differences. Bull Exp Biol Med. 2022;173(3):322–325. doi: 10.1007/s10517-022-05542-y. [DOI] [PubMed] [Google Scholar]
- Agarwal S, Pawlak N, Cucca A, Sharma K, Dobbs B, Shaw M, Charvet L, Biagioni M. Remotely-supervised transcranial direct current stimulation paired with cognitive training in Parkinson’s disease: An open-label study. J Clin Neurosci. 2018;57:51–57. doi: 10.1016/j.jocn.2018.08.037. [DOI] [PubMed] [Google Scholar]
- Ahmadizadeh MJ, Rezaei M, Fitzgerald PB. Transcranial direct current stimulation (tDCS) for post-traumatic stress disorder (PTSD): A randomized, double-blinded, controlled trial. Brain Res Bull. 2019;153:273–278. doi: 10.1016/j.brainresbull.2019.09.011. [DOI] [PubMed] [Google Scholar]
- Aksu S, Uslu A, İşçen P, Tülay EE, Barham H, Soyata AZ, Demirtas-Tatlidede A, Yıldız GB, Bilgiç B, Hanağası H, Woods AJ, et al. Does transcranial direct current stimulation enhance cognitive performance in Parkinson’s disease mild cognitive impairment? An event-related potentials and neuropsychological assessment study. Neurol Sci. 2022;43(6):4029–4044. doi: 10.1007/s10072-022-06020-z. [DOI] [PubMed] [Google Scholar]
- Al Qasem W, Abubaker M, Kvašňák E. Working Memory and Transcranial-Alternating Current Stimulation—State of the Art: Findings, Missing, and Challenges. Front Psychol. 2022;13 doi: 10.3389/fpsyg.2022.822545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Alam M, Truong DQ, Khadka N, Bikson M. Spatial and polarity precision of concentric high-definition transcranial direct current stimulation (HD-tDCS) Phys Med Biol. 2016;61(12):4506. doi: 10.1088/0031-9155/61/12/4506. [DOI] [PubMed] [Google Scholar]
- Alashram AR. The efficacy of transcranial random noise stimulation in treating tinnitus: A systematic review. European Archives of Oto-Rhino-Laryngology: Official Journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS): Affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery. 2024;281(12):6239–6252. doi: 10.1007/s00405-024-08858-9. [DOI] [PubMed] [Google Scholar]
- Alekseichuk I, Falchier AY, Linn G, Xu T, Milham MP, Schroeder CE, Opitz A. Electric field dynamics in the brain during multi-electrode transcranial electric stimulation. Nat Commun. 2019;10(1):2573. doi: 10.1038/s41467-019-10581-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Alexander ML, Alagapan S, Lugo CE, Mellin JM, Lustenberger C, Rubinow DR, Fröhlich F. Double-blind, randomized pilot clinical trial targeting alpha oscillations with transcranial alternating current stimulation (tACS) for the treatment of major depressive disorder (MDD) Transl Psychiatry. 2019;9(1):1–12. doi: 10.1038/s41398-019-0439-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Allaert J, De Raedt R, Sanchez-Lopez A, Baeken C, Vanderhasselt M-A. Mind the social feedback: Effects of tDCS applied to the left DLPFC on psychophysiological responses during the anticipation and reception of social evaluations. Soc Cogn Affect Neurosci. 2022;17(1):131–141. doi: 10.1093/scan/nsaa066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Alonso F, Mercadal B, Salvador R, Ruffini G, Bartolomei F, Wendling F, Modolo J. Biophysical modeling of the electric field magnitude and distribution induced by electrical stimulation with intracerebral electrodes. Biomed Phys Eng Express. 2023;9(4):045022. doi: 10.1088/2057-1976/acd385. [DOI] [PubMed] [Google Scholar]
- Ambrus GG, Antal A, Paulus W. Comparing cutaneous perception induced by electrical stimulation using rectangular and round shaped electrodes. Clin Neurophysiol. 2011;122(4):803–807. doi: 10.1016/j.clinph.2010.08.023. [DOI] [PubMed] [Google Scholar]
- Ambrus GG, Paulus W, Antal A. Cutaneous perception thresholds of electrical stimulation methods: comparison of tDCS and tRNS. Clin Neurophysiol. 2010;121(11):1908–1914. doi: 10.1016/j.clinph.2010.04.020. [DOI] [PubMed] [Google Scholar]
- Andrade SM, Cecília de Araújo Silvestre M, de França Tenório EÉ, Bezerra Sales Queiroz MH, de Jesus Santana K, Lima Holmes Madruga ML, Torres Teixeira Mendes CK, Araújo de Oliveira E, Bezerra JF, Barreto RG, Alves Fernandes da Silva SM, et al. Efficacy and safety of HD-tDCS and respiratory rehabilitation for critically ill patients with COVID-19 The HD-RECOVERY randomized clinical trial. Brain Stimulation. 2022;15(3):780–788. doi: 10.1016/j.brs.2022.05.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Angius L, Mauger AR, Hopker J, Pascual-Leone A, Santarnecchi E, Marcora SM. Bilateral extracephalic transcranial direct current stimulation improves endurance performance in healthy individuals. Brain Stimul. 2018;11(1):108–117. doi: 10.1016/j.brs.2017.09.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Antal A, Alekseichuk I, Bikson M, Brockmöller J, Brunoni AR, Chen R, Cohen LG, Dowthwaite G, Ellrich J, Floël A, Fregni F, et al. Low intensity transcranial electric stimulation: Safety, ethical, legal regulatory and application guidelines. Clin Neurophysiol. 2017;128(9):1774–1809. doi: 10.1016/j.clinph.2017.06.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Antal A, Boros K, Poreisz C, Chaieb L, Terney D, Paulus W. Comparatively weak after-effects of transcranial alternating current stimulation (tACS) on cortical excitability in humans. Brain Stimul. 2008;1(2):97–105. doi: 10.1016/j.brs.2007.10.001. [DOI] [PubMed] [Google Scholar]
- Antal A, Ganho-Ávila A, Assecondi S, Barbour T, Bjekić J, Blumberger DM, Bolognini N, Brunelin J, Chanes L, Dale M, Dubbioso R, et al. The consequences of the new European reclassification of non-invasive brain stimulation devices and the medical device regulations pose an existential threat to research and treatment: An invited opinion paper. Clin Neurophysiol. 2024;163:280–291. doi: 10.1016/j.clinph.2024.03.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Antal A, Kincses TZ, Nitsche MA, Bartfai O, Paulus W. Excitability changes induced in the human primary visual cortex by transcranial direct current stimulation: direct electrophysiological evidence. Invest Ophthalmol Vis Sci. 2004;45(2):702–707. doi: 10.1167/iovs.03-0688. [DOI] [PubMed] [Google Scholar]
- Antal A, Luber B, Brem A-K, Bikson M, Brunoni AR, Cohen Kadosh R, Dubljević V, Fecteau S, Ferreri F, Floël A, Hallett M, et al. Non-invasive brain stimulation and neuroenhancement. Clin Neurophysiol Pract. 2022;7:146–165. doi: 10.1016/j.cnp.2022.05.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Antal A, Sack AT, Bergmann TO, Bjekic J, Filipovic SR, Ganho-Ávila A, Moliadze V, Nitsche MA, Rossi S, Schuhmann T, Siebner H, et al. Note of concern regarding the sources of scientific evidence used to justify the Reclassification of Non-Invasive Brain Stimulation (NIBS) Devices without an intended medical purpose into Class III. Brain Stimul. 2025;18(2):103–108. doi: 10.1016/j.brs.2025.01.012. [DOI] [PubMed] [Google Scholar]
- Antal A, Terney D, Poreisz C, Paulus W. Towards unravelling task-related modulations of neuroplastic changes induced in the human motor cortex. Eur J Neurosci. 2007;26(9):2687–2691. doi: 10.1111/j.1460-9568.2007.05896.x. [DOI] [PubMed] [Google Scholar]
- Antonenko D, Fromm AE, Thams F, Kuzmina A, Backhaus M, Knochenhauer E, Li S-C, Grittner U, Floël A. Cognitive training and brain stimulation in patients with cognitive impairment: A randomized controlled trial. Alzheimer’s Res Ther. 2024;16(1):6. doi: 10.1186/s13195-024-01381-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Antonenko D, Thams F, Grittner U, Uhrich J, Glöckner F, Li SC, Floël A. Randomized trial of cognitive training and brain stimulation in non-demented older adults. Alzheimer’s Dementia: Transl Res Clin Intervent. 2022;8(1):e12262. doi: 10.1002/trc2.12262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Arnao V, Riolo M, Carduccio F, Tuttolomondo A, D’Amelio M, Brighina F, Gangitano M, Salemi G, Ragonese P, Aridon P. Effects of transcranial random noise stimulation combined with Graded Repetitive Arm Supplementary Program (GRASP) on motor rehabilitation of the upper limb in sub-acute ischemic stroke patients: A randomized pilot study. Journal of Neural Transmission (Vienna, Austria: 1996) 2019;126(12):1701–1706. doi: 10.1007/s00702-019-02087-9. [DOI] [PubMed] [Google Scholar]
- Ashrafi A, Mohseni-Bandpei MA, Seydi M. The effect of tDCS on the fatigue in patients with multiple sclerosis: A systematic review of randomized controlled clinical trials. J Clin Neurosci. 2020;78:277–283. doi: 10.1016/j.jocn.2020.04.106. [DOI] [PubMed] [Google Scholar]
- Assecondi S, Hu R, Kroeker J, Eskes G, Shapiro K. Older adults with lower working memory capacity benefit from transcranial direct current stimulation when combined with working memory training: A preliminary study. Front Aging Neurosci. 2022;14:1009262. doi: 10.3389/fnagi.2022.1009262. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Auvichayapat N, Auvichayapat P. Transcranial direct current stimulation in treatment of child neuropsychiatric disorders: Ethical considerations. Front Hum Neurosci. 2022;16 doi: 10.3389/fnhum.2022.842013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Babaeeghazvini P, Rueda-Delgado LM, Gooijers J, Swinnen SP, Daffertshofer A. Brain structural and functional connectivity: A review of combined works of diffusion magnetic resonance imaging and electro-encephalography. Front Hum Neurosci. 2021;15:721206. doi: 10.3389/fnhum.2021.721206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bahr-Hosseini M, Bikson M. Neurovascular-modulation: A review of primary vascular responses to transcranial electrical stimulation as a mechanism of action. Brain Stimul. 2021;14(4):837–847. doi: 10.1016/j.brs.2021.04.015. [DOI] [PubMed] [Google Scholar]
- Bartolomei F, Nica A, Valenti-Hirsch MP, Adam C, Denuelle M. Interpretation of SEEG recordings. Neurophysiologie Clinique = Clin Neurophysiol. 2018;48(1):53–57. doi: 10.1016/j.neucli.2017.11.010. [DOI] [PubMed] [Google Scholar]
- Bation R, Mondino M, Le Camus F, Saoud M, Brunelin J. Transcranial direct current stimulation in patients with obsessive compulsive disorder: A randomized controlled trial. Eur Psychiatry. 2019;62:38–44. doi: 10.1016/j.eurpsy.2019.08.011. [DOI] [PubMed] [Google Scholar]
- Battisti A, Lazzaro G, Ursumando L, D’Aiello B, Zanna V, Costanzo F, Vicari S, Menghini D. Examining tolerability, safety, and blinding in 1032 transcranial electrical stimulation sessions for children and adolescents with neuropsychiatric and neurodevelopmental disorders. Sci Rep. 2025;15(1):4560. doi: 10.1038/s41598-025-88256-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Baumann S, Mareš T, Albrecht J, Anders M, Vochosková K, Hill M, Bulant J, Yamamotová A, Štastný O, Novák T, Holanová P, et al. Effects of transcranial direct current stimulation treatment for anorexia nervosa. Front Psych. 2021;12:717255. doi: 10.3389/fpsyt.2021.717255. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Beanato E, Moon H-J, Windel F, Vassiliadis P, Wessel MJ, Popa T, Pauline M, Neufeld E, De Falco E, Gauthier B, Steiner M, et al. Noninvasive modulation of the hippocampal-entorhinal complex during spatial navigation in humans. Sci Adv. 2024;10(44):eado4103. doi: 10.1126/sciadv.ado4103. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Datta A. Guidelines for precise and accurate computational models of tDCS. Brain Stimul. 2012;5(3):430–431. doi: 10.1016/j.brs.2011.06.001. [DOI] [PubMed] [Google Scholar]
- Bikson M, Datta A, Elwassif M. Establishing safety limits for transcranial direct current stimulation. Clin Neurophysiol. 2009;120(6):1033–1034. doi: 10.1016/j.clinph.2009.03.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Datta A, Rahman A, Scaturro J. Electrode montages for tDCS and weak transcranial electrical stimulation: Role of “return” electrode’s position and size. Clin Neurophysiol. 2010;121(12):1976–1978. doi: 10.1016/j.clinph.2010.05.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Dmochowski J, Rahman A. The “quasi-uniform” assumption in animal and computational models of non-invasive electrical stimulation. Brain Stimul. 2013a;6(4):704–705. doi: 10.1016/j.brs.2012.11.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Esmaeilpour Z, Adair D, Kronberg G, Tyler WJ, Antal A, Datta A, Sabel BA, Nitsche MA, Loo C, Edwards D, et al. Transcranial electrical stimulation nomenclature. Brain Stimul. 2019;12(6):1349–1366. doi: 10.1016/j.brs.2019.07.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Ganho-Ávila A, Datta A, Gillick B, Joensson MG, Kim S, Kim J, Kirton A, Lee K, Marjenin T, Onarheim B, et al. Limited output transcranial electrical stimulation 2023 (LOTES-2023): Updates on engineering principles, regulatory statutes, and industry standards for wellness, over-the-counter, or prescription devices with low risk. Brain Stimul. 2023;16(3):840–853. doi: 10.1016/j.brs.2023.05.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Grossman P, Thomas C, Zannou AL, Jiang J, Adnan T, Mourdoukoutas AP, Kronberg G, Truong D, Boggio P, Brunoni AR, et al. Safety of Transcranial Direct Current Stimulation: Evidence Based Update 2016. Brain Stimul. 2016;9(5):641–661. doi: 10.1016/j.brs.2016.06.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Grossman P, Zannou AL, Kronberg G, Truong D, Boggio P, Brunoni AR, Charvet L, Fregni F, Fritsch B, Gillick B, et al. Response to letter to the editor: Safety of transcranial direct current stimulation: Evidence based update 2016. Brain Stimul. 2017;10(5):986–987. doi: 10.1016/j.brs.2017.06.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Inoue M, Akiyama H, Deans JK, Fox JE, Miyakawa H, Jefferys JGR. Effects of uniform extracellular DC electric fields on excitability in rat hippocampal slices in vitro. J Physiol. 2004;557(Pt 1):175–190. doi: 10.1113/jphysiol.2003.055772. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Name A, Rahman A. Origins of specificity during tDCS: Anatomical, activity-selective, and input-bias mechanisms. Front Hum Neurosci. 2013b;7:688. doi: 10.3389/fnhum.2013.00688. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bikson M, Paneri B, Mourdoukoutas A, Esmaeilpour Z, Badran BW, Azzam R, Adair D, Datta A, Fang XH, Wingeier B, Chao D, et al. Limited output transcranial electrical stimulation (LOTES-2017): Engineering principles, regulatory statutes, and industry standards for wellness, over-the-counter, or prescription devices with low risk. Brain Stimul. 2018;11(1):134–157. doi: 10.1016/j.brs.2017.10.012. [DOI] [PubMed] [Google Scholar]
- Bjekić J, Živanović M, Stanković M, Paunović D, Konstantinović U, Filipović SR. The subjective experience of transcranial electrical stimulation: A within-subject comparison of tolerability and side effects between tDCS, tACS, and otDCS. Front Hum Neurosci. 2024;18 doi: 10.3389/fnhum.2024.1468538. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bland NS, Sale MV. Current challenges: The ups and downs of tACS. Exp Brain Res. 2019;237(12):3071–3088. doi: 10.1007/s00221-019-05666-0. [DOI] [PubMed] [Google Scholar]
- Blaschke SJ, Vlachakis S, Pallast N, Walter HL, Volz LJ, Wiedermann D, Fink GR, Hoehn M, Aswendt M, Schroeter M, Rueger MA. Transcranial Direct Current Stimulation Reverses Stroke-Induced Network Alterations in Mice. Stroke. 2023;54(8):2145–2155. doi: 10.1161/STROKEAHA.123.042808. [DOI] [PubMed] [Google Scholar]
- Boccard-Binet S, Sen A. Safety of transcranial direct current stimulation in healthy participants. Epilepsy & behavior reports. 2020;15:100414. doi: 10.1016/j.ebr.2020.100414. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bolognini N, Souza Carneiro MI, Russo C, Vallar G. Transcranial direct current stimulation in stroke rehabilitation: Ready to move to randomized clinical trials and clinical practice? The issue of safety guidelines. Eur J Neurol. 2017;24(11):e78. doi: 10.1111/ene.13396. [DOI] [PubMed] [Google Scholar]
- Borrelli B, Sepinwall D, Ernst D, Bellg AJ, Czajkowski S, Breger R, DeFrancesco C, Levesque C, Sharp DL, Ogedegbe G, Resnick B, et al. A new tool to assess treatment fidelity and evaluation of treatment fidelity across 10 years of health behavior research. J Consult Clin Psychol. 2005;73(5):852–860. doi: 10.1037/0022-006X.73.5.852. [DOI] [PubMed] [Google Scholar]
- Borrione L, Cavendish BA, Aparicio LVM, Luethi MS, Goerigk S, Ramos MRF, Moran NKS, Carneiro AM, Valiengo L, Moura DO, de Souza JP, et al. Home-Use Transcranial Direct Current Stimulation for the Treatment of a Major Depressive Episode: A Randomized Clinical Trial. JAMA Psychiat. 2024;81(4):329–337. doi: 10.1001/jamapsychiatry.2023.4948. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bouaziz N, Luisada J-C, Jabri S, Andrianisaina P-S-K, Bellis A, Januel D. Moving to accelerated protocols of tDCS in catatonia: A case report. Front Psych. 2023;14:1302718. doi: 10.3389/fpsyt.2023.1302718. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bradley C, Nydam AS, Dux PE, Mattingley JB. State-dependent effects of neural stimulation on brain function and cognition. Nat Rev Neurosci. 2022;23(8):459–475. doi: 10.1038/s41583-022-00598-1. [DOI] [PubMed] [Google Scholar]
- Brambilla M, Dinkelbach L, Bigler A, Williams J, Zokaei N, Cohen Kadosh R, Brem AK. The Effect of Transcranial Random Noise Stimulation on Cognitive Training Outcome in Healthy Aging. Front Neurol. 2021;12:625359. doi: 10.3389/fneur.2021.625359. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Braun R, Klein R, Walter HL, Ohren M, Freudenmacher L, Getachew K, Ladwig A, Luelling J, Neumaier B, Endepols H, Graf R, et al. Transcranial direct current stimulation accelerates recovery of function, induces neurogenesis and recruits oligodendrocyte precursors in a rat model of stroke. Exp Neurol. 2016;279:127–136. doi: 10.1016/j.expneurol.2016.02.018. [DOI] [PubMed] [Google Scholar]
- Brevet-Aeby C, Mondino M, Poulet E, Brunelin J. Three repeated sessions of transcranial random noise stimulation (tRNS) leads to long-term effects on reaction time in the Go/No Go task. Neurophysiologie Clinique = Clin Neurophysiol. 2019;49(1):27–32. doi: 10.1016/j.neucli.2018.10.066. [DOI] [PubMed] [Google Scholar]
- Brunelin J, Mondino M, Gassab L, Haesebaert F, Gaha L, Suaud-Chagny M-F, Saoud M, Mechri A, Poulet E. Examining transcranial direct-current stimulation (tDCS) as a treatment for hallucinations in schizophrenia. Am J Psychiatry. 2012;169(7):719–724. doi: 10.1176/appi.ajp.2012.11071091. [DOI] [PubMed] [Google Scholar]
- Brunoni AR, Ekhtiari H, Antal A, Auvichayapat P, Baeken C, Benseñor IM, Bikson M, Boggio P, Borroni B, Brighina F, Brunelin J, et al. Digitalized transcranial electrical stimulation: A consensus statement. Clin Neurophysiol. 2022;143:154–165. doi: 10.1016/j.clinph.2022.08.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brunoni AR, Ferrucci R, Bortolomasi M, Vergari M, Tadini L, Boggio PS, Giacopuzzi M, Barbieri S, Priori A. Transcranial direct current stimulation (tDCS) in unipolar vs. Bipolar depressive disorder. Prog Neuropsychopharmacol Biol Psychiatry. 2011;35(1):96–101. doi: 10.1016/j.pnpbp.2010.09.010. [DOI] [PubMed] [Google Scholar]
- Brunoni AR, Ferrucci R, Fregni F, Boggio PS, Priori A. Transcranial direct current stimulation for the treatment of major depressive disorder: A summary of preclinical, clinical and translational findings. Prog Neuropsychopharmacol Biol Psychiatry. 2012;39(1):9–16. doi: 10.1016/j.pnpbp.2012.05.016. [DOI] [PubMed] [Google Scholar]
- Brunoni AR, Moffa AH, Sampaio-Júnior B, Gálvez V, Loo CK. Treatment-emergent mania/hypomania during antidepressant treatment with transcranial direct current stimulation (tDCS): A systematic review and meta-analysis. Brain Stimul. 2017;10(2):260–262. doi: 10.1016/j.brs.2016.11.005. [DOI] [PubMed] [Google Scholar]
- Buch ER, Santarnecchi E, Antal A, Born J, Celnik PA, Classen J, Gerloff C, Hallett M, Hummel FC, Nitsche MA, Pascual-Leone A, et al. Effects of tDCS on motor learning and memory formation: A consensus and critical position paper. Clin Neurophysiol. 2017;128(4):589–603. doi: 10.1016/j.clinph.2017.01.004. [DOI] [PubMed] [Google Scholar]
- Buchanan DM, Bogdanowicz T, Khanna N, Lockman-Dufour G, Robaey P, D’Angiulli A. Systematic review on the safety and tolerability of transcranial direct current stimulation in children and adolescents. Brain Sci. 2021;11(2):212. doi: 10.3390/brainsci11020212. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Buchanan DM, D’Angiulli A, Samson A, Maisonneuve AR, Robaey P. Acceptability of transcranial direct current stimulation in children and adolescents with ADHD: The point of view of parents. J Health Psychol. 2022;27(1):36–46. doi: 10.1177/1359105320937059. [DOI] [PubMed] [Google Scholar]
- Bulteau S, Sébille V, Fayet G, Thomas-Ollivier V, Deschamps T, Bonnin-Rivalland A, Laforgue E, Pichot A, Valrivière P, Auffray-Calvier E, Fortin J, et al. Efficacy of intermittent Theta Burst Stimulation (iTBS) and 10-Hz high-frequency repetitive transcranial magnetic stimulation (rTMS) in treatment-resistant unipolar depression: Study protocol for a randomised controlled trial. Trials. 2017;18(1):17. doi: 10.1186/s13063-016-1764-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Burkhardt G, Kumpf U, Crispin A, Goerigk S, Andre E, Plewnia C, Brendel B, Fallgatter A, Langguth B, Abdelnaim M, Hebel T, et al. Transcranial direct current stimulation as an additional treatment to selective serotonin reuptake inhibitors in adults with major depressive disorder in Germany (DepressionDC): A triple-blind, randomised, sham-controlled, multicentre trial. Lancet (London, England) 2023;402(10401):545–554. doi: 10.1016/S0140-6736(23)00640-2. [DOI] [PubMed] [Google Scholar]
- Cacciamani L, Tomer D, Mylod-Vargas MG, Selcov A, Peterson GA, Oseguera CI, Barbieux A. HD-tDCS to the lateral occipital complex improves haptic object recognition. Exp Brain Res. 2024;242(9):2113–2124. doi: 10.1007/s00221-024-06888-7. [DOI] [PubMed] [Google Scholar]
- Callai EMM, Zin LEF, Catarina LS, Ponzoni D, Gonçalves CAS, Vizuete AFK, Cougo MC, Boff J, Puricelli E, Fernandes EK, da Silva Torres IL, et al. Evaluation of the immediate effects of a single transcranial direct current stimulation session on astrocyte activation, inflammatory response, and pain threshold in naïve rats. Behav Brain Res. 2022;428:113880. doi: 10.1016/j.bbr.2022.113880. [DOI] [PubMed] [Google Scholar]
- Cancel LM, Arias K, Bikson M, Tarbell JM. Direct current stimulation of endothelial monolayers induces a transient and reversible increase in transport due to the electroosmotic effect. Sci Rep. 2018;8(1):9265. doi: 10.1038/s41598-018-27524-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cancel LM, Silas D, Bikson M, Tarbell JM. Direct current stimulation modulates gene expression in isolated astrocytes with implications for glia-mediated plasticity. Sci Rep. 2022;12(1):17964. doi: 10.1038/s41598-022-22394-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Caparelli-Daquer EM, Zimmermann TJ, Mooshagian E, Parra LC, Rice JK, Datta A, Bikson M, Wassermann EM. A pilot study on effects of 4 × 1 high-definition tDCS on motor cortex excitability; Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2012. pp. 735–738. Annual International Conference. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cappon D, den Boer T, Jordan C, Yu W, Lo A, LaGanke N, Biagi MC, Skorupinski P, Ruffini G, Morales O, Metzger E, et al. Safety and feasibility of tele-supervised home-based transcranial direct current stimulation for major depressive disorder. Front Aging Neurosci. 2022;13:765370. doi: 10.3389/fnagi.2021.765370. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Carlton CN, Bondy E, Riddle J, Mahan D, Daughters S, Schiller C, Frohlich F. Augmenting single-session behavioral activation for depression with delta-beta tACS: Preliminary investigation of a double-blind placebo-controlled randomized pilot clinical trial. J Mood Anxiety Disorders. 2025;11:100127. doi: 10.1016/j.xjmad.2025.100127. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cassarà AM, Newton TH, Zhuang K, Regel SJ, Achermann P, Pascual-Leone A, Kuster N, Neufeld E. Recommendations for the safe application of temporal interference stimulation in the human brain Part I: Principles of electrical neuromodulation and adverse effects. Bioelectromagnetics. 2025;46(2):e22542. doi: 10.1002/bem.22542. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Castellano-Tejedor C. Non-pharmacological interventions for the management of chronic health conditions and non-communicable diseases. Int J Environ Res Public Health. 2022;19(14):8536. doi: 10.3390/ijerph19148536. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Centeno M, Carmichael DW. Network Connectivity in Epilepsy: Resting State fMRI and EEG-fMRI Contributions. Front Neurol. 2014;5:93. doi: 10.3389/fneur.2014.00093. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Central Drugs Standard Control Organization. Circulars. n.d Retrieved April 30, 2025, from https://cdsco.gov.in/opencms/opencms/en/Acts-and-rules/Medical-Devices-Rules/
- Chang C-C, Huang C-C-Y, Chung Y-A, Im JJ, Lin Y-Y, Ma C-C, Tzeng N-S, Chang H-A. Online left-hemispheric in-phase frontoparietal theta tACS for the treatment of negative symptoms of Schizophrenia. J Pers Med. 2021;11(11):11. doi: 10.3390/jpm11111114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chao P-C, Chang C-C, Chang H-A. Hypomania induced by bifrontal transcranial direct current stimulation in a patient with bipolar depression. Psychiatry Investig. 2018;15(9):914–915. doi: 10.30773/pi.2018.08.11.1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Charvet LE, Kasschau M, Datta A, Knotkova H, Stevens MC, Alonzo A, Loo C, Krull KR, Bikson M. Remotely-supervised transcranial direct current stimulation (tDCS) for clinical trials: Guidelines for technology and protocols. Front Syst Neurosci. 2015;9:26. doi: 10.3389/fnsys.2015.00026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Charvet LE, Shaw MT, Bikson M, Woods AJ, Knotkova H. Supervised transcranial direct current stimulation (tDCS) at home: A guide for clinical research and practice. Brain Stimul. 2020;13(3):686–693. doi: 10.1016/j.brs.2020.02.011. [DOI] [PubMed] [Google Scholar]
- Charvet L, George A, Charlson E, Lustberg M, Vogel-Eyny A, Eilam-Stock T, Cho H, Best P, Fernandez L, Datta A, Bikson M, et al. Home-administered transcranial direct current stimulation is a feasible intervention for depression: An observational cohort study. Front Psych. 2023a;14:1199773. doi: 10.3389/fpsyt.2023.1199773. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Charvet L, Goldberg JD, Li X, Best P, Lustberg M, Shaw M, Zhovtis L, Gutman J, Datta A, Bikson M, Pilloni G, et al. Home-based transcranial direct current stimulation paired with cognitive training to reduce fatigue in multiple sclerosis. Sci Rep. 2025;15(1):4551. doi: 10.1038/s41598-025-88255-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Charvet L, Harrison AT, Mangold K, Moore RD, Guo S, Zhang J, Datta A, Androulakis XM. Remotely supervised at-home tDCS for veterans with persistent post-traumatic headache: A double-blind, sham-controlled randomized pilot clinical trial. Front Neurol. 2023b;14:1184056. doi: 10.3389/fneur.2023.1184056. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Charvet L, Shaw M, Dobbs B, Frontario A, Sherman K, Bikson M, Datta A, Krupp L, Zeinapour E, Kasschau M. Remotely supervised transcranial direct current stimulation increases the benefit of at-home cognitive training in multiple sclerosis. Neuromodul Soc. 2018;21(4):383–389. doi: 10.1111/ner.12583. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chen J-J, Zeng B-S, Wu C-N, Stubbs B, Carvalho A-F, Brunoni AR, Su K-P, Tu Y-K, Wu Y-C, Chen T-Y, Lin P-Y, et al. Association of central noninvasive brain stimulation interventions with efficacy and safety in tinnitus management: A meta-analysis. JAMA OtolaryngolHead Neck Surg. 2020;146(9):801–809. doi: 10.1001/jamaoto.2020.1497. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chen Y, Ou Z, Hao N, Zhang H, Zhang E, Zhou D, Wu X. Transcranial direct current stimulation in the management of epilepsy: A meta-analysis and systematic review. Front Neurol. 2024;15:1462364. doi: 10.3389/fneur.2024.1462364. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chenot Q, Hamery C, Lepron E, Besson P, De Boissezon X, Perrey S, Scannella S. Performance after training in a complex cognitive task is enhanced by high-definition transcranial random noise stimulation. Sci Rep. 2022;12(1):4618. doi: 10.1038/s41598-022-08545-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cherney LR, Babbitt EM, Wang X, Pitts LL. Extended fMRI-guided anodal and cathodal transcranial direct current stimulation targeting perilesional areas in post-stroke aphasia: A pilot randomized clinical trial. Brain Sci. 2021;11(3):306. doi: 10.3390/brainsci11030306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chhatbar PY, Chen R, Deardorff R, Dellenbach B, Kautz SA, George MS, Feng W. Safety and tolerability of transcranial direct current stimulation to stroke patients—A phase I current escalation study. Brain Stimul. 2017;10(3):553–559. doi: 10.1016/j.brs.2017.02.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Chhatbar PY, Kautz SA, Takacs I, Rowland NC, Revuelta GJ, George MS, Bikson M, Feng W. Evidence of transcranial direct current stimulation-generated electric fields at subthalamic level in human brain in vivo. Brain Stimul. 2018;11(4):727–733. doi: 10.1016/j.brs.2018.03.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cho N, Squair JW, Aureli V, James ND, Bole-Feysot L, Dewany I, Hankov N, Baud L, Leonhartsberger A, Sveistyte K, Skinnider MA, et al. Hypothalamic deep brain stimulation augments walking after spinal cord injury. Nat Med. 2024;30(12):3676–3686. doi: 10.1038/s41591-024-03306-x. [DOI] [PubMed] [Google Scholar]
- Choy O, Tan G, Wong YC. The effect of multi-session prefrontal cortical stimulation on aggression: A randomized, double-blind, parallel-group Trial. Life. 2023;13(8):8. doi: 10.3390/life13081729. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ciechanski P, Carlson HL, Yu SS, Kirton A. Modeling transcranial direct-current stimulation-induced electric fields in children and adults. Front Hum Neurosci. 2018;12:268. doi: 10.3389/fnhum.2018.00268. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ciechanski P, Cheng A, Lopushinsky S, Hecker K, Gan LS, Lang S, Zareinia K, Kirton A. Effects of Transcranial Direct-Current Stimulation on Neurosurgical Skill Acquisition: A Randomized Controlled Trial. World Neurosurg. 2017;108:876–884.:e4. doi: 10.1016/j.wneu.2017.08.123. [DOI] [PubMed] [Google Scholar]
- Cody SL, Fazeli PL, Crowe M, Kempf M-C, Moneyham L, Stavrinos D, Vance DE, Heaton K. Effects of speed of processing training and transcranial direct current stimulation on global sleep quality and speed of processing in older adults with and without HIV: A pilot study. Appl Neuropsychol Adult. 2020;27(3):267–278. doi: 10.1080/23279095.2018.1534736. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cogiamanian F, Vergari M, Pulecchi F, Marceglia S, Priori A. Effect of spinal transcutaneous direct current stimulation on somatosensory evoked potentials in humans. Clin Neurophysiol. 2008;119(11):2636–2640. doi: 10.1016/j.clinph.2008.07.249. [DOI] [PubMed] [Google Scholar]
- Cohen Kadosh R, Ciobotaru D, Karstens M, Nguyen V. Personalized home based neurostimulation via AI optimization augments sustained attention. Npj Digital Medicine. doi: 10.1038/s41746-025-01744-6. (In press) [DOI] [PMC free article] [PubMed] [Google Scholar]
- Couture M, Desbeaumes Jodoin V, Bousseau E, Sarshoghi A, Nitsche MA, Blumberger DM, Bolduc C, Weissman CR, Appelbaum LG, Daskalakis ZJ, Poorganji M, et al. Spaced Transcranial Direct Current Stimulation for Major Depression. Am J Psychiatry. 2025;182(3):276–284. doi: 10.1176/appi.ajp.20240083. [DOI] [PubMed] [Google Scholar]
- Cunillera T, Brignani D, Cucurell D, Fuentemilla L, Miniussi C. The right inferior frontal cortex in response inhibition: A tDCS-ERP co-registration study. Neuroimage. 2016;140:66–75. doi: 10.1016/j.neuroimage.2015.11.044. [DOI] [PubMed] [Google Scholar]
- da Silva R de MF, Brunoni AR, Miguel EC, Shavitt RG. Transcranial direct current stimulation for Obsessive-Compulsive Disorder: Patient selection and perspectives. Neuropsychiatr Dis Treat. 2019;15:2663–2669. doi: 10.2147/NDT.S184839. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dagan M, Herman T, Harrison R, Zhou J, Giladi N, Ruffini G, Manor B, Hausdorff JM. Multitarget transcranial direct current stimulation for freezing of gait in Parkinson’s disease. Mov Disord. 2018;33(4):642–646. doi: 10.1002/mds.27300. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dakwar-Kawar O, Berger I, Barzilay S, Grossman ES, Cohen Kadosh R, Nahum M. Examining the Effect of Transcranial Electrical Stimulation and Cognitive Training on Processing Speed in Pediatric Attention Deficit Hyperactivity Disorder: A Pilot Study. Front Hum Neurosci. 2022;16:791478. doi: 10.3389/fnhum.2022.791478. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dakwar-Kawar O, Mairon N, Hochman S, Berger I, Cohen Kadosh R, Nahum M. Transcranial random noise stimulation combined with cognitive training for treating ADHD: A randomized, sham-controlled clinical trial. Transl Psychiatry. 2023;13(1):1–11. doi: 10.1038/s41398-023-02547-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Daoud M, Medina-Villalon S, Garnier E, Bratu IF, Damiani G, Salvador R, Wendling F, Ruffini G, Bénar C, Pizzo F, Bartolomei F. Epileptogenic zone characteristics determine effectiveness of electrical transcranial stimulation in epilepsy treatment. Brain Communications. 2025;7(1):fcaf012. doi: 10.1093/braincomms/fcaf012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Daoud M, Salvador R, El Youssef N, Fierain A, Garnier E, Biagi MC, Medina Villalon S, Wendling F, Benar C, Ruffini G, Bartolomei F. Stereo-EEG based personalized multichannel transcranial direct current stimulation in drug-resistant epilepsy. Clin Neurophysiol. 2022;137:142–151. doi: 10.1016/j.clinph.2022.02.023. [DOI] [PubMed] [Google Scholar]
- dArsonval JA. Recherches d’électrotherapie: La voltaisation sinusodale. Arch Physiol Pathol. 1892;4:69–80. [Google Scholar]
- Dashtelei AA, Nitsche MA, Salehinejad MA, Habibi AH, Bakhtyiari J, Khatoonabadi AR. Adjunctive transcranial direct current stimulation to improve swallowing functions in Parkinson’s disease. EXCLI J. 2024;23:95–107. doi: 10.17179/excli2023-6496. [DOI] [PMC free article] [PubMed] [Google Scholar]
- DaSilva AF, Volz MS, Bikson M, Fregni F. Electrode positioning and montage in transcranial direct current stimulation. Journal of Visualized Experiments: JoVE. 2011;51:2744. doi: 10.3791/2744. [DOI] [PMC free article] [PubMed] [Google Scholar]
- daSilva Morgan K, Schumacher J, Collerton D, Colloby S, Elder GJ, Olsen K, Ffytche DH, Taylor J-P. Transcranial direct current stimulation in the treatment of visual hallucinations in Charles Bonnet Syndrome: A randomized placebo-controlled crossover trial. Ophthalmology. 2022;129(12):1368–1379. doi: 10.1016/j.ophtha.2022.06.041. [DOI] [PubMed] [Google Scholar]
- Datta A, Bansal V, Diaz J, Patel J, Reato D, Bikson M. Gyri-precise head model of transcranial direct current stimulation: Improved spatial focality using a ring electrode versus conventional rectangular pad. Brain Stimul. 2009a;2(4):201–207.:e1. doi: 10.1016/j.brs.2009.03.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Datta A, Bikson M, Fregni F. Transcranial direct current stimulation in patients with skull defects and skull plates: High-resolution computational FEM study of factors altering cortical current flow. Neuroimage. 2010;52(4):1268–1278. doi: 10.1016/j.neuroimage.2010.04.252. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Datta A, Elwassif M, Bikson M. Bio-heat transfer model of transcranial DC stimulation: Comparison of conventional pad versus ring electrode; Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2009b. pp. 670–673. Annual International Conference. [DOI] [PubMed] [Google Scholar]
- Daughters SB, Yi JY, Phillips RD, Carelli RM, Fröhlich F. Alpha-tACS effect on inhibitory control and feasibility of administration in community outpatient substance use treatment. Drug Alcohol Depend. 2020;213:108132. doi: 10.1016/j.drugalcdep.2020.108132. [DOI] [PMC free article] [PubMed] [Google Scholar]
- de Andrade VP, Dos Santos AM, Seguro LPC, Yuki EFN, Lopes MRU, Grecco MV, Borba EF, Greve JMA, Shinjo SK. Neuromodulation with aerobic exercise reduces fatigue in systemic lupus erythematosus: A randomised, sham-controlled, double-blind study. Clin Exp Rheumatol. 2024;42(7):1451–1458. doi: 10.55563/clinexprheumatol/yvrdgm. [DOI] [PubMed] [Google Scholar]
- Deans JK, Powell AD, Jefferys JGR. Sensitivity of coherent oscillations in rat hippocampus to AC electric fields. J Physiol. 2007;583(Pt 2):555–565. doi: 10.1113/jphysiol.2007.137711. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Declaration of Geneva. Wikipedia. 2025. https://en.wikipedia.org/w/index.php?title=Declaration_of_Geneva&oldid=1276976951 .
- Demchenko I, Tailor I, Chegini S, Yu H, Nezhad FG, Rueda A, Kever A, Krishnan S, Datta A, Meltzer JA, Graham SJ, et al. Human Applications of Transcranial Temporal Interference Stimulation: A Systematic Review. medRxiv: The Preprint Server for Health Sciences. 2025:2025.05.16.25327804. doi: 10.1101/2025.05.16.25327804. [DOI] [Google Scholar]
- Demina A, Petit B, Meille V, Sauvaget A, Lebon F, Trojak B. Combination of non-invasive brain stimulation with mindfulness-based interventions for anxiety and depression symptoms: Systematic review and meta-analysis of randomized controlled trials. Eur Arch Psychiatry Clin Neurosci. 2024 doi: 10.1007/s00406-024-01928-3. [DOI] [PubMed] [Google Scholar]
- Diedrich L, Kolhoff HI, Bergmann C, Bähr M, Antal A. Boosting working memory in the elderly: Driving prefrontal theta–gamma coupling via repeated neuromodulation. GeroScience. 2024a;47(2):1425–1440. doi: 10.1007/s11357-024-01272-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Diedrich L, Kolhoff HI, Chakalov I, Vékony T, Németh D, Antal A. Prefrontal theta—Gamma transcranial alternating current stimulation improves non-declarative visuomotor learning in older adults. Sci Rep. 2024b;14(1):4955. doi: 10.1038/s41598-024-55125-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Directorate-General for Health and Food Safety. Guidance on classification of medical devices. 2021. https://health.ec.europa.eu/latest-updates/mdcg-2021-24-guidance-classification-medical-devices-2021-10-04_en .
- Dittert N, Hüttner S, Polak T, Herrmann MJ. Augmentation of fear extinction by transcranial direct current stimulation (tDCS) Front Behav Neurosci. 2018;12:76. doi: 10.3389/fnbeh.2018.00076. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dmochowski JP, Datta A, Bikson M, Su Y, Parra LC. Optimized multi-electrode stimulation increases focality and intensity at target. J Neural Eng. 2011;8(4):046011. doi: 10.1088/1741-2560/8/4/046011. [DOI] [PubMed] [Google Scholar]
- Dobbs B, Pawlak N, Biagioni M, Agarwal S, Shaw M, Pilloni G, Bikson M, Datta A, Charvet L. Generalizing remotely supervised transcranial direct current stimulation (tDCS): Feasibility and benefit in Parkinson’s disease. J Neuroeng Rehabil. 2018;15(1):114. doi: 10.1186/s12984-018-0457-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dondé C, Neufeld NH, Geoffroy PA. The impact of Transcranial Direct Current Stimulation (tDCS) on bipolar depression, mania, and euthymia: A systematic review of preliminary data. Psychiatry Q. 2018;89(4):855–867. doi: 10.1007/s11126-018-9584-5. [DOI] [PubMed] [Google Scholar]
- Draaisma LR, Wessel MJ, Moyne M, Morishita T, Hummel FC. Targeting the frontoparietal network using bifocal transcranial alternating current stimulation during a motor sequence learning task in healthy older adults. Brain Stimul. 2022;15(4):968–979. doi: 10.1016/j.brs.2022.06.012. [DOI] [PubMed] [Google Scholar]
- Dragon K, Abdelnaim MA, Weber FC, Heuschert M, Englert L, Langguth B, Hebel T, Schecklmann M. Treating depression at home with transcranial direct current stimulation: A feasibility study. Front Psych. 2024;15 doi: 10.3389/fpsyt.2024.1335243. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Duffy MJ, Feltman KA, Kelley AM, Mackie R. Limitations associated with transcranial direct current stimulation for enhancement: considerations of performance tradeoffs in active-duty soldiers. Front Hum Neurosci. 2024;18:1444450. doi: 10.3389/fnhum.2024.1444450. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dumel G, Bourassa M-E, Charlebois-Plante C, Desjardins M, Doyon J, Saint-Amour D, De Beaumont L. Motor learning improvement remains 3 months after a multisession anodal tDCS intervention in an aging population. Front Aging Neurosci. 2018;10:335. doi: 10.3389/fnagi.2018.00335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dundas JE, Thickbroom GW, Mastaglia FL. Perception of comfort during transcranial DC stimulation: effect of NaCl solution concentration applied to sponge electrodes. Clin Neurophysiol. 2007;118(5):1166–1170. doi: 10.1016/j.clinph.2007.01.010. [DOI] [PubMed] [Google Scholar]
- D’Urso G, Toscano E, Sanges V, Sauvaget A, Sheffer CE, Riccio MP, Ferrucci R, Iasevoli F, Priori A, Bravaccio C, de Bartolomeis A. Cerebellar transcranial direct current stimulation in children with autism spectrum disorder: A pilot study on efficacy, feasibility, safety, and unexpected outcomes in tic disorder and epilepsy. J Clin Med. 2021;11(1):143. doi: 10.3390/jcm11010143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Edwards D, Cortes M, Datta A, Minhas P, Wassermann EM, Bikson M. Physiological and modeling evidence for focal transcranial electrical brain stimulation in humans: A basis for high-definition tDCS. Neuroimage. 2013;74:266–275. doi: 10.1016/j.neuroimage.2013.01.042. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ekhtiari H, Ghobadi-Azbari P, Thielscher A, Antal A, Li LM, Shereen AD, Cabral-Calderin Y, Keeser D, Bergmann TO, Jamil A, Violante IR, et al. A checklist for assessing the methodological quality of concurrent tES-fMRI studies (ContES checklist): A consensus study and statement. Nat Protoc. 2022;17(3):596–617. doi: 10.1038/s41596-021-00664-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- El Jamal C, Harrie A, Rahman-Filipiak A, Iordan AD, DaSilva AF, Ploutz-Snyder R, Khadr L, Vesia M, Bikson M, Hampstead BM. Tolerability and blinding of high-definition transcranial direct current stimulation among older adults at intensities of up to 4 mA per electrode. Brain Stimul. 2023;16(5):1328–1335. doi: 10.1016/j.brs.2023.08.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Erb C, Eckert S, Gindorf P, Köhler M, Köhler T, Neuhann L, Neuhann T, Salzmann N, Schmickler S, Ellrich J. Electrical neurostimulation in glaucoma with progressive vision loss. Bioelectron Med. 2022;8(1):6. doi: 10.1186/s42234-022-00089-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Esmaeilpour Z, Marangolo P, Hampstead BM, Bestmann S, Galletta E, Knotkova H, Bikson M. Incomplete evidence that increasing current intensity of tDCS boosts outcomes. Brain Stimul. 2018;11(2):310–321. doi: 10.1016/j.brs.2017.12.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Esmaeilpour Z, Schestatsky P, Bikson M, Brunoni AR, Pellegrinelli A, Piovesan FX, Santos MMSA, Menezes RB, Fregni F. Notes on human trials of transcranial direct current stimulation between 1960 and 1998. Front Hum Neurosci. 2017;11 doi: 10.3389/fnhum.2017.00071. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Esposito M, Mauri P, Panizza L, Mazza V, Miniussi C, Brignani D. Baseline levels of alertness influence tES effects along different age-related directions. Neuropsychologia. 2021;160:107966. doi: 10.1016/j.neuropsychologia.2021.107966. [DOI] [PubMed] [Google Scholar]
- Ester-Nacke T, Berti K, Veit R, Dannecker C, Salvador R, Ruffini G, Heni M, Birkenfeld AL, Plewnia C, Preissl H, Kullmann S. Network-targeted transcranial direct current stimulation of the hypothalamus appetite-control network: A feasibility study. Sci Rep. 2024;14(1):11341. doi: 10.1038/s41598-024-61852-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ester-Nacke T, Veit R, Thomanek J, Book M, Tamble L, Beermann M, Löffler D, Salvador R, Ruffini G, Heni M, Birkenfeld AL, et al. Repeated net-tDCS of the hypothalamus appetite-control network enhances inhibitory control and decreases sweet food intake in persons with overweight or obesity. Brain Stimul. 2025;18(3):863–874. doi: 10.1016/j.brs.2025.04.013. [DOI] [PubMed] [Google Scholar]
- Evans ID, Palmisano S, Loughran SP, Legros A, Croft RJ. Frequency-dependent and montage-based differences in phosphene perception thresholds via transcranial alternating current stimulation. Bioelectromagnetics. 2019;40(6):365–374. doi: 10.1002/bem.22209. [DOI] [PubMed] [Google Scholar]
- Faria P, Fregni F, Sebastião F, Dias AI, Leal A. Feasibility of focal transcranial DC polarization with simultaneous EEG recording: Preliminary assessment in healthy subjects and human epilepsy. Epilepsy Behav. 2012;25(3):417–425. doi: 10.1016/j.yebeh.2012.06.027. [DOI] [PubMed] [Google Scholar]
- Fazeli PL, Woods AJ, Pope CN, Vance DE, Ball KK. Effect of transcranial direct current stimulation combined with cognitive training on cognitive functioning in older adults with HIV: A pilot study. Appl Neuropsychol Adult. 2019;26(1):36–47. doi: 10.1080/23279095.2017.1357037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fehér KD, Morishima Y. Concurrent electroencephalography recording during transcranial alternating current stimulation (tACS) J Visual Exp: JoVE. 2016;107:e53527. doi: 10.3791/53527. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fehér KD, Nakataki M, Morishima Y. Phase-dependent modulation of signal transmission in cortical networks through tACS-induced neural oscillations. Front Hum Neurosci. 2017;11:471. doi: 10.3389/fnhum.2017.00471. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fehér KD, Nakataki M, Morishima Y. Phase-synchronized transcranial alternating current stimulation-induced neural oscillations modulate cortico-cortical signaling efficacy. Brain Connect. 2022;12(5):443–453. doi: 10.1089/brain.2021.0006. [DOI] [PubMed] [Google Scholar]
- Ferrucci R, Cortese F, Priori A. Cerebellar tDCS: how to do it. Cerebellum (London, England) 2015;14(1):27–30. doi: 10.1007/s12311-014-0599-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fertonani A, Ferrari C, Miniussi C. What do you feel if I apply transcranial electric stimulation? Safety, sensations and secondary induced effects. Clin Neurophysiol. 2015;126(11):2181–2188. doi: 10.1016/j.clinph.2015.03.015. [DOI] [PubMed] [Google Scholar]
- Fertonani A, Miniussi C. Transcranial electrical stimulation: what we know and do not know about mechanisms. Neuroscientist. 2017;23(2):109–123. doi: 10.1177/1073858416631966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fertonani A, Pirulli C, Bollini A, Miniussi C, Bortoletto M. Age-related changes in cortical connectivity influence the neuromodulatory effects of transcranial electrical stimulation. Neurobiol Aging. 2019;82:77–87. doi: 10.1016/j.neurobiolaging.2019.07.009. [DOI] [PubMed] [Google Scholar]
- Feurra M, Blagovechtchenski E, Nikulin VV, Nazarova M, Lebedeva A, Pozdeeva D, Yurevich M, Rossi S. State-dependent effects of transcranial oscillatory currents on the motor system during action observation. Sci Rep. 2019;9(1):12858. doi: 10.1038/s41598-019-49166-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Filipas L, Gallo G, Meloni A, Luzi L, Codella R. Effects of bilateral dorsolateral prefrontal cortex high-definition transcranial direct-current stimulation on time-trial performance in cyclists with type 1 diabetes mellitus. Brain Stimul. 2022;15(5):1292–1299. doi: 10.1016/j.brs.2022.09.005. [DOI] [PubMed] [Google Scholar]
- Finocchi C, Ferrari M. Female reproductive steroids and neuronal excitability. Neurol Sci. 2011;32(Suppl 1):S31–S35. doi: 10.1007/s10072-011-0532-5. [DOI] [PubMed] [Google Scholar]
- Fischer DB, Fried PJ, Ruffini G, Ripolles O, Salvador R, Banus J, Ketchabaw WT, Santarnecchi E, Pascual-Leone A, Fox MD. Multifocal tDCS targeting the resting state motor network increases cortical excitability beyond traditional tDCS targeting unilateral motor cortex. Neuroimage. 2017;157:34–44. doi: 10.1016/j.neuroimage.2017.05.060. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fisher DB, Fried P, Ruffini G, Ripolles O, Ketchabaw T, Santarnecchi E, Pascual-Leone A, Fox MD. Network-targeted non-invasive brain stimulation with multifocal tdcs. Brain Stimul. 2017;10(2):411–412. doi: 10.1016/j.brs.2017.01.219. [DOI] [Google Scholar]
- Flynn M, Campbell IC, Schmidt U. Concurrent self-administered transcranial direct current stimulation and attention bias modification training in binge eating disorder: Feasibility randomised sham-controlled trial. BJPsych Open. 2024;10(4):e118. doi: 10.1192/bjo.2024.54. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Food and Drug Administration. Regulatory Controls. FDA; 2018. Mar 27, https://www.fda.gov/medical-devices/overview-device-regulation/regulatory-controls . [Google Scholar]
- Fox MD, Greicius M. Clinical applications of resting state functional connectivity. Front Syst Neurosci. 2010;4:19. doi: 10.3389/fnsys.2010.00019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- França C, de Andrade DC, Teixeira MJ, Galhardoni R, Silva V, Barbosa ER, Cury RG. Effects of cerebellar neuromodulation in movement disorders: A systematic review. Brain Stimul. 2018;11(2):249–260. doi: 10.1016/j.brs.2017.11.015. [DOI] [PubMed] [Google Scholar]
- Fregni F, Nitsche M, Loo CK, Brunoni A, Marangolo P, Leite J, Carvalho S, Bolognini N, Caumo W, Paik N, Simis M, et al. Regulatory Considerations for the Clinical and Research Use of Transcranial Direct Current Stimulation (tDCS): Review and recommendations from an expert panel. Clin Res Regul Aff. 2015;32(1):22–35. doi: 10.3109/10601333.2015.980944. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fresnoza S, Christova M, Feil T, Gallasch E, Körner C, Zimmer U, Ischebeck A. The effects of transcranial alternating current stimulation (tACS) at individual alpha peak frequency (iAPF) on motor cortex excitability in young and elderly adults. Exp Brain Res. 2018;236(10):2573–2588. doi: 10.1007/s00221-018-5314-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fritsch B, Mayer M, Reis J, Gellner A-K. Safety of ipsilesional anodal transcranial direct current stimulation in acute photothrombotic stroke: Implications for early neurorehabilitation. Sci Rep. 2024;14(1):2501. doi: 10.1038/s41598-024-51839-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Froehlich F, McCormick DA. Endogenous electric fields may guide neocortical network activity. Neuron. 2010;67(1):129–143. doi: 10.1016/j.neuron.2010.06.005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Frohlich F, Riddle J. Conducting double-blind placebo-controlled clinical trials of transcranial alternating current stimulation (tACS) Transl Psychiatry. 2021;11(1):284. doi: 10.1038/s41398-021-01391-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fu Y, Kaneko M, Tang Y, Alvarez-Buylla A, Stryker MP. A cortical disinhibitory circuit for enhancing adult plasticity. Elife. 2015;4:e05558. doi: 10.7554/eLife.05558. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Galan-Gadea A, Salvador R, Bartolomei F, Wendling F, Ruffini G. Spherical harmonics representation of the steady-state membrane potential shift induced by tDCS in realistic neuron models. J Neural Eng. 2023;20(2):026004. doi: 10.1088/1741-2552/acbabd. [DOI] [PubMed] [Google Scholar]
- Gálvez V, Alonzo A, Martin D, Mitchell PB, Sachdev P, Loo CK. Hypomania induction in a patient with bipolar II disorder by transcranial direct current stimulation (tDCS) J ECT. 2011;27(3):256–258. doi: 10.1097/YCT.0b013e3182012b89. [DOI] [PubMed] [Google Scholar]
- Gamage NN, Liao W-Y, Hand BJ, Atherton PJ, Piasecki M, Opie GM, Semmler JG. Theta-gamma transcranial alternating current stimulation enhances ballistic motor performance in healthy young and older adults. Neurobiol Aging. 2025;152:1–12. doi: 10.1016/j.neurobiolaging.2025.04.010. [DOI] [PubMed] [Google Scholar]
- Ganho-Ávila A, Gonçalves ÓF, Guiomar R, Boggio PS, Asthana MK, Krypotos AM, Almeida J. The effect of cathodal tDCS on fear extinction: A cross-measures study. PLoS One. 2019;14(9):e0221282. doi: 10.1371/journal.pone.0221282. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ganho-Ávila A, Vieira Figueiredo D, Vagos P. Online cognitive therapy for social anxiety disorder in adolescence: A clinical case study using the CT@TeenSAD. Clin Case Stud. 2022;21(6):533–551. doi: 10.1177/15346501221091519. [DOI] [Google Scholar]
- Gaugain G, Quéguiner L, Bikson M, Sauleau R, Zhadobov M, Modolo J, Nikolayev D. Quasi-static approximation error of electric field analysis for transcranial current stimulation. J Neural Eng. 2023;20(1):016027. doi: 10.1088/1741-2552/acb14d. [DOI] [PubMed] [Google Scholar]
- Gaynor AM, Pergolizzi D, Alici Y, Ryan E, McNeal K, Ahles TA, Root JC. Impact of transcranial direct current stimulation on sustained attention in breast cancer survivors: evidence for feasibility, tolerability, and initial efficacy. Brain Stimul. 2020;13(4):1108–1116. doi: 10.1016/j.brs.2020.04.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gebodh N, Esmaeilpour Z, Adair D, Chelette K, Dmochowski J, Woods AJ, Kappenman ES, Parra LC, Bikson M. Inherent physiological artifacts in EEG during tDCS. Neuroimage. 2019;185:408–424. doi: 10.1016/j.neuroimage.2018.10.025. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gebodh N, Miskovic V, Laszlo S, Datta A, Bikson M. Frontal HD-tACS enhances behavioral and EEG biomarkers of vigilance in continuous attention task. Brain Stimul. 2024;17(3):683–686. doi: 10.1016/j.brs.2024.05.009. [DOI] [PubMed] [Google Scholar]
- Geffen A, Bland N, Sale MV. Effects of slow oscillatory transcranial alternating current stimulation on motor cortical excitability assessed by transcranial magnetic stimulation. Front Hum Neurosci. 2021;15:726604. doi: 10.3389/fnhum.2021.726604. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gehrman PR, Bartky EJ, Travers C, Lapidus K. A fully remote randomized trial of transcranial alternating current stimulation for the acute treatment of major depressive disorder. J Clin Psychiatry. 2024;85(2):23m15078. doi: 10.4088/JCP.23m15078. [DOI] [PubMed] [Google Scholar]
- Gellner A-K, Frase S, Reis J, Fritsch B. Direct current stimulation increases blood flow and permeability of cortical microvasculature in vivo. Eur J Neurol. 2023;30(2):362–371. doi: 10.1111/ene.15616. [DOI] [PubMed] [Google Scholar]
- George A, McConathey E, Vogel-Eyny A, Galletta E, Pilloni G, Charvet L. Feasibility of home-based transcranial direct current stimulation combined with personalized word retrieval for improving naming in primary progressive aphasia. Front Neurol. 2025;16:1543712. doi: 10.3389/fneur.2025.1543712. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ghazi-Noori A-R, Woodham RD, Rezaei H, Sharif MS, Bramon E, Ritter P, Bauer M, Young AH, Fu CHY. Home-based transcranial direct current stimulation in bipolar depression: An open-label treatment study of clinical outcomes, acceptability and adverse events. Int J Bipolar Disorders. 2024;12(1):30. doi: 10.1186/s40345-024-00352-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gholamali Nezhad F, Martin J, Tassone VK, Swiderski A, Demchenko I, Khan S, Chaudhry HE, Palmisano A, Santarnecchi E, Bhat V. Transcranial alternating current stimulation for neuropsychiatric disorders: A systematic review of treatment parameters and outcomes. Front Psych. 2024;15:1419243. doi: 10.3389/fpsyt.2024.1419243. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Goede LL, Oxenford S, Kroneberg D, Meyer GM, Rajamani N, Neudorfer C, Krause P, Lofredi R, Fox MD, Kühn AA, Horn A. Linking invasive and noninvasive brain stimulation in Parkinson’s disease: A randomized trial. Mov Disord. 2024;39(11):1971–1981. doi: 10.1002/mds.29940. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gomez-Tames J, Sugiyama Y, Laakso I, Tanaka S, Koyama S, Sadato N, Hirata A. Effect of microscopic modeling of skin in electrical and thermal analysis of transcranial direct current stimulation. Phys Med Biol. 2016;61(24):8825. doi: 10.1088/1361-6560/61/24/8825. [DOI] [PubMed] [Google Scholar]
- Gong C, Long Y, Peng X-M, Hu H, Chen J, Xiao L, Zhong Y-B, Wang M-Y, Luo Y. Efficacy and safety of noninvasive brain stimulation for patients with cerebellar ataxia: A systematic review and meta-analysis of randomized controlled trials. J Neurol. 2023;270(10):4782–4799. doi: 10.1007/s00415-023-11799-8. [DOI] [PubMed] [Google Scholar]
- Gouveia FV, Silk E, Davidson B, Pople CB, Abrahao A, Hamilton J, Ibrahim GM, Müller DJ, Giacobbe P, Lipsman N, Hamani C. A systematic review on neuromodulation therapies for reducing body weight in patients with obesity. Obes Rev. 2021;22(10):e13309. doi: 10.1111/obr.13309. [DOI] [PubMed] [Google Scholar]
- Gregersen F, Göksu C, Schaefers G, Xue R, Thielscher A, Hanson LG. Safety evaluation of a new setup for transcranial electric stimulation during magnetic resonance imaging. Brain Stimul. 2021;14(3):488–497. doi: 10.1016/j.brs.2021.02.019. [DOI] [PubMed] [Google Scholar]
- Gregoret L, Zamorano AM, Graven-Nielsen T. Multifocal tDCS targeting the motor network modulates event-related cortical responses during prolonged pain. J Pain. 2023;24(2):226–236. doi: 10.1016/j.jpain.2022.09.010. [DOI] [PubMed] [Google Scholar]
- Greinacher R, Buhôt L, Möller L, Learmonth G. The time course of ineffective sham-blinding during low-intensity (1 mA) transcranial direct current stimulation. Eur J Neurosci. 2019;50(8):3380–3388. doi: 10.1111/ejn.14497. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Griffiths C, da Silva KM, Mobee-Dumi A, Gill AK, Ibrahim D. Transcranial Direct Current Stimulation (tDCS) and Online Wellbeing Training Used at Home for Perinatal and Maternal Loss Patients with a Diagnosis of Depression: Depression, Real World-Functioning, and Quality of Life Outcomes. Open Journal of Depression. 2025;14(2):2. doi: 10.4236/ojd.2025.142003. [DOI] [Google Scholar]
- Grigutsch LS, Haverland B, Timmsen LS, Asmussen L, Braaß H, Wolf S, Luu TV, Stagg CJ, Schulz R, Quandt F, Schwab BC. Differential effects of theta-gamma tACS on motor skill acquisition in young individuals and stroke survivors: A double-blind, randomized, sham-controlled study. Brain Stimul. 2024;17(5):1076–1085. doi: 10.1016/j.brs.2024.09.001. [DOI] [PubMed] [Google Scholar]
- Grønli OK, Rasmussen D, Ingrid A, Per M, Bystad M. A four-month home-based tDCS study on patients with Alzheimer’s disease. Neurocase. 2022;28(3):276–282. doi: 10.1080/13554794.2022.2100710. [DOI] [PubMed] [Google Scholar]
- Grover S, Wen W, Viswanathan V, Gill CT, Reinhart RMG. Long-lasting, dissociable improvements in working memory and long-term memory in older adults with repetitive neuromodulation. Nat Neurosci. 2022;25(9):1237–1246. doi: 10.1038/s41593-022-01132-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guerra A, Asci F, Zampogna A, D’Onofrio V, Berardelli A, Suppa A. The effect of gamma oscillations in boosting primary motor cortex plasticity is greater in young than older adults. Clin Neurophysiol. 2021;132(6):1358–1366. doi: 10.1016/j.clinph.2021.01.032. [DOI] [PubMed] [Google Scholar]
- Guerra A, Colella D, Giangrosso M, Cannavacciuolo A, Paparella G, Fabbrini G, Suppa A, Berardelli A, Bologna M. Driving motor cortex oscillations modulates bradykinesia in Parkinson’s disease. Brain J Neurol. 2022;145(1):224–236. doi: 10.1093/brain/awab257. [DOI] [PubMed] [Google Scholar]
- Guerra A, Pogosyan A, Nowak M, Tan H, Ferreri F, Di Lazzaro V, Brown P. Phase Dependency of the Human Primary Motor Cortex and Cholinergic Inhibition Cancelation During Beta tACS. Cerebral Cortex (New York, NY: 1991) 2016;26(10):3977–3990. doi: 10.1093/cercor/bhw245. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guerra A, Ranieri F, Falato E, Musumeci G, Di Santo A, Asci F, Di Pino G, Suppa A, Berardelli A, Di Lazzaro V. Detecting cortical circuits resonant to high-frequency oscillations in the human primary motor cortex: A TMS-tACS study. Sci Rep. 2020;10(1):7695. doi: 10.1038/s41598-020-64717-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guerra A, Suppa A, Bologna M, D’Onofrio V, Bianchini E, Brown P, Di Lazzaro V, Berardelli A. Boosting the LTP-like plasticity effect of intermittent theta-burst stimulation using gamma transcranial alternating current stimulation. Brain Stimul. 2018;11(4):734–742. doi: 10.1016/j.brs.2018.03.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guidetti M, Arlotti M, Bocci T, Bianchi AM, Parazzini M, Ferrucci R, Priori A. Electric Fields Induced in the Brain by Transcranial Electric Stimulation: A Review of In Vivo Recordings. Biomedicines. 2022;10(10):2333. doi: 10.3390/biomedicines10102333. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guidetti M, Ferrucci R, Vergari M, Aglieco G, Naci A, Versace S, Pacheco-Barrios K, Giannoni-Luza S, Barbieri S, Priori A, Bocci T. Effects of Transcutaneous Spinal Direct Current Stimulation (tsDCS) in Patients With Chronic Pain: A Clinical and Neurophysiological Study. Front Neurol. 2021;12:695910. doi: 10.3389/fneur.2021.695910. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guidetti M, Giannoni-Luza S, Bocci T, Pacheco-Barrios K, Bianchi AM, Parazzini M, Ionta S, Ferrucci R, Maiorana NV, Verde F, Ticozzi N, et al. Modeling Electric Fields in Transcutaneous Spinal Direct Current Stimulation: A Clinical Perspective. Biomedicines. 2023;11(5):1283. doi: 10.3390/biomedicines11051283. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Guleyupoglu B, Schestatsky P, Edwards D, Fregni F, Bikson M. Classification of methods in transcranial electrical stimulation (tES) and evolving strategy from historical approaches to contemporary innovations. J Neurosci Methods. 2013;219(2):297–311. doi: 10.1016/j.jneumeth.2013.07.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Haberbosch L, Datta A, Thomas C, Jooß A, Köhn A, Rönnefarth M, Scholz M, Brandt SA, Schmidt S. Safety aspects, tolerability and modeling of retinofugal alternating current stimulation. Front Neurosci. 2019;13 doi: 10.3389/fnins.2019.00783. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Harrie A, Jamal CE, Padgett M, Rahman-Filipiak A, Hampstead BM. 66 Tolerability of HD-tDCS at Total Amplitudes of 2mA to 10mA in Older Adults. J Int Neuropsychol Soc. 2023;29(s1):473–474. doi: 10.1017/S1355617723006161. [DOI] [Google Scholar]
- Hausman HK, Alexander GE, Cohen R, Marsiske M, DeKosky ST, Hishaw GA, O’Shea A, Kraft JN, Dai Y, Wu S, Woods AJ. Primary outcome from the augmenting cognitive training in older adults study (ACT): A tDCS and cognitive training randomized clinical trial. Brain Stimul. 2023;16(3):904–917. doi: 10.1016/j.brs.2023.05.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hausman HK, Alexander GE, Cohen R, Marsiske M, DeKosky ST, Hishaw GA, O’Shea A, Kraft JN, Dai Y, Wu S, Woods AJ. tDCS reduces depression and state anxiety symptoms in older adults from the augmenting cognitive training in older adults study (ACT) Brain Stimul. 2024;17(2):283–311. doi: 10.1016/j.brs.2024.02.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hayward KS, Brauer SG, Ruddy KL, Lloyd D, Carson RG. Repetitive reaching training combined with transcranial Random Noise Stimulation in stroke survivors with chronic and severe arm paresis is feasible: A pilot, triple-blind, randomised case series. J Neuroeng Rehabil. 2017;14(1):46. doi: 10.1186/s12984-017-0253-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- He J, Tang Y, Lin J, Faulkner G, Tsang HWH, Chan SHW. Non-invasive brain stimulation combined with psychosocial intervention for depression: A systematic review and meta-analysis. BMC Psychiatry. 2022;22(1):273. doi: 10.1186/s12888-022-03843-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Heinitz S, Reinhardt M, Piaggi P, Weise CM, Diaz E, Stinson EJ, Venti C, Votruba SB, Wassermann EM, Alonso-Alonso M, Krakoff J, et al. Neuromodulation directed at the prefrontal cortex of subjects with obesity reduces snack food intake and hunger in a randomized trial†. Am J Clin Nutr. 2017;106(6):1347–1357. doi: 10.3945/ajcn.117.158089. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Helfrich RF, Knepper H, Nolte G, Strüber D, Rach S, Herrmann CS, Schneider TR, Engel AK. Selective modulation of interhemispheric functional connectivity by HD-tACS shapes perception. PLoS Biol. 2014;12(12):e1002031. doi: 10.1371/journal.pbio.1002031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hemmerich K, Lupiáñez J, Martín-Arévalo E. HD-tDCS mitigates the executive vigilance decrement only under high cognitive demands. Sci Rep. 2024;14(1):7865. doi: 10.1038/s41598-024-57917-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Herpich F, Melnick MD, Agosta S, Huxlin KR, Tadin D, Battelli L. Boosting Learning Efficacy with Noninvasive Brain Stimulation in Intact and Brain-Damaged Humans. J Neurosci. 2019;39(28):5551–5561. doi: 10.1523/JNEUROSCI.3248-18.2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Herzog R, Berger TM, Pauly MG, Xue H, Rueckert E, Münchau A, Bäumer T, Weissbach A. Cerebellar transcranial current stimulation—An intraindividual comparison of different techniques. Front Neurosci. 2022;16:987472. doi: 10.3389/fnins.2022.987472. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hodaj H, Payen JF, Hodaj E, Sorel M, Dumolard A, Vercueil L, Delon-Martin C, Lefaucheur JP. Long-term analgesic effect of trans-spinal direct current stimulation compared to non-invasive motor cortex stimulation in complex regional pain syndrome. Brain Commun. 2023;5(4):fcad191. doi: 10.1093/braincomms/fcad191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Holczer A, Vékony T, Klivényi P, Must A. Frontal two-electrode transcranial direct current stimulation protocols may not affect performance on a combined flanker Go/No-Go task. Sci Rep. 2023;13(1):11901. doi: 10.1038/s41598-023-39161-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hommel KA, Hente E, Herzer M, Ingerski LM, Denson LA. Telehealth behavioral treatment for medication nonadherence: A pilot and feasibility study. Eur J Gastroenterol Hepatol. 2013;25(4):469–473. doi: 10.1097/MEG.0b013e32835c2a1b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hsu S-P, Lu C-F, Lin B-F, Tang C-W, Kuo I-J, Tsai Y-A, Guo C-Y, Lee P-L, Shyu K-K, Niddam DM, Lee I-H. Effects of bihemispheric transcranial direct current stimulation on motor recovery in subacute stroke patients: A double-blind, randomized sham-controlled trial. J Neuroeng Rehabil. 2023;20(1):27. doi: 10.1186/s12984-023-01153-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hu Y, Jia Y, Sun Y, Ding Y, Huang Z, Liu C, Wang Y. Efficacy and safety of simultaneous rTMS-tDCS over bilateral angular gyrus on neuropsychiatric symptoms in patients with moderate Alzheimer’s disease: A prospective, randomized, sham-controlled pilot study. Brain Stimul. 2022;15(6):1530–1537. doi: 10.1016/j.brs.2022.11.009. [DOI] [PubMed] [Google Scholar]
- Huang WA, Stitt IM, Negahbani E, Passey DJ, Ahn S, Davey M, Dannhauer M, Doan TT, Hoover AC, Peterchev AV, Radtke-Schuller S, et al. Transcranial alternating current stimulation entrains alpha oscillations by preferential phase synchronization of fast-spiking cortical neurons to stimulation waveform. Nat Commun. 2021;12(1):3151. doi: 10.1038/s41467-021-23021-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Huang Y, Liu AA, Lafon B, Friedman D, Dayan M, Wang X, Bikson M, Doyle WK, Devinsky O, Parra LC. Measurements and models of electric fields in the in vivo human brain during transcranial electric stimulation. eLife. 2017a;6:e18834. doi: 10.7554/eLife.18834. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Huang Y-Z, Lu M-K, Antal A, Classen J, Nitsche M, Ziemann U, Ridding M, Hamada M, Ugawa Y, Jaberzadeh S, Suppa A, et al. Plasticity induced by non-invasive transcranial brain stimulation: A position paper. Clin Neurophysiol. 2017b;128(11):2318–2329. doi: 10.1016/j.clinph.2017.09.007. [DOI] [PubMed] [Google Scholar]
- Hummel FC, Wessel MJ. Non-invasive deep brain stimulation: Interventional targeting of deep brain areas in neurological disorders. Nat Rev Neurol. 2024;20(8):451–452. doi: 10.1038/s41582-024-00990-8. [DOI] [PubMed] [Google Scholar]
- Hunold A, Ortega D, Freitag S, Link D, Antal A, Klee S, Haueisen J. Retinal phosphenes induced by transorbital electrical stimulation: influence of light adaptation, electrode montage, and view direction. Life. 2025;15(5):5. doi: 10.3390/life15050820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Im JJ, Jeong H, Bikson M, Woods AJ, Unal G, Oh JK, Na S, Park J-S, Knotkova H, Song I-U, Chung Y-A. Effects of 6-month at-home transcranial direct current stimulation on cognition and cerebral glucose metabolism in Alzheimer’s disease. Brain Stimul. 2019;12(5):1222–1228. doi: 10.1016/j.brs.2019.06.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Indahlastari A, Kasinadhuni AK, Saar C, Castellano K, Mousa B, Chauhan M, Mareci TH, Sadleir RJ. Methods to compare predicted and observed phosphene experience in tACS subjects. Neural Plast. 2018;2018:8525706. doi: 10.1155/2018/8525706. [DOI] [PMC free article] [PubMed] [Google Scholar]
- International Organization for Standardization. ISO 14971:2019. Medical devices—Application of risk management to medical devices. 2016. https://www.iso.org/obp/ui/en/#iso:std:iso:14971:ed-3:v1:en .
- International Organization for Standardization. Medical devices—Application of risk management to medical devices. ISO; 2019. https://www.iso.org/standard/72704.html . [Google Scholar]
- Jackson MP, Truong D, Brownlow ML, Wagner JA, McKinley RA, Bikson M, Jankord R. Safety parameter considerations of anodal transcranial Direct Current Stimulation in rats. Brain Behav Immun. 2017;64:152–161. doi: 10.1016/j.bbi.2017.04.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jafari E, Alizadehgoradel J, Pourmohseni Koluri F, Nikoozadehkordmirza E, Refahi M, Taherifard M, Nejati V, Hallajian A-H, Ghanavati E, Vicario CM, Nitsche MA, et al. Intensified electrical stimulation targeting lateral and medial prefrontal cortices for the treatment of social anxiety disorder: A randomized, double-blind, parallel-group, dose-comparison study. Brain Stimul. 2021;14(4):974–986. doi: 10.1016/j.brs.2021.06.005. [DOI] [PubMed] [Google Scholar]
- Jiang J, Truong DQ, Esmaeilpour Z, Huang Y, Badran BW, Bikson M. Enhanced tES and tDCS computational models by meninges emulation. J Neural Eng. 2020;17(1):016027. doi: 10.1088/1741-2552/ab549d. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jiang W-L, Cai D-B, Sun C-H, Yin F, Goerigk S, Brunoni AR, Zhao X-W, Mayes TL, Zheng W, Xiang Y-T. Adjunctive tDCS for treatment-refractory auditory hallucinations in schizophrenia: A meta-analysis of randomized, double-blinded, sham-controlled studies. Asian J Psychiatr. 2022a;73:103100. doi: 10.1016/j.ajp.2022.103100. [DOI] [PubMed] [Google Scholar]
- Jiang X, Dahmani S, Bronshteyn M, Yang FN, Ryan JP, Gallagher RC, Damera SR, Kumar PN, Moore DJ, Ellis RJ, Turkeltaub PE. Cingulate transcranial direct current stimulation in adults with HIV. PLoS One. 2022b;17(6):e0269491. doi: 10.1371/journal.pone.0269491. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Johari K, Riccardi N, Malyutina S, Modi M, Desai RH. HD-tDCS of primary and higher-order motor cortex affects action word processing. Front Hum Neurosci. 2022;16:959455. doi: 10.3389/fnhum.2022.959455. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Johnson L, Alekseichuk I, Krieg J, Doyle A, Yu Y, Vitek J, Johnson M, Opitz A. Dose-Dependent Effects of Transcranial Alternating Current Stimulation on Spike Timing in Awake Nonhuman Primates. bioRxiv. 2019:696344. doi: 10.1126/sciadv.aaz2747. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Juras L, Martinčević M, Konstantinović U, Filipović SR, Vranić A, Bjekić J. Multi-session tDCS over the posterior parietal cortex and associative memory. PLoS One. 2025;20(1):e0318593. doi: 10.1371/journal.pone.0318593. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jwa A. Early adopters of the magical thinking cap: A study on do-it-yourself (DIY) transcranial direct current stimulation (tDCS) user community. J Law Biosci. 2015;2(2):292–335. doi: 10.1093/jlb/lsv017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Karimi F, Cassarà AM, Capstick M, Kuster N, Neufeld E. Safety of non-invasive brain stimulation in patients with implants: A computational risk assessment. J Neural Eng. 2025;22(1):016039. doi: 10.1088/1741-2552/ad8efa. [DOI] [PubMed] [Google Scholar]
- Kasten FH, Duecker K, Maack MC, Meiser A, Herrmann CS. Integrating electric field modeling and neuroimaging to explain inter-individual variability of tACS effects. Nat Commun. 2019;10(1):5427. doi: 10.1038/s41467-019-13417-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kasten FH, Negahbani E, Fröhlich F, Herrmann CS. Non-linear transfer characteristics of stimulation and recording hardware account for spurious low-frequency artifacts during amplitude modulated transcranial alternating current stimulation (AM-tACS) Neuroimage. 2018;179:134–143. doi: 10.1016/j.neuroimage.2018.05.068. [DOI] [PubMed] [Google Scholar]
- Kaufmann E, Hordt M, Lauseker M, Palm U, Noachtar S. Acute effects of spaced cathodal transcranial direct current stimulation in drug resistant focal epilepsies. Clin Neurophysiol. 2021;132(7):1444–1451. doi: 10.1016/j.clinph.2021.03.048. [DOI] [PubMed] [Google Scholar]
- Kaye AD, Ridgell S, Alpaugh ES, Mouhaffel A, Kaye AJ, Cornett EM, Chami AA, Shah R, Dixon BM, Viswanath O, Urits I, et al. Peripheral nerve stimulation: A review of techniques and clinical efficacy. Pain Ther. 2021;10(2):961–972. doi: 10.1007/s40122-021-00298-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ke Y, Liu S, Chen L, Wang X, Ming D. Lasting enhancements in neural efficiency by multi-session transcranial direct current stimulation during working memory training. npj Sci Learn. 2023;8(1):1–13. doi: 10.1038/s41539-023-00200-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kessler SK, Minhas P, Woods AJ, Rosen A, Gorman C, Bikson M. Dosage considerations for transcranial direct current stimulation in children: A computational modeling study. PLoS One. 2013;8(9):e76112. doi: 10.1371/journal.pone.0076112. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Khadka N, Bikson M. Role of skin tissue layers and ultra-structure in transcutaneous electrical stimulation including tDCS. Phys Med Biol. 2020;65(22):225018. doi: 10.1088/1361-6560/abb7c1. [DOI] [PubMed] [Google Scholar]
- Khadka N, Bikson M. Neurocapillary-Modulation. Neuromodulation. 2022;25(8):1299–1311. doi: 10.1111/ner.13338. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Khadka N, Borges H, Zannou AL, Jang J, Kim B, Lee K, Bikson M. Dry tDCS: Tolerability of a novel multilayer hydrogel composite non-adhesive electrode for transcranial direct current stimulation. Brain Stimul. 2018;11(5):1044–1053. doi: 10.1016/j.brs.2018.07.049. [DOI] [PubMed] [Google Scholar]
- Khadka N, Poon C, Cancel LM, Tarbell JM, Bikson M. Multi-scale multi-physics model of brain interstitial water flux by transcranial Direct Current Stimulation. J Neural Eng. 2023;20(4) doi: 10.1088/1741-2552/ace4f4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kibret B, Premaratne M, Sullivan C, Thomson RH, Fitzgerald PB. Electroconvulsive therapy (ECT) during pregnancy: Quantifying and assessing the electric field strength inside the foetal brain. Sci Rep. 2018;8(1):4128. doi: 10.1038/s41598-018-22528-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kim J, Park S, Kim H, Roh D, Kim DH. Home-based, Remotely Supervised, 6-Week tDCS in Patients With Both MCI and Depression: A Randomized Double-Blind Placebo-Controlled Trial. Clin EEG Neurosci. 2024;55(5):531–542. doi: 10.1177/15500594231215847. [DOI] [PubMed] [Google Scholar]
- Kitatani R, Koganemaru S, Maeda A, Mikami Y, Matsuhashi M, Mima T, Yamada S. Gait-synchronized oscillatory brain stimulation modulates common neural drives to ankle muscles in patients after stroke: A pilot study. Neurosci Res. 2020;156:256–264. doi: 10.1016/j.neures.2019.11.001. [DOI] [PubMed] [Google Scholar]
- Knopman DS, Laskowitz DT, Koltai DC, Charvet LE, Becker JH, Federman AD, Wisnivesky J, Mahncke H, Van Vleet TM, Bateman L, Kim D-Y, et al. RECOVER-NEURO: Study protocol for a multi-center, multi-arm, phase 2, randomized, active comparator trial evaluating three interventions for cognitive dysfunction in post-acute sequelae of SARS-CoV-2 infection (PASC) Trials. 2024;25(1):326. doi: 10.1186/s13063-024-08156-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Knotkova H, Hamani C, Sivanesan E, Le Beuffe MFE, Moon JY, Cohen SP, Huntoon MA. Neuromodulation for chronic pain. Lancet. 2021;397(10289):2111–2124. doi: 10.1016/S0140-6736(21)00794-7. [DOI] [PubMed] [Google Scholar]
- Ko D-K, Lee H, Kim D-I, Park Y-M, Kang N. Transcranial direct current stimulation improves heart rate variability: A systematic review and meta-analysis. Prog Neuropsychopharmacol Biol Psychiatry. 2024;134:111072. doi: 10.1016/j.pnpbp.2024.111072. [DOI] [PubMed] [Google Scholar]
- Kozlov M, Horner M, Kainz W, Weiskopf N, Möller HE. Modeling radio-frequency energy-induced heating due to the presence of transcranial electric stimulation setup at 3T. Magma (New York NY) 2020;33(6):793–807. doi: 10.1007/s10334-020-00853-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Krause MR, Vieira PG, Csorba BA, Pilly PK, Pack CC. Transcranial alternating current stimulation entrains single-neuron activity in the primate brain. Proc Natl Acad Sci. 2019;116(12):5747–5755. doi: 10.1073/pnas.1815958116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Krebs C, Peter J, Wyss P, Brem A-K, Klöppel S. Transcranial electrical stimulation improves cognitive training effects in healthy elderly adults with low cognitive performance. Clin Neurophysiol. 2021;132(6):1254–1263. doi: 10.1016/j.clinph.2021.01.034. [DOI] [PubMed] [Google Scholar]
- Kreuzer PM, Poeppl TB, Rupprecht R, Vielsmeier V, Lehner A, Langguth B, Schecklmann M. Daily high-frequency transcranial random noise stimulation of bilateral temporal cortex in chronic tinnitus—A pilot study. Sci Rep. 2019;9(1):12274. doi: 10.1038/s41598-019-48686-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kronberg G, Bikson M. Electrode assembly design for transcranial Direct Current Stimulation: A FEM modeling study; Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2012. pp. 891–895. Annual International Conference. [DOI] [PubMed] [Google Scholar]
- Kuck A, Stegeman DF, van Asseldonk EHF. Modeling Trans-Spinal Direct Current Stimulation in the Presence of Spinal Implants. IEEE Transactions on Neural Systems and Rehabilitation Engineering: A Publication of the IEEE Engineering in Medicine and Biology Society. 2019;27(5):790–797. doi: 10.1109/TNSRE.2019.2900377. [DOI] [PubMed] [Google Scholar]
- Kumar S, Marchina S, Langmore S, Massaro J, Palmisano J, Wang N, Searls DE, Lioutas V, Pisegna J, Wagner C, Shinde A, et al. Fostering eating after stroke (FEASt) trial for improving post-stroke dysphagia with non-invasive brain stimulation. Sci Rep. 2022;12(1):9607. doi: 10.1038/s41598-022-14390-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kuo H-I, Bikson M, Datta A, Minhas P, Paulus W, Kuo M-F, Nitsche MA. Comparing cortical plasticity induced by conventional and high-definition 4 × 1 Ring tDCS: A Neurophysiological Study. Brain Stimul. 2013;6(4):644–648. doi: 10.1016/j.brs.2012.09.010. [DOI] [PubMed] [Google Scholar]
- Kurtin DL, Alania K, Rhodes E, Vincent S, Violante IR, Grossman N. Task-related changes in resting state connectivity are affected by temporal interference (TI) stimulation. Brain Stimul. 2025;18(3):937–947. doi: 10.1016/j.brs.2025.04.010. [DOI] [PubMed] [Google Scholar]
- Kurzeck AK, Dechantsreiter E, Wilkening A, Kumpf U, Nenov-Matt T, Padberg F, Palm U. Transcranial Direct Current Stimulation (tDCS) for Depression during Pregnancy: Results from an Open-Label Pilot Study. Brain Sci. 2021;11(7):947. doi: 10.3390/brainsci11070947. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kvašňák E. Perception and pain thresholds of tDCS and tACS. Physiol Res. 2019;68(Suppl 4):S427–S431. doi: 10.33549/physiolres.934381. [DOI] [PubMed] [Google Scholar]
- Laakso I, Mikkonen M, Koyama S, Hirata A, Tanaka S. Can electric fields explain inter-individual variability in transcranial direct current stimulation of the motor cortex? Sci Rep. 2019;9(1):626. doi: 10.1038/s41598-018-37226-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Laakso I, Tanaka S, Koyama S, De Santis V, Hirata A. Inter-subject Variability in Electric Fields of Motor Cortical tDCS. Brain Stimul. 2015;8(5):906–913. doi: 10.1016/j.brs.2015.05.002. [DOI] [PubMed] [Google Scholar]
- Lafon B, Henin S, Huang Y, Friedman D, Melloni L, Thesen T, Doyle W, Buzsáki G, Devinsky O, Parra LC, Liu AA. Low frequency transcranial electrical stimulation does not entrain sleep rhythms measured by human intracranial recordings. Nat Commun. 2017;8(1):1199. doi: 10.1038/s41467-017-01045-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lamoš M, Bočková M, Missey F, Lubrano C, de Araújo E Silva M, Trajlínek J, Studnička O, Daniel P, Carron R, Jirsa V, Chrastina J, et al. Noninvasive Temporal Interference Stimulation of the Subthalamic Nucleus in Parkinson’s Disease Reduces Beta Activity. Movement Disorders: Official Journal of the Movement Disorder Society. 2025 doi: 10.1002/mds.30134. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Laurin A, Nard N, Dalmont M, Bulteau S, Bénard C, Bonnot O, Winer N, Dupont F, Apter G, Terranova-Commessie F, Guillin O, et al. Efficacy and Safety of Transcranial Electric Stimulation during the Perinatal Period: A Systematic Literature Review and Three Case Reports. J Clin Med. 2022;11(14):14. doi: 10.3390/jcm11144048. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lawrence RM, Bridgeford EW, Myers PE, Arvapalli GC, Ramachandran SC, Pisner DA, Frank PF, Lemmer AD, Nikolaidis A, Vogelstein JT. Standardizing human brain parcellations. Sci Data. 2021;8(1):78. doi: 10.1038/s41597-021-00849-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Leaver AM. Perceptual and Cognitive Effects of Focal Transcranial Direct Current Stimulation of Auditory Cortex in Tinnitus. Neuromodulation: Journal of the International Neuromodulation Society. 2025;28(1):136–145. doi: 10.1016/j.neurom.2024.06.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Leduc S. Production du sommeil et de l’anesthésie générale et locale par les courants électriques. CR Acad Sci. 1902;135:199. [Google Scholar]
- Lee J, Lee CW, Jang Y, You JS, Park YS, Ji E, Yu H, Oh S, Ryoo HA, Cho N, Park JY, et al. Efficacy and safety of daily home-based transcranial direct current stimulation as adjunct treatment for bipolar depressive episodes: Double-blind sham-controlled randomized clinical trial. Front Psych. 2022;13:969199. doi: 10.3389/fpsyt.2022.969199. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lefaucheur J-P. Transcranial magnetic stimulation. Handb Clin Neurol. 2019;160:559–580. doi: 10.1016/B978-0-444-64032-1.00037-0. [DOI] [PubMed] [Google Scholar]
- Leffa DT, Grevet EH, Bau CHD, Schneider M, Ferrazza CP, da Silva RF, Miranda MS, Picon F, Teche SP, Sanches P, Pereira D, et al. Transcranial Direct Current Stimulation vs Sham for the Treatment of Inattention in Adults With Attention-Deficit/Hyperactivity Disorder: The TUNED Randomized Clinical Trial. JAMA Psychiat. 2022;79(9):847–856. doi: 10.1001/jamapsychiatry.2022.2055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lench DH, Christopher P, Gavioli M, Nytes G, Ikonomidou C, Villegas MA, Gillick BT. Long-term outcome safety assessment after teleneuromodulation in children with cerebral palsy. Brain Stimul. 2024;17(3):633–635. doi: 10.1016/j.brs.2024.04.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lenoir C, Jankovski A, Mouraux A. Anodal Transcutaneous Spinal Direct Current Stimulation (tsDCS) Selectively Inhibits the Synaptic Efficacy of Nociceptive Transmission at Spinal Cord Level. Neuroscience. 2018;393:150–163. doi: 10.1016/j.neuroscience.2018.10.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li LM, Violante IR, Leech R, Ross E, Hampshire A, Opitz A, Rothwell JC, Carmichael DW, Sharp DJ. Brain state and polarity dependent modulation of brain networks by transcranial direct current stimulation. Hum Brain Mapp. 2019;40(3):904–915. doi: 10.1002/hbm.24420. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Li Z, Schoonjans E, Allaert J, De Smet S, Kappen M, Houfflyn J, Ottaviani C, De Raedt R, Pulopulos MM, Vanderhasselt M-A. Unraveling the temporal interplay of slow-paced breathing and prefrontal transcranial direct current stimulation on cardiac indices of autonomic activity. Psychophysiology. 2024;61(11):e14650. doi: 10.1111/psyp.14650. [DOI] [PubMed] [Google Scholar]
- Liebetanz D, Koch R, Mayenfels S, König F, Paulus W, Nitsche MA. Safety limits of cathodal transcranial direct current stimulation in rats. Clin Neurophysiol. 2009;120(6):1161–1167. doi: 10.1016/j.clinph.2009.01.022. [DOI] [PubMed] [Google Scholar]
- Liu A, Vöröslakos M, Kronberg G, Henin S, Krause MR, Huang Y, Opitz A, Mehta A, Pack CC, Krekelberg B, Berényi A, et al. Immediate neurophysiological effects of transcranial electrical stimulation. Nat Commun. 2018;9(1):5092. doi: 10.1038/s41467-018-07233-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu J, Tong K, Lin Y, Lee VWH, So KF, Shih KC, Lai JSM, Chiu K. Effectiveness of Microcurrent Stimulation in Preserving Retinal Function of Blind Leading Retinal Degeneration and Optic Neuropathy: A Systematic Review. Neuromodulation. 2021;24(6):992–1002. doi: 10.1111/ner.13414. [DOI] [PubMed] [Google Scholar]
- Liu Q, Liu Y, Zhang Y. Effects of cerebellar non-invasive stimulation on neurorehabilitation in stroke patients. An updated systematic review. Biomedicines. 2024;12(6):6. doi: 10.3390/biomedicines12061348. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lo O-Y, Charest S, Margulis H, Lipsitz L, Manor B. Feasibility and safety of sequential transcranial direct current stimulation and physical therapy in older adults at risk of falling: A randomized pilot study. Arch Rehabil Res Clin Transl. 2023;5(4):100288. doi: 10.1016/j.arrct.2023.100288. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lolas F. Brain polarization: Behavioral and therapeutic effects. Biol Psychiatry. 1977;12(1):37–47. [PubMed] [Google Scholar]
- Looi CY, Lim J, Sella F, Lolliot S, Duta M, Avramenko AA, Cohen Kadosh R. Transcranial random noise stimulation and cognitive training to improve learning and cognition of the atypically developing brain: A pilot study. Sci Rep. 2017;7(1):4633. doi: 10.1038/s41598-017-04649-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Louviot S, Tyvaert L, Maillard LG, Colnat-Coulbois S, Dmochowski J, Koessler L. Transcranial Electrical Stimulation generates electric fields in deep human brain structures. Brain Stimul. 2022;15(1):1–12. doi: 10.1016/j.brs.2021.11.001. [DOI] [PubMed] [Google Scholar]
- Lu P, Hanson NJ, Wen L, Guo F, Tian X. Transcranial Direct Current Stimulation Enhances Muscle Strength of Non-dominant Knee in Healthy Young Males. Front Physiol. 2021;12:788719. doi: 10.3389/fphys.2021.788719. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Luckey AM, McLeod SL, Robertson IH, To WT, Vanneste S. Greater Occipital Nerve Stimulation Boosts Associative Memory in Older Individuals: A Randomized Trial. Neurorehabil Neural Repair. 2020;34(11):1020–1029. doi: 10.1177/1545968320943573. [DOI] [PubMed] [Google Scholar]
- Luft CDB, Pereda E, Banissy MJ, Bhattacharya J. Best of both worlds: Promise of combining brain stimulation and brain connectome. Front Syst Neurosci. 2014;8 doi: 10.3389/fnsys.2014.00132. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Luo H, Ye X, Cai H-T, Wang M, Wang Y, Liu Q, Xu Y, Mao Z, Cai Y, Hong J, Zhang C, et al. Frequency-specific and state-dependent neural responses to brain stimulation. Mol Psychiatry. 2025;30(7):2880–2890. doi: 10.1038/s41380-025-02892-7. [DOI] [PubMed] [Google Scholar]
- Luo Y, Yang H, Yan X, Wu Y, Wei G, Wu X, Tian X, Xiong Y, Wu G, Wen H. Transcranial Direct Current Stimulation Alleviates Neurovascular Unit Dysfunction in Mice With Preclinical Alzheimer’s Disease. Front Aging Neurosci. 2022;14:857415. doi: 10.3389/fnagi.2022.857415. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ma Y, Jiao F, Batsikadze G, Yavari F, Nitsche MA. The impact of the left inferior frontal gyrus on fear extinction: A transcranial direct current stimulation study. Brain Stimul. 2024;17(4):816–825. doi: 10.1016/j.brs.2024.07.004. [DOI] [PubMed] [Google Scholar]
- Maas RPPWM, van de Warrenburg BPC, Schutter DJLG. Cathodal skin lesions in a tattoo following transcranial direct current stimulation. Brain Stimul. 2021;14(2):284–286. doi: 10.1016/j.brs.2021.01.011. [DOI] [PubMed] [Google Scholar]
- Madhavan S, Deshmukh S, Cummings M, Doshi A, Rezania K, Freels S, Sawa G. Home-Based Tele-tDCS in Amyotrophic Lateral Sclerosis: Feasibility, Safety, and Preliminary Efficacy. Annals of Clinical and Translational Neurology. 2025a doi: 10.1002/acn3.70038. n/a (n/a) [DOI] [PMC free article] [PubMed] [Google Scholar]
- Madhavan S, Deshmukh S, Cummings M, Doshi A, Rezania K, Freels S, Sawa G. Home-Based Tele-tDCS in Amyotrophic Lateral Sclerosis: Feasibility, Safety, and Preliminary Efficacy. Ann Clin Transl Neurol. 2025;12(5):1022–1033. doi: 10.1002/acn3.70038. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maiella M, Casula EP, Borghi I, Assogna M, D’Acunto A, Pezzopane V, Mencarelli L, Rocchi L, Pellicciari MC, Koch G. Simultaneous transcranial electrical and magnetic stimulation boost gamma oscillations in the dorsolateral prefrontal cortex. Sci Rep. 2022;12(1):19391. doi: 10.1038/s41598-022-23040-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Makovac E, Thayer JF, Ottaviani C. A meta-analysis of non-invasive brain stimulation and autonomic functioning: Implications for brain-heart pathways to cardiovascular disease. Neurosci Biobehav Rev. 2017;74(Pt B):330–341. doi: 10.1016/j.neubiorev.2016.05.001. [DOI] [PubMed] [Google Scholar]
- Malouin-Lachance A, Capolupo J, Laplante C, Hudon A. Does the Digital Therapeutic Alliance Exist? Integrative Review. JMIR Mental Health. 2025;12(1):e69294. doi: 10.2196/69294. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maltezou-Papastylianou C, Russo R, Wallace D, Harmsworth C, Paulmann S. Different stages of emotional prosody processing in healthy ageing-evidence from behavioural responses, ERPs, tDCS, and tRNS. PLoS One. 2022;17(7):e0270934. doi: 10.1371/journal.pone.0270934. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mancini M, Brignani D, Conforto S, Mauri P, Miniussi C, Pellicciari MC. Assessing cortical synchronization during transcranial direct current stimulation: A graph-theoretical analysis. Neuroimage. 2016;140:57–65. doi: 10.1016/j.neuroimage.2016.06.003. [DOI] [PubMed] [Google Scholar]
- Manojlović M, Bjekić J, Purić D, Filipović SR. High-density theta oscillatory-modulated tDCS over the parietal cortex for targeted memory enhancement. Clin Neurophysiol. 2025;170:80–90. doi: 10.1016/j.clinph.2024.12.004. [DOI] [PubMed] [Google Scholar]
- Manor B, Zhou J, Harrison R, Lo O-Y, Travison TG, Hausdorff JM, Pascual-Leone A, Lipsitz L. Transcranial Direct Current Stimulation May Improve Cognitive-Motor Function in Functionally Limited Older Adults. Neurorehabil Neural Repair. 2018;32(9):788–798. doi: 10.1177/1545968318792616. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Manto M, Argyropoulos GPD, Bocci T, Celnik PA, Corben LA, Guidetti M, Koch G, Priori A, Rothwell JC, Sadnicka A, Spampinato D, et al. Consensus Paper: Novel Directions and Next Steps of Non-invasive Brain Stimulation of the Cerebellum in Health and Disease. Cerebellum (London, England) 2022;21(6):1092–1122. doi: 10.1007/s12311-021-01344-6. [DOI] [PubMed] [Google Scholar]
- Marenna S, Huang S-C, Rossi E, Castoldi V, Comi G, Leocani L. Transcranial direct current stimulation as a preventive treatment in multiple sclerosis? Preclinical evidence. Exp Neurol. 2022;357:114201. doi: 10.1016/j.expneurol.2022.114201. [DOI] [PubMed] [Google Scholar]
- Marshall L, Helgadóttir H, Mölle M, Born J. Boosting slow oscillations during sleep potentiates memory. Nature. 2006;444(7119):610–613. doi: 10.1038/nature05278. [DOI] [PubMed] [Google Scholar]
- Martens G, Lejeune N, O’Brien AT, Fregni F, Martial C, Wannez S, Laureys S, Thibaut A. Randomized controlled trial of home-based 4-week tDCS in chronic minimally conscious state. Brain Stimul. 2018;11(5):982–990. doi: 10.1016/j.brs.2018.04.021. [DOI] [PubMed] [Google Scholar]
- Martin DM, Mohan A, Alonzo A, Gates N, Gbadeyan O, Meinzer M, Sachdev P, Brodaty H, Loo C. A pilot double-blind randomized controlled trial of cognitive training combined with transcranial direct current stimulation for amnestic mild cognitive impairment. J Alzheimer’s Disease: JAD. 2019;71(2):503–512. doi: 10.3233/JAD-190306. [DOI] [PubMed] [Google Scholar]
- Matsumoto H, Ugawa Y. Adverse events of tDCS and tACS: A review. Clin Neurophysiol Pract. 2017;2:19–25. doi: 10.1016/j.cnp.2016.12.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mattai A, Miller R, Weisinger B, Greenstein D, Bakalar J, Tossell J, David C, Wassermann EM, Rapoport J, Gogtay N. Tolerability of transcranial direct current stimulation in childhood-onset schizophrenia. Brain Stimul. 2011;4(4):275–280. doi: 10.1016/j.brs.2011.01.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- May ES, Hohn VD, Nickel MM, Tiemann L, Gil Ávila C, Heitmann H, Sauseng P, Ploner M. Modulating Brain Rhythms of Pain Using Transcranial Alternating Current Stimulation (tACS)—A Sham-Controlled Study in Healthy Human Participants. J Pain. 2021;22(10):1256–1272. doi: 10.1016/j.jpain.2021.03.150. [DOI] [PubMed] [Google Scholar]
- Mayur P, Howari R, Byth K, Vannitamby R. Concomitant transcranial direct current stimulation with ultrabrief electroconvulsive therapy: A 2-week double-blind randomized sham-controlled trial. J ECT. 2018;34(4):291–295. doi: 10.1097/YCT.0000000000000479. [DOI] [PubMed] [Google Scholar]
- Medical Devices Agency. Regulations and Approval/Certification of Medical Devices. (n.d.). Retrieved April 30, 2025, from https://www.pmda.go.jp/english/review-services/reviews/0004.html.
- Meiron O, Gale R, Namestnic J, Bennet-Back O, David J, Gebodh N, Adair D, Esmaeilpour Z, Bikson M. High-Definition transcranial direct current stimulation in early onset epileptic encephalopathy: A case study. Brain Inj. 2018;32(1):135–143. doi: 10.1080/02699052.2017.1390254. [DOI] [PubMed] [Google Scholar]
- Mellin JM, Alagapan S, Lustenberger C, Lugo CE, Alexander ML, Gilmore JH, Jarskog LF, Fröhlich F. Randomized trial of transcranial alternating current stimulation for treatment of auditory hallucinations in schizophrenia. Eur Psychiatry. 2018;51:25–33. doi: 10.1016/j.eurpsy.2018.01.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Menassa M, Stronks K, Khatmi F, Roa Díaz ZM, Espinola OP, Gamba M, Itodo OA, Buttia C, Wehrli F, Minder B, Velarde MR, et al. Concepts and definitions of healthy ageing: A systematic review and synthesis of theoretical models. eClinicalMedicine. 2023;56:101821. doi: 10.1016/j.eclinm.2022.101821. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mencarelli L, Menardi A, Neri F, Monti L, Ruffini G, Salvador R, Pascual-Leone A, Momi D, Sprugnoli G, Rossi A, Rossi S, et al. Impact of network-targeted multichannel transcranial direct current stimulation on intrinsic and network-to-network functional connectivity. J Neurosci Res. 2020;98(10):1843–1856. doi: 10.1002/jnr.24690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mercadal B, Salvador R, Biagi MC, Bartolomei F, Wendling F, Ruffini G. Modeling implanted metals in electrical stimulation applications. J Neural Eng. 2022;19(2):026003. doi: 10.1088/1741-2552/ac55ae. [DOI] [PubMed] [Google Scholar]
- Middag-van Spanje M, Nijboer TCW, Schepers J, van Heugten C, Sack AT, Schuhmann T. Alpha transcranial alternating current stimulation as add-on to neglect training: A randomized trial. Brain Commun. 2024;6(5):fcae287. doi: 10.1093/braincomms/fcae287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Milighetti S, Sterzi S, Fregni F, Hanlon CA, Hayley P, Murphy MD, Bundy DT, Nudo RJ, Guggenmos DJ. Effects of tDCS on spontaneous spike activity in a healthy ambulatory rat model. Brain Stimul. 2020;13(6):1566–1576. doi: 10.1016/j.brs.2020.08.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Minhas P, Bansal V, Patel J, Ho JS, Diaz J, Datta A, Bikson M. Electrodes for high-definition transcutaneous DC stimulation for applications in drug delivery and electrotherapy, including tDCS. J Neurosci Methods. 2010;190(2):188–197. doi: 10.1016/j.jneumeth.2010.05.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Minhas P, Bikson M, Woods AJ, Rosen AR, Kessler SK. Transcranial direct current stimulation in pediatric brain: A computational modeling study; Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2012. pp. 859–862. Annual International Conference. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Minhas P, Datta A, Bikson M. Cutaneous perception during tDCS: Role of electrode shape and sponge salinity. Clin Neurophysiol. 2011;122(4):637–638. doi: 10.1016/j.clinph.2010.09.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Minjoli S, Saturnino GB, Blicher JU, Stagg CJ, Siebner HR, Antunes A, Thielscher A. The impact of large structural brain changes in chronic stroke patients on the electric field caused by transcranial brain stimulation. NeuroImage Clinical. 2017;15:106–117. doi: 10.1016/j.nicl.2017.04.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miranda PC, Faria P, Hallett M. What does the ratio of injected current to electrode area tell us about current density in the brain during tDCS? Clin Neurophysiol. 2009;120(6):1183–1187. doi: 10.1016/j.clinph.2009.03.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miranda PC, Mekonnen A, Salvador R, Ruffini G. The electric field in the cortex during transcranial current stimulation. Neuroimage. 2013;70:48–58. doi: 10.1016/j.neuroimage.2012.12.034. [DOI] [PubMed] [Google Scholar]
- Miron J-P, Couture M, Desbeaumes Jodoin V, Bousseau E, Massé-Leblanc C, Bolduc C, Blumberger DM, Datta A, Nitsche MA, Lespérance P. Contact dermatitis following an intensive transcranial direct current stimulation protocol for major depressive disorder. Brain Stimul. 2023;16(5):1273–1275. doi: 10.1016/j.brs.2023.08.019. [DOI] [PubMed] [Google Scholar]
- Misselhorn J, Göschl F, Higgen FL, Hummel FC, Gerloff C, Engel AK. Sensory capability and information integration independently explain the cognitive status of healthy older adults. Sci Rep. 2020;10(1):22437. doi: 10.1038/s41598-020-80069-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Missey F, Acerbo E, Dickey A, Trajlinek J, Studnicka O, Lubrano C, De Araujo E Silva M, Brady E, Vsiansky V, Szabo JP, Dolezalova I, et al. Non-invasive Temporal Interference Stimulation of the Hippocampus Suppresses Epileptic Biomarkers in Patients with Epilepsy: Biophysical Differences between Kilohertz and Amplitude Modulated Stimulation. medRxiv: The Preprint Server for Health Sciences. 2025:2024.12.05.24303799. doi: 10.1016/j.brs.2025.11.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Miyake K, Yoshida M, Inoue Y, Hata Y. Neuroprotective effect of transcorneal electrical stimulation on the acute phase of optic nerve injury. Invest Ophthalmol Vis Sci. 2007;48(5):2356–2361. doi: 10.1167/iovs.06-1329. [DOI] [PubMed] [Google Scholar]
- Moirand R, Imbert L, Haesebaert F, Chesnoy G, Bediou B, Poulet E, Brunelin J. Ten sessions of 30 min tDCS over 5 days to achieve remission in depression: A randomized pilot study. J Clin Med. 2022;11(3):782. doi: 10.3390/jcm11030782. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Moliadze V, Schmanke T, Andreas S, Lyzhko E, Freitag CM, Siniatchkin M. Stimulation intensities of transcranial direct current stimulation have to be adjusted in children and adolescents. Clin Neurophysiol. 2015;126(7):1392–1399. doi: 10.1016/j.clinph.2014.10.142. [DOI] [PubMed] [Google Scholar]
- Monai H, Ohkura M, Tanaka M, Oe Y, Konno A, Hirai H, Mikoshiba K, Itohara S, Nakai J, Iwai Y, Hirase H. Calcium imaging reveals glial involvement in transcranial direct current stimulation-induced plasticity in mouse brain. Nat Commun. 2016;7:11100. doi: 10.1038/ncomms11100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Monastero R, Baschi R, Nicoletti A, Pilati L, Pagano L, Cicero CE, Zappia M, Brighina F. Transcranial random noise stimulation over the primary motor cortex in PD-MCI patients: A crossover, randomized, sham-controlled study. Journal of Neural Transmission (Vienna, Austria: 1996) 2020;127(12):1589–1597. doi: 10.1007/s00702-020-02255-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mondino M, Janin D, Galvao F, Brunelin J. High-frequency transcranial random noise stimulation for auditory hallucinations of schizophrenia: A case series. Biomedicines. 2022;10(11):2698. doi: 10.3390/biomedicines10112698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mondino M, Poulet E, Brunelin J. Moving to accelerated protocols of tDCS in schizophrenia: A case report. Brain Stimul. 2021;14(4):822–824. doi: 10.1016/j.brs.2021.05.006. [DOI] [PubMed] [Google Scholar]
- Mondino M, Thiffault F, Fecteau S. Does non-invasive brain stimulation applied over the dorsolateral prefrontal cortex non-specifically influence mood and emotional processing in healthy individuals? Front Cell Neurosci. 2015;9:399. doi: 10.3389/fncel.2015.00399. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Monte-Silva K, Kuo M-F, Hessenthaler S, Fresnoza S, Liebetanz D, Paulus W, Nitsche MA. Induction of late LTP-like plasticity in the human motor cortex by repeated non-invasive brain stimulation. Brain Stimul. 2013;6(3):424–432. doi: 10.1016/j.brs.2012.04.011. [DOI] [PubMed] [Google Scholar]
- Moret B, Camilleri Gorrieri R, Grassi M, Campana G. Cognitive exergame training and transcranial random noise stimulation effects on executive control in healthy young adults. Neuropsychology. 2021;35(5):568–580. doi: 10.1037/neu0000745. [DOI] [PubMed] [Google Scholar]
- Morimoto T, Miyoshi T, Matsuda S, Tano Y, Fujikado T, Fukuda Y. Transcorneal electrical stimulation rescues axotomized retinal ganglion cells by activating endogenous retinal IGF-1 system. Invest Ophthalmol Vis Sci. 2005;46(6):2147–2155. doi: 10.1167/iovs.04-1339. [DOI] [PubMed] [Google Scholar]
- Mosayebi-Samani M, Jamil A, Salvador R, Ruffini G, Haueisen J, Nitsche MA. The impact of individual electrical fields and anatomical factors on the neurophysiological outcomes of tDCS: A TMS-MEP and MRI study. Brain Stimul. 2021;14(2):316–326. doi: 10.1016/j.brs.2021.01.016. [DOI] [PubMed] [Google Scholar]
- Mrakic-Sposta S, Ferrucci R, Marceglia S, Vezzoli A, Parazzini M, Fiocchi S, Vergari M, Floro S, Deriz M, Priori A. Spinal tDCS and Oxy-inflammation in patients with multiple sclerosis: A pilot study. Brain Stimul. 2025;18(4):1131–1133. doi: 10.1016/j.brs.2025.05.111. [DOI] [PubMed] [Google Scholar]
- Müller D, Habel U, Brodkin ES, Weidler C. High-definition transcranial direct current stimulation (HD-tDCS) for the enhancement of working memory—A systematic review and meta-analysis of healthy adults. Brain Stimul. 2022;15(6):1475–1485. doi: 10.1016/j.brs.2022.11.001. [DOI] [PubMed] [Google Scholar]
- Murphy OW, Hoy KE, Wong D, Bailey NW, Fitzgerald PB, Segrave RA. Transcranial random noise stimulation is more effective than transcranial direct current stimulation for enhancing working memory in healthy individuals: Behavioural and electrophysiological evidence. Brain Stimul. 2020;13(5):1370–1380. doi: 10.1016/j.brs.2020.07.001. [DOI] [PubMed] [Google Scholar]
- Murugaraja V, Shivakumar V, Sivakumar P, Sinha P, Venkatasubramanian G. Clinical utility and tolerability of transcranial direct current stimulation in mild cognitive impairment. Asian J Psychiatr. 2017;30:135–140. doi: 10.1016/j.ajp.2017.09.001. [DOI] [PubMed] [Google Scholar]
- Nakazono H, Ogata K, Takeda A, Yamada E, Oka S, Tobimatsu S. A specific phase of transcranial alternating current stimulation at the β frequency boosts repetitive paired-pulse TMS-induced plasticity. Sci Rep. 2021;11(1):13179. doi: 10.1038/s41598-021-92768-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Natividade GR, de Araujo C, Fitz RC, Brietzke E, Schestatsky P, Gerchman F. Psychiatric profile and quality of life of subjects with excess weight treated with transcranial direct current stimulation combined with a hypocaloric diet. Nutr Neurosci. 2021;24(12):919–926. doi: 10.1080/1028415X.2019.1693319. [DOI] [PubMed] [Google Scholar]
- Neri F, Della Toffola J, Scoccia A, Benelli A, Lomi F, Cinti A, Smeralda CL, Romanella S, Giannotta A, Rossi S, Santarnecchi E. Neuromodulation via tRNS accelerates learning and enhances in-game performance at a virtual-reality first person shooter game. Comput Hum Behav. 2025;165 doi: 10.1016/j.chb.2024.108537. [DOI] [Google Scholar]
- Neuling T, Ruhnau P, Fuscà M, Demarchi G, Herrmann CS, Weisz N. Friends, not foes: Magnetoencephalography as a tool to uncover brain dynamics during transcranial alternating current stimulation. Neuroimage. 2015;118:406–413. doi: 10.1016/j.neuroimage.2015.06.026. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nikolin S, Alonzo A, Martin D, Gálvez V, Buten S, Taylor R, Goldstein J, Oxley C, Hadzi-Pavlovic D, Loo CK. Transcranial random noise stimulation for the acute treatment of depression: A randomized controlled trial. Int J Neuropsychopharmacol. 2020;23(3):146–156. doi: 10.1093/ijnp/pyz072. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nikolin S, Huggins C, Martin D, Alonzo A, Loo CK. Safety of repeated sessions of transcranial direct current stimulation: A systematic review. Brain Stimul. 2018;11(2):278–288. doi: 10.1016/j.brs.2017.10.020. [DOI] [PubMed] [Google Scholar]
- Ninot G, Agier S, Bacon S, Berr C, Boulze-Launay I, Bourrel G, Carbonnel F, Clément V, David M, Gerazime A, Gomez A, et al. La Plateforme CEPS: Une structure universitaire de réflexion sur l’évaluation des interventions non médicamenteuses (INM) Hegel. 2017;1(1):53–56. doi: 10.3917/heg.071.0053. [DOI] [Google Scholar]
- Nitsche MA, Doemkes S, Karaköse T, Antal A, Liebetanz D, Lang N, Tergau F, Paulus W. Shaping the effects of transcranial direct current stimulation of the human motor cortex. J Neurophysiol. 2007;97(4):3109–3117. doi: 10.1152/jn.01312.2006. [DOI] [PubMed] [Google Scholar]
- Nitsche MA, Niehaus L, Hoffmann KT, Hengst S, Liebetanz D, Paulus W, Meyer B-U. MRI study of human brain exposed to weak direct current stimulation of the frontal cortex. Clin Neurophysiol. 2004;115(10):2419–2423. doi: 10.1016/j.clinph.2004.05.001. [DOI] [PubMed] [Google Scholar]
- Nitsche MA, Paulus W. Excitability changes induced in the human motor cortex by weak transcranial direct current stimulation. J Physiol. 2000;527(3):633–639. doi: 10.1111/j.1469-7793.2000.t01-1-00633.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nitsche MA, Paulus W. Sustained excitability elevations induced by transcranial DC motor cortex stimulation in humans. Neurology. 2001;57(10):1899–1901. doi: 10.1212/wnl.57.10.1899. [DOI] [PubMed] [Google Scholar]
- Nitsche MA, Paulus W. Noninvasive brain stimulation protocols in the treatment of epilepsy: Current state and perspectives. Neurotherapeutics. 2009;6(2):244–250. doi: 10.1016/j.nurt.2009.01.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Noetscher GM, Yanamadala J, Makarov SN, Pascual-Leone A. Comparison of cephalic and extracephalic montages for transcranial direct current stimulation—A numerical study. IEEE Trans Biomed Eng. 2014;61(9):2488–2498. doi: 10.1109/TBME.2014.2322774. [DOI] [PubMed] [Google Scholar]
- Noury N, Hipp JF, Siegel M. Physiological processes non-linearly affect electrophysiological recordings during transcranial electric stimulation. Neuroimage. 2016;140:99–109. doi: 10.1016/j.neuroimage.2016.03.065. [DOI] [PubMed] [Google Scholar]
- Ogata K, Nakazono H, Ikeda T, Oka S, Goto Y, Tobimatsu S. After-effects of intermittent theta-burst stimulation are differentially and phase-dependently suppressed by α- and β-frequency transcranial alternating current stimulation. Front Hum Neurosci. 2021;15 doi: 10.3389/fnhum.2021.750329. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Omae E, Shima A, Tanaka K, Yamada M, Cao Y, Nakamura T, Hoshiai H, Chiba Y, Irisawa H, Mizushima T, Mima T, et al. Case report: An N-of-1 study using amplitude modulated transcranial alternating current stimulation between Broca’s area and the right homotopic area to improve post-stroke aphasia with increased inter-regional synchrony. Front Hum Neurosci. 2024;18 doi: 10.3389/fnhum.2024.1297683. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Opitz A, Falchier A, Yan C-G, Yeagle EM, Linn GS, Megevand P, Thielscher A, Deborah A, Milham MP, Mehta AD, Schroeder CE. Spatiotemporal structure of intracranial electric fields induced by transcranial electric stimulation in humans and nonhuman primates. Sci Rep. 2016;6(1):31236. doi: 10.1038/srep31236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Opitz A, Paulus W, Will S, Antunes A, Thielscher A. Determinants of the electric field during transcranial direct current stimulation. Neuroimage. 2015;109:140–150. doi: 10.1016/j.neuroimage.2015.01.033. [DOI] [PubMed] [Google Scholar]
- Opitz A, Yeagle E, Thielscher A, Schroeder C, Mehta AD, Milham MP. On the importance of precise electrode placement for targeted transcranial electric stimulation. Neuroimage. 2018;181:560–567. doi: 10.1016/j.neuroimage.2018.07.027. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pal N, Maire R, Stephan MA, Herrmann FR, Benninger DH. Transcranial direct current stimulation for the treatment of chronic tinnitus: A randomized controlled study. Brain Stimul. 2015;8(6):1101–1107. doi: 10.1016/j.brs.2015.06.014. [DOI] [PubMed] [Google Scholar]
- Palm U, Kumpf U, Behler N, Wulf L, Kirsch B, Wörsching J, Keeser D, Hasan A, Padberg F. Home use, remotely supervised, and remotely controlled transcranial direct current stimulation: A systematic review of the available evidence. Neuromodulation. 2018;21(4):323–333. doi: 10.1111/ner.12686. [DOI] [PubMed] [Google Scholar]
- Paneri B, Adair D, Thomas C, Khadka N, Patel V, Tyler WJ, Parra L, Bikson M. Tolerability of repeated application of transcranial electrical stimulation with limited outputs to healthy subjects. Brain Stimul. 2016;9(5):740–754. doi: 10.1016/j.brs.2016.05.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Parazzini M, Rossi E, Ferrucci R, Fiocchi S, Liorni I, Priori A, Ravazzani P. Computational model of cerebellar transcranial direct current stimulation; Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2013. pp. 237–240. Annual International Conference. [DOI] [PubMed] [Google Scholar]
- Pegado R, Melo M, Oehmen T, Kirsztajn GM, Silva-Filho E, Quintiliano A. Safety and feasibility of transcranial direct current stimulation in end-stage renal disease patients undergoing hemodialysis: An exploratory study. J Bras Nefrol. 2024;46(4):e20240010. doi: 10.1590/2175-8239-JBN-2024-0010en. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pereira HC, Sousa D, Simões M, Martins R, Amaral C, Lopes V, Crisóstomo J, Castelo-Branco M. Effects of anodal multichannel transcranial direct current stimulation (tDCS) on social-cognitive performance in healthy subjects: A randomized sham-controlled crossover pilot study. Prog Brain Res. 2021;264:259–286. doi: 10.1016/bs.pbr.2021.04.004. [DOI] [PubMed] [Google Scholar]
- Perera MPN, Bailey NW, Murphy OW, Mallawaarachchi S, Sullivan C, Hill AT, Fitzgerald PB. Home-based individualized alpha transcranial alternating current stimulation improves symptoms of obsessive-compulsive disorder: preliminary evidence from a randomized. Sham-Controlled Clinical Trial. Depression Anxiety. 2023;2023:9958884. doi: 10.1155/2023/9958884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Peterchev AV, Wagner TA, Miranda PC, Nitsche MA, Paulus W, Lisanby SH, Pascual-Leone A, Bikson M. Fundamentals of transcranial electric and magnetic stimulation dose: Definition, selection, and reporting practices. Brain Stimul. 2012;5(4):435–453. doi: 10.1016/j.brs.2011.10.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Peters D, Tran N, Adam T. In: Implementation research in health: A practical guide. Peters David, et al., editors. World Health Organization; 2013. Alliance for Health Policy and Systems Research, & World Health Organization. https://iris.who.int/handle/10665/91758 . [Google Scholar]
- Pikhovych A, Stolberg NP, Jessica Flitsch L, Walter HL, Graf R, Fink GR, Schroeter M, Rueger MA. Transcranial direct current stimulation modulates neurogenesis and microglia activation in the mouse brain. Stem Cells Int. 2016;2016:2715196. doi: 10.1155/2016/2715196. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pilloni G, Cho H, Tian TE, Beringer J, Bikson M, Charvet L. Immediate and differential response to emotional stimuli associated with transcranial direct current stimulation for depression: A visual-search task pilot study. Neuromodulation. 2024;27(4):759–765. doi: 10.1016/j.neurom.2023.07.006. [DOI] [PubMed] [Google Scholar]
- Pilloni G, Choi C, Coghe G, Cocco E, Krupp LB, Pau M, Charvet LE. Gait and functional mobility in multiple sclerosis: immediate effects of transcranial direct current stimulation (tDCS) paired with aerobic exercise. Front Neurol. 2020a;11:310. doi: 10.3389/fneur.2020.00310. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pilloni G, Choi C, Shaw MT, Coghe G, Krupp L, Moffat M, Cocco E, Pau M, Charvet L. Walking in multiple sclerosis improves with tDCS: A randomized, double-blind, sham-controlled study. Ann Clin Transl Neurol. 2020b;7(11):2310–2319. doi: 10.1002/acn3.51224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pilloni G, Pehel S, Ko T, Sammarco C, Charlson RE, Hanlon CA, Charvet L. Telehealth tDCS to reduce cannabis use: A pilot RCT in multiple sclerosis as a framework for generalized use. Drug Alcohol Depend. 2025;272:112706. doi: 10.1016/j.drugalcdep.2025.112706. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pilloni G, Vogel-Eyny A, Lustberg M, Best P, Malik M, Walton-Masters L, George A, Mirza I, Zhovtis L, Datta A, Bikson M, et al. Tolerability and feasibility of at-home remotely supervised transcranial direct current stimulation (RS-tDCS): Single-center evidence from 6,779 sessions. Brain Stimul. 2022;15(3):707–716. doi: 10.1016/j.brs.2022.04.014. [DOI] [PubMed] [Google Scholar]
- Pilloni G, Woods AJ, Charvet L. No risk of skin lesion or burn with transcranial direct current stimulation (tDCS) using standardized protocols. Brain Stimul. 2021;14(3):511–512. doi: 10.1016/j.brs.2021.03.006. [DOI] [PubMed] [Google Scholar]
- Pinto EF, Gupta A, Kulkarni GB, Andrade C. A randomized, double-blind, sham-controlled study of transcranial direct current stimulation as an augmentation intervention for the attenuation of motor deficits in patients with stroke. J ECT. 2021;37(4):281–290. doi: 10.1097/YCT.0000000000000769. [DOI] [PubMed] [Google Scholar]
- Pinto TP, Inaćio JC, de Aguiar E, Ferreira AS, Sudo FK, Tovar-Moll F, Rodrigues EC. Prefrontal tDCS modulates autonomic responses in COVID-19 inpatients. Brain Stimul. 2023;16(2):657–666. doi: 10.1016/j.brs.2023.03.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ploumitsakou M, Beanato E, Windel F, Nencha U, Maceira-Elvira P, di Natale M, Vassiliadis P, Léger B, Wessel M, Nouri A, Schaller K, et al. Effects of striatal transcranial temporal interference stimulation on motor learning after a traumatic brain injury. Brain Stimul. 2025;18(1):402–403. doi: 10.1016/j.brs.2024.12.559. [DOI] [Google Scholar]
- Poreisz C, Boros K, Antal A, Paulus W. Safety aspects of transcranial direct current stimulation concerning healthy subjects and patients. Brain Res Bull. 2007;72(4–6):208–214. doi: 10.1016/j.brainresbull.2007.01.004. [DOI] [PubMed] [Google Scholar]
- Pozdniakov I, Vorobiova AN, Galli G, Rossi S, Feurra M. Online and offline effects of transcranial alternating current stimulation of the primary motor cortex. Sci Rep. 2021;11(1):3854. doi: 10.1038/s41598-021-83449-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Prim JH, Ahn S, Davila MI, Alexander ML, McCulloch KL, Fröhlich F. Targeting the autonomic nervous system balance in patients with chronic low back pain using transcranial alternating current stimulation: A randomized, crossover, double-blind, placebo-controlled pilot study. J Pain Res. 2019;12:3265–3277. doi: 10.2147/JPR.S208030. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Priori A. Brain polarization in humans: A reappraisal of an old tool for prolonged non-invasive modulation of brain excitability. Clin Neurophysiol. 2003;114(4):589–595. doi: 10.1016/s1388-2457(02)00437-6. [DOI] [PubMed] [Google Scholar]
- Priori A, Berardelli A, Rona S, Accornero N, Manfredi M. Polarization of the human motor cortex through the scalp. Neuroreport. 1998;9(10):2257–2260. doi: 10.1097/00001756-199807130-00020. [DOI] [PubMed] [Google Scholar]
- Priori A, Ciocca M, Parazzini M, Vergari M, Ferrucci R. Transcranial cerebellar direct current stimulation and transcutaneous spinal cord direct current stimulation as innovative tools for neuroscientists. J Physiol. 2014;592(16):3345–3369. doi: 10.1113/jphysiol.2013.270280. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Proctor EK, Landsverk J, Aarons G, Chambers D, Glisson C, Mittman B. Implementation research in mental health services: An emerging science with conceptual, methodological, and training challenges. Adm Policy Ment Health. 2009;36(1):24–34. doi: 10.1007/s10488-008-0197-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pruvost-Robieux E, Benzakoun J, Turc G, Marchi A, Mancusi RL, Lamy C, Domigo V, Oppenheim C, Calvet D, Baron J-C, Mas J-L, et al. Cathodal transcranial direct current stimulation in acute ischemic stroke: pilot randomized controlled trial. Stroke. 2021;52(6):1951–1960. doi: 10.1161/STROKEAHA.120.032056. [DOI] [PubMed] [Google Scholar]
- Puonti O, Saturnino GB, Madsen KH, Thielscher A. Value and limitations of intracranial recordings for validating electric field modeling for transcranial brain stimulation. Neuroimage. 2020a;208:116431. doi: 10.1016/j.neuroimage.2019.116431. [DOI] [PubMed] [Google Scholar]
- Puonti O, Van Leemput K, Saturnino GB, Siebner HR, Madsen KH, Thielscher A. Accurate and robust whole-head segmentation from magnetic resonance images for individualized head modeling. Neuroimage. 2020b;219:117044. doi: 10.1016/j.neuroimage.2020.117044. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Qurat-Ul-Ain Ahmad Z, Ilyas S, Ishtiaq S, Tariq I, Nawaz Malik A, Liu T, Wang J. Comparison of a single session of tDCS on cerebellum vs. motor cortex in stroke patients: A randomized sham-controlled trial. Ann Med. 2023;55(2):2252439. doi: 10.1080/07853890.2023.2252439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Raco V, Bauer R, Norim S, Gharabaghi A. Cumulative effects of single TMS pulses during beta-tACS are stimulation intensity-dependent. Brain Stimul. 2017;10(6):1055–1060. doi: 10.1016/j.brs.2017.07.009. [DOI] [PubMed] [Google Scholar]
- Raco V, Bauer R, Olenik M, Brkic D, Gharabaghi A. Neurosensory effects of transcranial alternating current stimulation. Brain Stimul. 2014;7(6):823–831. doi: 10.1016/j.brs.2014.08.005. [DOI] [PubMed] [Google Scholar]
- Raco V, Bauer R, Tharsan S, Gharabaghi A. Combining TMS and tACS for closed-loop phase-dependent modulation of corticospinal excitability: A feasibility study. Front Cell Neurosci. 2016;10 doi: 10.3389/fncel.2016.00143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Radman T, Ramos RL, Brumberg JC, Bikson M. Role of cortical cell type and morphology in subthreshold and suprathreshold uniform electric field stimulation in vitro. Brain Stimulation. 2009;2(4):215–228.:228.e1-3. doi: 10.1016/j.brs.2009.03.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rajji TK, Bowie CR, Herrmann N, Pollock BG, Lanctôt KL, Kumar S, Flint AJ, Mah L, Fischer CE, Butters MA, Bikson M, et al. Slowing cognitive decline in major depressive disorder and mild cognitive impairment: A randomized clinical trial. JAMA Psychiat. 2025;82(1):12–21. doi: 10.1001/jamapsychiatry.2024.3241. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rathour RK, Kaphzan H. Dendritic effects of tDCS insights from a morphologically realistic model neuron. iScience. 2024;27(3):109230. doi: 10.1016/j.isci.2024.109230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Räty S, Borrmann C, Granata G, Cárdenas-Morales L, Schoenfeld A, Sailer M, Silvennoinen K, Holopainen J, De Rossi F, Antal A, Rossini PM, et al. Non-invasive electrical brain stimulation for vision restoration after stroke: An exploratory randomized trial (REVIS) Restor Neurol Neurosci. 2021;39(3):221–235. doi: 10.3233/RNN-211198. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Razza LB, De Smet S, Van Hoornweder S, De Witte S, Luethi MS, Baeken C, Brunoni AR, Vanderhasselt M-A. Investigating the variability of prefrontal tDCS effects on working memory: An individual E-field distribution study. Cortex. 2024;172:38–48. doi: 10.1016/j.cortex.2023.10.025. [DOI] [PubMed] [Google Scholar]
- Reato D, Rahman A, Bikson M, Parra LC. Effects of weak transcranial alternating current stimulation on brain activity-a review of known mechanisms from animal studies. Front Hum Neurosci. 2013;7:687. doi: 10.3389/fnhum.2013.00687. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Reckow J, Rahman-Filipiak A, Garcia S, Schlaefflin S, Calhoun O, DaSilva AF, Bikson M, Hampstead BM. Tolerability and blinding of 4×1 high-definition transcranial direct current stimulation (HD-tDCS) at two and three milliamps. Brain Stimul. 2018;11(5):991–997. doi: 10.1016/j.brs.2018.04.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Reinhart RMG, Nguyen JA. Working memory revived in older adults by synchronizing rhythmic brain circuits. Nat Neurosci. 2019;22(5):820–827. doi: 10.1038/s41593-019-0371-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rethwilm R, Schecklmann M, Agboada D, Ole Bergmann T, Seiberl W. Commentary: Do all studies using medical devices fall under the European Medical Device Regulation? Clin. Neurophysiol. 2024;167:239–240. doi: 10.1016/j.clinph.2024.09.012. [DOI] [PubMed] [Google Scholar]
- Richardson JD, Galletta EE, Charvet L, Shaw M. Feasibility of Remotely Supervised Transcranial Direct Current Stimulation (RS-tDCS) for People with Stroke-Induced and Progressive Aphasia. Aphasiology. 2023;37(7):1039–1063. doi: 10.1080/02687038.2022.2076279. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Riemann S, Mittelstädt M, Glatzki M, Zilges C, Wolff C, Niemann F, Roheger M, Floël A, Grothe M, Meinzer M. Information processing speed modulation by electrical brain stimulation in multiple sclerosis: Towards individually tailored protocols. Brain Commun. 2025;7(3):fcaf223. doi: 10.1093/braincomms/fcaf223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Riggall K, Forlini C, Carter A, Hall W, Weier M, Partridge B, Meinzer M. Researchers’ perspectives on scientific and ethical issues with transcranial direct current stimulation: An international survey. Sci Rep. 2015;5:10618. doi: 10.1038/srep10618. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Robinovitch LG. Sommeil électrique (inhibition des mouvements volontaires et de la sensibilité) par des courants électriques de basse tension et à interruptions modérément fréquentes, épilepsie électrique et électrocution. 1906. https://gallica.bnf.fr/ark:/12148/bpt6k6103187n .
- Rocke M, Fromm AE, Jansen N, Thams F, Trujillo-Llano C, Grittner U, Antonenko D, Floël A. Feasibility and pilot efficacy of self-applied home-based cognitive training and brain stimulation: A randomized-controlled trial. Brain Stimul. 2024;17(6):1241–1243. doi: 10.1016/j.brs.2024.10.009. [DOI] [PubMed] [Google Scholar]
- Roncero C, Mardigyan V, Service E, Singerman J, Whittaker KC, Friedman M, Chertkow H. Investigation into the effect of transcranial direct current stimulation on cardiac pacemakers. Brain Stimul. 2020;13(1):89–95. doi: 10.1016/j.brs.2019.08.010. [DOI] [PubMed] [Google Scholar]
- Rossi E, Marenna S, Castoldi V, Comi G, Leocani L. Transcranial direct current stimulation as a potential remyelinating therapy: Visual evoked potentials recovery in cuprizone demyelination. Exp Neurol. 2024;382:114972. doi: 10.1016/j.expneurol.2024.114972. [DOI] [PubMed] [Google Scholar]
- Rueger MA, Keuters MH, Walberer M, Braun R, Klein R, Sparing R, Fink GR, Graf R, Schroeter M. Multi-session transcranial direct current stimulation (tDCS) elicits inflammatory and regenerative processes in the rat brain. PLoS One. 2012;7(8):e43776. doi: 10.1371/journal.pone.0043776. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruf SP, Fallgatter AJ, Plewnia C. Augmentation of working memory training by transcranial direct current stimulation (tDCS) Sci Rep. 2017;7(1):876. doi: 10.1038/s41598-017-01055-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruffini G, Fox MD, Ripolles O, Miranda PC, Pascual-Leone A. Optimization of multifocal transcranial current stimulation for weighted cortical pattern targeting from realistic modeling of electric fields. Neuroimage. 2014;89:216–225. doi: 10.1016/j.neuroimage.2013.12.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruffini G, Salvador R, Castaldo F, Baleeiro T, Camprodon JA, Chopra M, Cappon D, Pascual-Leone A. Multichannel tDCS with advanced targeting for major depressive disorder: A tele-supervised at-home pilot study. Front Psych. 2024;15 doi: 10.3389/fpsyt.2024.1427365. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruffini G, Wendling F, Sanchez-Todo R, Santarnecchi E. Targeting brain networks with multichannel transcranial current stimulation (tCS) Curr Opin Biomed Eng. 2018;8:70–77. doi: 10.1016/j.cobme.2018.11.001. [DOI] [Google Scholar]
- Ruhnau P, Neuling T, Fuscá M, Herrmann CS, Demarchi G, Weisz N. Eyes wide shut: Transcranial alternating current stimulation drives alpha rhythm in a state dependent manner. Sci Rep. 2016;6:27138. doi: 10.1038/srep27138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruhnau P, Rufener KS, Heinze H-J, Zaehle T. Sailing in a sea of disbelief: In vivo measurements of transcranial electric stimulation in human subcortical structures. Brain Stimul. 2018;11(1):241–243. doi: 10.1016/j.brs.2017.09.015. [DOI] [PubMed] [Google Scholar]
- Rumpf J-J, Barbu A, Fricke C, Wegscheider M, Classen J. Posttraining alpha transcranial alternating current stimulation impairs motor consolidation in elderly people. Neural Plast. 2019;2019:2689790. doi: 10.1155/2019/2689790. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ruohonen J, Karhu J. tDCS possibly stimulates glial cells. Clin Neurophysiol. 2012;123(10):2006–2009. doi: 10.1016/j.clinph.2012.02.082. [DOI] [PubMed] [Google Scholar]
- Sabel BA, Kresinsky A, Cárdenas-Morales L, Haueisen J, Hunold A, Dannhauer M, Antal A. Evaluating current density modeling of non-invasive eye and brain electrical stimulation using phosphene thresholds. IEEE Trans Neural Syst Rehabil Eng. 2021;29:2133–2141. doi: 10.1109/TNSRE.2021.3120148. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sadeghihassanabadi F, Misselhorn J, Gerloff C, Zittel S. Optimizing the montage for cerebellar transcranial alternating current stimulation (tACS): A combined computational and experimental study. J Neural Eng. 2022;19(2) doi: 10.1088/1741-2552/ac676f. [DOI] [PubMed] [Google Scholar]
- Salehinejad MA, Abdi M, Dadashi M, Hallajian A-H, Sharifi K, Khadem A, Salvador R, Ruffini G, Nitsche MA. Optimized multichannel tDCS protocol for clinical use in patients with major depressive disorder. medRxiv. 2025:2025.02.12.25322157. doi: 10.1101/2025.02.12.25322157. [DOI] [Google Scholar]
- Salehinejad MA, Ghanavati E, Glinski B, Hallajian A-H, Azarkolah A. A systematic review of randomized controlled trials on efficacy and safety of transcranial direct current stimulation in major neurodevelopmental disorders: ADHD, autism, and dyslexia. Brain Behav. 2022;12(9):e2724. doi: 10.1002/brb3.2724. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Salehinejad MA, Nejati V, Mosayebi-Samani M, Mohammadi A, Wischnewski M, Kuo M-F, Avenanti A, Vicario CM, Nitsche MA. Transcranial direct current stimulation in ADHD: A systematic review of efficacy, safety, and protocol-induced electrical field modeling results. Neurosci Bull. 2020;36(10):1191–1212. doi: 10.1007/s12264-020-00501-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Salimpour Y, Liu C-C, Webber WR, Mills KA, Anderson WS. Subdural recordings from an awake human brain for measuring current intensity during transcranial direct current stimulation; Annual International Conference of the IEEE Engineering in Medicine and Biology Society; 2017. pp. 1110–1113. Annual International Conference. [DOI] [PubMed] [Google Scholar]
- Sampaio-Junior B, Tortella G, Borrione L, Moffa AH, Machado-Vieira R, Cretaz E, Fernandes da Silva A, Fraguas R, Aparício LV, Klein I, Lafer B, et al. Efficacy and Safety of Transcranial Direct Current Stimulation as an Add-on Treatment for Bipolar Depression: A Randomized Clinical Trial. JAMA Psychiat. 2018;75(2):158–166. doi: 10.1001/jamapsychiatry.2017.4040. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sánchez-León CA, Sánchez-López Á, Gómez-Climent MA, Cordones I, Cohen Kadosh R, Márquez-Ruiz J. Impact of chronic transcranial random noise stimulation (tRNS) on GABAergic and glutamatergic activity markers in the prefrontal cortex of juvenile mice. Prog Brain Res. 2021;264:323–341. doi: 10.1016/bs.pbr.2021.01.017. [DOI] [PubMed] [Google Scholar]
- Santana K, França E, Sato J, Silva A, Queiroz M, de Farias J, Rodrigues D, Souza I, Ribeiro V, Caparelli-Dáquer E, Teixeira AL, et al. Non-invasive brain stimulation for fatigue in post-acute sequelae of SARS-CoV-2 (PASC) Brain Stimul. 2023;16(1):100–107. doi: 10.1016/j.brs.2023.01.1672. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Santarnecchi E, Feurra M, Barneschi F, Acampa M, Bianco G, Cioncoloni D, Rossi A, Rossi S. Time course of corticospinal excitability and autonomic function interplay during and following monopolar tDCS. Front Psych. 2014;5:86. doi: 10.3389/fpsyt.2014.00086. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Saturnino GB, Antunes A, Thielscher A. On the importance of electrode parameters for shaping electric field patterns generated by tDCS. Neuroimage. 2015;120:25–35. doi: 10.1016/j.neuroimage.2015.06.067. [DOI] [PubMed] [Google Scholar]
- Saturnino GB, Siebner HR, Thielscher A, Madsen KH. Accessibility of cortical regions to focal TES: dependence on spatial position, safety, and practical constraints. Neuroimage. 2019;203:116183. doi: 10.1016/j.neuroimage.2019.116183. [DOI] [PubMed] [Google Scholar]
- Schaefer A, Kong R, Gordon EM, Laumann TO, Zuo XN, Holmes AJ, Eickhoff SB, Yeo BTT. Local-global parcellation of the human cerebral cortex from intrinsic functional connectivity MRI. Cereb Cortex. 2018;28(9):3095–3114. doi: 10.1093/cercor/bhx179. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schaper FLWVJ, Nordberg J, Cohen AL, Lin C, Hsu J, Horn A, Ferguson MA, Siddiqi SH, Drew W, Soussand L, Winkler AM, et al. Mapping lesion-related epilepsy to a human brain network. JAMA Neurol. 2023;80(9):891–902. doi: 10.1001/jamaneurol.2023.1988. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schilberg L, Engelen T, Ten Oever S, Schuhmann T, de Gelder B, de Graaf TA, Sack AT. Phase of beta-frequency tACS over primary motor cortex modulates corticospinal excitability. Cortex. 2018;103:142–152. doi: 10.1016/j.cortex.2018.03.001. [DOI] [PubMed] [Google Scholar]
- Schittkowski M, Pohlner J, Mercieca K, Grohmann C, Kröger L, Prokosch V, Lorenz K, Beck A, Haueisen J, Hunold A, Bernhard M, et al. Vision Restoration through transorbital electrical stimulation in Optic Neuropathy in patients with significant optic atrophy due to primary open-angle glaucoma—a randomised, controlled, double-blind, multicentre clinical trial: The VIRON study protocol. 2025 doi: 10.1136/bmjopen-2024-091705. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schlaug G, Cassarly C, Feld JA, Wolf SL, Rowe VT, Fritz S, Chhatbar PY, Shinde A, Su Z, Broderick JP, Zorowitz R, et al. Safety and efficacy of transcranial direct current stimulation in addition to constraint-induced movement therapy for post-stroke motor recovery (TRANSPORT2): A phase 2, multicentre, randomised, sham-controlled triple-blind trial. Lancet Neurol. 2025;24(5):400–412. doi: 10.1016/S1474-4422(25)00044-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schroeder PA, Ehlis AC, Wolkenstein L, Fallgatter AJ, Plewnia C. Emotional distraction and bodily reaction: modulation of autonomous responses by anodal tDCS to the prefrontal cortex. Front Cell Neurosci. 2015;9:482. doi: 10.3389/fncel.2015.00482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schuhmann T, Duecker F, Middag-van Spanje M, Gallotto S, van Heugten C, Schrijnemaekers AC, van Oostenbrugge R, Sack AT. Transcranial alternating brain stimulation at alpha frequency reduces hemispatial neglect symptoms in stroke patients. Int J Clin Health Psychol: IJCHP. 2022;22(3):100326. doi: 10.1016/j.ijchp.2022.100326. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schwippel T, Pupillo F, Feldman Z, Walker C, Townsend L, Rubinow D, Frohlich F. Closed-loop transcranial alternating current stimulation for the treatment of major depressive disorder: an open-label pilot study. Am J Psychiatry. 2024;181(9):842–845. doi: 10.1176/appi.ajp.20230838. [DOI] [PubMed] [Google Scholar]
- Sekhon M, Cartwright M, Francis JJ. Acceptability of healthcare interventions: An overview of reviews and development of a theoretical framework. BMC Health Serv Res. 2017;17(1):88. doi: 10.1186/s12913-017-2031-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sethi A, Pascual-Leone A, Santarnecchi E, Almalki G, Krishnan C. Transcranial random noise stimulation to augment hand function in individuals with moderate-to-severe stroke: A pilot randomized clinical trial. Restor Neurol Neurosci. 2023;41(5–6):193–202. doi: 10.3233/RNN-231314. [DOI] [PubMed] [Google Scholar]
- Shan Y, Wang H, Yang Y, Wang J, Zhao W, Huang Y, Wang H, Han B, Pan N, Jin X, Fan X, et al. Evidence of a large current of transcranial alternating current stimulation directly to deep brain regions. Mol Psychiatry. 2023;28(12):5402–5410. doi: 10.1038/s41380-023-02150-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Shaw M, Pilloni G, Charvet L. Delivering transcranial direct current stimulation away from clinic: remotely supervised tDCS. Mil Med. 2020;185(Suppl 1):319–325. doi: 10.1093/milmed/usz348. [DOI] [PubMed] [Google Scholar]
- Sheffield JG, Ramerpresad S, Brem A-K, Mansfield K, Orhan U, Dillard M, McKanna J, Plessow F, Thompson T, Santarnecchi E, Pascual-Leone A, et al. Blinding efficacy and adverse events following repeated transcranial alternating current, direct current, and random noise stimulation. Cortex. 2022;154:77–88. doi: 10.1016/j.cortex.2022.05.015. [DOI] [PubMed] [Google Scholar]
- Shenoy S, Bose A, Chhabra H, Dinakaran D, Agarwal SM, Shivakumar V, Narayanaswamy JC, Sivakumar PT, Venkatasubramanian G. Transcranial direct current stimulation (tDCS) for auditory verbal hallucinations in schizophrenia during pregnancy: A case report. Brain Stimul. 2015;8(1):163–164. doi: 10.1016/j.brs.2014.10.013. [DOI] [PubMed] [Google Scholar]
- Shpektor A, Nazarova M, Feurra M. Effects of transcranial alternating current stimulation on the primary motor cortex by online combined approach with transcranial magnetic stimulation. J Visual Exp: JoVE. 2017;127:55839. doi: 10.3791/55839. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sierawska A, Moliadze V, Splittgerber M, Rogge A, Siniatchkin M, Buyx A. First epileptic seizure and initial diagnosis of juvenile myoclonus epilepsy (JME) in a transcranial direct current stimulation (tDCS) study–Ethical analysis of a clinical case. Neuroethics. 2020;13(3):347–351. doi: 10.1007/s12152-020-09444-6. [DOI] [Google Scholar]
- Sierawska A, Prehn-Kristensen A, Moliadze V, Krauel K, Nowak R, Freitag CM, Siniatchkin M, Buyx A. Unmet Needs in Children With Attention Deficit Hyperactivity Disorder—Can Transcranial Direct Current Stimulation Fill the Gap? Promises and Ethical Challenges. Front Psych. 2019;10 doi: 10.3389/fpsyt.2019.00334. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Silva-Filho E, Pilloni G, Charvet LE, Fregni F, Brunoni AR, Bikson M. Factors supporting availability of home-based neuromodulation using remote supervision in middle-income countries; Brazil experience. Brain Stimul. 2022;15(2):385–387. doi: 10.1016/j.brs.2022.02.005. [DOI] [PubMed] [Google Scholar]
- Simpson EA, Saiote C, Sutter E, Lench DH, Ikonomidou C, Villegas MA, Gillick BT. Remotely monitored transcranial direct current stimulation in pediatric cerebral palsy: open label trial protocol. BMC Pediatr. 2022;22(1):566. doi: 10.1186/s12887-022-03612-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Simula S, Daoud M, Ruffini G, Biagi MC, Bénar C-G, Benquet P, Wendling F, Bartolomei F. Transcranial current stimulation in epilepsy: A systematic review of the fundamental and clinical aspects. Front Neurosci. 2022;16:909421. doi: 10.3389/fnins.2022.909421. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Simula S, Makhalova J, Pizzo F, Garnier E, Damiani G, Mercadal B, Chiara Biagi M, Salvador R, Medina-Villalon S, Ruffini G, Wendling F, et al. Impact of transcranial electrical stimulation on simultaneous stereoelectroencephalography recordings: A randomized sham-controlled study. Clin Neurophysiol. 2024;166:211–222. doi: 10.1016/j.clinph.2024.08.003. [DOI] [PubMed] [Google Scholar]
- Smeele SJ, Adhia DB, De Ridder D. Feasibility and safety of high-definition infraslow pink noise stimulation for treating chronic tinnitus-A randomized placebo-controlled trial. Neuromodulation. 2023;26(4):801–816. doi: 10.1016/j.neurom.2022.10.049. [DOI] [PubMed] [Google Scholar]
- Solomons CD, Shanmugasundaram V. Transcranial direct current stimulation: A review of electrode characteristics and materials. Med Eng Phys. 2020;85:63–74. doi: 10.1016/j.medengphy.2020.09.015. [DOI] [PubMed] [Google Scholar]
- Soterix Medical Inc. tDCS+ Monitoring Solutions – Transcranial Direct Current Stimulation. 2025. https://soterixmedical.com/research/monitoring/eeg .
- Splittgerber M, Japaridze N, Sierawska A, Gimenez S, Nowak R, Siniatchkin M, Moliadze V. First generalized tonic clonic seizure in the context of pediatric tDCS - A case report. Neurophysiol Clin. 2020c doi: 10.1016/j.neucli.2019.11.002. [DOI] [PubMed] [Google Scholar]
- Splittgerber M, Salvador R, Brauer H, Breitling-Ziegler C, Prehn-Kristensen A, Krauel K, Nowak R, Ruffini G, Moliadze V, Siniatchkin M. Individual baseline performance and electrode montage impact on the effects of anodal tDCS over the left dorsolateral prefrontal cortex. Front Hum Neurosci. 2020a;14:349. doi: 10.3389/fnhum.2020.00349. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Splittgerber M, Suwelack JH, Kadish NE, Moliadze V. The effects of 1 mA tACS and tRNS on children/adolescents and adults: investigating age and sensitivity to sham stimulation. Neural Plast. 2020b;2020:8896423. doi: 10.1155/2020/8896423. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sprugnoli G, Monti L, Lippa L, Neri F, Mencarelli L, Ruffini G, Salvador R, Oliveri G, Batani B, Momi D, Cerase A, et al. Reduction of intratumoral brain perfusion by noninvasive transcranial electrical stimulation. Sci Adv. 2019;5(8):eaau9309. doi: 10.1126/sciadv.aau9309. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sreeraj VS, Bose A, Shanbhag V, Narayanaswamy JC, Venkatasubramanian G, Benegal V. Monotherapy With tDCS for treatment of depressive episode during pregnancy: A case report. Brain Stimul. 2016;9(3):457–458. doi: 10.1016/j.brs.2016.03.007. [DOI] [PubMed] [Google Scholar]
- Steenbergen L, Sellaro R, Hommel B, Lindenberger U, Kühn S, Colzato LS. “Unfocus” on foc.us: Commercial tDCS headset impairs working memory. Exp Brain Res. 2016;234(3):637–643. doi: 10.1007/s00221-015-4391-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stett A, Schatz A, Gekeler F, Franklin J. Transcorneal electrical stimulation dose-dependently slows the visual field loss in retinitis pigmentosa. Transl Vis Sci Technol. 2023;12(2):29. doi: 10.1167/tvst.12.2.29. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Strube W, Kirsch B, Padberg F, Hasan A, Palm U. Transcranial direct current stimulation as monotherapy for the treatment of auditory hallucinations during pregnancy: A case report. J Clin Psychopharmacol. 2016;36(5):534. doi: 10.1097/JCP.0000000000000554. [DOI] [PubMed] [Google Scholar]
- Sudbrack-Oliveira P, Barbosa MZ, Thome-Souza S, Razza LB, Gallucci-Neto J, da Costa Lane Valiengo L, Brunoni AR. Transcranial direct current stimulation (tDCS) in the management of epilepsy: A systematic review. Seizure. 2021;86:85–95. doi: 10.1016/j.seizure.2021.01.020. [DOI] [PubMed] [Google Scholar]
- Tagami Y, Kurimoto T, Miyoshi T, Morimoto T, Sawai H, Mimura O. Axonal regeneration induced by repetitive electrical stimulation of crushed optic nerve in adult rats. Jpn J Ophthalmol. 2009;53(3):257–266. doi: 10.1007/s10384-009-0657-8. [DOI] [PubMed] [Google Scholar]
- Takahashi K, Glinski B, Salehinejad MA, Jamil A, Chang A-Y-C, Kuo MF, Nitsche MA. Induction and stabilization of delta frequency brain oscillations by phase-synchronized rTMS and tACS. Brain Stimul. 2024;17(5):1086–1097. doi: 10.1016/j.brs.2024.09.003. [DOI] [PubMed] [Google Scholar]
- Talsma LJ, Kroese HA, Slagter HA. Boosting cognition: effects of multiple-session transcranial direct current stimulation on working memory. J Cogn Neurosci. 2017;29(4):755–768. doi: 10.1162/jocn_a_01077. [DOI] [PubMed] [Google Scholar]
- Tatti E, Phillips AL, Paciorek R, Romanella SM, Dettore D, Di Lorenzo G, Ruffini G, Rossi S, Santarnecchi E. Boosting psychological change: Combining non-invasive brain stimulation with psychotherapy. Neurosci Biobehav Rev. 2022;142:104867. doi: 10.1016/j.neubiorev.2022.104867. [DOI] [PubMed] [Google Scholar]
- Tavakoli AV, Yun K. Transcranial alternating current stimulation (tACS) mechanisms and protocols. Front Cell Neurosci. 2017;11:214. doi: 10.3389/fncel.2017.00214. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Teixeira-Santos AC, Moreira CS, Pereira DR, Pinal D, Fregni F, Leite J, Carvalho S, Sampaio A. Working memory training coupled with transcranial direct current stimulation in older adults: A randomized controlled experiment. Front Aging Neurosci. 2022;14 doi: 10.3389/fnagi.2022.827188. [DOI] [PMC free article] [PubMed] [Google Scholar]
- The National Medical Products Administration. Rules for Classification of Medical Devices. n.d. Retrieved April 30, 2025, from https://english.nmpa.gov.cn/2019-10/11/c_415411.htm.
- Thomas F, Steinberg F, Pixa NH, Berger A, Cheng M-Y, Doppelmayr M. Prefrontal high definition cathodal tDCS modulates executive functions only when coupled with moderate aerobic exercise in healthy persons. Sci Rep. 2021;11(1):8457. doi: 10.1038/s41598-021-87914-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tokikuni Y, Watanabe A, Nakazono H, Miura H, Saito R, Miaowen D, Fuyama K, Takahashi K, Okada K, Sugawara K, Tohyama H, et al. Differing effectiveness of transcranial random noise stimulation and transcranial direct current stimulation for enhancing working memory in healthy individuals: A randomized controlled trial. J Neuroeng Rehabil. 2024;21(1):180. doi: 10.1186/s12984-024-01481-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tremblay S, Hannah R, Rawji V, Rothwell JC. Modulation of iTBS after-effects via concurrent directional TDCS: A proof of principle study. Brain Stimul. 2017;10(4):744–747. doi: 10.1016/j.brs.2017.03.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Trevizol AP, Shiozawa P, Albuquerque Sato I, da Silva ME, de Barros Calfat EL, Alberto RL, Cook IA, Cordeiro Q. Trigeminal nerve stimulation (TNS) for post-traumatic stress disorder: A case study. Brain Stimul. 2015;8(3):676–678. doi: 10.1016/j.brs.2015.02.008. [DOI] [PubMed] [Google Scholar]
- Tseng P-T, Chen Y-W, Zeng B-Y, Zeng B-S, Hung C-M, Sun C-K, Cheng Y-S, Stubbs B, Carvalho AF, Brunoni AR, Su K-P, et al. The beneficial effect on cognition of noninvasive brain stimulation intervention in patients with dementia: A network meta-analysis of randomized controlled trials. Alzheimer’s Res Ther. 2023;15(1):20. doi: 10.1186/s13195-023-01164-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tsui CT, Lal P, Fox KVR, Churchward MA, Todd KG. The effects of electrical stimulation on glial cell behaviour. BMC Biomed Eng. 2022;4(1):7. doi: 10.1186/s42490-022-00064-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Turi Z, Ambrus GG, Ho KA, Sengupta T, Paulus W, Antal A. When size matters: Large electrodes induce greater stimulation-related cutaneous discomfort than smaller electrodes at equivalent current density. Brain Stimul. 2014;7(3):460–467. doi: 10.1016/j.brs.2014.01.059. [DOI] [PubMed] [Google Scholar]
- Turi Z, Csifcsák G, Boayue NM, Aslaksen P, Antal A, Paulus W, Groot J, Hawkins GE, Forstmann B, Opitz A, Thielscher A, et al. Blinding is compromised for transcranial direct current stimulation at 1 mA for 20 min in young healthy adults. Eur J Neurosci. 2019;50(8):3261–3268. doi: 10.1111/ejn.14403. [DOI] [PubMed] [Google Scholar]
- Turski CA, Kessler-Jones A, Chow C, Hermann B, Hsu D, Jones J, Seeger SK, Chappell R, Boly M, Ikonomidou C. Extended Multiple-Field High-Definition transcranial direct current stimulation (HD-tDCS) is well tolerated and safe in healthy adults. Restor Neurol Neurosci. 2017;35(6):631–642. doi: 10.3233/RNN-170757. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Valiengo LCL, Goerigk S, Gordon PC, Padberg F, Serpa MH, Koebe S, Santos LAD, Lovera RAM, de Carvalho JB, van de Bilt M, Lacerda ALT, et al. Efficacy and safety of transcranial direct current stimulation for treating negative symptoms in Schizophrenia: A randomized clinical trial. JAMA Psychiat. 2020;77(2):121–129. doi: 10.1001/jamapsychiatry.2019.3199. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Valiengo LCL, Goulart AC, de Oliveira JF, Benseñor IM, Lotufo PA, Brunoni AR. Transcranial direct current stimulation for the treatment of post-stroke depression: Results from a randomised, sham-controlled, double-blinded trial. J Neurol Neurosurg Psychiatry. 2017;88(2):170–175. doi: 10.1136/jnnp-2016-314075. [DOI] [PubMed] [Google Scholar]
- Van Hoornweder S, Caulfield AK, Nitsche M, Thielscher A, Meesen LJ. Addressing transcranial electrical stimulation variability through prospective individualized dosing of electric field strength in 300 participants across two samples: The 2-SPED approach. J Neural Eng. 2022;19(5):056045. doi: 10.1088/1741-2552/ac9a78. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vandermeeren Y, Jamart J, Ossemann M. Effect of tDCS with an extracephalic reference electrode on cardio-respiratory and autonomic functions. BMC Neurosci. 2010;11(1):38. doi: 10.1186/1471-2202-11-38. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vasquez A, Fregni F. In: Transcranial Direct Current Stimulation in Neuropsychiatric Disorders: Clinical Principles and Management. 2nd. Brunoni A, Nitsche M, Loo C, editors. Springer International Publishing; 2021. Regulatory Aspects; pp. 383–392. [DOI] [Google Scholar]
- Vassiliadis P, Stiennon E, Windel F, Wessel MJ, Beanato E, Hummel FC. Safety, tolerability and blinding efficiency of non-invasive deep transcranial temporal interference stimulation: First experience from more than 250 sessions. J Neural Eng. 2024;21(2):024001. doi: 10.1088/1741-2552/ad2d32. [DOI] [PubMed] [Google Scholar]
- Vigod SN, Murphy KE, Dennis CL, Oberlander TF, Ray JG, Daskalakis ZJ, Blumberger DM. Transcranial direct current stimulation (tDCS) for depression in pregnancy: A pilot randomized controlled trial. Brain Stimul. 2019;12(6):1475–1483. doi: 10.1016/j.brs.2019.06.019. [DOI] [PubMed] [Google Scholar]
- Violante IR, Alania K, Cassarà AM, Neufeld E, Acerbo E, Carron R, Williamson A, Kurtin DL, Rhodes E, Hampshire A, Kuster N, et al. Non-invasive temporal interference electrical stimulation of the human hippocampus. Nat Neurosci. 2023;26(11):1994–2004. doi: 10.1038/s41593-023-01456-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Voarino N, Dubljević V, Racine E. tDCS for memory enhancement: analysis of the speculative aspects of ethical issues. Front Hum Neurosci. 2016;10:678. doi: 10.3389/fnhum.2016.00678. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vogelmann U, Pilloni G, Brunoni AR, Charvet L. How can we develop transcranial direct current stimulation into an effective at-home treatment tool for depression? Expert Rev Med Devices. 2024;21(10):883–886. doi: 10.1080/17434440.2024.2409767. [DOI] [PubMed] [Google Scholar]
- Vogelmann U, Stadler M, Soldini A, Chang K-Y, Chen M, Bulubas L, Dechantsreiter E, Plewnia C, Fallgatter A, Langguth B, Normann C, et al. A comparative analysis of technical data: At-home vs. in-clinic application of transcranial direct current stimulation in depression. Brain Stimul. 2025;18(4):1097–1105. doi: 10.1016/j.brs.2025.05.103. [DOI] [PubMed] [Google Scholar]
- Vöröslakos M, Takeuchi Y, Brinyiczki K, Zombori T, Oliva A, Fernández-Ruiz A, Kozák G, Kincses ZT, Iványi B, Buzsáki G, Berényi A. Direct effects of transcranial electric stimulation on brain circuits in rats and humans. Nat Commun. 2018;9(1):483. doi: 10.1038/s41467-018-02928-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vulić K, Bjekić J, Paunović D, Jovanović M, Milanović S, Filipović SR. Theta-modulated oscillatory transcranial direct current stimulation over posterior parietal cortex improves associative memory. Sci Rep. 2021;11(1):3013. doi: 10.1038/s41598-021-82577-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wandrey JD, Kastelik J, Fritzsche T, Denke C, Schäfer M, Tafelski S. Supplementing transcranial direct current stimulation to local infiltration series for refractory neuropathic craniocephalic pain: A randomized controlled pilot trial. Front Neurol. 2023;14:1069434. doi: 10.3389/fneur.2023.1069434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang HE, Scholly J, Triebkorn P, Sip V, Medina Villalon S, Woodman MM, Le Troter A, Guye M, Bartolomei F, Jirsa V. VEP atlas: An anatomic and functional human brain atlas dedicated to epilepsy patients. J Neurosci Methods. 2021;348:108983. doi: 10.1016/j.jneumeth.2020.108983. [DOI] [PubMed] [Google Scholar]
- Wang H, Wang K, Xue Q, Peng M, Yin L, Gu X, Leng H, Lu J, Liu H, Wang D, Xiao J, et al. Transcranial alternating current stimulation for treating depression: A randomized controlled trial. Brain J Neurol. 2022;145(1):83–91. doi: 10.1093/brain/awab252. [DOI] [PubMed] [Google Scholar]
- Wang H, Wang X, Jin J, Zhang W, Li Y, Liu Z, Yin T. Simultaneous stimulation using rTMS and tDCS produces the most effective modulation of motor cortical excitability in healthy subjects: A pilot study. Neurosci Lett. 2019;694:46–50. doi: 10.1016/j.neulet.2018.10.041. [DOI] [PubMed] [Google Scholar]
- Wang J, Yao X, Ji Y, Li H. Cognitive potency and safety of tDCS treatment for major depressive disorder: A systematic review and meta-analysis. Front Hum Neurosci. 2024;18 doi: 10.3389/fnhum.2024.1458295. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wang Y, Monai H. Transcranial direct current stimulation alters cerebrospinal fluid-interstitial fluid exchange in mouse brain. Brain Stimul. 2024;17(3):620–632. doi: 10.1016/j.brs.2024.04.009. [DOI] [PubMed] [Google Scholar]
- Warthen KG, Walker NC, Wicklund BD, Gonzalez MM, Ramirez N, Gee SC, Al-Dasouqi H, Madore MR. Neuromodulation of the cerebellum for motor applications: A systematic review. J Integr Neurosci. 2024;23(10):195. doi: 10.31083/j.jin2310195. [DOI] [PubMed] [Google Scholar]
- Wei J, Alamia A, Yao Z, Huang G, Li L, Liang Z, Zhang L, Zhou C, Song Z, Zhang Z. State-Dependent tACS effects reveal the potential causal role of prestimulus alpha traveling waves in visual contrast detection. J Neurosci. 2024;44(27):e2023232024. doi: 10.1523/JNEUROSCI.2023-23.2024. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Weise K, Wartman WA, Knösche TR, Nummenmaa AR, Makarov SN. The effect of meninges on the electric fields in TES and TMS. Numerical modeling with adaptive mesh refinement. Brain Stimul. 2022;15(3):654–663. doi: 10.1016/j.brs.2022.04.009. [DOI] [PubMed] [Google Scholar]
- Weller S, Nitsche MA, Plewnia C. Enhancing cognitive control training with transcranial direct current stimulation: A systematic parameter study. Brain Stimul. 2020;13(5):1358–1369. doi: 10.1016/j.brs.2020.07.006. [DOI] [PubMed] [Google Scholar]
- Wessel MJ, Beanato E, Popa T, Windel F, Vassiliadis P, Menoud P, Beliaeva V, Violante IR, Abderrahmane H, Dzialecka P, Park C-H, et al. Noninvasive theta-burst stimulation of the human striatum enhances striatal activity and motor skill learning. Nat Neurosci. 2023;26(11):2005–2016. doi: 10.1038/s41593-023-01457-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wexler A. A pragmatic analysis of the regulation of consumer transcranial direct current stimulation (TDCS) devices in the United States. J Law Biosci. 2015;2(3):669–696. doi: 10.1093/jlb/lsv039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wexler A. The practices of do-it-yourself brain stimulation: Implications for ethical considerations and regulatory proposals. J Med Ethics. 2016;42(4):211–215. doi: 10.1136/medethics-2015-102704. [DOI] [PubMed] [Google Scholar]
- Wexler A. In: Developments in Neuroethics and Bioethics. Bárd I, Hildt E, editors. Vol. 3. Academic Press; 2020. Chapter Five—Do-it-yourself and direct-to-consumer neurostimulation; pp. 127–155. [DOI] [Google Scholar]
- Wexler A, Reiner PB. The Routledge handbook of neuroethics. Routledge/Taylor & Francis Group; 2018. Home Use of tDCS: From “Do-It-Yourself” to “Direct-to-Consumer; pp. 271–284. [DOI] [Google Scholar]
- Wilkening A, Kurzeck A, Dechantsreiter E, Padberg F, Palm U. Transcranial alternating current stimulation for the treatment of major depression during pregnancy. Psychiatry Res. 2019;279:399–400. doi: 10.1016/j.psychres.2019.06.009. [DOI] [PubMed] [Google Scholar]
- Wischnewski M, Alekseichuk I, Opitz A. Neurocognitive, physiological, and biophysical effects of transcranial alternating current stimulation. Trends Cogn Sci. 2023;27(2):189–205. doi: 10.1016/j.tics.2022.11.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Wivatvongvana P, Soonthornthum C, Kitisak K. Correction: Intermittent theta-burst stimulation combined with transcranial direct current stimulation once weekly for treatment-resistant depression: a case report. J Med Case Reports. 2024;18(1):203. doi: 10.1186/s13256-024-04520-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woodham RD, Selvaraj S, Lajmi N, Hobday H, Sheehan G, Ghazi-Noori A-R, Lagerberg PJ, Rizvi M, Kwon SS, Orhii P, Maislin D, et al. Home-based transcranial direct current stimulation treatment for major depressive disorder: A fully remote phase 2 randomized sham-controlled trial. Nat Med. 2024;31(1):87–95. doi: 10.1038/s41591-024-03305-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woodham RD, Selvaraj S, Lajmi N, Hobday H, Sheehan G, Ghazi-Noori A-R, Lagerberg PJ, Rizvi M, Kwon SS, Orhii P, Maislin D, et al. Home-based transcranial direct current stimulation treatment for major depressive disorder: A fully remote phase 2 randomized sham-controlled trial. Nat Med. 2025;31(1):87–95. doi: 10.1038/s41591-024-03305-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woods AJ, Antal A, Bikson M, Boggio PS, Brunoni AR, Celnik P, Cohen LG, Fregni F, Herrmann CS, Kappenman ES, Knotkova H, et al. A technical guide to tDCS, and related non-invasive brain stimulation tools. Clin Neurophysiol. 2016;127(2):1031–1048. doi: 10.1016/j.clinph.2015.11.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Woods AJ, Cohen R, Marsiske M, Alexander GE, Czaja SJ, Wu S. Augmenting cognitive training in older adults (The ACT Study): Design and Methods of a Phase III tDCS and cognitive training trial. Contemp Clin Trials. 2018;65:19–32. doi: 10.1016/j.cct.2017.11.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- World Health Organization. WHO’s work on the UN Decade of Healthy Ageing (2021-2030) 2021. https://www.who.int/initiatives/decade-of-healthy-ageing .
- Wu D, Zhu Y, Wang Y, Liu N, Zhang P. Transcranial direct current stimulation of the prefrontal and visual cortices diversely affects early and late perceptual learning. Brain Behav. 2024;14(7):e3620. doi: 10.1002/brb3.3620. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Xia Y, Khalid W, Yin Z, Huang G, Bikson M, Fu BM. Modulation of solute diffusivity in brain tissue as a novel mechanism of transcranial direct current stimulation (tDCS) Sci Rep. 2020;10(1):18488. doi: 10.1038/s41598-020-75460-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Xia Y, Li Y, Khalid W, Bikson M, Fu BM. Direct Current Stimulation Disrupts Endothelial Glycocalyx and Tight Junctions of the Blood-Brain Barrier in vitro. Front Cell Dev Biol. 2021;9:731028. doi: 10.3389/fcell.2021.731028. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yang C-L, Qu Y, Huang J-P, Wang T-T, Zhang H, Chen Y, Tan Y-C. Efficacy and safety of transcranial direct current stimulation in the treatment of fibromyalgia: A systematic review and meta-analysis. Neurophysiologie Clinique = Clin Neurophysiol. 2024;54(1):102944. doi: 10.1016/j.neucli.2024.102944. [DOI] [PubMed] [Google Scholar]
- Yang T, Zhang J, Wang B, Zhang W, Xu M, Yang S, Liu H. Electrical stimulation to treat tinnitus: A meta-analysis and systemic review of randomized controlled trials. Therap Adv Chronic Dis. 2021;12:20406223211041069. doi: 10.1177/20406223211041069. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ybran Inc. YBRAIN. (n.d.). Retrieved April 30, 2025, from https://www.ybrain.com/en/
- Yin H, Yin H, Zhang W, Miao Q, Qin Z, Guo S, Fu Q, Ma J, Wu F, Yin J, Yang Y, et al. Transcorneal electrical stimulation promotes survival of retinal ganglion cells after optic nerve transection in rats accompanied by reduced microglial activation and TNF-α expression. Brain Res. 2016;1650:10–20. doi: 10.1016/j.brainres.2016.08.034. [DOI] [PubMed] [Google Scholar]
- Yoon M-J, Park HJ, Yoo YJ, Oh HM, Im S, Kim T-W, Lim SH. Electric field simulation and appropriate electrode positioning for optimized transcranial direct current stimulation of stroke patients: An in Silico model. Sci Rep. 2024;14(1):2850. doi: 10.1038/s41598-024-52874-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yu J, Lam CLM, Man ISC, Shao R, Lee TMC. Multi-session anodal prefrontal transcranial direct current stimulation does not improve executive functions among older adults. J Int Neuropsychol Soc: JINS. 2020;26(4):372–381. doi: 10.1017/S1355617719001206. [DOI] [PubMed] [Google Scholar]
- Yuan B, Tolomeo S, Yang C, Wang Y, Yu R. tDCS effect on prosocial behavior: A meta-analytic review. Soc Cogn Affect Neurosci. 2022;17(1):26–42. doi: 10.1093/scan/nsab067. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zaehle T, Beretta M, Jäncke L, Herrmann CS, Sandmann P. Excitability changes induced in the human auditory cortex by transcranial direct current stimulation: Direct electrophysiological evidence. Exp Brain Res. 2011;215(2):135–140. doi: 10.1007/s00221-011-2879-5. [DOI] [PubMed] [Google Scholar]
- Zaehle T, Rach S, Herrmann CS. Transcranial alternating current stimulation enhances individual alpha activity in human EEG. PLoS One. 2010;5(11):e13766. doi: 10.1371/journal.pone.0013766. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zanardi R, Poletti S, Prestifilippo D, Attanasio F, Barbini B, Colombo C. Transcranial direct current stimulation: A novel approach in the treatment of vascular depression. Brain Stimul. 2020;13(6):1559–1565. doi: 10.1016/j.brs.2020.08.013. [DOI] [PubMed] [Google Scholar]
- Zeng F-G, Tran P, Richardson M, Sun S, Xu Y. Human sensation of transcranial electric stimulation. Sci Rep. 2019;9(1):15247. doi: 10.1038/s41598-019-51792-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhang D-W, Moraidis A, Klingberg T. Individually tuned theta HD-tACS improves spatial performance. Brain Stimul. 2022a;15(6):1439–1447. doi: 10.1016/j.brs.2022.10.009. [DOI] [PubMed] [Google Scholar]
- Zhang K, Guo L, Zhang J, An G, Zhou Y, Lin J, Xing J, Lu M, Ding G. A safety study of 500 μA cathodal transcranial direct current stimulation in rat. BMC Neurosci. 2019;20(1):40. doi: 10.1186/s12868-019-0523-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhang M, Force RB, Walker C, Ahn S, Jarskog LF, Frohlich F. Alpha transcranial alternating current stimulation reduces depressive symptoms in people with schizophrenia and auditory hallucinations: A double-blind, randomized pilot clinical trial. Schizophrenia. 2022b;8(1):114. doi: 10.1038/s41537-022-00321-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhao H, Horn U, Freund M, Bujanow A, Gundlach C, Hartwigsen G, Eippert F. Assessing adverse effects and unspecific effects of transcutaneous spinal direct current stimulation (tsDCS) bioRxiv. 2023:2023.12.14.571694. doi: 10.1101/2023.12.14.571694. [DOI] [Google Scholar]
- Zhao Q, Han Y, Hu X-Y, Zhang S, Zhang L, Wang J, Zhang Q-Q, Tao M-S, Fang J, Yang J, Liu R-G, et al. Transcranial electrical stimulation for relief of peripartum mental health disorders in women undergoing cesarean section with combined spinal–epidural anesthesia: A pilot randomized clinical trial. Front Psych. 2022;13 doi: 10.3389/fpsyt.2022.837774. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhou J, Li D, Ye F, Liu R, Feng Y, Feng Z, Li R, Li X, Liu J, Zhang X, Zhou J, et al. Effect of add-on transcranial alternating current stimulation (tACS) in major depressive disorder: A randomized controlled trial. Brain Stimul. 2024;17(4):760–768. doi: 10.1016/j.brs.2024.06.004. [DOI] [PubMed] [Google Scholar]
- Zhou Y, Friston KJ, Zeidman P, Chen J, Li S, Razi A. The hierarchical organization of the default, dorsal attention and salience networks in adolescents and young adults. Cereb Cortex. 2018;28(2):726–737. doi: 10.1093/cercor/bhx307. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhou Y, Ma Y, Yu C, Chen Y, Ding J, Yu J, Zhou R, Wang X, Fan T, Shi C. Detection analysis of perioperative plasma and CSF reveals risk biomarkers of postoperative delirium of Parkinson’s disease patients undergoing deep brain stimulation of the subthalamic nuclei. Clin Interv Aging. 2022;17:1739–1749. doi: 10.2147/CIA.S388690. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhu R, Ma X, Wang Z, Hui Q, You X. Improving auditory alarm sensitivity during simulated aeronautical decision-making: The effect of transcranial direct current stimulation combined with computerized working memory training. Cognit Res: Princ Implic. 2025;10(1):11. doi: 10.1186/s41235-025-00620-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Živanović M, Bjekić J, Konstantinović U, Filipović SR. Effects of online parietal transcranial electric stimulation on associative memory: A direct comparison between tDCS, theta tACS, and theta-oscillatory tDCS. Sci Rep. 2022;12(1):14091. doi: 10.1038/s41598-022-18376-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zohny H. A defence of the welfarist account of enhancement. Perform Enhance Health. 2014;3(3):123–129. doi: 10.1016/j.peh.2015.09.002. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.