Commercial automated insulin delivery therapy can improve glucose outcomes by increasing glucose time-in-range (3·9–10·0 mmol/L; 70–180 mg/dL) and reducing hypoglycaemia (<3·9 mmol/L; <70 mg/dL) in people with type 1 diabetes. Commercial closed-loop systems have shown modest improvement in percentage time-in-range, with an increase of 5 percentage points for MiniMed 670G1 and of 10 percentage points for Control-IQ2 compared with open-loop therapy. New developments with closed-loop systems are likely to enable further improvements in glucose outcomes while reducing user burden for individuals able to access closed-loop systems. These developments are focused on addressing deficiencies with current closed-loop systems by, for example, reducing or eliminating the requirement of prandial insulin dosing, overcoming the slow kinetics of subcutaneous insulin compared with endogenous insulin production, and improving the handling of insulin before, during, and after exercise. These advances will ideally lead to improved performance and reduced patient burden, and enable more widespread usage.
Current closed-loop systems are hybrid systems. These systems are not fully automated and require user estimation of carbohydrate intake and manual insulin dosing before a meal. Pre-meal insulin dosing is important to prevent marked hyperglycaemia due to the slow kinetics of subcutaneous insulin absorption into the plasma, as compared with the rapid effects of endogenous insulin and incretin in normal postprandial physiology. An automated insulin control algorithm that is unaware of a meal will attempt to compensate for the postprandial hyperglycaemia by delivering additional insulin, which can potentially lead to subsequent hypoglycaemia following the meal. Compounding this problem is the fact that people might forget to dose prandial insulin entirely, dose the insulin after a meal, or misestimate carbohydrate consumption, which is common.3 Future closed-loop systems might benefit from a fully automated control algorithm that uses artificial intelligence to detect when a meal is consumed and deliver prandial insulin once detection occurs. This system will not replicate normal physiology, but has the potential to benefit those living with type 1 diabetes who have to consistently take prandial insulin. For example, an algorithm that detects a meal can either automatically deliver a proportion of the intended prandial insulin or prompt the user to dose.
Closed-loop systems are dependent on accurate continuous glucose monitoring, which measures interstitial glucose, to drive insulin delivery. The accuracy of continuous glucose monitoring has improved remarkably over the past 15 years and there is now an integrated continuous glucose monitoring standard for use in closed-loop systems. However, there are still some problems to overcome. There is inherently some delay in interstitial glucose measurements as compared with actual plasma glucose concentrations. This delay can be more pronounced during exercise and after meals when glucose concentrations are changing rapidly. However, data have shown that the accuracy of continuous glucose monitoring during exercise can be high.4 In addition, physically pressing on a continuous glucose monitoring device, such as lying on it while sleeping, can cause compression lows. After a hypoglycaemic episode, a continuous glucose monitoring system can have a delayed recovery; glucose can normalise before the device detects this rise. If continuous glucose monitoring underestimates blood glucose, this can cause prolonged inappropriate suspension of insulin and a risk of diabetic ketoacidosis. It is crucial that closed-loop systems mitigate this risk by limiting the time of insulin suspension. Additionally, interfering substances, including paracetamol, ascorbic acid, and urea, can cause certain continuous glucose monitors to overestimate blood glucose, potentially leading to the over-delivery of insulin, although many devices contain membranes to filter out select interfering substances.
Like standard insulin pump therapy, closed-loop systems are reliant on a functioning infusion set or cannula from a patch pump. Commercial pumps include pressure alarms to alert users of set failures, but these alarms have low sensitivity for detecting failures. Set failures can lead to marked hyperglycaemia and diabetic ketoacidosis if not intervened upon. Furthermore, if the system calls for the delivery of insulin, but the insulin is not delivered by the pump due to an error or partial occlusion, the system’s estimate of insulin-on-board will be inaccurate. If a problem with a pump occurs and the person takes insulin via a syringe, the closed-loop system will also have an inaccurate estimate of insulin-on-board, which could result in hypoglycaemia. There is a clear need for better infusion sets that are less prone to failure and for standard ways to manage set failures in the setting of closed-loop insulin delivery. Extended wear infusion sets are eagerly anticipated and a product with a 7-day wear time is now available in Europe5 and has been approved in the USA. Extended wear infusion sets could reduce scar tissue formation from the subcutaneous cannula, which can limit lifelong use of subcutaneous insulin infusion by decreasing the absorption of insulin and increasing the frequency of occlusions.6
Closed-loop systems might help to improve glycaemia during exercise, but several user-initiated actions are needed to maximise effectiveness. Future work is required to improve glycaemic outcomes during and following exercise, and the timing and type of exercise need to be carefully considered. For exercise within 2 h of a meal, meal insulin dosing should be reduced (appendix pp 1–2). All current closed-loop systems can set higher glucose targets for exercise, but these targets should be initiated 60–90 min before exercise onset. Although current closed-loop systems deliver only insulin, there might be a benefit to including additional hormones, such as glucagon to reduce hypoglycaemia in response to exercise and pramlintide to reduce postprandial glucose excursions; neither of these hormones are yet approved for use in pumps (appendix pp 1–2).
The burgeoning development of diverse insulin delivery algorithms, additional hormones, peripheral components, and standardised reporting of metrics has opened up new possibilities for closed-loop therapy.7,8 Although advances in closed-loop therapy could considerably improve health outcomes for people with type 1 diabetes, many cannot afford, or do not have access to, these systems.9 Addressing disparities in diabetes care will be another challenge to overcome in the years ahead. For those people who do not have access to closed-loop systems or choose not to use them, more sophisticated tools are needed to improve decision support systems for multiple daily injection therapy.10 Patients will soon have numerous choices of closed-loop systems and interoperable components. Patients and providers will need to consider which systems are best on the basis of an individual’s needs and characteristics. In this 100th year of insulin, we are optimistic for the prospects of closed-loop technologies to achieve further treatment advances for the millions of people worldwide with type 1 diabetes.
Supplementary Material
Acknowledgments
LMW has received research support from the Oregon Health & Science University Medical Research Foundation and a speaker’s honorarium from the Association of Diabetes Care and Education Specialists. PGJ has received travel honorarium and research support from Dexcom; has received research support and licensing revenue from Agamatrix; has patents issued and pending on continuous glucose monitoring and closed-loop areas; and is on the board of, and owns stock in, Pacific Diabetes Technologies. MCR has served in an advisory role for Zealand Pharma, Zucara Therapeutics, Xeris Pharmaceuticals, and Supersapiens; has received compensation as a speaker at scientific sessions, workshops, or both sponsored by Novo Nordisk, Eli Lilly, Sanofi, Insulet, and Medtronic Diabetes; has received in-kind research support from Dexcom; and owns stock and shares with Supersapiens and Zucara Therapeutics. DPZ has received speaker’s honoraria from Medtronic Diabetes, Ascensia Diabetes, and Insulet Canada, and research support from the Helmsley Charitable Trust and the International Society for Pediatric and Adolescent Diabetes Juvenile Diabetes Research Foundation Research Fellowship. JRC has served on advisory boards, data safety monitoring boards, or both, for Insulet, Novo Nordisk, and AstraZeneca, and owns stock in Pacific Diabetes Technologies; her institution, Oregon Health & Science University, has received research support from Dexcom. LMW, PGJ, and JCR time to write this Comment are supported by the National Institute of Diabetes and Digestive and Kidney Diseases of the National Institutes of Health (5R01DK120367–04).
Contributor Information
Leah M Wilson, Division of Endocrinology, Harold Schnitzer Diabetes Health Center, Oregon Health & Science University, Portland, OR 97239, USA.
Peter G Jacobs, Artificial Intelligence for Medical Systems Lab, Department of Biomedical Engineering, Oregon Health & Science University, Portland, OR 97239, USA.
Michael C Riddell, Muscle Health Research Centre, School of Kinesiology and Health Science, York University, Toronto, ON, Canada.
Dessi P Zaharieva, Department of Pediatrics, Stanford University, Stanford, CA, USA.
Jessica R Castle, Division of Endocrinology, Harold Schnitzer Diabetes Health Center, Oregon Health & Science University, Portland, OR 97239, USA.
References
- 1.Bergenstal RM, Garg S, Weinzimer SA, et al. Safety of a hybrid closed-loop insulin delivery system in patients with type 1 diabetes. JAMA 2016; 316: 1407–08. [DOI] [PubMed] [Google Scholar]
- 2.Brown SA, Kovatchev BP, Raghinaru D, et al. Six-month randomized, multicenter trial of closed-loop control in type 1 diabetes. N Engl J Med 2019; 381: 1707–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Gillingham MB, Li Z, Beck RW, et al. Assessing mealtime macronutrient content: patient perceptions versus expert analyses via a novel phone app. Diabetes Technol Ther 2021; 23: 85–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Guillot FH, Jacobs PG, Wilson LM, et al. Accuracy of the dexcom g6 glucose sensor during aerobic, resistance, and interval exercise in adults with type 1 diabetes. Biosensors (Basel) 2020; 10: E138. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Buckingham BA, Brazg RL, Bailey TS, et al. 100-lb: infusion set survival and performance during the Medtronic Extended-Wear Infusion Set (EWIS) pivotal trial. Diabetes 2021; 70 (suppl 1): 100. [Google Scholar]
- 6.Johansson UB, Amsberg S, Hannerz L, et al. Impaired absorption of insulin aspart from lipohypertrophic injection sites. Diabetes Care 2005; 28: 2025–27. [DOI] [PubMed] [Google Scholar]
- 7.Danne T, Nimri R, Battelino T, et al. International consensus on use of continuous glucose monitoring. Diabetes Care 2017; 40: 1631–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Shah VN, Garg SK. Standardized hybrid closed-loop system reporting. Diabetes Technol Ther 2021; 23: 323–31. [DOI] [PubMed] [Google Scholar]
- 9.Agarwal S, Schechter C, Gonzalez J, Long JA. Racial-ethnic disparities in diabetes technology use among young adults with type 1 diabetes. Diabetes Technol Ther 2021; 23: 306–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Tyler NS, Mosquera-Lopez CM, Wilson LM, et al. An artificial intelligence decision support system for the management of type 1 diabetes. Nat Metab 2020; 2: 612–19. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.